CareFusion Infant Flow SiPAP User manual

Infant Flow® SiPAP™

Operator’s manual

ii Infant Flow
®
This document is protected by United States and International Copyright laws.
This document may not be copied, reproduced, translated, stored in a retrieval system, transmitted in any form, or reduced to any electronic medium or machine-readable form, in whole or in part, without the written permission of CareFusion. Information in this document is subject to change without notice.
This document is for informational purposes only and should not be considered as replacing or supplementing the terms and conditions of the License Agreement.
© 2009–2010 CareFusion Corporation or one of its subsidiaries. All rights reserved. Infant Flow is a registered trademark and SiPAP is a trademark of CareFusion Corporation or one of its subsidiaries. All other trademarks are the property of their respective owners.
USA
CareFusion 22745 Savi Ranch Parkway Yorba Linda, California 92887-4668
Authorized European Representative
CareFusion Germany 234 GmbH Leibnizstrasse 7 97204 Hoechberg, Germany District Court Wuerzburg HRB7004
800.231.2466 tel +1.714.283.2228 tel
+49.931.4972.0 tel
+49.931.4972.423 fax
+1.714.283.8493 fax
carefusion.com
Literature number: 675–101–101 Revision K
675–101–101 Rev. K
Infant Flow® SiPAP™ iii

R evision History

Date Revision Changes
September 2003 675-101(3) Release
August 2004 D Release manual in VIASYS Respiratory Care template using
VIASYS Respiratory Care nomenclature. Revise part number list in Appendix B approved accessories.
November 2004 E
March 2005 F Updated the contact information.
May 2006 G Updated the company name.
Revised contact/ordering information.
Ch 4. Sec. 5.b revised transducer LED illumination conditions.
Appendix A corrected units from "Pm" to bpm.
Appendix E added dimension ranges to bonnet sizes.
Appendix E Was 467350 Transducer Assembly Is: 677-002 Transducer Interface.
Updated the Declaration of Conformity Notice.
Updated the Contact and Ordering Information.
Update the figures.
Added a Caution regarding back pressure. Added a Note regarding the Hudson RCI Humidification System.
Added the sentence “Ensure there is a minimum 8 LPM set
on the NCPAP/PRES Low Flow meter”
under “Two Point O
Changed step 8 regarding the nCPAP pressure.
Changed the second and third paragraphs under Changing a Control.
Sensor Calibration.
2
to the first paragraph
Added “Setting a Manual Breath.”
Added a note regarding the enabling of manual breath or back­up apnea breath.
Added a warning concerning infant flow consumables.
Added the statement “Disconnect the air and oxygen gas sources when the Infant Flow SiPAP
Removed Appendix E.
February 2009 H Changed Ti to T-High and Inspiratory Time to Time High.
Replaced reference to VIASYS Respiratory Care accessories with reference to Cardinal Health accessories.
Added “TM” superscript to “SiPAP”.
Added reference to AirLifeTM Infant nCPAP System Generator.
Removed “inspiratory time” or “inspiration time”.
Replaced “Inspiratory Time (Time High)” with “Time High
675–101–101 Rev. K
is not in use.”
iv Infant Flow
Date Revision Changes
(Thigh)”
Changed 1 cmH2O to 1.5 cmH2O; Added “or 60 psi” to clarify 4 bar.
Added the parts list for both Infant Flow® Products and AirLifeTM Products.
Added reference to Cardinal Health contact information on page v.
Added reference to AirLifeTM Infant nCPAP System accessories.
Added a warning about using an external oxygen monitor. Added reference to factory trained technician and Service
Manual P/N 675-120.
Added “®” (registered symbol) superscript to Infant Flow.
Updated CAUTION label: from “Back pressure from the humidifier chamber to some auto-feed water bags may occur.” To “Back pressure from some auto-feed humidifier chambers may cause the water bags to fill with air.”
®
Replaced Figure 5.
Add content concerning a depleted or damaged internal oxygen cell.
Added a warning about using an external oxygen monitor. Added content to explain fault code E5X.
Replaced “key” with “button”; clarified oxygen alarm by adding “the audible”; added clarification of the internal monitoring being disabled and that an external oxygen monitor must be used.
Added a Note regarding the 2nd Flow Meter being used for manual breath delivery;
Added hyphen in “T-High”.
Clarified the “Mode Select Screen”
Added “Directions for using the AirLifeTM Infant nCPAP System.
Changed 1 cmH2O to 1.5 cmH2O; Corrected low battery voltage level from 10 to 11.10.
Added “or trained biomedical engineer”.
Added a table entry for the oxygen monitor and alarms disable.
Changed 1 cmH2O to 1.5 cmH2O; corrected low battery voltage level from 10 to 11.10.
Updated Table 10.
Updated Table 11.
Clarified the meaning of T-High.
February 2010 J Revised to comply with the revised Medical Device Directive
2007/42/EC.
675–101–101 Rev. K
Infant Flow® SiPAP™ v
Date Revision Changes
March 2010 K Rebranded the manual to the CareFusion style.
Updated the part number table.
675–101–101 Rev. K
vi Infant Flow

Warranty

Infant Flow® SiPAPTM is warranted to be free from defects in material and workmanship and to meet the published specifications for One (1) year from date of shipment.
The liability of CareFusion (referred to as the Company) under this warranty is limited to replacing, repairing or issuing credit, at the discretion of the Company, for parts that become defective or fail to meet published specifications during the warranty period; the Company will not be liable under this warranty unless (A) the Company is promptly notified in writing by Buyer upon discovery of defects or failure to meet published specifications; (B) the defective unit or part is returned to the Company, transportation charges prepaid by Buyer; (C) the defective unit or part is received by the Company for adjustment no later than four weeks following the last day of the warranty period; and (D) the Company’s examination of such unit or part shall disclose, to its satisfaction, that such defects or failures have not been caused by misuse, neglect, improper installation, unauthorized repair, alteration or accident.
Any authorization of the Company for repair or alteration by the Buyer must be in writing to prevent voiding the warranty. In no event shall the Company be liable to the Buyer for loss of profits, loss of use, consequential damage or damages of any kind based upon a claim for breach of warranty, other than the purchase price of any defective product covered hereunder.
®
The Company warranties as herein and above set forth shall not be enlarged, diminished or affected by, and no obligation or liability shall arise or grow out of the rendering of technical advice or service by the Company or its agents in connection with the Buyer's order of the products furnished hereunder.
Limitation of Liabilities
This warranty does not cover normal maintenance such as cleaning, adjustment or lubrication and updating of equipment parts. This warranty shall be void and shall not apply if the equipment is used with accessories or parts not manufactured by the Company or authorized for use in writing by the Company or if the equipment is not maintained in accordance with the prescribed schedule of maintenance.
The warranty stated above shall extend for a period of One (1) year from date of shipment, with the following exceptions:
1. Components for monitoring of physical variables such as temperature, pressure, or flow are warranted for ninety (90) days from date of receipt.
2. Elastomeric components and other parts or components subject to deterioration, over which the Company has no control, are warranted for sixty (60) days from date of receipt.
3. Internal batteries are warranted for ninety (90) days from the date of receipt.
The foregoing is in lieu of any warranty, expressed or implied, including, without limitation, any warranty of merchantability, except as to title, and can be amended only in writing by a duly authorized representative of the Company.
675–101–101 Rev. K
Infant Flow® SiPAP™ vii

Contents

Revision History ......................................................................................................... iii
Warranty .................................................................................................................... vi
Contents ................................................................................................................... vii
List of Figures .......................................................................................................... viii
List of Tables ........................................................................................................... viii
Notices ....................................................................................................................... ix
Chapter 1 - Product Description ................................................................................. 1
Chapter 2 - Product Specifications............................................................................. 3
Chapter 3 - Summary of Warnings and Cautions....................................................... 7
Chapter 4 - Unpacking & Setup ............................................................................... 11
Chapter 5 - Operation .............................................................................................. 23
Chapter 6 - Operating Modes .................................................................................. 33
Chapter 7 - Alarms and Indicators ........................................................................... 35
Chapter 8 - Maintenance & Cleaning ....................................................................... 41
Chapter 9 – Explanation of Symbols ........................................................................ 43
Appendix A - Product Configurations ....................................................................... 49
Appendix B - Pneumatic Diagram ............................................................................ 51
Appendix C - Alarm Troubleshooting ....................................................................... 53
Appendix D - Fault Management ............................................................................. 57
Glossary ................................................................................................................... 63
Index ........................................................................................................................ 65
675–101–101 Rev. K
viii Infant Flow

List of Figures

Figure 1 – Stand unpacking and assembly ........................................................ 11
Figure 2 – Stand and Driver assembly ............................................................... 12
Figure 3 – Driver assembled with patient circuit and /humidifier ........................ 13
Figure 4 – Attaching the Abdominal Respiratory Sensor ................................... 14
Figure 5 – Flow Pressure Nomogram ................................................................ 15
Figure 6 – Front Panel ....................................................................................... 23
Figure 7 – Rear Panel ........................................................................................ 24
Figure 8 – Set Up Screen .................................................................................. 27
Figure 9 – Alarm set/confirm Screen ................................................................. 28
Figure 10 – Mode Select Screen ....................................................................... 29
Figure 11 – Parameter Adjust Screens .............................................................. 29
Figure 12 – Main Screen .................................................................................... 30
®
Figure 13 – Monitored Parameters Screen ........................................................ 30
Figure 14 – NCPAP ........................................................................................... 33
Figure 15 – BiPhasic .......................................................................................... 33
Figure 16 – BiPhasic tr....................................................................................... 34
Figure 17 - Flat Battery screen .......................................................................... 37

List of Tables

Table 1 – Functions and Accessories .................................................................. 2
Table 2 - Soft-key operation ............................................................................... 25
Table 3 – Parameter Default Value .................................................................... 27
Table 4 – Alarm Symbols and Indicators ........................................................... 38
Table 5 – Equipment Symbols ........................................................................... 43
Table 6 – Button Symbols .................................................................................. 45
Table 7 – Non-US Configuration Parameters .................................................... 49
Table 8 – US Configuration Parameters ............................................................ 49
Table 9 – Alarm Troubleshooting ....................................................................... 53
Table 10 – Fault Classification ........................................................................... 57
Table 11 – Fault Recovery ................................................................................. 58
675–101–101 Rev. K
Infant Flow® SiPAP™ ix

