Atec Agilent-E7401A, Agilent-E74002A, Agilent-E74003A, Agilent-E74004A, Agilent-E74005A User Manual

These specifications apply to the Agilent Technologies E7401A, E7402A, E7403A, E7404A and E7405A EMC analyzers.
Specifications
All specifications apply over 0° C to +55° C unless otherwise noted and are covered by the product warranty. The analyzer will meet its specifications when: it’s within the one year calibration cycle, AUTO ALIGN [ALL] is selected, stored a minimum 2 hours within the operating temperature range, turned on for at least 5 minutes, and Align Now RF has been run once every 24 hour period. Typical performance describes the level at which 80% of the units will meet or exceed with a 95% confidence level over 20 to 30° C, but is not covered in the product warranty. Characteristics describe expected product performance levels that are not covered in the product warranty.
Frequency specifications Frequency range
E7401A
50 9 kHz to 1.5 GHz
E7402A 9 kHz to 3.0 GHz
dc coupled 30 Hz
6
to 3.0 GHz
ac coupled 100 kHz6to 3.0 GHz
E7403A
dc coupled 9 kHz to 6.7 GHz dc coupled (Option UKB) 30 Hz6to 6.7 GHz ac coupled 100 kHz to 6.7 GHz Band 0 9 kHz to 3.0 GHz 1 2.85 GHz to 6.7 GHz
E7404A
dc coupled 9 kHz to 13.2 GHz dc coupled (Option UKB) 30 Hz6to 13.2 GHz ac coupled 100 kHz to 13.2 GHz Band LO harmonic = N 0 1- 9 kHz to 3.0 GHz (Option UKB) 30 Hz6to 3.0 GHz 1 1- 2.85 GHz to 6.7 GHz 2 2- 6.2 GHz to 13.2 GHz
E7405A 9 kHz to 26.5 GHz
Band LO harmonic = N 0 1- 9 kHz to 3.0 GHz 0 (Option UKB) 30 Hz6to 3.6 GHz 1 1- 2.85 GHz to 6.7 GHz 2 2- 6.2 GHz to 13.2 GHz 3 4- 12.8 GHz to 19.2 GHz 4 4- 18.7 GHz to 26.5 GHz
Frequency reference (Option 1D5)
Aging ±2 x 10-6/year ±1 x 10-7/year Temperature stability ±5 x 10
-6
±1 x 10
-8
Settability ±5 x 10
-7
±1 x 10
-8
Frequency readout accuracy
(start, stop, center, marker) ±(frequency indication
x frequency reference error
1
+ span accuracy + 15% of RBW + 10 Hz) + 1 Hz x N
4
Marker frequency counter
2
Accuracy
3
±(marker frequency x frequency reference error1+ counter resolution)
Counter Resolution Selectable from 1 Hz to 100 kHz
Frequency span
Range 0 Hz (zero span), 100 Hz x N4to
the range of the spectrum analyzer
Resolution 2 Hz x N
4
Accuracy (> 2000 sweep points)
Sweep type linear ±0.5% of span Sweep type log ±2% of span (characteristic)
Agilent
E7400 A-series EMC Analyzers
Data Sheet
Manufacturers > Agilent Technologies Categories > EMC Test Equipment > Conducted \ Radiated Emissions Test > EMI Analyzers \ EMC Receivers
Agilent E7405A EMC Analyzer Product Page
How-To Video
2
Sweep time
Range
Span > 0 Hz 1 ms to 4000 s Span = 0 Hz 10 µs15to 4000 s
(Option AYX) 50 ns15to 4000 s Accuracy ±1% Sweep trigger Free run, single, line, video,
