Nellcor Puritan Bennett Incorporated is an affiliate of Tyco Healthcare.
Nellcor, Oxiband, Durasensor, OxiCliq, Dura-Y, MAX-FAST, and O
XIMAX
are trademarks of Nellcor Puritan Bennett Incorporated or an affiliate.
Cet appareil ISM est conforme à la norme NMB-001 Canada.
To obtain information about a warranty, if any, contact Nellcor’s Technical
Services Department, or your local representative.
Purchase of this instrument confers no express or implied license under any
Nellcor Puritan Bennett patent to use the instrument with any sensor that is
not manufactured or licensed by Nellcor Puritan Bennett.
Covered by one or more of the following U.S. Patents and foreign
equivalents: 4,802,486; 4,869,254; 4,928,692; 4,934,372; 4,960,126;
5,078,136; 5,485,847; 5,743,263; 5,865,736; 6,035,223; 6,298,252; 6,463,310;
6,591,123; 6,675,031; 6,708,049; 6,801,797; and Re.35,122.
Contents
Safety Information ...................................................................... 1
Index ........................................................................................ 137
ivOXIMAX N-65 Operator’s Manual
Safety Information
Warnings
Warnings are identified by the WARNING symbol
shown above.
Warnings alert the user to potential serious outcomes,
such as death, injury, or adverse events to the patient or
user.
WARNING: The OXIMAX sensor uses the date and
time provided by the
oximeter when the sensor event record is recorded by
the sensor. The accuracy of the date/time is dependent
on the date/time already set in and provided by the
OXIMAX N-65.
OXIMAX N-65 handheld pulse
WARNING: Explosion hazard. Do not use the
XIMAXN-65 in the presence of flammable anesthetics
O
mixture with air, oxygen, or nitrous oxide.
WARNING: Chemicals from a broken LCD display
panel are toxic when ingested. Use caution when the
OXIMAX N-65 has a broken display panel.
OXIMax N-65 Operator’s Manual1
Safety Information
WARNING: Routinely monitor the patient to make
sure the O
XIMAX N-65 is functioning and the sensor is
correctly placed.
WARNING: Pulse oximetry measurements and pulse
signals can be affected by certain environmental
conditions, O
XIMAX sensor application errors, and
certain patient conditions. See the appropriate
sections of this manual for specific safety information.
WARNING: The use of accessories, sensors, and
cables other than those specified may result in
increased emission and/or create invalid readings of
the OXIMAXN-65.
WARNING: Failure to cover the OXIMAX sensor site
with opaque material in high ambient light conditions
may result in inaccurate measurements.
WARNING: Do not silence the audible alarm
function, or decrease the audible alarm volume, if
patient safety could be compromised.
WARNING: The OXIMAX N-65 is a prescription
device to be operated only by trained personnel. The
monitor is for attended monitoring only.
2OXIMax N-65 Operator’s Manual
Safety Information
WARNING: Dispose of batteries in accordance with
local ordinances and regulations.
WARNING: The OXIMAX N-65 is not
defibrillator-proof. However, it may remain attached
to the patient throughout defibrillation or while an
electrosurgical unit is in use. The measurements may
be inaccurate throughout the defibrillation, or use of
an electrosurgial unit, and shortly thereafter. To
avoid shock, the caregiver should not hold the
OXIMAX N-65 while using a defibrillator on a patient.
WARNING: Disconnect the OXIMAX N-65 and Nellcor
O
XIMAX sensor from the patient throughout magnetic
resonance imaging (MRI) scanning. Induced current
could potentially cause burns.
WARNING: To ensure accurate performance and
prevent device failure, do not subject the O
XIMAX
N-65 to extreme moisture, such as direct exposure to
rain. Such exposure may cause inaccurate
performance or device failure.
WARNING: Do not use an OXIMAX N-65, OXIMAX
sensor, or cables that appear damaged.
OXIMax N-65 Operator’s Manual3
Safety Information
Cautions
WARNING: Do not lift the OXIMAX N-65 by the
sensor or extension cable because the cable could
disconnect from the monitor and the monitor may
drop on the patient.
Cautions are identified by the CAUTION symbol shown
above.
Cautions alert the user to exercise care necessary for the
safe and effective use of the O
XIMAX N-65 handheld pulse
oximeter.
Caution: All combinations of equipment must be in
compliance with IEC Standard 60601-1-1 systems
requirements.
Caution: Federal law (U.S.A.) restricts this device to
sale by or on the order of a physician.
4OXIMax N-65 Operator’s Manual
Notes
Safety Information
Notes are identified by the Note symbol shown above.
Notes contain important information that may otherwise
be overlooked or missed.
OXIMax N-65 Operator’s Manual5
Safety Information
6OXIMax N-65 Operator’s Manual
Introduction
WARNING: Do not make any clinical judgments
based solely on the O
intended only as an adjunct in patient assessment. It
must be used in conjunction with clinical signs and
symptoms.
Intended Use for the OXIMAX N-65
The OXIMAX™ N-65 handheld pulse oximeter is
indicated for continuous or spot check monitoring of
functional arterial oxygen saturation (SpO
rate of adult, pediatric and neonatal patients in hospital,
hospital type facilities, transport and mobile
environments as well as in the home care environment.
How to Use this Manual
XIMAXN-65. The monitor is
2) and pulse
All users should read this manual thoroughly. More
experienced users of the monitor can refer directly to the
topics for the information they require.
The current copy of this manual is available on the
internet at:
This section identifies the symbols, controls, displays,
and indicators on the O
XIMAX N-65.
1 — %SpO2area of display
2 — Measured %SpO2
3 — Pulse beats per minute (BPM) area of display
4 — Measured BPM
OXIMax N-65 Operator’s Manual9
Symbols, Controls, Displays, and Indicators
5 — %SpO2 Lower Alarm Limitindicator
6 — %SpO
2 Upper Alarm Limit indicator
7 — Pulse Amplitude indicator (Blip bar)
8 — Time Colon time/date field separator
9 — Adjust Time mode indicator
10 — Adjust Volume mode indicator
11 — Set Limit mode indicator
12 — BPM Upper Alarm Limit
indicator
13 — BPM Limit Changed indicator
14 — Sensor Off Patient indicator
15 — Sensor Disconnected indicator
16 — %SpO
2 Limit Changed indicator
10OXIMax N-65 Operator’s Manual
Symbols, Controls, Displays, and Indicators
17 — Low Battery indicator
18 — Data In-Sensor indicator
19 — Alarm Volume Adjust indicator
20 — Pulse Beep Tone Volume Adjust indicator
21 — BPM Lower Alarm Limit indicator
22 — Print indicator
23 — Interference indicator
24 — Pulse Search indicator
25 — Data indicator
26 — Alarm Silenced indicator
OXIMax N-65 Operator’s Manual11
Symbols, Controls, Displays, and Indicators
27 — Power button
28 — Up Arrow button
29 — Backlight button
30 — Menu button
31 — Data (Record/Print) button
32 — Down Arrow button
33 — Alarm Silence button
12OXIMax N-65 Operator’s Manual
Front Panel Symbols
The following two symbols are located on the front panel
of the monitor.
Displays the measured percent of oxygen saturation.
Displays the measured pulse beats per minute.
Symbols, Controls, Displays, and Indicators
OXIMax N-65 Operator’s Manual13
Rear Panel Symbols
The following symbols are located on the rear panel of
the monitor.
