Thank you for showing your interest and buying our device
the BTL- 5000 SWT Series. We hope you will be satisfied
and thanks to you, this device will help many patients.
All of us at BTL wish you every success with your BTL- 5000
SWT Series. We pride ourselves on being as responsive as
possible to our customers’ needs. Your suggestions and
comments are always welcome since we believe an ongoing
relationship with our customers is critically important to our
future product line. Please call or email us your suggestions.
While we would like you to start using your equipment right
away, we encourage you to thoroughly read this manual in
order to fully understand the operational features of the
BTL-5000 SWT system.
Again, thank you for being a BTL customer. In the event of
any problem, or if you require service, please make an
initial call to your local distributor, who will decide whether
to refer the problem to our office.
PAGE 2 OF 47
CONTENTS
1 GENERAL CHARACTERISTICS OF THE DEVICE....................................................................................... 5
1.1 BTL – 5000 SWT System Series .................................................................................................................. 6
1.2 Shockwave and Its Character....................................................................................................................... 6
1.3.1 Ballistic Principle of the Shockwave Generation ......................................................................................7
1.4 Biological Effects of the Shockwave ............................................................................................................. 7
1.5 Advantages of the Shockwave Treatment .................................................................................................... 8
1.6 Possible Side Effects After the Shockwave Treatment................................................................................. 8
1.7 Indications for the Shockwave Treatment..................................................................................................... 8
1.8 Contraindications to the Shockwave Treatment ........................................................................................... 9
2 INSTRUCTIONS FOR USE .......................................................................................................................... 10
2.1 Front Panel of the BTL- 5000 SWT............................................................................................................. 10
2.2 Rear Panel of the BTL- 5000 SWT ............................................................................................................. 11
2.3 Assembly and Putting into Operation.......................................................................................................... 12
2.4 OPERATING THE DEVICE ........................................................................................................................ 14
2.6 Course of Therapy...................................................................................................................................... 20
2.6.1 Start - Interruption - End of Therapy.......................................................................................................20
3.1.1 Installation of Accessories......................................................................................................................26
3.1.2 Information about Accessories ...............................................................................................................26
PAGE 3 OF 47
3.1.3 Information about Connectors ................................................................................................................26
3.2 Unit Settings ............................................................................................................................................... 27
3.2.5 Setting of Display Contrast.....................................................................................................................28
3.2.6 Date and Time Setting............................................................................................................................28
3.2.7 Language Setting ...................................................................................................................................28
3.2.10 User Options ..........................................................................................................................................28
3.2.11 Style of Operation...................................................................................................................................29
3.2.12 Setting of HW Key..................................................................................................................................29
3.2.13 Unit Information......................................................................................................................................29
3.2.15 Service Functions...................................................................................................................................29
3.3 Specific Settings ......................................................................................................................................... 30
4 USER OPTIONS........................................................................................................................................... 32
4.2 User Sequences ......................................................................................................................................... 32
4.3 User Diagnoses/Programs.......................................................................................................................... 33
4.4 List of Recent Therapies............................................................................................................................. 33
5 APPLICATION OF SHOCKWAVES............................................................................................................. 34
7.1.2 Procedure of Worn-out Tube and Projectile Replacement .....................................................................39
7.2 Safety of the Device.................................................................................................................................... 42
The BTL -5000 SWT Series is a modern device enabling the application of therapies by using no invasive
shockwaves. Shockwaves are one of the most effective ways of treatment of pain connected to the
musculoskeletal system. Today, this is the second most frequent reason of absence from work.
The devices are equipped with a colour touch screen on the main unit, and considerably simplifies the use of the
device. The touch screen is equipped with a touch pen for easy operation of the device. A vertically positioned
device enables you to see the information on the screen clearly from different servicing positions. In addition, the
brightness of the display can be adjusted to suit the light conditions in the office or the health centre. The
information displayed on the screen will guide you throughout the whole therapy, simply adjust the parameters by
pressing the touch screen buttons and turning the knobs on the device.
Selecting a diagnosis from a list of alphabetically organized treatment protocols, or selecting a program, will make
an easy and efficient start to the therapy. You can adjust any treatment parameter manually by the simple use of
the touch screen buttons. Throughout the whole therapy, the device informs you about the therapeutical method
used, type of treatment, number of applied shocks, total and remaining shocks, frequency used, intensity and
other necessary data.
Save your time by using the predefined programs of the main unit. Based on detailed research and practical use
of the device, the well-organized predefined protocols will present recommendations for treatment of various
conditions.
The BTL- 5000 SWT allows you to insert the names of your clients and other relevant information into the internal
memory of the device and connect the client data with predefined or your own protocols. When your clients call
again, simply recall their name and apply the pre-set therapy.
With every BTL-5000 SWT you can purchase a trolley specially designed for the BTL products. Its design offers
maximum comfort of storage and use of the device. Four steady wheels provide simple movement of the device in
the office or the health care centre.
The trolley includes a range of accessory containers and baskets.
Please visit our corporate website at http://www.btlnet.com
services.
GENERAL CHARACTERISTICS OF THE DEVICE | PAGE 5 OF 47
for the latest information on BTL products and
1.1 BTL – 5000 SWT SYSTEM SERIES
The device consists of three units: the main unit, the compressor unit and the applicator.
Main system – BTL- 5000 unit, which includes the main microcomputer and software for control of the whole
equipment, including user encyclopaedia and the therapy guide.
Compressor – apart from the compressor there is also the complete electronics attached for the regulation of
the ballistic pressure.
Applicator – ergonomical applicator simplifies the „course of" the therapy, following the instructions of the
main unit.
The system enables the connection of two applicators. Each applicator can contain a different transmission
element, differing in its size and characteristic. Thereby the system features an advantage of service abilities for
two different therapies, with no need to change the transmission elements. However, only one applicator can be
active at a time. It is not possible to use two applicators together, at the same time.
1.2 SHOCKWAVE AND ITS CHARACTER
The shockwave is defined as a wave with a rapid increase of pressure in a very short time and gradual decline
with a phase of small negative pressure.
Shockwaves are aimed at the affected areas that are the cause of chronic pain. The influence of shockwaves
leads to for example dissolution of calcic deposits and better vascularisation, and in its final effect it causes relief
of pain.
Outside the client's body (extracorporeally), a pressure pulse with a high amplitude is generated, and its energy is
concentrated into the target area in the body . The pressure pulse without major losses travels through a water
medium to the client and penetrates soft tissue.
The pressure course of the shockwave in time expressively differs from the pressure course of the harmonic
sound wave. The shockwave is compared to the ultrasonic waves particularly characterized by a pressure jump
change, higher amplitude and non-periodicity.
In the shockwave, the positive amplitude is generally much bigger than the negative amplitude. The frequency
rate of the shockwaves is usually low (Hz units) and eventual cavitation (disturbance of material
consistency – rise of cavities) will relax. Consequently there is no threat of energy exhaustion on cavitations as in
the case of continuous ultrasound.
A substantial part of the shockwave energy penetrates into liquid (of the organism) with a great positive pressure
pulse. Its diffusion is only limited by the actual tissue absorption and eventual reflections on acoustic
inhomogeneities.
The shockwave is defined as a pressure pulse with these characteristics:
• High positive pressure amplitude – 10 to 1000 MPa = 100 to 1000 Bar (100x atmospheric pressure)
• Low negative pressure amplitude – 1 to 10 MPa
• Short time duration – 1 μs to 20 μs
• Rapid pressure increase – < 100 ns
• Broad frequency spectrum – 1 Hz to 1 MHz
GENERAL CHARACTERISTICS OF THE DEVICE | PAGE 6 OF 47
For therapeutical application, these values are lower, especially the maximum pressure amplitude, which is about
15 MPa, pulse length is 10 to 20 μs, and frequency of the applied shockwave is 1 to 25Hz. The treatment is
generally carried out without local anaesthetic and lasts about 15 to 30 minutes. During the first week after the
treatment, the client should avoid all activities that could excessively strain the treated area.
1.3 SHOCKWAVE GENERATION
Several types of generators were developed for the shockwave therapy, providing shockwaves with varied
characteristics. Shockwaves in their time span and spatial arrangement carry the imprint of the generation
method.
The BTL 5000 SWT uses the ballistic principle of the shockwave generation.
1.3.1 BALLISTIC PRINCIPLE OF THE SHOCKWAVE GENERATION
A pressure wave is formed via a projectile by accelerated compressed air, which is generated by an electronically
controlled ballistic-pressure compressor. Using elastic impact, the kinetic energy of the projectile is transferred
into a probe of the applicator and into the client's body via this probe. Consequently, during the treatment, the end
of the applicator must be in immediate contact with the skin and subcutaneous tissue.
1.4 BIOLOGICAL EFFECTS OF THE SHOCKWAVE
The shockwave effects mainly occur at sites, where there is a change in impedance, such as the bone-soft tissue
interface.
There is an improvement of regeneration and repair of tissues in the following areas and achievement of the
following effects:
xcellular effect – increase in transmittance of a cell membrane (improvement of ionic channels activity),
stimulation of cell division, stimulation of cellular cytokines production,
xreproduction of vessels in the area of tendons and muscles, improvement of blood circulation and MTB-
increase in concentration of growth factor beta 1 – chemotactic and mitogenic effect on osteoblasts,
x effect on nitrogen oxide system (bone healing and re-modelling),
x improvement of micro - circulation and metabolism,
x dissolution of calcic fibroblasts,
x support of collagen production,
x reduction of tissue tension,
x analgetic effect:
destruction of afferent nerves and nerve receptors,
CNS stimulants, sensed as pain; are also inflammation transmitter substances,
regression of pain caused by local ischemia,
gate control theory of pain.
GENERAL CHARACTERISTICS OF THE DEVICE | PAGE 7 OF 47
1.5 ADVANTAGES OF THE SHOCKWAVE TREATMENT
x by targeted operation of the shockwave, strain of surrounding tissues is quite insignificant
x the body is not strained by pharmaceuticals (except a short term effect of local anaesthesia)
x possibility to prevent from necessity of surgical intervention and relevant hazard
x thanks to ambulatory treatment the absence from work reduces to a minimum, as well as the absence from
training routine of sportsmen
xfor some indications (e.g. tennis elbow) there is basically no other effective treatment
1.6 POSSIBLE SIDE EFFECTS AFTER THE SHOCKWAVE TREATMENT
x erythema or swelling can temporarily occur in the treated area
x loss of bodily sensation or itching can temporarily occur in the treated area
x momentary increase of pain can appear immediately after the treatment
x haematoma
x petechia
x skin damage after previous corticoid therapy
x shockwave application can cause undesirable heart activity
1.7 INDICATIONS FOR THE SHOCKWAVE TREATMENT
x plantar fasciitis
x achillodynia – achillobursitis
x inflammations and calcification of shoulder joint tendons
x pain in the groin area
x epicondylitis (tennis and golf elbow)
x the apex patellae syndrome and tibial stress syndrome
x pain in the area of the hip and the iliotibial tract
x jumper's knee / patellar tendinitis
x pain in the hamstring insertions
x pain on the palmar side of the wrist
x exostoses of small hand joints in case of grade 1 arthrosis
x acupuncture
x pain trigger points - painful points in muscles
GENERAL CHARACTERISTICS OF THE DEVICE | PAGE 8 OF 47
1.8 CONTRAINDICATIONS TO THE SHOCKWAVE TREATMENT
x application to certain tissues (eyes and their surroundings, myocardium, spinal cord, gonad, kidneys, liver)
x coagulation of blood disorders
x polypus in the area of treatment
x pregnancy
x coagulation of blood disorders and the use of anticoagulants
x thrombosis
x tumour disease
x polyneuropathy
x acute inflammation
x growing cartilage in children
x therapy using corticoids
x inapplicable in areas and organs with possible gas content
x inapplicable on areas in proximity to large nerves, vessels, spinal cord and head
GENERAL CHARACTERISTICS OF THE DEVICE | PAGE 9 OF 47
2INSTRUCTIONS FOR USE
2.1 FRONT PANEL OF THE BTL- 5000 SWT
1. diag button to select diagnosis
2. prog button to select therapy program
3. man button to set therapy parameters manually
4. menu button (to set date, time, language, display contrast, sounds, user options, etc.)
