(Maintenance and service information support; troubleshooting, service assistance.)
Phone: 1300 729 258
Email: techservice-au@carefusion.com
iv Alaris System User Manual – with v9.33 Model 8015
Introduction
The Alaris™ PC unit section of this User Manual provides
procedures and information applicable to the Alaris System and
the PC unit. Each of the other major sections provides productspecific procedures and information.
The Alaris System is a modular system intended for adult,
pediatric, and neonatal care. It consists of the PC unit, the
Guardrails™ Suite MX, and up to four detachable infusion and/
or monitoring modules (channels). The Alaris Auto-ID module
can be included as a fifth module. The use of the Alaris System
is restricted to one patient at a time.
Guardrails Suite MX for the Alaris System brings a new level of
medication error prevention to the point of patient care. The
Guardrails Suite MX features medication dosing, concentration
delivery rate, and optional initial programming guidelines for up
to 30 patient-specific care areas, referred to as Profiles. Each
Profile contains a specific Drug Library, an IV Fluid library, and
channel labels, as well as instrument configurations
appropriate for the care area. Optional drug- or IV Fluid-specific
Clinical Advisories provide visual messages. Dosing limits for
each Guardrails drug entry or rate limits for each IV Fluid entry
can be a Hard Limit that cannot be overridden during infusion
programming and/or a Soft Limit that can be overridden, based
on clinical requirements.
WARNING
Read all instructions before using
the Alaris System.
CAUTION
nly
O
A Data Set is developed and approved by the facility’s
own multi-disciplinary team using the Guardrails Editor
software, the PC-based authoring tool. A Data Set is then
transferred to the Alaris System by qualified personnel. The
approved Data Sets are maintained by the Editor Software for
future updates and reference.
Information about an Alert that occurs during use is
stored within the PC unit, and can be accessed using the
Guardrails CQI Reporter software.
Documentation provided with Alaris System products
might reference product not present in your facility or not yet
available for sale in your area.
A superscript number (for example,
) identifies additional
information provided as a NOTE at the end of the procedure.
Alaris System User Manual – with v9.33 Model 8015v
Intended Use
Introduction (Continued)
The Alaris™ System with Guardrails™ Suite MX is intended for
use in professional healthcare facilities that utilize infusion
devices for the delivery of fluids, medications, blood, and blood
products. The Alaris System with Guardrails Suite MX is
intended to provide trained healthcare caregivers a way to
automate the programming of infusion parameters, thereby
decreasing the amount of manual steps necessary to enter
infusion data.
All data entry and validation of infusion parameters is
performed by the trained healthcare professional according to a
physician's order. The Alaris System with Guardrails Suite MX
is an interoperable system capable of communicating and
exchanging data accurately, effectively, securely, and
consistently with different information technology systems,
software applications, and networks, in various settings; and
exchanging data such that the clinical or operational purpose
and meaning of the data are preserved and unaltered.
Essential Performance
The Alaris infusion modules are designed to accurately deliver
the programmed amount of the medication or fluid over the
programmed time period. The Alaris infusion modules ensure
that an infusion is not being inadvertently delivered when the
user expects the system to be in a paused, stopped, or off
condition. The Alaris infusion modules employ measurement
systems to detect and alarm for conditions adverse to safe
administration of fluid. These include measurements of proper
infusion set loading (free flow detection), pressure (occlusion
detection), and air-in-line detection.
The EtC0
module will monitor and alarm when a patient
2
physiological condition is out of range.
The Sp0
module will monitor and alarm when a patient
2
physiological condition is out of range
viAlaris System User Manual – with v9.33 Model 8015
Introduction (Continued)
WARNINGS AND CAUTIONS:
Product-specific warnings and cautions, covered in the
applicable sections of this User Manual, provide information
needed to safely and effectively use the Alaris System.
WARNING
A statement that alerts the user to the
possibility of injury, death, or other serious adverse reactions
associated with the use or misuse of the device.
