Because of continuing product improvement, prices and specifications are subject to change without notice.
Changes to this manual, either in response to user input or to continuing product improvements, are
accomplished through reissue. If, in the normal use of this manual, errors, omissions, or incorrect data are
noted, please contact Edwards Technical Support or your local Edwards representative.
Edwards Technical Support
United States and Canada (24 hours) . . . . . . . . 800.822.9837 or tech_support@edwards.com
Outside the U.S. and Canada (24 hours). . . . . . 949.250.2222
The Edwards Lifesciences HemoSphere advanced monitor operator’s manual is comprised of eleven chapters,
eight appendices, and an index. Figures in this manual are intended for reference only and may not be an exact
replication of the screens as a result of continuous software improvements.
WARNINGRead this operator's manual carefully before attempting to use the Edwards
Lifesciences HemoSphere advanced monitor.
Refer to the instructions for use provided with each compatible accessory
before using it with the HemoSphere advanced monitor.
CAUTIONInspect all accessories and equipment for damage prior to use with the
HemoSphere advanced monitor. Damage may include cracks, scratches, dents,
exposed electrical contacts, or any signs that the housing may be compromised.
WARNINGTo prevent injury to patient or user, damage to platform, or inaccurate
measurements, do not use any damaged or non-compatible platform accessories, components or cables.
ChapterDescription
1Introduction: Provides an overview of the HemoSphere advanced monitor
2Safety and Symbols: Includes WARNINGS, CAUTIONS, and NOTES that are
found in the manual, as well as illustrations of labels found on the HemoSphere
advanced monitor and accessories
3Installation and Setup: Provides information about setting up the HemoSphere
advanced monitor and connections for the first time
and users of bedside monitors instructions for immediate monitor use
5Navigating the HemoSphere Advanced Monitor: Provides information on
monitoring screen views
6User Interface Settings: Provides information about the various display settings
including patient information, language and international units, alarm volume,
system time, and system date. It also provides instructions for selecting the
screen appearance
7Advanced Settings: Provides information on advanced settings including alarm
targets, graphical scales, serial port setup, and Demo Mode.
8Data Export and Connectivity: Provides information on monitor connectivity for
transferring patient and clinical data
9HemoSphere Swan-Ganz Module Monitoring: Describes procedures for setup
and operation of continuous cardiac output, intermittent cardiac output, and right
ventricular end diastolic volume monitoring using the Swan-Ganz module
10Oximetry Monitoring: Describes procedures for calibration and operation of
oximetry (oxygen saturation) measurement
11Help and Troubleshooting: Describes the Help menu and provides a list of
faults, alerts, and messages, with causes and suggested actions.
This manual describes the features and monitoring options of the Edwards Lifesciences HemoSphere
advanced monitor. The HemoSphere advanced monitor is a modular device that displays monitored data
obtained through Edwards hemodynamic technologies.
1
This manual has been prepared for use with the Edwards Lifesciences HemoSphere advanced monitor by
trained critical care clinicians, nurses, and physicians in any hospital environment where critical care is
administered.
This manual provides the operator of the HemoSphere advanced monitor with setup and operating
instructions, device interfacing procedures, and limitations.
1.2Indications For Use
1.2.1HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module
The HemoSphere advanced monitor when used with the HemoSphere Swan-Ganz module and Edwards
Swan-Ganz catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of
cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital
environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on
target patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for
each patient population.
15
Page 16
HemoSphere Advanced Monitor1 Introduction
1.2.2HemoSphere Advanced Monitor with HemoSphere Oximetry Cable
The HemoSphere advanced monitor when used with the HemoSphere oximetry cable and Edwards
oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of
venous oxygen saturation (SvO
and ScvO2) and derived hemodynamic parameters in a hospital
2
environment. Refer to the Edwards oximetry catheter indications for use statement for information on target
patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for
each patient population.
1.3Contraindications For Use
The HemoSphere advanced monitor has no contraindications for use.
1.4Intended Use Statement
The HemoSphere advanced monitoring platform is intended to be used by qualified personnel or trained
clinicians in a critical care environment in a hospital setting.
The HemoSphere advanced monitoring platform is intended for use with compatible Edwards Swan-Ganz
and oximetry catheters.
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and
a connected HemoSphere Swan-Ganz module are listed below in table 1-1. Only iCO, iCI, iSVR, and iSVRI
are available to the pediatric patient population.
Table 1-1 HemoSphere Swan-Ganz module available parameters list
AbbreviationDescriptionPatient population
COcontinuous cardiac output
sCOSTAT cardiac output
CIcontinuous cardiac index
sCISTAT cardiac index
EDVright ventricular end diastolic volume
sEDVSTAT right ventricular end diastolic volume
EDVIright ventricular end diastolic volume index
sEDVISTAT right ventricular end diastolic volume index
HR
avg
LVSWIleft ventricular stroke work index
PVRpulmonary vascular resistance
PVRIpulmonary vascular resistance index
RVEFright ventricular ejection fraction
sRVEFSTAT right ventricular ejection fraction
RVSWIright ventricular stroke work index
SVstroke volume
SVIstroke volume index
SVRsystemic vascular resistance
SVRIsystemic vascular resistance index
averaged heart rate
adult only
16
Page 17
HemoSphere Advanced Monitor1 Introduction
Table 1-1 HemoSphere Swan-Ganz module available parameters list (continued)
AbbreviationDescriptionPatient population
iCOintermittent cardiac output
iCIintermittent cardiac index
iSVRintermittent systemic vascular resistance
iSVRIintermittent systemic vascular resistance index
adult and pediatric
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring
with the HemoSphere advanced monitor and a connected HemoSphere oximetry cable are listed below in
table 1-2.
Table 1-2 HemoSphere oximetry cable available parameters list
AbbreviationDescriptionPatient population
SvO
ScvO
2
2
mixed venous oxygen saturation
central venous oxygen saturation
adult and pediatric
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring
with the HemoSphere advanced monitor and both a connected HemoSphere Swan-Ganz module and
oximetry cable are listed below in table 1-3.
Table 1-3 HemoSphere Swan-Ganz module with oximetry cable available parameters
list
AbbreviationDescriptionPatient population
DO
2
DO
Ioxygen delivery index
2
VO
2
VO
eestimated oxygen consumption when ScvO2 is
2
Ioxygen consumption index
VO
2
VO
Ieestimated oxygen consumption index when
2
oxygen delivery
oxygen consumption
being monitored
is being monitored
ScvO
2
adult and pediatric
WARNINGImproper use of the HemoSphere advanced monitor could present a hazard to the patient.
Carefully read the “warnings” section of this manual, located in chapter 2, before using the
platform.
The HemoSphere advanced monitor is intended for use only in patient assessment. This
instrument must be used in conjunction with a bedside physiological monitor and/or
patient clinical signs and symptoms. If hemodynamic values obtained from the device are
not consistent with the clinical presentation of the patient, consider troubleshooting before
initiating treatment options.
ECG signal input and all parameters derived from heart rate measurements have not been
evaluated for pediatric patients and are therefore not available for that patient population.
The HemoSphere advanced monitor is equipped with three technology expansion module slots (two
standard size and one large [L-Tech] size) and two cable ports. Module and cable connection points are
located on the left side panel. See figure 1-1.
Each module/cable is associated with a specific Edwards hemodynamic monitoring technology. Currently
available modules include the HemoSphere Swan-Ganz module, introduced below and in detail in chapter
9, HemoSphere Swan-Ganz Module Monitoring. Currently available cables include the HemoSphere oximetry
cable, introduced below and described in detail in chapter 10, Oximetry Monitoring.
1.5.1HemoSphere Swan-Ganz Module
The HemoSphere Swan-Ganz module enables continuous
cardiac output (CO) and intermittent cardiac output (iCO)
monitoring with an Edwards patient CCO cable and compatible
Swan-Ganz catheter. Right ventricular end diastolic volume
(EDV) monitoring is available with slaved in heart rate (HR
data from a bedside patient monitor. The HemoSphere Swan-Ganz module fits into a standard module slot.
For more information, see chapter 9, HemoSphere Swan-Ganz Module Monitoring. Table 1-4 lists the parameters
available while using the HemoSphere Swan-Ganz module.
stroke volume (SV)amount of blood ejected from the ventricles
stroke volume index (SVI)stroke volume relative to body surface area
systemic vascular
resistance (SVR)
systemic vascular
resistance index (SVRI)
intermittent assessment through the bolus
thermodilution method of the volume of
blood pumped by the heart measured in
liters per minute
intermittent cardiac output relative to body
surface area (BSA)
continuous assessment through advanced
thermodilution technology and algorithm
analysis of the percentage of blood volume
ejected from the right ventricle during
systole
continuous assessment of the volume of
blood in the right ventricle at the end of
diastole calculated by dividing stroke
volume (mL/beat) by RVEF(%)
with each contraction derived from CO
assessment and heart rate (SV = CO/HR x
1000)
(BSA)
a derived measure of impedance to blood
flow from left ventricle (afterload)
systemic vascular resistance relative to
body surface area (BSA)
Swan-Ganz thermodilution catheters
Swan-Ganz thermodilution catheters
Swan-Ganz CCOmbo V catheters
with ECG signal input
Swan-Ganz CCOmbo V catheters
with ECG signal input
Swan-Ganz CCO,CCOmbo, and
CCOmbo V catheters with ECG
signal input
Swan-Ganz CCO,CCOmbo, and
CCOmbo V catheters with ECG
signal input
Swan-Ganz CCO and CCOmbo
catheters with MAP and CVP analog
pressure signal input
Swan-Ganz CCO and CCOmbo
catheters with MAP and CVP analog
pressure signal input
catheter. The HemoSphere oximetry cable plugs into a
monitoring cable receptacle and can be used in combination
with other hemodynamic monitoring technologies. For more
information on oximetry monitoring, see chapter 10, Oximetry Monitoring. Table 1-5 lists the parameters available while using
the HemoSphere oximetry cable.
an estimate of the amount of oxygen used by the body per
minute (ScvO2 monitoring only)
the amount of oxygen used by the body per minute indexed
against body surface area (BSA)
an estimate of the amount of oxygen used by the body per
minute indexed against body surface area (BSA)
1.5.3Documentation and Training
Available documentation and training for the HemoSphere advanced monitor includes:
•HemoSphere Advanced Monitor Operator’s Manual
•HemoSphere Advanced Monitor Quick Start Guide
•HemoSphere Swan-Ganz Module Instructions for Use
•HemoSphere Oximetry Cable Instructions for Use
Instructions for Use are included with HemoSphere Advanced Monitor components. See table B-1,
“HemoSphere advanced monitor components,” on page 137. For more information on how you can receive
training or available documentation for the HemoSphere advanced monitor, contact your local Edwards
representative or Edwards Technical Support. See appendix F, System Care, Service and Support..
1.6Manual style conventions
Table 1-6 lists the style conventions used in this manual.
Table 1-6 Operator’s manual style conventions
ConventionDescription
Bold Bold text indicates a software term. This word or phrase will appear on the
screen as shown.
Bold buttonA button is a touch screen access point for the option appearing in bold.
For example, the Review button appears on screen as:
Oximetry Calibration icon
An arrow is shown between two on screen menu options that are selected
consecutively by the operator.
An icon is a touch screen access point for the menu or navigation graphic
shown. See table 2-1 on page 29 for full list of menu icons shown on the
HemoSphere advanced monitor.
Bold text with a menu icon indicates an icon that is paired with a software
term appearing on the screen as shown.
20
Page 21
HemoSphere Advanced Monitor1 Introduction
1.7Abbreviations Found in This Manual
Table 1-7 Acronyms, Abbreviations
AbbreviationDefinition
A/Danalog/digital
BSAbody surface area
BTblood temperature
arterial oxygen content
CaO
2
CIcardiac index
COcardiac output
CCOcontinuous cardiac output
CVPcentral venous pressure
DO
2
Ioxygen delivery index
DO
2
DPTdisposable pressure transducer
EDVend diastolic volume
EDVIend diastolic volume index
efuejection fraction unit
Hcthematocrit
HIShospital information systems
HGBhemoglobin
HRheart rate
HR
avg
iCOintermittent cardiac output
IECInternational Electrotechnical
ITinjectate temperature
LEDlight emitting diode
LVSWIleft ventricular stroke work index
MAPmean arterial pressure
MPAPmean pulmonary artery pressure
PApulmonary artery
PaO
2
PAWPpulmonary artery wedge pressure
POSTpower-on self test
PvO
2
RVEFright ventricular ejection fraction
RVSWIright ventricular stroke work index
sCISTAT cardiac index
sCOSTAT cardiac output
ScvO
2
sEDVSTAT end diastolic volume
sEDVISTAT end diastolic volume index
SpO
2
SQIsignal quality indicator
(used when describing certain SwanGanz catheters and patient CCO cable)
A warning advises against certain actions or situations that could result in personal injury or death.
WARNINGThis is how warnings appear throughout the text of this manual.
2.1.2Caution
A caution advises against actions or situations that could damage equipment, produce inaccurate data, or
invalidate a procedure.
CAUTIONThis is how cautions appear throughout the text of this manual.
2.1.3Note
A note draws attention to useful information regarding a function or procedure.
NOTEThis is how notes appear throughout the text of this manual.