5. user button (lists of clients, user diagnoses, programs and sequences, recent therapies, etc.)
6. touch screen
7. ON/OFF power switch (“ON” status indicated by the blue backlight)
8. enter key (to confirm selection or setting)
9. select knob (to select and set individual parameters)
10. esc key (to cancel selection or setting and return to the previous status)
11. time / stop knob (to set therapy time, and to start and stop therapy)
12. intensity knob (to set intensity)
13. hand rest of the applicator
14. shock transmitter of the applicator
15. shock transmitter screw cap of the applicator
16. connector of the applicator
17. guide mark of the connector
18. applicator button for therapy start
19. power activation indicator of the compressor
20. indicator signalling - ready to generate shockwaves ready
21. active tab indicator for the applicator output out A
22. active tab indicator for the applicator output out B
23. handles for easy transport of the compressor unit
BTL-5000 SWT USER’S MANUAL | PAGE 10 OF 47
2.2 REAR PANEL OF THE BTL- 5000 SWT
24. dummy plugs of the connectors
25. mains switch to switch the main unit on/off
26. socket for connection of the mains cable 230 V (or115 V) to the main unit
27. line fuse of the main unit
28. power switch 230V / 115V (bottom part of the main unit)
29. label with serial number and production date
30. com in - communication link input for the main unit (for servicing purposes only)
31. com in / out - communication link connector between the main unit and the compressor
32. type plate - showing device type, manufacturer and safety precautions and warning signs
33. ventilation louver
34. communication link input / output for the compressor unit
35. ON / OFF power switch for the compressor unit
36. socket to connect the mains cable 230 V (or 115 V) to the compressor unit
37. line fuses of the compressor unit
38. power switch 230V / 115V of the compressor unit
39. connectors for the shockwave applicators
40. input for connection of foot pedal
41. communication connectors (for servicing only)
42. catchment container
INSTRUCTIONS FOR USE | PAGE 11 OF 47
2.3 ASSEMBLY AND PUTTING INTO OPERATION
Inspect the package and in case of any damage report it to the delivery service or your distributor. Do not proceed
with installation or assembly if the package is damaged.
Unpack the device and place it on a stable horizontal surface suitable for the weight of the device. Always position
the unit out of direct sunlight. The device heats up therefore it must be placed away from a direct heat sources
such as radiators or room heaters. Cooling of the equipment is provided for by forced air circulation. Cooling vents
are located on the rear panel and at the bottom of the BTL-5000 SWT and must not be covered. Do not position
the device on a soft surface as it may obstruct the air flow to the bottom cooling vents. Do not put any objects
producing heat or containing water or other liquid on the top of the device. Do not place the device close to
equipment producing strong electromagnetic, electric or magnetic field (diathermy, X-rays, etc.), as the electronic
systems of the device could be undesirably influenced. In case of any questions please call your authorised BTL
distributor.
Retain the original packaging to ensure the eventual safe transportation of the device.
Attaching the accessory holder:
Remove the protective caps from the holes on a side of the device. Use a screwdriver to gently slide the blade
under the cap and take it out. Align the accessory holder and secure it with two screws from below.
Installation of supports on the compressor unit:
The supports are only installed if the compressor unit will
not sit on the cart (different legs are supplied with the cart).
The BTL-5000 e-stim SWT package includes compressor
unit legs, which prevent the unit from moving during
therapy.
Screw the supports according to the following picture.
.
INSTRUCTIONS FOR USE | PAGE 12 OF 47
1. Connection of the BTL -5000 SWT main unit with the compressor is accomplished by connecting the
communication cable to the input/output of the communication link on the relevant unit. Plug the
communication cable into the link connector on the main unit (31) and into the link connector (34) on the unit
of the compressor.
2. At first connect the shockwave compressor to the socket and then the main unit of the BTL -5000 SWT
device.
Plug the device directly into the mains socket. Do not use any multi-connection extension cables or
adapters.
Check the correct mains voltage - 230 V (or 115 V)
Before the first connection of the device to the mains, it is necessary to check if both the mains voltage
switches (26 and 28) are in the position corresponding to local voltage standards, (either in the position 230V
or 115V). For details, see the note “Switching the device to different mains voltage“
MAINTENANCE AND SAFETY INSTRUCTIONS. In case of any questions please call your BTL distributor.
3. Connect the supplied accessories step by step to the output connectors (39) on the rear panel of the
shockwave compressor as follows (item No. 3 on the picture):
in the chapter
Turn the connector of the applicator so the red spot on its end is in direct line with the red spot on the output
connector of the shockwave compressor and then plug the connector in.
INSTRUCTIONS FOR USE | PAGE 13 OF 47
WHEN DISCONNECTING THE CONNECTOR, hold the grooved end part of the
applicator of the connector with your fingers, slowly pull the connector towards you
and disconnect it carefully.
WARNING!!! DO NOT TURN THE WHOLE CONNECTED CONNECTOR BY
FORCE, AS IT CAN DAMAGE THE DEVICE!!!
The device detects the accessories, specifies the type and displays it on the screen
in the appropriate tab. In the case of connecting improper accessories by mistake, a
warning and help will display on the screen with instructions where to connect the
appropriate accessories.
4. Activate the ON / OFF power switch of the compressor unit (35).
5. Then turn the power supply switch to switch the main unit ON/ OFF (25)
6. Press the ON /OFF switch placed on the front panel of the main unit (7)
Note:
After switching on, the device will test all its internal circuits and functions (about 10 - 15 secs) and in the case of
any fault the system will signalize it and 'lock' itself into a secure mode. If this situation occurs, please call your
authorised BTL distributor.
2.4 OPERATING THE DEVICE
2.4.1 TOUCH SCREEN
The touch screen can display several graphic items, some of them are only informative, some can be pressed
and thereby activated. See the following basic items:
x soft buttons (can be pressed to change the indicated values )
x informative text
x channels tabs (switching between the channels and selection)
The items on the touch screen can be pressed by a finger or by using a special pointer with a soft tip, so - called
touch- pen, which is included in the accessories of the device. The touch screen must not come in contact with
any sharp objects, ball pens etc.
Selected channel
The BTL -5000 SWT Series allows the connection of up to two applicators, but only one can be active. Only one
so-called "selected channel" can be operated at a time. The selected channel has a light tab. All information on
the screen and all control elements relate to this channel. The most important information about therapies on the
other channels can be viewed on their tabs.
INSTRUCTIONS FOR USE | PAGE 14 OF 47
g
r
The BTL 5000 SWT does not allow the use of two applicators at the same time. Each applicator can contain
a different transmission item, differing in its size. This brings the advantage of applicability for two different
therapies without the need to change the transmission elements.
Soft buttons of the menu
Information
Channel marking:
W = shockwave
1 = No. of
Tab of the selected channel
enerato
2.4.2 NUMERIC KEYPAD
In addition to setting the numerical values with the select knob (9) on all the screens, you can also use the
„numeric keypad" for quicker setting.
Icon to open the window with the numeric keypad:
Information about connected accessories
Soft buttons (for therapy on
the selected channel)
Buttons with the same function
as (8) and (10)
Press the numeric keypad button to display the window with the numeric keypad for the parameter with the light
button (on the picture: the button - number of shocks - 2000). Set the required values and return to the previous
screen by pressing the enter (8) button. If you do not want to enter any values or change the selected parameter,
leave the window with the numeric keypad by pressing the esc (10) button.
If you enter a value out of the allowed range of values (stated above the entering box) or the device cannot set it,
then the value is rounded to the nearest lower allowed value.
INSTRUCTIONS FOR USE | PAGE 15 OF 47
2.5 THERAPY (SETTING PRINCIPLE)
2.5.1 DIAGRAM OF THE THERAPY SETTING PROCESS
ENTRY
SCREEN
THERAPY
INTERRUPTION
DIAGNOSIS
SELECTION
PROGRAM SELECTION
MANUÁL
SETTING
RUNNING
THERAPY
THERAPY START
THERAPY
INTERRUPTION
INSTRUCTIONS FOR USE | PAGE 16 OF 47
2.5.2 WELCOME SCREEN AND SELECTION OF CHANNELS, TABS AND ACCESSORIES
After switching the device ON, the entry screen includes information on the channel tabs with icons of connected
accessories. The number of channels - tabs dependents on the required configuration of the device. By pressing
the tab, information of the actual channel displays. Information adjusted on the „invisible channels" will definitely
remain saved. Almost the full screen is available for the "selected channel" - see the picture.
Switching the channels:
Tab of accessories connected to
the input out A (39)
Tab of accessories connected to
the input out B (39)
To switch between the connected accessories, press the W1 tab, see the picture. If both shockwave accessories
are connected, then the channel W1 displays the tab of accessories connected to the input out A (39), and by
pressing it, information on accessories connected to the input out B (39) will display.
Examples of information on the tabs:
Tab of channel W1 with no connected accessories
Tab of a selected channel W1, where the shockwave therapy can be applied with the connected
accessories BTL- 214-1 (SWT A)
Tab of a selected channel W1, where the shockwave therapy can be applied with the connected
accessories BTL- 214-1 (SWT B)
INSTRUCTIONS FOR USE | PAGE 17 OF 47
2.5.3 SETTING THERAPY PARAMETERS VIA THE ‘DIAG’ BUTTON
A list of diagnosis will display after pressing the diag (1) button. Each tab – channel – has its own assigned list of
diagnosis.
A fast search for diagnosis can be done by pressing the buttons with initial letters. For example, press the MNOP
button once and all diagnosis beginning with the letter M will display. According to of the number of presses, other
letters will display e.g. 2 presses of MNOP = N, 3 presses = O and 4 presses = P. The actual selected letter is
displayed in the box on the right, next to the buttons.
After finding the required diagnosis, select it by pressing the enter (8) button.
Information on therapy – encyclopaedia
2.5.4 SETTING THERAPY PARAMETERS VIA THE 'PROG' BUTTON
After pressing the prog (2) button, a screen will display, where a number of the required therapy program can be
entered. Numbers of the particular programs include prefix – a letter matching the selected therapy:
W – Shockwave therapy.
Switching between the two parts of the program number
For faster entry of the program number, you can also use the numeric keypad – for more details see the chapter
Numeric keypad.
INSTRUCTIONS FOR USE | PAGE 18 OF 47
2.5.5 SETTING THERAPY PARAMETERS MANUALLY (USER SETUP) VIA THE ‘MAN' BUTTON
The screen of the therapy parameters for the user (manual) setting will display by pressing the man (3) button. All
specifications of the therapy can be adjusted and possibly saved as the user program or the diagnosis.
By pressing the particular soft buttons, the actual menus and setup dialogs will appear. The majority of dialogs
include illustrations and symbols. See the option diagram of individual settings in this mode:
THERAPY START
INSTRUCTIONS FOR USE | PAGE 19 OF 47
2.5.6 THERAPY PARAMETERS SCREEN – ERGONOMIC, STANDARD AND EXPERT MODE
This screen always opens before the start of therapy, by pressing the diag (1) or prog (2) buttons (see Diagram
of the therapy setting process). The screen either shows the most important parameters of the therapy – you
have selected the ergonomic operation mode, or is shows all information on the therapy – you have selected the
expert or standard operation mode. In addition, the expert mode can modify all the parameters.
The differences between the modes are apparent from the following pictures:
Ergonomic mode Standard mode Expert mode
The operation mode can be set in the menu of the device by pressing menu (4) – for more details see the chapter
Operation mode. The expert operation mode can be entered from any mode by the quick selection – press the
man (3) button.
2.5.7 SETTING NUMBER OF SHOCKS
The total number of shocks for the therapy can be adjusted on screen of the therapy parameters, even during the
therapy course. The number is set either by pressing the button - number of shocks, or quick selection – by
turning the select (9) knob.
With the button number of shocks you can select "endless therapy" - see the picture
The endless therapy includes an infinite number of shocks, that can be applied at the start of the therapy
(end of therapy is not limited by marginal number of shocks).
