CareFusion Infant Flow Training Workbook

Infant Flow® LP nCPAP system
Clinical training workbook
Table of contents
Infant nasal CPAP ..............................................................................................1– 6
CPAP over view .....................................................................................................1–3
CPAP modalities .......................................................................................................4
Variable flow technology ..........................................................................................5
Self assessment ........................................................................................................6
Infant Flow
Infant Flow LP system ...............................................................................................7
Infant Flow SiPAP configurations ..............................................................................8
Infant Flow SiPAP display screen ............................................................................... 9
Infant Flow CPAP and circuit set-up ........................................................................ 10
Humidification and nCPAP ......................................................................................11
Airway temperature probe placement ................................................................... 11
Infant Flow SiPAP sensor calibration .......................................................................12
Infant Flow SiPAP set-up guide ......................................................................... 13 –14
Respiratory abdominal sensor ................................................................................. 15
Self assessment ......................................................................................................16
Infant Flow SiPAP nCPAP driver ................................................................... 17–22
Modes of operation .......................................................................................... 17–18
BiPhasic mode strategy ..................................................................................... 19 –20
SiPAP exercises and self assessment.................................................................. 21–22
®
SiPAP driver overview ............................................................... 7–16
Infant Flow LP generator assembly ............................................................. 23–27
Infant Flow LP generator ........................................................................................23
Infant Flow LP interfaces ................................................................................... 24–25
Infant Flow LP fixation devices ................................................................................ 26
Self assessment ......................................................................................................27
Infant Flow LP patient set-up...................................................................... 28– 46
Infant Flow LP interfaces ................................................................................... 28 –29
Fixation devices ......................................................................................................30
Headgear application ............................................................................................. 31
Generator assembly preparation .............................................................................32
Generator assembly and interface attachment to headgear ............................. 33–34
Bonnet application .................................................................................................35
Generator assembly and interface attachment to bonnet ................................. 36 –37
Bonnet application (alternative method 1) .............................................................. 38
Bonnet application (alternative method 2) ..............................................................39
Incorrect application of fixation device and generator assembly ....................... 40 – 41
Incorrect application of generator assembly and interface ................................ 42– 43
Final inspection of nasal interface placement .........................................................44
Self assessment and return demonstration ...................................................... 45–46
Routine nCPAP care ....................................................................................... 47– 49
Initiating and maintaining
effective nCPAP therapy
is a critical step in helping
respiratory- compromised
function. When used according to your facility’s treatment protocols and with this training workbook, the Infant Flow
LP nCPAP system can effectively deliver nCPAP therapy to help improve patient outcomes.
Frequently asked questions .........................................................................50–54
Self assessment ......................................................................................................54
Glossary ..........................................................................................................55 –56
References ............................................................................................................57
Infant nasal CPAP
Introduction
Worldwide each year, approximately 15 million (1 out of every 10) babies are born prematurely.1 Premature or
low-birth weight (LBW) infants are at a high risk for respiratory problems due to underdeveloped lungs. Common
neonatal respiratory conditions include apnea of prematurity, respiratory distress syndrome, transient tachypnea
of the newborn (TTN), meconium aspiration syndrome, pulmonary edema and post-extubation support. These
1,2
conditions are often associated with decreased pulmonary compliance and functional residual capacity (FRC).
Several of these infants will require respiratory support.
Respiratory distress syndrome (RDS) is a condition that strains normal respiration due to the lack of natural
surfactant production. Approximately 50% of neonates born at 26 to 28 weeks gestation and 30% of neonates
born at 30 to 31 weeks gestation develop RDS.
2
CPAP overview
What is surfactant?
Surfactant is a phospholipid, which reduces surface tension
to increase lung compliance.
Artificial surfactant may be given to help reduce surface
tension, increase compliance and improve ventilation.
Without additional respiratory assistance, many infants have
difficulty establishing the adequate functional residual
capacity (FRC) required to maintain normal respiration.
Respiratory support
Several options are available to help the clinician provide
respiratory support to the neonatal patient. Historically, the
initial treatment for infants with respiratory problems was
mechanical ventilation via an artificial airway. Intubation
presents a variety of challenges for any patient but
compounds problems with premature infants. Given the
potential complications of intubation, many physicians opt
for a less invasive approach for spontaneously breathing
infants that utilizes continuous positive airway pressure
(CPAP). As infants are preferential nose-breathers, nasal
CPAP (nCPAP) is the preferred method for treatment
delivery. CPAP enhances alveolar recruitment decreasing
pulmonary vascular resistance and intrapulmonary shunting,
stabilizes FRC and improves oxygenation. By increasing
surface area to alveolar gas exchange, CPAP decreases V/Q
mismatch. The goal of CPAP therapy is to maintain normal
lung volumes and oxygenation, while enabling the infant
3,4
to breathe on their own.
