The vb5 Portable Data Collector instrument is a portable single channel vibration data collector. It is ergonomically designed and lightweight for all day comfort. You can use this device for on-route as well as off-route data collection.
The vb5 Portable Data Collector provides recordings with up to 6,400 lines of resolution and up to 40 kHz F patented adaptive settling algorithm and 6Pack recording system offer quick, one-step data recording.
The data collector has plenty of storage and long battery life, and is backed by a five year warranty.
The vb5 Portable Data Collector is one of Bently Nevada hardware monitoring assets that work with System 1 Evolution software.
The vb5 Portable Data Collector offers the following features:
111M6572 Rev. F
max
l Single channel recordings l 6,400lines FFTresolution l Supports 40 kHz F
l 1 GB memory l ≥ 95 dB dynamic range l Spectrum and waveform recordings l Demodulation for early detection of rotating machinery
max
problems such as bearing faults
l Unique 6Pack recording system l Cable test mode l Option to add flex features such as balancing and
Remote Comms
l Upgradable Proflash system and free firmware updates
for 5 years
l Five-year warranty on the instrument hardware
vb5 Portable Data Collector Datasheet 111M6572 Rev. F

Specifications

Sensors
Sensor Input One channel Compatible
Sensor Types AC Coupled
Range
Connectors
Analog to Digital Conversion
Sensor Excitation Current
Sensor Detection
Tachometer Sensor (Optional)
Sensor Type
Laser Sensor Range
Tachometer Input
Supported Sensor Types
Power Supply to Sensor
TTL PulseRating
Speed Range
Accuracy ± 0.1 %
Accelerometer
16 V peak-peak Allows for ± 8 V sensor output swing (± 80 g)
1 x BNC (CH1)
Safety feature:Break-free inline connector
24-bit ADC
0 mA or 2.2 mA (configurable), 24 V maximum
2.2 mA required power for IEPE/ICP type accelerometer
Warns if short circuit or not connected
Laser sensor with reflective tape Sensor triggers on beam reflection
10 cm to 2 m nominal Range depends on size of reflective tape
Laser Tach, Contact, TTL Pulse Instrument has optically isolated input
5 V, 50 mA
3.5 V (4 mA) min 28 V (5 mA) max Off-state 0.8 V
10 RPM to 300,000 RPM (0.2 Hz to 5 kHz)
Pulse width at least 0.1 ms
Parameter Indication
> 1000 g (10,000 m/s2) > 1000
Maximum Levels (peak)
Dynamic Signal Range
Harmonic Distortion
Units
Magnitude and Cursors
Base Accuracy
High Frequency Attenuation
AC Coupling Attenuation
Attenuation Due to Integration
in/sec (25,000 mm/s) > 20 in (500 mm)
Effective limit is sensor sensitivity and output voltage
> 95 dB
typical at 400 line resolution Less than -70 dB typical
Other distortions and noise are lower
g or m/s2 or adB in/s or mm/s or vdB mil or mm or µm
0-peak, peak-peak or RMS Auto­scale by 1000x when required
US and SI options for amps and volts
Overall RMS value Waveform True pk-pk Dual cursors Harmonics
Digital readouts on chart ± 1% of readings approximately
0.1 dB
% of reading ≤ 0.1 dB 100 Hz to 10 kHz ≤ 3 dB >10
kHz to 40 kHz
Attenuation tolerances are in addition to base accuracy.
≤ 0.1 dB 10 Hz to <100 Hz ≤ 3 dB 1 Hz to <10 Hz
≤ 0.1 dB 1 Hz to <100 Hz ≤ 1.5 dB 0.2 Hz to <1 Hz
Values apply to single integration. (Acceleration to velocity)
Double the values for double integration (Acceleration to displacement)