Bacharach, Inc. warrants to Buyer that at the time of delivery
this Product will be free from defects in material and manufacture and will conform substantially to Bacharach Inc.’s applicable specifications. Bacharach’s liability and Buyer’s remedy
under this warranty are limited to the repair or replacement, at
Bacharach’s option, of this Product or parts thereof returned to
Seller at the factory of manufacture and shown to Bacharach
Inc.’s reasonable satisfaction to have been defective; provided
that written notice of the defect shall have been given by Buyer
to Bacharach Inc. within one (1) year after the date of delivery of
this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title
to this Product. Bacharach’s liability and Buyer’s remedy under
this warranty of title are limited to the removal of any title
defects or, at the election of Bacharach, to the replacement of this
Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND
ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING
WITHOUT LIMITATION THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE: AND (II) ANY OBLIGATION, LIABILITY, RIGHT,
CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER
OR NOT ARISING FROM BACHARACH’S NEGLIGENCE,
ACTUAL OR IMPLIED. The remedies of the Buyer shall be
limited to those provided herein to the exclusion of any and all
other remedies including, without limitation incidental or
consequential damages. No agreement varying or extending the
foregoing warranties, remedies or this limitation will be binding
upon Bacharach, Inc. unless in writing, signed by a duly
authorized officer of Bacharach.
Manufacturer’s name:Bacharach, Inc.
Manufacturer’s address:621 Hunt Valley Circle
New Kensington, PA 15068
European Operations:Bacharach Europe
Sovereign House, Queensway
Leamington Spa
Warwickshire CV31 3JR
England
Product name:Leakator® 10
conforms to the following product
specifications:
EMC:European Directive 89/336/EEC
EN 50081-1 (Emissions)
EN 50082-1 (Immunity)
WARNING!
Because this instrument is used to detect and monitor materials and
conditions which are listed by OSHA or others as potentially hazardous
to personnel and property, the information in this manual must be fully
understood and utilized to ensure that the instrument is operating
properly and is both used and maintained in the proper manner by
qualified personnel. An instrument that is not properly calibrated,
operated and maintained by qualified personnel is likely to provide
erroneous information, which could prevent user awareness of a potentially hazardous situation for the instrument user, other personnel and
property.
If, after reading the information in this manual, the user has questions
regarding the operation, application or maintenance of the instrument,
supervisory or training assistance should be obtained before use.
Factory assistance is available by calling 724-334-5000.
The Leakator 10 is an intrinsically safe, battery-powered
portable instrument designed to primarily detect the source of
combustible-gas leaks. The instrument is ideally suited for
heating service contractors, utility personnel, and other users
who are interested in pinpointing gas leaks, and testing gas
appliances in residential, commercial and industrial installations.
The instrument is supplied with a durable plastic carrying
case, 20" probe and instruction manual.
The Leakator 10 features:
• Ten bright-red LEDs and a speaker that provide both a
visual and audible indication to the presence of gas.
Figure 1. Leakator 10 and Accessories
Instruction 19-9167
1
Introduction
Leakator 10
• Three operation-status LEDs that show power on, sensor
operation, and low battery.
• A 20-inch flexible probe.
• Simple thumb wheel on/off and gain control allowing one-
handed operation.
• A solid-state sensor that has a typical life of 5 years.
• A battery capacity of 30 hours under normal use conditions.
• The ability to detect natural gas (methane) in concentrations
as low as 20 ppm. In addition, the following gases and vapors
are also detectable:
For safety reasons, the Leakator 10 must only be operated
and serviced by qualified personnel. Read and understand the contents of this instruction manual before
operating or servicing.
To prevent ignition of a hazardous atmosphere, do not
perform any maintenance work, such as replacing the
instrument’s batteries, sensor, or sensor-matching resistor, in an area classified as being hazardous.
2
Instruction 19-9167
Leakator 10
Technical Characteristics
2.0 TECHNICAL CHARACTERISTICS
Power ..............................Five C-size Alkaline Batteries.
Battery Life .................... Approximately 30 hours of con-
tinuous operation under normaluse conditions.
Sensor:
Type ............................Solid State, plug-in replacement.
Life Expectancy ........ Typically 5 years.
Calibration ................ No user calibration required.
Warm-Up Time ............... Approximately 10 seconds.
Duty Cycle ......................Continuous with no limitation.
Gas Indication ................ Visual: 10 red ultra-bright LEDs.
Audible: Variable speed ticking
sound (earphone provided for use
in noisy environments).
Status LEDs .................... Power On (green)
Sensor Failure (yellow)
Low Battery (yellow)
Instruction 19-9167
3
Technical Characteristics
Leakator 10
Weight .............................17.8 oz (0.5 kg) (w/o batteries)
Dimensions: .................... 8.5" x 2.25" x 1.75"
(21.6 x 5.7 x 4.4 cm)
Operating
Environment:
Position ......................Any
Temperature ............. 23° to 130°F (–5° to 54°C)
Humidity .................... 10% to 90% RH, non-condensing
Safety Approvals ............ UL Classified, Intrinsically Safe
for use in Class I, Division 1,
Groups A, B, C and D. CE Mark.
