• Deterioration in the respiratory condition, requiring
immediate ventilation
• Patient discomfort from prong/mask intolerance
• Nasal septal injury (e.g., columella necrosis)
• Blanching of the nares
• Dry mouth and airways
• Gastric ination
Columella necrosis
Nasal dilation
3
CPAP modalities
What are the treatment options?
A variety of technologies have been employed in nCPAP
delivery throughout the years.
Conventional CPAP (V-CPAP): Utilizes a traditional mechanical
ventilator to deliver a constant flow of gas. CPAP is created
by changing the expiratory port orifice size. The ventilator
equipment is comprehensive and expensive.
Bubble CPAP (B-CPAP): Utilizes a constant flow of heated
and humidified gas. The level of pressure is controlled by
the depth of the exhalation tube inserted into a water
container. The pressure can increase if condensate collects
in the tubing, the flow rate changes or the water evaporates
from the container. B-CPAP lacks system alarms and
imposes a higher WOB
inability to entrain flow during inspiration.
due to the constant flow and
(16 ,18 ,19)
High flow nasal cannula (HFNC): Has not been cleared by
the FDA for nasal CPAP delivery. HFNC utilizes a constant
flow of heated, humidified gas that potentially delivers a
positive distending pressure. The level of therapy cannot
be measured and fluctuates depending on body position,
oral leaks, nasal secretions and the size and weight of the
patient. HFNC does not contain critical alarms that ensure
the safe delivery of therapy.
Variable flow CPAP (VF-CPAP): Incorporates a generator that
redirects the heated and humidified gas flow away from the
patient during exhalation and allows air entrainment during
periods of high inspiratory effort.
provides the most stable pressure, even in the presence of
leaks up to 6 LPM.
Pandt and Associates demonstrated that the Infant Flow
variable flow technology delivered a consistent level of
CPAP with little fluctuations. In contrast, the conventional
CPAP did not reach the desired level of 5 cmH
O, and
2
the pressure fluctuated significantly throughout the
breath cycle.
9
Using a variable flow generator with a dedicated CPAP
driver provides a measurable therapy with system alarms
to help ensure safe and effective therapy.
Infant Flow
8 Lts flow, 5 cmH2O
System pressure
5 cmH2O
0 cmH
O
2
Time
Conventional CPAP
17 Lts flow, 5 cmH2O with 5 Lts reservoir bag
System pressure
5 cmH2O
By redirecting the gas, VF-CPAP offers a lower imposed
WOB and less expiratory resistance compared to other
nCPAP technologies. Because the pressure is created and
measured at the nares, the variable flow technology
Infant Flow variable ow
Inspiration: Gas flow converted
to pressure reducing the WOB
and maximizing the pressure
stability at the patient interface.
4
0 cmH
O
2
Moa, G., Nilsson, K. et al. Crit Care Med, 1988 , 16(12):1238 –1242 .
Expiration: Gas flow
flipped away from the nasal
prongs to the expiratory
tube. The residual gas
pressure provided by the
continuous gas flow creates
a stable CPAP throughout
the respiratory cycle.
Time
Variable ow technology
Inhalation
Dual jets
Flow directionFlow direction
Patient nare
What is variable ow technology?
10,11
The Infant Flow LP patented dual-jet variable flow
generator utilizes fluidic technology to deliver a
constant CPAP at the airway proximal to the infant’s
nares. Without moving parts or valves, the generator
provides consistent performance. The level of CPAP
created is proportional to the flow provided by the
driver; for example, 9 LPM creates approximately
5 cmH
O CPAP. The variable flow generator uses
2
Bernoulli’s Principle via injector jets directed toward
each nare. If the infant pulls additional flow, the
venturi action of the injector jets entrains additional
flow from either the source gas or exhalation tube
reservoir. During exhalation, the incoming gas flow
redirects away from the infant. This action is referred
to as the “fluidic flip.” By redirecting the gas, variable
flow nCPAP reduces the imposed WOB. The infant
can exhale freely and conserve precious calories for
development. In summary, the direction of gas flow
in variable flow devices depends on the patient’s
respiratory cycle. The flow “flips” away from the
nares when the infant exhales and then, “flips” back
as the exhalation phase ends. The response is almost
instantaneous as it occurs at the patient’s nares.
Fluidic flip
Exhalation
Exhaust tube
Vortice shedding
Patient nare
What is vortices technology?
