No p art o f this manu al may be reprod uced in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
U2761-90000
Edition
First Edition, April 30, 2008
Agilent Technologies, Inc.
3501 Stevens Creek Blvd.
Santa Clara, CA 95052 USA
Trademark Acknowledgements
Pentium is a U.S. registered trademark of
Intel Corporation.
Microsoft, Visual Studio, Windows, and MS
Windows are trademarks of Microsoft Corporation in the United States and/or other
countries.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to t he maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to
the federal government include only those
rights customarily provided to end user customers. Agilent provides this customary
commercial license in Software and technical data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer Software Documentation).
Safety Notices
CAUTION
A CAUTION notice denotes a haz-
ard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
IIU2761A User’s Guide
Page 3
Safety Symbols
Direct current (DC)Equipment protected throughout by
Alternating current (AC)Off (supply)
Both direct and alternating currentOn (supply)
Three-phase alternating currentCaution, risk of electric shock
Earth (ground) terminalCaution, risk of danger (refer to this
The following symbols on the instrument and in the documentation
indicate precautions which must be taken to maintain safe operation of
the instrument.
double insulation or reinforced
insulation
manual for specific Warning or Caution
information)
Protective conductor terminalCaution, hot surface
Frame or chassis terminalOut position of a bi-stable push control
EquipotentialityIn position of a bi-stable push control
U2761A User’s GuideIII
Page 4
General Safety Information
WARNING
CAUTION
The following general safety precautions must be observed during all
phases of operation of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates
safety standards of design, manufacture and intended use of the
instrument. Agilent Technologies Inc. assumes no liability for the
customer’s failure to comply with these requirements.
• Do not operate the product in an explosive atmosphere or in the
presence of flammable gases or fumes.
• Do not use the equipment if it does not operate properly. Have the
equipment inspected by qualified service personnel(s).
• Observe all markings on the instrument before connecting any wiring to
the instrument.
• Use the device with the cables provided.
• Repair or service that is not covered in this manual should only be
performed by qualified personnel(s).
IVU2761A User’s Guide
Page 5
Environment Conditions
CAUTION
This instrument is designed for indoor use and in the area with low
condensation. The table below shows the general environmental
requirements for this instrument.
Environment conditionsRequirements
Operating temperature0 °C to 50 °C
Operating humidity20 to 85% RH non-condensing
Storage temperature–20 °C to 70 °C
Storage humidity5 to 90% RH non-condensing
The U2761A USB modular function/arbitrary waveform generator
complies with the following safety and EMC requirements.
The CE mark is a registered trademark
of the European Community.This CE
mark shows that the product complies
with all the relevant European Legal
Directives.
ICES/NMB-001 indicates that this
ISM device complies with Canadian
ICES-001.
Cet appareil ISM est confomre a la
norme NMB-001 du Canada.
The CSA mark is a registered
trademark of the Canadian Standards
Association.
The C-tick mark is a registered
trademark of the Spectrum
Management Agency of Australia.
This signifies compliance with
the Australia EMC Framework
regulations under the terms of the
Radio Communication Act of 1992.
This instrument complies with the
WEEE Directive (2002/96/EC)
marking requirement. This affixed
product label indicates that you must
not discard this electrical/electronic
product in domestic household waste.
VIU2761A User’s Guide
Page 7
Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affixed product label indicates that you must not discard
this electrical/electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is shown as below:
Do not dispose in domestic household waste
To return this unwanted instrument, contact your nearest Agilent office, or
visit:
www.agilent.com/environment/product
for more information.
U2761A User’s GuideVII
Page 8
Declaration of Conformity
DECLARATION OF CONFORMITY
According to EN ISO/IEC 17050-1:2004
Generic example
Manufacturer’s Name:
Agilent Technologies Microwave Products (M) Sdn. Bhd
Manufacturer’s Address:
Bayan Lepas Free Industrial Zone,
11900, Bayan Lepas, Penang, Malaysia
Declares under sole responsibility that the product as originally delivered
Product Name:
Agilent USB Modular Function Generator
Models Number:
U2761A
Product Options:
This declaration covers all options of the above product(s)
complies with the essential requirements of the following applicable European Directives, and
carries the CE marking accordingly:
Low Voltage Directive (2006/95/EC)
EMC Directive (2004/108/EC)
and conforms with the following product standards:
EMC
Standard
StandardStandard
Standard Limit
LimitLimit
Limit
IEC 61326:2002 / EN 61326:1997+A1:1998+A2:2001+A3:2003
CISPR 11:1990 / EN55011:1990 Class A Group 1
IEC 61000-4-2:1995 / EN 61000-4-2:1995 4 kV CD, 8 kV AD
IEC 61000-4-3:1995 / EN 61000-4-3:1996 3 V/m, 80-1000 MHz
IEC 61000-4-4:1995 / EN 61000-4-4:1995 0.5 kV signal lines, 1 kV power lines
IEC 61000-4-5:1995 / EN 61000-4-5:1995 0.5 kV line-line, 1 kV line-ground
IEC 61000-4-6:1996 / EN 61000-4-6:1996 3 V, 0.15-80 MHz
IEC 61000-4-11:1994 / EN 61000-4-11:1994 1 cycle / 100%
The product herewith complies with the essential requirements of the Low Voltage Directive 2006/95/EC and the
EMC Directive 2004/108/EC and carries the CE Marking accordingly (European Union).
1
Performance Criteria:
A Pass - Normal ope ration, no effec t.
B Pass - Temporary degradation, self recoverable.
C Pass - Temporary degradation, operator intervention required.
D Fail - Not recoverable, component damage.
N/A – Not applicable
Notes:
Regulatory Information for Canada
ICES/NMB-001:2004
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est confomre à la norme NMB-001 du Canada.
Regulatory Information for Australia/New Zealand
This ISM device complies with Australian/New Zealand AS/NZS CISPR11:2004
U2761A User’s GuideIX
Page 10
In This Guide…
1Getting Started
In this chapter, you prepare your system for installation and configuration
to get started with the U2761A.
2Features and Functions
In this chapter, you will learn about the functions and features offered by
the U2761A.
3Characteristics and Specifications
In this chapter, you will observe the product characteristics and
specifications.