2.5.8 SETTING THERAPY INTENSITY
The intensity (power) of the shockwave therapy can be only adjusted on the screen of the therapy parameters,
even during the therapy course. The intensity is adjusted either by pressing the intensity button or quick selection
– by turning the intensity (12) knob.
To enter the quick mode of intensity setting, press the intensity knob and turn it at the same time – the adjusted
intensity changes much faster.
2.6 COURSE OF THERAPY
2.6.1 START - INTERRUPTION - END OF THERAPY
Start the therapy by pressing time / stop (11). The therapy can start only, if the channel screen displays the
therapy parameters and all the parameters are adjusted correctly. The device will signalize possible
discrepancies.
INSTRUCTIONS FOR USE | PAGE 20 OF 47
End of interrupted
therapy
Start of interrupted
Therapy parameters
screen
Therapy START
Running therapy
screen
therapy
Paused therapy
screen
Setting frequency,
number of shocks and
therapy intensity
Therapy interrupt. Adjusting number of
shocks and intensity
during interruption
Interrupted therapy – paused therapy – can be restarted by pressing the time / stop key and can be finished by
pressing esc (10).
In pause – interrupted therapy – the parameters of intensity and total number of applied shocks can be modified
by turning the time / stop(11) knob.
The intensity can be also changed during the therapy by turning the intensity (12) knob to the right – to increase
the intensity, or to the left – to decrease the intensity. For a quicker intensity setting, press the intensity (12) knob
while turning. The intensity setting will be much faster when pressing the knob.
Direction of intensity increase
intensity
The shockwaves therapy can also be started / interrupted by the button, located on the shockwave applicator
(18), or bythe foot pedal (40). The initial setup of the button or the pedal is adjusted to AUTO. The device
automatically recognizes the holdoff time of the button on the applicator, or the foot pedal and the selected
method of control. If you hold the button or the foot pedal for more than 1sec, then after its subsequent release
the therapy will stop (as in the continual mode, see the chapter APPLICATOR BUTTON MODE / FOOT PEDAL MODE). During the time of pressing the button or the foot pedal, generation of shockwaves will be in progress.
If the button of the applicator or the foot pedal is pressed for less than 1sec, then after its release the device will
be set to the option ON/OFF (see the chapter APPLICATOR BUTTON MODE / FOOT PEDAL MODE).
INSTRUCTIONS FOR USE | PAGE 21 OF 47
pp
pp
py
2.6.2 RUNNING THERAPY SCREEN
Name of selected therapy /
program and set parameters
No. of a
lied shocks
No. of remaining shocks
Time of a
lied thera
Pressing the time / stop (11) key can also start and interrupt the therapy – see the chapter Start - interruption - end of therapy.
*the maximum frequency value according to the relevant configuration of the device is 15 Hz or 22 Hz
(see more in the table in the chapter Device configuration).
2.7 THERAPY PARAMETERS
Each therapy can have different parameters. For each therapy, only the parameters that describe that therapy will
be displayed and these parameters can be adjusted in the manual mode, after pressing the man button. An exact
description of the parameters for individual therapies can be found in the relevant User's manual.
Adjusted frequency*
Adjusted no. of shocks
Adjusted intensity
Icon and name of connected
accessories
Symbolic description of
generated output
2.8 ENCYCLOPAEDIA
The encyclopaedia includes information on possible therapies and examples of areas for the shockwave
application. The device is always supplied with a hard copy of encyclopaedia. Its electronic format is included in
the device, and is available from most screens and menus.
Icon to open the encyclopaedia:
If you open the encyclopaedia after a selection of a particular diagnosis, the existent information to the selected
diagnosis will be displayed. Otherwise you will enter the contents of the encyclopaedia – the list of particular
diagnosis. You can browse through the encyclopaedia by turning select (9). After the selection of the required
diagnosis, you will enter the particular information about the diagnosis by pressing enter (8):
INSTRUCTIONS FOR USE | PAGE 22 OF 47
y
py
Information on diagnosis
Moving between
diagnosis
Moving between
information, if not displayed
on one screen
Loading parameters of selected
diagnosis to the parameter screen
- the device is read
for thera
Display of graphic information about
diagnosis (e.g. therapy area)
Display of other graphic information about
diagnosis (optional - skin, muscle, skeleton)
2.9 THERAPY SAVING
The user can select a particular operation after pressing the save button. Depending on a selected operation,
a form with the appropriate data will be revealed. An example of the procedure is displayed on the following
screens.
You can save the therapy always after setting the therapy parameters – i. e. from the screen of therapy parameters – see the chapter Therapy parameters screen - ergonomic, standard and expert mode.
The following information is saved for each shockwave therapy:
x Frequency
x Number of shocks
x Intensity
When saving the therapy, enter:
x Name of the diagnosis (therapy) – will be displayed in list of diagnosis under the diag (1) button
x Program number – will be displayed between programs under the prog (2) button
x Description, additional information – it is displayed in both lists „below the line"
INSTRUCTIONS FOR USE | PAGE 23 OF 47
A
You can choose the number of the saved program from the numbers 8000 to 8999. The device offers you the
lowest available number. And during saving it adds a letter of the corresponding compressor belonging to the
program. So the resulting number can be e.g.: W- 8001 for the shockwave therapy.
2.9.1 SAVE THERAPY AND ADD IT TO THE CLIENT DATA
All therapy parameters that are saved in this item of the menu are the same as the previous item included in the
therapy saving menu. The saved therapy will also appear in both lists – the list of diagnosis and the list of
programs.
In addition it will also display in the list of therapies assigned to the entered client.
See the following diagram displaying the whole dialog of the therapy saving.
Set therapy parametersSave therapy
Enter name of diagnosis
Enter program name Select client
Enter description
All information entered
- must be saved
fter saving - the therapy will be
displayed with the following lists
(See the next page)
INSTRUCTIONS FOR USE | PAGE 24 OF 47
The saved therapy can be found in these lists on the relevant channel:
List of diagnosis
List of programs
List of therapies of the selected client
2.9.2 END OF THERAPY / SHOCK GENERATION
On a regular basis the therapy ends after the application of the adjusted number of shocks. If, for whatever
reason, you want to end or interrupt the therapy before finishing the adjusted number of shocks, it is possible to
interrupt the therapy by pressing the time / stop key or the button on the applicator. To end the therapy press
esc.
The generation of shocks will finish by expiry of the adjusted number of shocks applied, or it is possible to finish
their generation by releasing (in a state of continual mode) / and pressing again the button of the applicator
(in a state of ON /OFF mode) see the chapter Shockwave setting - applicator button mode.
INSTRUCTIONS FOR USE | PAGE 25 OF 47
MENU
3 DEVICE M
After pressing menu (4) it is possible to browse the following offers by using the select knob:
x Accessories
x Encyclopaedia – see the chapter Encyclopaedia
x Unit settings
x Specific settings
3.1 ACCESSORIES
In this submenu you can select:
x Installation of accessories
x Information about connected accessories
x Information about the number of clients and connection of connectors on the rear panel of the device
3.1.1 INSTALLATION OF ACCESSORIES
Each connected accessory has the memory that includes identification data of the actual accessory. According to
this data, the device recognizes which accessory is connected, if it is compatible or not, and if the device can
work with the connected accessory or not. The memory also contains a serial number of the accessory.
As the memory contains a lot of information and its downloading lasts from 30 seconds to 2 minutes, there is an
installation of accessories provided for faster work of the device. After the installation, only the serial number of
the accessory is downloaded from the accessory memory and the other information is read from the device
memory.
During the installation process, follow the instructions on the screen. In particular:
x During the installation switch off therapies on all the compressors of the device
x Do not have other accessories connected than the one that is being installed (and connect it directly, not via
interface cables)
These guides are necessary for decrease of electromagnetic interference that could cause improper downloading
of the memory data.
3.1.2 INFORMATION ABOUT ACCESSORIES
This item of the menu displays the information about connected
accessories such as Name of the accessory, its serial number etc.
And mainly it displays for which generator – output / input – the
accessory has been intended and the number of already applied
shocks.
3.1.3 INFORMATION ABOUT CONNECTORS
In the BTL - 5000 SWT, this item of the menu only includes the specification of blind / connected connectors on
the rear panel of the main unit.
DEVICE MENU | PAGE 26 OF 47
3.2 UNIT SETTINGS
This submenu offers the options of setting and displaying these parameters:
x Password setting
x Sound setting
x Screen saver and auto switch-off
x Colour setting
x Setting of display contrast
x Date and time setting
x Language setting
x Operation mode
x Touch panel calibration
x User options
x Style of operation
x Setting of HW key
x Unit information
x Unlocking code
x Service functions
3.2.1 PASSWORD SETTING
This menu can change the code - password demanded by the device when it is
switched on. There is no further work with the device possible, without entering
this password. As standard, the device is supplied „unlocked" – with the password
switched off.
3.2.2 SOUND SETTING
This includes acoustic signalling of the pressed buttons, touch panel, etc. and accomplishment of certain
operations (therapy start, therapy interruption, end of therapy etc.). The device includes standard sounding as
set from the manufacturer, i.e. signalling the therapy operations. All audio tones can be switched off (without tones) or modified as required.
This item can edit individual audio
schemes, create new ones and modify
audio tones for each operation
separately.
The sound volume can be sets in the menu item User options.
DEVICE MENU | PAGE 27 OF 47
3.2.3 SCREEN SAVER AND AUTO SWITCH-OFF
This includes the setup of the screen saver type, activation time of the screen
saver, period of time during which the screen should be switched off and idle time
of the whole device.
3.2.4 COLOUR SETTING
The user can set colours of all elements displayed on the screen. There are 50
preset colour schemes available to choose from, or, if not satisfied with any of
them, the user can create and save his/her own colour schemes. In the custom
colour scheme, the user can step by step select individual elements.
3.2.5 SETTING OF DISPLAY CONTRAST
After selecting this offer, the optimum contrast (readability) of the display can be
set by turning the select (9) knob.
As the contrast of the screen depends on various factors, (such as
temperature), there is a fast and direct setting of the screen contrast, by using
the select (9) knob while simultaneously holding the enter (8) and esc (10)
keys.
3.2.6 DATE AND TIME SETTING
This allows the user to set the date and time in the device.
3.2.7 LANGUAGE SETTING
Preferred language of the texts displayed in the device. As standard, English is preset by the manufacturer.
3.2.8 OPERATION MODE
Select one of the three modes, see the chapter Therapy Parameters Screen – Ergonomic, Standard and
Expert Mode. The ergonomic mode is preset by the manufacturer as standard.
3.2.9 TOUCH PANEL CALIBRATION
If the buttons on the touch screen do not react when pressed, the touch screen needs calibration. The calibration
procedure is apparent from the screen of the device. It is advisable to use the touch- pen during calibration and
follow the instructions, displayed on the screen of the device.
Unsuccessful calibration can always be interrupted by pressing the esc button.
To verify the touch screen adjustments, use the "touch panel function test" feature.
3.2.10 USER OPTIONS
Here, you can set:
x Direction of movement in the menu when turning the select (9) knob
x Alphabetical order of diagnosis and certain menus (from A to Z / from Z to A)
x Position of the tab bar (up / down)
x Sound volume
DEVICE MENU | PAGE 28 OF 47
3.2.11 STYLE OF OPERATION
xNEW STYLE OF OPERATION
Use this option for main units of the first series (manufactured in 2001 and start of 2002). These devices have the
knobs (11) and (12) marked as time (11) and start / stop (12). It is necessary for these devices to select the
option new style of operation = no after download of the new version of firmware.
New units are preset by the manufacturer to a new operation mode (new style of operation = yes).
xEND OF THERAPY – SETTING ZERO INTENSITY / TIME
After the end of the therapy you can choose, if you want to leave the adjusted data of time and intensity on the
screen in zero values, or display the values of the last performed therapy.
xREPEAT SOUND FOR END
This item can switch on / off the function of repeated warning tone of the therapy end.
