CareFusion Alaris 8000, Alaris 8015, Alaris 8100 Service manual

4 (3)
CareFusion Alaris 8000, Alaris 8015, Alaris 8100 Service manual

Technical Service Manual

Alaris® PC Unit, Models 8000 and 8015

Alaris® Pump Module, Model 8100

Supports: Guardrails® Suite (v7)

Guardrails® Suite MX (v8)

December 2010

ALARM

I

NFU

S

E

S

TANDBY

SILENCE

 

SILENCE

 

SYSTEM

 

OPTIONS

ON

OPTIONS

1

2

3

 

1

2

3

4

5

6

 

4

5

6

7

8

9

ENTER

7

8

9

CLEAR

0

.

CANCEL

CLEAR

0

 

RATE (mL/h)

CHANNEL

SELECT

PAUSE

SYSTEM

ON CHANNEL

OFF

RESTART

ENTER

CANCEL

Model 8000

Model 8015

Model 8100

Alaris® Products

General Contact Information

CareFusion

San Diego, California carefusion.com/alaris

Customer Advocacy - North America

Clinical and technical feedback.

Phone: 888.876.4287

E-Mail: CustomerFeedback@carefusion.com

Technical Support - North America

Maintenance and service information support; troubleshooting.

United States:

Canada:

Phone:

Phone:

888.876.4287

800.387.8309

Customer Care - North America

Instrument return, service assistance, and order placement.

United States:

Canada:

Phone: 888.876.4287

Phone: 800.387.8309

 

 

Alaris® and Guardrails® are registered trademarks of CareFusion Corporation or one of its subsidiaries. Microstream® is a trademark of Oridion Medical 1987 Ltd. All other trademarks belong to their respective owners.

Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

 

TABLE OF CONTENTS

 

 

Chapter 1 - General Information

 

1.1

Introduction.......................................................................................................................................................................................

1-1

1.2

Alarms and Messages...............................................................................................................................................................

1-2

1.3

Battery Management System ...............................................................................................................................................

1-2

1.3.1

Battery and Charging Process.............................................................................................................................................

1-2

1.3.2

System On/Off................................................................................................................................................................................

1-3

1.3.3

Real Time Clock (RTC) with NVRam Control Lithium Battery-Backed RTC ..........................................

1-3

1.3.4

Battery Maintenance...................................................................................................................................................................

1-4

1.4

Battery Capacity Information.................................................................................................................................................

1-5

1.5

Pole Clamp Feature ....................................................................................................................................................................

1-6

1.6

Memory...............................................................................................................................................................................................

1-6

Chapter 2 - Checkout and Configuration

 

2.1

Introduction.......................................................................................................................................................................................

2-1

2.2

New Instrument Checkout.......................................................................................................................................................

2-1

2.3

Configuration Options and Setup - General ................................................................................................................

2-1

2.3.1

Configuration Notes ....................................................................................................................................................................

2-1

2.3.2

Configuration Setup Notes .....................................................................................................................................................

2-2

2.4

Configuration Setup - PC Unit..............................................................................................................................................

2-2

2.4.1

Access System Configuration Options ...........................................................................................................................

2-2

2.4.2

Alarm audio ......................................................................................................................................................................................

2-3

2.4.3

Anesthesia Mode..........................................................................................................................................................................

2-3

2.4.4

Battery meter...................................................................................................................................................................................

2-3

2.4.5

Clock setup.......................................................................................................................................................................................

2-4

2.4.6

Key click audio................................................................................................................................................................................

2-5

2.4.7

Max Pt. weight................................................................................................................................................................................

2-5

2.4.8

Patient ID Entry..............................................................................................................................................................................

2-6

2.4.9

PM reminder.....................................................................................................................................................................................

2-6

2.4.10

Profiles.................................................................................................................................................................................................

2-6

2.4.11

Tamper resist...................................................................................................................................................................................

2-7

2.5

Configuration Setup - Pump Module................................................................................................................................

2-7

2.5.1

Access System Configuration Options ...........................................................................................................................

2-7

2.5.2

Accumulated air.............................................................................................................................................................................

2-8

2.5.3

Air-in-line settings.........................................................................................................................................................................

2-8

2.5.4

Auto-restart attempts .................................................................................................................................................................

2-8

2.5.5

KVO rate adjust..............................................................................................................................................................................

2-9

2.5.6

Max rate..............................................................................................................................................................................................

2-10

Alaris® PC Unit

i

Alaris® Pump Module

 

Technical Service Manual

 

TABLE OF CONTENTS

Chapter 2 - Checkout and Configuration (Continued)

2.5.7

Max VTBI...........................................................................................................................................................................................

2-10

2.5.8

Pressure mode...............................................................................................................................................................................

2-11

2.5.9

SEC-PRI alert .................................................................................................................................................................................

2-12

2.5.10

Secondary.........................................................................................................................................................................................

2-13

2.6

Configuration Setup - Shared Infusion............................................................................................................................

2-13

2.6.1

Access System Configuration Options ...........................................................................................................................

2-13

2.6.2

Delay Options..................................................................................................................................................................................

2-13

2.6.3

Drug Calculation............................................................................................................................................................................

2-14

2.6.4

Multidose............................................................................................................................................................................................

2-15

2.6.5

Pressure dynamic.........................................................................................................................................................................

2-16

2.6.6

Volume/Duration............................................................................................................................................................................

2-16

Chapter 3 - Preventive Maintenance

3.1

Introduction.......................................................................................................................................................................................

3-1

3.2

Cleaning..............................................................................................................................................................................................

3-1

Chapter 4 - Principles of Operation

4.1

Introduction.......................................................................................................................................................................................

4-1

4.2

PC Unit ................................................................................................................................................................................................

4-1

4.2.1

Logic Board Assembly - Model 8000...............................................................................................................................

4-1

4.2.2

Logic Board Assembly - Model 8015 ...............................................................................................................................

4-3

4.2.3

Power Supply Board Assembly ...........................................................................................................................................

4-4

4.2.4

Power Requirements..................................................................................................................................................................

4-5

4.2.5

Mechanical........................................................................................................................................................................................

4-6

4.3

Pump Module ..................................................................................................................................................................................

4-7

4.3.1

Display Board Assembly ..........................................................................................................................................................

4-8

4.3.2

Logic Board Assembly...............................................................................................................................................................

4-9

4.3.3

Motor Controller Board Assembly......................................................................................................................................

4-12

4.3.4

Mechanical........................................................................................................................................................................................

4-12

4.4

Power Control Circuit .................................................................................................................................................................

4-13

4.5

Inter-Unit Communications Circuit and Connections.............................................................................................

4-13

4.5.1

Software .............................................................................................................................................................................................