XU2761A User’s Guide
Page 11
Contents
List of FiguresXV
List of TablesXVII
1 Getting Started1
Introduction2
Product at a Glance3
Product Outlook3
Product Dimensions 5
Dimensions Without Bumpers5
Dimensions With Bumpers6
Standard Purchase Items7
Inspection and Maintenance 8
Initial Inspection8
Electrical Check8
General Maintenance8
Installation and Configuration9
Installation 9
A. Check Your System 10
B. Install the IO Libraries Suite11
C. Install the Module Driver 12
D. Install the Agilent Measurement Manager13
E. Connect the Module to Your PC 15
F. Verify Your Module Connection 19
G. Launch Your Agilent Measurement Manager 21
55-Pin Backplane Connector Pin Configuration 23
Chassis Installation 24
U2761A User's GuideXI
Page 12
Contents
2 Features and Functions 25
Output Configuration26
Introduction26
Output Function 26
Output Frequency29
Output Amplitude 31
Output Units35
Output Termination36
Duty Cycle (Square Waves)38
Symmetry (Ramp Wave) 40
Output Control 41
Set Output Using SCPI Commands42
To S e l ec t A M49
Carrier Waveform 50
Carrier Frequency 51
Modulating Waveform52
Modulating Waveform Frequency53
Modulation Depth53
Generate AM Using SCPI Commands54
Frequency Modulation (FM) 55
To S e l ec t F M55
Carrier Waveform 56
Carrier Frequency 57
Modulating Waveform58
XIIU2761A User's Guide
Page 13
Modulating Waveform Frequency59
Frequency Deviation59
Generate FM Using SCPI Commands 60
Phase Modulation (PM) 61
To S e l ec t P M61
Carrier Waveform 62
Carrier Frequency 63
Modulating Waveform64
Modulating Waveform Frequency64
Phase Deviation65
Generate PM Using SCPI Commands 65
Frequency-Shift Keying (FSK) Modulation 66
To S e l ec t F S K M od ul at i on66
Carrier Waveform 67
FSK Carrier Frequency68
FSK “Hop” Frequency69
Generate FSK Modulation Using SCPI Commands70
To Select ASK Modulation76
Carrier Waveform 77
Carrier Frequency 78
ASK Rate 79
Generate ASK Modulation Using SCPI Commands 79
Frequency Sweep 80
U2761A User's GuideXIII
Page 14
Contents
To Select Sweep80
Start Frequency and Stop Frequency 81
Sweep Mode 82
Sweep Time 82
Sweep Trigger Source 83
Set Frequency Sweep Using SCPI Commands 84
Tri gg er in g85
Trigger Source Choices85
Internal Triggering 86
Manual Triggering 86
External Triggering 87
Trigger Input Signal 88
Trigger Output Signal 88
Set Triggering Using SCPI Commands90
Arbitrary Waveforms 91
To Create and Store an Arbitrary Waveform91
3 Characteristics and Specifications 95
Product Characteristics96
Product Specifications and Characteristics 97
Index103
XIVU2761A User's Guide
Page 15
List of Figures
Figure 1-1 55-pin backplane connector pin configuration 23
Figure 2-1 U2761A soft front panel 28
Figure 2-2 Top panel view of the U2761A 29
Figure 2-3 Panel view of the frequency section 30
Figure 2-4 Panel view of the amplitude section32
Figure 2-5 Panel view of the DC offset section34
Figure 2-6 Panel view of the Tools menu37
Figure 2-7 Square wave duty cycles 38
Figure 2-8 Panel view of the duty cycle section39
Figure 2-9 Ramp wave duty cycles40
Figure 2-10 Panel view of the symmetry section40
Figure 2-11 Panel view of the output section41
Figure 2-12 Pulse waveform45
Figure 2-13 Panel view of the pulse width section 46
Figure 2-14 AM waveform49
Figure 2-15 Panel view of AM50
Figure 2-16 Panel view of the Arbitrary waveform51
Figure 2-17 FM waveform55
Figure 2-18 Panel view of FM56
Figure 2-19 PM waveform61
Figure 2-20 Panel view of PM62
Figure 2-21 FSK modulation waveform66
Figure 2-22 Panel view of FSK67
Figure 2-23 PSK modulation waveform 71
Figure 2-24 Panel view of PSK72
Figure 2-25 ASK modulation waveform76
Figure 2-26 Panel view of ASK 77
Figure 2-27 Frequency sweep80
Figure 2-28 Panel view of sweep81
Figure 2-29 Trigger input pulse 88
Figure 2-30 Trigger output pulse88
Figure 2-31 Ramp waveform91
U2761A User’s GuideXV
Page 16
XVIU2761A User’s Guide
Page 17
List of Tables
Ta b l e 1 - 1 Synchronous Simultaneous Interface (SSI) connector
pin description23
Ta b l e 2 - 1 Output functions26
Ta b l e 2 - 2 Output frequency range 29
Ta b l e 2 - 3 Carrier frequency for AM 51
Ta b l e 2 - 4 Carrier frequency for FM 57
Ta b l e 2 - 5 Carrier frequency for PM 63
Ta b l e 2 - 6 Carrier frequency for FSK 68
Ta b l e 2 - 7 “Hop” frequency 69
Ta b l e 2 - 8 Carrier frequency for PSK 73
Ta b l e 2 - 9 Carrier frequency for ASK 78
U2761A User’s GuideXVII
Page 18
XVIIIU2761A User’s Guide
Page 19
Agilent U2761A USB Modular Function Generator
User’s Guide
1
Getting Started
Introduction 2
Product at a Glance 3
Product Dimensions 5
Standard Purchase Items 7
Inspection and Maintenance 8
Installation and Configuration 9
This chapter provides the introduction of the U2761A that
helps you get acquainted with the product and the product
outlook. This chapter also includes the installation and
configuration procedures that will help you get started with
the U2761A.
1Agilent Technologies
Page 20
1Getting Started
Introduction
The U2761A is a 20 MHz USB modular function generator
with Arbitrary waveform and pulse generation capabilities. It
can operate as a standalone or modular unit when used
with the U2781A USB modular instrument chassis.
The U2761A adopts the latest direct digital synthesis (DDS)
technology that digitally creates Arbitrary waveforms and
frequencies from a single, fixed source frequency. DDS offers
the precision of digitally control logic; reducing the
complexity of the generator while increasing the stability.
