Information and descriptions contained in this manual are the property of Bistos Co. , Ltd. and may not be copied,
reproduced, disseminated, or distributed without express written permissio n from Bistos Co., Ltd
Informatio n furnished by Bist os Co., Ltd is believed to be accurate and reliable . However, no responsibility is
assumed by Bistos for its use, or any infringeme nts of pate nts or o ther right s of third parties tha t may re sult fro m
its use. No license is granted by implication or otherwise under any patent or patent rights of Bistos.
1.1 Instructions for the safe operation and use of the device
1.2 Warnings
1.3 Shock hazard
1.4 General precautions on environment
1.5 Symbols
2 Introduction
2.1 General
2.2 Brief device description
2.3 Intended use
2.4 Operating principles
2.5 Essential performance
2.6 Accessaries and options
2.7 Appearance of device
2.7.1 Front view
2.7.2 Front view detail
2.7.3 Rear view
2.7.4 Side view
2.8 Description of each part
2.8.1 Control shell
2.8.2 Hood
2.8.3 Mattress tray
2.8.4 Stand
3 Install and connection
3.1 IV pole assembly
3.1.1 IV external monitor
3.1.2 IV plate
3.1.3 IV ringer pole
3.2 Air filter assembly
3.3 Connection of power and cable
3.3.1 Power connection
3.3.2 Cable connection
3.3.3 Sensor module connection
3.4 Placement of infant
3.5 Movement and installation
4 Operation
4.1 System start-up
4.2 LED
4.3 Key and knob operation
4.3.1 Key
4.3.2 Knob
4.4 Displays
BT-500 Operation Manual 2
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
4.5 Temperature measurement and control
4.6 Humidity Measurement and control
4.7 O
measurement
2
4.8 Weighing scale measurement
4.9 SpO
and Pulse rate measurement
2
4.10 Menu function
4.11 Pulse oximeter
4.12 External monitor
4.13 Shut down
4.14 Calibration of O
module
2
5 Alarms
5.1 System alarms
5.2 Temperature alarms
5.3 Humidity alarms
5.4 Oxygen alarms
5.5 Weighing scale alarms
5.6 Alarm self-test
6 Cleaning & Maintenance
6.1 General cleaning method and precautions
6.2 Hood
6.3 Shell, sensor module, scale module and basket
6.4 Water tank
6.5 Skin temperature sensor and SpO
sensor
2
6.6 Drain of residual water
6.7 Regular inspection
6.8 Battery replacement and disposal
6.9 Disposal of the device
7 Specification
8 Troubleshooting
8.1 General checking
8.2 Alarm message checking
9 Declaration on EMC
9.1 Electromagnetic emissions
9.2 Recomended separation distances between portable and mobile RF
communications equipment and the device
9.3 Electromagnetic immunity
Warranty
BT-500 Operation Manual 3
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
WARNING
Be informed that it may cause serious injury or death to the patient,
property damage, material losses against the “WARNING” sign.
CAUTION
Be informed that it may cause no harm in life but lead to injury
against the “CAUTION” sign.
Be informed that it may cause serious electrical shock to the patient
HAZARD” sign.
WARNING
Thoroughly read and understand the manual prior to use of the incubator.
ubator misuse may result in harm to an infant. Only properly trained
d approved for use in this
The total electrical current leakage of all items powered through the
120V AC/ 100V AC systems and less than 500uA for 230V AC systems.
1Safety
1.1 Instructions for the Safe Operation and Use of the device
Examine the incubator and any accessories periodically to ensure that the cables, line
cords and instruments do not have visible evidence of damage that may affect patient
safety or performance. The recommended inspection interval is once per week or less.
Do not use the incubator if there is any visible sign of damage.
Only the AC line cord supplied with the BT-500 is approved for use with the Unit.
Do not attempt to service the BT-500 incubator. Only qualified service personnel by
Bistos Co., Ltd. should attempt any needed internal servicing.
The BT-500 is not specified or intended for operation during the use of defibrillators or
during defibrillator discharge.
The BT-500 is not specified or intended for operation in the presence of electrosurgical
equipment.
The BT-500 is not specified or intended for operation in conjunction with any other type
of equipment except the specific devices that have been identified for use in this
Operator’s Manual.
Perform periodic safety testing to insure proper patient safety. This should include
leakage current measurement and insulation testing. The recommended testing interval
is once per year.
Do not operate the BT-500 incubator if it fails to pass the power on self-test procedure.
SHOCK HAZARD
1.2 Warnings
or operator, property damage, material losses against the “SHOCK
Failure to do so could result in personal injury or equipment damage.)
Inc
personnel should use the incubator as directed by an appropriately qualified
attending physician aware of currently known risks and benefits.
Use of accessories other than those listed an
product as original or replacement items may result in increased emissions
or decreased immunity.
incubator, including devices on the outlet strip, must be less than 300uA for
BT-500 Operation Manual 4
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Otherwise, personal injury or equipment damage could occur.
The use of accessory equipment not complying with the equivalent safety
resulting system. Consider the use of the accessory in the patient’s vicinity
and evidence that the safety certifications of the accessory have been
Devices connecting to the serial data port must be compliant with EN
2, the EMC requirement for Medical Devices. Failure to do so could
Medical electrical equipment needs special precautions regarding EMC and
provided in this manual. In addition, portable and mobile RF
temperature. Routinely monitor the infant’s rectal and/or axillary
temperature according to the attending physician’s orders or Nursery
decrease an infant’s
evaporative water loss, and may cause an increase in infant temperature.
weight, premature infants. The
attending physician should prescribe Temperature Control mode,
midity output level setting. Routinely monitor
the infant’s rectal and/or axillary temperature according to the attending
ling Limit line. Do not overfill. Water
incubator, use only skin temperature probes
from Bistos Co. Ltd. Using other probes could result in personal injury or
Never place the skin temperature probe under the infant or use it rectally.
requirements of this equipment may lead to a reduced level of safety of the
performed in accordance with the appropriate International Electrotechnical
commission (IEC) 60601-1 harmonized national standard. Personal injury or
equipment damage could occur.
60601-1result in personal injury or equipment damage.
needs to be installed and put into service according to the EMC information
communications equipment can effect medical electrical equipment.
The equipment shall not be used adjacent to or stack with other devices
unless verification of normal operation in the configuration in which it is to
be used can be achieved.
Use only Bistos recommended fuel cells for proper operation. Failure to do
so could result in personal injury or equipment damage.
Higher incubator relative humidity at any given temperature decreases an
infant’s evaporative heat loss, and may cause an increase in the infant
Standing Orders. Failure to do so could result in personal injury.
Higher relative humidity will, at any given time,
This effect is greatest n very low birth-
temperature setting, and hu
physician’s orders or Nursery Standing Orders. Failure to do so could result
in personal injury.
Fill the reservoir to the Maximum Fi
spillage may result, and personal injury could occur.
Use distilled water only (<10 ppm total dissolved solids). The use of sterile
water is not acceptable. Equipment damage could occur.
For proper operation of the
equipment damage.
BT-500 Operation Manual 5
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Personal injury could occur.
differentiate between an increase in core temperature with a
with the skin to provide accurate monitoring of the infant’s skin
temperature. When in skin mode, failure to maintain direct skin contact can
result in overheating. Routinely check the infant’s condition for correct
Do not use in the presence of flammable anesthetics. Personal injury or
enriched environment. Make sure that oxygen
when performing cleaning and maintenance procedures. Turn off or
If it is necessary to administer oxygen in an emergency, notify the attending
Administration of oxygen may increase the noise level for the infant within
entrations to verify delivery of the prescribed
oxygen concentration. Failure to do so could result in personal injury or
BT-500 cannot
cold skin (fever) and a low core and skin temperature (hypothermia). The
temperature of the infant is monitored separately.
When in skin mode, the skin temperature probe must be in direct contact
sensor attachment, and feel the infant’s skin for signs of overheating.
When an x-ray is taken through the hood, the hood could show up on the x-
ray as a radiolucent shadow and could result in incorrect diagnosis.
equipment damage could occur.
Keep matches, and all other sources of ignition, out of the room in which the
incubator is located. Textiles, oils, and other combustibles are easily ignited
and burn with great intensity in air enriched with oxygen. Personal injury or
equipment damage could occur.
Small quantities of flammable agents, such as ethyls and alcohol, left in the
incubator may cause a fire in connection with oxygen. Personal injury or
equipment damage could occur.
A fire and explosion hazard exists when performing cleaning or maintenance
procedures in an oxygensupply is turned Off and the oxygen hose to the incubator is disconnected
disconnect oxygen supplies during periods of non-use. Failure to do so could
result in personal injury or equipment damage.
physician immediately. Failure to do so could result in personal injury or
equipment damage.
Improper use of supplemental oxygen may be associated with serious side
effects including blindness, brain damage, and death. The risks vary with
each infant. The qualified attending physician should prescribe the method,
the concentration, and the duration of oxygen administration.
the infant incubator.
An oxygen analyzer shall be used separately when oxygen is delivered to the
infant.
Measure the oxygen conc
equipment damage.
BT-500 Operation Manual 6
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
If the patient’s arterial oxygen levels cannot be maintained when the oxygen
The oxygen concentration inspired by an infant does not accurately
that produces sparks in an incubator. Personal injury or equipment
Inspect gas/oxygen service components at regular service intervals for signs
lter could affect performance or cause carbon
commensurate with local conditions. Particularly, if the unit is used in an
y be
necessary. Failure to do so could result in infant injury or equipment
After each change of oxygen flow, allow at least 30 min to achieve new
projectiles if the gas is released rapidly due to damage or other causes.
are properly
internal relative humidity. Personal injury or equipment damage could
The use of infant seats, or other accessories within the incubator that can
control setting is set to maximum, the attending physician should prescribe
alternate means of oxygenation. Failure to do so could result in personal
injury or equipment damage.
determine the partial pressure of oxygen(pO2) in the blood. When deemed
advisable by the attending physician, measure blood pO
techniques. Failure to do so could result in personal injury or equipment
damage.
Disconnect the incubator from the hospital oxygen source when oxygen is
not in use. Failure to do so could result in personal injury or equipment
damage.
As oxygen use increases the danger of fire, do not place auxiliary equipment
damage could occur.
Use of anesthetic agents can interfere with oxygen analyzer accuracy.
of corrosion or damage. Failure to do so could result in personal injury or
equipment damage.
A dirty air intake micro fi
dioxide(CO2) build-up. Ensure that the filter is checked on a routine basis
by accepted clinical
2
unusually dusty environment, more frequent replacements ma
damage.
concentrations. Failure to do so could result in personal injury or equipment
damage.
Compressed gas cylinders, such as oxygen cylinders, can become hazardous
Securely fasten the cylinder. Failure to do so could result in personal injury
or equipment damage.
