GENERAL CHARACTERISTICS ...........................................................................................................................4
1
2 USER MANUAL ..................................................................................................................................................... 5
2.4 Medical Field Selection (push button [DIAG]) ............................................................................................8
2.4.1 User diagnoses ..........................................................................................................................................8
2.5.1 Density .......................................................................................................................................................9
2.5.2 Irradiated Area ...........................................................................................................................................9
2.5.4 Frequency ................................................................................................................................................10
2.5.5 Nogier (Frequencies according to Dr. Nogier) .........................................................................................10
2.5.6 EAV (Frequencies according to Dr. Voll) .................................................................................................10
2.5.7 Medical Fields ..........................................................................................................................................11
2.8.2 Patient List ............................................................................................................................................... 14
2.8.3 User diagnoses ........................................................................................................................................15
2.8.4 System Setup........................................................................................................................................... 17
2.9.1 Display during therapy .............................................................................................................................23
2.9.2 To start therapy........................................................................................................................................ 23
2.9.3 To interrupt therapy.................................................................................................................................. 23
2.9.4 To Continue Therapy ...............................................................................................................................23
2.9.5 To terminate therapy................................................................................................................................23
2.10 Power Supply: Mains and Battery...........................................................................................................23
2.11.3 ENT Attachment....................................................................................................................................... 24
2.11.4 Positioning an Attachment ....................................................................................................................... 25
2.13 Safety of Instrument.................................................................................................................................25
2.13.1 Remote control ( DOOR ).........................................................................................................................25
2.13.2 Use of Safety Accessories .......................................................................................................................25
2.16.1 Standard Accessories .............................................................................................................................. 27
4.2 Light; the principle of laser light and its specifics..................................................................................... 31
4.3 Types of lasers......................................................................................................................................... 32
4.3.4 Type of beam ...........................................................................................................................................34
4.3.5 Laser class (defined by the output)..........................................................................................................34
4.5 Biologic effects of a low-output laser (non-invasive laser) ....................................................................... 34
Page 2 of 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
4.6 Contradictions of non-invasive laser therapy ........................................................................................... 35
5 LIST OF THERAPIES...........................................................................................................................................36
5.1 Dentistry (Medical Field 03) .....................................................................................................................36
5.2 Dermatology (Medical Field 04) ...............................................................................................................36
5.3 Gynaecology (Medical Field 05)...............................................................................................................37
5.4 E.N.T. (Medical Field 06) .........................................................................................................................37
5.5 Rehabilitation (Medical Field 07)..............................................................................................................38
5.6 Neurology (Medical Field 08) ...................................................................................................................39
5.7 Orthopaedics (Medical Field 09) .............................................................................................................. 39
5.8 Sports Medicine (Medical Field 10).......................................................................................................... 40
5.9 Pediatrics (Medical Field 11)....................................................................................................................40
5.10 General Practitioner (Medical Field 12)....................................................................................................40
BTL-2000 is a portable therapeutic laser that employs red and infrared BTL laser probes of various
powers.
BTL-2000 is designed for easy use and is understandable even for those who are new to laser
technology. The laser beam irradiation strength is easy to regulate. Parameters can easily be
changed as desired and your own programmes can be created and saved. BTL-2000 software
allows continuous change of frequency, density, power, area and time. Considering all parameters
are correlated, when a parameter is changed the instrument automatically recalculates the
remaining parameters.
BTL-2000 provides preset programmes for the following medical fields: rehabilitation, orthopaedics
dentistry, gynaecology, neurology, dermatology, ENT, sports medicine, paediatrics, and general
practice. The use of the preset programmes can significantly speed up therapy. When a field and
diagnosis is chosen, the instrument will automatically set up the therapy parameters.
BTL-2000 excels with its high power and operates in both Continuous and Pulse modes. In the
pulse mode it is possible to modify the DUTY FACTOR of output frequency in the range of
10–90 %. BTL-2000 enables the user to precisely determine the necessary probe distance from
skin by setting the desired therapy surface area. BTL-2000 automatically tests laser probe
performance and corrects the power for optimal laser output. BTL-2000 is compatible with both
digital and analog probes.
BTL-2000 is especially suitable for acupuncture treatment. BTL-2000 has two modes for easy and
percise location of acupuncture points. While the instrument measures the changes in skin
resisitivity, visual and accoustic signals indicate acupuncture point locations. The instrument is also
pre-programmed with Nogier and EAV frequencies.
For dentistry, gynaecology and ENT special optical attachments enable the laser to treat difficult to
reach places. The optical attachments are easily attatched to the probe and ready for use.
BTL-2000 complies with all internationally recognised laser instrument specifications.
Page 4 z 45
BTL-2000
2 USER MANUAL
Upper Panel, Front Panel, Lower Panel
USER'S MANUAL & USER’S GUIDE
12
3
4
5
6
7
8
9
10
11
12
13
16
14
15
Page 5 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.1.1 Outward Appearance
BTL-2000 is housed in a metal varnished box. Weight and measurements are described in Technical Parameters3.
The lighted LCD display and push buttons are located on the front panel. Laser probe connector and mains power
switch are located on the upper panel. Connector for mains power/recharge adapter (CHARGE) cable, acupuncture
probe connector (ACUPUNCTURE), connector for remote blocking (REMOTE CONTROL) and input for measuring
laser power (PROBE TEST) are located on the lower panel.
Voltage converter switch (115/230) and mains power plug are located on the mains power adapter (BTL-282).
Manufacturer label and product label are located on the rear panel of the instrument case and mains power adapter.
Warning labels are located on the rear and upper panels of the instrument case and on the laser probe.
2.1.2 Functions of Push buttons, Connectors, and Display.
6. [MAN] push button sets therapy parameters for manual programming.
7. [MENU] push button returns instrument to System Menu.
8. [TAB] push button moves cursor to right when entering Safety Code, selects parameter settings [MAN], and moves cursor between medical field number and diagnosis number [PROG].
10. [ESC/STOP] push button terminates therapy, starts/stops timer, and returns display to previous setting.
11. [ + ] push button increases parameter values, selects settings, and moves cursor to the right.
12. [ - ] push button decreases parameter values, selects settings and moves cursor to the left deleting previous
character.
13. Connector for mains power/ recharge adapter cable (CHARGE).
14. Probe Test: Laser Optical Power Meter (TEST).
15. Acupuncture Probe Input (ACUPUNCTURE).
16. Remote Control Connector (DOOR).
2.1.3 Basic Operation Procedures
BTL 2000 is controlled by means of an interactive menu. Press the [ + ] or [ - ] push buttons to move between menu
settings. The selected setting is indicated by an arrow in the display’s left margin. The presence of an additional
arrow indicates that the menu exceeds four lines when selecting the last setting on the display. Press [ENTER] to
confirm selected setting. Press [ESC/STOP] to return to previous menu.
TAB
TAB
UVWXYZ
UVWXYZ
The lower margin of the push buttons contains a set of letters. For entering letters, choose the desired letter by the
number of push button presses relative to the placement of that character on the push button. For example in the
figure above, U is entered by pressing the [Tab] push button once, V is entered by pressing the [Tab] push button
twice, etc..
11 characters can be entered using each lettered push button. For entering non letter characters, choose the desired
character by the number of push button presses relative to the placement of that character in the table below. For
example in the figure below, “ ! “ is entered by pressing the [Tab] push button nine times, “ ? “ is entered by pressing
the [Tab] push button ten times, etc..
Page 6 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
DIAG
PROG
MAN
MENU
TAB
For entering sentences (name, diagnosis, etc.) the [ + ] push button moves cursor to the right, acting as a spacebar,
and the [ - ] push button moves the cursor to the left deleting previous characters.
2.2 Instrument Assembly and Initiation
Carefully unpack instrument and place on a sturdy and level surface (max. 15° angle). Place in cool, dry environment
out of reach from direct sunlight. Instrument generates heat when in use therefore it should not be placed in the
vicinity of other heat generating devices. This may cause overheating of instrument. The instrument is cooled by
natural air convection. Do not cover instrument with flammable materials. Do not place instrument in proximity of
devices that emit strong eletric, magnetic, or electro-magnetic fields (motors, transformers, X-rays, etc.). This can
cause electromagnetic interference. In case of any misunderstanding, please contact your BTL service
representative.
Before the INITIAL turning on of instrument, check the Voltage Converter located on the adapter to insure it
is in the correct position and in accordance with the mains voltage parameters. Voltage coversion is further
described in the chapter Maintenance 2.12. In case of any misunderstanding, please contact your BTL service
representative.
We recommend proper storage of the instrument box in case additional transport of instrument is necessary.
The instrument applies irradiation only, it does not apply any pharmaceuticals or substances.
Connecting instrument (DOOR connector on the lower panel) with the door contact is recommended. For further
information see Safety 2.13.