CAUTION
A statement that alerts the user to the
possibility of a problem with the device associated with the use
or misuse of the device.
DEFINED TERMS:
The following table identifies the defined terms used
throughout this document for certain trademarked products
and product features.
Product/FeatureDefined Term
Alaris™ Auto-ID moduleAuto-ID module
Alaris™ EtCO2 moduleEtCO2 module
Alaris™ PCA modulePCA module
Alaris™ PC unitPC unit
Alaris™ Pump modulePump module
Alaris™ SpO2 moduleSpO2 module
Alaris™ Syringe moduleSyringe module
Alaris™ System MaintenanceSystem Maintenance
Alaris™ Systems ManagerSystems Manager
Guardrails™ alertAlert
Guardrails™ clinical advisoryClinical Advisory
Guardrails™ CQI ReporterCQI Reporter
Guardrails™ data setData Set
Guardrails™ drug libraryDrug Library
Guardrails™ EditorEditor Software
Alaris System User Manual – with v9.33 Model 8015vii
Introduction (Continued)
Product/FeatureDefined Term
Guardrails™ hard limitHard Limit
Guardrails™ IV fluidIV Fluid
Guardrails™ limitLimit
Guardrails™ PCA pause
protocol
Guardrails™ soft limitSoft Limit
SmartSite™ needle-free valveNeedle-Free Valve
PCA Pause Protocol
Approved Parts Recommendation
CareFusion recommends the use of CareFusion manufactured
parts in the operation and maintenance of your CareFusion
equipment. Customer's use of repair or service parts, add-ons,
or disposables that are not approved by CareFusion is at
Customer's own risk and may void the product warranty
provided by CareFusion. Any 510(k) clearance from the Food
and Drug Administration (FDA) or regulatory approval secured
by CareFusion to market Alaris pumps was based on use of
only CareFusion manufactured parts and equipment. If nonCareFusion parts, add-ons or disposables are used for the
maintenance, repair or operation of your CareFusion
equipment, those parts were not validated by CareFusion for
safety and efficacy with our Alaris products, nor were they
included in the review and approval/clearance of the products.
Installation
Instruments are tested and calibrated before they are
packaged for shipment. To ensure proper operation after
shipment, it is recommended that an incoming inspection be
performed before placing the instrument in use.
Prior to placing the Alaris System in use:
1.Perform check-in procedure using System Maintenance
software.
2.Whether or not Profiles feature has been enabled
(see PC unit section, "System Options," "System
Configurations").
NOTE:
To enable the Profiles feature, a hospital-defined best-practice
Data Set must be uploaded to the PC unit.
viiiAlaris System User Manual – with v9.33 Model 8015
Alaris PC Unit
Model 8015
Alaris System User Manual – with v9.33 Model 8015Section 1
Power Off System .................................................................................................................................. 1-12
System Options...................................................................................................................................... 1-13
Clinician ID ..................................................................................................................................... 1-16
Power Down All Channels .............................................................................................................. 1-17
PC Unit SectionTable of Contents
Alaris System User Manual – with v9.33 Model 80151-iii
THIS PAGE
INTENTIONALLY
LEFT BLANK
Table of ContentsPC Unit Section
1-ivAlaris System User Manual – with v9.33 Model 8015
Introduction
Getting Started
This section of the User Manual provides PC unit (Model
8015) and Alaris
System instructions and information. It is used
in conjunction with:
•PC Unit/ Pump Module Technical Service Manual
•Product-specific sections of this User Manual
•System Maintenance software (and its instructions) for
Alaris System check-in, maintenance, and wireless configuration
The PC unit is the core of the Alaris System and provides
a common user interface for programming infusions and
monitoring, which helps to reduce complexity at the point of
care. The display uses color to clearly communicate critical
programming, infusion, monitoring and hospital-defined policy
information.
The wireless network card provides wireless communication
capability between the Alaris System and Alaris Systems
Manager. The combined use of the Alaris System and Alaris
Systems Manager is integrated into a facility’s existing network
infrastructure.