Notices

E MC Notic e
This equipment radiates and is susceptible to radio frequency energy. If not installed and used in accordance with the instructions in this manual, electromagnetic interference may result. The equipment has been tested and found to comply with the limits set forth in BS EN60601-1-2 for Medical Electrical Equipment Part 1-2: General requirements for safety-collateral standard. Electromagnetic compatibility – requirements and tests. These limits provide reasonable protection against electromagnetic interference when operated in the intended use environments (e.g. hospitals) described in this manual.
This device is also designed and manufactured to comply with the following standards:
Safety: UL 60601-1: 2003 Medical Electrical Equipment, Part 1: General
Requirements for Safety.
CAN/CSA C22.2 No 601.1-M90, Medical Electrical Equipment - Part 1: General Requirements for Safety including C22.2 No. 601.1S1-94 (IEC601-1, Amendment 1:1991) Supplement No. 1-94 to CAN/CSA
22.2 No. 601.1-M90
Electrical Safety:
Class 1 equipment
Contains type BF patient applied parts
Continuous Operation
MR I Notice
This equipment contains electromagnetic components whose operation can be affected by intense electromagnetic fields.
Do not operate this device in a MRI environment or in the vicinity of high-frequency surgical diathermy equipment, defibrillators, or short-wave therapy equipment. Electromagnetic interference could disrupt the operation of the device.
Intended Use Notice
The Infant Flow® SiPAP™, consisting of a Driver and Generator plus NCPAP Prongs and Masks, is intended for the provision of Bi-Level CPAP (SiPAP™) to produce a sigh. The system is for use in Hospitals, Hospital Type facilities and intra-Hospital transport environments and is indicated for the treatment of Newborn and Infant patients. The
Infant Flow personnel, under the direction of a physician.
®
SiPAP™ should only be operated by properly trained clinical
675–101–101 Rev. K
x Infant Flow
R egulatory Notice
Federal law restricts the sale of this device except by or on order of a physician.
Reuse of single-patient use accessories may degrade the performance of the product or cause cross contamination.
Clas sification
Type of Equipment: Medical Equipment, Class 1 and internally powered, IPX1 Protected, and uses type BF applied parts. Equipment is not suitable for use in presence of flammable anesthetics.
Declaration of Conformity Notice
This medical equipment complies with the Medical Device Directive, 93/42/EEC, and the following Technical Standards, to which Conformity is declared:
EN60601-1 and EN60601-1-2 EN 10993 EN 14971
®
EU Notified Body:
BSI (Reg. No. 0086)
Trade names:
®
Infant Flow
SiPAP™
Manufactured by:
CareFusion 22745 Savi Ranch Parkway Yorba Linda, California 92887-4668
If you have a question regarding the Declaration of Conformity for this product, please contact CareFusion.
675–101–101 Rev. K
Infant Flow® SiPAP™ 1

Chapter 1 - Product Description

Infant Flow® SiPAP™ provides a non-invasive form of respiratory support designed for infants in hospital environments such as Neonatal and Pediatric Intensive Care Units. It can also be used when transporting these patients within the hospital environment.
Infant Flow configuration. The Plus configuration provides NCPAP and time triggered BiPhasic modes with and without breath rate monitoring. The Comprehensive* configuration offers these features plus a patient triggered BiPhasic mode with apnea backup breaths. The Infant Flow LCD touch screen display, pressure time waveform graphics, integrated patient monitoring, alarms for high and low pressure and FiO battery power.
As a result of the unique, patented design, the Infant Flow System Generator has been proven to provide the most stable CPAP at the lowest work of breathing for patients compared to other devices performance of the Infant Flow expiratory flows. This system has been designed and tested to perform optimally when used only with accessories available from CareFusion. These accessories include circuits and generators, prong and mask patient interfaces and bonnets.
®
SiPAP™ is currently available in a Plus or Comprehensive*
®
SiPAP™ comes standard in all configurations with an
and up to 2 hours of backup
2
®
or AirLifeTM Infant nCPAP
(1)
®
Generator is irrespective of patient demand or
. The outstanding

Infant Flow® SiPAP™ Features

The expanded capabilities of the Infant Flow® SiPAP™ Plus and Comprehensive* configurations allow for applications to broader range of patients who may otherwise not be candidates for non-invasive respiratory support from NCPAP alone
NCPAP – continuous positive airway pressure based on clinician set pressure. Breath rate monitoring/alarm can be activated in this mode.
BiPhasic – time triggered pressure assists are delivered based on clinician set Time-High, rate and pressure criteria. Breath rate monitoring/alarm can be activated in this mode.
BiPhasic tr* – patient triggered pressure assists delivered based on clinician set Time-High and pressure criteria. Breath rate monitoring/alarm and Apnea backup breaths are automatically active in this mode.
Patented Infant Flow
®
Flow
Generator is a fluidic device for the generation of consistent infant nasal CPAP
with a low work of breathing compared to other devices
Fully integrated alarm package – Supply gases failure, High Patient Pressure, Low patient pressure, high and low delivered Oxygen concentration, change from AC to DC power source, low and flat battery charge status and Low breath rate/apnea alarm.
®
or AirLifeTM Infant nCPAP System Generator - The Infant
(2,3)
.
(1)
.
Battery Backup – Up to 2 hours of battery backup allows for intra-hospital transport. Clear indicators are provided for power supply in use (AC or DC), and battery charge level.
675–101–101 Rev. K
2 Chapter 1 - Product Description
Screen Lock – After 120 seconds of no screen inputs, the screen changes to the Locked Screen to prevent inadvertent changes. Upon activation of a high priority alarm the screen changes to an unlocked state to allow access to controls.
Table 1 – Functions and Accessories
Functions & Accessories Plus Comprehensive*
NCPAP
NCPAP with breath rate monitoring and alarm
BiPhasic
BiPhasic with breath rate monitoring and alarm
BiPhasic tr*
Internal Battery
Manual Breath
Apnea Back up rate
Screen lock
Prioritization of alarms
*Comprehensive configuration not available for sale in the United States
(1) Decreased imposed work with a new nasal continuous positive airway pressure device.
Klausner, James F., PhD, Lee, Amy., Hutchison, Alastair A., FRACP. Pediatric Pulmonology 22: 188-194; 1996
(2) A Prospective Randomized, Controlled Trial Comparing Synchronized Nasal Intermittent Positive
Pressure Ventilation versus Nasal Continuous Positive Airway Pressure as Modes of Extubation. Khalaf Nabeel, M., Brodsky Nancy, Hurley John, Bhandari Vineet. PEDIATRICS 108 (1): 13-17: 2001
(3) Efficacy of Nasal Intermittent Positive Pressure Ventilation in Treating Apnea of Prematurity. Lin
Chyi-Her, MD, Wang Shan-Tair, PhD, Lin Yuh-Jyh, MD, Yeh Tsu-Fuh, MD:Pediatric Pulmonolgy: 26 (5): 349-53; 1996
675–101–101 Rev. K
Infant Flow® SiPAP™ 3

Chapter 2 - Product Specifications

Modes

NCPAP
NCPAP with breath rate monitoring and low rate alarm
BiPhasic (time triggered)
BiPhasic (time triggered) with breath rate monitoring and low rate alarm
BiPhasic tr (patient triggered) with breath rate monitoring, low breath rate
alarm and apnea back up (Comprehensive models only)

Controls

Time High (T-High) – 0.1 – 3.0 seconds
Rate (R)
1-120 (Non-U.S. Configuration Parameters)
1-54 (U.S. Configuration Parameters)
Apnea Interval
(T
(TLBR) – 10-30 seconds; 5 second intervals (U.S. Configuration
NCPAP / Pres Low flow meter – 0-15 L/min, accuracy ± 15% of selected
output
Pres High flow meter – 0-5 L/min, accuracy ± 15% of selected output
Manual Breath – X 1
%O

Monitors

CPAP
PEEP
MAP
PIP
) – 10-30 seconds, 5 second intervals (Non-U.S. Configuration
apnea
Parameters)
Parameters)
– 21 -100%
2
%O
I:E ratio
Spontaneous rate (Rsp)
Battery charge level
675–101–101 Rev. K
2
4 Chapter 2 - Product Specifications

Alarms

High airway pressure – 3 cmH2O above measured airway pressure
Airway over-pressure limit alarm
maximum 11 cmH
maximum 15 cmH2O in patient triggered BiPhasic tr mode
Low airway pressure – 2 cmH
1.5 cmH
O if otherwise would be zero
2
High and Low delivered Oxygen concentration ±5% of setting. Minimum and maximum delivered FiO2 is 18 and 104% respectively.
Low breath rate alarm
Low battery charge level
Flat battery
Input gases failure
Alarm volume (electronic alarms) 70 dBa at 1 meter

Pneumatic Supply

Patient Gas Outlet: 15 mm standard taper fitting
Patient Pressure Input: 4.5 mm Luer taper fitting
Gas Supply: Nominal 4 bar or 60 psi, clean, dry medical air and oxygen
O in NCPAP and time triggered BiPhasic mode
2
O below measured airway pressure or
2
Range: 40.61 to 87 PSI; Maximum differential pressure 29 PSI
Manometer: Range 0 to + 20 cmH
Gas Connections: Standard DISS, NIST or Air Liquide connectors

Electrical Supply

Input Voltage: 100-230 VAC
Input Frequency: 50/60 Hz
Power Consumption: 50 VA maximum
Fuse Rating For 220 V nominal operation: “T” Type 2.5 A at 250 V
Device Housing Protection rating level: IPX1
Battery Working Time: 2 hours (from fully charged state)
Battery Charging Time: max. 16 hours
O, accuracy, ± 2% of span
2
675–101–101 Rev. K
Infant Flow® SiPAP™ 5

Atmospheric & Environmental

Temperature Range
Operating: 5 – 40° C
Storage: - 20 - 50° C
Relative Humidity -Operating: 0 – 95% non-condensing
Storage: 0 – 95% non-condensing

Physical

Dimensions (Driver only)-
(W x H x D) 26 x38 x 23.5 cm
(W x H x D) 10.25 x15 x 9.25 in
Weight (Driver only)-
8.8 kg
19.5 lb

Accessories

Silencer / Bacterial Filter - The additional resistance of the D1420/100
Silencer / Bacterial Filter and adaptor is less than 0.56 cmH and less than 0.40 cmH
O at 5 LPM.
2
O at 15 LPM,
2
675–101–101 Rev. K
6 Chapter 2 - Product Specifications
F&P 730
773386–102
Patient Circuit w/Generator (box of 20)
773387–101
Patient Circuit (box of 20)
773388–105
Patient Circuit RCI 16V (box of 20)
Prongs
Masks
777086–101
Nasal Mask - Small (box of 10)
006905
NCPAP Generator Kit
006915
Nasal Prongs - Medium
006920
Nasal Prongs - Large
Part No. Description
Infant Flow® Products
D1420/100 Silencer (box of 20)
11541–101 Patient Circuit Assembly (box of 20) 11541–102 Patient Circuit Assembly w/Generator (box of 20) 773386–101 Patient Circuit (box of 20)
F&P 850
12204–101 Patient Circuit (box of 20) 12204–102 Patient Circuit w/Generator (box of 20)
12233–102 Patient AirLife Circuit w/Generator (box of 20) 12233–101 Patient AirLife Circuit (box of 20)
RCI
773388–103 Patient Circuit w/Generator RCI 16V (box of 20) 773389–105 Patient Circuit RCI 21V (box of 20) 773389–104 Patient Circuit w/Generator RCI 21V (box of 20)
11513–101 Nasal Prongs - Small (box of 10) 11513–102 Nasal Prongs - Medium (box of 10) 11513–103 Nasal Prongs - Large (box of 10)
777086–102 Nasal Mask - Medium (box of 10) 777086–103 Nasal Mask - Large (box of 10) 777086–104
Nasal Mask Extra Large (box of 10)
AirLifeTM Products (U.S.A. only)
Prongs
006910 Nasal Prongs - Small
Masks
006925 Nasal Mask - Small 006930 Nasal Mask - Medium 006935 Nasal Mask - Large
675–101–101 Rev. K
Infant Flow® SiPAP™ 7

Chapter 3 - Summary of Warnings and Cautions

Please review the following safety information prior to operating the Infant Flow SiPAP™. Attempting to operate this equipment without fully understanding its features and functions may result in unsafe operating conditions.
Warnings and Cautions, which are general to the use of the device under all circumstances, are included in this section. Some Warnings and Cautions are also inserted within the manual where they are most meaningful.
Notes are also located throughout the manual to provide additional information related to specific features.
If you have a question regarding the installation, set up, operation, or maintenance of the device, contact CareFusion (see page v).