4-14

4.5.2

Unit Detection and Identification Circuitry ....................................................................................................................

4-14

4.5.3

IUI ...........................................................................................................................................................................................................

4-15

4.5.4

Module Detection and Logical Designation .................................................................................................................

4-16

4.5.5

Unit ID Assignment at System On .....................................................................................................................................

4-16

4.5.6

Module Attachment After System On ..............................................................................................................................

4-18

ii

Alaris® PC Unit

 

Alaris® Pump Module

 

Technical Service Manual

 

 

TABLE OF CONTENTS

 

 

Chapter 4 - Principles of Operation (Continued)

 

4.5.7

Module Detachment....................................................................................................................................................................

4-18

4.5.8

Communications Time-Out ....................................................................................................................................................

4-18

Chapter 5 - Corrective Maintenance

 

5.1

Introduction.......................................................................................................................................................................................

5-1

5.2

Disassembly / Reassembly ....................................................................................................................................................

5-1

5.3

PC Unit ................................................................................................................................................................................................

5-3

5.3.1

Battery Pack Assembly.............................................................................................................................................................

5-3

5.3.2

Latch Assembly..............................................................................................................................................................................

5-4

5.3.3

Power Cord.......................................................................................................................................................................................

5-5

5.3.4

Pole Clamp Assembly................................................................................................................................................................

5-6

5.3.5

Rear Panel (Model 8000) and Rear Panel Assembly............................................................................................

5-7

5.3.6

Rear Panel Assembly Parts...................................................................................................................................................

5-8

5.3.7

Model 8000: CI Board Accessory (Model 8012).......................................................................................................

5-10

5.3.8

Model 8015: Wireless Network Card Assembly........................................................................................................

5-11

5.3.9

Model 8000: Nurse Call Accessory (Model 8010)...................................................................................................

5-13

5.3.10

Model 8000: Nurse Call / CI Board Accessory..........................................................................................................

5-16

5.3.11

Handle..................................................................................................................................................................................................

5-17

5.3.12

Front and Rear Case Separation........................................................................................................................................

5-18

5.3.13

Model 8000: Display and Retainer ....................................................................................................................................

5-19

5.3.14

Model 8015: Display, Retainers, and Inverter Board..............................................................................................

5-20

5.3.15

IUI Connectors ...............................................................................................................................................................................

5-22

5.3.16

Rear Case Assembly Parts....................................................................................................................................................

5-23

5.3.17

Model 8015: Memory Card.....................................................................................................................................................

5-25

5.3.18

Chassis Assembly Parts ..........................................................................................................................................................

5-26

5.4

Pump Module ..................................................................................................................................................................................

5-28

5.4.1

Latch Assembly and Feet........................................................................................................................................................

5-28

5.4.2

IUI Connectors and Rear Case............................................................................................................................................

5-29

5.4.3

Door/Display Board Assembly .............................................................................................................................................

5-30

5.4.4

Display Board Assembly ..........................................................................................................................................................

5-31

5.4.5

Door Latch Assembly.................................................................................................................................................................

5-32

5.4.6

Platen Assembly............................................................................................................................................................................

5-33

5.4.7

Logic and Motor Controller Board Assemblies ..........................................................................................................

5-34

5.4.8

Motor, Air-in-Line (AIL) Sensor Assembly, and Bezel Assembly....................................................................

5-36

5.4.9

Membrane Frame Assembly and Pressure Sensors.............................................................................................

5-38

Alaris® PC Unit

iii

Alaris® Pump Module

 

Technical Service Manual

 

TABLE OF CONTENTS

Chapter 6 - Troubleshooting

6.1

Introduction.......................................................................................................................................................................................

6-1

6.2

Errors....................................................................................................................................................................................................

6-1

6.2.1

Error Numbering Scheme .......................................................................................................................................................

6-2

6.3

Maintenance Mode......................................................................................................................................................................

6-2

6.3.1

External Communications.......................................................................................................................................................

6-3

6.3.2

Calibration Parameters .............................................................................................................................................................

6-3

6.3.3

Clock Setup ......................................................................................................................................................................................

6-4

6.3.4

Battery Conditioning Test ........................................................................................................................................................

6-5

6.3.5

Keypad Test......................................................................................................................................................................................

6-6

6.3.6

Display Test......................................................................................................................................................................................

6-6

6.3.7

Display Error Log ..........................................................................................................................................................................

6-7

6.3.8

Display Contrast ............................................................................................................................................................................

6-8

Chapter 7 - Illustrated Part Breakdown

7.1

Introduction.......................................................................................................................................................................................

7-1

7.2

Illustrations........................................................................................................................................................................................

7-1

7.3

Parts List ............................................................................................................................................................................................

7-1

7.4

Ordering Parts ................................................................................................................................................................................

7-2

Figures

4-1

Block Diagram - Models 8000 and 8015 .......................................................................................................................

4-20

4-2

Block Diagram - Model 8100.................................................................................................................................................

4-21

7-1

Models 8000 and 8015 - Chassis Assembly...............................................................................................................

7-9

7-2

Model 8015 - Memory Card ...................................................................................................................................................

7-10

7-3

Models 8000 and 8015 - Rear Case Assembly.........................................................................................................

7-11

7-4

Models 8000 and 8015 - IUI Connectors and Backup Speaker .....................................................................

7-13

7-5

Model 8000 - LED Display Assembly ..............................................................................................................................

7-14

7-6

Model 8015 - LED Display Assembly...............................................................................................................................

7-15

7-7

Models 8000 and 8015 - Front to Rear Case Assembly......................................................................................

7-17

7-8

Models 8000 and 8015 - Handle Assembly.................................................................................................................

7-18

7-9

Models 8000 and 8015 - Rear Panel Assembly........................................................................................................

7-19

7-10

Model 8000 - Rear Plate Assembly ..................................................................................................................................

7-21

7-11

Model 8015 - Wireless Network Assembly...................................................................................................................

7-23

7-12

Model 8000 - Model 8010 Nurse Call Accessory ....................................................................................................

7-24

7-13

Model 8000 - Model 8012 Communications Interface Accessory ................................................................

7-25

7-14

Model 8000 - Communications Interface Board / Nurse Call Accessory.................................................

7-26

7-15

Models 8000 and 8015 - Pole Clamp Assembly.......................................................................................................

7-27

iv

Alaris® PC Unit

 

Alaris® Pump Module

 

Technical Service Manual

 

 

TABLE OF CONTENTS

 

 

Figures (Continued)

 

7-16

Models 8000 and 8015 - Power Cord Assembly and Labels ...........................................................................

7-28

7-17

Models 8000 and 8015 - Latch Assembly.....................................................................................................................