22
Page 23
HemoSphere Advanced Monitor2 Safety and Symbols
2.2 Warnings
The following are warnings that are used in the HemoSphere advanced monitor operator’s manual. They are
introduced in the manual where relevant to the function or procedure being described.
•Read this operator's manual carefully before attempting to use the Edwards Lifesciences
HemoSphere advanced monitor.
•Refer to the instructions for use provided with each compatible accessory before using it with the
HemoSphere advanced monitor.
•To prevent injury to patient or user, damage to platform, or inaccurate measurements, do not use
any damaged or non-compatible platform accessories, components or cables.
•Improper use of the HemoSphere advanced monitor could present a hazard to the patient.
Carefully read the “warnings” section of this manual, located in chapter 2, before using the
platform. (chapter 1)
•The HemoSphere advanced monitor is intended for use only in patient assessment. This instrument
must be used in conjunction with a bedside physiological monitor and/or patient clinical signs and
symptoms. If hemodynamic values obtained from the device are not consistent with the clinical
presentation of the patient, consider troubleshooting before initiating treatment options. (chapter 1)
•ECG signal input and all parameters derived from heart rate measurements have not been evaluated
for pediatric patients and are therefore not available for that patient population. (chapter 1)
•Shock hazard: Do not attempt to connect/disconnect system cables while hands are wet. Ensure
that hands are dry prior to disconnecting system cables. (chapter 3)
•Explosion Hazard! Do not use the HemoSphere advanced monitor in the presence of flammable
anesthetic mixture with air or with oxygen or nitrous oxide. (chapter 3)
•Make sure the HemoSphere advanced monitor is securely positioned or mounted and that all cords
and accessory cables are appropriately arranged to minimize the risk of injury to patients, users or
the equipment. (chapter 3)
•Do not stack additional equipment or items on top of the HemoSphere advanced monitor. (chapter
3)
•The HemoSphere advanced monitor must be positioned in an upright position to ensure IPX1
ingress protection. (chapter 3)
•Do not allow any liquids to splash onto the monitoring screen. Liquid buildup may disable the
touchscreen functionality. (chapter 3)
•Do not position the monitor so that it is difficult to access rear panel ports or power cord. (chapter
3)
•Equipment may be used in the presence of electrosurgery and defibrillators. Inaccurate parameter
measurements can be caused by factors such as electrocautery or electrosurgery unit interference.
(chapter 3)
•All IEC/EN 60950 equipment, including printers, to be positioned no closer than 1.5 meters to the
patient’s bed. (chapter 3)
23
Page 24
HemoSphere Advanced Monitor2 Safety and Symbols
•Make sure the battery is fully inserted and the battery door is properly latched. Falling batteries
could seriously injure patients or clinicians. (chapter 3)
•Only use Edwards approved batteries with the HemoSphere advanced monitor. Do not charge the
battery pack outside of the monitor. Doing so can damage the battery or injure the user. (chapter 3)
•To prevent any interruptions to monitoring during power loss, it is recommended to use the
HemoSphere advanced monitor with the battery inserted. (chapter 3)
•In cases of power failure and battery depletion, the monitor will go through a controlled shut off
procedure. (chapter 3)
•Do not use the HemoSphere advanced monitoring platform without an installed power cord entry
cover. Failure to do so may result in fluid ingress. (chapter 3)
•Do not use extension cords or multiple socket devices to connect the power cord. Do not use
detachable power cords other than the power cord provided. (chapter 3)
•To avoid the risk of electric shock, the HemoSphere advanced monitor can only be connected to a
supply mains with grounding (protective earth). Do not use three prong to two prong power
adaptors. (chapter 3)
•Grounding reliability can only be achieved when the instrument is connected to a receptacle marked
“hospital only”, “hospital grade”, or its equivalent. (chapter 3)
•Disconnect the monitor from the AC source by unplugging mains power cable from the AC Mains.
The On/Off button on the monitor does not disconnect the system from the AC mains supply.
(chapter 3)
•Only use HemoSphere advanced monitor accessories, cables and or components that have been
supplied and labeled by Edwards. Using other unlabeled accessories, cables and or components may
affect patient safety and measurement accuracy. (chapter 3)
•Upon initiation of a new patient session, the default high/low physiological alarm ranges should be
checked to ensure that they are appropriate for the given patient. (chapter 6)
•Perform New Patient or clear the patient data profile whenever a new patient is connected to the
HemoSphere advanced monitor. Failure to do so may result in previous patient data in the historical
displays. (chapter 6)
•The analog communication ports of the HemoSphere advanced monitor share a common ground
that is isolated from the catheter interface electronics. When connecting multiple devices to the
HemoSphere advanced monitor, all devices should be provided with isolated power to avoid
compromising the electrical isolation of any of the connected devices. (chapter 6)
•Risk and Leakage current of the final system configuration must comply with IEC 60601-1:2005/
A1:2012. It is the responsibility of the user to ensure compliance. (chapter 6)
•Accessory equipment connected to the monitor must be certified according to IEC/EN 60950 for
data-processing equipment or IEC 60601-1:2005/A1:2012 for electromedical equipment. All
combinations of equipment must be in compliance with IEC 60601-1:2005/A1:2012 systems
requirements. (chapter 6)
•When switching to a different bedside monitor, always check that the default values listed are still
valid. If necessary, reconfigure the voltage range and corresponding parameter range or calibrate.
(chapter 6)
24
Page 25
HemoSphere Advanced Monitor2 Safety and Symbols
•Do not turn off the audible alarms in situations in which patient safety could be compromised.
(chapter 7)
•Do not lower the alarm volume to a level that prohibits alarms from being adequately monitored.
Failure to do so could result in a situation where patient safety is compromised. (chapter 7)
•Visual and audible physiological alarms are activated only if the parameter is configured on the
screens as a key parameter (1-4 parameters displayed in parameter globes). If a parameter is not
selected and displayed as a key parameter, the audible and visual physiological alarms are not
triggered for that parameter. (chapter 7)
•Make sure that Demo Mode is not activated in a clinical setting to ensure that simulated data is not
mistaken for clinical data. (chapter 7)
•Do not use the HemoSphere advanced monitor as part of a Distributed Alarm System. The
HemoSphere advanced monitor does not support remote alarm monitoring/management systems.
Data is logged and transmitted for charting purposes only. (chapter 8)
•CO monitoring should always be discontinued when blood flow around the thermal filament is
stopped. Clinical situations where CO monitoring should be discontinued include, but are not
limited to: • Time periods when a patient is on cardiopulmonary bypass • Partial withdrawal of the
catheter so that the thermistor is not in the pulmonary artery • Removal of the catheter from the
patient (chapter 9)
•PACEMAKER PATIENTS – Rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon displayed heart rate.
Keep pacemaker patients under close surveillance. See table A-5 on page 134 for disclosure of the
pacemaker pulse rejection capability of this instrument. (chapter 9)
•
For patients requiring internal or external pacing support, the
HemoSphere advanced monitoring
platform should not be used to obtain heart rate and heart rate derived parameters under the
following conditions:
• pacer pulse synch output from bedside monitor includes the pacer pulse, however, the
characteristics are outside of the pacemaker pulse rejection capabilities specifications as listed in
table A-5.
• pacer pulse synch output characteristics from bedside monitor cannot be determined (chapter 9)
Note any discrepancies in heart rate (HRavg) with the patient monitor HR and ECG waveform
•
display when interpreting derived parameters such as SV, EDV, RVEF, and associated index
parameters. (chapter 9)
•Do not wrap the main body of the oximetry cable in fabric or place directly on the patient's skin
for long periods of time (>10 min). The surface does get warm (up to 45 °C) and needs to dissipate
heat to maintain its internal temperature level. A software fault will trigger if the internal
temperature exceeds its limits. (chapter 10)
•Before touching Yes to recall oximetry data, confirm that the displayed data matches the current
patient. Recalling incorrect oximetry calibration data and patient demographics will result in
inaccurate measurements. (chapter 10)
•Only use approved HemoSphere advanced monitor accessories, cables and or components that
have been supplied and labeled by Edwards. Using unapproved accessories, cables and or
components may affect patient safety and measurement accuracy. (appendix B)
25
Page 26
HemoSphere Advanced Monitor2 Safety and Symbols
•The HemoSphere advanced monitor contains no user-serviceable parts. Removing the cover or any
other disassembly will expose you to hazardous voltages. (appendix F)
•Shock or fire hazard! Do not immerse the HemoSphere advanced monitor, modules, or platform
cables in any liquid solution. Do not allow any fluids to enter the instrument. (appendix F)
•Explosion Hazard! Do not open battery, dispose of in fire, store at high temperature or short
circuit. It may ignite, explode, leak or get hot, causing serious personal injury or death. (appendix F)
•Use of accessories, sensors, and cables other than those specified may result in increased
electromagnetic emissions or decreased electromagnetic immunity. (appendix G)
•No modification of the HemoSphere advanced monitor is allowed. (appendix G)
•Portable and mobile RF communication equipment can potentially affect all electronic medical
equipment, including the HemoSphere advanced monitor. Guidance on maintaining appropriate
separation between communications equipment and the HemoSphere advanced monitor is
provided in table G-4. (appendix G)
2.3 Cautions
The following are cautions that are used in the HemoSphere advanced monitor operator’s manual. They are
introduced in the manual where relevant to the function or procedure being described.
•Federal (USA) law restricts this device to sale by or on the order of a physician.
•Inspect all accessories and equipment for damage prior to use with the HemoSphere advanced
monitor. Damage may include cracks, scratches, dents, exposed electrical contacts, or any signs that
the housing may be compromised.
•Always grasp the connector, not the cable, when connecting or disconnecting cables. Do not twist
or bend the connectors. Confirm that all sensors and cables are connected correctly and completely
before use. (chapter 3)
•To avoid corruption of data on the HemoSphere advanced monitor, always disconnect the patient
CCO cable and oximetry cable from the monitor before using a defibrillator. (chapter 3)
•Do not expose the HemoSphere advanced monitor to extreme temperatures. Refer to
environmental specifications in appendix A. (chapter 3)
•Do not expose the HemoSphere advanced monitor to dirty or dusty environments. (chapter 3)
•Do not obstruct the HemoSphere advanced monitor ventilation openings. (chapter 3)
•Do not use the HemoSphere advanced monitor in environments where strong lighting makes the
LCD screen difficult to view. (chapter 3)
•Do not use the monitor as a handheld device. (chapter 3)
•When moving the instrument, be sure to turn off the power and remove the connected power cord.
(chapter 3)
•When connecting the HemoSphere advanced monitor to external devices, refer to the external
device’s instruction manual for complete instructions. Verify proper operation of the system before
clinical use. (chapter 6)
26
Page 27
HemoSphere Advanced Monitor2 Safety and Symbols
•Only properly trained personnel should calibrate the HemoSphere advanced monitor analog ports.
(chapter 6)
•The accuracy of continuous SVR depends upon the quality and accuracy of the MAP and CVP data
transmitted from the external monitors. Since MAP and CVP analog signal quality from the
external monitor cannot be validated by the HemoSphere advanced monitor, actual values and the
values (including all derived parameters) displayed by the HemoSphere advanced monitor may not
be consistent. The accuracy of continuous SVR measurement, therefore, cannot be guaranteed. To
aid in determining the quality of the analog signals, regularly compare the MAP and CVP values
displayed on the external monitor to the values displayed on the physio relationship screen of the
HemoSphere advanced monitor. Refer to the external input device operator’s manual for detailed
information regarding accuracy, calibration, and other variables which may impact the analog
output signal from the external monitor. (chapter 6)
•Use a virus scan on any USB stick before inserting to prevent a virus or malware infection. (chapter
8)
•Restore Defaults replaces all settings with factory defaults. Any settings changes or customizations
will be permanently lost. Do not restore defaults while monitoring a patient. (chapter 8)
•Do not force the module into the slot. Apply even pressure to slide and click the module into place.
(chapter 9)
•Inaccurate cardiac output measurements may be caused by: • Incorrect placement or position of
the catheter • Excessive variations in pulmonary artery blood temperature. Some examples that
cause BT variations include, but are not limited to: * status post cardiopulmonary bypass surgery *
centrally administered cooled or warmed solutions of blood products * use of sequential
compression devices • Clot formation on the thermistor • Anatomical abnormalities (for example,
cardiac shunts) • Excessive patient movement • Electrocautery or electrosurgical unit interference
• Rapid changes in cardiac output (chapter 9)
•Refer to Appendix E to ensure computation constant is the same as specified in the catheter
package insert. If the computation constant differs, enter the desired computation constant
manually. (chapter 9)
•Sudden changes in PA blood temperature, such as those caused by patient movement or bolus drug
administration, may cause an iCO or iCI value to be computed. To avoid falsely triggered curves,
inject as soon as possible after the Inject message appears. (chapter 9)
•Make sure that the oximetry cable is securely stabilized to prevent unnecessary movement of the
attached catheter. (chapter 10)
•The catheter and the calibration cup must be dry for an accurate oximetry in vitro calibration. Flush
the catheter lumen only after the in vitro calibration has been completed. (chapter 10)
•Performing an in vitro calibration after the oximetry catheter has been inserted into the patient will
yield an inaccurate calibration. (chapter 10)
•The SQI signal is sometimes affected by the use of electrosurgical units. Attempt to distance
electrocautery equipment and cables from the HemoSphere advanced monitor and plug the power
cords into separate AC circuits if possible. If signal quality problems persist, call your local Edwards
representative for assistance. (chapter 10)
•Do not disconnect the oximetry cable while calibration or data recall are in process. (chapter 10)
27
Page 28
HemoSphere Advanced Monitor2 Safety and Symbols
•If the oximetry cable is being transferred from a HemoSphere advanced monitor to another
HemoSphere advanced monitor, check that the patient height, weight, and BSA are correct prior to
beginning monitoring. Re-enter patient data, if necessary. (chapter 10)
•Clean and store the instrument and accessories after each use. (appendix F)
•Do not pour or spray liquid on any portion of the HemoSphere advanced monitor, accessories,
modules, or cables. (appendix F)
•Do not use any disinfecting solution other than the types specified. (appendix F)
•DO NOT: Allow any liquid to come in contact with the power connector Allow any liquid to
penetrate connectors or openings in the monitor case or modules If any liquid does come in contact
with any of the above mentioned items, DO NOT attempt to operate the monitor. Disconnect
power immediately and call your Biomedical Department or local Edwards representative.
