CareFusion IVAC P6000, IVAC P7000, IVAC TIVA, IVAC TCI & TIVA Technical & Service Manual

s
IVAC® Syringe Pumps
Models P7000, P6000, TIVA, TCI & TIVA
Technical Service Manual
Contents
Chapter 1 Introduction & Start Up 3
Chapter 2 Configuration & Calibration 12
Chapter 3 Routine Maintenance 26
Chapter 4 Troubleshooting 35
Chapter 5 Circuit Descriptions 39
Chapter 6 Spare Parts Replacement Procedures 43
Appendix A Specifications 72
Appendix B Spare Parts Listing 80
Appendix C Fitting and Replacement Guidelines 87
Appendix D Configuration & Drug Protocol Records 90
Appendix E Disposal 98
Appendix F Service Contacts 101
Appendix G Document History 103
Chapter 1
Introduction and Start Up
Introduction 4
General Precautions 5
Front Panel, Controls and Indicators 6
Main Display 7
Loading a Syringe 8
Starting the Pump 9
Basic Features 10
Backoff Feature 11
Fast Start Feature (Model P7000 only) 11
Introduction and Start Up
Introduction
The IVAC® Syringe Pumps are a range of syringe pumps with features dedicated for use in a variety of areas within a hospital. The range includes these models:
The IVAC®
The IVAC®
applications
The IVAC®
The IVAC®
Two versions of this range of pumps have been manufactured which can be identified by the type of power on/off switch:
Mark 1 includes a mechanical power on/off switch on the end of the pump Mark 2 includes a soft power on/off button A on the front panel
Product Familiarity
Ensure that you are fully familiar with the syringe pump by carefully studying the Directions for Use (DFU) prior to operation and prior to attempting any repairs or servicing. As part of continuous improvement, product enhancements and changes are introduced from time to time.
P6000 Syringe Pump is a syringe pump suitable for critical care and general infusion applications
P7000 Syringe Pump is a variable pressure syringe pump suitable for critical care and general infusion
TIVA Syringe Pump is an anaesthesia syringe pump
TCI & TIVA Syringe Pump is an anaesthesia syringe pump incorporating Diprifusor
Purpose of this Manual
This Technical Service Manual describes how to set up, test and maintain the following IVAC® Syringe Pump models: P6000 P7000 TIVA TCI & TIVA
This manual is intended for use by personnel experienced in medical equipment testing and maintenance procedures .
Conventions Used in this Manual
BOLD
'Single quotes' Used to indicate cross-references made to another section of this manual. For example, see
Italics
Used for pump Display names, access codes, controls and indicators referenced in this manual, for example, SERVICE ACCESS menu, access code 251, OK softkey.
Chapter 2, 'Configuration and Calibration'.
Used for other documents or manuals. For example, refer to the relevant Directions for Use (DFU) for further information.
Wherever this symbol is shown a Hints & Tips note is found. These notes provide useful advice or information that may help to perform the task more effectively.
Wherever this symbol is shown a Toolbox note is found. These notes highlight an aspect of test or maintenance that is important to know about. A typical example is drawing attention to a software upgrade that you should check has been installed.
P7000, P6000, TIVA, TCI & TIVA 4/106 1000SM00012 Issue 8
General Precautions
Prior to using this pump, carefully read the Operating Precautions described in the Directions for Use
(DFU)
.
w
This pump contains static-sensitive components. Observe strict precautions for the protection of static sensitive components when attempting to repair and service the pump.
V
An explosion hazard exists if the pump is used in the presence of flammable anaesthetics. Exercise care to locate the pump away from any such hazardous sources.
B
Dangerous Voltage. An electrical shock hazard exists if the casing of the pump is opened or removed. Refer all servicing to qualified service personnel.
A
This pump is protected against the effects of high energy radio frequency emissions and is designed
M
to be fail safe if extremely high levels of interference are encountered. Should false alarm conditions be encountered, either remove the source of the interference or regulate the infusion by another appropriate means.
If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise suspected to have been damaged, remove it from service for inspection by qualified service personnel.
Introduction and Start Up
d
When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If the integrity of the external protective conductor in the installation or its arrangement is in doubt, the pump should be operated from the battery.
P7000, P6000, TIVA, TCI & TIVA 5/106 1000SM00012 Issue 8
Introduction and Start Up
Front Panel, Controls and Indicators
Front Panel
Front Panel - Model P6000 example
Controls and Indicators
H
I
j
k
N O
C
START
STOP
BATTERY
AC POWER
RATE KEYS/ CHEVRONS
BLANK SOFTKEYS
Press to start the infusion. The green LED will flash during infusion.
Press to stop/hold the infusion. The amber LED will be lit while on hold.
When illuminated, indicates that the pump is running on the internal backup battery. When flashing, indicates that the battery power is low, with less than 30 minutes of use remaining.
When illuminated, indicates that the pump is connected to an AC power supply and the battery is being charged.
Double or single for faster/slower, increase or decrease of values shown on main display.
Use in conjunction with the prompts shown on the display.
G
D
E
F
A
J
PURGE/ BOLUS
OPTION
PRESSURE
PRESSURE
ON/OFF
24H LOG
Press to purge the extension set during setup while the pump is on hold, or to deliver bolus at an increased rate while an infusion is running. See 'Basic Features' for further information.
Press to access optional features. See 'Basic Features' for further information.
(Model P6000, Model TIVA, Model TCI &TIVA). Press to display the pumping pressure and alarm level.
(Model P7000 only). Press to display the pumping pressure and alarm level.
Press once to switch the pump ON. Press and hold down for 3 seconds to switch the pump OFF.
Note: Mark I pumps have a mechanical ON/OFF switch on the side of the pump.
(Mark 1: Model P6000, Model TIVA, Model P7000). Press to display the 24 hour log of volume infused.
