Daktronics Venus 7000 User Manual

Page 1
Daktronics Venus® 7000 Controller Track Setup Guide
Figure 1: Venus 7000 Shell Menu
Figure 2: RTD Inputs
Figure 3: RTD Input 1 settings for a serial connection from the MEET MANAGER
Venus® is a registered trademark of Daktronics, Inc. FinishLynx™ is a trademark of the Lynx System Developers, Inc. Windows® and DOS® are registered trademarks of Microsoft, Inc. Hy-Tek Meet Manager® is a registered trademark of Hy-Tek ltd.
NOTE: This reference is intended as a guide to help configure the Venus 7000 controller to interface with Hy-Tek’s MEET MANAGER for track and FinishLynx photo finish system.
STEP 1 Configure RTD Inputs
Click on <Configure> shown in Figure 1. Next, select the RTD Inputs file tab as seen in Figure 2. To edit an input, highlight a number (i.e. 1:) and click on <Edit Input>.
STEP 2 Configure RTD Input 1
Choose Input 1 and configure it per Figure 3 for a serial connection from Hy-Tek’s MEET MANAGER.
Daktronics, Inc. PO Box 5128, 331 32nd Ave. Brooking, SD 57006 ED-12636 REV 1 www.daktronics.com tel (605) 697-4036 or (877) 697-1115 fax (605) 697-4444 e-mail: helpdesk@daktronics.com
Page 1 of 10
Page 2
Daktronics Track Venus® 7000 Controller COM Port Troubleshooting Guide
Figure 4: RTD Input 2 Settings for
Run Time from a serial comm. port on the Lynx Capture Station.
Figure 5: RTD Input 3 Settings for Results from a serial comm. port on the Lynx Capture System.
STEP 3 Identify how FinishLynx will connect to the Venus 7000.
Option 1 is through a serial comm. port. Option 2 is over
the network via a UDP/IP socket. One thing to consider when making this decision is that if you use serial ports and you want to send both results and running time from the FinishLynx capture computer, it is best to use two separate scoreboard outputs. If you use serial ports, then you need two serial ports. If you use the network, you just need to configure two UDP/IP scoreboard outputs, but no further hardware is required. If you decide to use the network, you will need the NCP plug-in from FinishLynx. It is possible to send both run time and results on the same serial port, but while the run time is counting on the scoreboard, there will be no results sent over. Either stop the run time manually, or by using the internal photocell in FinishLynx. After you have stopped the run time, when the next result is picked, it will be sent to the scoreboard.
STEP 4 Configure RTD Inputs 2, 3 & 4. Input 2 will be the running time from FinishLynx Capture
computer. Input 3 will be the start lists and results from FinishLynx Capture computer and Input 4 will be the start lists and results from FinishLynx Edit Computer.
Serial Port Connection Choose Input 2 and configure it per Figure 4. Choose
Input 3 and configure it per Figure 5. Choose Input 4
and configure it per Figure 6.
OR
Network via a UDP/IP socket connection
Choose Input 2 and configure it per Figure 7. Choose
Input 3 and configure it per Figure 8. Choose Input 4
and configure it per Figure 9.
Daktronics, Inc. PO Box 5128, 331 32nd Ave. Brooking, SD 57006 ED-12636 REV 1 www.daktronics.com tel (605) 697-4036 or (877) 697-1115 fax (605) 697-4444 e-mail: helpdesk@daktronics.com
Page 2 of 10
Page 3
Daktronics Track Venus® 7000 Controller COM Port Troubleshooting Guide
Figure 6: RTD Input 4 Settings for
Results from a serial comm. port on the Lynx Edit Station.
Figure 7: RTD Input 2 Settings for Run Time over a UDP/IP network socket from the Lynx Capture Station.
Figure 8: RTD Input 3 Settings for Results over a UDP/IP network socket from the Lynx Capture Station.
Figure 9: RTD Input 4 Settings for Results over a UDP/IP network socket from the Lynx Edit Station.
Daktronics, Inc. PO Box 5128, 331 32nd Ave. Brooking, SD 57006 ED-12636 REV 1 www.daktronics.com tel (605) 697-4036 or (877) 697-1115 fax (605) 697-4444 e-mail: helpdesk@daktronics.com
Page 3 of 10
Page 4
Daktronics Venus® 7000 Controller Track Request Guide
Figure 10: Start List
NOTE: The Track Request program is used to make requests for Start Lists, Results and Team scores from the Venus 7000 to the MEET MANAGER.
STEP 1 Installing Track Request.
Install the Track Request program onto the Venus 7000 using the Track Request.msi installation file from Daktronics.
STEP 2 Operation of Track Request. To open the Track Request program, double-click on the
icon found on the desktop of the Venus 7000 computer. Once you open the program, leave it open to keep it ready for the next request.
The parameters on the left side of the window are used to
select which information to display.
Round: determines if the data is from a preliminary,
semi-final, or final event. If there are no prelims or
semi-finals, the finals will be round 1. Event: is the event as numbered in the MEET
MANAGER.
