This document is protected by United States and International Copyright laws.
This document may not be copied, reproduced, translated, stored in a retrieval system, transmitted in any
form, or reduced to any electronic medium or machine-readable form, in whole or in part, without the
written permission of CareFusion. Information in this document is subject to change without notice.
This document is for informational purposes only and should not be considered as replacing or
supplementing the terms and conditions of the License Agreement.
800.231.2466 tel
+1.714.283.2228 tel
+1.714.283.8493 fax
carefusion.com
Literature number: 675–120 Revision G
+49.931.4972.0 tel
+49.931.4972.423 fax
675–120 Rev. G
Service Manual iii
September 2003
675-120(1)
Release
page 88.
Update list of service parts
Updated Table 6 starting on page 43
Revision History
Date Revision Changes
November 2004 B
December 2004 C
May 2005 D
Release manual in VIASYS Healthcare template using
VIASYS Healthcare Respir ator y Care nomenclature.
Revised per EO 27980. Removed the picture
from the title page.
Deleted the ESD warning from page 19.
Removed “O2 Sensor” and added the word “measured”
on page 30.
Changed “O2 sensor” to “fuel cell” on pages 39, 45, 62,
68, 69, 73, 86, and 87.
Replaced the O2 senor row on page 50.
Updated the error codes on pages 56 and 57.
Replaced the warning on page 61.
Changed the Fitting procedure on pages
64, 68, and 75.
Changed step 6 on page 65.
Added Addendum A – Oxygen Leak Test
Changed Transducer Assy. To Transducer Interface on
Revised per ECO 60329
Update address/contact info
Update battery remove/install procedure
Added note for fuel cell disposal
Update check valve assembly remove/install procedure
Update water trap & restrictor remove/install procedure
Delete redundant leak test
August 2008 E
675-120 Rev. G
Updated company information
Updated company information
Corrected Table 2 and updated Table 3
Updated Figure 6
Updated Figure 10
Updated the section “Diagnostics.”
ivInfant Flow
FRONT”
Model M675
Date Revision Changes
April 2010 F Changed Ti to T-Highthroughout
Figure 5 - Diagnostic screen updated
Spelling correction from “GRD” to “GND”
Added “Note: transducer must be attached.”
Changed “10 V” to “11.10 V”
Updated Table 6 - Error Codes (E50 - E54) and added
E55
Removed “Blender (Check Valves and Filter) Annually”
from table 7.
Added “, and two screws on the case-bottom”.
Added “and document L2879”
Replaced Ti symbol with T-High symbol
Changed “PL5” to “PL2”
Replaced Figure 3
Replaced Figure 4 and 5. Changed “value doesn’t” to
“value on the external oxygen analyzer doesn’t”.
Changed “467352” to “68289”; changed “777245” to
“52700A”; removed items 677-005A, 675-311, S117635,
467461, 467460.
Removed item S117641; changed “675-310” to “M675A-
®
LP SiPAP Model M675
May 2011 G
Added instructions and specifications for LP SiPAP
675-120 Rev. G
Service Manual v
Warranty
The Infant Flow® LP SiPAP and SiPAP systems are warranted to be free from
defects in material and workmanship and to meet the published specifications for
One (1) year from date of shipment.
The liability of CareFusion, (referred to as the Company) under this warranty is
limited to replacing, repairing or issuing credit, at the discretion of the Company, for
parts that become defective or fail to meet published specifications during the
warranty period; the Company will not be liable under this warranty unless (A) the
Company is promptly notified in writing by Buyer upon discovery of defects or failure
to meet published specifications; (B) the defective unit or part is returned to the
Company, transportation charges prepaid by Buyer; (C) the defective unit or part is
received by the Company for adjustment no later than four weeks followin g t he last
day of the warranty period; and (D) the Company’s examination of such unit or part
shall disclose, to its satisfaction, that such defects or failures have not been caused
by misuse, neglect, improper installation, unauthorized repair, alteration or accident.
Any authorization of the Company for repair or alteration by the Buyer must be in
writing to prevent voiding the warranty. In no event shall the Company be liable to
the Buyer for loss of profits, loss of use, consequential damage or damages of any
kind based upon a claim for breach of warranty, other than the purchase price of any
defective product covered hereunder.
