Beurer GL32 Instructions For Use Manual

Page 1
GL32
GB
BLOOD SUGAR
MEASURING DEVICE
Step by step
E-mail: kd@beurer.de www.beurer-medical.de
mg/dL
instructions for use
Page 2
Page 3
Beurer GL32 mg/dL
1
Contents: Blood Sugar Measuring Device GL32 mg/dL
1 About Your Beurer Blood Sugar Measuring Device ............2
1.1 Delivery scope, replacement parts and accessories .................................2
1.2 Functions of the device .............................................................................3
1.3 Signs and symbols ....................................................................................4
2 Safety information and warnings...........................................5
3 Description of Device and Accessories ................................8
3.1 Lancing device and needle lancets ...........................................................8
3.2 Blood sugar measuring device .................................................................. 8
3.3 Test strips................................................................................................11
4 Using the Device for the First Time and Basic Settings....13
4.1 Inserting and replacing the batteries .......................................................13
4.2 Basic settings ..........................................................................................13
5 Measuring Your Blood Sugar...............................................14
5.1 Obtaining a blood sample........................................................................15
5.2 Measuring the blood sugar value ............................................................18
5.3 Evaluating the measured value ............................................................... 19
5.4 Function test using control solution ......................................................... 21
6 The Measured Value Memory ...............................................24
6.1 Displaying individual values ....................................................................24
6.2 Displaying average values ......................................................................25
6.3 Deleting the measured value memory.....................................................26
6.4 Transferring measured values to a PC.................................................... 27
7 Storage and Maintenance .....................................................29
8 Troubleshooting ....................................................................29
9 Technical Details GL32 mg/dL .............................................31
10 Guarantee and Customer Service........................................33
Page 4
Beurer GL32 mg/dL
2
1 ABOUT YOUR BEURER BLOOD SUGAR MEASURING DEVICE
Read these instructions for use and all accompanying information material carefully and thoroughly before performing your first blood sugar measurement. Store the operating instructions in a safe place where they can be accessed by other users.
1.1 Delivery scope, replacement parts and accessories
Inspect the Beurer GL32 mg/dL measuring system to ensure that the packaging is undamaged and the contents are complete:
A 1 GL32 mg/dL blood sugar measuring device B 10 test strips C 1 lancing device with AST cap for taking blood samples from alternative sites D 10 sterile needle lancets E 1 practical carry case F 2 x 1.5 V AAA alkaline batteries (contained in the carry case) These instructions for use, additional information material
If the box has sustained extensive damage or if any contents are missing, please return the system to the retailer.
The blood sugar measuring device, test strips and control solution (available to purchase separately) are all designed to be used together. You should therefore only use the test strips and control solution intended for use with this device.
Note Only use the manufacturer's original accessories.
Page 5
Beurer GL32 mg/dL
3
Replacement parts and additional accessories
Test strips, control solution and lancets are also available for purchase without a prescription.
Item REF
50 test strips REF 464.00 MEDIUM control solution REF 457.02 100 needle lancets REF 457.01 Beurer GL32, GL34, BGL60 PC kit
(connecting cable, drivers and test software)
REF 462.10
1.2 Functions of the device
This device is intended for measuring the blood sugar content of human blood. It is suit­able for self-testing and does not have to be encoded. Using this measuring device, you can quickly and easily:
display and save your measured values (mg/dL).
display your average blood sugar measurements from the last 7, 14, 21, 28, 60 and 90
days.
set the time in 24-hour format and set the date (day and month).
upload your stored measured values to a PC for evaluation (additional accessories
required).
The measuring device also features the following control functions:
Warning if the temperature is unsuitable.
Battery change reminder when battery power is low.
Note
This device is only to be used for regular monitoring and not for the diagno­sis of diabetes.
Discuss your insulin dose with your doctor.
Page 6
Beurer GL32 mg/dL
4
1.3 Signs and symbols
The symbols on the packaging and the nameplate of the measuring device and accesso­ries have the following meaning:
In-vitro diagnostics
Manufacturer
Serial number
Refer to instructions for use
Temperature range +4°C to +40°C
Not for reuse/For single use only
PCT: Certification for products exported to the Russian Federation and CIS countries
Expiry date
Green Dot: German recycling system
Maximum shelf life after opening in months
Contents sufficient for <n> tests
Lot number
/Art.-Nr.
Order number
Sterilised by radiation (lancets)
Unit of measure for blood sugar value
Caution, refer to accompa­nying documents
Biohazard, risk of infec­tion
The following symbols are used in the instructions for use:
WARNING Warning: Risk of injury or danger to health.
CAUTION Safety notice referring to possible damage to the measuring device/accessories.
Note Important information to be noted.
Page 7
Beurer GL32 mg/dL
5
2 SAFETY INFORMATION AND WARNINGS
Risk of infection
All components of the measuring device and accessories may come into contact with human blood, and are therefore a possible source of infection.