Thus, creating a stable, accurate output signal for clean, low
distortion sine waves with fast rise and fall time up to
20 MHz and linear Ramp waves up to 200 kHz.
Various features of the U2761A
• 20 MHz Sine and Square waveforms
• Hi-Speed 2.0, USBTMC 488.2 standards
• Sine, Square, Ramp, Triangle, Pulse, and DC waveforms
• 40 mVpp to 5 Vpp amplitude range for 50 Ω load, and
80 mVpp to 10 Vpp amplitude range for open circuit
• Pulse generation
2U2761A User’s Guide
Page 21
Product at a Glance
Bumpers
Product Outlook
Getting Started1
Top V ie w
U2761A User’s Guide3
Page 22
1Getting Started
USB indicator
Power indicator
10 MHz reference connector
Trigger in connector
Trigger out connector
Output connector
Fan Ventilation
Power inlet
USB inlet
Fastening hole for USB cable with locking mechanism
55-pin backplane connector
Front View
Rear View
4U2761A User’s Guide
Page 23
Product Dimensions
Top V ie w
105.00 mm
175.00 mm
18.00 mm
Front View
25.00 mm
Dimensions Without Bumpers
Getting Started1
U2761A User’s Guide5
Page 24
1Getting Started
Top V ie w
117.00 mm
180.00 mm
15.00 mm
Front View
41.00 mm
Dimensions With Bumpers
6U2761A User’s Guide
Page 25
Standard Purchase Items
Verify the following items for the standard purchase of the
U2761A. If there are any missing or mechanically damaged
items, contact the nearest Agilent Sales Office.
✔ 12 V, 2 A AC/DC power adapter
✔ Power cord
✔ USB Standard-A to Mini-B interface cable
✔ L-Mount kit (used with modular instrument chassis)
✔ Agilent Automation Ready-CD (contains the IO Libraries
When you receive your U2761A, inspect the unit for any
obvious damage such as broken terminals or cracks, dents,
and scratches on the casing that may occur during shipment.
If any damage is found, notify the nearest Agilent Sales
Office immediately. The front of this manual contains the
warranty information.
Keep the original packaging in case the U2761A has to be
returned to Agilent in the future. If you return the U2761A
for service, attach a tag identifying the owner and model
number. Also include a brief description of the problem.
Electrical Check
The U2761A USB Modular Function/Arbitrary Waveform
Generator Service Guide will provide the complete
verification and calibration procedures. The procedures will
verify to a high level of confidence that the U2761A is
operating in accordance with its specifications.
General Maintenance
Any repair that is not covered in your modular product manuals should
only be performed by qualified personnel.
1 Power off your module and remove the power cord and
I/O cable from your device.
2 Remove your module from the bumper casing.
3 Shake off any dirt that may have accumulated on the
module.
4 Wipe your module with a dry cloth and install the
bumper back in place.
8U2761A User’s Guide
Page 27
Installation and Configuration
NOTE
A. Check Your System
B. Install the IO Libraries Suite
C. Install the Module Driver
D. Install the Agilent Measurement Manager
E. Connect the Module to Your PC
F. Verify Your Module Connection
G. Launch Your Agilent Measurement Manager
Installation
Follow the step-by-step instructions shown in the following
flowchart to get started with the preparations and
installations of your U2761A.
Getting Started1
You need to install the IVI-COM driver if you are going to use the U2761A
with Agilent VEE Pro, LabVIEW, or Microsoft
®
Visual Studio®.
U2761A User’s Guide9
Page 28
1Getting Started
A. Check Your System
Prior to any installation or configuration, please ensure that
your PC meets the following minimum system requirements.
Processor 1.6 GHz Pentium® IV or higher
Operating system Windows® XP Professional or Home Edition
(Service Pack 1 or later), or Windows® 2000 Professional
(Service Pack 4 or later)
Browser Microsoft® Internet Explorer 5.01 or higher
Available RAM 512 MB or higher recommended
Hard-disk space 1 GB
Video Super VGA 800×600 (1024×768 recommended)
Prerequisites Agilent IO Libraries Suite 14.2 or higher
(version 15.0
version 2.1
2
patch
, Microsoft® .NET Framework version 1.1 and 2.0
1 Available on Agilent Automation-Ready CD.
1
recommended), Agilent T&M Toolkit Runtime
2
, Agilent T&M Toolkit Redistributable Package 2.1
2
2 Bundled with Agilent Measurement Manager application software installer.
10U2761A User’s Guide
Page 29
B. Install the IO Libraries Suite
NOTE
NOTE
The IO Libraries Suite 14.2 or higher is available on the
Agilent Automation-Ready CD that comes with the standard
purchase of the U2761A.
• If you do not have the Agilent Automation-Ready CD, obtain the IO
Libraries Suite 14.2 or higher at http://www.agilent.com/find/iolib.
• Disconnect any USB instruments or connectivity interface from your
PC.
1 Close all other applications on your PC, insert the Agilent
Automation-Ready CD into your CD-ROM drive, and follow
the instructions on your screen.
2 If the IO Libraries Suite installation does not start
automatically, go to Start > Run (on the Windows Start
menu) and type <drive>:\autorun\auto.exe where
<drive>
3 If you obtain the IO Libraries Suite from the web, save
the self-extracting zip file (*.exe) to any location on your
hard disk.
4 Double-click the installation file to launch the installation.
5 Follow the instructions on your screen to proceed with
the installation.
6 After the installation has completed, you will see the IO
Control icon on the Windows taskbar notification area as
shown below.
is your CD-ROM drive location.
Getting Started1
For detailed installation instructions, refer to the Agilent IO Libraries Suite
Getting Started Guide at http://www.agilent.com/find/iolib.
U2761A User’s Guide11
Page 30
1Getting Started
NOTE
C. Install the Module Driver
Please ensure that there are no instruments connected to the PC when
installing the driver.
1 Verify that your PC meets the minimum system
requirements as stated in “A. Check Your System” on
page 10.
2 Insert the Product Reference CD-ROM into your CD-ROM
drive.
3 The installer will automatically launch the Agilent Modular
Products Installation Menu. Click Hardware Driver to begin the
installation.
