BTL 4000 Topline User manual

BTL-4000 Topline Series
USER‘S GUIDE
GENERAL CHARACTERISTICS | Page 1 of 67
100IE20/12/2007
BEFORE YOU START
Take a moment to reflect on the advantages of the BTL-4000 Topline Electrical
Stimulation, Ultrasound, Laser and Magnet technology in your own clinic. The BTL-4000
Topline system has many benefits not available in other systems. For example, the touch
screen is a major step forward since it allows users to precisely monitor the therapy and
document and store patient data for later recall. A choice of therapy protocols offers
maximum flexibility in a variety of clinical applications.
The combination electrical stimulation / ultrasound / laser and magnet therapy system
also offers substantial benefits since it eliminates the need to purchase separate units.
We sincerely believe the latest BTL physiotherapy system is technically superior to any
other physiotherapy products available and will provide years of trouble-free and
profitable use.
All of us at BTL wish you every success with your BTL-4000 Topline system. We pride
ourselves on being as responsive as possible to our customers’ needs. Your suggestions
and comments are always welcome since we believe an ongoing relationship with our
customers is critically important to our future product line. Please call or email us your
suggestions.
While we would like you to start using your equipment right away, we encourage you to
thoroughly read this manual in order to fully understand the operational features of the
BTL-4000 Topline system.
Again, thank you for being a BTL customer. In the event of any problem, or if you require
service, please make an initial call to your local distributor, who will decide whether to
refer the problem to our office.
PAGE 2 OF 67
CONTENTS
1 GENERAL CHARACTERISTICS........................................................................................................................ 6
2 INSTRUCTIONS FOR USE ................................................................................................................................ 7
2.1 Top Panel ...................................................................................................................................................... 7
2.2 Rear Panel..................................................................................................................................................... 8
2.3 Front Panel .................................................................................................................................................... 8
2.4 Installation of Touch Pen Holder.................................................................................................................... 9
2.5 Unpacking And Assembly.............................................................................................................................. 9
2.6 Operating the Unit ....................................................................................................................................... 12
2.6.1 Touch Screen......................................................................................................................................... 12
2.6.2 Numerical Keyboard............................................................................................................................... 13
2.7 Therapy Setting ........................................................................................................................................... 14
2.7.1 Therapy Flow Chart................................................................................................................................ 14
2.7.2 Welcome Screen and Selection of Channels, Tabs and Accessories.................................................... 15
2.7.3 Setting Therapy Parameters Via Diag Option ........................................................................................ 16
2.7.4 Setting Therapy Parameters Via Prog Option ........................................................................................ 17
2.7.5 Setting Therapy Parameters Manually Via The ‘MAN' Button................................................................ 18
2.7.6 Therapy Parameters Screen – Ergonomic, Standard and Expert Mode ................................................ 19
2.8 Course of Therapy ....................................................................................................................................... 20
2.8.1 Start, Interrupt and End of Therapy........................................................................................................ 20
2.8.2 Running Therapy Screen ....................................................................................................................... 21
2.8.3 Electrotherapy – Setting During Therapy ............................................................................................... 21
2.8.4 Accessories / Applicators – Visual Signalization. ................................................................................... 22
2.8.5 Indication of Operation – Energy On Operation ..................................................................................... 24
2.9 Therapy Parameters.................................................................................................................................... 26
2.10 ENCYCLOPAEDIA ...................................................................................................................................... 26
2.11 Therapy Saving ........................................................................................................................................... 27
2.11.1 Save Therapy......................................................................................................................................... 27
2.11.2 Save Therapy and Add It To The Patient Data ...................................................................................... 28
2.11.3 CopYing the Settings of The Therapy Parameters Between Tabs of the Same Type............................ 29
2.12 Interconnection of Units ............................................................................................................................... 30
2.12.1 Interconnection of BTL 4000 Topline Puls (combi) and Vacuum Unit BTL Vac ..................................... 30
2.12.2 Interconnection of BTL-4000 Topline Puls and BTL-4000 Topline Sono................................................ 30
2.12.3 Interconnection of BTL-4000 Topline Puls, BTL-4000 Topline Sono and BTL Vac ................................ 31
2.12.4 Setup and Operation of Combined Therapy in INDIVIDUAL Devices .................................................... 32
2.12.5 Stopping Combined Therapy in INDIVIDUAL Devices........................................................................... 32
3 MENU BUTTON................................................................................................................................................ 33
3.1 Accessories ................................................................................................................................................. 33
3.1.1 Installation of Accessories...................................................................................................................... 33
3.1.2 Information About Accessories .............................................................................................................. 33
3.1.3 Connectors – Information....................................................................................................................... 33
3.2 Unit Settings ................................................................................................................................................ 34
3.2.1 Password Setting ................................................................................................................................... 34
3.2.2 Sound Setting ........................................................................................................................................ 34
3.2.3 Screen Saver and Auto Switch-Off ........................................................................................................ 34
3.2.4 Setting of Colours .................................................................................................................................. 35
3.2.5 Display Options...................................................................................................................................... 35
PAGE 3 OF 67