Physiologic effects of CPAP are
represented in the organizational chart on page 2.
1
Physiologic effects of nasal CPAP in neonates
3–4,10
Infant nasal CPAP
Splint open airways
Recruits alveoli
and prevents
alveoli collapse
Increases FRC
and lung volumes
Improves V/Q ratio
and increases
oxygenation
Decreases WOB
Nasal (nCPAP) is associated with improved respiratory mechanics and decreased chronic lung disease (CLD) rate.
Conserves surfactant
Maintains airway
Increases pharyngeal
cross section
patency
Decreases
obstructive apnea
Reduces upper
airway resistance
Stabilizes chest wall
and diaphragm
Improves
breathing pattern
and decreases WOB
Decreases
intrapulmonary
shunting
Improves V/Q ratio
Stretches lung
Stimulates the
J receptors and HIBR
Reduces central
and
obstructive apnea
and pleura
Stimulates lung
growth
What is nasal CPAP (nCPAP)?
nCPAP is the application of positive pressure to
the airways of a spontaneously breathing infant
throughout the respiratory cycle. nCPAP is a continuous
flow of gas administered through nasal prongs inserted
in the nares or by a nasal mask placed around the
perimeter of the nose. The positive pressure, usually
O to 8 cmH2O, acts as a splint, which can help
4 cmH
2
prevent alveoli collapse.
BiPhasic CPAP alternates between two levels of CPAP
at a set time interval. The infant can breathe at both
CPAP settings. The BiPhasic mode helps increase the
infant’s tidal volume and may stimulate the respiratory
drive center.
2
Advantages of CPAP
• Increases FRC
• Maintains and increases lung volume
• Improves lung compliance
• Reduces work of breathing (WOB) and
airway resistance
• Provides a noninvasive procedure
• Allows small airways to develop
• Promotes the use of natural surfactant
• Promotes easy application
• Provides cost effectiveness
• Helps prevent extubation failure in some infants
• Stabilizes the airway diaphragm and chest wall
• Decreases incidence of chronic lung disease (CLD)
Indications for use
2–5
• Abnormalities on physical examination
- Increased WOB
- Increased respiratory rate
- Intercostal and substernal recession
- Grunting and nasal aring
- Pale skin color
- Restlessness
• Deteriorating arterial/capillary blood gas values
(e.g., hypercapnea)
• Increased oxygen requirements to maintain a SaO
greater than 92% with FiO
> 60%
2
• Atelectasis and inltration
• Clinical conditions
- Apnea of prematurity
- Chest infections (e.g., pneumonia)
- Transient tachypnea of the newborn (TTN)
- Mild meconium aspiration
• Weaning/Post-extubation support
• Congenital malformations of the upper airway (cleft
palate, choanal atresia or tracheoesophageal stula)
• Congenital diaphragmatic hernia or untreated
bowel obstruction
• Poor respiratory drive unresponsive to CPAP therapy
(frequent apnea episodes associated with oxygen
desaturation and/or bradycardia)
What is work of breathing?
WOB describes the amount of effort required to breathe.
2
Any therapy that introduces incoming pressure to a
patient’s respiratory system potentially adds imposed
WOB. Infants with RDS experience elevated WOB levels,
and by expending additional effort to inhale and exhale
against pressurized gas, the infant consumes precious
calories overcoming the high WOB level. These calories
could otherwise be spent on vital recovery and growth
processes. In addition to helping the infant conserve energy,
a WOB reduction may reduce stress and anxiety levels.
