Model FL7060
Electric Field Probe
2MHz–60GHz
2–1000 V/m
User-Selectable X, Y, Z Axes
The FL7060 is a smart, fast, extremely accurate electric field probe that contains
an internal microprocessor to provide linearization, temperature compensation,
control, and communication functions. Noise reduction and temperature
compensation allow accurate measurements down to 2 V/m without zero
adjustment. Microprocessor based linearization technology provides a 54 dB
dynamic range. When rotated about its critical angle mount, the probe provides
typical isotropic response of ±1.5 dB to 60 GHz.
The FL7060 is laser powered to allow for continuous operation without
recharging or battery replacement.
Correction factors are provided with the probe. These factors can be loaded into
the Model FM7004 Field Monitor (sold separately) to automatically correct the
probe readings at user-specified frequencies. When correction factors are
applied, the true accuracy of the probe can be realized.
The FL7060 communicates through glass fiber optic cables, up to 100 meters
long, to the FI7000 interface. X, Y, Z, and isotropic readings can be returned
through an FI7000 in 20 msec.
NOTES: This probe requires an FI7000 for power and communication.
FM7004 is recommended for local monitoring and control.
083110
160 School House Road Souderton, PA 18964-9990 • 215-723-8181 • www.ar-worldwide.com Page 1 of 2
Amplitude Accuracy (field aligned with sensor axes)
Without correction factors applied ............................. 1.0 dB @ 10 MHz
With correction factors applied .................................. Typical expanded measurement uncertainty (95% confidence interval)
0.95 dB, 2 MHz–1 GHz
1.5 dB, 1 GHz–60 GHz
Response Time/ Sampling Rate (through FI7000) ............... 20 msec/up to 50 samples per second, USB and GPIB only
Isotropic Deviation (measured at the critical angle) ............. ±0.5 dB @ 10 MHz
±1.5 dB, 2 MHz–60 GHz (typical)
Operating Range ............................................................ 2–1000 V/m
Linearity, 2 to 1000 V/m .................................................. 0.5 dB
Temperature Stability.. ..................................................... +0.5 dB over operating temperature range
Damage Level ................................................................. 1200 V/m CW
Ranges ........................................................................... Single continuous range
Data returned from probe ................................................ X, Y, Z axes, and composite
Power Requirements ........................................................ Laser powered from FI7000 interface
Dimensions ..................................................................... Approx. 278 x 65 x 65 mm
Probe Head Diameter ...................................................... 65mm
Weight ........................................................................... 150g approx.
Operating Temperature Range ......................................... 10C to 40C (50F to 104F) @ 5% to 95% RH non-condensing
Fiber Optic Connectors ................................................... Two E2000 compact duplex connectors at 1 meter, includes fiber optic verification
Calibration Data ............................................................. Accredited Calibration Report supplied with probe
SPECIFICATIONS, FL7060
loop.
Page 2 of 2