(appendix F)
•Conduct periodic inspections of all cables for defects. Do not coil cables tightly when storing.
(appendix F)
•Do not steam, radiate, or EO sterilize the HemoSphere oximetry cable. Do not immerse the
HemoSphere oximetry cable. (appendix F)
•If any electrolytic solution, for example Ringer’s lactate solution, is introduced into the cable
connectors while they are connected to the monitor, and the monitor is turned on, the excitation
voltage can cause electrolytic corrosion and rapid degradation of the electrical contacts. (appendix
F)
•Do not immerse any cable connectors in detergent, isopropyl alcohol or glutaraldehyde. (appendix
F)
•Do not use a hot air gun to dry cable connectors. (appendix F)
•Recycle or dispose of the lithium-ion battery in accordance to all federal, state, and local laws.
(appendix F)
•The instrument has been tested and complies with the limits of IEC 60601-1-2. These limits are
designed to provide reasonable protection against harmful interference in a typical medical
installation. This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference to other
devices in the vicinity. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to other devices which can be
determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures: · Reorient or relocate the receiving device.
· Increase the separation between the equipment. · Consult the manufacturer for help. (appendix G)
28
Page 29
HemoSphere Advanced Monitor2 Safety and Symbols
2.4 User Interface Symbols
The following are icons that appear on the HemoSphere advanced monitor screen. For more information
about screen appearance and navigation, see chapter 5, Navigating the HemoSphere Advanced Monitor. Certain
icons will only appear while monitoring with a specific hemodynamic technology module or cable, as
specified.
.
Table 2-1 Monitor display symbols
SymbolDescription
Navigation Bar Icons
begin CO monitoring
(HemoSphere Swan-Ganz module)
stop CO monitoring with CO countdown timer
(see CO Countdown Timer and STAT CO on
page 100)
(HemoSphere Swan-Ganz module)
monitor screen selection
clinical actions menu
settings menu
snapshot (screen capture)
silence audible alarms
Table 2-1 Monitor display symbols (continued)
SymbolDescription
event review
patient CCO cable test
(HemoSphere Swan-Ganz module)
Menu Navigation Icons
return to main monitoring screen
return to previous menu
cancel
scroll to select item on vertical list
vertical page scroll
horizontal scroll
alarms paused (silenced) with countdown
timer (See Silence Audible Alarms on page
The platform shall provide display of continuous CO and intermittent CO with a compatible Swan-Ganz
catheter according to the specifications provided in appendix A. The platform shall provide display of SvO
ScvO
with a compatible oximetry catheter according to the specifications provided in appendix A. The
2
platform shall provide alarm, alert, indicator, and/or system status when unable to provide accurate
measurement of the applicable hemodynamic parameter. For more information, see Essential Performance Characteristics on page 132.
Examine the shipping container for any signs of damage that may have occurred during transit. If any
damage is detected, photograph the package and contact Edwards technical support for assistance. Perform
a visual inspection of the packaging contents for damage. Damage may include cracks, scratches, dents or
any signs that the monitor, modules or cable housing may be compromised. Report any evidence of external
damage.
3
3.1.1Packaging Contents
The HemoSphere advanced monitoring platform is modular and therefore packaging configurations will
vary depending upon the kit ordered. The HemoSphere advanced monitoring system, which is the base kit
configuration, contains the HemoSphere advanced monitor, mains power cord, power entry cover,
HemoSphere battery pack, two expansion modules, one L-Tech expansion module, a quick start guide and
a USB stick containing this operator’s manual. See table 3-1. Additional items that may be included and
shipped with other kit configurations include the HemoSphere Swan-Ganz module, patient CCO cable, and
HemoSphere oximetry cable. Disposable and accessory items may be delivered separately. It is
recommended that the user confirm the receipt of all ordered equipment. Refer to appendix B: Accessories,
for a full list of available accessories.
HemoSphere Advanced Monitor3 Installation and Setup
3.1.2Required Accessories for Platform Modules and Cables
The following tables identify accessories required to display specific monitored and calculated parameters
for the specified hemodynamic technology module or cable:
Table 3-2 Accessories required for monitoring parameters with HemoSphere Swan-
Ganz module
Monitored and calculated parameters
Required AccessoryCOEDVRVEFSVRiCOSV
patient CCO cable
ECG cable
analog pressure input cable(s)
injectate temperature probe
Swan-Ganz thermodilution catheter
Swan-Ganz CCO catheter or SwanGanz CCOmbo catheter
Swan-Ganz CCOmbo V catheter
••••••
•••
•
•
•
••••
••••••
NOTENot all parameters can be monitored or calculated in pediatric patients. See table 1-1 on
page 16 for available parameters.
Table 3-3 Accessories required for monitoring parameters with HemoSphere oximetry
cable
Monitored and
calculated
parameters
Required AccessoryScvO
PediaSat oximetry catheter or compatible
central venous oximetry catheter
Swan-Ganz oximetry catheter
•
SvO
2
2
•
WARNINGShock hazard: Do not attempt to connect/disconnect system cables while hands are
wet. Ensure that hands are dry prior to disconnecting system cables.
CAUTIONAlways grasp the connector, not the cable, when connecting or disconnecting cables.
Do not twist or bend the connectors. Confirm that all sensors and cables are connected
correctly and completely before use.
To avoid corruption of data on the HemoSphere advanced monitor, always disconnect
the patient CCO cable and oximetry cable from the monitor before using a defibrillator.
34
Page 35
HemoSphere Advanced Monitor3 Installation and Setup
visual alarm indicator
power button
3.2 HemoSphere Advanced Monitor Connection Ports
The following monitor views illustrate the connection ports and other key features of the front, rear, and
side panels of the HemoSphere advanced monitor.
3.2.1Monitor Front
Figure 3-1 HemoSphere advanced monitor front view
35
Page 36
HemoSphere Advanced Monitor3 Installation and Setup
Figure 3-3 HemoSphere advanced monitor right panel
3.2.4Monitor Left Panel
Figure 3-4 HemoSphere advanced monitor left panel (shown with no
modules)
37
Page 38
HemoSphere Advanced Monitor3 Installation and Setup
3.3 HemoSphere Advanced Monitor Installation
3.3.1Mounting Options and Recommendations
The HemoSphere advanced monitor should be placed on a stable flat surface or securely mounted on a
compatible stand, according to your institution’s practices. A roll stand for the HemoSphere advanced
monitor is available as an optional accessory. See Additional Accessories Description on page 138 more
information. Contact your local Edwards representative for recommendations on additional mounting
options.
WARNINGExplosion Hazard! Do not use the HemoSphere advanced monitor in the presence
of flammable anesthetic mixture with air or with oxygen or nitrous oxide.
Make sure the HemoSphere advanced monitor is securely positioned or mounted and
that all cords and accessory cables are appropriately arranged to minimize the risk of
injury to patients, users or the equipment.
Do not stack additional equipment or items on top of the HemoSphere advanced
monitor.
The HemoSphere advanced monitor must be positioned in an upright position to
ensure IPX1 ingress protection.
Do not allow any liquids to splash onto the monitoring screen. Liquid buildup may
disable the touchscreen functionality.
Do not position the monitor so that it is difficult to access rear panel ports or power
cord.
Equipment may be used in the presence of electrosurgery and defibrillators. Inaccurate
parameter measurements can be caused by factors such as electrocautery or electrosurgery unit interference.
All IEC/EN 60950 equipment, including printers, to be positioned no closer than 1.5
meters to the patient’s bed.
CAUTIONDo not expose the HemoSphere advanced monitor to extreme temperatures. Refer to
environmental specifications in appendix A.
Do not expose the HemoSphere advanced monitor to dirty or dusty environments.
Do not obstruct the HemoSphere advanced monitor ventilation openings.
Do not use the HemoSphere advanced monitor in environments where strong lighting
makes the LCD screen difficult to view.
Do not use the monitor as a handheld device.
38
Page 39
HemoSphere Advanced Monitor3 Installation and Setup
3.3.2Battery Installation
Open the battery door (figure 3-3) and insert the battery into the battery bay, ensuring pack is fully inserted
and seated. Close the battery door and ensure that the latch is securely fastened. Follow instructions below
to connect the power cord and then fully charge the battery. Do not use a new battery pack as a power source
until it has been fully charged.
NOTETo ensure that the battery charge level displayed on the monitor is accurate, please
condition the battery before first use. For information on battery maintenance and
conditioning, see Battery Maintenance on page 155.
The HemoSphere battery pack is intended as a backup power source during power-loss
and can only support monitoring for a limited time period.
WARNINGMake sure the battery is fully inserted and the battery door is properly latched. Falling
batteries could seriously injure patients or clinicians.
Only use Edwards approved batteries with the HemoSphere advanced monitor. Do not
charge the battery pack outside of the monitor. Doing so can damage the battery or
injure the user.
To prevent any interruptions to monitoring during power loss, it is recommended to
use the HemoSphere advanced monitor with the battery inserted.
In cases of power failure and battery depletion, the monitor will go through a controlled
shut off procedure.
3.3.3Connecting Power Cord
Before connecting the power cord to the rear panel of the monitor, ensure that the power entry cover is
installed:
1 If the power entry cover is already installed, remove the two screws (figure 3-5) that attach the
power entry cover to the rear panel of the monitor.
2 Connect the detachable power supply cord. Ensure that the plug is seated securely.
3 Attach the power cord entry cover over the plug by routing the power cord through the cover
opening and then pressing the cover and gasket up against the rear panel of the monitor, aligning
the two screw holes.
4 Reinsert the screws to fasten the cover onto the monitor.
5 Plug power cord into a hospital grade outlet.
WARNINGDo not use the HemoSphere advanced monitoring platform without an installed power
cord entry cover. Failure to do so may result in fluid ingress.
39
Page 40
HemoSphere Advanced Monitor3 Installation and Setup
Figure 3-5 HemoSphere power entry cover - screw locations
3.3.3.1 Equipotential Connection
This monitor MUST be grounded during operation (Class I equipment according to IEC 60601-1). If a
hospital grade or three-prong receptacle is not available, a hospital electrician must be consulted to ensure
proper grounding. An equipotential terminal is provided on the rear panel of the monitor (figure 3-2) to be
connected to an equipotential grounding system (equipotential cable).
WARNINGDo not use extension cords or multiple socket devices to connect the power cord. Do
not use detachable power cords other than the power cord provided.
To avoid the risk of electric shock, the HemoSphere advanced monitor can only be
connected to a supply mains with grounding (protective earth). Do not use three prong
to two prong power adaptors.
Grounding reliability can only be achieved when the instrument is connected to a receptacle marked “hospital only”, “hospital grade”, or its equivalent.
Disconnect the monitor from the AC source by unplugging mains power cable from
the AC Mains. The On/Off button on the monitor does not disconnect the system
from the AC mains supply.
CAUTIONWhen moving the instrument, be sure to turn off the power and remove the connected
power cord.
3.3.4Connecting and Disconnecting a Hemodynamic Monitoring Module
The HemoSphere advanced monitor is shipped with two standard expansion modules and one L-Tech
expansion module. Before inserting a new monitoring technology module, remove the expansion module by
pressing the release button to unlatch and slide the blank module out.
Inspect the new module for external damage before installation. Insert the desired monitoring module into
the open slot by applying even pressure to slide and click the module into place.
3.3.5Connecting and Disconnecting a Hemodynamic Monitoring Cable
Both monitoring cable ports are equipped with a magnetic latch mechanism. Inspect the cable for damage
before connecting. A monitoring cable will snap into place when it is properly seated in the port. To
disconnect a cable, hold at the plug to pull it away from the monitor.
40
Page 41
HemoSphere Advanced Monitor3 Installation and Setup
3.3.6Connecting Cables from External Devices
The HemoSphere advanced monitor utilizes slaved-in monitored data to calculate certain hemodynamic
parameters. This includes data from the pressure input data ports and ECG monitor input port. All slavedin cable connections are located on the rear panel of the monitor (figure 3-2). See Required Accessories for Platform Modules and Cables on page 34 for a list of calculated parameters available with certain cable
connections. For more information on configuring the analog pressure ports, see Analog Pressure Signal Input
on page 75.
IMPORTANT NOTEThe HemoSphere advanced monitor is compatible with pressure
and ECG analog slave inputs from any external patient monitor that
has analog slave output ports which meet the signal input specifications identified in appendix A, table A-5 of this operator’s manual.
