Emerson HAS60E-IM-HW User Manual

HAS60E-IM-HW 06/2007
Instruction Manual
Chemiluminescence Detector CLD Analyzer Module (combined with NGA 2000 Platform, ML T, CA T 200 or TFID Analyzer)
st
Edition 06/2007
1
www.EmersonProcess.com
CLD Chemiluminescence Detector Instruction Manual
HAS60E-IM-HW
06/2007

ESSENTIAL INSTRUCTIONS

READ THIS P AGE BEFORE PROCEEDING!
Emerson Process Management (Rosemount Analytical) designs, manufactures and test s its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using and maintaining Emerson Process Management (Rosemount Analytical) products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson Process
Management (Rosemount Analytical) representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and
maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate
Instruction Manual and per applicable local and national codes. Connect all products
to the proper electrical and pressure sources.
T o ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson Process Management (Rosemount Analytical). Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR W ARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice. Misprints reserved.
1st Edition 06/2007 © 2007 by Emerson Process Management
Emerson Process Management GmbH & Co. OHG
Industriestrasse 1 D-63594 Hasselroth Germany T +49 (0) 6055 884-0 F +49 (0) 6055 884-209 Internet: www.EmersonProcess.com
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

Table of Contents

PREFACE ...........................................................................................................................P - 1
DEFINITIONS .....................................................................................................................P - 1
SAFETY INSTRUCTIONS WIRING AND INSTALLATION OF THIS APPARATUS.......P - 2
OPERATING AND MAINTAINING THIS APPARATUS .......................................................P - 3
SAFETY SUMMARY ........................................................................................................... P - 4
AUTHORIZED PERSONNEL ..............................................................................................P - 4
GASES AND GAS CONDITIONING (SAMPLE HANDLING) .............................................. P - 7
POWER SUPPLY................................................................................................................ P - 7
ELECTROSTATIC DISCHARGE ......................................................................................... P - 8
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE GAS
CYLINDERS ...................................................................................................................P - 9
DOCUMENTATION ...........................................................................................................P - 10
COMPLIANCES ................................................................................................................P - 10
SUITABILITY TESTS ........................................................................................................P - 10
GLOSSARY OF TERMS ................................................................................................... P - 11
ANALYZER SYSTEM ARCHITECTURE ........................................................................... A - 1
SECTION 1 DESCRIPTION AND SPECIFICATIONS ........................................................ 1 - 1
1-1 OVERVIEW ................................................................................................................ 1 - 1
1-2 TYPICAL APPLICATIONS .......................................................................................... 1 - 1
1-3 THEORY OF TECHNOLOGY ..................................................................................... 1 - 1
1-4 FEATURES ................................................................................................................. 1 - 1
1-5 SPECIFICATIONS ...................................................................................................... 1 - 5
a. General ................................................................................................................... 1 - 5
b. Physical .................................................................................................................. 1 - 5
c. Sample ................................................................................................................... 1 - 5
d. Gas Connections .................................................................................................... 1 - 5
ITable of ContentsEmerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW
NGA 2000 CLD
June 2007
SECTION 2 INSTALLATION .............................................................................................. 2 - 1
2-1 UNPACKING ............................................................................................................... 2 - 1
2-2 ASSEMBLY................................................................................................................. 2 - 1
2-3 LOCATION .................................................................................................................. 2 - 2
2-4 GASES ....................................................................................................................... 2 - 2
a. Gas Conditioning (Sample Handling) ..................................................................... 2 - 2
b. Connections............................................................................................................ 2 - 3
c. Specifications ......................................................................................................... 2 - 3
Zero Gas ..............................................................................................................................2 - 3
Span Gas .............................................................................................................................2 - 3
Ozonator Source Gas .......................................................................................................... 2 - 3
Sample Gas Pressure ..........................................................................................................2 - 3
Bypass Sample Gas Flow ....................................................................................................2 - 3
Leak Test ..............................................................................................................................2 - 4
Contaminants .......................................................................................................................2 - 4
2-5 ELECTRICAL CONNECTIONS .................................................................................. 2 - 4
SECTION 3 OPERATION .................................................................................................. 3 - 1
3-1 OVERVIEW ................................................................................................................ 3 - 1
3-2 STARTUP & INITIALIZATION ..................................................................................... 3 - 1
3-3 PREPARATIONS ........................................................................................................ 3 - 2
3-4 CALIBRATION PROCEDURE .................................................................................... 3 - 4
a. Calibration Setup .................................................................................................... 3 - 4
Calibration Gas List ..............................................................................................................3 - 4
Calibration Parameters ........................................................................................................ 3 - 5
b. Flow Balance Adjustment: ...................................................................................... 3 - 6
Zero Adjustment ...................................................................................................................3 - 6
Span Adjustment .................................................................................................................. 3 - 6
3-5 SYSTEM & NETWORK I/O MODULE CONTROLS (SETUP) - SYSTEM SIO .......... 3 - 8
a. Analog Output Setup .............................................................................................. 3 - 8
Output number: .................................................................................................................... 3 - 8
Choose signal source module... ........................................................................................... 3 - 8
Choose Signal... ...................................................................................................................3 - 8
Signal value for 0% (100%) output: ..................................................................................... 3 - 9
Output current: ..................................................................................................................... 3 - 9
Hold output during calibration: ............................................................................................. 3 - 9
b. Serial interface Setup ........................................................................................... 3 - 11
II Table of Contents Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
c. Relay Outputs Setup ............................................................................................ 3 - 12
Output number: .................................................................................................................. 3 - 12
Invert signal: .......................................................................................................................3 - 12
Choose source module... ...................................................................................................3 - 12
Choose signal... .................................................................................................................3 - 12
3-6 CONVERTER TEMPERATURE ADJUSTMENT ...................................................... 3 - 13
3-7 MEASUREMENT OF CONVERTER EFFICIENCY .................................................. 3 - 14
a. Test Setup for Measurement of Conversion Efficiency ......................................... 3 - 14
b. Test Procedure ..................................................................................................... 3 - 14
c. Subnormal Conversion Efficiency ......................................................................... 3 - 15
d. Replacement of Converter .................................................................................... 3 - 16
e. Capillaries ............................................................................................................. 3 - 16
f. TEA Scrubber ....................................................................................................... 3 - 16
SECTION 4 MAINTENANCE AND SERVICE .................................................................... 4 - 1
4-1 OVERVIEW ................................................................................................................ 4 - 1
4-2 FUSES ........................................................................................................................4 - 2
4-3 FANS .......................................................................................................................... 4 - 2
4-4 OZONATOR ................................................................................................................ 4 - 2
4-5 PRINTED CIRCUIT BOARDS .................................................................................... 4 - 2
4-6 CONVERTER ............................................................................................................. 4 - 4
4-7 DETECTOR DISASSEMBLY ...................................................................................... 4 - 5
a. Reaction Chamber Removal .................................................................................. 4 - 5
b. Reaction Chamber Installation ............................................................................... 4 - 5
c. Photodiode Removal .............................................................................................. 4 - 5
d. Photodiode Installation ........................................................................................... 4 - 5
4-8 LEAKAGE TEST ......................................................................................................... 4 - 7
a) Required Tools........................................................................................................ 4 - 7
b) Procedure ............................................................................................................... 4 - 7
SECTION 5 TROUBLESHOOTING ................................................................................... 5 - 1
5-1 OVERVIEW ................................................................................................................ 5 - 1
IIITable of ContentsEmerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW
NGA 2000 CLD
June 2007
SECTION 6 REPLACEMENT PARTS ............................................................................... 6 - 1
6-1 MATRIX ...................................................................................................................... 6 - 1
6-2 REPLACEMENT PARTS ............................................................................................ 6 - 2
SECTION 7 RETURN OF MATERIAL ............................................................................... 7 - 1
7-1 RETURN OF MATERIAL ............................................................................................ 7 - 1
7-2 CUSTOMER SERVICE............................................................................................... 7 - 1
7-3 TRAINING................................................................................................................... 7 - 1
LIST OF FIGURES AND TABLES ...................................................................................... L - 1
1 LIST OF FIGURES ..................................................................................................... L - 1
2 LIST OF TABLES ....................................................................................................... L - 2
IV Table of Contents Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

PREFACE

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the NGA 2000 CLD and the System Accessories of the NGA 2000 System.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS

The following definitions apply to WARNINGS, CAUTIONS and NOTES found throughout this publication.
WARNING
Highlights on operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION
Highlights on operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effective­ness.
NOTE
Highlights an essential operating procedure, condition or statement.
P - 1PrefaceEmerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW
NGA 2000 CLD
June 2007
IMPORTANT
SAFETY INSTRUCTIONS
WIRING AND INSTALLATION OF THIS APPARATUS
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non-EU states should also comply with the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points, internal and external, where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced. The integrity of all earth terminals must be maintained at all times.
3. To ensure safe operation of this equipment, connection to the mains supply should only be made through a circuit breaker which will disconnect breaker may also include a mechanically operated isolating switch. Circuit breakers or switches must comply with a recognized standard such as IEC947. All wiring must conform with any local standards.
all circuits carrying conductors during a fault situation. The circuit
4. Where equipment or covers are marked with the symbol to the right, hazardous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment — and then by trained service personnel only.
5. Where equipment or covers are marked with the symbol to the right, there is a danger from hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Certain surfaces may remain hot to the touch.
6. Where equipment or covers are marked with the symbol to the right, refer to the In­struction Manual for instructions.
7. Further graphical symbols used in this product:
Elektrostatic discharge (ESD)
Harmful (to Health)!
Explosion Hazard!
Toxic!
UV Source!
Disconnect from Mains!
All graphical symbols used in this product are from one or more of the following standards: EN61010-1, IEC417, and ISO3864.
P - 2 Preface Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

OPERATING AND MAINTAINING THIS APPARATUS

This instrument has left the factory in compliance with all applicable safety regulations.
To maintain this operating condition, the user must strictly follow the instructions and consider the warnings in this manual or provided on the instrument.
Before switching on the instrument, verify that the electrical supply voltage matches the instrument´s opera­ting voltage as set in the factory.
Any interruption in the instrument´s ground line, whether inside or outside the instrument, or removal or interruption of its ground line connection, could result in hazardous operating conditions. Intentionally interrup­ting the instrument´s protective ground is strictly prohibited.
Opening cover panels could expose voltage-carrying components. Connectors may also be under voltage. The instrument must be disconnected from all electrical supplies before attempting any calibrations, mainte­nance operations, repairs or component replacements requiring opening of the instrument. Any calibrations, maintenance operations, or repairs that need the instrument to be opened while connected to electrical supplies should be subject to qualified technicians familiar with the hazards involved only!
Use only fuses of the correct type and current ratings as replacements. Using repaired fuses and short circuiting of fuse holders is prohibited.
Observe all applicable regulations when operating the instrument from an auto-transformer or variac.
Substances hazardous to health may emerge from the instrument‘s exhaust.
Please pay attention to the safety of your operation personnel. Protective measures must be taken, if re­quired.
NOTE
Software compatibility is necessary for all NGA 2000 components in your system to work together. The version of your Platform’s software must be equal to or greater than the version of any other module(s) for successful compatibility.
You can locate the version of each NGA 2000 component as follows:
Platform Controller Board
Turn power ON. The display should show „Control Module V3. ...“. This is the software version.
Analyzer Module
See note on the name plate label located on the right side of the Analyzer Module case.
P - 3PrefaceEmerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW
NGA 2000 CLD
June 2007

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.