10,12
The patented Infant Flow LP generator is a new form of
variable ow that uses vortices technology to reduce the
imposed WOB during inhalation. Similar to the single- jet
technology, the ow entrainment reduces the WOB
on inhalation by meeting the patient’s inspiratory ow
demand and during exhalation gas ow ips away from
the patient reducing resistance.
Four low-momentum jets (two per nare) impinge inside
the generator to create a consistent and measurable
positive airway pressure within the generator head.
During inhalation, the dual jets entrain ow to meet
the patient’s inspiratory demand. During exhalation, the
jets easily deect to disrupt the gas ow. This disruption
of ow creates vortice shedding that spirals outwardly,
combining with the exhaled breath to create an
organized, efcient ow path toward the exhaust ports.
Exhaust
tube
Impinging
jets
Drive
line
Pressure
line
Patient
5
Self assessment
1. Describe RDS:
2. List three indications for nCPAP therapy:
3. List three benets of nCPAP therapy:
4. List three potential complications to nCPAP therapy:
8. Match the generator parts to the diagram:
Pressure line
Impinging jets
Exhaust tube
Drive line
Patient
c
d
b
a
e
Notes:
5. State four methods used to deliver nCPAP therapy:
6. Discuss the advantage of variable ow technology compared
to other CPAP modalities:
7. Discuss the importance of low work of breathing:
6
Infant Flow SiPAP
driver overview
Infant Flow LP system
The Infant Flow LP nCPAP system is a comprehensive system for delivering unique nCPAP therapy.
The system consists of:
• Infant Flow SiPAP driver
• Infant Flow LP generator assembly
• Infant Flow LP fixation—bonnet or headgear
• Infant Flow LP nasal interfaces—mask or prongs
This chapter discusses the set-up and operation of the Infant Flow SiPAP Plus driver.
Power LED
Indicates power on
and AC connection
Touch screen display
Adjusts settings and
displays patient parameters
and alarms.
Pressure low flow meter
Adjusts low CPAP level.
CPAP supplied to patient
directly relates to flow rate
from the nCPAP driver.
Alarm LED
Indicates alarm situation as visual and
audible notification color varies according
to alarm level
Note: Refer to the Infant Flow SiPAP
operator manual for additional instructions
on the set-up, operation and maintenance
of the Infant Flow SiPAP Plus driver.
Transducer LED
Indicates
transducer
connected.
% O
blender control
2
Adjusts FiO
Pressure high flow meter
Adjusts high CPAP level. CPAP
supplied to patient directly relates
to flow rate from nCPAP driver.
2.
Transducer
connection
Connects to
transducer interface.
Patient circuit
connection
Connects patient
breathing circuit to
gas outlet port.
Pressure line
connection
Connects patient
pressure line to device.
7
Infant Flow SiPAP congurations
The Infant Flow SiPAP driver is sold globally and is available
in different configurations. The two main models are Infant
Flow SiPAP Plus and Infant Flow SiPAP Comprehensive.
The Comprehensive model offers an additional ventilation
mode, BiPhasic trigger, which is not available in the U.S.
In select areas, additional languages or an international icon
overlay may be used in place of the English text. The operation
and maintenance of the Infant Flow SiPAP driver is the
same regardless of the specific configurations. Refer to the
Infant Flow SiPAP operator manual for more specific details.
DescriptionEnglish textICON symbol
Press to access the user calibration menu and language options.*
Press to return to the start-up screen.
Press to switch between the graphical and numerical monitoring screen.
Press to change the operation mode.
Press to deliver a manual breath. The breath delivers at the pressure high setting for
the set time high duration.
Indicates user should refer to the operator manual for additional information.
Indicates battery status and turns red if the battery charge is less than 40%.
Indicates the screen is locked. Press to unlock the screen.
Adjust the low flow rate setting for the baseline CPAP level.
Adjust the high flow rate setting for BiPhasic high CPAP level.
*Language option not available on all SiPAP models.
