Emerson Micro Motion 4200 Configuration And Use Manual

Micro Motion® 4200 Transmitters
Configuration and Use Manual
Configuration and Use Manual
MMI-20048166, Rev AA
April 2019
Safety messages
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion Customer Service department.
Emerson Flow customer service
Email:
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 704 136
Mexico +41 (0) 41 7686
111
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy 8008 77334 China +86 21 2892 9000
France 0800 917 901 India 800 440 1468
Central & Eastern +41 (0) 41 7686
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684
666
111
New Zealand 099 128 804
Japan +81 3 5769 6803
2
Configuration and Use Manual Contents
MMI-20048166 April 2019

Contents

Chapter 1 Before you begin............................................................................................................7
1.1 About this manual............................................................................................................................ 7
1.2 Related documents.......................................................................................................................... 7
1.3 Installation types.............................................................................................................................. 7
1.4 Communication tools and protocols.................................................................................................9
Chapter 2 Quick start................................................................................................................... 11
2.1 Applying power.............................................................................................................................. 11
2.2 Check meter status.........................................................................................................................11
2.3 Commissioning wizards..................................................................................................................12
2.4 Make a startup connection to the transmitter.................................................................................12
2.5 Set the transmitter clock................................................................................................................ 12
2.6 View the licensed features..............................................................................................................13
2.7 Set informational parameters.........................................................................................................13
2.8 Characterize the meter (if required)................................................................................................14
2.9 Verify mass flow measurement.......................................................................................................17
2.10 Verify the zero.............................................................................................................................. 17
Chapter 3 Introduction to configuration and commissioning........................................................19
3.1 Security and write protection......................................................................................................... 19
3.2 Work with configuration files..........................................................................................................21
Chapter 4 Configure process measurement..................................................................................23
4.1 Configure Sensor Flow Direction Arrow.......................................................................................... 23
4.2 Configure mass flow measurement................................................................................................ 24
4.3 Configure volume flow measurement for liquid applications.......................................................... 28
4.4 Configure Gas Standard Volume (GSV) flow measurement.............................................................32
4.5 Configure density measurement.................................................................................................... 36
4.6 Configure temperature measurement............................................................................................38
4.7 Configure Pressure Measurement Unit .......................................................................................... 40
4.8 Configure Velocity Measurement Unit ........................................................................................... 40
Chapter 5 Configure process measurement applications.............................................................. 43
5.1 Set up the API referral application .................................................................................................. 43
5.2 Set up concentration measurement............................................................................................... 56
Chapter 6 Configure advanced options for process measurement................................................ 69
6.1 Detect and report two-phase flow.................................................................................................. 69
6.2 Configure Flow Rate Switch ........................................................................................................... 70
6.3 Configure events............................................................................................................................ 71
6.4 Configure totalizers and inventories............................................................................................... 73
Configuration and Use Manual 3
Contents Configuration and Use Manual
April 2019 MMI-20048166
6.5 Configure logging for totalizers and inventories............................................................................. 76
6.6 Configure Process Variable Fault Action .........................................................................................76
Chapter 7 Configure device options and preferences....................................................................81
7.1 Configure the transmitter display................................................................................................... 81
7.2 Configure the transmitter's response to alerts................................................................................ 85
Chapter 8 Integrate the meter with the control system................................................................ 93
8.1 Configure the transmitter channels................................................................................................ 93
8.2 Configure the mA outputs.............................................................................................................. 94
8.3 Configure the Frequency Output.................................................................................................. 102
8.4 Configure the Discrete Output..................................................................................................... 106
Chapter 9 Configure digital communications............................................................................. 109
9.1 Configure HART communications ................................................................................................109
Chapter 10 Complete the configuration....................................................................................... 115
10.1 Test or tune the system using sensor simulation.........................................................................115
10.2 Enable or disable software write protection................................................................................116
Chapter 11 Transmitter operation................................................................................................119
11.1 View process and diagnostic variables........................................................................................ 119
11.2 View and acknowledge status alerts........................................................................................... 120
11.3 Read totalizer and inventory values............................................................................................ 122
11.4 Start, stop, and reset totalizers and inventories.......................................................................... 122
Chapter 12 Measurement support................................................................................................125
12.1 Use Smart Meter Verification (SMV)........................................................................................... 125
12.2 Zero the meter........................................................................................................................... 132
