Casella CEL CEL-272 User Manual

1 CEL Blue
1 Black
Register
Reduce the Artwork to 95%
Marks
CEL-272 PERSONAL SOUND EXPOSURE METER
Instructions
September 1990
PP60111
Issue: 2
for
archive Document
Discontinued Instrument
The CEL-272 Personal Sound Exposure Meter is a totally new instrument designed to fulfill the measurement requirements of the United Kingdom "Noise at Work Regulations" 1989, and the re­cently introduced ISO 1999: 1990 "Determination of Occupational Noise Exposure and Estimation of Noise Induced Hearing Impairment" which stipu­late standard methods for measuring personal sound exposure.
The CEL-272 complies with Draft IEC Document TC 29/WG 4, and substantially meets the require­ments of BS 6402 for Personal Noise Dose Me­ters. Where relevant, the performance of the instrument also satisfies IEC 804 for integrating av­eraging sound level meters.
Recent European noise exposure legislation men­tions three action levels. At levels below the 1st ac-
tion level (1.0 Pa hours) "reasonably practicable" noise reductions should be made. Between the 1st and 2nd action level (3.2 Pa sessments must be made, records kept, details given to employees, and hearing protection issued to those who ask for it. Above the 2nd and also at or above the peak action level (200 Pa - 140 dB for any period), hearing protection measures must be enforced by the employer.
The Personal Sound Exposure Meter is a mini­ature instrument intended to be worn by an individ­ual so that its microphone is subjected to as nearly as possible the same sound exposure as the wearer. It can be switched to give a reading in either Pa exposure or % noise dose (see box).
2
h - equivalent to 85 dB(A) for 8
2
h - 90 dB(A) for 8 hours) noise as-
2
h sound
2 CEL Blue
2 Black
Register Marks
DESCRIPTION
The instrument uses a1/4" electret microphone capsule installed in an energy absorbing housing that can be mounted close to the wearer's ear us­ing the clip supplied. The microphone is connected to the instrument by means of a slender integral ca­ble which is long enough for the instrument to be put in a convenient pocket or clipped to the wearer's waist-band or belt.
Signals from the microphone are amplified and passed through an A-weighting circuit that com­plies with the requirements of IEC 651. The output from the weighting circuit is squared and inte­grated over time to produce sound exposure val­ues in either Pa CEL-272, 100% noise dose corresponds to a sound level of 90 dB lasting for 8 hours - the 2nd action level.
The readings are shown on a custom-designed auto-ranging liquid crystal display (see figure). The
1
/2digit display also carries an indication of which
4 parameter is being measured and whether the bat­tery voltage has fallen below 4 V.
In addition, when a signal greater than the 200 Pa peak action level occurs, colons flash on the dis­play to indicate the overload. Similarly, if too many dose counts are recorded the store will overflow and a reading of 2000 will be flashed on the dis­play as an indication. These are latching functions that can be cleared only by resetting the instru­ment.
A pair of slide switches located beneath a lockable tamper-proof cover change the display between Sound Exposure or Noise Dose units. They can also switch the instrument to calibration mode which allows it to be calibrated by means of either a CEL-282 Acoustic Calibrator (IEC 942 Standard Class 2) or a CEL-284/2 Acoustic Calibrator (IEC Class 1).
A further switch position permits the display to be suppressed when required to provide security for the resulting sound exposure information.
The CEL-272 microphone may be replaced by a CEL-6100 Dummy Microphone that allows direct electrical calibration to be performed according to the requirements of the draft IEC Document. Please contact Lucas CEL for precise details of the procedure.
2
h or % Dose units. In the
CEL-272 Personal Sound Exposure Meter
CEL-6048 Microphone Coupler (for calibration), PP16022 Type AAA Manganese Alkaline
PP38524 1.5mm Allen Key, PP38054 Trimtool, PP60111 Instructions.
The following accessories can be ordered sepa­rately for calibrating the instrument.
CEL-282 Acoustic Calibrator Class 2, CEL-284/2AcousticCalibratorClass1, CEL-6100 Dummy Microphone.
(including cable, CEL Microphone, and Battery Holder),
Battery (4 off),
PREPARATION FOR USE
The instrument uses four type AAA Manganese Alkaline batteries that must be inserted in the Bat­tery Holder (supplied).
Zinc Carbon and rechargeable Nickel Cadmium batteries are NOT suitable.
To load batteries proceed as follows.
1. Use the 1.5mm Allen Key (PP38524)supplied to retract the locking screw which frees the instrument covers.
2. Slide the lower instrument cover out of the groove to expose the battery compartment behind the display.
3. Remove the Battery Holder.
4. Install four batteries in the Battery Holder according to the polarities indicated in the figure (and in the bottom of the Holder).
5. Insert the Battery Holder into the instrument according to the arrow mark with the arrow uppermost as shown.
Correct connection is achieved ONLY when the Holder is inserted with this orientation.
6. Check the batteries by sliding the Display switch to ON and the Battery switch to either
2
h or % Dose.
Pa
7. Replace batteries when Lo Bat is indicated.
SCHEDULE OF PARTS
A complete "CEL-272 Personal Sound Exposure Meter" consists of the following items.
8. Replace the cover and if required, refasten the screw.
The instrument is now ready for service.
3 CEL Blue
3 Black
Register Marks
Parameter Indicators
LO BAT
%DOSE
2
Pa h
Overload Colons
Autoranging Decimal Point
900044
2
Pa h
Batt Off
+
%
+
Adj
Display
C A L
+
+
On
Off
Loading...
+ 5 hidden pages