Emerson CAT 100, CAT 200 User Manual

Instruction Manual
HAS9xE-IM-ATEX
07/2005
Continuous Analyzer Transmitter
CAT 100 CAT 200
www.EmersonProcess.com
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson Process
Management (Rosemount Analytical) representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson Process Management (Rosemount Analytical). Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice. 2nd edition 07/2005 1st edition 07/2004
Emerson Process Management Manufacturing GmbH & Co. OHG
Industriestrasse 1 D-63594 Hasselroth Germany T +49 (0) 6055 884-0 F +49 (0) 6055 884-209 Internet: www.EmersonProcess.com
Instruction Manual
HAS9xE-IM-A TEX 07/2005
This instruction manual provides information about installing, operating and maintaining/servicing CA T series gas analyzers in hazardous (classified) areas and shall be read in conjunction with the standard analyzer instruction manual only!
This instruction manual covers all analyzer variations and and therefore may describe configurations and/or options not part of your specific instrument.
The following definitions apply to WARNINGS, CAUTIONS and NOTES found throughout this publication.
CA T

PREFACE

DEFINITIONS

Highlights an operation or maintenance procedure, practice, condition, state­ment, etc.
If not strictly observed, could result in injury , death, or long­term health hazards of personnel.
NOTE
Highlights an essential operating procedure, condition or statement.
Further graphical symbols may be used within this manual:
Elektrostatic discharge (ESD)
Explosion Hazard!
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc.
If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
Heavy Instrument!
Harmful (to Health)!
All graphical symbols used in this product are from one or more of the following standards: EN61010-1, IEC417, and ISO3864.
Emerson Process Management Manufacturing GmbH & Co. OHG P-1
Toxic!
Disconnect from Mains!
Instruction Manual
CA T

