December 2000
Issue 1
The ETM-2000 series
Easy Temperature Measurement
G Complete solutions
- Minimize the risk of errors
G Flexible solutions
- May be supplemented as needs change
G Data logging
- Collect and document data from critical measure
ments
G Computer software
- Analyze, present, and store critical measurements
G Industry protected
- The instrument is protected with a rubber boot
G Practical details
- Instruments may be carried on the belt or placed on
a surface
G A safe buy
- Built-in AMETEK quality and features such as
confidence, efficiency, technical support, and service
PRODUCT DESCRIPTION
As a result of the introduction of quality management systems
such as ISO-9000, HACCP, etc., as well as increased focus
on hygiene, environment and security, demands to minimizing
risks and errors have grown. To meet these demands you
need reliable and efficient measuring equipment.
AMETEK has composed a number of complete solutions for
temperature measurement (the ETM-2000 series) and for
pressure measurement (the EPM-2000 series). Each system
is a complete kit that is delivered in a practical, lightweight
carrying case, making it easy to carry around to the various
measuring sites. The configuration eliminates the chance of
forgetting probes, contact paste, or the operating manual.
With 5 different systems the ETM-2000 series (Easy Temperature Measurement) covers a broad variety of applications
within temperature measurement. The series has 3 systems
for thermocouples and 2 for Pt100 probes. Both Pt100 models
feature IP67 protection making them very suitable for use in
damp rooms in dairies, slaughterhouses and similar food
environments. 3 of the 5 systems feature ”Speed-Read” functions which calculate the temperature after just 14 seconds in
the medium. This saves the operator time when measuring
granulate, powder, and air. Data collection is an option in both
the Pt100 system and the thermocouple system. Results are
shown on the display or on a PC.
FUNCTIONS - ALL MODELS
Backlight: The display light is on as long as the key is
activated.
Display Hold: Freezes the reading – used when watching
the probe is more important than watching the indicator.
Auto switch-off: Auto switch-off after 12 minutes. Extends
battery life (auto switch-off does not operate in max/min mode
or logging mode. The user can also turn off this feature if not
required).
Set-up retention: The instrument switches on with the last
settings used.
FUNCTIONS - SPECIFIC MODELS
SpeedRead>>: This function calculates the temperature after
just 14 seconds in the medium, saving operator time.
A/B, A-B: Enables the user to select input channel A, channel
B, or the difference between them.
C/F: Toggle between Celsius and Fahrenheit.
Range lock: Enables the operator to set the instrument to 1°
resolution only.
Thermocouple selection: Change between the different
thermocouple types.
Max/Min: Stores maximum, minimum, and average tempera-
tures.
Logging on demand: Allows logging up to a maximum of
250 readings.
Pre-set interval logging: Allows readings to be stored auto-
matically at a pre-set time interval.
- 1 -
December 2000
Issue 1
ETM system 1
GENERAL INDUSTRY
Simple and sturdy instrument to be used wherever
temperature measurement is required bearings, lubricating oil, ambient air, liquids in
tanks, operational tests, and start of systems, etc.
Instrument: ETM-2000
Probes: General-purpose probe 2101013
Air probe 2301001
Flexible wire cable probe 2304001
Surface probe 2201002
Including: Contact paste, batteries, protective
rubber boot, carrying case with
foam inserts, and instructions
ETM system 2
GENERAL INDUSTRY WITH DATA COLLECTION
Same as system 1, but with more functions and
choice of probe types. Document your measurements with the data collection function, store min/
max values, or save time with the “SpeedRead“
function. PC analysis software may be added.
(EasyLink software and interface unit are not
included).
Instrument: ETM-2098
Probes: See system 1
Including: See system 1
ETM system 3
FOOD INDUSTRY
Precise Pt100 measuring instrument in an IP67
waterproof design. Check frozen meat, coldstorage rooms, fermentation processes, bunker
fuel tanks, pasteurization systems, ovens,
hotplates, etc.