Contraindications for use
2–5
• Severe cardiovascular instability
• Respiratory failure dened as pH < 7.25 and
> 60 mmHg torr
PaCO
2
Potential problems associated with CPAP therapy
3,5,6
Clinicians should be aware of the possible hazards and
complications associated with CPAP, and take the
necessary precautions to ensure safe and effective
applications, such as:
• Possible loss of prescribed pressure and decreased FiO
2
due to mouth breathing
• Increased intrathoracic pressure reducing venous return,
which may lower cardiac output
• Barotrauma leading to surgical emphysema/
pneumothoraces
• Aspiration
• Deterioration in the respiratory condition, requiring
immediate ventilation
• Patient discomfort from prong/mask intolerance
• Nasal septal injury (e.g., columella necrosis)
• Blanching of the nares
• Dry mouth and airways
• Gastric ination
Columella necrosis
Nasal dilation
3
CPAP modalities
What are the treatment options?
A variety of technologies have been employed in nCPAP
delivery throughout the years.
Conventional CPAP (V-CPAP): Utilizes a traditional mechanical
ventilator to deliver a constant flow of gas. CPAP is created
by changing the expiratory port orifice size. The ventilator
equipment is comprehensive and expensive.
Bubble CPAP (B-CPAP): Utilizes a constant flow of heated
and humidified gas. The level of pressure is controlled by
the depth of the exhalation tube inserted into a water
container. The pressure can increase if condensate collects
in the tubing, the flow rate changes or the water evaporates
from the container. B-CPAP lacks system alarms and
imposes a higher WOB
inability to entrain flow during inspiration.
due to the constant flow and
(16 ,18 ,19)
High flow nasal cannula (HFNC): Has not been cleared by
the FDA for nasal CPAP delivery. HFNC utilizes a constant
flow of heated, humidified gas that potentially delivers a
positive distending pressure. The level of therapy cannot
be measured and fluctuates depending on body position,
oral leaks, nasal secretions and the size and weight of the
patient. HFNC does not contain critical alarms that ensure
the safe delivery of therapy.
Variable flow CPAP (VF-CPAP): Incorporates a generator that
redirects the heated and humidified gas flow away from the
patient during exhalation and allows air entrainment during
periods of high inspiratory effort.
provides the most stable pressure, even in the presence of
leaks up to 6 LPM.
Pandt and Associates demonstrated that the Infant Flow
variable flow technology delivered a consistent level of
CPAP with little fluctuations. In contrast, the conventional
CPAP did not reach the desired level of 5 cmH
O, and
2
the pressure fluctuated significantly throughout the
breath cycle.
9
Using a variable flow generator with a dedicated CPAP
driver provides a measurable therapy with system alarms
to help ensure safe and effective therapy.
Infant Flow
8 Lts flow, 5 cmH2O
System pressure
5 cmH2O
0 cmH
O
2
Time
Conventional CPAP
17 Lts flow, 5 cmH2O with 5 Lts reservoir bag
System pressure
5 cmH2O
By redirecting the gas, VF-CPAP offers a lower imposed
WOB and less expiratory resistance compared to other
nCPAP technologies. Because the pressure is created and
measured at the nares, the variable flow technology
Infant Flow variable ow
Inspiration: Gas flow converted
to pressure reducing the WOB
and maximizing the pressure
stability at the patient interface.
4
0 cmH
O
2
Moa, G., Nilsson, K. et al. Crit Care Med, 1988 , 16(12):1238 –1242 .
Expiration: Gas flow
flipped away from the nasal
prongs to the expiratory
tube. The residual gas
pressure provided by the
continuous gas flow creates
a stable CPAP throughout
the respiratory cycle.
Time
Variable ow technology
Inhalation
Dual jets
Flow direction Flow direction
Patient nare
What is variable ow technology?
10,11
The Infant Flow LP patented dual-jet variable flow
generator utilizes fluidic technology to deliver a
constant CPAP at the airway proximal to the infant’s
nares. Without moving parts or valves, the generator
provides consistent performance. The level of CPAP
created is proportional to the flow provided by the
driver; for example, 9 LPM creates approximately
5 cmH
O CPAP. The variable flow generator uses
2
Bernoulli’s Principle via injector jets directed toward
each nare. If the infant pulls additional flow, the
venturi action of the injector jets entrains additional
flow from either the source gas or exhalation tube
reservoir. During exhalation, the incoming gas flow
redirects away from the infant. This action is referred
to as the “fluidic flip.” By redirecting the gas, variable
flow nCPAP reduces the imposed WOB. The infant
can exhale freely and conserve precious calories for
development. In summary, the direction of gas flow
in variable flow devices depends on the patient’s
respiratory cycle. The flow “flips” away from the
nares when the infant exhales and then, “flips” back
as the exhalation phase ends. The response is almost
instantaneous as it occurs at the patient’s nares.