4
Instruction 19-9167
Leakator 10
Battery Installation
3.0 BATTERY INSTALLATION
WARNING!
Do not replace batteries in a hazardous area.
CAUTION:
To maintain agency approval, use only
NEDA-14A type batteries or equivalent.
Remove battery cover. Install five (non-rechargeable) 1.5V
C-size alkaline batteries per Figure 2 – observe proper polarity. Then reinstall battery cover.
Instruction 19-9167
Figure 2. Battery Installation
5
Operation
Leakator 10
4.0 OPERATION
4.1 Turning ON the Instrument
Turn ON the instrument by rotating its thumb-wheel switch
(Fig 3) clockwise until a click is heard. Observe the following:
• The Power LED lights.
• The Fail LED will light, but should turn off in a few sec-
onds if the sensor is good and properly seated in its
socket.
• If the Low Bat. LED is on, replace the batteries per Sec-
tion 3.0 Battery Installation.
After the instrument stabilizes, rotate the Gain control in the
direction necessary to cause the bottom Gas Level LED to just
turn off. Be sure to always set this control in the same
environment you intend to test in.
The instrument is now ready for use.
4.2 Taking a Gas Reading
To verify that the instrument is operating, sample a known
combustible gas (e.g., a gas-air mixture from an unlit burner
of a natural-gas range). If no response is observed or heard,
refer to Section 6.0 Troubleshooting.
Important! The sensor becomes less sensitive
after being exposed to an excessive amount of
gas. Therefore, after testing the instrument as
described above, leave it on for several minutes
to restabilize the sensor.
6
Instruction 19-9167
Leakator 10
Operation
SENSOR
20"
EARPHONE
JACK
POWER ON
SENSOR FAILURE
LOW BATTERY
ON/OFF SWITCH &
GAIN CONTROL
COMBUSTIBLE GASES
Power
Fail
Low Bat.
Gain
Off
GAS
LEVEL
Figure 3. Jack, Control, and Indicators
Instruction 19-9167
7
Operation
Leakator 10
Take a gas reading by positioning the end of the instrument’s
flexible probe near the area to be sampled.
The presence of a combustible gas is indicated by a column of
ten red LEDs and a speaker. The number of lighted LEDs
gives visual indication of the relative gas level, while the
speaker produces clicking sounds, similar to a Geiger counter,
that increase in repetition rate as the gas concentration goes
up in relation to the instrument’s gain setting.
The instrument’s sensitivity to gas concentrations is adjusted
by means of its thumb-wheel Gain control. Clockwise rotation
increases sensitivity, while counterclockwise rotation decreases sensitivity. The Gain control is useful in pinpointing
large leaks by starting with the control set to light 1 or 2 LEDs,
and then gradually rotating the control counterclockwise as
the probe gets closer to the leak.
4.3 Using the Earphone
The earphone is an optional accessory that provides private
monitoring of the Leakator 10, and is recommended for use in
high-noise environments. The earphone plugs into the side of
the instrument, just above the thumb-wheel control.
4.4 Turning OFF the Instrument
Turn OFF the instrument by rotating the thumb-wheel control counterclockwise until a click is heard. There is no need
to purge the instrument with fresh air before turning it off.
8
Instruction 19-9167
Leakator 10
Maintenance
5.0 MAINTENANCE
WARNING!
Do not replace the batteries, sensor, or sensor
matching resistor in a hazardous area.
Since the Leakator 10 does not require calibration at regular
intervals, very little needs to be done to maintain the instrument in working order.
If the instrument has not been used for more than 3 months,
you may wish to turn it on and let it run in fresh air for several
minutes to keep the sensor at peak sensitivity.
If a problem should occur with your instrument, refer to
Section 6.0 Troubleshooting.
Detailed procedures of how to replace the sensor and other
components in the instrument are provided below.
5.1 Replacing the Sensor
Replacement sensors are classified at the factory according to
sensitivity, and are shipped with a matching resistor that
ensures the sensor will function properly when installed in
your instrument. It is important that the sensor and its
resistor be used together – DO NOT intermix resistors and
sensors.
Items required:
• Small and medium flat-blade screwdrivers
• Wire cutter
• Replacement sensor with matching resistor (P/N 19-0398)
Instruction 19-9167
9
Maintenance
Leakator 10
Procedure:
1. Turn OFF instrument.
2. Pry sensor out of its socket using a small flat-blade screwdriver (Fig 4) and discard.
3. Remove battery cover.
4. Remove old matching resistor (Fig 5) and discard.
5. Locate new sensor’s matching resistor; then cut and bend
the resistor leads using the forms that are molded into the
battery cover (Fig 6). Note that If the resistor leads are cut
too short, the instrument may not function.
6. Insert new matching resistor into printed circuit board,
and plug new sensor into socket at end of probe. Notice that
the sensor’s six pins and the socket are symmetrical and not
keyed, thus permitting the sensor to be plugged in two ways.