Oxygen levels within the incubator hood environment may be affected when
the access doors or access panels are opened. Make sure all hood access
door gaskets and tubing ports are properly installed. Any open gaps in the
incubator hood may reduce the incubator’s internal oxygen. Personal injury
could occur.
Make sure all hood access door gaskets and tubing ports
installed. Any open gaps in the incubator hood will reduce the incubator’s
occur.
BT-500 Operation Manual 7
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
alter the airflow pattern, may affect temperature uniformity, temperature
variability, the correlation of the incubator temperature reading to center
temperature, incubator hood temperature, and infant skin temperature.
ps placed over the top of the incubator hood may
interfere with upward travel of the vertical height adjustable stand. To
prevent this interference, always remove the phototherapy lamp prior to
stuffed animals, during clinical usage, patient safety and incubator
and ventilator tubing to the mattress with sufficient excess length to allow
Do not raise the hood at any time while the infant is in the incubator. Gain
access to the infant by the access panels and access doors. Failure to do so
tional rear) is open, the temperature
bator and patient together.
When moving the incubator within the same floor space, check that the
patient is secured safely in the unit and either remove or secure all loose
system components to prevent possible patient injury or equipment
ove involves varying floor heights or a complete floor level
change (i.e. thresholds, ramps, elevators), remove all items either not being
used or not necessary for the move, lower the VHA, IV poles and shelves to
mattress temperature and infant skin temperature. Personal injury could
occur.
Phototherapy units located too close to the incubator may affect hood wall
Personal injury of equipment damage could occur.
Phototherapy lam
positioning the stand.
If airflow passages are not kept clear of obstructions, such as blankets and
performance may be compromised.
To avoid overheating the infant due to direct radiation, do not position the
incubator in direct sunlight or under other sources of radiant heat.
Do not place surgical covers or blankets over the infant and warm air curtain
of side vents simultaneously. This may cause heat-induced injury and burns.
To prevent accidental disconnection, secure all patient leads, infusion lines,
for the full range of mattress height adjustment.
Only connect equipment to the serial port that complies with the relevant
IEC standard; and use data cables with plastic body connectors.
could result in personal injury or equipment damage.
When the front access panel (or op
display may not accurately reflect the incubator temperature. Do not leave
the front access panel (or optional rear) open longer than essential. Personal
injury could occur.
Positively secure all access panel latches to avoid accidental opening, Failure
For infant safety, do not leave the infant unattended when the access panels
to do so could result in personal injury or equipment damage.
are open. Personal injury could occur.
Always use two people when moving the incu
damage. If the m
BT-500 Operation Manual 8
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
their lowest position, place all drawers in their locked state, and remove all
Never place objects taller than the top of the wheel casters beneath the
etc., keep at least a 12” (30 cm) perimeter area clear around the vertical
timum incubator stability, always lock all stand wheels, Do not leave
and incubator connections must also be checked before adjusting the
while installing or removing medical gas tanks from the tank holder
Only one monitor shelf should be used per incubator. When using the
monitor shelf, always place the monitor in the center of the shelf, ensure
efore the responsibility of
Pulse rate measurement is based on the optical detection of a peripheral
accessories from the front and rear rail position.
incubator stand. Placement of objects there could interfere with the stability
of the vertical height adjustable stand. Personal injury of equipment damage
could occur.
To avoid possible tip-over or damage to adjacent carts, IV stands, shelves,
height adjustable stand.
For op
the unit unattended when parking on an incline. Failure to do so could result
in personal injury or equipment damage.
When raising or lowering the incubator, the operator should ensure that
both equipment and appendages are clear of the unit’s travel path. Patient
incubator height. Never place any objects on top of the drawer assembly
and always check before lowering the VH that there is sufficient clearance
between the incubator and stand assembly. Do not raise or lower the unit
assembly. Failure to do so could result in personal injury of equipment
damage.
The UART port is for debugging purposes only. It does not allow connections
with other devices.
Prior to placing the infant in the incubator, pre-warm the incubator to the
temperature prescribed by the attending physician, or according to nursing
protocol.
that the monitor fits within the border of the shelf, and avoid stacking
monitors on the shelf. Personal injury or equipment damage could occur.
Attach the incubator to the stand or the vertical height adjustable stand
using the bolts provided. Failure to do so could result in the incubator
separating from the stand if sufficiently tilted, particularly with the hood
open. Personal injury or equipment damage could occur.
This product has been validated with the accessories and options listed in
this manual and found to comply with all relevant safety and performance
requirements applicable to the device. It is ther
that person or organization who makes an unauthorized modification, or
incorporates an unapproved attachment to the device, to ensure that the
system still complies with those requirements.
A pulse oximeter should NOT be used as an apnea monitor
BT-500 Operation Manual 9
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
flow pulse and therefore may not detect certain arrhythmias. The pulse
oximeter to
Do not use the MS board pulse oximeter in the presence of flammable
anesthetics or other flammable substance in combination with air, oxygen
Do not remove the monitor cover except to replace the battery. An operator
may only perform maintenance procedures specifically described in this
e current must not exceed 100 microamperes; measure when an
Do not use Masimo oximetry sensors during MRI scanning as it could
Inaccurate measurements may be caused by significant levels of
Inaccurate measurements may be caused by intravascular dyes such as
Inaccurate measurements or loss of pulse signal may be caused by
ement of a sensor on an extremity with a blood pressure cuff, arterial
Loss of pulse signal can occur when there is arterial occlusion proximal to
n occur when the patient is in cardiac arrest or is in
oximeter should not be used as a replacement or substitute for ECG based
arrhythmia analysis.
A pulse oximeter is an early warning device. Use lab co-
completely understand the patient’s condition.
enriched environments, or nitrous oxide.
manual. Refer servicing to Masimo in repair of this equipment.
Leakag
external device is connected to the serial port.
potentially cause burns.
Inaccurate measurements may be caused by incorrect application or use.
dysfunctional hemoglobin (HbCO or MetHb).
indocyanine green or methylene blue.
Inaccurate measurements or loss of pulse signal may be caused by excessive
illumination.
Inaccurate measurements may be caused by excessive patient movement.
Inaccurate measurements may be caused by venous pulsation.
plac
catheter or intravascular line.
The MS board pulse oximeter can be used during defibrillation, but the
readings may be inaccurate for a short time.
Loss of pulse signal can occur when the sensor is too tight.
Loss of pulse signal can occur when the patient has hypotension, severe
vasoconstriction, severe anemia, or hypothermia.
the sensor.
Loss of pulse signal ca
shock.
BT-500 Operation Manual 10
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Use only Masimo sensors for SpO2 measurements.
Tissue damage can occur due to incorrect placement of sensor.
SHOCK
Unplug the unit from its power source prior to cleaning or maintenance. For
units equipped with an uninterruptible power supply(UPS) system, also
remove the battery pack prior to cleaning or maintenance. Failure to do so
al cleaning agents may be conductive and leave a residue that
Do not expose the unit to excessive moisture that would allow for liquid
Batteries can present a risk of electric shock. The following precautions
The total power of all equipment connected to the convenience outlet strip
on the pedestal / stand must me within the electrical requirements shown
on the rear of the pedestal / stand. Otherwise, personal injury or equipment
Make sure the Building power source is compatible with the electrical
incubator. Failure to do so could result in personal injury or equipment
plug an incubator power cord directly to an AC wall socket when the
incubator is mounted on a pedestal /stand. Always provide power to the
incubator by using the power cord coming directly from the pedestal /stand.
1.3 Shock hazards
HAZARD
could result in personal injury or equipment damage.
Some chemic
may permit a build-up of conductive dust or dirt. Do not allow cleaning
agents to contact electrical components, and do not spray cleaning solutions
onto any of these surfaces. Personal injury or equipment damage could
occur.
To ensure grounding reliability, plug the AC power cord only into a properly
grounded 3-wire hospital-grade or hospital-use outlet. Do not use extension
cords. If any doubt exists as to the grounding connection, do not operate the
equipment. Personal injury or equipment damage could occur.
pooling. Personal injury or equipment damage could occur.
Due to the risk of electrical shock hazard, only qualified personnel with
appropriate service documentation should service the unit.
should be taken when working on batteries: remove watches, rings or other
metal objects; use tools with insulated handles.
damage could occ u r.
specifications shown on the column of the pedestal / stand and on the
damage.
To prevent equipment damage or accidental power disconnections, do not
BT-500 Operation Manual 11
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Power off when the
could be damaged.
1.4 General precaution on environment
Do not keep or operate the equipment under the environment listed below.
Avoid placing in an area
exposed to moisture. Do not
touch the equipment with
wet hand.
Avoid placing in an area
where there is a high
variation of temperature.
Operating temperature
ranges from 20°C to 30°C.
Operating humidity ranges
from 0% to 95%.
Avoid exposure to direct
sunlight
Avoid in the vicinity of Electric
heater
Avoid placing in an area
where there is an excessive
humidity rise or ventilation
problem.
Avoid placing in an area
where chemicals are stored
or where there is in danger
of gas leakage.
Do not disjoint or
disassemble the equipment.
BISTOS Co., Ltd. does not
take responsibility of it.
Avoid placing in an area where
there is an excessive shock or
vibration.
Avoid dust and especially
metal material into the
equipment.
equipment is not fully
installed.
Otherwise, the equipment
BT-500 Operation Manual 12
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Symbol
Description
Used to identify safety information. Be well-known this information thoroughly before using
BT-500. During the operation, do not disconnect any cable.
ates the authorized representative in the European Community of
This symbol indicates the compliance with the essential requirements and provisions of the
This symbol indicates to not dispose the device together with unsorted municipal waste(for
EU only). The solid bar symbol indicates that mains adapter is put on the market after 13
1.5 Symbols
Indicate the warning for hot surface.
Type BF Applied part
Refer to operation manual. Read manual before placing the device.
Skin temperature sensor #1, to be connected to infant’s abdomen for baby (controlled) mode
Skin temperature sensor #2, to be connected to other than infant abdomen
IPX0 IPX0 Non-protected against ingress of water with harmful effects. (Device)
IPX1 IPX1 Protected against the vertically dripping water (Skin temperature sensor_2EA)
IPX2 IPX2 Protected against the dripping water (SpO2 sensor)
IPX6 IPX6 Protected against the powerful jetting (Foot switch_2EA)
Indicates the weight limit
This symbol indicates the manufacturer.
This symbol indicates the serial number of the device.
This symbol indic
manufacturer.
This symbol indicates to keep the device dry.
This symbol indicates to keep the correct upright position on the transport package.
This symbol indicates the device is fragile.
This symbol indicates the temperature limitation for operation, transport and storage.
This symbol indicates the humidity limitation for operation, transport and storage.
This symbol indicates the packing material is recyclable.
External Signal IN/OUT Port
Medical Device Directive 93/42/EEC as amended by 2007/47/EEC.
August 2005.
BT-500 Operation Manual 13
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
2 Introduction
2.1 General
This chapter provides a general description of the BT-500 infant incubator including.