The Company warranties as herein and above set forth shall not be enlarged,
diminished or affected by, and no obligation or liability shall arise or grow out of the
rendering of technical advice or service by the Company or its agents in connect ion
with the Buyer's order of the products furnished hereunder.
Limitation of Liabilities
This warranty does not cover normal maintenance such as cleaning, adjustment or
lubrication and updating of equipment parts. This warranty shall be void and shall not
apply if the equipment is used with accessories or parts not manufactured by the
Company or authorized for use in writing by the Company or if the equipment is not
maintained in accordance with the prescribed schedule of maintenance.
The warranty stated above shall extend for a period of One (1) year fr om date of
shipment, with the following exceptions:
1. Components for monitoring of physical variables such as temperature, pressure,
or flow are warranted for ninety (90) days from date of receipt.
2. Elastomeric components and other parts or components subject to deterioration,
over which the Company has no control, are warranted for sixty (60) days from
date of receipt.
3. Internal batteries are warranted for ninety (90) days from the date of receipt.
The foregoing is in lieu of any warranty, expressed or implied, including, without
limitation, any warranty of merchantability, except as to title, and can be amended
only in writing by a duly authorized representative of the Company.
675-120 Rev. G
viInfant Flow
®
LP SiPAP Model M675
Contents
Revision History ................................................................................ iii
Warranty ............................................................................................ v
Limitation of Liabilities ........................................................................ v
Notices ............................................................................................ viii
EMC Notice ............................................................................................................... viii
MRI Notice ................................................................................................................. viii
Intended Use Notice ................................................................................................... ix
Regulatory Notice ....................................................................................................... ix
Classification .............................................................................................................. ix
Declaration of Conformity Notice ................................................................................. x
This equipment radiates and is susceptible to radio frequency energy. If not installed
and used in accordance with the instructions in this manual, electromagnetic
interference may result. The equipment has been tested and found to comply with
the limits set forth in BS EN60601-1-2 for Medical Electrical Equipment Part 1-2:
General requirements for safety-collateral standard. Electromagnetic compatibility –
requirements and tests. These limits provide reasonable protection against
electromagnetic interference when operated in the intended use environments (e.g.
hospitals) described in this manual.
This device is also designed and manufactured to comply with the following
standards;
Safety:UL 60601-1: 2003 Medical Electrical Equipment, Part 1: General
Requirements for Safety.
CAN/CSA C22.2 No 601.1-M90, Medical Electrical Equipment - Part
1: General Requirements for Safety including C22.2 No. 601.1S1-94
(IEC601-1, Amendment 1:1991) Supplement No. 1-94 to CAN/CSA
22.2 No. 601.1-M90
®
LP SiPAP Model M675
With regards to Electrical Safety:
Class 1 equipment
Contains type BF patient applied parts
Continuous Operation
MRI Notice
This equipment contains electromagnetic components whose operation can be
affected by intense electromagnetic fields.
Do not operate this device in a MRI environment or in the vicinity of high-frequency
surgical diathermy equipment, defibrillators, or short-wave therapy equipment.
Electromagnetic interference could disrupt the operation of the device.
675-120 Rev. G
Service Manual ix
Intended Use Not ice
The Infant Flow® LP SiPAP system, consisting of a Driver and Generator plus
NCPAP Prongs and Masks, is a medical device intended for the provision of Bi-Level
CPAP to produce a sigh. This system is for use in Hospital, Hospital Type facilities
and intra-Hospital transport environments and is indicated for the treatment of
Newborn and Infant patients.
Operators of this equipment and Se rvice En gineers are required to read and thoroughly
understand the contents of this manual before using or maintaining the equipment.
This manual is intended for use by a competent, fully qualified Service
Engineer. It includes a description of the unit and how it w or ks. It also contains
operating and diagnostic procedures and maintenance instructions. For usage
of associated equipment, refer to the Manufacturer’s literature.
Regulatory No t ice
Federal law restricts the sale of this device except by or on order of a physician.
Classification
Type of Equipment: Medical Equipment, Class 1 and internally powered, IPX1
Protected, and uses type BF applied parts. Equipment is not suitable for use in
presence of flammable anesthetics.
