Brooks Instrument 1305 Operating Manual

Installation and Operation Manual
X-VA-1305-eng Part Number: 541B080AAG October, 2017
Model 1305 O-ring Seal Flowmeter
Model GT1305
5
Installation and Operation Manual
X-VA-1305-eng
Part Number: 541B080AAG
Model GT1305
October, 2017
Essential Instructions
Read before proceeding!
Brooks Instrument designs, manufactures and tests its products to meet many national and international standards. These products must be properly installed, operated and maintained to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, operating and maintaining Brooks Instrument products.
To ensure proper performance, use qualified personnel to install, operate, update, program and maintain the product.
Read all instructions prior to installing, operating and servicing the product. If this instruction manual is not the correct manual, please see back cover for local sales office contact information. Save this instruction manual for future reference.
WARNING: Do not operate this instrument in excess of the specifications listed in the Instruction and Operation Manual. Failure to heed
this warning can result in serious personal injury and / or damage to the equipment.
If you do not understand any of the instructions, contact your Brooks Instrument representative for clarification.
Follow all warnings, cautions and instructions marked on and supplied with the product.
WARNING: Prior to installation ensure this instrument has the required approval ratings to meet local and national codes. Failure to heed this warning can
result in serious personal injury and / or damage to the equipment.
Install your equipment as specified in the installation instructions of the appropriate instruction manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
Operation: (1) Slowly initiate flow into the system. Open process valves slowly to avoid flow surges. (2) Check for leaks around the flow meter inlet and outlet connections. If no leaks are present, bring the system up to the operating pressure.
Please make sure that the process line pressure is removed prior to service. When replacement parts are required, ensure that qualified people use replacement parts specified by Brooks Instrument. Unauthorized parts and procedures can affect the product's performance and place the safe operation of your process at risk. Look-alike substitutions may result in fire, electrical hazards or improper operation.
Ensure that all equipment doors are closed and protective covers are in place to prevent electrical shock and personal injury, except when maintenance is being performed by qualified persons.
WARNING: For liquid flow devices, if the inlet and outlet valves adjacent to the devices are to be closed for any reason, the devices must be completely
drained. Failure to do so may result in thermal expansion of the liquid that can rupture the device and may cause personal injury.
European Pressure Equipment Directive (PED)
All pressure equipment with an internal pressure greater than 0.5 bar (g) and a size larger than 25mm or 1" (inch) falls under the Pressure Equipment Directive (PED).
The Specifications Section of this manual contains instructions related to the PED directive.
Products described in this manual are in compliance with EN directive 2014/34/EU.
All Brooks Instrument Flowmeters fall under fluid group 1.
Products larger than 25mm or 1" (inch) are in compliance with PED category I, II or III.
Products of 25mm or 1" (inch) or smaller are Sound Engineering Practice (SEP).
European Electromagnetic Compatibility (EMC)
The Brooks Instrument (electric/electronic) equipment bearing the CE mark has been successfully tested to the regulations of the Electro Magnetic Compatibility (EMC directive 2014/30/EU). Special attention however is required when selecting the signal cable to be used with CE marked equipment.
Quality of the signal cable, cable glands and connectors:
Brooks Instrument supplies high quality cable(s) which meets the specifications for CE certification. If you provide your own signal cable you should use a cable which is overall completely screened with a 100% shield. “D” or “Circular” type connectors used should be shielded with a metal shield. If applicable, metal cable glands must be used providing cable screen clamping. The cable screen should be connected to the metal shell or gland and shielded at both ends over 360 Degrees. The shield should be terminated to an earth ground. Card Edge Connectors are standard non-metallic. The cables used must be screened with 100% shield to comply with CE certification. The shield should be terminated to an earth ground. For pin configuration : Please refer to the enclosed Instruction Manual.
ESD (Electrostatic Discharge)
CAUTION: This instrument contains electronic components that are susceptible to damage by static electricity. Proper handling procedures must be observed
during the removal, installation or other handling of internal circuit boards or devices. Handling Procedure:
1. Power to unit must be removed.
2. Personnel must be grounded, via a wrist strap or other safe, suitable means before any printed circuit card or other internal device is installed, removed or adjusted.