Brief Device Description
Product Features
Model Configurations
2.2 Brief Device Description
A mains electricity (AC-powered) unit designed to provide an enclosed controlled environment to
maintain appropriate temperature and humidity levels mainly for premature infants and other
newborns who cannot effectively regulate their body temperature. It typically consists of a clear
removable plastic hood with a mattress. It typically includes a means to warm the infant such as
providing heated air; temperature controls that work automatically either by measuring the air
temperature or through a temperature sensor attached to the infant skin; and humidity controls. The
device is intended to use in a hospital.
2.3 Intended Use
BT-500 is an infant incubator for non-invasively measuring and showing graphically humidity, air
temperature, skin temperature, O
maintaining life of a premature baby or a precocious baby under 2kgs. This device is for use only by
trained medical personnel located in hospital. Also this device can be used in the all departments of
the hospital which offers a neonatal care service such as NICU(Neonatal Intensive Care Unit), special
nursery unit and pediatrics.
2.4 Operating Principles
Air and skin temperature measurement and Control: Internal cartridge heater raises the
temperature. The infant environmental temperature value is determined by air or skin
temperature that is measured by the sensor module. Through the main fan, the air is
circulated within the hood and controls the temperature.
Humidity measurement and control: The steam of humidity module vaporizes the water
particles. It control the humidity within the hood. The water is sterilized by boiling it to 100℃.
O
Module (Optional): The oxygen concentration in the hood is controlled when the oxygen
2
option was attached.
SpO2 measurement (Optional): The SpO
installed. The probe sensor is applied to the end of an infant’s finger. By measuring the
intensity of reflected light which is dependent on the concentration of dissolved oxygen in the
blood, the oxygen saturation can be determined.
Pulse rate measurement (PR): During the SpO
pulsates by the heartbeat. By measuring this pulsation, the device can determine the pulse
rate of the subject.
2.5 Essential performance
Air temperature measurement and incubator air temperature control.
Skin temperature measurement and infant Skin temperature control.
Humidity measurement and control inside the incubator.
An alarm occurs if the difference between the control temperature and the reading
Module, weight and SpO2. This data is intended to aid the
2
of infant can be measured when the option was
2
measurement, the light reflected by blood is
2
BT-500 Operation Manual 14
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
WARNING
Do not use the skin temperature control on babies who are in shock or who have high
Picture
Name
Description
Qty
Hold up the hood and be composed with
and humidity
medical equipment and items which
Measures temperature and humidity inside the
AC Power cord(AC Power cord for operating the
temperature is more than a certain level.
System must remain in a safe condition acc. IEC 60601-1, IEC 60601-2-19
temperatures
2.6 Accessories and options
Control shell
(Standard)
(Standard)
Fixed Stand
(Standard)
(Optional)
Partition
(Optional)
module
(Standard)
Hood
Basket
Basket
Sensor
instruments and parts that control the temperature
Made of double framed clear acrylic panel to watch
inside, and to minimize heat loss
1ea
1ea
Movable incubator cradle with wheels 1ea
Storage of
infant needs
1ea
Partition of Basket 1ea
hood and infant’s body temperature
1ea
Mattress tray
(Standard)
mattress
(Standard)
Baby desk with X-ray tray 1ea
Accommodate infant stably with bouncy mattress 1ea
Skin
temperature
sensor
Measures infant’s skin temperature 2ea
(Standard)
IV-pole
(Optional)
IV hanger. 1ea
AC power
code
(Standard)
equipment)
1ea
BT-500 Operation Manual 15
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Displays measured values from the control and
Masimo
(Optional)
Notes;
External LCD
Monitor
(Optional)
video of infant inside the hood.
1ea
CCD Camera
(Optional)
Masimo SpO2
sensor probe
(Optional)
Extension for
SpO2 sensor
IV plate
(Optional)
Shelf
(Optional)
Lift Stand
(Optional)
Takes video of infant inside the hood 1ea
Measures infant’s SpO
Extend sensor cable 1ea
Plate to place items which infant needs 1ea
Plate to place items which infant needs 1ea
Movable incubator cradle with wheels (VHAVariable Height Adjustable)
1ea
2
1ea
Weighing
Scale
(Optional)
Measures Infant’s weight
1ea
* The built-in air filter requires periodic replacement to maintain clean air. Periodic replacement of
the pads according to the maintenance schedule is recommended.
**Oxygen control module is also available as an optional component.
BT-500 Operation Manual 16
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Control Shell
① Console box
① Rear access door
2.7 Appearance of BT-500
2.7.1 Front View
①
② Hood
③ Sensor module
④ Moving Stand
⑤ Basket
⑥ IV pole
⑦ IV plate
⑧ External monitor
2.7.2 Front View Detail
2.7.3 Rear view
Figure 2-1. Front view
② Water tank draw
③ Tilting mechanism handle
④ Hood handle
⑤ Front access door
⑥ Baby desk with X-RAY tray
⑦ Weighing scale
⑧ Compatible mattress
Figure 2-2. Front view details
② AC power cord & connector
Figure 2-3. Rear view
BT-500 Operation Manual 17
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Sensor module, SpO2 sensor &
2.7.4 Side view
①
external communication port
② Main power switch
③ Main power AC connector
④ Incubator handle
2.8 Description of each part
Figure 2-4. Left view
Figure 2-5. Right view
BT-500 is composed with several parts. The control shell is the part which controls the entire device.
To measure the infant’s environment, the sensor module is needed. The hood is used to protect an
infant from the external environment and maintain the internal environment of hood to best
condition.
2.8.1 Control shell
The control shell part consists of console box and water tank.
BT-500 Operation Manual 18
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
WARNING
To prevent accidental disconnection, secure all access panel latches. Failure to do
Figure 2-6. Control Shell
The water tank has a 1 Liter capacity. The reservoir permits visual
inspection of the water level. It is located in a drawer in the front of
the shell. When the drawer is closed and the latching handle is
engaged, the reservoir is connected to heater module.
For more information about how to clean, see "Chapter 6 cleanliness
and maintenance”
2.8.2 Hood
The hood of BT-500 is an acrylic material. There is Access door in the front, rear and both sides of
hood.
Figure 2-8. Hood front view & Access door / side function
so could result in personal injury of equipment damage.
BT-500 Operation Manual 19
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Rotate the pawl latches, and open the front
latches until they are fully engaged.
WARNING
When raising or lowering the incubator, the operator should ensure that both
could result in personal injury of equipment damage.
2.8.3 Mattress tray
access panel.
Pivot the front access panel to the full open
position (hanging straight down).
Slide out the mattress tray to the fully extended
position (up to 22cm).
Carefully lean on the mattress tray to ensure it is
properly supported and provides a firm infant
platform.
Return the mattress tray.
Close the front access panel, and rotate both
2.8.4 Stand
Max. to 220mm
Weight limit : under 20kg
BT-500 has two types of stand as fixed and lifting. Following figure show the fixed stand.
In case of lifting type, you are able to adjust the vertical height using two sets of up/down arrow on
footswitch. (VHA - Variable Height Adjustable) This type is optional. The height of the stand can be
adjusted if necessary by stepping on an appropriate side of the pedal for height adjustment.
Figure 2-12. Fixed Stand (Standard)
equipment and appendages are clear of the unit’s travel path. Patient and
incubator connections must also be checked before adjusting the incubator
height. Never place any objects on top of the drawer assembly and always
check before lowering the VHA that there is sufficient clearance between the
incubators and stand assembly. Do not raise or lower the unit while installing
or removing medical gas tanks from the tank holder assembly. Failure to do so
Figure 2-13. Lifting Stand (Optional)
BT-500 Operation Manual 20
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
① IV ringer pole assembly
3 Install and Connection
Attention to follow direction for installing BT-500.
Use this device in 20~30 ℃ of environmental temperature and 0~90 % of humidity.
Check the connection of the AC power cord and then use this cord,
Caution to this device because of easy to break.
When connecting AC power cord, connect the one electrical outlet and infant incubator.
Install the main body on horizontal location.
Do not use the electrical cord connected to generate a noise.
Install away from dust or inflammable material
3.1 IV pole Assembly
The IV poles can be mounted as below. Check each Part name and number
② IV plate assembly
③ IV external monitor assembly
④ Shelf
BT-500 Operation Manual 21
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
LCD monitor assembly (1ea)
① IV plate bolt (1ea), hexagon wrench
Ringer pole (1ea)
3.1.1 IV External monitor
3.1.2 IV plate
3.1.3 IV ringer pole
①
② Machine screw M5x5 (3ea), Hexagon
wrench 2.5mm
③ IV LCD pipe, 25.4mm (1ea)
④ Machine screw M3x12 (4ea), pan head
⑤ IV LCD frame (1ea)
⑥ Lock washer M6 (2ea)
⑦ Screw M6x25 (2ea), Hexagon wrench
5mm
⑧ LCD monitor assembly cable (1ea)
② IV plate (1ea)
③ IV plate bushing, 2T (1ea)
④ IV plate support, 5T (1ea)
⑤ IV pole guide (2ea)
⑥ Socket screw M5x5 (8ea), hexagon wrench
2.5mm
⑦ Machine screw M3x12, pan head (4ea)
⑧ IV plate pole, 25mm (1ea)
⑨ Machine screw M3x18, pan head (4ea)
⑩ Spring washer M3 (4ea)
⑪
Lock washer M6 (2ea)
IV plate frame (1ea)
⑫
Screw M6x25 (2ea), hexagon wrench
⑬
5mm
①
② IV ringer pipe, 25.4mm (1ea)
③ Machine screw M3x12, pin head (4ea)
④ IV LCD frame (1ea)
⑤ Lock washer M6 (2ea)
⑥ Screw M6x25 (2ea), hexagon wrench
5mm
BT-500 Operation Manual 22
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Total assembly
WARNING
Air filter exchange period is once in 3 month. Please check frequently and
Connect 110V or 220V AC power cord to power,
3.2 Air Filter Assembly
①
② Micro filter (1ea)
③ Filter cover (1ea)
④ Filter cover bolt (2ea)
carefully that it is not dirty.
3.3 Connection of Power and Cable
3.3.1 Power Connection
and connect the code to the power input terminal
one side of the body below the BT-500 body. At
that time, make sure latch the cord by using the
locking device for prevent of unexpected
separation. Afterward, it is operated by pushing
the power switch located at the bottom of the
front of the device as shown.
When power is supplied normally, the power
indicator LED is lit and Self-test screen appears at
the same time.
BT-500 Operation Manual 23
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Connect all probes and cables to the sensor part in the
WARNING
When inserting or removing the connectors, be careful not to put or pull the
so could result in personnel injury or equipment damage.
Connect the sensor module as shown. Remove the
WARNING
Prior to placing the infant in the incubator, pre-warm the incubator to the
protocol.