8
Infant Flow SiPAP display screen
Mode, control settings and function buttons
Mode
indicator
Monitored
parameters
FiO
Alarm priority/alarm
silence indicator
2
Operational
information
Battery
charge
Pressure/ Time graphics
or monitored
parameters display
Control/Setting
indicators
Soft key color code
White letter
Key enabled
Faded letter
Key inactive
Yellow letter
Solid: Pending confirmation
Flashing: Low-priority alarm
Red letter
Flashing: High-priority alarm
Solid: Reduction in another parameter
caused by an adjustment
Alarm management
High priority
• Series of 10 tones sound every 10 seconds
• Parameters display, and limits flash red
Medium priority
• 3 audible tones sound every 15 seconds
• Parameters display, and limits flash yellow
Low priority
• 2 audible tones sound every 30 seconds
• Parameters display, and limits change to yellow
9
Infant Flow SiPAP and circuit set-up
Circuit set-up
1. Gather the nCPAP supplies:
• Infant Flow SiPAP driver
• Single-limb, heated breathing circuit
• Infant Flow LP generator kit
• Infant Flow LP xation device
• Humidier and chamber
• Sterile water bag
2. Attach the water chamber to the humidier and
connect it to the water feed system. Follow the
manufacturer instructions for the proper set-up.
3. Connect the gas delivery tubing (A) to the ow
driver outlet port (I) and humidier chamber port (II).
4. Connect the elbow connector on the heated
breathing circuit (B) to the humidier chamber.
Insert the heater wire plug into the wire socket.
Securely insert the temperature probe in the port on
the circuit elbow (III). Insert the second temperature
probe (IV) into the airway port at the distal end of
the breathing circuit.
5. Connect the non-heated section (C) to the drive line
of the generator assembly (D).
I
A
II
B
V
III
E
6. Connect the proximal pressure line (E) to the
proximal port on the driver (V) and the pressure
line on the generator.
Temperature probe
When inserting the temperature probe into the
circuit, ensure the probe tip is in the middle of the
gas stream. This allows the gas temperature to be
measured accurately. If the probe is not properly
seated, the temperature measurement accuracy may
be compromised, leading to excessive condensation.
Cover the temperature probe with a reective shield
when used under a radiant warmer or bilirubin light.
IV
C
D
10
Humidication and nCPAP
Humidication
Heated humidification is recommended for nCPAP therapy.
The normal functions of the nose and air passages of
the respiratory tract are too warm, moisten and filter
the inhaled gases before they reach the lungs. In normal
respiration, the nasal mucosa and upper airways provide
75% of the heat and moisture supplied to the smaller
airways and alveoli. By the time air reaches the alveoli,
the inspired gas warms to 37 °C at 100% relative humidity
13
With nCPAP, the upper airways are not bypassed,
(RH).
but the high gas flows may be drying to the airways,
especially to a neonate’s underdeveloped lung. Adequate
humidification is essential to maintain airway clearance,
optimize ventilation and improve patient comfort.
The International Organization for Standardization (ISO) and
American Association for Respiratory Care (AARC) clinical
practice guidelines recommend gas temperature
between 34 and 41 °C to provide a humidity level of
33 to 44 mgH
O/L with artificial airways.14 Be cautious
2
using higher temperatures, as condensation may reduce
the mucous viscosity and interfere with the mucous
clearance. Extended exposure of gas temperatures over
41 °C may cause cellular damage to the airways.
14
The higher temperature settings may not be required to
deliver adequate humidification, since an artificial airway
is not used with nCPAP. Start with a temperature setting
of 36 °C to 37 °C and adjust the humidifier settings to
maintain adequate humidification; if condensate occurs,
reduce the humidifier temperature setting.
Airway temperature probe placement
Open bed or crib
When the infant is placed on an open bed warmer or
crib, it is recommended to remove the unheated section.
This places the temperature probe next to the
generator assembly.
If the infant is under a radiant warmer or bilirubin light,
the temperature probe should be covered with a light
reflective shield to prevent heating the probe. If the probe
is not covered, it could interfere with the operation of the
humidifier and cause excessive condensation to form.
Isolette or incubator
When the infant is in an isolette or incubator, the
non-heated section should be used with the temperature
probe placed outside of the isolette. Make sure that the
rest of the unheated section remains in the isolette.
If condensation is observed, remove the non-
heated (A) section and place the temperature probe
inside the isolette.