These provide a convenient means to utilize information from a
patient monitor to calculate additional hemodynamic parameters
for display. This is an optional feature that does not impact the
HemoSphere advanced monitor’s primary function of monitoring
cardiac output (with the HemoSphere Swan-Ganz module) and
venous oxygen saturation (with the HemoSphere oximetry cable).
WARNINGOnly use HemoSphere advanced monitor accessories, cables and or components that
have been supplied and labeled by Edwards. Using other unlabeled accessories, cables
and or components may affect patient safety and measurement accuracy.
3.4 Initial Start Up
3.4.1Start Up Procedure
To turn on and off the monitor, press the power button located on the front panel. After turning on the
monitor, the Edwards screen is displayed followed by the Power-On Self Test (POST) screen. The POST
verifies the monitor meets basic operating requirements by exercising critical hardware components and is
performed each time the system is turned on. POST status message is displayed on the startup screen along
with system information such as serial numbers and software version numbers.
41
Page 42
HemoSphere Advanced Monitor3 Installation and Setup
Figure 3-6 Startup screen
NOTEIf the diagnostic tests detect an error condition, a system error screen will replace the
startup screen. See chapter 11: Troubleshooting or appendix F: System Care, Service and Support. Otherwise, call your Edwards Lifesciences representative for assistance.
3.4.2Select Language
Upon initial HemoSphere advanced monitor startup, language options are offered which affect the displayed
language, time and date formats, and units of measurement. The language selection screen appears after the
software has initialized and POST is complete. Selecting the language also sets the display units and the time
and date format to the default settings for that language (see appendix D: Monitor Settings and Defaults).
Each of the language-related settings can be changed later in the Date/Time screen of the Monitor Settings screen and in the language option through Monitor Settings
When the language selection screen appears, touch the desired language for use.
General.
Figure 3-7 Language selection screen
NOTEFigure 3-6 and figure 3-7 are examples of startup and language selection screens.
NOTEThe computation constant is automatically calculated according to injectate system
type, injectate volume and catheter size. If the computation constant is manually
entered, the injectate volume and catheter size selections are set to Auto.
•Bolus Mode: Auto or Manual
13 Touch t h e Start Set button.
14 If in automatic bolus mode, Wait appears highlighted ( ) until the thermal baseline is
achieved. If in manual bolus mode, Ready ( ) will appear highlighted when the thermal
baseline is achieved. Touch the Inject button first to start the bolus procedure.
15 When Inject becomes highlighted ( ), use a rapid, smooth, continuous method to inject
the bolus with the volume amount previously selected.
16 Computing is highlighted ( ) and then the resultant iCO measurement is displayed.
17 Repeat steps 14-16 up to six times as desired.
18 Touch t h e Review button and if necessary, edit the bolus series.
19 Touch t h e Accept button.
4.1.3Continuous End Diastolic Volume Monitoring
Follow steps 1-8 at the start of section 4.1 before proceeding.
9 Attach the thermistor and thermal filament Swan-Ganz volumetric catheter connections
(figure 4-1) to the patient CCO cable.
10 Verify that the catheter is properly inserted into the patient.
11 Connect one end of the ECG interface cable to the rear panel of the HemoSphere advanced
monitor and the other end to the bedside monitor’s ECG signal output.
12 Touch the start monitoring icon to begin CO/EDV monitoring.
13 A countdown clock will appear on the stop monitoring icon to indicate the time until the
first CO/EDV value. After approximately 6 to 9 minutes, when sufficient data has been obtained,
an EDV and/or RVEF value will appear in the configured parameter globe(s).
14 The time until the next CO measurement is displayed on the information bar. For longer time
spans between calculations, select STAT parameters (sCO, sEDV, and sRVEF) as key parameters.
sCO, sEDV, and sRVEF are fast estimates of CO, EDV, and RVEF.
15 Touch the stop monitoring icon to stop CO/EDV monitoring.
All monitoring functions are initiated by touching the appropriate area on the touch screen. The navigation
bar, located on the left side of the screen, includes various controls for stopping and starting monitoring,
scrolling and selecting screens, performing clinical actions, adjusting system settings, capturing screen shots,
and silencing alarms. The main components of the HemoSphere advanced monitor screen are shown below
in figure 5-1. The main window displays the current monitoring view or menu screen. For details on
monitoring view types, see Monitor Views on page 51 . For details on other screen features, see the referenced
sections in figure 5-1.
48
Page 49
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
Main WindowNavigation Bar
(section 5.2)
Parameter Globe
(section 5.3.1)
Information Bar
(section 5.5)
Status Bar
(section 5.6)
Figure 5-1 HemoSphere advanced monitor screen features
49
Page 50
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
Start/Stop
Monitoring
Monitor
Screen
Selection
Clinical
Actions
Settings
Snapshot
Silence
Audible
Alarm
5.2Navigation Bar
The navigation bar is present on most screens. Exceptions are the startup screen and screens indicating the
HemoSphere advanced monitor has stopped monitoring.
Figure 5-2 Navigation bar - HemoSphere Swan-Ganz module monitoring
Start CO Monitoring. While monitoring with the HemoSphere Swan-Ganz module, the start CO
monitoring icon allows the user to initiate CO monitoring directly from the navigation bar. See Continuous
Cardiac Output on page 98 .
Stop CO Monitoring. The stop monitoring icon indicates that CO monitoring using the HemoSphere Swan-
Ganz module is underway. The user can immediately stop monitoring by touching this icon.
Monitor Screen Selection. The monitor screen selection icon allows the user to select the desired number of
monitored parameters displayed and the type of monitoring view used to display them, which is highlighted
in color (see figure 5-3, “Example of monitoring screen selection window,” on page 52). When a monitoring
view screen is selected, that monitoring mode is immediately displayed.
To return to the most recent monitoring screen displayed, touch the cancel icon .
Clinical Actions. The clinical actions icon provides access to the following clinical actions:
•Derived Value Calculator
•Event Review
50
Page 51
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
•iCO (HemoSphere Swan-Ganz module)
•Patient CCO Cable Test (HemoSphere Swan-Ganz module)
•Oximetry Calibration (HemoSphere oximetry cable)
A description of Derived Value Calculator and Event Review can be found in this chapter (see section
5.4.1 on page 65 and section 5.4.2 on page 65 ). For the remaining clinical actions, refer to the specified
module or cable chapter for more information.
Settings. The settings icon provides access to configuration screens which include:
•Patient Data: See chapter 6: User Interface Settings
•Monitor Settings: See chapter 6: User Interface Settings
•Advanced Setup: See chapter 7: Alarms / Targets , chapter 7: Adjust Scales , and chapter 8: Data
Export and Connectivity Settings
•Export Data: See chapter 8: Data Export and Connectivity Settings
•Demo Mode: See chapter 7: Demo Mode
•Engineering: See chapter 7: Engineering
•Help: See chapter 11: On Screen Help
Snapshot. The snapshot icon captures an image of the screen at the current time. A USB stick attached to
one of the two USB ports (rear and right panels) of the HemoSphere advanced monitor is required to save
the image.
Silence Audible Alarms. This icon silences all alarms for two minutes. New physiological alarms are silenced
during the two minute period. Alarms will resume sounding after the two minutes have elapsed. Faults are
silenced until the fault is cleared and re-occurs. If a new fault occurs, the alarm sound will resume.
Audible Alarms Silenced. Indicates that alarms are temporarily silenced. A two minute countdown timer and
“Alarms Paused” appear.
Monitoring Pause Exit. When the silence audible alarms button is touched for 3 consecutive seconds, a
monitoring pause confirmation popup will appear asking the user to confirm suspension of monitoring
operations. This function is used when the user wishes to pause monitoring. After confirmation, the silence
audible alarm button on the navigation bar will switch to the monitoring pause exit button and a
“Monitoring Pause” banner will be displayed. To return to monitoring, touch the monitoring pause exit
button.
5.3Monitor Views
There are six monitoring views: graphical trend, tabular trend, graphical/tabular trend split screen,
physiology, cockpit, and physio relationship. Up to four monitored parameters can be displayed on these
screens at one time.
To select a monitoring view:
51
Page 52
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
1 Touch the monitor screen selection icon . The monitor screen selection menu contains icons
that are based upon the look of the monitoring screens.
Figure 5-3 Example of monitoring screen selection window
2 Touch the circled number, 1, 2, 3, or 4, that represents the number of key parameters to be
displayed on the monitoring screens.
3 Select and touch a monitor view button to display the key parameters in that screen format.
5.3.1Parameter Globes
Parameter globes are located on the right side of most monitoring screens. The cockpit monitoring view is
composed of larger format parameter globes which function identically as described below.
5.3.1.1Change Parameters
1 Touch the displayed parameter label located outside the globe to change it to a different parameter.
52
Page 53
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
2 A popup window will show the selected parameter highlighted in color and other parameters
currently being displayed outlined in color. Available parameters appear on the screen without
highlights. Figure 5-4 shows the popup window that will appear while selecting continuous
parameters and monitoring with the HemoSphere Swan-Ganz module.
Figure 5-4 Example of key parameter
selection popup window
3 Touch an available parameter to select the replacement parameter.
5.3.1.2Change Alarm/Target
The Alarms / Targets popup screen lets the user view and set up alarm and target values for the selected
parameter or enable/disable the audible alarm and target settings. Additionally, the target settings can be
adjusted with a numbered key pad or with the scroll buttons when a minor adjustment is needed. This popup
screen is accessed by touching anywhere inside a monitored parameter globe or through the parameter
settings screen. For more information, see Alarms / Targets on page 79.
NOTEThere is a two minute inactivity timer associated with the popup screen.
53
Page 54
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
Audible alarm
indicator
Continuous %
Change
Indicator
Target Status
Indicator
Parameter Value
Units
Parameter Name
5.3.1.3Status Indicators
The lantern at the top of each parameter globe indicates the patient’s current status. The color changes as
the patient’s status changes. The globes may display additional information:
Figure 5-5 Parameter globe
Fault. When a fault condition occurs, the fault message(s) will be displayed on the status bar until the fault
condition is cleared. When there is more than one fault, alert or alarm, the message is cycled every two
seconds.
When a fault condition occurs, parameter calculations are stopped, and each affected parameter globe
displays the last value, time, and date at which the parameter was measured.
Continuous % Change Indicator. This indicator displays the percentage of change, followed by the time
period over which it changed. See Time Intervals / Averaging on page 75 for configuration options.
Target Status Indicators. The colored indicator at the top of each monitoring globe indicates the patient’s
clinical status. For indicator colors and their clinical indications, See table 7-2, “Target status indicator
colors,” on page 81.
5.3.2Graphical Trend Monitoring View
The graphical trend screen displays the current status and history of monitored parameters. The amount of
history shown for monitored parameters can be configured by adjusting the time scale.
When the target range for the parameter is enabled, the graph color codes the plot line, green indicating
within the target range, yellow indicating the value is outside the target range but within the physiological
alarm range, and red indicating the value is outside the alarm range. When the target range is disabled for
54
Page 55
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
the parameter the plot line is white. The colors match those of the clinical target indicator (lantern) on the
key parameter globes in the graphical trend graph when targets are enabled for the parameter. The alarm
limits for each parameter are displayed as colored arrows on the graph y-axis.
Figure 5-6 Graphical trend screen
To change the time scale of a displayed parameter, touch outside of the plot area along the x or y-axis, and a
scale popup menu will appear. Touch the value side of the Graphical Trend Time button to select a
different time period.
5.3.2.1Graphical Trend Scroll Mode
Up to 72 hours of monitored parameter data can be viewed by scrolling back. The date appears above the
parameter data during scrolling. Two dates will appear when appropriate. To start scrolling, touch the
appropriate scroll mode button. Keep touching the scroll mode button to increase the scroll speed. The
screen will return to live mode two minutes after the scroll button has been touched, or if the back button
is touched. The scroll rate will appear below the scroll buttons.
Table 5-1 Graphical trend scroll rates
Scroll settingDescription
>>>Scrolls at two times the current time scale
>>Scrolls at the current time scale (one graph
width)
>Scrolls at half the current time scale (one-half
graph width)
While in scroll mode the user can scroll to data older than the current time scale displays.
NOTEIt is not possible to touch past the most recent data or before the oldest data. The graph
will scroll only as far as data is available.
55
Page 56
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
5.3.2.2Intervention Events
While in the graphical trend screen, selecting the intervention icon provides a menu of intervention
types, details and a notes section.
Figure 5-7 Graphical trend- intervention window
To ent er a New Intervention:
1 Select the Intervention type from the New Intervention menu on left.
2 Select Detail from right menu tab. Unspecified is set as a default.
3 Select the keyboard icon to enter notes (optional).
4 Touch the enter icon .
To enter a previously used Intervention:
1 Select the Intervention from the Recents list tab.
2 To add, edit, or remove a note, touch the keyboard icon .
3 Touch the enter icon .
56
Page 57
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
Table 5-2 Intervention events
Intervention IndicatorType
Intervention
(green)
Positional
(purple)
Fluids
(blue)
Custom
(gray)
Inotrope
Vasodilator
Vasopressor
PEEP
Passive Leg Raise
Trendelenburg
Red Blood Cells
Colloid
Crystalloid
Custom Event
After selection of the intervention type, markers indicating the intervention are visually displayed on all
graphs. These markers can be selected for more information. Upon touching the marker, an information
balloon will appear. See figure 5-8: "Graphical trend screen - intervention information balloon". The
information balloon displays the specific intervention, date, time, and notes pertaining to the intervention.
