Druck UPS III Operating Manual

GE
Sensing
Druck UPS-III Loop Calibrator
User manual -K0317

Approved Service Agents

For the list of service centres visit our web site:
www.gesensing.com
This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark.
This symbol, on the instrument, indicates that the user should refer to the user manual.
Do not dispose of this product as household waste. Use an approved organisation that collects and/or recycles waste electrical and electronic equipment. For more information:
Contact us at www.gesensing.com

UPS III Loop Calibrator

Introduction
The Druck UPS III Series of loop calibrators can supply power (source mode) and produce readings (measure mode) to perform f ield calibrations on 2-wire devices. The set-up menu enables the user to “source” or “measure” in either voltage or current and to perform continuity tests. These user instructions include the operation, safety instructions and installation requirements for the loop calibrator.

Specifications

Accuracy
Values in table below includes temperature effects 17°C to 27 °C
Outside these limits .................. 0.003%/°C(0.0015%/°F)
Calibration Reference................ 22°C ±1°C/RH45%±15%
Mode Range Accuracy Remarks
Source mA 0 to 24 mA* 0.01% rdg + 2 lsd V-max. 75V Source mA +
24V Measure mA 0 to 24 mA* 0.01% rdg + 2 lsd V-max. 75V Measure mA
+ 24V Measure V 0 to 60V* 0.02% rdg + 4 lsd R-measure
Continuity <100Ω ** -1 mA
* Resolution 0.001 lsd least significant digits ** Audio + visual rdg reading
Hart® communications...............menu selectable 220Ωloop resistor
Operating Temperature...... -10°C to 50°C (-14°F to 122°F)
Storage Temperature......... -20°C to 70°C (-4°F to 158°F)
Conforms to ......... EN61010, EN 61326-1(1997)+ A1(1998)
Electrical Power Supply
.................................... 4 x 1.5 V alkaline size AA or
Universal power supply {see accessories}
0 to 24 mA* 0.01% rdg + 2 lsd R-max 1k
0 to 24 mA* 0.01% rdg + 2 lsd R-measure
This loop calibrator meets the essential protection requirements of the relevant EEC directives.
at 20 mA
15
Ω
Ω
1M
Ω
1 K0317 Issue No. 3
Physical
Dimensions......................... 77 x 129 x 24 mm(3” x 5” x 1”)
Weight .................................................. 275 grams (9.7 oz.)
Terminals ...................................4 mm sockets {gold plated}
Case ....................................................... High impact ABS
Relative Humidity....................................................0 to 90%

Safety

This symbol, on the loop calibrator, indicates that the user should refer to the user guide or manual.
Batteries
Remove batteries from the loop calibrator immediately when discharged and before storage.
Dispose of batteries in accordance with local regulations and battery manufacturers' instructions.
When storing and transporting batteries make sure they cannot be short circuited.
Power Supply
The power supply for this loop calibrator can be the internal non-rechargeable batteries or the external Universal power supply unit (see accessories).
Battery life >
The display shows with low battery power.
Battery Replacement
Unscrew and remove the securing screw from the battery panel.
Replace the batteries, check the polarity of the batteries.
Refit and secure the battery panel.
Accessories
Assy 305 Test lead set 191-129 Power supply, Universal, 100-240 V a.c. 47-63Hz 38016 Carrying case 38023 Protective rubber boot
75 hours in measure mode 18 hours at 12 mA (source mode)
>
K0317 Issue No. 3 2

OPERATION

Keys
The key switches the calibrator on and off. Press and hold for 2 seconds.
The key changes the measure or source operating mode. Pressing the
keys makes menu selections, sets numerical values and controls step and ramp functions (up/down).
The select advanced functions shown on the bottom of the display. When no key is pressed for 10 minutes, the calibrator times out and switches off. To disable this automatic time out, select autpower down in the set-up menu.
3 K0317 Issue No. 3

Operating Modes

F1
F1
Serial number
VERSION
CAL. DATE
HART
BATTERY
Information
F2
F2
Resolution
adjust
F2
Contrast
adjust
F2
Range
4-20 mA/0-20mA
F2
Source unit
mA/%
F2
Hart® ON/OFF
F2
Aut power down
ON/OFF
F2
Decimal
.,/
F2
Access code
menu
F2
Calibration
menu
F2
F1
F2
F2
Pressing
the start-up sequence. Pressing , at this time, the display shows the information screen:
Pressing , at this time, the display shows the set-up screen:
switches the instrument on and the display shows
The calibrator can be used in two modes
measu re
or
source

Measure mode

The display shows the measured value; depending on the settings made in set-up and advanced settings:
When measuring current pressing enables linear or flow,
pressing enables mA or % (value of 4 to 20 mA or 0 to 20 mA).
When measuring voltage pressing changes the resolution between 0.00V and 0.000V.
To measure continuity the displays shows an open or closed switch symbol with an audible signal on switch closure.
K0317 Issue No. 3 4
.
Loading...
+ 14 hidden pages