Fluidic flip
Exhalation
Exhaust tube
Vortice shedding
Patient nare
What is vortices technology?
10,12
The patented Infant Flow LP generator is a new form of
variable ow that uses vortices technology to reduce the
imposed WOB during inhalation. Similar to the single- jet
technology, the ow entrainment reduces the WOB
on inhalation by meeting the patient’s inspiratory ow
demand and during exhalation gas ow ips away from
the patient reducing resistance.
Four low-momentum jets (two per nare) impinge inside
the generator to create a consistent and measurable
positive airway pressure within the generator head.
During inhalation, the dual jets entrain ow to meet
the patient’s inspiratory demand. During exhalation, the
jets easily deect to disrupt the gas ow. This disruption
of ow creates vortice shedding that spirals outwardly,
combining with the exhaled breath to create an
organized, efcient ow path toward the exhaust ports.
Exhaust tube
Impinging jets
Drive line
Pressure line
Patient
5
Self assessment
1. Describe RDS:
2. List three indications for nCPAP therapy:
3. List three benets of nCPAP therapy:
4. List three potential complications to nCPAP therapy:
8. Match the generator parts to the diagram:
Pressure line
Impinging jets
Exhaust tube
Drive line
Patient
c
d
b
a
e
Notes:
5. State four methods used to deliver nCPAP therapy:
6. Discuss the advantage of variable ow technology compared
to other CPAP modalities:
7. Discuss the importance of low work of breathing:
6
Infant Flow SiPAP driver overview
Infant Flow LP system
The Infant Flow LP nCPAP system is a comprehensive system for delivering unique nCPAP therapy.
The system consists of:
• Infant Flow SiPAP driver
• Infant Flow LP generator assembly
• Infant Flow LP fixation—bonnet or headgear
• Infant Flow LP nasal interfaces—mask or prongs
This chapter discusses the set-up and operation of the Infant Flow SiPAP Plus driver.
Power LED
Indicates power on and AC connection
Touch screen display
Adjusts settings and displays patient parameters and alarms.
Pressure low flow meter
Adjusts low CPAP level. CPAP supplied to patient directly relates to flow rate from the nCPAP driver.
Alarm LED
Indicates alarm situation as visual and audible notification color varies according to alarm level
Note: Refer to the Infant Flow SiPAP operator manual for additional instructions on the set-up, operation and maintenance of the Infant Flow SiPAP Plus driver.
Transducer LED
Indicates transducer connected.
% O
blender control
2
Adjusts FiO
Pressure high flow meter
Adjusts high CPAP level. CPAP supplied to patient directly relates to flow rate from nCPAP driver.
2.
Transducer connection
Connects to transducer interface.
Patient circuit connection
Connects patient breathing circuit to gas outlet port.
Pressure line connection
Connects patient pressure line to device.
7
Infant Flow SiPAP congurations
The Infant Flow SiPAP driver is sold globally and is available
in different configurations. The two main models are Infant
Flow SiPAP Plus and Infant Flow SiPAP Comprehensive.
The Comprehensive model offers an additional ventilation
mode, BiPhasic trigger, which is not available in the U.S.
In select areas, additional languages or an international icon
overlay may be used in place of the English text. The operation
and maintenance of the Infant Flow SiPAP driver is the
same regardless of the specific configurations. Refer to the
Infant Flow SiPAP operator manual for more specific details.
Description English text ICON symbol
Press to access the user calibration menu and language options.*
Press to return to the start-up screen.
Press to switch between the graphical and numerical monitoring screen.
Press to change the operation mode.
Press to deliver a manual breath. The breath delivers at the pressure high setting for
the set time high duration.
Indicates user should refer to the operator manual for additional information.
Indicates battery status and turns red if the battery charge is less than 40%.
Indicates the screen is locked. Press to unlock the screen.
Adjust the low flow rate setting for the baseline CPAP level.
Adjust the high flow rate setting for BiPhasic high CPAP level.
*Language option not available on all SiPAP models.