12.4 Set up pressure compensation................................................................................................... 134
12.5 Validate the meter......................................................................................................................139
12.6 Perform a (standard) D1 and D2 density calibration....................................................................141
Chapter 13 Maintenance..............................................................................................................145
13.1 Install a new transmitter license................................................................................................. 145
13.2 Reboot the transmitter...............................................................................................................146
13.3 Battery replacement...................................................................................................................146
Chapter 14 Log files, history files, and service files........................................................................147
14.1 Generate history log files............................................................................................................147
14.2 Generate service files..................................................................................................................148
14.3 SMV history and SMV log............................................................................................................149
14.4 Totalizer history and log............................................................................................................. 150
14.5 Generate service files..................................................................................................................151
Chapter 15 Troubleshooting........................................................................................................ 157
15.1 Overview.................................................................................................................................... 157
15.2 Status alerts, causes, and recommendations.............................................................................. 157
4 Micro Motion 4200 Transmitters
Configuration and Use Manual Contents
MMI-20048166 April 2019
15.3 Transmitter does not communicate........................................................................................... 174
15.4 API referral problems..................................................................................................................174
15.5 Concentration measurement problems......................................................................................175
15.6 Density measurement problems................................................................................................ 176
15.7 Discrete Output problems..........................................................................................................178
15.8 Flow measurement problems..................................................................................................... 179
15.9 Frequency Output problems.......................................................................................................181
15.10 Milliamp output problems........................................................................................................183
15.11 Temperature measurement problems......................................................................................186
15.12 Check power supply wiring.......................................................................................................187
15.13 Check sensor to transmitter wiring...........................................................................................188
15.14 Check grounding......................................................................................................................189
15.15 Perform loop tests....................................................................................................................189
15.16 Trim mA Output....................................................................................................................... 193
15.17 Using sensor simulation for troubleshooting............................................................................ 194
15.18 Check HART communications.................................................................................................. 194
15.19 Check Lower Range Value and Upper Range Value ...................................................................195
15.20 Check mA Output Fault Action .................................................................................................195
15.21 Check the scaling of the frequency output............................................................................... 195
15.22 Check Frequency Output Mode ............................................................................................... 195
15.23 Check Frequency Output Fault Action ......................................................................................196
15.24 Check the direction parameters............................................................................................... 196
15.25 Check the cutoffs..................................................................................................................... 196
15.26 Check for two-phase flow (slug flow)........................................................................................196
15.27 Check for radio frequency interference (RFI).............................................................................197
15.28 Check HART burst mode...........................................................................................................197
15.29 Check the drive gain................................................................................................................. 197
15.30 Checking process variables.......................................................................................................198
15.31 Check the pickoff voltage......................................................................................................... 201
15.32 Check for internal electrical problems...................................................................................... 202
15.33 Locate a device using the HART 7 Squawk feature.................................................................... 203
Appendix A Using the transmitter display..................................................................................... 205
A.1 Components of the transmitter display........................................................................................ 205
A.2 Access and use the display menu..................................................................................................206
Appendix B Using ProLink III with the transmitter......................................................................... 211
B.1 Basic information about ProLink III ...............................................................................................211
B.2 Connect with ProLink III ............................................................................................................... 212
Appendix C Using a Field Communicator with the transmitter...................................................... 215
C.1 Basic information about the Field Communicator ........................................................................215
C.2 Connect with the Field Communicator ........................................................................................ 215
Configuration and Use Manual 5
Contents Configuration and Use Manual
April 2019 MMI-20048166
Appendix D Channel combinations................................................................................................217
D.1 Rules for channel combinations................................................................................................... 217
D.2 Valid combinations for channel configuration..............................................................................217
Appendix E Concentration measurement matrices, derived variables, and process variables........ 219
E.1 Standard matrices for the concentration measurement application............................................. 219
E.2 Derived variables and calculated process variables....................................................................... 220
Appendix F Environmental compliance.........................................................................................223
F.1 RoHS and WEEE............................................................................................................................ 223
6 Micro Motion 4200 Transmitters
Configuration and Use Manual Before you begin
MMI-20048166 April 2019