3.2.12 SETTING OF HW KEY
The device can be remotely configured. For example, if you later decide to purchase the upgrade, the
manufacturer (distributor) will send you 64 -digit code (so - called HW key) that you will enter and the device will
reconfigure its hardware according to the entered key. For the actual HW configuration see the chapter Unit
information.
3.2.13 UNIT INFORMATION
This item shows certain information about the device – serial number, device type, firmware version, etc. It also
contains the information up to when the device will operate – so called "device validity". If the operation of the
device is time limited, this item indicates until which date the device stays fully functional.
3.2.14 UNLOCK CODE
If the operation of the device is time limited, then in this item you can enter the so-called "unlocking code" for
cancellation of the operational time limitation of the device.
3.2.15 SERVICE FUNCTIONS
xREPAIR OF FILES
This checks the file system in the device - system of saved information and repairs possible errors - deletes
empty files, etc. We recommend using this feature in the case of lack of memory, if the device rejects to save any
data, or if you are in doubt that some data may have been lost, etc.
xFILE SYSTEM FORMATTING
If the feature repair of damaged files did not help you, select this function. Unfortunately, together with this
formatting you will loose all the user data and user setup of the device.
xDELETE ACCESSORIES
This item will delete all the installed accessories – use it, when you install the accessories incorrectly – e.g. bad
„scattered" picture on the tab, the connected accessories cannot be detected (permanently displayed question mark) etc.
xDEFAULT SETTING WITHOUT LOSS OF USER DATA
Restoration of all functions of the device the way they were preset by the manufacturer. User data – clients,
therapies etc. will stay saved.
DEVICE MENU | PAGE 29 OF 47
3.3 SPECIFIC SETTINGS
3.3.1 SHOCKWAVE SETTING
xCOMPRESSOR MODE
It adjusts two different types of the mode of pressure control in the
system: - quick start
- silent mode
When using quick start, the pressure control in the system is activated during switching the device on, therefore
the therapy start is instantaneous. We recommend using this setting especially during frequent operations. The
disadvantage of this mode is a higher noise level of the device.
In silentmode the pressure control in the system is activated only immediately before the start of therapy - during
entry to the manual screen of the main unit and after leaving, the pressure stops being kept in the system. That
eliminates the noise of the device, if not in use. The disadvantage can be the prolonged start of the therapy –
according to pressure in the system the start can be extended for about 10 to 15 seconds. We recommend using
this setting for quieter operations.
xTHERAPY CONTROL
Here, you can set whether you want the therapy to be controlled by the button on the applicator (18) or using the
foot pedal connected to the unit of the compressor (40). After selecting the type of therapy control, you can
continue by setting the options of the applicator button mode, or the foot pedal mode (see the following chapter).
xAPPLICATOR BUTON MODE / FOOT PEDAL MODE
This enables setting of the three different types of modes for the applicator button, or the foot pedal (according to
setting in the previous chapter). You can choose from:
- AUTO mode
- Continual mode
- ON/OFF mode
When setting the AUTO mode the device automatically recognizes
the holdoff time of the button on the applicator (18), or pressing the
foot pedal (40) and the selected way of control. If you hold the
button or the foot pedal for more than 1sec, then after its
subsequent release the therapy will stop (as in the continual mode, see below). During the time of pressing the button, or the
foot pedal, generation of shockwaves will be in progress.
If the button of the applicator, or the foot pedal is pressed for less than 1sec, then after its release the device will
be set to the option ON/OFF (see below).
By selecting the continual mode you will set the generation of the adjusted shockwaves by only one press of the
button on the shockwave applicator (18), or the foot pedal (40), to end the generation of shockwaves, another
touch of the button on the applicator is necessary, or pressing the pedal.
During activated option of the ON/OFF mode it is necessary to hold the button of the applicator (18), or the foot
pedal (40) pressed throughout the time of the application of shockwaves. Releasing the button or the pedal stops
the generation of the shockwaves.
The setting is saved separately for the button of the applicator and for the foot pedal.
DEVICE MENU | PAGE 30 OF 47
3.3.2 APPLICATOR KIT REPLACEMENT WIZARD
If, after a certain time, the applicator stops generating shockwaves correctly, it is possible to change the applicator
kit, following the chapter Procedure of worn-out tube and roller replacement. Install this replaced kit according
to the displayed instructions.
Procedure of the applicator kit replacement on the device:
1. Connect the original applicator to the compressor.
2. Start the applicator kit replacement guide.
3. Now, disconnect the original applicator from the compressor and complete the replacement of the kit (see the
chapter Procedure of worn-out tube and roller replacement)
4. After completion of the kit replacement connect the applicator back to the same channel of the compressor
unit.
5. Finish the guide.
6. The device has successfully installed the replaced kit.
DEVICE MENU | PAGE 31 OF 47
4USER OPTIONS
Pressing the user (5) button will display the menu with items referring to the data saved by the user or referring to
certain „special" functions of the shockwave generator:
x Clients
x User sequences
x User diagnoses / programs
x Recent therapies
4.1 CLIENTS
This item allows you to enter, edit and delete information about the clients. A particular therapy can be assigned
to each client.
4.2 USER SEQUENCES
This feature - user sequences - is not active for the BTL- 5000 SWT Series.
USER OPTIONS | PAGE 32 OF 47
4.3 USER DIAGNOSES/PROGRAMS
It is possible to start the user therapies, adjust their parameters, names and descriptions, delete them and sort
them by using the buttons and offers on the screen. It is very similar to the creation of a new diagnosis / program
– see the chapter Therapy saving. On the tab of the compressor you can only see those therapies that were
created and belong to this tab.
4.4 LIST OF RECENT THERAPIES
This feature allows you to choose one of the recent accomplished therapies on the selected tab and after
pressing the load button you can restart it and view its parameters.
USER OPTIONS | PAGE 33 OF 47
5APPLICATION OF SHOCKWAVES
Application of shockwaves is provided by a spring-loaded applicator, which:
x Provides proper and exactly specified contact of the applicator with the treated tissue during the therapy
x Minimizes transmission of reverse shocks to the hand of the therapist thanks to an inbuilt air damper
The component part of the applicator is an ergonomic hand rest (13) used for a comfortable grip of the applicator
and prevents unwanted slipping of the hand from the holder of the applicator during the therapy. The ergonomic
hand rest spins and can be any time adjusted into the desired position.
The anatomical shape of the handle allows the therapist to grip the holder of the applicator comfortably in different
positions, exactly according to the type of the applied therapy and personal preferences of the therapist
(left-handed person, right-handed person, man, woman).
The ergonomic shape of the handle offers the therapist sufficient support during the application, thereby
preventing from the development of tiredness of the hand during the therapy.
Examples of holding the shockwave applicator
Examples illustrate the application of shockwaves in different places and various of ways of holding the applicator.
The device is not designed for use in connection with different accessories and other medical equipment except
those stated in this manual.
The following list includes a summary of all accessories that can be supplied with the device.
For detailed information on individual accessories see the enclosed leaflet.
Standard BTL accessories:
x 1x trolley for the BTL- 5000 SWT
x 1x shockwave applicator with multi-focus shock transmitter Ø 15mm
x 1x replaceable multi-focus shock transmitter Ø 9m
x 1x replaceable focusing shock transmitter Ø 15mm
x 1x exchangeable kit (includes a spare set of O-rings)
x 1x applicator holder
x 1x gel 1000 ml
x 1x power cord
x 1x communication cable
x 4x silent blocks for compressor unit
x 1x spare fuse T1.6A/250V for the main unit
x 2x spare fuse T6.3A/250V for the compressor unit
x 1x pen for the for touch screen
x 1x CD with the user's manual
x 2x dust cover of the main unit
x markers for output cables
Optional BTL accessories:
x transport case for the BTL- 5000 SWT
x additional shockwave applicator with focusing shock transmitter
x foot pedal
x gel 300 ml
ACCESSORIES | PAGE 36 OF 47
7MAINTENANCE AND SAFETY INSTRUCTIONS
Service inspection and measuring of all parameters of the device and relevant recalibration must be executed at
intervals shorter than 30 calendar months. The inspection is carried out by authorised BTL customer service on
the basis user's order. The manufacturer does not guarantee for the technical parameters and safe operation of
the product, if the inspection is not executed on the defined schedule.
Keep the device clean, do not store or use it for a long time in extremely dusty environments and do not spill any
liquid onto the surface. Before using check the device and its accessories (especially cables) for mechanical or
other damages. Do not use the device if damaged!
Cleaning the surface of the device and its parts:
Clean the BTL - 5000 SWT and its parts only with a lightly damp soft cloth. Use water or a 2 % detergent solution.
Never use agents containing alcohol, ammonia, petrol, thinners, etc. To clean the device, do not use abrasive
materials as they could damage the surface of the device and e.g. deteriorate metal parts of the applicator. There
are no aseptic parts of the device and there is no need to sterilize them.
Cleaning accessories getting in contact with the client (e.g. shock transmitters):
Cleanse after each client using agents approved by the competent health officer. You can use e.g. Sekusept,
Bacilol and other. For the cables of the accessories you can use e.g. Incidur spray etc. Surface of the shock
transmitter with the cap can be washed with warm water, however for the complete removal of gel from the shock
transmitter it is necessary to unscrew the shock transmitter and clean it.
Fuse replacement
The fuse of the main unit is set on the rear panel in a round black box (27). Fuses of the shockwave compressor
are placed on the rear panel of the unit in a round black box (37). Check the accuracy of the inserted fuses during
their replacement. The main unit (25) and the shockwave compressor (35) both use different types of fuses!
Before replacement, first make sure that the mains switches of the main unit (25) and the shockwave compressor
(35) are in „0" position and disconnect the power cords from both units. Use a suitable screwdriver or a coin to
turn the segment of the fuse case to the left and remove the fuse. Insert a new fuse and finish the replacement
similar way, but turn it to the right. It is prohibited to insert a fuse with different indication, than mentioned on the sign above the fuse box.
This operation can be accomplished only by a person familiarized with this procedure!
Switching the device to different mains voltage
The witch of the main unit (28) is placed under the device in a round black box. The switch of the shockwave
compressor (38) is placed on the rear panel of the unit. Before switching over, firstly make sure that the power
switches (25, 35) are in the „0" position and disconnect the power cords from the device. Use a suitable
screwdriver or a coin to turn the segment of the switches so sign on the switch with the arrow pointing to,
correspond to the voltage of the local mains distribution system.
MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 37 OF 47
This operation can be accomplished only by person familiarized with this procedure!
If in doubt contact the authorised BTL customer service. The manufacturer is not responsible for any damage
incurred by incorrect switchover.
Transport and storage
We recommend retaining the original packaging of the device to ensure its maximum protection during
transportation. Disconnect the mains cables and the accessories cables. Avoid heavy shocks. The device must
be stored or transported under the conditions defined in the chapter Technical parameters.
7.1.1 SHOCK TRANSMITTER REPLACEMENT PROCEDURE
The shock transmitter (14) can be replaced by another one as necessary; three shock transmitters are the part of
the BTL - 5000 SWT standard accessories:
x 1x replaceable multi-focus shock transmitter Ø 15mm
x 1x replaceable multi-focus shock transmitter Ø 9mm
x 1x replaceable focusing shock transmitter Ø 15mm
Replacement procedure:
1. Unscrew the screw cap of the shock transmitter by hand (15).
2. Remove the shock transmitter from the applicator (14) (including both O - rings).
O-rings
3. Insert the selected shock transmitter „as it is" to the applicator (including the mounted O - rings).
4. Put the shock transmitter's screw cap of the applicator back on and screw it tightly by hand (15).
MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 38 OF 47
7.1.2 PROCEDURE OF WORN-OUT TUBE AND PROJECTILE REPLACEMENT
If the applicator stops working correctly after a certain time, it is possible to change the worn-out system of tube
and projectile by using the exchangeable kit that is part of the BTL - 5000 SWT package.
Do not use damaged applicators; otherwise you may cause risk of injury to the operating staff or the client.
The packaging of the spare kit includes:
x 1x exchangeable kit
x 1x key to unscrew / screw the cartridge screw cap
x 1x plastic covering of the exchangeable kit
x 1x set of spare O-rings
Replacement procedure:
1. Firstly, according to the chapter Applicator kit replacement wizard start
guide of the kit replacement.