Symbols, Controls, Displays, and Indicators
1 — Label
2 — Battery Cover
See Instructions for Use
Type BF Applied Part - Not defibrillator proof
Resistant to liquid ingress
Rx Prescription only
Date of Manufacture
European union approval
14OXIMax N-65 Operator’s Manual
Canadian/USA certification mark (electrical safety)
Russian regulatory approval
Serial number
Label Symbols
These symbols display on the monitor’s labels.
Keep away from heat
Temperature limitation
Symbols, Controls, Displays, and Indicators
Protect from moisture
Fragile
Keep upright
Humidity limitation
OXIMax N-65 Operator’s Manual15
Controls
Symbols, Controls, Displays, and Indicators
This section describes the controls on the front panel of
the monitor. The controls are activated by pressing the
button that corresponds to that control. For example,
press the Alarm Silence button to control the audible
alarm.
When the monitor is turned on, if you press a button,
except the Power button, it results in either a valid or
invalid key tone. See Audible Indicators on page 24. If a
tone fails to emit when you press a button, contact
qualified service personnel.
The following monitor display shows the buttons
described in this section.
The Alarm Silence button. Silences the audible alarms.
If you press the button when no audible alarm is active,
alarms during the selected alarm silence duration are
silenced. In all cases, the alarm silence times out after the
selected alarm silence duration and auditory alarm
capability reactivates.
Alarms that occur throughout the Power-On-Self-Test
(POST) cannot be silenced.
16OXIMax N-65 Operator’s Manual
Symbols, Controls, Displays, and Indicators
Press and hold the Alarm Silence button for three
seconds to display the Alarm Silence Duration menu,
which allows you to adjust the alarm silence interval
from 30, 60, 90, or 120 seconds, or to OFF (alarms
disabled). See Set Alarm Silence Duration on page 53.
The Down Arrow button. Sequentially decreases the
parameter under adjustment by one decrement. If you
press and hold the button for more than three seconds it
may cause the decrement to repeat.
Press the Down Arrow button during normal operation
to decrease the pulse beep volume.
The Data (Record/Print) button. Stores the currently
shown SpO
2 and BPM values (snap-shot data.) You can
print the stored snap-shot data, such as a single-event
report, snap-shot, and/or sensor-event data by pressing
the Data button in print mode.
The Menu button. Press repeatedly during normal
operation to display the parameters. Parameters you can
set include:
•high and low SpO
2 limits
•high and low BPM limits
•alarm volume
•pulse beep volume
•data print
If you do not press any buttons for approximately 30
seconds, the monitor returns to normal operation and
incorporates the selected parameter value.
See Menu Structure on page 107 for additional
information about setting parameters.
OXIMax N-65 Operator’s Manual17
Symbols, Controls, Displays, and Indicators
Alarms and alarm icons are enabled when parameters are
set and display in the event of an alarm condition.
Examples of alarms and alarm icons are:
•Sensor Disconnected
•Pulse Search
•Interference
•Low Battery
•Alarm Silenced indicator
If you press the Menu button while Power-On-Self-Test
(POST) is activated, the Time/Date set displays.
Repeatedly pressing the Menu button displays the
following time/date parameters you can set:
•Hour
•Minute
•Day
•Month
•Year
After the time/date parameters are set, you can return to
the POST display.
The Backlight button. Press to toggle the backlight ON
or OFF. The backlight remains on for approximately 10
minutes.
The Up Arrow button. While in menu mode, press
repeatedly to increase a parameter by one increment.
Press and hold the button for more than three seconds to
repeat the increment.
Press the Up Arrow button during normal operation
while monitoring to increase the pulse beep volume.
18OXIMax N-65 Operator’s Manual
Note: Normal operation means:
• the monitor is turned on
• a sensor is connected to the monitor
• the sensor is applied to the patient
• the patient’s %SpO
percentage) and pulse rate readings (BM) are
being reported
• no error conditions exist
The Power button. Press to toggle the monitor power ON
or OFF. The Power button has a raised protrusion
(bump) at its center and a gloss surface finish for tactile
differentiation from other buttons.
Displays and Indicators
The monitor display as shown below includes:
•a Pulse Amplitude blip bar
•functional icons
•current measured %SpO
Symbols, Controls, Displays, and Indicators
2 (oxygen saturation
2
•pulse rate
OXIMax N-65 Operator’s Manual19
Symbols, Controls, Displays, and Indicators
Decimal points after the %SpO2 or pulse rate indicate
that the respective limits have been changed from the
power-on default values.
There are various matrices within the signal processing
algorithm. Some of these are used to assess the severity
of conditions presented to the monitor in SpO
2 and pulse
rate measurements on a patient. Individual matrices, or
combinations of them, are used to drive the icon
indicators on the monitor front panel.
The signal processing algorithm increases the amount of
data required for measurements of SpO
2 and pulse rate
dependent on the measurement conditions. Throughout
normal measurement conditions, the averaging time is six
to seven seconds.
The signal processing algorithm extends the amount of
data required beyond seven seconds for measurement
conditions in which incoming signal quality is degraded,
such as:
•low perfusion
•active patients
•ambient light
•electromagnetic interference (EMI)
If the resulting dynamic averaging time exceeds 20
seconds, the Pulse Search indicator appears and SpO
2
and Pulse Rate displays continue updating every second.
As these conditions extend, the amount of data required
continues to increase.
If the dynamic averaging time reaches 40 seconds, the
Pulse Search indicator flashes and the SpO
2 and pulse
rate displays flashing zeros to indicate a loss-of-pulse
condition.
20OXIMax N-65 Operator’s Manual
Symbols, Controls, Displays, and Indicators
Selected display elements, such as, icons or numerals,
may flash. There are three flash rates. See Table 1.
Table 1: Flash Rates
PriorityHertzDuty Cycle
High1.4 Hz to 2.8 Hz20% to 60%
Medium0.4 Hz to 0.8 Hz20% to 60%
LowN/AConstant on
%SpO
2 display:
•shows the oxygen saturation level of functional
hemoglobin
•shows two dashes throughout Sensor Disconnected
and Sensor Off Patient conditions
•flashes the SpO
2 value when the SpO2 is outside
the alarm limits
•shows a decimal point (.) after the SpO
2 value
(98.0), if alarm limits have changed from their
power-on defaults
Pulse Amplitude indicator (blip bar). Indicates the
dynamic pulse amplitude and rate. As the detected pulse
becomes stronger, more bars light with each pulse. The
reverse is true for weak pulses.
When setting alarm volume, the Pulse Amplitude blip
bar reflects the alarm volume setting.
OXIMax N-65 Operator’s Manual21
Symbols, Controls, Displays, and Indicators
Pulse Rate display:
•shows the pulse rate in beats per minute
•flashes throughout loss-of-pulse alarms and when
the pulse rate is outside of the alarm limits
•shows two dashes throughout Sensor Disconnected
and Sensor Off Patient conditions
•shows pulse rates outside of the pulse rate range
(0, 20 to 300 bpm) as the closest value within the
range
•shows a decimal point (.) after the BPM value
(123.), if alarm limits have changed from their
power-on defaults
Low Battery indicator. Flashes when 15 or fewer
minutes of battery capacity remain. The indicator
displays constantly when the battery capacity reaches
critical condition at which time the monitor indicates an
error condition and shuts down.