WARNING
This measuring device must display the blood sugar content in mg/dL.The unit of measure mg/dL follows the blood sugar value. If your device does not display mg/dL, contact customer services im­mediately. You risk damaging your health if you perform a blood su­gar measurement using an unfamiliar unit of measure, since values may be interpreted incorrectly and cause the wrong corrective meas­ures to be taken.
This measuring device must only be used by one person. There is a risk of infection if the same device is used by more than one person.
The lancing device is intended for self-testing. Never share the lancing device or needle lancet with another person (risk of infection).
Use a new sterile needle lancet for each blood sample (for single use only).
Measuring
WARNING
The measured values you determine are useful for information purposes
only – they are not intended to replace consultation with your doctor. Dis­cuss your blood sugar values with your doctor on a regular basis. Do not change any aspects of your treatment unless instructed by your doctor.
Dehydration or loss of fluids, for example, through sweating, can lead to
incorrect results.
A very high or very low haematocrit value (proportion of red blood cells)
can lead to incorrect measurements. With a very high haematocrit value (over 60%), the displayed blood sugar value may be too low. If you have a very low haematocrit value (below 20%), the blood sugar value may be too high. If you do not know your haematocrit value, ask your doctor.
Do not use the test strips to take blood sugar measurements of newborn
infants.
Metabolites such as uric acid, ascorbic acid, acetaminophen (paracetamol),
dopa, methyldopa, L-dopa, and tolbutamide do not influence the results if they are present within the physiological value range.
Lipemia effects: Elevated blood triglyceride levels of up to 2000 mg/dL
rarely influence the results. Above this level, however, the blood sugar test may be affected.
Page 8
Beurer GL32 mg/dL
6
Only use fresh capillary whole blood. Do not use serum or plasma.
Use capillary blood obtained without squeezing the puncture site. When
squeezed, the blood is diluted with lymph from the tissues and therefore leads to an incorrect measurement result.
Do not use the test strips at an altitude above 3275 m.
Note
The Beurer GL32 mg/dL measuring system is designed for measuring capillary whole blood.
Storage and maintenance
WARNING
Store the measuring device and accessories out of the reach of young children. Small parts such as the needle lancets, batteries or test strips may be harmful if swallowed. If parts are swallowed, seek medical advice immediately.
The test strip container contains a desiccant that may cause skin or eye irritations if inhaled or swallowed. Keep the container out of reach of young children.
Batteries/Saving measured values
WARNING
Batteries may be fatal if swallowed. Therefore, always store batteries out of the reach of young children. If a battery is swallowed, seek medical advice immediately.
Do not throw batteries into a fire: Risk of explosion.
CAUTION
Remove the batteries if they have run out or if the device is not used for a long period of time. This prevents any damage that may be caused by leaking batteries.
Do not charge the batteries or reactivate them using other means. Do not disassemble or short-circuit the batteries.
Do not use rechargeable batteries.
Page 9
Beurer GL32 mg/dL
7
Note
The stored blood sugar values remain saved during a battery change. The date and time are retained during battery replacement and if the batteries are run down.
Each time you change the batteries, use two batteries of the same type, the same brand and the same capacity. Alkaline batteries should be used wherever possible.
Repairs
Note
Never open the device. Opening the device invalidates the guarantee.
Do not perform any repairs on the device yourself, otherwise correct func-
tioning of the device can no longer be guaranteed.
If repairs are required, please contact customer service.
Disposal
WARNING
When disposing of materials from the measuring device, always comply with the generally applicable guidelines for handling blood. Carefully dis­pose of all blood samples and materials with which you have come into contact, in order to prevent injury and infection of others.
After use, dispose of the test strips and lancets in a sharp-proof container.
Note
Used and fully discharged batteries should be disposed of in designated collection containers, at a suitable recycling facility, or via an electrical re­tailer. You are legally required to dispose of batteries properly. The following symbols are found on batteries that contain harmful sub­stances:
Pb = Battery contains lead Cd = Battery contains cadmium Hg = Battery contains mercury.
Please dispose of the device in accordance with EU directive 2002/96/EC – WEEE (Waste Electrical and Electronic Equip­ment). If you have any questions, contact the local authority re­sponsible for waste disposal.
Page 10
Beurer GL32 mg/dL
8
3 DESCRIPTION OF DEVICE AND ACCESSORIES
3.1 Lancing device and needle lancets
1
2
3
4
5
6
1: Cap
2: Protective disk for the lancet
3: Sterile needle lancet
4: Lancet holder
5: Release button
6: Tensioning device
3.2 Blood sugar measuring device
Front view Rear view
1: Test strip slot
2: LCD display
3: Back button
4: ON/OFF/Forwards button
5: PC interface with cover
1: Two alkaline batteries of type
1.5 V AAA
2: "SET" button for making settings
Page 11
Beurer GL32 mg/dL
9
Display symbols for the blood sugar measurement:
1: Code symbol (not functional)
2: Arrow symbol for test strips
3: Value display
4: Low blood sugar value (not functional)
5: High blood sugar value (not functional)
6: Test strip symbol
7: Speaker (not functional)
8: Blood droplet symbol
6
9: Blood sugar unit mg/dL
10: Blood sugar unit mmol/L (not functional)
11: Ketone measurement recommended (not functional)
12: Memory deleted symbol
13: Memory symbol
Page 12
Beurer GL32 mg/dL
10
Other display symbols:
14: Temperature symbol
15: Date: Day-Month
16: Time: Hours-minutes
17: Battery change symbol
18: CTL for control solution
19: DAY AVG for average value over 7, 14, 21, 28, 60 or 90 days
Note
This blood sugar measuring device is supplied with the blood sugar unit mg/dL.
