The following safety precautions apply to both operating and maintenance personnel and must be followed during all
phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
• Read and understand the safety and operational information in this manual.
• Apply all the listed safety precautions.
• Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument
at an incorrect line voltage will void the warranty.
• Make all connections to the instrument before applying power.
• Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design,
manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply
with these requirements.
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy available and the
voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is
a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested,
which denes the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the
mains supply. The voltages in the environment are typically derived from a limited-energy transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains
supply at a standard wall outlet or similar sources. Example measurement environments are portable
tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains
installation of a building. Examples are measurements inside a building’s circuit breaker panel
or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary
power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specied in this manual
for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the
instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating
of the measuring system.
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be
115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your
country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This
instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power
cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power
cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts
(including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the
recommended three-conductor AC line power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of potential only
and is not to be used as a safety ground. Do not operate in an explosive or ammable atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
• In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided particulates.
• In relative humidity conditions outside the instrument’s specications.
• In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense
on the instrument.
• In air temperatures exceeding the specied operating temperatures.
• In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.
• In environments with restricted cooling air ow, even if the air temperatures are within specications.
• In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is
0∘C to 40∘C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by this instrument
may be outside specications if the instrument is used in non-oce-type environments. Such environments may include
rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise,
strong electric elds, or strong magnetic elds.
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the
instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated,
and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the
instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases,
or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not
for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support
device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must
be made by qualied service-trained maintenance personnel who are aware of the hazards involved when the instrument’s
covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may
exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect
all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there
are no hazardous voltages present on any conductors by measurements with a properly-operating voltage-sensing device
before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the
measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt
any service or adjustment unless another person capable of rendering rst aid and resuscitation is present.
Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit
exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of the instrument’s fuse
requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses.
Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specied in this manual or on
the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a re. Failure
to use the specied fuses will void the warranty.
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to
B&K Precision for service and repair to ensure that safety and performance features are maintained.
For continued safe use of the instrument
• Do not place heavy objects on the instrument.
• Do not obstruct cooling air ow to the instrument.
• Do not place a hot soldering iron on the instrument.
• Do not pull the instrument with the power cord, connected probe, or connected test lead.
• Do not move the instrument when a probe is connected to a circuit being tested.
Safety Symbols
SymbolDescription
indicates a hazardous situation which, if not avoided, will result in death or serious injury.
indicates a hazardous situation which, if not avoided, could result in death or serious injury
indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
This is the In position of the power switch when instrument is ON.
This is the Out position of the power switch when instrument is OFF.
is used to address practices not related to physical injury.
4. Do not install unit near any heat source or heating emitting devices.
5. Prevent the power cord from being walked on or pinched.
6. Unplug this unit during lightning storms or when unused for long periods of time.
1.1 Precautions
1. The unit must be used within its specied range.
The rated input voltages can be found on the rating label at the back of the unit. Before plugging into the AC supply,
check with the rating label.
2. This unit has a built-in Tracking O.V.P. (Over Voltage Protection) feature. In the event of the output voltage
becoming 10% greater than the set value, the O.V.P. will be triggered and the output power will be cut o and
>FAULT< warning will appear.
When you get this warning, switch o the unit and remove all loading.
If you switch the unit back on again it should resume normal operation. In the event that this problem persists contact
the manufacturer.
3. This unit has a built-in buzzer, which will sound when Over Temperature/Overload/Over Voltage Protection has been
triggered. When you get this warning, switch the unit o and remove all loading.
Check your load and output voltage setting.
Allow the unit to cool down for 30 minutes.
If you switch the unit back on it should resume normal operation.
In the event this problem persists contact B&K Precision for assistance.
4. Only use the supplied software and optional accessories with this unit.
5. Refer all servicing to manufacturer.
Warning! The maximum output voltage for Model 1698B is up to 60Vdc. It may be hazardous to touch metal part
of the terminals.
2. Presson the numeric keypad within 3 seconds to enter upper voltage limit adjustment menu. The rst line
says “Over” and the second if the voltage limit setting.
.
3. Use the numeric keypad
4. Pressto conrm.
toto input upper voltage limit.
4.7 Enable Output at Power Up
This feature limits the upper level setting of output voltage to prevent inadvertent setting of high voltage, which may
damage your application. The value of this upper voltage range limit will be retained until further reset.
1. Press.
2. Press the up arrow keywithin 3 seconds to enable output on power up.
4.8 Disable Output at Power Up
1. Press.
2. Press the down arrow key
within 3 seconds to disable output on power up.
4.9 LCD Brightness
1. Press.
2. Presson numeric keypad within 3 seconds to enter brightness menu.
3. Use the rotary knobto raise or lower brightness from 0 to 9.
2. Use the numeric keypadthruto select a preset voltage and current value.
3. Useandto adjust preset voltage and current values.
4. Presswithin 3 seconds to conrm or pressto exit program setting.
5.2 Setting Timed Program
The unit can be programmed to operate up to 20 timed subprograms (0-19 STEP as shown in the display). Each
subprogram is capable of running a preset operation period of 1 second to 99 minutes and 99 seconds with its own preset
voltage and current level. The timed subprogram can be set to run in sequence repeatedly from 1 to 9999 cycles or
innity run. You can run the unit through the sequence of subprograms for the input cycles number unless interrupted
by pressing the CLEAR key.