The power will be ON following the pressing of the mains power switch on the upper panel. When the battary is low, a
“Low battery“ warning will be signalled on the display. It is then necessary to connect the mains power adapter to
recharge the battery.
When turning the instrument on you MUST NOT press any buttons on the front panel. This will generate the
error sign “Error 100 – Keyboard button jammed”. It is then necessary to turn off the instrument for minimum
of three seconds to allow it to reset. (See NOTE).
It is now necessary to enter a code to unlock the instrument. The manufacture code is preset as 0000. If it is not
modified you can continue by pressing the [ENTER] push button. It is possible to change the code using the
instructions in the chapter User diagnoses 2.4.1. Once the accurate code has been entered, the display shows th
main menu. If there are any other reactions after the turning on of the instrument do not attach laser probe and
contact BTL service technician before further use.
NOTE: After turning the instrument on and before the entering of the code, the instrument is testing its internal
functions (∼ 0.5 sec.). If the instrument finds an internal problem, you will see the error notice “Error“ and the
instrument will then be locked in Safe Mode. In this case, it is necessary to contact your BTL service representative. If
the error notification “Error 100 – Keyboard button jammed“ is displayed, it is either an operating error or it is
possible a button is jammed. In the second case, try to release the affected button or contact your BTL service
representative.
A B C D E 1 2 . , : ;
F G H I J 3 4 + - * /
K L M N O 5 6 ( ) < >
P Q R S T 7 8 @ # $ &
U V W X Y Z 9 0 ! ? ´
2.3 Starting Operation
Switch the instrument on by pressing the power switch on the upper panel. Enter the safety code on initial screen.
BTL-2000
T HER APEUTICLASER
E N T E R CODE: 0000
Page 7 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
Using the [ + ] and [ - ] arrows, insert the safety code. Use the [TAB] push button to move cursor to the next position.
Once the code has been entered, press the [ENTER] push button. The manufacture code is preset as 0000. Press
the [ESC/STOP] push button to delete inserted values and to return to the first position.
If the code is entered correctly, the display will show menu which can be preset using the [Sart with] function. If the
code is not correct, the cursor will move to the first position of the code.
BTL-2000
UNIVERSAL
D I A G PROG
M A N MENU
Use the relative push buttons on the instrument front panel to select from the four programmes on the Main Menu.
2.4 Medical Field Selection (push button [DIAG])
When pressing the [DIAG] push button the instrument will enter the Medical Field Menu and it is then possible to
select a preset medical field, Acupuncture frequency (Nogier or EAV), or user created programmes.
According to user specialization, there is only one medical field pre-programmed in the instrument. Additional medical
fields are available but not included.
User diagnoses
Nogier
EAV
Dentistry
Dermatology
Gynaecology
ENT
Rehabilitation
Neurology
Orthopaedics
Sports medicine
Paediatrics
General Practitioner
Example of medical field menu:
> U se r diagnoses
N o g i er
E A V
D e n t istry
Use the [ + ] or [ - ] arrows to move between medical fields and press the [ENTER] push button to select desired
medical field. Press the [ESC/STOP] push button to return to the previous menu.
2.4.1 User diagnoses
This setting allows searching for and selecting a user diagnoses with preset therapeutic parameters.
If there are no user diagnoses in the memory, “Memory free” will appear on the display.
If there are less than 6 user-created diagnoses in memory, a list of these diagnoses in alphabetical order will appear
on the display. If there are 6 or more user-created diagnoses in memory, the following diplay will appear:
U S E RDIAGNOSES
F i r s t char:_
Page 8 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
Enter the first letter of desired diagnosis name and press the [ENTER] push button to enter the list of diagnoses. Use
the [ + ] and [ - ] arrows to move between diagnoses names and press the [ENTER] push button to select desired
diagnosis. After selecting desired diagnosis, preset therapy parameters will be displayed. Press the [ESC/STOP]
push button to return to previous menu.
D e n s . _0.9J/cm
2
F r e qu ency6.36Hz
A r e a 1.0cm2Pwr30mW
T i m e 1:39Ready
Density shows set irradiation density value (J/cm
Frequency shows set frequency (Hz)
Area shows set area of irradiated area (cm2)
Power shows set laser beam output (mW)
Time shows set therapy duration (min/sec)
Use the [ + ] or [ - ] arrows to adjust therapy parameters. Use the [TAB] push button to move between parameters.
Start therapy by pressing the probe push button.
For saving the adjusted parametersuse the [MENU] push button to enter patient file or user diagnoses file.
Use the [ESC/STOP] push button to return to previous menu.
2
)
2.5 Adjusting therapy parameters
When adjusting preset therapy parameters, use the [MENU] push button to enter patient files, diagnoses, or user
settings.
2.5.1 Density
When the density is adjusted, the instrument automatically recalculates therapy time and power parameters.
2.5.2 Irradiated Area
When using a divergent probe, use the following table for the probe distances from skin:
Irradiated area
[cm2]
0.5 1.5
1 2
1.5 2.5
2 3
2.5 3.5
3 4
When using a divergent probe, depth of penetration changes according to the distance of the probe from skin and it
is necessary to deduct the desired depth of penetration from the displayed distance of probe.
When using a convergent probe, distance of the probe from the skin is not a crucial factor. The distance is not
measured because the beams are parallel.
Distance of probe from skin
[cm]
2.5.3 Power
When the power is adjusted, the instrument automatically recalculates the duration of exposure. Power value could
be adjusted within range 20% to 100% of maximal probe output. Output power and the duration of therapy has an
inverse relationship. An increase in power output will result in a decrease in therapy duration. Density will remain the
same. The instrument calculates output power in tenths of mW, but the display only shows whole numbers. Maximum
and minimum power depends on the type of probe used, preset density and area.
Page 9 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.5.4 Frequency
When the frequency is adjusted, the instrument automatically recalculates therapy time and power parameters.
BTL-2000 offers pre-programmed EAV and Nogier frequencies.
2.5.5 Nogier (Frequencies according to Dr. Nogier)
After entering Nogier freqencies, the list of frequencies will be displayed. Use [ + ] and [ - ] arrows to move between
the frequencies. Press the [ENTER] push button to select required frequency. Press the [ESC/STOP] to return to
previous menu.
> U 1 .14Hz
A 2 .28Hz
B 4 .56Hz
C 9 .12Hz
When required frequency is selected, therapy parameters will be displayed:
D e n s . _2.0J/cm
2
F r e q u ency1.14Hz
A r e a 1.0cm2Pwr30mW
Ti m e 1:23Ready
Use the [ + ] and [ - ] arrows to adjust therapy parameters. Use the [TAB] push button to move between therapy
parameters. Density and area values are preset. Frequencies can be modified only in the range of Nogier’s
frequencies. Power and time are automatically recalculated. Start therapy by pressing probe push button.
2.5.5.1 Table of frequencies for laser acupuncture according to Dr. Nogier
Frequency [Hz] Use and Application Area
U 1.14 Universal – whole body
A 2.28 Chaotic changes – tissues and transitional epithelium, body openings (mouth, nose,
eyes, …)
B 4.56 Nutritive effect – chest and abdominal cavity
C 9.12 Mesenchyma – limbs, gluteal area, muscles and ligaments, kidneys, genitals
D 18.3 Metabolism – joints of cerebral symmetric organs
E 36.5 Nerve tissue –peripheral parts of spine, back and top parts of head and neck
F 73.0 Psychosomatic disorders – subcortical areas, lower parts of the face
G 146.0 Emotional diseases – brain cortex, nervous system, frontal and temporal parts of head,
lateral parts of nose
2.5.6 EAV (Frequencies according to Dr. Voll)
After entering EAV, a list of frequencies will be displayed. Use the [ + ] and [ - ] arrows to move between the
frequencies. Press the [ENTER] push button to select required frequency. Press the [ESC/STOP] push button to
return to previous menu. When required frequency is selected, therapy parameters will be displayed:
D e n s . _2.0J/cm
2
F r e q u Ency1.70Hz
A r e a 1.0cm2Pwr30mW
T i m e 1:23Ready
Use the [ + ] and [ - ] arrows to adjust therapy parameters. Use the [TAB] push button to move between therapy
parameters. Density and area values are preset. Frequencies can be modified only in the range of EAV frequencies.
Power and time are automatically recalculated. Start therapy by pressing probe push button.
Press the [ESC/STOP] push button to return to previous menu.