Terms

WARNINGS identify conditions or practices that could result in serious adverse
CAUTIONS identify conditions or practices that could result in damage to the
NOTES identify supplemental information to help you better understand how
®
reactions or potential safety hazards.
driver or other equipment.
the driver works.

Warnings

Infant Flow® SiPAP™ is intended for use by a trained practitioner, under the
direct supervision of a qualified physician.
When the Infant Flow
care professional should be in attendance at all times to react to an alarm or other indications of a problem.
Always have an alternate means of ventilation available whenever the Infant
Flow
Do not attach the Generator to the patient until User Verification and initial set
up into NCPAP mode is complete.
Water in the air supply can cause malfunction of this equipment.
The operator should not touch the electrical connectors of the Infant Flow®
SIPAP™ or its accessories, and the patient simultaneously.
An audible alarm indicates an anomalous condition and should never go
unheeded.
®
®
SiPAP™ is in use.
SiPAP™ is connected to a patient, a trained health
675–101–101 Rev. K
8 Chapter 3 - Summary of Warnings and Cautions
Anti-static or electrically conductive hoses or tubing should not be used within the patient circuit.
If a mechanical or electrical problem is recognized while operating the Infant
®
Flow
SiPAP™, it must be removed from use and referred to qualified service personnel for servicing. Using inoperative equipment may result in patient injury.
Prior to patient application, ensure that all User Verification testing and calibration procedures are successfully completed. User Verification testing and calibration procedures must be done off patient.
The
indicates a connection between the Transducer Assembly and the driver. It does not indicate attachment or correct positioning of the Abdominal Respiratory Sensor.
Under certain conditions (minimum supply pressure and maximum gas demand, including auxiliary output) output flow rates and therefore pressure delivered to the generator may be reduced.
The Pres High
flow meter must be adjusted to zero when not required for the
patient.
Whenever a patient is attached to respiratory care equipment, constant attendance is required by qualified personnel. The use of an alarm or monitoring system does not give absolute assurance of warning for every malfunction that may occur in the system. In addition, some alarm conditions may require immediate attention.
Nasal CPAP treatment in general can cause nasal irritation, septal distortion, skin irritation and pressure necrosis. Adherence to the recommended usage instructions for the Infant Flow
®
SiPAP™ and AirLifeTM Infant nCPAP System
accessories may reduce the incidence of these complications.
It is strongly recommended that regular monitoring for gastric distention be carried out for patients receiving non-invasive ventilatory support. Refer to your facility’s policy and procedure for further guidance.
This device exhausts O
during normal operation. Oxygen vigorously
2
accelerates combustion. To avoid fire hazard, do not place flammable materials or sources of heat close to the exhaust.
The Abdominal Respiratory Sensor is used only to enable features associated with certain modes from the Infant Flow
®
SiPAP™. When using the Abdominal Respiratory Sensor, always use an additional, external device for monitoring of the respiratory rate and detection of apneic episodes as well as an appropriate monitor for continuous SaO
®
If the Infant Flow
SiPAP™ driver is shelf mounted, ensure that the driver is
monitoring.
2
stable and that all circuit tubing, hoses and cables are restrained to avoid hazard of toppling.
Check that the water trap is empty before use and empty it frequently during use.
Do not block or restrict the exhaust port located on the instrument back panel. Equipment malfunction may result.
675–101–101 Rev. K
Infant Flow® SiPAP™ 9
Do not use the equipment without the expiratory tubing connected to the
generator.
Only use the supplied AC cable to connect to the power supply.
The Transducer LED indicator on the front panel of the driver only signifies
connection to the driver. It does not indicate connection to or proper positioning of the Abdominal Respiratory Sensor.
Do not overload the pole and stand.
Oxygen vigorously accelerates combustion. To avoid explosion hazard, do
not use any instrument or other equipment that may have been exposed to oil or grease contamination.
When a low gas supply alarm occurs, the oxygen concentration delivered to
the patient will differ from that set on the %O
control.
2
A source gas failure will change the FiO2 and may result in patient injury.
The functioning of this equipment may be adversely affected by the operation
of other equipment nearby, such as high frequency surgical (diathermy) equipment, defibrillators, short-wave therapy equipment, “walkie-talkies”, or cellular phones.
Due to possible explosion hazard, the Infant Flow
®
SiPAP™ should not be
used in the presence of flammable anesthetics.
Electric shock hazard – Do not remove any of the Infant Flow
®
SiPAP™ covers or panels. Refer all servicing to an authorized CareFusion service technician or factory trained technician (see Service Manual P/N 675-120).
A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation. Upon loss of protective earth ground, all conductive parts including knobs and controls that may appear to be insulated can render an electrical shock. To avoid electrical shock, plug the power cord into a properly wired receptacle, use only the power cord supplied with the ventilator, and make sure the power cord is in good condition.
The Infant Flow
®
SiPAP™ is designed to ensure that the user and patient are not exposed to excessive leakage current per applicable standards. However, this cannot be guaranteed when external devices are attached to the driver. In order to prevent the risk of excessive enclosure leakage current from external equipment attached to the driver, isolation of the protective earth paths must be provided to ensure proper connection. This isolation should ensure that the cable shields are isolated at the peripheral end of the cable.
When the Infant Flow internal oxygen monitor is disabled, the Infant Flow
®
SiPAPTM unit is connected to a patient, and the
®
SiPAPTM unit must be
used with an external oxygen monitor.
675–101–101 Rev. K
10 Chapter 3 - Summary of Warnings and Cautions

Cautions

Before use, verify that this equipment has been authorized for use by qualified technical service personnel.
Ensure that the voltage and installed fuses are set to match the voltage of the wall outlet, or damage may result.
A battery that is fully drained (i.e. void of any charge) may cause damage to the driver and should be replaced.
All accessory equipment that is connected to the driver should comply with CSA/IEC601/ETL.
Although failure of any of the above tests will not prevent the ventilator from functioning, it should be checked to make sure it is operating correctly before use on a patient.
The Infant Flow CareFusion accessories. Only accessories approved for use by CareFusion should be used. If in doubt, please contact your local sales representative.
®
SiPAP™ has been designed and tested using only

Notes

Employ safe lifting procedures when assembling the unit.
Do not sterilize the driver. The internal components are not compatible with
sterilization techniques.
Do not submerge the driver or pour cleaning liquids over or into the driver.
Following each alarm verification test, ensure that control settings and alarm
limits are reset as instructed before proceeding to the next test.
CareFusion cannot ensure product performance as stated in this manual with the use of Non-CareFusion accessories.
675–101–101 Rev. K
Infant Flow® SiPAP™ 11

Chapter 4 - Unpacking & Setup

Assembly and physical setup

Optional
Accessory
Figure 1 – Stand unpacking and assembly
675–101–101 Rev. K
12 Chapter 4 - Unpacking & Setup
Infant Flow
SiPAP Driver
Optional
Accessory
Figure 2 – Stand and Driver assembly
675–101–101 Rev. K
Infant Flow® SiPAP™ 13

Attaching a patient circuit

Infant Flow
SiPAP Driver
Optional
Accessory
Figure 3 – Driver assembled with patient circuit and /humidifier
Note
We recommend between 96.8 °F (36 °C) and 98.6 °F (37 °C) but never higher than
98.6 °F (37 °C) for inspired gases.
675–101–101 Rev. K
14 Chapter 4 - Unpacking & Setup
CAUTION
Back pressure from some auto-feed humidifier chambers may cause the water bags to fill with air. Ensure that the humidifier chambers are adequately filled according to the manufacturer’s instructions.
Note
When the Hudson RCI Humidification System is being used with Infant Flow® SiPAP
TM
, it is recommended that the standard compliance column be used.

Attaching the Abdominal Respiratory Sensor

Figure 4 – Attaching the Abdominal Respiratory Sensor
1. Connect the Transducer Assembly to the front panel of the driver (Fig. 3)
2. Connect Abdominal Respiratory Sensor to the Transducer interface.
3. Apply gentle compression to sensor. Verify function with illumination of LED on transducer interface.
4. Apply sensor with suitable tape (Fig 4).
a. Pressure line perpendicular to tape
b. Sensor between umbilicus and xiphisternum
c. Placement on the side of the abdomen may be necessary
5. Verify correct placement
a. Observe spontaneous breathing
b. Transducer LED illuminates on expiration; Front panel Transducer LED
illuminates on inspiration
675–101–101 Rev. K
Infant Flow® SiPAP™ 15

Flow / Pressure Relationship

The Infant Flow® SiPAP™ is subject to a direct relationship between the controlled enriched gas flow and airway pressure. A nomogram illustrating the relationship between constant airway pressure and flow settings is shown in Figure 5. For example, 8 L/min gas flow provides approximately 5 cmH
2
O.
Note
Individual devices have a tolerance of up to ± 10% from that illustrated in the nomogram and in particular, at pressures below 2 cmH
2
O.
Flow Pressure Nomogram (for reference only) to show typical flow pressure relationships. This is not meant to establish actual product performance.
Figure 5 – Flow Pressure Nomogram
675–101–101 Rev. K
16 Chapter 4 - Unpacking & Setup

User Verification Test

WARNING
Do not attach Generator to the patient until User Verification and initial set up into NCPAP mode is complete.
CAUTION
Although failure of any of the above tests will not prevent the ventilator from functioning, it should be checked to make sure it is operating correctly before use on a patient.

