The IVAC® PCAM® Syringe Pump is designed to provide a small, self-administered dose of analgesic, as and when the patient demands
it by activating a hand operated button. The clinician can select limits for various parameters, including the size of each individual
dose, the minimum time between doses and the number of doses allowed during a period. In parallel to the patient controlled
operation the pump allows the clinician to set an automatic loading dose which will be delivered at the onset of treatment. Similarly a
continuous background infusion which is delivered irrespective of the patients demands for analgesia, can also be selected.
In addition, the pump will automatically record valuable information about each patients treatment and their individual demands for
analgesia. This allows further analysis of the frequency with which analgesia is being requested, the total dose delivered etc.
Product Familiarity
Prior to operation of the pump and prior to attempting any repairs or servicing, carefully read the Directions for Use (DFU)
As part of continuous improvement, product enhancements and changes are introduced from time to time.
Purpose of this Manual
This Technical Service Manual describes how to set up, test and maintain the IVAC® PCAM® Syringe Pump. This manual is intended for
use by personnel experienced in medical equipment testing and maintenance procedures .
Conventions Used in this Manual
BOLDUsed for pump Display names, access codes, controls and indicators referenced in this manual, for
example, GENERAL OPTIONS menu, access code 251, LOCK 1 keyswitch.
'Single quotes'Used to indicate cross-references made to another section of this manual. For example, see Chapter
2, 'Configuration & Calibration'.
underlineUsed to indicate links to another section of this manual.
ItalicsUsed to refer to other documents or manuals. For example, refer to the relevant Directions for Use (DFU)
for further information. Also used for emphasis, for example, ...position the narrow end of the tool...
Wherever this symbol is shown a Hints & Tips note is found. These notes provide useful advice or
information that may help to perform the task more effectively.
Wherever this symbol is shown a Toolbox note is found. These notes highlight an aspect of test or
maintenance that is important to know about. A typical example is drawing attention to a software
upgrade that should be checked that it has been installed.
Prior to using this pump, carefully read the Operating Precautions described in the Directions for Use (DFU).
w
This pump contains static-sensitive components. Observe strict precautions for the protection of static
sensitive components when attempting to repair and service the pump.
An explosion hazard exists if the pump is used in the presence of flammable anaesthetics. Exercise care to
locate the pump away from any such hazardous sources.
Dangerous Voltage. An electrical shock hazard exists if the casing of the pump is opened or removed. Refer
all servicing to qualified service personnel.
This pump is protected against the effects of high energy radio frequency emissions and is designed to
be fail safe if extremely high levels of interference are encountered. Should false alarm conditions be
M
encountered, either remove the source of the interference or regulate the infusion by another appropriate
means.
If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise
suspected to have been damaged, remove it from service for inspection by qualified service personnel.
the pump is running on the internal
backup battery. When flashing,
indicates that the battery power is
low, with less than 30 minutes of use
remaining.
AC POWERWhen illuminated, indicates that the
pump is connected to an AC power
supply and the battery is being
charged.
PLUS/MINUS
BUTTONS
ARROW
BUTTONS
Use to move cursor and to increase
or decrease values shown on main
display.
Use as softkeys in conjunction with
the prompts shown on the display.
For example, to select the CALIBRATE
option.
W
HISTORYPress to display PCA demands
J
PRINTPress to print patient history.
LOCK 1Insert key into LOCK 1
OFF
SET
RUN
LOCK 2Insert key into LOCK 2 and
PURGE/
BOLUS
Press and hold both buttons to
purge the extension set during
set up. See 'Basic Features' for
further information.
and drug infused history
graphs, 24 hour review and
event log.
Note: A suitable printer must be
connected to the pump.
keyswitch and turn key to
switch between OFF, SET and
RUN positions.
OFF - Turns the power off.
SET - Use to select or modify
protocols and to access
configuration and test routines.
RUN - Use to start the infusion.
Note: Switching from RUN
mode to SET mode without
first pressing the STOP button
automatically stops the
infusion.
turn key clockwise to open the
syringe cover.
