Emerson 3414, 3412, 3411 User Manual

Part Number 3-9000-759Revision D
DanielTM 3410 Series Gas Ultrasonic Flow Meters
Installation Manual
June 2014

Daniel customer service

Location Tel ephon e number Fax number
North America/Latin America +1.713.467.6000 +1.713.827.4805

Daniel Customer Service +1.713.827.6413 +1.713.827.6312

USA (toll free) +1.888.356.9001 +1.713.827.3380
Asia Pacific (Republic of Singapore) +65.6777.8211 +65.6777.0947.0743
Europe (Stirling Scotland, UK) +44 (0)1786.433400 +44 (0)1786.433401
Middle East Africa (Dubai, UAE) +971 4 8118100 +971 4 8865465
E-mail
Customer Service: tech.service@emersonprocess.com
Customer Support: daniel.cst.support@emerson.com
Asia-Pacific: danielap.support@emerson.com
Europe: DanielEMA.CST@EmersonProcess.com
Return Material Authorization (RMA)
A Return Material Authorization (RMA) number must be obtained prior to returning any equipment for any reason. Download the RMA form from the Support Services web page by selecting the link below.
www2.emersonprocess.com/EN-US/BRANDS/DANIEL/SUPPORT-SERVICES/Pages/Support-Services.aspx?

Signal words and symbols

This is a safety alert symbol. It is used to alert you to potential physical injury hazards. Obey all safety messages that follow this symbol to avoid possible injury or death.
Safety alert symbol
Danger indicates a hazardous situation which, if not avoided, will result in death or serious injury.
Warning indicates a hazardous situation which, if not avoided, could result in death or serious injury.
Caution indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
Caution indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
Pay special attention to the following signal words, safety alert symbols and statements:

Important

Important is a statement the user needs to know and consider.
Tip
Tip provides information or suggestions for improved efficiency or best results.

Note

Note is a “general by-the-way” content not essential to the main flow of information.

Important safety instructions

Installing, operating or maintaining a Daniel product improperly could lead to serious injury or death from explosion or exposure to dangerous substances. To reduce this risk:
Comply with all information on the product, in this manual, and in any local and national
codes that apply to the product.
Do not allow untrained personnel to work with this product.
Use Daniel parts and work procedures specified in this manual.
Daniel Measurement and Control, Inc. (Daniel) designs, manufactures and tests products to function within specific conditions. Because these products are sophisticated technical instruments, it is important that the owner and operation personnel strictly adhere both to the information printed on the product and to all instructions provided in this manual prior to installation, operation, and maintenance.
Daniel also urges you to integrate this manual into your training and safety program.
BE SURE ALL PERSONNEL READ AND FOLLOW THE INSTRUCTIONS IN THIS MANUAL AND ALL NOTICES AND PRODUCT WARNINGS.
Product owners (Purchasers):
Use the correct product for the environment and pressures present. See technical data
or product specifications for limitations. If you are unsure, discuss your needs with your Daniel representative.
Inform and train all personnel in the proper installation, operation, and maintenance of
this product.
To ensure safe and proper performance, only informed and trained personnel should
install, operate, repair and maintain this product.
Verify that this is the correct instruction manual for your Daniel product. If this is not
the correct documentation, contact Daniel at 1-713-827-6314. You may also download the correct manual from:
http://www.daniel.com
Save this instruction manual for future reference.
If you resell or transfer this product, it is your responsibility to forward this instruction
manual along with the product to the new owner or transferee.
ALWAYS READ AND FOLLOW THE INSTALLATION, OPERATIONS, MAINTENANCE AND
TROUBLESHOOTING MANUALS AND ALL PRODUCT WARNINGS AND INSTRUCTIONS.
Do not use this equipment for any purpose other than its intended service. This may
result in property damage and/or serious personal injury or death.
Product Operation Personnel:
To prevent personal injury, personnel must follow all instructions of this manual prior to
and during operation of the product.
Follow all warnings, cautions, and notices marked on, and supplied with, this product.
Verify that this is the correct instruction manual for your Daniel product. If this is not
the correct documentation, contact Daniel at 1-713-827-6314. You may also download the correct manual from:
http://www.daniel.com
Read and understand all instructions and operating procedures for this product.
If you do not understand an instruction, or do not feel comfortable following the
instructions, contact your Daniel representative for clarification or assistance.
Install this product as specified in the INSTALLATION section of this manual per
applicable local and national codes.
Follow all instructions during the installation, operation, and maintenance of this
product.
Connect the product to the appropriate pressure and electrical sources when and
where applicable.
Ensure that all connections to pressure and electrical sources are secure prior to and
during equipment operation.
Use only replacement parts specified by Daniel. Unauthorized parts and procedures can
affect this product's performance, safety, and invalidate the warranty. “Look-a-like” substitutions may result in deadly fire, explosion, release of toxic substances or improper operation.
Save this instruction manual for future reference.

Notice

THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. ALL SALES ARE GOVERNED BY DANIEL'S TERMS AND CONDITIONS, WHICH ARE AVAILABLE UPON REQUEST. WE RESERVE THE RIGHT TO MODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.
DANIEL DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND MAINTENANCE OF ANY DANIEL PRODUCT REMAINS SOLELY WITH THE PURCHASER AND END-USER.
TO THE BEST OF DANIEL'S KNOWLEDGE THE INFORMATION HEREIN IS COMPLETE AND ACCURATE. DANIEL MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABIL­ITY AND FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL DANIEL BE LIABLE FOR ANY INCIDENTAL, PUNITIVE, SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, LOSS OF REVENUE OR USE AND COSTS INCURRED INCLUDING WITHOUT LIMITATION FOR CAPITAL, FUEL AND POWER, AND CLAIMS OF THIRD PARTIES.
PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY BE TRADEMARKS/REGISTERED TRADEMARKS OF THESE COMPANIES

Warranty and Limitations

1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein, Daniel Measurement & Control, Inc. (“Daniel”) warrants that the licensed firmware embodied in the Goods will execute the programming instructions provided by Daniel, and that the Goods manufactured by Daniel will be free from defects in materials or workmanship under normal use and care and Services will be performed by trained personnel using proper equipment and instrumentation for the particular Service provided. The foregoing warranties will apply until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Daniel, whichever period expires first. Consumables and Services are warranted for a period of 90 days from the date of shipment or completion of the Services. Products purchased by Daniel from a third party for resale to Buyer ("Resale Products”) shall carry only the warranty extended by the original manufacturer. Buyer agrees that Daniel has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products. If Buyer discovers any warranty defects and notifies Daniel thereof in writing during the applicable warranty period, Daniel shall, at its option, correct any errors that are found by Daniel in the firmware or Services or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Daniel to be defective, or refund the purchase price of the defective portion of the Goods/Services. All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources or environmental conditions, accident, misuse, improper installation, modification, repair, use of unauthorized replacement parts, storage or handling, or any other cause not the fault of Daniel are not covered by this limited warranty, and shall be at Buyer's expense. Daniel shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by Daniel. All costs of dismantling, reinstallation and freight and the time and expenses of Daniel's personnel and representatives for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Daniel. Goods repaired and parts replaced by Daniel during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Daniel and can be amended only in a writing signed by Daniel. THE WARRANTIES AND REMEDIES SET FORTH ABOVE ARE EXCLUSIVE. THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. Buyer acknowledges and agrees that corrosion or erosion of materials is not covered by this warranty.
2. LIMITATION OF REMEDY AND LIABILITY PERFORMANCE. THE REMEDIES OF BUYER SET FORTH IN THIS AGREEMENT ARE EXCLUSIVE. IN NO EVENT, REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (WHETHER BASED IN CONTRACT, INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERWISE), SHALL DANIEL'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED OR SERVICES PROVIDED BY DANIEL GIVING RISE TO THE CLAIM OR CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL DANIEL'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAMAGES. THE TERM “CONSEQUENTIAL DAMAGES” SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF ANTICIPATED PROFITS, REVENUE OR USE AND COSTS INCURRED INCLUDING WITHOUT LIMITATION FOR CAPITAL, FUEL AND POWER, AND CLAIMS OF BUYER'S CUSTOMERS.
: DANIEL SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN
Daniel Series 3410 Gas Ultrasonic Meter Installation Manual PrefaceTable of Contents
3-9000-759 Rev D June 2014

Contents

Preface

Daniel customer service
Signal words and symbols
Important safety instructions
Notice
Warranty and Limitations
Section 1: Introduction
1.1 Typical applications ........................................................................................ 1
1.2 Features and benefits ...................................................................................... 2
1.3 Acronyms, abbreviations and definitions .......................................................... 3
1.4 Daniel MeterLink software ...............................................................................6
1.5 Daniel 3410 Series meter design ...................................................................... 7
1.6 Meter specifications .................................................................................... 12
1.7 Pre-installation considerations....................................................................... 18
1.8 Safety .......................................................................................................... 18
1.9 Daniel 3410 Series Certifications and Approvals .............................................. 20
1.10 FCC compliance .......................................................................................... 21
1.11 References.................................................................................................. 22
Section 2: Mechanical installation
2.1 Meter piping, lifting and mounting ................................................................ 23
2.2 Meter components........................................................................................ 25
2.3 Piping recommendations .............................................................................. 29
2.4 Meter safety for hoist rings and lifting slings................................................... 33
2.4.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges ..... 34
2.4.2 Appropriately rated lifting slings ........................................................................ 40
2.5 Mounting requirements in heated or cooled pipelines...................................... 42
Section 1: Introduction i
PrefaceTable of Contents Daniel Series 3410 Gas Ultrasonic Meter Installation Manual
June 2014 3-9000-759 Rev D
Section 3: Electrical installation
3.1 Cable length TTL mode ..................................................................................43
3.2 Cable length Open Collector mode .................................................................43
3.3 Grounding meter electronics housing ............................................................. 44
3.4 Conduit seals ................................................................................................45
3.4.1 Startup for systems using explosion-proof conduit .............................................46
3.4.2 Startup for systems that use flame-proof cable ..................................................50
3.5 Wiring and I/O ..............................................................................................51
3.5.1 CPU Module labeling and LED indicators.............................................................52
3.6 I/O connections ............................................................................................57
3.6.1 Frequency/Digital outputs ..................................................................................59
3.6.2 Analog input settings..........................................................................................62
3.6.3 Analog output settings .......................................................................................63
3.6.4 Digital Input........................................................................................................63
3.6.5 DHCP server switch settings ...............................................................................63
3.6.6 Configuration protect switch settings.................................................................64
3.6.7 External power source connection and fuse ........................................................65
3.7 Security seal installation ................................................................................66
3.7.1 Seal Transmitter Electronics Enclosure................................................................66
3.7.2 Base Enclosure Security Seals..............................................................................68
3.7.3 Transducer assembly security seals.....................................................................70
3.7.4 Sealing the unit...................................................................................................71
Section 4: Configuration
4.1 Daniel MeterLink Setup..................................................................................73
4.2 Field Setup Wizard ........................................................................................74
4.3 Using AMS Device Manager to configure the meter ..........................................80
4.4 Using a Field Communicator to configure the meter ........................................ 95
4.5 Security seals for the meter (optional) ............................................................98
ii Section 3: Electrical installation
Daniel Series 3410 Gas Ultrasonic Meter Installation Manual PrefaceTable of Contents
3-9000-759 Rev D June 2014
Appendix A: Engineering drawings
A.1 3410 Series engineering drawings.................................................................................... 99
Appendix B: Open source licenses
B.1 GNU General Public License ............................................................................................ 102
B.2 GNU Lesser General Public License.................................................................................. 113
B.3 BSD Open Source License................................................................................................ 117
B.4 M.I.T License ................................................................................................................... 118
B.5 Zlib License ..................................................................................................................... 120
Appendix C: Index
C.1 Manual Index .................................................................................................................. 121
Section 4: Configuration iii
PrefaceTable of Contents Daniel Series 3410 Gas Ultrasonic Meter Installation Manual
June 2014 3-9000-759 Rev D
iv Section 4: Configuration
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual List of Tables
3-9000-759 Rev D June 2014

List of Tables

Table 1-1 Daniel 3410 Series Gas Ultrasonic Meter acronyms, abbreviations and definitions ............ 3
Table 1-3 Performance specifications.............................................................................................. 13
Table 1-2 Meter specifications......................................................................................................... 13
Table 1-4 Transducer specifications................................................................................................. 15
Table 2-1 Hoist ring part number lookup table ................................................................................ 38
Table 2-2 Hoist Ring Lookup Table for Daniel 3414 Gas Meters........................................................ 39
Table 2-3 Hoist Ring Lookup Table for Daniel 3411 or 3412 Gas Meters........................................... 39
Table 3-1 Configurations for open collector frequency outputs....................................................... 43
Table 3-2 CPU Module labeling and LED functions ........................................................................... 53
Table 3-3 Ethernet cable to PC communication............................................................................... 54
Table 3-4 Serial Port A parameters................................................................................................... 56
Table 3-5 Frequency/Digital Outputs possible configurations.......................................................... 61
Table 3-6 DHCP server switch settings............................................................................................. 63
Table 3-7 Configuration protect switch settings .............................................................................. 64
Table 4-1 Local display labels, descriptions and valid units............................................................... 77
Table B-1 Open source licences .............................................................................................101
List of Tables v
List of Tables Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
vi List of Tables
Daniel 3410 Series Gas Ultrasonic Meter Installation Manual List of Figures
3-9000-759 Rev D June 2014