Page 13
Beurer GL32 mg/dL
11
3.3 Test strips
Front view Rear view
1: Slit for blood collection
2: Hold here
3: Contacts
The rear side of the test strip is marked with an 'X'.
Insert the test strip into the device so that the contacts are visible in the slit. Ensure that the front side of the test strip is facing you.
Read the following information about handling and storing your test strips carefully. The test strips can only be guaranteed to return accurate results if you follow all instructions carefully.
WARNING Each test strip must only be used once and only for one patient.
Page 14
Beurer GL32 mg/dL
12
Handling test strips
Note
Before using the first test strip, ensure that the seal on the container is unbroken. If the seal is damaged, do not use the test strips.
Once you have removed the test strip, tightly close the container again immediately.
Do not use test strips if the expiry date has passed. The use of expired test strips can lead to inaccurate measurements. The expiry date is located on the container next to the egg timer symbol
.
The test strips can be stored for three months once the container is ope­ned. Make a note of when this time will run out (date of opening + 3 months) on the label
. The shelf life decreases if the expiry date is rea­ched before the end of the 3 months (see the date next to the egg timer symbol
).
Do not use the test strips if either of these dates has expired (
/ ).
Any part of the test strip may be handled with dry, clean hands.
Use the test strips immediately upon removal from the container.
Do not bend, cut, or otherwise modify the test strip.
Do not use test strips that have come into contact with liquids for the blood
sugar measurement.
Storing your test strips
Note
Store test strips in a cool, dry place between +4°C and 40°C. Protect test strips against direct sunlight or heat. Do not store test strips in the fridge.
Test strips may be stored at a relative humidity between 10 % and 85 %.
Only store test strips in their original container - never use another con-
tainer.
Page 15
Beurer GL32 mg/dL
13
4 USING THE DEVICE FOR THE FIRST TIME AND BASIC SETTINGS
4.1 Inserting and replacing the batteries
Note
Two batteries are included in the delivery scope of your blood-sugar measur­ing device. The batteries are stored inside the case.
1
Remove the lid of the battery compartment on the rear of the device (see "3.2 Blood sugar measuring device", page 8).
2
If you are replacing the batteries, first remove both the old batteries. The device retains the date and time when the batteries are replaced. If the replacement takes a long time, you may have to adjust the date and time afterwards (see "4.2 Basic settings", page 13 ).
3
Insert two new batteries of type alkaline AAA 1.5 V. It is important to ensure that the batteries are inserted with the correct poles according to the diagram. Do not use rechargeable batteries.
4
Carefully replace the lid on the battery compartment.
When the battery change symbol
appears, the battery is almost empty. Replace
both batteries as soon as possible. The screen displays E-b if the batteries are so flat that no further measurements are possible.
4.2 Basic settings
1
The measuring device must be switched off. Remove the lid of the battery compartment on the rear of the device (see "3.2 Blood sugar measuring device", page 8).
2
Press the "SET" button. The year display flashes.
Page 16
Beurer GL32 mg/dL
14
3
Setting the date and time
Note
The date and time must always be set. Otherwise, you cannot accurately save measured values with a date and time or call them up later.
The time is displayed in 24-hour format, for exam­ple: 20:32.
Set the year (the calendar runs until 2049) by pressing the ON/OFF/Forwards button
. Press "SET" to confirm the year. The month display flashes. Repeat the above procedure for the month, day, hours, and minutes.
4
"dEL" is displayed and the memory symbol flashes. (Caution: All stored measured values will be deleted, if you now press the ON/OFF/Forwards button
twice. For more informa­tion, see "6.3 Deleting the measured value memory" on page 26.) To finish making the settings without deleting the measured values, press the "SET" button. "OFF" is displayed briefly and the measuring device switches off automatically. Carefully replace the lid on the battery compartment.
5 MEASURING YOUR BLOOD SUGAR
WARNING
If you drop the lancing device while a needle lancet is inserted, pick it up carefully and dispose of the lancet.
CAUTION
Only use the lancing device with needle lancets made by the same manu­facturer. The use of other needle lancets may impair the function of the lancing device.
If the lancing device is manufactured by a third party, consult their instruc­tions for use.