4 If the menu does not launch automatically, go to Start >
Run (on the Windows Start menu) and type <drive>:\
Driver\Hardware\setup_Hw.exe where <drive>
your CD-ROM drive location. Click OK to begin the
installation.
5 Follow the instructions on the screen and click Next to
proceed.
12U2761A User’s Guide
is
Page 31
6 Click Install to begin the installation. Follow the
instructions on the screen to proceed with the
installation.
7 Click Finish when the installation has completed.
D. Install the Agilent Measurement Manager
1 If you have done “C. Install the Module Driver” on
page 12, proceed to Step 2. If not, close all other
applications on your PC and insert the Product Reference CD-ROM into your CD-ROM drive.
2 Click Measurement Manager on the U2761A Installation Menu to
begin the installation.
Getting Started1
3 If the installation menu does not appear after a few
seconds, go to Start > Run and type <drive>:\
Application\Modular Instruments Measurement
Manager\setup.exe where <drive> is your CD-ROM
drive location.
4 Click OK to begin the installation.
5 If you do not have any of the prerequisites installed, the
InstallShield Wizard software prerequisite will appear.
U2761A User’s Guide13
Page 32
1Getting Started
NOTE
6 Click OK to begin the installation of the listed missing
prerequisites.
7 Once the above installation has completed, installation of
the Measurement Manager software will proceed as
normal.
8 The Measurement Manager InstallShield Wizard dialog will
appear. Click Next to begin.
9 Read the License Agreement and select I accept the terms in
the License Agreement to proceed. You may click Print to
print a hardcopy of the Agilent License Terms for
reference. Click Next to proceed.
10 Fill in the Customer Information Form accordingly and
click Next.
11 Click Next to install to the specified folder or click Change
to install to a different folder.
12 Click Install to begin the installation of the Measurement
Manager.
13 Click Finish when the installation has completed.
14 A shortcut to this software will be created on your
desktop.
USING THE LICENSED MATERIALS INDICATES YOUR ACCEPTANCE OF
THE LICENSE TERMS. IF YOU DO NOT AGREE TO ALL OF THESE TERMS,
using
YOU MAY RETURN ANY UNOPENED LICENSED MATERIAL FOR A FULL
REFUND. IF THE LICENSED MATERIALS ARE BUNDLED OR PRELOADED
WITH ANOTHER PRODUCT, YOU MAY RETURN THE ENTIRE UNUSED
PRODUCT FOR A FULL REFUND.
14U2761A User’s Guide
Page 33
E. Connect the Module to Your PC
NOTE
Ensure that the Agilent Measurement Manager is installed before
proceeding.
1 After the installations have completed, connect the power
cord to the AC/DC power adapter. The AC/DC power
adapter requirements are 100 to 240 VAC, 50/60 Hz, with
an output voltage of +12 VDC.
2 Insert the DC output plug from the AC/DC power adapter
to the power jack on the rear panel of the U2761A.
3 Connect the U2761A to any USB port on your PC with
the bundled USB cable.
4 Your PC will automatically detect the connected unit and
the Found New Hardware Wizard window will appear. Select
Yes, this time only and click Next to proceed.
Getting Started1
U2761A User’s Guide15
Page 34
1Getting Started
5 Select Install the software automatically (Recommended) and
click Next.
6 A warning message will appear on the Hardware Installation
window. Click Continue Anyway to proceed with the
installation of the U2761A.
16U2761A User’s Guide
Page 35
Getting Started1
NOTE
If you do not wish to receive similar warning messages in the future,
follow the instructions below.
• Go to Start > Control Panel and double-click System.
• Select the Hardware tab and click Driver Signing on the Drivers panel. The Driver Signing Options dialog box will appear.
• Select Ignore to disable the warning message.
7 Click Finish to complete the installation.
U2761A User’s Guide17
Page 36
1Getting Started
8 The Assign USB device alias window will appear. Each time
the U2761A is plugged in, this dialog box will appear. To
disable this dialog, select the Never show this dialog option
in the Show this dialog panel and click
OK.
9 For modules other than the U2300A Series, U2500A
Series, U2600A Series, and U2781A, the system will
perform a firmware version check on your connected
module.
a If the module firmware version is the same as the
installed version on the PC, it will not perform any
firmware download and the U2761A is now ready for
use.
b If the module firmware version differs from the
installed version on the PC, the following message box
will appear. Firmware versions (V1.00 and V1.01) in the
figure are for illustration purpose only, it might vary
depending on the device and PC firmware version.
18U2761A User’s Guide
Page 37
Getting Started1
NOTE
c Click Ye s to begin the firmware download. The
following message box will appear indicating the
download in progress.
Ensure that you do not remove the USB and power connection until the
firmware download has completed.
d Your U2761A is ready for use once the firmware
download has completed.
F. Verify Your Module Connection
The Agilent Connection Expert is one of the utilities in the
IO Libraries. The Connection Expert configures the
connected instruments and enables communication. It is able
to automatically detect the U2761A devices plugged into the
PC.
1 Go to Start > All Programs > Agilent IO Libraries Suite > Agilent
Connection Expert to launch the Connection Expert.
2 The detected U2761A will be visible on the Instrument I/O
on this PC explorer pane. Right-click on the U2761A
instrument on the explorer pane.
U2761A User’s Guide19
Page 38
1Getting Started
3 A context menu will appear as shown below and select
Send Commands To This Instrument.
4 The Agilent Interactive IO dialog box will appear. Click Send &
Read to send the *IDN? default command. The instrument’s response should appear in the Instrument
Session History panel.
5 If the Connection Expert can successfully communicate
with the U2761A, this indicates that the instrument is
installed correctly.
20U2761A User’s Guide
Page 39
G. Launch Your Agilent Measurement Manager
NOTE
• The IO Control will launch automatically when you start your PC.
• Launching the Measurement Manager without the IO Control running
will cause failure of the Measurement Manager to detect or establish
any connection with the U2761A connected to your PC.
• To run the IO Control, go to Start > All Programs > Agilent IO
Libraries Suite > Utilities > IO Control.
1 Double-click the Measurement Manager software icon on
your desktop or go to Start > All Programs > Agilent > Modular
Products > Agilent Measurement Manager to launch the
software.