2. Unscrew the screw cap of the shock transmitter by hand (15).
3. Remove the shock transmitter from the applicator (14) (including both O - rings).
O-rings
MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 39 OF 47
4. Using the key (part of the packaging) unscrew the screw cap of the cartridge, follow the pictures.
5. Remove the worn-out kit from the cartridgeof the applicator.
6. Take out a new replaceable kit from the plastic cover.
MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 40 OF 47
7. Carefully insert the kit "as it is" into the cartridge of the applicator in the direction shown in the picture (do not
touch the rear cover behind the spring, as it is coated with a thin layer of lubricant from production!).
The front cover of the kit must be adjusted in position shown in the picture and inserted in one level with the
edge of the outer case of the applicator (push the kit a tiny bit inside the case of the applicator). The replaced
kit must move freely!
8. After the insertion of the kit, firmly screw thescrew cap of the cartridge back using the key.
9.
When replacing the applicator kit we recommend replacing the O-rings on the applicator
attachment too. Spare O-rings are included in the accessories. Then put the applicator attachment
(14) with both O-rings replaced back in.
10. Insert the shock transmitter back to the applicator (14) (including both O - rings).
11. Put the screw cap of the shock transmitter back on and screw it tightly by hand (15).
12. Connect the applicator back to the BTL - 5000 SWT device and finish the installation of the new kit according
to the chapter Applicator kit replacement wizard.
MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 41 OF 47
7.2 SAFETY OF THE DEVICE
ATTENTION!
Only accessories connected to the connectors marked as follows can be applied to the client. The client's part is
of BF type. Attention, the values of power, electric current or voltage can exceed the safe values of the power,
current and voltage.
The device is equipped with a protective system against the connection of different equipment other than supplied
by the manufacturer; consequently it does not allow for operation with equipment from other manufacturers.
Do not apply the shockwave to the areas with the body organs containing gases, or close to big nerves,
vessels, spinal cord or the head!
The device does not use any pharmaceuticals or matters, which would be applied or were an integral part of the
device.
The connectors marked with these signs can be connected with the foot pedal for initiation of the therapy. The
foot pedal is placed under the bottom part of the shockwave compressor.
General safety precautions:
x Carefully read the service instructions before the first activation of the device.
x The device should only be used under the control of a physician who determined the indication.
x All personnel, using this device must be knowledgeable in the servicing technique, maintenance and
verification of the device and about the principles of safe work with the device.
xElectrical mains connected to the device must be performed according to the existing valid standards
(IEC 364) and must be certificated according to these standards. If you are not sure, that the power mains
supply is in perfect order, carry out a technical revision by an engineering inspector.
xCrosscheck, whether the parameters of the power mains supply correspond with the requirements of the
device according to the chapter Technical parameters and if the mains voltage switch on both units of the
device is switched to the correct voltage according to the parameters of the power mains.
xThe device is designed to work in an environment defined in the chapter Technical parameters. It must not
be used in an environment with threats of e.g. explosion hazard, or water penetration. The device must not be
used in connection with flammable anaesthetics, or oxidative gases (O2, N2O etc.).
xPlace the device out of direct sunlight and out of reach of strong electromagnetic fields, to avoid undesirable
interactions. In case of undesirable interactions place the device away from source of disturbance, or contact
the authorised BTL service centre.
xBefore every use, examine the device properly (loose cables, damaged cable insulation, the surface of
accessories, functions of the display, operating elements etc.) and in case of any discrepancy stop using the
device and contact the authorised BTL service centre. If the behaviour of the device shows any divergences
from the function procedure described in the instruction manual, stop using the device and contact the
authorised BTL service centre.
MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 42 OF 47
xIf the device shows any signs of defects, or you are in doubt about its correct operation, stop the therapy
immediately. If you cannot detect the source of doubts even after careful reading through the instruction
manual, appeal to the authorised service department of the BTL device without delay. If the device is used
contrary to this instruction manual, or it is used despite the device showing functional divergences from this
instruction manual, then the user is responsible for any damage caused by the device!
xDo not disassemble the device in any case. Dismantling and unlocking the safety cover can cause danger of
electric shock. Eventual replacement of the lithium battery and the air micro filter must by carried out by
authorised BTL service centre.
xAll materials and parts that come into direct contact with the client's body (detergents used for cleaning the
shock transmitter ...), must meet the appropriate standards, relevant to irritation, allergization, toxicity
genotoxicity, carcinogenity according to the standards ISO 10993 - 1, ISO 10993 - 3, ISO 10993 5. The user
himself is responsible for all these materials and parts that were not supplied by the BTL device distributor.
xDo not connect the connectors used for the accessories and the other connectors with anything other than
the equipment intended by the service instructions, as it can cause danger of electric shock and serious
damage of the device!
xThe device does not radiate or does not use any toxic materials during its operation, storage and transport
under the given conditions.
xWhen transferring the device from a cold environment into warm, wait before plugging it into the power mains,
until the temperature levels (2 hours minimum).
x Before starting the therapy, check whether all the adjusted parameters correspond to your requirements.
x The shock wave is used as the contact therapy – Do not in any case apply on damaged skin!
x Do not terminate the application by switching off the mains switch, but by pressing the time / stop (11) key.
When switching the device off and on, there is a time delay necessary of at least 3 seconds.
xWhen disposing of the device after years of operation, it is necessary to remove the lithium battery. Dispose
of it in a proper way – do not put the battery into the communal waste. The device can then be disposed of in
the usual way. The device does not contain any toxic materials that could damage the environment when
disposed of in the ecological way.
x It is not allowed to use the device and accessories other than according to the instruction manual.
x This device must be kept out of reach of the children.
x The device does not include any components to be reparable by the user, except the replacement kit. Do not
remove any covers from the device. Repairs must always be provided by the authorised BTL service centre.
xDo not unplug the applicator during therapy.
7.3 WARRANTY CONDITIONS
The manufacturer provides a 24 months guarantee for the device from the date of purchase mentioned in the
certificate of warranty, and 6 months for the accessories and the expendable supplies (guarantee does not apply
to common wear of the parts).
Conditions of the guarantee determine, if the device is not used in compliance with this instruction manual, or
unprofessional intervention to the device is carried out by third parties. In case of malfunction of the device always
contact the authorized BTL service centre.
MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 43 OF 47
y
8 TECHNICAL PARAMETERS
Identification of the deviceBTL - 5000 SWT BASIC / BTL - 5000 SWT POWER
Operating conditions
Ambient temperature + 10 °C to+ 40 °C
Relative humidity 30 % to 75 %
Atmospheric pressure 700 hPa to 1060 hPa
Position vertical – on legs
Type of operation permanent
Transport and storage conditions
Ambient temperature - 10 °C to + 55 °C
Relative humidity 25 % to 85 %
Atmospheric pressure 650 hPa to 1100 hPa
Position of the main unit any
Position of the compressor basic
Storage time max. 1 year
Additional conditions transport only in the supplied packing
Power suppl
Main unit max. Input 80VA
Compressor maximum input 550VA
Mains voltage (switched)~ 100 V to 120 V (115 V nominal), AC
Frequency 50 Hz to 60 Hz
Equipment protection class
External exchangeable fuse 1xT1.6A / 250V, tube safety fuse 5 x 20 mm, according to IEC 127-2
II (according to IEC 536)
2xT6.3A / 250V, tube safety fuse 5 x 20 mm, according to IEC 127-2
Power switch according to IEC 601 - 1 at the back of the device, position 0 and I
Design
Weight – main unit only max. 5.3 kg
Weight – compressor unit max. 20 kg
Weight - applicator max. 900 g including cable
Main unit dimensions (w x h x d) 230 x 390 x 260 mm
Compressor unit dimensions (w x h x d) 330 x 270 x 320 mm
Dimensions - applicator (w x h x d) 40 x 280 x 140 mm
IP code according to EN 60 529
IP 20
Display units
Graphic colour touch screen diag. 5.7" / 14.5 cm, resolution 320x240 pixels (BTL - 5000 SWT Basic)
diag. 8.4" / 21.5 cm, 640x480 pixels (BTL - 5000 SWT Power)
Indicators 1x blue for the main unit
1x green, 1x yellow, 2x blue for the compressor
Classification
Applied part type BF
Class according to MDD 93/42/EEC IIb
~ 200 V to 240 V (230 V nominal), AC
(main unit)
(compressor unit)
TECHNICAL PARAMETERS | PAGE 44 OF 47
Adjustable values
Shock intensity max. 3 / 5* Bar (maximum instantaneous value)
Shock frequency max. 15 Hz / 22 Hz*
Therapy period 0 – 9999 shocks
*only for BTL- 5000 SWT POWER
Shockwave specifications
Maximum positive pressure max. 15.8 MPa
Maximum negative pressure max. 0.9 MPa
Energy density
Steps of adjustable values
Intensity 0.1 Bar
Frequency 1 Hz
Total number of pulses per therapy dynamically - according to size from 1 to 50 shocks
0.25 mJ/mm
2
8.1 APPLICABLE STANDARDS
Name IEC, EN, ISO, MDD
Medical electric device. Part 1: General requirements for safety IEC 601-1
Appendices to EN 601-1 A2, A11 a A12
Medical electric device. Part 1-1: Systems IEC 60601-1-1
Medical electric device. Part 1: EMC - requirements and examinations IEC 60601-1-2
ISM – radio interference characteristics EN 55011
EMC – electrostatic discharge IEC 61000-4-2
EMC – radiated electromagnetic field IEC 61000-4-3
EMC – fast electric transient effects/pulse groups IEC 61000-4-4
EMC – shock pulse IEC 61000-4-5
Medical electric device. Part 4: Programmable electric medical systems IEC 60601-1-4
Medical device - risk analysis ISO 14971
Bio-compatibility of medical device ISO 10 993-1
Statutory rules for determination of technical requirements to sanitary engineering assets
Particular requirements for the safety of equipments for extracorporeally induced lithotripsyIEC 60601-2-36
Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and
muscle stimulators
Clinical investigation of medical devices for human subjects
NV 181/2001 Col.
MDD 93/42/EEC
IEC 60601-2-10
ISO 14155
TECHNICAL PARAMETERS | PAGE 45 OF 47
8.2 MANUFACTURER
This product is manufactured in accordance with the EU Medical Devices Directive by:
BTL Industries Ltd.
161 Cleveland Way
Stevenage
Hertfordshire
SG1 6BU
United Kingdom
E-mail: sales@btlnet.com
http://www.btlnet.com
For service, please contact the service department at service@btlnet.com
.
No part of this manual may be reproduced, saved in a research centre or transferred by any means incl. electronic, mechanic,
photographic or other records without previous approval from BTL Industries Limited
BTL Industries Limited operates a policy of continuous development. Therefore, it reserves the right to make changes and
improvements to the products without prior notice.
The content of this document is provided "as it is ". Except as required by applicable law, no warranties of any kind, either
expressed or implied, are made in relation to the accuracy, reliability or contents of this document. BTL Industries Limited
reserves the right to revise this document or withdraw it at any time without prior notice.
TECHNICAL PARAMETERS | PAGE 46 OF 47
9DEVICE CONFIGURATION
Type:
Number of shockwave therapies 1 1
Display, diag.