This key lock is located
on the left side of the
pump
2. Squeeze the finger grips together on the plunger holder and slide the mechanism to the left.
3. Lift the syringe clamp and rotate to the left.
4. Insert the syringe into the slots on the plunger holder (see Figure 1).
5. Squeeze the finger grips on the plunger holder and slide the mechanism to the right until the syringe barrel flange locates into the
V slot (see Figure 2).
Ensure that the syringe is advanced until the syringe barrel flange touches the front of the V slot
closest to the syringe clamp. This is important to prevent delay at the start of the infusion.
6. Release the finger grips. Apply gentle pressure on the plunger holder to ensure that the drive is engaged.
7. Rotate the syringe clamp until it locks onto the syringe barrel (see Figure 2).
8. Check that the syringe plunger and syringe barrel flange are correctly located into their slots.
3. Load the syringe. See instructions in previous section.
4. Power the pump ON by switching LOCK 1 to the SET position.
5. NEW PATIENT? NO retains patient data then displays the last protocol used. YES clears previous patient data then displays preset
protocol A.
6. Check the protocol displayed. If required, select the NEXT PROTOCOL option to choose an alternative preset protocol, or select the MODIFY PROTOCOL option to adjust the current protocol. See 'Modifying a Preset Protocol' below for further information.
7. Switch LOCK 1 to the RUN position and remove the key.
8. CONFIRM PROTOCOL. Select OK.
9. CONFIRM SYRINGE. Select OK.
10. Purge (cover must be open): Press and hold the
11. Close the cover.
12. Connect the pump to test equipment as required (see Chapter 2, 'Configuration & Calibration' and Chapter 3, 'Routine
Maintenance').
buttons together.
W
13. Press the
H button to start operation.
Modifying a Preset Protocol
1. Switch LOCK 1 to the SET position. PROTOCOL SUMMARY is displayed.
2. Select the MODIFY PROTOCOL option (this option is not available when disabled). The current protocol parameters/values are
listed.
3. Use the
R Q buttons to move up/down the list of parameters. To change a parameter, highlight it and select ALTER. Use the R
Q buttons to toggle the values then select OK to confirm and return to the PROTOCOL SUMMARY. Note: A modified protocol
has no preset letter in the top left corner of the PROTOCOL SUMMARY.
Note: See 'Preset Protocol Setup' in Chapter 2 for details of protocol parameters.
If enabled, an additional 'generic' drug with parameter limits set to maximum values can be selected
when modifying protocols. Indicated by the drug name XXX DRUG, the drug can be selected
from the preset list of drug names.
buttons together to deliver a limited volume of fluid in order to purge
W
the extension line prior to being conected to the patient.
The purge feature cannot be activated when the cover is closed
Ensure the extension line is disconnected from the patient before purging the
line
Alarms are not disabled during a purge operation
Press and hold the T button for 2 seconds then enter the pre-programmed clinician over-ride
code to use this feature. Clinician over-ride can be used in RUN mode to administer an additional
bolus dose or a continuous background infusion of a limited dose and duration, for example, during the
PCA lock out period. It can also be used in SET mode to allow modification of the pre-set PCA Protocol
when this option has been disabled.
For further information, see 'Access Codes' in Chapter 2.
If the over-ride code is incorrectly entered more than three times, the event
CLIN. ACCESS TAMPER is logged in the event log and a warning appears on the
Display
Delivery of the clinician over-ride continuous infusion will automatically halt
while a Patient or Clinician over-ride bolus is being administered
To cancel clinician over-ride during delivery, press the STOP button then select
YES
Pressing the J button provides records of patient history and events since NEW PATIENT was last
selected:
Press x 1 to display an hour-by-hour record of the number of good/failed PCA demands and
the total drug infused over the last 24 hours
Press x 2 to display a graph of the good/failed PCA demands over the last 24 hours
Press x 3 times to view a graph of the total drug infused over the last 24 hours
Press x 4 times to view the event log
Print
Pressure
Maximum Dose
With a suitable printer connected (see 'Printer Set Up' on the following page), pressing the
button provides printouts of Patient History, Protocol Summary and the Event Log. Refer to the
DFU for detailed printing instructions.
Notes:
1) Access to the full Event Log can be enabled by entering access code 794.