1 Before you begin

1.1 About this manual

This manual helps you configure, commission, use, maintain, and troubleshoot Micro Motion 4200 transmitters.
Important
This manual assumes that the following conditions apply:
The transmitter has been installed correctly and completely according to the instructions in the
transmitter installation manual
The installation complies with all applicable safety requirements
The user is trained in government and corporate safety standards

1.2 Related documents

See the approval documentation shipped with the transmitter, or download the appropriate documentation from the Micro Motion web site (www.emerson.com):
Micro Motion 4200 2-Wire Transmitter: Installation Manual
Micro Motion 4200 2-Wire Transmitter: Product Data Sheet
Micro Motion 4200 2-Wire Transmitter: Safety Manual for Safety Instrumented Systems
Micro Motion ProLink III User Manual
Sensor installation manual, which is shipped with the sensor
FMEDA report for Coriolis Flowmeter with the 4200 Transmitter, Prepared for Emerson by exida.com LLC

1.3 Installation types

The 4200 transmitter was ordered and shipped for one of two installation types. The fifth character of the transmitter number indicates the installation type.
Figure 1-1: Installation type indication for 4200 transmitters
The number is located on the device tag on the side of the transmitter.
Table 1-1: Installation types for 4200 transmitters
Code Description
I Integral mount
C Remote mount
Configuration and Use Manual 7
Before you begin Configuration and Use Manual
April 2019 MMI-20048166
Figure 1-2: 4200 transmitter -- Integral mount
A. Conduit openings
B. Clamping ring C. Sensor case D. Transmitter housing cover (hidden from view)
The transmitter is installed directly on the sensor.
The connections between the transmitter and sensor are 9-wire, and do not require field wiring on the integral mount version.
The I/O connections consist of 2 channels, each channel being 2-wire. Power must be supplied to Channel A for the transmitter to operate, while Channel B connections are optional.
8 Micro Motion 4200 Transmitters
Configuration and Use Manual Before you begin
MMI-20048166 April 2019
Figure 1-3: 4200 transmitter -- Remote mount
A. Transmitter housing cover
B. Clamping ring C. Junction box
The transmitter is installed remotely from the sensor. The 9-wire connection between the sensor and transmitter must be field wired. Power supply and I/O must be field wired to the transmitter. The sensor connection is in the junction box.

1.4 Communication tools and protocols

You can use several different communications tools and protocols to interface with the transmitter, use different tools in different locations, or use different tools for different tasks.
Tool
Display Not applicable
ProLink III HART
Field Communicator HART
For information about how to use the communication tools, see Using ProLink III with the transmitter and
Using a Field Communicator with the transmitter in this manual.
Supported protocols
"Factory Use Only" port in non-hazardous areas
Configuration and Use Manual 9
Before you begin Configuration and Use Manual
April 2019 MMI-20048166
Note
Some configuration and administrative procedures can also be performed through the display menus. However, for complete access to transmitter functions, Micro Motion recommends setting up and using an administrative connection.
Tip
You may be able to use other communications tools, such as AMS Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this manual. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at
www.emerson.com.
10 Micro Motion 4200 Transmitters
Configuration and Use Manual Quick start
MMI-20048166 April 2019

2 Quick start

2.1 Applying power

The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement.
Procedure
1. Verify that the cables are connected to the transmitter as described in the installation manual.
2. Verify that all transmitter and sensor covers and seals are closed.
WARNING
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are tightly closed. For hazardous area installations, applying power while housing covers are removed or loose can cause an explosion.
3. Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. During this period, the Transmitter
Initializing alert is active. The diagnostic routines should complete in approximately 30 seconds.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power has been off long enough to allow components to reach ambient temperature, allow the electronics to warm up for approximately 10 minutes before relying on process measurements. During this warm-up period, you may observe minor measurement instability or inaccuracy.
When the flowmeter has completed its power-up sequence, if the default settings are in effect:
The display will show the current mass flow rate and measurement unit.
If there are any active fault or informational alarms, the alert banner displays until the alert has been
manually acknowledged.
If the alert has been acknowledged but is still active, the alert icon displays above the menu button, and
the Alert List menu appears at the top of the main menu.

2.2 Check meter status

Check the meter for any error conditions that require user action or that affect measurement accuracy.
Procedure
Wait approximately 10 seconds for the power-up sequence to complete. Immediately after power-up, the transmitter runs through diagnostic routines and checks for error
conditions. During the power-up sequence, the Transmitter Initializing alert is active. This alert should clear automatically when the power-up sequence is complete.
Configuration and Use Manual 11
Quick start Configuration and Use Manual
April 2019 MMI-20048166

2.3 Commissioning wizards

The transmitter menu includes a Guided Setup to help you move quickly through the most common configuration parameters. ProLink III also provides a commissioning wizard.
By default, when the transmitter starts up, the Guided Setup menu is offered. You can choose to use it or not. You can also choose whether or not Guided Setup is displayed automatically.
To enter Guided Setup upon transmitter startup, choose Yes at the prompt.
To enter Guided Setup after transmitter startup, choose Menu > Startup Tasks.
To control the automatic display of Guided Setup, choose Menu > Configuration > Guided Setup.
For information on the ProLink III commissioning wizard, see the Micro Motion ProLink III User Manual.
As the commissioning wizards are self guided, they are not documented in detail.

2.4 Make a startup connection to the transmitter

For all configuration tools except the display, you must have an active connection to the transmitter to configure the transmitter.
Procedure
Identify the connection type to use, and follow the instructions for that connection type in the appropriate appendix.
Communications tool
ProLink III HART Using ProLink III with the transmitter
Field Communicator HART Using a Field Communicator with the
Connection type to use Instructions
transmitter

2.5 Set the transmitter clock

Display
ProLink III Device Tools > Configuration > Transmitter Clock
Field Communicator Configure > Manual Setup > Clock
The transmitter clock provides timestamp data for alerts, service logs, history logs, and all other timers and dates in the system. You can set the clock for your local time or for any standard time you want to use.
Tip
You may find it convenient to set all of your transmitter clocks to the same time, even if the transmitters are in different time zones.
Menu > Configuration > Time/Date/Tag
Procedure
1. Select the time zone that you want to use.
2. If you need a custom time zone, select Special Time Zone and enter your time zone as a difference
from UTC (Coordinated Universal Time).
3. Set the time appropriately for the selected time zone.
12 Micro Motion 4200 Transmitters
Configuration and Use Manual Quick start
MMI-20048166 April 2019
Tip
The transmitter does not adjust for Daylight Savings Time. If you observe Daylight Savings Time, you must reset the transmitter clock manually.
4. Set the month, day, and year.
The transmitter tracks the year and automatically adds a day for leap years.

