Do not alter the ground connection. Without the protective ground connection, all accessible
conductive parts (including control knobs) can render an electric shock. The power jack and mating
plug of the power cable meet IEC safety standards.
To avoid electrical shock hazard, disconnect power cord before removing covers. Refer servicing to
qualified personnel.
Before connecting the line cord to the AC mains, check the rear panel AC line voltage indicator.
Applying a line voltage other than the indicated voltage can destroy the AC line fuses. For continued
fire protection, replace fuses only with those of the specified voltage and current ratings.
Safety Summary
The following safety precautions apply to both operating and maintenance personnel and must be
observed during all phases of operation, service, and repair of this instrument. Before applying
power, follow the installation instructions and become familiar with the operating instructions for this
instrument.
If this device is damaged or something is missing, contact the place of purchase immediately.
This manual contains information and warnings that must be followed to ensure safe operation as
well as maintain the meter in a safe condition.
GROUND THE INSTRUMENT
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical
ground. This instrument is grounded through the ground conductor of the supplied, three-conductor
ac power cable. The power cable must be plugged into an approved three-conductor electrical
outlet. Do not alter the ground connection. Without the protective ground connection, all accessible
conductive parts (including control knobs) can render an electric shock. The power jack and mating
plug of the power cable must meet IEC safety standards.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes. Operation of any
electrical instrument in such an environment constitutes a definite safety hazard.
KEEP AWAY FROM LIVE CIRCUITS
Instrument covers must not be removed by operating personnel. Component replacement and
internal adjustments must be made by qualified maintenance personnel. Disconnect the power
cord before removing the instrument covers and replacing components. Under certain conditions,
even with the power cable removed, dangerous voltages may exist. To avoid injuries, always
disconnect power and discharge circuits before touching them.
DO NOT SERVICE OR ADJUST ALONE
Do not attempt any internal service or adjustment unless another person, capable of rendering first
aid and resuscitation, is present.
DO NOT SUBSTITUTE PARTS OR MODIFY THE INSTRUMENT
Do not install substitute parts or perform any unauthorized modifications to this instrument. Return
the instrument to B&K Precision for service and repair to ensure that safety features are maintained.
WARNINGS AND CAUTIONS
WARNING and CAUTION statements, such as the following examples, denote a hazard and
appear throughout this manual. Follow all instructions contained in these statements.
A WARNING statement calls attention to an operating procedure, practice, or condition, which, if
not followed correctly, could result in injury or death to personnel.
A CAUTION statement calls attention to an operating procedure, practice, or condition, which, if not
followed correctly, could result in damage to or destruction of part or all of the product.
WARNING:
WARNING:
CAUTION:
3
This product uses components which can be damaged by electro-static discharge (ESD). To avoid
damage, be sure to follow proper procedures for handling, storing and transporting parts and
subassemblies which contain ESD
This product is subject to Directive 2002/96/EC of the
Refer to the user manual for warning information to
damage to instrument.
On (Supply). This is the AC mains
instrument.
Off (Supply). This is the AC mains
CAUTION:
-sensitive components.
Compliance Statements
Disposal of Old Electrical & Electronic Equipment (Applicable in the Europea n
Union and other European countries with separate collection syst ems)
European
Parliament and the Council of the European Union on
waste
electrical and electronic equipment (WEEE) , and in
jurisdictions
adopting that Directive, is marked as being put on the
market after August 13, 2005, and shou ld no t be
disposed of as unsorted
municipal waste. Please utilize your local WEEE
collection
facilities in the disposition of this product and
otherwise observe all applicable requirements.
Safety Symbols
avoid hazard or personal injury and prevent
Chassis (earth ground) symbol.
On (Power). This is the In position of the power
switch when instrument is ON.
Off (Power). This is the Out position of the power
switch when instrument is OFF.
connect/disconnect switch at the back of the
connect/disconnect switch at the back of the
instrument.
4
CE Declaration of Conformity
The model 4007B and 4013B meet the requirements of 2006/95/EC Low Voltage
Directive and 2004/108/EC Electromagnet Compatibility Directive with the following
standards.
Low Voltage Directive
- EN61010
Safety requirements for electrical equipment for measurement, control,
and laboratory use.