P7000, P6000, TIVA, TCI & TIVA 6/106 1000SM00012 Issue 8
Main Display
Introduction and Start Up
Model P6000
Infusion Rate
Infusion Status
Model P7000 (Dedicated mode)
Infusion Status
Infusion Rate
Dose Rate
Volume Infused
Syringe Type/Drug Name
IVAC 50
Volume Infused Option
Drug Name/Protocol
Current Time
Pressure Icon
VTBI Status
VTBI Option
Pumping Pressure
Model TIVA
Infusion Rate
Dose Rate
Model TCI & TIVA
Calculated Concentration
Infusion Status
Infusion Status
Syringe Type or Drug Name/Protocol
Volume Infused
Volume Infused
Volume Infused
7.5ml
22.5mg
Review Option
Current Time
Review Information
Review Option
Effect Site Concentration
Decrement Time
Target Concentration
Trend Graph
Note that the Cumulative Drug Mass Infused is also displayed (directly beneath Volume Infused) on pumps with software version V3R2 and above.
P7000, P6000, TIVA, TCI & TIVA 7/106 1000SM00012 Issue 8
Introduction and Start Up
Loading a Syringe
1. Squeeze the finger grips together on the plunger holder and slide the mechanism to the left.
2. Lift the syringe clamp and rotate to the left.
3. Insert the syringe into the slots on the plunger holder (see Figure 1). For Model Diprifusor tag is positioned towards the display to allow the prefilled syringe tag to be read.
4. Squeeze the finger grips on the plunger holder and slide the mechanism to the right until the syringe barrel flange locates into the V slot (see Figure 2).
Ensure that the syringe is advanced until the syringe barrel flange touches the front of the V slot closest to the syringe clamp. This is important to prevent delay at the start of the infusion.
5. Release the finger grips. Apply gentle pressure on the plunger holder to ensure that the drive is engaged.
6. Rotate the syringe clamp until it locks onto the syringe barrel (see Figure 2).
7. Check that the syringe plunger and syringe barrel flange are correctly located into their slots.
TCI & TIVA only - ensure that the
Syringe inserted into plunger holder slots
Syringe barrel
Syringe clamp
Syringe barrel flange
Syringe plunger
Syringe clamp shown
locked onto syringe barrel
Plunger holder
Finger grips
V slot
Syringe barrel flange
Figure 1.
Figure 2.
P7000, P6000, TIVA, TCI & TIVA 8/106 1000SM00012 Issue 8
Introduction and Start Up
Loading a Syringe (continued)
Guide to handling the pressure disc
1. Insert the pressure disc into pressure disc support slots as shown in Figure 1.
2.
Push pressure disc securely home as shown in Figure 2.
Pressure disc support slots
Figure 1.
When removing the pressure disc pull it with your finger inside the disc recess as shown in Figure 3. above.
Figure 2. Figure 3.
Pressure disc
Disc recess
Starting the Pump
1. Connect pump to AC Mains.
Important: To verify continued operation on battery power, prior to beginning an infusion, remove the pump from the AC mains, confirm the pump continues to function on battery power and then
reconnect the pump to the AC mains.
2. Press the A button to switch pump ON (Mark II), or switch on power switch on side of pump (Mark I).
3.
CLEAR DATA? NO retains previous data. YES clears previous infusion data.
4. Load syringe. See instructions in previous section.
5. Insert pressure disc into pressure transducer (Model P7000 only). See 'Guide to handling the pressure disc' above.
6. Confirm syringe.
7. Change the rate if necessary using the
8. Purge: Press the
9. Connect the pump to test equipment as required (see Chapter 2, 'Configuration & Calibration' and Chapter 3, 'Routine Maintenance').
10. Press the
H
G button followed by the PURGE softkey.
button to start operation.
NO keys.
P7000, P6000, TIVA, TCI & TIVA 9/106 1000SM00012 Issue 8
Basic Features
Introduction and Start Up
Pressure Level
Purge/Bolus
24 Hour Log
Option Button
Press the E button (On Model P7000, press the F button). Line pressure occlusion alarm level and current pressure level are shown on a graph. Use the
Press the and hold down the two
During PURGE/BOLUS, the pressure limit alarms are temporarily increased to maximum levels.
Press the
button. If required, use the M buttons to set bolus dose required then press
G
PURGE/BOLUS softkeys together to deliver.
button. The 24 hour log of volume infused is shown. 24 Hour Log is also available
J
M keys to adjust alarm level.
via the D button.
Press the
DRUGS AND DOSING (or DRUG NAME)
button to display optional features:
D
Allows the pump to be set up for use with a specific drug and/or dosing protocol. Follow Drugs and Dosing instructions as per relevant DFU.
SET VTBI*
Set a specific volume to be infused and set rate at end of VTBI (not available on Model TCI & TIVA or Model TIVA).
SET VTBI OVER TIME*
Specify a VTBI and delivery time (not available on Model TCI & TIVA or Model TIVA).
SET BY DOSERATE*
Set rate in doserate increments (mg/kg/h). The flow rate will be calculated from the doserate.
SET BY ml/h*
Set rate in flow rate increments (ml/h). The doserate will be calculated from the flow rate.
CLEAR VOLUME
INDUCTION*
MULTIDOSE*
RATE LOCK*
24H LOG
EVENT LOG*
Clear the displayed volume infused. Set an induction volume (P7000 only).
Set multidose volumes (P7000 only).
Prevent rate being changed once infusion has started.
Shows volume infused over 24 hours.
Displays the event log. Holds up to 800 individual events.
* these options are not displayed/available when feature is disabled.
Note that option names may vary and the Model
TCI & TIVA and Model TIVA pumps have
additional model-specific features. For additional information, refer to the relevant DFU.
P7000, P6000, TIVA, TCI & TIVA 10/106 1000SM00012 Issue 8
Introduction and Start Up
Backoff Feature
BACKOFF is a configurable feature, enabled/disabled via the CONFIGURATION menu, access code 251. When enabled, BACKOFF is automatically activated every time an occlusion occurs. The pump action reverses and pumps backwards to
release the pressure which has built up in the infusion system. This minimises the post occlusion bolus.
How BACKOFF works
When an occlusion occurs, the pump continues pumping until the pressure alarm level is reached and the pump stops. Note that an occlusion can be due to a clamp or stopcock being left closed, kinked line, infiltration etc. During this time, a bolus volume of fluid builds up and is stored between the occlusion and the syringe, because it is unable to be infused into the patient. BACKOFF prevents the accumulated volume of fluid from being released into the patient as a post occlusion bolus.