675-120 Rev. G
xInfant Flow
Declaration of Conformity Notice
This medical equipment complies with the Medical Device Directive, 93/42/EEC, and
the following Technical Standards, to which Conformity is declared:
Council Directive(s): MDD 93/42/EEC Annex II (excluding section 4)
Safety: EN 60601-1, EN 794-1
EMC: EN 60601-1-2:2001
Conformity Assessment: MDD Annex II
Quality System: ISO 13485
EU Notified Body: BSI (Reg. No. 0086)
Device Classification: IIb
CareFusion
22745 Savi Ranch Parkway
Yorba Linda, CA 92887, USA
If you have a question regarding the Declaration of Conformity for this product,
please contact CareFusion.
675-120 Rev. G
Service Manual 1
Chapter 1 – Product Description
The Infant Flow® LP SiPAP and SiPAP systems are non-invasive forms of respiratory
support designed for use in hospital environments such as Neonatal and Pediatric
Intensive Care Units. They can also be used when transporting patients within the
hospital environment.
The Infant Flow
Comprehensive configuration. The Plus configuration provides NCPAP and time
triggered, BiPhasic modes with and without breath-rate monitoring. The
Comprehensive configuration offers these features plus patient BiPhasic mode with
apnea backup breaths. The Infant Flow
standard in all configurations with an LCD touch screen display, pressure time
waveform graphics, integrated patient monitoring, alarms for high and low pressure
and FiO
and up to 2 hours of backup battery power.
2
As a result of the unique, patented design, the Infant Flow
systems have been proven to provide the most stable CPAP at the lowest work-ofbreathing for patients compared to other devices. The outstanding performance of
the Infant Flow
expiratory flows. This system has been designed and tested to perform optimally
when used only with accessories available from CareFusion. These accessories
include circuits and generators, prong and mask patient interfaces and bonnets.
®
LP SiPAP and SiPAP systems are currently available in the Plus or
®
LP SiPAP and SiPAP systems come
®
LP SiPAP and SiPAP
®
LP SiPAP and SiPAP systems is irrespective of patient demand or
Infant Flow® LP SiPAP and SiPAP Features
The expanded capabilities of the Infant Flow® LP SiPAP and SiPAP, Plus and
Comprehensive configurations, allow for applications to a broader range of patients
who may otherwise not be candidates for non-invasive respiratory support.
NCPAP. Allows for continuous positive airway pressure based on clinician set
pressure. Breath rate monitoring/alarm can be activated in this mode.
BiPhasic. Allows for time triggered pressure assists to be delivered based on
clinician set inspiratory time, rate, and pressure criteria. Breath rate monitoring/alarm
can be activated in this mode.
BiPhasic tr*. Allows for patient triggered pressure assists to be delivered based on
clinician set inspiratory time and pressure criteria. Breath rate monitoring/alarm, and
Apnea backup breaths are automatically active in the mode.
Patented Infant Flow
device for the generation of consistent infant nasal CPAP with a low work of
breathing. The Infant Flow LP system has a lower driving pressure than other
variable-flow devices.
Fully integrated alarms packages. Supply gases failure, High Patient Pressure,
Low patient pressure, high and low delivered Oxygen concentration, change from AC
to DC power source, low and flat battery charge status and Low breath rate/apnea
alarm.
®
LP Generator. The Infant Flow® LP Generator is a fluidic
Battery Backup. Up to 2 hours of battery backup allows for intra-hospital transport.
Clear indicators are provided for power supply in use (AC or DC), and battery charge
level.
675-120 Rev. G
2Infant Flow
NCPAP
•
•
NCPAP with breath rate monitoring
and alarm
• •
BiPhasic
•
•
and alarm
• •
BiPhasic tr*
•
Internal Battery
•
•
Manual Breath
•
•
Apnea Back up rate
•
Screen lock
•
•
Prioritization of alarms
•
•
LP. Refers to the low driving pressure of the Infant Flow LP Generator. The Infant
Flow LP SiPAP system is compatible only with the Infant Flow LP Generator. T he
prescribed nCPAP level will not be obtained if other variable flow generators are
used.
Screen Lock. After 120 seconds of no screen inputs, the screen changes to the
Locked Screen to prevent inadvertent changes. Upon activation of a high priority
alarm the screen changes to an unlocked state to allow for immediate interventions
as required.