AUTHORIZED PERSONNEL

To avoid loss of life, personal injury and damage to this equipment and on-site property, do not operate or service this instrument before reading and understanding this instruction manual and receiving appropriate training. Save these instructions.
WARNING
ELECTRICAL SHOCK HAZARD !
Do not operate without covers secure. Do not open while energized. Installation and/or servicing requires access to live parts which can cause death or serious injury.
Refer servicing to qualified personnel.
For safety and proper performace this instrument must be connected to a properly grounded three­wire source of power.
WARNING
TOXIC AND OXIDIZING GAS
This module generates ozone which is toxic by inhalation and is a strong irritant to throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor.
The module’s exhaust contains both ozone and nitrogen dioxide, both toxic by inhalation, and may contain other constituents of the sample gas which may be toxic. Such gases include various oxides of nitrogen, unburned hydrocarbons, carbon monoxide and other products of combustion reactions. Carbon monoxide is highly toxic and can cause headache, nausea, loss of conscious­ness, and death.
Avoid inhalation of the ozone produced within the module, and avoid inhalation of the sample and exhaust products transported within the module. Avoid inhalation of the combined exhaust prod­ucts at the exhaust fitting.
Keep all tube fittings tight to avoid leaks. The user is responsible for leakage testing only at the inlet and outlet fittings on the rear panel.
Connect rear exhaust outlet to outside vent with stainless steel or Teflon line. Check vent line and connections for leakage.
P - 4 Preface Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
WARNING
OVERBALANCE HAZARD
This analyzer module may tip instrument over if it is pulled out too far and the Platform is not properly supported.
CAUTION
Tampering with or unauthorized substitution of components may adversely affect the safety of this instrument. Use only factory documented/approved components for repair.
Because of the danger of introducing additional hazards, do not perform any unauthorized modifi­cation to this instrument!
WARNING
POSSIBLE EXPLOSION HAZARD
This equipment is not designed and should not be used in the analysis of flammable samples. Use of this equipment in this way could result in explosion or death.
WARNING
POSSIBLE EXPLOSION HAZARD
Ensure that all gas connections are made as labeled and described within this manual and leak free. Improper gas connections may cause explosion, serious injury or death.
CAUTION
HIGH TEMPERATURES !
While working at thermostated components inside the analyzer modules hot components may be accessible!
CAUTION
Do not interchange gas inlets and outlet! All gases must be conditioned before supplying! When supplying corrosive gases ensure that gas path components are not affected!
Exhaust lines must be installed in a descending way, need to be pressureless, frost-protected and in compliance with applicable legislative requirements!
P - 5PrefaceEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
WARNING
Before opening gas paths they must be purged with ambient air or neutral gas (N or harmful to health sample gas components!
Ultraviolet light from the ozone generator can cause permanent eye damage ! Do not look directly at the ultraviolet source ! Use of ultraviolet filtering glases is recommended.
) to avoid hazards caused by toxic, flammable, explosive
2
WARNING
INTERNAL UV SOURCE !
Instruction Manual
HAS60E-IM-HW
June 2007
WARNING
TOXIC CHEMICAL HAZARD!
The optional UV lamp contains mercury. Lamp breakage could result in mercury exposure ! Mercury is highly toxic if absorbed through the skin or ingested, or if vapors are inhaled.
Handle lamp assembly with extreme care. If the lamp is broken, avoid skin contact and inhalation in the area of the lamp or the mercury spill.
Immediately clean up and dispose of the mercury spill and lamp residue as follows:
Wearing rubber gloves and goggles, collect all droplets of mercury by means of a suction pump and aspirator bottle with a long capillary tube. (Alternately, a commercially available mercury spill clean-up kit is recommended.)
Carefully sweep any remaining mercury and lamp debris into a dust pan. Carefully transfer all mercury, lamp residue and debris into a platic bottle which cab be tightly capped.
Label and return to hazardous material reclamation center. Do not place in the trash, incinerate or flush down the sewer.
Cover any fine droplets of mercury in non-accessible crevices with calcium polysulfide and sulfur dust.
P - 6 Preface Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

GASES AND GAS CONDITIONING (SAMPLE HANDLING)

WARNING
Take care of the safety instructions applicable for the gases (sample gases, test gases and ozonator air)!
CAUTION
PRESSURIZED GAS
This module requires periodic use of pressurized gas. See General Precautions for Handling and Storing High Pressure Gas Cylindes, page P-6.
CAUTION
EXTERNALLY RESTRICT SAMPLE FLOW TO LESS THAN 2,200 cc/min.
No restrictor is provided in the inlet of this module. For those users who cannot externally limit sample flow, contact your local service or sales office.

POWER SUPPLY

CAUTION
Verify the power voltage at site of installation corresponds to the analyzer module´s rated voltage as given on the nameplate label!
Verify the safety instruction given by power supply unit manufacturer !
CAUTION
The mains socket has to be nearby the power supply unit and easily accessible! Disconnecting from power requires unplugging the power connector!
To comply with the CE mark requirements use only power supply units of type SL10 or equivalent units. Equivalent units must provide SELV output voltages!
Verify proper polarity when connecting DC 24 V operated analyzer modules !
P - 7PrefaceEmerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW
NGA 2000 CLD
June 2007

ELECTROSTATIC DISCHARGE

CAUTION
The electronic parts of the Analyzer Module can be irreparably damaged if exposed to electrostatic discharge (ESD).
The instrument is ESD protected when the covers have been secured and safety precautions observed. When the housing is open, the internal components are not ESD protected anymore.
Although the electronic parts are reasonable safe to handle, you should be aware of the following consid­erations:
Best ESD example is when you walked across a carpet and then touched an electrical grounded metal doorknob. The tiny spark which has jumped is the result of electrostatic discharge (ESD).
You prevent ESD by doing the following:
Remove the charge from your body before opening the housing and maintain during work with opened housing, that no electrostatic charge can be built up.
Ideally you are opening the housing and working at an ESD - protecting workstation. Here you can wear a wrist trap.
However, if you do not have such a workstation, be sure to do the following procedure exactly:
Discharge the electric charge from your body. Do this by touching a device that is grounded electrically (any device that has a three - prong plug is grounded electrically when it is plugged into a power recep­tacle).
This should be done several times during the operation with opened housing (especially after leaving the service site because the movement on a low conducting floors or in the air might cause additional ESDs).
P - 8 Preface Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
GENERAL PRECAUTIONS FOR HANDLING AND STORING
HIGH PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association´s „Handbook of Compressed Gases“ published in 1981.
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 52 °C (125 °F). A flame should never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
P - 9PrefaceEmerson Process Management GmbH & Co.OHG
Instruction Manual
9
6
HAS60E-IM-HW
NGA 2000 CLD
June 2007

DOCUMENTATION

The following CLD instruction materials are available. Contact Customer Service Center or the local represen­tative to order.
HAS60E-IM-HW Instruction Manual NGA 2000 CLD (this document)
HAS60E-IM-SW39 Software Manual NGA 2000 CLD
90002496 Instruction Manual NGA 2000 Platform

COMPLIANCES

This product may carry approvals from several certifying agencies, including the Canadian Standards Association (which is also an OSHA accredited, Nationally Recognized Testing Laboratory), for use in non-hazardous, indoor locations.
Emerson Process Management has satisfied all obligations from the European Legislation to harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC. Recommendation (May 1993).
NAMUR
This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New Zealand.
N

SUITABILITY TESTS

P - 10 Preface Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

GLOSSARY OF TERMS

Analyzer Module
The module that contains all sensor/detector components for development of a Primary Variable signal; includes all signal conditioning and temperature control circuitry.
Backplane
The interconnect circuit board which the Controller Board, Power Supply, Analyzer Module power and network cables, I/O Modules and Expansion Modules plug into.
Control Module
The Operator Interface plus the Controller Board.
Controller Board
The computer board that serves as the Network Manager and operate the Display and Keypad.
Distribution Assembly
The Backplane and the card cages that hold I/O and Expansion Modules.
I/O Module
A circuit board that plugs into the Backplane from the rear of the Platform. Has a connector terminal for communication with external data acquisition devices and provides an input/output function.
Power Supply
Any of a variety of components that provides conditioned power to other NGA 2000 components, from the Power Supply Board that plugs into the front of the Backplane in a stand-alone instrument to several larger ones that can power larger collections of modules and components.
P - 11PrefaceEmerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW
NGA 2000 CLD
June 2007
Primary Variable
The measured species concentration value from an Analyzer Module.
Secondary Variable
Data placed on the network by a module regarding current status, e.g., sample flow, temperature and pres­sure.
Softkeys
The five function softkeys located below the front panel display; they assume the function displayed directly above each on the display, a function dictated by software.
System
Any collection of Analyzer Module(s), Platform(s) and I/O Module(s).
P - 12 Preface Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

ANALYZER SYSTEM ARCHITECTURE

The NGA 2000 CLD is available as a "stand-alone ana­lyzer" or as a "blind" Analyzer Module (AM). The CLD analyzer module can be part of the stand-alone analyzer or a component of an analyzers system (Fig. A-1).
The NGA 2000 system made it possible, to configure the CLD as a flexible "stand-alone analyzer" consisting of a CLD "Analyzer Module", a Platform (complete with front panel display/operator interface), and input/output (I/O) modules.
The “analyzer module” is a “blind” analysis unit but re­tains all the advanced design features. The AM variant is designed for integration as part of a NGA 2000 analy­sis system or special customer developed networks.
The platform/MLT´s front panel can act as operator in­terface for a stand-alone analyzer or as the a central interface for multiple Analyzer Modules. In multi anal­yzer systems, this feature eliminates duplication of the display/operator interface. In addition to the obvious op­erational benefits there are significant cost and system packaging advantages not possible with conventional analyser configurations.
This flexible network communication architecture is shown in the schematic of Fig. A-2.
½ 19" MLT
Analyzer
ROSEMOUNT NGA 2000NGA 2000
Platform with MLT AM
or
19" MLT Analyzer
CLD
FID
PMD
NGA Network
½ 19" MLT
Analyzer
ROSEMOUNT NGA 2000NGA 2000
MLT MLT
ROSEMOUNT NGA 2000NGA 2000
CLD FID
NGA Network
ROSEMOUNT NGA 2000NGA 2000
ROSEMOUNT NGA 2000NGA 2000
ROSEMOUNT NGA 2000NGA 2000
ROSEMOUNT NGA 2000NGA 2000
NGA NetworkNGA Network
Platform with MLT AM
19" MLT Analyzer
CLD
FID PS
or
CLD PS

Figure A-1: From separate analyzers to analyzer system

NGA Network
NGA Network
PS
A - 1Analyzer System ArchitectureEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
The modular configurable bi-directional network offers the following options:
Stand-alone analyzers (Single devices)
analyzer modules in a platform including
optional inputs and outputs (SIO/DIO).
Platform
Simple interconnection of analyzer modules to an
analyzer system based on one of the three struc­tures - see below.
These structures can be distinguished by acting of the host
with platform as host including system inputs
and outputs (SIO/DIO)
with MLT/TFID/CAT 200 analyzer as controller
including system inputs and outputs (SIO/DIO)
with customer owned specific control units
(not described in this manual, consult factory)
For combination possibilities of NGA 2000 I/O´s see table A-1.
Analyzer Module
AC DC
Network Cable
Analyzer Module
Analyzer Module
Network Cable
24VDC Cable

Figure A-2: Example of NGA cabling

Power supply
24VDC Cable
DC
AC
A - 2 Analyzer System Architecture Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
Based on a platform, MLT or TFID analyzer the sche­matic on Figure A-3 illustrates the simplicity of a net­worked system which incorporates AM’s, such as Chemi­luminesence Detectors, MLT's (NDIR/UV/VIS plus Oxy­gen or TCD) and Flame Ionisation Detectors.
The system I/O modules (SIO, DIO) of the platform (or MLT/TFID analyzer) support all integrated analyzer mod­ules with analog, digital and serial interfaces as well as relay outputs.
Analyzer Modules (AMs)
Other system functionality includes links to associated sample handling (PLC) and Data Acquisition Systems such as WinControl.
Local I/O are existing to MLT, TFID and CAT 200 analyz­ers only and support the corresponding analyzer mod­ule only.
Local I/Os via Internal System Bus
DIO SIO
DIO SIO
DIO SIO
O
2
PMD
DIO SIO
8 Digital Inputs
HC
HFID
System I/Os via Internal System Bus
24
Digital
Outputs
Solenoid
Valves
NO
WCLD
Printer or PLC
x
3
Relay
Outputs
NO/NO
CLD
RS 232 /
x
19" MLT 3/4
Analyzer
RS 485
Personal
Computer
ppm O
2
TO2
Alternatively
19" Platform/ TFID Analyzer
Alternatively
HC
FID
½ 19" MLT
Analyzer
HC
TFID
NGA Network
NGA 2000ROSEMOUNT
NGA 2000ROSEMOUNT
NGA 2000ROSEMOUNT
CO/NO/SO /
EO - MLT
2
Alternatively to Platform
or MLT/TFID Analyzer
MLT 2
Analyzer
PO
2
2
MLT
Field PC
Workstation
NGA 2000ROSEMOUNT
Figure A-3: Example/Possibilities of NGA Analyzer Systems
A - 3Analyzer System ArchitectureEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
System unit SIO/DIO-Configuration
Instruction Manual
HAS60E-IM-HW
June 2007
CLD/FID/HFID analyzer module (AM):
without front panel,
i.e. without control unit (host)
can be combined with
a platform, a MLT analyzer, a TFID analyzer; a CAT 200 analyzer or a customer developed control unit
Platform (Control Module Software):
Control unit with front panel
Without measurement channels
CLD analyzer
CLD analyzer module into a platform with
front panel
CLD analyzer module combined with
MLT/TFID/CAT 200 analyzer
No local CLD/FID/HFID I/O’s
1 SIO and up to 4 DIO's (or 5 DIO's)
can be installed in the platform (CM I/O’s)
SIO and DIO’s can be configured
for all AM channels connected to the platform
1 SIO and 4 DIO’s (or 5 DIO’s) can be
installed in the platform
1 SIO and 1 DIO (or 2 DIO’s) can be
installed in the MLT/TFID/CAT 200 analyzer (CM I/O)
SIO and DIO can be configured
for all AM’s connected to the MLT/TFID/CAT 200 analyzer

Table A-1: Possibilities of NGA 2000 I/O combinations

A - 4 Analyzer System Architecture Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
SECTION 1
DESCRIPTION AND SPECIFICATIONS

1-1 OVERVIEW

This manual describes the Chemiluminescence (CLD) Analyzer Module of Emerson Process Mangement´s NGA 2000 Series of gas analysis components (See Fig­ure 1-2, Figure 1-3 and Figure 1-4).
The CLD Analyzer Module is designed to continously determine the concentration of Nitric Oxide and oxides of Nitrogen (NO plus Nitrogen Dioxide [NO
]) in a flow-
2
ing gaseous mixture. The concentration is expressed in parts-per-million.
The CLD Analyzer Module is designed as a slide-in module (if configured in stand-alone instrument fashion), removable from the front of the Platform, with gas con­nections made from the rear. All electronics relative to sample detection and conditioning are included in this module.