When enabled, the Alaris Systems Manager allows the
exchange of information between the Alaris Systems Manager
and the Alaris System. The PC unit can be operated
manually or in concert with the information exchanged with the
Alaris Systems Manager. If communication with the wireless
network is interrupted (for example, out of range), the Alaris
System can be used, as intended, in the manual mode.
WARNING
Read all instructions, including those
for the attached module(s) and
applicable accessories, before using
the Alaris System.
CAUTION
nly
O
PC Unit SectionGetting Started
Alaris System User Manual – with v9.33 Model 80151-1
Introduction (Continued)
Alarms and Alerts: See "Appendix A - Troubleshooting and
Maintenance" for specific PC unit alarms and alerts.
Contraindications: None known.
Electromagnetic Environment: See "Appendix" section
of this User Manual ("Regulations and Standards,"
"Compliance").
Getting StartedPC Unit Section
1-2Alaris System User Manual – with v9.33 Model 8015
General Setup and Operation
Attach and Detach Module
Modules can be attached to either side of the PC unit or to
either side of another module. The process to attach or detach
is the same for either side, whether attaching/detaching
to/from a PC unit or another module.
An individual hospital/facility can choose to permanently attach
modules. To remove permanently attached modules, contact
qualified service personnel.Inserting a finger or other object
into the IUI connector, when the module is attached to the PC
unit, could result in electrical shock.
WARNING
Attach Module
The Alaris System is designed to operate a maximum of
four infusion or monitoring modules. Modules added in excess
of four are not recognized by the system. The Auto-ID module
can be included as a fifth module. A module can be attached
in any position; however, when mounted on an IV pole, it is
recommended that a balanced configuration be maintained.
Application of adhesive tape or other materials to the sides of
the PC unit and modules can prevent proper latching.
1.Position free module at a 45° angle, aligning IUI
connectors.
IUI Connector
Inserting a finger or other object into
the IUI connector, when the module
is attached to the PC unit, could
result in electrical shock.
Aligned IUI Connectors
Top View
45°
IUI Connectors Not Aligned
Front View
Top View
PC Unit SectionGeneral Setup and Operation
Alaris System User Manual – with v9.33 Model 80151-3
Attach and Detach Module (Continued)
Attach Module (Continued)
2.Rotate free module down against PC unit or attached
module until release latch snaps in place.
WARNING
When properly secured/snapped, the
release latch provides a very secure
connection between modules. If not
properly latched, a module can be
dislodged during operation.
WARNING
Failure to perform these operations
can result in improper instrument
operations.
WARNING
Failure to follow these instructions
may result in potential hazards
associated with damaged IUI
connectors.
Detach Module
1.Ensure that module is powered off before detaching.
2.Push module release latch and then rotate module up
and away from PC unit or attached module (opposite to
motion shown in "Attach Module" procedure) to disengage
connectors.
•Alaris System reidentifies and shows appropriate
module identification (A, B, C, or D), from left to right.
•Appropriate module position(s) (A, B, or C) for
remaining module(s) appear on Main Display.
Release Latch
General Setup and OperationPC Unit Section
1-4Alaris System User Manual – with v9.33 Model 8015
Attach and Detach Module (Continued)
Add Module While System is Powered On
Add module as described in "Attach Module."
•System tests module, causing all LED segments and
indicator lights of displays to illuminate briefly.
•Appropriate module identification display (A, B, C, or D)
illuminates. Modules are always labeled left to right, so if
a module is added to left of other modules, all modules are
reidentified. Module re-identification does NOT interrupt or
affect infusion or monitoring on active modules.
•Module positions (A, B, C, or D) appear on Main Display.
•If any of the following conditions are observed, affected
module must be removed from use and inspected by
qualified personnel:
◦LED segments are not illuminated on displays during
power-on test.
◦Indicator lights do not illuminate.
◦Appropriate module identification does not appear.
PC Unit SectionGeneral Setup and Operation
Alaris System User Manual – with v9.33 Model 80151-5
Start-Up
Power On System
1.Connect PC unit to an external AC power source.
