Emerson I-0021, NGA 2000 User Manual

Rosemount Analytical
NGA 2000
Software Manual
MLT Analyzer
ML T Analyzer Module
(combined with NGA 2000 platform /
MLT analyzer / TFID analyzer
or customer-developed control unit)
2nd Edition 07/98
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Catalog No.: 90 003 482
Managing The Process Better
Rosemount Analytical
This Operation Manual includes information about the operation of the instrument. Information about the additional indications and notes regarding maintenance, troubleshooting and repair are found in the accompanying Maintenance & Operation Manual.
T roubleshooting, component replacement and internal adjustments must be made by qualified service personnel only.
Fisher-Rosemount GmbH & Co does not take responsibility for an y omissions or errors in this manual. Any liability for direct or indirect damages, which might occur in connection with the deliv ery or the use of this manual, is expressly e xcluded to the extend permitted by applicable law.
This instrument has left the works in good order according to safety regulations. T o maintain this operating condition, the user must strictly follow the instructions and consider the warnings in this manual or provided on the instrument.
Misprints and alterations reserved
©
1998 by FISHER-ROSEMOUNT GmbH & Co. (ETC/PAD)
st
1
Edition: 04/98
2nd Edition: 07/98
Read this operation manual carefully before attempting to operate the analyzer ! For expedient handling of reports of defects , please include the model and serial number which can be read on the instrument identity plate.
Fisher - Rosemount GmbH & Co.
European T echnology Center
Industriestrasse 1
D - 63594 Hasselroth • Germany
Phone + 49 (6055) 884-0
T elefax + 49 (6055) 884-209
Internet: http://www.processanalytic.com
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

Contents

1 Introduction 1 - 1
2 Structure of Menus 2 - 1
3 Startup and Operation, General Notes and Main Menu 3 - 1
3.1 Starting and Initializing...................................................................................3 - 1
3.2 Display and Function...................................................................................... 3 - 2
3.3 "TAG" and Operating Keys ............................................................................3 - 2
3.4 Lines and Softkey Functionality .....................................................................3 - 3
3.5 Important Functions of the Softkeys...............................................................3 - 4
3.6 Entering/Changing of Variables......................................................................3 - 5
3.7 Starting a Function.........................................................................................3 - 6
3.8 Main Menu .....................................................................................................3 - 7
4 Basic Controls and Analyzer Module Calibration 4 - 1
4.1 Analyzer Channel Status................................................................................ 4 - 3
4.1.1 Status Details – Failures...............................................................................4 - 5
– Maintenance requests ........................................................4 - 7
– Function controls ................................................................4 -11
– Measurements/Alarms........................................................4 -13
– Events.................................................................................4 -15
– Acknowledge and clear failures..........................................4 -19
– Acknowledge and clear maintenance requests ..................4 -23
– Acknowledge and clear function controls............................4 -27
4.1.2 Operational Settings.......................................................................................4 -31
4.2 Single Component Display - Change of Channel...........................................4 -33
4.3 Multi Component Display - Change of Channel.............................................4 -35
4.4 Calibration Procedure Status......................................................................... 4 -37
4.5 Zero Calibration..............................................................................................4 -39
4.6 Span Calibration.............................................................................................4 -43
4.7 Flow Zero Gas................................................................................................ 4 -47
4.8 Flow Span Gas...............................................................................................4 -51
4.9 Flow Sample Gas...........................................................................................4 -55
4.10 Flow Test Gas................................................................................................4 -59
4.11 Close all Valves.............................................................................................. 4 -63
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
I
5 Analyzer and I/O-Module Expert Configuration 5 - 1
5.1 Analyzer Module Setup 5 - 3
5.1.1 Calibration Parameters...................................................................................5 - 5
– Span gases.........................................................................................5 - 6
– Tolerances ..........................................................................................5 - 7
– Calibration procedure setup................................................................5 - 9
– Time controlled calibration..................................................................5 -12
– Calibration...........................................................................................5 -15
– Advanced calibration methods............................................................5 -18
– Zero gases..........................................................................................5 -20
5.1.2 Alarm Parameters...........................................................................................5 -21
5.1.3 Range Parameters .........................................................................................5 -25
– Begin and end of ranges.....................................................................5 -27
– Response times (t90) ...........................................................................5 -28
– Autoranging control.............................................................................5 -29
5.1.4 Cross Interference Compensation..................................................................5 -31
5.1.5 Linearization...................................................................................................5 -33
5.1.6 Programmable Logic Control (PLC)................................................................5 -37
5.1.7 Programmable Calculator...............................................................................5 -45
5.1.8 Measurement Display Configuration...............................................................5 -49
5.1.9 Acknowledgement of Status Reports..............................................................5 -52
5.1.10 Concentration Measurement Parameters.......................................................5 -54
5.1.11 Peak Measurement ........................................................................................5 -55
5.1.12 Differential Measurement ...............................................................................5 -57
5.1.13 Gasflow Setup................................................................................................5 -59
5.1.14 Pressure Compensation.................................................................................5 -60
5.1.15 Flow Measurement.........................................................................................5 -62
5.1.16 Temperature Measurement............................................................................5 -63
5.1.17 Loading/Saving Configuration Parameters.....................................................5 -64
5.1.18 Inputs and Outputs.........................................................................................5 -67
– Local SIO ............................................................................................5 -68
– Local DIO............................................................................................5 -74
– Signal codes........................................................................................5 -75
5.1.19 Delay and Average.........................................................................................5 -79
5.1.20 Special Functions...........................................................................................5 -81
5.1.21 AK-Protocol Communication ..........................................................................5 -82
II
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
5.2 I/O Module Controls 5 -83
5.2.1 SIO Module 5 -84
5.2.2 DIO Module(s) 5 -93
5.3 I/O Module Setup 5 -99
6 System Configuration 6 - 1
6.1 Diagnostic Menus........................................................................................... 6 - 3
6.1.1 Control Module Diagnostics................................................................. 6 - 4
6.1.2 Analyzer Module Diagnostics .............................................................. 6 - 5
6.2 Date and Time................................................................................................6 - 6
6.3 Security Codes...............................................................................................6 - 7
6.4 Network Module Binding ................................................................................6 - 9
6.5 System Reset.................................................................................................6 -11
6.6 System Modules.............................................................................................6 -12
7 Display Controls 7 - 1
Index
Supplement: System Calibration
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
III
IV
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
1 Introduction
This software manual describes step by step how to operate successfully with the ½ 19" and 19" MLT analyzer module and analyzer of the NGA 2000 Series. Chapter two shows the structure of the MLT software menus. Chapter three describes the display and the keyboard of the analyzer. Chapter four describes the basic controls with detailed illustrations. So you can easily compare the actual display of the analyzer module with the illustrations of the manual. Chapter five describes the expert configurations of the analyzer module and of the Input/ Output Modules (I/O modules). Chapter six describes the system configuration. The layout of both chapters is not as detailed as in chapter four. Normally, the way to a certain menu of the MLT software is described with the software catchwords you have to press to reach this menu. You will find the illustration of the corresponding LCD screen at the end of the catchword listing. After that you can read the meaning of the functions and variables of each expert or system configuration menu. Some contents of the expert configurations are not important for each customer. It depends on the configuration of your NGA 2000 system, relative to the following components:
Control Module CMAnalyzer Module AMInput/Output Modules I/O's (SIO = Standard I/O, DIO = Digital I/O,
System Auto Calibration I/O, Analog Output with 3 Alarms I/O, Auto Calibration I/O)
You can distinguish the following system units and SIO/DIO configurations:
System Unit SIO/DIO-Configuration
MLT analyzer module (AM):
without front panel,
i.e. without control unit
can be combined with a platform,
an MLT analyzer, a TFID analyzer or a customer developed control
1 local SIO and 1 local DIO can be
installed in the MLT analyzer module
SIO and DIO can be configured
for the channels of the analyzer module only
unit
Platform (CM Software):
Control unit with front panel
Without measurement channels
1 SIO and up to 4 DIO's can be
installed in the platform (CM I/O)
SIO and DIO can be configured
for all channels combined with the platform
Section
Page
5.1.18
p. 5-67
5.2
p. 5-83
MLT analyzer (CM and MLT AM software):
Analyzer with front panel
CM and AM software in the same
analyzer, i.e. all functions of the control unit and of the AM are combined in one controller board
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
1 SIO and 1 DIO can be installed
in the MLT analyzer (CM I/O)
SIO and DIO can be configured
for all channels combined with the MLT analyzer
NGA 2000
5.2
p. 5-83
1 - 1
The following illustrations shall make plain the connection between the hardware configu­ration and the software setup of the modules:
NGA 2000 System via Platform
(see 5.1.18)
(see 5.1.18)
Local
I/O's
SIO DIO
SIO DIO
(additional
manuals)
(additional
manuals)
(additional
manuals)
NGA 2000 System via MLT Analyzer
Analyzer
Modules (AM's)
TFID
MLT
CLD
PMD
FID
Control
Module (CM)
Platform
I/O
Modules
1 SIO
4 DIO's
max.
Other I/O's
(see 5.2.1)
(see 5.2.2)
(see 5.1.18)
(see 5.1.18)
1 - 2
Local
I/O's
SIO DIO
SIO DIO
(additional
manuals)
(additional
manuals)
(additional
manuals)
Analyzer
Modules (AM's)
MLT
TFID
CLD
PMD
FID
NGA 2000
Control
Module (CM)
MLT Analyzer
System I/O
Modules
1 SIO
1 DIO
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
(see 5.2.1)
(see 5.2.2)
NGA 2000 System via TFID Analyzer
1 Introduction
(see 5.1.18)
(see 5.1.18)
Local
I/O's
SIO DIO
SIO DIO
(additional
manuals)
(additional
manuals)
(additional
manuals)
Analyzer
Modules (AM's)
TFID
MLT
CLD
PMD
FID
Control
Module (CM)
TFID Analyzer
System I/O
Modules
1 SIO
1 DIO
(see 5.2.1)
(see 5.2.2)
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
1 - 3
1 - 4
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
enu lines ending with three points (...) are followed by submenus with further functions and set-ups. All set-ups in the menu "Basic controls" are valid for the At each menu point of the "Expert configuration" and "System configuration" you can find which functionality will be set up:
nalyzer Module), CM (
odule), I/O (
utput Module).
-
-
-
-
NGA 2000
Analyzer module
calibration & basic
controls... (AM)
Section 4 Section 5 Section 6 Section 7
Calibration procedure status... Zero calibration of all ranges
Span calibration of the current range Current gas valve setting:
- Zero gas flow
- Span gas flow
- Sample gas flow
- Test gas flow
- Closing all valves
Main Menu
Analyzer and I/O-
module expert
configuration...
- Analyzer module controls... (AM)
- I/O module controls...
(I/O's: SIO, DIO)
- Analyzer module setup...
(AM)
- I/O module setup...
(Other micro processor controlled I/O's)
Section 3.8
System configuration
and diagnostics...
- System calibration... (CM)
- Diagnostic menus...
(AM, CM, I/O's: SIO, DIO and other I/O's)
- Date and time... (CM)
- Security codes... (Lock of each level)
- Network module binding...
(AM, CM, I/O's: SIO, DIO and other I/O's)
- System reset... (CM)
- System modules... (CM)
- System tag (CM)
Display controls...
- Display controls of the front panel LCD
- Front panel control (number of digits)
- Auxiliary lines (for the single component display)
2 Structure of Menus
Notes: * M
* *
AM (A
2 - 1
Control M
analyzer module (AM).
Input-/O
2 - 2
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