2 The Measurement Manager welcome screen will appear.
Getting Started1
U2761A User’s Guide21
Page 40
1Getting Started
NOTE
3 The Select USB Device dialog box will appear displaying the
connected U2761A devices. To start the application, select
a U2761A device and click OK to establish the connection.
For more information on how to use the Measurement Manager, refer to
the Agilent Measurement Manager help file.
22U2761A User’s Guide
Page 41
55-Pin Backplane Connector Pin Configuration
GND
NC
GND
TRIG4
nBPUB
NC
11
GND
NC
TRIG3
GND
CLK10M
NC
10
GND
NC
GND
TRIG5
GND
NC
9
GND
NC
STAR_TRIG
NC
8
TRIG2
GND
GND
NC
GND
TRIG6
GA2
NC
7
GND
NC
TRIG1
GND
GA1
NC
6
GND
NC
GND
TRIG7
GA0
NC
5
GND
NC
TRIG0
GND
NC
NC
4
GND
VBUS
GND
+12 V
+12 V
+12 V
3
GND
GND
GND
+12 V
+12 V
+12 V
2
GND
USB_D
–
USB_D+
GND
+12 V
+12 V
1
F
E
D
C
B
A
The 55-pin backplane connector is used when the U2761A
module is inserted into the U2781A USB modular instrument
chassis. For more details, refer to the Agilent U2781A USB Modular Instrument Chassis User's Guide.
Ta b l e 1 - 1 Synchronous Simultaneous Interface (SSI) connector pin
description
SSI timing signalFunctionality
GNDGround
NCNot connected
VBUSUSB bus power sensing input
USB_D+, USB_D–USB differential pair
TRIG0~TRIG7Trigger bus
+12 V+12 V power with 4 A current
nBPUBUSB backplane input detect
CLK10M10 MHz clock source
STAR_TRIGStar trigger
GA0,GA1,GA2Geographical address pin
U2761A User’s Guide23
Page 42
1Getting Started
Chassis Installation
The L-mount kit is to be installed to your U2761A module.
The following instructions describe the simple procedure of
installing the L-mount kit and your module in the U2781A
chassis.
1 Unpack the L-mount kit from its packaging.
2 Remove your U2761A module from the bumper casing.
3 Using a Phillips screwdriver, fasten the L-Mount kit to
your U2761A mo du le.
4 Insert your U2761A module into the U2781A chassis with
the 55-pin backplane connector positioned at the bottom
of the module.
5 Once you have slotted the module into the chassis, tighten
the screws of the L-mount kit to secure the connection.
24U2761A User’s Guide
Page 43
Agilent U2761A USB Modular Function Generator
User’s Guide
2
Features and Functions
Output Configuration 26
Pulse Waveform 45
Amplitude Modulation (AM) 49
Frequency Modulation (FM) 55
Phase Modulation (PM) 61
Frequency-Shift Keying (FSK) Modulation 66
Phase-Shift Keying (PSK) Modulation 71
Amplitude-Shift Keying (ASK) Modulation 76
Frequency Sweep 80
Tr ig g er i ng85
Arbitrary Waveforms 91
In this chapter, you will learn about the functions and
features of fered by the U2761A.
25Agilent Technologies
Page 44
2Features and Functions
Output Configuration
Introduction
Output Function
This section contains the information to help you configure
the U2761A for outputting waveforms. You may not need to
change some of the parameters discussed here, but they are
provided so that you will have accessibility when needed.
The U2761A can output five standard waveforms (Sine,
Square, Ramp, Triangle, and Pulse), and DC. You can select
one of the three built-in Arbitrary waveforms or creat e your
own custom waveforms. In addition, you can internally
modulate Sine, Square, Ramp, Triangle, and Arbitrary
waveforms using AM, FM, PM, FSK, PSK, or ASK. The linear
or logarithmic frequency sweeping is available for Sine,
Square, Ramp, Triangle, and Arbitrary waveforms.
The table below shows which output functions are allowed
with modulation and sweep. Each “V” indicates a valid
combination. If you change to a function that is not
applicable for modulation, or sweep; then the modulation or
mode will be disabled.
Ta b l e 2 - 1 Output functions
SineSquareRampTrianglePulseDCArbitrary
AM, FM, PM, FSK, PSK, ASK CarrierVVVVV
AM, FM, PM Internal ModulationVVVVV
FSK, PSK, ASK Internal ModulationV
Sweep ModeVVVVV
26U2761A User’s Guide
Page 45
Features and Functions2
Function Limitation
If you change to a function where the maximum frequency is
less than the current function, the frequency will be
adjusted to the maximum value for the new function.
For example, if you are currently outputting a 20 MHz sine
wave and then change to the Ramp function, the U2761A
will automatically adjusts the output frequency to 200 kHz
(the upper limit for Ramp).
Amplitude Limitation
If you change to a function where the maximum amplitude
is less than the current function, the amplitude will
automatically be adjusted to the maximum value for the new
function. This may occur when the output units are Vrms or
dBm due to the differences in crest factor for the various
output functions.
For example, if you output a 2.5 Vrms Square wave (into
50 Ω) and then change to the Sine wave function, the
U2761A will automatically be adjusted the output amplitude
to 1.768 Vrms (the upper limit for Sine wave in Vrms).
U2761A User’s Guide27
Page 46
2Features and Functions
Waveform pattern selection
Waveform parameters
Soft Front Panel Operation
The following figure shows the soft front panel of the
U2761A.
28U2761A User’s Guide
Figure 2-1U2761A soft front panel
Page 47
To select a function, click any of the functions on the top
panel as shown below. When a function is selected, the
button will be illuminated.
Figure 2-2Top panel view of the U2761A
Click Arb to output Arbitrary waveform. Select other
Arbitrary waveform choices from the drop down list
Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|PULSe|DC|USER}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
Output Frequency
Features and Functions2
.
As shown below, the output frequency range depends on the
function currently selected. The default frequency is 1 kHz for all functions.
Ta b l e 2 - 2 Output frequency range
FunctionMinimum frequencyMaximum frequency
Sine1 μHz20 MHz
Square1 μHz20 MHz
Ramp, Triangle1 μHz200 kHz
Pulse500 μHz5 MHz
DCNot applicableNot applicable
Arbitrary1 μHz200 kHz
U2761A User’s Guide29
Page 48
2Features and Functions
Function Limitations
If you change to a function where the maximum frequency is
less than the current function, the frequency will be
adjusted to the maximum value for the new function.