Power-on Check

This test is run automatically on power up of the driver and automatically performs the following checks:
Flash ROM
Hardware Inputs/Outputs
Audible and visual alarms indicators
Test and calibration of pressure sensor
Test of dump valve
The unit carries out a full functional check during this time. If unsuccessful, the screen remains darkened and the warning bar remains on. In this case, check for the following;
Power Supply not connected
Battery voltage low
If the checks are successful, the screen changes to Power Up Screen. After two seconds, the screen changes to Power Up Check Screen.
During the Power Up check:
Screen image shown in negative
Warning bar comes on for one second
Transducer Assembly LED comes on for one second
Audible alarm sounds for one second
Dump valve is tested
Pressure is set to zero
675–101–101 Rev. K
Infant Flow® SiPAP™ 17
After two seconds screen changes to Set Up Screen. Alarm limits are disabled and a flashing question mark appears under the NCPAP / Pres Low flow meter screen indicator.

Two Point O2 Sensor Calibration

Enter the Calibration Screen from the Set Up Screen by pressing the calibration button on the lower right hand corner of the touch screen. Ensure there is a minimum 8 LPM set on the NCPAP/PRES Low Flow meter. In addition, ensure there is a minimum of 3 LPM set on the NCPAP/High Flow meter. Adjust the %O 21%. Allow the %O associated flashing button.
display to stabilize. Confirm the calibration by touching the
2
control to
2
Adjust the %O touching the associated flashing button. Return to the Start up Screen by pressing the Exit button.
control to 100%. Allow the %O2 display to stabilize. Confirm by
2
Note
If O2 calibration fails, a red “X” is shown. Refer to the Service Manual.
If the internal oxygen cell is depleted or damaged, it may not be possible to calibrate the O button. This will disable oxygen monitoring and the audible oxygen alarms until the device is powered off. Whenever the device is operating with oxygen monitor and alarms disabled, a fault code E5x displays, and measured FiO
sensor. The internal oxygen monitor may be disabled using the Disable O2
2
displays as dashes.
2
WARNING!
When the Infant Flow® SiPAPTM unit is connected to a patient, and the internal oxygen monitor is disabled, the Infant Flow® SiPAPTM unit must be used with an external oxygen monitor.
If calibration is attempted, and fails, or if the oxygen cell fails while the device is in normal use, a Fault Code E5X displays, as tabulated in appendix D, and a high priority alarm is indicated visually and audibly. To enable continued operation, the internal oxygen monitoring may be disabled by pressing and holding the alarm mute / reset button for 3 seconds. This disables the internal oxygen monitor and alarms and clears the alarm condition. The E5X code remains to indicate that the oxygen monitor is inoperative. An external oxygen monitor must be used.
675–101–101 Rev. K
18 Chapter 4 - Unpacking & Setup

Leak Test

1. Have the patient circuit and generator assembled as shown in Fig 3.
2. Connect the patient interface (prong or mask) to the generator (see Chapter 5, Step by Step Fixation) and occlude the opening to the patient.
3. If not powered up already, switch on the power to the driver.
4. Adjust the NCPAP / Pres Low flow meter to 8 L/min. Verify that the measured pressure is 5 ± 1 cmH
5. Adjust %O2 control as prescribed for the current patient. Verify that the blender setting, and the measured oxygen value, are within 3%. Touch the associated flashing screen icon to confirm.
6. Adjust the Pres High flow meter as prescribed for the current patient. Touch the associated flashing screen icon to confirm.
7. Connect the Transducer Interface to the front panel of the driver if breath monitoring is desired in treatment. Touch the associated flashing screen icon to confirm.
8. The display screen changes to the Alarm Set/Confirm Screen. Press the NCPAP button or Alarm Mute/Reset button to set alarms and begin monitoring.
9. Monitored parameter for CPAP should be 4-5 cmH leaks or blockages, (including the humidification system).
10. Remove the occlusion to the patient interface. The monitored CPAP display should be 0-2 cmH
O. Touch the associated flashing screen icon to confirm.
2
O. If not, check circuit for
2
O. If not, check that the interface is not still occluded.
2
675–101–101 Rev. K
Infant Flow® SiPAP™ 19
Alarm Test Initial Settings
2
nCPAP System Patient Circuit
Pres High flow meter
For Step 9 Use the settings provided below

Alarms Test

WARNING
Prior to patient application, ensure that all User Verification testing and calibration procedures are successfully completed. User Verification testing and calibration procedures must be done off patient.
NOTE
Following each alarm verification test, ensure that control settings and alarm limits are reset as instructed before proceeding to the next test.
Air Supply Pressure
O
Supply Pressure
Patient Circuit
Generator
NCPAP / Pres Low flow meter
% O2
Mode
Rate
T-High
Tapnea (Non-U.S. Configuration) / T
(U.S. Configuration)
LBR
> 30 psig (2.1 bar)
> 30 psig (2.1 bar)
Infant Flow® or AirLifeTM Infant
Infant Flow® or AirLifeTM Infant nCPAP System Generator
8 L/min (for delivery of 5 cmH2O)
30%
3 L/min
NCPAP
30 bpm
0.3 sec
20 sec
675–101–101 Rev. K
20 Chapter 4 - Unpacking & Setup
Perform the Alarms Test on the Infant Flow® SiPAP™ using the following steps and the initial settings provided above.
1. Make appropriate connections for air and O to appropriate AC outlet. Attach patient circuit, generator and patient interface (mask or prong) as shown in Figure 3. Occlude the opening to the patient.
2. Power up the driver and allow Power On Check to complete.
3. Low airway pressure alarm: From NCPAP operating mode, with alarms set, remove occlusion from opening to patient. Verify that the low airway pressure alarm activates. Restore the patient interface occlusion and press the Alarm Mute / Silence button for 3 seconds to reset the alarms.
4. High airway pressure alarm: Adjust the NCPAP / Pres Low flow meter to 11 L/min. Verify that the high airway pressure alarm activates. Return the NCPAP / Pres Low flow meter to 8 L/min and press the Alarm Mute / Silence button for 3 seconds to reset the alarms.
5. High %O
Alarm: Adjust the % O2 control to 35%. Verify that the High %O
2
alarm activates. Return the O2 control setting to 30%. Reset alarms by pressing the Alarm Mute / Reset button for 3 seconds.
6. Low % O
Alarm: Adjust the % O2 control to 25%. Verify that the Low %O
2
alarm activates. Return the O2 control setting to 30%. Reset alarms by pressing the Alarm Mute / Reset button for 3 seconds.
7. Loss AC Alarm: Disconnect the AC power cord from the wall outlet. Verify that the Loss AC alarm activates. Reconnect the AC power cord. Clear the alarm by pressing the Alarm Mute / Reset button.
8. High Circuit Pressure Alarm: Increase nCPAP pressure to 11.1 cmH increasing the NCPAP/PRES Low Flow meter. Verify that the High Circuit pressure alarm activates. Return NCPAP/PRES Low Flow meter to 8 LPM and press the Alarm Mute/Silence button for three seconds to reset the alarms.
9. Low Breath Rate (Apnea) Alarm: Select and confirm BiPhasic+Apnea/LBR (U.S. configuration). Using the abdominal sensor, manually tap the abdominal sensor to simulate a spontaneous breath rate. The default mandatory breath rate should be left alone. No alarms should be present. Change the mandatory rate control setting to 1 bpm and stop tapping the abdominal sensor. Verify that the Low Breath rate alarm activates after the default interval of 20 seconds. Resume simulating spontaneous breath rate, turn the rate control to the default setting and clear the alarm by pressing the Alarm Mute/Reset button for 3 seconds. Note: a transducer must be connected to perform the alarm check.
gas supply. Connect power cord
2
2
2
O by
2
675–101–101 Rev. K
Infant Flow® SiPAP™ 21
Two Point O2 Sensor Calibration
Patient Circuit Leak test
2
Low O
2
Alarm

Infant Flow® SiPAP™ User Verification Test Checklist

Driver Serial Number:_____________________ Test Date:_________________
TEST PASS FAIL
Automated Tests
Power On Check
Manual Tests
Manual Alarms Checks
Low Airway Pressure Alarm
High Airway Pressure Alarm
High O
Loss AC Alarm
High Circuit Pressure Alarm
Low Breath Rate (Apnea) Alarm
Alarm
Signature of tester:_______________________________________________
Title___________________________________________________________
675–101–101 Rev. K
22 Chapter 4 - Unpacking & Setup
Infant Flow®
SiPAP™
675–101–101 Rev. K
Infant Flow® SiPAP™ 23

Chapter 5 - Operation

Front Panel Indicators and Controls

The front panel consists of a LCD touch screen display with key pad, separate flow meter controls for adjustment of NCPAP /Pres Low and Pres High and a %O blender control. Patient circuit connections are located along the bottom panel. LEDs along the top of the front panel indicate power on, connection to wall AC, active alarms and Transducer Interface connection to the driver. An ambient light sensor is located under the front panel to adjust the backlight of the screen display in high and low light environments
2
Figure 6 – Front Panel
675–101–101 Rev. K
24 Chapter 5 - Operation

Rear Panel

Figure 7 – Rear Panel
675–101–101 Rev. K
Infant Flow® SiPAP™ 25
Table 2 - Soft-key operation
Description Example
A button which is enabled.
A button which is inhibited due to non-availability of the designated feature or pending acknowledgement of an active alarm condition.
A selected mode or control pending confirmation is visually highlighted and intermittently flashes between yellow and white text.
While a button is pressed the edges are highlighted to provide a pressed appearance.
When there is an active alarm associated with a measured value the measured value concerned is displayed with RED FLASHING text. The associated limit value (if any) is displayed in RED.
When an alarm that is associated with a measured value is resolved, the device remains in a LOW priority alarm state, with the measured value displayed in YELLOW FLASHING text and the associated limit displayed in YELLOW, until the alarms are cleared by the operator.
When parameter adjustments cause a reduction in another parameter to maintain requirements for minimum breath interval, the reduced parameter is displayed in RED for 15 seconds

Changing a control

When a control such as Time-High is selected, increase and decrease buttons appear. The control and the displayed value for the selected parameter are highlighted. Use the decrease or increase keys to adjust the parameter. You may accept the action by pressing the control button again. If no action is taken, the new parameter will take effect after 15 seconds.
In normal treatment screens, parameter changes take immediate effect.
When you change the mode, such as NCPAP to BiPhasic, press the flashing mode button to accept the change.
If no screen interactions occur for a period of 120 seconds and there are not active alarms, then the screen goes to a ‘locked’ state to prevent inadvertent entries. To unlock the screen, press the screen lock button. In the case of a high priority alarm, the screen immediately unlocks to allow access to controls.
CAUTION!
When changing a control, use only your finger. Damage to the touch screen could result if a pen or similar item is used to make changes.
675–101–101 Rev. K
26 Chapter 5 - Operation

Increase / Decrease buttons

Pressing the ‘increase’ or ‘decrease’ buttons causes a currently-selected control to be changed to the next valid greater or lesser value. Each press of the increase or decrease button is accompanied by an audible click. If the control limit is reached an audible beep sounds to alert the operator.
Displays of calculated values (such as I:E ratio) dependent on a control setting change will change with acceptance of the parameter change.