Page 10 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.5.6.1 Table of EAV frequencies EAV (according to Dr.Voll)
Frequency [Hz] Use and Application Area
1.20 Tachycardia
1.70 Abscess, furunculosis
1.75 Psychostimulative and sympaticostimulative effect
9.20 Eczema due to renal insufficiency, diastolic hypertension, renal insufficiency
9.35 Paralysis
9.40 Cystitis, disorders of prostate, disordes of colon, bronchitis, eczema due to disorders of
lungs or vessles, gout, joint ache, gastric disorders (including ulcer), parsis, varices,
haemorrhoids
9.45 Tonsillitis, angina pectoris, disorders of hypophysis and gonads
After entering the required medical field, the following menu will be displayed:
DENTISTRY
F i r s t char:_
Enter the first letter of desired diagnosis name and press the [ENTER] push button to enter the list of diagnoses. Use
the [ + ] and [ - ] arrows to move between diagnoses names and press the [ENTER] push button to select desired
diagnosis. After selecting desired diagnosis, preset therapy parameters will be displayed. Press the [ESC/STOP]
push button to return to previous menu.
Page 11 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
D e n s . _2.0J/cm
2
F r e q u ency5.00Hz
Ar e a 1.0cm2Pwr30mW
T i m e 1:23Ready
Use the [ + ] and [ - ] arrows to adjust therapy parameters. Use the [TAB] push button to move between therapy
parameters. Start therapy by pressing probe push button.
Press the [ESC/STOP] push button to return to previous menu.
2.6 Programme selection (push button [PROG])
When pressing the [PROG] push button the instrument will enter the Preset Programme Menu. It is then possible to
select a preset therapy programme. A progamme number consists of a medical field number (first two numbers) and
a diagnosis number (second two numbers).
P r o g 0400
A c n e
D en s . 0.4J/cm
2
F r e qu ency5.50Hz
Use the [ + ] and [ - ] arrows to set desired programme number. Use the [TAB] push button to move between the
medical field number and diagnosis number. Press the [ENTER] push button to enter therapy parameters of desired
programme. Press the [ESC/STOP] push button to return to previous menu. When the desired programme has been
entered, therapy parameters will be displayed.
D en s . _0.4J/cM
2
F r e qu ency5.00Hz
A r e a 1.0cm2PwR16mW
Ti m e 0:30Ready
Use the [ + ] and [ - ] arrows to adjust therapy parameters. Use the [TAB] push button to move between therapy
parameters. Start therapy by pressing the probe push button. Press the [ESC/STOP] push button to return to
previous menu.
2.7 Manual setting (push button [MAN])
When pressing the [MAN] push button, the instrument will enter the Therapy Prameters Menu and it is then possible
to set therapy parameters manually.
D en s . _1.0J/cm
2
F r e qu encyContinual
Ar e a 1.0cm2Pwr30mW
Ti m e 0:33Ready
Use the [ + ] and [ - ] arrows to adjust therapy parameters. Use the [TAB] push button to move between therapy
parameters. Start therapy by pressing the probe push button. Press the [ESC/STOP] push button to return to
previous menu.
Page 12 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.8 User menu (push button [MENU])
When pressing the [MENU] push button, the instrument will enter the System Menu and it is then possible to select
from the following list.
Patient search
Patient list
User diagnoses
System setup
Acupuncture
Probe test
> Pa t i ent search
Pa t i ent list
U se r diagnases
Sy s t em setup
Press the [ + ] or [ - ] arrows to move between displayed settings. Press the [ENTER] push button to select desired
setting. Press the [ESC/STOP] push button to return to previous menu.
2.8.1 Patient search
Enter Patient Search when searching for patient file with saved therapy parameters. If there is no patient file in
memory, “Memory free” will be shown on display. If there are less than 6 patient files in memory, a list of these
patient files in alphabetical order will appear on the display. If there are 6 or more patient files in memory, the
following menu will be displayed:
PA T IENT LIST
F i r s t char:_
Type the first letter of desired patient name and press the [ENTER] push button to enter the list of patients. Use the
[ + ] and [ - ] arrows to move between patient names and use the [ENTER] push button to select desired patient.
After selecting desired patient, patient therapy parameters will be displayed.
D e n s . _0.9J/cm
2
F r e qu encYContinual
A r e a 1.0Cm2Pwr30mW
T i m e 1:39Ready
Use the [ + ] and [ - ] arrows to adjust therapy parameters. Use the [TAB] push button to move between therapy
parameters. Start therapy by pressing the probe push button.Press the [ESC/STOP] push button to return to
previous menu.
Page 13 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.8.2 Patient List
When entering Patient List, the following list will be displayed:
> N ew patient
E d i t patient
De l e te patient
Use the [ + ] and [ - ] arrows to move between displayed settings. Press the [ENTER] push button to select desired
file. Press the [ESC/STOP] push button to return to previous menu.
2.8.2.1 New Patient
When creating a new patient, enter New Patient by using the [ENTER] push button. After entering New Patient, the
following menu will be displayed:
Pa t i e nt name:
_
Enter the patient name and press the [ENTER] push button to save it in memory.
After pressing the [ENTER] push button, the therapy parameters will be displayed.
D e n s . _0.9J/cm
2
F r e qu encyContinual
A r e a 1.0cm2Pwr30mW
T i m e 1:39Edit
Use the [ + ] and [ - ] arrows to adjust therapy parameters. Use the [TAB] push button to move between therapy
parameters. Press the [ENTER] push button to save adjusted data. After pressing the [ENTER] push button, “Data
saved“ will be shown on the display.Press the [ESC/STOP] push button to return to previous menu.
2.8.2.2 Edit Patient
Enter Patient Search when searching for patient file with saved therapy parameters. If there is no patient file in
memory, “Memory free” will be shown on display. If there are less than 6 patient files in memory, a list of these
patient files in alphabetical order will appear on the display. If there are 6 or more patient files in memory, the
following menu will be displayed:
PA T IENT LIST
F i r s t char:_
Type the first letter of desired patient name. Use the [ENTER] push button to enter the list of patients in alphabetical
order. Use the [ + ] and [ - ] arrows to move between patient names. Use the [ENTER] push button to select desired
patient. After selecting desired patient, patient therapy parameters will be displayed.
Page 14 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
D e n s . _0.9J/cm
2
F r e qu encyContinual
A r e a 1.0cm2Pwr30mW
T i m e 1:39Edit
Use the [ + ] and [ - ] arrows to adjust therapy parameter. Use the [TAB] push button to move between therapy
parameters. Press the [ENTER] push button to save adjusted data. After pressing the [ENTER] push button, “Data
saved“ will be shown on the display.Press the [ESC/STOP] push button to return to previous menu.
2.8.2.3 Delete Patient
Enter Patient Search when searching for patient file with saved therapy parameters. If there is no patient file in
memory, “Memory free” will be shown on display. If there are less than 6 patient files in memory, a list of these
patient files in alphabetical order will appear on the display. If there are 6 or more patient files in memory, the
following menu will be displayed:
PA T IENT LIST
F i r s t char:_
Type the first letter of desired patient name and press the [ENTER] push button to enter the list of patients in
alphabetical order. Use the [ + ] and [ - ] arrows to move between patient names and the [ENTER] push button to
select desired patient. After selecting desired patient, following setting menu will be displayed:
D A T A WILLBEDELETED
Co n t inue?Yes
Press the [ENTER] push button to delete data. After pressing the [ENTER] push button “Data deleted“ will be shown
on the display. Use the [ + ] and [ - ] arrows to move between the Yes or No settings. Press the [ESC/STOP] push
button to return to previous menu.
2.8.3 User diagnoses
When entering User diagnoses, the following list will be displayed:
> N ew D Ignosis
E d i t Diagnosis
De l e TE diagnosis
Use the [ + ] and [ - ] arrows to move between settings. Press the [ENTER] push button to select desired setting.
Press the [ESC/STOP] push button to return to previous menu.
Page 15 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.8.3.1 New Diagnosis
When creating a new diagnosis, enter New Diagnosis by using the [ENTER] push button. After entering New
Diagnosis the following menu will be displayed:
Th e r a py name:
_
Enter the diagnosis name and press the [ENTER] push button to save it in memory.
After pressing the [ENTER] push button, therapy parameters will be dispayed.
D e n s . _0.9J/cm
2
F r e qu encyContinual
A r e a 1.0cm2Pwr30mW
T i m e 1:39Edit
Use the [ + ] and [ - ] arrows to adjust therapy parameters. Use the [TAB] push button to move between therapy
parameters. Press the [ENTER] push button to save adjusted data. After pressing the [ENTER] push button, “Data
saved“ will be shown on the display.Press the [ESC/STOP] push button to return to previous menu.
2.8.3.2 Edit diagnosis
If there are no diagnoses in memory, “Memory free” will be shown on display. If there are less than 6 diagnoses in
memory, a list of these patient files in alphabetical order will appear on the display. If there are 6 or more patient files
in memory, the following menu will be displayed:
U SE R DIAGNOSES
Fi r s t char:_
Type the first letter of desired diagnosis name and press the [ENTER] push button to enter the list of diagnoses. Use
the [ + ] and [ - ] arrows to move between diagnoses and the [ENTER] push button to select desired diagnosis. After
selecting diagnosis, diagnosis parameters will be displayed.
