Ametek ETC-125 A, ETC- 400 A, ETC- 400 R Operating Manual

User- and Reference manual
Temperature Calibrator
JOFRA ETC-125 A / 400 A / 400 R
Copyright 2005 AMETEK DENMARK A/S (123943-UK)
About this manual….
The user manual sets out the operating instructions for the
instrument. It is designed to provide a quick reference guide for use in the field.
The reference manual is aimed at users who are familiar with Ametek calibrators, as well as those who are not. The manual is divided into 10 chapters which describe how to set up, operate, service and maintain the calibrator. The technical specifications are described and accessories may be ordered from the list of accessories.
Safety symbols
This manual contains a number of safety symbols designed to draw your attention to instructions which must be followed when using the instrument, as well as any risks involved.
Warning
Events which may compromise the safe use of the instrument and result in considerable personal or material damage.
Caution…
Events which may compromise the safe use of the instrument and result in slight personal or material damage.
Note…
123943 03 29-09-2005
Special situations which demand the user’s attention.
List of contents
.................................................. User manual, English
.........................................Reference manual, English
29-09-2005 123943 03
User manual
Temperature Calibrator
JOFRA ETC-125 A / 400 A / 400 R
Copyright 2005 AMETEK DENMARK A/S (123943-UK)
FIG. 1
1
N
O
115 V
2
5
3 4
29-09-2005
FIG. 2
14
CALIBRATION INSTRUMENTS
AUTO STEP
ESC MENU
5
4
3
6
2
C
min
C F
F
1
ETC
C
F
7
8
9
F
C
min
10
13
29-09-2005
12
11
FIG. 3
AUTO STEP
ESC MENU
ESC
°C
°C
MENU
29-09-2005
FIG. 4
AUTO STEP
AUTO STEP
29-09-2005
FIG. 5
ESC MENU
ESC MENU
ESC MENU
ESC MENU
ESC MENU
ESC MENU
29-09-2005
ESC MENU
29-09-2005
List of contents
1.0 Introduction..................................................................... 2
2.0 Safety instructions ......................................................... 3
3.0 Operating the calibrator................................................. 6
3.1 Before use ...........................................................................6
3.2 Keyboard .............................................................................8
3.3 Display.................................................................................8
3.4 Connections ........................................................................9
3.5 Calibrator functions - overview ............................................9
3.6 Selecting the set-temperature .............................................9
3.7 AUTO STEP ......................................................................10
3.8 MENU................................................................................11
4.0 After use........................................................................ 12
123943 03 1
1.0 Introduction
ETC-calibrators are temperature calibrators designed to calibrate temperature sensors.
Read this manual carefully before using the instrument and make sure that all safety instructions and warnings are observed.
2 123943 03
2.0 Safety instructions
Read this manual carefully before using the instrument!
In order to avoid any personal injuries and/or damage to the instrument all safety instructions and warnings must be observed.
Disposal – WEEE Directive
These calibrators contain Electrical and Electronic circuits and must be recycled or disposed of properly (in accordance with the WEEE Directive 2002/96/EC).
Warning……
About the use:
The calibrator must not be used for any purposes other
than those described in this manual.
The calibrator has been designed for interior use only
and should not be used in hazardous areas, where vapour or gas leaks, etc. may constitute a danger of explosion.
The calibrator must be kept free within an area of 20
cm on all sides and 1 metre above the calibrator.
Never use heat transfer fluids such as silicone, oil,
paste, etc.
These fluids may penetrate the calibrator and cause
damage or create poisonous fumes.
123943 03 3
About the insertion tubes:
Never leave hot insertion tubes which have been
removed from the calibrator unsupervised – they may constitute a fire hazard.
Caution – Hot surface
Do not touch the well, the insertion tube or the grid
plate, as the calibrator is heating up / has been heated up – they may be very hot
Do not touch the tip of the sensor when it is removed
from the insertion tube/well – it may be very hot.
4 123943 03
Caution…
About the use:
Do not use the instrument if the ventilator is out of
order.
About the well, insertion tube and grid plate:
The well and the insertion tube must be clean before
use.
The insertion tube must never be forced into the well.
The well could be damaged as a result, and the insertion tube may get stuck.
The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause verdigris on the
insertion tube inside the instrument. There is a risk that the insertion tube may be stuck if this is allowed to happen.
Note…
The product liability only applies if the instrument is subject to a manufacturing defect. This liability becomes void if the user fails to follow the instructions set out in this manual or uses unauthorised spare parts.
123943 03 5
3.0 Operating the calibrator
3.1 Before use
Warning
The calibrator must not be used for any purposes other
than those described in this manual.