1. Press
2. Useto select a step to be modied.
3. Use the upand downkeys to navigate between setting voltage, current, and time
4. Press Enterto conrm and exit.
then presson the numeric keypad to enter program setting menu.
5.3 Run Timed Programming
1. Pressthen pressto select step program to run
2. Useto select the number of steps to run (2 to 20) and press.
Ex. Selecting step 3 means run from step 1 to step 3
3. Input the desired number of cycles from 0000 to 9999 using the numeric keypad or rotary knob, pressto
conrm and begin timed program. Leave this setting on 0000 for innite cyclic run. Pressto
terminate the program.
The purpose of recalibration is to reduce the dierence between the set values and the displayed values on the LCD
display. Recalibration is only required when this dierence is greater than 0.1 V for voltage or -0.01 A / +0.02 A for
current.
6.2 Troubleshooting
Keypad and Dial do not work Check key lock symbol for lock state, unlock unit by pressing
thenkey, otherwise switch the unit o and then back on to see if the problem persists.
No output power Check the output for on/o symbol on the display. Press
Cannot get a high voltage
setting within the
rated maximum
“OUT OF RANGE”
keeps appearing
When keying in operations
the unit keeps exiting and
does not save inputs
then.
Check the upper voltage limit setting by pressingthen. Adjust the
maximum voltage limit with the
Check if the setting is within the rated range. If this occurs during voltage setting, refer to troubleshooting solution to the previous problem.
Only 10 seconds are allowed for inputting and 3 seconds for operation mode settings.
The PC software provides remote communication, data logging, front panel emulation, and timed programming capabilities
using the USB or RS-485 interfaces. The software supports Windows 7, 8, 8.1, 10.
Note: Do not connect both USB and RS-485 simultaneously.
7.1 USB Driver Installation
Figure 7.1 Download Link
1. Click “CP210X_USB_Driver.zip” to begin the download
2. Whem complete, right click on the .zip folder and click “Extract All…”
3. Depending on your system right-click and run the 32-bit or 64-bit installer as administrator to begin installation, see
5. Select a destination location for the software. Figure 7.5
6. Select a location for the software shortcut.Figure 7.6
7. Check the box if you would like to create a desktop shortcut and click “Next” to continue. Figure 7.7
8. Click “Install” to begin installation.Figure 7.8
7.3 Multi-Unit Control
7.3.1 Connect Multiple Power Supplies to PC with RS-485
Use the RS-485 interface at the back of the power supply to daisy-chain up to 31 power supplies, see Figure 7.9. The
USB to RS-485 adapter shown in Figure 7.10 below is required.
7.4 Software Setup and Conguration
Connect the instrument to the PC with the supplied USB cable and power on the instrument. Ensure the drivers are
installed and the device is connected. To install the drivers, refer to Section 7.1 above for driver installation instructions.
Also, the software uses the “SCPI” protocol, this must be enabled, see Section 7.4.2. With the instrument connected,
open the PC software. Figure 7.11 shows the screen to expect when the software starts. If there is already a saved
connection the software will automatically connect to the instrument.
Figure 7.9 Connection diagram for multiple power supply control
Figure 7.10 Connection diagram between USB adapter and RS-485 connectors
• Type in a connection name, select USB from the drop down, and select the correct COMM port assigned by your PC.
Check the windows the device manager to determine the correct COMM port. In this case Windows has assigned
COM port 12. Click “Save” and the black display will light up to indicate a successful connection. See Figure 7.15
• If using RS-485, enter the three digit ID displayed on the RS-485 setting on the instrument.
− To set the RS-485 ID on the instrument press
or numeric keypad entry to edit ID. Pressto save and exit.
If everything has gone well and the power supply is connected the software should look similar to Figure 7.16.
Display Panel
thenand use theand
7.4.2 Enable SCPI protocol
Pressthento enter SCPI setting. Useto move the selector down next to the
“n” and pressto save and exit. See Figure 7.17
7.4.3 Internal Timed Program
Congure up to 20 user dened voltage, current, and duration steps save to the instruments internal memory, or read
parameters already stored in the internal memory for edit.
1. Click on the “Internal Timed Program” tab
2. Double-click on the desired cell and use the slider to set voltage and current values.
3. Set the number of cycles from 1 to 9999 and clickto start, or clickto save the program into the
instruments internal memory for future recall.
External Timed Program is completely controlled by the PC. The PC counts the step time and changes the specic
voltage and current levels of the power supply. Congure up to 20 user dened voltage, current, and duration steps.
1. Click on the “External Timed Program” tab. See Figure 7.20
2. Double-click on the desired cell and use the slider to set voltage and current values. Figure 7.21
3. Set step duration with the up/down buttons. The time can be set from 0 up to 9 hours 59 minutes and 59 seconds.
If time set left at 0 the step will be skipped. See Figure 7.22
4. Set the number of cycles from 0-999. Leave this setting at 0 for continuous loop.
5. Click
to start running the cycle.
7.4.5 Internal Preset Memory
Preset up to 9 voltage and current combinations into memory for quick output. Set voltage and current with the slider
Data Log allows can be used to view present or stored output data, see Figure 7.24. Data can be saved and exported
as .CSV spreadsheet le or sent to a printer.
7.4.7 Data Log Sampling Time and Voltage Upper Limit (UVL) Setting
Under the “Setting” tab:
• Adjust the data log sampling time. 1S means the voltage, current, and power reading will be recorded every 1 second.
• Set the Voltage Upper Limit (UVL) and general output setting and timed programming setting will not exceed this