List of Figures

Figure 1-1 Daniel MeterLink download and registration ................................................................ 6
Figure 1-2 Daniel 3414 Gas Ultrasonic Flow Meter design ............................................................. 8
Figure 1-3 Daniel 3412 Gas Ultrasonic Flow Meter design ............................................................. 9
Figure 1-4 Daniel 3411 Gas Ultrasonic Flow Meter design .......................................................... 10
Figure 1-5 Transmitter electronics enclosure with local display and glass endcap ....................... 10
Figure 2-1 Daniel 3414 Flow Meter assembly............................................................................... 26
Figure 2-2 Daniel 3412 Flow Meter assembly .............................................................................. 27
Figure 2-3 Daniel 3411 Flow Meter assembly .............................................................................. 28
Figure 2-4 Transmitter electronics enclosure with optional local display and glass endcap .......... 28
Figure 2-5 Piping recommendations uni-directional without flow conditioner ............................ 30
Figure 2-6 Piping recommendations Uni-directional with flow conditioner ................................. 31
Figure 2-7 Piping recommendations Bi-directional flow with flow conditioner............................. 31
Figure 2-8 Meter end flange with tapped flat-counterbore hole for hoist ring ............................. 34
Figure 2-9 Safety approved hoist ring and non-compliant eye bolt ............................................. 35
Figure 2-10 90 Degree angle between slings ................................................................................. 36
Figure 2-11 Incorrect sling attachment.......................................................................................... 37
Figure 2-12 Correct sling attachment ........................................................................................... 40
Figure 2-13 Incorrect sling attachment.......................................................................................... 41
Figure 3-1 Internal Transmitter Electronics Enclosure chassis ground .......................................... 44
Figure 3-2 External ground lug .................................................................................................... 45
Figure 3-3 Electronics field wiring - upper terminal block, switches, ground lug .........................47
Figure 3-4 Transmitter electronics field wiring lower terminal block .......................................... 48
Figure 3-5 CPU Module labeling and LED indicators ..................................................................... 52
Figure 3-6 PC to meter serial connection wiring .......................................................................... 57
Figure 3-7 CPU Module I/O connections ...................................................................................... 58
Figure 3-8 CPU Module - Frequency/Digital outputs common ground ........................................ 62
Figure 3-9 CPU Module power source connections ..................................................................... 65
Figure 3-10 Transmitter electronics enclosure security latch ......................................................... 66
Figure 3-11 Transmitter Electronics Enclosure security seals ......................................................... 67
List of Figures vii
List of Figures Daniel 3410 Series Gas Ultrasonic Meter Installation Manual
June 2014 3-9000-759 Rev D
Figure 3-12 Base Enclosure wire seal installation 68
Figure 3-13 Base Enclosure security seals 69
Figure 3-14 Transducer assembly security seal70
Figure 4-1 AMS Device Description search81
Figure 4-2 AMS file download complete 82
Figure 4-3 AMS Device Manager 83
Figure 4-4 AMS Device Manager - Overview 83
Figure 4-5 AMS Device Manager - Guided Setup 84
Figure 4-6 AMS Device Manager - Service Tools All Variables status indicators86
Figure 4-7 Display Meter K-Factors86
Figure 4-8 AMS Device Manager - Configure Manual Setup87
Figure 4-9 Gating configuration parameter Edge gated, active high 89
Figure 4-10 Gating configuration parameter Edge gated, active low89
Figure 4-11 Gating configuration parameter State gated, active high89
Figure 4-12 Gating configuration parameter State gated, active low89
Figure 4-13 Configure Flow Analysis Alert 90
Figure 4-14 AMS Device Manager - Service Tools Alerts 91
Figure 4-15 Configuration changes dialog91
Figure 4-16 AMS Device Manager - Service Tools 92
Figure 4-17 AMS Device Manager - Service Tools All Variables 93
Figure 4-18 AMS Device Manager - Service Tools Trends 94
Figure 4-19 3414 transmitter field wiring conduit entries 96
Figure 4-20 Field Communicator wiring diagram for the 3410 Series electronics 97
viii List of Figures
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
11

Section 1: Introduction

Daniel 3410 Series Ultrasonic Gas Flow Meters have various configurations that meet a broad
TM
range of customer requirements. Each meter comes fully assembled from Daniel Measurement and Control, Inc.
Refer to the following documents for additional details:
P/N 3-9000-763 Daniel MeterLink Software for Gas and Liquid Ultrasonic Flow Meters
Quick Start Manual
P/N 3-9000-761 HART® Field Device Specification Manual
P/N 3-9000-769 Daniel 3410 Series Ultrasonic Gas Flow Meter Maintenance and
Troubleshooting Manual
Daniel 3414 Gas Ultrasonic Flow Meter technology provides custody transfer gas measurement with the 3414 Model four-path (eight transducers) chordal design.
Model 3412 dual-path (four transducers) chordal design and Model 3411 single-path (two transducers) chordal design provides check metering, pipeline balance, storage measurement, production, or wet gas applications.
See Section 1.2 for advantages of Daniel 3410 Series Ultrasonic Gas Flow Meters.

1.1 Typical applications

Daniel 3410 Series Ultrasonic Gas Flow Meters have various configurations that meet a broad range of customer requirements. Each meter comes fully assembled from Daniel. The technology can be applied to custody transfer, allocation measurement, and check metering applications such as:
Custody transfer
Power plants
Large industrial users
Production
Underground storage sites
Offshore
Allocation measurement
Typical applications 1
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D

1.2 Features and benefits

Proven long term stability
Field proven reliability
No line obstruction
No pressure loss
No moving parts
Low maintenance
Bi-directional measurement
Extractable transducers
Extensive self diagnostics
Immediate alarm reporting
Continuous Flow Analysis
Abnormal profile Blockage Internal bore buildup Liquids present in the gas meter Reverse Flow Speed of Sound comparison error
Auto-detected ASCII/RTU Modbus communications protocol
Low power consumption
Sophisticated noise reduction
Internet-ready communications
Ethernet access
On-board LED status indicators
Analog pressure and temperature inputs
TM
Communication via Emerson’s AMS
Device Manager and Field Communicator
API Chapter 21 compliant event and data logging (gas meters)
®
Daniel MeterLink (a Windows
-based interface software)
Local Display (optional)
For other features and benefits refer to the product datasheet:
http://www2.emersonprocess.com/EN-US/BRANDS/DANIEL/FLOW/Pages/Flow.aspx
2 Features and benefits
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014

1.3 Acronyms, abbreviations and definitions

Table 1-1 Daniel 3410 Series Gas Ultrasonic Meter acronyms, abbreviations and definitions
Acronym or abbreviation Definition
°degree (angle)
o
C
o
F
ADC analog-to-digital converter
AI Analog Input
AMS® Device Manager Asset Management Software - Device Manager
AO Analog Output
ASCII MODBUS A Modbus protocol message framing format in which ASCII characters are used to
boolean A type of data point that can only take on values of TRUE or FALSE (generally TRUE is
bps Bits Per Second (baud rate)
cPoise centipoise (viscosity unit)
CPU Central Processing Unit
CTS Clear-to-Send; the RS-232C handshaking signal input to a transmitter indicating that
DAC Digital-to-Analog Converter
Daniel MeterLink
DI Digital Input
DO Digital Output
DHCP Dynamic Host Configuration Protocol
dm
ECC Error Correction Code
EEPROM Electrically-Erasable, Programmable Read-Only Memory
Flash non-volatile, programmable read-only memory
FODO output that is user configurable as either a Frequency or Digital Output
HART® Communication Protocol Highway Addressable Remote Transducer communications protocol
hr hour (time unit)
Hz Hertz (cycles per second, frequency unit)
I/O Input/Output
TM
degrees celsius (temperature unit)
degrees fahrenheit (temperature unit)
delineate the beginning and end of the frame. ASCII stands for American Standard Code for Information Interchange.
represented by a value of 1, FALSE is represented by a value of 0)
it is okay to transmit data – i.e., the corresponding receiver is ready to receive data. Generally, the Request-to-Send (RTS) output from a receiver is input to the Clear-to­Send (CTS) input of a transmitter.
Daniel Ultrasonic Meter interface software
-1
decimeter (10
meters, length unit)
Acronyms, abbreviations and definitions 3
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
Table 1-1 Daniel 3410 Series Gas Ultrasonic Meter acronyms, abbreviations and definitions
Acronym or abbreviation Definition
IS Intrinsically Safe
K Kelvin (temperature unit)
kHz
LAN Local Area Network
LED Light-emitting Diode
m meter (length unit)
m3/d
3
m
/h
m3/s
mA milliamp (current unit)
MAC Address Media Access Control (Ethernet Hardware Address -EHA)
kilohertz (103 cycles per second, frequency unit)
cubic meters per day (volumetric flow rate)
cubic meters per hour (volumetric flow rate)
cubic meters per second (volumetric flow rate)
microinch (μinch)
micron
microinch (10-6 in)
micrometer (10
-6
m)
MMU Memory Management Unit
MPa
megapascal (equivalent to 106 Pascal) (pressure unit)
N/A Not Applicable
Nm
3
/h
normal cubic meters per hour
NOVRAM Non-Volatile Random Access Memory
Pa Pascal, equivalent to 1 newton per square meter (pressure unit)
Pas Pascal Second (viscosity unit)
PC Personal Computer
PFC Peripheral Field Connection (Board)
P/N Part Number
PS Power Supply (board)
psi pounds per square inch (pressure unit)
psia pounds per square inch absolute (pressure unit)
psig pounds per square inch gage (pressure unit)
RRadius of meter
rad radian (angle)
RAM Random Access Memory
RTS Request-to-Send; the RS-232C handshaking signal output by a receiver when it is
ready to receive data
RTU MODBUS A Modbus protocol framing format in which elapsed time between received charac-
ters is used to separate messages. RTU stands for Remote Terminal Unit.
4 Acronyms, abbreviations and definitions
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Table 1-1 Daniel 3410 Series Gas Ultrasonic Meter acronyms, abbreviations and definitions
Acronym or abbreviation Definition
s second (time unit, metric)
SDRAM Synchronous Dynamic Random Access Memory
sec second (time unit, U.S. Customary)
TCP/IP Transmission Control Protocol/Internet Protocol
time_t seconds since Epoch (00:00:00 UTC Jan. 1, 1970) (time unit)
UDP User Datagram Protocol
U.L. Underwriters Laboratories, Inc. - product safety testing
and certification organization
V Volts (electric potential unit)
W Watts (power unit)
Acronyms, abbreviations and definitions 5
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D

1.4 Daniel MeterLink software

Daniel MeterLink configuring your meter, collecting logs and reports and monitoring the meter health and alarm statuses. Daniel MeterLink may be downloaded at no charge from:
Figure 1-1 Daniel MeterLink download and registration
TM
software has robust features for setting communications parameters,
http://www.daniel.com/um2.htm
6 Daniel MeterLink software
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014

Procedure

1. From the right panel under Quick Links, click the MeterLink Registration and Download link.
2. Click the Order Now button to complete the Online registration form.
3. Click Next to go to the order confirmation page.
4. Click Complete Order. You will receive a conformation e-mail with a hyperlink directing you to the download site. Click the link provided.
5. Click Save.
Refer to the Daniel MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual (P/N 3-9000-763) for installation instructions and setup for initial communications. You may download the manual from the Daniel MeterLink web page:
http://www2.emersonprocess.com/en-US/brands/daniel/Flow/ultrasonics/Pages/MeterLink.aspx

1.5 Daniel 3410 Series meter design

Daniel 3410 Series Gas Ultrasonic Flow Meters are designed to accurately measure products in applications where reliable performance is critical, by measuring the difference in signal transit time with and against the flow across one or more measurement path(s). A signal transmitted in the flow direction travels faster than one transmitted against the flow direction. Each measurement path is defined by a transducer pair in which each transducer alternately acts as transmitter and receiver. The meter uses transit time measurements and transducer location information to calculate the mean velocity.
Daniel 3410 Series meter design 7
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
A. Transmitter electronics enclosure (explosion-proof) Optional - Local Display with glass endcap (Figure 1-4) B. Base electronics enclosure (intrinsically safe) C. Meter body with transducer assemblies (T-11, T-12, T-21 or T-22) (intrinsically safe)
A.
B.
C.
June 2014 3-9000-759 Rev D
Computer simulations of various velocity profiles demonstrate that multiple measurement paths provide an optimum solution for measuring asymmetric flow.Daniel 3414 Gas Ultrasonic Flow meters utilize four cross-bore, parallel-plane measurement paths that offer a high degree of accuracy, repeatability, bi-directional measurement and superior low-flow capabilities without the compromises associated with conventional technologies. These features make the Daniel 3414 the best choice for custody transfer applications.
Figure 1-2 Daniel 3414 Gas Ultrasonic Flow Meter design
8 Daniel 3410 Series meter design
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
A. Transmitter electronics enclosure (explosion-proof) Optional - Local Display with glass endcap (Figure 1-4) B. Base electronics enclosure (intrinsically safe) C. Meter body with transducer assemblies (T-11, T-12, T-21 or T-22) (intrinsically safe)
A.
B.
C.
3-9000-759 Rev D June 2014
Daniel 3412 Gas Ultrasonic Flow meters utilize two-path in-line (four transducers) measurement paths and are designed to measure the difference in signal transit time with and against the flow across one or more measurement path(s). The two paths are configured at right angles to one another in a “bulls-eye” arrangement.
Figure 1-3 Daniel 3412 Gas Ultrasonic Flow Meter design
Daniel 3410 Series meter design 9
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
A. Transmitter electronics enclosure (explosion-proof) Optional - Local Display with glass endcap (Figure 1-4) B. Base electronics enclosure (intrinsically safe) C. Meter body with transducer assemblies (T-11, T-12, T-21 or T-22) (intrinsically safe)
A.
B.
C.
June 2014 3-9000-759 Rev D
Daniel 3411 Gas Ultrasonic Flow meters are single-path (two transducer) Gas Ultrasonic Flow Meter and is referred to as a bounce-path (as the signal is bounced off the meter body) or a centerline path (as it goes through the centerline of the meter body) meter. The bounce-path method simplifies construction of the meter and makes the meter less susceptible to interference from pipeline liquids.
Figure 1-4 Daniel 3411 Gas Ultrasonic Flow Meter design
The Daniel Gas Ultrasonic Flow Meter design is available with an optional glass endcap and a local display.
Figure 1-5 Transmitter electronics enclosure with local display and glass endcap
10 Daniel 3410 Series meter design
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
All Daniel ultrasonic flow meter’s U.L. safety listing is accomplished through the combination of an explosion-proof transmitter electronics enclosure that houses the CPU module, Power Supply board, I.S. Barrier board, Backplane board and optional LCD Display board.
Note: The optional LCD Display requires firmware v1.04 or later and Uboot version, January 31,
2013.
The Base Electronics Enclosure that houses the Acquisition Module. Intrinsically safe transducers and cable assemblies are designed for Class 1, Division 1, Groups C and D areas without need of further protection when installed in accordance with the field wiring diagram (refer to Daniel drawing DMC - 005324 (see Appendix A).
Daniel 3410 Series meter design 11
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
CONTENTS MAY BE UNDER PRESSURE
When the meter is under pressure, DO NOT attempt to remove or adjust the transducer holder.
Attempting to do so may release pressurized gases, resulting in serious injury or equipment damage
CONTENTS MAY BE HAZARDOUS
The meter must be fully depressurized and drained before attempting to remove the transducer holder. If gas or fluid begins to leak from the transducer holder, stop immediately and reinstall the holder.
Failure to do so may cause serious injury or equipment damage.
A. Transducer holder
A.
ESCAPING GASES OR LIQUIDS HAZARD
The purchaser of the meter is responsible for the selection of Daniel components/seals and materials compatible with the chemical properties of gas flow measurement.
Failure to select suitable meter components/seals may cause escaping gases or liquids, resulting in injury or equipment damage.
June 2014 3-9000-759 Rev D