2.Press
SYSTEM ON key.
3.System self test begins:
•Diagnostics test causes all LED display segments
and Status Indicator lights of attached module(s) to
illuminate briefly.
•Power Indicator illuminates.
•Appropriate module identification (A, B, C, or D) is
displayed on attached module(s).
•An audio tone sounds.
•If PM Reminder option is enabled and scheduled
preventive maintenance is due,
REMINDER screen appears.
•At completion of system-on test,
MAINTENANCE
New Patient? screen
appears.
•If either of the following conditions is observed, PC unit
or affected attached module must be removed from
use and inspected by qualified personnel:
◦System fails any part of self test.
◦Main Display does not appear backlit, appears
irregular, or has evidence of a row of pixels not
functioning properly.
NOTE:
Previous infusion parameters are automatically cleared after
8 hours.
The self test provides the clinician with verification of the
operational safety and correct functioning of alarms for the
Alaris System.
General Setup and OperationPC Unit Section
1-6Alaris System User Manual – with v9.33 Model 8015
Start-Up (Continued)
Respond to Maintenance Reminder
If the
Preventive Maintenance (PM) Reminder option is enabled
and the PC unit or an attached module is due for preventive
maintenance, a
at power up. If necessary, the reminder can be temporarily
bypassed by pressing the
1.Notify the appropriate facility personnel when a
MAINTENANCE REMINDER occurs and remove instrument
requiring maintenance (see "Attach and Detach Module").
2.If Alaris System was powered off to replace PC unit,
reinitiate start-up process.
OR
MAINTENANCE REMINDER message appears
CONFIRM soft key.
MAINTENANCE REMINDER
Module(s) due for routine
B
preventive maintenance:
Module A:
YYYY-MM-DD
CONFIRM
If an attached module (such as a Pump module) was
powered off and removed,
MAINTENANCE REMINDER
display reflects removal of that module. To continue
start-up process, press
CONFIRM soft key.
Adjust Display Contrast
1.Press
DISPLAY CONTRAST soft key.
MAINTENANCE REMINDER
Module(s) due for routine
B
preventive maintenance:
CONFIRM
Midtown Hospital
NEW PATIENT ?
“Yes” Clears Previous
Patient Data
>Select Yes or No
DISPLAY
CONTRST
Yes
No
PC Unit SectionGeneral Setup and Operation
Alaris System User Manual – with v9.33 Model 80151-7
Start-Up (Continued)
Adjust Display Contrast (Continued)
2.To adjust display for optimum viewing, use Lighter/Darker
soft keys.
3.To return to main screen, press
CONFIRM soft key.
Select New Patient and Profile Options
The following procedures assume the Profiles feature is
enabled.
1.Select required
NEW PATIENT? option.
•To indicate programming is for a new patient and clear
all stored patient parameters from memory, press Yes
soft key.
•To confirm programming is for same patient and retain
all stored patient parameters, press No soft key.
System Options
Display Contrast
®
>Adjust Display to
Desired Contrast
CONFIRM
Midtown Hospital
NEW PATIENT ?
“Yes” Clears Previous
Patient Data
Lighter
Darker
Yes
No
◦Last used Profile is displayed.
◦If Profiles feature is disabled, main menu appears.
2.Accept or change current Profile:
•To accept current Profile, press Yes soft key.
Main screen appears.
•To change Profile, press No soft key and continue with
next step.
Profile selection screen appears.
>Select Yes or No
DISPLAY
CONTRST
Midtown Hospital
Adult ICU
Adult ICU ?
“Yes” Confirms Same
Profile
>Select Yes or No
Yes
No
General Setup and OperationPC Unit Section
1-8Alaris System User Manual – with v9.33 Model 8015
Start-Up (Continued)
Select New Patient and Profile Options (Continued)
3.To select a Profile, press corresponding left soft key.
To view additional choices, press
4.To confirm Profile selection, press
PAGE DOWN soft key.