3. Printed circuit cards must be transported in a conductive container. Boards must not be removed from protective enclosure until immediately before installation. Removed boards must immediately be placed in protective container for transport, storage or return to factory.
Comments
This instrument is not unique in its content of ESD (electrostatic discharge) sensitive components. Most modern electronic designs contain components that utilize metal oxide technology (NMOS, SMOS, etc.). Experience has proven that even small amounts of static electricity can damage or destroy these devices. Damaged components, even though they appear to function properly, exhibit early failure.
Installation and Operation Manual
X-VA-1305-eng Part Number: 541B080AAG October, 2017
Dear Customer , We appreciate this opportunity to service your flow measurement and control requirements with a Brooks
Instrument device. Every day, flow customers all over the world turn to Brooks Instrument for solutions to their gas and liquid low-flow applications. Brooks provides an array of flow measurement and control products for various industries from biopharmaceuticals, oil and gas, fuel cell research and chemicals, to medical devices, analytical instrumentation, semiconductor manufacturing, and more.
The Brooks product you have just received is of the highest quality available, offering superior performance, reliability and value to the user. It is designed with the ever changing process conditions, accuracy requirements and hostile process environments in mind to provide you with a lifetime of dependable service.
We recommend that you read this manual in its entirety. Should you require any additional information concerning Brooks products and services, please contact your local Brooks Sales and Service Office listed on the back cover of this manual or visit www.BrooksInstrument.com
Y ours sincerely , Brooks Instrument
Model GT1305
WARNING
GLASS TUBE
EXPLOSION
HAZARD
Plastic protective sleeve must remain over glass tube.
Fasten meter windowssecurely. Do not operate above pressure and temperature
limits. Avoid pressure and flow surges. Do not service or repair while pressurized. Read and understand instruction manual. Failure to comply could result in serious personal
injury or property damage.
WARNING
GLASS TUBE EXPLOSION HAZARD
Protective sleeve Fasten meter windows securely. Failure to comply could result in serious personal
injury or property damage.
must
remain
over
glass
tube.
Model GT1305
Installation and Operation Manual
X-VA-1305-eng
Part Number: 541B080AAG
October, 2017
THIS PAGE WAS INTENTIONALLY
LEFT BLANK
Installation and Operation Manual
X-VA-1305-eng Part Number: 541B080AAG October, 2017
Paragraph Page
Number Number
Section 1 Introduction
1-1 Description .........................................................................................................................................1-1
1-2 Design Features.................................................................................................................................1-1
1-3 Specifications .....................................................................................................................................1-1
1-4 Optional Equipment............................................................................................................................1-4
Section 2 Installation
2-1 Receipt of Equipment.........................................................................................................................2-1
2-2 Recommended Storage Practice .......................................................................................................2-1
2-3 Return Shipment ................................................................................................................................2-2
2-4 T ransit Precautions ............................................................................................................................2-2
2-5 Removal from Storage ....................................................................................................................... 2-2
2-6 Installation ..........................................................................................................................................2-2
Model GT1305
Contents
Section 3 Operation
3-1 Operating Instructions ........................................................................................................................3-1
Section 4 Maintenance and Cleaning
4-1 General ..............................................................................................................................................4-1
Section A Essential Instructions.............................................................................................................A-1
Warranty, Local Sales/Service Contact Information....................................................................... Back Cover
Figures
Figure Page
Number Number
1-1 Model 1305 Dimensions.....................................................................................................................1-4
2-1 T ypical Byp ass Installation .................................................................................................................2-3
2-2 Panel Cutouts.....................................................................................................................................2-4
3-1 T ypical Byp ass Installation .................................................................................................................3-2
3-2 Reading Edge of Floats......................................................................................................................3-2
Tables
Table Page
Number Number
1-1 Product S pecifications........................................................................................................................1-3
1-2 Capacities ..........................................................................................................................................1-3
1-3 Pressure and Temperature Ratings ...................................................................................................1-3
i
Contents
Model GT1305
Installation and Operation Manual
X-VA-1305-eng
Part Number: 541B080AAG
October, 2017
THIS PAGE WAS INTENTIONALLY
LEFT BLANK
ii
Installation and Operation Manual
WARNING
X-VA-1305-eng Part Number: 541B080AAG October, 2017
1-1 Description
1-2 Design Features
Section 1 Introduction
Model GT1305
The Brooks® Model 1305 glass tube flowmeters provide a low cost means of flow rate indication. They are designed for applications where the metering accuracy requirements are not severe. These meters are ideal for purge type service, seal oil systems, bearing lubrication and cooling water indication. Model 1305 is available with vertical inlets and outlets and standard materials of construction only .