2) Continuous printing can be configured, see 'General Options (251)'.
When enabled, this icon is shown on the Display. It provides a visual indicator of current
pumping pressure and pressure level at which the alarm will operate.
When enabled, this icon is shown on the Display. It provides a visual indication of the
H
amount of drug administered during the limit period (as shown to the left of the icon).
If the dose limit reaches the alarm level, the bottle icon will appear full, the pump will
stop infusing and the message Max Dose Limit is displayed. The icon will flash until the
dosing is less than the maximum dose limit. Clinician over-ride is always available.
To set up a printer, fit a suitable printer with a serial interface cable and connect to the pump. See list of recommended printers below.
See also PRINT button functions on previous page.
Canon Bubblejet (use cable 5000SP00008). Note: A serial to parallel adapter is required.
Continuous Printing
To configure continuous printing of events as they occur, enable the CONTINUOUS PRINT option in GENERAL OPTIONS.
Patient Hand Set
The electronic patient hand set is designed to be ambidextrous and suitable for both adult and paediatric use.
The indicator light can be configured to show when the PCA dose is available or being delivered. Should the clinical
situation require it, the indicator light can be disabled. The PCA button will provide feed-back on all, or just good
demands. See Chapter 2, 'General Options' for HANDSET MODE configuration details.
The patient handset does not contain any latex.
The hand set connector is a latching (but non-locking) connector. To remove, hold the body of the
connector and pull away from the pump. If required, the pump can be operated in continuous or
clinician over-ride modes without the hand set connected.
Note: An alarm warning will operate if the hand set is disconnected from the pump while it is in
operation, or when the handset is connected to the pump with the PCA button depressed.
The syringe pump software contains a number of configuration and test routines that can be accessed using a technical access code as
shown in the table below.
CodeTitleDescription
111*HOSPITAL NAMEConfigure name of hospital/ward to be displayed when pump is powered on and
when pump is in 'sleep' mode (when DISPLAY SLEEP is enabled).
To set the hospital name, use the
S button to adjust, then select OK to store.
the
123SELF TESTSelf test routine begins from the start. See Chapter 3, 'Routine Maintenance' for
further information.
124SELF TESTSelf test routine begins at internal PSU voltage test.
125SELF TESTSelf test routine begins at display test.
126SELF TESTSelf test routine begins at declutch test.
127SELF TESTSelf test routine begins at handset test.
167COMMS LEARN MODEConfiguration set via comms interface.
R Q buttons to toggle through characters and
168COMMS TEACH MODEConfiguration output to another device (pump).
243SYRINGE SIZE CALIBRATIONSyringe size measurement calibration. See 'Calibration Procedures' on the following
pages for instructions.
251CONFIGURATION OPTIONS
MENU
359BUILD CONFIGURATIONConfigure language and syringe range.
376SERVICE LOGReview and clear service log errors, hours of battery use and hours of pump on
501**MASS DOSINGEnable mass dosing. Drugs and protocols use mass dose mode.
502**VOLUME DOSINGEnable volume dosing. Drugs and protocols use a mix of mass and volume.
611TOTAL MEMORY CLEARReset memory. Note that the pump will require full calibration.
717OCCLUSION CALIBRATIONOcclusion calibration. See 'Calibration Procedures' on the following pages for
794FULL EVENT LOGEnable access to full Event Log.
835MODIFY CLINICIAN CODEChange the 3-digit clinician over-ride code.
***This option is not available on pumps with software version V2R8 and below.
These options are no longer in use on pumps with software version V3R2 and above. See DOSE MODE in 'Drugs and Safety
Limits' and MIX MASS & VOL MODES in 'General Options (251)'.
Configuration of drug limits, options, protocols and real-time clock. See
'Configuration Options (251)' on the following pages for further details.
time. To reset the service log to zero press the RESET softkey.
instructions.
Entering an Access Code
Each menu (and certain individual options) has its own three-digit technical access code which is entered using the following
procedure:
1. Hold down the
2. When the display shows ACCESS CODE 0 0 0, release the H button.
3. Enter the required access code "XXX" using the
move through the digits).