The calibrator is designed for interior use only and
should not be used in risk-prone areas, where vapour or gas leaks, etc. may constitute an explosives hazard.
Never use heat transfer fluids such as silicone, oil,
paste, etc.
These fluids may penetrate the calibrator and cause
damage or create poisonous fumes.
The calibrator must be kept clear within an area of 20
cm on all sides and 1 metre above the calibrator.
Caution – Hot surface
Do not touch the well, the insertion tube or the grid plate,
as the calibrator is heating up / has been heated up – they may be very hot.
6 123943 03
Follow the instructions below before using the calibrator (cf. Fig. 1):
1. Place the calibrator on an even horizontal surface away from all draughts.
Caution…
2. Check that the voltage shown on the label above the power
3. Plug in the cable below (pos. 3) the power control switch
4. The selected insertion tube is inserted into the calibrator
Do not use the instrument if the ventilator is out of order. Ensure a free supply of air to the ventilator (pos. 5).
control switch is identical to the mains voltage (ETC-400 only). If not, the instrument should be returned to Ametek Denmark A/S or the distributor for replacement.
(pos. 4) and check that the earth connection is present.
(ETC-125 A only).
Caution…
The well and the insertion tube must be clean before
5. Place the sensor (pos. 1, Fig. 1).
use.
The insertion tube must never be forced into the well.
The well could be damaged as a result, and the insertion tube may get stuck.
123943 03 7
3.2 Keyboard
The keys on the keyboard activate the following functions (cf. Fig. 2):
POS Description
c
d
e
f
g
ENTER button used to accept chosen options.
UP ARROW button used to adjust temperature values
(value increases) and to select menu options.
DOWN ARROW button used to adjust temperature values (value decreases) and to select menu options.
ESC/MENU button used to escape or to activate the menu system (hold button down for min. 2 seconds).
AUTO STEP button used to activate AUTO STEP. The function is used to switch between a series of set­temperatures automatically.
3.3 Display
The various segments of the display are used to indicate the following (cf. Fig. 2):
POS Description
h
i
Used to display Read-temperature and parameters in the menu system.
Celsius temperature unit for top display.
j
k
l
8 123943 03
Fahrenheit temperature unit for top display.
Fahrenheit temperature unit for bottom display.
Celsius temperature unit for bottom display.
Minute time unit for bottom display.
Used to display set-temperature, time-until-stable and parameter values in the menu system.
AUTO STEP symbol used to indicate that the function is active (symbol flashes repeatedly).
Check mark displayed when the calibrator is stable.
3.4 Connections
The instrument is designed for the following connections (cf. Fig. 1):
POS Description
d
Connection of RS232 cable
3.5 Calibrator functions - overview
The instrument’s functions are divided into hierarchical groups. See the key diagram in Fig. 3.
3.6 Selecting the set-temperature
)
)
)
The calibrator will now work towards the new set-temperature.
Press or flashes (the starting point is the last chosen set-temperature
even if the instrument has been turned off).
Press or to select the required temperature.
Press
return to the previous value.
to accept the change or
The current set-temperature
.
ESC MENU
to cancel and
123943 03 9
3.7 AUTO STEP
The AUTO STEP function (cf. Fig. 4) is used to step automatically between a range of different set-temperatures.
AUTO
. The instrument displays the number of set-
STEP
.
)
Press temperature
)
)
)
)
)
)
)
Press or to select the required number of steps.
Press
will flash.
Press or to select the required temperature.
Press
temperature will flash. This process will be repeated until the last value has been accepted. The extra
wish the calibrator to remain at every step will flash.
Press or to set the required number of minutes.
Press
activated.
Press after the last set-temperature to end the function
ESC
or to leave the function at any time.
MENU
to accept your selection. The first set-temperature
to accept your selection. The next set-
for which you
to accept your selection. The function will be
10 123943 03
3.8 MENU
The MENU function (cf. Fig. 5) is used to modify the SETUP parameters.
ESC
)
)
)
Hold down will appear on the display.
Press
Press or to toggle between the SETUP
parameters:
MENU
for approx. 2 seconds. The word
. The first SETUP parameter will be displayed.
: Temperature unit °C or °F.
well is stable before the check mark symbol is displayed.
decimal.
)
)
)
)
Press
change. The current value will flash.
Press or to select the required value.
Press to accept your selection or
return to the previous value.
Once you have changed all SETUP parameters as required,
cancel the function by pressing
: The highest permissible temperature for
the calibrator.
: Extra time which must elapse once the
: Temperature resolution of 0 or 1
to select the SETUP parameter you wish to
ESC
to cancel and
MENU
ESC
twice.
MENU
123943 03 11
Loading...
+ 53 hidden pages