Table 1 - Functions and Accessories
Functions & Accessories Plus Comprehensive*
®
LP SiPAP Model M675
BiPhasic with breath rate monitoring
*Comprehensive configuration not available for sale in the United States
CAUTION!
The Infant Flow® LP SiPAP and SiPAP systems have been designed and tested
as complete systems using Infant Flow® accessories. Only accessories approved
for use should be used. If in doubt, please contact your local CareFusion
representative.
WARNING!
The Infant Flow® LP SiPAP system is compatible only with the Infant Flow
LP Generator. The prescribed nCPAP level will not be obtained if other
variable flow generators are used.
675-120 Rev. G
Service Manual 3
Chapter 2 – Product Specifications
Modes
• NCPAP
• NCPAP with breath rate monitoring and low rate alarm
• BiPhasic (t im e triggered)
• BiPhasic (t im e triggered) with breath rate monitoring and low rate alarm
• Relative Humidit y -Operating: 0 – 90% non-condensing
• Storage: 0 – 90% non-condensing
Physical
•Dimensions (driver only) -(W x H x D) 26 x38 x 23.5 cm /
10.25 x15 x 9.25 in
•Weight (driver only)-8.8 kg / 19.5 lb
675-120 Rev. G
6Infant Flow
®
Infant Flow
LP SiPAP an d SiPAP
Model M675
®
LP SiPAP Model M675
675-120 Rev. G
Service Manual 7
Chapter 3 – Warnings and Cautions
Please review the following safety information before operating the Infant Flow® LP
SiPAP or SiPAP system. Attempting to operate this equipment without fully
understanding its features and functions may result in unsafe operating conditions.
Warnings and Cautions, which are general to the use of the device under all
circumstances, are included in this section. Some Warnings and Cautions are also
inserted within the manual where they are most meaningful.
Notes are also located throughout the manual to provide additional information
related to specific features.
If you have a question regarding the installation, set up, operation, or maintenance of
the device, contact technical support at CareFusion.
Terms
WARNINGS identify conditions or practices that could result in serious adverse
reactions or potential safety hazards.
CAUTIONSidentify conditions or practices that could result in damage to the
ventilator or other equipment.
NOTESidentify supplemental information to help you better understand how
Warnings
•Whenever a patient is attached to respiratory care equipment, constant
attendance is required by qualified personnel. The use of an alarm or
monitoring system does not give absolute assurance of warning for every
malfunction that may occur in the system. In addition, some problems may
require immediate attention.
•The gas blender incorporated in this product is designed to mix medical
grade air and oxygen only. Do not modify the inlets to accommodate other
source gases such as anesthetic gases.
•Check that the water trap is empty before use and empty it frequently during
use.
•Liquid water or other contaminants in either gas supply, particularly the air
supply, may cause malfunction of this equipment and equipment connected
to it.
•When f illing a humidifier, do not move the stand. Moving or transporting the
stand while refilling may cause the stand and equipment to over balance.
the ventilator works.
•The Infant Flow LP SiPAP system is compatible only with the Infant Flow LP
675-120 Rev. G
Generator. The prescribed nCPAP level will not be obtained if other variable
flow g ener ators are used.
8Infant Flow
• Do not use conductive patient circuits with the Infant Flow® LP SiPAP Driver.
• Nasal CPAP can cause nasal irritation, septal distortion, skin irritation and
pressure necrosis. Observe the usage guidelines to minimize these
complications.
®
LP SiPAP Model M675
•This device exhausts O
during normal operation. Oxygen vigorously
2
accelerates combustion. To avoid fire hazard, do not place flammable
materials or sources of heat close to the exhaust.
• Do not use the equipment without the exhaust tube fitted (refer to Figure 2).
• To reduce trip hazard, always ensure cable and tubes are restrained away
from walking areas.
•The Abdominal Respiratory Sensor will not detect all forms of apnea.
Independent monitoring should always be used with this device.
•If the unit is shelf mounted, ensure that the unit is stable and that hoses and
cables are restrained to avoid hazard of toppling.
•This equipment is not suitable for use in the presence of a flammable
anesthetic mixture.
•The NCPAP Pres High flowmeter must be adjusted to zero when not required
for the patient.