For example, if you are currently outputting a 20 MHz sine
wave and then change to the Ramp function, the U2761A
will automatically adjusts the output frequency to 200 kHz
(the upper limit for Ramp).
Duty Cycle Limitations
For Square waveforms, the U2761A may not be able to use
the full range of duty cycle values at higher frequencies as
shown below.
• 20% to 80% (frequency
• 40% to 60% (frequency
If you change to a frequency that cannot produce the
current duty cycle, the duty cycle is automatically adjusted
to the maximum value for the new frequency.
≤ 10 MHz)
> 10 MHz)
For example, if you currently have the duty cycle set to 70%
and then change the frequency to 12 MHz, the U2761A will
automatically adjusts the duty cycle to 60% (the upper limit
for t his frequency).
Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the
desired frequency value and select the unit from the drop
down list. To set the waveform period instead, select Period
.
Figure 2-3Panel view of the frequency section
30U2761A User’s Guide
Page 49
Remote Interface Operation
Vpp2V
maxVoffset
–()×≤
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
Output Amplitude
The default amplitude is 1 Vpp (into 50 Ω) for all functions.
Offset Voltage Limitations
The relation between the output amplitude and offset voltage
is shown below. Vmax is the maximum peak voltage for the
selected output termination (5 V for a 50 Ω load or 10 V for
a high-impedance load).
Features and Functions2
FREQuency {<frequency>}
Limits Due to Output Termination
If you change the output termination setting, the value of
the output amplitude will be adjusted (and no error will be
generated).
For example, if you set the amplitude to 5 Vpp and then
change the output termination from 50 Ω to “high
impedance”, the amplitude value will double to 10 Vpp. If
you changefrom “high impedance” to 50 Ω, the displayed
amplitude value will drop to half. For more information, see
“Output Termination” on page 36.
U2761A User’s Guide31
Page 50
2Features and Functions
Limits Due to Units Selection
In some cases, the amplitude limits are determined by the
output units selected. This may occur when the units are
Vrms or dBm due to the differences in crest factor for the
various output functions.
For example, if you output a 2.5 Vrms Square wave (into
50 Ω) and then change to the Sine wave function, the
U2761A will automatically adjusts the output amplitude to
1.768 Vrms (the upper limit for Sine wave in Vrms).
You can set the output amplitude in Vpp, Vrms, or dBm. For
more information, see “Output Units” on page 35.
You cannot specify the output amplitude in dBm if the
output termination is set to “high impedance”.
Soft Front Panel Operation
On the Amplitude panel as shown in Figure 2- 4, input the
desired amplitude value and select the unit from the drop
down list.
Figure 2-4Panel view of the amplitude section
32U2761A User’s Guide
Page 51
Remote Interface Operation
V
offsetVmax
V
pp
2
---------–≤
Or, you can set the amplitude by specifying a high level and
low level using the following commands.
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
DC Offset Voltage
The default offset is 0 V for all functions.
Limits Due to Amplitude
The relation between the offset voltage and output amplitude
is shown below. Vmax is the maximum peak voltage for the
selected output termination (5 V for a 50 Ω load or 10 V for
a high-impedance load).
Features and Functions2
VOLTage {<amplitude>}
VOLTage:HIGH {<voltage>}
VOLTage:LOW {<voltage>}
If the specified offset voltage is not valid, the U2761A will
automatically adjusts it to the maximum DC voltage allowed
with the specified amplitude.
U2761A User’s Guide33
Page 52
2Features and Functions
Soft Front Panel Operation
On the Offset panel as shown in Figure 2-5, input the
desired offset value and select the unit from the drop down
list.
Figure 2-5Panel view of the DC offset section
Remote Interface Operation
VOLTage:OFFSet {<offset>}
Or, you can set the offset by specifying a high level and low
level using the following commands.
VOLTage:HIGH {<voltage>}
VOLTage:LOW {<voltage>}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
34U2761A User’s Guide
Page 53
Output Units
Features and Functions2
This configuration applies to output amplitude only. At
power-on, the units for output amplitude are volts
peak-to-peak.
• The output units consist of Vpp, Vrms, or dBm. The
default unit is Vpp.
• The unit setting is stored in volatile memory. The units
are set to “Vpp” upon power-off or after a remote
interface reset.
• The output units for amplitude cannot be set to dBm if
the output termination is currently set to
“high impedance”. The units are automatically converted
to Vpp.
Soft Front Panel Operation
On the Amplitude panel as shown in Figure 2- 4, input the
desired amplitude value and select the unit from the drop
down list.
Remote Interface Operation
VOLTage:UNIT {VPP|VRMS|DBM}
U2761A User’s Guide35
Page 54
2Features and Functions
Output Termination
This configuration applies to output amplitude and offset
voltage only. The U2761A has a fixed series output
impedance of 50 Ω to the device output connector. If the
actual load impedance is different from the specified value,
the amplitude and offset levels will be incorrect.
• The range of the output termination is 1 Ω to 10 kΩ, or
Infinite. The default value is 50
• The output termination setting is stored in volatile
memory and upon power-off or after a remote interface
reset, the setting will return to a default value.
• If you specify a 50 Ω termination but are actually
terminating into an open circuit, the actual output will be
twice the value specified.
For example, if you set the offset to 100 mVDC (and
specify a 50 Ω load) but are terminating the output into
an open circuit, the actual offset will be 200 mVDC.
• If you change the output termination setting, the output
amplitude and offset levels are automatically adjusted (no
error will be generated).
For example, if you set the amplitude to 5 Vpp and then
change the output termination from 50 Ω to “high
impedance”, the amplitude value will double to 10 Vpp. If
you changefrom “high impedance” to 50 Ω, the displayed
amplitude value will drop to half.
• You cannot specify the output amplitude in dBm if the
output termination is currently set to “high impedance”.
The units are automatically converted to Vpp.
Ω
.
36U2761A User’s Guide
Page 55
Features and Functions2
Soft Front Panel Operation
Click Tools and select Waveform Gen as shown in the following.
Figure 2-6Panel view of the Tools menu
Then, select the Output Setup tab, input the desired load
impedance value on the Impedance Load panel and select
the unit from the drop down list, or select High Z for
high impedance load.
