Outcome assessment form .................................................................................................41
Useful information .............................................................................................................42
ii
Indications and contraindications
Indications
The 3100B High Frequency Oscillatory Ventilator is indicated for use in the ventilatory
support and treatment of selected patients weighing 35 kilograms and greater with
acute respiratory failure.
Contraindications
The 3100B High Frequency Oscillatory Ventilator has no specific contraindications.
1
Identifying patients for high-frequency oscillatory ventilation
Patients with ALI or ARDS, weighing 35 kilograms or greater, who are currently
failing on conventional ventilation with a protective lung strategy, will benefit from
HFOV. The following criteria are generally used for determining the feasibility of
using HFOV.
• FiO2 ≥ 60, PEEP ≥ 10 with a P/F ratio < 200
• Plateau pressure > 30 cmH2O
• Presence of bilateral infiltrates on the chest x-ray consistent with ARDS
• Oxygenation index > 24
Note:
Several clinical papers and randomized control trials have
demonstrated that the earlier application of HFOV on patients with
severe ARDS may be important for successful outcomes.
2
The Multicenter Oscillatory Ventilation for Acute Respiratory Distress Syndrome Trial1
(MOAT2) excluded severe COPD and asthma patients from the RCT trial of the 3100B
HFOV. High frequency oscillatory ventilation is recognized as less effective in diseases
with increased airway resistance and using it in such cases may potentially result in air
trapping and hyperinflation.
3
Things to consider before placing the patient on HFOV
• Hemodynamic status:
– The patient should be hemodynamically stable with a mean arterial pressure
of at least 75 mmHg
– If mean arterial pressure is less than 75 mmHg, consider fluids and/or
vasopressors to optimize the hemodynamic status before starting
the oscillator
• Obtain ABG
– Ideally, pH should be greater than 7.2
– If pH is less than 7.2, consider correcting with a buffer for the transition
• Patient’s sedation status:
– You may need to consider sedation and neuromuscular blockades for
transition—due to the fixed Bias Flow of the device, patients are unable to
actively breathe and maintain a stable airway pressure and lung volume
– Once transitioned, patients may be able to be maintained on sedation only
4
• Ensure the patient has had a recent chest x-ray
• Consider the type of mattress the patient is on—you should have the ability to
firm up the mattress if needed
• Verify whether the patient requires an off-unit procedure such as a CT scan or
MRI; if this is the case, consider doing the off-unit procedure before placing
the patient on HFOV
• If using an in-line suction catheter, ensure the fit to the circuit and suction
patient before placing on HFOV
• Give a brief explanation of HFOV to the family and patient to prepare them
for noise, chest wiggle, etc.
• If lung recruitment maneuvers are not part of your ventilation protocol, speak
with the physician to consider Treatment with Oscillation and Open Lung
Strategy (TOOLS)
2
5
Pre-use checklist
1. Connect the source gases to system.
2. Connect the power to system.
3. Check that the patient circuit support is installed on the system.
4. Connect the patient circuit and the humidifier to the system.
5. Connect the patient circuit control and pressure sense lines to the system.
6. Turn on the power.
7. Check that the source gas lights are off.
8. Check that the start/stop light is off.
9. Check that the alarm silence light is on.
10. Perform the patient-circuit calibration described in the following section.
11. Perform verification performance.
12. Perform the alarm-check procedure as described in the Operators Manual.
6
13. Preset the flow, frequency, % inspiratory time, power and running mean
airway pressure.
14. Set the Max Paw and Min Paw switches.
15. Set the blender and humidifier controls for the desired operation.
16. Remove the stopper from the patient circuit and connect it to the
patient’s ET tube.
7
Patient circuit calibration
Perform the patient circuit calibration procedure before ventilating a patient.
Each circuit that is used on the oscillator must be calibrated. The circuit calibration
procedure verifies the circuit is leak-free and will hold pressure. Perform this
procedure before placing a patient on the 3100B HFOV and anytime a circuit
component is changed.
1. Insert the stopper in the patient circuit wye and turn on the bias flow gas.
2. Rotate the ADJUST control to Max.
3. Set the Max Paw Alarm to 59 cmH2O.
4. Set the bias flow to exactly 20 LPM (the middle of the ball is at the line—you may
need to bend down to see this accurately.).
5. Depress and hold RESET (Oscillator OFF).
8
6. Observe the mean pressure display and adjust the patient circuit calibration screw
for a reading of 39 to 43 cmH2O.
a. Before adjusting the calibration screw, confirm there are no leaks, the bias
flow is at 20 LPM and the circuit is set up correctly. See the troubleshooting
guide for more information.
b. Use caution when adjusting the calibration screw. Do not over tighten or
apply excessive force because equipment damage may occur.
9
Ventilator performance check
The ventilator performance check ensures the 3100B HFOV is functioning properly.
Perform this procedure before placing a patient on the 3100B HFOV. Insert the
stopper in the patient circuit wye and turn on both gas sources.
1. Turn the Adjust control to the 12 o’clock position.
2. Set the bias flow at 30 LPM.
3. Pressurize the system by pressing and holding Reset and Adjust for a mean
pressure of 29 to 31 cmH2O.
4. Set the frequency to 6.0 Hz, % I–time to 33 and press START/STOP to start
the oscillator.
5. Set the power to 6.0.
6. Observe the following parameters using the appropriate altitude range and verify
they fall within the ranges specified.
10
Verify the following parameters according to the altitude of your hospital.
Altitude (feet)
0 to 2,00026 to 34113 to 135
2,000 to 4,00026 to 34104 to 125
4,000 to 6,00026 to 3499 to 115
6,000 to 8,00026 to 3486 to 105
mPAW (cmH2O)P (cmH2O)
Note:
See troubleshooting on pg. 22 for additional information.
11
Initial settings and management
1. Set bias flow between 25 to 40 LPM.
Patients with severe air-leak syndrome or cuff leak may require higher set Bias
Flow to achieve the desired mPaw.
2. Set the initial mean airway pressure (mPaw) at 5 cmH2O pressure above the
conventional ventilator mPaw.
a. You may consider a recruitment maneuver first if the patient is extremely
hypoxic by applying 40 cmH2O for 40 to 60 seconds.
b. If oxygenation worsens, increase mPaw in 3 to 5 cmH2O increments every
30 minutes until the maximum setting is reached.
Note:
Oxygenation typically may worsen in the first 30 minutes of
recruitment in severe ARDS.
12
Loading...
+ 33 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.