3 Startup and Operation, General Notes and Main Menu

3.1 Starting and Initializing

After switching on the MLT analyzer or analyzer module (in a platform or part of a NGA network), the initialization procedure will be performed. A self control of the analyzer modules or the analyzer is running. You can see a sequence of several displays. They show the status of initialization, revision notes of the MLT software and the tag:
(C) 1998 FISHER-ROSEMOUNT Analytical
NGA-2000 Control Module/Version 3.2.1
Initializing Network
Initializing network interface
LCDReset Abort
F1 F2 F3 F4 F5
If you press the F1 key during the initializing, you will reset the LCD brightness and contrast to factory settings (see also section 7). Pressing the F3 key will abort the network initializing. Then you will have no connection to any analyzer module. Only the menus of the platform will be available. At the end of the initializing procedure you can see the single component display of channel one (see illustration on next page). It is the origin to all the other channels, menus and submenus. The instructions of the basic controls (chapter four) are all beginning with the single com­ponent display. The actual display might differ from the shown one because the customer can configure it according to his requirements (see section 5.1.8 p. 5-49 and section 7).
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
3 - 1

3.2 Display and Function

The LCD screen shows all measurement values of the analyzer and all customer instructions. You can operate with five function keys, four arrow keys (cursors) and the enter key. The function of each key depends on:
the type of analyzer/analyzer module usedthe optional auxiliary modules (e.g. I/O boards) usedthe individual menu displayed
In case of power failure all customer specific module parameters are saved by a battery­powered buffer.

3.3 "TAG" and Operating Keys

At the top left of each menu page you will find the tag of the current channel.
Typical tags:
MLT/CH1/R1: MLT Analyzer or Analyzer Module / CHannel 1 / Range 1TFID-R1: Thermo FID Analyzer or Analyzer Module - Range 1
In this manual you will find normally "TAG" as general name. But in the specific MLT menu pages you will find "MLT".
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Function Keys:
keys without defined functionsThe current function depends on
the menu selected
The softkey legend is shown on
the display above the key
Range: 1
0.0 100.0
Cursor keys:
↑↑ -key / ↓↓ -key:
Line up / line down
within the same menu
Alteration of numbers,
variables or digits
-key / -key:
Moving back/forwards
between the pages of a menu
Selection of digits
Enter Key:
To confirm a previously entered value
(variable)
To start a selected function
(Alternative: -key)
To go into a menu (via menu line)
3 - 2
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
3 Startup and Operation, General Notes and Main Menu

3.4 Lines and Softkey Functionality

Lines can be selected by the ↓↓ -key or the ↑↑ -key. The selected line is displayed white on black. You have four different types of lines in the menu:
Menu line... / Menu Softkey...
Line/Softkey lettering ending with three dots.You will go to a submenu/further menu by pressing the softkey resp. by pressing the
88 -key or the -key in the menu line.
Function line / Function Softkey !
Line/Softkey lettering ending with an exclamation-mark.You will start a function (e.g. calibration) by pressing the softkey resp. by pressing the
88 -key or the -key in the function line.
Line of variables:
Line ending with a colon.Display of module parameters (variables).Some parameters can be changed (e.g. begin of range), some parameters display only
a status (e.g. temperature) and cannot be changed. These variables will be displayed below a line within the menu.
Text line
Line without any punctuation marks.Only display of informations.
The following illustrations shall make plain the functionalities of lines and softkeys des­cribed above:
Menu Line
Function Lines
Lines of Variables
(These here cannot
be changed)
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
F1 F2 F3 F4 F5
37.50 ppm
Valves...
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
3 - 3
Text Line
TAG
-- Calibration Procedure Status --
Procedure status: Ready Remaining procedure time: 0 s Current/expected gas flow: Samplegas Concentration in span gas units: 37.50 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Fri 05-29-1998 13:32:06 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
F1 F2 F3 F4 F5
37.50 ppm
Function Softkey
Menu Softkeys
3.5 Important Functions of the Softkeys Display
Change from the single component display to the multi component display.F1 in the single component display.
Measure
Change from all menus and submenus to the single component display of the channel
selected.
F1.
Status (see section 4.1 p. 4-3!)
Change to the menu "Analyzer Channel Status":
Display of the most important parameters and information about the status of the current channel or module.
If available: F2.
Main (see section 3.8 p. 3-7!)
Change from the single component display to the main menu.F3 in the single component display.
Channel
Scrolling through the channels in the same menu. In the main menu and the single
component display you can move among all channels of the connected analyzers and analyzer modules. In the submenus you can only move among the channels of the current analyzer or analyzer modules.
If available: F3 (F4 in the single component display).
3 - 4
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
3 Startup and Operation, General Notes and Main Menu
Lock
Lock of all three operation levels from the main menu, if a security code is activated in
the system configuration level (see section 6.3 p. 6-7).
F4 in the main menu.
Calib (see section 4.4 p. 4-37 and 5.1.1 p. 5-15!)
Change from the single component display to the menu "Analyzer module calibration".F5 in the single component display.
MFG Data (see section 3.8 p. 3-7/8!)
Change from the main menu to the menu "Module Manufacturing Data":
Further submenus are available with informations about the control module and analyzer module data, such as address of the manufacturer, serial number of the modules or the software and hardware revisions.
F5 in the main menu.
Back
Moving back to the last menu page selected (Alternative: -key) or
reset of a changed but not confirmed parameter to the former value.
If available: F4 for moving back, F2 for reset.
More
Changing to a further menu page.If available: F5.