Remote Interface Operation
OUTPut:LOAD {<ohms>|INFinity|}
U2761A User’s Guide37
Page 56
2Features and Functions
20% Duty Cycle80% Duty Cycle
Duty Cycle (Square Waves)
The duty cycle of a Square wave represents the amount of
time per cycle that the Square wave is at a high level
(assuming that the waveform is not inverted).
Figure 2-7Square wave duty cycles
Refer to “Pulse Waveform” on page 45 for the information on
the duty cycle for Pulse waveforms.
Duty Cycle
• 20% to 80% (frequency ≤ 10 MHz)
• 40% to 60% (frequency > 10 MHz)
The duty cycle is stored in volatile memory where the duty
cycle is set to default 50% upon power-off or after a remote
interface reset.
The duty cycle setting is stored when you change from
Square wave to another function. Thus, when you return to
the Square function, the previous duty cycle is used.
Limits Due to Frequency
If the Square wave function is selected and you change to a
frequency that cannot produce the current duty cycle, the
duty cycle is automatically adjusted to the maximum value
for the new frequency.
For example, if you currently have the duty cycle set to 70%
and then change the frequency to 12 MHz, the U2761A will
automatically adjusts the duty cycle to 60% (the upper limit
for t his frequency).
38U2761A User’s Guide
Page 57
Features and Functions2
The duty cycle setting does not apply to a Square waveform
used as the modulating waveform for AM, FM, or PM. A 50%
duty cycle is always used for a modulating Square waveform.
The duty cycle setting applies only to a Square waveform
carrier.
Soft Front Panel Operation
After selecting Square at the top panel as shown in
Figure 2- 2, on the Duty Cycle panel below, input the desired
duty cycle value.
Figure 2-8Panel view of the duty cycle section
Remote Interface Operation
FUNCtion:SQUare:DCYCle{<percent>}
The APPLy command automatically sets the duty cycle to
50%.
U2761A User’s Guide39
Page 58
2Features and Functions
0% Symmetry100% Symmetry
Symmetry (Ramp Wave)
This configuration applies to Ramp wave only. Symmetry
represents the amount of time per cycle that the Ramp wave
is rising (assuming that the waveform is not inverted).
Figure 2-9Ramp wave duty cycles
• The symmetry percentage is stored in volatile memory
where the symmetry is set to default 100% upon power-off
or after a remote interface reset.
• The symmetry setting is stored when you change from
Ramp wave to another function. Thus, when you return to
the Ramp function, the previous symmetry is used.
• If you select a Ramp waveform as the modulating
waveform for AM, FM, or PM, the symm etry setting does
not apply.
Soft Front Panel Operation
After selecting Ramp at the top panel as shown in
Figure 2- 2, on the Symmetry panel below, input the desired
symmetry value.
Figure 2-10 Panel view of the symmetry section
40U2761A User’s Guide
Page 59
Output Control
Features and Functions2
Remote Interface Operation
FUNCtion:RAMP:SYMMetry{<percent>}
The APPLy command automatically sets the symmetry to
100%.
You can disable or enable the soft front panel output
control. By default, the output is disabled at power-on. When
enabled, the Output button is illuminated.
If an excessive external voltage is applied to the device
output connector, the output will be disabled. To re-enable
the output, remove the overload from the output connector
and click Output to enable the output.
Soft Front Panel Operation
Click Output to enable or disable the output as shown
in Figure 2- 11.
Figure 2-11 Panel view of the output section
U2761A User’s Guide41
Page 60
2Features and Functions
Set Output Using SCPI Commands
Remote Interface Operation
OUTPut {OFF|ON}
The APPLy command overrides the current setting and
automatically enables the output control.
The following SCPI commands show a sample procedure of
generating output.
Example 1, To output a DC voltage
-> *CLS; *RST//To reset the U2761A to default
power-on state, this command
can be ignored if this operation
is not required.
-> FUNC DC//Sets the output to DC.
-> VOLT:OFFS 1//Sets the output DC offset voltage
to 1 VDC.
-> OUTP:LOAD INF//Changes the output termination
to infinity.
-> OUTP ON //Turns on output.
42U2761A User’s Guide
Page 61
Features and Functions2
Example 2, To output a Sine wave
-> *CLS; *RST//To reset the U2761A to default
power-on state, this command can
be ignored if this operation is not
required.
-> FUNC SIN//Sets the output to Sine wave.
-> VOLT 5 VPP//Sets the output amplitude to 5 Vpp.
-> FREQ 1000//Sets the output frequency to 1 kHz.
-> VOLT:OFFS 0//Sets the output offset to 0.
-> OUTP ON //Turns on output.
Example 3, To output a Square wave
-> *CLS; *RST//To reset the U2761A to
default power-on state,
this command can be
ignored if this operation
is not required.
-> APPL:SQU 2000,5 VPP,0//Sets the output to Square
wave, frequency 2 kHz,
amplitude
5 Vpp, offset 0.
-> FUNC:SQU:DCYC 30//Changes the duty cycle to
30%.
-> OUTP ON //Turns on output.
U2761A User’s Guide43
Page 62
2Features and Functions
Example 4, To output a Ramp wave
-> *CLS; *RST//To reset the U2761A to default
power-on state, this command
can be ignored if this operation
is not required.
-> FUNC RAMP// Sets the output to Ramp wave.
-> VOLT 5 VPP//Sets the output amplitude to
5 Vpp.
-> FREQ 10000//Sets the output frequency to
10 kHz.
-> VOLT:OFFS 0//Sets the output offset to 0.
-> FUNC:RAMP:SYMM 50 //Changes the symmetry to 50%.
-> OUTP ON //Turns on output.
44U2761A User’s Guide
Page 63
Pulse Waveform
Pulse width
Period
90%90%
50%50%
10%
10%
Rise timeFall time
Features and Functions2
As shown below, a Pulse waveform consists of a period, a
pulse width, a rising edge, and a falling edge.
Figure 2-12 Pulse waveform
Pulse Period
The range of the pulse period is from 200 ns to 2000 s. The
default value is 1 ms. The U2761A will adjusts the pulse
width as necessary to accommodate the specified period.