Page 17
Beurer GL32 mg/dL
15
5.1 Obtaining a blood sample
Preparing to take the blood sample
1
Choose a site on the body from where you wish to take the blood sample. You can use the lancing device to take a blood sample from the fingertip or alterna­tive sites such as the palm of the hand, forearm, upper arm, thigh, or calf. We recommend that you take the blood sample from the fingertip. To make the blood sample as pain-free as possible, do not take the blood directly from the centre of the fingertip, but instead from slightly to the side of the centre.
WARNING
If you suspect hypoglycaemia: Always take blood from the fingertip. This is because changes in blood sugar levels are mo-
re rapidly detected in blood from the fingertip.
Measurement at the fingertip and measurement at other body sites (AST) can lead to considerably different measured values. Always consult your doctor before starting to test at alternative sites.
2
Have the following components ready: Measuring device, container of test strips, lancing device, and a sterile needle lancet. If obtaining a blood sample from a site other than the fingertip, you will also need the AST cap.
3
Before taking the sample, wash your hands using soap and warm water. In addi­tion to optimal hygienic conditions, this is also ensures good circulation of blood through the fingertips. Dry your hands carefully. Also ensure that your lancing site is hygienically clean if taking a blood sample from an alternative site (AST).
WARNING If you wipe the lancing site with alcohol, ensure that the site has
dried completely before beginning the measurement.
Taking the blood sample
WARNING
Change the lancing site for every test, e.g. use a different finger or the other hand. Repeated lancing in the same location may lead to infections, loss of sensation, or scarring.
Do not use the AST cap to take a blood sample from the finger.
Never squeeze the finger to obtain a larger droplet of blood. When squee-
zed, the blood is diluted with lymph from the tissues, which may lead to an incorrect measurement result.
Note that inadequate circulation at the puncture site, for example caused by cold or illness, can lead to inaccurate measurements.
Page 18
Beurer GL32 mg/dL
16
1
Twist to remove the cap from the lancing device.
2
Place a sterile needle lancet in the lancing device and push the lancet firmly into place.
3
Remove the protective disc from the lancet by twisting while holding the shaft of the lancet firmly. Keep the protective disc for safe disposal of the used needle lancet after you have obtained the blood sample.
4
You will need to use a different cap depending on the site from which you want to take the sample:
Fingertip: Cap (grey) Other body sites: AST cap (transparent)
Place the selected cap on the lancing device and twist it into place.
5
Five different puncture depths can be selected on the lancing device:
1 to 2: Delicate or thin skin
3: Normal skin
4 to 5: Thick or rough skin
Turn the cap in the appropriate direction until the arrow points to the required puncture setting.
6
Pull back the tensioning device until you hear it click into place. If it does not engage, the lancing device may have unintentionally already been tensioned when inserting the needle lancet and could already be engaged.
Page 19
Beurer GL32 mg/dL
17
7
The lancing device can now be used to take the blood sample. Ensure that the blood remains in a droplet form and is not smeared.
Blood sample from the fingertip Place the lancing device firmly against the finger, slightly to the side of the centre of the fingertip. Press the release button. Lift the lancing device from the finger. A round blood droplet of at least 0.5 microlitres (approx. 1.2 mm, original size: z) is required.
Blood sample from an alternative site (AST) Choose a site that is soft, not next to a bone, with no visible veins and without excessive body hair. Warm the puncture site to ensure good circulation, for exam­ple by lightly massaging the area. Place and hold the lancing device against the puncture site for a few seconds and press the release button. Continue to hold the lancing device against your skin until a round droplet of blood has formed under the cap. Maintain the pressure until the blood droplet has reached a size of at least 0.5 microlitres (approx. 1.2 mm, original size: z). Lift the lancing device carefully away from the skin.
8
If insufficient blood was collected, increase the punc­ture depth and repeat steps 5 to 7.
9
Untwist the cap and carefully remove it from the lanc­ing device.
10
Place the protective disk you kept earlier flat on a hard surface. Pierce the protective disk with the tip of the needle so that the needle is covered.
Page 20
Beurer GL32 mg/dL
18
11
Carefully remove the needle lancet from the lancing device and discard the lancet in a suitable sharp-proof container. Carefully dispose of all blood samples and materials with which you have come into contact. This prevents injury and prevents infection spreading to other people.
12
Twist the (grey) cap back in place.
5.2 Measuring the blood sugar value
1
Hold the measuring device so that the display is facing you.
2
Insert a test strip, contacts first, into the device. Ensure that the front side is facing you. You can handle any part of the test strip with clean, dry hands.
3
The device switches on automatically and shows the start display, shortly followed by the checking symbol “
“ . As soon as the flashing blood droplet
symbol 6 is displayed, the device is ready to perform measurements. Hold the blood collection slit (at the point of the test
strip) against the blood droplet. Do not press the puncture site (fingertip or other body part) against the test strip. The blood must not be smeared. The blood is sucked into the slit.