Following an occlusion alarm:
Pumping mechanism reverses, it pumps backwards
Accumulated volume of fluid (post occlusion volume) is ‘taken back’ into the syringe and not infused into the
patient
Post occlusion volume is deducted from the VOLUME (Volume Infused) value shown on the Display
Post occlusion volume is deducted from the infused value in the 24H LOG as this volume has not been infused into
the patient
BACKOFF time and date is recorded in the EVENT LOG
Clinical benefits of BACKOFF
Reduces the pressure in the extension set due to the plunger of the syringe moving backwards
Removes the post occlusion volume of fluid from the system, it is not delivered to the patient as a bolus on
removal of the occlusion. This prevents the patient from receiving an unintentional and unnecessary bolus
Fast Start Feature (Model P7000 only)
FAST START is a configurable feature of the Model P7000 syringe pump, enabled/disabled via the CONFIGURATION
menu, access code 251. When enabled, FAST START automatically reduces the 'mechanical slack' between the plunger mechanism and syringe at the start of an infusion. The mechanical slack is a very small gap (barely visible) due to the type of syringe and mechanics of the pump, between the back of the syringe plunger and the plunger drive mechanism at the start of an infusion.
How FAST START works
At the start of an infusion, the pump starts at a faster rate for a short period of time. This action minimises the start up delay but may not eliminate it completely.
Notes:
1) The pump can be heard to pump faster than the set infusion rate and can also be felt if a hand is rested on the pump
at the start of an infusion.
2) The faster rate is 50% of the maximum syringe rate, syringe rate ranges are listed in Appendix A, 'Specifications'.
Purge recommendation: Even with FAST START available, it is still recommended to perform a PURGE with the pump before
connecting the extension set to the patient and starting the infusion, especially at low flow rates i.e. below 5ml/h.
Clinical benefits of FAST START
The fluid is delivered to the patient much sooner than if fast start was not available
The delay in delivering the drugs following the changes of syringes is reduced greatly
P7000, P6000, TIVA, TCI & TIVA 11/106 1000SM00012 Issue 8
Chapter 2
Configuration and Calibration
Access Codes 13
Entering an Access Code 13
Dedication Options (901/902) 14
Configuration Options (251) 14
Drug Setup (251) 15
General Options (251) 17
Calibration Procedures 20
Clearing Internal RAM (611) 25
Configuration and Calibration
Access Codes
The syringe pump software contains a number of configuration and test routines that can be accessed using a technical access code. The majority of routines are 'menu' driven and are accessed by entering a technical access code as shown in the table below.
Code Title Description
123 SELF TEST Self test routine begins from the start. See Chapter 3, 'Routine Maintenance' for
further information 124 SELF TEST Self test routine begins at power supply measurement 125 SELF TEST Self test routine begins at declutch test 126 SELF TEST Self test routine begins at pressure disc test 127 SELF TEST Self test routine begins at beam test 128 SELF TEST Self test routine begins at nurse call test 167 COMMS LEARN MODE Configuration set via comms interface 168 COMMS TEACH MODE Configuration output to another device (pump) 176 LINEAR DIAGNOSTIC Linear potentiometer test 222 SELF TEST Self test routine begins at display test 223 SELF TEST Self test routine begins at keypad button test 243 SYRINGE SIZE CALIBRATION Syringe size measurement calibration. See 'Calibration Procedures' on the following
pages for instructions 251 CONFIGURATION MENU Configuration of drugs, options, syringes and real-time clock. See 'Configuration
Options (251)' on the following pages for further details 253 LINEAR POT CALIBRATION Linear potentiometer calibration. See 'Calibration Procedures' on the following
pages for instructions 263 PRESSURE CALIBRATION Line pressure calibration. See 'Calibration Procedures' on the following pages for
instructions 359 LANGUAGE Configure language 376 SERVICE ACCESS MENU Review service log, errors and hours of use. See 'Service Access Options' in Chapter
3 for further details 611 TOTAL MEMORY CLEAR Reset memory. Note that the pump will require full calibration 717 BEAM OCCLUSION CALIBRATION Beam occlusion calibration. See 'Calibration Procedures' on the following pages for
instructions 901 DEDICATION MODE Set pump to operate in fully dedicated mode. Model P7000 only, see 'Dedicated
Options (901/902)' below for further details 902 DEDICATION MODE Set pump to semi-dedicated mode. Model P7000 only, see 'Dedicated Options
(901/902)' below for further details
Entering an Access Code
Each menu (and certain individual options) has its own three-digit technical access code which is entered using the following procedure:
1. Hold down the
2. Enter the required access code “xxx” using the the digits).
3. When the required code is shown on the display, press the
P7000, P6000, TIVA, TCI & TIVA 13/106 1000SM00012 Issue 8
H button and turn the pump ON. The main display shows 000.
M keys in conjunction with the NEXT softkey (to move through
OK softkey to confirm.
Configuration and Calibration
Dedication Options (901/902)
The Model P7000 pump can be operated in one of two modes:
Fully dedicated mode
Operating the pump in this mode will remind a user that a pressure disc must be fitted at the start of any infusion. In this mode occlusion pressures are always displayed in
Semi-dedicated mode
mmHg. To set, enter access code 901.
Operating the pump in this mode will require a user that a pressure disc must be fitted only when drugs and dosing features are used. When a pressure disc is not in use, pressure levels L-0 to L-10 will be displayed. To set, enter access code
902.
Configuration Options (251)
Enter the access code 251 (see 'Entering an Access Code' on the previous page for instructions). The CONFIGURATION menu is displayed:
CONFIGURATION MENU
Option* Description
DRUG SETUP Set drug names and protocols. See 'Drug Setup' on the following pages for instructions. GENERAL OPTIONS See 'General Options' on the following pages for further details. CLOCK SET Set the current date and time.
To set the clock, use M and NEXT softkey to adjust then OK to store.
HOSPITAL NAME Enables hospital/ward name to be displayed during the power-up sequence.
To set the hospital name, use M to toggle through characters, NEXT softkey to go to next character then OK to store.
ENABLE SYRINGES Configure the type and size of syringes permitted for use.