Touching the edit button allows the user to edit intervention time, date, and note. Touching the exit button
closes the balloon.
NOTEThe intervention information balloon has a 2 minute time out.
Intervention Editing. The time, date, and associated note for each intervention can be edited after initial
entry:
1 Touch the intervention event indicator associated with the intervention to be edited.
2 Touch the edit icon located on the information balloon.
3 To change the time of the selected intervention, touch on Time Adjust, and enter the updated
time on keypad.
4 To change the date, touch on Date Adjust, and enter the updated date on keypad.
5 Touch the keyboard icon to enter or edit notes.
6 Touch the enter icon .
Figure 5-8 Graphical trend screen - intervention information balloon
57
Page 58
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
5.3.3 Tabular Trends
The tabular trends screen displays selected key parameters and their history in a tabular format.
Figure 5-9 Tabular trend screen
1 To change the interval between values, touch inside the table.
2 Select a value on the Tabular Increment popup.
Figure 5-10 Tabular increment popup
5.3.3.1Tabular Trend Scroll Mode
Up to 72 hours of data can be viewed by scrolling back. The scroll mode is based on the number of cells.
Three scroll speeds are available: 1x, 6x, and 40x.
58
Page 59
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
While the screen scrolls, the date appears above the table. If the time period overlaps two days, both dates
will appear on the screen.
1 To start scrolling, touch and hold one of the gray arrows. The scroll rate will appear above the
scroll icons.
Table 5-3 Tabular trend scroll rates
SettingTimeSpeed
1Xone cellSlow
6Xsix cellsModerate
40Xforty cellsFast
2 To exit scroll mode, stop touching the scrolling arrow or touch the return icon .
NOTEThe screen will return to live mode two minutes after the last touch of the scroll arrow
icon or if the return icon is touched.
5.3.4Graphical/Tabular Trends Split
The graphical/tabular trends split screen displays a combination of the graphical trend and tabular trend
monitoring views. This display is useful for viewing the current status and history of selected monitored
parameters in graphical format and other selected monitored parameters in tabular format at the same time.
If two key parameters are selected, the first key parameter is displayed in graphical trend format, and the
second in tabular trend format. Key parameters can be changed by touching the parameter label located on
the parameter globe. If more than two key parameters are selected, the first two parameters are displayed in
graphical trend format, and the third and fourth — if a fourth is selected — are displayed in tabular trend
format. The time scale for data displayed on any key parameter graphical trend view(s) is independent of the
time scale displayed on the tabular trend view(s). For more information on the graphical trend view see
Graphical Trend Monitoring View on page 54. For more on tabular trend view, see Tabular Trends on page 58.
59
Page 60
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
5.3.5Physiology Screen
The Physiology screen is an animation depicting the interaction between the heart, blood, and vascular
system.
Continuous parameter values are displayed in association with the animation.
Figure 5-11 Physiology screen
In the physiology screen the image of the beating heart is a visual representation of the heart rate and is not
an exact representation of beats per minute. Key features of this screen are numbered shown in figure 5-11.
This example is of the continuous physiology screen during active monitoring with the HemoSphere SwanGanz module and slaved-in ECG, MAP, and CVP signals.
1 ScvO
/SvO2 parameter data and signal quality indicator (SQI) are displayed here while the
2
HemoSphere oximetry cable is connected and actively monitoring venous oxygen saturation
2 Cardiac output (CO/CI) is indicated on the arterial side of the vascular system animation. The
blood flow animation rate will adjust based on the CO/CI value and the low/high target ranges
selected for that parameter.
3 Systemic Vascular Resistance, indicated in the center of the vascular system animation, is available
while monitoring CO/CI and utilizing MAP and CVP analog pressure signal inputs from a
connected patient monitor, as SVR =[(MAP-CVP)/CO]*80. The level of constriction shown in
the vessel will adjust based on the derived SVR value, and the low/high target ranges selected for
that parameter.
NOTEThe alarms/targets settings can be adjusted through the Alarms / Targets setting screen
(see Alarms / Targets Setup Screen on page 82 ) or by selecting the desired parameter as
a key parameter, and accessing the parameter Alarms/Targets popup window by
touching inside of the parameter globe.
4 From the continuous mode, touch the clock/waveform icon at the upper left, to go to the
intermittent physiology screen. This button only appears when there is historical intermittent data
available. See 5.3.5.1 Historic Physiology Screen, below.
5 Touch the syringe to go to the iCO screen to shoot bolus cardiac output.
60
Page 61
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
5.3.5.1Historic Physiology Screen
The historic physiology screen displays both intermittent bolus data and a snapshot of continuous data
overlaid on a visual representation of the heart and circulatory system. The circulatory system has several
variations to illustrate the patient’s condition at the time of the bolus set — for example, the vessels constrict.
Up to 36 historic physiology records may be viewed via the horizontal tabs composite along the top of the
screen.
5.3.6Cockpit Screen
This monitoring screen, shown in figure 5-12, displays large parameter globes with the values of the
parameter being monitored. Cockpit parameter globes graphically indicate alarm/target ranges and values,
and utilize needle indicators to show where the current parameter value falls. Similar to standard parameter
globes, the value within the globe will flash when the parameter is alarming.
Figure 5-12 Cockpit monitoring screen
The key parameter globes shown on the cockpit screen display a more complex target and alarm indicator
than the standard parameter globe. The full display range of the parameter is used to create a gauge from the
graphical trends minimum to maximum settings. A needle is used to indicate the current value on the circular
gauge scale. When target ranges are enabled, red (alarm zone), yellow (warning target zone), and green
(acceptable target zone) are used to indicate the target and alarm regions within the circular gauge. When
target ranges are not enabled, the circular gauge area is all gray in color and target or alarm indicators are
removed. The value indicator arrow changes to indicate when the values are out of the gauge scale limits.
5.3.7Physio Relationship
The physio relationship screen depicts the balance between oxygen delivery (DO2) and oxygen consumption
(VO
). It automatically updates as parameter values change so the values are always current. The connecting
2
lines highlight the relationship of the parameters to each other.
61
Page 62
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
5.3.7.1Continuous and Historical Modes
The physio relationship screen has two modes: continuous and historical. When in continuous mode, the
intermittent and derived values are always displayed as unavailable.
Figure 5-13 Physio relationship screen
1 The vertical lines above and below the parameters appear in the same color as the parameter
lantern.
2 Vertical lines that directly connect two parameters will appear in the same color as the parameter
lantern below (for example, between SVRI and MAP).
3 The horizontal lines are the same color as the line above them.
4 The left bar appears after a bolus set is performed. Touch the clock/waveform icon to display
historical data (see figure 5-13).
5 Touch the iCO icon to open the thermodilution new set configuration screen.
62
Page 63
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
NOTEBefore a thermodilution set is performed and before any values are entered (see 5.3.7.2
Parameter Boxes below)
the clock/waveform and iCO icons do not appear. Only the
available continuous parameters are displayed.
Figure 5-14 Historic physio relationship data screen
NOTEThe historic physio relationship screen displays most of the parameters available on the
system at a point in time. The screen displays lines connecting the parameters,
highlighting the relationship of the parameters to each other. The historic physio
relationship screen displays the configured (1-4) key parameters on the right hand side
of the screen. There is a horizontal tab composite at the top that allows the user to
navigate through the database of historic records. The record times correspond to
thermodilution bolus sets and derived value calculations.
The historic physio relationship screen allows the user to enter parameters used to
calculate derived parameters DO
and VO2, on only the most recent record. The values
2
entered are for the time of the record and not the current time.
The historic physio relationship screen is accessed through the clock/waveform icon
on the continuous physio relationship screen. Touch the return icon to return to
the continuous physio relationship screen. There is no 2 minute time-out for this
screen.
To calc u lat e DO
oxygen is required. For the historic physio relationship screen, a PaO
of zero (0) is used. To calculate DO
and VO2, the partial pressure of arterial (PaO2) and venous (PvO2)
2
and PvO2 value
2
and VO2 using values other than zero (0) for PaO2
2
and PvO2, use the Derived Value Calculator (see section 5.4.1 on page 65 ).
63
Page 64
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
Parameter
UnitsValue
Target
status
indicator
Faulted or
value not
available
5.3.7.2Parameter Boxes
Each small parameter box displays:
•Parameter name
•Parameter units
•Parameter value (if available)
•Clinical target status indicator (if a value is available)
If the parameter is in a fault state, the value is blank, indicating it is or was unavailable at the time of the
display.
Figure 5-15 Physio relationship parameter boxes
5.3.7.3Setting Targets and Entering Parameter Values
To change the target settings or enter a value, touch a parameter to bring up the target/enter popup. The
physio relationship target/enter popup will be displayed when the following physio relationship small
parameter boxes are touched:
•HGB
•SpO
•SvO
•CVP (when analog pressure signal input not configured)
•MAP (when analog pressure signal input not configured)
•HRavg (when ECG signal input not configured)
2
/ScvO2 (when no HemoSphere oximetry cable measurement is available)
When the value is accepted, a new time-stamped historic physio relationship record is created. It includes:
The historic physio relationship screen is shown with the newly created record; you can then enter the rest
of the manually entered values to calculate any derived values.
•Current continuous parameter data
•The entered value and any derived calculated values.
64
Page 65
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
5.4Clinical Actions
Most options on the clinical actions menu are related to the current monitoring mode (e.g., while monitoring
with the HemoSphere Swan-Ganz module). The following clinical actions are available across all monitoring
modes.
5.4.1Derived Value Calculator
The Derived Value Calculator allows the user to compute certain hemodynamic parameters and provides
a convenient way to display these parameters for one-time calculation.
Calculated parameters include: DO
1 Touch the clinical actions icon Derived Value Calculator icon .
2 Enter the required values and the derived calculations will automatically display.
3 Touch the home icon to return to the monitoring screen.
, VO2, SVR, LVSWI and RVSWI.
2
5.4.2Event Review
Use Event Review to view parameter-related and system events that occurred during monitoring. Up to 72
hours of events are recorded in order with the most recent event at the top.
3 Touch the home icon to return to the monitoring screen.
The following events are included in the clinical event review log.
Table 5-4 Reviewed events
EventLog time
CO Monitoring StartedWhen CO Monitoring is started
CO Monitoring StoppedWhen the user or system stops CO monitoring
CO Cable Test PassedWhen the Patient CCO Cable Test was performed and
passed
Draw BloodThe Draw option is selected in the In vivo Calibration
Draw Screen
HGB UpdateOximetry cable update completes following the HGB
update process
iCO Bolus PerformedWhen an iCO bolus is performed
In vitro CalibrationWhen oximetry cable update completes following the
In vitro calibration process
In vivo CalibrationWhen oximetry cable update completes following In
vivo calibration process
Light Out of RangeWhen the Oximetry Light Range Fault occurs
Monitoring PausedActive monitoring paused to prevent audible alarms
and parameter monitoring
Monitoring ResumedNormal monitoring resumed. Audible alarms and
parameter monitoring are active
65
Page 66
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
TimeDate
Battery
Monitoring
Mode
Averaged
Heart Rate
Blood
Temperature
HIS
Status
Wi-Fi
Status
Lock
Screen
Table 5-4 Reviewed events (continued)
EventLog time
Oximetry DisconnectedAn oximetry cable disconnection is detected
Recall Oximetry DataWhen recalled oximetry calibration data is accepted by
the user
System Restart RecoveryWhen the system has resumed monitoring without
being prompted following a power cycle
Time ChangeThe system clock is updated
5.5Information Bar
The information bar appears on all active monitoring screens and most clinical action screens. It displays the
current time, date,monitoring mode, battery status, and the lock screen symbol. While monitoring with the
HemoSphere Swan-Ganz module, the CO countdown, blood temperature, and slaved in heart rate may also
be displayed. When the monitor has a HIS or Wi-Fi connection, the status will be displayed. See table 8-1
on page 91 for Wi-Fi status symbols and table 8-2 on page 92 for HIS connectivity status symbols. Figure 517 shows an example of an information bar while monitoring with the HemoSphere Swan-Ganz module
with a slaved in ECG heart rate.
Figure 5-17 Information Bar - HemoSphere Swan-Ganz Module
NOTEFigure 5-17 is an example of an information bar with U.S. standard defaults. To see the
defaults for all languages, see table D-6, “Language default settings,” on page 148.
5.5.1Battery
The HemoSphere advanced monitor allows for uninterrupted monitoring during power loss when the
HemoSphere battery pack is installed. Battery life is indicated on the information bar by the symbols shown
in table 5-5. For more information on battery installation, see Battery Installation on page 39. To ensure that
the battery charge status displayed on the monitor is correct, it is recommended to perform periodic checks
of battery health through battery conditioning. For information on battery maintenance and conditioning,
see Battery Maintenance on page 155.
Table 5-5 Battery status
Battery
symbol
Indication
The battery has greater than
50% charge remaining.
The battery has less than 50%
charge remaining.
66
Page 67
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
Table 5-5 Battery status (continued)
Battery
symbol
Indication
The battery has less than 20%
charge remaining.
The battery is charging and
connected to mains power.
The battery is fully charged and
connected to mains power.
The battery is not installed.