Client's records x x
Encyclopaedia x x
Encyclopaedia with anatomical
pictures
Predefined diagnosis 13 13
Preset programs 13 13
Language versions x x
Sound schemes x x
Colour schemes x x
Recent therapy x x
Electrotherapy upgrade - x
Laser therapy upgrade - x
Magnetic therapy upgrade - x
Ultrasound therapy upgrade - x
Additional applicator* - optional
Shockwave parameters :
Pressure 1.5 - 3 Bar 1.5 - 5 Bar
Frequency 1 - 15 Hz 1 - 22 Hz
Individual shocks mode x x
Continual mode x x
Burst mode - x
* Possible simultaneous connection of two applicators
........ exostoses of small hand joints in case of grade 1 arthrosis ........................................................................ 5
........ pain in the area of the hip and the iliotibial tract............................................................................................ 6
........ pain in the groin area........................................................................................................................................ 7
........ pain in the hamstring insertions...................................................................................................................... 8
........ pain on the palmar side of the wrist................................................................................................................ 9
type: continual
pressure: 2
frequency: 10
number of shocks: 2000
scar area: no
lying prone, leg supported under the ankle
preferably within 5-10 days
3 – 5 sessions
Manual localization of painful points by means of palpation along the Achilles
tendon. ~Application of gel. ~Start the therapy with 400 shocks, 2.5 bar, 5 Hz
around the most painful point, subsequently treat the most painful point with
approx. 1000 shocks, always longitudinally along the tendon, then apply
successively on additional painful points in the calf. Increase the energy
gradually up to 3 bar according to the patient's reaction, last 400 pulses apply
with a frequency of 10 Hz. If needed, the number of shocks may be increased
by additional 2,000 during one application.
We recommend ultrasound examination of the Achilles tendon, especially in
the patients who have formerly received local administration of
corticosteroids.
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1108)
type: continual
pressure: 2.5
frequency: 10
number of shocks: 2000
scar area: no
Lying prone, leg supported under the ankle. Manual localization of the painful
point by means of palpation.
preferably within 5-10 days
3 – 5 sessions
Manual localization of painful points by means of palpation around the heel
and on the sole. ~Application of gel. ~Start the therapy with 400 shocks, 2.5
bar, 10 Hz around the most painful point, subsequently treat the most painful
point with approx. 1000 shocks, then apply successively in the surrounding
tissues including the plantar aponeurosis and the Achilles tendon. Increase
the energy gradually up to 3.5 bar according to the patient's reaction, last 400
pulses apply with a frequency of 15 Hz.
If needed, the number of shocks may be increased to 2,500 during one
application.
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1107)
W-0109
program: W-0109 exostoses of small hand joints in case of grade 1 arthrosis
type: continual
pressure: 2
frequency: 10
number of shocks: 1500
scar area: no
lying on back or sitting, arms in the right-angled flexion in the elbow,
comfortably supported, preferably with a soft elastic pad
preferably within 5-10 days
3 – 5 sessions
Palmar manual localization of painful points. ~Application of gel. ~Start the
therapy with 400 shocks, 2 bar, 5 Hz around the most painful points,
subsequently treat the most painful points with approx. 1000 shocks, last 400
pulses can be applied on the palm with a frequency of 10 Hz.
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1116)
W-0104
program: W-0104 pain in the area of the hip and the iliotibial tract
type: continual
pressure: 2.5
frequency: 10
number of shocks: 2000
scar area: no
lying on side
preferably within 5-10 days
3 – 5 sessions
Manual localization of painful points in the trochanter major region and along
the iliotibial tract. ~Application of gel. ~Start the therapy with 400 shocks, 2.5
bar, 10 Hz around the most painful point, subsequently treat the most painful
point with approx. 1000 shocks, then apply successively along the iliotibial
tract, last 400 pulses apply with a frequency of 15 Hz. ~If needed, the number
of shocks may be increased by additional 2,000 during one application.
Caution: Aim the shockwave at soft tissues only, avoid direct application on
the bone!
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1111)
type: continual
pressure: 2.5
frequency: 10
number of shocks: 2000
scar area: no
lying on back, lower extremity supported in slight semiflexion both in the hip
joint and knee joint and in slight abduction
preferably within 5-10 days
3 – 5 sessions
Manual localization of painful points in the groin region and along the muscles
on the front and inner side of the thigh. ~ Application of gel. ~ Start the
therapy with 400 shocks, 2.5 bar, 10 Hz around the most painful point,
subsequently treat the most painful point with approx. 1000 shocks, then
apply successively over the inner and front side of the thigh, last 400 pulses
apply with a frequency of 15 Hz. ~ If needed, the number of shocks may be
increased by additional 2,000 during one application.
Caution: Avoid application on the inguinal channel (passage of large vessels
and nerve bundles, lymph nodes).
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1113)
W-0105
program: W-0105 pain in the hamstring insertions
type: continual
pressure: 2.5
frequency: 10
number of shocks: 2000
scar area: no
lying prone
preferably within 5-10 days
3 – 5 sessions
Manual localization of painful points in the area of tuber ischiadicum and
along the hamstrings. ~Application of gel. ~Start the therapy with 400 shocks,
2.5 bar, 10 Hz around the most painful point, subsequently treat the most
painful point with approx. 1000 shocks, then apply successively over the rear
side of the thigh, last 400 pulses apply with a frequency of 15 Hz.
If needed, the number of shocks may be increased by additional 2,000 during
one application.
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1112)
W-0110
program: W-0110 pain on the palmar side of the wrist
type: continual
pressure: 2
frequency: 10
number of shocks: 1500
scar area: no
lying on back or sitting, arms in the right-angled flexion in the elbow,
comfortably supported, preferably with a soft elastic pad
preferably within 5-10 days
3 – 5 sessions
Palmar manual localization of painful points. Application of gel. Start the
therapy with 400 shocks, 2 bar, 5 Hz around the most painful point,
subsequently treat the most painful point with approx. 1000 shocks, last 400
pulses apply with a frequency of 10 Hz.
Caution: Avoid the area of the passage of vessels and nerve bundles on the
wrist!
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1117)
type: continual
pressure: 3
frequency: 10
number of shocks: 2000
scar area: no
lying on back or sitting
preferably within 5-10 days
3 – 5 sessions
Manual localization of painful points in soft tissues around the shoulder joint
by means of palpation, localization of painful points located in deeper
structures by means of shockwaves. Application of gel. Start the therapy with
400 shocks, 3 bar, 10 Hz around the most painful point, subsequently treat
the most painful point with approx. 1000 shocks, then apply successively in
the entire shoulder girdle, last 400 pulses apply with a frequency of 15 Hz.
We recommend performing ultrasound localization of calcifications.
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1114)
type: continual
pressure: 2
frequency: 10
number of shocks: 2000
scar area: no
lying on back, supported knee
preferably within 5-10 days
3 – 5 sessions
Manual localization of painful points by means of palpation on the edges of
the patella and the insertion of m.quadriceps femoris. ~Application of gel.
~Start the therapy with 400 shocks, 2 bar, 5 Hz around the most painful point,
subsequently treat the most painful point with approx. 1000 shocks, then
apply successively on additional painful points on the thigh. Increase the
energy gradually up to 3 bar according to the patient's reaction, last 400
pulses apply with a frequency of 10 Hz.
If needed, the number of shocks may be increased by additional 2,000 during
one application.
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1109)
type: continual
pressure: 2
frequency: 10
number of shocks: 2000
scar area: no
lying on back or sitting, arms in the right-angled flexion in the elbow,
comfortably supported, preferably with a soft elastic pad
preferably within 5-10 days
3 – 5 sessions
Manual localization of painful points in the area of radial/ulnar epicondyle and
along the related muscle groups of the forearm. ~Application of gel. ~Start the
therapy with 400 shocks, 2 bar, 5 Hz around the most painful point,
subsequently treat the most painful point with approx. 1000 shocks, then
apply successively in the related muscle groups of the forearm, last 400
pulses apply with a frequency of 10 Hz.
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1115)
type: continual
pressure: 1.5
frequency: 10
number of shocks: 2000
scar area: no
lying on back, supported knee
preferably within 5-10 days
3 – 5 sessions
Manual localization of painful points along the tibia. ~Application of gel. ~Start
the therapy with 400 shocks, 1.5 bar, 5 Hz around the most painful point,
subsequently treat the most painful point with approx. 1000 shocks, then
apply successively on additional painful points. Increase the energy gradually
up to 2.5 bar according to the patient's reaction, last 400 pulses apply with a
frequency of 10 Hz. ~If needed, the number of shocks may be increased by
additional 500 during one application.
Caution: Aim the shockwave at soft tissues only, avoid direct application on
the bone!
Painful points are indicated by the dotted areas in the picture. The arrows
indicate the direction of movement of the applicator on related areas.
(1110)
3.1.1 Initial Logo of Device................................................................................................................................4
3.1.2 Prefix of Device Name..............................................................................................................................4
3.1.5 View the Opening Screen.........................................................................................................................6
3.1.6 System Language ....................................................................................................................................7
3.2 Special Setting.............................................................................................................................................. 7
3.3 Configuration - Information ........................................................................................................................... 8
3.3.3 Serial Number ..........................................................................................................................................9
3.3.5 Diagnoses and Programs.......................................................................................................................10
3.3.6 File System ............................................................................................................................................11
3.3.7 RTC Time and Date ...............................................................................................................................11
3.4.1 Time of Usage........................................................................................................................................12
3.5.3 System Language ..................................................................................................................................13
3.5.4 Setting of HNDP and VNDP ...................................................................................................................14
3.5.6 Final Check ............................................................................................................................................15
3.6 Service Functions ....................................................................................................................................... 18
3.6.1 Format and Scandisk .............................................................................................................................18
3.6.3 Final Check ............................................................................................................................................18
3.6.4 PC Bridge...............................................................................................................................................18
3.6.5 Ventilator Test ........................................................................................................................................18
3.6.7 Buttons Test ...........................................................................................................................................19
3.6.8 Knobs Test .............................................................................................................................................19
3.6.9 RTC Time and Date ...............................................................................................................................20
In this mode it is possible to perform the distributor localization of the device, change the initial logo of the device,
set the prefix of the device's name, change the address on the opening screen and set the system language in
which the device will communicate with the user by default. Then it is possible to find out some technical
parameters of the device, such as number and types of individual generators, versions of all components used in
the device, number of diagnoses in the device, current limitations, etc. In addition it is possible to put the entire
device into the demo mode. This mode serves mainly for service setting and testing of the device.
2ENTERING THE SERVICE MODE
Press the "ON/OFF" button to switch the device off
Press the "enter" button and while holding it pressed switch the device on by the "ON/OFF" button
After there appears the dialog box requesting for a service code or an access password, release the "enter"
button. If the request for entering a password is enabled in the device, it is necessary to enter the password
first. (If you do not know the password, enter the universal one: "00000000").
Upon the request "enter service code" enter the code "7913" and press the "enter" button to confirm.
After the device's opening screen appears, press the "menu" button – if the correct service code has been
entered, the last menu item is "service functions – special".
Select the "service functions – special" item and press "enter" to enter the device service setting mode.
SERVICE MODE | PAGE 3 OF 21
3DESCRIPTION OF INDIVIDUAL SERVICE MENU ITEMS
3.1 DISTRIBUTOR LOCALIZATION
3.1.1 INITIAL LOGO OF DEVICE
This item serves for change of logo on the opening screen of the device. It provides selection of several preset
pictures and the "no logo" option in case of which the opening screen will not display any picture. For selection
use the arrows or the "select" knob – the preview of the selected picture is in the upper part of the screen. To
confirm selection of the displayed picture press "enter", to return without changes press "esc".
Note:
Possible additional pictures (logos) can be added to the device according to the supplied pattern; in case of any
questions please contact the service department.
3.1.2 PREFIX OF DEVICE NAME
This item serves for change of prefix in front of the type name of the device. The current value is displayed in the
upper part of the screen. To change the prefix press the button under the line where the current value is
displayed. After pressing there appears a keyboard where the user prefix can be entered. To confirm press
"enter", to return without changes press "esc".
Example:
BTL-5000 SWT POWER:
"BTL-" = prefix of name, can be changed;
"5000 SWT POWER" = type name of the device, cannot be changed
Note:
The prefix may be whatever, it is limited to one text line, longer text is not displayed. After switch-on of the
national keyboard it is also possible to enter special characters.
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 4 OF 21
3.1.3 INITIAL ADDRESS
This item serves for change of the address displayed on the opening screen. The current address is displayed in
the upper part of the screen; if you want to change it, press the button containing the item to be changed below
the line – then there appears a keyboard where the user address, company name, phone number etc. can be
entered. To confirm press "enter", to return without changes press "esc".