2.6 View the licensed features

Display Menu > About > Licenses > Licensed Features
ProLink III Device Tools > Device Information > Licensed Features
Field Communicator Overview > Device Information > Licenses
You can view the licensed features to ensure that the transmitter was ordered with the required features.
Licensed features are purchased and available for permanent use. The options model code represents the licensed features.
A trial license allows you to explore features before purchasing. The trial license enables the specified features for a limited number of days. This number is displayed for reference. At the end of this period, the feature will no longer be available.
To purchase additional features or request a trial license, document the Unique ID Number and current license key from your transmitter and contact customer service. To enable the additional features or trial license, you will need to install the new license on the transmitter.

2.7 Set informational parameters

Display
ProLink III Device Tools > Configuration > Informational Parameters
Field Communicator Configure > Manual Setup > Device
You can set several parameters that identify or describe the transmitter and sensor. These parameters are not used in processing and are not required.
Procedure
1. Set informational parameters for the transmitter.
a) Set Transmitter Serial Number to the serial number of your transmitter.
The transmitter serial number is provided on the metal tag that is attached to the transmitter housing.
Menu > Configuration > Device Information
b) Set Descriptor to any desired description of this transmitter or measurement point.
c) Set Message to any desired message.
d) Verify that Model Code (Base) is set to the base model code of the transmitter.
Configuration and Use Manual 13
Quick start Configuration and Use Manual
April 2019 MMI-20048166
The base model code completely describes your transmitter, except for the features that can be licensed independently. The base model code is set at the factory.
e) Set Model Code (Options) to the options model code of the transmitter.
The options model code describes the independent features that have been licensed for this transmitter. The original options model code is set at the factory. If you license additional options for this transmitter, Micro Motion will supply an updated options model code.
2. Set informational parameters for the sensor.
a) Set Sensor Serial Number to the serial number of the sensor connected to this transmitter.
The sensor serial number is provided on the metal tag that is attached to the sensor case.
b) Set Sensor Material to the material used for the sensor.
c) Set Sensor Liner to the material used for the sensor liner.
d) Set Flange Type to the type of flange that was used to install the sensor.
Do not set Sensor Type. Sensor Type is set or derived during characterization.

2.8 Characterize the meter (if required)

Display
ProLink III Device Tools > Calibration Data
Field Communicator Configure > Manual Setup > Characterization
Characterizing the meter adjusts your transmitter to match the unique traits of the sensor it is paired with. The characterization parameters (also called calibration parameters) describe the sensor’s sensitivity to flow, density, and temperature. Depending on your sensor type, different parameters are required.
Values for your sensor are provided on the sensor tag or the calibration certificate.
If your transmitter was ordered with a sensor, it was characterized at the factory. However, you should still
verify the characterization parameters.
Perform a characterization whenever you replace a core processor.
Procedure
1. Optional: Specify Sensor Type as Curved Tube.
Note
Unlike earlier transmitters, the 4200 derives Sensor Type from the user-specified values for FCF and K1 in combination with an internal ID.
2. Set the flow calibration factor: FCF (also called Flow Cal or Flow Calibration Factor). Be sure to include
all decimal points.
3. Set the density characterization parameters: D1, D2, TC, K1, K2, and FD. (TC is sometimes shown as
DT.)
4. Apply the changes as required by the tool you are using.
Menu > Configuration > Sensor Parameters
14 Micro Motion 4200 Transmitters
Configuration and Use Manual Quick start
MMI-20048166 April 2019
The transmitter identifies your sensor type, and characterization parameters are adjusted as required:
If Sensor Type changed from Straight Tube to Curved Tube, five characterization parameters are
removed from the list.
If Sensor Type did not change, the list of characterization parameters does not change.