WARNINGTo prevent any interruptions to monitoring during power loss, always use the
HemoSphere advanced monitor with the battery inserted.
In cases of power failure and battery depletion, the monitor will go through a controlled
shut off procedure.
5.5.2Lock Screen
If the monitor is being cleaned or moved, lock the screen. For cleaning instructions refer to Cleaning the
Monitor and Modules on page 152. The screen will automatically unlock once the internal timer has counted
down.
1 Touch the lock screen icon.
2 Touch the time that the screen will remain locked on the Lock Screen popup.
Figure 5-18 Lock screen
3 A large lock icon will appear to the right of the information and status bar.
4 To unlock the screen, touch and hold the large lock icon .
67
Page 68
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
5.6Status Bar
The status bar appears at the bottom of all active monitoring screens. It displays faults, alarms, alerts, some
warnings and notifications. When there is more than one fault, alert or alarm, the message is cycled every
two seconds.
Figure 5-19 Status bar
5.7Monitor Screen Navigation
There are several standard navigational procedures on the screen.
5.7.1Vertical Scrolling
Some screens will have more information than fits on the screen at one time. If vertical arrows appear on a
review list, touch the up or down arrow to see the next set of items.
If selecting from a list, the vertical scroll arrows move up or down one item at a time.
5.7.2Navigation Icons
There are some buttons that always perform the same function:
Home. The home icon takes you to the most recently viewed monitoring screen and stores any modification
made to data on the screen.
Return. The return icon takes you to the previous menu screen and stores any modification made to data on
the screen.
Cancel. The cancel icon causes any entries to be discarded.
On some screens, for example Patient Data, there is no cancel button. As soon as a patient’s data is entered,
it is stored by the system.
List buttons. Some of the screens have buttons that appear to be split in two.
68
Page 69
HemoSphere Advanced Monitor5 Navigating the HemoSphere Advanced Monitor
backspace
cancel
enter
decimal
cancel
enter
backspace
cursor
left
cursor
right
In these cases, touching anywhere on the button reveals a list of selectable items. The right side of the button
displays the current selection.
Value button. Some screens have square buttons as shown below. Touch the button to display a keypad.
Toggle button. When an option exists between two choices, such as on/off, a toggle button appears.
Touch on the opposite side of the button to switch the choice.
Keypad. Touch the keys on the keypad to enter numeric data.
Keyboard. Touch the keys on the keyboard to enter alphanumeric data.
After the system is turned on, the user has the option to either continue monitoring the last patient or to
start monitoring a new patient. See figure 6-1 below.
NOTEIf data for the last patient monitored is 12 hours or older, the only option is to start a
new patient.
6
Figure 6-1 New or continuing patient screen
6.1.1New Patient
Starting a new patient clears all previous patient data. The alarm limits and continuous parameters are set to
their default values.
WARNINGUpon initiation of a new patient session, the default high/low physiological alarm
ranges should be checked to ensure that they are appropriate for the given patient.
70
Page 71
HemoSphere Advanced Monitor6 User Interface Settings
The user has the option of entering a new patient upon initial startup of the system or while the system is
running.
WARNINGPerform New Patient or clear the patient data profile whenever a new patient is
connected to the HemoSphere advanced monitor. Failure to do so may result in
previous patient data in the historical displays.
1 After turning on the monitor, the new or continuing patient screen appears (figure 6-1). Touch
New Patient
and continue to step 6.
OR
If the monitor is already on, touch the settings icon and continue to step 2.
2 Touch Patient Data button.
3 Touch New Patient button.
4 Touch Ye s button on the confirmation screen to start a new patient.
5 The New Patient Data screen appears. See figure 6-2.
Figure 6-2 New Patient Data screen
6 Touch the enter key on the keypad/keyboard to save each patient demographic selection value
and return to the patient data screen.
7 Touch Patient ID button and use the keyboard to enter the patient’s hospital ID.
8 Touch Height button and use the keypad to enter the patient’s height. The unit default for your
language is at the upper right of the keypad. Touch it to change the unit of measurement.
9 Touch Age and use the keypad to enter the patient’s age.
10 Touch Weight and use the keypad to enter the patient’s weight. The unit default for your language
is at the upper right of the keypad. Touch it to change the unit of measurement.
11 Tou ch Gender and select Male or Female.
12 The BSA is calculated from the height and weight using the DuBois formula.
71
Page 72
HemoSphere Advanced Monitor6 User Interface Settings
13 Touch the home icon and refer to instructions for starting monitoring with the desired
hemodynamic monitoring technology.
NOTEThe home icon is disabled until all patient data is entered.
6.1.2Continue Monitoring Patient
If the last patient’s data is less than 12 hours old, the patient’s demographics and patient ID will be displayed
when the system is turned on. When monitoring of the last patient is continued, the patient’s data is loaded
and the trend data is retrieved. The most recently viewed monitoring screen is displayed. Touch Continue Same Patient.
6.1.3View Patient Data
1 Touch settings icon .
2 Touch Patient Data button to see patient data. The screen will also include a New Patient
button.
3 Touch the return icon to return to the settings screen.
6.2 Monitor Settings
The Monitor Settings screen allows the user to change several monitor related settings.
Figure 6-3 Monitor Settings
NOTEThe screen will return to the monitoring view after two minutes of inactivity.
6.2.1General Monitor Settings
The General Monitor Settings are those that affect every screen. These are the display language, units used,
alarm volume, and snapshot sound.
72
Page 73
HemoSphere Advanced Monitor6 User Interface Settings
The HemoSphere advanced monitor interface is available in several languages. A language selection screen
appears the first time the HemoSphere advanced monitor is started. See figure 3-7, “Language selection
screen,” on page 42. The language screen will not appear again, but the display language can be changed at
any time.
The selected language determines the default time and date format. These can also be changed
independently of the language selected.
NOTEIf power is lost and restored to the HemoSphere advanced monitor, the system settings
prior to the power loss, including alarm settings, alarm volume, target settings,
monitoring screen, parameter configuration, language and unit selection, are automatically restored to last configured settings.
6.2.1.1 Change Language
1 Touch the settings icon .
2 Touch the Monitor Settings button.
3 Touch General button.
Figure 6-4 General Monitor Settings
4 Touch the value section of the Language button and select the desired display language.
5 Touch the home icon to return to the monitoring screen.
NOTESee appendix D for all language default settings.
6.2.2Change Date and Time Display
English (US) dates default to MM/DD/YYYY, and the time defaults to a 12 Hour clock.
When an international language is selected, the date defaults to the format found in appendix D: Monitor Settings and Defaults, and the time defaults to a 24 hour clock.
73
Page 74
HemoSphere Advanced Monitor6 User Interface Settings
1 Touch the settings icon .
2 Touch Monitor Settings button.
3 Touch Date / Time button.
Figure 6-5 Date / Time Settings
4 Touch the value section of the Date Format button and touch the desired format.
5 Touch the value section of the Time Format button and touch the desired format.
6 Touch the home icon to return to the monitoring screen.
6.2.2.1 Adjust Date or Time
The system time can be reset if necessary. When the time or date is changed, trended data is updated to
reflect the change. Any retained data is updated to reflect the time change.
NOTEThe time clock of the HemoSphere advanced monitor does not automatically adjust for
daylight saving time (DST). This adjustment needs to made using the following instructions.
1 Touch the settings icon .
2 Touch Monitor Settings button.
3 Touch Date / Time.
4 To change the date, touch the value section of the Date Adjust button and enter the date on the
keypad.
5 To change the time, touch the value section of the Time Adjust button and enter the time.
6 Touch the home icon to return to the monitoring screen.
74
Page 75
HemoSphere Advanced Monitor6 User Interface Settings
6.2.3Monitoring Screens Settings
From the Monitoring Screens settings screen, the user can set physiology and physio relationship
monitoring screen options.
1 Touch the settings icon .
2 Touch the Monitor Settings button.
3 Touch the Monitoring Screens button.
4 Select the Indexed or Non-Indexed toggle for parameters in the physiology and physio
relationship screens.
6.2.4Time Intervals / Averaging
The Time Intervals / Averaging screen lets the user select the continuous % change time interval.
NOTEThe screen will return to the monitoring view after two minutes of inactivity.
1 Touch the settings icon .
2 Touch Parameter Settings button.
3 Touch Time Intervals / Averaging button.
4 Touch the right side of the Continuous % Change Interval value button and touch one of the
following time interval options:
•None•15 min
•5 min•20 min
•10 min•30 min
5 Touch home icon to return to the monitoring screen.
6.2.5Analog Pressure Signal Input
While performing CO monitoring, the HemoSphere advanced monitor can also calculate SVR by utilizing
analog pressure signal inputs from a connected patient monitor.
NOTEConnecting to external input devices allows additional information to be displayed. For
example, when MAP and CVP are available continuously from a bedside monitor, SVR
is displayed if configured in a parameter globe. MAP and CVP are displayed on the
physio relationship and physiology monitoring screens.
75
Page 76
HemoSphere Advanced Monitor6 User Interface Settings
WARNINGThe analog communication ports of the HemoSphere advanced monitor share a
common ground that is isolated from the catheter interface electronics. When
connecting multiple devices to the HemoSphere advanced monitor, all devices should
be provided with isolated power to avoid compromising the electrical isolation of any
of the connected devices.
Risk and Leakage current of the final system configuration must comply with IEC
60601-1:2005/A1:2012. It is the responsibility of the user to ensure compliance.
Accessory equipment connected to the monitor must be certified according to IEC/
EN 60950 for data-processing equipment or IEC 60601-1:2005/A1:2012 for electromedical equipment. All combinations of equipment must be in compliance with IEC
60601-1:2005/A1:2012 systems requirements.
CAUTIONWhen connecting the HemoSphere advanced monitor to external devices, refer to the
external device’s instruction manual for complete instructions. Verify proper operation
of the system before clinical use.
Once the bedside monitor has been configured for the desired parameter output, connect the monitor via
an interface cable to the selected analog input port on the HemoSphere advanced monitor.
NOTEA compatible bedside monitor must provide an analog output signal.
Please contact your local Edwards representative to obtain the correct HemoSphere
advanced monitor analog input interface cable for your bedside monitor.
The following procedure describes how to configure the analog input ports of the HemoSphere advanced
monitor.
1 Touch the settings icon .
2 Touch Monitor Settings button.
3 Touch Analog Input button.
4 Select MAP from the Parameter list button for the numbered analog port where MAP is
connected (1 or 2). The default setting values for MAP will be displayed.
NOTEIf an analog signal is not detected on the selected port, “Not Connected” will be
displayed below the Port list button.
When an analog input connection or disconnection is first detected, a brief notification
message will be displayed on the status bar.
5 Select CVP in the Parameter list button for the numbered analog port where CVP is connected.
The default setting values for CVP will be displayed.
76
Page 77
HemoSphere Advanced Monitor6 User Interface Settings
NOTEThe same parameter may not be configured on more than one analog input at the same
time.
6 If the default values are correct for the bedside monitor being used, touch the home icon .
If the default values are not correct for the bedside monitor being used (refer to the bedside
monitor operator’s manual), the user can modify the voltage range, full scale range, or perform the
calibration option described in section 6.2.5.1 of this chapter.
Touch the Full Scale Range value button to change the displayed full scale signal value. Table 61 below shows the allowable input values for full scale range based on the selected parameter.
Table 6-1 Analog input parameter ranges
ParameterFull Scale Range
MAP0 to 510 mmHg (0 kPa to 68 kPa)
CVP0 to 110 mmHg (0 kPa to 14.6 kPa)
NOTEA voltage reading of zero is automatically set to a minimum pressure reading of 0
mmHg (0 kPa). The Full Scale Range represents the full scale signal or maximum
pressure reading for the selected Volta ge R a ng e.
Touch the Vo lt age R an ge list button to change the displayed voltage range. The selectable
voltage ranges available for all of the parameters are:
•0 - 1 volts
•0 - 5 volts
•0 - 10 volts
•Custom (see 6.2.5.1: Calibration)
WARNINGWhen switching to a different bedside monitor, always check that the default values
listed are still valid. If necessary, reconfigure the voltage range and corresponding
parameter range or calibrate.
6.2.5.1 Calibration
The calibration option is required when default values are incorrect, or the voltage range is not known. The
calibration process configures the HemoSphere advanced monitor with the analog signal received from the
bedside monitor.
NOTEIf the default values are correct, do not calibrate.
77
Page 78
HemoSphere Advanced Monitor6 User Interface Settings
CAUTIONOnly properly trained personnel should calibrate the HemoSphere advanced monitor
analog ports.
1 Touch the settings icon .
2 Touch Monitor Settings button.
3 Touch Analog Input button.
4 Select the desired port number (1 or 2) from the Port list button and corresponding parameter
(MAP or CVP) from the Parameter list button.
5 Select Custom from the voltage value popup screen. The Analog Input Custom Settings screen
will appear.
6 Simulate a full scale signal from the bedside monitor to the selected analog input port on the
HemoSphere advanced monitor.
7 Set the maximum parameter value equal to the full-scale signal value.
8 Touch Calibrate Maximum button. The Maximum A/D value will appear on the Analog
Input Custom Settings screen.
NOTEIf an analog connection is not detected, the Calibrate Maximum and Calibrate
Minimum buttons will be disabled and the Maximum A/D value will be displayed as
Not Connected.