•Under extr eme conditions (minimum supply pressure and maximum gas
demand, including auxiliary output) output flow rates and delivered pressure
may be reduced.
• Only use the supplied AC cable to connect to the power supply.
• Do not attach the Generator to the patient until the initial set up is complete.
• The
indicates a connection between the transducer interface and the
unit. It does not indicate correct positioning of the Abdominal Respiratory
Sensor.
• Calibration m ust only be done when the unit is not connected to the patient.
• Verif y that the displayed value for delivered FiO
corresponds to the value set
2
on the blender. Refer to Faults and Indications.
•Oxygen vigor ously accelerates combustion. To avoid explosion hazard, do
not use any instrument or other equipment that may have been exposed to oil
or grease contamination.
675-120 Rev. G
Service Manual 9
Cautions
•Federal Law (USA) restricts this device to sale by or on the order of a
physician.
•The precision gas blender in this product may become non-functional or
damaged if used without the protective water trap and filters provided.
•The power switch on this unit does not isolate the external power supply.
Disconnect the power supply cable to ensure complete isolation.
•Before use, verif y that this equipment has been authorized for use by a
qualified person.
•The Infant Flow
tested as complete systems using Infant Flow
accessories approved for use should be used. If in doubt, please contact your
local CareFusion representative.
•Where the integrit y of the external protective earth conductor is in doubt, the
equipment shall be powered by its internal power source (battery).
•Do not immerse any part of the IFSD in water or sterilize it with gas or steam.
®
LP SiPAP and SiPAP systems have been designed and
®
SiPAP accessories. Only
•Ensure patient breathing circuit is replaced at regular intervals.
675-120 Rev. G
10Infant Flow
®
Infant Flow
LP SiPAP and SiPAP
Model M675
®
LP SiPAP Model M675
675-120 Rev. G
Service Manual 11
Chapter 4 – System Construction
CAUTION!
Where the integrity of the external protective earth conductor is in doubt the
equipment shall be powered by its internal power source (battery).
The IFSD is AC powered with an integral rechargeable DC battery that provides
power for up to two hours without any interruption of performance or function. If the
AC power supply fails or is disconnected, the IFSD automatically switches to battery
power and gives an audio and visual alarm.
The IFSD is enclosed in a case with Operator controls and input connectors on the
front and rear panel. The front panel is shown in Figure 1. The back panel is shown
in Figure 2. The case incorporates non slip feet for table top use or must be fitted to
a dedicated stand. The major components within the casing are:
• a gas module
• an electronics module
• a front panel module
• a patient trigger module
• a firmware module
675-120 Rev. G
12Infant Flow
®
LP SiPAP Model M675
Figure 1: IFSD Front Panel
Gas Module
The function of the gas module is to take air and oxygen, blend them into the
required mixture and deliver this mixture to the patient at the prescribed flow rat e.
The gas module also measures the oxygen concentration, measures the patient
pressure, provides an auxiliary gas outlet (optional) and provides switched Biphasic flow.
The main components are an air/oxygen blender, a flow manifold, a vent valve, an
exhaust manifold with alarm whistle, NCPAP Pres low and NCPAP Pres high flowmeters,
and a valve/sensor PCB. The inlet gas connections are on an interc han geable i nlet bl ock
to allow for different gas fittings. The exhaust manifold discharges gas to the outside of
the case and is positioned away from the electrical connectors and switch to reduce any
potential explosive hazard.
675-120 Rev. G
Service Manual 13
Leak Test
Electronics Module
The function of the electronics module is to power the unit either by AC mains supply
or DC emergency battery supply, to control the gas module and read the gas module
sensors. The main components are a power supply unit, a rechargeable batt ery, a
main processor PCB, LED PCB, Valve/Sensor PCB and a LCD screen (touch
screen). The LCD screen includes a back-light which is always on when the IFSD is
powered. The touch screen displays information and receives inputs from the Operator
via the touch screen keyboard.
Figure 2: IFSD Back Panel
Port
675-120 Rev. G
14Infant Flow
®
LP SiPAP Model M675
Figure 3: IFSD Internal Components
675-120 Rev. G
Service Manual 15
Front Panel Module
The function of the front panel module is to house the gas and electrical connections
to the patient, Operator controls and indicators. The module consists of a front panel
plate, the touch screen with key pad, flowmeters and FiO
and indicators and an ambient light sensor. The backlight on the touch screen is
decreased if the ambient light sensor detects a low ambient light level and increased
if it detects a high ambient light level.