2.8.1 Sample sensor tags

Figure 2-1: Tag on older curved-tube sensors (all sensors except T-Series)
Figure 2-2: Tag on newer curved-tube sensors (all sensors except T-Series)
Configuration and Use Manual 15
Quick start Configuration and Use Manual
April 2019 MMI-20048166

2.8.2 Flow calibration parameters (FCF, FT)

Two separate values are used to describe flow calibration: a 6-character FCF value and a 4-character FT value. They are provided on the sensor tag.
Both values contain decimal points. During characterization, these are entered as a single 10-character string. The 10-character string is called either Flowcal or FCF.
If your sensor tag shows the FCF and the FT values separately and you need to enter a single value, concatenate the two values to form the single parameter value, retaining both decimal points.
Concatenating FCF and FT
FCF = x.xxxx FT = y.yy Flow calibration parameter: x.xxxxy.yy

2.8.3 Density calibration parameters (D1, D2, K1, K2, FD, DT, TC)

Density calibration parameters are typically on the sensor tag and the calibration certificate.
If your sensor tag does not show a D1 or D2 value:
For D1, enter the Dens A or D1 value from the calibration certificate. This value is the line-condition
density of the low-density calibration fluid. Micro Motion uses air. If you cannot find a Dens A or D1 value, enter 0.001 g/cm3.
For D2, enter the Dens B or D2 value from the calibration certificate. This value is the line-condition density
of the high-density calibration fluid. Micro Motion uses water. If you cannot find a Dens B or D2 value, enter 0.998g/cm3 .
If your sensor tag does not show a K1 or K2 value:
For K1, enter the first 5 digits of the density calibration factor. In this sample tag, this value is shown as
12500.
For K2, enter the second 5 digits of the density calibration factor. In this sample tag, this value is shown as
14286.
Figure 2-3: K1, K2, and TC values in the density calibration factor
16 Micro Motion 4200 Transmitters
Configuration and Use Manual Quick start
MMI-20048166 April 2019
If your sensor does not show an FD value, contact customer service.
If your sensor tag does not show a DT or TC value, enter the last 4 characters of the density calibration factor. In the sample tag shown above, the value is shown as 4.44.

2.9 Verify mass flow measurement

Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available method.
Procedure
Read the value for Mass Flow Rate on the transmitter display, which is the default initial display.
Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the Process Variables
panel.
Connect to the transmitter with the Field Communicator and read the value for Mass Flow Rate.
Online > Overview > Mass Flow Rate
Postrequisites
If the reported mass flow rate is not accurate:
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues.

2.10 Verify the zero

Display
ProLink III Device Tools > Calibration > Smart Zero Verification and Calibration > Verify Zero
Field Communicator Service Tools > Maintenance > Calibration > Zero Calibration > Perform Zero Verify
Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one of the following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Prerequisites
Important
Do not verify the zero or zero the meter if a high-severity alert is active. Correct the problem, then verify the zero or zero the meter. You may verify the zero or zero the meter if a low-severity alert is active.
Menu > Service Tools > Verification & Calibration > Meter Zero > Zero Verification
Procedure
1. Prepare the meter:
Configuration and Use Manual 17
Quick start Configuration and Use Manual
April 2019 MMI-20048166
a) Allow the meter to warm up for at least 20 minutes after applying power.
b) Run the process fluid through the sensor until the sensor temperature reaches the normal
process operating temperature.
c) Stop flow through the sensor by shutting the downstream valve, and then the upstream valve if
available.
d) Verify that the sensor is blocked in, that flow has stopped, and that the sensor is completely full
of process fluid.
2. Start the zero verification procedure, and wait until it completes.
3. If the zero verification procedure fails:
a) Confirm that the sensor is completely blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
b) Verify that the process fluid is not flashing or condensing, and that it does not contain particles
that can settle out.
c) Repeat the zero verification procedure.
d) If it fails again, zero the meter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Related information
Zero the meter
18 Micro Motion 4200 Transmitters
Configuration and Use Manual Introduction to configuration and commissioning
MMI-20048166 April 2019

3 Introduction to configuration and commissioning

3.1 Security and write protection

The transmitter has several features that can help to protect it against intentional or unintentional access and configuration changes.
When enabled, the software setting Write Protection prevents any configuration changes. When
enabled, a lock icon displays at the top of the home screen of the display.
When enabled, the display option Display Security prevents any configuration changes being made from
the display unless the display password is entered. Display Security does not prevent configuration changes from other interfaces.

3.1.1 Enable or disable software write protection

When enabled, Write-Protection prevents changes to the transmitter configuration. You can perform all other functions, and you can view the transmitter configuration parameters.
Write protection is enabled by toggling the physical write protect (dip) switch (identified by a lock icon) located behind the display module.
Figure 3-1: Write Protect (Dip) Switch Behind the Display Module
Configuration and Use Manual 19
Introduction to configuration and commissioning Configuration and Use Manual
April 2019 MMI-20048166
Figure 3-2: Write Protect on the Display (Upper Right Corner)
You cannot change write protection from any host configuration tool.
Note
Write protecting the transmitter primarily prevents accidental changes to configuration, not intentional changes. Any user who can make changes to the configuration can disable write protection.