8
Infant Flow SiPAP display screen
Mode, control settings and function buttons
Mode indicator
Monitored parameters
FiO
Alarm priority/alarm silence indicator
2
Operational information
Battery charge
Pressure/ Time graphics or monitored parameters display
Control/Setting indicators
Soft key color code
White letter
Key enabled
Faded letter
Key inactive
Yellow letter
Solid: Pending confirmation
Flashing: Low-priority alarm
Red letter
Flashing: High-priority alarm
Solid: Reduction in another parameter
caused by an adjustment
Alarm management
High priority
• Series of 10 tones sound every 10 seconds
• Parameters display, and limits flash red
Medium priority
• 3 audible tones sound every 15 seconds
• Parameters display, and limits flash yellow
Low priority
• 2 audible tones sound every 30 seconds
• Parameters display, and limits change to yellow
9
Infant Flow SiPAP and circuit set-up
Circuit set-up
1. Gather the nCPAP supplies:
• Infant Flow SiPAP driver
• Single-limb, heated breathing circuit
• Infant Flow LP generator kit
• Infant Flow LP xation device
• Humidier and chamber
• Sterile water bag
2. Attach the water chamber to the humidier and
connect it to the water feed system. Follow the
manufacturer instructions for the proper set-up.
3. Connect the gas delivery tubing (A) to the ow
driver outlet port (I) and humidier chamber port (II).
4. Connect the elbow connector on the heated
breathing circuit (B) to the humidier chamber.
Insert the heater wire plug into the wire socket.
Securely insert the temperature probe in the port on
the circuit elbow (III). Insert the second temperature
probe (IV) into the airway port at the distal end of
the breathing circuit.
5. Connect the non-heated section (C) to the drive line
of the generator assembly (D).
I
A
II
B
V
III
E
6. Connect the proximal pressure line (E) to the
proximal port on the driver (V) and the pressure
line on the generator.
Temperature probe
When inserting the temperature probe into the
circuit, ensure the probe tip is in the middle of the
gas stream. This allows the gas temperature to be
measured accurately. If the probe is not properly
seated, the temperature measurement accuracy may
be compromised, leading to excessive condensation.
Cover the temperature probe with a reective shield
when used under a radiant warmer or bilirubin light.
IV
C
D
10
Humidication and nCPAP
Humidication
Heated humidification is recommended for nCPAP therapy.
The normal functions of the nose and air passages of
the respiratory tract are too warm, moisten and filter
the inhaled gases before they reach the lungs. In normal
respiration, the nasal mucosa and upper airways provide
75% of the heat and moisture supplied to the smaller
airways and alveoli. By the time air reaches the alveoli,
the inspired gas warms to 37 °C at 100% relative humidity
13
With nCPAP, the upper airways are not bypassed,
(RH).
but the high gas flows may be drying to the airways,
especially to a neonate’s underdeveloped lung. Adequate
humidification is essential to maintain airway clearance,
optimize ventilation and improve patient comfort.
The International Organization for Standardization (ISO) and
American Association for Respiratory Care (AARC) clinical
practice guidelines recommend gas temperature
between 34 and 41 °C to provide a humidity level of
33 to 44 mgH
O/L with artificial airways.14 Be cautious
2
using higher temperatures, as condensation may reduce
the mucous viscosity and interfere with the mucous
clearance. Extended exposure of gas temperatures over
41 °C may cause cellular damage to the airways.
14
The higher temperature settings may not be required to
deliver adequate humidification, since an artificial airway
is not used with nCPAP. Start with a temperature setting
of 36 °C to 37 °C and adjust the humidifier settings to
maintain adequate humidification; if condensate occurs,
reduce the humidifier temperature setting.
Airway temperature probe placement
Open bed or crib
When the infant is placed on an open bed warmer or
crib, it is recommended to remove the unheated section.
This places the temperature probe next to the
generator assembly.
If the infant is under a radiant warmer or bilirubin light,
the temperature probe should be covered with a light
reflective shield to prevent heating the probe. If the probe
is not covered, it could interfere with the operation of the
humidifier and cause excessive condensation to form.
Isolette or incubator
When the infant is in an isolette or incubator, the
non-heated section should be used with the temperature
probe placed outside of the isolette. Make sure that the
rest of the unheated section remains in the isolette.
If condensation is observed, remove the non-
heated (A) section and place the temperature probe
inside the isolette.