Alarm Silenced indicator. Displays when audible alarms
have been silenced. It flashes when the audible alarms are
disabled.
Interference indicator. Lights when the N-65 algorithm
detects the incoming signal quality is degraded.
Degradations can be caused by ambient light, electrical
noise, electrosurgical interference, patient activity, or
other causes. An intermittently lit Interference indicator
is common during patient monitoring, and indicates that
the N-65 algorithm is dynamically adjusting the amount
of data required for measuring SpO
2 and Pulse Rate.
When lit continuously, it indicates the N-65 algorithm
has extended the amount of data required for measuring
SpO
2 and Pulse Rate and, consequently, fidelity in
tracking rapid changes in these values may be reduced.
See Front Panel Description on page 9.
22OXIMax N-65 Operator’s Manual
Symbols, Controls, Displays, and Indicators
Pulse Search indicator. Displays before initial
acquisition of a pulse signal and throughout prolonged
challenging monitoring conditions. The Indicator
flashes throughout a loss-of-pulse signal.
Data In-Sensor indicator. Displays when the sensor
contains patient alarm events. The indicator flashes when
the sensor memory is full.
Upper Alarm Limit indicator. Indicates that the
displayed value is the upper alarm limit for SpO
2 or Pulse
Rate.
Lower Alarm Limit indicator. Indicates that the
displayed value is the lower alarm limit for SpO
2 or Pulse
Rate.
Data indicator. Displays when the monitor is in the store
snap-shot data or data print mode. The indicator flashes
when snap-shot data prints.
Print indicator. Displays when the monitor is in print
mode. The indicator flashes when the monitor prints
single-event, snap-shot, or patient alarm events data,
which is stored in the sensor.
Sensor Disconnected indicator. Displays when the
patient sensor is disconnected from the monitor.
Alarm Volume Adjust indicator. Displays when the
monitor is in the alarm volume adjust menu.
Pulse Beep Tone Volume Adjust indicator. Displays
when the monitor is in the pulse beep tone volume adjust
menu.
Adjust Time indicator. Displays when the monitor is in
the time/date set menu.
OXIMax N-65 Operator’s Manual23
Set Limit mode indicator. Displays when the monitor
alarm limit values, or pulse rate values, are being
adjusted.
Adjust Volume mode indicator. Displays when the
monitor alarm volume levels are being adjusted.
Sensor Off Patient indicator. Displays when the monitor
detects the sensor has come off the patient.
SNAP-SHOT DATA ID indicator. An alphanumeric
display that shows the number of the current captured
patient (SpO
ID to the data captured. The ID number is shown in the
%SpO
2 area of the display.
Audible Indicators
The monitor generates auditory signals for use as alarms,
status indicators, and feedback. See Audible Indicators
on page 121.
Symbols, Controls, Displays, and Indicators
2) and pulse rate (PR) values. It assigns an
24OXIMax N-65 Operator’s Manual
Set up the O XIM AX N-65
WARNING: To ensure patient safety, do not place the
XIMAX N-65 in any position that might cause it to fall
O
on the patient.
WARNING: As with all medical equipment, carefully
route patient cables to reduce the possibility of patient
entanglement or strangulation.
WARNING: Ensure that the speaker is clear of any
obstruction and that the speaker holes are not
covered. Failure to do so could result in an inaudible
alarm tone.
WARNING: Disconnect the OXIMAX N-65 and Nellcor
XIMAX sensor from the patient throughout magnetic
O
resonance imaging (MRI) scanning. Induced current
could potentially cause burns.
WARNING: To ensure accurate performance and
prevent device failure, do not subject the O
XIMAX
N-65 to extreme moisture, such as direct exposure to
rain. Such exposure may cause inaccurate
performance or device failure.
OXIMax N-65 Operator’s Manual25
Set up the OXIMAX N-65
WARNING: Do not use an OXIMAX N-65, OXIMAX
sensor, or cables that appear damaged.
WARNING: Do not lift the OXIMAX N-65 by the
sensor or extension cable because the cable could
disconnect from the monitor and the monitor may
drop on the patient.
WARNING: Use only the Nellcor extension cable
DEC-4 with the O
that is intended for computer use to the O
sensor port. Do not connect any device other than a
Nellcor-approved O
sensor connector. Do not use the DEC-8 with the
monitor.
XIMAX N-65. Do not attach any cable
XIMAX
XIMAX sensor to the OXIMAX
26OXIMax N-65 Operator’s Manual
List of Components
QuantityItem
1OXIMAXN-65 handheld pulse oximeter
4Alkaline “AA” size, 1.5-volt batteries
Set up the OXIMAX N-65
Connect O
1Nellcor O
pack
1Compact disk (O
operator’s manual (applicable to country of
sale)
1Quick guide adhesive label
XIMAX sensor or sensor assortment
XIMAXN-65 manuals) and/or
XIMAX Sensor to the Monitor
WARNING: Pulse oximetry readings and pulse
signals can be affected by certain ambient
environmental conditions, O
XIMAX sensor application
errors, and certain patient conditions. See the
appropriate sections of the manual for specific safety
information:
•Safety Information on page 1.
•OXIMAX Sensor Performance Considerations on
page 89.
•OXIMAX Sensors and Accessories on page 71.
Caution: Use only Nellcor-approved O
XIMAX sensors
and, if needed, the DEC-4 extension cable.
OXIMax N-65 Operator’s Manual27
Set up the OXIMAX N-65
Connect a Nellcor OXIMAX SpO2 sensor to the monitor
SpO
2 sensor port. You may use an extension cable to
provide more distance between the monitor and the
sensor. Use only the DEC-4 extension cable available
from Nellcor.
1
— SpO2OXIMAX Sensor Port
Inaccurate Monitor Measurements
Physiological conditions, medical procedures, or external
agents that may interfere with the monitor’s ability to
detect and display accurate measurements include:
•incorrect application of the O
•placement of the O
XIMAX sensor on an extremity
with a blood pressure cuff, arterial catheter, or
intravascular line
•excessive patient activity
•intravascular dyes, such as indocyanine green or
methylene blue
•externally applied coloring agents, such as nail
polish, dye, or pigmented cream
28OXIMax N-65 Operator’s Manual
XIMAX sensor
Set up the OXIMAX N-65
•failure to cover the OXIMAX sensor site with opaque
material in high ambient light conditions
•venous pulsation
•dysfunctional hemoglobin
•low perfusion
Loss-of-pulse signal can occur for the following reasons:
•the O
XIMAX sensor is applied too tightly
•defibrillation
•a blood pressure cuff is inflated on the same
extremity as the one with the O
XIMAX sensor
attached
•there is arterial occlusion proximal to the O
XIMAX
sensor
•poor peripheral perfusion
•loss of pulse/cardiac arrest
High ambient light sources that can interfere with the
performance of an SpO
2 OXIMAX sensor are:
•surgical lights (especially those with a xenon light
source)
•bilirubin lamps
•fluorescent lights
•infrared heating lamps
•direct sunlight
OXIMax N-65 Operator’s Manual29
Set up the OXIMAX N-65
30OXIMax N-65 Operator’s Manual
Battery Operation
WARNING: Dispose of battery in accordance with
local ordinances and regulations.
Battery Power
The OXIMAX N-65 uses batteries to power the monitor. A
new set of batteries provides between 15 to 40 hours of
operation depending on the battery type.