• Capacity ranges 0.8 to 25 gpm; 3 to 50 scfm
• O-ring seal construction
• Sturdy bolted side-plate construction
• Standard NPT connections
• Vertical inlets and outlet s standard
1-3 Specifications (Reference Tables 1-1 and 1-2)
Do not operate this instrument in excess of the specifications listed below. Failure to heed this warning can result in serious personal injury and/or damage to the equipment.
Glass metering tubes are designed for operation up to the maximum operating pressures and temperatures as specified herein. Due to the inherent brittle characteristics of glass and conditions beyond our control, tube breakage could result even within specified operating conditions. Do not use glass tube meters with fluids that are toxic, or chemically react with glass such as water above 140°F, steam, alkalis, fluorine, hydrofluoric acid, or molten metal. Failure to heed warning can result in serious personal injury and/or damage to the equipment.
WARNING
1-1
Section 1 Introduction
Model GT1305
Installation and Operation Manual
X-VA-1305-eng
Part Number: 541B080AAG
October, 2017
WARNING
GLASS TUBE
EXPLOSION
HAZARD
Plastic protective sleeve must remain over glass tube.
Fasten meter windows securely. Do not operate above pressure and temperature
limits. Avoid pressure and flow surges. Do not service or repair while pressurized. Read and understand instruction manual. Failure to comply could result in serious personal
injury or property damage.
WARNING
GLASS TUBE EXPLOSION HAZARD
Protective sleeve Fasten meter windows securely. Failure to comply could result in serious personal
injury or property damage.
must
remain
over
glass
tube.
1-2
Installation and Operation Manual
X-VA-1305-eng Part Number: 541B080AAG October, 2017
Table 1-1 Product Specifications
Capacities and PrCapacities and Pr
Capacities and Pr
Capacities and PrCapacities and Pr
Max Max
WW
orking Prorking Pr
Max
W
orking Pr
Max Max
WW
orking Prorking Pr
PED CategoryPED Category
PED Category SEP
PED CategoryPED Category
StandarStandar
d Flow d Flow
Standar
d Flow
StandarStandar
d Flow d Flow
Optional Flow Optional Flow
Optional Flow
Optional Flow Optional Flow
RepeatabilityRepeatability
Repeatability 1% Full Scale
RepeatabilityRepeatability
Scale Length
Scale Length
Scale Length 75mm (3”)
Scale LengthScale Length
Scale GrScale Gr
Scale Gr
Scale GrScale Gr
AmbienAmbien
t t
Ambien
t
AmbienAmbien
t t
OperOper
ating Fluid ating Fluid
Oper
ating Fluid
OperOper
ating Fluid ating Fluid
ConnectionsConnections
Connections Standard: Female NPT
ConnectionsConnections
Connection OrienConnection Orien
Connection Orien
Connection OrienConnection Orien
DimensionsDimensions
Dimensions See Dimensions Figure 1-1
DimensionsDimensions
Materials of ConstructionMaterials of Construction
Materials of Construction
Materials of Construction
Materials of Construction
Metering Metering
Metering
Metering Metering PrPr
otective otective
Pr
otective
PrPr
otective otective
WindowWindow
Window Scratch resistant, UV stabilized polycarbonate
WindowWindow Float & Float StopsFloat & Float Stops
Float & Float Stops 316 stainless steel
Float & Float StopsFloat & Float Stops HousingHousing
Housing Aluminum side plates
HousingHousing End FittingsEnd Fittings
End Fittings Brass, 316 stainless steel
End FittingsEnd Fittings O-ringsO-rings
O-rings Buna-N, Viton®-A fluoroelastomers
O-ringsO-rings BoltsBolts
Bolts 300 series stainless steel
BoltsBolts
essuressur
e Dre Dr
opsops
essur
e Dr
ops See Capacities Table 1-2
essuressur
e Dre Dr
opsops
essuressur