4. When the required code is shown, select the ENTER option to confirm.
R Q buttons in conjunction with the S button (to select the NEXT option to
Configuration & Calibration
Configuration Options (251)
Enter the access code 251 (see 'Entering an Access Code' on the previous page for instructions). The CONFIGURATION OPTIONS menu
is displayed:
CONFIGURATION OPTIONS MENU
OptionDescription
DRUG NAMES AND SAFETY LIMITSSet drug names and limits. See the next section for further details.
GENERAL OPTIONSSee 'General Options (251)' for further details.
PROTOCOL DEFAULT SET-UPSet default protocol and alter preset protocols. See 'Preset Protocol Set Up' for further
details.
CLOCK SETSet the internal clock.
To set the clock, use the
store.
The internal clock is the reference against which patient history and
events are stored. Patient history should always be recorded and if
required, printed prior to changing the clock.
Changing the clock will automatically:
Reset the time and date against which all new patient history is
Clear previous patient data. Only NEW PATIENT? YES will be
R Q buttons and the S button to adjust, then select OK to
stored and may affect the presentation of the history graphs.
1. Enter the access code 251 to display the CONFIGURATION OPTIONS menu.
2. Select DRUG NAMES AND SAFETY LIMITS using the
3. Select NEXT DRUG to choose another drug or select MODIFY DRUG to modify the current drug, as detailed below. Alternatively,
select QUIT to return to the CONFIGURATION OPTIONS menu.
4. Step through each DRUG option (see table below) and modify as required.
Use the R Q buttons to toggle/select a value then select OK to confirm and continue to the next option.
Use the S button at any time to go back to the previous option.
5. When set up is complete, switch LOCK 1 to the OFF position.
DRUG optionDescription
DRUG NAMEEnter drug name.
Use the RQ buttons in conjunction with the S button (to go to the next drug
name letter).
DOSE MODE*Select the dose mode:
VOLUMEVolume mode. Drugs and protocols use volume based units. Not
MASSMass mode. Drugs and protocols use mass based units only.
MINIMUM: DRUG CONC.Set the minimum drug concentration between 1 μg/ml and 999 μg/ml or 1.0 mg/ml
and 99.9 mg/ml. OFF is also available in Volume mode.
RQ buttons then select ENTER. DRUG 1 is displayed.
DRUG NAMES AND SAFETY LIMITS
available when MIX MASS & VOL MODES is disabled.
MAXIMUM: DRUG CONC.Set the maximum drug concentration between 1 μg/ml and 999 μg/ml or 1.0 mg/
ml and 99.9 mg/ml. Maximum value cannot be set below the minimum value.
MINIMUM: LOCKOUT PERIODSet the minimum lockout period (0 - 180 minutes).
MAXIMUM: LOCKOUT PERIODSet the maximum lockout period (0 - 180 minutes). Maximum period cannot be set
below minimum period.
MINIMUM: PCA DOSESet the minimum PCA dose. In Mass mode - between 0 μg and 999 μg or 1.0 mg
and 99.9 mg. In Volume mode, between 0.0 ml and 99.9 ml.
MAXIMUM: PCA DOSESet the maximum PCA dose. In Mass mode between 0 μg and 999 μg or 1.0 mg and
99.9 mg. Volume mode between 0.0 ml and 99.9 ml.
MAXIMUM: CONTINUOUSSet the maximum continuous dose. In Mass mode - between 0 μg/h and 999 μg/h
or 1.0 mg/h and 99.9 mg/h. In Volume mode - between 0.1 ml/h and 25.0 ml/h
(V2R8 or below), between 0.1 ml/h and 35.0 ml/h (V3R2 or above and for syringes
>50ml).
MAXIMUM: LOADING DOSESet the maximum loading dose. In Mass mode - between 0 μg and 999 μg or 1.0
mg and 99.9 mg. In Volume mode - between 0.0 ml and 99.9 ml.
MAXIMUM: MAX LIMITSet the maximum limit. In Mass mode - between 0 μg and 999 μg or 1.0 mg and
999 mg. In Volume mode - between 0.0 ml and 999 ml.