1.6 Meter specifications

12 Meter specifications
Consult your Daniel Sales and Service representative to ensure you purchase the correct components and seals for your application.
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
Table 1-3 Performance specifications
Meter size 4” to 24” 30” 36”
Qmin(ft/s) 222
Qt (ft/s) 10 8.5 7.5
Qmax (ft/s) 100 85 75
3-9000-759 Rev D June 2014
Specifications for Daniel 3410 Series Gas Ultrasonic Flow Meters are below:
Table 1-2 Meter specifications
Daniel 3411, 3412 and 3414 meter specifications
Meter type Number of paths
3411 Daniel single path (two transducer) or center-line (bounce) design
3412 Daniel two path (four transducer) center-line (bounce) design
3414 Daniel four path (eight transducer) chordal design
Ultrasonic type
Transit-time based measurement
Spool piece with integral mount transducers
Enclosure materials ASTM B26 Gr A356.0 T6 Aluminum
— 100% conversion coated and exterior coated with a polyurethane enamel
ASTM A351 Gr CF8M Stainless Steel —Passivated
Optional Local Display with a glass endcap on transmitter enclosure
Meter Performance
Linearity
Model 3414 Four-path meter chordal design± 0.3% of measured value over a 100:1 turndown 3-100 ft/s; 0.3 to 30 m/s) including lab
uncertainty
Flow calibrated accuracy is ± 0.1% of reading relative to lab over entire flow calibration
range (Q
min
- Q
max
)
Model(s) 3411 single-path or 3412 two-pathFlow calibrated accuracy is ± 0.5% of reading relative to labAccuracy is typically ±1.5% of actual volume flow
1. Does not take into consideration changes in wall roughness and installation effects.
Repeatability
Velocit y range
±0.05% of reading in the specified velocity range from 5% to 100% (Q
100 ft/s (30 m/s) with over-range)
1
(without flow calibration)
1
)
max
125 fps (38 m/s) on some line sizes
Meter meets or exceeds AGA9 (2007) performance specifications
Body and Flange Sizes and Pressure rating range
U.S. Customary Units - Meter sizes 4, 6, 8, 10, 12, 16, 18, 20, 24, 30, and 36 (inches)
ANSI pressure classes 300, 600, 900 and 1500 (per ANSI B16.5)
Carbon Steel
316 Stainless Steel
Metric Units - Meter sizes DN - 100, 150, 200, 250, 300, 400, 450, 500, 600, 700, 750, 900
PN 50, 100, 150, 200
Carbon Steel
316 Stainless Steel
Maximum Pressures
Dependent on operating temperature
Meter bore
Meter specifications 13
Schedule 20, 30, 40, 60, 80, 100, 120, 140, 160, STD, XS, LW
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
Table 1-2 Meter specifications
Meter Performance
Flange types ANSI classes - 300, 600, 900 and 1500 (per ANSI B16.5) Specific Gravity 0.35 to 1.50 Accuracy Limits Model 3414 accuracy limits (AGA 9 compliant) are:
± 1% without a flow calibration (10” and smaller line sizes)
± 0.7% without a flow calibration (for 12” and larger line sizes)
±0.1% with a flow calibration
Model(s) 3411 and 3412 accuracy limits are:
± 1.5% without a flow calibration
Minimum operating pressure
100 psig (7 bar)
Electronic specifications
Power Meter
10.4 VDC to 36 VDC
11 W power consumption (15 W maximum)
\
14 Meter specifications
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
Table 1-4 Transducer specifications
Transducer t ype Temperature range Mount and holder type
T-11 -20 °C to +100 °C (-4 °F to 212 °F) Standard mounts/Holders, NBR O-ring
Inconel mounts/316L Holders, NBR O-ring Inconel Mounts/Inconel Holders/FKM O-ring
T-12 -20 °C to +100 °C (-4 °F to 212 °F) Standard mounts/Holders, NBR O-ring
Inconel mounts/316L Holders, NBR O-ring Inconel Mounts/Inconel Holders/FKM O-ring
T-2 1
1
-20 °C to +100 °C (-4 °F to 212 °F) Standard mounts/Holders, NBR O-ring Inconel mounts/316L Holders, NBR O-ring Inconel Mounts/Inconel Holders/FKM O-ring
T-2 2
2
-50 °C to +100 °C (-58°F to 212 °F) Standard mounts/Holders, NBR O-ring Inconel mounts/316L Holders, NBR O-ring Inconel Mounts/Inconel Holders/FKM O-ring
1. T-21 transducers use W-01 transformers
2. T-22 transducers use W-02 transformers
3-9000-759 Rev D June 2014
Table 1-2 Meter specifications
Electronic specifications (Continued)
Transducer, mounts and holders
Note: The process temperature must not exceed the operating temperature range of the transducers.
Note: T-11 and T-21 transducers are designed for 12 inch and larger meters. T-12 and T-22 transducers are designed for 4"
through 10 meters.
Note: T-11 and T-21 transducers are used for all meter sizes for Models 3411 and 3412.
Note: The ultrasonic transducers are not intended for use across boundary walls of different hazardous area classifications. The
transmitter electronics cannot be remote mounted from Division 1 classification to a Division 2 area to meet an area classification.
Meter specifications 15
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
Table 1-2 Meter specifications
Communications specifications
Connectivity protocols One serial RS-232/RS-485 port (115 kbps baud rate) (Modbus RTU/ASCII)
(1) Serial Port A
(RS-232/RS-485 Full Duplex/RS-485 Half Duplex)
One Ethernet Port (TCP/IP) 100 Base
Up to 10 Mbps (internal connection) 100Mbps (external connection)
Modbus TCP, TCP/IP
Device compatibility Daniel Ultrasonic flow meters are compatible with nearly every commercially available flow
computer. Examples: FloBoss 103, FloBoss S600 flow computer, ROC 107,
Digital, analog, and frequency inputs
Digital Input(s) (Selectable)
Analog Input(s) (2) 4-20 mA
(1) Single polarity
Four pulse configurations available
AI-1 Temperature
AI-2 Pressure
Note: The analog-to-digital conversion accuracy is within ±0.05% of full scale over the operating
temperature range.
Note: AI-1 and AI-2 are electronically isolated and operate in sink mode. The input contains a series
resistance so HART® Communicators can be connected to configure sensors.
A 24 Volt DC power output is available to provide power to the sensors.
16 Meter specifications
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Table 1-2 Meter specifications
Digital, analog, and frequency outputs
Frequenc y/Digital Output(s)
The meter has user-configurable selections for either a frequency output or digital status (FODO) (Also
see Section 3.6.1).
(3) Frequency/Digital Outputs
FODO1 (four possible output configurations)
FODO2(eight possible output configurations)
FODO3(eight possible output configurations)
Frequency or Digital Output parameter pairs ( Frequency or Digital Outputs (FODO 1) source selections:
see Section 3.6.1)
(FO1A, DO1A, FO1B, DO1B)
Frequency or Digital Outputs (FODO 2) source selections
(FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B)
Frequency or Digital Outputs (FODO 3) source selections
(FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B)
Mode options:
Open Collector (requires external excitation supply voltage and pull-up resistor)
TTL (internally powered by the meter 0-5 VDC signal)
Channel B Phase options:
Lag forward, Lead reverse (Phase B lags Phase A while reporting forward flow, leads Phase A
while reporting reverse flow)
Lead forward, Lag reverse (Phase B leads Phase A while reporting forward flow, lags Phase A
while reporting reverse flow)
Phase A and Phase B output (based on flow direction)
Reverse flow - output only reports flow in the reverse direction. For frequency outputs, Phase B
of the output is 90 degrees out of phase with Phase A.
Forward flow - output only reports flow in the forward direction. For frequency outputs, Phase
B of the output is 90 degrees out of phase with Phase A.
Absolute - output reports flow in both directions. For frequency outputs, Phase B of the output
is 90 degrees out of phase with Phase A.
Bidirectional - output reports flow on Phase A only in the forward direction and on Phase B
only in the reverse direction.
Maximum frequency for the frequency outputs
1000Hz
5000Hz
Analog Output(s)
Meter specifications 17
(1) 4-20 mA independently configurable analog output (HART)
(1) 4-20 mA independently configurable analog output (conventional)
The analog output zero scale offset error is within ±0.1% of full scale and gain error is within ±0.2% of full scale. The total output drift is within ±50 ppm of full scale per °C.
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
An “X” signifies the user should contact Daniel Measurement and Control, Inc. for information on the dimensions of the flameproof joints.
June 2014 3-9000-759 Rev D

1.7 Pre-installation considerations

Pipeline equipment code compliance, ANSI, ASME, etc.
Proper Inlet/outlet meter tube piping for reasonable stable flow to the settling chamber
(first meter tube spool upstream of the meter).
Electrical safety compliance; UL, CSA, ATEX, IECEx etc.
Civil and structural good practices compliance
Contractual agreements or governmental compliance (or both)
In-situ performance test procedures
Field tested meter health check and flow dynamics diagnostics
Data collection and retention procedures

1.8 Safety

The Daniel 3410 Series Gas Ultrasonic Flow Meter is suitable for use in U.L. Class 1, Division 1, Group C and D hazardous locations.
Refer to the 3410 Series Systems Wiring Diagram, Sheet 3 (P/N DMC -005324) for the certification tag (see Appendix A).
18 Pre-installation considerations
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
EXPLOSION OR FIRE HAZARD
Conduit runs must have a sealing fitting within 18 inches (457 mm) of the enclosure to reduce the risk of an explosion or a fire.
During operation, keep covers tight.
During equipment maintenance, disconnect power before opening transmitter or base
electronics. Clean cover joints before replacing.
DO NOT substitute meter components. Component substituting may compromise the intrinsic
safety.
Failure to do so may result in severe injury to personnel or cause damage to the equipment.
3-9000-759 Rev D June 2014
Daniel 3410 Series Liquid Ultrasonic Meters are INMETRO certified. Refer to the 3410 Series Liquid Ultrasonic Flow Meter Tag, INMETRO Certification drawing DMC - 006224.
Certificate number: NCC 11.0163 X Marking: --Ex d ia IIB T4 Gb IP66 W Electrical parameters: Refer to Section 1.6, Table 1-1
Special conditions for safe use
Explosion proof joint dimensions are compliant with the Brazilian Association of technical standard: ABNT NBR IEC 60079-1, Table 3.
The enclosure for the explosion proof transmitter and intrinsically safe barrier must be remote mounted (refer to Section 1.6, Table 1-2)
o
40
F (60 oC) (refer to Section 1.6, Table 1-2)
if the operating temperature exceeds 1
Cable length (refer to Section 1.6, Table 1-2)
Safety 19
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D

1.9 Daniel 3410 Series Certifications and Approvals

Daniel 3410 Series Gas Ultrasonic Flow Meters have electrical, metrology, intrinsic safety and Pressure Equipment Directive certifications and approvals by the agencies listed below. Refer to the nameplate tag on the meter body, the wiring diagram (P/N DMC - 005324) in Appendix A and observe all safety precautions. Daniel 3410 Series Gas Ultrasonic Flow Meters operate within the pressure and temperature range of the device (also see Section 1.6 for meter specifi­cations). Daniel 3410 Series Gas Ultrasonic Flow Meters are approved to the ATEX Directive 94/9/EC.

