Alaris™ Gateway Workstation v1.2
Introduction
Introduction
The Alaris™ Gateway Workstation v1.2 (hereinafter referred to as Workstation) has been designed as a modular system to provide a
communications gateway between an Alaris Infusion Pump (hereinafter referred to as Pump) and any Patient Data Management System
(PDMS), Patient Monitoring (PM) System, Hospital Information System (HIS) or Clinical Information System (CIS) that requires access to
the infusion data retained within the Pump.
Workstation Hardware Options
There are three different configuration options for the Workstation. To identify its configuration, refer to the labels on the front and back
of the Workstation.
Option 1. Basic or Auxiliary Workstation
REF
80203UNS01-32
SN:
13507025317.04.2014
This is an entry level Workstation with organization and central AC power supply functionality. It does not support PDMS connectivity. It
may be used as an Auxiliary Workstation to increase the number of Pumps needed at the bedside.
T
Option 2. Primary wired Workstation
REF
80203UNS02-32
SN:
13507025317.04.2014
This option is designed to accommodate connectivity to third party systems via a hospital wired network infrastructure. A Primary wired
Workstation can be coupled with one Auxiliary Workstation to send infusion data for up to 22 Pumps to third party systems.
T
Option 3. Primary wireless Workstation
REF
80203UNS03-32
SN:
13507025317.04.2014
This option is designed to accommodate connectivity to third party systems via a hospital wired or wireless network infrastructure. A
Primary wireless Workstation can be coupled with one Auxiliary Workstation to send infusion data for up to 22 Pumps to third party
systems.
1000DF00742 Issue 2
3/30
T
Alaris™ Gateway Workstation v1.2
Introduction
Workstation Software Options
There are different software versions that can be ordered for each Primary Workstation.
The software version of Workstation can be identified by the labelling on the front. Please contact local CareFusion representative for
further information about Workstation versions, upgrade options and product availability.
This software version enables connectivity to third party systems
and meets the latest compliance standards.
This software version enables connectivity to third party systems.
®
Alaris
Gateway Workstation
The Alaris DS Docking Station, although similar in appearance to the
A
Workstation, does not have any connectivity features and does not
work with the Alaris Communication Engine. The Alaris DS Docking
Station can be identified by the number of vertical MDI tiles, as it
has an even number compared to the Workstation which has an odd
number and that the AC power Cable socket is at the front.
Note: Please make sure to refer to the correct Workstation Directions For Use to familiarise with the specific Workstation version.
Features:
• Central management system for multiple Pumps
• Medical Device Interface (MDI) – a unique mounting mechanism providing data communications and AC power to the Pump
• Reduced cable clutter with the use of a single AC power inlet
• Efficient organisation of multiple infusion lines and configurations
• Battery back-up in the event of power supply interruption
• High visibility Pump Alarm Location Beacon assists with the location of Pumps in an alarm state, when fitted
• Nurse call interface for all Pumps attached to the Workstation MDI Tiles
The Workstation supports optional upgrades to enhance the data communication interfaces and to support software for connections to
such client / server systems.
1000DF00742 Issue 2
4/30
Alaris™ Gateway Workstation v1.2
Introduction
Intended Use
The Alaris Gateway Workstation v1.2 is intended to be used to provide mounting, power and communications support to the Alaris
Infusion Pumps range within the operating environment specified in this Directions For Use (DFU). In such environments, the
Workstation may be exposed to the following range of therapies:
Fluid therapy, blood transfusions, parenteral feeding, drug therapy, chemotherapy, dialysis and anaesthesia. The Workstation is designed
not to directly impact or affect the infusion delivery process.
The Workstation is designed to work with the following Alaris Infusion Pumps:
Alaris CC Guardrails Syringe Pump80033UND1-G v3.1.4YesYesYesYes
Alaris GH Syringe Pump (with Plus Software) MK38002MED01v4.1.4YesYe sYesYes
Alaris CC Syringe Pump (with Plus Software) MK38003MED01v4.1.4YesYe sYesYes
Alaris GH Guardrails Syringe Pump (with Plus Software) MK38002MED01-G v4.1.4YesYesYe sYe s
Alaris CC Guardrails Syringe Pump (with Plus Software) MK38003MED01-G v4.1.4YesYesYe sYe s
Alaris GH Syringe Pump (with Plus Software) MK48002MED01
8002TIG01
Alaris CC Syringe Pump (with Plus Software) MK48003MED01
8003TIG01
Alaris GH Guardrails Syringe Pump (with Plus Software) MK48002MED01-G
8002TIG01-G
Alaris CC Guardrails Syringe Pump (with Plus Software) MK48003MED01-G
8003TIG01-G
Alaris GW Volumetric Pump2504xxxx1*v5r1F
Alaris GP Volumetric Pump80263UN01v1.7.18YesYes***YesYes
Alaris GP Guardrails Volumetric Pump80263UN01-Gv1.9.2YesYesYesYes
Alaris GP Volumetric Pump (with Plus Software)9002MED01
9002TIG01
Alaris GP Guardrails Volumetric Pump (with Plus Software)9002MED01-G
9002TIG01-G
Alaris VP Plus Guardrails Volumetric Pump9003MED01-G
9003TIG01-G
Alaris SE Pump7131xxxxxx*
7231xxxxxx*
v4.1.8YesYesYesYes
v4.1.8YesYesYesYes
v4.1.8YesYesYesYes
v4.1.8YesYesYesYes
(v51F) **
v5r2A
(v52A) **
v2.1.14
v2.1.15
v2.1.14
v2.1.15
v1.1.28YesYesYesYes
v2.79
v2.80
v4.54
v8.53
BeaconNurse Call
High
(red)
YesYesYesYes
YesNoYesNo
YesYesYesYes
YesYesYesYes
YesYesYesYes
Medium
(amber)
HighMedium
* xxxx and xxxxxx denotes the language and country specific model codes
** The display on the Alaris GW Volumetric Pump can only show four characters so software version will be shown without the letter r
*** The medium (amber) level alarm will not be triggered for a Rate Lock alarm for the Alaris GP Volumetric Pump
Note: The single and dual channel Alaris SE Pump can be connected to the serial interface of the Workstation however can not be
fitted onto the MDI tile.