Incompatible Control Settings

When a change to one control requires a change to a separate control to avoid an incompatible control setting, the required change is made automatically by the driver software.
If the adjusted control setting is restored prior to 15 seconds elapsed time or prior to pressing any other control, then the required change action is reversed.
For example in BiPhasic mode with T-High = 2.0, as R is increased above 28 b/min the constraint on minimum T-Low can be met only through a reduction in T-High. If R is increased to 29, then T-High shall reduce automatically to 1.9s. If R is then
immediately reduced to 28, the previous setting for T-High shall be restored.

Parameter default value on change of mode

Some controls are active in more than one operating mode. In these instances, there is a separate default value for operating modes as illustrated with the following table. Settings that are changed by the operator in one specific mode will be maintained if the mode is changed to another mode within the same mode group. All defaults shall revert to factory default on power-cycling or software restart.

Setting a Manual Breath

The manual-breath function is available in CPAP, Biphasic, and Biphasic tr modes. For manual breath to be active when the manual button is selected, the pre-use pressure-high check has been completed and the pressure high-flow meter is set for preferred manual breath. One manual breath is delivered per button-press.
Note:
The Pressure High Flow Meter must have flow above zero in order to deliver manual breaths.
675–101–101 Rev. K
Infant Flow® SiPAP™ 27
apnea
LBR
Table 3 – Parameter Default Value
Mode
Parameter
T-High Default for NCPAP apply Default for BiPhasic apply Default for
Rate Default for BiPhasic apply
Rb Default for
T
/ T
NCPAP NCPAP + rate
monitoring
This setting applies to all modes: system-wide default applies to all modes
BiPhasic BiPhasic + rate
monitoring
BiPhasic tr
BiPhasic tr
BiPhasic tr

User Interface Display

Screen Displays

1. Set Up Screen – The Set Up screen prompts the user to confirm settings for base line pressure level (NCPAP / Pres Low), %O High) controls and confirmation of connection of the Transducer Interface (XDCR) to the driver.
, upper level pressure (Pres
2
Figure 8 – Set Up Screen
Adjust the NCPAP / Pres Low flow meter counter clockwise to increase the control to the required flow rate and clockwise to decrease the flow. Touch the associated flashing icon to confirm. The icon changes to a check mark and the next button flashes.
Set the %O
as prescribed. Touch the associated flashing button to confirm. The
2
icon changes to a check mark and the next button flashes.
Adjust the Pres High flow meter as desired for delivery of BiPhasic, BiPhasic tr or manual breaths using counterclockwise turns to increase and clockwise turns to decrease the flow. Touch the associated flashing button to confirm. The icon changes to a check mark and the next button flashes.
675–101–101 Rev. K
28 Chapter 5 - Operation
If breath rate monitoring is desired, attach the Transducer Interface and the abdominal sensor. Refer to Chapter 4 for instructions on application of the Abdominal Respiratory Sensor. Touch the flashing button to confirm.
If an alarm is activated as a result of any of the settings, the button displays a flashing “X”. Alarm conditions must be cleared and all settings must be confirmed with a green check mark before other screens can be entered.
2. Alarm set/confirm Screen
Figure 9 – Alarm set/confirm Screen
Touch the NCPAP button or the Alarm Mute / Reset button for 3 seconds to set the alarms and to move to the next screen. If either button is not touched within 2 minutes, the alarm limits will be set automatically. When the alarm limits have been set, the screen display changes to the Mode Select Screen with the driver operating in NCPAP mode.
Note:
Press High Flow meter must be checked through the Start-up screen and be set during operation. This enables manual breath or back-up apnea breath, where applicable, to be active.
675–101–101 Rev. K
Infant Flow® SiPAP™ 29
Figure 10 – Mode Select Screen
3. Mode Select Screen - Here the operator can select the desired mode of
operation. Once selected, the operator has the ability to adjust the screen controls for the mode selected. Only the relevant controls available for the selected mode are visible.
To make a change to a control, touch the control. Both the control and its
associated numeric display highlight and the adjust buttons appear. Press up or down buttons to adjust the setting as desired. Confirm the change by pressing the control again.
4. Parameter Adjust Screen – During normal operation, active controls for the current operating mode can be adjusted by touching the control, using the increase or decrease arrows to make the adjustments and then pressing the control again to confirm the change.
Figure 11 – Parameter Adjust Screens
5. Main screen– The Main Screen provides the operator with displays of current
mode of operation, alarm status, battery charge status, monitored parameters and a pressure time graphic display. Active controls are available for adjustment in this screen.
675–101–101 Rev. K
30 Chapter 5 - Operation
Figure 12 – Main Screen
6. Monitored Parameter Screen – This screen is accessed by pressing the
change screen button. The monitored parameters screen displays measured values and control settings. Adjustments to controls active for the currently selected mode are possible from this screen. To return to the main screen press the screens button again.
Figure 13 – Monitored Parameters Screen
675–101–101 Rev. K
Infant Flow® SiPAP™ 31

Step by Step Fixation for Infant Flow System Generators

Following this fixation technique closely helps to ensure:
Enhanced stability of the Generator
Minimal disturbance to the infant
WARNING
Do not attach Generator to the patient until User Verification and initial set up into NCPAP mode is complete.
Patient Interfaces from CareFusion accommodate a wide range of patients. Application of an incorrectly sized prong, mask or bonnet will affect stability of the generator. The clinician may consider alternating the use of prong and mask interfaces at set intervals for a single patient in order to change pressure points on the infant’s face and reduce the risk of skin breakdown.
1. Measure for prong/mask size using the nose guide. Connect the interface to the generator.
2. Measure for bonnet size from the middle of the forehead to the nape of the neck and then back to the middle forehead. DO NOT use a “head
circumference” measurement to determine bonnet size.
3. Loosely weave Generator straps through the buttonholes. Begin from the inside of the colour coded buttonhole. Place the Generator on top of the bonnet above the central Velcro strip.
4. Place the bonnet onto the infant’s head, checking that the ears are in a normal position. Ensure the bonnet is pulled well down over the ears and down to the nape of the neck.
Switch on the power to the driver and complete Set Up Screen steps to enter
NCPAP mode with the prescribed settings for the current patient.
5. Lift the Generator from the top of the bonnet and bring towards the nose. Gently insert the nasal prongs/mask into position while supporting the Generator. Secure the generator straps horizontally across the infant’s cheeks.
Do not over tighten.
6. Secure all three tubes from the Generator with the central Velcro strip. Split the inspiratory and pressure lines and secure with secondary Velcro strips. Tie the open end of the bonnet if desired.
7. Final check:
Nose in neutral position; eyes visible; ears not folded
Desired upper and lower pressure levels and FiO
Infant settles quickly after fixation
675–101–101 Rev. K
are delivered
2
32 Chapter 5 - Operation

Every hour

Repeat checks listed above in Final Check.

Every 3-4 hours

Loosen the generator straps and release the tubes from the central Velcro strip. The nasal area can be cleaned with warm sterile water. Do not apply creams or ointments.
Ensure that:
Nasal prongs/mask is not occluded with mucus/water droplets
Patient prongs/mask and bonnet continue to fit appropriately.
Re-apply the generator as described above.

Directions for using the AirLife™ Infant nCPAP System

Please refer to P/N 36-5569 included in the AirLifeTM Infant nCPAP Fixation Device for the Directions for Use.
WARNING!
CareFusion consumables are specifically designed to be used with Infant Flow® Drivers and are the only consumables validated for use with Infant Flow® devices.
675–101–101 Rev. K
Infant Flow® SiPAP™ 33

Chapter 6 - Operating Modes

NCPAP

The Nasal CPAP mode can be enabled to have breath rate monitoring displayed (NCPAP +Apnea Mode, or NCAP +LBR Mode), or the system can operate without having the breath rate monitoring displayed (NCPAP Mode). Breath rate monitoring requires the use of the Transducer Assembly (part number 677-002) and the Abdominal Respiratory Sensor (part number 467349).
Figure 14 – NCPAP

BiPhasic

Allows for time triggered pressure assists, with or without breath rate monitoring and adjustable low breath rate alarm, delivered based on clinician set Time High (T-High) criteria, rate and pressure settings.
Figure 15 – BiPhasic
675–101–101 Rev. K
34 Chapter 6 - Operating Modes

BiPhasic tr*

Allows for patient triggered pressure assists with breath rate monitoring enabled, adjustable apnea time interval, apnea alarm and adjustable apnea back up rate. The upper level pressure is delivered based on operator set Time High (T-High) and pressure settings.
Figure 16 – BiPhasic tr
* This mode available in Comprehensive configurations only.
675–101–101 Rev. K
Infant Flow® SiPAP™ 35

Chapter 7 - Alarms and Indicators

Audible and visual indications are given to alert the Operator to specified conditions that affect operation. The electronic alarm limits are automatically set after two minutes without the necessity of Operator inputs.
Alarms can be manually set at any time if required (i.e. after settings change or patient disconnect/ reconnect) by pressing the Alarm Mute / Reset button for 3 seconds.

Alarm Priority

On activation of a MEDIUM or HIGH priority alarm, a locked screen will automatically revert to the unlocked display. The alarm status indicator flashes intermittently, based on the current, highest alarm priority. Distinct audible alarms represent a HIGH, MEDIUM or LOW priority alarm.
Measured parameters and alarm limits associated with a high or medium priority alarm condition flash RED (HIGH priority) or YELLOW (MEDIUM priority) to denote alarm condition and priority.

Silencing audible alarms

Pressing the Alarm Mute / Reset button will silence active alarms for up to 30 seconds. If a new high priority alarm condition occurs during the alarm silence time period, the silence will be cancelled to alert the operator of the new alarm condition.

Resetting alarms

Press the Alarm Mute / Reset button for 3 seconds to clear resolved and LOW Priority alarms and to reset alarm limits (i.e. after a control setting change). Where the alarm cause remains, the appropriate alarm will immediately reoccur.
Audible alarm priority
High Priority – a series of 10 tones (2 groups of 3 tones followed by a pause and 2
more tones) every 10 seconds
Medium Priority – three audible tones every 15 seconds
Low Priority – two audible tones every 30 seconds
675–101–101 Rev. K
36 Chapter 7 - Alarms and Indicators

Alarm Types

The following alarm systems are provided with the Infant Flow® SiPAP™. Electronic alarms are set after 2 minutes of operation without operator intervention although the operator can manually set or reset them if required. Refer to Appendix C for information on troubleshooting alarms.

Supply Gases Failure

If the differential pressure between the two inlet gases falls outside of the limit of 30 PSI (2.0 bar) or one gas fails completely, an alarm will sound and the gas at the higher pressure only will be delivered to the patient.