Standards

•US
Canada
•Europe
- Explosive Atmospheres (ATEX)
- International Electrotechnical Commission (IECEx)
- Pressure Equipment Directive (PED via BSI)
- Electromagnetic Compatibility (EMC)
- International Organization of Legal Metrology (OIML)

Approval Agencies

•UL
•ULC
•DEMKO
•INMETRO
•NEPSI
•GOSTR
IMPORTANT
Please consult Daniel Customer Service for the complete metrology approvals list.
20 Daniel 3410 Series Certifications and Approvals
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
3-9000-759 Rev D June 2014

1.10 FCC compliance

This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Par t 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
FCC compliance 21
Section 1: Introduction Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D

1.11 References

[1] Gould Modbus Protocol Reference Guide, Rev. B, PI-MBUS-300
[2] Measurement of Fuel Gas By Turbine Meters, American Gas Association, Transmission
Measurement Committee Report No. 7, Second Revision, April 1996 (also referred to as AGA7)
[3] Compressibility Factors of Natural Gas and Other Related Hydrocarbon Gases, American
Gas Association, Transmission Measurement Committee Report No. 8, Second Edition, Second Printing, July 1994 (also referred to as AGA8)
[4] Speed of Sound in Natural Gas and Other Related Hydrocarbon Gasses, Report 10, First
Edition, May 2003 (also referred to as AGA10)
[5] Manual of Petroleum Measurement Standards, Chapter 21 — Flow Measurement Using
Electronic Metering Systems, Section 1 — Electronic Gas Measurement, American Gas Association and American Petroleum Institute, First Edition, September 1993
[6] AGA Report No. 9, Measurement of Gas by Multipath Ultrasonic Meters, Second Edition
(April 2007)
22 References

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation

SURFACE TEMPERATURE HAZARD
The meter body and piping may be extremely hot or cold.
Wear appropriate personal protective equipment when coming in contact with the meter. Failure to do so may result in injury.
CUTTING HAZARD
Sharp edges may be present on the meter.
Wear appropriate personal protective equipment when working on the meter. Failure to do so may cause serious injury.
TRANSPORTATION HAZARD
When moving the meter, do not insert the forks of a forklift into the bore.
inserting the forks may cause the meter to become unstable, resulting in injury or damage to the bore and sealing face.
TRIPPING HAZARD
Clear all obstacles or obstructions from the work area when transporting, installing or removing the meter.
Failure to clear the work area may cause injury to personnel.
3-9000-759 Rev D June 2014
Section 2: Mechanical installation

2.1 Meter piping, lifting and mounting

Refer to the following sections for piping recommendations, lifting with hoist rings and slings, mounting in heated or cooled pipelines and safety warnings and precautions.
Meter piping, lifting and mounting 23
Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
CRUSHING HAZARD
Do not remove flange stabilizers.
Attempting to do so may allow the meter to roll, resulting in serious injury or equipment damage.
A.
A. Flange stabilizers
ESCAPING GASES OR LIQUIDS HAZARD
The purchaser of the meter is responsible for the selection of Daniel components/seals and materials compatible with the chemical properties of gas flow measurement.
Failure to select suitable meter components/seals may cause escaping gases or liquids, resulting in injury or equipment damage.
June 2014 3-9000-759 Rev D
Consult your Daniel Sales and Service representative to ensure you purchase the correct components and seals for your application.
24 Meter piping, lifting and mounting
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation
CONTENTS MAY BE UNDER PRESSURE
When the meter is under pressure, DO NOT attempt to remove or adjust the transducer holder.
Attempting to do so may release pressurized gases, resulting in serious injury or equipment damage
CONTENTS MAY BE HAZARDOUS
The meter must be fully depressurized and drained before attempting to remove the transducer holder. If gas or fluid begins to leak from the transducer holder, stop immediately and reinstall the holder.
Failure to do so may cause serious injury or equipment damage.
A. Transducer holder
A.
3-9000-759 Rev D June 2014

2.2 Meter components

Daniel 3410 Series Gas Ultrasonic Flow Meters are assembled, configured and tested at the factory. The meter components include the transmitter electronics enclosure, the base
1
electronics enclosure and the meter body with transducer assemblies.
1. Refer to the 3-9000-744 Split Clamp Extractor Tool Operation Manual to remove the transducer holders while the
meter is pressurized.
Meter components 25
Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
EXPLOSION OR FIRE HAZARD
Conduit runs must have a sealing fitting within 18 inches (457 mm) of the enclosure to reduce the risk of an explosion or a fire.
During operation, keep covers tight.
During equipment maintenance, disconnect power before opening transmitter or base electronics. Clean cover joints before replacing.
DO NOT substitute meter components. Component substituting may compromise the intrinsic safety.
Failure to do so may result in severe injury to personnel or cause damage to the equipment.
A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board
B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and transducer assemblies and cables
A.
B.
C.
D.
D. Flange stabilizers
Backplane board, (Optional: glass endcap for Local Display)
June 2014 3-9000-759 Rev D
The 3414 four path ultrasonic meter components are shown below.
Figure 2-1 Daniel 3414 Flow Meter assembly
26 Meter components
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation
A.
B.
C.
A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board
B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and transducer assemblies and cables D. Flange stabilizers
Backplane board. (Optional: glass endcap for Local Display)
D.
3-9000-759 Rev D June 2014
The 3412 dual path ultrasonic meter components are shown below.
Figure 2-2 Daniel 3412 Flow Meter assembly
Meter components 27
Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
A.
B.
C.
A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board
B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and transducer assemblies and cables D. Flange stabilizers
Backplane board. (Optional: glass endcap for Local Display)
D.
A.
B.
A. Transmitter electronics enclosure with glass endcap B. Local display
June 2014 3-9000-759 Rev D
The 3411 single path ultrasonic meter components are shown below.
Figure 2-3 Daniel 3411 Flow Meter assembly
Figure 2-4 Transmitter electronics enclosure with optional local display and glass endcap
28 Meter components
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation
BURST HAZARD
Before pipeline cleaning and maintenance ("pigging operations"), remove straightening vanes or flow conditioners. Failure to do so may cause excessive pressure in the meter system, resulting in serious injury/ death or equipment damage.
3410 Series Gas Ultrasonic Flow Meter with flow conditioner for uni-directional flow
3410 Series Gas Ultrasonic Flow Meter with flow conditioner for bi-directional flow
Flow conditioner: Daniel Profiler or CPA 50E Straightening device
Straightening device
Flow conditioner: Daniel Profiler or CPA 50E
SUNSHIELD PROTECTION
Install a sunshield to prevent prolonged exposure to direct sunlight in extreme climates.
Failure to shield the meter may result in exceeding the process temperature range and damage transmitter electronics.
3-9000-759 Rev D June 2014

2.3 Piping recommendations

Sunshields, provided by the customer, may be required to prevent exceeding the process fluid temperature when the meter is mounted in a location with extremely hot climates.
Piping recommendations 29
Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
For optimal flow measurement conditions, Daniel suggests the piping configurations below. Regardless of the configuration selected, the user agrees to accept full responsibility for the site piping design and installation.
June 2014 3-9000-759 Rev D
Flow conditioning is recommended for best measurement results
Honed or un-honed meter tube(s)
Flow direction (unidirectional or bidirectional)
Correct meter size selection - too low may cause poor flow stability (thermal
convection or too fast may cause erosion problems and resonance, cracks or failure of probes or thermowells (approximately 0.3 to 30 m/sec or 1 to 100 ft/sec).
Space availability for meter lengths (to allow inlet piping customization)
Concentric alignment pins or flange concentricity technique considerations
IMPORTANT
The bore of the mating piping should be within 1% of the meter inside diameter.
Figure 2-5 Piping recommendations uni-directional without flow conditioner
30 Piping recommendations
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation
Flow conditioner: Daniel Profiler or CPA 50E Straightening device
Straightening device
Flow conditioner: Daniel Profiler or CPA 50E
To access the product datasheet from the Daniel products page (above link), select the Daniel Gas Ultrasonic Flow Meter link, click the Documentation tab, expand the Data Sheets ­Bulletins - Catalogs tab, then select the Data Sheet.
3-9000-759 Rev D June 2014
Figure 2-6 Piping recommendations Uni-directional with flow conditioner
Figure 2-7 Piping recommendations Bi-directional flow with flow conditioner
All pipe lengths are minimum:
D = Nominal pipe size in inches (i.e. 6” pipe size; 10 D = 60 in)
P = Pressure measurement location
T = Temperature measurement location
http://www2.emersonprocess.com/en-US/brands/daniel/Flow/ultrasonics/Pages/Ultrasonic­Series-3400.aspx
The meter is provided with dowel pins to align the meter body bore with the bore of the
mating piping.
Piping recommendations 31
Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
FAULTY METER INSTALLATION
Correctly install the equipment.
If meter bodies are mounted or oriented differently than specified above, debris or gas may collect in the transducer ports which could adversely affect the transducer signals, or cause equipment damage.
June 2014 3-9000-759 Rev D
Daniel 3410 Series Ultrasonic Gas Flow Meters should be mounted in horizontal piping
with the chord paths horizontal
Normally, the meter body is installed so that the electronics assembly is on the top of
the meter. If there is insufficient space above the piping for this arrangement, the meter can be ordered with extra long transducer cables for remote mounting or the meter housing can be installed with the electronics assembly on the bottom.
The mating piping should include temperature measurement connections located a
minimum of three nominal pipe diameters length down stream of the meter, or per AGA Report No. 9.
32 Piping recommendations
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation
DANGER TO PERSONNEL AND EQUIPMENT
Lifting a Daniel Ultrasonic Meter with other equipment
The following lifting instructions are for installation and removal of the Daniel Ultrasonic Meter ONLY. The instructions below do not address lifting the Daniel ultrasonic meter while it is attached, bolted, or welded to meter tubes, piping, or other fittings.
Using these instructions to maneuver the Daniel Ultrasonic Meter while it is still attached, bolted, or welded to a meter tube, piping, or other fitting may result in equipment damage, serious injury, or death.
The operator must refer to their company's hoisting and rigging standards, or the "DOE-STD­1090-2004 Hoisting and Rigging" standard if such company standards do not exist, for lifting and maneuvering any assembled meter tube and associated piping.
CRUSHING HAZARD
During meter installation or removal, always place the unit on a stable platform or surface that supports its assembled weight.
Failure to do so could allow the meter to roll, resulting in serious injury or equipment damage.
Prior to lifting the unit, refer to the Daniel 3414, 3412, or 3411 Gas Ultrasonic Flow Meter nameplate or outline dimensional (general arrangement) drawing for the assembled weight.
3-9000-759 Rev D June 2014

2.4 Meter safety for hoist rings and lifting slings

A Daniel Gas Ultrasonic Flow Meter can be safely lifted and maneuvered into and out of a meter run for installation or service by obeying the following instructions.
Meter safety for hoist rings and lifting slings 33
Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
A. Plug Bolt
B. Flat Counterbore Surface
A.
B.
June 2014 3-9000-759 Rev D
When lifting a Daniel Ultrasonic Meter by itself, Daniel recommends two methods. These methods are:
Using appropriately rated Safety Engineered Swivel Hoist Rings installed in the Daniel
Ultrasonic Meter end flanges.
Using appropriately rated lifting slings positioned at designated areas of the Daniel
Ultrasonic Meter.
Both methods must be used in conjunction with all appropriate company hoisting and rigging standards or the DOE-STD-1090-2004 HOISTING AND RIGGING standards do not exist. Refer to the following sections for more information on these two methods.
standard if such company