3.6 Entering/Changing of Variables

88 -key
If you have already selected a line of variables (displayed white on black) and press the
88 -key, only the parameter will be selected and can be changed.
If you press the 88 -key again, the new value will be confirmed.
↑↑ -key / ↓↓ -key
Function depends on the variable selected: - Changing the parameter values
- Scrolling among variables selected
- Changing of digits or characters
Increasing or decreasing of numbers.
-key / -key
Selection of digits within a number.For some variables you can change the quantity of digits or characters.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
3 - 5

3.7 Starting a function

If you press the 88 -key or the → -key in a function line, you will be asked to confirm the order in the following way:
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
F1 F2 F3 F4 F5
37.50 ppm
If you will press the F2 key, the function will start immediately.If you will press the F4 key, you will go back to the last menu page.
Note: If you don't want to be asked before the start of each function, you can configure it in the menu "Measurement Display Configuration" in the expert configuration level (see section 5.1.8 p. 5-49). Select "No" in the line "Display confirmation menus". Then, each function will start directly after the order is done and no confirmation will be required.
3 - 6
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
3 Startup and Operation, General Notes and Main Menu
To setup see 6.2 p. 6-6
3.8 Main Menu If you press the F3 key (Main...) or the → -key in any single component display, you
will change to the "Main Menu". From there you can change to all operating levels of your MLT analyzer or analyzer module to set up and control the parameters of measurement, calibration and data transfer ! Via the F5 key (MFG Data) you can change to several submenus, where you will find a lot of important data about the control module (MLT analyzer or platform) and the analyzer module, such as service address or serial number !
Ramifications from the Main Menu:
TAG
-- Main Menu --
Analyzer module calibration & basic controls...
Analyzer and I/O-module expert configuration... System configuration and diagnostics... Display controls... Time & Date: 14:01:45 25 May 1998
System tag: Fisher-Rosemount
Measure Status... Channel Lock... MFG Data
F1 F2 F3 F4 F5
F1
F2
F3
Changing to the single component display of the current channel
Changing to the menu "Analyzer Channel Status" of the current channel
Scrolling through all channels of the connected analyzers and analyzer modules
95.00 ppm
+
+
+
See chapter 4 ! See chapter 5 ! See chapter 6 ! See chapter 7 !
Factory Setting
Section 3.1 p. 3-1/2 !
Section 4.1 p. 4-3 !
See channel tag !
F4
F5
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Lock of any operating level by security code
Changing to the menu "Module Manufacturing Data"
NGA 2000
+
+
Section 6.3 p. 6-7 !
See next pages !
3 - 7
Ramifications from the menu "Module Manufacturing Data":
1. Control module data:
TAG
Control module data...
Analyzer module data...
Measure <<< Back... >>>
-- Module Manufacturing Data --
F1 F2 F3 F4 F5
TAG
(C) Copyright Fisher-Rosemount Analytical Inc., 1998
Manufactured by:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, CA 92807-1802 /USA
Tel: (714) 986-7600
FAX: (714) 577-8739
Measure Back...
F1 F2 F3 F4 F5
2. Analyzer module data:
95.00 ppm
95.00 ppm
TAG
(C) Copyright Fisher-Rosemount GmbH & Co, 1998
Manufactured by:
Fisher-Rosemount GmbH & Co
Industriestrasse 1
D-63594 Hasselroth / Germany
Tel. (+49) 6055 884-0
FAX. (+49) 6055 884-209
Measure Or... Back... More...
F1 F2 F4 F5
TAG
-- Control Module Version Information --
Serial number: CM 4711 Manufacturing date: 14.05.1998
Hardware revision: ACU02 R: 3.2.2 Software revision: 3.2.1/P001 Revision date: May 29 1998 Revision time: 12:35:00
Phrase dictionary version: P001/1 Language: English
Measure Back...
F3
95.00 ppm
95.00 ppm
F1 F2 F3 F4 F5
TAG
Control module data...
Analyzer module data...
Measure <<< Back... >>>
-- Module Manufacturing Data --
95.00 ppm
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Version Information --
Serial number: AM 4712 Manufacturing date: 14.05.1998
Hardware revision: ACU02 R: 3.2.2 Software revision: 3.2.1 / P001 Revision date: May 29 1998 Revision time: 12:41:18
Measure Back...
95.00 ppm
More... Measure Back...
F1 F2 F3 F4 F5
TAG
(C) Copyright Fisher-Rosemount GmbH & Co, 1998
Manufactured by:
Fisher-Rosemount GmbH & Co
Industriestrasse 1
D-63594 Hasselroth / Germany
Tel. (+49) 6055 884-0
FAX. (+49) 6055 884-209
Measure Back... More...
95.00 ppm
F1 F2 F3 F4 F5
TAG
Measurement system: PSV-System RAM-memory: 242748 Bytes Local SIO module installed: Enabled Serial interface adaptor: RS-232 Heater installed: No Local DIO module installed: 2 Sensor system revision:
-- Hardware Configuration --
Channel
95.00 ppm
F1 F2 F3 F4 F5
3 - 8
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4 Basic Controls and Analyzer Module Calibration
In the chapter "Basic Controls and Analyzer Module Calibration" the most important mea­surement and calibration functions of your MLT analyzer or analyzer module are described.
All steps are figured with detailed illustrations and operation instructions. In the left column you can see display and keyboard of the NGA front panel. The keys you have to press are illustrated in black. In the right column you can read the instructions and notes. All instructions will begin with any single component display and will end with the correspond­ing single component display after the setups are done. So you can easily compare the actual display of the analyzer or analyzer module with the illustrations of the manual.
Example: You want to change from the single component display of channel 1(CO2) to
the single component display of channel 2 (CO).
Picture one shows the starting situation: single component display of CO
Picture two shows the result you get if you press the F4 key (Channel):
single component display of CO.
Left column: Right column:
Display and keyboard Instructions and notes
MLT25/CH1/R2
2.50 %CO2
0 5.00
Maintenance-Requests: No Any_Alarms: No Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
MLT25/CH2/R2
95.00 ppm CO
Range: 2
0.0 100.0Temperature: 25.0 C
Change to the single
component display of another channel
F4
Press
Example: Changing from CO2 (Channel 1) to CO (Channel 2)
Next instruction
or step
.
2
0 250
Temperature: 25.0 C Maintenance-Requests: No Any_Alarms: No Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Range: 2
0.0 100.0
NGA 2000
4 - 1
4 - 2
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
37.50 ppm CH4
0.00 50.00
Range: 1
4.1 Analyzer Channel Status
Change to the
menu "Analyzer Channel Status"
F2
Press
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
In the menu "Analyzer Channel Status" you can find status informations of the current channel. Via the menu lines "Status details..." and "Operational settings..." you can change to further submenus. (see 4.1.1 p. 4-5...30 and
4.1.2 p. 4-31/32)
Change to the
single component display
F1
Press
Note: Via the F2 key you can change to submenus containing the primary and secondary raw measurements.
F1 F2 F3 F4 F5
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back in the single
component display
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 3
4 - 4
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm CH4
Failures
Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the "List
of Possible Failures (page 1 of 2)"
Press
or
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 5
MLT25/CH2/R2
-- List of Possible Failures(1/2) --
One or more failures: No Configuration replaced by factory setting: No Chopper fail: No Raw signal too high: No Detector signal communication failed: No Raw signal too low: No (reserved)
(reserved)
Temperature measurement: No Invalid pressure measurement: No
Measure Back... More...
37.50 ppm
F1 F2 F3 F4 F5
MLT25/CH2/R2
-- List of Possible Failures(2/2) --
External Input: No Linearization underflow: No Linearization overflow: No
37.50 ppm
Change to the
second menu page
F5
Press
Change to the
single component display
F1
Press
Measure Back...
F1 F2 F3 F4 F5
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back in the single
component display of the current channel
4 - 6
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm CH4
Maintenance Requests
Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Maintenance requests..."
Press or as often as necessary to get the menu line "Maintenance requests..." white on black.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 7
TAG
-- Status Details --
Failures...
Maintenance requests...
Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
37.50 ppm
Change to the
"List of Possible Maintenance Requests (page 1 of 2)"
Press
Measure Back...
F1 F2 F3 F4 F5
MLT25/CH2/R2
-- List of Possible Maintenance Requests(1/2) --
One or more maintenance requests: No Too much hours of operation: No Zero calibration - deviation too high: No Span calibration - deviation too high: No Measurement too noisy for zero calibration: No Measurement too noisy for span calibration: No
Measure Back...
37.50 ppm
More...
F1 F2 F3 F4 F5
or
Change to the
second menu page
F5
Press
MLT25/CH2/R2
-- List of Possible Maintenance Requests(2/2) --
External Input: No
Measure Back...
F1 F2 F3 F4 F5
4 - 8
37.50 ppm
NGA 2000
Change to the
single component display
F1
Press
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm CH4
Maintenance Requests
Back in the single
component display of the current channel
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 9
4 - 10
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm CH4
Function Controls
Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Function controls..."