3.13.1 Selecting a Standard Waveform .......................................................... 16
3.13.2 Setting the Output ................................................................................ 16
3.13.3 Using Voltage Offset ........................................................................... 16
3.13.4 Storing and Recalling a Waveform Generator Setup ........................... 16
5
4007B/4013B – Operating Manual
Model
4007B
4013B
Frequency Characteristics
Sine
0.1 Hz – 7 MHz
0.1 Hz – 12 MHz
Square
0.1 Hz – 7 MHz
0.1 Hz – 12 MHz
Triangle
0.1 Hz – 1 MHz
Accuracy
0.01% ± 0.186 Hz
Resolution
5 digits or 100 mHz
Output Characteristics
Amplitude Range
10 mV to 10 Vp-p (into 50 Ω)
20 mV to 20 Vp-p (open circuit)
Resolution
3 digits (1000 counts)
Amplitude Accuracy
± 2 % ± 20 mV of programmed output from 1.01 V – 10 V
Flatness
± 1 dB to 12 MHz
Offset Range
± 4.5 V (into 50 Ω)*
Offset Resolution
10 mV, 3 digits
Offset Accuracy
± 2 % ± 10 mV (into 50 Ω)
Output Impedance
50 Ω
Output Protection
Protected against short circuit or accidental voltage practically
connector
Waveform Characteristics
1 Introduction
1.1 Introduction
This manual contains information required to operate the B&K Precision model 4007B and
4013B DDS Function Generators. This section covers the instrument’s general description,
specifications and characteristics.
1.2 Description
The 4007B and 4013B are versatile high performance function generators. Implemented
using a DDS (direct digital synthesis) architecture, these instruments generate stable and
precise sine, square, and triangle waveforms. Both models also provide linear and logarithmic
sweep for users needing sweep capability. An auxiliary TTL output at the generator’s
frequency is available to synchronize external devices. Up to 9 fr ont panel settings can be
saved to memory storage.
1.3 Specifications
available in electronic laboratories, applied to the main output
6
Harmonic Distortion
DC – 100 kHz, -55 dBc
1MHz – 7 MHz, -30 dBc
DC – 100 kHz, -55 dBc
1MHz – 12 MHz, -30 dBc
Square Rise/Fall Time
< 20 ns (10% to 90% at full amplitude into 50 Ω)
Variable Duty Cycle
20% to 80 %, up to 1 MHz for Square waveform
Symmetry Accuracy at 50%
< 1 %
Sweep Characteristics
Sweep Shape
Linear or Logarithmic, up or down
Sweep Time
20 ms to 100 s
Input and Output
Sync Out
TTL pulse at programmed freq uency; 50 Ω source impedance
General
Memory Storage
Store up to 9 instrument settings
Power Requirements
100 V – 240 V AC ± 10% (90 V – 264 VAC)
Max. Power Consumption
25 VA
Operating Temperature
0 °C – 50 °C
Storage Temperature
-10 °C – 70 °C
Humidity
95% RH, 0 °C – 30 °C
Dimensions
213 mm x 88 mm x 210 mm (WxHxD)
Weight
Approx. 2 kg
Safety Standards
EN55011 for radiated and conducted emissions
EN61010
100 kHz – 1 MHz, -45 dBc
EN55082
100 kHz – 1 MHz, -45 dBc
*Depends on amplitude setting.
Note: All specifications apply to the unit after a temperature stabilization time of 15 minutes
over an ambient temperature range of 23 °C ± 5 °C. Specifications are subject to change
without notice.
To ensure the most current version of this manual, please download the latest version here:
http://www.bkprecision.com/search/4013B
For current up-to-date product in for mation, please visit www.bkprecision.com
7
4007B/4013B – Operating Manual
2 Installation
2.1 Introduction
This section contains installation information, power requirements, initial inspection and signal
connections for the 4007B and 4013B signal generators.
2.2 Package Contents
Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items
from the shipping carton, and check for any obvious signs of physical damage that may have
occurred during transportation. Report any damage to the shipping agent immediately. Save
the original packing car t on for possi bl e fut ur e res hi pm ent. Every generator is shipped with the
following contents:
• 4007B / 4013B DDS function generator
• AC Power Cord
• Manual contained on CD
• Certificate of Calibration
Verify that all items above are included in the shipping container. If anything is missing,
please contact B&K Precis ion.
2.4 Instrument Mounting
The 4007B and 4013B Function Generators are intended for bench use. The instrument
includes a front feet tilt mechanism for optimum panel viewing angle. The instrument does not
require special cooling when operated within conventional temperature limits. It may be
installed in a closed rack or test station if proper air flow can assure removing about 15 W of
power dissipation.