3.3.2 Cable Connection
hood and left and rear part of stand as shown.
cable by force or twist it. Be sure to confirm the position, direction and shape
and to pull lightly the locking metal part of the end of the cable. Failure to do
3.3.3 Sensor Module Connection
connector of the sensor module. While pulling the
both guide locks, pull the sensor module out of the
hood.
3.4 Placement Infant
To place an infant in the incubator, perform the following:
1) Pre-warm the incubator.
2) Rotate the pawl latches and open the front access door of hood.
3) Place the infant on the center of the mattress carefully.
4) Close the access door and ensure the pawl latches are fully engaged.
temperature prescribed by the attending physician, or according to nursing
BT-500 Operation Manual 24
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
WARNING
Never move the incubator when there is an infant in the incubator.
damage.
To place the BT-500 in desired place,
CAUTION
Install the incubator on a horizontal and stable place. Step on the two
when there is difficult to disconnect from the supply mains.
3.5 Movement and lock
To move the position of incubator, perform the following:
1) Identify that infant is not in incubator.
2) Turn off all accessories plugged into the stand receptacles and remove accessories and items not
in use during movement.
3) In case of VHA, adjust the stand to its lowest position prior to move the equipment.
4) Coil up the power cord and secure it.
Always push or pull the incubator forward or backward in a straight line
along the length of the stand (from the ends). Lateral or angular movement
(across the width) can result in inadvertent tip-over if the wheels encounter
any obstacle. Personnel injury or equipment damage could occur.
Always use two people when moving the incubator. When moving the
incubator within the same floor space, And remove or secure all loose
system components to prevent possible patient injury or equipment
damage. If the move involves varying floor heights or a complete floor level
change (i.e. thresholds, ramps, elevators), remove all items either not being
used or not necessary for the move, lower the VHA, IV poles and shelves to
their lowest position, place all drawers in their locked state, and remove all
accessories from the front and rear rail position.
Always close and latch drawers when not in use and particularly when the
incubator is being moved.
For optimum stability, always lower the incubator to its lowest position prior
to transport. Make sure that items placed on the monitor shelf are properly
secured. Failure to do so could result in personal injury or equipment
you should lock the two casters on the
stand. To lock a caster, lower the
stopper on the caster to the locking
position. To unlock a caster, raise the
stopper.
stoppers to lock the casters securely. To move the incubator to another
place, be sure to unlock the casters.
The AC power plug is a means to isolate its circuits electrically from the
supply mains on all poles simultaneously. Do not place the device in an area
BT-500 Operation Manual 25
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
connection may differ depending on the
4 Operation
4.1 System Start-up
1) Turn on the power switch in the bottom
left of the BT-500 front.
2) Then check the following logo is
displayed in the main LCD.
3) Make sure the system check screen is
being output. Make sure each piece of
equipment is functioning properly. The
options and modules connected or not.
BT-500 Operation Manual 26
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Control panel
4) Ensure that the various items properly
output to the main display screen of the
product. If the O
control option has
2
been added to the main display screen
it can be changed to control the O
concentration inside the incubator.
2
4.2 LED
The following symbol s are f u nctions and descriptions of LEDs on the control panel.
SymbolNameDescription
[Mute]
[Keylock] Lock the Key, used to unlock it. When key is locking, LED is on.
[Override] In override mode, LED is on.
[AC power]
If the key beep is on, whenever you push any keys the beep sounds.
To use control panel, you must unlock the key lock function first. This is the protective method of
misuse or unintended operation. To unlock the key locking function, you should use [Keylock]
When alarm situation is activated, using this key you can off the
sound for a certain time. In this case, LED is off.
In case of AC power cord is disconnected by accident, LED is
blinking.
Table 4-1. Function of LED
key.
When you push [Keylock] key, LED is off. Then you can go into mode setting and setting of control
parameters.
BT-500 Operation Manual 27
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Symbol
Name
Description
Use this key to select the parameter for set-up.
Use this key to silent during a period of time in case of
Display window
Action
Description
Control
Rotate CW
+0.1 ℃ or +1 %
Rotate CCW
-0.1 ℃ or -1 %
Menu
Push
Select options / Finish selecting options
Rotate(CW/CCW)
Shift options and change setting value
4.3 Key and knob operation
4.3.1 Key
[Parameter]
Use this key to select the temperature, humidity
(option : select the oxygen concentration)
[Mode] Use this key to select control mode.
[Up] Use this key to raise the parameter’s value.
[Down] Use this key to lower the parameter’s value.
[Enter]
Use this key to enter the parameter’s value to control mode.
[Menu] Use this key to enable the use of Knob to select menu options.
[Mute]
the alarm situation.
[Keylock] Use this key to Lock or unlock.
Table 4-2. Use of Key
4.3.2 Knob
Not only Knob is used for changing the setting value, but also it can be used for shifting and changing
the Menu options. For changing the setting value, when the setting is complete, setting value is
reflected to control by pushing. For using the purpose of Menu option, you should [Menu]
when key locking is released.
Following table shows how to handle knob in each case.
setting value
4.4 Displays
BT-500 provides the one display screen and the one dynamic Menu window together. Figure 4-6 is the
basic layout of BT-500.
Push Reflect setting value.
Table 4-3. Use of Knob
BT-500 Operation Manual 28
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Figure 4-6. Main operation display
Figure 4-7. Basic screen of Main display (O2 control available)
① Air temperature region
This region displays the measured air temperature in the hood. Celsius(℃) or
Fahrenheit(℉) can be displayed according to set. In case of module failure, connection failure
and senor failure is indicated by “--.-”.
Note: In this region, current heater power is displayed by gauge
② Skin temperature region
This region displays the patient’s skin temperature in the hood. Celsius(℃) or Fahrenheit(℉)
can be displayed according to set. The case of module failure, connection failure and senor
failure is indicated by “--.-”.
③ Humidity region
This region displays the measured relative humidity in the hood. Correct unit
is %RH and % is displayed only.
Note: In this region, current water level is displayed by gauge.
&
.
BT-500 Operation Manual 29
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
In this region, according to control mode,
especially air mode, user is able to set the
user is able to set the patient’s skin
temperature in the hood. At this time,
setting value that is lower than outer
temperature. Also user can set the relative
humidity in the hood.
This region displays the lists of items regarding
value and PR(pulse rate)
with unit(%, bpm). When the signal is not
Temperature
Setting Value
Humidity
Setting Value
④ Setting region
temperature in the hood. In case of skin mode,
equipment does not control the temperature
⑤ SpO2 region
the SpO2. SpO2
extracted from SpO2 waveform are displayed
setting value Pulse rate value
SpO
2
measured, it is indicated as “---”.
⑥ O2 region
This region displays the O2 concentration by the % unit in the hood When the O2 sensor has
malfunction or disconnection, it is indicated as “-- ”. When the O2 sensor reaches the
saturating concentration, it is indicated as “**”.
⑦ Scale region
This region displays weight of infant. When the scale module has malfunction or
disconnection, it is indicated as ” -.---”
⑧ Status region
This region displays the Date/Time, patient ID, control mode icon, override mode icon,
Keylock icon, and Sound icon. In control mode, there are air mode and skin mode and
configuration color is changed by modes status. In override mode, air mode, if setting
temperature is above 37.1 ℃, skin mode, if setting temperature is above 37.6 ℃, the
setting can be executed with [keylock] key. In case of sound icon,
Press the [Mute key]
Following table indicates the function of icon used in the status region.
Icon Name Description
Air mode icon Control mode status : Air mode
Skin mode icon Control mode status : Skin mode
Humidity mode icon Control mode status: Humidity mode
Trend icon Trend indication
O2 icon O2 indication
SpO2 icon SpO2 indication
button to pause the alarm sound in case of the alarm situation.
BT-500 Operation Manual 30
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Air/Skin temperature
Override mode icon
1) To unlock the key, push the [keylock] key. (red underline
key or rotate CW the knob to raise the setting
or rotate CCW the knob to
hood is controlled automatically.
PR icon Pulse rate indication
Weight icon Weighing scale indication
Key lock icon Key is locking.
Key unlock icon Key input is enable..
Air/Skin temperature Override mode On.
Alarm on icon Alarm Sound on.
Alarm off icon Alarm Sound off.
4.5 Temperature Measurement and Control
Measurement: Measurement of air temperature and skin temperature are starting at the same
time of BT-500 power on. In case of skin temperature, especially, metal part of skin temperature
sensor probe must be attached to the infant’s abdomen.
Setting and Control
① Air mode: The measured value is displayed on air temperature region in upper left of screen
and the setting value is displayed on setting region in upper right of screen with blue.
on setting value)
2) Push the [Up] key or rotate the knob CW to raise the
setting temperature from 23.0℃ to 37.0℃ in 0.1℃
increments. In temperature override mode, push the [Up]
temperature from 37.1℃ to 39.0℃.
Push the [Down] key
lower the setting temperature from 39.0℃ to 23.0℃ in
0.1℃ increments.
3) To confirm the setting value and lock the key, push the
[keylock] key.
4) According to this value, the temperature of inside of the
Note: When key locking is released, LED is off and key lock icon is removed. Also, if no keys are
pushed or knob is rotated within 10s of selecting display, it is automatically locked.
② Skin mode: The measured value is displayed on skin temperature region in upper middle of
screen and the setting value is displayed on setting region in upper right of screen with
yellow.
BT-500 Operation Manual 31
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
1) To unlock the key, push the [keylock] key. (red underline on
CW the knob to raise the setting temperature from 37.6˚ to
When using two skin temperature sensors, it is displayed as
1) To unlock the key, push the [keylock] key. (red underline on
hood is controlled automatically.
WARNING
Premature neonates below 30 weeks may need humidity over 80% for long
water vapor on the wall because of possible bacterial colonization.
setting value) By pushing [Mode] key , change control
mode status.
2) Push the [Up] key
setting temperature from 35.0˚ to 37.5˚ in 0.1˚ increments.
or rotate the knob CW to raise the
In temperature override mode, push the [Up] key or rotate
39.0˚.
Push the [Down] key or rotate CCW the knob to lower
the setting temperature from 39.0˚ to 35.0˚ in 0.1˚
increments.
3) To confirm the setting value and lock the key, push the
[keylock] key.
4) To maintain the infant’s body temperature, the
temperature of inside of the hood is controlled
automatically.
5)
shown. In order to use skin mode as the control mode, skin
temperature 2 should be removed.
4.6 Humidity Measurement and Control
Measurement: Humidity measurement starts at the same time when BT-500 is powered on.
Setting and Control: The measured value is displayed on humidity region in upper middle of
screen and the setting value is displayed on setting region in upper right of screen with sky-blue.
setting value) By pushing [Parameter] key, change the
parameter.
2) Push the [Up]
or [Down] key or rotate the knob to
raise or lower the setting humidity from 40% to 95% in 1%p
increments.