Heat: is the heat as numbered in the MEET MANAGER. Multi: is used for decathlons, heptathlons and
pentathlons. Example, the Event# is the number of the decathlon, the Multi# is the number of the sub-event within the decathlon like the shot put. This is numbered in the MEET MANAGER. The Multi: field will not affect the lists unless the Event: field is a
multi-disciplined event. Page: will select which lines of data are displayed. The
page size is set in the MEET MANAGER. If the size is set to 8 and you have 24 competitors, page 1 will show place 1-8, page 2 will show 9-16 and page 3 will show 17-
24. If you enter page 0, the competitors will cycle 1-8, 9­16, 17-24 automatically. The page time is configured in the MEET MANAGER.
The radio buttons under the Request Type: are used to
select which type of data you would like. Choose Start
List, Result List, Women’s Scores, Men’s Scores, or Awards. Awards are similar to the results, except they
will only display the athletes that received awards in that event.
STEP 3 Start Lists.
Refer to Figure 10. To request a Start List, fill in the
appropriate Round, Event, Heat, Multi and Page. Next click on the Start List radio button under the Request Type:. Now press the large <REQUEST> button to request the selected information from MEET MANAGER.
STEP 4 Results Lists.
Refer to Figure 11. To request Results, fill in the
appropriate Round, Event, Heat, Multi and Page. To select a final result for multiple heats, enter 99 for the Heat. Next click on the Result List radio button under the Request Type:. Now press the large <REQUEST> button to request the selected information from MEET MANAGER.
Daktronics, Inc. PO Box 5128, 331 32nd Ave. Brooking, SD 57006 ED-12636 REV 1
www.daktronics.com tel (605) 697-4036 or (877) 697-1115 fax (605) 697-4444 e-mail: helpdesk@daktronics.com
Page 4 of 10
Page 5
Daktronics Venus® 7000 Controller Track Request Guide
Figure 11: Results List
Figure 12: Multi-Event List
Figure 13: Team Scores List
STEP 5 Multi-event Results.
Decathlons, heptathlons and pentathlons have multiple disciplines all in one event. To display each discipline, the event number must be chosen for the multi-discipline event. For example, if the decathlon is event 40, enter 40 into the Event: field and then change the Multi: field to the discipline desired. If the 100 meter dash is the 1st discipline of a decathlon, set Event: to the proper event number and the Multi: to 1, then click request. Refer to Figure 12. The Multi: field will not affect the lists unless the Event: field is a multi-disciplined event.
STEP 6 Team Score.
Refer to Figure 13. To display team scores, select the
Women’s Scores or the Men’s Scores and click on
<REQUEST>. When requesting team scores, the values for
Round:, Event:, Heat: and Multi: are ignored. However, the Page: field operates normally.
Daktronics, Inc. PO Box 5128, 331 32nd Ave. Brooking, SD 57006 ED-12636 REV 1
www.daktronics.com tel (605) 697-4036 or (877) 697-1115 fax (605) 697-4444 e-mail: helpdesk@daktronics.com Page 5 of 10
Page 6
Daktronics Venus® 7000 Controller Track Request Guide Troubleshooting
If you request an event or heat that does not exist, nothing
will happen.
A Hy-Tek sequence has 5 frames. Frame 1 is for the start
list, 2 is for results, 3 and 4 are teams scores and 5 is awards. Frames 2 and 5 are typically identical and frames 3 and 4 are identical.
If the results are flashing, be sure the Venus 7000
controller sequence Hy-Tek is running in “selected frame” mode and not in “continuous” mode.
When using the track request program, Port is associated
to the Venus 7000 controller input that the MEET MANAGER is connected to. The Port defaults to the
Venus 7000 controller Input 1. The COM Port and protocol settings for the input must be changed in the Venus 7000 controller configuration menu.
Verify that you have the Alpha Scoreboard option in your
Hy-Tek MEET MANAGER for track. Configure MEET MANAGER to use the Daktronics Fixed Digit output. Next identify and configure the comm. port on the Hy-Tek computer that will connect to the Venus 7000. Last set the baud rate in MEET MANAGER to 9600.
You can also push data from the MEET MANGER by
pressing Ctrl-F1 while in the Run menu of MEET MANAGER. A menu will pop-up similar to the Track Request.
Refer to Daktronics Dwgs B-127739 and B-267220 for
sample system riser diagrams.
Refer to Daktronics ED-14511 for instructions on setting
up MEET MANAGER and FinishLynx.
Daktronics, Inc. PO Box 5128, 331 32nd Ave. Brooking, SD 57006 ED-12636 REV 1
www.daktronics.com tel (605) 697-4036 or (877) 697-1115 fax (605) 697-4444 e-mail: helpdesk@daktronics.com
Page 6 of 10
Page 7
Daktronics Track Venus® 7000 Controller COM Port Troubleshooting Guide
Figure 14: RTD Input Diagnostic menu with no data Figure 15: RTD Input Diagnostic menu with Data
STEP 1 Choose the RTD input to test
From the Venus 7000 Shell Menu (refer to Figure 1 in the
Track Setup Guide), click on <DIAGNOSTICS>. In the
Diagnostics menu, select the RTD file folder tab.