Instrument: ETM-2024
Probes: General-purpose probe 121908
Air probe 120909
Needle shaped probe 121910
Including: Batteries, rubber protection boot,
carrying case with foam inserts,
and instructions
ETM system 4
FOOD INDUSTRY WITH DATA COLLECTION
Same as system 3, but with more functions.
Document your measurements with the data
collection function, store min/max values or save
time with the “SpeedRead” function. PC analysis
software may be added. (EasyLink software and
interface unit are not included).
Instrument: ETM-2084
Probes: See system 3
Including: See system 3
FUNCTIONAL COMPARISON
System 1 2 3 4 5
Probe type K KTJNRS Pt100 Pt100 KTJNRS
Dual Input l
SpeedRead ll l
°C and °F llll
Hold ll l l l
Fixed resolution llll
MAX/MIN reading llll
Average ll
Data logging ll
Adjustment from front ll
Temperature range:
Pt100: -150 to 800°C (-238 to 1472°F)
Type K: -200 to 1350°C (-392 to 2462°F)
Type T: -250 to 400°C (-418 to 752°F)
Type J: 150 to 750°C (302 to 1382°F)
Type N: -200 to 1300°C (-392 to 2372°F)
Type R & S: 0 to 1750°C (32 to 3182°F)
PHYSICAL SPECIFICATIONS
Operating temperature: ..................... –10°C to 50°C (14°F to 122°F)
Model ETM-2000, ETM-2029, ETM-2098: ................................. IP65
Model ETM-2024, ETM-2084:.................................................... IP67
Dimensions: ............................ 155 x 67 x 40 mm (6.1 x 2.6 x 1.6 in)
Weight: ........................................................................ 180 g (6.4 oz)
Shipping details complete systems:
Dimensions: ...................... 395 x 350 x 100 mm (15.5 x 13.8 x 4 in)
Weight app.: ................................................................... 2 Kg (4.4 lb)
CE label ....................................................................................... YES
FUNCTIONAL SPECIFICATIONS
Accuracy model ETM-2000, ETM-2029, ETM-2098
K, T, J, N above –100°C (-148°F): ............ 0.1% rdg ±0.2°C (0.36°F)
K, T, J, N below –100°C (-148°F): ............ 0.5% rdg ±0.2°C (0.36°F)
R, S: .............................................. Better than 0.1% F.S. ±0.2% rdg.
Accuracy model ETM-2024, ETM-2084
Pt100: ..................................................... ±0.2% rdg ± 0.1°C (0.18°F)
(All accuracies are ±1 digit)
The above mentioned specification is obtained by combining the
accuracy and any uncertainty due to the measurement process.
However exclusive of the inaccuracy of the probe.
“SpeedRead”: This function reduces the accuracy from
Class 0.5 to 2 (CEN TC141/WG12 N42E applies).
ETM system 5
HEATING AND VENTILATION
Sturdy general purpose system for temperature
measurements in energy installations. Adjustment/
control, measurements in thermometer wells or
direct on pipes (through the insulation), flue gas
temperatures on small oil-fired furnaces, radiator
surface and room temperature measurements, etc.
Instrument: ETM-2029
Probes: General-purpose probe 2101013
Flue gas 2501023
Flexible probe 2304001
Surface probe 2202001.
Including: See system 1
Resolution: .............................. 0.1° up to 199.9°C (392°F) or else 1°
Battery type: ................................................ 2 AA or equivalent cells
Battery life: ......................................................... Typically 500 hours
Auto switch-off: ............................................................... 12 minutes
Low battery indication: .......................... Symbol appears on display
Cold junction compensation: .............................. Yes (system 1,2,5)
PROBE SPECIFICATIONS
Type K (IEC 584 Class 2): ........... ±2.5°C (±3.5°F) or 0.0075 x T
(Whichever is greatest)
Type Pt100 (IEC 751 Class A): ................... ± 0.15°C + 0.002 x T
(±0.27°F + 0.002 x T
- 2 -
actual
actual
actual
)