Safety Instructions

HAS9xE-IM-A TEX
07/2005
SAFETY INSTRUCTIONS
INTENDED USE ST ATEMENT
CA T series gas analyzers are intended to be used as analyzers for industrial purposes. They must not be used in medical, diagnostic or life support applications, and no independent agency certifications or approvals are to be implied as covering such applications!
SAFETY SUMMARY
If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.
AUTHORIZED PERSONNEL
T o avoid explosions, loss of life, personal injury and damage to this equipment and on-site property , do not install, operate or service this instrument before reading and understanding this instruction manual and receiving appropriate training. Save these instructions.
ADDITIONAL LITERA TURE
This manual covers aspects specific for using CAT analyzers in hazardous (classified) areas, only .
To install, start-up, operate and maintain/service the instrument in a safe manner it is MANDAT O RY to read all additional instruction manuals shipped together with the instrument!
The following instruction manuals are available and/or referenced within this manual at hand:
CAT 100:
HAS1xE-IM-HS Instruction manual BINOS 100 series
CAT 200:
HAS3xE-IM-HW Hardware instruction manual NGA ML T series HAS3xE-IM-SW39 Instruction NGA ML T software V 3.9.x
(Software instruction manual depends on currently installed analyzer softwareof the internal NGA M LT module)
Contact your local service center or sales office when missing documents. Save all instructions for future use!
Emerson Process Management Manufacturing GmbH & Co. OHGP-2
Instruction Manual
HAS9xE-IM-A TEX 07/2005
Any addition, substitution, or replacement of components installed on or in this device, must be certified to meet the hazardous area classi­fication that the device was certified to prior to any such component addition, substitution, or replacement. In addition, the installation of such device or devices must meet the requirements specified and defined by the hazardous area classification of the unmodified device. Any modifications to the device not meeting these requirements, will void the product certification(s).
CA T
Safety Instructions
EXPLOSION HAZARD BY MODIFICATION
PARTS INTEGRITY AND UPGRADES
T ampering with or unauthorized substitution of components may adversely affect the safety of this instrument. Use only factory approved components for repair .
Because of the danger of introducing additional hazards, do not perform any unauthorized modification to this instrument.
Contact Rosemount Analytical Inc., Customer Service Center for Return Authorization.
POSSIBLE EXPLOSION HAZARD Do not open instrument when energized. Ensure that external circuitry is disconnected or de-energized before
opening the instrument. Ensure that all gas connections are made as labeled and are leak free.
Improper gas connections could result in explosion and death.
Emerson Process Management Manufacturing GmbH & Co. OHG P-3
CA T
Instruction Manual
HAS9xE-IM-A TEX
07/2005
Safety Instructions
POSSIBLE EXPLOSION HAZARD
The CAT analyzer may utilize not only sample gas but one or more pressurized carrier gases and/or calibration gases.
If an internal flowmeter is not used, an external flowmeter is required for flow control. Legislative requirements and instructions for installation in hazardous (classified) areas must be considered.
POSSIBLE EXPLOSION HAZARD
When installing these instruments all legislative requirements must be taken into consideration! Take care of the special conditions given within the instruction drawing in the manual on hand!
HOW TO ST A Y IN COMPLIANCE WITH THE
EUROPEAN DIRECTIVE 94/9/EC („A TEX“) WHEN PERFORMING
GAS ANAL YSIS WITHIN A FLAMEPROOF ENCLOSURE.
Special conditions apply to using a flameproof enclosure analyzer under the scope of the "European Directive for Equipment used in Explosive Atmosphere" (Directive 94/9/EC; A TEX). To stay compliant to the directive please consider the following clarification sheet released by the European ATEX Notified Body Group:
Emerson Process Management Manufacturing GmbH & Co. OHGP-4
Instruction Manual
HAS9xE-IM-A TEX 07/2005
CA T
Safety Instructions
Emerson Process Management Manufacturing GmbH & Co. OHG P-5
Instruction Manual
CA T
Compliances
HAS9xE-IM-A TEX
07/2005
Compliances
This product is certified by several agencies for the use in hazardous (classified) areas. The following certification marks apper on the product‘s nameplate label:
Zone classification: USA Class I Zone 1 AEx d e m IIB + H2 T4X Canada Ex d e m IIB + H2 T4X European Union A TEX, category 2, Zone 1, EEx d e m IIB + H2 T4X
USA/Canada
Certified by the Canadian Sta ndards Association, an „OSHA Nationally Recognized Testing Labor­tory“ (NRTL) for USA and Kanada
European Union (EU)
Conforms with the provisions of the EMC Directive 89/336/EEC, Potentially Explosive Atmospheres Directive 94/9/EC and CE Directive 93/68/EEC.
EC ATEX Type Examination Certificate: LCIE 00 ATEX 6009 X. EMERSON Process Management submitted samples for type examination of conformance to a independent Notified Body (Laboratoire Central des Industries Electriques; LCIE), which issued a type examination certificate.
Analyzer markings:
The appendix contains a copy of the type examination certificate.
Emerson Process Management Manufacturing GmbH & Co. OHGP-6
Instruction Manual
HAS9xE-IM-A TEX 07/2005
CA T
Table of Content s
PREFACE ...............................................................................................................P-1
DEFINITIONS ...............................................................................................................P-1
SAFETY INSTRUCTIONS..............................................................................................P-2
CHAPTER 1 INST ALLA TION....................................................................................... 1-1