external, delay, offset, and gate (Option 1D6)
Delay trigger range 1 µs to 400 s
Sweep (trace) point range 101 to 8192
Span = 0 Hz 2 to 8192
Resolution bandwidth 10 Hz to 3 MHz (-3 dB) in
1-3-10 sequence
16
5 MHz (-3 dB) bandwidth 200 Hz
16,
9 kHz, 120 kHz, 1 MHz
(-6 dB) EMI bandwidths
1 MHz (impulse) EMI bandwidth
Option 1D5 Adds 1 Hz and 3 Hz Accuracy
10 Hz to 300 MHz (-3 dB) ±10%
1Hz and 3 Hz (Option 1D5) ±10%
1 kHz to 3 MHz (-3 dB) ±15%
5 MHz (-3 dB) ±30%
200 Hz (-6 dB) ±10%
9 kHz to 120 kHz (-6 dB) ±20%
1 MHz (-6 dB) ±10%
1 MHz (impulse) ±15% Selectivity (characteristic)
10 Hz to 300 Hz (-3 dB) < 5:1 (-60 dB/-3 dB)
(Digital, approximately Gaussian-shaped)
1 kHz to 3 MHz (-3 dB) < 5:1 (-60 dB/-3 dB)
(approximately Gaussian-shaped)
200 Hz (-6 dB) < 3:1 (-40 dB/-6 dB)
(Digital, Kaizer Windows)
9 kHz, 120 kHz,
1 MHz (-6 dB) < 10:1 (-60 dB/-6 dB)
(approximately Gaussian-shaped)
1 MHz (impulse) < 10:1 (-60 dB/-6 dB)
(approximately Gaussian-shaped)
Video bandwidth range 30 Hz to 3 MHz6in 1-3-10
sequence 1, 3, 10 Hz for RBW’s < 1 kHz
Stability
Noise sidebands (1 kHz RBW, 30 Hz VBW and sample detector)
E7401A
1 kHz na 79 dBc/Hz
(Option 1D5) 10 kHz -93 dBc/Hz -95 dBc/Hz > 20 kHz -100 dBc/Hz -102 dBc/Hz > 30 kHz -104 dBc/Hz -106 dBc/Hz > 100 kHz -113 dBc/Hz -116 dBc/Hz
E7402/03/04/05A
1 kHz na 78 dBc/Hz
(Option 1D5) 10 kHz -90 dBc/Hz
21
-94 dBc/Hz
21
> 20 kHz -100 dBc/Hz
21
-105 dBc/Hz
21
> 30 kHz -106 dBc/Hz
21
-112 dBc/Hz
21
> 100 kHz -119 dBc/Hz
21
-122 dBc/Hz
21
> 1 MHz -125 dBc/Hz
21
-127 dBc/Hz
21
> 5 MHz -127 dBc/Hz
21
-129 dBc/Hz
21
> 10 MHz ≤ -131 dBc/Hz
21
-136 dBc/Hz
21
Residual FM
1 kHz RBW, 1 kHz VBW 150 x N4Hz pk-pk in 100 ms Option 1D5 100 x N4Hz pk-pk in 100 ms 10 Hz RBW, 10 Hz VBW 2 x N4Hz pk-pk in 20 ms
System-related sidebands
30 kHz offset from CW signal -65 dBc + 20 Log N
4
Amplitude specifications Amplitude range
Measurement range Displayed average noise level (DANL)
to maximum safe input level
Input attenuator range
E7401A 0 to 60 dB, in 5 dB steps E7402A, 03A, 04A 0 to 65 dB (75 dB6), in 5 dB steps E7405A 0 to 65 dB, in 5 dB steps
-70
Typical performance Spec
-90
dBc/Hz
-110
-130 1k 10k 100k 1M
Offset (Hz)
3
Maximum safe input level
Average continuous power
E7401A (input attenuator 15 dB)
+30 dBm (1 W) (input attenuator 5 dB)
E7402A/03A/04A/05A +30 dBm (1 W)
Peak pulse power (input attenuator 30 dB)
E7402A/03A/04A/05A +50 dBm (100 W) E7401A +30 dBm (1 W)
dc
E7401A, E7402A 100 Vdc E7402A (Option UKB) 0 Vdc (dc coupled)
50 V (ac coupled)
E7403A, E7404A 0 Vdc (dc coupled)
50 V (ac coupled)
E7405A 0 Vdc
(Option UKB) 0 Vdc (dc coupled)
50 V (ac coupled)
1 dB gain compression (total power at input mixer5)