MAXIMUM: CLINICIAN BOLUSSet the maximum clinician bolus dose. In Mass mode - between 1 μg and 999 μg or
1.0 mg and 99.9 mg. In Volume mode - between 0.1 ml and 99.9 ml.
* This option is not available on pumps with software version V2R8 and below.
Note: For pumps with software version V2R8 or earlier, the options may vary, or will not be available. Refer to the
Nurse Call feature enabled (hardware feature allowing pump to communicate
with the hospital's nurse call system, typically linked to central nurse's station).
The number of protocols available for use is configured in GENERAL OPTIONS.
1. Enter the access code 251 to display the CONFIGURATION OPTIONS menu.
2. Select PROTOCOL DEFAULT SET-UP using the
3. Select NEXT PROTOCOL to choose another protocol or select MODIFY PROTOCOL to modify the current protocol, as detailed
below. Alternatively, select QUIT to return to the CONFIGURATION OPTIONS menu.
4. Use the
To enter a parameter, highlight it and select ALTER. Where relevant, the limits set in DRUG NAMES AND SAFETY LIMITS are
Use the R Q buttons to toggle/select a value then select OK to confirm. Alternatively, use CANCEL to quit.
Select OK at any time to return to PROTOCOL DEFAULT.
5. When set up is complete, switch LOCK 1 to the OFF position.
R Q buttons to move up/down the list of parameters (see table below).
displayed.
PROTOCOL parameterDescription
DRUG NAME
Select the drug. Use the
set up in DRUG NAMES AND SAFETY LIMITS).
If enabled, a generic drug is available, indicated by MASS DRUG or VOL DRUG
(only available when MIX MASS & VOL MODES option is enabled).
Important: The parameter limits of the generic drug are automatically set to the maximum
value.
RQ buttons then select ENTER. PROTOCOL DEFAULT A is displayed.
PROTOCOL DEFAULT SET-UP
R Q buttons to step through the names of the available drugs (as
DRUG CONC.Set the drug concentration between the minimum/maximum limits set for the selected drug.
PCA DOSESet the PCA dose between the minimum/maximum limits set for the selected drug.
LOCKOUT PERIODSet the lockout period between the minimum/maximum limits set for the selected drug.
OCCLUSION LEVELSet the occlusion level (L0 to L10).
CONTINUOUS*Set the continuous rate, below the maximum continuous rate limit set for the selected drug,
or 35ml/h (for syringes >50ml), 20ml/h (for syringes <50 ml and all syringes on pumps with
V2R8 or below), whichever is lowest.
LOADING DOSE*Set the loading dose, below the maximum loading dose set for the selected drug.
MAX LIMIT*Set maximum dose, below the maximum max dose set for the selected drug.
LIMIT DURATION*Set the maximum cumulative period (1 to 8 hours).
PCA DELIVERY*Set the PCA delivery rate. STAT rate 100 ml/h (80 ml/h for 20 ml syringes), or set by delivery
time (1 to 60 minutes).
* These options are not available if disabled.
Note: For pumps with software version V2R8 or earlier, the options may vary, or will not be available. Refer to the
To calibrate the syringe size detection system, follow the two-point calibration procedure described below.
Calibration tools required:
1000TG00055 (Syringe Sizing Spacer)
Calibration procedure:
Enter the access code 243.
Fit calibration tool into position on the pump and close the clamp, following Steps 1 - 3 below.
At each step, CALIBRATE is displayed if value is within tolerances.
Select CALIBRATE to store calibration point. Select NEXT to continue to the next screen.
Notes:
1) If CALIBRATE is not displayed, check for correct positioning of calibration tool. If calibration cannot be performed, repairs to
pump may be necessary.
2) The calibration values shown below are for illustrative purposes only.
1000TG00055
Step 1:
Step 2:
Position the narrow end of the calibration
tool under the syringe clamp
Re-position the calibration tool with the
wide end under the syringe clamp
243
243
243
SYRINGE CALIBRATION
Clamp down
SYRINGE CALIBRATION
Clamp down
SYRINGE CALIBRATION
Clamp down
Fit spacer
(40.. 070) : 060
CALIBRATE
(40.. 070) : 060
NEXT
(40.. 070) : 060
(170.. 210) : 186
CALIBRATE
243
SYRINGE CALIBRATION
Clamp down
Fit spacer
Difference
(40.. 070) : 060
(170.. 210) : 186
126
Step 3: Complete calibration
Switch LOCK 1 to the OFF position to complete the
calibration sequence.