High Airway Pressure

A HIGH priority audible and visual high pressure alarm activates when pressure reaches 3 cmH

Airway Over-Pressure Limit Alarm

A HIGH priority audible and visual high pressure alarm activates at 11 cmH2O during NCAP and time triggered BiPhasic modes and 15 cmH activation of this alarm, the patient circuit pressure drops to near zero. Pressure is restored after 3 seconds, and reduces to near zero again should the condition causing the alarm remain.
O above the measured airway pressure.
2
O in BiPhasic tr mode. Upon
2

Low Airway Pressure

A HIGH priority audible and visual low pressure alarm activates if pressures fall to 2 cmH otherwise be less than zero.
O below the measured airway pressure or at 1.5 cmH2O, if this would
2

High and Low % O2

HIGH priority audible and visual alarms are provided at ± 5% of the measured FiO2 with an upper maximum limit of 104% and a lower minimum limit of 18%.
A low hazard warning occurs at 18% Oxygen or below.

Low Battery Charge

If the battery charge falls below 40% the battery charge indicator changes from green to red as a warning indicator. In this instance, plug the driver into an approved AC power source.

Low Battery Voltage

If the battery voltage falls to < 11.10 V for 5 seconds a MEDIUM priority audible and visual alarm is activated. In this instance, plug the driver into an approved AC power source.
675–101–101 Rev. K
Infant Flow® SiPAP™ 37

Flat Battery

If the battery charge is too low to reliably power the analogue and valve driver circuits, the unit enters a safe ‘flat battery’ screen, until it is either switched off, or plugged into a suitable external power source. The screen display will go completely blank when the battery charge is too low to power it. While sufficient power is available, audible and visual indication of the high-priority alarm is maintained.
Figure 17 - Flat Battery screen

Operation without Electrical Power (No AC or DC power)

The Infant Flow® SiPAP™ will continue to deliver NCPAP flow only as set from the NCPAP / Pres Low flow meter and the set %O DC power. In this mode, visual indications and audible alarm warnings are not given except for the supply gases failure alarm.
in the event of a total loss of AC and
2

Error code indication

When a unit error is active, and this does not cause complete device failure, then a non-mutable HIGH priority alarm is activated with the error code displayed in flashing RED text in the upper right hand of the display screen, alternating with any currently displayed mode information. Refer to Appendix D for a listing of error codes. Remove driver from service and refer to a qualified service technician.
675–101–101 Rev. K
38 Chapter 7 - Alarms and Indicators

Alarm Symbols and Indicators

The following displays are shown within the graphical display. As needed displays in this table are shown separately as Domestic US configuration displays (left-hand column) and non-US configuration displays (right-hand column).
Table 4 – Alarm Symbols and Indicators
Indicator Meaning
Battery status/charge level; Green indicates full charge; red indicates charge < 40%)
External power source not connected
Flat battery
Battery fault (battery unable to hold charge) or supply fault
Respiratory transducer interface has become detached during breath monitoring
Indication during pre-use checks that the NCPAP / Pres Low flow meter should be set as desired and pressure verified
Indication during pre-use checks that the Pres High flow meter should be set as desired and pressure verified
Indication during pre-use checks and/or calibration that % O2 should be set and verified.
675–101–101 Rev. K
Infant Flow® SiPAP™ 39
Indicator Meaning
Indication during pre-use checks that operator should attach the respiratory sensor (cross indicates that transducer assembly is not connected)
Indication during pre-use checks that the operator should attach the respiratory sensor (indicates that transducer assembly is connected)
Does not verify attachment of sensor to patient
Refer to manual
Power has failed; re-connect external power source
Display of % O2 measured value and associated alarms
Display of NCPAP airway pressure measured value and alarms (NCPAP modes only)
Display of mean airway pressure measured value and alarms (BiPhasic and BiPhasic tr* modes only)
I:E ratio
Spontaneous breathing rate
Set parameter T-High
Set parameter Rate (BiPhasic modes only; mandatory rate)
Set parameter RB (BiPhasic tr* mode only; backup rate)
Set parameter low breath rate or apnea alarm timeout
675–101–101 Rev. K
40 Chapter 7 - Alarms and Indicators
Indicator Meaning
Display of measured airway pressure graph. With breath monitoring active, spontaneous breaths are indicated in yellow, below delivered airway pressure graph.
Device fault (fault code will be indicated). Refer to manual. Contact qualified service technician.
Note
Provision of labeling in this manual for any function should not be taken as evidence that the function is available. For example parameter RB relates to BiPhasic tr* mode, not currently approved for use in the US.
675–101–101 Rev. K
Infant Flow® SiPAP™ 41

Chapter 8 - Maintenance & Cleaning

Cleaning

Examine the exterior of the case and the stand for damage and dirt. If necessary clean the unit and stand. If damage to either is apparent, always seek qualified Technical advice.
Clean the exterior surfaces of the driver, Transducer Assembly and stand with a mild soap or liquid disinfectant solution. Do not use cleaning agents that contain abrasives. Make sure that cleaning agents do not enter the driver through patient connections or other ports.
CAUTION
Do not immerse any part of the Infant Flow® SiPAP™ driver in water or sterilize it with gas or steam.

Maintenance

No special maintenance is required by the operator other than that listed below. There are no operator serviceable parts. The unit must only be maintained and serviced by an approved service supplier or trained biomedical engineer. Only parts approved by CareFusion may be used in this unit. Refer to the Service Manual or your Service Supplier for an approved service parts list
WARNING
Oxygen vigorously accelerates combustion. To avoid explosion hazard, do not use any instrument or other equipment that may have been exposed to oil or grease contamination.
Calibrate the oxygen analyzer regularly. Calibration of the oxygen analyzer must be done with the unit off patient.
Regularly check and empty the water trap accessed from the rear panel of the driver enclosure. Push the button on the bottom of the water trap to release the water into a suitable waste receptacle.
Disconnect the air and oxygen gas sources when the Infant Flow use.
®
SiPAP™ is not in
675–101–101 Rev. K
42 Chapter 8 - Maintenance & Cleaning
Storage and Battery Care
Store the unit in a clean dry location. Make sure that all connections and ports are suitably covered to prevent the ingress of dirt, moisture and foreign objects. If the unit is not being used for a long period of time, remove the battery (refer to the Service Manual or your Service Technician).
Dispose of scrap units in accordance with the local regulations. Refer to the Service Manual or your Service Supplier.
675–101–101 Rev. K
Infant Flow® SiPAP™ 43
IEC 60878
IEC 30878
IEC 60878

Chapter 9 – Explanation of Symbols

Equipment Symbols

The following symbols may be referenced on the Infant Flow® SiPAP™ driver or in accompanying documentation.
Table 5 – Equipment Symbols
Symbol
Symbol #5031 IEC
Source / Compliance
Symbol #03-02 IEC 60878
Symbol #5016 IEC 60417
60417
Symbol #5019 IEC 60417
Symbol #01-20
Meaning
Indicates ATTENTION, consult ACCOMPANYING DOCUMENTS
This symbol indicates a FUSE.
This symbol indicates DIRECT CURRENT (DC)
This symbol indicates protective EARTH (ground).
Symbol #5021 IEC 60417
Symbol # 01-24 IEC 60878
Symbol # 5333 IEC 60417
Symbol #02-03 IEC 60878
Symbol #5032 IEC 60417
Symbol #01-14
Symbol #5007 IEC 60417 Symbol #01-01 IEC 60878
Symbol #5008 IEC 60417 Symbol #01-02
MDD Directive 93/42/EEC
This symbol indicates the EQUIPOTENTIAL connection used to connect various parts of the equipment or of a system to the same potential, not necessarily being the earth (ground) potential (e.g., for local bonding).
This symbol indicates TYPE B equipment, which indicates equipment that provides a particular degree of protection against electric shock, particularly with regards to allowable leakage current and reliability of the protective earth connection.
This symbol is located on the rating plate. It indicates the equipment is suitable for alternating current.
Indicates ON (Power)
Indicates OFF (Power)
CE Mark
CareFusion Symbol
675–101–101 Rev. K
This symbol indicates an INTERNAL BATTERY FUSE
44 Chapter 9 – Explanation of Symbols
ISO 15223:2000
Symbol
2.5A/T 250 V 101010
XDCR
Source / Compliance
ISO 7000:2004 (2616)
ISO 15223:2000 (3.11)
EN 980:2003 (5.7.3)
N/A
CareFusion Symbol
IEC 60878:1988 (01-41)
ISO 7000:2004 (2301)
IEC 60878:1988 (02-03)
Meaning
Electrical AC inlet
Operating temperature range of unit
Fuse holder and fuse rating
Transducer Assembly
Warning Bell
Type BF patient applied part
ETL Mark and Registration Number
Intertek Group
ISO 15223: 2000 (3.13)
EN 980:2003 (4.4)
IEC 60878 (03-02)
ISO 15223:2000 (3.1.4)
EN 980:2003 (4.4)
(3.12)
EN 980:2003 (4.3)
ISO 15223:2000 (3.2)
EN 980:2003 (4.2)
Year of Manufacture
Read Accompanying Documents
Unique Batch Number Identifier
Use Before Expiry Date shown Year-Month
Single Use Only - Do NOT Re-use
675–101–101 Rev. K
Infant Flow® SiPAP™ 45
Symbol
Source / Compliance
ISO 15223:2000 (3.8)
ISO 15223:2000 (3.8)
Meaning
Keep Dry
Keep Away from Heat

Symbols used on buttons

The following symbols are used to label user input areas within the graphical display. As needed displays in this table are shown separately as Domestic US configuration displays (left-hand column) and non-US configuration displays (right-hand column).
Table 6 – Button Symbols
Symbol Description
High Priority Alarm Active, red flashing
Medium Priority Alarm Active, yellow flashing
Low Priority Alarm Active, yellow, does not flash.
No alarms are present, green, does not flash
Active alarm silenced
Adjust BiPhasic rate
Adjust BiPhasic tr* backup rate
Adjust apnea alarm timeout
Adjust low breath rate alarm timeout
Adjust BiPhasic, BiPhasic tr* on time, and NCPAP manual breath function
675–101–101 Rev. K
46 Chapter 9 – Explanation of Symbols
Symbol Description
Decrease / Increase currently selected parameter
Go to mode select screen.
Nasal CPAP mode
Nasal CPAP mode with breath rate monitoring
BiPhasic mode
BiPhasic mode with breath rate monitoring
BiPhasic tr* mode with breath rate monitoring
Manual Breath. Single BiPhasic cycle at current
. One BiPhasic
2
settings for T-High, Pres High and % O cycle is delivered regardless of button press duration
Toggle between Main Screen and Monitored Parameter Screen
Go to user calibration screen
Confirm
Wait
Completed
Action has failed
Press to un-lock keypad
Warning message. To clear, press any of the three icons.
Oxygen monitor and alarms disable
675–101–101 Rev. K
Infant Flow® SiPAP™ 47
Note
Provision of labeling in this manual for any function should not be taken as evidence that the function is available. For example parameter R not currently approved for use in the US.
relates to BiPhasic tr* mode,
B
675–101–101 Rev. K
48 Chapter 9 – Explanation of Symbols
Infant Flow®
SiPAP™
675–101–101 Rev. K
Infant Flow® SiPAP™ 49
3
High
1
15%
1