1-2 TYPICAL APPLICATIONS

The CLD Analyzer Module has specific applications in the following areas:
Oxides of Nitrogen emissions from the combus­tion of fossil fuels in
Vehicle engine exhaust
Incinerators
Boilers
Gas appliances
Turbine exhaust
Nitric acid plant emissions
De-NO
control system
x
Nitric oxide emissions from decaying organic material (e.g., landfills).
The reactions involved are:
NO + O
NO
As NO and O
NO
3
* NO
2
mix in the reaction chamber, the intensity
3
* + O
2
+ red light
2
2
of the emitted red light is measured by a photodiode and is proportional to the concentration of NO in the original gas sample.
To measure NO
(NO + NO2), any NO2 in the sample is
x
reduced to NO (at < 95 % efficiency) by being continously passed through a heated bed of vitreous carbon (this occurs before the sample gas is presented to the ozone). Any NO initially present in the sample passes through this converter stage unchanged before being routed to the reaction chamber.
The photodiode generates a DC current, which is then amplified, conditioned and expressed on the network as the Primary Variable.
Exhaust
NO Molecule
NO
Ozone
O3 Molecule
O2 Molecule
NO2 Molecule
(excitated)
NO2 Molecule
Figure 1-1: Function Principle of
CLD Measurement

1-3 THEORY OF TECHNOLOGY

The CLD Analyzer Module uses the chemiluminescence method of detection. This technology is based on the reaction of NO with ozone (O
) to produce NO2 and oxy-
3
gen (O2). Some of the NO2 molecules thus produced are in an electronically excited state (NO2* - the "*“ refers to the excitation). These revert immediately to the ground state, with emission of photons (essentially, red light).

1-4 FEATURES

Among the features included in the CLD Analyzer Mod­ule are:
1) ozonator air loss shutoff and
2) NO/NO
mode capability.
x
1 - 1Description and SpecificationsEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
Bulkhead Reducer
1/4T-1/8T SS
Ozone Air
Sample
Bulkhead Reducer
1/4T - 1/8T SS
Exhaust
Bulkhead
1/4T SS
Connector
Cross
1/8T BRS
Male Connector
1/8T - 3/8NPT BRS
Flow Balance
Valve
903207
Tee Assembly
1/8T - 1/4MPT
Tee
1/4T-1/8FPT SS
1/4T SS
1/8T - 1/8NPT BRS
Male Connector 1/8T - 3/8NPT BRS
Pressure Sensor 30 psig Air 655254
1/8T - 1/8MPT KYNAR
Elbow
Back
Pressure
Regulator
(see Table 1-1)
Reducer 1/4T - 1/8T SS
Elbow 1/8T - 1/4MPT SS
Connector
0.116 ID Viton BLK
Pressure Switch 655215
Bypass
Flow Sensor
902931
Elbow 1/8T - 1/4MPT SS
Air Restrictor 430 cc/min. @ 12 psig 658157
Connector Glass
BLUE
Elbow 1/8T - 1/8MPT KYNAR
Tee 1/8T KYNAR
NOx to NO
Converter
655250
Top
Tee 1/8T
KYNAR
Cross
1/8T SS
1/8T - 3/8NPT SS
Pump Capillary
1/8 OD TEFLON
(see Table 1-1)
Male Connector
Ozone Generator
657719
Connector Glass BLUE
Vent Capillary
70 cc/min. 5 psig
634398
Te e
1/8T
KYNAR
Sample Pressure Sensor
15 psig 655253
Elbow 1/8T SS
1/8 OD SS Elbow 1/8T SS
1/8 OD SS
Reduction Union 1/8T - 1/16T SS
Reaction Chamber
Ozone
Exhaust
Sample
Sample Capillary
(see Table 1-1)

Figure 1-2: Flow Diagram - CLD Analyzer Module with Bypass Flow Sensor

Brief Description
Part Number
Sample Capillaries
Sample Capillary 200 cc/min. @ 5 psig 659658
Sample Capillary 200 cc/min. @ 2 psig 660404
Sample Capillary 70 cc/min. @ 5 psig 659657
Sample Capillary 70 cc/min. @ 2 psig 660403
Pump Capillaries
Pump Capillary 200 cc/min. @ 5 psig 657473
Pump Capillary 200 cc/min. @ 2 psig 660405
Back Pressure Regulators
Back Pressure Regulator Brass 5 psig 655269
Back Pressure Regulator Stainless Steel 5 psig 659063
Back Pressure Regulator Brass 2 psig 660400
Back Pressure Regulator Stainless Steel 2 psig 660401

Table 1-1: Components depending on Module Configuration

1 - 2 Description and Specifications Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
Bulkhead Reducer
1/4T-1/8T SS
Ozone Air
Sample
Bulkhead Reducer
1/4T - 1/8T SS
Cross
1/8T BRS
Male Connector
1/8T - 3/8NPT BRS
Flow Balance
Valve
903207
Tee Assembly
1/8T - 1/4MPT
Connector
1/8T - 1/8NPT BRS
Male Connector 1/8T - 3/8NPT BRS
Pressure Switch 655215
Pressure Sensor 30 psig Air 655254
Back Pressure Regulator
(see Table 1-1)
Exhaust
Bulkhead
1/4T SS
0.116 ID Viton BLK
Connector Glass
Elbow 1/8T - 1/4MPT SS
Air Restrictor 430 cc/min. @ 12 psig 658157
NOx to NO
Converter
655250
Top
BLUE
Tee 1/8T
Cross
1/8T SS
Tee 1/8T KYNAR
1/8T - 3/8NPT SS
KYNAR
Male Connector
Ozone Generator
657719
Connector Glass BLUE
Vent Capillary
70 cc/min. 5 psig
634398
Pump Capillary
1/8 OD TEFLON
(see Table 1-1)
Sample Pressure Sensor
15 psig
655253
Tee 1/8T
KYNAR
Te e 1/8T KYNAR
Union 1/8T SS
1/8 OD SS
Union 1/8T SS
1/8 OD SS
Reduction Union
1/8T - 1/16T SS
Reaction Chamber
Ozone
Exhaust
Sample
Sample Capillary
(see Table 1-1)

Figure 1-3: Flow Diagram - CLD Analyzer Module without Bypass Flow Sensor

1 - 3Description and SpecificationsEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Driver Board
Instruction Manual
HAS60E-IM-HW
June 2007
Inlet/Outlet
Fittings
Fan
NO2 to NO Converter
3/2-way Solenoid Valve
Ozone Generator
Signal Board
Sample Backpressure
Regulator (under Cover)
Detector
Flow Sensor
Power Supply Board
Flow Balance Needle Valve

Figure 1-4: CLD Analyzer Module - Top View

1 - 4 Description and Specifications Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007

1-5 SPECIFICATIONS

a. General

Measurement Species: ........................ NO, NOx
Ranges: .............................................. 0 to 5 ppm through 0 to 10,000 ppm NO, NOx
Repeatability: ...................................... ± 0.5 % of fullscale (at constant temperature)
Min. Detectable Level: ......................... 0.1 ppm
Noise: .................................................. < 1 % of fullscale, peak to peak
Linearity: ............................................. ± 1 % of fullscale
Response Time: .................................. < 1 sec. for 90 % of fullscale for ranges of 25 ppm or greater
t90 Response ...................................... 1-2 sec.
Zero Drift: (at constant temperature) .... <± 1 % of fullscale/24 hours, <± 2 % of fullscale/week
Span Drift: (at constant temperature) ... <± 1 % of fullscale/24 hours, <± 3 % of fullscale/week
Effect of Temperature: ......................... < 2 % of fullscale (over any 10 °C ambient temperature variation interval for
Ambient Temperature: ......................... 0 °C to 45 °C (32 °F to 113 °F)
Rated Power: ...................................... 24V DC 150W
NGA 2000 CLD
< 3 sec. for 90 % of fullscale for ranges of less than 25 ppm
a rate of change no greater than 10 °C/hour)

b. Physical

Case Classification: ............................. General purpose for installation in weather-protected area
Dimensions: ........................................ See Figure 2-5: Outline and Mounting Dimensions
Weight:................................................ 8.1 kg (18 lbs.)
Mounting: ............................................ Inside a Platform or custom-installed in a panel
Max. Length of LON Cable: ................. 1,600m (1 mile) between Analyzer Module and Platform

c. Sample

Temperature: ....................................... 0 °C to 45 °C (32 °F to 131 °F)
Total Flow Rate: ................................... (Externally measured) 900 to 2,200 cc/min. with backpressure regulator
Particles: ............................................. Filtered to < 2 microns
Dewpoint: ............................................ 5.5 °C below ambient temperature, no entrained liquid
Materials in contact with Sample: ........ Stainless steel, Teflon, glass, brass and neoprene
Optional: ............................ Stainless steel, Teflon, glass and Kynar
Ozonator Gas: ..................................... Clean, dry air or oxygen; flow rate: 1 l/min. maximum;

d. Gas Connections

Ozone Air: ........................................... 1/4" O.D. tube fitting, stainless steel
Exhaust: .............................................. 1/4" O.D. tube fitting, stainless steel
Sample In: ........................................... 1/4" O.D. tube fitting, stainless steel
pressure adjusted to 1,344 hPa (5 psig) or to 1,138 hPa (2 psig)
pressure: 689 hPa to 1,034 hPa-gauge (10 to 15 psig); maintain a constant pressure ± 34 hPa (± 0.5 psig)
See the Preface section of the Platform manual for specifications regarding Platform-related components (e.g., case dimensions) and the I/O Module manual for specifications regarding I/O (e.g., relay outputs).
1 - 5Description and SpecificationsEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
1 - 6 Description and Specifications Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
SECTION 2
INSTALLATION
WARNING
Before starting to install this equipment, read the "Essential instructions" on the inside cover and the Safety Summary beginning on page P-2. Failure to follow the safety instructions could re­sult in serious injury or death.

2-1 UNPACKING

If the Chemiluminescence (CLD) Analyzer Module is received as a separate unit, carefully examine the ship­ping carton and contents for signs of damage. Immedi­ately notify the shipping carrier if the carton or contents is damaged. Retain the carton and packing material un­til all components associated with the CLD Analyzer Module are operational.
NGA 2000 CLD

2-2 ASSEMBLY

If the CLD Analyzer Module requires assembly with other components (e.g., the Platform and associated I/O Mod­ules), do so at this time.
To install the CLD Analyzer Module into a Platform:
1. Loosen the six fastening screws for the front panel of the Platform, hold the handles, and swing the front panel to the farest right.
2. Following the guides on the bottom left and bot­tom center of the Platform, carefully slide the CLD Analyzer Module halfway into place
3. Lift the spring-loaded pins on the front of the CLD Analyzer Module, and carefully slide in the rest of the distance.
If the module and Platform are difficult to assemble, remove the module, ensure the top cover of the module is firmly seated on the hold-down screws, and repeat the assembly procedure.
4. Secure the module in position by releasing the pins, which seat in the available holes in the bottom of the case (see Figure 2-1, below).
5. Connect network cable and power cable to the Analyzer Module (refer to Section 2-5 for electrical connections).
6. After startup and calibration have been performed, secure the front panel of the Platform with the six screws provided.
Pin Seats
Analyzer Module Guides

Figure 2-1: Analyzer Module Installation into Instrument Platform (view without front panel)

2 - 1InstallationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007

2-3 LOCATION

Install the CLD Analyzer Module in a clean, weather-pro­tected, vibration-free location free from extreme tempera­ture variations and moisture. For best results, install the instrument near the sample stream to minimize sample transport time.
The analyzer should be installed within ±15° of horizon­tal.
NOTE
Unrestricted air flow to the rear of the Analyzer Mod­ule is critical to its performance and reliablilty.
Operating ambient temperature is 0 °C to 45 °C (32 °F to 113 °F). Temperature change should not exceed 10 °C (18 °F) per hour. The same temperature range restric­tions apply to the location of the air and span gas cylin­ders.