3.2.6 Date and Time Setting ........................................................................................................................... 35
3.2.7 Language Setting................................................................................................................................... 35
3.2.8 Operation Mode ..................................................................................................................................... 35
3.2.9 Touch Panel Calibration......................................................................................................................... 35
3.2.10 User Options .......................................................................................................................................... 36
3.2.11 Style of Operation .................................................................................................................................. 36
3.2.12 Setting of HW Key.................................................................................................................................. 36
3.2.13 Unit Information...................................................................................................................................... 36
3.2.13 Unlock Code .......................................................................................................................................... 36
3.2.14 Service Functions .................................................................................................................................. 36
3.3 Special Settings........................................................................................................................................... 37
4 USER SETTINGS OPTION............................................................................................................................... 38
4.1 Clients.......................................................................................................................................................... 38
4.2 User Sequences .......................................................................................................................................... 39
4.2.1 Creating New Sequence ........................................................................................................................ 39
4.2.2 Parameters of Sections in Sequence..................................................................................................... 41
4.2.3 Saving New Sequence........................................................................................................................... 41
4.3 User Diagnoses / Programs......................................................................................................................... 42
4.4 List of Recent Therapies.............................................................................................................................. 42
5 ACCESSORIES ................................................................................................................................................ 43
5.1 Power Supply Adapter 60W / Adapter 90W................................................................................................. 43
5.2 Accumulator................................................................................................................................................. 43
5.3 Lithium Battery............................................................................................................................................. 44
5.4 Accessories Common for All Units .............................................................................................................. 44
5.5 Accessories for Electrotherapy .................................................................................................................... 44
5.6 Accessories for Ultrasound Therapy............................................................................................................ 45
5.7 Accessories for Laser Therapy .................................................................................................................... 45
5.8 Accessories for Magnetotherapy ................................................................................................................. 45
6 MAINTENANCE AND SAFETY INSTRUCTIONS ............................................................................................ 46
6.1 Safety .......................................................................................................................................................... 47
6.2 Contraindications......................................................................................................................................... 51
6.2.1 Contraindications for Electrotherapy ...................................................................................................... 51
6.2.2 Contraindications for Ultrasound............................................................................................................ 52
6.2.3 Contraindications for Laser Therapy ...................................................................................................... 52
6.2.4 Contraindications for Magnetotherapy ................................................................................................... 53
6.3 Useful Addresses......................................................................................................................................... 53
6.4 Warranty ...................................................................................................................................................... 53
7 TECHNICAL PARAMETERS ........................................................................................................................... 54
7.1 Technical Parameters of The BTL-4000 Topline Series Devices ................................................................ 54
7.2 Technical Parameters of Power Supply Adapter 60W / Adapter 90W ......................................................... 55
7.3 Basic Parameters of Electrotherapy Generator ........................................................................................... 56
7.4 Basic Parameters of Ultrasound Generator ................................................................................................. 56
7.5 Basic Parameters of Laser Generator ......................................................................................................... 57
7.6 Basic parameters of Magnet Generator....................................................................................................... 57
7.7 Technical Parameters of Ultrasound Heads ................................................................................................ 58
7.8 Technical Parameters of Laser Probes........................................................................................................ 58
7.9 Technical Parameters of Laser Clusters...................................................................................................... 59
7.10 Technical parameters of Magnetic Applicators ............................................................................................ 60
7.11 Applicable Standards................................................................................................................................... 60
7.12 Interconnection of Devices .......................................................................................................................... 61
PAGE 4 OF 67
7.13 Manufacturer ............................................................................................................................................... 61
8 UNITS CONFIGURATIONS.............................................................................................................................. 62
8.1 Table of Configurations of the Combined Devices BTL 4000 Topline Series .............................................. 62
8.2 Table of Configurations of the Electrotherapy Devices BTL-4000 Topline Puls........................................... 65
8.3 Table of Configurations of the Ultrasound Therapy Devices BTL-4000 Topline Sono ................................. 66
8.4 Table of Configurations of the Laser Therapy Devices BTL-4000 Topline Laser......................................... 66
8.5 Table of Configurations of the Magnetotherapy Devices BTL-4000 Topline Magnet ................................... 67
PAGE 5 OF 67
1 GENERAL CHARACTERISTICS
The BTL-4000 Topline Series offers advanced and well designed physiotherapy units for professional use.