7-29

7-18

Models 8000 and 8015 - Battery Pack Assembly....................................................................................................

7-30

7-19

Model 8015 - Communications Board.............................................................................................................................

7-31

7-20

Model 8015 - Inverter Board..................................................................................................................................................

7-32

7-21

Models 8000 and 8015 - Isolated RS-232 Board.....................................................................................................

7-33

7-22

Models 8000 and 8015 - IUI Boards.................................................................................................................................

7-34

7-23

Model 8000 - Logic Board.......................................................................................................................................................

7-35

7-24

Model 8015 - Logic Board .......................................................................................................................................................

7-36

7-25

Models 8000 and 8015 - Power Supply Board ..........................................................................................................

7-37

7-26

Model 8100 - Membrane to Bezel Assembly ..............................................................................................................

7-41

7-27

Model 8100 - Motor and AIL Sensor Assembly.........................................................................................................

7-42

7-28

Model 8100 - Logic Board and Motor Controller Board Assembly................................................................

7-43

7-29

Model 8100 - Platen Assembly ............................................................................................................................................

7-44

7-30

Model 8100 - Door and Display Board Assembly ....................................................................................................

7-45

7-31

Model 8100 - Rear Case and IUI Connectors ............................................................................................................

7-47

7-32

Model 8100 - Latch and Feet Assembly.........................................................................................................................

7-48

7-33

Model 8100 - Label Locations ..............................................................................................................................................

7-49

7-34

Model 8100 - AIL Board............................................................................................................................................................

7-50

7-35

Model 8100 - Display Board...................................................................................................................................................

7-51

7-36

Model 8100 - Logic Board .......................................................................................................................................................

7-52

7-37

Model 8100 - Motor controller Board................................................................................................................................

7-53

Tables

 

 

1-1

Defined Terms ................................................................................................................................................................................

1-2

1-2

Battery Trip Points........................................................................................................................................................................

1-4

1-3

Abbreviations, Acronyms, Symbols ..................................................................................................................................

1-7

4-1

Power Sources...............................................................................................................................................................................

4-5

4-2

Power Supply Board Voltages..............................................................................................................................................

4-6

4-3

IUI Logic..............................................................................................................................................................................................

4-17

4-4

IUI Signals.........................................................................................................................................................................................

4-17

5-1

Required Materials, Supplies and Tools.........................................................................................................................

5-2

5-2

Torque Values - Models 8000 and 8015 ........................................................................................................................

5-39

5-3

Torque Values - Model 8100..................................................................................................................................................

5-40

5-4

Level of Testing Guidelines - Models 8000 and 8015 ...........................................................................................

5-41

5-5

Level of Testing Guidelines - Model 8100 .....................................................................................................................

5-42

6-1

Technical Troubleshooting Guide.......................................................................................................................................

6-9

Alaris® PC Unit

v

Alaris® Pump Module

 

Technical Service Manual

 

TABLE OF CONTENTS

List of Tables (Continued)

6-2

Subsystem Codes ........................................................................................................................................................................

6-18

6-3

Failure Codes..................................................................................................................................................................................

6-19

6-4

Error Codes......................................................................................................................................................................................

6-22

7-1

Parts List - Models 8000 and 8015 ...................................................................................................................................

7-3

7-2

Parts List - Model 8100.............................................................................................................................................................

7-38

vi

Alaris® PC Unit

 

Alaris® Pump Module

 

Technical Service Manual

Chapter 1 – GENERAL INFORMATION

CAUTION

1.1

INTRODUCTION

To avoid damaging the keypad, do not use sharp objects (such as, pens, pencils) to activate switches.

CAUTION

Any attempt to service this product by anyone other than an authorized CareFusion Service Representative, while the product is under warranty, may invalidate the warranty.

This manual describes how to service the PC Unit (Models 8000 and 8015) and Pump Module (Model 8100). It is used in

conjunction with the following Alaris® System documents and software:

Alaris® System Directions for Use (DFU)

Maintenance software and user manual

This manual is intended for personnel experienced in analysis, troubleshooting and repair of analog/digital microprocessorbased electronic equipment.

Reference the Alaris® System DFU for a product introduction, detailed setup and operation procedures, definitions (including precaution definitions), specifications, and other information related to the use of the Alaris® System.

If the PC Unit or Pump Module requires service while under warranty, it must be serviced only by CareFusion authorized service personnel. Reference the "Warranty" and "Service Information" sections in the Alaris® System DFU.

Important: The SYSTEM ON key becomes the "Power Off" key anytime the Alaris® System is in a watchdog state. The watchdog state is identified by:

constant loud audio tone that cannot be silenced

nonresponsive keypad

flashing red arrow above SYSTEM ON key

Alaris® PC Unit

1-1

Alaris® Pump Module

Technical Service Manual

GENERAL INFORMATION

Table 1-1. Defined Terms

The following table identifies the defined terms used throughout this document for certain trademarked products and product features.

Product / Feature

 

 

Defined Term

Alaris® Auto-ID module

Auto-ID Module

Alaris® EtCO

2

module

EtCO

2

Module

 

 

 

 

 

Alaris® PCA module

PCA Module

Alaris® PC point-of-care unit

PC Unit

Alaris® PC unit

PC Unit

Alaris® Pump module

Pump Module

Alaris® SpO

module

SpO

2

Module

2

 

 

 

 

 

Alaris® Syringe module

Syringe Module

Guardrails® data set

Data Set

1.2ALARMS AND MESSAGES

Alarm messages are displayed on the scrolling Message Display bar. Reference the Alaris® System DFU for detailed information.

An audio alarm and the Alarm Status Indicator flash red when an alarm limit is met or exceeded. All alarms can be temporarily silenced by pressing the SILENCE key on the PC Unit.

1.3BATTERY MANAGEMENT SYSTEM

This section contains general information on the battery management system. Included is information on how the power supply processor monitors and maintains the battery, controls the system on/off for the rest of the instrument and provides support functions for the main processor.

The battery management system consists of the power supply processor IC and various sensors and signal processing circuits.

The power supply processor performs the following functions:

Controls battery charger.

Provides a battery status "battery gauge".

Monitors battery voltage and temperature.

Controls instrument power source (on/off function).

The power supply processor communicates with the main processor via a serial data channel. The main processor issues commands to the battery manager which then responds with status information and data using this channel.

1.3.1Battery and Charging Process

The battery pack is a 10-cell (1.2V per cell), high capacity nickel metal hydride (NiMH) type, rated at 12 volts and 4 amp-hours (with a minimum of 500 charge cycles). It has a built-in temperature sensor which allows the battery manager to monitor battery temperature.