Low Battery Indicator
The Low Battery indicator flashes and a low priority
alarm sounds when approximately 15 minutes of
operation remains available. Replace the batteries. See
Battery Installation on page 33.
OXIMax N-65 Operator’s Manual31
Battery Operation
Caution: 15 minute indicator for remaining battery
operating time is approximate and is based on
Alkaline AA batteries. Remaining operating time may
be different for other types of batteries.
Caution: Periodically check the battery for corrosion.
If the O
XIMAX N-65 is to be stored for three months or
longer, remove the batteries from the monitor before
storage.
Critical Battery Indication
When the batteries are critically low, the monitor:
•displays an error message (Er 521)
•sounds a high priority alarm
•shuts down
Replace the batteries and restart the monitor. See Battery Installation on page 33.
32OXIMax N-65 Operator’s Manual
Battery Installation
To install batteries:
1. Turn off the power.
2. Pull the battery compartment latch downward toward
the bottom of the monitor, and remove the battery
access door.
3. Install four AA size batteries as shown in Figure 1.
Battery Operation
OXIMax N-65 Operator’s Manual33
Battery Operation
4. Replace the battery access door.
1 — Battery compartment latch
Figure 1: Battery Installation
WARNING: Explosion hazard. Do not use the
XIMAX N-65 in the presence of flammable anesthetics
O
mixed with air, oxygen, or nitrous oxide.
WARNING: Dispose of batteries in accordance with
local ordinances and regulations.
34OXIMax N-65 Operator’s Manual
O XIM AX N-65 Operation
Introduction
The parameters of the OXIMAX N-65 are set at the factory
according to Table 2. The parameters may be set on an
individual basis by the clinician.
Note: This section contains detailed information on
setting parameters. For quick reference information, see
Set Menu Parameters on page 108 and Set Time/Date
Parameters on page 110.
Parameters remain in effect until the monitor is turned
off.
Table 2: Parameter Ranges and Defaults
Parameter
%SpO2 Upper
Alarm Limit
%
SpO2 Lower
Alarm Limit
Pulse Rate Upper
Alarm Limit
Pulse Rate Lower
Alarm Limit
OXIMax N-65 Operator’s Manual35
Ranges/
Selections
Lower Alarm
Limit plus 1 to
100%
20% to Upper
Alarm Limit
minus 1
Lower Alarm
Limit plus 1 to
250 bpm
30 bpm to Upper
Alarm Limit
minus 1
Factory
Defaults
100%
85%
170 bpm
40 bpm
OXIMAX N-65 Operation
Table 2: Parameter Ranges and Defaults (Continued)
Parameter
Alarm Silence
Duration
Alarm Volume1 to 1010
Pulse Beep
Vo l u m e
Turn On the OXIMAX N-65
Discussion
Verify the monitor works properly and is safe to use.
Proper operation of the monitor is verified each time it is
turned on as described in this procedure. The verification
procedure Power-On-Self-Test (POST) takes
approximately 10 seconds to complete. When the
monitor is turned on, POST automatically tests the
monitor circuitry and functions.
Ranges/
Selections
Alarms 30, 60,
90, 120 seconds,
OFF
0 to 1010
Factory
Defaults
30 seconds
Caution: During the Power-On-Self-Test (POST)
immediately after power-up, confirm that all display
segments and icons are shown and the monitor
speaker sounds a one-second tone.
WARNING: If you do not hear the POST pass tone,
do not use the OXIMAX N-65.
36OXIMax N-65 Operator’s Manual
When the monitor is turned on, the backlight remains on
and the display cycles through the following sequence as
POST takes place:
Procedure
OXIMAX N-65 Operation
•All display graphics are shown for three seconds
and the backlight is turned on.
•The display goes blank (all display elements off)
for one second.
•The software version number shows for 3 seconds
as a three digit number in the right number field
and two dashes in the left number field.
•The current time of day is shown in 24 hour
format.
•Successful completion of POST is announced by a
POST pass tone. A failed POST is announced by a
high-priority alarm tone.
To turn on the monitor:
1. Press the Power button to turn on the monitor.
The backlight remains on throughout POST.
OXIMax N-65 Operator’s Manual37
OXIMAX N-65 Operation
All display numbers and icons show for three
seconds. The backlight is on while all numbers and
icons are shown.
2. The display goes blank for one second.
38OXIMax N-65 Operator’s Manual
OXIMAX N-65 Operation
3. The software version number appears for 3 seconds
as a three digit number in the right number field and
two dashes in the left number field.
Note: The software version 001 is an example.
Check the monitor for the software version
installed. Write down the software version
number and have it available prior to requests
for technical assistance.
4. The current time displays in a 24-hour format.
OXIMax N-65 Operator’s Manual39
OXIMAX N-65 Operation
5. If the monitor detects a problem, an error tone sounds
and an error code (Er) and the error number displays.
See Troubleshooting on page 93.
6. Upon successful completion of POST, the monitor
sounds a one-second tone to indicate it has passed the
test.
WARNING: Ensure the speaker is clear of
any obstructions and the speaker holes are
not covered. Failure to do so could result in
an inaudible alarm tone.
Note: The POST pass tone also functions as an
audible confirmation that the speaker performs
properly.
40OXIMax N-65 Operator’s Manual
OXIMAXSensor Attached
When an OXIMAX sensor contains patient event alarm
data, and is attached to the monitor, the Data In-Sensor
indicator is shown.
OXIMAX N-65 Operation
The monitor shows two dashes in the %SpO
2 and Pulse
Rate displays as it searches for a valid pulse. For optimal
performance allow the monitor to search and lock onto a
pulse for approximately 10 seconds.
When a valid pulse is detected, the monitor begins
monitoring and displays patient parameters.
Movement of the blip bar indicates that real-time data is
displayed. Continue to listen for the pulse beep tone as
the monitor is used. If it does not sound with each pulse,
it indicates one of the following:
•pulse beep volume is off
•speaker/audio has malfunctioned
•signal is corrupted
•O
XIMAX N-65 has stopped functioning
OXIMax N-65 Operator’s Manual41
OXIMAX N-65 Operation
When an OXIMAX sensor is attached to the monitor and
applied to a patient, and loses the pulse signal, the
monitor displays [- - / - -] (two dashes and two dashes)
and displays the Pulse Search icon as shown on the
monitor.
re
No OXIMAX Sensor Attached
Upon successful completion of POST, the monitor
sounds a one-second tone indicating that it has passed
POST.
42OXIMax N-65 Operator’s Manual
The monitor displays [- - / - -] (two dashes and two
dashes) and the Pulse Search indicator does not display
when the monitor fails to detect an O
invalid sensor.
Backlight On/Off
The backlight turns off after approximately 10 seconds.
With the monitor turned on, press the Backlight button
to turn the monitor backlight on or off.
Caution: When adjusting any menu
parameter, the SpO
display, but they continue to record.
Adjust Pulse Beep Volume
OXIMAX N-65 Operation
XIMAX sensor or an
2 and BPM do not
Discussion
There are two ways to adjust the pulse beep volume:
•during monitoring
•using the menu structure
OXIMax N-65 Operator’s Manual43
OXIMAX N-65 Operation
Procedure — During Monitoring
To change the pulse beep volume during monitoring:
Press the Up Arrow or Down Arrow button while
monitoring to increase or decrease the pulse beep
volume.