2.5 Power Requirements
The 4007B and 4013B can be operated from any source of 90V to 264V AC, frequency from
48Hz to 66Hz. The maximum power consumption is 25 VA. Use a slow blow fuse of 1A,
UL/CSA approved as indicated on the rear panel of the instrument.
The instrument power fuse is located in the AC input plug. To access the fuse, first
disconnect the power cord and then remove the fuse cartridge.
8
Fuse box slit
Fuse box
Check/Remove Fuse
2.6 Fuse Replacement
There is a 1A, 250V rated slow blow fuse at the AC input. Should the fuse ever get blown,
follow the steps below to replace:
1. Locate the fuse box next to the AC input connector in the rear panel.
2. With a small flat blade screwdriver, insert into the fuse box slit to pull and slide out the fuse box
as indicated below.
3. Check and replace fuse if necessary.
2.7 Grounding Require me nts
For the safety of operating personnel, the instrument must be grounded. The central pin on
the AC plug grounds the instrument when properly connected to the ground wire and plugged
into proper receptacle.
WARNING
TO AVOID PERSONAL INJURY DUE TO SHOCK, THE THIRD WIRE EARTH
GROUND MUST BE CONTINUOUS TO THE POWER OUTLET. BEF ORE
CONNECTION TO THE POWER OUTLET, EXAMINE ALL CABLES AND
CONNECTIONS BETWEEN THE UNIT AND THE FACILITY POWER FOR A
CONTINUOUS EARTH GROUND PATH.
THE POWER CABLE MUST MEET IEC SAFETY STANDARDS.
2.8 Signal Connections
Use RG58U 50Ω or equivalent coaxial cables for all input and output signals to and from the
instrument.
9
4007B/4013B – Operating Manual
3 Operating Instructions
3.1 General Description
This section describes the displays, controls and connectors of the function generator.
All controls for the instrument local operation are located on the front panel.
1 Power Button Power ON/OFF
2 LCD Display
3 Function Keys F1 – F4 function keys
4 Waveform Buttons Select Sine, Ramp/Triangle, Square waveform shape
5 Rotary Knob
6 Cursor Keys
7 Output ON/OFF Enable/Disable Output
8 SWEEP Button Selects Sweep functio n menu
9 UTIL Button Utility menu
10 Units Keys
11 Numeric Keypad
10
3.2 Display Window
The function generator has a graphic LCD display that can display up to 124 x 32 dots. When you poweron the unit sine waveform is selected and current settings appear in the display. The bottom displays a
menu (selectable with function keys) that corresponds to the function, parameter or mode display selected.
3.3 Front Panel Controls
The front-panel controls select, display, and change parameter, function, and mode settings.
Use the numerical keypad, rotary input knob and the cursor movement keys to enter data into the
waveform generator.
To change a setting:
1. Press the key that leads to the item to change.
2. Move cursor using cursor keys to the appropriate position in the numeric field (if applicable).
3. Use the rotary input or the numerical keyboard to change the value of the displayed item. Changes take
effect immediately.
3.4 Connectors
The function generator has two BNC connectors on the front panel where you can connect coaxial cables.
These coaxial cables serve as carrier lines for output signals delivered from the function generator.
Output
Use this connector to transfer the main output signal from the function generator.
SyncOut
Use this connector to output a positive TTL sync pulse generated at each waveform cycle.
3.5 Output Connections
The waveform generator output circuits operate as a 50 Ω voltage source working into a 50 Ω
load. At higher frequencies, non terminated or improperly terminated output causes
aberrations on the output waveform. In addition, loads less than 50 Ω reduce the waveform
amplitude, while loads more than 50 Ω increa s e w av eform am pl it ud e.
Excessive distortion or aberrations caused by improper termination are less noticeable at
lower frequencies, especially with sine and triangle waveforms. To ensure waveform integrity,
follow these precautions:
1. Use good quality 50 Ω coaxial cable and connectors.
2. Make all connections tight and as short as possible.
3. Use good quality attenuators if it is necessary to reduce waveform amplitudes applied to
sensitive circuits.
