Bronkhorst E-5700 User Manual

Instruction manual
Power supply
and readout units
E-5700 series
Doc. no.: 9.17.009G Date: 25-07-2011
ATTENTION
Please read this instruction manual carefully before installing and operating the instrument.
BRONKHORST HIGH-TECH B.V.
BRONKHORST HIGH-TECH B.V.
Even though care has been taken in the preparation and publication of the contents of this manual, we do not assume legal or other liability for an y inaccuracy, mistake, mis­statement or any other error of whatsoever nature contained herein. The material in this manual is for information purposes only, and is subject to change without notice.
Bronkhorst High-Tech B.V.
July 2011
Warranty
The products of Bronkhorst High-Tech B.V. are warranteed against defects in material and workmanship for a period of three years from the date of shipment, provided they are used in accordance with the ordering specifications and the instructions in this manual and that they are not subjected to abuse, physical damage or contamination. Products that do not operated properly during this period may be repaired or replaced at no charge. Repairs are normally warranteed for one year or the balance of the original warranty, whichever is the longer. See also paragraph 9 of the Conditions of Sales.
The warranty includes all initial and latent defects, random failures, and indeterminable internal causes.
It excludes failures and damage caused by the customer, such as contamination, improper electrical hook-up, dropping etc.
Re-conditioning of products primarily returned for warranty service that is partly or wholly judged non-warranty may be charged for.
Bronkhorst High-Tech B.V. prepays outgoing freight charges when any part of the service is performed under warranty, unless otherwise agreed upon beforehand. However, if the product has been returned collect to Bronkhorst High-Tech B.V., these costs are added to the repair invoice. Import and/or export charges, foreign shipping methods/carriers are paid for by the customer.
BRONKHORST HIGH-TECH B.V.
TABLE OF CONTENTS
1 Introduction
1.1 General description.............................................................................................................................page 5
1.2 Model configuration.............................................................................................................................page 5
1.3 Power supply.......................................................................................................................................page 6
1.4 In/output signals..................................................................................................................................page 6
1.4.1 Rearpanel connectors.........................................................................................................................page 6
1.4.2 Connection to measuring and controlling devices..............................................................................page 6
1.4.3 Connection to remote equipment........................................................................................................page 6
1.4.3.1 Analog input/output signals.................................................................................................................page 7
1.4.4 EMC and cables..................................................................................................................................page 7
1.5 Specifications ......................................................................................................................................page 8
1.5.1 Housings..............................................................................................................................................page 8
1.5.2 Specifications ......................................................................................................................................page 9
2 Installation...................................................................................................................................page 10
3 Recommended spare parts .................................................................................................page 11
BRONKHORST HIGH-TECH B.V.
1 INTRODUCTION
1.1 General description
The Bronkhorst HIG H-TECH B.V. basic m odule E-5700 has been designed for powering 2 therm al Mass Flow Controllers or Pressure Controllers with 0 - 5 (10 V) output/setpoint. The output signal is shown on a 3½ -digit voltmeter as 0 ... 100%. Either channel 1 or 2 can be selected by means of a two position switch on the front. The internal/external setpoint switch provides the possibility to either select the required setpoints locally by means of the 10-turn setpoint potentiometers on the front, or use external (computer) signals. The available modular built systems are described in the model configuration.
1.2 Model configuration
E - 5 7 N N - A A A
Housing Code
½ 19” table top (42TE) 1 ½ 19” rack (42TE) 3 table top cassette (14TE) 5 panel mount cassette (14TE) 6
front view rear view
Model E-5752 E-5762
1
No. of channels Code
2 - channel 2 4 - channel 4 6 - channel 6
Output signal (Sensor)
0 - 5 Vdc 0 - 10 Vdc
INT
EXT
2
Code
A B
Output signal (Module)
0 - 5 Vdc* 0 - 10 Vdc* 0 - 20 mA 4 - 20 mA Other/Specify
* Voltage output signal of E-5700 only in
Code
A B C D Z
Ext. Setpoint signal (Module)
0 - 5 Vdc 0 - 10 Vdc 0 - 20 mA 4 - 20 mA Other/Specify
combination with identical voltage output signal of sensor.
Code
A B C D Z
Model E-5714 E-5734
INT
EXT
1
INT
EXT
1
2
2
Model E-5716 E-5736
EXT
1
INT
EXT
1
2
INT
EXTINT
1
2
2
9.17.009 page 5
BRONKHORST HIGH-TECH B.V.
1.3 Power supply
Each E-5700 housing incorporates one power supply. System setup in such that maxim al 6 channels are available. T he max . number of instrum ents depends on the individual power requirement of each instrument. The power input incorpor ates an on/of f switch and a f use. F or extra pr otection the power supply has a separate internal fuse.
1.4 In/output signals
1.4.1 Rearpanel connectors
I/O I/O
ch. 2 ch. 2
ch. 1 ch.1
1.4.2 Connection to measuring and controlling devices
The instrument connectors 1 and 2 are female sub-miniature 9-pin D-connectors. The pin designation is according to the BRONKHORST HIGH-TECH B.V. standard for analog instruments.
