Bronkhorst IQ+FLOW Quick Start

Quick Installation Guide
Doc nr.: 9.17.074B Date: 14-04-2015
SCOPE OF THIS GUIDE
1. Check IQ+FLOW properties
Before installing it is important to read the attached label and check:
Instrument type:
Temperature
IQ(F/M) - Gas Flow IQP - Pressure
Output/Setpoint
A – 0…5 Vdc B – 0…10 Vdc F – 0…20 mA G – 4…20 mA
2. Check test-pressure
T B n S s
3. Check if system piping is clean
F U R r
IQ+FLOW® digital mass flow/pressure meters and controllers for gases are one of the smallest chip based instruments of its kind. IQ
+
FLOW® mass flow meters/controllers can be applied for measuring and fast control of gas flows up to
5000 mln/min, for applications with pressure conditions up to 10 bar (145 psi) and temperatures between 5 and 50 °C
+
(41 to 122 °F). IQ
FLOW® pressure meters/controllers can be applied for pressure ranges from 0,01…0,5 bar up to
0,2…10 bar. The instruments can either be operated in analog or digital mode (RS232 or RS485).
+
This guide covers the short-form instructions for IQ
FLOW® instruments regarding:
start-up mounting zeroing operation
+
This manual will help you start-up your IQ
FLOW® in 10 steps. More information can be found in documents listed below:
Instruction manual IQ+FLOW® series (document nr. 9.17.045)
Consists information for basic and advanced operation, more detailed product information and instructions for troubleshooting
FlowPlot Manual (document nr. 9.17.030) Hook-up diagram IQ+FLOW® Single-channel (document nr. 9.16.101) Hook-up diagram IQ+FLOW® Multi-channel (document nr. 9.16.090)
These documents can be downloaded from the website: www.bronkhorst.com
or can be sent by e-mail on request.
Starting-up
- gas (IQF) red label or
- pressure (IQP) yellow label Flow rate Fluid to be measured / controlled Up- and downstream pressure Input and output signals
IQ+FLOW® instruments are designed for dry, clean, inert and non-explosive gases. Do not use the instruments for gases that do not belong to this category.
he tested pressure is stated on the instrument with a red-coloured sticker. efore installation, make sure that the test pressure is in accordance with ormal safety factors for your application. If there is no Pressure Testing ticker on the device or if the test pressure is incorrect, the instrument
hould not be mounted in the process line and be returned to the factory.
or reliable measurement always make sure that the fluid stream is clean.
se filters to ensure a clean, moisture- and oil- free gas stream.
ecommended pore-size: 7 µm. If back flow can occur, a downstream filter is
ecommended too. Be aware of the pressure drop caused by using filters.
2
9.17.074
4a. Mount / install instrument properly
F
or IQ+FLOW® the upright position is preferred. When using an IQ+FLOW®
i p v c a
Install the IQ+FLOW® instrument in the line, in accordance with the direction of the FLOW arrow. The arrow for f fittings. Special types of fittings are available on request.
4b. Compression type fittings
F f h
4c. 10-32 UNF fittings
T o O l f t d
4d. Mounting downported instruments
F and that they are dry and free from dirt or dust.
7a. Analog operation (For Single-channel versions only):
F R
7b. Digital RS232 operation
D u R B F
nstrument in up- or downward position make sure to “zero” the instrument
rior to use (see step 10). Avoid installation in close proximity of mechanic ibration and/or heat sources. The housing of the instrument is according to lass IP40, which means that the instrument is suitable for indoor (dry)
pplications, like laboratories or well protected (OEM) housings.
low direction is indicated on the body of the instrument. If applicable follow the guidelines of the supplier of the
or leak tight installation of compression type fittings make sure that the tube is inserted up to the shoulder in the
itting body and that no dirt or dust is present on tube, ferrules or fittings. Tighten the nut finger tight; while
olding the instrument, then tighten the nut one turn.
ighten the 10-32 UNF fittings according to the instructions of the supplier
f the fittings.
nly use 1/16” tubing with a straight and clean cut without burrs to ensure
eak tightness. Preferably deburr the tubing prior to installation. A new
errule connection must be made for each new adapter to ensure leak­ightness and minimum dead volume, due to variances in the adapter
imensions.
or downported instruments make sure that the seals are present at the bottom, that the surfaces are undamaged
5. Leak check
Check the system for leaks before applying pressure. Especially if toxic or other dangerous fluids are used!