D e n s . _0.9J/cm
2
F r e qu encyContinual
A r e a 1.0cm2Pwr30mW
T i m e 1:39Edit
Use the [ + ] and [ - ] arrows to adjust therapy parameters. Use the [TAB] push button to move between therapy
parameters. Press the [ENTER] push button to save adjusted data. After pressing the [ENTER] push button, “Data saved“ will be shown on the display. Press the [ESC/STOP] push button to return to previous menu.
Page 16 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.8.3.3 Delete Diagnosis
If there are no diagnoses in memory, “Memory free” will be shown on display. If there are less than 6 diagnoses in
memory, a list of these patient files in alphabetical order will appear on the display. If there are 6 or more patient files
in memory, the following menu will be displayed:
U SE R DIAGNOSES
Fi r s t char:_
Type the first letter of desired diagnosis name and press the [ENTER] push button to enter the list of diagnoses. Use
the [ + ] and [ - ] arrows to move between diagnoses and the [ENTER] push button to select desired diagnosis. After
selecting diagnosis, following menu will be displayed:
D A T A WILLBEDELETED
Co n t inue?Yes
Press the [ENTER] push button to delete data. After pressing the [ENTER] push button “Data deleted“ will be shown
on display. Use the [ + ] and [ - ] arrows to move between the Yes or No settings. Press the [ESC/STOP] push button
to return to previous menu.
2.8.4 System Setup
When entering System Setup, the following list will be displayed:
Entry Code Changes
Backlight
DF (duty factor) Change
Acupuncture Programme
Keyboard Beep
Start with
Language
Door Activity
Battery Formatting
Default Settings
> En t r y code change
Ba c k light
DF c hange
A cu p . progRamme
Press the [ + ] or [ - ] arrows to move between displayed settings. Press the [ENTER] push button to select desired
setting. Press the [ESC/STOP] push button to return to previous menu.
Page 17 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.8.4.1 Entry Code Change
The manufacture code is preset as 0000. It is possible to change the code by entering Entry Code Change.
After entering Entry Code Change, the following menu will be displayed:
BTL-2000
T E R A PEUTICLASER
EN T E R CODE:0000
Enter the preset safety code and press the [ENTER] push button. After entering the preset safety code, the following
setting menu will be displayed:
BTL-2000
T E R A PEUTICLASER
N EW C ODE: 0000
Use the [ + ] and [ - ] arrows to enter the new code. Use the [TAB] push button to move cursor to next position. Once
the code has been entered, press the [ENTER] push button. After pressing the [ENTER] push button “Data saved“
will be shown on display. Press the [ESC/STOP] push button to return to previous menu.
2.8.4.2 Backlight
When entering Backlight, the following setting menu will be displayed:
Ba c k l ight sEtting
0 % 50%100%
M M M M M MMMMMMMMMMM
Use the [ + ] and [ - ] arrows to adjust backlight intensity. Use the [ENTER] push button to store new data. After
pressing the [ENTER] push button “Data saved“ will be shown on display. Press the [ESC/STOP] push button to
return to previous menu.
Backlight intensity is relative to the energy consumed; the operation time of the battaries is influenced by the
backlight intensity. We recommend not setting this intensity too high.
2.8.4.3 DF change
If the instrument is operating in pulse mode (it means the frequency is not set to 0) it is possible to adjust the Duty
Factor of the output signal. Duty factor (%) defines the ratio of impulse length to period length. The preset value is
80%, and can vary from 10% to 90%. The instrument recalculates the correlating time and power in response to the
changing of the mean value of the laser beam. DF value in frequency mode has significant influence on the
calculated time. If the time is too long, increase the DF and the time length will automatically decrease. The following
figure compares two different duty factors with other conditions being constant. If DF is higher, the laser irradiates a
higher mode (average) and the time length shortens.
Page 18 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.8.4.3.1 Comparison of Two Different Duty Factors
After entering the DF change setting, the following menu will be displayed:
D u t y factor
8 0 %
Use the [ + ] and [ - ] arrows to adjust Duty Factor. Use the [ENTER] push button to store new data. Press the
[ESC/STOP] push button to return to previous menu.
2.8.4.4 Acupuncture Programme
After entering the Acupuncture Programme setting, the therapy paramters will be displayed:
D e n s . _0.9J/cm
2
F r e qu encyContinual
A r e a 1.0cm2Pwr30mW
T i m e 1:39Edit
Use the [ + ] and [ - ] arrows to adjust therapy parameters. Use the [ENTER] push button to store new data. After
pressing the [ENTER] push button “Data saved“ will be shown on display. Press the [ESC/STOP] push button to
return to previous menu.
2.8.4.5 Keyboard Beep
After entering the Keyboard Beep setting, the following menu will be displayed:
Ke y b o ard
Sw i tched on
Use the [ + ] and [ - ] arrows to move between the ON and OFF settings. Use the [ENTER] push button to store new
data. Press the [ESC/STOP] push button to return to previous menu.
2.8.4.6 Start With
After entering the Start With setting, the following list will be displayed. From this list it is possible to preset the
instruments introductory menu. This menu appears after the entering of the safety code.
Page 19 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
Main menu
Diagnoses – Selection according to branches and diagnoses
Program – Selection from the programs menu
Manual – Manual setting
Menu – Selection from menu
Pacient searching
> Ma i n
D i a g
P r o g
M a n
Use the [ + ] and [ - ] arrows to move between Start With settings. Use the [ENTER] push button to store new data.
After pressing the [ENTER] push button “Data saved“ will be shown on display. Press the [ESC/STOP] push button
to return to previous menu.
2.8.4.7 Language
After entering the Language setting, the following list will be displayed. From this list it is possible to preset the user
language.
English Czech… (according to default set)
> C e s k Y
E n g l Ish
Use the [ + ] and [ - ] arrows to move between Cesky / English settings. Use the [ENTER] push button to store new
data. Press the [ESC/STOP] push button to return to previous menu.
2.8.4.8 Door Activity
After entering the Door Activity setting, the following menu will be displayed. This programme allows the blocking of
the instrument while the door is open.
D o o r Activity
Cl o Sed
Use the [ + ] and [ - ] arrows to move between the Closed and Open settings. Use the [ENTER] push button to store
new data. Press the [ESC/STOP] push button to return to previous menu.
Page 20 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.8.4.9 Battery Formatting
Use this setting for the discharging of the NiCd accumulators.
Instrument batteries are NiCd accumulators, therefore it is necessary to format them approximately once
every two months according to frequency of use. This function prolongs the lifetime of the accumulators.
Ba t t e ry formatting
Co n t inue?Yes
Press the [ + ] or [ - ] arrows to move between the Yes and No settings. Use the [ENTER] push button to store new
data. Press the [ESC/STOP] push button to return to previous menu.
Warning! After entering the Yes setting, formatting will begin and the only possibility of terminating formatting is to
switch off the instrument. If the mains adapter is not connected, the instrument emits an accoustic signal and
“Connect adapter“ will be displayed.
After discharching, accumulators are automatically recharched. This process lasts approximately 4 hours. In the case
of power supply interruption for a period greater than 1 minute, all formatting processes will be restarted after
restoration of power supply.
2.8.4.10 Default settings
After entering Default Settings, the following menu will be displayed. From this menu it is possible to restore
the instruments default settings.
R e s t o re settings
C o n t inue?Yes
Press the [ + ] or [ - ] arrows to move between the Yes and No settings. Use the [ENTER] push button to store new
data. Press the [ESC/STOP] push button to return to previous menu.
Default setting are as follows:
Keyboard Beep - on
Door Activity - close
Backlight - maximal
Language - English
Acupuncture Program - default values
Battery Capacity - default values
Preset user files, user diagnoses and safety code are not deleted by the restoring of default settings.
2.8.5 Acupuncture
The BTL-2000 instrument offers two methods for locating acupuncture points:
1) Locating acupuncture points using probe with acupuncture attachment (not included in standard accesories):
Skin resistance is measured by an appropriate course of tension 0.5 V (at 600 kΩ). An acupuncture point can
be recognized by a local decrease in skin resistance and can be seen in % on the bar-graph and is
accompanied by an accoustic signal. Unscrew the brass cover from the probe head and replace it with the
acupuncture attatchment. On the lower panel of the instrument, connect the probe with acupuncture attachment
to Probe Output (2) and the chrome-plated patient pin to the Acupuncture Output (15). After the acupuncture
point is located, press the push button on the probe to begin therapy. This is only possible with Nogier
frequencies – see chapter “Performing Therapy”.