Depending on the required configuration, each system can consist of up to four units - for electrotherapy,
ultrasound, laser or magnet therapy treatment.
The touch-controlled display considerably simplifies the use of the unit. The touch screen is equipped with a touch
stylus for more convenient operation. The vertically positioned case of the instrument enables you to see the
information on the screen clearly and from different positions. In addition, the display's brightness can be adjusted
to fit the light conditions in the office. The information displayed on the screen will guide you throughout the whole
therapy. Simply adjust the parameters by pressing the touch screen buttons and turn the main knob to set the
intensity.
The modular design of the BTL equipment allows you to build the combination you require. Combine an
electrotherapy unit of your choice with either ultrasound, laser or magnet all in a single unit. This can save
considerable money in your physiotherapy investment
Selecting a diagnosis from a list of alphabetically organized treatment protocols, or selecting a program, will make
an easy and efficient start of the therapy. Naturally, you can adjust any treatment parameter manually by the
simple use of the touch screen buttons. Throughout the whole therapy, the display informs you about the
remaining therapy time, channel and therapeutic method used, the type of therapy applied, attached accessories,
and other necessary data.
If several accessories are attached to your unit at the same time, you can easily recognize the accessory required
for a specific treatment. Select a treatment on the display (electrotherapy, ultrasound, laser or magnet), and the
control light on the corresponding accessory (electrotherapy cable, ultrasound head, laser probe / cluster or
magnet disc / double disc / linear or solenoid applicator) will start blinking to indicate that this accessory should be
used.
Save time by using the pre-programming of the BTL-4000 Topline units. Based on detailed research and practical
use of the units, the well-organized pre-programmed protocols will give you recommendations for treating various
conditions. The unit also includes up to 500 free lots to define your own protocols. Moreover, you can recall the
last 20 treatments.
Add the names of your patients and other relevant information to the unit's internal memory and connect the
patient data with pre-programmed or your own protocols. When your patients call again, simply recall their name
and apply the pre-set therapy.
With every BTL unit, you can purchase a cart specially designed for BTL products. Its versatile design allows you
to conveniently store and use 1 or 2 physiotherapy units and a vacuum unit. The cart includes a range of
accessory trays and baskets. Four well-built and steady castors ensure easy movement of the unit in the office or
hospital.
Please visit our corporate website at http://www.btlnet.com
services.
GENERAL CHARACTERISTICS | Page 6 of 67
for the latest information on BTL products and
2 INSTRUCTIONS FOR USE
2.1 TOP PANEL
1. - 6 outputs for patient cables on the rear panel of the device, see chapter Rear Panel
7. display
8 diag/prog button, for quick selection of diagnosis or programs
9 man button, for manual setting of all therapy parameters
10 menu button, to set date, time, language, display contrast, sounds, user options, etc.
11 enter button to confirm selection or setting
12 esc button to cancel selection or setting and return to the original setting
13 select knob to select individual parameters
INSTRUCTIONS FOR USE | PAGE 7 OF 67
2.2 REAR PANEL
1 -6 patient outputs – for exact configuration, see table Configuration of Output Connectors
14 mains switch for switching the device on/off – positions I / 0.