2-4 GASES

a. Gas Conditioning (Sample Handling)

NOTE
All gases must be conditioned before supplying!
The gases must be
- dry
- free of dust (filtered for particulates down to two microns)
- free of aggressive components affecting gas paths materials (e.g. by corrosion)
- free of Ammonia etc. in order to prevent crystallin formation
before entering the Analyzer Module.
The gases should have a dew point 5 °C (13 °F) below coldest ambient temperature.
CAUTION
EXTERNALLY RESTRICT SAMPLE/CALIBRATION
GAS FLOW TO LESS THAN 2,200 cc/min.
MAXIMUM NO
LEVEL IN PARTS
PER MILLION
800 Air
2,500 Air
10,000 Oxygen
GAS SUPPLIED TO
x
REAR PANEL AIR
INLET

Table 2-1: Gas Specifications

Damage to internal components may occur if this flow level is exceeded.
No restrictor is provided in the sample inlet of this module. For those users who cannot externally limit sample flow, contact your local service or sales office.
SETTIN G ON OZONE
PRESSURE GAUGE
(pressure values: gauge)
690 hPa (10 psig); provides flow of approximately 500 cc/min. to ozone generator
690 hPa (10 psig); provides flow of approximately 1,000 cc/min. to ozone generator.
1035 hPa (15 psig); provides flow
of approximately 1,000 cc/min to ozone generator.
SETTING ON SAMPLE
PRESSURE GAUGE
(pressure values: gauge)
344 hPa (5 psig); provides flow of approximately 200 cc/min. to reaction chamber
344 hPa (5 psig); provides flow of approximately 70 cc/min. to reaction chamber.
103 hPa (1.5 psig); procides
flow of approximately 20 cc/min. to reaction chamber.
2 - 2 Installation Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
Exhaust Ozonator Air
Exhaust Fan Sample IN Intake Fan

Figure 2-2: CLD Rear Panel Connections

b. Connections

Connect inlet and outlet lines for sample, ozonator air, and exhaust to appropriately labeled fittings on the rear panel (see Figure 2-2) , each of which is a 1/4-inch fer­rule-type compression fitting.
Stainless steel or Teflon tubing is recommended for the sample line. Exhaust tubing should be 1/4 inch (6.3 mm) or larger, and made of stainless steel or Teflon.
Zero/span gases should be introduced at the SAMPLE IN fitting at normal sample inlet flow rate/pressure.

c. Specifications

Each gas should be supplied from a cylinder equipped with a clean, non-corrosive, two-stage regulator. A shut­off valve is recommended.
Zero Gas
Nitrogen (N
) is recommended for use as zero gas.
2
Alternatively synthetic air or NOx free ambient air can be used.
Calibration Gas
A mixture of NO in a background of nitrogen is recom­mended as span gas. For maximum accurancy, the con­centration of NO in the span gas should be about 80 % to 100 % of full scale range.
Ozonator Source Gas
For analyzers with ranges less than 0 to 2,500 ppm NO free ambient air should be used for generation of the ozone required for the chemiluminescence reaction. For ranges higher than 0 to 2,500 ppm, NOx free oxygen is required. See Table 2-1 for correct pressure settings.
When using ambient air an external air conditioning unit (LAM) can optionally be used (contact your local service or sales office).
CAUTION
At no time should ozonator gas pressure exceed 2,070 hPa-gauge (30 psig). Damage to internal com­ponents may occur if this pressure level is ex­ceeded.
Sample Gas Pressure
See Table 2-1 for correct pressure settings.
Bypass Sample Gas Flow
Bypass sample gas flow rate should be between 700 and 2,000 cc/min. with backpressure regulator pressure (see Capillary Pressure in „Current Measurement Pa­rameters“ menu, which can be viewed by selecting the PARMS softkey in "Main Menu") adjusted to 344 hPa (5 psig).
x
2 - 3InstallationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
Leak Test
The CLD Analyzer Module is completely tested at the factory for gas leakage (leakage rate - 7.5 hPa/min. with He or - 2.5 hPa/min. with N
). The user is responsible for
2
leakage testing at the inlet and outlet fittings on the rear panel minimum twice a year (see Section 4-8).
Contaminants
White crystal deposits on the windows of the reaction chamber, and the clogging of capillaries and vent are usually due to sample contaminates such as ammonia reacting with the high ozone levels and NO components.
To eliminate the contaminates, the sampling system should be reworked or a preventive maintenance pro­gram developed (if dropout is not excessive). Another source of crystalline formation is contaminated air.
NOTE
Contamination is a result of a not properly work­ing sample handling systrem and is not be cov­ered by Emerson Process Management warranty.

2-5 ELECTRICAL CONNECTIONS

NOTE
Electrical installation must be in compliance with the requirements of NAMUR and DIN VDE and/or any applicable national or local codes (like local electricity supply enterprises (ESE)).
Refer to figure 2-3. Two electrical connections are re­quired on the Analyzer Module: POWER and NET­WORK. On the Analyzer Module, two NETWORK con­nectors are available, either of which is appropriate for:
1. Interconnection with the Backplane of the Plat­form. (See Instruction Manual for the NGA 2000 Platform).
2. "Daisy-chaining" with other NGA 2000 compo­nents.
Connect Analyzer Module POWER to a 24 VDC, 10 A power source, either the Platform or external power source.
Regulator Adjust Slot
Metering Valve
SAMPLE
REGULATOR
FLOW
BALANCE
24 V
Polarity
1 + 2 ­3 GND
T 6A
~
250 V

Figure 2-3: Front Panel Controls and Electrical Connections

LON
LON
Network
1
2
Connection “1”
Network Connection “2”
Power Connection
Fuse
2 - 4 Installation Emerson Process Management GmbH & Co.OHG
Instruction Manual
CABLE
DIGITAL
I/O
655246
3
HAS60E-IM-HW June 2007
NGA 2000 CLD
Optional with Bypass Flow
1
Sensor Configuration
POWER SUPPLY BOARD-LON/POW ER
MODULE ASSEMBLY 657510
J2
J1
LON/PWR MODULE 656761
J5
3 2 1
RED
CABLE, CONTR OLLER PWR 90303 4-V1
NC
J6
2
1
1
POWER SUPPLY
2
BOARD 657520
J13
8 7
1
2
J1
10
9
3
12
J11
1
10
J3
6
1
7
12
RED
HEATERS
DETECTOR
CONVERTER HEATER JACKET 655228
J3
E3
E2
E1
ORN
GRN
BLK
GRN
BLK
1
GRN
J2
2
ORN
3
WHT
THERM OSTAT
2
WHT
CONVERTER TEMPERATURE SENSOR 655282
CONVERTER
655250
NC
1
J1
J4
1
J2
COMPUTER ANALYSIS BOARD 658350
1
J3
CABLE, INTERNAL NETWORK 903035-V1
HARNESS, SIGNAL POWER 655252
CABLE, FLOW SENSOR 655384
FLOW
RED
SENSOR
902931
BLK
WHT
1
HARNESS, DRIVER POWER 655243
HEATER/THERMOSTAT ASSEMBLY 655235 PHOTODIODE
THERMISTOR 655216
SOCKET
NO/NOx
WHT
WHT
SHLD
OZONATOR
POWER SUPPLY
657716
SHIELD
GRN
SOLENOID 655263
JUMPER 657162
ASSEMBLY
655218-V1
THERMO STAT 657298
HARNESS, EXPANSION I/O 655249
1 1 1
J5 J6
1
J7
2
OZONATOR
657719
TRANSISTOR 2SD1308 655264
SAMPLE PRESSURE SENSOR 655253
E C
B
B = YEL C = RED E = BLU
PRESSURE SWITCH (NO) 4-15 PSIG 655215
YEL (COM)
BRN (NO)
RED
SAMPLE PRESSURE SENSOR 655254
RED
WHT
J5
J3
J6
BLK
WHT
SIGNAL BOARD 655580
5 4 3
J8
2 1
J9
J1
J13
1 2 3 4 5 6 7 8 9
ORN
WIRE*
BRN
*COAX CABLE
1 2 3 6
1 4 5 6
GRN
WHT
RED
BLK
RED
SHLD*
GRN
YEL
WHT
BLU
BLK
WHT
RED
J11
8
J2
GRN
WHT
BLK
CABLE, ANALOG I/O 903033
BLK
BLK
7 6
1
,
WHT
RED
RED
4 3 1
J4
J3
J8
J10
DRIVER BOARD
655620
3
J11
1
J2 J6
10 1 2
2
4
BLK
3
J2
2
RED
1
4
BLK
3
J1
2 1
RED
WHT
2
1
WHT
BRN
2
1
BRN
4
ORN
3
J9
2
ORN
1
1 2 3 4 5 6 7 8
BLU
5 4
RED
3
J5
2
YEL
1
4 3
J12
BRN
2 1
YEL
12
10
CABLE, DIAGNOSTICS 903032-V1
24
2
Brief Description Part Number
Detector 200 cc/min. @ 5 psig 659754
Detector 200 cc/min. @ 2 psig 42716203
Detector 70 cc/min. @ 5 psig 42716204
Detector 70 cc/min. @ 2 psig 42716205
FAN 655245
FAN 655245

Figure 2-4: CLD Wiring Diagram

2 - 5InstallationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
[
]
[23].9[23].9[33]
4.3
[109]
2.5 [64]
1.5 [38]
4.1
[104]
8.4
[213]
6.1
[155]
8.2
[208]
[18]
Instruction Manual
HAS60E-IM-HW
June 2007
.4
[10]
.7
.8
[20]
2.8
[71]
.5
[13]
1.1 [28]
4.1
4.1
104
6.0
[152]
17.5
[445]
1.3 [33]
1.3
Dimensions:
INCHES
[MM]

Figure 2-5: Outline and Mounting Dimensions

2 - 6 Installation Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
SECTION 3
OPERATION

3-1 OVERVIEW

Once the CLD has been correctly assembled and in­stalled, the analyzer is ready for operation.
Before operating the system, verify that the Leak Checks have been performed in accordance with Section 2-4.
In this section, all operations for starting up the analyzer are explained. For more detailed information about soft­ware screens see associated Software Manual.
For the remainder of this section, Analyzer Module inter­connection with a Platform or some interfacing compo­nent is assumed. Display and Keypad information refers to that which the user can expect to see and do with regard to the Front Panel of the Platform.
Depending from the software version that is installed, menu layout can change, whereas the principle of op­eration always stays the same.
This instruction manual is based on menus of software version 3.7.1.

3-2 STARTUP & INITIALIZATION

the startup sequence will interrogate the network to lo­cate and identify all components on the network. The user will have to bind appropriate combinations of com­ponents after the startup sequence. See the Platform manual for instructions on binding combinations of mod­ules.
Pressing the F1 key during initializing will reset the CLD brightness and contrast to factory settings. Pressing the F3 softkey will abort the network initializing, aborting any connection to other analyzers. In that case, only the menus of the local analyzer will be available.
At the end of the initializing routine the "measure" screen will display. This screen is the access to all other chan­nels, menus and submenus. The actual display may dif­fer from that shown depending on any custom configu­ration.
After the warm-up period (about one hour for the CLD Analyzer Module), all modules are completely functional. Establish that correct ozonator air pressure and sample flow rate are within specifications (see Section 1-5). Cali­brate and adjust converter efficiency, and begin opera­tion as the following sections indicate.
Apply LON connection and power to the CLD Analyzer Module. If it is associated with a Platform, do this by plugging in the Platform to a power source. The Plat­form has no ON/OFF power button. Once power has been supplied to the Platform, the CLD Analyzer Module will be energized.
After switching on the CLD, the analyzer will begin its booting procedure which is apparent on the CLD screen. The first part of the initialization procedure is a self check of the software and analyzer components. Various dis­plays will show the status of the initialization including revision notes, "Initializing network interface", "Search­ing for nodes", "Scanning Module 2: CLD, 12 % Com­plete", and "Calculating bindings".
If the user´s system contains only one Analyzer Module, all system components, the Controller Board and the network "self-install" (bind together) during initial startup. If the system contains more than one Analyzer Module,
3 - 1OperationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007

3-3 PREPARATIONS

After performing start-up and initialization that is described in Section 3-2, operating variables must be adjusted, i.e. the Analyzer Module is to be calibrated.
In the following, analyzer function control and calibration procedure is described step-by-step.
CLD
7.50
0.00 10.00Range 1
Sample Flow:
Sample press:
Ozonator: OFF-PRESS. SW.
Converter temp:
Display Status... Main... Channel BasicCal
1300 ml/min
4.0 hPa
30.9 C
200
50.0
150.0
ppm NO
1500
490.0
500.0

Figure 3-1: Measure Mode Display

After starting up the Analyzer Module, the Measure Mode Display is displayed as shown in figure 3-1.
To now check back physical parameters of the CLD Ana­lyzer Module with the values defined in your Test Data Sheet, that you received with the analyzer, you have to select the "Physical Measurements"-Menu.
Enter the diagnostics menu "Physical Measurements" as follows, using the softkeys F1 through F5:
The menu "Physical Measurements" lets you monitor physical measurement parameters of the CLD analyzer.
During warm-up time of the analyzer, you can observe temperature values (sensor temperature, block tempera­ture...) of internal components.
After warm up, check
Block Temperature
Detector Temperature
Converter Temperature
with the operating values given by yout Test Data Sheet. You will find an excerpt of a Test Data Sheet as an ex­ample on the next page in figure 3-3.
NOTE:
As long as operating temperatures are not yet reached by the internal components of the ana­lyzer, it is not ready for operation. Warm-up can last up to one hour.
After warm-up or during warm-up procedure of the ana­lyzer you can connect all gas supply lines to the back of the analyzer.
Supply gases at the pressures given in the Test Data Sheet and restrict flow by an external flow limiter to 1.3 through 1.5 l/min.
Internal sample gas and ozonator pressures:
Main (Menu)...
Expert Controls and Setup...
Analyzer Module Setup...
Physical Measurements...
Line#1
Sample Capillary Pressure: Ozone Supply Pressure: Ozonator Status: Ozonator Power: Bypass Flow: Converter temperature: Ozonator temperature: Sensor temperature: Block temperature: Capillary Flow Rate:
Pressure limits...
Pressure limits...
HOME ESCAPE MORE HISTORY INFO
Physical Measurements
0.000 ppm
0.7 hPa
4.0 hPa
OFF - PRES. SW.
Enabled
2 ml/min
199.9 ml/min
320 C
51.5 C
1.1 C
51.5 C
You will find internal physical paramters in the Test Data Sheet that you have received with the ana­lyzer. Externally supply gases at the given pres­sures with an external adjustment.
Option Bypass Flow:
Depending on measuring capillary, bypass flow should by 900 - 2,000 cc/min. If no bypass is in­stalled, restrict flow externally to 1.3 - 1.5 l/min.