11
4007B/4013B – Operating Manual
4. Use termination or impedance-matching devices to avoid reflections.
5. Ensure that attenuators and terminations have adequate power handling capabilities.
If there is a DC voltage across the output load, use a coupling capacitor in series with the
load. The time constant of the coupling capacitor and load must be long enough to maintain
pulse flatness.
ImpedanceMatching
If the waveform generator is driving a high impedance, such as a 1 MΩ input impedance
(paralleled by a stated capacitance) of an oscilloscope vertical input, connect the transmission
line to a 50 Ω attenuator, a 50 Ω termination and t o the osc i ll os cop e input. The attenuator
isolates the input capacitance of the device and terminates the waveform generator properly.
3.6 MENU Keys
These keys select the main menus for displaying or changing a parameter, function or mode.
3.6.1 WAVEFORM Keys
The keys select the waveform output and displays the waveform parameter menu (frequency,
amplitude and offset).
Sine Menu
F1: FREQ – (Frequency) Selects and displays the frequency. Change the frequency setting
using the cursor keys, rotary knob or numerical keys.
F2 : AMPL – (Amplitude) Selects and displays the Amplitude. Change the amplitude setting
using the cursor keys, rotary knob or numerical keys.
F3:OFST – (Offset) Selects and displays the Offset parameter. Change the offset by using
the cursor keys, rotary dial or numerical keys. If a certain setting cannot be produced, the
waveform generator will display a “Setting Conflict” message.
Amplitude and Offset Interaction:
Amplitude and offset settings interact and are bound by hardware restrictions. In order to
obtain the desired waveform, the following amplitude and of fset hardware limitations must be
considered:
12
Output Voltage
Range
Constraints of Amplitude +
Offset
(Vp-p)/2 + |offset| <= 5 volts
(Vp-p)/2 + |offset| <= 0.5 volts
(Vp-p)/2 + |offset| <= 0.05 volts
The offset voltage has three ranges as follows:
1.01 volt to 10.00 volts
0.101 volt to 1 volt
0.010 volt to 0.100 volt
F4:SYM – (Symmetry) When the Square or Triangle waveforms are selected, the
SYMMETRY(Duty cycle) parameter is available. Change the symmetry(Triangle) or duty
cycle(square) by using the cursor keys, rotary dial or numerical keys. If a certain setting
cannot be produced, the waveform generator will display a warning message.
SquareMenu
3.6.2 UTILITY Key
Utility Menu
F1: RECALL – Recalls a previously stored front-panel setup from the selected buffer. Change
the buffer number by using the rotary input knob. Valid storage buffer numbers are from 1 to
9.
Buffer 0 is the factory default setup.
F2:STORE – Stores the current front-panel setup to the specified storage buffer. Change the
buffer number by using the data keys or the rotary input knob. Valid storage buffer numbers
range from 1 to 9.
F3: OUT-ON – Selects the OUTPUT state from a power-up. Select ON to enable or OFF to
disable the output from a pow er-up.
13
4007B/4013B – Operating Manual
F4: INTENSITY – Adjusts the intensity of the LCD screen. Valid range is 1 – 9; 9 giving th e
highest intensity and contrast.
3.6.3 SWEEP Key
Selects the Sweep Mode and allows entering sweep parameters Sweep Start, Sweep Stop
and Sweep Rate.
To select the sweep mode, press SWEEP, then press the function key that corresponds to
the desired Sweep menu option , as show n:
Sweep Menu
F1: ON/OFF – Enables/disables the sweep function. Select between Sweep ON or OFF.
F2: LIN – Selects the Linear sweep shape.
F3: LOG – Selects the Logarithmic sweep shape.
F4: SET – Defines the Sweep Start and Stop frequencies.
Set Sweep Menu
3.7 ON Key
Use this key to control the main output signal. When the output is active, the key will lit. By
default, this will be ON (enabled) from a power-up. This can be changed by changing the
OUT-ON settings from the utility menu.
3.8 Cursor Keys
Use these keys to move the cursor (when visible) either left or right. They are used in
conjunction with the rotary input knob to set the step size of the rotary input knob.
3.9 Rotary Input Knob
14
Use this knob to increase and decrease numeric values. The cursor indicates the low-order
Key Function
Value
Description
Function
Sine
Output Waveform
Frequency
1.0000 kHz
Waveform Frequency
Amplitude
5.00 V
Peak-to-peak output amplitude
Offset
0.00 V
Zero DC offset
Output
OFF
Output disabled
Sweep
OFF
Sweep function disabled
Message Text
Description
Out of range
The set value is out of the instrument’s limits
Setting Conflict
Settings conflict with another parameter setting or value.