The Pulse Amplitude (blip bar) increments or decrements
as a relative indicator of the current volume. Attempted
adjustments outside the range generate an invalid key
tone.
The minimum pulse rate volume is none, or OFF (no blip
bar segments shows.) The maximum pulse rate volume is
ten (ten segments.)
The Beep Tone Volume Adjust indicator flashes when
the pulse beep volume is adjusted to zero.
Note: When the monitor times-out in 30
seconds, the parameter is set and the monitor
display returns to the monitoring mode.
44OXIMax N-65 Operator’s Manual
Procedure — Using Menu Structure
Note: For quick reference information on using
the menu structure to set parameters, see Set Menu Parameters on page 108.
To adjust the pulse beep volume using the menu
structure:
1. From the main monitoring screen, press the Menu
button five times until the pulse beep volume level is
shown and the monitor sounds.
OXIMAX N-65 Operation
2. Press the Up Arrow button or the Down Arrow
button until the desired tone level is heard.
3. Press the Menu button three times to set the tone
volume and return to normal operation.
OXIMax N-65 Operator’s Manual45
OXIMAX N-65 Operation
Adjust Alarm Volume
Discussion
When the Alarm Volume display shows, adjust the alarm
volume by pressing the Up Arrow or Down Arrow
button. The Pulse Amplitude (blip bar) increments or
decrements as a relative indicator of the current alarm
volume. Attempted adjustments outside the range
generate an invalid key tone.
Note: When the monitor times-out in 30
seconds, the parameter is set and the monitor
display returns to the normal mode.
Procedure
To adjust the alarm volume:
WARNING: Do not silence the audible alarm
function, or decrease the audible alarm volume, if
patient safety could be compromised.
46OXIMax N-65 Operator’s Manual
OXIMAX N-65 Operation
1. From the main monitor screen, press the Menu
button six times until the alarm volume screen
appears.
2. Press the Up Arrow or Down Arrow button to
increase or decrease alarm volume.
3. Press the Menu button two times to set the alarm
level and return to the monitor display screen.
OXIMax N-65 Operator’s Manual47
OXIMAX N-65 Operation
Set Time and Date
Discussion
When you make the month entry, the monitor checks the
day selection to ensure the date is valid. If the day
selection is not valid for the month selected, the menu
display returns to the day selection display.
WARNING: The OXIMAX sensor uses the
date and time provided by the O
XIMAX N-65
when the patient alarm events are recorded
to the memory chip in the sensor. The
accuracy of the date/time used for patient
alarm events depends on the date/time
already set in, and provided by the monitor.
Note: When the monitor times-out in 30
seconds, the parameter is set and the monitor
display returns to the normal mode.
When you make the year entry, the monitor checks the
day and month selections to ensure the selections are
valid. If the day or month selection is not valid for the
year selected, the menu display returns to the day
selection display.
Examples of invalid dates:
•30 February
•31 February
•31 April
•31 June
•31 September
•31 November
48OXIMax N-65 Operator’s Manual
Procedure
To set the time and date:
1. From the main monitor screen, press the Power
2. Press the Power button to turn the monitor on.
3. While the monitor is in the POST mode, press the
OXIMAX N-65 Operation
•29 February on a non-leap year
button to turn the monitor off.
Menu button until the set hours display shows. The
hours display (13) will flash.
4. Press the Up Arrow button or the Down Arrow
button until the desired hours appear.
OXIMax N-65 Operator’s Manual49
OXIMAX N-65 Operation
5. Press the Menu button to set the hours and show the
minutes set display. The minutes display (45) flashes.
6. Press the Up Arrow button or the Down Arrow
button until the desired minutes appear.
7. Press the Menu button to set the minutes and show
the day set display. The day (29) flashes.
50OXIMax N-65 Operator’s Manual
OXIMAX N-65 Operation
8. Press the Up Arrow button or the Down Arrow
button until the desired day is shown.
9. Press the Menu button to set the day and show the
month set display. The month (7) flashes.
10. Press the Up Arrow button or the Down Arrow
button until the desired month is shown.
OXIMax N-65 Operator’s Manual51
OXIMAX N-65 Operation
11. Press the Menu button to set the month and display
the year set display. The year (2004) will flash.
12. Press the Up Arrow button or the Down Arrow
button until the desired year is shown.
13. Press the Menu button to set the year and return to
normal operation.
52OXIMax N-65 Operator’s Manual
Set Alarm Silence Duration
Discussion
When the Alarm Silence Duration display shows, you
can adjust the alarm silence duration of the high,
medium, and low-priority alarms.
A low battery low-priority alarm cannot be silenced.
Note: When the monitor times-out in 30
seconds, the parameter is set and the monitor
returns to the normal mode.
Procedure
To set the alarm silence duration:
1. From the main monitor screen, press the Alarm
Silence button until SEC or OFF appears on the
pulse rate area of the display. Then, release the
button.
OXIMAX N-65 Operation
WARNING: Do not silence the audible
alarm function, or decrease the audible
alarm volume, if patient safety could be
compromised.
OXIMax N-65 Operator’s Manual53
OXIMAX N-65 Operation
Available alarm silence durations are OFF (alarm
disabled), 30, 60, 90, and 120 seconds.
2. Press the Up Arrow button or the Down Arrow
button until the desired alarm silence duration is
shown.
3. Press the Alarm Silence button to set the shown
alarm silence duration and return to normal
operation.
Disable Audible Alarms
Discussion
When the alarm silence duration is set to OFF the
monitor produces no audible alarms. The Alarm Silence
indicator flashes while the alarm silence duration is set to
OFF.
54OXIMax N-65 Operator’s Manual
Procedure
To disable audible alarms:
1. From the main monitor screen, press the Alarm
OXIMAX N-65 Operation
Note: When the monitor times-out in 30
seconds, the parameter is set and the monitor
display returns to the normal mode.
WARNING: Do not silence the audible
alarm function, or decrease the audible
alarm volume, if patient safety could be
compromised.
Silence button until OFF displays in the pulse rate
area of the monitor. Release the button.
Available Alarm silence durations are OFF, 30, 60,
90, and 120 seconds.
OXIMax N-65 Operator’s Manual55
OXIMAX N-65 Operation
2. Press the Alarm Silence button to set the alarm
silence duration to OFF and return to normal
operation.
The Alarm Silence indicator flashes and an audible
reminder sounds every three minutes while the alarm
silence duration is set to OFF.
Set Alarm Limits
Discussion
The Alarm Limit display allows you to adjust the upper
and lower saturation and pulse rate limits.
Press and hold the Up Arrow or Down Arrow buttons to
scroll rapidly through the limit values.
The Alarm Limit Changed indicator displays anytime
an alarm limit changes. See Alarm Limit Changed Indicator on page 61.
Note: When the monitor times-out in 30
seconds, the parameter is set and the monitor
display returns to the normal monitoring.
56OXIMax N-65 Operator’s Manual
Procedure
To set alarm limits:
1. From the main monitor screen, press the Menu
button once. The SpO
OXIMAX N-65 Operation
2 lower alarm limit displays.
The %SpO
The upper value of the %SpO
limited to the %SpO
2 low alarm limit range is 20% to 99%.
2 low alarm limit is
2 upper alarm limit. The %SpO2
low alarm limit cannot be set equal to, or higher than,
the %SpO
2 upper alarm limit.