To enable syringes:
1. Use M to select, MODIFY to enable/disable the selected syringe type then OK to store.
2. Turn the pump OFF and return to service or select the next CONFIGURATION option.
ENABLE UNITS Configure the dose units permitted for use (Model P7000 only).
To enable units:
1. Use M to select, MODIFY to enable/disable the selected dose then OK to store.
2. Turn the pump OFF and return to service or select the next CONFIGURATION option.
* Note: For software versions earlier than version V3R2, the options may vary, or will not be available. Refer to the
relevant IVAC® Syringe Pump DFU for comprehensive information.
P7000, P6000, TIVA, TCI & TIVA 14/106 1000SM00012 Issue 8
Drug Setup (251)
Configuration and Calibration
Drug Setup (Model TCI & TIVA/Model
TIVA)
1. Enter the access code 251 to display the CONFIGURATION menu.
2. Select
3. Select the drug name using the
DRUG SETUP using the M keys and press the OK softkey. The drug name list is displayed.
M keys and press the OK softkey. Alternatively, press QUIT to exit and return to
the CONFIGURATION menu.
4. To use a drug, it must be enabled by pressing the
QUIT.
5. Step through each
Press the OK softkey to confirm each option.
Press the BACK key at any time to go back to the previous option and make changes
DRUG SETUP option To adjust, use these keys...
DRUG NAME
CONCENTRATION UNITS
DEFAULT CONCENTRATION
MINIMUM CONCENTRATION
MAXIMUM CONCENTRATION
DOSE UNITS
INDUCTION DOSE
INDUCTION TIME
MAINTENANCE RATE
BOLUS DOSE
DRUG SETUP option (see table below) to setup or modify the drug name and protocol.
To toggle through characters use single chevrons To go to first/last character use double chevrons
NEXT to go to next drug name letter.
M M M or OFF M or OFF
Important: If the DEFAULT CONCENTRATION, the MINIMUM CONCENTRATION and the MAXIMUM CONCENTRATION are equal, the TIVA mode start-up sequence will bypass the CONCENTRATION option
M then OK to select and continue. M
Note: If the setting is less than 0.01 the TIVA mode start-up sequence will bypass the INDUCTION option
M M M
Note: If the setting is less than 0.01 the BOLUS setting is turned OFF. The Bolus feature is disabled in TIVA mode.
YES softkey at ENABLE DRUG. To return to the drug name list press
and
and .
.
6. At the CONFIRM option, review the drug setup data displayed then press the OK softkey to confirm. The DRUG SETUP menu is redisplayed.
P7000, P6000, TIVA, TCI & TIVA 15/106 1000SM00012 Issue 8
Configuration and Calibration
Drug Setup (251) (continued)
Drug Setup (Model P6000/Model P7000)
1. Enter the access code 251 to display the CONFIGURATION menu.
2. Select
3. Select the drug name using the
DRUG SETUP using the M keys and press the OK softkey. The drug name list is displayed.
M keys and press the OK softkey. Alternatively, press QUIT to exit and return to
the CONFIGURATION menu.
4. To use a drug, it must be enabled by pressing the
QUIT.
YES softkey at ENABLE DRUG. To return to the drug name list press
5. Step through each DRUG SETUP option (see table below) to setup or modify the drug name and protocol.
Press the OK softkey to confirm each option.
Press the BACK key at any time to go back to the previous option and make changes
DRUG SETUP option To adjust, use these keys...
DRUG NAME
DOSE UNITS
DEFAULT DOSERATE
MINIMUM DOSERATE
MAXIMUM DOSERATE
CONCENTRATION UNITS
DEFAULT CONCENTRATION
MINIMUM CONCENTRATION
MAXIMUM CONCENTRATION
BOLUS RATE
MAXIMUM BOLUS
PRESSURE ALARM
To toggle through characters use single chevrons To go to first/last character use double chevrons
NEXT to go to next drug name letter.
M M M or OFF M or OFF M M M M M M M
and
and .
.
6. At the
CONFIRM option, review the drug setup data displayed then press the OK softkey to confirm. The DRUG SETUP
menu is redisplayed.
P7000, P6000, TIVA, TCI & TIVA 16/106 1000SM00012 Issue 8
Configuration and Calibration
General Options (251)
1. Enter the access code 251 to display the CONFIGURATION MENU.
2. Select
Option Description
BACKOFF Enable/Disable Backoff feature (feature where pump action reverses to release infusion
AC FAIL ALARM Enable/disable AC Fail alarm (activated when AC power disconnected and pump
NEOI WARNING Set the Near End Of Infusion (NEOI) warning time between 1 min. and 15 mins. (NEOI
EOI POINT Enable/disable Keep Vein Open at End of Infusion (KVO at EOI). Enabled: Pump will
BEAM ALARM Set the default beam occlusion alarm level between level 0 (L0) and level 7 (L7).
WEIGHT Set a default patient weight between 0.1kg and 150kg. Factory default is 70kg.
PURGE RATE Set the purge rate volume between 100ml/h and 500ml/h. (Rate used during PURGE
HANDS FREE BOLUS Enable/disable Bolus function in a 'hands-free' way—no need to hold down buttons
DEFAULT BOLUS Set the default bolus value between 0.1ml and 25ml.
NURSE CALL Enable/disable the Nurse Call feature (hardware feature which allows the pump to
NURSE CALL INVERTED Enable/disable inversion of Nursecall hardware output. Enabled: Nurse Call hardware
COMMS PUMP ADDRESS Set the communications address of the pump.
COMMS MONITOR ONLY Enable/disable pump remote control via the comms link. Enabled: Control of pump
COMMS ODD PARITY Enable/disable odd communication parity bit generation.
COMMS ASCII Enable/Disable ASCII communications mode.