1000DF00742 Issue 2
5/30
Alaris™ Gateway Workstation v1.2
Introduction
The pumps listed above, except the Alaris GW Volumetric Pump, will activate the Pump Alarm Location Beacon and
A
To access and configure the software installed on the Workstation, use a standard Web browser and an Ethernet connection either over
a network or by directly linking to the Workstation from a client PC.
The Workstation Web Browser User Interface is validated on Windows Internet Explorer 10.
The software is provided under and is subject to a licence from CareFusion.
nurse call for high and medium priority alarms.
The Alaris GW Volumetric Pump will not activate the Pump Alarm Location Beacon and nurse call for medium priority
alarms, however will activate the Pump Alarm Location Beacon and nurse call for high priority alarms.
1000DF00742 Issue 2
6/30
Alaris™ Gateway Workstation v1.2
Introduction
Workstation Identification
The Workstation can be confirmed by viewing the serial number label on the rear of the Workstation and checking the SKU is
80203UNSxx, where xx indicates the connectivity option. The two or three numbers at the end of the SKU identifies the Workstation
configuration, see below for all available configurations.
The user must be thoroughly familiar with the Workstation described in this manual prior to use.
Please refer to the relevant Directions For Use for correct operation of the Pumps.
All illustrations used in this manual show typical settings and values which may be used in setting up the functions of the Workstation.
These settings and values are for illustrative use only. The complete range of settings and values are shown in the Specifications section.
The illustrations in this DFU show example configurations and equipment that might not be available to all markets and regions. Please
contact the local affiliate office for further information.
It is important to ensure that you only refer to the most recent version of the Directions for Use and Technical Service
A
Conventions used in this manual
BOLDUsed for Display names, software commands, controls and indicators referenced in this manual, for
'Single quotes'Used to indicate cross-references made to another section of this manual.
ItalicsUsed to refer to other documents or manuals and also used for emphasis.
WarningA warning is an alert to a potential hazard which could result in serious personal injury and/or product
CautionA caution is an alert to a potential hazard which could result in minor personal injury and/or product
NoteNotes contain supplementary information or emphasize a point or procedure.
A
Manual for your CareFusion products. These documents are referenced on www.carefusion.com. Copies can be
obtained by contacting your local CareFusion representative.
example, Battery Indicator, PURGE, ON/OFF button.
damage if proper procedures are not followed.
damage if proper procedures are not followed.
Important Information: Wherever this symbol is shown an Important note is found. These notes
highlight an aspect of use that is important for the user to be aware of when operating the Pump.
UserAccesses the product at the patient to administer treatment and monitors the products use.
Qualified Service Personnel Service and repair the product. Upload and download data to product.
1000DF00742 Issue 2
8/30
Controls and Indicators
Controls
SymbolDescription
Alaris™ Gateway Workstation v1.2
Controls and Indicators
a
ON/OFF Button - Press once to switch the Workstation on. Press and hold for two seconds to switch the
Workstation off. In the event that the system needs to be reset, press and hold for at least four seconds, then
press again to switch the Workstation on.
a
Indicators
SymbolDescription
j
j
k
k
d
Battery Indicator - When illuminated the Workstation is operating from internal battery; when flashing the
battery power is low and auto power down is imminent.
AC Power Indicator - When illuminated the Workstation is connected to the AC power supply and the battery is
being charged.
A Status Indicator - Provides a visual indication of the internal software activity.
B Status Indicator - Provides a visual indication of communication activity of the network within the
Workstation.
ON Status Indicator - When illuminated the Workstation is powered on.
w
System Fault Indicator - The Workstation will illuminate this indicator when an internal fault is present and
detected.