Appendix A - Product Configurations

Non-US Configuration Parameters

Table 7 – Non-US Configuration Parameters
Parameter Min Max Accuracy Units Default
Set Oxygen concentration, %O2 21 100
NCPAP / Pres Low flow rate 0 15
Pres High flow rate 0 5
BiPhasic / BiPhasic tr* on time, T-
0.1 3.0
±
±15%
± 15%
± 0.005
% N/A
L/min N/A
L/min N/A
seconds .3 sec
BiPhasic rate, R (mandatory rate) 1 120
BiPhasic tr* backup rate, Rb (apnea backup rate)
Apnea timeout, Tapnea 10 30
1 120
± 0.5
± 0.5
±
bpm 30 bpm
bpm 10 bpm
seconds 20 sec
Note
BiPhasic tr mode not currently available in the United States. In non-US configurations, T-High automatically reduces at higher R and Rb rate settings to maintain a minimum off time of 100 milliseconds

US Configuration Parameters

Table 8 – US Configuration Parameters
Parameter Min Max Accuracy Units Default
Set Oxygen concentration, %O2 21 100
NCPAP / Pres Low flow rate 0 15
Pres High flow rate 0 5
BiPhasic on time, T-High 0.1 3.0 *
BiPhasic rate, R (Mandatory rate) 1 54
Low Breath Rate timeout, T
10 30
LBR
.
±3
±
± 15%
± 0.005
± 0.5
±
% N/A
L/min N/A
L/min N/A
seconds 1.0 sec
bpm 10 bpm
Seconds 20 sec
Note
In US configurations, T-High automatically reduces at higher R and Rb rate settings to maintain a minimum off time of 1.0 seconds.
675–101–101 Rev. K
50 Appendix A - Product Configurations
Infant Flow®
SiPAP™
675–101–101 Rev. K
Infant Flow® SiPAP™ 51

Appendix B - Pneumatic Diagram

675–101–101 Rev. K
52 Appendix B - Pneumatic Diagram
Infant Flow®
SiPAP™
675–101–101 Rev. K
Infant Flow® SiPAP™ 53

Appendix C - Alarm Troubleshooting

Table 9 – Alarm Troubleshooting
Alarm
%O2 < 18% High
%O2 > 104% High
High %O2
(> 5% above setting for 15
seconds).
Low %O2
(> 5 % below setting for 15
seconds).
Over pressure
(> 11 cmH in NCPAP and BiPhasic modes)
Over pressure
(> 15 cmH in BiPhasic tr* mode)
Low battery charge
(< 40%).
Priority Possible Cause
O
2
required.
O
2
High
required.
Blender setting changed.
Supply gas failure
Water trap overflow
High
Blender setting changed.
Supply gas failure
Water trap overflow
High
O
2
Flow rate set too high.
Occlusion of exhalation limb
Blocked silencer/bacteria filter
High
O
2
Flow rate set too high.
Occlusion of exhalation limb
Blocked silencer/bacteria filter
Warning indication.
Battery status indicator changes from green to red.
calibration
calibration
Actions
Restore FiO2 level to above the minimum limit
Press Alarm Reset for 3 seconds.
Recalibrate O
as soon as
2
practicable.
Restore FiO2 level to below the maximum limit
Press Alarm Reset for 3 seconds.
Recalibrate O
as soon as
2
practicable.
Press Alarm Mute to silence the alarm
Correct delivered oxygen concentration
Press Alarm Reset for 3 seconds to set new limits
Press Alarm Mute to silence the alarm
Correct delivered oxygen concentration
Press Alarm Reset for 3 seconds to set new limits
Check exhaust tube / filter
Reduce flow rate to achieve
pressure below high pressure limit
Press Alarm reset for 3 seconds to set new limits
Check exhaust tube / filter
Reduce flow rate to achieve
pressure below high pressure limit
Press Alarm reset for 3 seconds to set new limits
Connect external power
675–101–101 Rev. K
54 Appendix C - Alarm Troubleshooting
Alarm
Priority Possible Cause
Battery fault High (Cannot be
reset)
Low battery
Medium
voltage < 11.10 V for
5 seconds).
AC power
High
disconnected
High
High NCPAP / Pres Low
(CPAP > 3
O
cmH
2
above set
for 15 seconds).
Low NCPAP
High / Pres Low
( CPAP < 2
O
cmH
2
below set
for 15 seconds)
or
< 1.5 cmH
O
2
at any time).
High
High BiPhasic or BiPhasic tr* pressure
(MAP > 3
O
cmH
2
above
set for 15 seconds).
BiPhasic or BiPhasic tr*)
High (Cannot be
silenced) mode fails to operate as set.
Battery disconnected
Battery failing to hold charge
Battery disconnected
Battery failing to hold charge
AC power disconnected
NCPAP / Pres Low setting change
Circuit disconnect / reconnect
NCPAP / Pres Low setting change
Circuit disconnect / reconnect
Circuit leak
Pres High setting
change
Circuit disconnect / reconnect
See description of error code displayed
Actions
Push Alarm Mute button for 3 seconds to silence alarm
Refer to Service Engineer.
Push Alarm Mute button for 3
seconds to silence alarm
Connect AC power
Push Alarm Mute button for 3
seconds to silence alarm
Reconnect the AC power.
Push Alarm Mute button for 3
seconds to silence and reset alarm limits
Reset alarm limits after setting change and patient circuit disconnect / reconnect
Push Alarm Mute button for 3 seconds to silence and reset alarm limits
Reset alarm limits after setting change and patient circuit disconnect / reconnect
Check for leaks in patient circuit
Push Alarm Mute button for 3
seconds to silence and reset alarm limits
Reset alarm limits after setting change and patient circuit disconnect / reconnect
Revert to nCPAP mode
Refer to Service Technician
675–101–101 Rev. K
Infant Flow® SiPAP™ 55
rate
fails completely
power
Alarm
Low breath rate
Flow meter fault.
Gas Supply failure
Oxygen cell calibration error.
Electrical fault.
Water trap blocked
Software fault
Software not running with
unit connected to
Priority Possible Cause
High
Plus: audible / visual alarms only
Comprehensive : audible / visual alarms and backup
N/A
N/A
High
(Cannot be reset)
High
(Cannot be reset)
Rr = 0 for > low breath (apnea) interval timeout
No flow indications
Flow can’t be
adjusted.
Differential pressure between the two inlet gases falls outside of the limit of 30 PSI (2.0 bar) or one gas
Oxygen cell incorrectly calibrated, damaged or depleted
AC power LED does not match screen icon.
Full or leaking
Filter blocked
Loss of wall
pressure
Imbalance in gas supply
See description of error code displayed
See description of error code displayed
Actions
Push Alarm Mute button once to silence alarm
Restore patient breathing.
Check placement / connection of
abdominal Respiratory Sensor
Refer to Service Technician.
Check gas inlet supplies
Check inlet water trap
Refer blender to service technician
Calibrate or replace oxygen cell.
Refer to Service Manual.
Refer to Service Technician
Refer to Service Technician
Refer to Service Technician
Refer to Service Technician
*BiPhasic tr mode not currently available in the United States
675–101–101 Rev. K
56 Appendix C - Alarm Troubleshooting
Infant Flow®
S iPAP™
675–101–101 Rev. K
Infant Flow® SiPAP™ 57
and BiPhasic tr*)
worst error code

Appendix D - Fault Management

The general philosophy when handling a software detectable fault condition is to still allow a basic level of treatment to be applied to the patient - with over pressure protection, oxygen alarms and apnea monitoring (where possible), but inhibiting the higher level features of the unit (such as BiPhasic modes).

Fault classification

Each fault condition is classified according to the severity ratings (where means not available under software control,
(
) means may or may not be available depending on other severity rating 3 and 4
conditions, which may occur individually or simultaneously):
Table 10 – Fault Classification
means available under software control, and
Measurements
Control Modes and Features
Severity
1
(major)
2
3
Classification
Un-usable
Re­stricted
Un­trigger­able
Impact On Unit Functionality
Unit is inoperable under software control, but can still be used in an un­powered pneumatic mode
Unit functionality is restricted to NCPAP modes only
Patient trigger functions not available (NCPAP and BiPhasic with breath monitoruing on
Reporting Mechanism
A list of error codes are presented to the user via the "Fault lockout" display
Where applicable, error codes listed on mode selection screen. Status bar mode alternates with worst error code
Where applicable, error codes listed on mode selection screen. Status bar mode alternates with
O
O2 %
2
CmH
NCPAP
+apnea
Biphasic
tr
biphasic
Unit will not
4a
(minor)
No backup
operate on battery when the external power is removed
Battery status appears as if flat battery
() () ()
675–101–101 Rev. K
58 Appendix D - Fault Management
worst error code
error code.
Clas-
Software
Corrective
4b
(minor)
5
(minor)
6
(minor)
Severity
Classification
Spurious
Spurious
No oxygen monitor
Impact On Unit Functionality
Spurious software exception trapped
Non-fatal error trapped
Oxygen monitor and alarm functions are not available.
Reporting Mechanism
Software restarts, and status bar extended mode alternates with
High-priority alarm; status bar extended mode alternates with worst error code
High priority alarm may be cleared by operator reset; status bar extended mode alternates with
Measurements
O2 %
Control Modes and Features
O
2
CmH
NCPAP
() () ()
+apnea
Biphasic
tr
biphasic