TCI Enable/disable TCI option. Enabled: TCI option is activated (TCI hardware must be
DISPLAY TCI DECREMENT TIME
Key:
= available option
= unavailable option
GENERAL OPTIONS using the M keys and press the OK softkey. The GENERAL OPTIONS MENU is displayed:
GENERAL OPTIONS MENU (Model
Software version: V3R2
system pressure build-up and minimise post occlusion bolus following a pressure alarm). Enabled: Backoff feature ON (This does not apply when the pump is in TCI mode). Disabled: Backoff feature OFF.
operating on battery power). Enabled: AC Fail alarm ON. Disabled: AC Fail alarm OFF.
warning signals the syringe is almost empty allowing time to change syringe).
switch to run at the KVO rate 1.0ml/h (or the current set rate if this is lower) at the EOI point. Disabled: Pump will stop at the EOI point.
operation).
during bolus infusion. (Bolus is a function which administers a controlled volume of fluid or drug at an increased rate). Enabled: Hands Free Bolus function allowed. Disabled: Hands Free Bolus function not allowed.
communicate with the hospital's nurse call system, typically linked to central nurse's station). Enabled: Nurse Call Hardware enabled. Disabled: Nurse Call hardware disabled.
output is inverted. Disabled: Nurse Call hardware output normal.
via monitor only. Disabled: Remote control of pump via comms link allowed. Note that monitor control is always possible.
fitted). Disabled: TCI option not activated. If enabled, the display shows the TCI Decrement Time Icon (TCI hardware must be
fitted). * For pumps with software versions earlier than version V3R2, the options may vary, or will not be available. Refer to the relevant IVAC® Syringe Pump DFU for comprehensive information.
Note: For default settings, refer to Appendix D, 'Configured Options and Drug Protocol Records'.
TIVA/Mode
TCI & TIVA)
l
TIVA* TCI
&TIVA*
   
 
 
 
P7000, P6000, TIVA, TCI & TIVA 17/106 1000SM00012 Issue 8
Configuration and Calibration
General Options (251) (continued)
GENERAL OPTIONS MENU (Model P6000/Model P7000)
Software version: V3R2
Option Description
AUTOSAVE Enable/disable Autosave feature (option to retain previous settings when pump is
switched on). Enabled: Confirmation is requested at power up to clear or to use previous patient data. Disabled: Patient information is cleared on power up.
BACKOFF Enable/disable Backoff feature (feature where pump action reverses to release infusion
system pressure build-up and minimise post occlusion bolus following a pressure alarm). Enabled: Backoff feature ON. Disabled: Backoff feature OFF.
FAST START AND BACKOFF
SET VTBI OVER TIME Enable/disable Volume to be Infused (VTBI) over time option (allows the setting of a
VTBI Enable/disable the Volume to be Infused (VTBI) option (allows the setting of a fixed
INDUCTION Enable/disable the INDUCTION option (option to set an induction volume to be
MULTIDOSE Enable/disable the MULTIDOSE option (feature allowing the delivery of a prescribed dose
RATE LOCK Enable/disable RATE LOCK option (anti-tamper feature preventing rate changes, bolus/
QUIET MODE Enable/disable Quiet Mode feature (keypress tones, low priority alarms and power down
PRESSURE ALARM Set pressure alarm value between 1mmHg and 750mmHg. Default is 300mmHg.
MAX. PRESSURE ALARM Set maximum pressure limit value between 1mmHg and 750mmHg. Default is
AUTO PRESSURE ALARM Enable/disable Auto Set Pressure (AUTO) option (option to adjust the pressure alarm
AC FAIL ALARM Enable/disable AC Fail alarm (activated when AC power disconnected and pump
RATE TITRATION Enable/disable ability to adjust (titrate) infusion while infusion is running. Enabled: rate
NEOI WARNING Set the Near End Of Infusion (NEOI) warning time between 1 min. and 15 mins. (NEOI
EOI POINT Set the End Of Infusion (EOI) volume as a percentage of the syringe volume between
KVI at EOI Enable/disable Keep Vein Open at End of Infusion (KVO at EOI). Enabled: Pump will
BEAM ALARM Set the default beam occlusion alarm level between level 0 (L0) and level 7 (L7).
WEIGHT Set a default patient weight between 0.1kg and 150kg. Factory default is 70kg.
PURGE RATE Set the purge rate volume between 100ml/h and 500ml/h. (Rate used during PURGE
BOLUS Enable/disable Bolus function. (Function which administers a controlled volume of fluid
Enable/disable Fast Start and Backoff features (Faststart feature - drive slack fast start at start of an infusion, Backoff feature - see above). Enabled: Fast Start and Backoff ON. Disabled: Fast Start and Backoff OFF.
fixed volume to be delivered over a fixed period of time). Enabled: VTBI OVER TIME option available via the D button. Disabled: VTBI OVER TIME option not available.
volume to be delivered). Enabled: VTBI option available via the D button. Disabled: VTBI option not available.
delivered). Enabled: INDUCTION option available via the D button. Disabled: INDUCTION option not available.
to be repeated over a specific period). Enabled: MULTIDOSE option available via the D button. Disabled: MULTIDOSE option not available.
purge operations and pump powerdown). Enabled: RATE LOCK option available via the
button. Disabled: RATE LOCK option not available.
D
sequence tones are muted). Enabled: Quiet mode feature ON. Disabled: Quiet Mode is OFF.
750mmHg.
level by a set pressure value above the measured in-line pressure, 15 mins. after start of infusion). Enabled: AUTO option available via the F button. Disabled: AUTO option not
available.
operating on battery power). Enabled: AC Fail alarm ON. Disabled: AC Fail alarm OFF.
can be changed while the pump is infusing, without putting pump on hold. Disabled: START button must be pressed to confirm new rate when rate changes are made while infusing using the rate adjust keys.
warning signals the syringe is almost empty allowing time to change syringe).
0.5% and 2.0%.
switch to run at the KVO rate 1.0ml/h (or the current set rate if this is lower) at the EOI point. Disabled: Pump will stop at the EOI point.
operation).
or drug at an increased rate). Enabled: Bolus function allowed. Disabled: Bolus function not allowed.
P6000* P7000*
 
 
P7000, P6000, TIVA, TCI & TIVA 18/106 1000SM00012 Issue 8
Configuration and Calibration
General Options (251) (continued)
GENERAL OPTIONS MENU (Model P6000/Model P7000)
Software version: V3R2
Option Description
DEFAULT BOLUS Set the default bolus rate value between 10ml/h and 1200 ml/h.
CAP BOLUS RATE Set a maximum permissible bolus rate value between 10 ml/h and 1200ml/h.
CAP RATE Sets maximum permissible infusion rate value between 1ml/h and 1200ml/h.