2.4.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges

Daniel Ultrasonic meters come equipped with a tapped hole located on the top of each meter body end flange. A flat machined surface surrounds each tapped hole. This feature provides complete surface contact ONLY between the meter flange and an OSHA compliant Safety Engineered Swivel Hoist Ring as shown in Figure 2-9.
Operators SHALL NOT use Eye Bolts (see Figure 2-9) in the Daniel Ultrasonic Meter flange tapped holes to aid in lifting or maneuvering the unit.
Operators SHALL NOT use other Hoist Rings that do not fully seat flush with the counter bore on the top of the meter flanges.
Figure 2-8 Meter end flange with tapped flat-counterbore hole for hoist ring
34 Use of appropriate safety engineered swivel hoist rings in meter end flanges
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation
Eye bolt
Safety engineered swivel hoist ring
3-9000-759 Rev D June 2014
Figure 2-9 Safety approved hoist ring and non-compliant eye bolt
Safety precautions using safety engineered swivel hoist rings
Read and follow the Safety Precautions listed below:
1. Meters must only be lifted by personnel properly trained in the safe practices of rigging
and lifting.
2. Remove the plug bolts installed in the tapped holes on the top of the flanges. Do not
discard the bolts as they must be reinstalled once the lifting operation is complete to prevent corrosion of the tapped holes.
3. Make sure the tapped holes on the meter are clean and free of debris before installing
the hoist rings.
4. Use only the safety engineered swivel hoist rings that are rated for lifting the meter. Do
not use any other type of hoist rings with the same screw size or heavy duty hoist rings. The meter tapping and counter bore size are suitable only for the hoist rings specified by Daniel.
5. When installing a hoist ring, make sure the base surface of the hoist ring fully contacts
the machined flat surface of the tapped hole. If the two surfaces do not come in contact then the hoist ring will not hold its full rated load. Torque the hoist ring attachment bolts to the limit indicated on the hoist rings.
Use of appropriate safety engineered swivel hoist rings in meter end flanges 35
Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
6. After installation of the hoist rings, always check that the ring rotates and pivots freely
in all directions.
7. NEVER attempt to lift the meter using only one hoist ring.
8. Always use separate slings to each hoist ring. NEVER reeve one sling through both hoist
rings. The slings must be of equal length. Each sling must have a load rating that equals or exceeds the hoist ring load rating. The angle between the two slings going to the hoist rings must never exceed 90 degrees or the load rating of the hoist rings will be exceeded.
Figure 2-10 90 Degree angle between slings
36 Use of appropriate safety engineered swivel hoist rings in meter end flanges
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation
3-9000-759 Rev D June 2014
9. NEVER allow the slings to contact the electronics enclosure. Damage to the enclosure
may occur. Use a spreader bar with the slings to prevent contact with the electronics enclosure and the base enclosure (see Figure 2-12). If the slings do come in contact with the electronic enclosure then remove the two bolts holding the enclosure to its base and temporarily remove the head from the meter during the lifting operation. You will need to unplug the cable from J3 on the Acquisition Module. Two screws hold this cable in place.
Once the lifting operation is complete, reattach and secure the electronics cable to J3 on the Acquisition Module, return the electronics enclosure to its original position, replace the bolts, and secure the enclosure in place.
Lifting the meter with the upper enclosure installed but without the bolts installed, may cause the electronics to fall and cause personal injury or equipment damage.
Figure 2-11 Incorrect sling attachment
10. NEVER apply shock loads to the meter. Always lift the meter gradually. If shock loading
ever occurs, the hoist ring must be inspected per manufacturer's recommendations prior to any further service. If a proper inspection cannot be performed, discard the hoist ring.
11. NEVER lift with any device, such as hooks, chains, or cables that could create side pulls
that could damage the ring of the hoist ring.
Use of appropriate safety engineered swivel hoist rings in meter end flanges 37
Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
12. NEVER lift more than the ultrasonic meter assembly including electronics and
transducers with the hoist rings. The only exception that safe is to lift the meter with one ASME B16.5 or ASME B16.47 blind flange bolted to each end flange of the meter. NEVER use the hoist rings on the meter to lift other components such as meter tubes, piping or fittings attached to the meter. Doing so will exceed the load rating of the hoist rings.
13. Remove the hoist rings from the meter after lifting is completed and store them in an
appropriate case or container per their manufacturer's recommendation.
14. Apply heavy lubricant or anti-seize to the threads of the plug bolts and reinstall the plug
bolts to keep the tapped holes free of debris and to prevent corrosion.
How to obtain safety engineered swivel hoist rings
TA list of approved manufacturers of safety engineered hoist rings is below:
American Drill Bushing Company(
Carr Lane Manufacturing Company (
www.americandrillbushing.com)
www.carrlane.com)
Select an approved supplier from the list below. These vendors can supply the safety­engineered hoist rings. This is not intended to be a complete list.
Fastenal (
Reid Tools
www.fastenal.com)
(www.reidtool.com)
The appropriate hoist rings can also be purchased directly from Daniel. The following table provides part numbers for reference:
Table 2-1 Hoist ring part number lookup table
Daniel part number
1-504-90-091 3/8”-16UNC, 1000 lb. 23053 CL-1000-SHR-1
1-504-90-092 1/2”-13UNC, 2500 lb 23301 CL-23301-SHR-1
1-504-90-093 3/4”-10UNC, 5000 lb. 23007 CL-5000-SHR-1
1-504-90-094 1”-8UNC, 10000 lb. 23105 CL-10000-SHR-1
1-504-90-095 1-1/2”-6UNC, 24000 lb. 23202 CL-24000-SHR-1
1. Note: The part numbers include only one hoist ring. Two hoist rings are required per meter.
1
Hoist ring thread size & load rating
1
American Drill Bushing Co. P/N
1
Carr Lane Manufacturing Co. P/N
1
38 Use of appropriate safety engineered swivel hoist rings in meter end flanges
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation
3-9000-759 Rev D June 2014
What size safety engineered swivel hoist ring do you need?
To determine the size of the hoist rings required for your meter, use the appropriate table below. Look down the column that matches the ANSI rating of your meter. Find the row that contains
your meter size. Follow the row to the end to find the appropriate hoist ring part number.
Table 2-2 Hoist Ring Lookup Table for Daniel 3414 Gas Meters
1
ANSI 300 ANSI 600 ANSI 900 ANSI 1500 Daniel Part Number
4” to 10” 4” to 8” 4” to 8” 4” to 6” 1-504-90-091
12” to 18” 10” to 16” 10” to 12” 8” to 10” 1-504-90-092
20” to 24” 18” to 20” 16” to 20” 12” 1-504-90-093
30” to 36” 24” to 30” 24” 16” to 20” 1-504-90-094
36” 30" to 36” 24" to 36” 1-504-90-095
1. 4” to 6” 45 degree meters and 8” to 36” 60 degree meters.
Table 2-3 Hoist Ring Lookup Table for Daniel 3411 or 3412 Gas Meters
ANSI 300 ANSI 600 ANSI 900 ANSI 1500 Daniel Part Number
4” to 10” 4” to 8” 4” to 8” 4” to 6” 1-504-90-091
12” to 18” 10” to 16” 10” to 12” 8” to 10” 1-504-90-092
20” to 24” 18” to 20” 16” to 20” 12” 1-504-90-093
30” to 36” 24” to 30” 24” 16” to 20” 1-504-90-094
36” 30" to 36” 24" to 36” 1-504-90-095
Use of appropriate safety engineered swivel hoist rings in meter end flanges 39
Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D

2.4.2 Appropriately rated lifting slings

The following instructions are intended to provide general guidelines for using proper lifting slings when lifting a Daniel 3410 Series Gas Ultrasonic Flow Meter by itself. These instructions are intended to be followed in addition to your company's standards or the DOE-STD-1090­2004 Hoisting and Rigging standard if such company standards do not exist.
Safety precautions using appropriate rated lifting slings
1. Meters must only be lifted by personnel properly trained in the safe practices of rigging
and lifting.
2. NEVER attempt to lift the meter by wrapping slings around the electronics enclosure.
3. NEVER attempt to lift the meter using only one sling around the meter. Always use two
slings wrapped around each end of the body as shown below. A choker style sling is recommended.
Figure 2-12 Correct sling attachment
40 Appropriately rated lifting slings
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation
3-9000-759 Rev D June 2014
4. Visually inspect the slings prior to use for any signs of abrasion or other damage. Refer
to the sling manufacturer's procedures for proper inspection of the particular sling you are using.
5. Only use slings with ratings that exceed the weight to be lifted. Reference your
company's standards for safety factors that must be included when calculating the load rating.
6. NEVER allow the slings to contact the electronics enclosure or the transducer cabling.
Damage to the meter may occur. If the slings do come in contact with the electronics enclosure, then remove the two bolts holding the enclosure to its base and temporarily remove the head from the meter during the lifting operation (Remove the two bolts holding the enclosure to its base and unplug the cable from the Acquisition Module. Two screws hold this cable in place.) Use a spreader-bar on the slings to prevent contact with the electronics.
7. Once the lifting operation is complete, reattach and secure the electronics cable to J3
on the Acquisition Module, return the electronics enclosure to its original position, replace the bolts, and secure the enclosure in place. Lifting the meter with the upper enclosure installed but with out the bolts installed, may cause the electronics to fall and cause personal injury or electronics damage.
Figure 2-13 Incorrect sling attachment
8. NEVER apply shock loads to the meter. Always lift the meter gradually. If shock loading
ever occurs, the slings must be inspected per manufacturer's procedures prior to being placed in any further service.
Appropriately rated lifting slings 41
Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
SURFACE TEMPERATURE HAZARD
The meter body and piping may be extremely hot or cold.
Wear appropriate personal protective equipment when coming in contact with the meter. Failure to do so may result in injury.
June 2014 3-9000-759 Rev D

2.5 Mounting requirements in heated or cooled pipelines

The ambient operating temperature of the Daniel 3410 Series Gas Ultrasonic Flow Meter
o
electronics (i.e. Flameproof enclosure and Intrinsically safe base enclosure) is -40
o
F (+60 oC). If the meter is installed into a pipeline which is heated or cooled outside this
+140 temperature range it is necessary to remove the electronics housing from the meter body (i.e. Spool piece acting as process fluid conduit) and mount it next to the meter body on a pipe stand or other rigid structure.
Extended length transducer cables (P/N 2-3-3400-194, 15 ft. long) shall be used to connect the Daniel 3410 Series Gas Ultrasonic Flow Meter electronics to the transducers installed in the meter body. The process temperature must also not exceed the operating temperature range of
the transducers. T-11, T-12 and T-21 transducers have an operating range from -4 to 212 212
o
F (+100
o
F (+100
o
C). T-22 transducers have an operating range from -58 oF (-50
o
C).
F (-40 oC) to
o
F (-20 oC)
o
C) to
42 Mounting requirements in heated or cooled pipelines
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
ΩΩΩ
3-9000-759 Rev D June 2014

Section 3: Electrical installation

3.1 Cable length TTL mode

The maximum cable length is 2000 feet when the Digital Output “TTL” mode is selected.

3.2 Cable length Open Collector mode

For the Digital Output “open collector” mode, the maximum cable length depends on the cable parameters, pull-up resistance used, the maximum frequency to output, and frequency input parameters being driven. The following table provides estimated cable lengths for different pull­up resistor values and different Max Frequency settings in the meter using the following cable parameters. The table also provides an estimated cable voltage drop which indicates how much voltage will be across the cabling and effectively indicates to what voltage level the frequency input can be pulled down to by the frequency output.
If the voltage drop is higher than the voltage required for the frequency input to see a low state, then the configuration will most likely not work for your system. Performance of frequency outputs will vary from this table with setup and frequency input being driven.
Table 3-1 Configurations for open collector frequency outputs
Cable
Cable
Length (2 Conductors) Capacitance Resistance Resistance Frequency Current (2 Conductors)
(x1000ft) nF (Hz) (A) VDC
0.5 16.8 10.00 1000 1016.8 5000 0.024 0.397
1 33.6 20.00 1000 1033.6 1000 0.023 0.780
2 67.2 40.00 1000 1067.2 1000 0.022 1.511
4 134.4 80.00 1000 1134.4 1000 0.021 2.843
0.5 16.8 10.00 500 516.8 5000 0.046 0.780
1 33.6 20.00 500 533.6 5000 0.045 1.511
1.7 57.12 34.00 500 557.12 5000 0.043 2.461
6.5 218.4 130.00 500 718.4 1000 0.033 7.296
resistance
The 22 AWG wire characteristics:
Capacitance = 20 pF/ft or 20 nF/1000 ft (between two wires)
Resistance = 0.0168 Ohms/ft or 16.8 Ohms/1000 ft
Pull-up voltage = 24 VDC
Cable
Pull-up resistance
Tot al
Maximum frequency
Sink
Cable voltage drop
Cable length TTL mode 43
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
The internal grounding terminal shall be used as the primary equipment ground. The external terminal is only a supplemental bonding connection where local authorities permit or require such a connection. DO NOT connect digital grounds to the ground lugs.
A. Transmitter Electronics Enclosure ground lug
A.
June 2014 3-9000-759 Rev D

3.3 Grounding meter electronics housing

The meter electronics should be internally grounded for intrinsically safe operations. Connect a wire to the chassis ground lug installed inside the Transmitter Electronics Enclosure as the primary ground. A secondary ground is located outside of the Transmitter Electronics Enclosure (see Figure 3-2).
Figure 3-1 Internal Transmitter Electronics Enclosure chassis ground
44 GROUNDING METER ELECTRONICS HOUSING
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
A.
A. External ground lug
EXPLOSION HAZARD
To reduce the risk of an explosion or fire, conduit runs must have a sealing fitting connected within 457.2 mm (18 inches) of the enclosure. Substitution of components may impair intrinsic safety of the meter.
Failure to keep covers tight during operation may result in serious injury or death.
EXPLOSION HAZARD
Substitution of components may impair the intrinsic safety and cause ignition of flammable or combustible atmospheres. Disconnect power before servicing.
Failure to remove power and use Daniel approved components may cause serious injury.
3-9000-759 Rev D June 2014
Figure 3-2 External ground lug

3.4 Conduit seals

Conduit seals are required for meter installations in hazardous environments. Adhere to safety instructions to protect personnel and equipment.
Conduit seals 45
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
HAZARDOUS VOLTAGE INSIDE
Do not open the Transmitter Electronics Enclosure when an explosive gas atmosphere is present. Disconnect equipment from supply circuit before opening the enclosure.
Failure to remove power may result in serious injury or death.
June 2014 3-9000-759 Rev D

3.4.1 Startup for systems using explosion-proof conduit

1. Assemble conduit to the Transmitter Electronics Enclosure. A conduit seal fitting is
required within 18 inches (457 mm) of the enclosure.
2. Check to make certain that all power to field wiring is turned OFF.
3. Remove the end cap nearest the conduit entry to gain access to the transmitter
electronics.
4. Pull the wires into the electronics enclosure.Complete the field connection wiring as
shown in see Figure 3-3 and Section 3.5.
46 STARTUP FOR SYSTEMS USING EXPLOSION-PROOF CONDUIT
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
A.
A. Conduit wiring entry (four entries)
Switches:
C. Upper terminal block
D. FODO Group 2
B.
C.
G.
B.
FODO2 GND2 FODO3
E. Analog Out (Current 4-20mA)
AO2+ AO2 -
1. Port A
2. DHCP
3. WRITE PROT.
Analog In (AO1) Analog Input 1 (Temperature) TT+ TT -
PT+ PT -
Analog In (AO2) Analog Input 2 (Pressure)
D.
E.
F.
F. Analog In
Upper terminal block (continued)
G. Ground lug
3-9000-759 Rev D June 2014
5. Complete the field connection wiring and apply electrical power to the system.
Figure 3-3 Electronics field wiring - upper terminal block, switches, ground lug
Startup for systems using explosion-proof conduit 47
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
F.
G.
H.
A. Upper terminal block B. FODO Group 1 connections
FODO1 GND1 DI 1
E. Ethernet
Ethernet (orange and white wire) Ethernet (orange wire) Ethernet (green and white wire) Ethernet (green wire)
C. AO1 AO1+ AO1 -
D. Serial COMs (RS-232, RS-485) RS-232: RTS, TX, RX, CTS RS-485: TX+, TX -, RX+, RX - (4-wire Full Duplex )
B.
C.
D.
E.
A.
RS-485: TX+, TX- (2-wire Half Duplex)
F. 24V loop power (4-24mA inputs/outputs) G. Power In (104VDC - 36VDC) H. Fuse cover
June 2014 3-9000-759 Rev D
Figure 3-4 Transmitter electronics field wiring lower terminal block
48 STARTUP FOR SYSTEMS USING EXPLOSION-PROOF CONDUIT
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
3-9000-759 Rev D June 2014
6. Set or configure the meter operating parameters using Daniel MeterLink. For additional
installation information refer to the system wiring diagram (see Appendix A), Daniel
MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual
(P/N 3-9000-763) and use the Daniel MeterLink Field Setup Wizard to complete the configuration.
7. Verify the field connections are working correctly. Allow the system to run for the time
specified by the customer (usually one week) and an electrician has fully tested the connections. After the Acceptance Test is witnessed and approved, seal the conduit.
8. Power down the system and apply the sealing compound to the conduit and allow to
set in accordance with manufacturer specifications.
9. If required, install the security latches and wire seals on the Transmitter Electronics
Enclosure end caps (see Section 3.7 and Figure 3-11).
10. If required, install the wire seals through the socket head bolts on the Base Enclosure
(see Section 3.7, Figure 3-12 and Figure 3-13).
11. Re-apply electrical power to the system.
Startup for systems using explosion-proof conduit 49
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
HAZARDOUS VOLTAGE INSIDE
Do not open the Transmitter Electronics Enclosure when an explosive gas atmosphere is present. Disconnect equipment from supply circuit before opening the enclosure.
Failure to remove power may result in serious injury or death.
June 2014 3-9000-759 Rev D