3.1.2 Configure security for the display

Display
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Display Menus
You can configure a display password, and require the operator to enter the password to make any changes to configuration through the display, or to access alert data through the display.
The operator always has read-only access to the configuration menus.
Procedure
1. Enable or disable display security as desired.
Menu > Configuration > Security > Display Security
20 Micro Motion 4200 Transmitters
Configuration and Use Manual Introduction to configuration and commissioning
MMI-20048166 April 2019
Option Description
Enabled When an operator chooses an action that leads to a configuration change, they are
prompted to enter the display password.
Disabled When an operator chooses an action that leads to a configuration change, they are
prompted to activate ⇦⇧⇩⇨. This is designed to protect against accidental changes to configuration. It is not a security measure.
2. If you enabled display security, enable or disable alert security as desired.
Option Description
Enabled If an alert is active, the alert symbol is shown above the Menu button on the display but
the alert banner is not displayed. If the operator attempts to enter the alert menu, they are prompted to enter the display password.
Disabled If an alert is active, the alert symbol is shown in the upper right corner of the display and
the alert banner is displayed automatically. No password or confirmation is required to enter the alert menu.
Restriction
You cannot disable display security and enable alert security.
If you did not enable display security, alert security is disabled and cannot be enabled.
If both display security and alert security are enabled, and you disable display security, alert security
is disabled automatically.
3. Set the display password to the desired value.
Default: AAAA
Range: Any four alphanumeric characters
If you enable display security but you do not change the display password, the transmitter will post a Configuration alert.

3.2 Work with configuration files

3.2.1 Save a configuration file using ProLink III

You can save the current transmitter configuration to your PC. The ProLink PC file format is supported.
Procedure
To save the current configuration to your PC, in ProLink III format:
a) Choose Device Tools > Configuration Transfer > Save Configuration.
b) Select On my computer in ProLink III file format and click Next.
c) Click Save.
d) Select the configuration parameters to be included in this file.
Configuration and Use Manual 21
Introduction to configuration and commissioning Configuration and Use Manual
April 2019 MMI-20048166
To save a backup file, select all parameters.
To save a replication file, select all parameters except device-specific parameters.
e) Click Save.
f) Browse to the desired location, then enter the name for this configuration file.
g) Set the file type to ProLink configuration file.
h) Click Start Save.
The configuration file is saved to the specified location as yourname.pcfg.

3.2.2 Load a configuration file using ProLink III

You can load a configuration file to the transmitter's working memory. The PC file formats are supported: the ProLink III PC file format is supported.
Note
When you use ProLink III format for configuration files, you can specify configuration parameters individually or by groups. Therefore, you can use this format for both backup and replication.
Procedure
To load a file in ProLink III format from the PC:
a) Choose Device Tools > Configuration Transfer > Load Configuration.
b) Select On my computer in ProLink III file format and click Next.
c) Select the parameters that you want to load.
d) Click Load.
e) Set the file type to Configuration file.
f) Navigate to the file you want to load, and select it.
g) Click Start Load.
The parameters are written to working memory, and the new settings become effectively immediately.

3.2.3 Restore the factory configuration

Display
ProLink III Device Tools > Configuration Transfer > Restore Factory Configuration
Field Communicator Service Tools > Maintenance > Reset/Restore > Restore Factory Configuration
A file containing the factory configuration is always saved in the transmitter's internal memory, and is available for use.
Menu > Configuration > Restore Factory Configuration
This action is typically used for error recovery or for repurposing a transmitter.
If you restore the factory configuration, the real-time clock, the audit trail, the historian, and other logs are not reset.
22 Micro Motion 4200 Transmitters
Configuration and Use Manual Configure process measurement
MMI-20048166 April 2019

4 Configure process measurement

4.1 Configure Sensor Flow Direction Arrow

Display Menu > Configuration > Process Measurement > Flow Variables > Flow Direction
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Sensor Direction
Sensor Flow Direction Arrow is used to accommodate installations in which the Flow arrow on the sensor does not match the majority of the process flow. This typically happens when the sensor is accidentally installed backwards.
Sensor Flow Direction Arrow interacts with mA Output Direction, Frequency Output Direction, and Totalizer Direction to control how flow is reported by the outputs and accumulated by the totalizers and
inventories.
Sensor Flow Direction Arrow also affects how flow is reported on the transmitter display and via digital communications. This includes ProLink III and the Field Communicator.
Figure 4-1: Flow arrow on sensor
A. Flow arrow
B. Actual flow direction
Procedure
Set Sensor Flow Direction Arrow as appropriate.
Option
With Arrow The majority of flow through the sensor matches the flow arrow on the sensor. Actual
Configuration and Use Manual 23
Description
forward flow is processed as forward flow.
Configure process measurement Configuration and Use Manual
April 2019 MMI-20048166
Option Description
Against Arrow The majority of flow through the sensor is opposite to the flow arrow on the sensor. Actual
forward flow is processed as reverse flow.
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow direction or the setting of Sensor Flow Direction Arrow. Sensor Flow Direction Arrow controls only whether actual flow is processed as forward flow or reverse flow

4.2 Configure mass flow measurement

The mass flow measurement parameters control how mass flow is measured and reported. The mass total and mass inventory are derived from the mass flow data.