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 5 OF 21
3.1.4 DEFAULTS
The device provides preset addresses, so that it is not necessary to fill in always the full address. It is also
possible to quickly and completely delete the entire address. After selection of the address the program requires
confirmation and then the selected address shall be set. To view or edit the address follow the above-described
process. To exit the screen press "enter" or "esc".
3.1.5 VIEW THE OPENING SCREEN
This item serves for clear displaying of the full opening screen as it will look like on the device's display. To exit
the screen press "enter" or "esc".
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 6 OF 21
3.1.6 SYSTEM LANGUAGE
This item serves for setting of the language which the device switches to if the user selects to delete all data or to
restore the factory settings. The screen displays a select menu containing all languages that are loaded in the
device. The current language has a checkmark to the left. For selection use the arrows and then press "enter"; to
exit the menu without change of language press "esc".
3.2 SPECIAL SETTING
Items from this submenu are relating to the Electro generator.
Default mode function enables to select as a default unit for electro generator the voltage or current unit.
Slowly change intensity: determines the speed of setting pírustku intensity
Electro generator maximal steps of intensity determines how many mA will be changed by one slight turning of knob. Velikost jednoho kroku pidání intensity
Note: If your device does not include electro generator, do not use this function.
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 7 OF 21
3.3 CONFIGURATION - INFORMATION
3.3.1 MASTER
This item contains information about all major electronic components in the device. It is designed especially for
service purposes and for possible communication with the service department. For movement in the displayed
page use the arrows or the "select" knob. To return to the previous screen press "enter" or "esc".
3.3.2 SLAVES
This item contains information about all generators installed in the device. Here it is possible to find out which
generators are installed in the device, what are their software versions, what is their temperature, etc. It is
possible to install 4 generators at the most. To return to the previous screen press "enter" or "esc".
Note:
All the connected generators are displayed here, regardless of the variant of the device, and the number of the
displayed generators need not correspond to the number of generators which the device works with in the
standard mode. For example, if the device is in variant "N = non-upgradable" it works always with one permitted
generator, regardless of how many they actually are in the device.
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 8 OF 21
3.3.3 SERIAL NUMBER
This item contains information about the serial number of the device, the loaded program version, the device
number in the network and the time limits of operation. The entire screen is also accessible from the standard
mode via the menu. To return to the previous screen press "enter" or "esc".
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 9 OF 21
3.3.4 LIMITATION
This item displays information about the set limitation of the device. It contains limitation of all possible items. The
whole information consists of 8 screens with similar structure. For smooth switching among the screens use the
buttons with symbols ">>" or "<<".
Structure of information: The header of the table displays titles of the columns. The first column displays what the
limitation refers to (type of current or therapy, number of generators, etc.). The second column labelled "HW"
displays limitation determined by the hardware key of the device. The third column displays limitation determined
by the software used in the device. The fourth column labelled "Act" displays the resulting (current) limitation
which is set according to the previous two columns and according to the generators which are installed in the
device. The value "YES" indicates that the device supports the respective type, the value "NO" indicates that the
item is forbidden.
To return to the previous screen press "esc".
3.3.5 DIAGNOSES AND PROGRAMS
This item displays information of all diagnoses and programs in the device. To return to the previous screen press
"esc".
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 10 OF 21
3.3.6 FILE SYSTEM
This item displays information about the file systems in the device. The information is displayed separately for
each file system; for switching among them use the ">>" or "<<" buttons. The "E" file system is designed for
saving of all user data, the "S" file system serves for system purposes and is not accessible to the user.
The information about free space is available also to the user, in the service functions during standard operation.
This information serves mostly for service purposes. To return to the previous screen press "esc".
3.3.7 RTC TIME AND DATE
This item displays information about the internal time of the device. This value cannot be changed and serves
only for information. It can differ from the actual date and time. For setting date and time use menu during
standard operation of the device.
This information serves for service purposes. To return to the previous screen press "esc".
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 11 OF 21
3.4 DEVICE CONFIGURATION
3.4.1 TIME OF USAGE
This item serves for checking the operational time limitation of device when the operation of the device is time
limited.
3.5 FACTORY SETTINGS
3.5.1 SYST. TP CALIBRATION
This item serves for factory calibration of the touch panel. The values saved during this calibration are used if the
user selects return to the factory settings or file system formatting. The process of calibration is the same as in the
standard mode of the device. After start of the calibration first touch in the upper right corner of the display (follow
the displayed instructions), then in the lower left corner (again follow the displayed instructions). Then it is
possible to check the successfulness of the calibration – when touching the display there should always appear a
small white square and the tip of the touch pen shall be inside it. It is necessary to test particularly all four corners
of the display. If calibration is OK, confirm and save it by touching in the centre of the display; if it is not, press
"esc" to cancel the calibration mode without saving changes and start calibration again.
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 12 OF 21
3.5.2 SYST. CONTRAST SETTING
This function serves for factory setting of the display contrast. The stored value is used if the user selects
restoration of the factory settings or file system formatting. The contrast setting is the same as in the standard
mode of the device (except in case of fast setting by simultaneous pressing of "esc"+"enter" and turning of the
"select" knob).
After selection of this function the screen displays the bar graph with the current value. To change contrast turn
the "select" knob. The change of contrast appears immediately on the display. The set contrast should make the
display well readable. To save the set value press "enter". To return to the previous screen without saving the
new value of contrast press "esc".
3.5.3 SYSTEM LANGUAGE
This item serves for setting of the language which the device switches to if the user selects to delete all data or to
restore the factory settings. The screen displays a select menu containing all languages that are loaded in the
device. The current language has a checkmark to the left. For selection use the arrows and then press "enter"; to
exit the menu without change of language press "esc".
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 13 OF 21
3.5.4 SETTING OF HNDP AND VNDP
This function serves for setting of frequency characteristics of the display. The used default values need not be
changed. Change of these values affects the quality of displaying and the refresh frequency of the display.
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 14 OF 21
3.5.5 BURN-IN
This function serves for burn-in of completely assembled and set device. The burn-in process is controlled by an
internal program. The program is designed so that the device is loaded as much as possible, but, at the same
time, that the load corresponds to the real use of the device. From the view of reliability, the most critical aspect is
periodical heating and cooling of the device. The burn-in program works in periods. During each period one half of
the device (= one half of all generators) is switched on at full power and the other half is switched off. After
expiration of the period the halves swap. The total time of burn-in is not limited and depends on the operator. The
display informs about the current status (number of passed periods and total time of burn-in). Before starting, the
operator can change duration time of the period. To start/stop this function press the respective button.
3.5.6 FINAL CHECK
This function serves for final testing of most of the functions and modules in the device. The function runs step by
step and the operator has to successively fulfil all requirements described on the display of the device. If all
requirements of a particular point in the final check are fulfilled, press the "pass" button, otherwise press the "fail"
button; for moving among the individual points of final check use the buttons with symbols ">>" and "<<". The
"reset" button serves for putting the point into the initial state, e.g. if you want to check the point again. After
completion of all points of the final check there is displayed an overview of the test results; this overview is saved
in the memory of the device together with the date of check. To leave the final check press "esc".
The user can view the results of the last check as well as those of all the previous checks.
The "serial number" item is only information for the operator, in case of filling in the paper report.
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 15 OF 21
SERIAL NUMBER
START
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 16 OF 21
RESULTS OF THE LAST FINAL CHECK
LIST OF RESULTS OF FINAL CHECK
MEASURE OUTPUT CURRENT, VOLTAGE AND POWER
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 17 OF 21
3.6 SERVICE FUNCTIONS
3.6.1 FORMAT AND SCANDISK
The functions work in the same way as the similar service functions in the standard operation.
Attention!
Function "format" deletes all user data and settings in the device.
3.6.2 BURN-IN
Similar to the function in factory settings (see above).
3.6.3 FINAL CHECK
Similar to the function in factory settings (see above).
3.6.4 PC BRIDGE
A specialized function for control of internal generators in the device directly from the PC. The device switches
over to the mode in which it only forwards data from PC to the respective generator. This function is used for
development and service purposes.
3.6.5 VENTILATOR TEST
A separate function for testing of the ventilator in the device. After pressing of this button the ventilator is switched
on at full speed/switched off. The ventilator status must correspond to the indication on the display.
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 18 OF 21
3.6.6 SPEAKER TEST
This function serves for testing of correct connection of the speaker. Pressing of this button starts/stops playing of
the tune from the speaker. The playing status must correspond to the indication on the screen.
3.6.7 BUTTONS TEST
This item serves for testing of all buttons on the device. The screen displays images of all buttons of the device
(including those on the knobs). After pressing a button there should appear its name and its image should be
highlighted. To pass the test, no one button image may remain inactive. To end testing press the button on the
screen (not the "esc" button as usually).
3.6.8 KNOBS TEST
This item serves for testing of all knobs on the device. The screen displays images of all knobs of the device.
After turning a knob in one and then in the other direction there should appear the knob's name, and its image
with the respective direction should be highlighted. To pass the test, no one knob image may remain inactive.
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 19 OF 21
3.6.9 RTC TIME AND DATE
This item serves for verification of functionality of the device's real time internal clock. Similar to the function in
configuration - information settings (see above).
3.7 SD CARD SERVICES
VIEW ENCY PICTURES - this items serves for browsing all pictures saved in SD card.
CREATE FILE LISTS TO SD CARD - function is intended ONLY for the production. It scans the structure of SD
card, creates the list of all files on SD card and then this list saves to SD card (this list is used when the device
starts).
DELETE FILE LISTS FROM DISC L - function is intended ONLY for the production. It deletes the file list from the
disc L.
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 20 OF 21
3.8 DELETE ALL KEY FILES
This function deletes all files with HW key in the device.
3.9 REGIONAL SETTING
This item presets the default setting of device from which the device is starting (US market and the rest of the
world).
DESCRIPTION OF THE SERVICE MENU - SERVICE MODE | PAGE 21 OF 21
BTL-5000 SWT
TROUBLESHOOTING
PAGE 1 OF 2
101VS01/12/2008EN
Fault description Checking out / Error localization Result Action / Remedy
is the main switch of each device in the "I" position? no swtch the main switch in the "I" position
After pressing the ON/OFF switch, nothing happens
The applicator gives strange noise, generation of
shocks is irregular
Low energy of the applied shocks at the max. setting
Therapy cannot be started
Is the compressor unit connected with the main unit by the
communication cable?
is the 115V/230V switch in the proper position? no
are the fuses of the main unit OK? no replace the fuses
is the nut of the bumper tight? no tighten the nut of the bumper
is the nut of the replacement kit tight?
yes replace the tube-and-roller kit
is the nut of the bumper tight? no tighten the nut of the bumper
is the nut of the replacement kit tight?
yes replace the tube-and-roller kit
is the generation of shocks irregular? no replace the tube-and-roller kit
is the applicator connected properly? no connect the applicator
is the correct channel selected? no select the channel of the active accessories
is there any visible damage of pins of the applicator connector? yes replace the applicator
connect the compressor unit by the communication
no
cable with the main unit
switch the switch over to the proper position
corresponding to the local mains
no tighten the nut of the replacement kit
no tighten the nut of the replacement kit
The applicator does not generate shocks
is the set intensity too low? yes increase the intensity to the maximum
are there leaks of the compressed air in the applicator connector? yes replace the applicator
is the way of control of therapy in the Specific Settings of the unit
correct?
the valve in the applicator responses, but shocks are generated at the
maximum intensity setting?
the cumber of the shocks applied increases, but the applicator does not
generate shocks
does the applicator respond to pressing of the button? yes replace the applicator
PAGE 2 OF 3
yes replace the applicator
correct the way of control of therapy in the Specific
no
Settings
no replace the tube-and-roller kit
Error code Fault description checking out / error localization result action/remedy
connect the compressor unit by the communication
1400 Unexpectedly shutoff
Is the compressor unit connected with the main unit by the
communication cable ?
No
cable with the main unit
Yes resta rt the ma i n unit
1402 Pressure up the system f ailed
1403
1404
1405
1406
Internal EEPROM error – counter has not
been written
Internal EEPROM error – the structure for
the new tube has not been initialized
Internal EEPROM error – the configuration
has not been written
Internal EEPROM error – reading
con figuration fail ed
1407 Internal EEPROM error – reading failed
1408 Internal voltage out of range 22V – 26V
Do you hear any signs of leaking the air between compressor
and proportional valve?