6. Electrical connection
Electrical connections must be made with a standard cable or according to the IQ
+
FLOW® hook-up diagram. These diagrams can be found on the “Documentation and software tools” CD-ROM and at the download section of www.bronkhorst.com. IQ
+
FLOW® instruments are powered with +15…+24 Vdc.
or analog operation refer to the “Hook-up diagram IQ+FLOW®” or use an J45 loose-end cable (7.03.419) to connect the required signals.
igital operation over RS232 can be established when sing the following setup. Using a RS232 cable or a USB­S232 converter with a PC will allow you to use (free) ronkhorst® software for Windows, such as FlowDDE and
© 2015 Bronkhorst High-Tech B.V. 3
lowPlot.
7c. Bus / Digital RS485 operation
8a. Micro-switch operation (For Single-channel versions only)
U a f a a t
8b. Rotary switch operation (For Multi-channel versions only)
S s
= 1 c
With digital operation over RS485 a bus-system with multiple instruments can be set up. Refer to the “Instruction Manual IQ
+
FLOW® series” for possible systems.
Multi-functional switch
sing the micro-switch on the instruments, several actions can be monitored
nd started. The green LED is used for status indication. The red LED is used
or errors/warnings/messages. The micro-switch can be used to start several
ctions, such as auto-zero, restore factory settings and bus-initialisation ctions, if applicable. See specific zero-procedure below and section 3.3 of
he Instruction Manual for more details.
Status LED
Error/Warning LED
elect the communication type and Baud rate with the “COMM. TYPE”
witch. Select the node address with the “MSD” and “LSD” switch (e.g. MSD
and LSD = 9 selects node 19 for channel 1, but also node 20 and 21 for hannels 2 and 3). See section 3.3 of the Instruction Manual for more details.
Unless specified otherwise, the IQ+FLOW® instrument is set for RS232 communication at 38k4 baud on node address 03 by default.
Warm-up time:
Let the IQ
+
FLOW® warm-up for at least 30 minutes for best accuracy.
Do not apply pressure until electrical connections are made. When applying pressure to the system, avoid pressure shocks in the system and increase pressure gradually. Also decrease pressure gradually when required.
9. Purging
In systems for use with toxic or other dangerous fluids, purging for at least 30 minutes with a dry, inert gas (like Nitrogen or Argon) is absolutely necessary before use. After use with toxic or other dangerous fluids, complete purging is also required before exposing the system to air.
10. Zeroing (optional)
The zero point of each instrument is factory adjusted. If required the zero point can be re-adjusted over RS232 or by means of using the micro-switch. Procedure for zeroing by-micro switch (not for pressure meter/controller):
Warm-up, pressure up the system and fill the instrument according to the process conditions. Make sure no flow is going through the instrument by closing valves near the instrument. The setpoint must be zero. Press micro-switch and hold it. After a short time the red LED will go ON and OFF, then the green LED will go ON.
At that moment release the micro-switch.
The zeroing procedure will start at that moment and the green LED will blink fast. The zeroing procedure waits
for a stable signal and saves the zero. If the signal is not stable, zeroing will take long and the nearest point to zero is accepted. The procedure will take approximately 10 seconds.
When the indication is showing 0% signal and the green indication LED is burning continuously again, the zeroing
action was successful.
Ready
Your IQ
+
FLOW® instrument is now ready for operation.