3) Push the [Enter] button for reflecting set-up and push the
[keylock]
4) According to this value, the relative humidity of inside of the
key to lock.
periods of time. In this high humidity environment, water vapor
condensation could build up inside hood surface and prolonged
condensation can lead to rain-out or watery surface underneath mattress.
It is highly recommended that user should prevent water vapor
condensation by controlling temperature and humidity and clean out
BT-500 Operation Manual 32
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Measurement: O2 concentration in the hood begins automatically
in upper right of screen with green.
1) To unlock the key, push the [keylock] button. (red underline
value automatically.
WARNING
Oxygen flow rate cannot guarantee the accuracy of oxygen concentration
Oxygen concentration should be continuously measured by calibrated
An oxygen analyzer shall be used separately when oxygen is delivered to
equipment that produces sparks in an incubator. Personal injury or
damage.
Measurement: Weighing measurement starts at the
Kg or pound unit according to set record.
4.7 O2 Measurement and Control
starting operation at the same time. The measured value is
displayed on O2 region in the lower of screen with green.
Setting and Control: The measured value is displayed on O2 region
in the lower of screen; setting value is displayed on setting region
on setting value) By pushing [Parameter] key, change the
parameter.
2) To unlock the key, push the [Keylock] key
. (red underline
on setting value)
3) To change the setting mode, push the [Parameter] key
.
Setting mode is Temperature, Humidity, Oxygen sequentially.
4) To set the key, push the [Up]
, [Down] key or
control knob key to set the O2 concentration unit 1%
5) Press the [Enter] key
the [Keylock] key
to reflect the setting and then press
to lock.
6) Adjust the O2 concentration in accordance with the setting
in the incubator.
oxygen analyzer.
Administration of oxygen may increase the noise level for the infant within
the infant incubator.
the infant.
Use of anesthetic agents can interfere with oxygen analyzer accuracy.
As oxygen use increases the danger of fire, do not place auxiliary
equipment damage could occur.
Disconnect the incubator from the hospital oxygen source when oxygen is
not in use. Failure to do so could result in personal injury or equipment
4.8 Weighing Scale Measurement
same time when BT-500 is powered on. Measured data
continuously displayed on lower part of screen in white
BT-500 Operation Manual 33
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Measurement: To measure SpO2,
screen in white(SpO2) and blue(PR).
Functions of system, alarm, trend, and modules are able to be set and
After pushing the [keylock] key to unlock,
rotation of the Knob is used to shift between menu options and pushing is
used to reflect the setting value. When the cursor is located on arrow, you
You can change Temperature unit, Weight
around with noise level more than 66dB.
The silence period of alarm can be set
You can set the delay time 0,5,10 and
levels.
4.9 SpO2 and PR(pulse rate) Measurement
apply the probe to the patient’s toe
or earlobe. The exact applied part is
decided according to the medical
opinion. The measured value is
displayed on SpO2 region in lower of
4.10 Menu functions
changed by using Menu options.
press the [menu] key to go to the menu function.
To change and move to other category, use Control Knob. CW or CCW
can move to other menu category tabs.
① System
unit, Language, Time and Warm-up mode.
There are 3 different kind of warm-up mode
Silent, Normal, and Fast Mode.
Silent: Warm-up time will take 1 hour
around with noise level less than 45dB.
Normal: Warm-up time will take 45Min
around with noise level less than 55dB.
Fast: Warm-up time will take 25Min.
② Alarm
to 5, 10, 20 30, and 60 minutes. For
the silence period, alarm sound is
silent and silent time is displayed on
main display.
SpO2 Alarm delay refers to the delay
time after an alarm condition occurs,
an alarm sounds until the occurrence.
15 seconds.
And you can set PR Beep volume
level. There are 5 different volume
BT-500 Operation Manual 34
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Trend period is set as 9, 18, 36, and 72
Trend display
The calibration of O2(21%, 100%, weight is
available. Select the O2 setting mode or measured
xternal monitor in
the monitor mode function. Read the chapter 4.11
Below table is the description of SpO2 setting
This mode is provides the best combination of sensitivity and
the majority of patients.
This mode should be used for the sickest patients, where
clinician and patient contact is continuous.
③ Trend
hours and it is displayed on main display.
Also each trend display data are possible to
enable or disable: air temperature, skin
temperature, humidity and oxygen. These
trend display datum are displayed on main
operation display with setting period.
④ Modules
mode of main display. And select the Cam mode,
Graph mode, Masimo mode of e
to understand explanation of each monitor mode.
In addition, set the Cam. Hor. Mirror to reverse
the display from side to side.
⑤ SpO2
menu. Please refer to below table to set SpO2
settings. Changed setting will be apply on SpO
and display on external monitor, please refer to
4.11 for external monitor explanations.
Items Description
sensitivity
Normal
2
probe-off detection performance. This mode is recommended for
Algorithm
mode
Maximum
sensitivity
APOD This mode is the least sensitive in picking up a reading on patients
obtaining a reading is most difficult. Maximum Sensitivity is
designed to interpret and display data for even the weakest of
signals. This mode is recommended during procedures and when
BT-500 Operation Manual 35
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
with low perfusion but has the best detection for prove-off
risk of the sensor becoming detached. (Pediatric, combative, etc. )
2 seconds
4 seconds
8 seconds
10 seconds
12 seconds
14 seconds
16 seconds
SpO2
limit
BT-500 provide the simple
conditions. This mode is useful for patients that are at particular
averaging
⑥ Help
information about the
equipment. The information
is as follow:
parts of full set BT-500
main control panel and
how to use and change
FastSat
ON This mode is always on for 2-4 and 4-6 averaging modes.
OFF -
SpO2
alarm
PR Alarm
limit
Low If the SpO2 value is lower than this value, an alarm is occurring.
High If the PR value is higher than this value, an alarm is occurring.
Low If the PR value is lower than this value, an alarm is occurring.
main display
mode and parameters
SpO2 averaging time
BT-500 Operation Manual 36
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
WARNING
⑦ Scale : If weighing scale module is connected, a menu of weight calibration appears. Press
the [keylock] key
to unlock and press the [Menu] key more than 2 seconds then,
the menu will be displayed.
4.11 Pulse O xim et er
• Pulse oximeter of BT-500 must be operated by qualified personnel only.
Pulse oximetry is a continuous and non-invasive method of measuring the level of arterial oxygen
saturation in blood. The measurement is taken by placing a sensor on a patient, usually on the
fingertip for adults, and the hand or foot for neonates. The sensor is connected to the pulse oximetry
instrument with a patient cable. The sensor collects signal data from the patient and sends it to the
instrument. The instrument displays the calculated data in two ways: 1) as a percent value for arterial
oxygen saturation (SpO
), and, 2) as a pulse rate (PR).
2
- Operating Principles
: The BT-500 MS board pulse oximeter is based on three principles:
1. Oxyhemoglobin and deoxyhemoglobin differ in their absorption of red and infrared light
(spectrophotometry).
2. The volume of arterial blood in tissue and the light absorbed by the blood changes during
the pulse (plethysmography).
3. Arterio-venous shunting is highly variables and that fluctuating absorbance by venous
blood is a major component of noise during the pulse.
The pulse oximeter of BT-500 as well as traditional pulse oximetry determines SpO
by passing red and
2
infrared light into a capillary bed and measuring changes in light absorption during the pulsatile cycle.
BT-500 Operation Manual 37
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Red and infrared light-emitting diodes (LEDs) in oximetry sensors serve as the light sources, a
photodiode serves as the photodetector.
Traditional pulse oximetry assumes that all pulsations in the light absorbance signal are caused by
oscillations in the arterial blood volume. This assumes that the blood flow in the region of the sensor
passes entirely through the capillary bed rather than through any arterio-venous shunts. The
traditional pulse oximeter calculates the ratio of pulsatile absorbance (AC) to the mean
absorbance(DC) at each of two wavelengths, 660 nm and 905 nm:
S(660) = AC(660) / DC(660)
S(905) = AC(905) / DC(905)
The oximeter then calculates the ratio of these two arterial pulse-added absorbance signals:
R = S(660) / S(905)
This value of R is used to find the saturation SpO
in a look-up table built into the oximeter’s software.
2
The values in the look-up table are based upon human blood studies against a laboratory co-oximeter
on healthy adult volunteers in induced hypoxia studies.
The Masimo SET MS board pulse oximeter assumes that arterio-venous shunting is highly variable and
that fluctuating absorbance by venous blood is the major components of noise during the pulse. MS
board decomposes S(660) and S(905) into an arterial signal plus a noise component and calculates the
ratio of the arterial signals without the noise:
S(660) = S1 + N1
S(905) = S2 + N2
R = S1 / S2
Again, R is the ratio of two arterial pulse-added absorbance signals and its value is used to find the
saturation SpO
in an empirically derived equation into the oximeter’s software. The values in the
2
empirically derived equation are based upon human blood studies against a laboratory co-oximeter on
healthy adult volunteers in induced hypoxia studies.
The above equations are combined and a noise reference (N’) is determined:
N’ = S(660) – S(905) x R
If there is no noise N’ = 0: then S(660) = S(905) x R which is the same relationship for the traditional
pulse oximeter.
The equation for the noise reference is based on the value of R, the value being seeked to determine
the SpO
between 1% and 100% and generates an N’ value for each of there R-values. The S(660) and S(905)
signals are processed with each possible N’ noise reference through and adaptive correlation
canceler(ACC) which yields an output power for each possible value of R(i.e., each possible SpO
1% to 100%). The result is a Discrete Saturation Transform (DST™) plot of relative output power versus
possible SpO
. The MS board software sweeps through possible values of R that corresponds to SpO2 values
2
from
2
value as shown in the following figure where R corresponds to SpO2 = 97%:
2
BT-500 Operation Manual 38
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
WARNING
Do not remove the grounding conductor from the power plug on any
must be intact and undamaged.
The DST plot has two peaks: the peak corresponding to the higher saturation is selected as the SpO2
value. This entire sequence is repeated once every two seconds on the most recent four seconds of
raw data. The MS board SpO
therefore corresponds to a running average of arterial hemoglobin
2
saturation that is updated every two seconds.
- Grounding
Connect the oximeter only to a three-wire, grounded, hospital-grade receptacle. The three-conductor
plug must be inserted into a properly wired three-wire receptacle; if a three-wire receptacle is no
available, a qualified electrician must install one in accordance with the governing electrical code.
circumstances.
Do not use extension cords or adapters of any type. The power cord and plug
If there is any doubt about the integrity of the protective earth conductor arrangement, operate the
oximeter on internal battery power until the AC power supply protective conductor is fully functional.
- Patient Isolation
To ensure patient electrical isolation, connect only to other equipment with electronically isolated
circuits.
Note: Do not connect to an electrical outlet controlled by a wall switch or dimmer.
- Cabling entanglement/strangulation
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient
entanglement or strangulation.