1-2 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

GENERAL INFORMATION

1.3BATTERY MANAGEMENT SYSTEM

(Continued)

1.3.1Battery and Charging Process

(Continued)

Self-resetting thermal fuse at 70° C.

Self-resetting current limit sense at 5A.

The battery charge circuit charges the battery with a constant current of 2A whenever the power supply processor turns the charger on. The power supply processor regulates average charge current by turning the charger on and off with the appropriate duty ratio. The battery charge cycle consists of 4 modes: fast charge, float charge, terminates charge, and top-up charge.

Fast Charge: Fast charge is initiated whenever the battery is between 10° C and 27° C, and has been discharged by more than 200 amp-seconds through actual use or self discharge. Leaving the instrument unplugged for a day would cause about 200 amp-seconds of self discharge. The charge current is a continuous 2A. The end of a fast charge is detected when the temperature of the battery is at least 30° C and increases at a rate of 0.7° C/min. (or 1.4A depending on system load at 0.6° C/min.) above its temperature at start of charge, or when the battery voltage declines by 50 mV below its peak value, or total charge time exceeds 2.6 hours.

Top-Up Charge: The top-up charge phase begins at the end of the fast charge phase and finishes adding the last few percent of charge to the battery and balances individual cell charges. This phase charges at an average rate of 0.4A.

Float Charge: The float charge phase begins at the end of the top-up phase and helps maintain a fully charged battery. This phase charges at an average rate of 225 mA and 1 second per minute.

Terminates Charge: The power supply processor does not allow charging to begin unless the temperature is 10 - 40° C. The power supply processor terminates charge if battery temperature drops below 10° C or rises above 55° C.

1.3.2System On/Off

The power supply processor provides the interface between system on/off switch and the main processor. When the instrument is off, the power supply processor interprets either power switch as a turn on command and applies power to the rest of the instrument, informing the main processor the switch was pressed. Once power is on,

further presses of a power switch are passed on to the main processor which determines the appropriate response under the existing conditions. If the response is to turn the power off, the main processor requests that the power supply processor remove power from the rest of the instrument.

If an error has been detected which causes the watchdog to be in alarm, a push of either power switch will immediately cause the power to be turned off, without intervention by the power supply processor.

1.3.3Real Time Clock (RTC) with NVRam Control Lithium Battery-Backed RTC

RTC switches SRAM voltage to prevent memory loss during a power loss.

Alaris® PC Unit

1-3

Alaris® Pump Module

Technical Service Manual

GENERAL INFORMATION

1.3BATTERY MANAGEMENT SYSTEM

(Continued)

1.3.4Battery Maintenance

CAUTION

Use only batteries approved by CareFusion, due to battery manager requirements and the thermostat contained in the battery assembly. If the instrument has been in storage, connect it to AC power before turning it on. One refresh cycle is usually sufficient to restore battery capacity. If necessary, repeat the procedure at 24-hour intervals, 2 or 3 times, to increase capacity.

Several features are included in the battery manager to help properly maintain the battery.

A battery capacity measurement is available in diagnostic mode.

A special circuit removes all load from battery when voltage falls too low, preventing damage from over-discharge due to long-term storage.

NiMH batteries can be stored no more than 6 months (3 months in instrument) with no load but will self-discharge from a charged state in about 100 days. This does not damage the battery as it would if it were

a lead-acid type battery. Connect the instrument to AC to recharge batteries.

Disposal

There are no federal, state or local laws managing the disposal of NiMH batteries. NiMH batteries are recyclable. To learn about recycling batteries at INMETCO, visit their website at http://www.inmetco.com.

Table 1-2. Battery Trip Points

Battery Voltage

 

Instrument Response

 

 

 

• 30 minutes left on gauge

Couloumb Control (I.t)

• instrument continues to function

warning tone activated

 

 

low battery warning

 

 

 

• instrument does not pump

11.20V

constant alarm

 

• low battery alarm (depletion)

 

 

 

• 1 minute or longer (nominal 5 minutes) after low battery alarm

11.0V

backup speaker activated

 

• instrument shutdown (5 minutes after alarm)

 

 

10.8V

disconnect battery

 

 

 

1-4 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

GENERAL INFORMATION

1.4BATTERY CAPACITY INFORMATION

All batteries have specific conditions under which they are guaranteed to meet their published specifications. Deviations from these conditions typically result

in a reduction of available capacity. Manufacturers of NiMH batteries rate capacities, usually expressed in Ah (Ampere-hours), based on a specified "ideal" charge and discharge condition as well as the use of a new battery.

An ideal charge cycle starts with a fully discharged battery charged at C/10 (C is rated capacity in Ah) constant current for 15 hours while at room temperature. For example, a 1.8 Ah battery would be charged for 15 hours at 180 mA constant current with a room temperature of 23° C.

The ideal discharge starts with a fully charged battery under a C/5 constant current load at room temperature, discharging to

a cell voltage of 0.9V. The rated capacity is then calculated as the time to discharge divided by 5. A 1.8 Ah cell would be discharged at 360 mA constant current and not reach 0.9V for at least 5 hours. A given battery type has different capacities based on the load. For example, a battery rated at 1.8 Ah at a 360 mA load may have only 1.6 Ah at a 1600 mA load.

There are many conditions which can affect the battery capacity. The following

conditions have the most practical impact on battery capacity delivered in this instrument.

Battery Alarm Voltage: The battery alarm voltage is the voltage at which the instrument stops operating and generates an alarm indicating the instrument needs to be connected to AC power. Under perfect conditions, a battery of 10 cells connected in series reaches the end

of discharge at 9.0V; however, cells are not perfectly matched so some will reach 0.9V before others. The problem occurs when a cell in series with other cells can go below 0.9V and actually go into cell reversal, which permanently damages the particular cell. On the

other hand, increasing the alarm voltage to compensate for imperfectly matched cells results in reduced run times with available capacity. The user sees this as premature low battery warnings and alarms. The instrument deals with this by increasing the alarm voltage to guarantee the battery is not damaged and reduce the assumed capacity to below that printed on the battery. The battery gauge is intended to show the minimum run time left on the battery taking all these factors into account.

Charge Rate: The ideal charge rate requires 15 hours to get to full charge, which is undesirable from the user’s perspective. The instrument provides a multiphase charge cycle which results in about 80% capacity in the first 2 hours after fast charge. The next charge phase, top-up, is designed to finish the charge and to bring all individual cells to the fully charged state, essentially rematching them. See the "Battery and

Charging Process" section in this chapter for "Fast Charge" and "Top-Up Charge" information. If the top-up charge is not completed, the cell mismatch is not reduced and the cumulative capacity reduction occurs. Top-up is a 3-hour charge but the elapsed time to complete it may be over 5 hours, as the charger is turned on and off to keep the battery cool during that time.