DISPLAY EVENT LOG Enable/disable EVENT LOG option on OPTIONS menu. (Event Log shows record of timed
and dated pump events such as power on/off, alarms, rate changes). Enabled: EVENT LOG option available via the OPTIONS menu (D button). EVENT LOG option not available via the OPTIONS menu (D button). Note that the EVENT LOG
option is always available via the SERVICE ACCESS menu.
LOG LAST PATIENT ONLY Enable/disable display of all patient data in Event Log. Enabled: Event Log displays last
patient data only (since new patient selected). Disabled: Event Log shows all patient data.
NURSE CALL Enable/disable the Nurse Call feature (hardware feature which allows the pump to
communicate with the hospital's nurse call system, typically linked to central nurse's station). Enabled: Nurse Call Hardware enabled. Disabled: Nurse Call hardware disabled.
NURSE CALL INVERTED Enable/disable inversion of Nursecall hardware output. Enabled: Nurse Call hardware
output is inverted. Disabled: Nurse Call hardware output normal.
COMMS PUMP ADDRESS Set pump address for use in communications. Range is 1-254.
COMMS MONITOR ONLY Enable/disable pump remote control via the comms link. Enabled: Control of pump
via monitor only. Disabled: Remote control of pump via comms link allowed. Note that monitor control is always possible.
COMMS ODD PARITY Enable/disable odd communication parity bit generation.
COMMS ASCII Enable/disable ASCII communications mode.
Key:
= available option
= unavailable option
* For pumps with software versions earlier than version V3R2, the options may vary, or will not be available. Refer to the relevant IVAC® Syringe Pump DFU for comprehensive information.
Note: For default settings, refer to Appendix D, 'Configured Options and Drug Protocol Records'.
P6000* P7000*
     
 
 
P7000, P6000, TIVA, TCI & TIVA 19/106 1000SM00012 Issue 8
Configuration and Calibration
Calibration Procedures
The calibration procedure for each of the four sensors within the pumps are described in this section. Calibration is only necessary if in testing, the sensor performs outside of specification or if a full memory clear has been carried out (in which case all calibration stages must be completed). Refer to Chapter 3, 'Routine Maintenance' for test procedures.
Syringe Size Calibration (243)
Calibration tools required: 1000TG00010 (50ml Spacer A) and 1000TG00011 (100ml Spacer B)
Analysis of process variation shows that a 2 point calibration system is not always sufficient to define the syringe size detection system across the 5ml to 100ml range. The effect is only observed when using 100ml syringes, which may not be recognised. If this occurs, the pump should be recalibrated using a 3 point calibration.
Calibration procedure:
Enter the access code 243.
Fit calibration tool into position on pump as shown below in Steps 1-3 and close the clamp. At each step, CAL is
displayed if value is within tolerances.
CAL softkey to store calibration point.
Press
Notes:
1) If CAL is not displayed, check for correct positioning of calibration tool. If calibration cannot be performed, repairs to pump may be necessary.
2) The calibration values shown on the pump displays are for illustrative purposes only.
Power down to complete the calibration sequence (Step 4).
Step 1
Step 3
Step 2
Step 4
Confirmatory Check - To confirm that the syringe sizing calibration has been performed correctly, select a syringe (preferably 50ml), load and confirm the correct syringe type. Check that the correct syringe size is detected and displayed.
P7000, P6000, TIVA, TCI & TIVA 20/106 1000SM00012 Issue 8
Configuration and Calibration
Calibration Procedures (continued)
Linear Pot Calibration (253)
Calibration tools required: 0000TG00059 (105mm SPACER)
Calibration procedure:
Enter the access code 253.
Fit calibration tool in position on pump as shown in Steps 1-2 below. At each step CAL is displayed if value is within
tolerance.
CAL softkey to store calibration point.
Press
Notes:
1) If
CAL does not appear in the display, check for correct positioning of calibration tool. If calibration cannot be
performed, repairs to pump may be necessary.
2) The calibration values shown on the pump displays are for illustrative purposes only.
Power down to complete the calibration sequence.
Step 1
Step 2
P7000, P6000, TIVA, TCI & TIVA 21/106 1000SM00012 Issue 8
Configuration and Calibration
Calibration Procedures (continued)
Beam Occlusion Calibration (717)
To convert Kilograms of Force (kgf) to Newtons (N) multiply by 9.806650. For example 10 kgf =
98.07N.
Excessive force will damage the plunger mechanism. Do not apply more than 10 kgf ±0.05
kgf to the plunger mechanism at any time.
Calibration tools required: 0000TG00020 (or 0000TG00200) and 0000JG00014
Calibration procedure:
Enter the access code 717.
Fit calibration tools and position plunger as shown in Steps 1-3 below, zero the gauge dial. At each step press CAL
softkey when required calibration force is reached.
Notes:
1) If
CAL does not appear in display, check for correct positioning of tool. If calibration cannot be performed, repairs
to the pump may be necessary.
2) The calibration values shown on the pump displays are for illustrative purposes only.
Allow 30 seconds for pressure to stabilise before pressing CAL softkey.
For best results, before pressing CAL softkey, strive for mid-range force value.
Power down to complete the calibration sequence.
Step 1 0kgf
Step 3
Step 2 4kgf
P7000, P6000, TIVA, TCI & TIVA 22/106 1000SM00012 Issue 8
Configuration and Calibration
Calibration Procedures (continued)
Line Pressure Calibration (263) P7000 only
Calibration tools required:
Pressure gauge (Range 0-1400 mmHg) (Tolerance +/- 2mmHg)
P7000 Dedicated pressure disc infusion set (for example, G30402)
50ml Luer-lock syringe
Calibration procedure:
1. Enter access code
2. Without a pressure disc infusion set fitted press
3. Load pressure disc infusion set into transducer (as shown in Step 2) then connect infusion set to syringe and gauge.
4. Using syringe, apply pressure required (as shown in Step 3 overleaf). Press pressure is displayed on pressure gauge (Step 3).
Step 1 0mmHg (Without set fitted)
263.
CAL softkey (Step 1).