CONFIRM soft key.
Main screen appears.
Patient ID Entry Feature
The option to enter and display a 16-character alphanumeric
patient identifier is always available. The instrument can be
configured to automatically display the Patient ID Entry
screen during start-up or to provide access only through the
Systems Options menu (see "System Options").
If Yes was selected to indicate programming for a new patient,
perform one of following steps:
•If patient identifier is not required, press
CONFIRM or EXIT
soft key.
•To manually enter patient identifier, use numeric data entry
keys and/or alpha speed keys.
◦An alphanumeric identifier, of up to 16 characters, can
be entered.
◦Press soft key next to a letter group to list letters in
that group. Press soft key next to an individual letter to
enter that letter.
Midtown Hospital
Profiles
Adult ICU
Adult General Care
Neonatal
Peds ICU
Neonatal ICU
>Press CONFIRM
Patient ID Entry
A
B
C
D
E
_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
>Enter Patient ID and Press
CONFIRM
EXIT
CONFIRM
CONFIRM
1 of 2
PAGE
DOWN
PAGE
DOWN
A-E
F-J
K-O
P-T
U-Y
◦To access letter "Z" and special characters (hyphen,
underscore, space), press
◦To clear an entire entry, press
◦To back up a single character at a time, press
PAGE DOWN soft key.
CLEAR key.
CANCEL
key.
•To scan barcode on patient identification band,
see"Alaris Auto-ID Module Model 8600", section 6 of this
User Manual.
PC Unit SectionGeneral Setup and Operation
Alaris System User Manual – with v9.33 Model 80151-9
A
p
Start-Up (Continued)
Adjust Audio Volume
1.Press AUDIO ADJUST soft key.
WARNING
The user should check that the
current alarm preset is appropriate
prior to use on each patient.
WARNING
hazard can exist if different Alarm
Presets are used for the same or
similar equipment in a single area,
such as an Intensive Care Unit or a
Cardiac O
erating Theatre.
Midtown Hospital
Adult ICU
VTBI = 250.0 mL
VOLUME
INFUSED
AUDIO
ADJUST
2.To change volume to desired level, press either Louder
or Softer soft key. To sample alarm loudness level, press
Test soft key.
3.To return to PC unit screen, press
MAIN SCREEN soft key.
After 30 seconds without a key press, Main Display
appears.
NOTE:
The Minimum Audio Volume defaults to level 1. Levels 1-5 may be
set per profile. The user is able to lower the device AUDIO
VOLUME to the minimum limit set. When the minimum AUDIO
VOLUME LEVEL has been reached, the Louder/Softer soft keys
are unavailable and the PC unit emits an illegal key press audio.
CAUTION
Setting the audio volume to the
lowest level will lower all system
alarms, including secondary alarms
such as End of Infusion.
Audio Volume Adjust
Test
3
Softer
>Change Setting or
Cancel
Louder
MAIN
SCREEN
General Setup and OperationPC Unit Section
1-10Alaris System User Manual – with v9.33 Model 8015
Start-Up (Continued)
Lock/Unlock Tamper Resist
1.Initiate operation of applicable module.
2.Press and hold Tamper Resist Switch, on back of PC unit,
for 3 to 4 seconds (see "General Information," "Features
and Displays," "Operating Features, Controls, Indicators").
•An advisory tone (if Key Click Audio is enabled)
and a three-second
PANEL LOCKED prompt on Main
Display confirm activation.
•When Tamper Resist is active, keypad panel is locked;
however, clinician can:
◦Silence audio alarm.
◦View volume(s) infused.
◦View and test audio alarm setting.
◦View selected parameters on attached modules.
Any other key press results in a visual
PANEL LOCKED
prompt and, if Key Click Audio is enabled, an illegal
key–press audio advisory.
3.To unlock keypad panel, press and hold Tamper Resist
Switch for 3 to 4 seconds.
An advisory tone (if Key Click Audio is enabled) and a
three-second
PANEL UNLOCKED prompt on Main Display
confirm activation.