3.4.2 Startup for systems that use flame-proof cable

1. Check to make certain that all field wiring power is turned OFF.
2. Remove the end cap nearest the cable entries to gain access to the transmitter
electronics.
3. Install the cable and cable gland.
4. Complete the field connection wiring and apply electrical power to the system.
5. Set or configure the meter operating parameters using Daniel MeterLink. For additional
installation information refer to the system wiring diagram (see Appendix A), Daniel
MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual
(P/N 3-9000-763) and use the Daniel MeterLink Field Setup Wizard to complete the configuration.
6. Verify the field connections are working correctly. Allow the system to run for the time
specified by the customer (usually one week) and an electrician has fully tested the connections. After the Acceptance Test is witnessed and approved, seal the conduit.
7. Power down the system and apply the sealing compound to the conduit and allow to
set in accordance with manufacturer specifications.
8. If required, install the security latches and wire seals on the Transmitter Electronics
Enclosure end caps (see Section 3.7 and Figure 3-11).
9. If required, install the wire seals through the socket head bolts on the Base Enclosure
(see Section 3.7, Figure 3-12 and Figure 3-13).
10. Re-apply electrical power to the system.
50 STARTUP FOR SYSTEMS THAT USE FLAME-PROOF CABLE
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
If not using Ethernet, a full duplex serial connection is necessary for Daniel MeterLink to communicate with a Daniel 3410 Series Gas Ultrasonic Meter.
3-9000-759 Rev D June 2014

3.5 Wiring and I/O

Daniel MeterLink uses the TCP/IP protocol to communicate with the Daniel 3410 Series Ultrasonic Gas Flow Meter electronics instead of Modbus ASCII or RTU. The TCP/IP protocol only
works across either Ethernet, RS-485 full duplex (4-wire) or RS-232. Daniel MeterLink communicate with multiple meters if they are multi-dropped using 4-wire, full duplex RS-485 mode. The meter electronics is HART capable and provides communication flexibility with Daniel 3410 Series Gas Ultrasonic Flow Meters.
Note: Port B for RS-485 full duplex communication is not supported.
The HART® output provides communication with other field devices (e.g., Field Communicator and AMS™ Device Manager software) and ultimately, communicates key diagnostic information through PlantWeb® architecture.
can
The Daniel 3410 Series Gas Ultrasonic Flow Meter electronics auto-detects the protocol used and automatically switches between TCP/IP, Modbus ASCII, and Modbus RTU so it is not necessary to make any meter configuration changes to the protocol.
Wiring and I/O 51
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
A.
B.
C.
D.
E.
A. Acquisition/Measurement mode B. Power C. RX (RS-485/RS-232) - receiving data D. LED 4 - not used E. LED 5 - not used F. TX (RS-485/RS-232) - transmitting data (RS-485 2-wire use TX+ and TX-)
F.
G.
G. Link (Eth1 Link) - user Ethernet connection
June 2014 3-9000-759 Rev D

3.5.1 CPU Module labeling and LED indicators

The meter’s metrology mode and the status of the data transfer from the Acquisition Module to the CPU Module are indicated via light-emitting diode (LED) status indicators. The WRITE PROT. switch protects the meter’s configuration,
Figure 3-5 CPU Module labeling and LED indicators
52 CPU MODULE LABELING AND LED INDICATORS
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
3-9000-759 Rev D June 2014
Table 3-2 CPU Module labeling and LED functions
CPU Module label or LED
Function Switch position indicator or LED
WRITE PROT.
Write-protect mode - with switch in the ON
position (default setting) protects configuration and firmware overwrites.
To write configuration changes or download
firmware to the meter change the switch to the OFF position
DHCP
Dynamic Host Protocol Server - enables you to
communicate with a Daniel meter that is not connected to a network.
When the CPU Module switch is in the ON
position, the meter is enabled to act as a DHCP server for a single DHCP client connected to the Ethernet port using a crossover cable. This should be used for peer to peer connections only.
When the connection is made, select to use the
Meter Name in the meter instead of the Meter Directory Name in order to keep all log files and configurations separate from each meter.
PORT A
PORT A override - RS-232 serves as an override
during meter commissioning to establish com­munications and in the event the user cannot communicate with the meter due to an inadver­tent communication configuration change. The override period is for two minutes.
Supports:auto-detected ASCII (Start bit 1, Data Bit 7,
Parity Odd/Even, Stop Bit 1)
RTU (Start Bit 1, Data Bit 8, Parity none, Stop
Bit 1).
Modbus protocols
RS-232 Baud rate=19,200
Modbus ID=32
MEAS System color indicates metrology mode
Acquisition mode
Measurement mode
PWR
LED 4
LED 5
RX
TX
LINK
3.3V Power Indicator Solid Green
Not used
Not used
RX signal (Port A for RS485 or RS232 )
communication) receiving data
TX signal (Port A for RS485; 2-wire or 4-wire) or
RS232 communication) transmitting data
ETH1Link user Ethernet connection Solid green
Switch position
ON - (default setting) enables write-
protection of the configuration and firmware
OFF - enables writing configuration
changes or downloading firmware
Switch position
ON - the meter is enabled to act as a
DHCP server for a single DHCP client
OFF - disables the DHCP server
Switch position
ON - enables RS-232 PORT A override
OFF - (default setting) disables
RS-232 PORT A override
LED status
Red flashing LED
Solid red the Acquisition Module is not
communicating with the CPU Module
Green flashing LED
Flashing green (when receiving data)
Flashing green (when transmitting
data)
CPU Module labeling and LED indicators 53
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
RESTRICTED ETHERNET AND SERIAL CONNECTIVITY USAGE
Failure to restrict Ethernet and communication access to the Daniel 3410 Series Gas Ultrasonic Flow Meter can result in, among other things, unauthorized access, system corruption, and/or data loss.
User is responsible for ensuring that physical access and Ethernet or electronic access to the Daniel 3410 Series Gas Ultrasonic Flow Meter is appropriately controlled and any necessary security precautions are implemented; such as, establishing a firewall, setting password permissions and/or implementing security levels.
RX+
TX+
TX -
RX -
June 2014 3-9000-759 Rev D
Ethernet communications
The Ethernet port IP address, subnet mask, and gateway address are software-configurable. In addition, a meter can be configured to act as a DHCP (Dynamic Host Configuration Protocol) server to assign an IP address to a PC or laptop running Daniel MeterLink. The DHCP server facility is not intended to act as a general purpose DHCP server for a wider network. To this end, no user control is provided over the class or range of IP addresses the unit provides. A standard twisted pair (Cat-5) cable should be used for Ethernet wiring.
It is strongly recommended that the meter be configured using an independent (off-network) single host. After configuration of the Daniel 3410 Series Gas Ultrasonic Flow Meter, the DHCP option must be turned off if used on a LAN/WAN.
Table 3-3 Ethernet cable to PC communication
Ethernet communication
Wire color CPU
White w/Orange Stripe TX+
Orange w/White Stripe TX -
White w/Green Stripe RX+
Green w/White Stripe RX -
Use ethernet cable, Daniel P/N 3-3400-079, to connect the PC to the meter.
A DIN 41612 48-pin connector is the interface from the CPU Module to the Field Connection Board (male end located on the back of the Field Connection Board).
54 CPU MODULE LABELING AND LED INDICATORS
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
3-9000-759 Rev D June 2014
Modbus TCP
If the meter firmware supports Modbus TCP slave functionality, the following controls will be available.
Modbus TCP unit identifier: Enter the Modbus TCP unit identifier here. Valid values are 0-255.
Enable alternate Modbus TCP port: The standard TCP port for Modbus TCP is port 502. This port is always enabled in a meter that supports Modbus TCP. By selecting this option, you can also enable Modbus TCP communications on a secondary TCP port specified by Alternate Modbus TCP por t.
Alternate Modbus TCP port:
Enter the alternate TCP port number here after selecting Enable alternate modbus TCP port. Valid port numbers are from 1 to 65535. The meter will not allow some port numbers that are either used by the meter or are defined port numbers for other protocols. Daniel MeterLink™ will prompt you if it was not able to write the specified port number to the meter.
CPU Module labeling and LED indicators 55
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
Serial connections
Use a serial cable, Daniel P/N 3-2500-401, to connect to a PC running Daniel MeterLink. The cable is designed for RS-232 communications which is the serial Port A default configuration (see Appendix A field wiring diagram, Daniel Drawing DMC - 005324). The DB-9 end of the cable plugs directly into the PC running Daniel MeterLink. The three wires on the other end of the cable connect to the CPU Module RS-485/RS-232 terminals. The RED wire goes to RX, the WHITE wire goes to TX, and the BLACK wire goes to COM (see Table 3-6 for Port A wiring). RS-485, 2-wire connection on Port A, uses TX+ and TX- on the CPU Module and has a ground wire.
When Beldon wire No. 9940 or equivalent is used, the maximum cable length for RS-232 communications at 9600 bps is 88.3 meters (250 ft.) and the maximum cable length for RS-485 communication at 57600 bps is 600 meters (1970 ft.).
Port A supports a special override mode which forces the port to use known communication values (19200 baud, address 32, RS-232). Note that the protocol is auto-detected. This mode is expected to be used during meter commissioning (to establish initial communication) and in the event that the user cannot communicate with the meter (possibly due to an inadvertent communication configuration change). Alternately, when using Daniel MeterLink with an Ethernet port, use Ethernet cable, Daniel P/N 3-3400-079, to connect the PC.
Table 3-4 Serial Port A parameters
Port/Communication Description Common features
Port A (Standard)
RS-232
RS-485 Half Duplex
RS-485 Full Duplex
RS-485
1
(2-wire
communication on Port A.)
Typically used for general communications with
a flow computer, RTU (Modbus slave) and radios.
RS-485 2-wire (Half Duplex) connected to
+ and TX -
TX Special override mode to force port configura-
tion to known settings.
Supports RTS/CTS handshaking with software-
configurable RTS on/off delay times.
Factory default is RS-232, Address 32, 19200
baud.
Communications via Daniel MeterLink
using RS-232 or RS-485 Full Duplex
Software configurable Modbus
Address (1-247)
Auto-detects TCP/IP and ASCII or RTU
Protocol
ASCII Protocol:
Start Bits = 1, Data Bits = 7
Parity: odd or even 1, Stop Bits = 1Baud Rates: 1200, 2400, 9600,
19200, 38400, 57600, 115000 bps
RTU Protocol:
Start Bits = 1, Data Bits = 8
Parity: none, Stop Bits = 1Baud Rates: 1200, 2400, 9600,
19200, 38400, 57600, 115000 bps
Ethernet
Preferred port for diagnostic communication
via Daniel MeterLink
Modbus TCP/IP, Modbus TCP
10 Mbps/100 Mbps
1. RS-485 2-wire connections use TX+ and TX - on the CPU Module
2
2
2
2
56 CPU MODULE LABELING AND LED INDICATORS
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
If not using Ethernet, a full duplex serial connection is necessary for Daniel MeterLink to communicate with a Daniel 3410 Series Gas Ultrasonic Flow Meter.
RX
TX
COM
DB-9 Connector Female
3-9000-759 Rev D June 2014
2. Denotes auto-detected protocols
Figure 3-6 PC to meter serial connection wiring

3.6 I/O connections

The Daniel 3410 Series Gas Ultrasonic Flow Meter provides the I/O connections on the CPU Module.
I/O connections 57
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
A.
B.
C.
D.
A. Frequency/Digital Output 2 B. Frequency/Digital Output 3 C. Analog Output 2 - 4-20mA output D. Analog Input - HART temperature and pressure connections
June 2014 3-9000-759 Rev D
Figure 3-7 CPU Module I/O connections
58 I/O CONNECTIONS
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
3-9000-759 Rev D June 2014