Figure 3-2: Physical Measurements Display

3 - 2 Operation Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

Figure 3-3: Excerpt of a Test Data Sheet with values that are to be compaired with physical measurements.

3 - 3OperationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007

3-4 CALIBRATION PROCEDURE

The CLD analyzer module may require periodic calibra­tion with known zero and span gases in order to main­tain a desire level of analytical accurancy. It is recom­mended, after initial startup, that the CLD Analyzer Mod­ule is calibrated at least once every eight hours.
This practice should continue until evidence indicates that some other interval is more appropriate depending on the analytical accurancy required.
Calibration is the process of flowing known zero or span calibration gas into the analyzer for a specified period (averaging time), after which the analyzer will automati­cally set its zero or span factors so that the concentra­tion measurement equals the calibration gas value. A limit can be set, beyond which any attempt by the ana­lyzer to reset its concentration measurement will cause a warning alarm. In this case, user intervention would be required to reset the alarm and attempt another calibra­tion.

a. Calibration Setup

Calibration Gas List
This menu is used to set the concentration values of the calibration gases for each range.
Main (Menu)...
Analyzer and I/O expert controls & setup...
Analyzer module setup...
Calibration gas list...
Line#1
Calibration Gas List
Zero gas - range 1:
NO Span gas - range 1: NOx Span gas - range 1:
Zero gas - range 2: NO Span gas - range 2: NOx Span gas - range 2:
0.000 ppm
0.00 ppm
10.00 ppm
10.00 ppm
0.00 ppm
19.90 ppm
19.90 ppm
MORE...
Line#1
Zero gas - range 3:
NO Span gas - range 3: NOx Span gas - range 3:
Zero gas - range 4: NO Span gas - range 4: NOx Span gas - range 4:
Calibration...
HOME ESCAPE INFO
Calibration Gas List
MORE
0.000 ppm
0.00 ppm
101.0 ppm
101.0 ppm
0.00 ppm
250.0 ppm
250.0 ppm
Figure 3-5: Calibration Gas List
Channels 3 and 4
If not yet done, put in the ozonator and zero gas concen­trations that you supply to the analyzer. See gas cylinder certification for exact values.
In case that measuring ranges differ from ordering code, put in the measuring ranges:
Main (Menu)...
Analyzer and I/O, expert controls & setup...
Analyzer module setup...
Gas measurement parameters...
Range settings...
Line#1
Range Settings
Minimum Range: Maximum Range:
Range 1 lower limit:
Range 1 upper limit: Range 2 lower limit: Range 2 upper limit: Range 3 lower limit: Range 3 upper limit: Range 4 lower limit: Range 4 upper limit:
Pressure limits...
HOME ESCAPE INFO
0.000 ppm
9.9 ppm
999.0 ppm
0.0 ppm
10.2 ppm
0.0 ppm
24.9 ppm
0.0 ppm
100.0 ppm
0.0 ppm
500.0 ppm

Figure 3-6: Range Settings Menu

Calibration...
HOME ESCAPE INFO
MORE
In case that only one measuring range is in use,
NOTE
we suggest to set all other measuring ranges to the same value to prevent calibration failure. When
Figure 3-4: Calibration Gas List
Channels 1 and 2
3 - 4 Operation Emerson Process Management GmbH & Co.OHG
doing so, set also all test gas values to the same value.
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
Calibration Parameters
This menu provides various parameter settings for all calibration performed from Basic or Expert modes.
Main (Menu)...
Analyzer and I/O expert controls & setup...
Analyzer module setup...
Calibration parameters...
Line#1
Calibration adjustment limits:
Calibration averaging time: Calibration failure alarm: Cal failure error allowed:
Calibration time out: Zero ranges: Span ranges:
HOME ESCAPE INFO
Calibration Parameters
CAL NO/NOx
0.000 ppm
Disabled
50 %
SEPARATELY SEPARATELY

Figure 3-7: Calibration Parameters Display

9 s
Ye s
59 s
Calibration time out:
Sets how long the analyzer will wait for the signal to sta­bilize before issuing a Warning.
Zero (Span) ranges:
Used to select wether to calibrate ranges "TOGETHER" or "SEPARATELY". If together, zeroing or spanning will go through each range one by one. If the change re­quired is too great, it will fail and send an alarm if warn­ing alarms are enabled. In this case, Disable Calibration Adjustment Limits and try again. First check that the cali­bration gases are correct. If non-zero gases are used, or the changes are great, zero and span may have to be repeated a few times.
In case that a system calibration is used, set all ranges to "SEPARATELY".
Please note that Software will accept only span gas val­ues that are a factor of 10-110% of the measuring range. From that, it is possible, that not all of the four measuring ranges can be calibrated altogether. In that case you will have to set the option to "SEPARATELY".
Calibration adjustment limits:
Set to "Disable" to recover from a calibration failure.
Calibration averaging time:
Set the time used by the analyzer to average its reading during calibration. A longer time will give a better calibra­tion.
When using a system calibration, take care that averag­ing time is long enough for the analyzer to reach a settled reading. Otherwise, calibration may fail !
Calibration failure alarm:
When turned on ("yes"), issues a warning if the analyzer has to change its calibration by more than the Cal Fail­ure Error, if warning alarms are enabled.
Cal failure error allowed:
The percentage by which the calibration can change before an alarm is triggered if the Calibration Failure Alarm is enabled.
In case that a calibration is not possible because the difference of display to the true value is too great, switch of the "CalCheck" option.
3 - 5OperationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007

b. Flow Balance Adjustment:

To adjust the Flow Balance Valve, select:
Main (Menu)...
Analyzer basic controls (calibration) & setup...
Line#1
Measurement range number:
Range upper limit: Automatic range change control: Ranges with valid calibration: Calibration status: Span gas concentration:
Status: Measurement Mode: Ozonator status: Ozonator power:
Pressure limits...
HOME NO/NOx INFO
Basic Controls
ZERO SPAN
0.000 ppm
24.9 ppm Disabled
READY
10.0 ppm
NORMAL
NOx mode
OFF -PRES. SW.
Enabled

Figure 3-8: Basic Controls Menu

To adjust the flow balance, select the measuring range that is really in use and do a Zero- / Span adjust in NO mode.
1&2
Span Adjustment
After that, do a Span adjust in the same measuring range.
For that, push the F4 softkey in the Basic Controls menu, so that the Analyzer Span menu turns up:
Line#1
1
Are you sure ?
You must have span gas flowing through the analyzer.
Calibration time: Measurement range number: Span ranges: Calibration status: Error message for last span:
HOME NO/NOx INFO
Analyzer Span
SPAN
0.000 ppm
24 s
1
SEPARATELY
READY
CAL OK

Figure 3-10: Analyzer Span Menu

Push the softkey F4 again to start the spanning sequence. Be sure to have span gas flowing through your analyzer and to have it purged before.
Zero Adjustment
In the Basic Controls menu, push the F3 softkey and enter the Analyzer Zero menu.
Line#1
Are you sure ?
You must have zero gas flowing through the analyzer.
Calibration time: Measurement range number: Zero ranges: Calibration status: Error message for last zero:
.
HOME ESCAPE INFO
Analyzer Zero
ZERO
0.000 ppm
SEPARATELY
READY
CAL OK

Figure 3-9: Analyzer Zero Menu

Push the F3 softkey again, to start the zeroing process. Be sure to have zero gas flowing through your analyzer and to have it purged it from sample gas before.
When the zeroing process is finished, "Calibration sta­tus" will turn to "READY".
24 s
When the spanning sequence is finished, the "Calibra­tion status" will turn to "READY".
When having completed the zero- and span adjust in NO mode, switch over to NOx mode by pressing the F2 softkey in the Basic Controls menu. Observe the mea­suring value after switching into NOx mode:
If the measuring value stays constant (maybe af­ter a short peak immediately after switching into NOx mode), your flow balance is perfectly adjusted
1
and your analyzer is ready for measurement.
If the measuring value changes after switching to NOx mode, turn the flow balance valve carefully until the difference between measuring values in NO mode and NOx mode is minimized.
To make sure that your analyzer is correctly calibrated, switch back to NO mode, if necessary, repeat zero- and span adjustment and check for an existing difference between measurement in NO mode and in NOx mode.
3 - 6 Operation Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NOTE:
If you are using a span gas that has different concentrations of NO and NO
, measuring value
x
changes when switching between measuring modes NO and NOx, obey the following laws:
Measuring value in NO mode:
Concentration of NO in span gas
Measuring value in NOx mode:
Concentrations of (NO+NO
) in span gas
2
NGA 2000 CLD
3 - 7OperationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
3-5 SYSTEM & NETWORK I/O MODULE
CONTROLS (SETUP) - SYSTEM SIO
To adjust SIO functions, select
Main (Menu)...
Analyzer and I/O, expert controls & setup...
System & network I/O module controls...
System SIO module...
Line#1
Analog output setup...
Serial interface setup...
Relay outputs setup...
Module installed...
.
MEASURE
-- System SIO Module --
Back...
0.000 ppm

Figure 3-11: System SIO Module Menu

System SIO module menu allows you to adjust different SIO functions, like analog outputs and serial interfaces.
YES
Output number:
Choose the desired analog output (1-8) to set the pa­rameters. The number of outputs depends on the ana­lyzer configuration as 2, 4, 6 or 8.
Choose signal source module...
Select the "Analyzer Modules" submenu by selecting the "Choose signal source module..." line and pressing the Return () softkey.
Select the tag of the desired reference channel with the orsoftkeys and then press the or softkey. The display will return to the previous menu automatically and the selected reference channel will be displayed in the "Source module:" line.
The available selections may be different depending on the installed modules.
Line#1
-- Analyzer Modules --
0.000 ppm
Control Module: 0.0
1000723 Line#1: 1.0
CLD: 1.0
MLT/CH3
1

a. Analog Output Setup

In the System SIO Module menu (Figure 3-11) select "Analog output setup..."
Line#1
Output number:
Choose signal source module... Choose signal... Signal value for 0% output: Signal value for 100% output: Output current: Hold output during calibration: Signal name: Current signal value: Source module:
Measure More...
-- Analog Output Setup --
Back...
0.000 ppm
0.00
100.00
0...20 mA
????
-10.00
Line#1

Figure 3-12: Analog Output Setup Menu

No
HOME NO/NOx INFO
ZERO SPAN

Figure 3-13: Analyzer Modules Menu

Choose Signal...
1
Select the "Signals" submenu by selecting the "Choose Signal..." line and pressing the softkey. (The list of sig­nals will depend on the module chosen.)
Press the F5 softkey to go to additional menus to choose the Primary Variable signal for the analog output. The Primary Variable is the actual NO or NOx concentration.
See Section 5.2.1.1 of CLD Software Manual for com­plete list of signals.
The signal chosen here will be applied to the analog out­put (1-8) chosen above.
3 - 8 Operation Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
Line#1
Measure >>>
-- Signals --
<<< Back...
0.000 ppm
Sample flow:
Sample press.:
Ozonator:
Converter temp.:
Block temp.:
Detector temp.:
NO/NOx:
Noise level:
>>>
Line#1
Measure >>>
-- Signals --
Measurement mode:
<<< Back...
0.000 ppm
t90 time:
Output delay time:
Calibration status
Linearizer:
Operational state:
Health:
Interference:

Figure 3-14: Signals Menu

NOTE
If the measurement range is changed, the settings done in this menu will revert back to the standard values of the range. The output values can be changed permanently in the menu "Range Set­tings."
NOTE
The signal range of the analog output should be less than the smallest range of the channel. Oth­erwise the analog output may exhibit excessive noise.
Output current:
Select the desired output range in the "Output current range:" line. The options are 0...20 mA or 4...20 mA.
Hold output during calibration:
Enable this option to hold the analog output to the last value during calibration.
Pressing the F5 (More...) softkey changes the to the submenus "Output Signal if Assigned Module Fails" and "Fine Adjustment."
Signal value for 0% (100%) output:
It is possible to set the signal value for 0 % output and for 100% output so as to output only a portion of the entire range.
Example:
Range from 0 to 1,000 ppm
0% value to be 400 ppm, 100% value to be 700 ppm.
Analog output normally: 0 V = 0 ppm, 10 V = 1,000 ppm
After changing the output scaling: 0 V = 400 ppm, 10 V = 700 ppm.
Move the cursor to the "Signal value for 0 % output:" line and adjust the value to 400. Then change to the "Signal value for 100 % output:" line and adjust the value to 700.
Line#1
Output(s) value on analyzer failure:
Output number: Operation mode: Fine adjustment for 0% output: Fine adjustment for 100% output:
Measure More...
-- Output Signal If Assigned Module Fails --
BeginOfRange - 10%
-- Fine Adjustment --
Back...
0.000 ppm
Normal
4096
819
Figure 3-15: Output Signal If Assigned Module
Fails Menu
1
3 - 9OperationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
Output(s) value on analyzer failure:
Choose the desired signal level to cause a failure condi­tion. The choices are:
Acutal
BeginOfRange
EndOfRange
BeginOfRange-10%
BeginOfRange+10%
Output number:
Choose the output number (1-8) for setting the fine ad­justment.
Operation mode:
Normal: The absolute measurement signal will be sent to the analog output.
Adjust 0V: Used to set the display equal to the analog output for 0 V and 0 mA. Life zero signals (4 - 20 mA and 2 - 10 V) are set automatically and cannot be adjusted.
Adjust 10V: Used to set the display equal to the analog output for 10V und 20mA.
Source module: The name of the module chosen in the "Choose signal source module" menu.
Pressing the F5 (More...) softkey changes to the submenu "Special Scaling for Concentration Signal"
Line#1
Output #1:
Output #2: Output #3: Output #4: Output #5: Output #6: Output #7: Output #8:
Measure More...
-- Special Scaling for Concentration Signal -­(Scaling is the same as range limits)
Back...
0.000 ppm
Ye s
Ye s Ye s Ye s Ye s Ye s Ye s Ye s
Figure 3-16: Special Scaling for Concentration
Signal Menu
This menu allows for the setting of each of the 8 outputs to be the same as the range limits "Yes" or as to set on the previous menus.
Pressing the F5 (More...) softkey changes to the submenu "Analog Output Updates per Second."
Select the "Fine adjustment for 0% output" and/or "Fine adjustment for 100 % output" lines with the ↵ or → softkey. Adjust to the desired value with the ↑ or ↓ softkey and confirm with the ↵ softkey.
The range of values are:
3,000 to 6,000 for 0 % (default 4096)
600 to 1,000 for 100 % (default 819)
The last three lines of the "Analog Output Setup" menu are displayed only for configuration values of the analog output.
Signal name: The name of the signal chosen the "Choose signal" menu.
Current signal value: The current value of the variable.
Line#1
Output #1: Output #2: Output #3: Output #4: Output #5: Output #6: Output #7: Output #8:
Measure
-- Analog Output Updates Per Second --
Back...
0.000 ppm
Figure 3-17: Analog Output Updates per
Second Menu.
This menu allows for the setting of the update rate for each of the 8 outputs.
1
0 0 0 0 0 0 0 0
3 - 10 Operation Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