A
11
Infant Flow SiPAP sensor calibration
Infant Flow
SiPAP
Pres High
NCPAP/Pres Low
5
L/min
70
60
50
L/min
14
Infant Flow
SiPAP
Pres High
NCPAP/Pres Low
L/min
L/min
8
Infant Flow
SiPAP
Pres High
NCPAP/Pres Low
L/min
L/min
8
100
90
Infant Flow
SiPAP
Pres High
NCPAP/Pres Low
4
5
L/min
80
70
60
50
40
30
21
L/min
8
10
12
14
Two-point oxygen sensor calibration
Two-point oxygen sensor calibration should be performed
before initially using the Infant Flow SiPAP driver and with
each circuit change. To avoid unwanted alarms, occlude the
9 LPM prior to turning on the Infant Flow SiPAP driver on.
When the Infant Flow SiPAP driver is turned on, a power on
self-check automatically performs.
prongs or mask and set the low pressure flowmeter to
To perform two-point calibration:
1. Press the CAL button to enter the calibration menu.
2. Set the pressure low flowmeter to 9 LPM and the pressure high flowmeter to 2 to
3 LPM. Note: The pressure high flowmeter must be set during set-up to enable the
manual breath button.
3. Adjust the oxygen control to 21%. Allow time for the reading in the oxygen display
window to stabilize.
4. Press the flashing question mark button located under the 21% icon. The question
mark changes to a static hourglass. When calibration is complete, a static green
check mark icon appears and the oxygen display window reads 21%.
40
80
12
10
14
12
30
10
90
30
21
8
6
4
2
60
50
40
21
4
100
3
2
1
70
5
80
90
4
100
5. Adjust the oxygen control to 100%. Allow time for the reading in the oxygen display
window to stabilize.
6. Press the flashing question mark button located under the 100% icon. The question
mark changes to a static hourglass icon. When calibration is complete, the hourglass
icon changes to a static checkmark. The oxygen display window
reads 100%.
7. If oxygen calibration fails, a red X displays on the button of the screen, the alarm
sounds and an Error code displays on the top-left corner. Turn the driver off and
then, back on. Repeat the calibration procedure.
Disable the oxygen sensor
The internal oxygen sensor may be disabled by pressing the O2 disable button on the
calibration screen. This disables oxygen monitoring and the audible oxygen alarm. An
error code displays to indicate the oxygen monitor is inoperative. An external oxygen
monitor must be used whenever the oxygen sensor is disabled.
Leak test
While occluding the patient interface, set a flow of 9 LPM on the pressure low flowmeter.
A CPAP of 5 cmH
O ± 1 should display on the SiPAP screen. If pressure is not reached,
2
check the system for leaks. Release the occlusion, and the displayed pressure should be
≤ 2 cmH
does not fall, check the circuit for occlusions.
12
O. Wait for 15 seconds, and a disconnect alarm should sound. If the pressure
2
14
12
10
60
50
70
40
30
21
5
80
90
4
100
Infant Flow SiPAP set-up guide
Set-up menu screen
1. Adjust the pressure low flowmeter until the desired
nCPAP pressure displays on the screen. Press the flashing
question mark icon, which changes to a static checkmark
to confirm the setting.
2. Adjust the oxygen control dial to set the desire FiO
2
Press the flashing question mark icon. A checkmark
appears to confirm the setting.
3. Adjust the pressure high flowmeter until the pressure
displays 2 to 3 cmH
O above the set nCPAP pressure.
2
Press the flashing question mark icon. A static checkmark
appears to confirm the setting.
4. To use the low breathing rate/apnea monitor, connect
the transducer interface to the Infant Flow SiPAP driver.
Press the flashing question mark under the infant
respiratory sensor icon. This will change to a static
checkmark to confirm the setting. This does not confirm
that you want to use the respiratory monitoring option
but ensures that all modes are available for later use.
%.
Alarm set/conrm screen
Press the nCPAP button or alarm bar for three seconds
to set the alarm limits and move to the next screen. If no
button is pressed within two minutes, the alarm limits
automatically set and the screen changes to the mode
select screen.
Set-up menu screen
5. After completing the above steps, the screen changes
and displays the nCPAP mode. The infant can now be
connected to the Infant Flow SiPAP system.
Mode select screen
All available modes display at the bottom of the screen.
1. For low breath rate/apnea modes, attach the abdominal
respiratory sensor to the transducer and properly place
it on the infant’s abdomen.
2. Select the desired mode of operation by pressing the
corresponding button (mode select screen). The parameter
adjust screen displays, and the new mode displays in the
upper-left corner (parameter adjust screen).
Alarm set/confirm screen
3. Make the desired setting changes, and press the
selected mode to confirm the settings and activate
the new mode.