Alaris® PC Unit

1-5

Alaris® Pump Module

Technical Service Manual

GENERAL INFORMATION

1.4BATTERY CAPACITY INFORMATION

(Continued)

Cycle Life and Aging: Batteries wear out as they get older and go through many charge/discharge cycles; the chemicals and materials used to construct the cell break down. The instrument deals with this by assuming that a battery will continually reduce

capacity at a rate equivalent to 30% over 4 years and continually reduce capacity at a rate equivalent to 30% per 200 full charge cycles. These calculated values are used to reduce the run time displayed on the battery gauge.

Partial Discharge/Recharge: When a battery is partially discharged and then charged for less than the full time, differences between individual cell capacities result in cells completing charge at different times. If the full charge sequence is not completed, the cell mismatch becomes progressively

greater. This is viewed by the user as low apparent run times and premature low battery warning and alarms. The problem is cumulative in that the mismatch increases for every partial cycle. The lowered capacity is not permanent but may require 2-3 full discharge/charge cycles to recover. The instrument

deals with this by reducing the run time displayed based on a limited history of partial cycles.

Temperature During Charge: As the effective ambient temperature of the battery increases, the amount of charge that the battery will accept decreases. At an ambient temperature of 35° C, an enclosed battery will temporarily accept only about 90% of the charge it would otherwise accept at 23° C.

1.5POLE CLAMP FEATURE

The PC Unit’s pole clamp adapts to a wide variety of surfaces, to provide versatility. The pole clamp features include:

Ergonomically designed knob.

Accommodates diameters from 5/8 to 13/8 inches (15.9 to 34.9 mm).

Vertical or horizontal orientation, allowing it to adapt to both IV poles and bed rails.

If pole-mounted, the system should be mounted in an upright position during operation, to provide maximum stability and display readability.

1.6 MEMORY

Model 8000:

Flash 4.5 MB (nonvolatile): It stores boot and application codes for the PC Unit.

RAM 4 MB (volatile): It stores the Data Set, CQI logs, event logs, error logs, battery logs, and patient data. It is volatile but has a lithium battery on the Logic Board to maintain memory when no other power is applied.

Model 8015:

Flash 16 MB (nonvolatile): It stores boot and application codes for the PC Unit, CQI logs, event logs, error logs, and battery logs.

CF Flash 64 MB (nonvolatile): It stores audio tones, the Data Set, and "restorable" patient data.

SDRAM 128 MB (volatile): It executes transferred application code from flash. It stores patient-specific program and instantaneous information.

1-6 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

GENERAL INFORMATION

Table 1-3. Abbreviations, Acronyms, Symbols

Various abbreviations, acronyms and symbols are used throughout this manual. This table defines those that are not commonly known or easily recognized.

*"active low" logic signal

/not - true when signal is low

AIL

air-in-line

AKB

PC Unit keypad processor

DFU

Directions for Use

DS

display

FDSA

Fluid Delivery Subassembly

IUI

inter-unit interface

NMD

nonmaskable interrupt

PSP

power supply processor

PWN

pulse width N (where "N" is width of pulse)

SEC-PRI

secondary-to-primary

TH

thermistor

UCS

upper chip-select

VDAC

voltage DAC

VEE

voltage emitter

VRAW

voltage raw (unregulated voltage)

WDI

watchdog input

WDO

watchdog output

Alaris® PC Unit

1-7

Alaris® Pump Module

Technical Service Manual

GENERAL INFORMATION

THIS PAGE

INTENTIONALLY LEFT BLANK

1-8 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

Chapter 2 – CHECKOUT AND CONFIGURATION

CAUTIONS

2.1 INTRODUCTION

Should an instrument be dropped or severely jarred, remove it from use immediately. It should be thoroughly tested and inspected by qualified service personnel to ensure proper function prior to reuse.

Keep the Pump Module door closed when the instrument is not in use.

NOTE: The PC Unit does not feature a learn/teach function or a default drug list.

This chapter describes PC Unit and Pump Module initial setup and configuration. Due to product changes over time, configurations described in this chapter may differ from the instrument being serviced.

2.2NEW INSTRUMENT CHECKOUT

Prior to placing a new instrument in use, perform a check-in procedure using the applicable maintenance software.

When powering up the instrument, verify the instrument beeps and all display LED segments flash. This confirms that the instrument has performed its self test and is operating correctly. During operation, the instrument continually performs a self test, and alarms and displays a message if it detects an internal malfunction.

Contact CareFusion authorized service personnel if the instrument has physical damage, fails to satisfactorily pass the startup sequence, fails a self test, or continues to alarm.

2.3CONFIGURATION OPTIONS AND SETUP - GENERAL

Reference the Alaris® System DFU for the following information:

System Settings

Pump Module Settings

2.3.1Configuration Notes

Changes to factory default values are retained after a power cycle.

If Factory Default is Yes, then all configuration settings are set to their factory default.

Alaris® PC Unit

2-1

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.3CONFIGURATION OPTIONS AND SETUP - GENERAL (Continued)

2.3.1Configuration Notes (Continued)

If Factory Default is No, then 1 or more of the configuration settings has been changed. If desired, Factory Default can be selected and set to Yes, which will set all configuration settings to their factory default.

With Profiles feature enabled, settings are configured independently for each profile. A hospital-defined, best-practice Data Set must be loaded to enable Profiles feature. Date and Time is a system setting and is the same in all profiles.

If Guardrails® software is being used to load Data Sets into PC Unit, leave

configuration settings at factory defaults. When a Data Set is loaded, it overrides configuration settings because each profile has its own individual configuration settings.

2.3.2Configuration Setup Notes

Disabling Profiles option disables loaded Data Set and allows a configuration option to be changed.

Pressing EXIT soft key while in a

System Configuration - Module screen immediately powers system down, with no "Powering Down" display.

Pressing EXIT soft key while in a System Config screen returns display to main

System Configuration - Module screen.

Pressing CONFIRM soft key while in a System Configuration option screen:

-accepts existing setting or setting change

-displays next option setting screen (if applicable) or returns display to

System Config screen

Pressing PC Unit’s CANCEL key while in a System Configuration option screen:

-leaves setting unchanged

-returns display to System Config screen

2.4CONFIGURATION SETUP - PC UNIT

2.4.1Access System Configuration Options

1.Hold OPTIONS key at power up.

System Configuration - Module

Factory default:

No

Shared Infusion Settings

PC Unit

Pump Module

SPO2 Module

>Select an Option

or EXIT

EXIT

PAGE

DOWN

 

2.Press PC Unit soft key. To view additional options, press PAGE DOWN soft key.