4.2.1 Configure Mass Flow Measurement Unit

Display Menu > Configuration > Process Measurement > Flow Variables > Mass Flow Settings > Units
ProLink III Device Tools > Configuration > Process Measurement > Flow > Mass Flow Rate Unit
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Unit
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow rate. The default unit used for mass total and mass inventory is derived from this unit.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
Default: g/sec (grams per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Define a special measurement unit for mass flow
Display
ProLink III Device Tools > Configuration > Process Measurement > Flow > Mass Flow Rate Unit > Special
Field Communicator Configure > Manual Setup > Measurements > Optional Setup > Special Units > Mass Special Units
Menu > Configuration > Process Measurement > Flow Variables > Mass Flow Settings > Units > SPECIAL
24 Micro Motion 4200 Transmitters
Configuration and Use Manual Configure process measurement
MMI-20048166 April 2019
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Procedure
1. Specify Base Mass Unit.
Base Mass Unit is the existing mass unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Mass Flow Conversion Factor as follows:
a) x base units = y special units
b) Mass Flow Conversion Factor = x ÷ y
4. Enter Mass Flow Conversion Factor.
The original mass flow rate value is divided by this value.
5. Set Mass Flow Label to the name you want to use for the mass flow unit.
6. Set Mass Total Label to the name you want to use for the mass total and mass inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Defining a special measurement unit for mass flow
You want to measure mass flow in ounces per second (oz/sec).
1. Set Base Mass Unit to Pounds (lb).
2. Set Base Time Unit to Seconds (sec).
3. Calculate Mass Flow Conversion Factor: a. 1 lb/sec = 16 oz/sec
b. Mass Flow Conversion Factor = 1 ÷ 16 = 0.0625
4. Set Mass Flow Conversion Factor to 0.0625.
5. Set Mass Flow Label to oz/sec.
6. Set Mass Total Label to oz.

4.2.2 Configure Flow Damping

Display
ProLink III Device Tools > Configuration > Process Measurement > Flow > Flow Rate Damping
Field Communicator Configure > Manual Setup > Measurements > Flow > Flow Damping
Configuration and Use Manual 25
Menu > Configuration > Process Measurement > Flow Variables > Flow Damping
Configure process measurement Configuration and Use Manual
April 2019 MMI-20048166
Flow Damping controls the amount of damping that will be applied to the measured mass flow rate. It affects flow rate process variables that are based on the measured mass flow rate. This includes volume flow rate and gas standard volume flow rate.
Flow Damping also affects specialized flow rate variables such as temperature-corrected volume flow rate (API referral) and net mass flow rate (concentration measurement). It is not applied to the flow rate received via the frequency input.
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value specifies the time period, in seconds, over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
Default: 0.64 seconds
Range: 0 seconds to 60 seconds
Note
If a number greater than 60 is entered, it is automatically changed to 60.
Tip
A high damping value makes the process variable appear smoother because the reported value changes
slowly.
A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
The combination of a high damping value and rapid, large changes in flow rate can result in increased
measurement error.
Whenever the damping value is non-zero, the reported measurement will lag the actual measurement
because the reported value is being averaged over time.
In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the reported value.
The transmitter automatically rounds off any entered damping value to the nearest valid value. Therefore,
the recommended damping value for gas applications should be 3.2 seconds. If you enter 2.56, the transmitter will round it off to 3.2.
For filling applications, Micro Motion recommends using the default value of 0.04 seconds.
26 Micro Motion 4200 Transmitters
Configuration and Use Manual Configure process measurement
MMI-20048166 April 2019
Effect of flow damping on volume measurement
Flow damping affects volume measurement for liquid volume data. Flow damping also affects volume measurement for gas standard volume data. The transmitter calculates volume data from the damped mass flow data.
Interaction between Flow Damping and mA Output Damping
In some circumstances, both Flow Damping and mA Output Damping are applied to the reported mass flow value.
Flow Damping controls the rate of change in flow process variables. mA Output Damping controls the rate of change reported via the mA Output. If mA Output Process Variable is set to Mass Flow Rate, and both Flow Damping and mA Output Damping are set to non-zero values, flow damping is applied first, and the added damping calculation is applied to the result of the first calculation.