4
WARNING
Hold the blood collection slit of the test strip against the blood droplet until the slit is completely filled and you hear a beep. Withdrawing the test strip from the blood droplet before the beep may lead to incorrect measurements.
5
Once the slit is filled with blood, the device performs the blood sugar meas­urement. The measuring device counts down for 6 seconds, and the result of the measurement is then shown in the display. Read the measured value. For an explanation of the measured value and possible actions, see the next chapter "5.3 Evaluating the measured value", page 19. If an error message is displayed, refer to chapter “8 Troubleshooting" on page 29.
Page 21
Beurer GL32 mg/dL
19
6
Remove the test strip from the device and dispose of it carefully in accor­dance with the applicable guidelines in order to prevent infection of other people.
Note
If the device does not begin the measurement, do not subsequently apply
more blood. Instead, remove the test strip and end the test procedure. Start again using a new test strip.
If the test strip is already inserted in the device and you do not apply any blood to the test strip within three minutes, the device switches itself off. In this case, remove the test strip and reinsert it into the slit. The device will switch on again automatically.
If you experience difficulties applying blood to the test strip correctly, con­tact customer service.
5.3 Evaluating the measured value
Your blood sugar measuring device can process values between 20 and 600 mg/dL. The warning message "Lo" is displayed for measured values below 20 mg/dL. The warning message "Hi" is displayed for measured values above 600 mg/dL.
Note If you suspect that the blood sugar values are incorrect, first repeat the test
and, if necessary, perform a function test using the control solution. If you consistently have reason to doubt the results, contact your doctor.
The following table shows a classification of blood sugar values according to the guide­lines of the German Diabetes Association (DDG).
Time of blood sugar measurement
Normal blood sugar values
Suspected diabetes
Diabetes
Fasting
Below 100 mg/dL
100 to 110 mg/dL
Above 110 mg/dL
2 hours after eating
Below 140 mg/dL
140 to 200 mg/dL
Above 200 mg/dL
Source: Deutsche Diabetes Gesellschaft (DDG) 2008
Page 22
Beurer GL32 mg/dL
20
Critical blood sugar values
Display Blood sugar Action
Lo = Severe hypoglycaemia (low blood sugar)
Below 20 mg/dL
Immediate medical attention requi­red.
Low
Below 70 mg/dL
Eat a suitable snack or small meal. Follow your doctor's advice.
High
Fasting: Above 100 mg/dL
2 hours after eating: Above 140 mg/dL
If this high value is still present 2 hours after the last meal, this may indicate hyperglycaemia (high blood sugar). Discuss with your doctor any possible action to take in this case.
High, possible ketones
Above 240 mg/dL
Ketone test recommended. Consult your doctor.
Hi = Severe hyperglycaemia (very high blood sugar).
Above 600 mg/dL
Repeat the measurement with a new test strip. If the display shows the same result again: Seek medi­cal advice immediately.
Page 23
Beurer GL32 mg/dL
21
5.4 Function test using control solution
The control solution is used for checking the whole blood sugar measuring system. This enables you to determine whether the measuring device and the test strips are functioning correctly together and whether the test is performed correctly. It is very important that you perform these tests using the control solution to ensure accurate measuring results.
CAUTION Never use control solutions made by other manufacturers. The correct func­tioning of your Beurer blood sugar measuring device can only be checked using the Beurer MEDIUM control solution.
When is it recommended to perform a test using the control solution?
After opening a new test strip container.
If you suspect that the measuring device or the test strips may be defective.
If your blood sugar measurement values do not correspond to the way you are feeling..
If you have dropped the measuring device or it has been subjected to other mechani-
cal strain.
Every time you suspect that the blood sugar results may be incorrect. It is sufficient to perform a single control solution test if the result is within the recom­mended range.
WARNING Do not ingest the control solution. The control solution is only to be used for
function tests and is for external use only.
Note
Store the control solution in a tightly sealed container at room temperature below 30°C. Do not refrigerate.
Store the control solution out of reach of young children.
Do not store the control solution in the medicine cabinet. It may be mis-
taken for medicine to be consumed.
Preparations
Do not use the control solution once the expiry date has passed. Using control solution that is out-of-date may lead to incorrect function test results. The expiry date is located on the bottle next to the egg timer symbol
.
The control solution can be kept for three months once the bottle is open. Make a note of when this time will run out (date of opening + 3 months) on the label
. The shelf life decreases if the expiry date is reached before the end of the 3 months (see the date next to the egg timer symbol
).
Page 24
Beurer GL32 mg/dL
22
Do not use the control solution if either of these dates has expired (
/ ).
Allow the measuring device, test strips and control solution to reach room temperature (+20°C to +25°C).
Performing a function test using the control solution
1
Hold the measuring device so that the display is facing you.
2
Insert a test strip into the slit on the measuring device, contacts first. Ensure that the front side of the test strip is facing towards you (see "3.3 Test strip", page 11).