BT-500 Operation Manual 39
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
WARNING
Do not place the monitor or external power supply in any position that might
or patient cable; use only the handle on the monitor
Air temperature, skin temperature, humidity,
Air temperature, skin temperature, humidity,
Plethysmograph, and SIQ). If there’s some alarms,
Masimo mode”.
cause it to fall on the patient. Do not lift the monitor by the power supply cord
- RS-232 & Alarms
RS-232 System Interconnection. Consult IEC-601-1-1 for system interconnection guidance. The specific
requirements for system interconnection are dependent upon the device connected to the MS board
pulse oximeter and the relative locations of each device from the patient, and the relative location of
the connected device to the medically used room containing the MS board pulse oximeter. In all
circumstance the MS board pulse oximeter must be connected to a grounded AC power supply. The
MS board pulse oximeter is referred to as an IEC 60601/F device in the summary of situations table
contained in IEC 60601-1-1.
Check alarm limits each time the MS board pulse oximeter is used to ensure that they are appropriate
for the patient being monitored.
4.12 External Monitor
BT-500 uses external 7” Color TFT LCD monitor that displays measured values from the control and
video of infant inside the hood. Cam mode, Graph mode and Massimo mode are shown as below.
(1) Cam Mode
oxygen concentration and Weighing scale are
displayed on external monitor with real-time
video of infant.
There’s no information for the Masimo board. If
there’s some alarms, it would display message
“Check alarm on Masimo mode”.
(2) Graph Mode
oxygen concentration and Weighing scale, SpO2
and PR are displayed on external monitor with
real-time video of infant. Also, graphs of each
parameter are displayed.
There’s no information for the Masimo board. It
just displays basic parameter (SpO
it would display message “Check alarm on
, PR,
2
BT-500 Operation Manual 40
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Air temperature, skin temperature, humidity,
(3) Masimo Mode
oxygen concentration and Weighing scale are
basically displayed. SpO2 and PR are mainly
displayed.
- Parameter Region
- Information Region
- PPG and SIQ Region
BT-500 Operation Manual 41
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
RED
RED
RED
YELLOW
RED
RED RED
RED
RED
RED
RED
RED
RED
RED
- Alarm messages on Alarm Region
Messages display Messages display
- Board failure & Diagnostic Failure code Region
Display when the event occurred. If no failure, it would be blanked. Each description about error
codes is explained in Section 8.
4.13 Shut down
To ensure safe terminating operation of BT-500, follow as below.
1) Turn off the Power switch to power down the incubator.
2) Turn off the main power source.
4.14 Calibration of O2 module
Follow the direction to set and control O
Calibration of O
concentration 21 %: Calibrate O2 concentration to atmospheric oxygen
2
concentration as the standard.
1) Move the position of the sensor module to O
occurred. But if sensor module relocates normal position after calibration, it operates
normally.
2) To unlock the key, push the [keylock]
concentration for improving accuracy in the hood of BT-500.
2
calibration position. At this time, alarm can be
2
button. (red underline on setting value)
3) Press the [Menu]
button to enter the menu category. Select the O2 Calibration category
in [modules] catego r y.
4) Handle the knob to select the 21 % Calibration category. Push the knob to compete the
calibration.
5) If it is indicated as “Done” massage, calibration completes the work.
Otherwise, it is indicated as “Fail” massage, check the O
sensor and atmospheric oxygen
2
whether normal concentration.
6) Move the sensor module to normal position.
BT-500 Operation Manual 42
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Calibration of O2 concentration 100 %: Calibrate O2 concentration to 100 %. Calibration of O2
concentration 100 % should be performed after Calibration of O
concentration 21% is completed.
2
1) Move the position of the sensor module to O
calibration position. Sensor module and “O2
2
calibration kit” place tight. At this time, alarm can be occurred. But if sensor module relocates
normal position after calibration, it operates normally.
2) Connect “O
Pressure of O
3) Turn "Sealing valve" clockwise to close contact with the sensor module and the "O
calibration kit”to be injected 100% O2 concentration. Set oxygen flow to 0 PLM.
2
is between 40PSI(275kPa) and 50PIS(344kPa)
2
2
Calibration Kit".
4) Be careful not to obstruct the "O
5) Inject oxygen into “O
calibration kit” At this time, the flow rate set between 5 LPM and 7
2
LPM (Litter Per Minute). Inject oxygen from 5 minute to 6minute to recognize O
calibration kit" of ”O2 outlet”
2
sensor fully.
2
6) To unlock the key, push the [keylock] button. (red underline on setting value)
7) Press the [Menu]
button to enter the menu category. Select the O2 Calibration category
in [modules] category.
8) Handle the knob to select the 21 % Calibration menu. Push the knob to compete the
calibration.
9) If it is indicated as “Done” massage, calibration completes the work.
Otherwise, it is indicated as “Fail” massage, check the O
sensor and atmospheric oxygen
2
whether normal concentration.
10) Move the sensor module to normal position.
BT-500 Operation Manual 43
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Alarm of BT-500 indicates the failure condition of
Followings are types of alarm message.
System alarms
Situation of pressing the key button
more than 1 minute
Problem with not working or
malfunction of impeller
Required to remove the skin
mode control.
5 Alarms
equipment. According to each alarm condition,
alarm messages are displayed on the front LCD. At
the same time, alarm sound is caused and sensor
module’s LED blinks. Also all setting values for
alarm condition do not changed by any interrupt
such as power on and off.
Alarm pop-up window is activated in the center of
the display as shown in case of the alarm situation
as shown.
5.1 System alarms
This alarm is activated when disorder is occurred during operation of equipment.
It is possible there will be a delay of 1 second to determine and check an accurate alarm condition.
Messages Display Sound
O O O Problem with main motor
O O O Problem with main heater
O O O Problem with water heater
O O O
O O O
O O O
O O O
O O O
O O O
LED
flash
Descriptions
Problem with air humidity
measurement sensor in the hood
Problem with air temperature
measurement sensor in the hood
Problem with power of impeller
motor
Problem with skin temperature
probe
O O O
O O O
O O O
Table 5-1. System alarm message
temperature sensor 2 during skin
When the sensor module is
disconnected
Sensor module is not placed in the
proper position.
BT-500 Operation Manual 44
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Temperature alarms
LED
flash
If the temperature measured is
2.5 ℃
If the measured temperature is
1.5 ℃
If the temperature measured is
2.5 ℃
If the measured temperature is
1.5 ℃
1) In case the measured
temperature is over 37 ℃.
1) In case the measured
temperature is over 37 ℃.
1) In case the measured
temperature is under 37 ℃.
When Hood doors are open or close, BT-500 shows the condition through icons instead of alarm
sound.
Front and rear doors are open.
Front and rear doors are close open
Door open/close icon
Table 5-2. Door Open/Close Icon
Front door is open.
Rear door is open.
5.2 Temperature alarms
This alarm is activated when either disorder is occurred measuring air temperature in the hood or
body temperature and difference of constant figure and over in setting temperature value.
Messages Display Sound
O O O
O O O
lower than the set temperature over
higher than the set temperature over
Descriptions
O O O
O O O
O O O
O O O
lower than the set temperature over
higher than the set temperature over
temperature reaches more than
℃ when the setting
38
temperature is under 37 ℃.
2) In case the measured
temperature reaches more than
when the setting
39 ℃
temperature reaches more than
℃ when the setting
38
temperature is under 37 ℃.
2) In case the measured
temperature reaches more than
39 ℃
when the setting
O O O
temperature reaches 37.5 ℃
±0.5 ℃
when the setting
BT-500 Operation Manual 45
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
2) In case the measured
37 ℃.
Humidity alarms
LED
flash
There is no water in the internal
water tank.
Oxygen alarms
LED
flash
If the measured values of the two O2
certain concentration
Measured O2 concentration is lower
5 %
Measured O2 concentration is higher
5 %
Weighing Scale alarms
LED
flash
Weigh in excess of 10kg is on the
mattress.
temperature reaches more than
℃ ±0.5 ℃ when the
39.5
setting temperature is over
Table 5-3. Temperature alarm message
5.3 Humidity alarms
This alarm is activated when humidity in the hood is lower than setting value or there is no water in
the tank.
Messages Display Sound
O O O
O O O
Table 5-4. Humidity alarm message
Measured value is lower than setting
value over 5%p.
Descriptions
5.4 Oxygen alarm
The alarm will be activated when there are differences between setting O2 concentration range and
O2 concentration inside of hood. It may be activated when there is malfunction on O2 control module.
Messages Display Sound
O O O
O O O If O2 Control module is disconnected
O O O
In case there is problem with power
of O2 control module.
Sensor is a difference more than a
Descriptions
5.5 Weighing Scale alarms
This alarm is activated when the measured weight is not in the measurable range which means an
object other than infant is placed on the mattress.
Messages Display Sound
O O O
O O O
Table 5-5. Oxygen alarm message
O X X
Table 5-6. Weighing alarm message
than setting O2 concentration over
than setting O2 concentration over
Descriptions
BT-500 Operation Manual 46
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
WARNING
Do not place the monitor or external power supply in any position that might
before being used again.
5.6 Alarm self-test
The self-testing is necessary for the alarm system to check whether the performance of the audible
and visible alarms has degraded over time.
System alarm
After removing the sensor module from the top of hood, check the followings. The message
is displayed on LCD with the audible alarm. Also, the red LED flashes on
the sensor moduel connector.
Power fail alarm
If the power cable is disconnected with power on, the LCD screen turns off and the red LED
flashes with a continuous audible alarm.
Note: A fully charged battery should supply the power failure alarm for approximately 30 minutes.
If the alarm is tested for the full 10 minutes, BT-500 must be run at least 3 hours to recharge
the battery before it is used to the patient. Total recharge time is 5 hours when the main
power switch is on.
cause it to fall on the patient. Do not lift the monitor by the power supply cord
or patient cable; use only the handle on the monitor
Do conduct Alarm self-test whether the alarm function is working properly
BT-500 Operation Manual 47
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
WARNING
Do not immerse the sensor in water, solvents, or cleaning solutions. Do not
Do not immerse the patient cable in water, solvents, or cleaning solutions (the
ter proof). Do not sterilize by irradiation,
Incubator should be thoroughly cleaned and disinfected after each baby is
cause an infection of patient.
6 Cleaning & Maintenance
This chapter contains instructions for the care and cleaning of the BT-500 unit and its accessories.
The BT-500 requires proper care and preventive maintenance. This ensures consistent operation and
maintains the high level of performance necessary.
6.1 General cleaning method and precautions
(1) Cleaning Methods
You can maintain the cleanliness using various methods. Though, to avoid the damages or
contamination, use the following recommended methods. .
Clean up with warm water and detergent.
Clean completely the incubator after each baby is discharged and before being used again.
If you use the materials (unauthorized materials) that can cause damage to the device, it is
impossible to get free repair service although the device is in warranty period.
Do not use steam cleaning the all parts of incubator. The excessive moisture can cause
damage.