3.6.1 Frequency/Digital outputs

The meter has three user-configurable selections for configuring either a Frequency output or Digital output (FODO).
FODO1 (four possible parameter configurations)
FODO2 (eight possible parameter configurations)
FODO3 (eight possible parameter configurations)
Frequency or Digital Outputs (FODO 1) source
FO1A, DO1A, FO1B, DO1B
Frequency output 1A is based on frequency content (Actual - Uncorrected Flow Rate)
Frequency output 1B is based on frequency content and Frequency 1 B Phase
Digital output 1A is based on Digital output1A content (Frequency1A Validity or Flow
Direction)
Frequency or Digital Outputs (FODO 2) source
FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
Frequency output 1A is based on frequency content (Actual - Uncorrected Flow Rate)
Frequency output 1B is based on frequency content and Frequency 1B Phase
Frequency output 2A is based on frequency content (Actual - Uncorrected Flow Rate)
Frequency output 2B is based on frequency content and Frequency 2B Phase
Digital output 1A is based on Digital output1A content (Frequency 1Validity and Flow
Direction)
Digital output 1B is based on Digital output1B content (Frequency 1 Validity and Flow
Direction)
Digital output 2A is based on Digital output 2A content (Frequency 2 Validity and Flow
Direction)
Digital output 2B is based on Digital output 2B content (Frequency 2 Validity and Flow
Direction)
Frequency/Digital outputs 59
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
Frequency or Digital Outputs (FODO 3) source
FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
Frequency output 1A is based on frequency content (Actual - Uncorrected Flow Rate)
Frequency output 1B is based on frequency content and Frequency 1B Phase
Frequency output 2A is based on frequency content (Actual - Uncorrected Flow Rate)
Frequency output 2B is based on frequency content and Frequency 2B Phase
Digital output 1A is based on Digital output 1A content (Frequency 1 Validity and Flow
Direction)
Digital output 1B is based on Digital output 1B content (Frequency 1Validity and Flow
Direction)
Digital output 2A is based on Digital output 2A content (Frequency 2 Validity and Flow
Direction)
Digital output 2B is based on Digital output 2B content (Frequency 2 Validity and Flow
Direction)
Mode options
Open Collector (requires external excitation supply voltage and pull-up resistor)
TTL (internally powered by the meter 0-5 VDC signal)
Channel B Phase options:
Lag forward, Lead reverse (Phase B lags Phase A while reporting forward flow, leads Phase
A while reporting reverse flow)
Lead forward, Lag reverse (Phase B leads Phase A while reporting forward flow, lags Phase
A while reporting reverse flow)
Phase A and Phase B output (based on flow direction)
Reverse flow - output only reports flow in the reverse direction. For frequency outputs,
Phase B of the output is 90 degrees out of phase with Phase A.
Forward flow - output only reports flow in the forward direction. For frequency outputs,
Phase B of the output is 90 degrees out of phase with Phase A.
Absolute - output reports flow in both directions. For frequency outputs, Phase B of the
output is 90 degrees out of phase with Phase A.
Bidirectional - output reports flow on Phase A only in the forward direction and on Phase
B only in the reverse direction.
60 FREQUENCY/DIGITAL OUTPUTS
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
FODO1
FODO2
FODO3
FO1A FO1B DO1A DO2B
FO2A FO2B DO2A DO2B
3-9000-759 Rev D June 2014
Maximum frequency for the frequency outputs
1000Hz
5000Hz
Table 3-5 Frequency/Digital Outputs possible configurations
Frequency/Digital output Source configuration
Frequency /Digital Output 1
1
Frequency output 1A
Frequency output 1B
Digital output 1A
Digital output 1B
Frequency /Digital Output 2
2
Frequency output 1A
Frequency output 1B
or
Digital output 1A
Digital output 1B
Frequency /Digital Output 3
2
Frequency output 2A
Frequency output 2B
Digital output 2A
Digital output 2B
1. Solid blue line denotes valid selection for Frequency/Digital Output 1.
2. Black dashed -line denotes valid selections for Frequency/Digital Output 2 and
Frequency/Digital Output 3.
Frequency/Digital outputs 61
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
A.
A. FODO1 and Digital input1 - shared common ground (Group 1) B. FODO2 and FODO3 - shared common ground (Group 2)
B.
June 2014 3-9000-759 Rev D
Output for FODO1 and Digital Output1 (Group 1 on the CPU Module) share a common ground and have 50V isolation. FODO2 and FODO3 (Group 2 on the CPU Module) share a common ground and have 50V isolation. This allows an output to be connected to a different flow computer. The outputs are opto-isolated from the CPU Module and have a withstand voltage of at least 500V rms dielectric.
Figure 3-8 CPU Module - Frequency/Digital outputs common ground

3.6.2 Analog input settings

The Daniel 3410 Series Gas Ultrasonic Flow Meter has the capability to sample analog temperature (Analog Input 1) and pressure (Analog Input 2) with 4-20 mA signals. These analog input signals are configured to sink. The two independent analog input circuits are configured for conventional 4-20 mA service. Also, 24VDC isolated power supply connection is provided for an external power source. Refer to the Field wiring diagram DMC - 005324 in Appendix A.
62 ANALOG INPUT SETTINGS
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
DHCP
Switch 2
3-9000-759 Rev D June 2014

3.6.3 Analog output settings

The Daniel 3410 Series Ultrasonic Gas Flow Meter provides two 4-20 mA analog output signals that are software configurable for either sink or source current (see Appendix A DMC - 005324).
Full HART® functionality is provided so that any commercially available HART® transmitter which meets the specifications of the HART® Communications Foundation can be connected to the Daniel® Liquid Ultrasonic Flow Meter.
Analog Output1 (AO1) is user-configurable as a 4-20mA output and has HART
capabilities.
Analog Output 2 (AO2) is user-configurable as a conventional 4-20 mA output.

3.6.4 Digital Input

The Daniel 3410 Series Gas Ultrasonic Flow Meter provides one digital input that can be used as a general purpose input. The digital input must be configured via the Daniel MeterLink
Tools|Edit/Compare Configuration screen.

3.6.5 DHCP server switch settings

The meter can be configured to act as a DHCP server. The DHCP server is enabled/disabled via CPU Module DHCP switch as follows:
Table 3-6 DHCP server switch settings
CPU Module switch
DHCP server disabled
OFF ON
DHCP server enabled
Analog output settings 63
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
WRITE PROT.
Switch 3
June 2014 3-9000-759 Rev D

3.6.6 Configuration protect switch settings

The meter’s configuration parameters and firmware can be protected against changes via the CPU Module WRITE PROT. switch as follows:
Table 3-7 Configuration protect switch settings
CPU Module switch
Configuration protected
ON (default setting)
Configuration unprotected
OFF
64 CONFIGURATION PROTECT SWITCH SETTINGS
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
A.
B.
A. Power In connector (main power) B. 24V LOOP POWER C. 2 Amp fuse (used for the main power input)
C.
3-9000-759 Rev D June 2014

3.6.7 External power source connection and fuse

Located inside the Transmitter Electronics Enclosure is a connector for a user-provided external power source, a 2 Amp fuse and a 24V loop power connection for ultrasonic meter analog outputs, temperature transmitter or pressure transmitter devices. The current is limited to 88mA.
Figure 3-9 CPU Module power source connections
External power source connection and fuse 65
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
A.
A. Transmitter Electronics Enclosure end cap. Optional glass endcap for Local Display B. Security latch
B.
June 2014 3-9000-759 Rev D

3.7 Security seal installation

Security seals protect the integrity of the meter metrology and prevent tampering with transducer assemblies. The following sections detail how to properly seal the Daniel 3410 Series Gas Ultrasonic Flow Meter after commissioning. The security seal wires are commercially available.
Be sure to set the WRITE PROT. switch on the CPU Module to the ON position prior to sealing the enclosure.

3.7.1 Seal Transmitter Electronics Enclosure

Use the following instructions to install the security seal wires on the Transmitter Electronics Enclosure.
Figure 3-10 Transmitter electronics enclosure security latch
1. Rotate the end cap clockwise fully closing and compressing the end cap seal. Install the
Security latch using a 3mm Allen wrench.
66 SECURITY SEAL INSTALLATION
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
A.
A. Transmitter Electronics Enclosure end cap B. Security wire seals
B.
3-9000-759 Rev D June 2014
2. Install the security seal wire into and through one of the two holes in the end cap.
Choose holes that minimize counterclockwise rotation of the end cap when the security wire is taut (maximum wire diameter .078 inch; 2.0mm).
Figure 3-11 Transmitter Electronics Enclosure security seals
3. Adjust the security wire, removing all slack and thread into the lead seal.
4. Crimp lead seal and cut wire ends to remove excess wire.
Seal Transmitter Electronics Enclosure 67
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
A.
A. Base Enclosure cover
B.
B. Security wire seals
June 2014 3-9000-759 Rev D

3.7.2 Base Enclosure Security Seals

Use the following instructions to install the security seal wire on the Base Enclosure.
Procedure
1. Install security wire seal into and through the hole in the socket head screw on the Base
Enclosure cover (maximum wire diameter .078 inch; 2.0mm).
Figure 3-12 Base Enclosure wire seal installation
2. Position the wire to prevent counterclockwise rotation of the screws when the seal wire
68 BASE ENCLOSURE SECURITY SEALS
is taut.
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
A.
A. Transmitter Electronics Enclosure
D.
B. Security wire sealse C. Transmitter Electronics endcap security latch
B.
C.
D. Base Enclosure
3-9000-759 Rev D June 2014
3. Feed the security wire beneath the Transmitter Electronics Enclosure and through the
adjacent socket head screw. Twist the wire, removing all slack and seal.
Figure 3-13 Base Enclosure security seals
4. Cut wire ends to remove excess wire.
Base Enclosure Security Seals 69
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
A. B.
C.
A. Transducer cable nut B. Transducer cable connector C. Security wire seal
June 2014 3-9000-759 Rev D

3.7.3 Transducer assembly security seals

Use the following instructions and Figure 3-14 to install the security seal wire on the transducer assembly.
1. Rotate the transducer cable nut (Item A) clockwise, compressing the seal on the
transducer cable connector.
2. Insert a security seal wire into one of the two holes in the transducer cable connector
(Item B) and feed through one of the two holes in the transducer cable nut (Item A).
Note: Choose holes that minimize counterclockwise rotation of the transducer cable
nut when the security wire (Item C) is taut.
3. Adjust the security wire, removing all slack and thread into the lead seal.
4. Cut wire ends to remove excess wire.
Figure 3-14 Transducer assembly security seal
70 TRANSDUCER ASSEMBLY SECURITY SEALS
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation
3-9000-759 Rev D June 2014

3.7.4 Sealing the unit

The unit should be properly sealed with a sealing compound after electrical connections have been tested according to the customer's Best Practices schedule. Some areas require a witnessed Acceptance Test for the installed system and require that the meter run for a predetermined length of time (approximately one to two weeks) before the unit is sealed. This allows time to verify all electrical connections are correct, that the meter is accurately measuring flow and that the meter meets the customer’s installation requirements.
See Section 3.4.1 and Section 3.4.2
Sealing the unit 71
Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
72 SEALING THE UNIT
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration
3-9000-759 Rev D June 2014

Section 4: Configuration

After the mechanical and electrical installation is complete and connectivity is established, use the Daniel MeterLink
(P/N 3-9000-763) to setup software communications with the meter.

4.1 Daniel MeterLink Setup

1. Review the software operating system, hardware and peripheral requirements.
2. Follow the installation instructions for your operating system (Windows® XP, Windows
Vista®, Windows® 7 or Windows® 8).
3. Configure a direct connection driver for first time modem configuration for Daniel
MeterLink communications.
4. Select the Daniel MeterLink
Registration Wizard to keep you informed about updates and technical support. Registration options include:
Register by e-mail
Register by phone
Software for Gas and Liquid Ultrasonic Meters Quick Start Manual
desktop icon and complete the information in the
Register later (remind me)
Register later using the Help|Resister Program menu (don’t remind me)
5. Select
6. Connect to your meter. If your meter is not shown in the list, select
7. Save the meter configuration file, collect a Maintenance log and Waveforms to
8. Run the Field Setup Wizard.
File|Program Settings and customize the user-preferences (e.g. User name,
Company name, display units, Liquid Meter volume units and other interface settings)
Edit Meter Directory
and setup the connections properties.
document the meter’s “As Found” settings.
Daniel MeterLink Setup 73
Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D