3
The device switches on automatically and shows the start display, shortly followed by the checking symbol “
“. As soon as the flashing blood droplet
symbol 6 is displayed, the device is ready to perform measurements.
4
Press the ON/OFF/Forwards button to switch to control mode. In control mode, the measured value is not stored and hence does not distort your measurement statistics.
5
"CTL" appears in the display.
6
Shake the control solution well before use. Unscrew the cap and squeeze out a drop of solution. Wipe away the first drop and squeeze out an additional drop.
7
To prevent the remaining control solution in the bottle from becoming con­taminated through the tip of the bottle in contact with the test strip, do not apply the drop directly to the test strip. Instead, apply the drop to a clean surface, and then apply the drop from there to the blood collection slit on the test strip. The solution is sucked into the slit. Wipe the tip of the bottle using a clean, dry paper tissue.
8
Once the slit is filled with solution, the device begins the measurement. The device counts down for 6 seconds, and the result of the measurement is then shown in the display.
9
Check whether the result is within the specified results range for the control solution. This results range is printed on the test strip container.
Page 25
Beurer GL32 mg/dL
23
Expected results
At room temperature, the measurement results for 95% of all control solution tests should lie within the results range printed on the test strip container.
WARNING The results range printed on the test strip container applies only for the control
solution. This is not a recommended value for your blood sugar level.
If measurement results are outside the specified range, check the following possible causes:
Cause Remedy
The drop of control solution was not discarded.
The tip of the bottle was not wiped clean.
The bottle was not shaken vigourously enough.
Eliminate the cause and repeat the test.
Control solution has expired or is con­taminated.
Repeat the test using a new bottle of control solution.
The control solution, test strip, or measur­ing device is too warm or too cold.
Allow the measuring device, test strips and control solution to reach room temperature (+20°C to +25°C) and
repeat the test. Test strip is damaged. Repeat the test using a new test strip. Test strip is out of date. Open a new container of test strips.
Repeat the test. There is a problem with the measuring
device.
Contact customer service.
WARNING If you persistently obtain results outside of the specified range when testing
with the control solution, do not continue using the system to determine your blood sugar content. Contact customer service.
Page 26
Beurer GL32 mg/dL
24
6 THE MEASURED VALUE MEMORY
Your blood sugar value is stored automatically each time you take a measurement. Ex­ception: if "CTL" is activated during a measurement with control solution, the value is not stored. The measured value memory can store up to a maximum of 448 values. Once this number is reached, the measured value always replaces the oldest value in the memory. You can display the average value for the last 7, 14, 21, 28, 60 and 90 days, as well as displaying each value individually.
Note
If measured values are already stored in the memory and you then reset the date, the average values are calculated according to the new time pe­riod.
"---" indicates that the measured value memory is empty. "OFF" is then displayed briefly and the measuring device switches off automatically.
6.1 Displaying individual values
You can display individual values for the last 448 measurements. The most recent value is displayed first, and the oldest value is displayed last. The measuring device also dis­plays the date and time of the measurement. The number of the measurement is dis­played briefly before each measured value.
1
The measuring device must be switched off. Press the ON/OFF/Forwards button
.
2
The start display appears briefly. The device then displays the date and time, and the memory symbol
flashes. Press the
ON/OFF/Forwards button again
.
Page 27
Beurer GL32 mg/dL
25
3
The number of the measurement flashes briefly (fig. 1) and the most recent value is displayed (fig. 2). Each time you press the ON/OFF/Forwards button
, the previous measured value is displayed. You can display a maximum of 448 previous measured values. After the last available individual value, the device displays "End" followed by "OFF".
Note
You can use the back button to return to meas- ured values that have already been displayed.
You can cancel the process at any time. To exit the process, press and hold the ON/OFF/Forwards button
for three seconds.
The device displays "End" followed by "OFF".
Fig. 1
Fig. 2
4
The measuring device switches off automatically.
6.2 Displaying average values
You can display the average blood sugar value from the last 7, 14, 21, 28, 60 and 90 days. The device also displays how often you measured your blood sugar during the period selected.
1
The measuring device must be switched off. Press and hold the ON/OFF/Forwards button
for three seconds.
2
The start display appears briefly. The device then displays the date and time, and the memory symbol
. Once the symbols "DAY" and "AVG" are flashing, you can release the ON/OFF/Forwards but­ton
.
Page 28
Beurer GL32 mg/dL
26
3
The average value for 7 days is displayed, together with the num­ber of measured values recorded within this period (in the diagram: 14 measured values within the last 7 days).
Press the
button repeatedly to display the average value for 14, 21, 28, 60 and 90 days and the corresponding number of measurements. After the last available average value, the device displays "End" followed by "OFF".
Note
You can cancel the process at any time. To exit the process, press and hold the ON/OFF/Forwards button
for three seconds. The device displays "End"
followed by "OFF".
4
The measuring device switches off automatically.
6.3 Deleting the measured value memory
1
The measuring device must be switched off. Remove the lid of the battery compartment on the rear of the measuring device.