15 socket for connection of external power supply Adapter 60W / Adapter 90W
16 warning label with parameters of power supply and input of the device
2.3 FRONT PANEL
17 display of the remaining time of therapy for the active channel
18 display of the current intensity for the active channel
19 start button, for starting or pausing the therapy on the active channel
20 stop button, for pausing or stopping the therapy on the active channel
21 on/off switch – serves for switching the device on/off
22 service connector under a cover and the type, manufacture and warning labels (placed on the bottom
cover of the device)
INSTRUCTIONS FOR USE | PAGE 8 OF 67
2.4 INSTALLATION OF TOUCH PEN HOLDER
Lean the upper part of the holder against the upper part of any selected gap between the ribs on the rear panel of
the device. Press the holder to plug the parts indicated on the picture up to stop.
2.5 UNPACKING AND ASSEMBLY
Inspect the box for damage and report any damage to the carrier and your distributor. Do not proceed with
installation and assembly if the box is damaged.
Unpack the equipment and place it on a stable horizontal surface suitable for the equipment's weight. Always
position the unit out of direct sunlight as this may make the touch screen difficult to read. Always position the unit
away from direct heat sources such as radiators or a room heater. Cooling of the equipment is provided by forced
air circulation. Cooling vents are located on the rear panel and at the bottom of the equipment and must not be
covered. Place the equipment so that the free space behind the rear side is at least 10 cm. Do not position the
equipment on a soft surface which may obstruct air flow to the bottom cooling vents. Do not put any heat-
producing devices or objects containing water or other liquid on the equipment. Do not place the equipment close
to devices producing strong electromagnetic, electric or magnetic field (diathermy, X-rays, etc.), as equipment
electronics could be undesirably influenced. In case of any questions, please call your distributor or service agent.
The same installation conditions should be observed for the Adapter 60W / Adapter 90W power supply adapter.
Retain the original packaging to ensure future possible transportation of the device.
Plug the device into the mains socket by means of the Adapter 60W / Adapter 90W power supply adapter
(see chapter Technical Parameters).
PLUG THE POWER SUPPLY ADAPTER DIRECTLY INTO THE MAINS SOCKET; DO NOT USE ANY MULTI-
CONNECTION EXTENSION CABLE OR ADAPTOR.
In case of any questions, please call your distributor or service agent.
Switching the device on:
Connect the power supply adapter to the device and plug its mains cable into the mains socket, switch the I/ 0
rocker switch (14) on the rear panel to the I position and finally press the ON/OFF switch (21) on the front panel.
The system will then run a self-test. If the self-test finds no faults, the screen will display the equipment type and it
is ready for use - see note below.
INSTRUCTIONS FOR USE | PAGE 9 OF 67
Connection of accessories
Connect the accessories to the output connectors (1) - (6) on the rear panel in this way:
Put in the cable connector and secure the fluted ring by pressing and turning in clockwise. ATTENTION WHEN
DISCONNECTING THE CONNECTOR, first of all, it is necessary to take by the fingers the fluted ring NOT THE
WHOLE connector. TURN THE FLUTED RING ANTICLOCKWISE and then after releasing the ring, disconnect
the connector by pulling the fluted ring still held in the fingers!
ATTENTION!!! DO NOT TURN THE WHOLE CONNECTOR USING FORCE, THE DEVICE COULD BE
DAMAGED.
For configuration of the output connectors (1) to (6), see the table Configuration of Output Connectors
The unit will automatically detect the accessories type and display it on the screen. If a wrong one is connected,
the equipment will not operate and the screen will display a warning and help about where to connect which
accessory.
Recharging the accumulator
The device accumulator is sold in half-charged status. That is why we recommend formatting the accumulator
after purchase of the device: connect the device to the mains via the external adapter for at least 48 hours without
interruption, with the mains rocker switch (14) in position I. The device will be recharged and the accumulator
will be properly formatted. A properly formatted accumulator enables longer operation of the device without mains
supply. For details, see the chapter Accumulator.
Restart of the device - reset
If for any reason (electromagnetic interference, etc.) the device stops responding to the user's commands and the
message "please wait..." with small moving squares is not displayed, the device can be put into the initial state by
simultaneously pressing the man (9) and enter (11) buttons. After this, the device immediately goes into the initial
status.
Note:
After switching on, the unit tests for about 10 - 15 secs all internal functions. If any fault exists, the screen will
display a warning. If any fault that compromises patient safety exists, the system will 'lock' itself into ‘secure’
mode. If this situation occurs, please call your local distributor for service advice.