Fig. 1: Installation drawing and analyzer dimensions, inch (mm) ...........................................................1-2
CHAPTER 2 ST ARTUP ............................................................................................... 2-1
CHAPTER 3 SERVICE AND MAINTENANCE.............................................................. 3-1
APPENDIX ...............................................................................................................A-1
A-1 EC Declaration of Conformity.................................................................................................A-2
A-2 EC Type Examination Certificates ..........................................................................................A-2
LIST OF ILLUSTRA TIONS
CHAPTER 1 INST ALLA TION
Fig. 1: Installation drawing and analyzer dimensions, inch (mm) ...........................................................1-2
Emerson Process Management Manufacturing GmbH & Co OHG TOC-1
CA T
Instruction Manual
HAS9xE-IM-A TEX
07/2005
Emerson Process Management Manufacturing GmbH & Co OHGTOC-2
Instruction Manual
HAS9xE-IM-A TEX
07/2005
Installing and wiring this instrument must comply with all relevant national legislative requirements and regulations.
Consider all safety instructions within this on hand manual and all associa­ted analyzer instruction manuals!
Take care of the special instructions at the installation drawing at the next page!
CA T
Chapter 1
Installation
POSSIBLE EXPLOSION HAZARD
POSSIBLE EXPLOSION HAZARD
Installing this instrument requires opening the enclosure and working at the open instrument. This is permitted only when both no hazardous atmos­phere is present and the instrument and connected external circuitry are de-energized!
Depending on the local regulation this may require a competent hot work supervisor to ssue a hot work permit.
More detailled instructions about installing these instruments are given by the associated analyzer instruction manual.
Emerson Process Management Manufacturing GmbH & Co. OHG 1-1
CA T
Instruction Manual
HAS9xE-IM-A TEX
07/2005
1 Installation
3 signal cable
glands
4 installation
drill-holes
A . EEx e junction box B. Threads for cable glands for analog and digital
signal cables (M16 x 1,5)
C. 11 gas fittings (number depending on appli-
cation; flame arrestors inside gas inlets and outlets).
D. Optional sample handling system, design
depending on applicational needs.
Note: Junction box must be protected by fuse supply, which has a breaking capacity adjusted to the short circuit of the equipment (10 amps / 250 VAC).
Fig. 1: Installation drawing and
analyzer dimensions, inch (mm)
Emerson Process Management Manufacturing GmbH & Co. OHG1-2
Instruction Manual
HAS9xE-IM-A TEX 07/2005
Chapter 2
Once the CA T analyzer is installed correctly and in accordance to the instructions given in chapter „Installation“ it is prepared for startup.
POSSIBLE EXPLOSION HAZARD Ensure all covers, plugs and housing parts are in place and secured
properly before supplying mains and signal voltages!
CA T
Startup
Ensure all requirements given by the clarification sheet for performing gas analysis within a flameproof enclosure are considered BEFORE supplying gases (see page S-6)!
This sheet gives instructions, too, for the sequence of supply­ing gases during process and analyzer startup .
After all safety aspects are followed and checked the instrument may be powered and operated according the instructions given in the related instrument‘s instruction manual.
Emerson Process Management Manufacturing GmbH & Co. OHG 2-1
CA T
Instruction Manual
HAS9xE-IM-A TEX
07/2005
2 Startup
Emerson Process Management Manufacturing GmbH & Co. OHG2-2
Instruction Manual
HAS9xE-IM-A TEX
07/2005
Inspection, maintenance and service must be carried out considering all related standards (e.g. EN 60079-17 “Inspection and maintenance of electrical installations in hazardous areas (other than mines)”
Service or replacement of safety related components or requiring to open the instrument are permitted only if no hazardous atmosphere is present and both the instrument and connected circuitry are de-energized!
CA T
CHAPTER 3
SERVICE AND MAINTENANCE
POSSIBLE EXPLOSION HAZARD
POSSIBLE EXPLOSION HAZARD
Depending on the local regulation this may require a competent hot work supervisor to ssue a hot work permit.
POSSIBLE EXPLOSION HAZARD
After maintenance or replacement of parts concerning explosion protection an authority on explosion protection has to verify that the analyzer still meets the requirements for explosion protection before it is switched on again.
Parts essential for explosion protection must not be repaired, they have to be replaced if defective!
The authority has to issue a certificate for this and/or attach a test label to the equipment before startup after maintenance or replacement of parts.
Emerson Process Management Manufacturing GmbH & Co. OHG 3- 1
CA T
Instruction Manual
HAS9xE-IM-A TEX
07/2005
3 Service and Maintenance
FLAMMABLE GASES - POSSIBLE EXPLOSION HAZARD Leaks may cause explosion when measuring flammable gases! When measuring flammable gases it is recommended to perform a leak test
on all gas paths, connections and components before startup or applying power. Leak tests should be carried out on a 2 month‘s regular basis and after repair/maintenance.
See the analyzer instruction manual for instructions on how to carry out leak tests.
FLAMMABLE GASES - POSSIBLE EXPLOSION HAZARD When measuring flammable gases it is recommended to purge the system
with air or an inert gas, e.g. nitrogen, prior to opening the enclosure. Violation may cause an explotion!
More detailled instructions about servicing and maintaining CAT analyzers are subject of the associated analyzer instruction manuals.
Emerson Process Management Manufacturing GmbH & Co. OHG3-2
Instruction Manual
HAS9xE-IM-A TEX
07/2005