Fault recovery / action

If a detectable fault condition occurs (either before treatment begins or while being applied) the software will respond in the following way:
Table 11 – Fault Recovery
E## Fault condition Consequence
Program memory
­checksum error
Battery too flat (<6.5V) to operate LCD,
-
analogue and valve driver circuits (no external power) Battery too flat (<10.25V) to operate analogue and valve driver
­circuits but
sufficient for LCD driver (no external power)
Software corrupt ­execution inhibited
No user interface display
Sensor readings invalid
sification
Unusable
Unusable
Unusable
Response
Hardware held in permanent reset condition with alarm bar lit (status LED on)
Hardware held in reset condition with alarm bar lit (status LED off) until external power applied
User lockout: "Plug in external power" prompt
Action Required
Service: Reload software
User: Plug in external power
User: Plug in external power
675–101–101 Rev. K
Infant Flow® SiPAP™ 59
calibration data
and Flow)
flat)
"E##" alarm
circuits
rail)
prompt
sense)
E## Fault condition Consequence
Unable to
Non-volatile
E10
memory fault
Calibration data
E11
lost
retrieve/set unit configuration and
Sensor readings invalid
Configuration DIP settings and/or PT PRESENT
E12
different to non­volatile
Possible incomplete unit
set-up performed configuration record Charged battery voltage too low
E20
(<11V) when
Battery capacity
low under test load External supply voltage too low
E21
(<14V) to charge battery (battery
Battery will not
charge
Clas-
sification
Unusable
Unusable
Unusable
No backup
No backup
Software Response
User lockout: Error "E##" prompt
User lockout: Error "E##" prompt
User lockout: Error "E##" prompt
Battery fault icon flashes and "E##" alarm
Battery low alarm continues even through external power applied -
Corrective Action Required
Service: Fix or replace PCB
Service: Low level calibration
, Pressure
(O
2
Service: Perform set-up procedure
Service: Fix Battery or charger
User: Plug in correct external supply Service: Fix PSU
Analogue supply
E22
rails out of limits
Valve driver
E23
supply rails out of limits Hardware 'safe-
E24
start' watchdog disabled Pressure sensor
E30
fault (ADC hits
Zero valve not
E31
connected (via sense) Zero valve
E32
activation fault (via sense) Unable to auto-
E33
zero pressure sensor Dump valve not
E41
connected (via
Dump valve
E42
activation fault (via sense)
Unreliable sensor
readings
Valve operations
unreliable
Unusable
Unusable
Valves disabled Unusable
Pressure sensor
readings invalid
Unusable
Pressure sensor
readings
Unusable unreliable Pressure sensor readings
Unusable unreliable Pressure sensor readings
Unusable unreliable
No over pressure protection
No over pressure protection
Restricted
Restricted
User lockout: Error "E##" prompt User lockout: Error "E##" prompt User lockout: Error "E##" prompt User lockout: Error "E##"
User lockout: Error "E##" prompt User lockout: Error "E##" prompt User lockout: Error "E##" prompt
Restricted mode: Error "E##" alarm
Restricted mode: Error "E##" alarm
Service: Fix circuits
Service: Fix valve supply rail
Service: Fix reset/safe-start circuits
Service: Fix sensor/circuits
Service: Fix valve/circuits
Service: Fix valve/circuits
Service: Fix valve/sensor/circ uits
Service: Fix valve/circuits
Service: Fix valve/circuits
675–101–101 Rev. K
60 Appendix D - Fault Management
or high gain)
blender.
timeout)
blender.
report Apnea
E## Fault condition Consequence
Oxygen sensor
E50
fault (ADC hits rail) Oxygen sensor can not be
E51
recalibrated by user (bad offset
Oxygen sensor calibrates but the
E52
fuel cell is worn­out (low gain)
Oxygen sensor readings invalid
Possible fuel cell, electronic, blender or gas supply fault
Oxygen sensor readings unreliable
Oxygen sensor
Oxygen sensor readings unreliable
E53
too noisy to calibrate (calibration
Oxygen calibration may be invalid (O
E54
reading below 17% or above
2
Oxygen sensor readings unreliable
104% detected)
Oxygen sensor
E55
disabled by the operator
BiPhasic valve
E61
not connected (via sense) BiPhasic valve
E62
activation fault (via sense)
Oxygen sensor readings unreliable
BiPhasic modes unusable
BiPhasic modes unusable
Apnea and/or
PT transducer
­disconnected
patient trigger unusable/interrup ted
PT module fault
E70
(PTRDY or CAN bus failure) No breath signal from PT module
E71
although CAN data does not
Apnea and patient trigger unusable
Patient may be in apnea but PT module dysfunctional?
Clas-
sification
No oxygen
monitor
No oxygen
monitor
No oxygen
monitor
No oxygen
monitor
No oxygen
monitor
No oxygen
monitor
Restricted
Restricted
Un-
triggerable
Un-
triggerable
Untriggera
ble
Software Response
High priority alarm; Error "E##" prompt
High priority alarm; Error "E##" prompt
High priority alarm; Error "E##" prompt
High priority alarm; Error "E##" prompt
High priority alarm; Error "E##" prompt
Error “E##” prompt
Restricted mode: Error "E##" alarm
Restricted mode: Error "E##" alarm
NCPAP and BiPhasic modes with breath monitoring on, inhibited (or low breath rate alarm given if treatment started) Reduced functionality ­"E##" alarm
Reduced functionality ­"E##" alarm
Corrective Action Required
Service: Fix sensor/circuits.
User: Check gas supplies. Service: Fix sensor/circuits/
User: Check gas supplies. Service: Replace sensor. User: Check gas supplies. Service: Fix sensor/circuits/
User: recalibrate the oxygen cell.
User: re-power the device to re­enable oxygen monitoring
Service: Fix valve/circuits
Service: Fix valve/circuits
User: Reconnect PT transducer
Service: Fix PT/circuits
Service: Fix PT/circuits
No trigger signal
E72
from PT module
BiPhasic tr* mode inoperable.
Untriggera
ble
Reduced functionality ­"E##" alarm
Service: Fix PT/circuits
675–101–101 Rev. K
Infant Flow® SiPAP™ 61
E## Fault condition Consequence
Spurious software interrupt, XTAL
(E9
fails, stack
0) overflow/underflo
w, CPU Class B exception
Abnormal hardware,
E90
software or watchdog reset
Internal software
E91
error detected
Unknown error
E99
detected
Software interrupted and restarts (possibly during treatment)
Software restarts possibly during treatment
Software unreliable
Software unreliable
[1] Error codes in parentheses (brackets) are generated as an indirect consequence of the problem.
* Biphasic tr mode not currently available in the United States
Clas-
sification
Spurious
Spurious
Unusable
Unusable
Software Response
Hardware reinitialized (disabled) with alarm bar lit and beeper sounding to identify root cause
Software restarts
- "E##" alarm
User lockout: Error "E##" prompt User lockout: Error "E##" prompt
Corrective Action Required
Software: Fix persistent exceptions
Software: Fix persistent exceptions Service: Fix abnormal reset
Software: Fix software error
Software: Fix software error

Fault code display screen

The fault lockout screen shall incorporate item ref. (as appropriate to build) and shall display a list of all active fault codes. Faults not resulting in user lockout shall result in indication on the status bar.
675–101–101 Rev. K
62 Appendix D - Fault Management
Infant Flow®
SiPAP™
675–101–101 Rev. K
Infant Flow® SiPAP™ 63
Term
Meaning
Apnea
Temporary inability to breathe.
mandatory)
CPAP
Continuous Positive Airway Pressure
mask
NCPAP+Apnea
NCPAP with Low Breath Rate monitoring (non-US labeling)
Rate
Mandatory rate (per minute); active in BiPhasic mode
RSP
Patient’s spontaneous respiratory rate (per minute)
s / sec
Seconds
US labeling
Labeling using English text in place of symbols and/or icons
PEEP
Positive End-Expiratory Pressure
PIP
Peak Inspiratory Pressure

Glossary

LBR Low Breath Rate
Bpm
Breaths per minute (applies to each of spontaneous, triggered and
Generator
BiPhasic
BiPhasic+LBR BiPhasic with Low Breath Rate monitoring (US labeling)
BiPhasic tr*
BiPhasic tr+Apnea* NCPAP Nasally applied CPAP NCPAP+LBR NCPAP with Low Breath Rate monitoring (US labeling)
RB
T
/ T
apnea
T-High Length of time (in seconds) for a sigh or manual breath (Time High).
Non-US labeling
LBR
Patient attachment for delivering CPAP, used with nasal prongs or
Time triggered, time cycled pressure assists at two separate pressures levels.
Patient triggered, time cycled pressure assists at two separate pressure levels. *This mode currently not available in the United States. BiPhasic tr* with Low Breath Rate monitoring (non-US labeling). *This mode currently not available in the United States.
Backup ventilator rate (in BiPhasic mode during apnea alarm, per minute; non-US labeling)
Apnea Interval (non-US labeling) or Low Breath Rate (LBR) monitor alarm time (US-labeling); both in seconds This mnemonic may also be associated with an alarm icon
Labeling using non-linguistic symbols in place of English text wherever possible
Pres Low
Pres High
675–101–101 Rev. K
Adjustable lower baseline pressure level control in BiPhasic and BiPhasic tr modes Adjustable upper pressure level control in BiPhasic and BiPhasic tr modes
64
Infant Flow®
SiPAP™
675–101–101 Rev. K
Infant Flow® SiPAP™ 65

Index

fixation, 31
A
Absominal Respiratory Sensor, 14 Accessories, 2 airway pressure, 15 alarm priority, 35 Alarm set/confirm Screen, 28 Alarm Symbols, 38 alarms, 4 Alarms, 35 Alarms Test, 19 assembly, 11
B
Battery Backup, 1 Battery Care, 42 BiPhasic, 1, 33 BiPhasic tr, 1, 34 button symbols, 45
Flat Battery alarm, 37 front panel, 23 Fully integrated alarm package, 1
G
gas flow, 15
H
High Airway Pressure alarm, 36 High and Low % O High Patient Circuit Pressure alarm, 36
alarm, 36
2
I
incompatible control settings, 26 increase buttons, 26 Indicators and Controls, 23 intended use, ix
C
calibration
O2 sensor, 17
cautions, 7 Circuit Occlusion Alarm, 20 classification, x Cleaning, 41 configuration
Comprehensive, 1 Plus, 1
Configurations, 49 control buttons, 25 controls, 3
D
decrease buttons, 26
E
electrical supply, 4 electromagnetic components, ix EMC, ix environmental requirements, 5 Error code indication alarm, 37
F
fault lockout screen, 61 faults, 57 features, 1
L
Leak Test, 18 Low Airway Pressure alarm, 36 Low Battery Charge alarm, 36 Low Battery Voltage alarm, 36
M
Main Screen, 29 Maintenance, 41 manual breath
setting, 26
Mode Confirm Screen, 29 modes, 3 Monitored Parameter Screen, 30 monitors, 3 MRI, ix
N
NCPAP, 1, 33 notes, 7
O
O2 sensor
calibration, 17
Operating Modes, 33 Operation without Electrical Power alarm, 37
675–101–101 Rev. K
66 Index
P
Parameter Adjust Screen, 29 parameter default values, 26 Patented Infant Flow patient circuit, 13 physical specifications, 5 Pneumatic Diagram, 51 pneumatic supply, 4 power-on check, 16
®
Generator, 1
R
radio frequency energy, ix Rear Panel, 24 reset alarm limits, 35
S
safety information, 7 Screen Displays, 27 Screen Lock, 2 Set Up screen, 27
silence active alarms, 35 Silencer / Bacterial Filter, 5 Soft-key operation, 25 specifications, 3 Storage, 42 Supply Gases Failure alarm, 36 symbols, 43
T
Troubleshooting, 53
U
User Interface, 27
User Verification Test, 16 User Verification Test Checklist, 21
W
warnings, 7 warranty, vi
675–101–101 Rev. K
Loading...