The green Power light is always on when AC power is connected to the unit.
control, patient connectors
2
The Alarm Warning Bar
Patient Trigger Module (P lus and Comprehensive Models)
The patient trigger module consists of a PCB which plugs into the main processor
PCB. Its function is to detect patient breaths and apnea and give this information to the
main processor in the electronics module. The main processor uses the signals from the
patient trigger module to instruct the biphasic pressure control to provide a timed sigh
to the patient.
Firmware Module
The firmware module is the unit’s embedded software. Its function is to instruct the
microprocessor how to control the unit and to interact with the Operator.
Touch Screen
The touch screen provides the Operator with a series of screens with icons to enable
settings, calibration and fault diagnosis. The Start Up Screen is shown in Figure 4.
The display includes a status bar which incorporates a battery status, mode
indicator, alarm button and patient trigger indicator. The center part of the display
shows icons which relate to the function being selected or performed. The display also
includes a key pad with six keys. The icon in each key changes depending on the
function being performed.
flashes red to indicate an alarm.
Battery Status and Chargi ng Indicator
If the battery status shows three bars or less, the display flashes alternately between
red bars on a white background and a pink background.
Alarm Button
The alarm button alerts the Operator to fault conditions. An audible alarm is activated
at the same time. In the unalarmed condition the button is green
high-priority alarm condition, the button flashes
heard. If the alarm button is pressed (to silence the audible alarm), the button
changes to
flashes
condition, the button is solid
675-120 Rev. G
. In the
red and an audible alarm is
and flashes red. In a medium-priority alarm condition, the button
yellow and an audible alarm is heard. In a low-priority alarm
yellow with an audible alarm. If any alarm
16Infant Flow
condition resets itself, the yellow alarm bar remains to alert the clinician of a previous
problem.
Caution/Information
The Caution/Information icon alerts the Operator to read this manual. It flashes
during an alarm condition.
Mode Indicator
The Mode Indicator shows four question marks when in Start up or Adjust and
changes to show the applicable mode in use (e.g. NCPAP).
®
LP SiPAP Model M675
Figure 4: Touch Screen Display – Start Up
Patient Trigger Indicator (Plus and Comprehensive Models)
The Patient Trigger indicator changes to when the transducer interface is
connected and ready to be connected to the patient and indicates a
transducer interface is not connected.
when the
675-120 Rev. G
Service Manual 17
Alarm Conditions
Audible and visible alarm indications are given to alert the Operator to specified
conditions that affect the operation of the unit. The electronic alarm limits are
automatically set after two minutes without Operator inputs but the Operator can
manually set the alarm parameters for certain conditions if required. The IFSD has
the following alarms:
•an audible W histle which sounds a constant tone when the gas imbalance
limits are exceeded.
•an audible Alarm which sounds a constant tone or two different intermittent
tones; medium or high level (medium level sounds beep.beep.beep every 15
seconds and high level sounds beep .beep .beep beep beep every 10 seconds).
•a visual Warning Bar which flashes red continuously, flashes intermittently or
comes on continuously.
•an Al arm Button on the touch screen status bar which flashes continuously.
Other indications are given to indicate the status of the condition. Refer to Table 2for
alarms and indications for specified conditions.
Diagnostic Screen
A diagnostic screen is provided for low level calibration, test procedures for valves
and an error log. Figure 5 shows the screen display. Labeling of Diagnostic screen
will be the same regardless ofconfiguration.
Figure 5: Diagnostic Screen
675-120 Rev. G
18Infant Flow
Box
Description
PCB. Position 1 is on and 0 is off.
INDICATOR
contrast.
successfully completed and a if the calibration fails.
Box
Description
S/W
The S/W box shows the version and serial number of the loaded firmware.
(E2P).
PT
The PT box shows the status of the Patient Trigger module.
inspiratory time, and breath rate (controlled by DIP 1).
(BLL) and if the backlight is operative (BLF - 1 for fail, O for on).
(BAT), the battery charge status (CHG) and the temperature (TMP).
8.2 and 10 V).