9 Repeat the process to calibrate the minimum parameter value.
10 Touch th e Accept button to accept the displayed custom settings and return to the Analog Input
Screen.
11 Repeat steps 4-10 to calibrate another port if needed, or touch the home icon to return to
the monitoring screen.
CAUTIONThe accuracy of continuous SVR depends upon the quality and accuracy of the MAP
and CVP data transmitted from the external monitors. Since MAP and CVP analog
signal quality from the external monitor cannot be validated by the HemoSphere
advanced monitor, actual values and the values (including all derived parameters)
displayed by the HemoSphere advanced monitor may not be consistent. The accuracy
of continuous SVR measurement, therefore, cannot be guaranteed. To aid in determining the quality of the analog signals, regularly compare the MAP and CVP values
displayed on the external monitor to the values displayed on the physio relationship
screen of the HemoSphere advanced monitor. Refer to the external input device
operator’s manual for detailed information regarding accuracy, calibration, and other
variables which may impact the analog output signal from the external monitor.
There are two types of alarms on the HemoSphere advanced monitor:
1 Physiological alarms: These are set by the clinician and signify the upper and/or lower alarm
ranges for configured key continuous parameters.
7
2 Technical alarms: This alarm signifies a device fault or alert. Technical alarms are non-latching.
Alarms occur with either Medium or High priority. Only parameters that are displayed (key parameters) will
have active visual and audible alarms.
For physiological parameters CO/CI, sCO/sCI, SV/SVI, and ScvO
priority is medium and the lower alarm (red zone) priority is high. For the physiological parameters SVR/
SVRI, EDV/sEDV, EDVI/sEDVI, and RVEF/sRVEF the alarm priority is always medium. See Alarm Priorities on page 147.
Among technical alarms, faults are of medium priority, and will halt operation of the related monitoring
activity. Alerts are of low priority and will not halt any monitoring activity. As faults are of higher priority
than alerts, alerts will not be alarmed if there are any active faults.
All alarms have an associated text displayed on the status bar. The system will actively cycle through every
active alarm text on the status bar. In addition, alarms will generate the visual alarm indicator shown in
table 7-1 below. For additional information, see table 11-1 on page 119.
Table 7-1 Visual alarm indicator colors
Alarm priorityColorLight pattern
HighredFlashing ON/OFF
MediumyellowFlashing ON/OFF
LowyellowSolid ON
/SvO2 the upper alarm (red zone)
2
79
Page 80
HemoSphere Advanced Monitor7 Advanced Settings
The visual alarm indicator will indicate the highest active alarm priority. The audible tone associated with the
highest priority active alarm will be played. Where the priority levels are the same, physiological alarms take
priority over faults and alerts. All technical alarms are generated once detected by the system; there is no
inherent delay in alarms from the point of detection. For physiological alarms, the delay is the amount of
time it takes to calculate the next physiological parameter:
•HemoSphere Swan-Ganz module continuous CO and associated parameters: varies, but is typically
around 57 seconds (See CO Countdown Timer and STAT CO on page 100).
•Oximetry: 2 seconds
All alarms are logged and stored for the given patient and can be accessed via the Data Download function
(see Data Download on page 89). The Data Download log is cleared when initiating a new patient (see New Patient on page 70). The current patient can be accessed from up to 12 hours following a system power-off.
7.1.1Silence Alarms
7.1.1.1 Physiological Alarms
Physiological alarms can be silenced directly from the monitoring screen by touching the silence audible
alarms icon . The physiological alarm audio tone is silenced for two minutes. No audio tone for any
physiological alarm will be emitted during this two minutes, including new physiological alarms triggered
during this time. If a technical alarm is generated during this two-minute period, the audio silence will be
cleared, allowing alarm audio tones to resume. The user can also manually clear the two-minute period by
pressing the alarm silence button again. Once the two-minute period has elapsed, active physiological alarms
will resume audio sound.
If the physiological alarm is a medium priority, the visual alarm indicator (blinking yellow) is also disabled
for two minutes. A high priority visual alarm indicator (blinking red) cannot be disabled. For information on
physiological alarm priorities, see Alarm Priorities on page 147.
NOTEPhysiological parameters can be configured to have no alarms. See sections 7.1.5
and7.1.7.
WARNINGDo not turn off the audible alarms in situations in which patient safety could be
compromised.
7.1.1.2 Technical Alarms
During an active technical alarm, the user can silence the alarm and clear the visual alarm indicator (medium
and low priority) by touching the silence audible alarms icon . The visual alarm indicator and audio
tone will remain inactive unless another technical or physiological alarm condition triggers, or the original
technical alarm resolves and re-triggers.
7.1.2Set Alarm Volume
The alarm volume ranges from low to high with a default of medium. It applies to physiological alarms,
technical faults, and alerts. Alarm volume can be changed at any time.
80
Page 81
HemoSphere Advanced Monitor7 Advanced Settings
1 Touch the settings icon .
2 Touch Monitor Settings button.
3 Touch General button.
4 Touch the right side of the Alarm Volume list button to select the desired volume.
5 Touch the home icon to return to the monitoring screen
WARNINGDo not lower the alarm volume to a level that prohibits alarms from being adequately
monitored. Failure to do so could result in a situation where patient safety is
compromised.
7.1.3Set Targets
Targets are visual indicators (lanterns) set by the clinician to indicate if the patient is in the ideal target zone
(green), warning target zone (yellow), or alarm zone (red). The use of target zone ranges can be enabled or
disabled by the clinician. Alarms (high /low) differ from target zones in that the alarm parameter value
flashes and has a audible alarm.
Parameters that can “Alarm” are indicated by a bell icon in the Alarms / Targets settings screen. High/
low alarms by default also become the ranges for the red caution zone for that parameter. Parameters which
DO NOT have the ability to set a high/low alarm will not have a bell icon in the Alarms/Targets settings
screen for that parameter but can still have target ranges set.
Table 7-2 Target status indicator colors
ColorIndication
GreenAcceptable – Green target zone is considered
an ideal range for parameter as set by the
clinician.
YellowYellow target zone is considered a warning
range and visually indicates that the patient has
exited the ideal range but has not entered the
alarm or caution range as set by the clinician.
81
Page 82
HemoSphere Advanced Monitor7 Advanced Settings
Table 7-2 Target status indicator colors
ColorIndication
RedRed alarm and/or target zones can be
considered “Alarm” parameters indicated by a
bell icon in the Alarms /Targets settings
screen. High/low alarms by default also
become the range for the red caution zone for
that parameter. Parameters which DO NOT
have the ability to set a high/low alarm will not
have a bell icon in the Alarms /Targets settings
screen for that parameter but can still have
target ranges set. Ranges for the alarm and/or
target zone are to be set by the clinician.
GrayIf a target is not set, the status indicator
appears as gray.
7.1.4 Alarms / Targets Setup Screen
The Alarms / Targets Setup Screen allows the clinician to view and set up alarms and targets for each key
parameter. From the Alarms/Targets screen, located within the Advanced Setup settings menu, the user
can adjust targets and enable/disable audible alarms. Any features accessed through the Advanced Setup
settings menu are passcode protected and should only be altered by experienced clinicians. The settings for
each key parameter are displayed in a parameter box. The currently configured key parameters are the first
set of key parameters displayed. The remaining key parameters are displayed in a defined order. The
parameters also indicate what the target ranges are based on: Custom Default, Edwards Default, and
Modified.
Table 7-3 Target defaults
Default nameDescription
Custom DefaultA custom default target range was set for the parameter and the
parameter target range has not been modified from that default.
Edwards DefaultThe parameter target range has not been changed from the
original settings.
ModifiedParameter target range was changed for this patient.
NOTEVisual and audible alarm settings are only applicable to parameters being displayed.
To modify Alarms / Targets:
1 Touch the settings icon .
2 Touch Advanced Setup button and enter the required password.
6 The defaults can be displayed as Indexed or Non-Indexed. Choose the desired format on the
Set all parameters according to: toggle button.
7 Touch the parameter of interest.
8 Touch the value button for each target setting and enter the desired value. The corresponding
indexed or non-indexed value for that parameter will be set automatically.
9 Continue steps 7 and 8 for each parameter. Touch the right or left arrow at the bottom of the
screen to display the next or previous set of parameters.
10 When all desired parameters have been modified, touch Confirm All.
84
Page 85
HemoSphere Advanced Monitor7 Advanced Settings
7.1.7Configure Targets and Alarms for One Parameter
The Alarms/Targets popup lets the user set up alarm and target values for the selected parameter. The user
can also enable or disable the audible alarm. Adjust the target settings by using the numbered keypad or by
using the scroll buttons when a minor adjustment is needed.
1 Touch inside a globe to open the alarms/targets popup for that parameter. The alarms/targets
popup is also available on the physio relationship screen by touching a parameter box.
2 To disable the audible alarm for the parameter, touch the Audible Alarm icon at the top
right of the popup.
NOTEParameters that DO NOT have the ability to set a high/low alarm will not have an
Audible Alarm icon on the Alarms / Targets popup.
3 To disable visual targets for the parameter, touch the Ta rg et enabled icon at the top left of
the popup. The target indicator for that parameter will appear gray.
4 Use the arrows to adjust the zone settings or touch the value button to open a numeric keypad.
Figure 7-3 Set individual parameter alarms
and targets
5 When the values are correct, touch the enter icon .
6 To cancel, touch the cancel icon .
WARNINGVisual and audible physiological alarms are activated only if the parameter is configured
on the screens as a key parameter (1-4 parameters displayed in parameter globes). If a
parameter is not selected and displayed as a key parameter, the audible and visual physiological alarms are not triggered for that parameter.
85
Page 86
HemoSphere Advanced Monitor7 Advanced Settings
7.2 Adjust Scales
The graphical trend data fills the graph from left to right with the most recent data at the right. The
parameter scale is on the vertical axis with the time scale on the horizontal.
Figure 7-4 Graphical trend screen
The scales setup screen allows the user to set up both the parameter and time scales. The key parameters are
at the top of the list. Use the horizontal scroll buttons to see additional parameters.
1 Touch the settings icon .
2 Touch Advanced Setup button and enter the required password.
NOTEThe screen will return to the monitoring view after two minutes of inactivity.
86
Page 87
HemoSphere Advanced Monitor7 Advanced Settings
4 For each parameter, touch the Lower button to enter the minimum value to appear the vertical
axis. Touch the Upper button to enter the maximum value. Use the horizontal scroll icons
to view additional parameters.
5 Touch the right side of the Graphical Trend Time value button to set the total amount of time
displayed on the graph. The options are:
•3 minutes•1 hour•12 hours
•5 minutes•2 hours (default)•18 hours
•10 minutes•4 hours•24 hours
•15 minutes•6 hours•48 hours
•30 minutes
6 Touch the right side of the Tabular Increment value icons to set the amount of time to each
tabbed value. The options are:
•1 minute (default)•30 minutes
•5 minutes•60 minutes
•10 minutes
7 To advance to the next set of parameters, touch the arrow at the bottom left.
8 Touch the home icon to return to the monitoring screen.
7.3 Serial Port Setup
Use the Serial Port Setup screen to configure the serial port for digital data transfer. The screen displays
until the return icon is touched
Figure 7-6 Tabular increment popup
87
Page 88
HemoSphere Advanced Monitor7 Advanced Settings
1 Touch the settings icon .
2 Touch Advanced Setup button and enter the required password.
3 Touch Serial Port Setup button.
4 Touch the list button of any serial port setup parameter to change the default value shown.
5 Touch the return icon when configuration of serial port settings is complete.
NOTEA RS232 9 pin serial port is available for real time communication to support patient
monitoring systems through the IFMout protocol.
7.4Demo Mode
Demonstration Mode is used to display simulated patient data to assist in training and demonstration.
Demonstration mode displays data from a stored set and continually loops through a predefined data set.
During Demo Mode, the HemoSphere advanced monitoring platform user interface retains the same
functionality as a fully operational platform. Simulated patient demographics must be entered to
demonstrate Swan-Ganz technology functions. The user can touch the controls as if a patient was being
monitored.
When Demo Mode is entered, trended data and events are cleared from being displayed and saved for return
to patient monitoring.
1 Touch settings icon .
2 Touch the Demo Mode button.
NOTEWhen the HemoSphere advanced monitoring platform runs in Demo Mode, all
audible alarms are disabled
3 Touch Ye s on the Demo Mode confirmation screen.
4 See chapter 9: HemoSphere Swan-Ganz Module Monitoring for details on monitoring with the
HemoSphere Swan-Ganz module.
5 The HemoSphere advanced monitoring platform must be restarted prior to monitoring a patient.
WARNINGMake sure that Demo Mode is not activated in a clinical setting to ensure that simulated
data is not mistaken for clinical data.
7.5Engineering
The engineering menu can only be operated by a system engineer and is password protected. If an error is
encountered, start by referring to chapter 11: Troubleshooting .
The Export Data screen lists a number of data export features of the HemoSphere advanced monitor. This
screen is password protected. From this screen clinicians can export diagnostic reports, delete monitoring
sessions, or export monitoring data reports. For more on exporting monitoring data reports, see below.
8.1.1Data Download
The Data Download screen allows the user to export monitored patient data to a USB device in Windows
Excel XML 2003 format.
NOTEThe screen will return to the monitoring view after two minutes of inactivity.