Confirmatory Check - To confirm that the syringe sizing calibration has been performed correctly,
select a syringe (preferably 50ml), load and confirm the correct syringe type. Verify that the correct
syringe size is detected and displayed.
To set the pump occlusion alarm level, follow the calibration procedure described below. To test the occlusion alarm levels, see
'Occlusion Alarm Levels Test' in Chapter 3.
To convert Kilograms of Force (KgF) to Newtons (N) multiply by 9.806650. For example 10 KgF =
98.07N.
Excessive force will damage the plunger mechanism. Do not apply more than 10 KgF ±0.05
KgF to the plunger mechanism at any time.
Calibration tools required:
0000TG00020 (shown) or 0000TG00200 and 0000JG00014
Calibration procedure:
Enter the access code 717.
Select the OK option to start the drive, following Steps 1-5 below.
Notes:
1) If CALIBRATE does not appear in display, check for correct positioning of tool. If calibration cannot be performed, repairs to the
pump may be necessary.
2) The calibration values shown on the pump displays are for illustrative purposes only.
Replacement Power Supply PCBs are supplied with the charging voltage set to the correct value. The procedure below can be followed
in order to check the DC voltage setting if deemed necessary.
Equipment required: Oscilloscope
1. Ensure that the pump is swtiched OFF and is disconnected from the AC power supply.
2. Access the pump, see Chapter 6, 'Spare Parts Replacement Procedures' for instructions.
3. Disconnect the battery connector from the Power Supply PCB and connect the oscilloscope to PL3 (0V to pin 1).
4. Set the oscilloscope range to 0 to 7 VDC.
5. Connect the pump to AC power supply.
6. Adjust RV1 on the Power Supply PCB until the peak voltage level displayed on the oscilloscope is 7.0 ± 0.1 VDC.
7. Reseal RV1.
8. Disconnect the AC power supply, remove the oscilloscope connections and refit the battery connector.
Clearing Internal RAM (611)
Warning: Do not clear the RAM unless absolutely necessary because all the calibration and
configuration in the pump will be cleared.
If the internal RAM or its associated battery is replaced on the Control PCB, or if the pump fails with a 'CODE 5' RAM error it will be
necessary to do the following:
Clear the internal RAM:
1. Enter the access code 611.
2. Select ENTER and wait for the RAM to be cleared.
3. When the message RAM CLEARED appears, switch LOCK 1 to the OFF position.
Fully calibrate pump. Perform each of the two calibration procedures, as described in this chapter.
Reconfigure the pump:
Set Configuration, Drugs and Safety Limits and Protocols (it may be possible to use the Teach/Learn facility, see Chapter 3,
'Routine Maintenance' for instructions)
Enter the access code 359 and set syringe range and language
Enter the access code 376 and set service date
Carry out Performance Verification Procedure (PVP). See Chapter 3, 'Routine Maintenance' for instructions.
Potential Equalisation Terminal Resistance Test (PE Test) 35
Battery Maintenance 36
Physical Inspection and Clean 36
Performance Verification Procedure 37
Routine Maintenance
Introduction
For routine maintenance, the following tests and performance verification procedure should be performed in addition to the tasks
described in the section 'Physical Inspection and Clean'.
Refer to the relevant DFU for the recommended routine maintenance period.
Self-Test Procedure (123)
The self-test procedure is designed to allow confirmation of many of the pump functions, defaults and calibrations without requiring
internal inspection.
1. Enter the access code 123. See 'Entering an Access Code' in Chapter 2 for instructions.
2. The pump now proceeds through a series of tests. Press the NEXT softkey to move to the next test.
Refer to the table below for details of each test.
Important: If the pump fails the test sequence at any stage, it should be taken out of service and inspected by a qualified service
engineer.
TestDisplayDescription/Action
Review software setupSoftware revision
Program CRC
Language
Hospital name
Displays software version, program CRC, language and hospital name.