INSTRUCTIONS FOR USE | PAGE 10 OF 67
Tab. 2.1 Configuration of output connectors
Type
Output 1 Output 2 Output 3 Output 4 Output 5 Output 6
BTL-4610 Topline E1 BTL-4615 Topline E1 BTL-4620 Topline E2 E1 BTL-4625 Topline E2 E1 BTL-4810S Topline S1A S1B E1 BTL-4815S Topline S1A S1B E1 BTL-4810L Topline L1A L1B DOOR E1 BTL-4815L Topline L1A L1B DOOR E1 BTL-4810M2 Topline E1 M1 M2 BTL-4815M2 Topline E1 M1 M2 BTL-4626Topline E1A E1B E2A E2B BTL-4628Topline E1A E1B E2A E2B BTL-4816S Topline E1A E1B S1B S1A BTL-4818S Topline E1A E1B S1B S1A BTL-4816L Topline E1A E1B DOOR L1A BTL-4818L Topline E1A E1B DOOR L1A BTL-4816M2 Topline E1A E1B M1 M2 BTL-4818M2 Topline E1A E1B M1 M2 BTL-4800SL Topline S1A S1B DOOR L1A BTL-4920 Topline M1 M2 BTL-4800LM2 Topline L1A DOOR M1 M2 BTL-4800SM2 Topline S1A S1B M1 M2 BTL-4110 Topline L1A L1B DOOR ACUP BTL-4710 Topline S1A S1B E input E output BTL-4210 Topline T1A T1B E input E output BTL-4640 Topline E2B E2A E1B E1A BTL-4645 Topline E2B E2A E1B E1A BTL-4820S Topline S1A S1B E2 E1 BTL-4825S Topline S1A S1B E2 E1 BTL-4820L Topline L1A DOOR E2 E1 BTL-4825L Topline L1A DOOR E2 E1 BTL-4820M2 Topline E2 E1 M1 M2 BTL-4825M2 Topline E2 E1 M1 M2
* units 4610 Topline and 4615 Topline can be supplied or upgraded with the HVT module
Notes:
E1 connector for connection of electrotherapy accessories (BTL-236-1(2), BTL vac) to E1 generator
E2 connector for connection of electrotherapy accessories (BTL-236-1(2), BTL vac) to E2 generator
E1A, E1B connector for connection of electrotherapy accessories (BTL-236-1(2), BTL vac) to E1 generator
(adaptor)
E2A, E2B connector for connection of electrotherapy accessories (BTL-236-1(2), BTL vac) to E2 generator
(adaptor)
E input connector for electrotherapy input on the ultrasound unit for combined therapy
E output connector for electrotherapy output on the ultrasound unit for combined therapy
L1A, L1B connector for connection of laser probe/cluster BTL-448/BTL-445 to L1 generator
S1A, S1B connector for connection of ultrasound heads BTL-237 (e.g. 1 cm
) to U1 generator
M1, M2 connector for connection of magnetotherapy applicators (BTL-239) to M1 generator
Door connector for sensor of open door
Acup connector for acupuncture electrode
How many patients and which outputs you can simultaneously connect to can be seen on the display after
pressing the menu button in menu / accessories / connectors – information.
INSTRUCTIONS FOR USE | PAGE 11 OF 67
2.6 OPERATING THE UNIT
2.6.1 TOUCH SCREEN
The touch screen may be operated by finger touch or using the special soft tip stylus that is supplied with the unit.
DO NOT TOUCH THE SCREEN WITH A SHARP OBJECT OR PEN, AS THIS MAY CAUSE PERMANENT
DAMAGE.
Select required parameters by pressing:
1. 3D buttons.
2. Bright tabs of the selected channel (in the lower left corner of the screen) to switch between connected
accessories, such as ultrasound heads or laser accessories.
3. The dark tab of the required channel (in the lower left corner of the screen) to switch between channels.
Touch screen buttons:
The touch screen buttons have three dimensional (3D) shading and may be pressed with the finger or special
stylus. To confirm the requested changes or values, press enter. To cancel the changes, press escape (esc).
Selected channel:
Although most of the configurations allow the running of several therapies at one time, only one channel can be
controlled at once. The tab of this selected channel is shaded light. All information on the screen and all controls
are related to this channel. The most important information about the therapies on the other channels is visible on
their tabs.