Page 21 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
NOTE: The acupuncture attachment decreases probe aperture approximately 20% resulting in a reduced probe
output. The instrument automatically calculates this reduction and corrects the output to a preset laser beam
density.
2. Acupuncture Set (not included in standard accesories)
This set contains acupuncture probe with spring-mounted needle, chrome-plated patient pin, and connection
cable. This connection cable consists of a connector on one end, identical to the connector of the laser probe
and on the other end two different color banana jacks. For measuring acupuncture points, initially connect the
connector cable end to Probe Output (2). Attach the reference chromium-plated electrode to the red banana
plug, which the patient will hold tight in his/her hand. Attach the acupuncture probe with needle to the green
banana plug. This probe contains a special spring that ensures good contact with the skin without using
excessive pressure. The instrument will then detect the acupuncture set and begin to measure skin resistance.
The resistance of the skin is measured using 0.5 V (600 kΩ). This level of generated electricity is directly
proportional to the resistance of the skin. The Bar graph displays skin resistance in %. As you begin to pass the
spring mounted needle over the skin, the acupuncture points are defined by a local decrease in voltage
resistance, therefore noticeable by an increased of frequency on the Bar graph and an increase in acoustic
tone. When the acupuncture point is located, disconnect the needle probe, replace it with the laser probe and
desired therapy can be started. The procedure of therapy is described in chapter “Performing therapy”.
A CUPUNCTURE
Pr o b e modAuto
0 % %30%
Press the [ + ] a [ - ] arrows to move between levels of instruments resistance sensitivity. The automatic mode
(AUTO) is then changed to (1), (2), (3) or (4). Setting (4) represents the most sensitive setting for patients with very
dry skin. Setting (1) represents the least sensitive setting for patients with very wet skin. Use the [ENTER] push
button to store new data. The therapy parameters will then be displayed:
D e n s . _0.9J/cm
2
F r e qu encyConTinual
A r e a 1.0cm2Pwr30mW
T i m e 1:39REady
Press the [ + ] or [ - ] arrows to adjust therapy parameters. Press the [TAB] push button to move between therapy
parameters. Press the [ENTER] push button to store new data. After pressing the [ENTER] push button “Data
saved“ will be shown on display. Press the [ESC/STOP] push button to return to previous menu.
2.8.6 PROBE TEST: Optical Power of Probe
This function measures probe output and the status of the probe diode. BTL-2000 automatically tests the probe
performance and corrects the power for optimal output (only for the BTL probes).
The warning messages ATTENTION and LASER IRRADIATION will be displayed followed by the message PROBE TEST. Probe parameters will be automatically displayed. Connect probe without cover to the TEST enter on the lower
panel of the instrument. Remove the brass cover and insert probe into laser optical power meter “TEST” (14). Using
probe push button to begin laser emmision. Display will show actual measured probe power, which is dependent on
the angle of the probe. Terminate laser emmision using probe push button.
PROBE TEST
T y p e 685nm/30mW
O u t put26mW
Page 22 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.9 Therapy
2.9.1 Display during therapy
The instrument continuously checks for laser probe connection. When a probe is connected the parameters of laser
therapy will be displayed. In the case that probe is not connected, no therapy parameters will be displayed.
Never
disconnect probe or switch off instrument while irradiation is in process!
The instrument displays the probe status in the lower right corner of display:
No probe - no probe connected
Ready - instrument ready for irradiation
Wait - period before Ready mode
In use - therapy in process
Pause - therapy interrupted by probe push-button (press STOP to cancel)
Door - remote control; open door
Edit - editing data (names , diagnoses, …)
2.9.2 To start therapy
Irradiation can begin when a probe is attached and Ready is signalled in right lower corner on the display. The
display shows preset parameters. Take care when handling the probe in order to avoid the unwanted pressing of
probe push button. If all parameters are properly set for therapy, hold the probe as you would a pencil and point the
laser aperture at the area you want to irradiate. Press the push button on the probe (fig. 1, Appendix A) for one
second. When irradiation begins, release the push button and irradiation will continue automatically. Time countdown
for the therapy begins when the laser is activated. In use is signalled. Each second of irradiation is indicated by a
slight ticking sound emitted by the instrument. According to the appropriate therapy, either hold the probe in one
place or move it over the designated area at the appropriate distance. When set therapy time has elapsed, the laser
automatically stops irradiating and returns to its initial mode.
In case of an emergency situation, switch the laser off by pressing the [ESC/STOP] push button.
2.9.3 To interrupt therapy
2.9.3.1 With the Probe Push button
If the probe push button is pressed during irradiation therapy, the laser switches off and the message “Pause“ will be
signalled on the display and the instrument will enter PAUSE mode. Continuation of the therapy is possible after
pressing the push button on the probe (for two seconds).
Note: Parameters cannot be changed in PAUSE mode. To adjust parameters, press the [ESC/STOP] push button
and adjust the parameters as needed. Press the probe push button again to reactivate irradiation.
2.9.3.2 Emergency [ESC/STOP] push button
The [ESC/STOP] push button functions as an emergency stop. When pressed the laser switches off immediately and
therapy stops. Irradiation cannot begin again until the “Wait“ mode (~ 2 sec.) in completed and the instrument has
returned to its initial mode. Pressing the [ESC/STOP] push button during a therapy session returns the instrument to
its initial mode.
2.9.4 To Continue Therapy
If the therapy session is interrupted by pressing the probe push button, press again (for 1 second) to resume therapy.
Keep the push button pressed until the probe begins irradiating.
2.9.5 To terminate therapy
Press the probe push button to interrupt therapy and then press the [ESC/STOP] push button. The instrument will
then return to the same parameters before therapy was initiated.
2.10 Power Supply: Mains and Battery
The instrument may be supplied from built-in batteries or from a mains power adapter. Batteries, when fully charged,
allow the instrument to function for 6 - 10 hours depending on the number of therapies and probe outputs. If batteries
begin to be low, the battery signal in the upper right corner will be displayed. From this moment the instrument
guarantees approximately 15 minutes of functioning. This period may be shorter if therapy is performed. When
batteries are empty the instrument will automatically switch off. If you want to continue irradiation, attach the mains
power adapter and switch the instrument off and then on. The instrument is now being supplied by the mains
power adapter and the batteries are being recharged.
Recharging is indicated by the battery sign with an arrow (not displayed in some menus). Battery charging is occuring
any time the adapter is attached to the instrument, even if the instrument is switched off. There is no risk of overcharging the batteries, when the batteries are fully charged, the charging mode switches to a waiting mode (the
Page 23 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
battery sign with an arrow disappears). The recharging cycle of totally empty bateries lasts approximately 260
minutes.
You may continuously switch between mains adapter and batteries; recharging connector may be attached or
removed during the use of instrument, without having to switch the instrument off. Nevertheless we recommend to
discharge the accumulators completely and the to recharge them. The instrument can be charged overnight without
any danger. Use only the original adapter, BTL-282, supplied by the manufacturer.
Accumulators used are NiCd and it is necessary discharge them occasionally (once in 14 days or once in
month, depending on how often the instrument is used). Have the instrument switched on until it switches off
(e.g., have the instrument on during night and attach the adapter in the morning). This will prolong lifetime of
accumulators.
2.10.1 Sparing Batterires
Display lighting decreases continuously when the instrument is not in use. Press any push button on the instrument
front panel (not on the probe!) to renew lighted display.
2.11 Optical Attachments
Optical attachments direct laserbeam to treat difficult to reach places. Three optical attachments for dentistry,
gynecology and ENT are included with the BTL-2000.
2.11.1 Dentistry Attachment
When using a 685 nm wavelength with the dentistry attachment, it is necessary to increase the density of a preset
therapy by 20% and by 10% when using a 830 nm wave length.
2.11.2 Gynaecological Attachment
When using a 685 nm wavelength with the gynaecology attatchment, it is necessary to increase the density of the
preset therapy by 30 % and by 20 % when using a 830 nm wavelength.
2.11.3 ENT Attachment
When using a 830 nm or 685 nm wavelength with the ENT attatchment, it is necessary to increase the density of the
preset therapy by 20 %.
Sterilize attachments before each use (water temperature 120°C, time 10 minutes). Do not sterilize plastic cap.
Note: The attachments are not included in standard accesorries.
Page 24 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.11.4 Positioning an Attachment
Unscrew the brass cover from the probe head and replace it with plastic cap. Do not tighten plastic cap until optical
attachment is added. Once optical attatchment is added, tighten plastic cap to end of probe. To change optical
attachments, loosen plastic cap by one turn, exchange optical attachments and retighten plastic cap.
2.12 Maintenance
It is recommended to use and/or store instrument in a dry, dust free environment. Instrument should never be placed
in water. Before every use check instrument for damages and if there are damages present do not use.