Pin number description
1 not used 2 0 - 5 (10) Vdc output signal 3 0 - 5 (10) Vdc setpoint signal 4 0 Valve 5 + Valve 6 - 15 Vdc power supply 7 + 15 Vdc power supply 8 0 V/common 9 ground/shield
1.4.3 Connection to remote equipment
The female in/out (sub-miniature 9-pin) D-connector 3 has the following pin configuration.
Pin number description
1 Ext. output ch. 1 2 Ext. output ch. 2 3 not used 4 Ext. input ch. 1 5 Ext. input ch. 2 6 not used 7 not used 8 0 V/common 9 ground/shield
page 6 9.17.009
le).
BRONKHORST HIGH-TECH B.V.
1.4.3.1 Ana log input/output signa ls Analog input signals should be connected to pin 4 (+) for channel 1 and/or pin 5 (+) for channel 2 and
0 Vdc/common. Analog output signals are available at pin 1 (+) for channel 1 and/or pin 2 (+) for channel 2 and 0 Vdc/common. Signals are according to one of the Bronkhorst HIGH-TECH B.V. standards. The model configuration contains a code, describing the input/output signals.
Notes: a. Max. load current output : 300 Ohm (sourcing output) b. Min. load voltage output: 10 kOhm c. Input load resistance (voltage) 1 MOhm d. Input load resistance (current) 61,4 Ohm (sinking input)
1.4.4 EMC and cables
The system setup described in this manual carries the CE-mark. Therefore it has to comply with the EMC requirements as are valid for this kind of equipment.
However compliance with the EMC requirements is not possible without the use of proper cables and connector assemblies.
For good results Bronkhorst HIGH-T ECH B.V. can provide standard cables. O therwise follow the guidelines as stated below.
Fold the shield of the cable bac k over the cable (the shield must be around t he cab
20 mm
Wind a copper tape ar ound the shield
Solder a black wire on the tape and connect to pin 9 of connector
copper tape
shielded cable
e.g. LAPP LiYCY
8 mm
other wires
D-connector housing metalized
connector
black wire
(shield)
Note: When connec ting the system to other devices ( e.g. I/O to PLC), be sure that the integrity of the shielding is not affected. Do not use unshielded wire terminals.
9.17.009 page 7
+1.0
1.5 Specifications
1.5.1 Housings
14 TE table top cassette
BRONKHORST HIGH-TECH B.V.
14 TE panel mount
14TE
96
42 TE ½ 19" table top
96
90
230 233
144
+0.8
92
mm
Panel cut out
mm
42 TE ½ 19" rack mount
page 8 9.17.009
1.5.2 Specifications
Power supply : 100 ... 240 Vac, 50 ... 60 Hz, 40 VA Signal input : consult model configuration Signal output : consult model configuration Connectors : tw o 9-pin subminiature D-connectors (female) for instrument connection
one 9-pin subminiature D-connector (female) for external output signal and external setpoint signal including mating connector
Fuse : consult system label Front : - one 3½ -digit LED indicator for indication 0 ... 100,0 %
- two setpoint potentiometers with micrometer knob setting
- one 2-position switch for DVM channel selection
- one 2-position switch for internal/external setpoint mode Setpoint potmeter : two 10-turn dial with 5 kOhm potentiometers Operating temp. : 0 - 50 °C ambient Housing : IP - 20
Note : Never block ventilation holes. Operation at high ambient temperatures over extended periods may lead to reduction of the operating
lifetime of your power supply / readout system.
BRONKHORST HIGH-TECH B.V.
9.17.009 page 9
BRONKHORST HIGH-TECH B.V.
2 INSTALLATION
Receipt of equipment:
Check the outside packing for damage incurred during shipment. Should the packing box be damaged, then the loc al carrier mus t be notified at once regarding his liability, if so required. At the same time a report should be submitted to Bronkhorst HIGH-TECH B.V. Ruurlo Holland
Remove the envelope c ontaining the pack ing lis t; carefully remove the equipm ent f rom the pac king box. do not discard spare or replacem ent parts with the packing m aterial and inspect the contents f or damaged or m issing parts.
Return shipment: When returning material: always describe the problem and if possible the work to be done, in a covering letter.
Important: Clearly note, on top of package the custom clearance number of Bronkhorst High-Tech B.V. namely:
NL801989978B01
Service: If this equipment is not properly serviced, serious personal injury and/or damage to the equipment could result. It is therefore important that servicing is performed by trained and qualified service personnel. Bronkhorst HIGH-TECH B.V. have a trained staff of servicemen available.
Installation: By connecting the cable to the 100 ... 240 Vac source, the readout/control unit can operate. The DVM shows
000.0 (only if no external meter or controller has been connected to the housing). If the LED's of display do not light on, please check fuse in 100 ... 240 Vac connection plug. Depending on the execution the units can:
1) power external sensors Eurostyle: +15 Vdc/350 mA
-15 Vdc/50 mA
2) display signals 0 - 5(10) Vdc = 0 - 100.0 % reading
3) power command setpoint signals 0 - 5(10) Vdc to external controller.
page 10 9.17.009
BRONKHORST HIGH-TECH B.V.
3 RECOMMENDED SPARE PARTS
Partnumber:
Power supply 100 .. 240 Vac 1.16.027 Display -digit 1.08.014 2-position switch 1.14.002 Potmeter 5 kOhm 1.12.001 10-turn dial 1.13.003
9.17.009 page 11
Loading...