Midtown Hospital
Adult ICU
VTBI = 250.0 mL
PANEL LOCKED
VOLUME
INFUSED
Midtown Hospital
Adult ICU
VTBI = 250.0 mL
AUDIO
ADJUST
PANEL UNLOCKED
VOLUME
INFUSED
AUDIO
ADJUST
PC Unit SectionGeneral Setup and Operation
Alaris System User Manual – with v9.33 Model 80151-11
Power Off System
Press and hold CHANNEL OFF key until a beep is heard
(approximately 1.5 seconds) and then release to initiate power
down.
•During power off sequence, Main Display flashes
Powering Down.
•To interrupt power down sequence, quickly press any key
(except
Once all attached modules are powered off, PC unit
automatically powers down.
SYSTEM ON) on PC unit.
Powering Down
General Setup and OperationPC Unit Section
1-12Alaris System User Manual – with v9.33 Model 8015
System Options
Display Contrast
1.Press OPTIONS key.
2.Press Display Contrast soft key.
3.Adjust display and return to main screen (see Start-Up,"
Adjust Display Contrast" procedure).
SILENCE
OPTIONS
1
2
4
5
7
8
CLEAR
0
System Options 1 of 3
Display Contrast
Patient ID
Clinician ID
Power Down All Channels
Anesthesia Mode
>Select an Option or
EXIT
EXIT
SYSTEM
ON
3
6
ENTER
9
CANCEL
PAGE
DOWN
Patient ID
Enter
1.Press
2. Press Patient ID
PC Unit SectionGeneral Setup and Operation
Alaris System User Manual – with v9.33 Model 80151-13
OPTIONS key.
soft key.
System Options 1 of 3
Display Contrast
Patient ID
Clinician ID
Power Down All Channels
Anesthesia Mode
>Select an Option or
EXIT
EXIT
PAGE
DOWN
System Options (Continued)
Patient ID (Continued)
Enter (Continued)
3.Scan or manually enter patient identifier:
•To manually enter patient identifier, use numeric data
entry keys and/or alpha speed keys.
◦An alphanumeric identifier, of up to
16 characters,
can be entered.
◦Press soft key next to a letter group to list letters
in that group. Press soft key next to an individual
letter to enter that letter.
◦To access letter "Z" and special characters
(hyphen, underscore, space), press
PAGE DOWN
soft key.
◦To clear an entire entry, press
CLEAR key.
◦To back up a single character at a time, press
CANCEL key.
•To scan barcode on patient identification band, see
AUTO-ID section of this User Manual.
4. To verify correct entry, press
CONFIRM soft key.
Patient ID Entry
CONFIRM
A-E
F-J
K-O
P-T
U-Y
PAGE
DOWN
A
B
C
D
E
123456789CD_ _ _ _ _
>Enter Patient ID and Press
CONFIRM
EXIT
General Setup and OperationPC Unit Section
1-14Alaris System User Manual – with v9.33 Model 8015
SILENCE
Modify
System Options (Continued)
Patient ID (Continued)
1. Press
OPTIONS key.
2.Press Patient ID soft key.
3.To clear entire entry, press
OR
CLEAR key.
To back up a single character at a time, press CANCEL
key.
4.To enter modified patient identifier, use numeric data entry
keys and/or alpha speed keys.
•An alphanumeric identifier, of up to 16 characters, can
be entered.
•Press soft key next to a letter group to list letters in
that group. Press soft key next to an individual letter to
enter that letter.
OPTIONS
1
4
7
CLEAR
SYSTEM
ON
2
3
5
6
ENTER
9
8
0
CANCEL
OR
•To access letter "Z" and special characters (hyphen,
underscore, space), press
5.To verify correct entry, press
New Patient ID Entry verification screen appears.
PC Unit SectionGeneral Setup and Operation
Alaris System User Manual – with v9.33 Model 80151-15
To retain original (old) Patient ID, press No soft key.