Indication of
connected accessory
Channel marking:
E – electrotherapy 1 – number of E
generator
Information about other generator possibilities: combined therapies, HVT etc...
Name of screen for the selected channel
3D buttons
(for therapy on the
selected channel)
Information about connected accessory: (e.g.): 1 – 1 one cable 4 – 4 electrodes
Buttons with the same function as (11) and (12)
Tab of the unselected channel
Tab with running therapy of the unselected channel
Tab of the selected channel
INSTRUCTIONS FOR USE | PAGE 12 OF 67
2.6.2 NUMERICAL KEYBOARD
In addition to setting the numerical value with the select (13) button, you can also use the “numerical keyboard”.
Press this icon to open the window with the numerical keyboard:
Set numerical values of the parameter that has been selected - the “orange” button in the left picture above. Enter
the value and press enter (11) to return to the original screen. Press esc (12) to exit the screen. If you set a value
that exceeds the allowed value range (allowed value range is stated at the top of the screen), or if the unit cannot
set the required value, the value will be rounded as close as possible to the allowed value.
INSTRUCTIONS FOR USE | PAGE 13 OF 67
2.7 THERAPY SETTING
2.7.1 THERAPY FLOW CHART
INSTRUCTIONS FOR USE | PAGE 14 OF 67
2.7.2 WELCOME SCREEN AND SELECTION OF CHANNELS, TABS AND ACCESSORIES
The welcome screen, accessed upon power up, displays channel tabs and icons showing which accessories are
connected. The number of channels displayed depends on the unit configuration. The following diagram shows
that almost the entire display is available for the selected channel.
If more accessories are connected to one generator (e.g. two ultrasound heads connected to a single ultrasound
generator), pressing the channel tab of the generator will switch between accessories.
The colour of the selected channel tab is light grey.
Examples of information on tabs:
Tab of channel E2, which is selected and has no accessories
Tab of selected channel E1 with the possibility to apply HVT therapy and with electrotherapy
accessory BTL-236-1 with two electrodes. HVT generator can be added only to units 4610 Topline
and 4615 Topline
Tab of channel E1 (which is not selected), with electrotherapy accessory BTL-236-2 and possibility to
apply combined therapy
Tab of non selected generator M1 with connected disc applicator BTL-239-2.
Tab of selected generator U1 with connected ultrasound heads. Press the tab
to switch between the 1cm
and 4cm2 heads.
Tab of selected generator L1 with connected laser probe and cluster. Press
the tab to switch between the 830nm probe and the cluster.
INSTRUCTIONS FOR USE | PAGE 15 OF 67
2.7.3 SETTING THERAPY PARAMETERS VIA DIAG OPTION
Press the diag/prog (8) button once to display a list of diagnoses / therapy protocols. Each channel tab has its
own list of therapy protocols. The letter in front of each number corresponds to the type of selected therapy:
E – electrotherapy; U – ultrasound; L – laser; M – magnetotherapy. For example, the channel tab of the
ultrasound generator includes a list of therapy protocols for ultrasound treatment. If the tab depicts an HVT (only
for
BTL-4610 Topline and BTL-4615 Topline units) or combined therapy symbols (see the chapter Welcome Screen
and Selection of Channels, Tabs and Accessories), the list includes protocols for HVT or combined therapies.
To find a therapy protocol fast, press the button with the initial letter of the required protocol. The selected letter
depends on how many times the button is pressed. For example, after pressing the MNOP button once, there are
listed protocols starting with the first letter, M. Pressing MNOP twice = N, three times = O and four times = P. The
currently selected letter is displayed in the box to the left of the buttons.
To select the found required diagnosis, press the enter button (11). If the protocol has more therapies, they are
listed after selecting the protocol.
See the chapter: Therapy Parameters
Screen – Ergonomic, Standard and
INSTRUCTIONS FOR USE | PAGE 16 OF 67
Saved by user protocols can easily be find in the “*user diagnoses / programs” item, which appears at the top
(bottom according to the chosen sorting type) of the diagnoses list.
2.7.4 SETTING THERAPY PARAMETERS VIA PROG OPTION
Press the diag/prog (8) button twice to set the required program number. The program numbers generally
correspond to the program numbers used in the traditional BTL physiotherapy line. The letter in front of each
number corresponds to the type of therapy selected: E – electrotherapy; U – ultrasound; L – laser;
M – magnetotherapy.