CAL softkey when required calibration
Step 2 Load set
To pressure gauge
50ml Luer-lock syringe
Transducer
P7000, P6000, TIVA, TCI & TIVA 23/106 1000SM00012 Issue 8
Configuration and Calibration
Calibration Procedures (continued)
Line Pressure Calibration (263) P7000 only continued
Calibration procedure (continued):
5. Release pressure but do not remove extension set as shown in Step 4.
6. Power down to complete the calibration sequence.
Step 3 400mmHg ± 2mmHg
Step 4 0mmHg (Set fitted, pressure released)
P7000, P6000, TIVA, TCI & TIVA 24/106 1000SM00012 Issue 8
Configuration and Calibration
Clearing Internal RAM (611)
Warning: Do not clear the RAM unless absolutely necessary, because all calibration and configuration in the pump will be cleared.
If the internal RAM or its associated battery is replaced on the Control PCB, or if the pump fails with an 'ER5 RAM' error it will be necessary to do the following:
Clear the internal RAM:
1. Power up the pump holding down the
2. Enter the access code
3. Press
ENTER and wait for the RAM to be cleared.
4. When the message
Fully calibrate pump. Perform each of the four calibration procedures, as described in this chapter.
Reconfigure the pump:
Set Configuration, drugs and protocol (it may be possible to use the Teach/Learn facility, see Chapter 3, 'Routine
611.
RAM CLEARED appears, switch the pump off.
H button.
Maintenance' for instructions).
Enter access code 251 and set time.
Enter access code 359 and set language.
Enter access code 376 and set service date.
Enter access code 901 to set to dedicated (Model P7000 only).
Carry out Performance Verification Procedure (PVP). See Chapter 3, 'Routine Maintenance' for instructions.
P7000, P6000, TIVA, TCI & TIVA 25/106 1000SM00012 Issue 8
Routine Maintenance
Introduction 27
Chapter 3
Self-Test Procedure (123) 27
Upgrading Software 28
Event Log Download 29
Teach Learn 29
Linear Speed Test 30
Drive Occlusion Test 30
Line Pressure Test (Model P7000 only) 31
Potential Equalisation Terminal Resistance Test (PE Test) 31
Battery Maintenance 31
Linear Diagnostics Test 32
Physical Inspection and Clean 33
Service Access Options (376) 33
Performance Verification Procedure 34
Routine Maintenance
Introduction
For routine maintenance, the following tests and performance verification procedures should be performed in addition to the tasks described in the section 'Physical Inspection and Clean'.
Refer to the relevant
DFU for the recommended routine maintenance period.
Self-Test Procedure (123)
The self-test procedure is designed to allow confirmation of many of the pump functions, defaults and calibrations without requiring internal inspection.
1. Enter the access code
2. The pump now proceeds through a series of tests. Press the
Refer to the table below for details of each test.
Important: If the pump fails the test sequence at any stage, it should be taken out of service and inspected by a qualified
service engineer.
Test Display Description/Action
Review software setup SOFTWARE REVISION
Review calibration data setup
Internal PSU and mains voltage
Audible Alarms AUDIO: ALARM Check loud alarm sound beeps.
Visual: Display DISPLAY TEST Check that the display displays an even graduated grey tone. Visual: Backlight BACKLIGHT FULL / DIM / OFF The display starts the test on FULL backlight, goes DIM and then turns
Visual: LED Indicators LEDS: FLASHING Check that the STOP, START and BATTERY LED’S are flashing. Note: the
Touch panel buttons KEY TEST: 00 Press buttons in turn from START (01) to PURGE (13).
Declutch switch DECLUTCH: 1 Squeeze plunger holder finger grips and check that the display
Plunger detector SYRINGE PLUNGER: 0 Press plunger plate button and check display switches from 0 (No
Motor encoder MOTOR / ENCODER : 1 Motor is pulsed while encoders are tested. Motor moves forwards and
Linear potentiometer LINEAR POT : _ _ . _ _V Declutch the transmission and slide plunger holder to the far left, check
Syringe size detection SYRINGE POT : _ _ . _ _ V Lift the syringe clamp and check that the values displayed increase
123. See 'Entering an Access Code' in Chapter 2 for instructions.
NEXT button to move to the next test.
Displays software version, program CRC, serial number and language. PROGRAM CRC SERIAL NUMBER LANGUAGE TCI LANGUAGE (TCI & TIVA only)
SYRINGE REVIEW Displays syringe, beam and linear calibration figures. Pressure
calibration shown on P7000 only. PSU VOLTS: _ . _ _ V
REG VOLTS: _ . _ _ V
AUDIO: WATCHDOG Check loud alarm sounds continuously. AUDIO: KEYPAD Check soft alarm sounds continuously.
PSU VOLTS: On battery supply typically: 5.50V - 6.50V
PSU VOLTS: On mains supply: 6.50V - 7.50V.S
OFF. At the end of the test it turns back on to FULL.
AC LED does not flash, it remains constant.
H 01 C 09 I 02 E or F 10
03, 04
N
O
05, 06
A or J 11
D 12 C 07 G 13 C 08
alternates between 1 (engaged) and 0 (disengaged - finger grips squeezed together).
syringe fitted) to 1 (Syringe plunger fitted).
backwards as encoders pass.
the value displayed (approx. 00.19V). Declutch and slide the plunger holder to the far right and check the value displayed (approx. 03.00V).
within the normal range (approx. 0.05V and 3.00V).
P7000, P6000, TIVA, TCI & TIVA 27/106 1000SM00012 Issue 8
Routine Maintenance
Self-Test Procedure (123) (continued)
Test Display Description/Action
Pumping pressure detection BEAM FORCE : _ . _ _ V Remove the syringe and confirm that the value displayed is within
normal range (± 0.05V). Gently press back on the plunger holder and watch the value increase.
Pressure disc detection (P7000 only)
Pressure sensor test (P7000 only)
Nursecall NURSECALL: ON Check for audible clicks of the relay. Comms COMMS _ Self-test, transmit/receive link back check.
PRESSURE DISC: 0 With an extension set fitted the value should be 0. Without an
extension set the value should be 1.
PRESSURE: _._ _V Check the values displayed are in approximate range of +0.30V to
+0.60V.
DONE - SWITCH OFF The final screen displays DONE - SWITCH OFF. Providing the pump
passed all the tests it can be powered OFF and put back into service. Note that if the pump fails the test sequence at any stage, it should be taken out of service and inspected by a qualified service engineer.