Review syringe dataSyr range
Syr cal
Occ cal
Internal PSU voltageInternal PSU voltage:
_ . _ _ V
Audible: AlarmAudible alarmCheck that the alarm sounds continuously.
Visual: Display and
Backlight
Touch panel buttonsPress S1Press buttons in turn from START (01) to PURGE (10).
Display test
Backlight
Displays syringe range, syringe and occlusion calibration figures.
On mains supply: 9.00V approximately
On battery supply: 5.20V - 6.50V
Check that the display dots are an even tone as they all go on full
backlight, then to dim, then turn off.
H 01S 06
I02J07
R 03V 08
Q 04G09
S 05T 10
Visual: LED IndicatorsLEDS flashingCheck that the STOP and START LED’S are flashing.
Declutch switchDeclutch:Squeeze plunger holder finger grips and check that the display
alternates between 1 (engaged) and 0 (disengaged - finger grips
squeezed together).
Near End of InfusionEOI opto:Starting with an empty, extended 50ml syringe on the pump, squeeze
the finger grips and move the plunger holder slowly to the left. Check the
display switches from 0 to 1. The position at which the display changes (EOI
point) depends on the type of syringe in use. See 'Setting the EOI Position' in
Chapter 2. for values.
Linear GridGrid opto:Squeeze the plunger holder grips together and slide the mechanism
to the right. Slowly move the syringe plunger to the left and check the
display alternates between 0 (OPTO over slot) and 1 (OPTO over bar).
Plunger DetectorPlunger opto:Press plunger plate button. Check the display switches from 1 (no syringe
plunger fitted) to 0 (syringe plunger fitted).
Motor EncoderMotor/encoders:Motor is pulsed while encoders are tested. Motor moves forwards and
backwards as encoders pass.
Cover DetectCover detect:Open and close the cover. Check that the display changes from 0 (cover
The Event Log can be downloaded directly to a PC using Microsoft HyperTerminal.
Note: The procedure below uses Microsoft HyperTerminal - Microsoft Windows XP Professional.
Equipment required:
RS232 cable, 9-pin D type
PC running Microsoft Windows
Cabling connections:
GND PWR1 TXD2 RXD2 ALM IN
5 4 3 2 15 4 3 2 1
6 7 8 9
NC NC CTS1 TXD1
PumpPC
MaleFemale
Pin 5
Pin 8
Pin 9
GNDPin 5GND
CTS1Pin 4PWR1
TXD1Pin 2RXD2
Pump Set Up
1. Connect the RS232 cable to the serial port of the pump and the COM port of the PC.
2. Enter access code 794. See 'Entering an Access Code' in Chapter 2 for instructions. This enables access to the full Event Log.
3. Turn the pump OFF.
4. Turn the pump ON. Select NO to retain previous patient data.
5. Press the
to be downloaded.
J button four times to display the Event Log. Use the RQ buttons to position the cursor at the start point of events
Set Up HyperTerminal on your PC
1. Open HyperTerminal on your PC. Click Start, point to All Programs, point to Accessories, point to Communications, and then
click HyperTerminal. Enter a HyperTerminal connection name, for example, PCAM ELDU, select an icon and click OK.
2. Select COM1 and click OK (leaving Country/region and Area code as default). Ensure COM1 is not already pre configured, for
example, pre configured for a hand-held PC. Configure COM1 Properties as shown below, click Apply then OK.
3. On the File menu, select Properties. Select the Settings tab then configure settingsas shown below. Click OK to close the
Properties window.
4. On the Transfer menu, select Capture Text. Enter file name and desired location for data to be saved to (example, D:\DATA\PCAM
EVENTLOG). Click Start.
5. Press the
button on the pump.
6. Once finished, the Event Log data will be displayed on the PC screen.
7. Save and Close HyperTerminal.
8. Locate the saved file (example D:\DATA\PCAM EVENTLOG) and open it using Wo rdPad. Save the WordPad document as a unicode
text document:
On the WordPad File menu, select Save As. Enter a File name (i.e. pump serial number). Select file type (Unicode Text Document)
from the drop-down box then click Save.