The socket, however, is wired so that it doesn’t matter which
way the sensor is installed.
7. Install battery cover.
8. Test instrument response (refer to Section 4.2).
Pry sensor from its socket
using a small flat-blade
screwdriver inserted
through access hole.
Figure 4. Removing Sensor
10
Instruction 19-9167
Leakator 10
FORM
CLIP
(2 ENDS)
CAUTION:
Use Only C-Size 1.5 x 5
NEDA 14A or Equivalent
Cut Matching Resistor
leads to correct length by
inserting each lead into
form as shown and cutting
here.
NOTE:
The plastic clip
guide length represents the
minimum lead length. If the
resistor leads are cut too
short, the instrument may
not function.
Optimum lead length is
11/16".
After cutting resistor leads,
use this form to bend leads
as shown.
MATCHING
RESISTOR
Maintenance
Figure 5. Location of Matching Resistor
Figure 6. Cutting and Bending Resistor Leads using Forms
Molded into Battery Cover
Instruction 19-9167
11
Maintenance
Leakator 10
5.2 Replacing the Printed Circuit Board
Items required:
• Medium flat-blade screwdriver
• Medium Phillips screwdriver
• Replacement printed circuit board (P/N 19-0418)
Procedure:
(See Figure 7)
1. Turn OFF instrument and lay it face down on work area.
2. Remove battery cover and batteries.
3. Remove sensor-matching resistor.
4. Remove four screws securing rear case.
5. Carefully lift rear case and flexible arm assembly away
from instrument, being careful not to pull wires attached to
printed circuit board. Lay rear case face down on work area.
6. Carefully remove printed circuit board from front case.
7. Unplug connectors J2 and J3 from printed circuit board.
8. Plug connectors J2 and J3 into new printed circuit board;
then install board into front case.
9. Install flexible arm assembly and rear case using the
screws that were removed in Step 4.
10. Replace sensor-matching resistor.
11. Replace batteries and battery cover.
12. Test instrument response (refer to Section 4.2).
12
Instruction 19-9167
Leakator 10
Maintenance
5.3 Replacing the Speaker
Items required:
• Medium flat-blade screwdriver
• Medium Phillips screwdriver
•11/32" Nut driver
• Replacement speaker (P/N 19-0400)
Procedure:
(See Figure 7)
1. Turn OFF instrument; then remove battery cover and
batteries.
2. Remove sensor-matching resistor.
3. Remove four screws securing rear case; then carefully lift
rear case and flexible arm assembly away from instrument,
being careful not to pull wires attached to printed circuit
board. Lay rear case face down on work area.
4. Carefully remove printed circuit board from front case;
then unplug speaker connector J2 from printed circuit board.
5. Remove two nuts and washers securing speaker to rear
case; then remove speaker.
6. Install new speaker and secure in place using the nuts and
washers that were removed in Step 5. CAUTION! Do not
overtighten the nuts!
7. Replace flexible arm assembly and rear case using the
screws that were removed in Step 3.
1. Turn OFF instrument and lay it face down on work area.
2. Remove battery cover and batteries.
3. Remove sensor-matching resistor.
4. Remove four screws securing rear case.
5. Carefully lift rear case and flexible arm assembly away
from instrument, being careful not to pull wires attached to
printed circuit board. Lay rear case face down on work area.
6. Carefully remove printed circuit board from front case.
7. Unplug sensor connector J3 from printed circuit board.
8. Remove flexible probe assembly and remove sensor.
9. Plug connector on new flexible probe assembly into printed
circuit board connector J3.
14
Instruction 19-9167
Leakator 10
Maintenance
10. Replace printed circuit board, new flexible probe assem-
bly, and rear case using the screws that were removed in
Step 4.
11. Replace sensor.
12. Replace sensor-matching resistor.
13. Replace batteries and battery cover.
14. Test instrument response (refer to Section 4.2).
Instruction 19-9167
15
Troubleshooting
Leakator 10
6.0 TROUBLESHOOTING
WARNING!
Do not perform any maintenance work in a
hazardous area.
Because of the Leakator 10’s advanced design, trouble with
the instrument can usually be diagnosed by looking at the
LEDs as seen through the front-case. See Figure 3.
If the Fail LED lights, check that the sensor is seated and J3
is connected to the printed circuit board. If still no change,
replace the sensor per Section 5-1 Replacing the Sensor.
If the Low-Bat. LED lights, replace the batteries per Section 3.0 as soon as possible. The instrument should operate,
however, for several hours before the batteries go dead.
If the instrument doesn’t respond to a combustible gas, perform the following in the order presented until the problem is
corrected:
- First ensure that all connectors are securely in place on the
printed circuit board, and that the sensor is firmly seated
into its socket.
- Replace sensor, even though the Sensor-Fail LED isn’t on,
per Section 5.1 Replacing the Sensor.
- Replace circuit board per Section 5.2 Replacing the Printed