2
Press the "SET" button slowly six times.
3
"dEL" is displayed and the memory symbol flashes. Press the ON/OFF/Forwards button
.
4
The "dEL" symbol is now also flashing. To delete the measured value memory, press the ON/OFF/Forwards button
.
Note
To switch off the device without deleting the meas­ured values, press the "SET" button.
5
"---" is displayed briefly, followed by the symbol.
6
Press the ON/OFF/Forwards button . "OFF" is displayed briefly and the measuring device switches off. Carefully replace the lid on the battery compartment.
Page 29
Beurer GL32 mg/dL
27
6.4 Transferring measured values to a PC
The Beurer GL32 mg/dL measuring system features a PC interface that you can use to upload your stored blood sugar measurement values to a PC. Using special software, you can evaluate your measured values on the PC and use the evaluations for monitoring your blood sugar levels.
The connection cable is available from your retailer as the accessory set "Beurer GL32, GL34, BGL60 PC kit" (see "1.1 Delivery scope, replacement parts and accessories", page 2). The connection cable is supplied together with a CD containing the test software for evaluating your measurement results, which enables you and your doctor to better moni­tor your blood sugar levels. For more information, please refer to the instructions included with the accessory set. These contain all the information you require for the data transfer.
CAUTION Only use the original data connection cable supplied by Beurer for data trans-
fer. Otherwise, your measuring device or PC may be damaged.
Preparations
Place the blood sugar measuring device and the Beurer GL32, GL34, BGL60 PC kit next to the PC ready for use.
Install the evaluation
software on your PC as described in the instructions for the
Beurer GL32, GL34, BGL60 PC kit.
Transferring your measurement data
1
Remove the rubber cover from the PC interface on the blood sugar measuring device.
2
The measuring device must be switched off. Insert the flat USB connector of the connection cable into a USB port on your PC. Insert the round jack plug into the PC interface port on your measuring device.
3
"PC" appears in the display of the measuring device. The measuring device is now ready for data transfer.
4
Follow the information on data transfer and evaluation in the software and in the operating instructions for the Beurer GL32, GL34, BGL60 PC kit.
Page 30
Beurer GL32 mg/dL
28
Note
You cannot perform blood sugar measurements while the blood sugar measuring device is connected to a PC.
Effective evaluation of the measured values is only possible if you have set the date and time correctly (see "4.2 Basic settings", page 13).
The measured values remain stored on the measuring device after transfer to the PC.
Disconnecting the device from the PC
1
Pull the jack plug from the blood sugar measuring device.
2
The measuring device switches off automatically.
Page 31
Beurer GL32 mg/dL
29
7 STORAGE AND MAINTENANCE
Storing your device
After use, store the measuring device in the supplied case.
Note
Keep these instructions for use in a safe place.
If you do not use the device for a prolonged period of time, remove the
batteries.
Maintenance
Clean the surface of the device using a damp cloth (water or mild detergent). Dry the device using a lint-free cloth.
Note The measuring device is made from precision components. The accuracy of the measured values and the service life of the device depend on careful handling:
Protect the device against impact and do not drop it.
Protect the device against damaging influences such as moisture, dirt, dust,
blood, control solution or water, extremes of temperature and direct sunlight.
Do not store close to strong magnetic fields, wireless systems or mobile telephones.
8 TROUBLESHOOTING
Problem: Unexpected message on the display
No. Cause Action
E-b Flat batteries Replace all batteries. E-F Test strip was removed from the
device during measurement
Repeat the blood sugar measurement using a new test strip
E-t Temperature of the measuring
environment, measuring device or test strip was outside the permissible range
Repeat the test with a new test strip once the measuring environment, measuring device, and test strip have reached room temperature (+20°C to +25°C).
A used or contaminated test strip was inserted
Insert an unused, clean test strip and repeat the measurement
E-U
Blood was applied to the test strip too early
Repeat the test using a new test strip. Do not apply the blood until the blood droplet symbol 6 flashes in the display.
Page 32
Beurer GL32 mg/dL
30
Problem: Device does not turn on.
Cause Action
Flat batteries Replace the batteries. Battery missing or incorrectly inserted Check whether the batteries are inserted
correctly (see "4.1 Inserting and replacing the batteries", page 13).
Test strip was not fully inserted or inserted on the wrong side
Insert the test strip, contacts first, into the slit on the device. Ensure that the front side of the test strip is facing towards you (see "3.3 Test strip", page 11).
Device is defective Contact customer service.
Problem: The test does not start after you insert the test strip in the device and apply blood.
Cause Action
Insufficient quantity of blood Repeat the test with a new test strip and
a larger blood droplet. Test strip is defective Repeat the test using a new test strip. Blood was applied while the device was
turned off
Repeat the test and apply the blood
when 6 is flashing. The basic settings of the device have
been changed and the change was not completed (see "4.2 Basic settings", page 13).