System Config - PCU 1 of 2

Alarm audio:

Profile 1

Anesthesia Mode: Disabled

Battery meter:

Disabled

Clock setup:

09:00

Key click audio:

Enabled

>Select an Option or EXIT

PAGE

EXIT DOWN

-- Continued Next Page --

2-2 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.4CONFIGURATION SETUP - PC UNIT

(Continued)

2.4.1Access System Configuration Options

(Continued)

System Config - PCU 2 of 2

Max Pt. weight: 500 kg

Patient ID Entry: Disabled

PM reminder:

Disabled

Profiles: Disabled

Tamper resist: Disabled

>Select an Option or EXIT

PAGE

UP EXIT

2.4.2 Alarm audio

There are 3 different alarm profiles that can be selected to help differentiate between audio alarms in certain hospital environments.

Audio alert patterns (±5%):

Profile 1: 500 ms ON, 1500 ms OFF,

500 ms ON

Profile 2: 100 ms ON, 500 ms OFF,

400 ms ON

2.To change profile selection, press soft key next to desired profile. To hear a sample of selected audio alarm, press Test soft key.

3.To accept setting, press Confirm soft key.

2.4.3Anesthesia Mode

The Anesthesia Mode allows access to additional drugs in each profile that is appropriate to anesthesiology. This mode also features permanent pause and the ability to set higher air-in-line settings.

1.After accessing System Config - PCU options display, press Anesthesia Mode soft key.

System Configuration - PCU

Anesthesia Mode

Enable

 

 

Disable

 

 

 

 

 

Allows use of the Anesthesia

Mode feature.

Profile 3: 400 ms ON, 500 ms OFF,

400 ms ON

1.After accessing System Config - PCU options display, press Alarm audio soft key.

System Configuration - PCU

 

Profile

Alarm audio profile

1

selection

Profile

 

2

Select the desired alarm

Profile

audio patterns. Press Test

3

to hear a sample of the

Confirm

selected profile.

 

 

Test

 

 

2.To change setting, press soft key next to applicable option (Enable or Disable).

2.4.4Battery meter

When Enable is selected, the approximate time remaining on the battery is displayed (in a numeric value). The run time depends on the battery charge/condition and the load on the battery at the time. Run time may also be affected by the operating mode, rate, monitoring options and back pressure. The battery meter reading may take up

to 5 minutes after starting an infusion to stabilize.

Alaris® PC Unit

2-3

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.4CONFIGURATION SETUP - PC UNIT

(Continued)

2.4.4Battery meter (Continued)

1.After accessing System Config - PCU options display, press Battery meter soft key.

System Configuration - PCU

Battery meter

Enable

 

 

Disable

 

 

 

 

 

Allows the pump to display remaining battery run time to the user.

2.To change setting, press soft key next to applicable option (Enable or Disable).

2.4.5Clock setup

This option is used to set the time and date.

1.After accessing System Config - PCU options display, press Clock setup soft key.

System Configuration - PCU

Clock setup

 

Current time:

 

Change

09:00

Time

 

 

Current date:

Change

1998-12-15

Date

 

>Confirm Time-of-Day and

Date

2.To change time:

a. Press Change Time soft key.

System Configuration - PCU

Clock setup

 

 

Current time:

 

 

Change

 

__:__

 

Time

 

 

 

 

Current date:

 

Change

 

1998-12-15

 

Date

 

 

 

 

>Enter Time-of-Day

b. Enter time.

System Configuration - PCU

Clock setup

 

 

Current time:

 

 

Change

 

09:30

 

Time

 

 

 

 

Current date:

 

Change

 

1998-12-15

 

Date

 

 

 

 

>Press CONFIRM

CONFIRM

3.To change date:

a. Press Change Date soft key.

System Configuration - PCU

Clock setup

 

Current time:

 

Change

09:30

 

Time

 

 

 

Current date:

 

Change

yyyy-mm-dd

 

Date

 

 

>Enter Current Date

CONFIRM

CONFIRM

2-4 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.4CONFIGURATION SETUP - PC UNIT

(Continued)

2.4.5Clock setup (Continued)

b. Enter date.

System Configuration - PCU

Clock setup

 

Current time:

 

Change

09:30

Time

 

 

Current date:

Change

2001-01-29

Date

 

>Press CONFIRM

CONFIRM

4.To accept settings, press CONFIRM soft key.

2.4.6Key click audio

This option is used to Enable or Disable the keypress audio feedback.

1.After accessing System Config - PCU options display, press Key click audio soft key.

System Configuration - PCU

Key click audio

Enable

 

 

 

Disable

Provides an audible sound with each keypress.

2.To change setting, press soft key next to applicable option (Enable or Disable).

2.4.7Max Pt. weight

This option is used to set the maximum patient weight.

1.After accessing System Config - PCU options display, page 2, press Max Pt. weight soft key.

System Configuration - PCU

Maximum patient weight adjustment

No

Maximum weight:

 

Change

500 kg

value

 

 

Confirm

 

 

2.To change maximum weight:

a. Press Change value soft key.

System Configuration - PCU

Maximum patient weight adjustment

Maximum weight: Change

_ _ _ _kg

value

 

b. Enter new weight.

System Configuration - PCU

Maximum patient weight adjustment

Maximum weight:

Change

_100 kg

value

 

 

Confirm

 

 

Alaris® PC Unit

2-5

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.4CONFIGURATION SETUP - PC UNIT

(Continued)

2.4.7Max Pt. weight (Continued)

3.To accept setting, press Confirm soft key.

2.4.8Patient ID Entry

Enabling the Patient ID feature allows an alphanumeric patient identifier to be entered in the Patient ID Entry screen that displays after responding Yes to New Patient?. When the Patient ID feature is disabled, the Patient ID Entry screen can only be accessed through the Systems Options menu.

1.After accessing System Config - PCU options display, page 2, press Patient ID Entry soft key.

System Configuration - PCU

Patient ID Entry

Enable

 

 

Disable

Allows the patient ID to be assigned at system start up.

1.After accessing System Config - PCU options display, page 2, press PM reminder soft key.

System Configuration - PCU

Preventive

Enable

Maintenance reminder

 

Disable

 

 

 

 

 

Notifies user at power on when modules are due for routine preventive maintenance.

2.To change setting, press soft key next to applicable option (Enable or Disable).

2.4.10Profiles

A profile is a unique set of system configuration settings and best-practice guidelines for a specific patient population or patient type. Profile settings are established prior to system implementation. The Profiles option can be enabled only when a Data Set is loaded. Disabling the Profiles option disables the loaded Data Set.