Main screen appears with old Patient ID.
Clinician ID
1.Press
OPTIONS key.
2.Press Clinician ID soft key.
Patient ID Entry
Patient ID
123456789CD
will be changed to
234567891EF
Is this correct?
>Press Yes or No
System Options 1 of 3
Display Contrast
Patient ID
Clinician ID
Power Down All Channels
Anesthesia Mode
>Select an Option or
EXIT
EXIT
Yes
No
PAGE
DOWN
3.Scan or manually enter clinician identifier:
To manually enter clinician identifier, use numeric data
entry keys and/or alpha speed keys.
•An alphanumeric identifier, of up to 16 characters, can
be entered.
•Press soft key next to a letter group to list letters in
that group. Press soft key next to an individual letter to
enter that letter.
•To access letter "Z" and special characters (hyphen,
underscore, space), press
•To clear an entire entry, press
•To back up a single character at a time, press
PAGE DOWN soft key.
CLEAR key.
CANCEL
key.
General Setup and OperationPC Unit Section
1-16Alaris System User Manual – with v9.33 Model 8015
System Options (Continued)
Clinician ID (Continued)
4. To verify correct entry, press CONFIRM soft key.
Power Down All Channels
1.Press
OPTIONS key.
2. Press Power Down All Channels soft key.
Clinician ID Entry
CONFIRM
PAGE
DOWN
PAGE
DOWN
A-E
F-J
K-O
P-T
U-Y
A
B
C
D
E
123456789CD_ _ _ _ _
>Enter Clinician ID and Press
CONFIRM
EXIT
System Options 1 of 3
Display Contrast
Patient ID
Clinician ID
Power Down All Channels
Anesthesia Mode
>Select an Option or
EXIT
EXIT
3.Press Yes soft key.
During power off sequence, Main Display flashes
POWERING DOWN.
System Options
Power Down
All Channels?
>Press Yes or No
Yes
No
PC Unit SectionGeneral Setup and Operation
Alaris System User Manual – with v9.33 Model 80151-17
System Options (Continued)
Anesthesia Mode
When the Anesthesia Mode is enabled and then the pause
feature is used - the module remains in an indefinite pause until
restarted.
When Anesthesia Mode is enabled:
• A channel can be paused indefinitely without an alarm.
•The air-in-line associated with the Profile can be set up to
500 micro liters.
•All limits are set to Soft.
•Limit Checking mode is set to Smart.
•Key-press audio is turned off.
•Auto-Restart for Anesthesia Mode is set to 9 and is not
configurable.
•Panel Lock through Tamper Resist Mode or Authorized
User Mode is not available.
•Guardrails drug list defaults to drugs designated by Editor
Software as anesthesia only. All Guardrails drugs in a
Profile can be viewed by pressing ALL DRUGS soft key.
•Bolus Dose is automatically available for:
WARNING
When the Alaris System is set up for
use in Anesthesia Mode, it is
important to select the Profile that
corresponds with the care area the
patient will be taken to when the
Anesthesia Mode is discontinued.
This ensures that the Alaris System
will be in the correct Profile following
the use of the Anesthesia Mode.
◦Guardrails drugs that have Bolus Dose limits defined
◦generic drug calculation setup
•Anesthesia Mode, alternating with other required prompts,
is displayed in prompt bar of Main Display.
•Callback audio for paused module is permanently silenced.
•Review of drug calculation setup page is omitted when
restoring a stopped drug calculation.
•Clinical Advisories are not displayed.
NOTE:
If an infusion is paused in regular mode and then the anesthesia
mode is enabled - the device will alarm at the 2 minute warning.
Enable
1.Press
OPTIONS key.
System Options 1 of 3
Display Contrast
Patient ID
Clinician ID
Power Down All Channels
Anesthesia Mode
>Select an Option or
EXIT
EXIT
PAGE
DOWN
General Setup and OperationPC Unit Section
1-18Alaris System User Manual – with v9.33 Model 8015
Loading...
+ 412 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.