STEP 2 Test Hy-Tek Meet Manager
In the Input: data box, select Input 1. If the RTD Diagnostic menu shows Input 1 under the Input: data box, click on <REFRESH>. Refer to Figure 14.
STEP 4 Test Analysis
If data was received, numbers and names will appear on the right side of the colon as seen in Figure 15. This data should also be displayed on your scoreboard display system. The data on the left side of the colons are the Item
Number ranges.
STEP 3 Test Analysis
After clicking <REFRESH>, examine the data in the message
box on the bottom left of the window as seen in Figure
14. If only the numbers on the left side of the colons are displayed, you have received no data. Try sending data from MEET MANAGER again. If no data is received (check data by clicking <REFRESH>), inspect cable routing and connections.
Daktronics, Inc. PO Box 5128, 331 32nd Ave. Brooking, SD 57006 ED-12636 REV 1
www.daktronics.com tel (605) 697-4036 or (877) 697-1115 fax (605) 697-4444 e-mail: helpdesk@daktronics.com
Page 7 of 10
Page 8
Venus® 7000 Controller Track & Field Real Time Data
Figure 16: Insert Data Frame
Figure 17: Frame Properties
STEP 1 Create Data Frame When creating a new RTD sequence, a Data Frame must
be added. To do this, open the Frame menu, then select
Insert and then Data Frame. Refer to Figure 16 below
STEP 3 Hy-Tek sequences
Rename your file “Hy-Tek”. From there, build a
Important Field Type: select RTD
frame similar to Figure 19. Frame #1 will be for Start Lists. To start the RTD Template, click <PROPERTIES> on the bottom of the window. A Frame Properties window will appear.
Frame Properties Data Box Description: (Refer to Figure 17) Field: applies box number Row &Col: Horizontal & Vertical position Font: Size of letters and numbers Overlay: Outline used for video background
RTD Field: Individual field properties Input: COM Port Source Item #: Serial bit on data stream Length: Number of characters after selected item
number
Justification: Centering or spacing left or right Field Name: Automatically sets item number length
for selected choice. (Refer to STEP 4)
Sample Text: Text that will be seen on Sequence Designer but not on the Scoreboard
STEP 2 Delete Graphic Frame
After adding a data frame, there will be two frames.
Frame 1 must be deleted. On the bottom of the window, the frame must be Frame 1 of 1 Type: Data and not Type: Graphic. Refer to Figure 16. To delete a frame,
click on , or right-click on the frame to be deleted. Highlight Frame option, and then select Delete. Click on
<DELETE CURRENT>.
Daktronics, Inc. PO Box 5128, 331 32nd Ave. Brooking, SD 57006 ED-12636 REV 1
www.daktronics.com tel (605) 697-4036 or (877) 697-1115 fax (605) 697-4444 e-mail: helpdesk@daktronics.com
Page 8 of 10
Page 9
Venus® 7000 Controller Track & Field Real Time Data
Figure 18: Data Frame Parameters
Figure 19: Hy-Tek Frame 1
Figure 20: Hy-Tek Frames 2 and 5
STEP 4 Verify the Data Frame Parameters Refer to Figure 18. These will need to be verified for
each frame. For the Hy-Tek sequence, Input 1, frames 1, 2 and 5 will be set to hy track ln.itf and Frames 3 and 4 will be set to hy track scr.itf.
STEP 5 Start creating RTD fields Pick the Frame Properties file tab, next select Field
Type: RTD. Now the RTD Field should display that fields parameters. Click on the Field Name, choose the desired information for that field. After entering each
RTD field and it’s respective parameters, the frame should
appear similar to Figure 19.
STEP 6 Additional Hy-Tek frames The Hy-Tek sequence should consist of five frames: Start
List, Results List, Women’s Scores, Men’s Scores, and Awards. Refer to Figure 19, Figure 20 and Figure 21 for
frame setup.
NOTE: Women’s Score and Men’s Score frames are duplicates. Once one frame is created, it can be copies to the other frame.
Daktronics, Inc. PO Box 5128, 331 32nd Ave. Brooking, SD 57006 ED-12636 REV 1
www.daktronics.com tel (605) 697-4036 or (877) 697-1115 fax (605) 697-4444 e-mail: helpdesk@daktronics.com
Page 9 of 10
Page 10
Venus® 7000 Controller Track & Field Real Time Data
Figure 21: Hy-Tek Frames 3 and 4 Figure 22: Lynx Start
Figure 23: Lynx Results
STEP 7 FinishLynx RTD Sequences Following the same instructions as the Hy-Tek in
STEP 3, Create a Lynx Start list sequence. After that is
completed, create a separate sequence Lynx Results list. Refer to Figure 22 and Figure 23.
Daktronics, Inc. PO Box 5128, 331 32nd Ave. Brooking, SD 57006 ED-12636 REV 1
www.daktronics.com tel (605) 697-4036 or (877) 697-1115 fax (605) 697-4444 e-mail: helpdesk@daktronics.com
Page 10 of 10
Loading...