Test Certificate
Phototherapy Unit 4000
Location of appaMU*:
File No..:
Edition: 01.97
6141.420
1.
1.1
1.2
1.3
C= Check cmditim
L = Check fa leakage
Check condition
Housing The unit must not be free from contaminations which may reduce the
safety. C
Plexiglass guard 2M 21335
On/off switch “white” 2M 21333
C
C
serial No.:
stamp: InAce No. or
delivery No.:
I
I
I
I I I I I I
I
I
1.4
1.5
1.6
1.7
1.8
2. Check function
2.1
On/off switch “blue” 2M 21332
Elapsed time meter 2M 21329 or elapsed time meter with LCD display 1840460.
Mains connection 2M 21327 with fuses T1,0A1815083orfusesT1.6A1815105
(only for 90/100 V unit).
Power cord 1807331
Visual inspection of lamp mount including provided clamps in the refector.
Enter operating hours
C
C
C
C
V
I
I
I I
I I
I
I I I I
I I I I
I I 1
I I I I
I I I I
I
I
I
I
Dr
2.2 *
D3
n/a
n/a
n/a
n/a
n/a
n/a
n/a
Radiation source lighting blue 2M 21010
(4 or 6 pieces), replace after 1000 hours of operation. Note: Normally 4 blue and 2 white lamps are mounted. To obtain a higher bilirubin­effective radiant intensity the 2 white light lamps can be replaced with blue light lamps.
Date of next replacement:
2.3
2.3.1
2.3.2
3.
3.1
3.2 Equivalent leakage current
Function
Switch on “blue” on/off switch: 4 blue lamps 2M 21010 light up and elapsed time meter is running.
Switch on “white” on /off switch: 2 white lamps 2M 21009 or 2 blue lamps 2M 21010 light up
Electrical safety check
Protective conductor resistance
Test value: R < 0.2 R
Switch on device with both switches.
Follow up measurements may exdceed the first-measurement value in the Test Certificate by max. 50%, however the stated test value must not be exceeded.
0
0
I
I
I I
I
I
Test value: IA 5 1 .O mA
Note: The first-measured value must be transferred to a new Test Certificate.
First-measured value: IA
4. Stand 2 different versions are possible:
4.1 Stand 2M 16760 (from PT 600/600)
4.1.1 Height adjustment
4.1.2 Castors 2M 15116 (2 pieces)
4.1.3 Locking castors 2M 20761 (2 pieces)
4.2 Stand 2M 21190 (new)
$
0
CO
0
co co
I
I
I
I I
I
I I
I
I
I