e* at 250°Fe* at 250°F
essur
e* at 250°F
essuressur
e* at 250°Fe* at 250°F
AccurAccur
acyacy
Accur
acy ±10% Full Scale from 100% to 10% of scale reading
AccurAccur
acyacy
AccurAccur
acyacy
Accur
acy ±5% Full Scale from 100% to 10% of scale reading
AccurAccur
acyacy
aduationsaduations
aduations Std.: On-tube printed scale (Linear scale, direct reading gpm water or scfm air)
aduationsaduations
TT
emperemper
aturatur
T
emper
TT
emperemper
TT
T
TT
e Limitse Limits
atur
e Limits 33 to 125°F (1°C to 52°C)
aturatur
e Limitse Limits
TT
emperemper
T
emper
TT
emperemper
tationtation
tation Vertical inlets and outlets
tationtation
ubeube
ube Borosilicate glass
ubeube TT
ube Sleeveube Sleeve
T
ube Sleeve UV stabilized polycarbonate
TT
ube Sleeveube Sleeve
/121°C/121°C
/121°C See Pressure and Temperature Ratings Table 1-3
/121°C/121°C
Opt.: Detachable printed scale (Choice of direct reading units, millimeter or percentage of maximum flow with factor tag)
aturatur
e Limitse Limits
atur
e Limits See Pressure and Temperature Ratings Table 1-3
aturatur
e Limitse Limits
Section 1 Introduction
Model GT1305
Table 1-2 Capacities
WATER AIR @ STP
MET ER PRESSURE D ROP PRESSURE D ROP
SIZE TUBE FLOAT GPM INCHES W.C. FLOAT SCFM INCHES W.C.
8
10
NOTE: ALL AIR FLOWS ARE AT 14.7 PSIA AND 70 DEG F
R-8M-75-1 8-RJ-23 2.50 13.0 8-RJ-23 10.00 15.0
R-10M-75-3 10-RJ-83 35.00 35.0
8-RV-8 0.80 8.0 8-RV-8 3.00 9.0 8-RJ-10 1.50 9.0 8-RJ-10 6.00 11.0
8-RJ-30 3.50 19.0 8-RJ-30 14.00 22.0 8-RJ-39 5.00 34.0 10-RJ-80 12.00 18.0 10-RJ-37 24.00 18.0
10-RJ-180 21.00 71.0 10-RJ-90 50.00 71.0
Table 1-3 Pressure and Temperature Ratings
Pr Pr
essuressur
Pr
Pr Pr
MeterMeter
Meter
MeterMeter
SizeSize
Size
SizeSize
#8 200 13.8 SEP 250 121
#10 150 10.3 SEP 250 121
Pr Pr
essuressur
Pr
essur
Pr Pr
essuressur
MeterMeter
Meter
MeterMeter
SizeSize
Size
SizeSize
#8 145 10 SEP 250 121
#10 94 6.5 SEP 250 121
e Rating for Female e Rating for Female
essur
e Rating for Female
essuressur
e Rating for Female e Rating for Female
Max. OperMax. Oper
Max. Oper
Max. OperMax. Oper
PrPr
essuressur
ee
Pr
essur
e
PrPr
essuressur
PrPr
Pr
PrPr
essuressur
essur
essuressur
ee
ee
e
ee
PSIG Bar °F °C
e Rating for Female e Rating for Female
e Rating for Female
e Rating for Female e Rating for Female
Max. OperMax. Oper
Max. Oper
Max. OperMax. Oper
PSIG Bar °F °C
ThrThr
Thr
ThrThr
atingating
ating
atingating
ThrThr
eaded 1305 Units eaded 1305 Units
Thr
eaded 1305 Units
ThrThr
eaded 1305 Units eaded 1305 Units
atingating
ating
atingating
eaded 1305 Unitseaded 1305 Units
eaded 1305 Units
eaded 1305 Unitseaded 1305 Units
Max. OperMax. Oper
Max. Oper
PEDPED
PED
PEDPED
PEDPED
PED
PEDPED
Max. OperMax. Oper
TT
T
TT
With CRN With CRN
With CRN
With CRN With CRN
Max. OperMax. Oper
Max. Oper
Max. OperMax. Oper
TT
T
TT
emperemper
emper
emperemper
emperemper
emper
emperemper
aturatur
atur
aturatur
ApprAppr
Appr
ApprAppr
aturatur
atur
aturatur
atingating
ating
atingating
ee
e
ee
ovaloval
oval
ovaloval
atingating
ating
atingating
ee
e
ee
1-3
Section 1 Introduction
Model GT1305
Top View
"C"
Outlet Connection
"A"
Installation and Operation Manual
X-VA-1305-eng
Part Number: 541B080AAG
October, 2017
"D"
"B"
Scale
Length
Inlet Connection
"A"
Front View
Approximate Dimensions 1305 Flow Indicators
Meter Scale "A"
Size Length (Conn. Size) "B" "C" "D"
8 75 mm 1/2" NPT 7-3/4" (196.85) 2-3/8" (60.33) 2-3/8" (60.33)
10 75 mm 1" NPT 7-3/4" (196.85) 2-3/8" (60.33) 2-3/8" (60.33)
Side View
DIMENSIONS INCHES (mm)
1-4
Figure 1-1 Model 1305 Dimensions
Installation and Operation Manual