b. Serial interface Setup

The submenu "Serial interface Setup" is used to set the parameters for data transfer between the analyzer and external devices. The choices in this menu depend on the configuration of the analyzer. The full specification of the serial interface is described in its own manual.
In the System SIO Module menu (Figure 3-11) select "Serial interface setup..."
Line#1
-- Serial Interface Setup --
Baud rate:
Data bits: Stop bits: Parity: Echo mode: Handshake: Transmission delay: Type of installed serial interface: Communication protocol: Special protocol definitions...
Pressure limits...
Measure
Back...
0.000 ppm
19200
Disabled
Xon/Xoff
RS232

Figure 3-18: Serial Interface Setup Menu

None
AK
Line#1
Device address (RS-485 only): 1
Measure More...
-- AK Protocol Deifinitions --
Back...
0.000 ppm

Figure 3-19: AK Protocol Definitions Menu

8 1
0
The value can range from 1 to 50.
Options:
Baud rate: 300, 1200, 2400, 4800, 9600, 19200
Data bits: 7, 8
Stop bits: 1, 2
Parity: None, Even, Odd
Echo mode: Enabled, Disabled
Handshake: None, Xon, Xoff
Transmission delay: 0...100.
Type of installed serial interface: RS232, RS485/2w, RS485/4w, RS485/4w bus, None.
Communication protocol: AK, MODBUS RTU, None (not applicable to CLD)
NOTE
The "special protocol definitions..." line accesses a submenu for setting the parameters of the AK and MODBUS RTU (not available yet) communi­cation protocols.
3 - 11OperationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007

c. Relay Outputs Setup

There are three relays on the SIO board. The contact logic can be set with a jumper on the SIO board to select NO (normally open) or NC (normally closed). Full details of the SIO board are contained in its own manual.
In the System SIO Module menu (Figure 3-11) select "Relay outputs setup..."
Line#1
Output number:
Invert signal: Choose source module... Choose signal...
Signal comes from: Signal name: Actual status:
Measure
-- Relay Outputs Setup --
Back...
0.000 ppm
Disabled

Figure 3-20: Relay Outputs Setup Menu

Output number:
Corresponds to the relay number 1-3.
???? ????
Off
Choose signal...
Line#1
1
Pressure limits...
Measure >>>
-- Choose Signal --
Maintenance request
<<< Back...
0.000 ppm
1
Function control
Failure
Cal. In progress
Zero in progress
Span in progress
Zero failed
Span failed

Figure 3-22: Choose Signal Menu

Choose desired signal for the relay output number (1-3) being configured.
The list of signals will depend on the chosen module. If available, press the >>> (F5) softkey for additional sig­nals.
The three lines displayed at the bottom of the "Relay Outputs Setup" menu show the current status of the se­lected relay output.
Invert signal:
"Disabled" signal is normal, "Enabled" signal is inverted.
Choose source module...
Line#1
Measure >>>
-- Choose Source Module --
1000723 Line#1: 1.0
<<< Back...
0.000 ppm
Control Module: 0.0

Figure 3-21: Choose Source Module Menu

Choose desired source module for the relay output num­ber (1-3) being configured.
The list of modules will depend on the installed modules.
Signal comes from: The module chosenfrom the "Choose Source Module" menu.
Signal name: The signal chosen from the "Choose Sig­nal" menu.
Actual status: The current status of the signal; Off or On.
3 - 12 Operation Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

3-6 CONVERTER TEMPERATURE ADJUSTMENT

The vitreous carbon converter used in this analyzer mod­ule must be checked periodically to assure that it is work­ing at peak efficiency. The efficiency of the converter is typically 95 % to 98 %, that is, 95 % to 98 % of the nitro­gen dioxide introduced to the module is reduced to nitric oxide. That is well above the 90 % minimum required by several Environmental Protection Agency.
Two conditions reduce the efficiency of the converter:
The converter is operating at too low a tempera­ture and the efficiency drops or
The converter is operating at too high a tempera­ture and the nitrogen dioxide is reduced to nitro­gen, which is not detectable by the chemilumines­cence reaction.
Initially, an interval of one week between converter effi­ciency checks is recommended because high tempera­ture operation changes conditions inside the converter. The active surface area of the vitreous carbon increases through use. Initially, when the surface area is low, the temperature at which converter efficiency peaks is rela­tively high. This peak temperature moves downscale as surface area increases, and less external energy is re­quired to cause adequate conversion.
The nominal range of converter operational tempera­ture is 300 °C to 400 °C (572 °F to 752 °F). The current converter temperature can be viewed in the "Physical measurement" menu. To access the converter tempera­ture adjustment:
Main (Menu)...
Analyzer and I/O expert controls & setup...
Analyzer module setup...
Physical measurements parameters...
Follow this procedure to optimize the operating tempera­ture of the converter:
1. Power up the module and allow it to stabilize at operating temperature (about one hour).
2. Check the Converter Temperature on the Physi­cal Measurements menu (see Section 3-3). Note the value for future reference.
3. Introduce a calibration gas of known (NO
) con-
2
centration into the analyzer and note concentra­tion value determined when the full response has been achieved.
4. Change Converter Setpoint in the "Temperature Control" menu to 300 °C. Allow module 15 min­utes to stabilize, recheck the concentration value and note the value for later use.
5. Increase the Converter Setpoint value by 20 °C, wait 15 minutes, and note the concentration value. Repeat this step until either a converter efficiency of between 95 % and 98 % is obtained or the final 20 °C increment produces an efficiency increase of less than one percent.
6. Decrease the Converter Setpoint value by 5 °C, which places the converter at a temperature suit­able for low ammonia interference and efficient NO conversion.
7. Recheck the Converter Temperature value in the "Physical Measurements" menu, and compare it to the initially recorded value.
NOTE
Converter temperature is not a direct measure of converter efficiency. Temperature measurement is for reference purposes only.
2
3 - 13OperationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
3-7 MEASUREMENT OF CONVERTER
EFFICIENCY
It is the responsibility of the user to measure efficiency of the NO
-to-NO converter during initial startup and there-
2
after at intervals appropriate to the application (normally once a month).
The reactant material used in the converter provides the optimum combination of high conversion efficiency and low ammonia interference. Unlike most competitive ana­lyzers, the NGA 2000 CLD Analyzer Module utilizes a reactant material that gradually becomes more efficient at a given temperature. Thus, after a period of use, op­eration at a lower temperature setpoint than initially re­quired is possible.
a. Test Setup for Measurement of Conversion
Efficiency
A typical setup for measurement of conversion efficiency is shown in figure 3-23. The test setup includes:
A cylinder on nitric oxide standard gas consisting of NO in N
.
2
The concentration on NO in the standard gas should be about the full-scale value of the range under test. The test sample supplied to the analyzer should contain a concentration of NO comparable to that in the samples that are to be analyzed. Alternatively, a higher concen­tration NO standard may be used if the test setup in­cludes provision for diluting it appropriately with zero air. Suitable standard gases are available from various sup­pliers. Stainless steel cylinders are commonly used, but specially treated aluminium is preferred for low parts­per-million NO samples.
An ozone generator utilizing an ultraviolet lamp, not a corona discharge.
A corona discharge ozone generator is undesirable be­cause it may produce oxygen atoms, which can then combine with atmospheric nitrogen to form NO. The re­sult can erroneuosly high value for the measured con­version efficiency.

b. Test Procedure

1. Measure converter temperature in the sub-menu structure. Note present reading as a reference for compari­son with subsequent readings.
2. Lower converter temperature to 300 °C by setting control parameters, and wait 15 minutes for tem­perature equilibration.
3. Connect a Converter Efficiency Tester (e.g. Model
958) to the CLD Module (see figure 3-23), and fol­low Steps 4 through 17 below:
4. Attach the NO/N
supply to C2, the air supply to
2
C1, and the CLD Module inlet fitting to C3.
5. With the variable transformer off, switch the CLD Module to NO mode, and close valve MV1.
6. Open valve MV2 until the CLD Module SAMPLE Pressure Gauge reaches operating pressure and the BYPASS flowmeter indicates some bypass flow. Wait until stable readings are obtained by the CLD Module.
7. Zero and span the Analyzer output to indicate the value of the NO concentration being used. This value should be about 80 % of full-scale. Record this concentration.
8. Open valve MV1 (air supply metering valve) and adjust to blend enough air to lower the NO con­centration (as noted above) about 10 %. Record this concentration.
9. Power up the ozonator, and increase its supply voltage until NO concentration noted in Step 8 is reduced to about 20 percent of the concentration noted in Step 7. Wait for stabilization. NO
is now
2
being formed from the NO+O3 reaction. There must always be at least 10 percent un-reacted NO at this point. Record this concentration.
10. Switch the CLD Module to NO
mode. Total NO
x
concentration is now output to the network for dis­play. Record this concentration.
11. Turn off the ozonator, and allow the Analyzer read­ing to stabilize. Total NO
concentration of the di-
x
lute NO span gas initially used is displayed. Record this concentration.
x
3 - 14 Operation Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
12. Close valve MV1. The NO concentration should be equal to or greater than the reading in Step 7. This indicates whether the NO contains any NO2.
13. Calculate the efficiency of the NO converter by sub­stituting the concentrations obtained during the test in the equation below:
%efficiency = 1 + x 100
(b-a) (c-d)
Where:
a = recorded concentration in Step 10
b = recorded concentration in Step 11
c = recorded concentration in Step 8
d = recorded concentration in Step 9
In the example in figure 3-23 the following calcula­tions would apply:
%efficiency = 1 + x 100 = 92%
(80-85) (80-20)
16. Reset converter temperature setpoint 5 °C lower. Converter temperature is now set to the front edge of the plateau on the efficiency-vs.-temperature curve (see figure 3-23). This setting should pro­vide the optimum combination of high conversion efficiency and low ammonia interference.
17. Wait 15 minutes for temperature equilibration, and check converter temperature. Compare present temperature with original value. Normally, converter temperature should be in the range of 300 °C
to 400 °C (572 °F to 752 °F).

c. Subnormal Conversion Efficiency

If measured conversion efficiency of between 95 % and 98 % is unobtainable within the normal temperature range, the most probable cause is depletion of the cata­lytic material within the converter.
However, before concluding that the converter is defec­tive, ensure that the conversion efficiency measurement is accurate. Though the measured efficiency is less than the 95 % to 98 % range, the actual efficiency may be somewhat higher.
Efficiency checks should be made on each ana­lyzer range, using an NO span gas concentration appropriate to the instrument range.
NOTE
In the initial measurement, after lowering the tem­perature setpoint in Step 2, the efficiency will nor­mally be less than 92 %
14. Reset converter temperature setpoint 20 °C higher, wait 15 minutes for temperature equilibration, and measure conversion efficiency by repeating Steps 3 through 13. Conversion efficiency should be improved.
15. Repeat Step 14 until:
a) 95 % to 98 % efficiency is attained or
b) the final 20 °C converter temperature adjust-
ment yields an increase in efficiency of less than 1 %.
An apparent subnormal efficiency can be due to a prob­lem external to the Analyzer Module, perhaps located either within the test setup or between it and the Ana­lyzer Module. Check the following:
1. Leakage
2. Loss of NO
between test setup and Analyzer
2
Module. Such loss can occur by reaction with a rubber diaphragm in a pressure reagulator or flow controller. Stainless steel diaphragms are pre­ferred. Loss can also occur during passage through filter media.
3 - 15OperationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007

d. Replacement of Converter

If the subnormal conversion efficiency is real, and not due to measurement error introduced by the test setup, the converter must be replaced. See section 4-6.
The usual case of converter failure is destruction of a large part of the catalytic material be excessive heat. This is due either to an excessively high temperature setpoint or failure of the converter temperature control circuitry.