RS232
The RS232 box shows factory set data.
values.
(OSEN) and the software offset (OFFS) and gain (GAI N) values.
for the zero valve and the Z sensor (ZSEN) voltage.
for the dump valve and the D sensor (DSEN) voltage.
for the pressure assist valve and the P sensor (PSEN) voltage.
Table 2 - Diagnostics Screen Boxes
DIPS The DIPS box shows the position of the DIP switches on the main processor
®
LP SiPAP Model M675
CONTRAST
CALIBRATION
INDICATOR
The contrast indicator box is for use in the factory for setting the screen
The calibration indicator box shows
during calibration, a when
The following display boxes show if the values are within t he set parameters, a
if the values are outside the set parameters, a if the function is not
calibrated and a
if the function is disabled on the DIP switch.
Table 3 - Additional Diagnostics Screen Boxes
H/W The H/W box shows the status of the watchdog t i mer (WDG) and the EEPROM
CONFIG The CONFIG box shows the softw are con figurati on based on the DIP swit ch
settings and requirements. LAB = INTL or US (controlled by DIP 3), which is the
Comprehensive or Plus model; tBiP = 1 or 0 (controlled by DIP 2 and 5), which
indicates Triggered BiPhasic is enabled (1) or disabled (2); APB = 1 or 0
(controlled by D IP 2 ), w hi ch in dicates that Apnea BiPhasic is enabled (1) or
disabled (0 ); The re maining sy mbols provi de the de fault and r ange s for
LCD The LCD box shows the ambient light sensor voltage (AMB), the back light level
PSU The PSU box shows the external power supply voltage (EXT), the battery voltage
RAILS The RAILS box shows the ground voltage (GND), and the control voltages (6.5,
PRESSURE The PRESSURE box shows cm H2O pressure value, the voltage from the
pressure sensor (PSEN) and the software offset (OFFS) and gain (GAIN)
OXYGEN The OXYGEN box shows the FiO2 %, the voltage from the oxygen sensor
ZVALVE The ZVALVE box shows if the Z DRIVE is operative (ZDRV - 1 for fail, O for on)
DVALVE The DVALVE box shows if the D DRIVE is operative (DDRV - 1 for fail, O for on)
PVALVE The PVALVE box shows if the P DRIVE is operative (PDRV - 1 for fail, O for on)
675-120 Rev. G
Service Manual 19
Chapter 5 – Theory of Operation
Gas Flow
With the oxygen and air connections connected and the power switch ed on, oxygen
and air at 386 kPa (56 psig) flows to the blender. The air passes thr ough a water trap
with an integral filter where any moisture in the air is removed. Oxygen and air are
filtered before entering a non-return valve in the blender. The blender mixes the
oxygen and air in the proportion set by the position of the FiO
blender supplies blended gas between 2.5 and 30 LPM.
The blender incorporates an alarm/bypass module. If a single gas supply fails or
there is an imbalance between the inlet gas pressures exceeding 200 kPa (30 psig),
the module directs the flow through a reed whistle to warn the Operator of the
condition and at the same time connects the inlet gas at the hig her pressure to the
blender outlet.
The blended gas is filtered before passing through a flow control to an oxygen
analyzer. The oxygen analyzer utilizes a galvanic fuel cell to measure and display
the measured delivered oxygen concentration.
control valve. The
2
A vent valve on the exhaust manifold incorporates a solenoid operated pilot valve
and a large orifice spool valve. The vent valve is normally held in the open position
by a spring to vent high flow rates with low pressure drops so that the patient
pressure is minimal. The electrically actuated pilot valve is operated by blended gas
so that the vent valve continues to operate if one gas supply is lost.
Blended gas flows through an electronically operated dump valve which is normally
in the closed position. If over pressurization occurs, the sensors detect the condition
and the controller signals the dump valve to open, sounds an alarm buzzer and at
the same time closes the vent valve to rapidly reduce the patient pressure to near
zero. When the over pressure condition stops, the controller signals the vent valve to
open and the dump valve to close.
The NCPAP flowmeter is set by the Operator to the required flow. The flow rate is
shown on the gauge on the flowmeter. When the flowmeter control knob is turned
fully clockwise, the flow of gas is turned off.
675-120 Rev. G
Loading...
+ 67 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.