Remove the lid of the battery compartment
and press the "SET" button repeatedly
until "OFF" is displayed. Close the battery
compartment and repeat the test. Device is defective Contact customer service.
Note If the problem persists, contact customer service.
Page 33
Beurer GL32 mg/dL
31
9 TECHNICAL DETAILS GL32 mg/dL
Dimensions 96 x 46 x 20 mm Weight 69 g (including batteries) Electrical power 2 x 1.5 V AAA alkaline batteries Battery life More than 1000 measurements
Measured value memory
448 measured values with date/time Forwards and backwards scrolling through the memory Data retained during battery replacement
Average values For 7, 14, 21, 28, 60, 90 days Automatic switch-off 3 minutes after the last operation
Storage/transport temperature
Temperature: +4°C – +40°C Relative humidity: 10 – 85%
Operating ranges
Temperature: +10°C – +40°C Relative humidity: 10 – 85% (non-condensing)
Glucose measure­ment range
20 – 600 mg/dL
Coding No encoding of the device is required ("codefree") Blood sample Capillary whole blood Blood quantity 0.5 microlitres
Measurement duration
Approx. 6 seconds
Calibration Plasma Test method Amperometric biosensor Application Suitable for self-testing System function test Each time the device is switched on
Page 34
Beurer GL32 mg/dL
32
EMC
This device complies with European standard EN 61326 and is subject to specific preven­tive measures in terms of electromagnetic compatibility. Note that portable and mobile high-frequency communications equipment may interfere with this device. Further details can be requested by contacting the customer service address provided below.
Note regarding function of test strips
The test strips enable a quantitative measurement of the glucose content of capillary whole blood. When the slit for collecting blood comes into contact with a drop of blood, it fills automatically through simple capillary action. The blood is sucked into the absorbent slit of the test strip and the measuring device measures the sugar level in the blood. The test is based on the measurement of an electric current caused by a chemical reac­tion between the glucose and the test strip reagent. The device analyses this current. The strength of the current varies according to the glucose content in the blood sample. The results are shown in the display of the measuring device. Only a small quantity of blood is required (0.5 microlitres) and the measurement takes approx. 6 seconds. The strips can detect blood sugar levels in the range 20 to 600 mg/dL.
Chemical components of the test strip sensor
Glucose oxidase 10%
Electron shuttle 50%
Enzyme protection 8%
Non-reactive components 32%
Note regarding function of the control solution
The control solution contains a fixed concentration of glucose that reacts with the test strip. A test with the control solution is performed in the same way as a blood test, but the control solution is used instead of blood. The results of the control solution test must lie within the specified results range. This results range is printed on every test strip con­tainer.
Chemical composition of the control solution
The control solution is a red-coloured solution with a D-glucose content of less than 0.2%.
Ingredient Percentage concentration D-glucose 0.05 0.19% Salts 1.4% Active ingredient for viscosity regulation 15.0%
Standards
The Beurer GL32 mg/dL measuring system for self-testing complies with the following European Directives and standards: IVD (98/79/EC), EN 61010-1, EN 61010-2-101, EN 13640, EN ISO 15197, MDD (93/42/EC).
Page 35
Beurer GL32 mg/dL
33
Comparison between measured values and laboratory values
Performance characteristics: Accuracy and precision The blood sugar results were compared with the YSI 2300 laboratory device. At a concen­tration of < 75 mg/dL, 95% of values were within +/- 15 mg/dL, while at a sugar concen­tration 75 mg/dL,
> 98% were within 20% of the reference values. The CV (coefficient of
variation) (%) is < 5 %. The blood sugar measuring device is therefore comparable with a laboratory system.
10 GUARANTEE AND CUSTOMER SERVICE
Guarantee
This product comes with a 3-year guarantee for material and manufacturing faults. The guarantee does not apply:
in the case of damage caused by improper use
to wearing parts
to deficiencies of which the customer was aware at the time of purchase
to personal negligence on the part of the customer
in the case of third-party intervention
This guarantee does not affect your statutory rights. In order to make a claim within the warranty period, the customer is required to provide proof of purchase. Claims must be made within a period of 3 years from the date of purchase to BEURER GmbH, Söflinger Str. 218, 89077 Ulm, Germany. In the case of claims against the guarantee, the customer has the right to have the product repaired by us or in a workshop authorised by us. Fur­ther rights (of the guarantee) remain unaffected.
Customer service address
If you have any questions, please contact customer service:
Lifestyle Marketing International Ltd. P. O. Box 584 WN1 9EX WIGAN Phone: +44 870 879 0812 E-mail: customerservice@lifestylemi.com
Brandlinx Direct Limited Hainault House Baldonnell Business Park Dublin 22 Phone: +353 1 412 3606 E-mail: sales@brandlinx.ie
OUR COMMITMENT TO YOU: We aim to satisfy our customers by providing high-quality healthcare products and the best customer service. If you are not completely satisfied with this product, please contact customer service.