Press or as often as necessary to get the menu line "Function controls..." white on black.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 11
TAG
-- Status Details --
Failures... Maintenance requests...
Function controls...
Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
MLT25/CH2/R2
-- List of Possible Function Controls(1/1) --
One or more function controls: No Request from external input: No Concentration raw signal simulation: No No sample gas flowing: No Calibration in progress: No State - Warmup: No State - Pause: No State - Standby: No Secondary measurement simulation: No External input: No
Measure Back...
37.50 ppm
Change to the
"List of Possible Function Controls"
Press
or
Change to the
single component display
F1
Press
F1 F2 F3 F4 F5
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back in the single
component display of the current channel
4 - 12
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm CH4
Measurements/Alarms
Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Measurements/ Alarms..."
Press or as often as necessary to get the menu line "Measurements/Alarms..." white on black.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 13
TAG
-- Status Details --
Failures... Maintenance requests... Function controls...
Measurements/Alarms...
Events... Acknowledge and clear failures !
Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Measurements/Alarms --
Concentration...
Concentration average... Flow... Pressure... Temperature... Calculator-1... Calculator-2... Calculator-3... Calculator-4...
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the
submenu "Measurements/ Alarms"
Press
or
Change to the single
component display
F1
Press
Options:
With the ↑↑ or the ↓↓ -key
you can change to each line in the menu.
Press the 88 -key in the
line selected to change to the corresponding sub­menu: There you will find the status of the four
alarms Note: If the hardware of the menu point selected is not available, a corresponding message will appear!
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
4 - 14
Range: 1
0.0 100.0
NGA 2000
Back in the single
component display of the current channel
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm CH4
Events
Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Events..."
Press or as often as necessary to get the menu line "Events..." white on black.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 15
TAG
-- Status Details --
Failures... Maintenance requests... Function controls... Measurements/Alarms...
Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
MLT25/CH2/R2
-- Analyzer Module Events (1/2) --
Failure: --.--.-­Maintenance request: --.--.-­Function control/Service: --.--.-­Last configuration lost: --.--.-­Concentration alarm: --.--.-­Flow alarm: --.--.-­Linearization underflow: --.--.-­Linearization overflow: --.--.-­Calculator alarms active: --.--.-­Local analog output out of range: --.--.--
Measure Status... Channel Back... More...
37.50 ppm
Change to the
submenu "Analyzer
Module Events
(page 1 of 2)"
Press
or
Change to the
second menu page
F5
Press
Note: If any event has happened,
you will find a report of date and time in the corresponding line of this menu.
F1 F2 F3 F4 F5
MLT25/CH2/R2
-- Analyzer Module Events (2/2) --
Zero calibration started: --.--.-­Zero calibration finished: --.--.-­Span calibration started: --.--.-­Span calibration finished: --.--.-­Concentration average alarm: --.--.-­Temperature alarm: --.--.-­Pressure alarm: --.--.-­New linearization calculated: --.--.-­Sensor calibration: --.--.--
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the
single component
display
F1
Press
4 - 16
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm CH4
Events
Back to the single
component display of the current channel
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 17
4 - 18
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm CH4
Acknowledge and Clear Failures
Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: Yes Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Notes: If you have solved the reasons for the failures reported, you should start this function. The menu "List of Possible Failures" will be ready for new reports !
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Acknowledge and clear failures !"
Press or as often as necessary to get the menu line "Acknowledge and clear failures !" white on black.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 19
TAG
-- Status Details --
Failures... Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures !
Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
Start the function
Press
or
Note: Starting this function here is only possible, if it is enabled in the menu "Acknowledge­ment of Status Reports" (see 5.1.9 p. 5-52) !
Confirm the order
F2
Press to start the function immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Status Details".
F1 F2 F3 F4 F5
TAG
- S U C C E S S -
- The selected function has been started/executed ­(Wait a moment...)
37.50 ppm
F1 F2 F3 F4 F5
Confirmation
message of the function start appears
This message will be dis­played after the function has been started. Then the display will jump automatically to the menu "Status Details".
4 - 20
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
F1 F2 F3 F4 F5
37.50 ppm
Acknowledge and Clear Failures
Change to the
single component display of the channel selected
F1
Press
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back to the single
component display of the current channel after clearing of failures
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 21
4 - 22
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm CH4
Acknowledge and Clear Maintenance Requests
Change to the menu
"Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: Yes Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Note: If you have solved the reasons for the maintenance requests reported, you should start this function. The menu "List of Possible Maintenance Requests" will be ready for new reports !
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Acknowledge and clear maintenance requests !"
Press or as often as necessary to get the menu line "Acknowledge and clear maintenance requests !" white on black.
90003482(2) [NGA-e (MlT-Software 3.2.X)] 07/98
NGA 2000
4 - 23
TAG
-- Status Details --
Failures... Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures !
Acknowledge and clear maintenance requests !
Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Start the function
Press
or
Note: Starting this function here is only possible, if it is enabled in the menu "Acknowledge­ment of Status Reports" (see 5.1.9 p. 5-52) !
Confirm the order
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
- S U C C E S S -
- The selected function has been started/executed ­(Wait a moment...)
37.50 ppm
F2
Press to start the function immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Status Details".
Confirmation
message of the function start appears
This message will be dis­played after the function has been started. Then the display will jump automatically to the menu "Status Details".
F1 F2 F3 F4 F5
4 - 24
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
F1 F2 F3 F4 F5
37.50 ppm
Acknowledge and Clear Maintenance Requests
Change to the
single component display of the channel selected
F1
Press
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back to the single
component display of the current channel after clearing of maintenance requests
90003482(2) [NGA-e (MlT-Software 3.2.X)] 07/98
NGA 2000
4 - 25
4 - 26
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
37.50 ppm CH4
Acknowledge and Clear Function Controls
Change to the menu
"Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Notes: If you have solved the reasons for the function controls reported, you should start this function. The menu "List of Possible Function Controls" will be ready for new reports !
Change to the
submenu "Status Details"
Press
or
F1 F2 F3 F4 F5
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Acknowledge and clear function controls !"
Press or as often as necessary to get the menu line "Acknowledge and clear function controls!" white on black.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 27
TAG
-- Status Details --
Failures... Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests !
Acknowledge and clear function controls !
Measure Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
Start the function
Press
or
Note: Starting this function here is only possible, if it is enabled in the menu "Acknowledge­ment of Status Reports" (see 5.1.9 p. 5-52) !
Confirm the order
F2
Press to start the function immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Status Details".
F1 F2 F3 F4 F5
TAG
- S U C C E S S -
- The selected function has been started/executed ­(Wait a moment...)
37.50 ppm
F1 F2 F3 F4 F5
Confirmation
message of the function start appears
This message will be dis­played after the function has been started. Then the display will jump automatically to the menu "Status Details".
4 - 28
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
4.1.1 Analyzer Channel Status - Status Details
TAG
-- Status Details --
Failures...
Maintenance requests... Function controls... Measurements/Alarms... Events...
Acknowledge and clear failures ! Acknowledge and clear maintenance requests ! Acknowledge and clear function controls !
Measure Back...
F1 F2 F3 F4 F5
37.50 ppm
Acknowledge and Clear Function Controls
Change to the
single component display of the channel selected
F1
Press
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back to the single
component display of the current channel after clearing of function controls
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 29
4 - 30
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

