Casella CEL CEL-224 User Manual

Casella USA 17 Old Nashua Road #15 Amherst, NH 03031-2839
CEL-224
Sound Level Meter
www.casellausa.com info@casellausa.com
Handbook
Page 1 of 4 tel: (800) 366-2966
25 Jan 2006 fax: (603) 672-8053
Casella USA 17 Old Nashua Road #15 Amherst, NH 03031-2839
Introduction
In line with the requirements of the new consolidated revision of the International Standard for sound level meters (IEC 651), the CEL-224 has been designed as a low cost full specification Type 3 instrument. It also meets the British Standard BS 3489 and the American Standard ANSI S1.4(2)A in addition to other national standards. The instrument will be found to be a valuable tool by all those charged with a responsibility to control either industrial or environmental noise. A pressed metal case is used that has been styled to be compact and provide the correct acoustic characteristics whilst also being convenient to use.
Schedule of Parts
The instrument is normally provided as a complete measurement kit under the number CEL-224/K1 and includes the following individual items:
CEL-214 Sound level meter CEL-184 Acoustic calibrator CEL-2843/2 Coupler CEL-2962 Foam windscreen 6F22 9V batteries (3 off) D3380 Kit case RS1 Calibration screwdriver
Preparation of the Instrument
The battery compartment is located in the rear. of the instrument and is accessible when the two retaining screws have been removed. The batteries should be connected to the two flying leads and placed in the battery compartment before replacing the cover. To check the batteries the ON/OFF switch should be set to the BATT position and the meter should then indicate within the scale banded 'battery' if they have sufficient charge to operate the instrument correctly.
The batteries must be removed if they become discharged, or if the instrument is not going to be used for some time. As an alternative to the primary cells provided, nickel cadmium rechargeable batteries type RS591-089 may be used with this instrument. A suitable charger for these NiCad batteries is the type RS691-067 that must be specified separately.
Calibration
In line with current instrument practice all types of Sound Level Meters should be regularly calibrated to ensure the accuracy of the reading obtained. For this purpose, the CEL-184 Acoustic Calibrator is available which should be used in conjunction with the CEL-2843/2 coupler. To check the calibration of the CEL-224 the CEL-2843/2 coupler should be placed over the microphone ensuring that it is correctly seated against the shoulder in the microphone coupler. The calibrator may be then placed over the coupler and microphone assembly, here again, ensuring that both the '0' rings are passed and the coupler and calibrator are correctly seated. Switch on both the calibrator and the sound level meter, the latter having its range selector set to the 70dB(A) range. The calibration level is 90dB(A) and this should cause an indication of 20 on the scale. Should the instrument have drifted, due to environĀ­mental influences, it may be reset by means of the CAL control located through the aperture in the front panel of the instrument. Using a suitable screwdriver the level may be reset to the correct reading. If large changes in the calibration level are observed, the meter should be returned to our service department or your local agent.
www.casellausa.com Page 2 of 4 tel: (800) 366-2966 info@casellausa.com
25 Jan 2006 fax: (603) 672-8053
Loading...
+ 2 hidden pages