1.After accessing System Config - PCU options display, page 2, press Profiles soft key.

2.To change setting, press soft key next to applicable option (Enable or Disable).

2.4.9PM reminder

Enabling the PM reminder feature allows a MAINTENANCE REMINDER message to appear when the PC Unit or attached module is due for routine scheduled preventive maintenance.

System Configuration - PCU

Profiles

Enable

 

 

Disable

Allow the user to select from an institution defined list of Profiles

2.To change setting, press soft key next to applicable option (Enable or Disable).

2-6 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

999 mL/h
PAGE
DOWN
1 mL/h

CHECKOUT AND CONFIGURATION

2.4CONFIGURATION SETUP - PC UNIT

(Continued)

2.4.11 Tamper resist

When Enable is selected, this mode can be activated or deactivated by pressing and holding the Tamper Resist Switch on the back of the PC Unit for 3 to 4 seconds. Reference the Alaris® System DFU for more detailed information.

Prior to accessing the Tamper resist mode, all modules that are to be operating must be set up.

System Configuration - PCU

Tamper resistant mode

Enable

 

 

Disable

 

 

 

Allows use of the tamper

Auth

Resistant Mode feature.

user

 

Change

 

override

 

 

1.After accessing System Config - PCU options display, page 2, press Tamper resist soft key.

2.To change setting, press soft key next to applicable option (Enable or Disable).

2.5CONFIGURATION SETUP - PUMP MODULE

2.5.1Access System Configuration Options

1.Hold OPTIONS key at power up.

System Configuration - Module

Factory default:

No

Shared Infusion Settings

PC Unit

Pump Module

SPO2 Module

>Select an Option

or EXIT

EXIT

PAGE

DOWN

 

2.Press Pump Module soft key. To view additional options, press PAGE DOWN soft key.

System Config - Pump 1 of 2

Accumulated air: Enabled

Air-in-line settings: 75microliter -

Auto-restart attempts: 0

KVO rate adjust:

Max rate:

>Select an Option or EXIT

EXIT

System Config - Pump 2 of 2

Max VTBI:

9999 mL

Pressure mode: Pump

SEC-PRI alert: Enabled

Secondary: Disabled

>Select an Option or EXIT

PAGE

EXIT

UP

 

Alaris® PC Unit

2-7

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.5CONFIGURATION SETUP - PUMP MODULE (Continued)

2.5.2Accumulated air

When Disable is selected, the air-in-line system detects a bolus size of »50, 75 or 250 microliters only. When Enable is selected, the air-in-line system detects a bolus size of »50, 75 or 250 microliters and a number of small boluses of a predetermined number, depending on the bolus size selected in the Air-in-line settings option.

1.After accessing System Config - Pump options display, press Accumulated air soft key.

System Configuration - Pump

Accumulated Enable

Air-in-line

Disable

Detects the presence of multiple small air bubbles.

2.To change setting, press soft key next to applicable option (Enable or Disable).

2.5.3Air-in-line settings

There are 3 different air-in-line bolus settings that can be selected; 50, 75 or 250 microliters. This is the amount of air allowed to pass through the detector before an air-in- line alarm sounds.

1.After accessing System Config - Pump options display, press Air-in-line settings soft key.

System Configuration - Pump

Air-in-line detection

50

 

threshold

75

(in microliters)

 

Sets the upper limit

250

for a single bolus of air

 

 

to pass without alarm

 

 

 

 

 

 

 

2.To change setting, press soft key next to applicable option.

2.5.4Auto-restart attempts

This setting determines the number of attempts (0 to 9) the instrument will make to restart the infusion (following detection of a patient-side occlusion) before it alarms "Occlusion".

1.After accessing System Config - Pump options display, press Auto-restart attempts soft key.

System Configuration - Pump

Auto-Restart Mode

Increase

 

 

 

Attempts:

Decrease

 

0

Confirm

Determines the number of attempts to restart the infusion following detection of a patient side occlusion before the pump will alarm.

2-8 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.5CONFIGURATION SETUP - PUMP MODULE (Continued)

2.5.4Auto-restart attempts (Continued)

2.To set number of attempts, press

Increase or Decrease soft key.

When the maximum number of attempts is selected, Increase is grayed-out (disabled).

System Configuration - Pump

Auto-Restart Mode

Increase

 

Attempts:

Decrease

3

 

Confirm

 

 

 

Determines the number of attempts to restart the infusion following detection of a patient side occlusion before the pump will alarm.

3.To accept setting, press Confirm soft key.

2.5.5KVO rate adjust

This option is used to select the KVO (Keep Vein Open) rate (0.1 to 20 mL/h). This determines the rate of fluid flow after "Infusion Complete" has occurred.

1.After accessing System Config - Pump options display, press KVO rate adjust soft key.

System Configuration - Pump

KVO rate adjustment

KVO rate:

Change

KVO rate

1 mL/h

Confirm

 

 

 

2.To change KVO rate:

a. Press Change KVO rate soft key.

System Configuration - Pump

KVO rate adjustment

KVO rate:

Change

KVO rate

_ _ mL/h

 

 

 

 

 

 

 

b. Enter desired rate.

System Configuration - Pump

KVO rate adjustment

KVO rate:

Change

KVO rate

5 mL/h

Confirm

 

 

 

3.To accept setting, press Confirm soft key.

Alaris® PC Unit

2-9

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.5CONFIGURATION SETUP - PUMP MODULE (Continued)

2.5.6Max rate

This option is used to select the maximum rate (0.1 to 999 mL/h).

1.After accessing System Config - Pump options display, press Change value soft key.

System Configuration - Pump

Maximum rate adjustment

Maximum rate:

Change

999 mL/h

value

Confirm

 

 

 

2.To change rate:

a. Press Change value soft key.

System Configuration - Pump

Maximum rate adjustment

Maximum

Change

rate:

value

_ _ _ mL/h

 

b. Enter desired rate.

System Configuration - Pump

Maximumrate

 

adjustment

 

Maximum rate:

Change

99.9 mL/h

value

Confirm

 

3.To accept setting, press Confirm soft key.

2.5.7Max VTBI

This option is used to select the maximum VTBI adjustment rate (0.1 to 9999 mL).

1.After accessing System Config - Pump options display, page 2, press Max VTBI soft key.

System Configuration - Pump

Maximum VTBI adjustment

Maximum VTBI:

Change

9999 mL

value

 

Confirm

2-10 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.5CONFIGURATION SETUP - PUMP MODULE (Continued)

2.5.7Max VTBI (Continued)

2.To change rate:

a. Press Change value soft key.