A
11
Infant Flow SiPAP sensor calibration
Infant Flow
SiPAP
Pres High
NCPAP/ Pres Low
5
L/min
70
60
50
L/min
14
Infant Flow
SiPAP
Pres High
NCPAP/ Pres Low
L/min
L/min
8
Infant Flow
SiPAP
Pres High
NCPAP/ Pres Low
L/min
L/min
8
100
90
Infant Flow
SiPAP
Pres High
NCPAP/ Pres Low
4
5
L/min
80
70
60
50
40
30
21
L/min
8
10
12
14
Two-point oxygen sensor calibration
Two-point oxygen sensor calibration should be performed
before initially using the Infant Flow SiPAP driver and with
each circuit change. To avoid unwanted alarms, occlude the
9 LPM prior to turning on the Infant Flow SiPAP driver on.
When the Infant Flow SiPAP driver is turned on, a power on
self-check automatically performs.
prongs or mask and set the low pressure flowmeter to
To perform two-point calibration:
1. Press the CAL button to enter the calibration menu.
2. Set the pressure low flowmeter to 9 LPM and the pressure high flowmeter to 2 to
3 LPM. Note: The pressure high flowmeter must be set during set-up to enable the
manual breath button.
3. Adjust the oxygen control to 21%. Allow time for the reading in the oxygen display
window to stabilize.
4. Press the flashing question mark button located under the 21% icon. The question
mark changes to a static hourglass. When calibration is complete, a static green
check mark icon appears and the oxygen display window reads 21%.
40
80
12
10
14
12
30
10
90
30
21
8
6
4
2
60
50
40
21
4
100
3
2
1
70
5
80
90
4
100
5. Adjust the oxygen control to 100%. Allow time for the reading in the oxygen display
window to stabilize.
6. Press the flashing question mark button located under the 100% icon. The question
mark changes to a static hourglass icon. When calibration is complete, the hourglass
icon changes to a static checkmark. The oxygen display window
reads 100%.
7. If oxygen calibration fails, a red X displays on the button of the screen, the alarm
sounds and an Error code displays on the top-left corner. Turn the driver off and
then, back on. Repeat the calibration procedure.
Disable the oxygen sensor
The internal oxygen sensor may be disabled by pressing the O2 disable button on the
calibration screen. This disables oxygen monitoring and the audible oxygen alarm. An
error code displays to indicate the oxygen monitor is inoperative. An external oxygen
monitor must be used whenever the oxygen sensor is disabled.
Leak test
While occluding the patient interface, set a flow of 9 LPM on the pressure low flowmeter.
A CPAP of 5 cmH
O ± 1 should display on the SiPAP screen. If pressure is not reached,
2
check the system for leaks. Release the occlusion, and the displayed pressure should be
≤ 2 cmH
does not fall, check the circuit for occlusions.
12
O. Wait for 15 seconds, and a disconnect alarm should sound. If the pressure
2
14
12
10
60
50
70
40
30
21
5
80
90
4
100
Infant Flow SiPAP set-up guide
Set-up menu screen
1. Adjust the pressure low flowmeter until the desired
nCPAP pressure displays on the screen. Press the flashing
question mark icon, which changes to a static checkmark
to confirm the setting.
2. Adjust the oxygen control dial to set the desire FiO
2
Press the flashing question mark icon. A checkmark
appears to confirm the setting.
3. Adjust the pressure high flowmeter until the pressure
displays 2 to 3 cmH
O above the set nCPAP pressure.
2
Press the flashing question mark icon. A static checkmark
appears to confirm the setting.
4. To use the low breathing rate/apnea monitor, connect
the transducer interface to the Infant Flow SiPAP driver.
Press the flashing question mark under the infant
respiratory sensor icon. This will change to a static
checkmark to confirm the setting. This does not confirm
that you want to use the respiratory monitoring option
but ensures that all modes are available for later use.
%.
Alarm set/conrm screen
Press the nCPAP button or alarm bar for three seconds
to set the alarm limits and move to the next screen. If no
button is pressed within two minutes, the alarm limits
automatically set and the screen changes to the mode
select screen.
Set-up menu screen
5. After completing the above steps, the screen changes
and displays the nCPAP mode. The infant can now be
connected to the Infant Flow SiPAP system.
Mode select screen
All available modes display at the bottom of the screen.
1. For low breath rate/apnea modes, attach the abdominal
respiratory sensor to the transducer and properly place
it on the infant’s abdomen.
2. Select the desired mode of operation by pressing the
corresponding button (mode select screen). The parameter
adjust screen displays, and the new mode displays in the
upper-left corner (parameter adjust screen).
Alarm set/confirm screen
3. Make the desired setting changes, and press the
selected mode to confirm the settings and activate
the new mode.