4.1.2 Analyzer Channel Status - Operational Settings

TAG
37.50 ppm CH4
Change to the
menu "Analyzer Channel Status"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings... Hours of operations: 164
Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F2
Press
Change to the line
"Operational settings..."
Press once to get the line "Operational settings..." white on black.
F1 F2 F3 F4 F5
TAG
-- Analyzer Channel Status --
Status details...
Operational settings...
Hours of operations: 164 Operation status: Ready Events: No Alarms: No Failures: No Maintenance requests: No Function control/Service: No
Measure Channel Back...
RawMeas
37.50 ppm
F1 F2 F3 F4 F5
Change to the
menu "Analyzer Operation Settings"
Press
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 31
TAG
-- Analyzer Operation Settings --
Remote control: Enabled Range change control: Manual Range: 1 Range upper limit: 50.00 ppm Span gas concentration: 50.00 ppm t90-time: 2.00 s Hours of operation: 164 Last re-start occured: 8:40:35 June 22, 1998 Actual zero gas concentration: 0.00 ppm
Auto-start procedures...
Measure Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Auto-Start Procedures --
Position in auto-start list: 1
Channel tag: ­Procedure type: ­Interval mode: Never Start time: ­Start date: -
Time & Date: 16:03:25 June 22, 1998
Measure Back...
37.50 ppm
Change to the
submenu "Auto­Start Procedures"
Press
Change to the
single component display
Note: In the menu "Auto-Start Procedures" you can control the status of the three kinds of time controlled calibrations. (see 5.1.1 p. 5-12/13)
F1 F2 F3 F4 F5
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Back to the single
component display of the current channel
4 - 32
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
MLT25/CH1/R2
2.50 %CO2
4.2 Single Component Display
Change of Channel
Change to the single
component display of another channel
0 5.00
Maintenance-Requests: No Any_Alarms: No Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
MLT25/CH2/R2
Range: 2
0.0 100.0Temperature: 25.0 C
95.00 ppm CO
0 250
Temperature: 25.0 C Maintenance-Requests: No Any_Alarms: No Operation: Ready
Range: 2
0.0 100.0
F4
Press
Example: Changing from CO2 (Channel 1) to CO (Channel 2)
Change back to the
single component display of the starting channel
F4
Press
as often as necessary to get the display of the channel you want
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
MLT25/CH1/R2
2.50 %CO2
0 5.00
Maintenance-Requests: No Any_Alarms: No Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 2
0.0 100.0Temperature: 25.0 C
Note: You can reach any existing channel by pressing the F4 key for several times.
Single component
display of the starting channel appears
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 33
4 - 34
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
MLT25/CH1/R2
2.50 %CO2
4.3 Multi Component Display
Change of Channel
Change to the

multi component display

0 5.00
Maintenance-Requests: No Any_Alarms: No Operation: Ready
Display Status... Main... Channel Calib...
Range: 2
0.0 100.0Temperature: 25.0 C
F1 F2 F3 F4 F5
2.50
95.00
333.0
150.0
20.00
Select Status... Tags Off
MLT25/CH1
%CO2
MLT25/CH2
ppm CO
MLT25/CH3
ppm SO2
MLT25/CH4
ppm NO
MLT25/CH5
%O2
2]
0.00
0.00 250.00
0.00 500.00
0.00 150.00
0.00 100.00
[
[2]
[2]
[2] F.S.
[2]
5.00
LCDReset
F1 F2 F3 F4 F5
F1
Press
Note: You can change to the multi component display from each single component display.
Enable the "selecting
symbol": >
F1
Press
Notes:
In each bargraph you will
find the begin and end of range of the corresponding channel. (F.S. = full scale)
The number in parentheses
shows the number of the selected range.
Option: With the F3 key you can fade out or in the tags.
or
2.50
95.00
333.0
150.0
20.00
Select Status... Tags Off
MLT25/CH1
%CO2
MLT25/CH2
ppm CO
MLT25/CH3
ppm SO2
MLT25/CH4
ppm NO
MLT25/CH5
%O2
0.00
0.00 250.00
0.00 500.00
0.00 150.00
0.00 100.00
F1 F2 F3 F4 F5
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
2]
[
[2]
[2]
[2] F.S.
[2]
5.00
LCDReset
NGA 2000
Select any channel
Press or as often as necessary to
put the > - mark into the line you want to select.
Example: Change from CO2 (Channel 1) to CO (Channel 2)
4 - 35
2.50
95.00
333.0
150.0
20.00
Select Status... Tags Off
MLT25/CH1
%CO2
MLT25/CH2
ppm CO
MLT25/CH3
ppm SO2
MLT25/CH4
ppm NO
MLT25/CH5
%O2
2]
0.00
0.00 250.00
0.00 500.00
0.00 150.00
0.00 100.00
[
[2]
[2]
[2] F.S.
[2]
LCDReset
F1 F2 F3 F4 F5
MLT25/CH2/R2
95.00 ppm CO
5.00
Change to the single
component display of the channel selected
F1
Press
Note: With the F5 key you can enable the LCD-brightness and contrast of the factory settings (see also section 7).
Single component
display of the channel selected appears
0 250
Temperature: 25.0 C Maintenance-Requests: No Any_Alarms: No Operation: Ready
Display Status... Main... Channel Calib...
Range: 2
0.0 100.0
F1 F2 F3 F4 F5
4 - 36
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
37.50 ppm CH4
4.4 Calibration Procedure Status
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
37.50 ppm
Valves...
F1 F2 F3 F4 F5
F5
Press
Change to the
menu "Calibration Procedure Status"
Press
Options: Via the F5 key you can change to a submenu, where you can close all valves or set up each valve separately with:
zero gas or
span gas or
sample gas or
test gas.
(See sections 4.7 to 4.11, p. 4-47...)
TAG
-- Calibration Procedure Status --
Procedure status: Ready Remaining procedure time: 0 s Current/expected gas flow: Samplegas Concentration in span gas units: 37.50 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Fri 05-29-1998 13:32:06 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
F1 F2 F3 F4 F5
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
37.50 ppm
NGA 2000
Change to the menu
page "Calibration Deviations"
F5
Press
4 - 37
TAG
-- Calibration Deviations --
Deviation from zero: 0.34 ppm Sum of zero deviations: 4.41 ppm
Deviation from span: 0.82 ppm Sum of span deviations: 7.57 ppm
Measure Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
37.50 ppm CH4
Change to the single
component display of the current channel
F1
Press
Options:
With the F3 key you can
change to further available channels to check their "Calibration Deviations".
With the F4 key you can
go back to the menu "Calibration Procedure Status".
Back to the single
component display
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
4 - 38
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
37.50 ppm CH4
0.00 50.00
Range: 1
4.5 Zero Calibration
Change to the
menu "Analyzer Module Calibration"
F5
Press
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
37.50 ppm
Valves...
Caution:
Before starting zero calibration, realize that zero gas is available ! (See also section 5.1.1, p. 5-5... !)
Note: The zeroing of all measure­ment ranges of the same channel is running simulta­neously.
Change to the
line "Start zero calibration procedure !"
Press once to get the line "Start zero calibration procedure !" white on black.
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure !
Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
37.50 ppm
Valves...
F1 F2 F3 F4 F5
Start the zero
calibration
Press
or
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 39
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Calibration Procedure Status --
Procedure status: Purging1-Wait Remaining procedure time: 8 s Current/expected gas flow: Zerogas Concentration in span gas units: 37.50 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Fri 05-29-1998 13:32:06 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
37.50 ppm
F1 F2 F3 F4 F5
Confirm the order for
zero calibration
F2
Press
to start
the zeroing immediately.
Option: Press the F4 key if you want to cancel the order.
Notes:
The display of this message
depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
The 3 following illustrations
show the displays you can see after starting the zeroing procedure.
Zeroing:
1. Purging-Wait
Notes:
The procedure time de-
pends on the parameters entered for purge time (see section 5.1.1 p. 5-9).
The purge time must be
long enough to get a stable signal before calibration.
You can cancel the run-
ning calibration procedure at any time with the F2 key.
TAG
-- Calibration Procedure Status --
Procedure status: Zeroing-Wait Remaining procedure time: 97 s Current/expected gas flow: Zerogas Concentration in span gas units: 3.13 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Fri 05-29-1998 13:32:06 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
F1 F2 F3 F4 F5
4 - 40
3.13 ppm
NGA 2000
Zeroing:
2. Zeroing-Wait
Note: The procedure time de­pends on the parameters entered for stability time and averaging time. (see section 5.1.1 p. 5-9)
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
-- Calibration Procedure Status --
Procedure status: Ready Remaining procedure time: 0 s Current/expected gas flow: Samplegas Concentration in span gas units: 0.00 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Mon 06-08-1998 09:21:34 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
0.00 ppm
F1 F2 F3 F4 F5
4.5 Zero Calibration
Change to the
single component display of the current channel
F1
Press
Options:
F4 key: Changing back
to the menu "Analyzer Module Calibration".
F5 key: Changing to
the menu "Calibration Deviations".
TAG
0.00 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Single component
display appears after zeroing is done
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 41
4 - 42
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
37.50 ppm CH4
4.6 Span Calibration
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
37.50 ppm
Valves...
F1 F2 F3 F4 F5
F5
Press
Caution:
Before starting span calibration, realize that spangas with correct concentration is available ! (See also section 5.1.1, p.5-5... !)
Change to the line
"Start span calibration procedure !"
Press twice to get the line "Start span calibration procedure !" white on black.
Notes:
Normally, all measurement
ranges of the same chan­nel will be calibrated simul­taneously.
To calibrate seperately you
have to change the para­meters (see 5.1.1 p. 5-11).
TAG
-- Analyzer Module Calibration --
Calibration procedure status... Start zero calibration procedure !
Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
F1 F2 F3 F4 F5
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
37.50 ppm
Valves...
NGA 2000
Start the