4. If no selection is made within two minutes and no
alarms sound, the screen locks to prevent entries. The
mode buttons go blank, except for the last button on
the right (locked screen). To unlock the screen, press
the lock icon.
Mode select screen
Parameter adjust screen
Locked screen
13
Infant Flow SiPAP set-up guide (continued)
18.0
16.0
Pressure (cmH
O)
Flow (LPM)
Parameter adjust screen
1. To change the settings during set-up and normal
operation, touch the desired parameter button.
2. Press the up or down arrows to adjust the parameter
to the desired setting.
3. Confirm the change by re-pressing the parameter button.
The main screen displays.
Note: nCPAP and BiPhasic pressure levels are set by
adjusting the flow.
Main screen and monitored parameter screen
1. To monitor therapy, use the main screen or
monitored parameter screen. The main screen
graphically displays the delivered pressure. The
monitored parameter screen displays numerical
values for the delivered pressure.
Parameter adjust screen
14
2. Press the Change Screen button to switch the
screen display.
Alarm reset/silence
1. Press the alarm bar to silence the active alarms
for 30 seconds.
2. Press the alarm bar for three seconds to clear resolved
and low-priority alarms and to reset alarm limits. Smart
alarm technology automatically sets high pressure, low
pressure and % oxygen thresholds.
14.0
12.0
10.0
2
8.0
6.0
4.0
2.0
4.06.08.010.012.014.016.0
Main screen
Monitor parameter screen
Main screen with active alarm
Flow pressure nomogram
The Infant Flow SiPAP LP system is subject to a direct
relationship between the controlled gas flow and airway
pressure. For example, 9 LPM of gas flow provides
approximately 5 cmH
O CPAP.
2
Tip:Manual breath: The high pressure flowmeter must be
set to deliver a manual sigh/breath during CPAP. The boost
in pressure delivers for the time high that was entered during
the set-up process.
Respiratory abdominal sensor
Respiratory abdominal sensor (optional)
For use only with the Infant Flow SiPAP Plus and
Comprehensive nCPAP drivers. The respiratory abdominal
sensor enables the clinician to monitor for apnea/low breath
rate in both nCPAP and BiPhasic modes. The accessories
include the reusable transducer and single-patient-use
abdominal sensor. In the BiPhasic trigger mode, the
respiratory abdominal sensor and transducer allow
patient-triggered pressure assists with breath rate
monitoring (not available in the U.S.).
Respiratory transducer connection
1. Connect the transducer cable to the transducer port
on the front panel.
2. Connect the abdominal sensor pressure line to the
transducer interface.
3. Compress the sensor pad gently, repeating this several
times while observing the transducer LED.
Infant set-up
To apply the sensor to the infant using suitable tape
(figure 1):
1. Visually identify the optimum outward movement of
the abdomen during inspiration. When the infant
breathes, the most movement is between the lowest
rib and the abdomen.
2. If the infant is supine, place the capsule midway
between the umbilicus and xiphisternum, which is the
notch at the center of the two lower ribs. On larger
infants, an alternative site is the upper chest to detect
intercostal movement.
3. If the infant is prone, place the sensor laterally over the
lower rib and abdomen. The sensor tubing should be
directed over the back.
4. Tape the sensor firmly into position using a non-
allergenic microprobe tape. Position the sensor line
perpendicular to the tape. Only use tape that is
approved by your facility’s protocol.
5. Verify correct placement. The transducer LED should
illuminate on expiration, and the SiPAP front panel LED
on inspiration.
6. If the LED does not illuminate, try repositioning the
sensor and adding a second piece of tape making an
“X” over the sensor.
Figure 1: Abdominal sensor placement
15
Self assessment
1. Where should the circuit airway temperature probe be
placed if the infant is in an isolette/incubator?
2. If condensation occurs in the breathing circuit, what
should you do?
3. Explain how to disable the oxygen sensor:
4. Demonstrate how to reset the alarm limits when the
device is in operation:
7. While in nCPAP, you press the manual breath button,
but nothing happens. What would prevent a manual
breath from being delivered?
8. When should you perform an oxygen sensor
calibration on the Infant Flow SiPAP?
Notes:
5. To deliver a CPAP of 5 cmH
6. Where is the best placement for the respiratory
abdominal sensor?
16
O, what would the ow rate be?
2
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.