4. If no selection is made within two minutes and no
alarms sound, the screen locks to prevent entries. The
mode buttons go blank, except for the last button on
the right (locked screen). To unlock the screen, press
the lock icon.
Mode select screen
Parameter adjust screen
Locked screen
13
Infant Flow SiPAP set-up guide (continued)
18.0
16.0
Pressure (cmH
O)
Flow (LPM)
Parameter adjust screen
1. To change the settings during set-up and normal
operation, touch the desired parameter button.
2. Press the up or down arrows to adjust the parameter
to the desired setting.
3. Confirm the change by re-pressing the parameter button.
The main screen displays.
Note: nCPAP and BiPhasic pressure levels are set by
adjusting the flow.
Main screen and monitored parameter screen
1. To monitor therapy, use the main screen or
monitored parameter screen. The main screen
graphically displays the delivered pressure. The
monitored parameter screen displays numerical
values for the delivered pressure.
Parameter adjust screen
14
2. Press the Change Screen button to switch the
screen display.
Alarm reset/silence
1. Press the alarm bar to silence the active alarms
for 30 seconds.
2. Press the alarm bar for three seconds to clear resolved
and low-priority alarms and to reset alarm limits. Smart
alarm technology automatically sets high pressure, low
pressure and % oxygen thresholds.
14.0
12.0
10.0
2
8.0
6.0
4.0
2.0
4.0 6.0 8.0 10.0 12.0 14.0 16.0
Main screen
Monitor parameter screen
Main screen with active alarm
Flow pressure nomogram
The Infant Flow SiPAP LP system is subject to a direct
relationship between the controlled gas flow and airway
pressure. For example, 9 LPM of gas flow provides
approximately 5 cmH
O CPAP.
2
Tip: Manual breath: The high pressure flowmeter must be
set to deliver a manual sigh/breath during CPAP. The boost
in pressure delivers for the time high that was entered during
the set-up process.
Respiratory abdominal sensor
Respiratory abdominal sensor (optional)
For use only with the Infant Flow SiPAP Plus and
Comprehensive nCPAP drivers. The respiratory abdominal
sensor enables the clinician to monitor for apnea/low breath
rate in both nCPAP and BiPhasic modes. The accessories
include the reusable transducer and single-patient-use
abdominal sensor. In the BiPhasic trigger mode, the
respiratory abdominal sensor and transducer allow
patient-triggered pressure assists with breath rate
monitoring (not available in the U.S.).
Respiratory transducer connection
1. Connect the transducer cable to the transducer port
on the front panel.
2. Connect the abdominal sensor pressure line to the
transducer interface.
3. Compress the sensor pad gently, repeating this several
times while observing the transducer LED.
Infant set-up
To apply the sensor to the infant using suitable tape
(figure 1):
1. Visually identify the optimum outward movement of
the abdomen during inspiration. When the infant
breathes, the most movement is between the lowest
rib and the abdomen.
2. If the infant is supine, place the capsule midway
between the umbilicus and xiphisternum, which is the
notch at the center of the two lower ribs. On larger
infants, an alternative site is the upper chest to detect
intercostal movement.
3. If the infant is prone, place the sensor laterally over the
lower rib and abdomen. The sensor tubing should be
directed over the back.
4. Tape the sensor firmly into position using a non-
allergenic microprobe tape. Position the sensor line
perpendicular to the tape. Only use tape that is
approved by your facility’s protocol.
5. Verify correct placement. The transducer LED should
illuminate on expiration, and the SiPAP front panel LED
on inspiration.
6. If the LED does not illuminate, try repositioning the
sensor and adding a second piece of tape making an
“X” over the sensor.
Figure 1: Abdominal sensor placement
15
Self assessment
1. Where should the circuit airway temperature probe be
placed if the infant is in an isolette/incubator?
2. If condensation occurs in the breathing circuit, what
should you do?
3. Explain how to disable the oxygen sensor:
4. Demonstrate how to reset the alarm limits when the
device is in operation:
7. While in nCPAP, you press the manual breath button,
but nothing happens. What would prevent a manual
breath from being delivered?
8. When should you perform an oxygen sensor
calibration on the Infant Flow SiPAP?
Notes:
5. To deliver a CPAP of 5 cmH
6. Where is the best placement for the respiratory
abdominal sensor?
16
O, what would the ow rate be?
2
Loading...
+ 42 hidden pages