span calibration

Press
or
4 - 43
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
TAG
-- Calibration Procedure Status --
Procedure status: Purging1-Wait Remaining procedure time: 4 s Current/expected gas flow: Spangas-1 Concentration in span gas units: 37.50 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Mon 06-08-1998 09:21:34 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
37.50 ppm
F1 F2 F3 F4 F5
Confirm the order for
span calibration
F2
Press
to start
the spanning immediately.
Option: Press the F4 key if you want to cancel the order.
Notes:
The display of this message
depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
The 3 following illustrations
show the displays you can see after starting the spanning procedure.
Spanning:
1. Purging-Wait
Notes:
The procedure time de-
pends on the parameters entered for purge time (see 5.1.1 p. 5-9).
The purge time must be
long enough to get a stable signal before calibration.
You can cancel the run-
ning calibration procedure at any time with the F2 key.
TAG
-- Calibration Procedure Status --
Procedure status: Spanning-Wait Remaining procedure time: 108 s Current/expected gas flow: Spangas-1 Concentration in span gas units: 43.57 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Mon 06-08-1998 09:21:34 Last span calibration was: Fri 05-29-1998 13:37:23 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
F1 F2 F3 F4 F5
4 - 44
43.57
NGA 2000
Spanning:
2. Spanning-Wait
Note: The procedure time de­pends on the parameters entered for stability time and averaging time. (see 5.1.1 p.5-9)
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
-- Calibration Procedure Status --
Procedure status: Ready Remaining procedure time: 0 s Current/expected gas flow: Samplegas Concentration in span gas units: 50.00 ppm
---------------------- Results ---------------------­Last zero calibration: Success Last span calibration: Success Last zero calibration was: Mon 06-08-1998 09:21:34 Last span calibration was: Mon 06-08-1998 09:28:13 Successful zero+span calibrated ranges: 1+2+3+4
Measure Cancel ! Back... More...
50.00 ppm
F1 F2 F3 F4 F5
4.6 Span Calibration
Change to the
single component display of the current channel
F1
Press
Options:
F4 key: Changing back
to the menu "Analyzer Module Calibration".
F5 key: Changing to
the menu "Calibration Deviations".
TAG
50.00 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1 F.S.
0.0 100.0
Single component
display appears after spanning is done
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 45
4 - 46
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
37.50 ppm CH4
4.7 Flow Zero Gas
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
37.50 ppm
Valves...
F5
Press
Caution:
Before starting zero gas flow, realize that zero gas is available ! (See also section 5.1.1, p.5-5... !)
Change to the
menu "Set Gas Valves"
F5
Press
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
Start the zero
gas flow
Press
or
Starting this function will switch on the zero gas valve as current gas valve.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 47
TAG
-- Confirmation Required --
37.50 ppm
Confirm the order to
flow zero gas
F2
Press to start the zero gas flow immediately.
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
F1 F2 F3 F4 F5
TAG
- S U C C E S S -
- The selected function has been started/executed ­(Wait a moment...)
37.50 ppm
Option: Press the F4 key if you want to cancel the order and go back to the menu "Set Gas Valves".
Note: The display of this message depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
Confirmation
message of the function start appears
Note: This message will be displayed after the function has been started. Then the display will jump automatic­ally to the menu "Set Gas Valves".
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Zerogas Operation status: Ready
Measure Status... Channel Back...
0.00 ppm
F1 F2 F3 F4 F5
Change to the single
component display of the current channel
Options:
Start the gasflow with
span gas, sample gas or test gas.
Close all valves.
F3 key: Changing to
another available channel to execute gas flow.
F4 key: Go back to the
menu "Analyzer Module Calibration" to start zeroing or spanning.
4 - 48
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
0.00 ppm CH4
4.7 Flow Zero Gas
Single component
display appears during the zero gas flow is running
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 49
4 - 50
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
37.50 ppm CH4
4.8 Flow Span Gas
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
37.50 ppm
Valves...
F5
Press
Caution:
Before starting span gas flow, realize that spangas is available ! (See also section 5.1.1, p. 5-5... !)
Change to the
menu "Set Gas Valves"
F5
Press
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Flow span gas !"
Press once to get the line "Flow span gas !" white on black.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 51
TAG
Flow zero gas !
Flow span gas !
Flow sample gas ! Flow test gas !
-- Set Gas Valves --
37.50 ppm
Start the span
gas flow
Press
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
or
Starting this function will switch on the span gas valve as current gas valve.
Confirm the order to
flow span gas
F2
Press to start the span gas flow immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Set Gas Valves".
Note: The display of this message depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
TAG
- S U C C E S S -
- The selected function has been started/executed -
(Wait a moment...)
F1 F2 F3 F4 F5
4 - 52
37.50 ppm
NGA 2000
Confirmation
message of the function start appears
Note: This message will be displayed after the function has been started. Then the display will jump automatic­ally to the menu "Set Gas Valves".
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Spangas-1 Operation status: Ready
Measure Status... Channel Back...
50.00 ppm
F1 F2 F3 F4 F5
4.8 Flow Span Gas
Change to the single
component display of the current channel
Options:
Start the gasflow with
zero gas, sample gas or test gas.
Close all valves.
F3 key: Changing to
another available channel to execute gas flow.
F4 key: Go back to the
menu "Analyzer Module Calibration" to start zeroing or spanning.
TAG
50.00 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1 F.S.
0.0 100.0
Single component
display appears during the span gas flow is running
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 53
4 - 54
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
0.00 ppm CH4
4.9 Flow Sample Gas
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Zerogas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
0.00 ppm
Valves...
F5
Press
Change to the
menu "Set Gas Valves"
F5
Press
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Zerogas Operation status: Ready
Measure Status... Channel Back...
0.00 ppm
F1 F2 F3 F4 F5
Change to the line
"Flow sample gas !"
Press or as often as necessary to get the line "Flow sample gas !" white on black.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 55
TAG
Flow zero gas ! Flow span gas !
Flow sample gas !
Flow test gas !
-- Set Gas Valves --
0.00 ppm
Start the sample
gas flow
Press
Close all valves !
Valve position: Zerogas Operation status: Ready
Measure Status... Channel Back...
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
0.00 ppm
F1 F2 F3 F4 F5
or
Starting this function will switch on the sample gas valve as current gas valve.
Confirm the order to
flow sample gas
F2
Press to start the sample gas flow immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Set Gas Valves".
Note: The display of this message depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
TAG
- S U C C E S S -
- The selected function has been started/executed -
(Wait a moment...)
F1 F2 F3 F4 F5
4 - 56
0.00 ppm
NGA 2000
Confirmation
message of the function start appears
Note: This message will be displayed after the function has been started. Then the display will jump automatic­ally to the menu "Set Gas Valves".
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
4.9 Flow Sample Gas
Change to the single
component display of the current channel
Options:
Start the gasflow with
zero gas, span gas or test gas.
Close all valves.
F3 key: Changing to
another available channel to execute gas flow.
F4 key: Go back to the
menu "Analyzer Module Calibration" to start zeroing or spanning.
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Single component
display appears during the sample gas flow is running
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 57
4 - 58
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
37.50 ppm CH4
4.10 Flow Test Gas
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
37.50 ppm
Valves...
F5
Press
Starting this function will switch an external valve and flow the MLT with test gas.
Change to the
menu "Set Gas Valves"
F5
Press
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Flow test gas !"
Press or as often as necessary to get the line "Flow test gas !" white on black.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 59
TAG
Flow zero gas ! Flow span gas ! Flow sample gas !
Flow test gas !
-- Set Gas Valves --
37.50 ppm
Start the function to
flow test gas
Press
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
or
Confirm the order
to flow test gas
F2
Press to start the test gas flow immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Set Gas Valves".
Note: The display of this message depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
TAG
- S U C C E S S -
- The selected function has been started/executed -
(Wait a moment...)
F1 F2 F3 F4 F5
4 - 60
37.50 ppm
NGA 2000
Confirmation
message of the function start appears
Note: This message will be displayed after the function has been started. Then the display will jump automatic­ally to the menu "Set Gas Valves".
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Testgas Operation status: Ready
Measure Status... Channel Back...
20.00 ppm
F1 F2 F3 F4 F5
4.10 Flow Test Gas
Change to the single
component display of the current channel
Options:
Start the gasflow with
zero gas, span gas or sample gas.
Close all valves.
F3 key: Changing to
another available channel to execute gas flow.
F4 key: Go back to the
menu "Analyzer Module Calibration" to start zeroing or spanning.
TAG
20.00 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Single component
display appears during the test gas flow is running
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 61
4 - 62
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
37.50 ppm CH4
4.11 Close all Valves
Change to the
menu "Analyzer Module Calibration"
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
Range: 1
0.0 100.0
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Calibration --
Calibration procedure status...
Start zero calibration procedure ! Start span calibration procedure !
Time & Date: 12:50:55 June 16, 1998 Valve position: Samplegas Expected zero gas: 0.00 ppm Expected span gas: 50.00 ppm Span gas name: Methane
Measure Status... Channel Back...
37.50 ppm
Valves...
F5
Press
Change to the
menu "Set Gas Valves"
F5
Press
F1 F2 F3 F4 F5
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
Change to the line
"Close all valves !"
Press or as often as necessary to get the line "Close all valves !" white on black.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 63
TAG
Flow zero gas ! Flow span gas ! Flow sample gas ! Flow test gas !
-- Set Gas Valves --
37.50 ppm
Start the function to
close all valves
Press
Close all valves !
Valve position: Samplegas Operation status: Ready
Measure Status... Channel Back...
F1 F2 F3 F4 F5
TAG
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
Yes Back...
37.50 ppm
F1 F2 F3 F4 F5
or
Confirm the order
to close all valves
F2
Press to close all valves immediately.
Option: Press the F4 key if you want to cancel the order and go back to the menu "Set Gas Valves".
Note: The display of this message depends on the setup in the expert configurations. (see 5.1.8 p. 5-49)
TAG
- S U C C E S S -
- The selected function has been started/executed -
(Wait a moment...)
F1 F2 F3 F4 F5
4 - 64
37.50 ppm
NGA 2000
Confirmation
message of the function start appears
Note: This message will be displayed after the function has been started. Then the display will jump automatic­ally to the menu "Set Gas Valves".
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
TAG
-- Set Gas Valves --
Flow zero gas !
Flow span gas ! Flow sample gas ! Flow test gas !
Close all valves !
Valve position: AllClosed Operation status: Ready
Measure Status... Channel Back...
37.50 ppm
F1 F2 F3 F4 F5
4.11 Close all Valves
Change to the single
component display of the current channel
Options:
Start the gasflow with
zero gas, span gas, sample gas or test gas.
F3 key: Changing to
another available channel to execute gas flow.
F4 key: Go back to the
menu "Analyzer Module Calibration" to start zeroing or spanning.
TAG
37.50 ppm CH4
0.00 50.00
Failures: No Maintenance-Requests: No Temperature: 20.0 C Operation: Ready
Display Status... Main... Channel Calib...
F1 F2 F3 F4 F5
Range: 1
0.0 100.0
Single component
display appears in the standby mode: All valves are closed
Note: After the valves are closed you can start again the measurement with the "Flow sample gas" order (see 4.9).
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 65
4 - 66
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