System Configuration - Pump

Maximum VTBI adjustment

Maximum VTBI:

Change

_ _ _ _ mL

value

 

 

Selectable: Occlusion limit of 50 to

525 mmHg (approximately 1 to 10.2 psi), in

25 mmHg increments, can be selected.

1.After accessing System Config - Pump options display, page 2, press Pressure mode soft key.

System Configuration - Pump

Pressure mode

Pump

selection

 

Selectable

Pressure mode:

 

Pump

Change

 

Lock Status

 

 

Lock status:

 

Unlocked

 

CONFIRM

b. Enter desired rate.

System Configuration - Pump

Maximum VTBI adjustment

Maximum VTBI:

Change

999.9 mL

value

Confirm

 

 

 

 

 

 

 

 

 

3.To accept setting, press Confirm soft key.

2.5.8Pressure mode

There are 2 pressure modes available that determine the patient-side occlusion limit. The selected mode can be Unlocked or Locked after being confirmed.

Pump: Instrument will occlude at approximately 10 psi pressure.

2.To change setting:

a.Ensure lock status is Unlocked.

b.Press soft key next to applicable option (Pump or Selectable).

If Selectable mode is selected, following screen displays.

System Configuration

Selectable Max

 

Up

Pressure:

 

Down

525 mmHg

 

 

 

 

 

 

 

 

Selectable

 

Up

Default Pressure:

 

Down

300 mmHg

 

 

 

 

 

 

 

 

 

 

 

 

CONFIRM

Alaris® PC Unit

2-11

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.5CONFIGURATION SETUP - PUMP MODULE (Continued)

2.5.8Pressure mode (Continued)

3.To change pressure setting:

a.Press Up or Down soft key next to applicable option.

When maximum pressure setting is selected, Up is grayed-out (disabled). When minimum pressure setting is selected, Down is grayed-out (disabled).

c.To lock selection, press Change Lock Status soft key.

System Configuration - Pump

Pressure mode

Pump

selection

 

 

Selectable

Pressure mode:

 

Change

Selectable

Lock Status

350 mmHg

 

Lock status:

 

Locked

 

CONFIRM

System Configuration

Selectable Max

 

Up

Pressure:

 

 

 

Down

400 mmHg

 

 

 

 

 

 

 

 

Selectable

 

Up

Default Pressure:

 

Down

350 mmHg

 

 

 

 

 

 

 

 

 

 

 

 

CONFIRM

4.To accept setting, press CONFIRM soft key.

2.5.9SEC-PRI alert

When Enable is selected, an alarm sounds when the instrument switches from Secondary to Primary mode.

1.After accessing System Config - Pump options display, page 2, press SEC-PRI alert soft key.

b.To accept setting, press CONFIRM soft key.

System Configuration - Pump

Pressure mode

Pump

selection

 

 

 

Selectable

 

 

Pressure mode:

Change

Selectable

Lock Status

 

350 mmHg

 

Lock status:

 

Unlocked

 

CONFIRM

System Configuration - Pump

Secondary to

Enable

primary transition

 

 

 

 

alert

Disable

 

 

 

Causes six audio tones to sound when the infusion transitions from secondary to primary.

2.To change setting, press soft key next to applicable option (Enable or Disable).

2-12 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.5CONFIGURATION SETUP - PUMP MODULE (Continued)

2.5.10 Secondary

The Secondary mode is used to deliver 2 different fluids sequentially through the same line.

1.After accessing System Config - Pump options display, page 2, press

Secondary soft key.

System Configuration - Pump

Secondary mode

Enable

 

 

 

 

Disable

 

 

 

 

Allows use of the secondary feature.

2.To change setting, press soft key next to applicable option (Enable or Disable).

2.6CONFIGURATION SETUP - SHARED INFUSION

2.6.1Access System Configuration Options

1.Hold OPTIONS key at power up.

System Configuration - Module

Factory default:

No

Shared Infusion Settings

PC Unit

Pump Module

SPO2 Module

>Select an Option

or EXIT

EXIT

PAGE

DOWN

 

2.Press Shared Infusion Settings soft key.

System Config - Shared

Delay Options: Disabled

Drug Calculation: Disabled

Multidose: Disabled

Pressure dynamic:Disabled

Volume/Duration: Disabled

>Select an Option

or EXIT

EXIT

2.6.2Delay Options

Enabling Delay Options allows the system to be programmed to delay the start of an infusion a) for up to 120 minutes or b) for

a specific time up to 23 hours 59 minutes. A callback for a programmed delay can be scheduled to give an alert Before an

infusion is to be initiated, After an infusion is completed, Before and After an infusion, or to give no alert (None).

In the following procedure, Delay Options is changed to Enable. Changing the option setting to Disable disables the Callback soft key, as displayed in step 1.

1.After accessing System Config - Shared options display, press Delay Options soft key.

System Configuration - Shared

Enable

Delay Options

Disable

Callback:

None Callback

Allows use of the Delay Options and Callback features.

CONFIRM

Alaris® PC Unit

2-13

Alaris® Pump Module

Technical Service Manual

CHECKOUT AND CONFIGURATION

2.6CONFIGURATION SETUP - SHARED MODULE (Continued)

2.6.2Delay Options (Continued)

2.To change setting, press soft key next to applicable option (Enable or Disable).

System Configuration - Shared

Enable

Delay Options

Disable

Callback:

None Callback

Allows use of the Delay Options and Callback features.

CONFIRM

3.To change callback setting: a. Press Callback soft key.

System Configuration - Shared

Callback Options

 

Before

 

 

 

Before

None

& After

After

 

 

 

 

None

 

 

CONFIRM

b.Press soft key next to applicable setting (following example illustrates Before as having been selected).

System Configuration - Shared

Callback Options

 

 

Before

 

 

 

 

 

Before

Before

& After

After

 

 

 

 

 

 

 

None

 

 

 

CONFIRM

4.To accept setting, press CONFIRM soft key.

2.6.3Drug Calculation

Enabling Drug Calculation allows:

entry of drug dose (correct flow rate to achieve desired dose is automatically calculated),

OR

entry of flow rate (corresponding drug dose is automatically calculated).

In the following procedure, the Drug Calculation option is changed to Enable. Changing the option setting to Disable disables the Bolus Dose soft keys, as displayed in step 1.

1.After accessing System Config - Shared options display, press Drug Calculation soft key.

System Configuration - Shared

Drug Calculation

Enable

Allows use of the

 

Drug Calculation

Disable

feature.

 

 

 

Bolus Dose

Enable

Allows use of the

 

Bolus Dose

Disable

feature.

Disable

CONFIRM

2-14 Alaris® PC Unit

Alaris® Pump Module

Technical Service Manual

Loading...
+ 160 hidden pages