50 MHz 0 dB 6.7 GHz -3 dB13.2 GHz -5 dB
Displayed average noise level (dBm) (Input terminated, 0 dB attenuation, sample-detector) 1 kHz RBW; 30 Hz VBW 10 Hz RBW; 1 Hz VBW 1 Hz RBW; 1 Hz VBW (Option 1D5)
1 kHz 10 Hz 1 kHz 10 Hz 1 Hz RBW RBW w/preamp w/preamp Option 1D5
on on, typical w/preamp
on, typical
E7401A
400 kHz to 10 MHz ≤ -115 ≤ -134 ≤ -150 ≤ -155 ≤ -165 10 MHz to 500 MHz ≤ -119 ≤ -138 ≤ -154 ≤ -156 ≤ -166 500 MHz to 1 GHz ≤ -117 ≤ -136 ≤ -152 ≤ -156 ≤ -166 1 GHz to 1.5 GHz ≤ -114 ≤ -133 ≤ -150 ≤ -155 ≤ -165
E7402A
30 Hz to 9 kHz
22
(Option UKB) na ≤-93 na na na 9 kHz to 100kHz22na -109 na na na 100 kHz to 1 MHz22na -135 na na na 1 MHz to 10 MHz22≤-117 ≤-136 na ≤-152 ≤-162 10 MHz to 1 GHz ≤-117 ≤-136 ≤-15219≤-156 ≤-166 1 GHz to 2 GHz ≤-116 ≤-135 ≤–15319≤-156 ≤-166 2 GHz to 3 GHz ≤-114 ≤-133 ≤–15119≤-154 ≤-164
E7403A, 04A, 05A
30 Hz to 9 kHz
22
(Option UKB) na ≤ -93 na na na 9 kHz to 100kHz22na -1096na na na 100 kHz to 1 MHz22na -1356na na na 1 MHz to 10 MHz22≤ -1176≤ -1376na -155 -165 10 MHz to 1 GHz ≤ -116 ≤ -135 ≤ -15119≤ -157 ≤ -167 1 GHz to 2 GHz ≤ -116 ≤ -131 ≤ –15119≤ -155 ≤ -165 2 GHz to 3 GHz ≤ -112 ≤ -131 ≤ –14919≤ -152 ≤ -162 3 GHz to 6 GHz ≤ -112 ≤ -131 na ≤ -138 na 6 GHz to 12 GHz ≤ -111 ≤ -130 na ≤ -137 na 12 GHz to 22 GHz ≤ -107 ≤ -126 na ≤ -134 na 22 GHz to 26.5 GHz ≤ -106 ≤ -125 na ≤ -132 na
Display range
Log Scale 0.1, 0.2, 0.5 dB/division and 1 to 20 dB/division
in 1 dB steps; ten divisions displayed
RBW 1kHz 0 to -85 dB from reference level is calibrated
RBW 300 kHz 0 to -120 dB from reference level is calibrated Linear scale 10 divisions Scale units dBm, dBmV, dBµV, dBµA, Amps, Volts and Watts
Marker readout resolution
Log scale
0 to -85 dB 0.04 dB 0 to -120 (RBW 300 Hz 0.04 dB
Linear scale 0.01% of reference level
Fast sweep times for zero span (Option AYX) Log Scale
0 to -85 dB 0.3 dB
Linear 0.3 dB of reference level
Frequency response (10 dB input attenuation)
Absolute7Typical Relative
flatness
8
E7401A
9 kHz to 1.5 GHz ±0.5 dB na ±0.5 dB
E7402A/03A/04A/05A
30 Hz to 3 GHz
6
(Option UKB) ±0.5 dB na ±0.5 dB
9 kHz to 3 GHz ±0.46 dB ±0.14 dB ±0.5 dB
3.0 GHz to 6.7 GHz ±1.5 dB ±0.39 dB ±1.3 dB
6.7 GHz to 13.2 GHz ±2.0 dB ±0.68 dB ±1.8 dB
13.2 GHz to 26.5 GHz ±2.0 dB ±0.86 dB ±1.8 dB
Input attenuation switching uncertainty at 50 MHz
Attenuation setting
0 dB to 5 dB ±0.3 dB 10 dB Reference 15 dB ±0.3 dB 20 to 60 dB (E7401A) ±(0.1 dB + 0.01 x attenuator setting) 20 to 65 dB ±(0.1 dB + 0.01 x attenuator setting)
Absolute amplitude accuracy Typical
At reference settings
13
±0.34 dB ±0.13 dB E7401A ±0.30 dB ±0.10 dB Preamp on
16
±0.37 dB ±0.14 dB
Overall amplitude accuracy9±(0.54 dB + absolute
frequency response)
Loading...