Cleaning the surface of the instrument and its accessories:
Surfaces of the instrument and accessories should be cleaned using a cloth dampened with water or a diluted
detergent. Never use chemicals that contain alcohol, ammonia, benzene, thinners, etc..
After each application of a therapy, it is recommended that the laser probe lens be wiped with a cotton cloth.
Recommended antiseptic agents: 0.25% - 1% Lysoformin, 2% Sekusept pulver, 2% Chloramin B or 0.5% Persteril.
Converting Network Voltage
Electricity converter is located on the front panel of the adapter in a black round case. Before converting network
voltage, make sure the mains power switch is turned off. Unplug the power cable from the instrument and from the
electrical wall socket. Be sure that the fuses in the instrument are for the voltage needed. See manufacturer label
under fuses. In a few cases the correct fuses for the power desired were already installed. With the appropriate
screwdriver rotate the switch until the arrow shows the same value as the electrical network. Only qualified personnel
should perform this procedure! In case of any misunderstanding, please contact your BTL service representative.
The producer of the instrument is not responsible for any damages caused by incorrect voltage conversion.
Probe durability
During use period there is aging of laser diode. BTL-2000 measures and displays the laser power on the display. The
instrument automatically adjusts to maximum power according to the state of the diode. When the power decreases,
due to probe malfunction or diode aging, to less than 40% total power, we recommend replacement of the probe.
Instrument manages actual radiation output according to the power of the laser diode. Due to power decrement of
optical interfaces inside the probe, the actual power output is lower than stated on the probe label which
characterizes the type of probe laser diode.
It is recommended to have the instrument and laser probe tested approximately every 30 months. This must
be done by the manufacturer based on the request of the user.
Transport and Storage
To avoid damage, transport instrument only in its original packaging. Remove power cable, probe and patient
electrodes before packing. Avoid rough handling. Store and transport instrument only under the defined conditions
specified in technical parameters.
If item is not functional or repairable it is possible to liquidate under certain circumstances. Item does not contain
toxic materials. Disposal of item will not lead to environmental degradation.
2.13 Safety of Instrument
Do not connect non BTL laser probes.
The internal electronics of the BTL-2000 have received the BF Level of Safety certification. The laser and
acupuncture probes have also received the BF Level of Safety certification. The producer of these products
guarantees the quality of this certification and the quality of these products under the circumstances that only BTL
lasers and acupuncture probes are used with this instrument.
2.13.1 Remote control ( DOOR )
There is remote control connector on the lower panel of the instrument (DOOR) (see chapter Outward Appearance)
for attachment of safety circuit. If the signal on the connector is active (e.g., open door), the instrument will not
irradiate. If you attempt to start therapy with the probe push button, the laser will not be turned on and the instrument
will go to Pause mode. The display will show Door accompanied by a sound signal. Any 4.5 - 30 V safety device can
be attached to the connector. These are attached between the "door contact signal" and "zero potential" connectors.
Electric current needed is approximately 10 mA. Description of Lineouts is on Figure No.3, Appendix A.
2.13.2 Use of Safety Accessories
Both therapist and patient are obligated to wear safety eyewear to protect their eyes from laser beam. Safety eyewear
must comply with applicable rules and requirements.
Page 25 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
2.13.3 Safety Warnings
• Before the initial turning on of the instrument, read manual carefully and observe all operation instructions.
• Check if parameters of electricity correspond with the technical parameters of the instrument and if the voltage
switch on the rear panel of instrument is in the correct position in accordance with mains power parameter. If you
are not sure whether the electrical network is in optimal condition, an inspection by a certified electrical inspector
is required.
•Instrument should be used in proper work environment. It should not be used in an environment where there is
danger of explosion and/or water damage.
• During use, storage and transport of BTL-2000, no toxic radiation is emitted.
• The manufacturer or distributor must instruct every user of BTL-2000 regarding service, maintenance, safety
principles, and operation of instrument.
• Always place the instrument on a solid, horizontal surface that does not exceed 15 degrees. Cooling vents,
located on the rear panel, must at all times be free of obstruction.
• Do not place instrument in proximity of devices that emit strong electric, magnetic, or electro-magnetic fields.
(motors, transformers, x rays, etc.). This can cause magnetic interference. In case of any interference, place the
instrument further from its source or contact your BTL service representative.
• Do not place instrument in direct sunlight.
• Only use the instrument if no mechanical or water damage is noticed.
• If instrument is used in discrepancy with this manual or despite functional mistakes noted in this
manual, the user is responsible for all damages caused by instrument.
• Danger! Removing instrument casing might result in electrical shock!
• All materials and accessories which are in direct contact with patient must be followed by standard health
practices to avoid skin harm, allergies, toxicity, genotoxicity, and carcinogenic effects in accordance with ISO
10993-1, ISO 10993-3, ISO 10993-5.
•The user is responsible for the use of materials and accessories that are not supplied by BTL. This includes the
cleaning and disinfecting liquids.
•It is advisable to thoroughly check the instrument once a month for loose cables, damaged diodes and damaged
display functions.
•It is recommended to have the laser callibrated and the instrument checked for internal malfunctions
approximately every 30 months. This must be done by the manufacturer based on the request of the
user.
• After application of therapy it is necessary to disinfect probe.
• Do not irradiate sensitive areas such as eyes, head, thyroid gland and other endocrine glands.
• During therapy it is necessary for the patient and the doctor to use the supplied safety eyewear.
• Application of therapy is terminated upon pressing the [ESC/STOP] pushbutton not the mains power switch.
• Between the turning off and turning on of instrument, it is necessary to wait a minimum of 3 seconds.
• Avoid using the instrument in areas that contain flammable and explosive compounds.
• Display warning labels outside the immediate laser workplace. Connect the remote door-blocking device if
necessary. Workplace must comply with hygienic standards.
• Do not disconnect the probe or turn the instrument off while irradiating. Do not attatch non-BTL accessories to
laser probe or acupuncture connectors.
• Handle probes with care! Probes are not water-resistant.
Page 26 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
• Do not remove instrument casing.
• This instrument emits a class 3 B laser beam. During therapy with this instrument, you must follow all
instructions written in this user manual. Improper therapy manipulation can cause dangerous irradiation
and damage to eyesight! The manufacturer will not accept responsibility of damage due to improper use
of the intrument.
For all damages in this case, the user is responsible.
•Protect your workplace against direct hit of laser beam.
2.14 Useful addresses
The product is manufactured in accordance with the EU Medical Devices Directive by :
BTL Industries Ltd.
Suite 401 Albany House
324-326 Regents Street
London, W1B 3BL United Kingdom
E-mail: sales@btlnet.com
http://www.btlnet.com
For service, please contact service department at service@btlnet.com
2.15 Warranty
The Manufacturer of this product warrants the product to be free from defects in workmanship and material for a
period of twelve months after the date of shipment from the factory. This warranty excludes any disposable items and
accessories, including, but not limited to cables or leads, power cords and electrodes. The manufacturer agrees to
correct such defects without charge, or at its option to replace the item with a comparable model. To register and be
eligible for warranty service, you must send or fax the fully completed warranty registration form within 30 days of
installation. All costs of shipment are the responsibility of the purchaser. Damage to any part such as by accident or
misuse or improper installation or by use of any accessories or abrasive material not produced by the Manufacturer is
not covered by this warranty. Because of varying climatic conditions, this warranty does not cover any changes in
finish, including rusting, pitting, corrosion, tarnishing or peeling. Servicing performed by unauthorized persons render
this warranty invalid. There is no other express warranty. The Manufacturer hereby disclaims any and all warranties,
including but not limited to, those of merchantability and fitness for a particular purpose to the extent permitted by
law. The duration of any implied warranty which cannot be disclaimed is limited to the time period as specified in the
express warranty. The Manufacturer shall not be liable for incidental, consequential, or special damages arising out
of, or in connection with product use or performance except as may be otherwise accorded by law.
This warranty may differ from the warranty terms and conditions provided by your supplier and by applicable laws in
your country.
.
2.16 Accessories
Do not connect non BTL accessories to the instrument.