Empty location
Attempt to recall settings from an empty storage location.
position of the displayed value which changes when you rotate the knob (for straight numeric
entries only). For other types of data, the whole value changes when you rotate the knob.
3.10 Power-On Settings
Table 3-2 lists the factory default settings or selected after RECALL 0.
Table 1 – Power-on Default Settings
3.11 Memory
The waveform generator uses non-volatile flash memory for storing the front panel settings.
Up to 9 front panel settings can be stored (not including default settings).
3.12 Displaying Errors
The waveform generator displays error messages when front-panel setti ng s are either inv alid
or may produce unexpected results.
Table 2 – Error Messages
3.13 Quick Start
This section explains how to generate various waveforms and modify the output waveform.
* Generating a waveform output
* Modifying waveform output
* Storing and recalling a waveform generator setup
15
4007B/4013B – Operating Manual
3.13.1 Selecting a Standard Waveform
You can select several standard waveforms as: sine, triangle, square. Creating a standard
waveform requires selecting the waveform type, parameters and their settings that define the
waveform.
Generating a standard waveform requires the following:
* Selecting the waveform
* Setting the output frequency
* Setting the output amplitude and offset
3.13.2 Setting the Output
To set the output channel, press the Output ON key. The key will lit indicating the output is
enabled.
3.13.3 Using Voltage Offset
Through the offset parameter you can add a positive or negative DC level to the output
waveform.
To set voltage offset:
1. Select a waveform to displ ay its menu.
2. Press F3 :OFST to display the offset setting.
3. Use the rotary input knob or the numerical keys to set the voltage offset.
To turn the voltage offset OFF, repeat the steps above, but set the offset voltage level to 0.
3.13.4 Storing and Recalling a Waveform Generator Setup
You can store up to 9 front-panel setups inside the internal non-volatile flas h me mory. When
you recall a stored setup, the front-panel settings change to match the settings in the stored
setup.
Storing Setups
To store the front-panel setup:
1. Press UTILITY to display the menu.
2. Press F2:STORE to select the Store mode.
3. Use the rotary input knob to select a buffer number. Vali d buffer numbers r ange from 1 to 9.
Buffer 0 is a read-only buffer that contains the default setup/power-on settings.
The function generator will overwrite and store settings into a buffer that had settings
previously stored inside without a warning.
16
Recalling Setups
To recall stored fro nt-panel setup:
1. Press UTILITY to display the menu.
2. Press F1:RECALL to select the Recall mode.
3. Use the rotary input knob to select a buffer number. Valid buffers numbers range from 0 to 9.
Buffer 0 is a read-only buffer that contains the default power-on setup.
17
SERVICE INFORMATION
Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #. Return the product
in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any
leads, probes, conn ectors and accessories that you are using with the device.
Non-Warranty Service: Please go the support and service s ection on our website www.bkprecision.com to obtain a RMA #. Return the
product in the original packaging to the address below. Clearly state on the RMA the performance problem and return any leads, probes,
connectors and accessories that you are using with the device. Customers not on an open account must include payment in the form of a
money order or credit card. For the most current repair charges please refer to the service and support section on our website.
Return all merchandise to B&K Precision Corp. with pre-paid shipping. The flat-rate repair charge for Non-Warranty Service
does not include return shipping. Return shipping to locations in North America is included for Warranty Service. For overnight
shipments and non-North American shipping fees please contact B&K Precision Corp.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
www.bkprecision.com
714-921-9095
Include with the returned instrument your complete return shipping address, contact name, phone number and
description of problem.
LIMITED THREE-YEAR WARRANTY
B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from
defects in workmanship and materials for a period of three years from date of purchase.
B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned
product must be accompanied by proof of the purchase date in the form of a sales receipt.
To help us better serve you, please complete the warranty registration for your new instrument via our website
www.bkprecision.com
Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of
unauthorized alterations or repairs. The warranty is void if the serial number is altered, defaced or removed.
B&K Precision Corp. shall not be liable for any consequential damages, including without limitation damages resulting from
loss of use. Some states do not allow limitations of incidental or consequential damages. So the above limitation or exclusion
may not apply to you.
This warranty gives you specific rights and you may have other rights, which vary from state-to-state.