Combined therapies
symbol of the ultrasound head (see the chapter Welcome Screen and Selection of Channels, Tabs and
Accessories). They are in the positions E-35xx - E-39xx.
Your own therapy protocols
electrotherapy, U-80xx - U-89xx for ultrasound therapies, L-80xx - L-89xx for laser therapies and
M-80xx - M-89xx for magnetotherapy.
Your own sequences
U-95xx - U-99xx for ultrasound therapies, L-95xx - L-99xx for laser therapies and M-95xx - M-99xx for
magnetotherapy.
Programs recommended for diagnoses can be found in the User’s Guide.
E+U are listed among E programs and can be found on the electrotherapy tab with the
(except sequences) can be saved under the following numbers: E-80xx - E-89xx for
are saved under the following numbers: E-95xx - E-99xx for electrotherapy,
For fast program number selection, use the numerical keyboard, see the chapter Numerical Keyboard for
details.
INSTRUCTIONS FOR USE | PAGE 17 OF 67
2.7.5 SETTING THERAPY PARAMETERS MANUALLY VIA THE ‘MAN' BUTTON
Press the man (9) button to select manual setting for the therapy. You may store manual settings for use at a
later time.
Press individual 3D buttons to open menus and setting screens. The majority of screens are accompanied by
illustrating pictures and symbols. See the example below:
INSTRUCTIONS FOR USE | PAGE 18 OF 67
2.7.6 THERAPY PARAMETERS SCREEN – ERGONOMIC, STANDARD AND EXPERT MODE
This screen opens after loading a diagnosis or a program via pressing the diag/prog (8) button respectively once
or twice (see the chapter Therapy Flow Chart), before the start of therapy. The screen shows either the most
important therapy parameters (you have selected the ergonomic mode) or all information about the therapy (you
have selected the expert or standard mode). In addition, in expert mode you can modify all parameters.
The differences between modes are best seen here:
ergonomic mode standard mode
expert mode
Set the operation mode via the menu (10) button – refer to the chapter Operation Mode. For fast switching to
expert mode from any other operation mode, press the man (9) button.
SETTING THERAPY TIME
Press the time screen button on the therapy parameters screen to set the required time. Use the select knob or
the numeric keyboard in the top left corner.
INTENSITY SETTING
For the ultrasound, laser and magnet therapies, the intensity (output) can be set only on the therapy parameters
screen and only when the therapy is not running. To set the intensity, press the intensity button.
Electrotherapy: Intensity is set during therapy and can only be adjusted by turning the select (13) knob.
INSTRUCTIONS FOR USE | PAGE 19 OF 67
y
2.8 COURSE OF THERAPY
2.8.1 START, INTERRUPT AND END OF THERAPY
To start a therapy on the selected channel, press the start (19) button. The therapy can start only if the therapy
parameters screen is displayed. To pause the therapy, press the start (19) button again or the stop (20) button.
The paused therapy can be resumed by again pressing the start (19) button or stopped by pressing the esc (12)
button.
While the therapy is paused, you can adjust the time (except for laser therapies and all types of sequences) by
the select (13) knob.
For electrotherapy, you can adjust the intensity during therapy running by turning the select (13) knob to the
right (to increase intensity) or to the left (to decrease intensity).
Direction of
intensit
Laser therapy can also be started / paused by the start / stop button on the laser probe.
START / STOP
INSTRUCTIONS FOR USE | PAGE 20 OF 67
2.8.2 RUNNING THERAPY SCREEN
Therapy description
Time remaining till the
Indication whether
monitoring of the
contact between
therapy applicator and
Indication of sound
Symbolic description
2.8.3 ELECTROTHERAPY – SETTING DURING THERAPY
SETTING THE INTENSITY IN 4-ELECTRODE THERAPIES
Making a therapy with four electrodes, you can set different intensities between each pair of electrodes. In case of
4-pole interference, the output intensity is set by the select (13) knob on both channels at the same time (the
screen button 1+2 is pressed) or on each channel separately (the screen button 1 or 2 is pressed) according to
your ordered configuration.
Icon and info for the
Intensity
Info for the contact
between the ultrasound
head and the patient
If the intensity on one of the channels is set to zero, the therapy is terminated.
INSTRUCTIONS FOR USE | PAGE 21 OF 67