4.2.3 Configure Mass Flow Cutoff

Display Menu > Configuration > Process Measurement > Flow Variables > Mass Flow Settings > Low Flow Cutoff
ProLink III Device Tools > Configuration > Process Measurement > Flow > Mass Flow Cutoff
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Cutoff
Mass Flow Cutoff specifies the lowest mass flow rate that will be reported as measured. All mass flow rates below this cutoff will be reported as 0.
Procedure
Set Mass Flow Cutoff to the value you want to use.
Default: A sensor-specific value set at the factory. If your transmitter was ordered without a sensor, the
default may be 0.0.
Recommendation: 0.5% of maximum flow rate of the attached sensor. See the sensor specifications.
Important
Do not use your meter for measurement with Mass Flow Cutoff set to 0.0 g/sec. Ensure that Mass Flow Cutoff is set to the value that is appropriate for your sensor.
Effect of Mass Flow Cutoff on volume measurement
Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the actual mass data rather than the reported value.
Volume flow has a separate Volume Flow Cutoff that is not affected by the Mass Flow Cutoff value.
Interaction between Mass Flow Cutoff and mA Output Cutoff
Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported via the mA Output. If mA Output Process
Configuration and Use Manual 27
Configure process measurement Configuration and Use Manual
April 2019 MMI-20048166
Variable is set to Mass Flow Rate, the mass flow rate reported via the mA Output is controlled by the higher of the two cutoff values.
Mass Flow Cutoff affects all reported values and values used in other transmitter behavior (e.g., events defined on mass flow).
mA Output Cutoff affects only mass flow values reported via the mA Output.
Cutoff interaction with mA Output Cutoff lower than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
mA Output Cutoff: 10 g/sec
Mass Flow Cutoff: 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0 will be used in all internal processing.
Cutoff interaction with mA Output Cutoff higher than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
mA Output Cutoff: 15 g/sec
Mass Flow Cutoff: 10 g/sec
Result:
If the mass flow rate drops below 15 g/sec but not below 10 g/sec:The mA Output will report zero flow.
The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
4.3 Configure volume flow measurement for liquid
applications
The volume flow measurement parameters control how liquid volume flow is measured and reported. The volume total and volume inventory are derived from volume flow data.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.
Related information
Configure Volume Flow Type for liquid applications
28 Micro Motion 4200 Transmitters
Configuration and Use Manual Configure process measurement
MMI-20048166 April 2019
Configure Volume Flow Measurement Unit for liquid applications Configure Volume Flow Cutoff

4.3.1 Configure Volume Flow Type for liquid applications

Display Menu > Configuration > Process Measurement > Flow Variables > Volume Flow Settings > Flow Type >
ProLink III Device Tools > Configuration > Process Measurement > Flow > Volume Flow Type > Liquid Volume
Field Communicator Configure > Manual Setup > Measurements > Optional Setup > GSV > Volume Flow Type > Liquid Volume
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
If you are using the API referral application, you must set Volume Flow Type to Liquid. Gas standard volume measurement is incompatible with the API referral application.
Restriction
If you are using the concentration measurement application, you must set Volume Flow Type to Liquid.
Gas standard volume measurement is incompatible with the following applications:
Concentration measurement
Procedure
Set Volume Flow Type to Liquid.
Liquid
4.3.2 Configure Volume Flow Measurement Unit for liquid
applications
Display
ProLink III Device Tools > Configuration > Process Measurement > Flow > Volume Flow Rate Unit
Field Communicator Configure > Manual Setup > Measurements > Flow > Volume Flow Unit
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the volume flow rate. The unit used for the volume total and volume inventory is based on this unit.
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
Default: l/sec (liters per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Configuration and Use Manual 29
Menu > Configuration > Process Measurement > Flow Variables > Volume Flow Settings > Units
Configure process measurement Configuration and Use Manual
April 2019 MMI-20048166
Options for Volume Flow Measurement Unit for liquid applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement Unit, plus one user-defined measurement unit. Different communications tools may use different labels for the units.
Define a special measurement unit for volume flow
Display Menu > Configuration > Process Measurement > Flow Variables > Volume Flow Settings > Units >
ProLink III Device Tools > Configuration > Process Measurement > Flow > Volume Flow Rate Unit > Special
Field Communicator Configure > Manual Setup > Measurements > Optional Setup > Special Units > Volume Special Units
SPECIAL
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Procedure
1. Specify Base Volume Unit.
Base Volume Unit is the existing volume unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Volume Flow Conversion Factor as follows:
a) x base units = y special units
b) Volume Flow Conversion Factor = x ÷ y
4. Enter Volume Flow Conversion Factor.
The original volume flow rate value is divided by this conversion factor.
5. Set Volume Flow Label to the name you want to use for the volume flow unit.
6. Set Volume Total Label to the name you want to use for the volume total and volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Defining a special measurement unit for volume flow
You want to measure volume flow in pints per second (pints/sec).
1. Set Base Volume Unit to Gallons (gal).
2. Set Base Time Unit to Seconds (sec).
3. Calculate the conversion factor:
a. 1 gal/sec = 8 pints/sec
b. Volume Flow Conversion Factor = 1 ÷ 8 = 0.1250
4. Set Volume Flow Conversion Factor to 0.1250.
30 Micro Motion 4200 Transmitters
Loading...
+ 194 hidden pages