5 Analyzer and I/O-Module Expert Configuration

If you press the 88 -key or the → -key in the line "Analyzer and I/O-module expert configuration..." of the "Main Menu", you will change to the following display:
TAG
-- Expert Module Configuration --
Analyzer module controls...
I/O module controls... Analyzer module setup...
I/O module setup...
Measure Channel Back...
F1 F2 F3 F4 F5
37.50 ppm
From the menu "Expert Module Configuration" you can change to several submenus to set up parameters for the measurement and calibration of your analyzer and analyzer module. Besides you can set up the configuration for auxiliary modules. Which part of these menus are important for you depends on the configuration of your NGA 2000 system. In the following table you will find a short overview about the contents of the menus:
Menu Important Contents Section/Page
Analyzer module controls... See "Analyzer module setup" and note below! I/O module controls...
Configuration of the SIO and of the DIO's
installed in a platform or
5.2
p. 83 - 98
in an MLT analyzer (auxiliary I/O's)
Analyzer module setup...
Measurement set up and controlCalibration set up and controlConfiguration of the SIO and of the DIO's
5.1
p. 3 - 82
installed in the MLT analyzer module (Local SIO/DIO)
Programmable logic control (PLC)Programmable calculators
I/O module setup...
Set up for micro processor controlled
5.3 / p. 99
auxiliary I/O modules
Note: All submenus of "Analyzer module controls..." are the same like those of "Analyzer module
setup...", if you use an MLT analyzer or analyzer module (resp. TFID analyzer or analyzer module)! If your analyzer is not such a type, the submenus of "Analyzer module controls..." and "Analyzer module setup..." will look different, partially. Look at the corresponding software manuals.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 1
Structure of chapter five:
At the begin of any chapter you will find the way to a certain submenu of the MLT software starting from the line "Analyzer and I/O-module expert configuration" in the "Main Menu". The way will be described with the software catchwords, that you have to enter one after another to reach the corresponding submenu. At the end of the catchword listing you will find the illustration of the LCD screen. After that you can read the set up instructions and explanations. Sometimes the illustration will be completed by sketches or further menu pictures.
Example: You will set up the calibration parameters for the calibration procedure of the
analyzer module.
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Calibration parameters
↓↓
Calibration procedure setup
↓↓
TAG
-- Calibration Procedure Setup --
Purge time: 10 s
Procedure times-out after: 120 s Analog output during calibration: Tracking Span ranges: Together
Valve position: Samplegas
Stability time: 30 s Averaging time: 5 s
Measure Channel Back...
F1 F2 F3 F4 F5
In the menu "Calibration Procedure Setup" you can set up the parameters for the cali­bration procedure of the zeroing and spanning ...
95.00 ppm
Further explanations and instructions will follow !
5 - 2
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
5.1 Analyzer Module Setup
Main Menu — Analyzer and I/O-module expert configuration
↓↓

Analyzer module setup

↓↓
TAG
-- Analyzer Module Setup and Controls (1/3) --
Calibration parameters...
Alarm parameters... Range parameters... Cross interference compensation... Linearization... Programmable logic control (PLC)... Programmable calculator... Measurement display configuration... Acknowledgement of status reports... Concentration measurement parameters...
Measure ManData Channel Back... More...
95.00 ppm
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Setup and Controls (2/3) --
Peak measurement...
Differential measurement... Gasflow setup... Pressure compensation... Flow measurement... Temperature measurement... Loading/saving configuration parameters... Inputs and outputs... Delay and average... Special functions...
Measure Channel Back...
95.00 ppm
More...
F1 F2 F3 F4 F5
TAG
-- Analyzer Module Setup and Controls (3/3) --
AK-protocol communication...
Measure Back...
F1 F2 F3 F4 F5
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
95.00 ppm
5 - 3
From the menu "Analyzer Module Setup and Controls" you can change to all menus and submenus of the expert configuration level via the corresponding menu lines.
Changing to the menus and submenus:
Select the menu line you want with the ↓↓ -key or the ↑↑ -key.
If necessary: Change to a further menu page with the F5 key (More...)
Change to the corresponding menu/submenu with the 88 -key or the -key.
Menu "Load Factory Configuration":
If you press the F2 key (ManData) in the menu "Analyzer Module Setup and Controls (1/3)", you will change to the following menu:
TAG
-- Load Factory Configuration --
- BE CAREFUL with this function -
Replace current configuration with factory settings !
Measure Back...
F1 F2 F3 F4 F5
95.00 ppm
With the function of this menu you can delete the RAM data and load the factory settings from the Flash-EPROM.
Caution: This function is irreversible after starting (and confirmation) ! Your RAM settings will be deleted !
Note:
The same function is available in the menu "Load/Save Analyzer Module Configuration" ! (see section 5.1.17 p. 5-64)
Starting the function:
Press the 88 -key or the -key or
cancel and go back to the last menu page with the F4 key (Back...) or the ← -key.
If asked: Confirm with the F2 key (Yes) or
cancel and go back to the last menu page with the F4 key (Back...) or the -key.
Further Ramifications via softkeys:
F1
F3
5 - 4
Change to the single component display of the current channel ! Change to the other available channels
of an MLT analyzer or analyzer module !
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Chapter 4 !
+
See channel tag !
+
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓

Calibration parameters

↓↓
TAG
-- Calibration Parameters --
Span gases...
Tolerances... Calibration procedure setup... Time controlled calibration... Calibration... Advanced calibration methods... Zero gases...
Measure Channel Back...
F1 F2 F3 F4 F5
37.50 ppm
From the menu "Calibration Parameters" you can change to several submenus to set up the zero and span gas calibration parameters and to start the different calibration methods.
Change to the submenus:
Select the menu line you want with the ↓↓ -key or the ↑↑ -key.Change to the corresponding submenu with the 88 -key or the → -key.
Alternatives:
F1: Change to the single component display !F3: Change to the other available channels of an MLT analyzer or
analyzer module !
F4: Go back to the last menu page !
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 5
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Span gases
Main Menu — Analyzer and I/O-module expert configuration
Analyzer module setup
Calibration parameters
Span gases
↓↓ ↓↓ ↓↓ ↓↓
MLT25/CH2/R2
-- Span Gas Definition --
Actual span gas concentration: 250.0 ppm
Spangas range -1: 50.00 ppm Spangas range -2: 250.0 ppm Spangas range -3: 2500 ppm Spangas range -4: 10000 ppm
Span gas unit: ppm "ppm" --> "mg/Nm3" conversion factor: 1.000
Concentration in span gas units: 95.00 ppm
Measure Back...
F1 F2 F3 F4 F5
95.00 ppm
In the menu "Span Gas Definition" you can set up the default value for each range of the current channel, the actual span gas concentration and the desired span gas unit. The concentration of each range should be a value between 70 and 110% of the end of range value.
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the 88 -key or the → -key.
Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the 88 -key or
cancel and go back to the last value with the F2 key.
The last two lines of variables: These lines are only information lines. The variables cannot be changed.
Note:
The "ppm "mg/Nm3" conversion factor can be set up in the menu "General Concentration Measurement Setup" (see section 5.1.10 p. 5-54) !
5 - 6
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Tolerances
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Calibration parameters
↓↓
Tolerances
↓↓
TAG
-- Tolerances --
Max. zero calibration deviation: 20.00 %
Max. span calibration deviation: 20.00 % Check deviation: Disabled Stability tolerance - range 1: 100.00 % Stability tolerance - range 2: 100.00 % Stability tolerance - range 3: 100.00 % Stability tolerance - range 4: 100.00 %
Last zero calibration: Success Last span calibration: Success
Measure Channel Back...
F1 F2 F3 F4 F5
95.00 ppm
In the menu "Tolerances" you can set up the parameters of the stability controlled calibration procedure (see p. 5-9/10) in one channel. Besides, you can disable the stability control and the maximum calibration deviation (deviation control deviation tolerance). If you want to set up the parameters of the other channels of an MLT analyzer or analyzer module, press the F3 key.
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the 88 -key or the → -key.
Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the 88 -key or
cancel and go back to the last value with the F2 key.
Lines "Last zero calibration/Last span calibration": The calibration status of the last calibration will be displayed in these two lines.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 7
Lines of variables "Max. zero calibration deviation" / "Max. span calibration deviation": The deviation between two zero or span gas concentrations will be determined during the
stability time of the calibration procedure (see menu "Calibration Procedure Setup" p. 5-9/10). The maximum calibration deviation is the deviation tolerance expressed as percentage of the current end of range value. The calibration procedure will only be possible within this range. If you choose a very small deviation tolerance, perhaps the calculated deviation is out of range during the whole calibration procedure and the calibration will run into time out. If you choose a very large deviation tolerance, the calibration will also be done in ranges with a little stability. Value is usually: 10% or 20%
Line of variables "Check deviation":
Enabled: The stability and tolerance control during the calibration procedure will be
based on the values of the menu "Tolerances".
Disabled: The calibration procedure is running without stability and tolerance control.
Any Calibration will be accepted.
Example:
End of range: 1000 ppmMax. calibration deviation: 20% (equivalent 200 ppm)Desired value: 990 ppmDisplay: 720 ppmDeviation: 250 ppm (more than 200 ppm!)Result: Max. calibration procedure time over limit;
Calibration will be canceled !
Potential solutions:
Adjust 100% in the line "Max. zero calibration deviation"
resp. "Max. span calibration deviation" or select "Disabled" in the line "Check deviation".
Result: Any Calibration will be accepted!
5 - 8
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Calibration procedure setup
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Calibration parameters
↓↓
Calibration procedure setup
↓↓
TAG
-- Calibration Procedure Setup --
Purge time: 10 s
Procedure times-out after: 120 s Analog output during calibration: Tracking Span ranges: Together
Valve position: Samplegas
Stability time: 30 s Averaging time: 5 s
Measure Channel Back...
F1 F2 F3 F4 F5
95.00 ppm
In the menu "Calibration Procedure Setup" you can set up the parameters for the cali­bration procedure of the zeroing and spanning in one channel. If you want to set up the parameters of the other channels of an MLT analyzer or analyzer module, press the F3 key.
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the 88 -key or the → -key.
Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the 88 -key or
cancel and go back to the last value with the F2 key.
Lines of variables "Purge time" / "Procedure times-out after" and "Stability time" / "Averaging time" – Informations about the calibration procedures setup:
The measurement signal should be stable in a certain range before zeroing and spanning. Therefore the analyzer cell should be purged enough with zero gas or span gas before calibration. Then, after the purge time, the stability time is running. During the stability time the average over two signals will be calculated. The averaging time determines the time difference between these two signals.
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 9
If the deviation of the two values is less then the allowed maximum calibration deviation (see menu "Tolerances", p. 5-7/8), the calibration may start. If not, the stability control is starting again until a stable measurement is possible. The time in the line "Procedure times-out after" is the limit for the whole procedure. The following figure shows the procedure of the stability controlled zero and span gas calibration:
Start Span Calibration
Concentration
Start Zero
Calibration
Averaging
over T
T
i
i
Purge
Time
Stability
Time
Stability
Time
Stability
Time
Stability
Time
Time
Note:
The stability time and the averaging time are factory settings and cannot be changed by the customer. If you want to use other values, ask your service !
Line of variables "Analog output during calibration": Here, you can determine if you want to hold the analog output and the limits of a local SIO
in the MLT analyzer module during calibration. The set-ups for a platform or MLT analyzer SIO have to be done in the menu "Analog Outputs" (see 5.2.1 p. 5-87).
Tracking: The analog output signal is following to the signal during the whole calibra-
tion procedure.
Holding: During the whole calibration procedure the analog output signal is fixed to the
last signal before calibration. This adjustment may be used for instance to control limits with the analog output.
Sample Gas Zeroing Spanning Sample Gas
5 - 10
Analog Output
Tracking
Analog Output
Holding
0V
NGA 2000
xV
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Calibration procedure setup
Line of variables "Span ranges":
Together:
All measurement ranges of the same channel will be calibrated together. This is the usual calibration mode.
Separately:
Each measurement range will be calibrated separately.
Line of variables "Valve position":
Options:
SamplegasZerogasPurgegasTestgasSpangas-1...-4LinearizerSpangasAllClosedOther-Proc.Basic-Status
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 11
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Time controlled calibration
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Calibration parameters
↓↓
Time controlled calibration
↓↓
TAG
-- Time Control --
Zero calibration - Day of week: Never
- Hour: 0
- Minute: 0 Span calibration - Day of week: Never
- Hour: 0
- Minute: 0 Zero+Span calibration - Day of week: Never
- Hour: 0
- Minute: 0
Measure Channel Back...
F1 F2 F3 F4 F5
37.50 ppm
ShowList...
In the menu "Time Control" you can determine the starting time of the zero gas calibration resp. the span gas calibration for one channel. If you want to determine the values of the other channels of an MLT analyzer or analyzer module, press the F3 key.
Notes:
The time controlled calibration is only possible if the span gas comes from a valve gear!
If the valve gear is not available you have to select "Never" in all lines containing "Day of week".
A zero gas calibration should be done before starting a span gas calibration. Therefore
the starting time of the zero gas calibration should be different from the starting time of the span gas calibration. If you select the option "Zero+Span calibration" the zeroing will run before the spanning, automatically.
Set up time parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the 88 -key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Options for day of week: Monday, ..., Sunday, Each day, Never
for hour: 0, 1, 2, 3, ..., 23 for minute: 0, 1, 2, 3, ..., 59
Confirm the new value with the 88 -key or cancel and go back to the last value with F2.
5 - 12
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Further Parameters:
Via the F5 key (ShowList) you can change to the following submenu:
TAG
-- Auto-Start Procedures --
Position in auto-start list: 1
Erase current position from list ! Erase all positions from list !
Channel tag: ­Procedure type: ­Interval mode: Never Start time: ­Start date: -
Measure Back...
F1 F2 F3 F4 F5
37.50 ppm
In this submenu you can determine some further parameters of the time controlled calibration.
Lines "Position in auto-start list", "Erase current position from list !" and "Erase all positions from list !":
If you have disabled a time controlled calibration in the menu "Time Control" you have to reset the memory to avoid further calibrations by time control. Starting the function "Erase current position from list !" will reset the memory for the position selected in the line "Position in auto-start list".
There are three positions belonging to the following lines in the menu "Time Control": Position 1: Zero calibration
Position 2: Span calibration Position 3: Zero+Span calibration
Starting the function "Erase all positions from list !" will delete the set-ups for all kinds of time controlled calibration in the menu "Time Control".
The last five lines of variables: These lines will display the parameters of the position selected in the line "Position in
auto-start list".
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 13
5 - 14
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Loading...