2.16.1 Standard Accessories
Adapter with mains power cable (BTL-282)
BTL-2000 – control instrument
Warning label Laser Workplace
2.16.2 Additional Accessories
Laser probe with desired power and wavelength
Acupuncture set (electrode for patient, locating needle probe)
Optical attachments for dentistry, gynaecology and ENT
Acupuncture attachment for laser probe
Connector for Remote Control
Safety eyewear
Page 27 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
3 TECHNICAL PARAMETERS
3.1 Technical parameters BTL-2000
General Parameters
- Weight 700g ±50g
- Dimension 234 x 132 x 33 mm ±3%
- Cover (by EN 60529) IP20
- Class (by EN 60601-1) II (with inner power source and additional external source)
- Class (by 93/42EHS) Iib
- Type of applied part BF
- Display LCD display with backlight
- Modes of operations Continual
Operating conditions:
- Temperature 10 °C to + 40 °C
- Relative humidity 30 % to 75 %
- Atmospheric pressure 700 hPa to 1060 hPa
- Position N/A
Storage conditions
- Temperature - 10 °C to + 55 °C
- Relative humidity 25 % to 85 %
- Atmospheric pressure 650 hPa to 1100 hPa
- Position N/A
- Maximum Duration 1 year
Transport conditions
- Temperature - 10 °C to + 55 °C
- Relative humidity 25 % to 85 %
- Atmospheric pressure 650 hPa to 1100 hPa
- Position N/A
- Maximum Duration 1 year
Power supply
- external Adapter BTL-282 (see 3.2 Technical parameters ADAPTER
BTL-282)
Battery Supply
- battery type NiCd accumulator
- number 8
- total voltage 9.6 V
- capacity 950 mA
- min. operational time 1.5 h (probe 300 mW + backlight)
Voltage Insulation
- Mains Power to Patient Circuit
Insulation
- Frequency * *** 0.0 – 10.0 Hz, step 0.1
11.0 – 100.0 Hz, step 1
110 – 1000 Hz, step 10 1100 – 5000 Hz, step 100
- Standard Deviation ± 3 % of default value
- Density * 0.1 – 10.0 J/cm2, step 0.1
11.0 – 99.0 J/cm2, step 1
1.5 kV
Page 28 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
- Standard Deviation ±20% (according to ČSN EN 60601-2-22)
- Area * 0.1 – 10.0 cm2, step 0.1
11.0 – 99.0 cm2, step 1
- Standard Deviation see beam divergency
- Power Input * 1.0 – 999 mW, step 1
- Standard Deviation ±20% (ČSN EN 60601-2-22)
- Duty factor ** 10 – 90 %, step 10%
- Standard Deviation ±1% of DF range
Time of Therapy * 00:05 –99:59, step 1min/1sec ± 3%
- Standard Deviation ±3% of adjusted value
Locating Acupuncture Points Measuring resistance of skin
- Maximum Power of Electrode 0.5 V (0.45V typical)
- Voltage Characteristics DC
- Maximum Electrical Current between
Electrodes
- Measurement Ranges 4 manual, 3 automatic
- Indicators Optic
Measuring Laser Output
- Max. Laser Power Input
- Measuring Ranges
- Standard Deviation ±15% (ČSN EN 60601-2-22)
Laser Probe
- Laser Beam Divergence
(Divergent Probe)
- Laser Beam Divergence
(Convergent Probe)
- Laser Class
- In Compliance with ČSN EN 60825-1 ČSN EN 60601-2-22
Laser Modes
- Continual
- Pulse
BTL is Designed to Comply with ČSN EN 60601-1 Appendix ČSN EN 60601-1 A2 Appendix ČSN EN 60601-1 A12 ČSN EN 33 4200 ČSN EN 60825-1 Appendix ČSN EN 60825-1 A11 ČSN EN 60601-2-22 ČSN EN 60601-1-2
Emitting laser indicator Displayed In use, sound
Ready for emission indicator Displayed Ready, sound
Not ready for emission indicator Displayed Wait (safety pause), sound
Other safety installations warning labels on cover and probe warning label on workspace door remote control connector
Power switch positions 0/1 (1 = •)
5µΑ (typical 0.5 to 3.5 µA)
500 mW / max. 15 s
2 – automatic switching
0.28 rad ±0.05 rad
0.015 rad ±0.05 rad
III.B; III.A, Visible and Invisible light spectrum
Page 29 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
3.2 Technical parameters ADAPTER BTL-282
General Parameters
- Weight – equipment incl. supply cable 0,3 kg ± 10 % (0,2 kg without supply cable)
- Dimensions (l x w x h) 135 x 72 x 32 mm ±3 %
- Cover grade according to EN 60529 IP40
- Mode of operation Continual
- Indication of operation
Operating Conditions
- Ambient temperature + 0 °C to + 40 °C
- Relative humidity 30 % to 75 %
- Atmospheric pressure 700 hPa to 1060 hPa
- Position Horizontal on legs
Storage and transport conditions
- Ambient temperature - 10 °C to + 55 °C
- Relative humidity 25 % to 85 %
- Atmospheric pressure 650 hPa to 1100 hPa
- Position N/A
Power supply
- Voltage AC 110V to 230V (±10% tj. ~99V to ~253V )
- Frequency 50 Hz – 60 Hz
- Input Max. 42 VA
- Class according to IEC 536 II
- Inlet type Mini 2 poles
- Fuses (IEC 60127-3/IV) Internal T500mA (can be exchanged only by the manufacturer)
Insulation barriers:
- mains – output (output cable, plastic
cover)
Output parameters
- output voltage 13V ± 0,5V; ripple 100mVpp at consumption of 1,5ADC
- output current 1,5A ; short-time (max. 10sec) 2A
Descripton of pictograms on the equipment applied part of the type BF
equipment with double insulation
*) The stated values are maximum. The actual values depend on the type of the connected laser generator and on the purchased
configuration of the equipment.
**) Adjustable only in the pulse mode; in the continuous mode it is always 100 %.
***) The zero frequency means the continuous mode of the laser operation.
****) Frequency is standardly adjustable within the range 0 – 1000 Hz. Upgrade of the frequency range up to 5000 Hz is a part
of the above-standard accessories of the equipment.
Green LED on the right, next to the output connector
4 kV
Page 30 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
4 USER’S GUIDE
4.1 Introduction
This guide is intended to make you familiar with performing non-invasive laser beam therapy. At the same time, it will
give you general instructions on how to perform non-invasive laser beam therapy in clinical practice.
4.2 Light; the principle of laser light and its specifics
According to classic, relative or quantum physics, the nature and qualities of light can be interpreted in a number of
ways. However, the basic nature of light itself as either a particle/corpuscular or an electromagnetic/wave mechanism
remains unreachable for our senses and is incomparable to any other levels of reality. Therefore, we can only accept
the conclusions and deductions of numerous experiments.
The specifics of laser radiation:
• totally monochromatic: maintaining only one wavelength (e.g., HeNe laser 632.8nm)
LASER 780nm BULB
Intensity
400
500 600
700 800
• totally polarized: spatial orientation of the waves on a defined area is absolute
nm
Intensity
400
500 600
700 800
nm
Linear polarization
• totally coherent: wavelength orientation in time is absolute, maxima and minima of all waves are identical in time
Page 31 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
Coherent Non-coherent
A laser beam of a high intensity therefore represents the “death ray” of science fiction authors; a laser beam can
destroy material or matter. However, a controlled laser beam in low intensities not exceeding 500 mW shows minor
effect on living organisms and tissues, but can have significant stimulative effects.
The word LASER stands for Light Amplification by Stimulated Emission of Radiation.
The first theoretical postulates on a laser beam were laid down by Albert Einstein in the early twentieth century,
describing the theoretical possibilities of a stimulated emission of light in 1916-1917. However, the first laser
instrument was not constructed until 1960, following other important discoveries in the area of the so called quantum
radiotechnique (N.G. Basov and A.M. Prochorov of the former USSR and C.H. Townes of the USA received a Nobel
Prize for these discoveries in 1964).
The medical application of laser light followed soon after. At first, only high intensity laser beams were utilized. Lasers
of this type provide a source of energy that, with its thermal effect, can destroy, cut, evaporate and cauterize tissue.
E. Mester, a Hungarian scientist, was the first to introduce the stimulative effect of a low intensity laser. He conducted
his experiments - which are still of a great importance - in the late sixties.
4.3 Types of lasers
Lasers can be divided according to several criteria:
4.3.1 Source
• solid-substance lasers (ruby laser)
• liquid-based lasers
• gas lasers – helium/neon, argon lasers
• semi-conductor (GaAs, GaAIAs) -- the most important form of non-invasive lasers of our time.
The advantages of semi-conductor lasers include a significant miniaturization of the source, robustness and
imminstrumenty to damage and the possibility to construct a laser instrument of any wavelength. Nevertheless, some
lasers still remain at experimental level.
Laser beam penetration through tissue is to a great extent dependent on the wavelength. In general, green and blue
lasers are still being developed and their clinical use is extremely limited. Most commonly used are beams in the
wavelength of red light. After 2-3mm penetration through living tissue, their intensity is reduced to half. This
predetermines red laser beams to be used especially in mucose and skin disease therapy. Since the output is
absorbed by a very small column of tissue, devices of 20mW and smaller output are usually fully adequate.
Infra-red lasers intensity is reduced to half after 2-3cm penetration through tissue. These lasers are used especially
in treating deep structures (joints, muscles, attachments). Since the output is absorbed by the column of tissue and
we want the intensity absorbed by a target area deep in the tissue to be sufficient, we recommend devices of at least
30 mW output.