2. Press and hold the Up Arrow or Down Arrow
buttons to scroll rapidly through the values.
3. Press the Menu button to set the limit value. The
SpO
2 upper alarm limit displays.
OXIMax N-65 Operator’s Manual57
OXIMAX N-65 Operation
The %SpO2 upper alarm limit range is 21% to 100%.
The lower value of the %SpO
limited to the %SpO
upper alarm limit cannot be set equal to or lower than
the %SpO
4. Press the Up Arrow or Down Arrow buttons to
increase or decrease the upper alarm limit.
2 upper alarm limit is
2 low alarm limit. The %SpO2
2 low alarm limit.
5. Press the Menu button to set the limits value. The
BPM lower alarm limit displays.
The pulse rate low alarm limit range is 30 to 249. The
upper value of the pulse rate low alarm limit is
limited one number lower than the pulse rate upper
alarm limit. The pulse rate low alarm limit cannot be
set equal to or higher than the pulse rate upper alarm
limit.
58OXIMax N-65 Operator’s Manual
OXIMAX N-65 Operation
6. Press the Up Arrow or Down Arrow buttons to
increase or decrease the lower limit.
7. Press the Menu button to set the limit value. The
BPM upper alarm limit displays.
The pulse rate upper alarm limit range is 31 to 250.
The lower value of the pulse rate upper alarm limit is
limited to one number above the pulse rate low alarm
limit. The pulse rate upper alarm limit cannot be set
equal to or lower than the pulse rate low alarm limit.
OXIMax N-65 Operator’s Manual59
OXIMAX N-65 Operation
8. Press the Up Arrow or Down Arrow button to
increase or decrease the upper pulse rate limit.
9. Press the Menu button to confirm the alarm limit
settings.
10. Press the Menu button until you return to normal
operation.
Note: Limit changes are in effect only as long
as the monitor remains on. When it is turned
off, the default limits are restored. When the
monitor is turned on, the default limits will be
in effect.
60OXIMax N-65 Operator’s Manual
Alarm Limit Changed Indicator
Alarm limits that have changed from the default values
are identified by a decimal point (.) after the shown value
(%SpO
The Alarm Limit Changed indicator displays anytime
an alarm limit changes.
2 or BPM).
OXIMAX N-65 Operation
Record Snap-Shot Data
Discussion
The monitor contains an internal memory that can store
50 patient data records (snap-shots). You can print data
records. Data are retained in the monitor memory while it
remains on and cleared when it is turned off, or powers
itself off. If the data are cleared, they are not available to
print. Replacement of the monitor batteries clears patient
data.
The Data indicator flashes at a medium priority rate
when the monitor patient memory is full.
OXIMax N-65 Operator’s Manual61
OXIMAX N-65 Operation
Procedure
To record data:
1. From the main monitor screen, press the Data button
to capture a record of the patient data (snap shot).
The captured data displays.
The number (1) in the SpO
2 field is the identification
(ID) number of captured snap-shots.
When you press the Data button, and there is no
empty event memory location available, the monitor:
• displays the last ID number assigned (50)
•the Data icon continues to flash at the medium
priority rate
• an invalid key tone sounds
• the monitor returns to the normal mode after
approximately three seconds
62OXIMax N-65 Operator’s Manual
Print Data
Discussion
OXIMAX N-65 Operation
The monitor can print data when used with a Citizen PD22T portable printer, which is available from Nellcor
Customer Services at 1.800.635.5267, or your local
Nellcor representative.
Caution: Ensure that the printer model
number contains a “T,” which indicates that
the printer has been configured for use with
the O
XIMAX N-65.
Note: Read the entire user’s manual for the
Citizen PD-22T portable printer prior to
operating the printer with the monitor.
The monitor must be linked via infrared (IR) to a
compatible printer to print.
Note: The monitor remains on if there are
patient alarm event data stored.
Print data contains one or more of the following:
•a summary report, if currently connected to a
patient
The summary report includes the date, current
time, duration of the current monitoring session,
and the minimum, maximum, and mean of the
SpO
2 and pulse rate.
•a snap-shot report, if any are stored
OXIMax N-65 Operator’s Manual63
OXIMAX N-65 Operation
•sensor event data (patient alarm event data), if a
The monitor prints all available data as shown in
Figure 2.
sensor with patient event data is connected to the
monitor
Figure 2: Combined Printed Data
64OXIMax N-65 Operator’s Manual
OXIMAX N-65 Operation
Print Mode
If you press the Data button in the print menu when no
data is stored, and the device is not connected to a
patient, an invalid key tone sounds and the monitor
transitions to normal mode.
The Print-mode indicator (page icon) flashes at the
medium rate throughout print mode.
When the print function is complete, the display returns
to the normal mode. If the monitor times-out before the
Data button is pressed, the display returns to the normal
mode.
The Menu button is disabled during print mode.
If there is no communication with the printer for
30 seconds, print mode is aborted and the display returns
to normal mode.
After Printing Data
The summary report data is purged after it is printed or
on time-out due to printer communication failure.
Snap-shot or sensor-event data is not purged after
printing or time-out due to printer communication failure.
Instead, it is purged from the monitor’s memory at power
off. Sensor-event data is retained in the sensor and loads/
reloads to the monitor memory at power on.
Alarm annunciation and patient monitoring is disabled
during active printing when the monitor is
communicating with printer.
OXIMax N-65 Operator’s Manual65
OXIMAX N-65 Operation
If snap-shot data are stored in the monitor memory, the
Data icon shows on the print data display. If
sensor-event data are stored in sensor memory, the Data In-Sensor icon shows on the print data display. If both
types of data are stored, both icons show. If no data are
stored, neither icon shows and the display is blank.
The monitor and sensor does not differentiate between
patients. Use caution when reviewing the report as the
list of snap-shots may contain data from more than one
patient. Also, if a single-use sensor is used more than
once, the sensor trend report section can contain data
from more than one patient.
Note: The snap-shot list is deleted when the
monitor is turned on. When first connecting a
single-use sensor, look for the Data In-Sensor
icon.
66OXIMax N-65 Operator’s Manual
Procedure
To print data:
1. Align the printer and the monitor.
2. Orient the printer and monitor as shown in Figure 3.
OXIMAX N-65 Operation
The alignment of the infrared ports of the printer and
the monitor must not exceed two feet and must not be
closer than six inches (15 cm).
Figure 3: Printer Alignment
OXIMax N-65 Operator’s Manual67
OXIMAX N-65 Operation
3. From the main monitor screen, press the Menu
button seven times. The print display is shown.
4. Press the Data button to start printing. During
printing, the stored snap-shot number displays for
each snap-shot.
68OXIMax N-65 Operator’s Manual
O XIM A X Sensor Event
Record
Event Records
WARNING: The OXIMAX sensor uses the date and
time provided by the O
event record is recorded by the sensor. The accuracy
of the date/time is dependent on the date/time already
set in and provided by the monitor.
XIMAX N-65 when the sensor
The adhesive O
XIMAX sensors are capable of storing
patient alarm events in the memory chip. The
sensor-event data record allows alarm event history to
travel with the patient on the sensor’s memory chip for
quick assessment at every point of care where O
XIMAX
monitors are used.
The patient alarm data are stored (recorded) with the
monitor limit/threshold settings that were active at the
time of the event on the recording monitor. These events
can be viewed on the next O
XIMAX enabled monitor to
which the patient is connected when the patient moves to
a new point of care.