X-VA-1305-eng Part Number: 541B080AAG October, 2017
2-1 Receipt of Equipment
Section 2 Installation
Model GT1305
When the equipment is received, the outside packing case should be checked for damage incurred during shipment. If the packing case is damaged, the local carrier should be notified at once regarding his liability. A report should be submitted to the nearest Brooks Instrument location listed on the Global Service Network page on our website:
BrooksInstrument.com/GlobalSupportCenters
Remove the envelope containing the packing list. Carefully remove the instrument from the packing case. Make sure spare parts are not discarded with the packing materials. Inspect for damaged or missing parts.
Remove the envelope containing the packing list. Carefully remove the instrument from the packing case. Make sure spare part s are not discarded with the packing materials. Inspect for damaged or missing parts.
2-2 Recommended Storage Practice
2-3 Return Shipment
If intermediate or long-term storage of equipment is required, it is recommended that the equipment be stored in accordance with the following:
a. Within the original shipping container . b. S tored in a sheltered area, preferably a warm, dry, heated
warehouse.
c. Ambient temperature of 70° F (21° C) nominal, 109° F (43° C)
maximum, 45° F (7° C) minimum.
d. Relative humidity 45% nominal, 60% maximum, 25% minimum.
Upon removal from storage a visual inspection should be conducted to verify the condition of equipment is "as received".
Prior to returning any instrument to the factory for any reason, visit our website for instructions on how to obtain a Return Materials Authorization Number (RMA #) and complete a Decontamination Statement to accompany it: BrooksInstrument.com/Service. All instruments returned to Brooks also require a Material Safety Data Sheet (MSDS) for the fluid(s) used in the instrument. Failure to provide this information will delay processing of the instrument. Instrument must have been purged in accordance with the following:
2-1
Section 2 Installation
CAUTION
CAUTION
Model GT1305
2-4 Transit Precautions
2-5 Removal from Storage
2-6 Installation
Installation and Operation Manual
X-VA-1305-eng
Part Number: 541B080AAG
October, 2017
To safeguard against damage during transit, transport the instrument to the installation site in the same container used for transport ation from the factory if circumstances permit.
Upon removal of the instrument from storage, a visual inspection should be conducted to verify its "as-received" condition.
A. Location For proper operation the Model 1305 must be mounted within 6 degrees of true vertical, with the inlet connection at the bottom and the outlet at the top. The use of a level is recommended to assure vertical positioning. Piping must be adequately supported to prevent undue strain on the flowmeter.
B. Piping Arrangement It is recommended that bypass piping be installed around the flowmeter so it may be isolated from the process line for servicing and cleaning. Refer to Figure 2-1 for a typical installation.
Do not allow the float to fall out of the metering tube. A damaged float will affect the accuracy of the meter. Be careful not to break the tube by pulling on it at an extreme angle or applying excessive force.
Failure to drain the flowmeter when isolated in a bypass loop may result in tube breakage caused by thermal expansion of the process liquid.
2-2
C. Panel Mounting If panel mounting is used refer to Figure 2-2 for the required cutout.