1 Touch the settings icon .
2 Touch the Export Data button.
3 Enter password when prompted in Export Data Password popup window.
4 Make sure an approved Edwards USB device has been inserted.
CAUTIONUse a virus scan on any USB stick before inserting to prevent a virus or malware
infection.
5 Touch the Data Download button.
Monitoring Data. To generate a spreadsheet of monitored patient data:
89
Page 90
HemoSphere Advanced Monitor8 Data Export and Connectivity Settings
1 Touch the value side of the Interval button and select the frequency of the data to download. The
shorter the frequency, the greater the amount of data. Options are:
•20 seconds (default)
•1 minute
•5 minutes
2 Touch the Start Download button.
NOTEDo not disconnect the USB device until the “Download complete” message appears.
If a message appears stating that the USB device is out of space, insert a different USB
device and restart the download.
All monitored patient data may be cleared by the user. Touch the Clear All button and confirm to clear.
8.2 Clear Data and Settings
The Clear Data and Settings screen allows the user to restore the factory defaults. For more information
on factory defaults, see below.
8.2.1Restore Factory Defaults
When the defaults are restored, the HemoSphere advanced monitor stops all functions and restores the
system to a factory default state.
CAUTIONRestore Defaults replaces all settings with factory defaults. Any settings changes or
customizations will be permanently lost. Do not restore defaults while monitoring a
patient.
1 Touch the settings icon .
2 Touch the Advanced Setup button.
3 Enter Advanced Setup Password. See service manual for clinician passcode.
4 Touch Clear Data and Settings button.
5 Touch Restore Factory Defaults button.
6 A confirmation screen appears. Touch Ye s to continue.
7 Turn the monitor power off and then follow the start-up process.
8.3 Wireless Settings
The HemoSphere advanced monitor can connect to available wireless networks.
1 Touch the settings icon .
2 Touch the Advanced Setup button and enter password. See service manual for clinician
passcode.
90
Page 91
HemoSphere Advanced Monitor8 Data Export and Connectivity Settings
3 Touch the Wireless button.
4 Select the desired wireless network from the list of available connections and enter the password
if required.
NOTEDo not connect to an unrecognized or unsecured network. See Cyber Security on page 94.
Wi-Fi connection status is indicated on the information bar by the symbols shown in table 8-1.
Table 8-1 Wi-Fi connection status
Wi-Fi SymbolIndication
very high signal strength
medium signal strength
low signal strength
very low signal strength
8.4 HIS Connectivity
The HemoSphere advanced monitor has the ability to interface with the Hospital Information
Systems (HIS) to send and receive patient demographics and physiological data. The
HemoSphere advanced monitor supports Health Level 7 (HL7) messaging standard and implements
Integrating Healthcare Enterprise (IHE) profiles. HL7's version 2.6 messaging standard is the most
commonly used means for electronic data exchange in the clinical domain. Use a compatible interface to
access this feature. The HemoSphere advanced monitor HL7 communication protocol, also referred to as
HIS Connectivity, facilitates the following types of data exchanges between the HemoSphere advanced
monitor and external applications and devices:
•Sending of physiological data from the HemoSphere advanced monitor to the HIS and/or medical
devices
•Sending of physiological alarms and device faults from the HemoSphere advanced monitor to the
HIS
no signal strength
no connection
91
Page 92
HemoSphere Advanced Monitor8 Data Export and Connectivity Settings
•HemoSphere advanced monitor retrieval of patient data from the HIS.
Figure 8-1 HIS- Patient query screen
HIS connectivity status is indicated on the information bar by the symbols shown in Table 8-2.
Table 8-2 HIS connectivity status
HIS symbolIndication
Connection to all configured HIS actors are good.
Unable to establish communication with configured HIS
actors.
Patient ID is set to “Unknown” in all outbound HIS
messages.
Intermittent errors are occurring in communications with
configured HIS actors.
Persistent errors are occurring in communications with
configured HIS actors.
8.4.1Patient Demographic Data
The HemoSphere advanced monitor, with HIS Connectivity enabled, can retrieve patient demographics data
from enterprise application. Once the HIS Connectivity feature is enabled, touch the Query button. The
Patient Query screen allows the user to search for a patient based on name, patient ID or room and bed
information. The Patient Query screen can be used to retrieve patient demographics data when starting a
new patient or to associate the patient physiological data being monitored on the HemoSphere advanced
monitor with a patient record retrieved from HIS.
92
Page 93
HemoSphere Advanced Monitor8 Data Export and Connectivity Settings
Once a patient is selected form the query results, patient demographics data is displayed in the New Patient
Data screen.
Figure 8-2 HIS- New patient data screen
The user can enter or edit patient height, weight, age, gender, room and bed information on this screen. The
selected or updated patient data can be saved by touching the home icon . Once patient data is saved,
the HemoSphere advanced monitor generates unique identifiers for the selected patient and sends out this
information in outbound messages with physiological data to the enterprise applications.
8.4.2Patient Physiological Data
The HemoSphere advanced monitor can send monitored and calculated physiological parameters in
outbound messages. Outbound messages can be sent to one or more configured enterprise applications.
Continuously monitored and calculated parameters with the HemoSphere advanced monitor can be sent to
the enterprise application.
8.4.3Physiological Alarms and Device Faults
The HemoSphere advanced monitor can send physiological alarms and device faults to configure HIS.
Alarms and faults can be sent to one or more configured HIS. Statuses of individual alarms including change
in states are sent out to the enterprise application.
For more information on how to receive access to HIS Connectivity, contact your local Edwards
representative or Edwards Technical Support.
WARNINGDo not use the HemoSphere advanced monitor as part of a Distributed Alarm System.
The HemoSphere advanced monitor does not support remote alarm monitoring/
management systems. Data is logged and transmitted for charting purposes only.
93
Page 94
HemoSphere Advanced Monitor8 Data Export and Connectivity Settings
8.5 Cyber Security
This chapter outlines ways in which patient data can be transferred to and from the HemoSphere advanced
monitor. It is important to note that any facility using the HemoSphere advanced monitor must take
measures to protect the privacy of a patients personal information in accordance with country specific
regulations, and consistent with the facility’s polices for managing this information. Steps that can be taken
to safeguard this information and the general security of the HemoSphere advanced monitor include:
•Physical Access: Limit use of the HemoSphere advanced monitor to authorized users.
•Active Use: Users of the monitor should take measures to limit patient data storage. Patient data
should be removed from the monitor after a patient is discharged and patient monitoring has ended.
•Network Security: The facility must take measures to ensure the security of any shared network
to which the monitor may be connected to.
•Device Security: Users should only use Edwards approved accessories. In addition, ensure that any
connected device is free of malware.
The use of any HemoSphere advanced monitor interface outside of its intended purpose could pose cyber
security risks. No HemoSphere advanced monitor connections are meant to control the operations of
another device. All available interfaces are shown in HemoSphere Advanced Monitor Connection Ports on page 35
and specifications for these interfaces are listed in table A-5, “HemoSphere advanced monitor technical
specifications,” on page 134.
8.5.1HIPAA
The Health Insurance Portability and Accountability Act of 1996 (HIPAA), introduced by the U.S.
Department of Health and Human Services, outlines important standards to protect individually identifiable
health information. If applicable, these standards should be followed during monitor use.
The HemoSphere Swan-Ganz module is compatible with all approved Edwards Swan-Ganz pulmonary
artery catheters. The HemoSphere Swan-Ganz module acquires and processes signals to and from a
compatible Edwards Swan-Ganz catheter for CO, iCO and EDV/RVEF monitoring. This section provides
an overview of the HemoSphere Swan-Ganz module connections. See figure 9-1.
4 Connect the compatible Swan-Ganz catheter to the patient CCO cable. See table 9-1 below for
available parameters and required connections.
Table 9-1 Available HemoSphere Swan-Ganz module parameters and required connections
ParameterRequired connectionSee
COthermistor and thermal filament connectionContinuous Cardiac Output on page 98
iCOthermistor and injectate (bath or in-line) probeIntermittent Cardiac Output on page 101
EDV/RVEF
(SV)
SVRthermistor and thermal filament connection
thermistor and thermal filament connection
*HR slaved-in by HemoSphere advanced monitor
*MAP and CVP slaved-in by HemoSphere advanced
monitor
EDV/RVEF Monitoring on page 107
SVR on page 110
5 Follow the necessary directions for monitoring. See Continuous Cardiac Output on page 98,
Intermittent Cardiac Output on page 101 or EDV/RVEF Monitoring on page 107.
9.1.1Patient CCO Cable Test
To test the integrity of the Edwards patient CCO cable, perform a cable integrity test. It is recommended to
test the integrity of the cable as part of a troubleshooting process. This does not test the injectate temperature
probe connection of the cable.
To access the patient CCO cable test window, touch the clinical actions icon
Tes t icon . Refer to figure 9-2 for numbered connections.
Figure 9-2 Patient CCO cable test connections
Patient CCO Cable
1 Attach the patient CCO cable to the inserted HemoSphere Swan-Ganz module .
2 Attach the patient CCO cable thermal filament connector and thermistor connector to their
mating test ports on the HemoSphere Swan-Ganz module.
3 Touch the Start button to begin the cable test. A progress bar will appear.
4 Replace the patient CCO cable if it fails the cable test.
5 Touch the enter icon when the cable has passed. Disconnect the patient cable thermal
filament connector and thermistor connector from the HemoSphere Swan-Ganz module.
9.2Continuous Cardiac Output
The HemoSphere advanced monitor measures cardiac output continuously by introducing small pulses of
energy into the blood stream and measuring blood temperature via a pulmonary artery catheter. The
maximum surface temperature of the thermal filament used to release these pulses of energy within the
blood is 48 °C. Cardiac output is computed using proven algorithms derived from the conservation of heat
principles, and indicator dilution curves that are obtained by cross-correlation of energy input and blood
temperature waveforms. After initialization, the HemoSphere advanced monitor continuously measures and
displays the cardiac output in liters per minute without operator calibration or intervention.
9.2.1Connecting the Patient Cables
1 Connect the patient CCO cable to the inserted HemoSphere Swan-Ganz module as previously
described in section 9.1.
2 Attach the catheter end of the patient cable to the thermistor and thermal filament connectors on
the Swan-Ganz CCO catheter. These connections are emphasized as numbers and in
figure 9-3 on page 99.
3 Verify that the CCO catheter is properly inserted into the patient.
Figure 9-3 CO connection overview
9.2.2Initiating Monitoring
WARNINGCO monitoring should always be discontinued when blood flow around the thermal
filament is stopped. Clinical situations where CO monitoring should be discontinued
include, but are not limited to:
• Time periods when a patient is on cardiopulmonary bypass
• Partial withdrawal of the catheter so that the thermistor is not in the pulmonary
artery
• Removal of the catheter from the patient
When the system is properly connected, touch the start monitoring icon to begin CO monitoring. The
CO countdown timer will appear on the stop monitoring icon. After approximately 3 to 6 minutes, when
sufficient data has been obtained, a CO value will appear in the parameter globe. The CO value displayed on
the screen will be updated approximately every 60 seconds.
NOTENo CO value will be displayed until sufficient time-averaged data is available.
In some situations where patient conditions create large changes in pulmonary artery blood temperature
over several minutes, the monitor may take longer than 6 minutes to obtain an initial CO measurement.
When CO monitoring is in progress, updating of the CO measurement may also be delayed by unstable
pulmonary artery blood temperature. The last CO value and measurement time will be displayed in place of
an updated CO value. Table 9-2 shows the alert/fault messages that appear on the screen at different time
points while the signal stabilizes. Refer to table 11-6, “HemoSphere Swan-Ganz module CO faults/alerts,”
on page 123 for more information on CO faults and alerts.
Table 9-2 Unstable thermal signal time lapse for CO alert and fault messages
Alert COFault CO
Condition
Monitoring Commencing: minutes from
commencement without CO measurement
Monitoring in Progress: minutes from last
CO update
Signal
Adapting —
Continuing
61530
na620
Unstable Blood
Temp. —
Continuing
Thermal Signal
Loss
A fault condition terminates monitoring. A fault condition could result from migration of the catheter tip
into a small vessel preventing the thermistor from accurately sensing the thermal signal. Check catheter
position and reposition the catheter, if necessary. After verifying patient status and catheter position, CO
monitoring may be resumed by touching the start monitoring icon .
9.2.4CO Countdown Timer and STAT CO
The CO countdown timer is located on the stop monitoring icon . This timer alerts the user as to when
the next CO measurement will take place. The time to the next CO measurement varies from 60 seconds to
3 minutes or longer. A hemodynamically unstable thermal signal may delay CO calculations. For longer time
spans between CO measurements, the STAT CO is available. The STAT CO (sCO) is a fast estimate of the
CO value and is updated every 60 seconds. Select sCO as a key parameter to view STAT CO values. Select
CO and sCO as key parameters while viewing the graphical/tabular trends split screen and CO monitored
data is graphically plotted alongside tabular/numerical data for STAT values of sCO. See Graphical/Tabular Trends Split on page 59.
CAUTIONInaccurate cardiac output measurements may be caused by:
•Incorrect placement or position of the catheter
•Excessive variations in pulmonary artery blood temperature. Some examples that
cause BT variations include, but are not limited to:
* status post cardiopulmonary bypass surgery
* centrally administered cooled or warmed solutions of blood products
* use of sequential compression devices
•Clot formation on the thermistor
•Anatomical abnormalities (for example, cardiac shunts)
100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.