An event occurs when the %SpO
2 value exceeds either
the upper or lower alarm limit for at least 15 seconds.
The monitor records patient alarm events to the memory
chip in the sensor every 5 minutes. The maximum
number of events that can be stored in an O
XIMAX sensor
is typically 100.
OXIMax N-65 Operator’s Manual69
OXIMAX Sensor Event Record
Event records can be viewed only after an OXIMAX sensor
containing patient alarm events has been connected to an
O
XIMAX monitor capable of displaying sensor event
records. The O
sensor event records, but does support printing sensor
event records. Event records are designed to view patient
events from prior areas of care or transport (history),
while monitor trend should be used to view data or events
from a patient currently being monitored.
The Data In-Sensor indicator displays when the sensor
contains patient alarm events. The Data In-Sensor
indicator flashes when the sensory memory is full.
XIMAX N-65 does not support viewing
Recording and viewing of O
only available on O
XIMAX enabled monitors. The OXIMAX
XIMAX sensor-event data is
sensors may function on older technology monitors, but
the O
XIMAX sensor event record feature is not available.
70OXIMax N-65 Operaotr’s Manual
O XIM AX Sensors and
Accessories
Select an OXIMAX Sensor
WARNING: The OXIMAX sensor uses the date and
time provided by the O
event record is recorded by the sensor. The accuracy
of the date/time is dependent on the date/time already
set in and provided by the monitor.
WARNING: Failure to cover the OXIMAX sensor site
with opaque material in high ambient light conditions
may result in inaccurate measurements.
WARNING: Before use, carefully read the applicable
O
XIMAX sensor Directions For Use, including all
warnings, cautions, and instructions.
XIMAX N-65 when the sensor
WARNING: Do not use a damaged OXIMAX sensor or
extension cable. Do not use an O
exposed optical components.
WARNING: Use only Nellcor-approved OXIMAX
sensors and extension cables with the O
Other sensors or extension cables may cause
improper monitor performance.
OXIMax N-65 Operator’s Manual71
XIMAX sensor with
XIMAX N-65.
OXIMAX Sensors and Accessories
WARNING: Do not attach any cable to the OXIMAX
sensor port connector that is intended for computer
use.
WARNING: Tissue damage can be caused by
incorrect application or duration of use of an O
sensor. Inspect the O
directed in the OXIMAX sensor Directions For Use.
WARNING: Pulse oximetry readings and pulse
signals can be affected by ambient environmental
conditions, O
patient conditions.
WARNING: Do not immerse or wet the OXIMAX
sensor as this may damage the sensor.
XIMAX
XIMAX sensor site periodically as
XIMAX sensor application errors, and
WARNING: Do not lift the monitor by the OXIMAX
sensor or extension cable because the cable could
disconnect from the monitor, causing the monitor to
drop on the patient.
72OXIMax N-65 Operator’s Manual
OXIMAX Sensors and Accessories
Caution: The OXIMAX Sensor Disconnected icon and
associated alarm indicate that either the O
XIMAX
sensor is disconnected or the wiring is faulty. Check
XIMAX sensor connection and, if necessary,
the O
replace the O
XIMAX sensor, extension cable, or both.
For additional information on monitor measurements, see
Inaccurate Monitor Measurements on page 28.
For a complete current list of all O
applicable to the O
When selecting an OXIMAX sensor consider the following:
•patient weight and activity
•adequacy of perfusion
•available O
XIMAX sensor sites
•need for sterility
•anticipated duration of monitoring
OXIMax N-65 Operator’s Manual73
OXIMAX Sensors and Accessories
For additional information refer to Table 3 or contact
your local Nellcor Representative. Refer to OXIMAX Sensor Performance Considerations on page 89 for
additional information on O
Table 3: Nellcor Oximetry Sensor Models and Patient
OXIMAX SensorModel
XIMAX sensor performance.
Weights
Patient Size
> = greater than
< = less than
OXIMAX MAX- FAST
adhesive forehead sensor,
single-patient use
Table 3: Nellcor Oximetry Sensor Models and Patient
Weights (Continued)
OXIMAX SensorModel
Patient Size
> = greater than
< = less than
OXIMAX Dura-Y
®
D-YS>1 kg (>2.2 lbs)
multisite sensor, reusable
For use with the Dura-Y
sensor:
Ear clip (Reusable,
D-YSE>30 kg (66 lbs)
nonsterile)
Pedi-Check
TM
pediatric
D-YSPD3 kg to 40 kg
spot-check clip (Reusable,
nonsterile)
OXIMAX Sensor Features
OXIMAX sensor features are different for OXIMAX sensors
XIMAX sensor type, such as adhesive, recycled, and
by O
reusable. The O
O
XIMAX sensor plug.
Tabl e 4 : OXIMAX Sensor Features
Feature
XIMAX sensor type is located on the
Adhesive
Sensors
(6.6 lbs to 88 lbs)
Recycled
Sensors
Reusable
Sensors
OXIMAXSensor Event Record
Ye sN oN o
Sensor MessagesYesYesYes
Sensor ID MessageYesYesYes
76OXIMax N-65 Operator’s Manual
Biocompatibility Test
Biocompatibility testing has been conducted on Nellcor
O
XIMAX sensors in compliance with ISO 10993-1,
Biological Evaluation of Medical Devices, Part 1:
Evaluation and Testing. The O
the recommended biocompatibility testing and are,
therefore, in compliance with ISO 10993-1.
Optional Accessories
The optional accessories offered with the monitor are:
•rubber boot with clip, page 78
•functional thermoformed holster, page 79
•carrying case, page 80
•water-resistant jacket, page 81
•printer, page 82
•thermal paper, page 83
•DEC-4 extension cable, page 84
OXIMAX Sensors and Accessories
XIMAX sensors have passed
•transport boot, page 85
OXIMax N-65 Operator’s Manual77
OXIMAX Sensors and Accessories
Rubber Boot With Clip
This accessory protects the monitor.
1 — Front view2 — Rear view
Figure 4: Rubber Boot with Clip
78OXIMax N-65 Operator’s Manual
OXIMAX Sensors and Accessories
Functional Thermoformed Holster
This accessory protects the monitor.
1 — Front view2 — Rear view
Figure 5: Functional Thermoformed Holster
OXIMax N-65 Operator’s Manual79
OXIMAX Sensors and Accessories
Carrying Case
The Nylon carrying case is equipped with compartments
for:
•the monitor
•Operator’s manual
•compact disk containing all manuals
•sensors
and,
•a carrying strap that is adjustable from 71 cm to
135 cm (28 inches to 53 inches)
Figure 6: Carrying Case
80OXIMax N-65 Operator’s Manual
Water-Resistant Jacket
The water-resistant jacket is made of clear plastic to
facilitate use of the monitor in inclement weather or
similar conditions.
Figure 7: Water-Resistant Jacket
OXIMAX Sensors and Accessories
OXIMax N-65 Operator’s Manual81
OXIMAX Sensors and Accessories
Infrared Printer
The Citizen PD-22T portable printer is used to print
selected data from the monitor. The monitor uses the
printer IrDA infrared interface. Refer to the User’s
Manual supplied with the printer for details on the
PD-22T printer.