Visible spectrum
Infra-red radiation,
microwaves, radio
4.3.3 Operation mode
• pulse lasers: short intervals of radiation emission -- hundreds of ns and top function of the laser is in
instruments or tens of watts.
• continuous lasers: uninterrupted emission of radiation.
• lasers with modulated operation: duration of the pulse depends on frequency and employed substance. Top
function of the laser is in instruments through hundreds of mW.
Power
Power
MODULATED PULSE
Mean
Mean
Power
CONTINUOUS
Time
Time
Mean
Time
Producers usually offer various alternatives, including the possibility to combine various modes modulated in
accordance with various parameters.
Page 33 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
4.3.4 Type of beam
• convergent: a point beam
• divergent: a widening beam that allows irradiation of a larger area without the need to move the probe over the
tissue being treated
• laser scanners: devices for irradiation of large areas
4.3.5 Laser class (defined by the output)
Values described bellow are only for user orientation.
• Class 2: less than 1mW
• Class 3A (3R, 3M): less than 5mW, output density less than 25 W/m
• Class 3B: less than 500mW
4.3.6 Use
• simple pen lasers: usually feature only a very limited choice of frequency modes. The advantage of this type of
laser is its accuracy.
• pocket lasers: designed to fit in the pocket of a physician's gown, and with the probe connected by a cable with
the body. These lasers are most common. They may employ probes of tens of mW output.
• table devices: usually connected with the mains, and with the probe connected with the device by a cable. The
advantages of this type usually include a wide choice of frequency modes and the possibility to set many different
parameters.
2
4.4 Basic physical quantities
• wavelength: one of the basic parameters of a laser beam is that it is always monochromatic. Only one
wavelength measured in nanometers (nm).
• output: usually measured in mW. A declared output is either top, i.e. amplitude of pulse, or mean, i.e. average
output of the laser. The use of the optic attachment or light conductor causes approximately a 10% loss of output.
It is always necessary to know the output level at the active point of the probe. In lasers where the beam is
transmitted by a light conductor, final output can markedly differ from source output! Every transference through a
media (air, air-light conductor optic, light conductor optic-air) lowers output.
The mean output also depends on the chosen mode (continuous or pulse). The microprocessor unit checks which
mode is employed when counting the therapy time.
• Modulation frequency: declared in the number of oscillations per second - Hertz (Hz). Sometimes added are the
characteristics of the pulses regarding how much % constitutes the active state time (employed substance or
modulation class) and the width of the pulse. Numeric frequencies are used most commonly (5Hz stimulative,
10Hz analgesic effect), Nogier frequencies are also used often (previously mainly in acupuncture), and Bahr
frequencies (acupuncture) and Voll frequencies are also used.
• Power density (dose): (J/cm
2
) the degree of consistency measured by the quantity of energy emmited per unit
area.
Below you can see a formula for the calculation of emitted dose of energy. (this number is a dose emitted to skin
surface, not deeper structures).
mean radiation output (W) x application time (s)
emitted dose of energy in J/cm
emitted surface area (cm
• DNA synthesis stimulation (via phytohemaglutinin stimulation)
• increased fibroblast activity (as far as keloids are concerned activated fibroblasts are able to perform a re-
absorption of fibrin)
• phagocytosis activation
• activation of the Na/K pump on the membrane
• activation of metabolic processes in the cellula: partly through the Na/K pump and Ca transport affection, partly
through a direct activation of the mitochondrial system
• local level changes of important mediators -- inflammatory (histamin substances, prostaglandins) or e.g.,
endorphins.
When performing an in vitro experiment, no difference can be seen between laser beams applied in continuous and
pulse modes. However, when performing an in vivo experiment, we observe marked differences between a
continuous beam and a beam adjusted to a certain defined mode. In physical therapy, this phenomenon can be
observed when performing electrotherapy or applying ultrasound. It seems that, as far as a complex organism with its
numerous information systems and feedbacks is concerned, a physical quantity modulated to a certain frequency
features one additional quality -- specific information transmitted by this frequency.
The effect of laser beam in acupuncture is totally specific. A laser beam affects tissue in a way fully comparable with
the effect of an inserted needle -- but without any pain or complications related to needle introduction. The skin above
the acupuncture point is a light conductor for the laser beam. We find it interesting that in laser acupuncture, in
contrast with other forms of laser therapy, the wavelength of a laser beam is not of a big importance (however, red
lasers are slightly more effective) and the output of the device is of no importance at all (we can even use a penlaser). We conclude that, as far as this method of application is concerned, it is not the effect of a physical quantity
itself, but the above-mentioned transmission of information that plays the push button role.
Many treatments include several effects of laser light at a time (e.g., treatment of crural ulcer combine analgesic,
stimulative, as well as bactericidal effects).
4.6 Contradictions of non-invasive laser therapy
• direct irradiation of glands with intrinsic secretion
• direct irradiation of the eye. Irreversible changes of retina may occur.
• direct irradiation of malignities and potential pre-cancerous sites
• febrile patients
• use of pulse modes or infra-red lenses on patients with seizure diseases such as epilepsy
• application to the abdomen area during gravidity
Cicatrix recens (2) 01 4.0 5.0 2nd part of therapy
Furunculus 02 2.0 1.7
Herpes (1) 03 3.0 Cont 1st part of therapy
Herpes (2) 04 3.0 5.0 2nd part of therapy
Otitis media acuta (1) 05 5.0 Cont 1st part of therapy
Otitis media acuta (2) 06 5.0 5.0 2nd part of therapy
Otitis media chronica 07 7.0 5.0
Perforatio membranae
part of therapy" and "2nd part of therapy" should be understood as two parts of one therapy session.
5.11 Acupuncture
Laser beam has a specific effect on acupuncture points, adequate to use of an acupuncture needle. The laser
acupuncture has several advantages: it is non-invasive, painless (most patients do report some non-pleasant
sensations when acupuncture needle is used), and saves time (usually, several tens of seconds are enough).
Very useful is modulation by Nogier frequencies.
For all programmes the following parameters have been preset:
- Duty factor 80%
- Area 1 cm
2
- Power is calculated according to the probe used
Density
2
[J/cm
Freq.
]
[Hz]
Comment
st
part of therapy
Page 41 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
6 APPENDIX A
Laser Aperature
Illustration 1 - Description of LASER PROBE
Direction of Laser Beam
Direction of Laser Beam
Laser Control Probe
Laser Activity Indicator
Power Cord
Laser Aperture Label
Illustration 2 – Description of REMOTE CONTROL
Page 42 z 45
BTL-2000
7 APPENDIX B
USER'S MANUAL & USER’S GUIDE
Maximum outputs:
60mW/685nm, 400mW/830nm
Laser classified by:
IEC 601-2-22:1995
IEC 60825-1:1993/A2:2001
VISIBLE AND INVISIBLE
LASER RADIATION
AVOID EXPOSURE TO BEAM
CLASS 3B LASER PRODUCT
Label 1 – This Warning Label is located on rear panel of instrument case. This label states that laser is
classified by European Norms. This instrument emits a class 3B laser beam. This laser emits visible
and invisible radiation.
STOP
STOP
STOP
STOP
STOP
LASER APERTURE
CLASS 3B
400mW 830nm
LASER APERTURE
CLASS 3B
300mW 830n m
LASER APERTURE
CLASS 3B
200mW 830nm
LASER APERTURE
CLASS 3B
100mW 830nm
LASER APERTURE
CLASS 3B
50mW 830 nm
STOP
STOP
LASER APERTURE
CLASS 3B
30mW 685nm
LASER APERTURE
CLASS 3B
50mW 685n m
Label 2 – This Class and Aperture Label is located on Laser Probe. This label states the laser class, power
and wavelength.
Page 43 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
LASER RADIATION
CLASS III. B
Label 3 – Laser Radiation Class III warning label: use in laser therapy room.
Page 44 z 45
BTL-2000
USER'S MANUAL & USER’S GUIDE
This product is manufactured in accordance with the EU Medical Devices Directive by:
BTL Industries Ltd.
Suite 401 Albany House
324-326 Regents Street
London, W1B 3BL United Kingdom
E-mail: sales@btlnet.com
http://www.btlnet.com
For service, please contact service department at service@btlnet.com
BTL Industries Limited operates a policy of continuous development. Therefore, it reserves the right to make changes and
improvements to the Product described in this manual without prior notice.
The contents of this document is provided "as is". Except as required by applicable law, no warranties of any kind, either expressed
or implied, are made in relation to the accuracy, reliability or contents of this document. BTL Industries Limited reserves the right to
revise this document or withdraw it at any time without prior notice.
.
Page 45 z 45
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.