Please maintain cables with no dust and soil. And clean cables with wet fabric (water
temperature is about 40˚C / 104℉). Please clean cables using clinical alcohol once in a
week.
Do not immerse the device or sensor in liquid or a cleanser. Also all liquid must not be
permitted to enter in the device or sensor.
When cleaning the surface of the air circulation well, take care to prevent liquids from
entering the motor. Then, dry with a clean cloth or paper towel.
When you clean the sensor module, inner and outer wall of the hood, do not use alcohol that
can cause crazing (small stress crack) of the clear acrylic.
sterilize by irradiation, steam, or ethylene oxide.
patient cable connectors are not wa
steam, or ethylene oxide.
discharged and before being used again. If patient is cared for in an incubator
for more than 7 days without the device being cleaned or disinfected, it may
(2) Maintenance Methods
Once a month, rub the main body and accessories using soft fabric with alcohol. Do not use
thinner, lacquer, ethylene or oxidant.
After using the device, please store the device where the place has -20˚C~+60˚C of
temperature and 0%~95% of humidity.
Restrict the device be used by doctors and nurses only.
6.2 Hood
Keep the external surface clean and free of dust, dirt, and residual liquids. Clean with a damp cloth
using mild soap and water or hospital approved, nonabrasive disinfectants.
BT-500 Operation Manual 48
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
WARNING
Turn off the BT-500 and unplug the BT-500 from the AC power source and
hood.
WARNING
Turn off the BT-500 and unplug the BT-500 from the AC power source. Do not
Be careful not to have a burn when clean the radiator of shell. Check the
temperature of the radiator before cleaning.
CAUTION
Take extra care when cleaning the display surfaces, which are sensitive to
WARNING
Do not remove and clean the humidity module in operation.
and the water
Do not autoclave the components of device, except for water tank specifically
identified in this chapter.
WARNING
All Masimo sensors and accessories listed to be available for device.
detach all accessories before cleaning. Do not immerse the unit in water or
allow liquids to enter the case.
Do not use alcohol for cleaning. Alcohol can cause crazing of the clear acrylic
6.3 Shell, Sensor module, Scale module, Basket
Keep the external surface clean and free of dust, dirt, and residual liquids. Clean with a damp cloth
using mild soap and water or hospital approved, nonabrasive disinfectants.
immerse the unit in water or allow liquids to enter the case.
Do not allow liquids to enter when clean the console box. Failure to do so could result
in personal injury or equipment damage.
rough handling. Rub the lens that covers them with a soft, dry cloth.
6.4 Water tank
Maintain inside of the water tank clean from the dust, the residue, the remaining liquid. Use the mild
soap, water, non-abrasive sanitizer approved in the hospital and wipe with damp cloth. Also, the
water tank can be disinfected by using autoclave
The humidifier reservoir (water tank) should be cleaned
changed every day.
.
6.5. Skin temperature sensors and SpO2 sensors.
Keep the external surface clean and free of dust, dirt, and residual liquids. Clean with a damp cloth
using mild soap and water or hospital approved, nonabrasive disinfectants
BT-500 Operation Manual 49
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
The residual water in the humidity module can
cause bacteria multiplication or equipment damage
shown in the figure. Place the water tank more
WARNING
The humidity module can be sufficiently hot to cause burns.
WARNING
The operation of each part should be checked every time before the device is
ng it immediately, and contact our distributor in local
6.6 Drain of residual water
when it is frozen. When the infant leave the
hospital or the equipment not being used, remove
the residual water in the humidity module as
follows:
1) To cool down the humidity module, turn off
the BT-500 and remove water tank and wait
about 60 minutes.
2) Remove the basket from the bottom side as
than 1L capacity below the drain valve.
3) Remove the drain valve from shell bottom
using spanner.
4) Waiting until no more water has fall down
from the drain and turn on the BT-500.
5) Press down the [Parameter] key
than 5 seconds and the system will ask the
password. Enter [0,0,1,1] to start the heater
operated to remove the moisture from
humidity module. For drying out the humidity
module, the heater will be operating about
4minutes and 40 seconds. Be careful the
heated vapor can be expelled at the drain
valve.
6) The BT-500 will be reset. Turn the power off
and assemble the drain valve as reverse order.
The hot water can be poured from drain valve.
6.7 Regular Inspection
Similar to most medical equipment, BT-500 has to be inspected periodically on an annual basis in
general. Refer to the service manual which is supported by Bistos Co.,Ltd. about inspectional items.
Regular inspection must be performed by the company’s technician. The user and operator must not
disjoint or remodel the device.
The operation of each function should be checked every three-month in
used to patient.
general.
more
If any defect should be detected in an inspection, indicate on the unit that it is
out of order, stop usi
area.
BT-500 Operation Manual 50
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
CAUTION
The internal battery must be handled by the company’s technician only. Do
In order to comply with EU Directive 2002/96/EC on Waste Electrical and
Electronic Equipment (WEEE): This product may contain material which
dous to human health and the environment. DO NOT
to be RECYCLED in accordance with local regulations, contact your local
6.8 Battery Replacement and Disposal
If a loss of battery operation run time is noticed, the battery could possibly require replacement.
When disposing or replacing of internal Li-ion battery, adhere to all applicable laws regarding
recycling. Avoid storing battery above 140°F. If clothing or skin comes in contact with material from
inside the battery, immediately wash with plenty of clean water.
not attempt to open the BT-500.
The internal battery is consumables. Therefore the operation time by the battery can be decreased. If
the operation time is not long enough, please contact service center and change the battery. If this
system is used with not sufficient operating time by the internal battery, it is possible to be shut down
the system because of the lack of the internal battery’s capacity. This situation can cause not intended
stop of measuring and monitoring function.
6.9 Disposal of the BT-500
When disposing of the BT-500, adhere to all applicable laws regarding recycling. If you are not able to
dispose the BT-500 or you need a help for disposing the BT-500, please contact us. In the case of
there are no appropriate ways to dispose, we will pick up the BT-500 for you.
DISPOSAL
could be hazar
DISPOSE of this product as unsorted municipal waste. This product needs
authorities for more information. This product may be returnable to your
distributor for recycling - contact the distributor for details.
BT-500 Operation Manual 51
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Physical characteristics
Standard
Dimension: 1,024(W)x690(D)x1,354(H)mm
Weight: Approx. 99.3 Kg
Full loaded
(including IV pole, IV plate, plate, external LCD
stand, O2 control module and check valve)
Dimension: 1,441(W)x797(D)x1,500(H)mm
Body (Shell + Hood)
Dimension: 1,024(W)x604(D)x773(H)mm
Weight: Approx. 52.7 Kg
Fixed stand
Dimension: 1,009.5(W)x815(D)x645(H)mm
Weight: Approx. 36 Kg
Lifting stand
Dimension: 1,441(W)x797(D)x1,500(H)mm
Weight: Approx. 40.5 Kg
IV plate
Dimension: 479(W)x402(D)x756.4(H)mm
Weight: Approx. 5 Kg
Plate
Dimension: 402(W)x302(D)
Weight: Approx. 11 Kg
IV pole
Dimension: 300(W)x310(D)x718 ~ 1,068(H)mm
Weight: Approx. 2.5 Kg
External 7” color TFT LCD
Dimension: 364.7(W)x35(D)x839(H)mm
Swivel angle: +180˚/ -140
Shelf
Dimension: 219(W)x334(D)x36(H)mm
Weight limit: approx. 3 Kg
Basket
Dimension: 464.2(W)x490.5(D)x241.3(H)mm
Weight limit: approx. 10 Kg
Basket partition
Dimension: 391(W)x205(D)x53(H)mm
Weight: Approx. 0.45 Kg
Weighing scale
Dimension: 810(W)x400(D)x46.6(H)mm
CCD camera (NTSC or PAL)
Dimension: 69.1(W)x46(D)x38(H)mm
Resolution: 510x492 pixel
Sensor module
Dimension: 157(W)x119.2(D)x64.5(H)mm
Weight: Approx. 270 g
Sensor module & CCD camera assembly
Dimension: 194(W)x119.2(D)x81(H)mm
Weight: Approx. 310 g
SpO2 probe
Wire length: 1,000 mm
Weight: approx. 30 g
SpO2 probe extension
Wire length: 3,080 mm
Weight: approx. 30 g
Mattress
Dimension: 727(W)x377(D)x27(H)mm
O2 control module
Dimension: 300(W)x170(D)x180(H)mm
Weight: Approx. 760 g
Check valve
Dimension: 330(W)x365(D)x65(H)mm
Weight: Approx. 460 g
Skin temperature: control and measurement of skin temperature of the infant
Control range
Override mode: 37.6 ℃ ~ 39.0 ℃
Measurement range
25.0 ℃ ~ 45.0 ℃
Accuracy
Accuracy of skin temperature sensor
Humidity: control and measurement of relative humidity in the hood
Control system
Steam humidifier
Control range
40 %RH ~ 95 %RH
Measurement range
15 %RH ~ 99 %RH
Accuracy
Autoclavable water box
O2: measurement of O2 in the hood
Measurement range
18 % ~ 100 %
Accuracy
O2: servo control of O2 in the hood
Control range
21 % ~ 65 %
Accuracy
Max inlet pressure
120 psi
Hose diameter
8mm, 9.5mm, 11mm, 13mm (Use only oxygen hose)
Weighing scale: measurement of weight of the infant
Measurement range
0.000 Kg ~10.000 Kg
Accuracy
Pulse rate(PR): measurement of pulse rate of the infant
Measurement range
30 bpm ~ 240 bpm
Error tolerance
± 3 digit
Resolution
1 bpm
SpO2: measurement of SpO2 of the infant
Measurement range
1 % ~ 100 %
Accuracy (Without movement noise)
70 % ~ 100 %, ± 3 digit
0 % ~ 69 %, unspecified
Resolution
1 %
Power (AC)
Internal battery
Input
AC 100/240, 50/60Hz
Li-ion rechargeable battery (3.7V, 2600mAh)
Consumption
1200VA
5 hours to full charge when main power switch is on.