Figure 8: Printer
82OXIMax N-65 Operator’s Manual
Thermal Paper
The thermal paper is for the Citizen PD-22T infrared
printer. Refer to the User’s Manual supplied with the
printer for details on the thermal paper.
OXIMAX Sensors and Accessories
Figure 9: Thermal Paper
OXIMax N-65 Operator’s Manual83
OXIMAX Sensors and Accessories
DEC-4 Extension Cable
The DEC-4 extension cable provides 1.2 m (4 ft.) of
cable extension between the monitor and the sensor.
1 — DEC-4 extension cable
2 — Sensor connector
Figure 10: DEC-4 Extension Cable
84OXIMax N-65 Operator’s Manual
Transport Boot
This accessory is used in the transport environment for
compliance with ISO 9919 outside of a hospital, such as
ambulance transport.
OXIMAX Sensors and Accessories
1 — Front view2 — Rear view
Figure 11: Transport Boot
OXIMax N-65 Operator’s Manual85
OXIMAX Sensors and Accessories
86OXIMax N-65 Operator’s Manual
Performance
Considerations
WARNING: Pulse oximetry readings and pulse
signals can be affected by ambient environmental
conditions, O
patient conditions. See the appropriate sections of the
manual for specific safety information, such as Safety
Information on page 1 and OXIMAX Sensors and
Accessories on page 71.
Performance Verification
The performance of the OXIMAX N-65 is verified by
following the procedures outlined in the Performance
Verification section of the monitor service manual.
Qualified service personnel should perform these
procedures before using the monitor for the first time in a
clinical setting.
XIMAX sensor application errors, and
OXIMAXN-65 Performance Considerations
This section describes patient conditions that can affect
the monitor’s measurements.
OXIMax N-65 Operator’s Manual87
Performance Considerations
Dysfunctional Hemoglobins
Dysfunctional hemoglobins, such as,
carboxyhemoglobin, methemoglobin, and
sulfhemoglobin, are unable to carry oxygen. SpO
readings may appear normal; however, a patient may be
hypoxic because less hemoglobin is available to carry
oxygen. Further assessment beyond pulse oximetry is
recommended.
Anemia
Anemia causes decreased arterial oxygen content.
Although SpO
patient may be hypoxic. Correcting anemia can improve
arterial oxygen content. The monitor may fail to provide
an SpO
2 if hemoglobin levels fall below 5 gm/dl.
Saturation
2
2 readings may appear normal, an anemic
The monitor displays saturation levels between 1% and
100%.
Pulse Rates
The monitor displays pulse rates between 20 and 300
beats per minute. The sensor accuracy ranges do not
apply to pulse rates above 250 bpm. Detected pulse rates
below 20 are shown as 0.
88OXIMax N-65 Operator’s Manual
Performance Considerations
OXIMAXSensor Performance Considerations
WARNING: Pulse oximetry readings and pulse signal
can be affected by ambient conditions, O
application errors, and patient conditions.
WARNING: Tissue damage can be caused by
incorrect application or inappropriate duration of use
of an SpO
site as directed in the O
2 OXIMAX sensor. Inspect the OXIMAX sensor
XIMAX sensor directions for
use.
WARNING: Use only Nellcor-approved OXIMAX
sensors and extension cables. Do not use cables more
than 4 feet in length. Use only the DEC-4 extension
cable or only the OXIMAX sensor.
XIMAX sensor
Inaccurate measurements can be caused by:
•incorrect application of the O
•placement of the O
XIMAX sensor on an extremity
XIMAX sensor
with a blood pressure cuff, arterial catheter, or
intravascular line
•excessive patient activity
•intravascular dyes, such as indocyanine green or
methylene blue
•externally applied coloring, such as nail polish or
pigmented cream
•failure to cover the O
XIMAX sensor site with opaque
material in high ambient light conditions
•venous pulsation
•dysfunctional hemoglobin
OXIMax N-65 Operator’s Manual89
Performance Considerations
•low perfusion
Loss-of-pulse signal can occur for the following reasons:
•the O
XIMAX sensor is applied too tightly
•defibrillation
•a blood pressure cuff is inflated on the same
extremity as the one with the O
XIMAX sensor
attached
•there is arterial occlusion proximal to the O
sensor
•poor peripheral perfusion
•loss of pulse/cardiac arrest
To use the O
•Select an appropriate O
XIMAX sensor:
XIMAX sensor.
•Apply the sensor as directed, and observe all
warnings and cautions presented in the Directions
For Use accompanying the O
XIMAX sensor.
•Clean and remove any substances, such as nail
polish, from the application site.
•Periodically check to ensure that the O
XIMAX
sensor remains properly positioned on the patient.
XIMAX
High ambient light sources that can interfere with the
performance of an SpO
2 OXIMAX sensor are:
•surgical lights (especially those with a xenon light
source)
•bilirubin lamps
•fluorescent lights
•infrared heating lamps
•direct sunlight
90OXIMax N-65 Operator’s Manual
Performance Considerations
To prevent interference from ambient light, ensure that
the O
XIMAX sensor is properly applied, and cover the
O
XIMAX sensor site with opaque material.
If interference due to patient activity presents a problem,
try one or more of the following to correct the problem:
•verify that the O
XIMAX sensor is properly and
securely applied
•move the O
•use an adhesive O
•use a new O
XIMAX sensor to another site
XIMAX sensor
XIMAX sensor with fresh adhesive
backing
•keep the patient still, if possible
If interference due to poor perfusion presents a problem,
consider using the MAX-R O
FAST O
XIMAX sensor. The MAX-R OXIMAX sensor
XIMAX sensor or the MAX-
obtains measurements from the nasal septal anterior
ethmoid artery, an artery supplied by the internal carotid.
These O
XIMAX sensors may obtain measurements when
peripheral perfusion is relatively poor.
OXIMax N-65 Operator’s Manual91
Performance Considerations
92OXIMax N-65 Operator’s Manual
Troubleshooting
WARNING: If you are uncertain about the accuracy
of any measurement, check the patient's vital signs by
alternate means; then make sure the O
functioning correctly.
WARNING: There are no user-serviceable parts
inside the O
removed by qualified service personnel.
Caution: Do not spray, pour, or spill any liquid on the
O
openings in the enclosure as this may damage the
monitor.
Error Codes
XIMAX N-65, its accessories, connectors, switches, or
XIMAX N-65. The cover should only be
XIMAX N-65 is
When the monitor detects an error condition, it may
display “Er” followed by the error code.
OXIMax N-65 Operator’s Manual93
Troubleshooting
See Table 5, "Error Codes", on page 94 for a list of error
codes and corrective actions.
When an error occurs, the unit:
•stops monitoring
•removes all information from the display and
displays the message “Er XXX”
•sounds a low priority alarm
When an error occurs, you must:
•Turn the monitor off.
•Wait 10 seconds and turn the monitor on.
•If the error appears, follow the action(s) listed in
Table 5 .
Table 5: Error Codes
Error
Code
10Check/replace sensor/extension cable.
11
1 — Replace batteries.
Action
2 — Notify service personnel.
17Check/replace sensor/extension cable.
19Check/replace sensor/extension cable.
273
1 — Restart the monitor.
2 — Set the time and date.
3 — Notify service personnel.
274Return the monitor for reprogramming.
275Check/replace sensor/extension cable.
276Replace with O
94OXIMax N-65 Operator’s Manual
XIMAX sensor.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.