4.2 Field Setup Wizard

1. Use the Field Setup Wizard-Startup in Daniel MeterLink and select the checkboxes that
allow proper configuration for your meter (Temperature, Pressure, Meter Corrections, Meter Outputs, Gas chromatograph setup, Continuous flow analysis, View local display setup, 4+4 8-path meter combine output and SOS limit percentage). Selections on this page will affect other configuration selections. Select setup.
2. Use General setup to configure the meter’s units system (U.S Customary or Metric
units) volume units, flow rate time, low flow cutoff, contract hour, enable reverse flow alarm, set meter time and notepad comments. Select Digital Outputs page.
Note: The Meter’s Units system configured on the General Page affect the units for the
optional Local Display items.
3. Set the Frequency/Digital Outputs Sources for either a frequency output or a digital
status. Select the Source for each Frequency/Digital output and select the desired drive Mode. The Mode options are Open Collector which requires an external excitation voltage and pull-up resistor or TTL mode which outputs a 0-5 VDC signal. Select continue to Frequency Outputs page.
Next to continue to General
Next to continue to Frequency/
Next to
Note: Frequency outputs 1 and Digital outputs 1 are paired together meaning the
Digital outputs 1 will report the status for the parameter for Frequency outputs 1. Similarly, Frequency outputs 2 and Digital outputs 2 are paired together. Additionally, each Frequency output has an A and B output phase.
4. Configure Frequency output 1 and Frequency output 2 content, flow direction, Channel
B phase, maximum frequency output (Hertz) and full scale volumetric flow rate. Select
Next to continue to Meter Digital Outputs.
5. Select the Meter Digital Output parameters for Digital output 1A, Digital output 1B,
Digital output 2A and Digital output 2B based on Frequency validity or flow direction.
if the output of the ultrasonic meter is reversed from what a flow computer is expecting, select Inverted Operation. This changes the digital output from a HIGH for a TRUE condition to output a LOW for a TRUE condition. Select
Next to continue to
Current Outputs.
6. Current Outputs are based on Uncorrected (Actual) flow rate) content, flow direction
(Forward, Reverse or Absolute) and Full scale volumetric flow rate used with output (20mA maximum). Alarm action parameters determines the state the output will drive during an alarm condition (High 20mA, Low - 4 mA, Hold last value, Very low - 3.5, Very high 20.5 mA or None). Select
Next to configure the HART® Output(s) parameters.
7. HART® Output parameters include four Dynamic process variables (Primary,
Secondary, Third and Fourth variable. The Primary variable is set to match the Content set for Current output 1. If a second current output is available, the Secondary variable is set to match the Content set for Current output 1) Identification and HART® units (volume units, flow rate time units, velocity units, pressure and temperature units).
Next to continue to the Meter Corrections page.
Select
74 Field Setup Wizard
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration
3-9000-759 Rev D June 2014
8. The Meter Corrections page allows you to configure the flow profile, temperature,
pressure and linear expansion corrections, pipe outside diameter The Young’s Modulus value (ratio of tensile stress to tensile strain) and the Poisson’s ratio value (the absolute ratio of the pipe material lateral strain over axial strain) for 3412 and 3411 Gas Ultrasonic Flow meters. Click
Next to continue to the Temperature and Pressure page.
9. Set the temperature and pressure scaling for analog inputs, enter fixed values, and set
alarm limits for both. The alarm limit selections are hold last output value or use fixed value.
Live temperature selections include minimum and maximum inputs or fixed
temperature.
Live pressure selections include minimum and maximum inputs, gage
(atmospheric pressure), absolute, or fixed pressure.
Click Next to continue to the Gas Chromatograph Setup page.
10. Select the settings below to configure a serial port as a Modbus Master to poll a gas chromatograph.
Serial Port:
select which serial port will be connected to the GC. While the port is
configured for communications to a GC, it will not act as a Modbus slave device for communications from Daniel MeterLink™ or a SCADA system.
GC protocol: select the protocol for which the GC is configured. The Daniel Gas
Ultrasonic meter uses 7 data bits, Even parity, and 1 stop bit for ASCII Modbus and 8 data bits, No parity, and 1 stop bit for RTU Modbus
GC baud rate: select the baud rate for which the GC is configured.
GC comms address: enter the Modbus ID of the GC.
GC stream number: enter the stream number for the gas composition the Daniel
Gas Ultrasonic meter will read.
GC heating value units: elect the units for which the heating value is configured
in the GC.
Use which gas composition on GC alarm: select which gas composition the
Daniel Gas Ultrasonic meter will use if the GC goes into alarm. If Fixed value is selected, the meter will start using the fixed gas composition stored in the meter. If Last good value is selected, the meter will use the last gas composition collected from the GC before the GC started to report alarms. Click Next to continue to the AGA8 page.
Field Setup Wizard 75
Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
11. Configure the properties necessary for the AGA8 calculations. This page is only displayed for Daniel Gas Ultrasonic meters if both temperature and pressure are set to Live or Fixed and Base condition correction is selected on the Startup Page. Configuration parameters include:
Calculations performed internally (by the meter) or Externally
AGA8 method - Gross Method 1, Gross method 2 or Detail
Base temperature and pressure
Specific gravity - reference temperature and pressure
Volumetric gross heating value and reference temperature
Molar density reference temperature and pressure
Flow Mass density, flow compressibility and Base compressibility
Gas composition inputs - components and mole percent
Click Next to continue to the Continuous Flow Analysis page, if View Continuous Flow Analysis setup was selected on the Startup page.
12. Configure the flow analysis parameters shown below.
Set low and high flow limits
Enable SOS comparison (requires AGA 8 Detail method)
Enable liquid detection and Profile factor limit
Enable Blockage. Enter the percent for Symmetry, Cross-flow, Chords A to D
turbulence
Enable Internal bore buildup
Click Next to continue to the Local display page, if View local display setup was selected
on the Startup page.
13. Configure the parameters for the local display. Use the drop-down arrow in the Display Items list box and select or modify the parameters that will be displayed; the Display items, the Display units and the Scroll delay.
76 Field Setup Wizard
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration
3-9000-759 Rev D June 2014

Display Items:

The Local Display’s labels and descriptions are shown below:
Table 4-1 Local display labels, descriptions and valid units
Local display labels, descriptions and valid units
QFLOW — Uncorrected volume flow rate
ACF – Actual Cubic Feet
ACM – Actual Cubic Meters
MACF – Thousand Actual Cubic Feet
MACM –Thousand Actual Cubic Meters
TDYVL — Current day’s forward uncorrected volume
+ACF – Actual Cubic Feet
+ACM – Actual Cubic Meters
+MACF – Thousand Actual Cubic Feet
+MACM –Thousand Actual Cubic Meters
TDYVL — Current day’s reverse uncorrected volume
-ACF – Actual Cubic Feet
-ACM – Actual Cubic Meters
-MACF – Thousand Actual Cubic Feet
-MACM –Thousand Actual Cubic Meters
YSTVL — Previous day’s forward uncorrected volume
+ACF – Actual Cubic Feet
+ACM – Actual Cubic Meters
+MACF – Thousand Actual Cubic Feet
+MACM –Thousand Actual Cubic Meters
YSTVL — Previous day’s reverse uncorrected volume
-ACF – Actual Cubic Feet
-ACM – Actual Cubic Meters
-MACF – Thousand Actual Cubic Feet
TOTVL — Forward uncorrected volume
+ACF – Actual Cubic Feet
+ACM – Actual Cubic Meters
+MACF – Thousand Actual Cubic Feet
+MACM –Thousand Actual Cubic Meters
TOTVL — Reverse uncorrected volume
-ACF – Actual Cubic Feet
-ACM – Actual Cubic Meters
-MACF – Thousand Actual Cubic Feet
-MACM –Thousand Actual Cubic Meters
QBASE — Corrected volume flow rate
SCF – Standard Cubic Feet
SCM – Standard Cubic Meters
MSCF – Thousand Standard Cubic Feet
MSCM – Thousand Standard Cubic Meters
Field Setup Wizard 77
Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
Table 4-1 Local display labels, descriptions and valid units
Local display labels, descriptions and valid units
TDYVL — Current days forward corrected volume
+SCF – Standard Cubic Feet
+SCM – Standard Cubic Meters
+MSCF – Thousand Standard Cubic Feet
+MSCM – Thousand Standard Cubic Meters
TDYVL — Current days reverse corrected volume
-SCF – Standard Cubic Feet
-SCM – Standard Cubic Meters
-MSCF – Thousand Standard Cubic Feet
-MSCM – Thousand Standard Cubic Meters
YSTVL — Previous days forward corrected volume
+SCF – Standard Cubic Feet
+SCM – Standard Cubic Meters
+MSCF – Thousand Standard Cubic Feet
+MSCM – Thousand Standard Cubic Meters
YSTVL — Previous days reverse corrected volume
-SCF – Standard Cubic Feet
-SCM – Standard Cubic Meters
-MSCF – Thousand Standard Cubic Feet
-MSCM – Thousand Standard Cubic Meters
TOTVL — Forward corrected volume
+SCF – Standard Cubic Feet
+SCM – Standard Cubic Meters
+MSCF – Thousand Standard Cubic Feet
+MSCM – Thousand Standard Cubic Meters
TOTVL — Reverse corrected volume
-SCF – Standard Cubic Feet
-SCM – Standard Cubic Meters
-MSCF – Thousand Standard Cubic Feet
-MSCM – Thousand Standard Cubic Meters
VEL — Average flow velocity
Ft/S – Feet per Second
M/S – Meters per Second
SOS — Average sound velocity
Ft/S – Feet per Second
M/S – Meters per Second
TEMP — Flow-condition temperature
DEGF – Degrees Fahrenheit
DEGC – Degrees Celsius
PRESS — Flow-condition pressure
PSI – Pound per square inch
MPA – Megapascals
FRQ1A — Frequency channel 1A
HZ – Hertz
FRQ1B — Frequency channel 1B
HZ – Hertz
78 Field Setup Wizard
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration
3-9000-759 Rev D June 2014
Table 4-1 Local display labels, descriptions and valid units
Local display labels, descriptions and valid units
KFCT1 — Frequency 1 K-factor
CF – Cubic Feet
CM –Cubic Meters
MCF – Thousand Cubic Feet
MCM –Thousand Cubic Meters
FRQ2A — Frequency channel 2A
HZ – Hertz
FRQ2B — Frequency channel 2B
HZ – Hertz
KFCT2 — Frequency 2 K-factor
CF – Cubic Feet
CM –Cubic Meters
MCF – Thousand Cubic Feet
MCM –Thousand Cubic Meters
AO1 — Analog Output 1 current
MA – Milliamperes
AO2 — Analog Output 2 current
MA – Milliamperes
Note: When connected to a meter with the local display option, reverse flow
direction is indicated with a minus sign (negative) before the value(s) shown on the local display.

Display units:

The Meter volume units displayed are either U.S. Customary or Metric. To modify the Display Units, configure the Meter units system in the Field Setup Wizard — General Page.
U.S. Customary volume unit selections are:
Cubic feet
Thousand cubic feet
Metric volume unit selections are:
Cubic meters
Display units preceded by a plus or minus sign indicate forward and reverse flow
direction, as shown in the table below.\
The local display Flow rate time units are modifiable by selecting the drop-down arrow
and clicking the time unit in the list box.
Valid flow rate time units selections are:
second
hour
day
Field Setup Wizard 79
Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D

Scroll delay:

The Scroll Delay is the time interval for the selected display items to be shown on the Local Display. The default scroll delay setting is five seconds. Click the spin box up or down arrow to increase or decrease the length of time an item displays.
1. Select
2. Save the meter configuration file, collect a Maintenance log and Waveforms to
Finish to write the configuration settings to the meter.
document the “As Left” settings.

4.3 Using AMS Device Manager to configure the meter

This procedure assumes you have AMS Device Manager installed on the host computer and have downloaded the latest Daniel Liquid Ultrasonic Meter Device Description (DD).
If not installed, click the link below to download the AMS device installation tool kit.
http://www2.emersonprocess.com/en-US/documentation/deviceinstallkits/Pages/devicein­stallkitsearch.aspx
80 Using AMS Device Manager to configure the meter
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration
3-9000-759 Rev D June 2014

Procedure - installing AMS Device Description

1. Use the link above to search for the Device Description (DD) for your Daniel 3810 Series
Liquid Ultrasonic Flow Meter.
2. Use the pull-down menu and select the Brand/Manufacturer -
3. Next select the Device,
4. Choose the
5. Next, select
6. Select
7. Select the
Device Revision1, from the pull-down menu.
HART from the Communication Protocol menu.
AMS Device Manager for the Host System.
Host System Revision 11.5.
Liquid 3810 Series from the pull-down menu.
8. Verify your search parameters are correct, as shown below.
Figure 4-1 AMS Device Description search
Emerson Daniel Industries.
9. Click
Using AMS Device Manager to configure the meter 81
Search Now.
Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
10. Click the Daniel Industries Liquid 3810 Series Rev 1 hyperlink. The file download dialog
displays. Click the
Save button to save the files to your host system. You may use the
default download location or change the directory.
11. AMS file download options
12. Click the
Save button to complete the file download.
Figure 4-2 AMS file download complete
13. Click
Open or Open Folder to view the downloaded files.
14. Establish power to the meter and wiring to Analog Input 1 for HART communication.
15. Start the AMS Device Manager using a laptop or PC.
16. Enter login credentials and click
OK to launch the application.
82 Using AMS Device Manager to configure the meter
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration
3-9000-759 Rev D June 2014
17. Click the Configure tab, and then select Guided Setup, Manual Setup or Alert Setup.
Figure 4-3 AMS Device Manager
Figure 4-4 AMS Device Manager - Overview
Using AMS Device Manager to configure the meter 83
Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D

AMS Device Manager - Guided Setup

The Guided setup wizard provides configuration parameter settings for the meter. The Guided Setup is a subset of the Manual Setup parameters.
Figure 4-5 AMS Device Manager - Guided Setup
Note: Before writing configuration changes to your meter, make sure you have saved
the Configuration file and Maintenance log.

Procedure

1. Disable the Write Protect switch in the CPU Module to write any of the following
configuration parameters to your meter.
2. Click the
Setup Units tab to configure the system units (U.S. Customary or Metric units),
Volume units, Flow rate time units, Velocity units, Pressure units and Temperature units. Click Apply to write the parameters to the meter.
3. Click the
Setup Outputs tab to configure the Device Variables Mapping, Units,
Frequency/Digital outputs, Frequency and Digital Outputs 1 and 2, Analog outputs, Digital Input, Pressure and Temperature.
a. Analog output 1 (HART) -
Rate and is a read only attribute). Configure
Upper range value and Alarm Action and view the HART Parameters Tag, Date,
Content (Primary Variable) displays Uncorrected Flow
Direction (flow), Lower Range value,
Descriptor, Message, Final Assembly Number Poll Address, Number of Response Preambles.
b. Analog Output 2 -
and has a read only attribute. Configure
Upper range value and Alarm Action. Map the Third and Four variables using the
Content (Secondary Variable) displays Uncorrected Flow Rate
Direction (flow), Lower Range value,
Manual Setup wizard. Selections include Uncorrected Volume Flow Rate, Pressure and Temperature.
84 Using AMS Device Manager to configure the meter
Loading...