APPENDIX

This chapter contains series CA T EC declarations of conformity and EC type examination certificates.
CA T
Emerson Process Management Manufacturing GmbH & Co. OHG A- 1
CA T

A-1 EC Declaration of Conformity

A-1 EC Declaration of Conformity
Instruction Manual
HAS9xE-IM-A TEX
07/2005
Emerson Process Management Manufacturing GmbH & Co. OHGA-2
Instruction Manual
HAS9xE-IM-A TEX
07/2005

A-2 EC Type Examination Certificates

A-2 EC Type Examination Certificates
CA T
Emerson Process Management Manufacturing GmbH & Co. OHG A- 3
CA T
Instruction Manual
HAS9xE-IM-A TEX
07/2005
A-2 EC Type Examination Certificates
Emerson Process Management Manufacturing GmbH & Co. OHGA-4
Instruction Manual
HAS9xE-IM-A TEX
07/2005
CA T
A-2 EC Type Examination Certificates
Emerson Process Management Manufacturing GmbH & Co. OHG A- 5
CA T
Instruction Manual
HAS9xE-IM-A TEX
07/2005
A-2 EC Type Examination Certificates
Emerson Process Management Manufacturing GmbH & Co. OHGA-6
Instruction Manual
HAS9xE-IM-A TEX
07/2005
CA T
A-2 EC Type Examination Certificates
Emerson Process Management Manufacturing GmbH & Co. OHG A- 7
CA T
Instruction Manual
HAS9xE-IM-A TEX
07/2005
A-2 EC Type Examination Certificates
Emerson Process Management Manufacturing GmbH & Co. OHGA-8
Instruction Manual
HAS9xE-IM-A TEX
07/2005
CA T
A-2 EC Type Examination Certificates
Emerson Process Management Manufacturing GmbH & Co. OHG A- 9
CA T
Instruction Manual
HAS9xE-IM-A TEX
07/2005
A-2 EC Type Examination Certificates
Emerson Process Management Manufacturing GmbH & Co. OHGA-10
CA T
Instruction Manual
HAS9xE-IM-A TEX
07/2005
Instruction Manual
HAS9xE-IM-A TEX
07/2005
EUROPE
Emerson Process Management GmbH & Co. OHG
Industriestrasse 1 63594 Hasselroth Germany T +49 (6055) 884-0 F +49 (6055) 884-209 Internet: www.emersonprocess.de
EUROPE, MIDDLE EAST , AFRICA Emerson Process Management Shared Services Limited
Heath Place Bognor Regis West Sussex PO22 9SH England T +44-1243-863121 F +44-1243-845354 Internet: www.emersonprocess.co.uk
CA T
NORTH AMERICA Emerson Process Management Rosemount Analytical Inc.
6565P Davis Industrial Parkway Solon, OH 44139-3922 USA T +1 (440) 914-1261 F +1 (440) 914-1271 Internet: www.emersonprocess.com
LATIN AMERICA Emerson Process Management Rosemount Analytical Inc. Latin America Headquarters