Upgrading Software
Upgrade of the Models P7000, TCI & TIVA and TIVA Syringe Pumps software to V3R2 or greater is mandatory at the next service.
Perform upgrades by acquiring the software upgrade kits specified in spare parts listings.
Equipment required: Software upgrade kit (includes EPROM fitting and removal instructions)
Software Upgrade Kits Available
Syringe Pump Model Part Number
Model P7000 1000SP00530
Model TIVA 1000SP00531
Model TCI & TIVA 1000SP00532
Note: The latest version of software for the Model P6000 syringe pump is 1000SP00529.
P7000, P6000, TIVA, TCI & TIVA 28/106 1000SM00012 Issue 8
Routine Maintenance
Event Log Download
A PC application known as the Event Log Download Utility (ELDU) (part number 1000SP00209) is available to download logs from IVAC® Syringe Pumps.
ELDU Operation
1. Click on
2. Click
3. Select
4. Select
5. Select
6. Select
- Required PC com port selected.
- Character mode and parity match IVAC
- Delay mode is Normal.
- Pump address matches
7. Click
8. Connect the RS232 cable.
9. Power up pump.
10. Click
11. Press
12. Select
ELDU icon on PC.
Accept to agree with Restrictions of Use and continue.
Configure from drop-down menu. Setup Pump and choose IVAC® as pump type. Settings to select log to be downloaded. Communications then check options are set as follows:
® pump configuration.
COMMS PUMP ADDRESS in GENERAL OPTIONS.
OK to confirm.
Download Log from main PC screen.
Close, when finished.
File from drop-down menu and save file. Log may be printed here as required.
Teach Learn
A pump can be configured/reconfigured by transferring data from one pump (Teacher pump) to another pump (Learner pump) via the serial port.
Equipment required
Two pumps, of the same model (for example, two Model
version of software
RS232 cable. For Models
P6000/P7000/TIVA use part number 1000SP01008. For Model TCI & TIVA use part number
6000SP00012
Procedure
1. Connect the Teacher pump to the Learner pump using the RS232 cable.
2. Switch both pumps ON. 3 On the Teacher pump, enter the access code The Teacher pump displays
4. When complete, select
Possible reasons for Teach/Learn failure:
Pumps are different models  Software versions are different  Loose cable connection or faulty RS232 cable
TEACHING and the Learner pump displays LEARNING.
OK.
168 and on the Learner pump, enter the access code 167.
Using Teach Learn to reconfigure a pump only transfers certain configuration settings. The following settings are not transferred and must be reset manually: CLOCK SET (251), SET LANGUAGE (359), SERVICE DATE (376), DEDICATED MODE (901/902), ENABLE UNITS (251) on P7000 only.
P7000 pumps). Note that both pumps must have the same
Once Teach Learn is complete, all configuration settings on the Learner pump, including DRUG SET UP, must be checked against the Teacher pump (original source).
P7000, P6000, TIVA, TCI & TIVA 29/106 1000SM00012 Issue 8
Routine Maintenance
Linear Speed Test
The linear accuracy of the pump can be verified by measuring the time the plunger holder takes to travel a specified distance. The distance travelled is measured using a dial indicator, mounted in place of the syringe, and the elapsed time can be measured using a stop watch.
Equipment required: 1000TG00080 and 0000JG00014
1. Declutch the drive mechanism and move the plunger holder to the right.
2. Fit the linear test gear 1000TG00080 to the pump and move plunger holder towards the dial gauge until it is a about 3mm clear of the probe. Fit a plunger detect spatula 0000JG00014.
3. Set the pump to run at a rate of 100ml/h, confirm syringe type and start it. Allow the pump to run until the plunger detect spatula touches the probe and then using a stop watch, time the travel over a distance of 15.00mm.
4. Using the values specified in table below, check that the pump is travelling at the correct speed. If the test values fall outside the stated limits (or the movements of the dial are jerky) then the pump requires further investigation.
The table below provides data for a combination of syringe types and sizes.
Syringe Type Size (ml) Flow Rate (ml/h) Expected Time (min:sec)
BD Plastipak 50 100.0 4m 55s +/-3s
IVAC 50 100.0 4m 57s +/-3s
Terumo 50 100.0 5m 57s +/-4s
B.Braun Omnifix 50 100.0 5m 29s +/-4s
Monoject 50 100.0 4m 55s +/-3s
Nipro 50 100.0 5m 55s +/-4s
Fresenius Injectomat 50 100.0 5m 49s +/-4s
Braun Perfusor 50 100.0 5m 30s +/-3s
Once 50 100.0 5m 20s +/-4s
Zeneca 50 100.0 5m 02s +/-3s
BD Precise 50 100.0 5m 50s +/-4s
BD Perfusion 50 100.0 5m 28s +/-4s
Notes:
1) It is only necessary to perform the test using one syringe type, 'BD Plastipak' is recommended.
2) Not all syringe types listed are specified for use with every model of pump.
Drive Occlusion Test
Test gear required: 0000TG00020 (or 0000TG00200) and 0000JG00014
Method 1. Semi-dedicated pumps
1. Fit the test gear. (For Model transducer). Confirm syringe type 'BD Plastipak 50ml'. Set the beam occlusion level to L-3. Set up a continuous rate of 100ml/h.
2. Run pump and check that the force at alarm is in the range of 2.2kgf to 3kgf. If it is out of range re-calibrate (see 'Beam Occlusion Calibration' in Chapter 2) and retest.
Method 2. Dedicated pumps (
1. Fit a pressure disc infusion set into the transducer. Set the line occlusion level to 220mmHg. Fit the test gear. Confirm syringe type 'BD Plastipak 50ml' (configure if necessary). Set up a continuous rate of 100ml/h.
2. Run pump and check that the force at alarm is in the range 3.1kgf to 3.9kgf. If it is out of range re-calibrate (see 'Beam Occlusion Calibration' in Chapter 2) and retest.
P7000, P6000, TIVA, TCI & TIVA 30/106 1000SM00012 Issue 8
P7000, set pump to semi-dedicated mode and remove pressure disc infusion set from
Model
P7000 only)