Is the compressor running?
Is the compressor unit connected with the main unit by the
communication cable ?
Are connections inside the compressor unit connected properly?
Does the supply adapter, inside the compressor unit, work well.
Does has the adapter output voltage 24V?
Yes
check all pressure connections and tightness between
compressor and proportional valve
No check the connection of compressor
No replace the compressor
Yes resta rt the ma i n unit
connect all connections in the compressor unit
No
properly
Yes replace the supply adapter inside the compressor unit
replace PCB 014-52GENPNEU
Yes
replace PCB 014-52MAI NS
Do you hear any signs of leaking the air from applicator and/or
connector?
Do y ou hear any signs of leakin g the air fr om compres sor un i t?
1409
Difference between set pressure and
output pressure of proportional valve is too
high.
PAGE 3 OF 3
Yes replace applicator
Yes
check and repair all pressure connections from
proportional valve to output connector.
No replace the proportional valve
BTL-5000
BTL 5000 SWT
BTL-4000 Topline
Process of Upload of
Application from PC
by serial line - for end
users
USER‘S MANUAL
PAGE 1 OF 5
101z3IE28/08/2008EN
CONTENTS
1 NEEDED EQUIPMENT AND TOOLS............................................................................................................. 3
2 CONTROLS OF FIRMWARELOADER.......................................................................................................... 3
3 PROCESS OF UPLOAD ................................................................................................................................ 4
4 CONNECTION TO PC.................................................................................................................................... 4
5 LOADING OF FIRMWARE............................................................................................................................. 5
6 RESET AND DISCONNECTION OF THE DEVICE........................................................................................ 5
PAGE 2 OF 5
1NEEDED EQUIPMENT AND TOOLS
program equipment "FirmwareLoaderMini.exe" version 5.91 or higher - hereinafter "FirmwareLoader"
serial loading cable
PC with MS Windows operating system and a serial port (it is possible to use a USB-COM adapter)
2CONTROLS OF FIRMWARELOADER
1 path and name of the file with new firmware 6 selection of communication port
2 button for starting of upload of new firmware 7 setting of communication speed
3 button for searching for a file with new firmware 8 button for check of correct connection of the device to PC
4 process of upload of new firmware 9 button for restart of the device
5 button for automatic searching 10 button for termination of the program FirmwareLoader
APPENDIX | PAGE 3 OF 5
3PROCESS OF UPLOAD
Switch the device off by the mains switch (rocker switch on the rear panel). Connect the mains cable to the device
and switch on the mains switch. Then switch the device on by longer holding of the ON/OFF button on the front
panel. After a while the button should light up and after releasing it the device should remain in the ON status
(shining). After some 15 seconds there should appear the opening screen of the application. If the device contains
the proper version of the application, do not continue. The version of the application is stated in the initial title on
the start-up screen (OS SP2001, ver XX.XX.XXXX) or in the menu item "unit settings" – "unit information".
4CONNECTION TO PC
Connect the loading cable to the "com out" connector on the rear panel. The other end of the cable connect to the
serial port of the PC – it is also possible to use the USB-COM adapter.
Switch the device off, press simultaneously the two lower knobs (INTENSITY, TIME/STOP) and the ENTER
button and, when holding them pressed, switch the device on. The display should remain dark and after a while
(approx. 10 seconds) the device should play a short tune – then you can release the buttons. Now the device is in
the mode for upload of firmware by external serial line.
Start the "FirmwareLoader" program on the PC.
We recommend setting the communication speed to 115200 bit/s (7).
Set the correct communication port (6) and communication speed (7). It is possible to use the "Auto find" function
(5) which will try to find the device in all available serial ports.
(Note: In some types of computer this function may not work reliably.)
Check the connection with the device by the "Check connection" button (8). If everything works OK, the dialog
box with the message "Unit present" must appear after pressing this button. If there appears the message "Unit
not found", the computer cannot connect to the device for a reason – it is necessary to find out why, try to repeat
the previous steps with other communication speed, check the cable, etc.
Note:
If the program does not find the device on the first try, it is suitable to press the "Check connection" button (8) for
several more times, the device could respond later.
APPENDIX | PAGE 4 OF 5
5LOADING OF FIRMWARE
!!! ATTENTION !!!
Loading of firmware is a very delicate process which under certain conditions can disable work with the
device. The loading program is designed to minimize the possibility of an error; however, some risk still
remains. Therefore it is recommended that you have some experience with computers before performing
this action.
Do not switch the device and/or the PC during loading – incorrectly loaded firmware could cause locking
of the device (this status could require even a servicing intervention of the manufacturer).
During loading do not move the window with the running application in the Windows environment,
otherwise there is a risk of a programming error – in such case restart programming.
Press the "Browse file" button (3) with a directory tree symbol to select the file with the application to be loaded to
the device. The selected file is displayed in the upper field (1). The program remembers the last setting.
Press the "Program file auto" button (2) to start loading of the program to the device. First there is detected
configuration of the device and consistency of the loaded file. This part is indicated by the message "Reading unit
information…" (4). If any incompatibilities or other problems are found, there appears an error message with the
description of the error. These errors are usually caused by incorrect or corrupted input file – in such case it is
necessary to get a new file.
After this testing the process of loading is started – it is indicated by the message "Programming…" (4). If any
problem appears during programming, an error message is displayed. During loading the program informs about
the current percentage status of completion. If an error occurs during programming, do not switch the device off
and restart programming. If the error repeats, contact the service department.
If programming passed OK, there appears the dialog box "Complete OK". Now you can reset and disconnect the
device.
6RESET AND DISCONNECTION OF THE DEVICE
If this work with the device is complete, disconnect it by pressing the "Restart unit" button (9). The device first
restarts and then the PC disconnects from it. After restart of the device there should appear the standard opening
screen of the application and it is possible to standardly work with the device. To check that the new version of
firmware is loaded in the device, see the "unit information" menu item.
APPENDIX | PAGE 5 OF 5
BTL-5000
BTL 5000 SWT
BTL-4000 Topline
Process of Upload - for
service
USER‘S MANUAL
PAGE 1 OF 6
101z3IE04/08/2008EN
CONTENTS
1 GENERAL INFORMATION............................................................................................................................ 3
2 NEEDED EQUIPMENT AND TOOLS............................................................................................................. 3
3 CONNECTION TO PC.................................................................................................................................... 3
4 UNIT HW INFO............................................................................................................................................... 5
5 LOADING OF FIRMWARE............................................................................................................................. 5
6 RESET AND DISCONNECTION OF THE DEVICE........................................................................................ 5
7 APPENDIX - SUPPLEMENTARY DESCRIPTION AND EXPLANATION...................................................... 6
PAGE 2 OF 6
1 PROCESS OF UPLOAD OF APPLICATION FROM PC BY
SERIAL LINE - FOR SERVICE
Application shall be loaded into the assembled equipment containing an application or the program loader.
2NEEDED EQUIPMENT AND TOOLS
program equipment "FirmwareLoaderService.exe" version 3.02 or higher - hereinafter "FirmwareLoader"
serial loading cable
PC with operating system and a COM port (it is possible to use a USB-COM adapter)
3CONNECTION TO PC
Switch the device off, press simultaneously the two lower knobs (INTENSITY, TIME/START/STOP) and the
ENTER button and, when holding them pressed, switch the device on. The display should remain dark and after a
while (approx. 10 seconds) the device should play a short tune. If so, the device is in the mode for upload of
firmware by external serial line.
Connect the loading cable to the "com out" connector on the rear panel. The other end of the cable connect to the
serial port of the PC – it is also possible to use the USB-COM adapter.
Start the "FirmwareLoader.exe" program on the PC.
We recommend setting the communication speed to 115200 bit/s.
The version of the loader can be found out after loading of HW Info (see below). If you are not sure you can leave
this item checked off, the loading will work even with the new devices, but it will be slower.
Set the proper communication port ("Settings" tab) and communication rate. It is possible to use the "Auto find"
function which will try to find the device in all available serial ports (only the ports really connected to the computer
are displayed).
To check the connection with the device at any time press the "Check connection" button in the program. If
everything works OK, the dialog box with the message "Unit present" must appear then.
If there appears the message "Unit not found", the computer cannot connect to the device for a reason – it is
necessary to find out why, try to repeat the previous steps with other communication rate, check the cable, etc.
PROCESS OF UPLOAD | PAGE 3 OF 6
The version of the loader can be found out after loading of HW Info (see below). If you are not sure you can leave
this item checked off, the loading will work even with the new devices, but it will be slower.
Set the proper communication port ("Settings" tab) and communication rate. It is possible to use the "Auto find"
function which will try to find the device in all available serial ports (only the ports really connected to the computer
are displayed).
To check the connection with the device at any time press the "Check connection" button in the program. If
everything works OK, the dialog box with the message "Unit present" must appear then.
If there appears the message "Unit not found", the computer cannot connect to the device for a reason – it is
necessary to find out why, try to repeat the previous steps with other communication rate, check the cable, etc.
PROCESS OF UPLOAD | PAGE 4 OF 6
4UNIT HW INFO
Switch over to the "Unit info" tab and press the "Reload" button in the upper left corner of the window.
The percentage status of communication should appear at the bottom of the window. After a while a table in the
program should list the hardware configuration of the tested device. On the "Unit info" tab you can see the
hardware configuration of the device. The key data are especially the version of the loader and the versions of
particular printed circuits in the equipment. Unit info can be updated (reloaded from the device) at any time by
pressing the "Reload" button.
5LOADING OF FIRMWARE
If the information in the Unit info is correct it is possible to start loading of firmware. Switch to the "Programming"
tab.
Press the button with a directory tree symbol to select the file with the application to be loaded to the device (30056autoXXX_C.bin or 300-56autoXXX_BW.bin). The selected file is displayed in the upper field. The program
remembers the last setting.
Press the "Program file auto" button to start the procedure which compares the selected file with data from the
device – if the device is compatible, this procedure modifies the loaded file according to parameters read from the
device and starts loading of the program to the device. If any problem occurs, there appears an error message
with possible description of the error. During loading the program informs about the current percentage status of
completion.
If programming passed OK, there appears the dialog box "Complete OK". Now you can reset and disconnect the
device.
To abort programming press the "Abort action" button. However, after starting a device with incompletely loaded
firmware there is a risk of its failure.
6RESET AND DISCONNECTION OF THE DEVICE
If this work with the device is complete, disconnect it by pressing the "Restart unit" button. The device first restarts
itself and then the PC disconnects from it.
PROCESS OF UPLOAD | PAGE 5 OF 6
7 APPENDIX - SUPPLEMENTARY DESCRIPTION AND
EXPLANATION
Loading is completely automatic, i.e. it is not necessary to check the versions in the device and the loaded file.
The program automatically modifies itself according to the parameters read from the device – i.e., if the program
finds out that the device contains another type of hardware than is set in the loaded file, the loaded file is
internally modified before loading so that it corresponds to the configuration of the device. The exception is files
for colour and b/w device which are so much different that they cannot be mutually substituted and modified. This
automatic control is fully functional if there have not been any major hardware changes in the device. If there have
been such changes, it is necessary to load the correct new configuration (i.e. file with the correct new hardware
configuration) at least once to the device. A typical example is replacement of the supply part in physiotherapy
devices by a more powerful one: in such case it is not possible to load the universal file to the device as usually,
but at the first loading after the replacement it is necessary to load to the device a specific file containing this
change. Next loading works standardly with the automatic configuration.
It is possible to upload to several devices at the same time – their number is limited only by the number of COM
ports on the PC. Each of these devices has its own FirmwareLoader session running.
During loading do not move the window with the running application in the Windows environment, otherwise there
is a risk of a programming error – in such case restart programming.
APPENDIX | PAGE 6 OF 6
BTL-5000 SWT
Service process of the
upload of application in
the SWT compressor
unit from PC via the
serial line
USER‘S MANUAL
PAGE 1 OF 6
100IE04/08/2008EN
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.