e. Capillaries

Replacement vent and pump capillaries should be in­stalled finger-tight. Use of a wrench can constrict capillariesm, thus chang­ing flow rate. Sample capillary is metal; use a wrench for tightening.

f. TEA Scrubber

The presence on NO
in the NO cylinders can cause
2
inaccurate converter efficiency values. The TEA Scrub­ber accessory can be used to remove residual NO2 from the NO cylinders. Use of this accessory allows a NO2­free NO calibration gas.
3 - 16 Operation Emerson Process Management GmbH & Co.OHG
Instruction Manual
A
HAS60E-IM-HW June 2007
NGA 2000 CLD
. TYPICAL TEST SETUP
FLOWMETER FM2
METERING VALVE MV2
CONNECTOR C2
STANDARD GAS: NO IN N2 BACKGROUND
CONNECTOR C3
OFF
ON
VAC
VARIABLE TRANSFORMER
T
CONNECTOR C1
ZERO AIR
OZONATOR UTILIZING ULTRAVIOLET LAMP
FLOWMETER FM1
METERING VALVE MV1
CAUTION: Externally limit sample flow rate to less than 2200 cc/min.
TO SAMPLE INLET OF CLD ANALYZER MODULE
FLOW APPROXIMATELY 3 LITERS PER MINUTE
ALL LINES AND FITTINGS STAINLESS STEEL OR TEFLON
B. TYPICAL TEST RESULTS
NO in
2
N
90
85
20
80
0
c
AIR ADDED
c-d
NO Mode
OZONATOR OFF
OZONATOR ON

Figure 3-23: Converter Test Setup

a
d
TO NOX MODE
NOX Mode
a-b
b
% Efficiency = 100
a - b
1+
( )
c - d
3 - 17OperationEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
Convert er
Efficiency
Abo ut 97 %
Final 2 degree high er adj ustm ent
Final 1 degree lower adjustment
Convert er Tem perat ure

Figure 3-24: Converter Temperature Adjustment

3 - 18 Operation Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
SECTION 4
MAINTENANCE AND SERVICE
WARNING
Before starting work, read the "Essential instruc­tions" on the inside cover and the Safety Summary beginning on page P-2. Failure to follow the safety instructions could re­sult in serious injury or death.
NGA 2000 CLD

4-1 OVERVIEW

The CLD Analyzer Module requires very little mainte­nance during normal operation.
The gas path system should be leak tested at least twice a year and after maintenance, replacement or repair of gas path parts.
Occasionally, the intake fan screen may require clean­ing, refer to section 4-3.
WARNING
Do not operate without covers secure.
Do not open while energized.
Disconnect power to the module(s) prior to replacing components.
This equipment should not be adjusted
or repaired by anyone except properly
qualified service personnel.
CAUTION
Tampering with or unauthorized substitution of components may adversely affect the safety of this instrument. Use only factory documented/ap­proved components for repair.
Because of the danger of introducing additional hazards, do not perform any unauthorized modifi­cation to this instrument!
Also, the detector´s reaction chamber and sapphire win­dow may require cleaning, refer to section 4-7.
White crystal deposits on the windows of the reaction chamber and plugging of capillaries and vent are usually due to sample contaminates such as ammonia reacting with the high ozone levels and NO components. To elimi­nate the contaminates, the sampling system should be reworked or a preventive maintenance program devel­oped (if dropout is excessive). Another source of crys­talline formation is contaminated air.
Several components may require replacement. These are discussed in the following sections.
Tag each connector and its location before disconnect­ing any wiring. This helps in reassembly.
4 - 1Maintenance and ServiceEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007

4-2 FUSES

The main power fuse may require replacement.
NOTE
Before replacing the fuse, remove power to the Analyzer Module.
See figure 2-3 for the location of the main power fuse
[T 6A 250 V (6x32 mm)], which protects 24 VDC input to the module.
NOTE
Use only fuses of the correct type and current rat­ings as replacements. Using repaired fuses and short circuiting of fuse holders is prohibited.

4-3 FANS

Refer to figure 4-1. To replace either rear panel fan, re­move the cover of the Analyzer Module and then the rear panel. Disconnect connectors and remove screws. Assemble in reverse order.

4-4 OZONATOR

4-5 PRINTED CIRCUIT BOARDS

CAUTION
The electronic parts of the Analyzer Module can be irreparably damaged if exposed to electrostatic discharge (ESD).
The instrument is ESD protected when the covers have been secured and safety precautions ob­served. When the housing is open, the internal components are not ESD protected anymore.
All four printed circuit boards can be replaced, if neces­sary. Refer to figure 4-1 for location of the Driver, Power Supply, Signal and Computer Boards.
To remove any PCB (except the Computer Board), dis­assemble the enclosure side first. Ribbon and other cables are long enough to allow the entire side to be folded out from the remainder of the components. This makes PCB removal much simpler.
Refer to figure 4-1. To replace the ozonator, remove the two large straps and all tie-wraps, and disconnect the one electrical connection. Reassemble in reverse order.
4 - 2 Maintenance and Service Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
Ozonator
Transist or
655264
Sample
Press ure
Sensor
655253
Driv er Board
655620
Ozone Press ure Sensor
655254
PCB I nsulato rs
Insulator
Com puter Analy sis
Board 658350
Signal
Board
655580
657719
NO/ NOx Solenoid Valv e
659477
Press ure Switch 655215
Ozonator Power Sup ply
657716
3
Thermost at 657298
4
Flow Sensor 902931
Glass Tube
Connec tors
Fan Guard
Conv ert er Ass embly 655250
See Figure 4-2 -
1
Fan Guar d
2
Fan 655245
Detector Assembly, see Figure 4-3
(PN dependin g on conf iguration)
EMI Shield
Detector Case Insulator
5
Flow Balance Metering Valv e 903207
Sample Regulator
(depending on
conf iguration,
see Table 1-1)
4 Optional Bypass Flow Sensor
Fuse, Power
903347
LON/ Power Mod ule
Power Supply Board
657520
Brief Description Part Number
Detector 200 cc/min. @ 5 psig 659754
Detector 200 cc/min. @ 2 psig 42716203
Detector 70 cc/min. @ 5 psig 42716204
Detector 70 cc/min. @ 2 psig 42716205
Detector assembly without capillary 659754X
3 Ground wires from ozonator shown
2 Flow direction of intake fan (shown) is into case. Flow direction of exhaust fan is out of case.
1 Fan guard between fan and EMI filter in on the intake fan only (shown).

Figure 4-1: CLD Module Assembly

5
4 - 3Maintenance and ServiceEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD

4-6 CONVERTER

CAUTION
HIGH TEMPERATURES !
While working at thermostated components inside the Analyzer Modules hot components may be accessible!
Converter complete (655250)
Instruction Manual
HAS60E-IM-HW
June 2007
Refer to figure 4-1 and figure 4-2. To replace the con­verter or temperature sensor, disconnect the two pneu­matic tubes and two electrical connections. Unlace the heater blanket, and remove the converter. Reassemble in reverse order, ensuring that the converter is oriented with the glass cloth at the bottom and the sensor is ori­ented correctly inside the heater jacket.
Heater Jacket
(655228-R1)
Temperature Sensor
(655282-R1)
Wrap with
Aluminium Foil
Converter Tube
(655227)
Connectors
(632784)

Figure 4-2: Converter Assembly

4 - 4 Maintenance and Service Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

4-7 DETECTOR DISASSEMBLY

Refer to figure 4-3.

a. Reaction Chamber Removal

Disconnect the stainless steel tubing lines at the fittings. Remove the (4) nuts holding the Detector Assembly to the chassis. Disconnect the plug from connector J1 on the Signal Board and remove the assembly from the chassis.
NOTE
Care should be taken to avoid getting heatsink compound on optical surfaces. If this substance is removed during the disassembly process, a zinc­oxide-filled, silicone grease (e.g. Dow Corning 340 or EG & G Wakefield Engineering´s Series 120 Thermal Joint Compound) be reapplied in the re­assembly of this component.
Although the heater and thermostat can be removed to facilitate handling, contact with the white heatsink com­pound can be minimized by leaving these items in place. Remove the (2) screws holding the top plate of the De­tector, and move the plate along the wires and away from the Detector.

c. Photodiode Removal

Remove the Detector Assembly as described above. In­vert the housing to access the mounting bracket. Re­move the (3) screws and shoulder washers from the bracket, insulating disk and bottom plate as a unit to mini­mize the spread of the heatsink compound.
Remove the (2) screws holding the lower section of the Detector Housing, then slide the section along the cable and remove.
Remove the (2) screws holding the socket, thermistor and photodiode in place, being careful not to lose the washers that are used as shims.
Grasp the socket and photodiode base while slowly ro­tating to separate the photodiode from the housing. Some friction will be felt as an O-ring is used around the photo­diode as a seal.

d. Photodiode Installation

To replace the photodiode, carefully remove the diode from the green socket, and replace with a new one. Be­fore mounting the new diode, the top cap of the enclo­sure should be temporarily removed and the (2) screws holding the Reaction Chamber loosened about two turns.
Remove the (2) screws holding the tube assembly in place. Hold the tubing with one hand while inverting the Detector Housing with the other, allowing the Reaction Chamber O-ring and window to be removed from below.

b. Reaction Chamber Installation

To reinstall, hold the housing in the inverted position while sliding the Reaction Chamber O-ring and window into position and the tubing into the slot in the housing. Hold the Reaction Chamber in place while rotating the hous­ing upright. Replace the hold-down screws.
NOTE
The procedure described above is for the purpose of maintaining the relative positions of windows and O-ring to the Reaction Chamber during instal­lation.
Replace the top cap and screws. Reverse the removal procedure to reinstall the Detector Assembly into the Analyzer Module.
This allows air which is trapped between the O-ring seals to escape when the diode is inserted. It also maintains the position of the O-ring and window in the upper com­partment.
The new photodiode should be slowly inserted into the housing while gradually rotating the body. This allows the O-ring to properly seat. Continue replacing screws, wash­ers, thermistors, etc., with the thicker shim (washer) on the opposite side of the socket from the thermistor.
Replace the lower section of the housing, then the bot­tom cover, insulator and bracket with the shoulder wash­ers and screws.
Re-tighten the screws in the Reaction Chamber (upper section). Replace the top cap and its screws.
To reinstall in the Analyzer Module, reverse the proce­dure for removal as indicated above.
4 - 5Maintenance and ServiceEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
e
e
w
Photodiode
Ozone
Sample
M3X0.5 x 16mm Screw (2) 3mm Spring W ash er (2)
Exhaust
Heater
*
Sapphir
Windo
Reaction Chamber
Thermistor Asse mbl y
Photodiode S ock
Asse mbl y
Detector Mounting Bracket
Heater
Ther mo stat
Instruction Manual
HAS60E-IM-HW
June 2007
M3X0.5 x 25mm Screw (2)
3mm Spring W ash er (2)
Detector Hea der
*
*
Retain er Gas ket
Reacti on Ch amber
Phot odi od e Cable
Insul ator (between Lower Cover and Mounting Br acket)
*
Heater/Thermostat Assembly 655235.
Phot odi ode Cas e
Ground
M3X0.5 x 16mm
Screw (3)
Photodiode
Tubing Cov er
Lower Cover
Nylon Shoulder
Washers ( 3)
Thermistor
655216
655258
O - Ring 876478
Phot odi ode Ass embl y
(see detail below)
M3X0.5 x 20mm Screw (2)
(see detail below)
3mm Spring W ash er (2)
Socket Assembly
Detector Cover
M3X0.5 x 16mm Screw (2)
3mm Spring W ash er (2)
Photodiode
Ground
Filterboard
O - Ring 854540
Sapphi re W ind ow
Cushio ning Gasket
4 - 6 Maintenance and Service Emerson Process Management GmbH & Co.OHG

Figure 4-3: Detector Assembly

Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

4-8 LEAKAGE TEST

The gas path system should be leak tested at least twice a year and after maintenance, replacement or repair of gas path parts.

a) Required Tools

Test Medium
external Manometer
Swagelok® Blind Unions (2)

b) Procedure

To perform aleakage testing, proceed as follows (see figure 4-4):
1. Close sample and ozonator air supply
2. Disconnect the gas connections
3. Close sample inlet and ozonator inlet with blind unions
4. Connect a pressure meter to the exhaust fitting
5. Connect test medium supply (N2 or He) to the pres­sure meter
6. Supply Analyzer Module with the test medium with a pressure of approx. 2,000 hPa (15 psig) and close supply . Since CLD internal it comes to equalization of pres­sure, supply is to open and to close repeatedly until manometer doesn’t show pressure changes.
7. Watch the manometer. Over a period of about 15 minutes the pressure drop may not be higher than 7.5 hPa/min. using Helium (He) or 2.5 hPa/min. using Nitrogen (N2)
If the specifications (see table 4-1) were adhered to, then the test is finished.
N
2
or
He
test medium
Gas source with
External Manometer:
15 psig = 1,000 hPa
Ozonator Air
Closed Gas Connections
P
Sample
Exhaust
CLD Analyzer Module

Figure 4-4: Principle Leakage Test Assembly

Test Medium Nitrogen (N2) Helium (He)
Test Pressure
Test Time
permissible
Drop in Pressure
15 psig / 2,000 hPa / 1 bar 15 psig / 2,000 hPa / 1 bar
15 min 15 min
2.5 hPa / min 7.5 hPa / min

Table 4-1: Leakage Test Specifications

4 - 7Maintenance and ServiceEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
If there is a leakage, then proceed as follows:
a) Using Helium (He) for test medium
Supply Analyzer Module with the test medium with a pressure of approx. 2,000 hPa (15 psig).
Look for the leak with a helium leak detector.
b) Using Nitrogen (N2) for test medium
Supply Analyzer Module with the test medium with a pressure of approx. 2,000 hPa (15 psig).
Liberally cover all fittings, seals, and other pos­sible sources of leakage with a suitable leak test liquid such as SNOOP. Bubbling or foam­ing indicates leakage.
Remove the source of leak.
Perform a leakage test once more as decribed obove.
Instruction Manual
HAS60E-IM-HW
June 2007
.
4 - 8 Maintenance and Service Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
SECTION 5
TROUBLESHOOTING

5-1 OVERVIEW

Liberally cover all fittings, seals, and other possible sources of leakage with a suitable leak test liquid such as SNOOP. Bubbling or foaming indicates leakage. Checking for bubbles will locate most leaks but could miss some, as some areas are inaccessible to the appli­cation of SNOOP. For positive assurance that system is leak free, perform one of the tests above.
NGA 2000 CLD
5 - 1TroubleshootingEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
5 - 2 Troubleshooting Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
SECTION 6
REPLACEMENT PARTS
CAUTION
Tampering with or unauthorized substitution of components may adversely affect the safety of this instru­ment. Use only factory documented/approved components for repair.
Because of the danger of introducing additional hazards, do not perform any unauthorized modification to this instrument!