Alarm sounds for 30 minutes while no power
Operation environment
Storage environment
Temperature
20 ~ 30°C (68 ~ 86°F)
Temperature
–20 ~ 60°C (4 ~ 140°F)
Humidity
0 ~ 95% non-condensing
Humidity
0 ~ 95% non-condensing
Air pressure
70~106 kPa
Air pressure
70~106 kPa
Altitude
0 - 2,000m(0 – 6,561.68 ft)
Altitude
0 - 2,000m(0 – 6,561.68 ft)
Other environment
<0.5% at a point 15cm above from the center of the
mattress
Air velocity over the mattress
<0.3 m/s
℃
± 0.5 ℃
Normal mode: 35.0 ℃ ~ 37.5℃
± 0.5℃
± 0.3 ℃
± 5 %p
± 5 %p
± 5 %p
± 50 g
Maximum carbon dioxide(CO2) concentration
BT-500 Operation Manual 53
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Normal mode: 45 minutes, <55 dBA
Fast mode: 30 minutes, >66 dBA
Under 45 dBA in environmental sound pressure level
40 dBA
Standard
Complies with IEC60601-1, IEC60601-1-2, IEC60601-2-19
Class I equipment & Internal powered equipment
Continuous operation
Type BF applied part
Skin temperature probe: IPX1
Foot switch: IPX6
Alarm signal sound pressure
Priority
High (All)
Measured sound pressure level
73 dBA
A-weighted background level
48 dBA
Warm-up time
Sound pressure level in hood
SpO2 probe: IPX2
Silent mode: 60 minutes, <45 dBa
≤
BT-500 Operation Manual 54
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Alarm
Messages
Main heater overheated caused by
motor or thermostat malfunction
Check the fuse F1 of BD-500-MAIN.
If not, replace the humidity module.
Check the fuse F1 of BD-500-MAIN.
If not, replace the humidity module.
8Troubleshooting
8.1 General Checking
When the following problems are happened, before contact the head office, please check following
measures.
1. When you can’t power on the device.
• Please check that an adaptor of the device is connected to the AC connector properly.
• Please check the LED in the front panel is green or yellow.
2. When values are not displayed
• Please check that sensor and extension cables are properly connected to sensor connector.
• Please check that sensor works properly after connect cables.
3. When values are not in proper range
• Please check if the hood is closed well.
• Please check the probe is attached properly.
• When setting the scale to zero or measuring weight, please check if any object is placed on
the measuring plate.
If the unit has trouble, check the possible cause in sequence from above.
8.2 Alarm Message Checking
If the BT-500 has some problem for operation, it display alarm messages as below.
- Alarm messages on Main Display
Cause Solution
Motor Failed Motor driver malfunction Replace the main board (BD-500-MAIN)
Overburned
Heater
Humidity Heater
failed
Heater system
failed
Stuck key Key or knob is stuck.
Air sensors
failure
Sensor mod
disconnected
Sensor module
not in position
the broken impeller of Main fan
Water heater of humidity module
malfunction
Main heater of shell malfunction
Air temperature sensor
malfunction
Disconnected the sensor module Connect the sensor module to shell.
Sensor module not in position on
hood
Stop the operation of BT-500 and contact
to the distributor or BISTOS.
If the fuse F1 is broken, replace it.
If the fuse F1 is broken, replace it.
Check key and knob if they are stuck.
Replace U3 of BD-500-CPU CTRL.
Replace CIR-500-sensor sub or CIR-500sensor
Check the position of the sensor module.
BT-500 Operation Manual 55
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Skin Probe
Disconnect
Fan power
failure
In Skin mode, Skin sensor probe is
disconnected from connector. .
Fan power system malfunction Replace the main board (BD-500-MAIN)
- Alarm messages on Alarm Region of External Monitor
Alarm Messages Solution
Bad Sensor ID offset
Open LEDs
Short detector
Interference detected
Shorted LEDs
No adhesive sensor
Check the sensor status. If you are not able to comprehend the
causes, connect the service personnel.
No cable
No sensor connected
Connect the skin sensor probe, or change
the control mode to air mode.
Sensor off patient
Too much ambient light
Unrecognized sensor
Low SpO2
High pulse rate
Check the patient status.
Low pulse rate
- Error codes of External Monitor
If the external monitor displays alarm error codes such as board failure codes and diagnostic failure
codes as below. For this, connect the service personnel immediately.
Board Failure Codes Meaning
32 DSP: Checksum failure
33 DSP: Program memory test failure
34 DSP: Data Memory test failure
35 DSP: Detector ADC Interrupt Test Failure
36 DSP: MCU Interrupt failure
Diagnostic failure Code Meaning
37 DSP : Diag queue Overrun
38 DSP : Hardware Status Failure
39 DSP : Raw Queue Overrun
40 DSP: MCU Watch Dog Failure
BT-500 Operation Manual 56
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
0001 LED Ground
0002 Reference Voltage
0004 Digital Voltage
0008 DSP Voltage
0010 Positive LED Voltage
0020 Red current level
0040 IR Current level
0080 Digital ground
0100 Positive Preamp voltage
0200 Preamp
0400 Positive Detector voltage
0800 Negative Detector voltage
1000 LED current
2000 Analog ground
4000 LED drive voltage
8000 Sensor ID
BT-500 Operation Manual 57
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
The essential performance of the BT-500, Infant
The BT-500 is suitable for use in all establishments
9 Declaration on EMC
BT-500 needs special precautions regarding EMC (Electromagnetic compatibility) and needs to be
used according to the EMC information provided in this user manual. Wireless communications
equipment such as wireless home network devices, mobile phones, cordless telephones and their
base stations, walkie-talkies can affect the BT-500 and should be kept at least 1 m away from the
equipment.
9.1 Electromagnetic emissions
The BT-500 is intended for use in the electromagnetic environment specified below. The customer
or the user of the BT-500 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
incubator, is to provide an enclosed controlled
environment to maintain appropriate temperature
RF emissions
CISPR 11
Group 1
and humidity levels for infant. So there is no
intentional or controlled RF emission for it’s
intended performance. Therefore, its RF emissions
are very low and are not likely cause any interference
with nearby electronic equipment.
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Class A
Class A
Complies
other than domestic, and may be used in domestic
establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purpose.
Warning: This BT-500 is intended for use by healthcare
professionals only. This equipment/system may cause
radio interference or may disrupt the operation of
nearby equipment. It may be necessary to take
mitigation measures, such as re-orienting or relocating
the BT-500 or shielding the location.
BT-500 Operation Manual 58
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
The BT-500 is intended for use in an electromagnetic environment in which radiated RF
Separation distance according to frequency of transmitter
800 MHz to 2.5 GHz
0.01
0.1
1
10
100
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected
9.2 Recommended separation distances between portable and mobile RF
communications equipment and the BT-500
disturbances are controlled. The customer or the user of the BT-500 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the BT-500 as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
[W]
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
150 kHz to 80 MHz
0.12 0.12 0.23
0.38 0.38 0.73
1.2 1.2 2.3
3.8 3.8 7.3
12 12 23
80 MHz to 800 MHz
[m]
by absorption and reflection from structures, objects and people.
BT-500 Operation Manual 59
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
IEC 60601
Electromagnetic environment -
Electrostatic
±8 ㎸
±8 ㎸
Floors should be wood, concrete
Electrical fast
Mains power quality should be
Surge
230 V~ 50 ㎐ Power
±1 kV line(s) to line(s)
Mains power quality should be
Voltage dips, short
100V~ 50 ㎐,
100V~ 50 ㎐,
Mains power quality should be
Power frequency
30A/m
30A/m
Power frequency magnetic fields
9.3 Electromagnetic immunity
The BT-500 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BT-500 should assure that it is used in such an environment.
Immunity test
discharge (ESD)
IEC 61000-4-2
transient/burst
IEC 61000-4-4
IEC 61000-4-5
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Test level
Contact
±2㎸, ±4㎸,
±8㎸, ±15㎸
Air
230 V~ 50 ㎐ Power
supply line,
±2 ㎸ AC,
100 ㎑ PRR
supply line,
±0.5㎸,±1㎸
L1 to L2
240 V~ 50 ㎐
Power supply line
0 % UT for 0.5 cycle
0% UT for 1 cycle
50 ㎐ :
70 % UT for 25 cycles
60 ㎐ :
70 % UT for 30 cycles
50 ㎐ :
0 % UT for 250 cycles
60 ㎐ :
0 % UT for 300 cycles
50 ㎐ :
70 % UT for 25 cycles
60 ㎐ :
70 % UT for 30 cycles
50 ㎐ :
0 % UT for 250 cycles
60 ㎐ :
0 % UT for 300 cycles
guidance
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
that of a typical commercial or
hospital environment.
that of a typical commercial or
hospital environment.
that of a typical commercial or
hospital environment. If the
user of the BT-500 infant
incubator requires continued
operation during power mains
interruptions, it is
recommended that the BT-500
infant incubator be powered
from an uninterruptible power
supply.
(50 Hz and 60 Hz)
magnetic field
IEC 61000-4-8
NOTE Uт is the a.c. mains voltage prior to application of the test level.
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
BT-500 Operation Manual 60
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
IEC 60601 test
level
Compliance
level
Electromagnetic environment
- guidance
Conducted RF
230 V~ 50 ㎐
230 V~ 50 ㎐
Portable mobile RF communications
following symbol :
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.
Immunity test
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Power supply line
3 Vrms,
150㎑ -80㎒
6 Vrms in ISM
Bands between
0.15㎒ and 80 ㎒
3 V/m, 10 V/m
80 ㎒ to 2.7 G㎐
80%, 1 ㎑ AM
RF Wireless
Comm.
(Refer to test
report clause
1.15)
Power supply line
3 Vrms,
150㎑ -80㎒
6 Vrms in ISM
Bands between
0.15㎒ and 80
㎒
3 V/m, 10 V/m
80 ㎒ to 2.7 G㎐
80%, 1 ㎑ AM
RF Wireless
Comm.
(Refer to test
report clause
1.15)
equipment should be used no closer to
any part of the BT-550, including
cables, than the recommended
separation distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended separation distance
150㎑ -80㎒
80 ㎒ to 2.7 G㎐
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a,
should
be less than the compliance level in
each frequency range
b
.
Interference may occur in the vicinity
of equipment marked with the
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the BT- BT-500 is used exceeds the applicable RF compliance
level above, the BT-350 should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as re-orienting or relocating the BT-500.
b
BT-500 Operation Manual 61
P/N : 500-ENG-OPM-EUR-R12
Bistos Co., Ltd.
2019.07
Product Name
Infant Incubator
Model Name
BT-500
Serial No.
Warranty Period
2 Years
Date of Purchase
Hospital:
Manufacture
Bistos Co., Ltd.
Product Warranty
Address:
Customer
Sales Agency
※ Thank you for purchasing BT-500.
※ This product is manufactured and passed through strict quality control and inspection.
※ Compensation standard concerning repair, replacement, refund of the product complies with
“Framework Act on Consumers” noticed by Fair Trade Commission of Republic of Korea.
Name:
Telephone:
Service Telephone and Fax. Numbers
Telephone: +82 31 750 0340
Fax: +82 31 750 0344
th
7
FL., A Bldg., Woolim Lions Valley 5-cha, 302,
Galmachi-ro, Jungwon-gu, Seongnam-si,
Telephone: + (32) 2. 732.59.54
Bistos Co., Ltd.
Gyeonggi-do, Korea
www.bistos.co.kr
bistos@bistos.co.kr
Obelis s.a
Bd. Général Wahis 53
1030 Brussels, BELGIUM
Fax.: + (32) 2.732.60.03
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.