6-1 MATRIX

Each analyzer is configured per the customer sales or­der.
To identify the configuration of an analyzer, locate the analyzer name plate label. The analyzer matrix appears on the analyzer name plate label.
Measuring Range (Fullsclae Range)
Analyzer Matrix
Serial Number
Figure 6-1: Name Plate Label
6 - 1Replacement PartsEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007

6-2 REPLACEMENT PARTS

658157 Air Restrictor, 430 cc/min. @ 12 psig 655269 Back pressure regulator,
Brass/Neoprene, 5 psig
659063 Back pressure regulator,
Stainless steel/Viton, 5 psig
660400 Back pressure regulator,
Brass/Neoprene, 2 psig
660401 Back pressure regulator,
Stainless steel/Viton, 2 psig
655246 Cable, 24-pin Flat ribbon,
Comp Bd.--Driver Bd.
903032-V1 Cable, 10-pin Flat ribbon,
Comp Bd.--Power Supply Bd--Driver Bd
903033 Cable, 24-pin Flat ribbon,
Comp Bd.--Signal Bd.
655249 Cable, 30-pin Flat ribbon,
Comp Bd.--Signal Bd.
903034-V1 Cable, 8-pin Flat ribbon,
Comp Bd.--Power Supply Bd. 634398 Capillary, Vent, 70 cc/min @ 5 psig 657473 Capillary, Pump, 200 cc/min. @ 5 psig 660405 Capillary, Pump, 200 cc/min. @ 2 psig 659658 Capillary, Sample, 200 cc/min. @ 5 psig 660404 Capillary, Sample, 200 cc/min. @ 2 psig 659657 Capillary, Sample, 70 cc/min. @ 5 psig 660403 Capillary, Sample, 70 cc/min. @ 2 psig 658350 Computer Board PCB 655250 Converter, complete 659754 Detector, compl., 200 cc/min. @ 5 psig 42716203 Detector, compl., 200 cc/min. @ 2 psig 42716205 Detector, compl., 70 cc/min. @ 5 psig 42716206 Detector, compl., 70 cc/min. @ 2 psig 655620 Driver Board PCB 655245 Fan 24 VDC (Exhaust and Intake) 846761 Fan Guard 655512 Filter, EMI, Fan 902931 Flow sensor, 400 cc/min. - 2,000 cc/min. 903347 Fuse, Main Power T 6A 250V 656761 LON/PWR Board PCB 903207 Needle valve, 1/8", Flow Balance 659494 Ozon Generator (Spiral lamp only)
657719 Ozon Generator 657716 Power Supply Ozon Generator 657720 Power Supply PCB 655253 Pressure sensor, 0-15 psig (Sample) 655254 Pressure sensor, 0-30 psig (Ozone) 655215 Pressure switch, Pressure 500 hPa 659287 PROM SW-Version 2.2.1 659287-R1 PROM SW-Version 2.3 659895-R1 PROM SW-Version 3.3.0 659895-R2 PROM SW-Version 3.3.1 659895-R3 PROM SW-Version 3.3.3 659895-R4 PROM SW-Version 3.3.4 659895-R5 PROM SW-Version 3.6 659895-R6 PROM SW-Version 3.7.0 659895-R7 PROM SW-Version 3.7.1 659895-R8 PROM SW-Version 3.7.2 659895-R9 PROM SW-Version 3.9.3 659895-R10 PROM SW-Version 3.9.4 655580 Signal Board PCB 659477 Solenoid valve assembly, 24 VDC, NO/ NOx 657298 Thermostat assembly Converter 80 °C 655264 Transistor assembly, NPN
655250 Converter Replacement Parts
655227 Glas tube, filled and conditioned 655228-R1 Heater Jacket 655282-R1 Temperature Sensor, Platinum
Detector Replacement Parts
655218-V1 Cable Photodiode 659754X Detector assembly (without Capillary) 655235 Heater/Thermostat Assembly 854540 O-Ring, Viton 0.739 ID 0.87 OD 876478 O-Ring, Viton 0.737 ID 0.94 OD 655258 Photodiode 655216 Temperature Sensor
6 - 2 Replacement Parts Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
SECTION 7
RETURN OF MATERIAL

7-1 RETURN OF MATERIAL

If factory repair of defective equipment is required, pro­ceed as follows:
1. Secure a return authorization from a Emerson Processs Management Sales Office or Represen­tative before returning the equipment. Equipment must be returned with complete identification in accordance with Emerson instructions or it will not be accepted.
2. In no event will Emerson be responsible for equip­ment returned without proper authorization and identification.
3. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping.
4. In a cover letter, describe completely:
a. The symptoms that determined the equipment
is faulty.
b. The environment in which the equipment was
operating (housing, weather, vibration, dust, etc.).
c. Site from where the equipment was removed.
d. Whether warranty or non-warranty service is
expected.
e. Complete shipping instructions for the return
of the equipment.
5. Enclose a cover letter and purchase order and ship the defective equipment according to instructions provided in the Emerson Return Authorization, pre­paid, to:
EMERSON Process Management
Process Analytic Division
Customer Service Center
USA: +1 (800) 433-6076
EMERSON Process Management
GmbH & Co. OHG
D-63594 Hasselroth, Germany
Industriestrasse 1
EU: +49 (6055) 884-0 Fax: -209
If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the conditions listed in the standard Emer­son warranty, the defective unit will be repaired or re­placed at Emerson´s option, and an operating unit will be returned to the customer in accordance with the ship­ping instructions furnished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.

7-2 CUSTOMER SERVICE

For order administration, replacement parts, application assistance, on-site or factory repair, service or mainte­nance contract information, contact:
EMERSON Process Management
Process Analytic Division
Customer Service Center
USA: +1 (800) 433-6076 EU: +49 (6055) 884-470

7-3 TRAINING

A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Cur­rent Operator and Service Training Schedule, contact the Technical Services Department at:
EMERSON Process Management
GmbH & Co. OHG
D-63594 Hasselroth, Germany
Industriestrasse 1
EU: +49 (6055) 884-470/-472 Fax: -469
EMERSON Process Management
Process Analytic Division
Customer Service Center
USA: +1 (800) 433-6076
7 - 1Return of MaterialEmerson Process Management GmbH & Co.OHG
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
7 - 2 Return of Materials Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD

LIST OF FIGURES AND TABLES

1 LIST OF FIGURES

Figure A-1: From separate analyzers to analyzer system ............................................................... A - 1
Figure A-2: Example of NGA cabling .............................................................................................. A - 2
Figure A-3: Example/Possibilities of NGA Analyzer System ........................................................... A - 3
Figure 1-1: Function Principle of CLD Measurement .......................................................................1 - 1
Figure 1-2: Flow Diagram - CLD Analyzer Module with Bypass Flow Sensor .................................1 - 2
Figure 1-3: Flow Diagram - CLD Analyzer Module without Bypass Flow Sensor ............................1 - 3
Figure 1-4: CLD Analyzer Module - Top View ..................................................................................1 - 4
Figure 2-1: Analyzer Module Installation into Instrument Platform (view without front panel) .........2 - 1
Figure 2-2: CLD Rear Panel Connections........................................................................................2 - 3
Figure 2-3: Front Panel Controls and Electrical Connections ..........................................................2 - 4
Figure 2-4: CLD Wiring Diagram .....................................................................................................2 - 5
Figure 2-5: Outline and Mounting Dimensions.................................................................................2 - 6
Figure 3-1: Measure Mode Display ..................................................................................................3 - 2
Figure 3-2: Physical Measurements Display ....................................................................................3 - 2
Figure 3-3: Excerpt of a Test Data Sheet with values that are to be compaired with .............................
physical measurements. ................................................................................................3 - 3
Figure 3-4: Calibration Gas List Channels 1 and 2 ..........................................................................3 - 4
Figure 3-5: Calibration Gas List Channels 3 and 4 ..........................................................................3 - 4
Figure 3-6: Range Settings Menu ....................................................................................................3 - 4
Figure 3-7: Calibration Parameters Display .....................................................................................3 - 5
Figure 3-8: Basic Controls Menu......................................................................................................3 - 6
Figure 3-9: Analyzer Zero Menu ...................................................................................................... 3 - 6
Figure 3-10: Analyzer Span Menu .....................................................................................................3 - 6
Figure 3-11: System SIO Module Menu .............................................................................................3 - 8
Figure 3-12: Analog Output Setup Menu ...........................................................................................3 - 8
Figure 3-13: Analyzer Modules Menu ................................................................................................3 - 8
Figure 3-14: Signals Menu .................................................................................................................3 - 9
Figure 3-15: Output Signal If Assigned Module Fails Menu............................................................... 3 - 9
Figure 3-16: Special Scaling for Concentration Signal Menu .......................................................... 3 - 10
Figure 3-17: Analog Output Updates per Second Menu. .................................................................3 - 10
Figure 3-18: Serial Interface Setup Menu ........................................................................................3 - 11
Figure 3-19: AK Protocol Definitions Menu ......................................................................................3 - 11
Figure 3-20: Relay Outputs Setup Menu .........................................................................................3 - 12
Figure 3-21: Choose Source Module Menu ..................................................................................... 3 - 12
Figure 3-22: Choose Signal Menu ...................................................................................................3 - 12
L - 1List of Figures and TablesEmerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW
NGA 2000 CLD
June 2007
Figure 3-23: Converter Test Setup ...................................................................................................3 - 17
Figure 3-24: Converter Temperature Adjustment ............................................................................3 - 18
Figure 4-1: CLD Module Assembly ..................................................................................................4 - 3
Figure 4-2: Converter Assembly ...................................................................................................... 4 - 4
Figure 4-3: Detector Assembly ........................................................................................................4 - 6
Figure 4-4: Principle Leakage Test Assembly ..................................................................................4 - 7
Figure 6-1: Name Plate Label ..........................................................................................................6 - 1

2 LIST OF TABLES

Table A-1: Possibilities of NGA 2000 I/O combinations ................................................................. A - 4
Table 1-1: Components depending on Module Configuration ........................................................ 1 - 2
Table 2-1: Gas Specifications .........................................................................................................2 - 2
Table 4-1: Leakage Test Specifications ..........................................................................................4 - 7
L - 2 List of Figures and Tables Emerson Process Management GmbH & Co.OHG
Instruction Manual
HAS60E-IM-HW June 2007
NGA 2000 CLD
Emerson Process Management GmbH & Co.OHG
NGA 2000 CLD Hardware
WORLD HEADQUARTERS ROSEMOUNT ANAL YTICAL EUROPE Emerson Process Management GmbH & Co. OHG
Industriestrasse 1 63594 Hasselroth Germany T 49 6055 884 0 F 49 6055 884209
Emerson Process Management Rosemount Analytical Inc.
6565 P Davis Industrial Parkway Solon, OH 44139 USA T 440.914.1261 Toll Free in US and Canada 800.433.6076 F 440.914.1271 e-mail: gas.csc@EmersonProcess.com
www.raihome.com
GAS CHROMAT OGRAPHY CENTER AND LATIN AMERICA
Emerson Process Management Rosemount Analytical Inc.
11100 Brittmoore Park Drive Houston, TX 77041 T 713 467 6000 F 713 827 3329
HAS60E-IM-HW
06/2007
EUROPE, MIDDLE EAST AND AFRICA Emerson Process Management
Shared Services Limited
Heath Place Bognor Regis West Sussex PO22 9SH England T 44 1243 863121 F 44 1243 845354
ASIA-PACIFIC Emerson Process Management
Asia Pacific Private Limited
1 Pandan Crescent Singapore 128461 Republic of Singapore T 65 6 777 8211 F 65 6 777 0947 e-mail: analytical@ap.emersonprocess.com
© Emerson Process Management GmbH & Co. OHG 2007
Loading...