Soft Front Panel Operation
After selecting Pulse at the top panel as shown in
Figure 2- 2, on the Frequency panel as shown in Figure 2- 3,
select Period, input the desired pulse period value,
and then select the unit from the drop down list.
U2761A User’s Guide45
Remote Interface Operation
PULSe:PERiod{<seconds>}
Page 64
2Features and Functions
Pulse Width
The pulse width represents the time from the 50% threshold
of the pulse’s rising edge to the 50% threshold of the next
falling edge.
• The range of the pulse width is 40 ns to <2000 s (see
restrictions below). The default pulse width is 500
• The minimum pulse width (Wmin) is affected by the
period.
Wmin = 40 ns for period ≤ 10 s
Wmin = 200 ns for period > 10 s, but ≤ 100 s
Wmin = 2 μs for period > 100 s, but ≤ 1000 s
Wmin = 20 μs for period > 1000 s
• The specified pulse width must also be less than the
difference between the period and the minimum pulse
width as shown in the equation below. The U2761A will
adjusts the pulse width as necessary to accommodate the
specified period.
Pulse Width
≤
Period – Wmin
μ
s.
Soft Front Panel Operation
After selecting Pulse at the top panel as shown in
Figure 2- 2, on the Width panel below, input the desired
pulse width value and select the unit from the drop down
list.
Figure 2-13 Panel view of the pulse width section
Remote Interface Operation
FUNCtion:PULSe:WIDTh{<seconds>}
46U2761A User’s Guide
Page 65
Pulse Duty Cycle
The pulse duty cycle is defined as:
Duty Cycle = 100
where the pulse width represents the time from the 50%
threshold of the rising edge of the pulse to the 50%
threshold of the next falling edge.
• Pulse duty cycle: >0% to <100% (see restrictions below).
The default is 50%.
• The specified pulse duty cycle must conform to the
following restrictions determined by the minimum pulse
width (Wmin). The U2761A will adjust the pulse duty
cycle as needed to accommodate the specified period.
and
where:
Features and Functions2
×
Pulse Width / Period
Duty Cycle
Duty Cycle
Wmin = 40 ns for period ≤ 10 s.
Wmin = 200 ns for period > 10 s, but ≤ 100 s.
Wmin = 2 μs for period > 100 s, but ≤ 1000 s.
Wmin = 20 μs for period > 1000 s.
≥
100 × Wmin / Period
≤
100 × (1 – Wmin / Period)
Soft Front Panel Operation
After selecting Pulse at the top panel as shown in
Figure 2- 2, select Duty Cycle as shown in
Figure 2- 13 and input the desired pulse duty cycle value on
the Duty Cycle panel.
Remote Interface Operation
FUNCtion:PULSe:DCYCle{<percent>}
U2761A User’s Guide47
Page 66
2Features and Functions
Generate Pulse Waveform Using SCPI Commands
Example 1
-> *CLS; *RST//To reset the U2761A to default
power-on state, this command
can be ignored if this operation
is not required.
-> FUNC PULS// Sets the output to Pulse wave.
-> VOLT 5 VPP//Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0//Sets the output offset to 0.
-> PULS:PER 1//Sets the pulse period to 1 s.
-> FUNC:PULS:DCYC 50 //Set the pulse duty cycle to 50%.
-> OUTP ON //Turns on output.
48U2761A User’s Guide
Page 67
Amplitude Modulation (AM)
Modulating frequency
AM depth
A modulated waveform consists of a carrier waveform and a
modulating waveform. An example of the AM waveform is
shown in Figure 2- 14. In AM, the amplitude of the carrier is
varied by the instantaneous voltage of the modulating
waveform. The amount of amplitude modulation is called the
modulation depth which refers to the portion of the
amplitude range that will be used by the modulation.
Features and Functions2
Figure 2-14 AM waveform
To Select AM
The U2761A allows only one modulation mode at a time. The
U2761A does not allow modulation when sweep is enabled.
U2761A User’s Guide49
Page 68
2Features and Functions
Soft Front Panel Operation
Click Mod and then select AM as shown in
Figure 2- 15. To output the AM waveform, configure the
settings for the carrier frequency, modulating frequency,
depth, output amplitude, offset voltage, and the desired
waveform.
Figure 2-15 Panel view of AM
Remote Interface Operation
AM:STATe {OFF|ON}
Carrier Waveform
The AM carrier waveform consists of Sine, Square, Ramp,
Triangle, or Arbitrary waveform. The default waveform is Sine wave. You cannot use Pulse or DC as the carrier
waveform.
Soft Front Panel Operation
For Arbitrary waveform, select Arb at the top panel
as shown in Figure 2-2 and then select the desired
waveform from the drop down list as shown in
Figure 2- 16.
50U2761A User’s Guide
Page 69
Features and Functions2
Figure 2-16 Panel view of the Arbitrary waveform
Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|USER}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
Carrier Frequency
The maximum carrier frequency depends on the selected
function as shown below. The default frequency is 1 kHz for
all functions.
Ta b l e 2 - 3 Carrier frequency for AM
FunctionMinimum frequencyMaximum frequency
Sine1 μHz20 MHz
Square1 μHz20 MHz
Ramp, Triangle1 μHz200 kHz
Arbitrary 1 μHz200 kHz
U2761A User’s Guide51
Page 70
2Features and Functions
Modulating Waveform
Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the
desired frequency value and select the unit from the drop
down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
The modulating waveform consists of Sine, Square, Ramp,
NegativeRamp(Nramp), Triangle, or Arbitrary waveform.
The default modulating waveform is Sine wave.
• Square has 50% duty cycle
• Ramp has 100% symmetry
• Triangle has 50% symmetry
• Negative Ramp has 0% symmetry
Soft Front Panel Operation
Select the desired modulating waveform from the drop down
list as shown in Figure 2- 15.
Remote Interface Operation
AM:INTernal:FUNCtion
{SINusoid|SQUare|RAMP|NRAMp|TRIangle|USER}
52U2761A User’s Guide
Page 71
Modulating Waveform Frequency
NOTE
The range of the modulating waveform frequency is 2 mHz
to 20 kHz. The default modulating waveform frequency is 100 Hz.
Soft Front Panel Operation
On the Mod Frequency panel as shown in Figure 2- 15, input
the desired modulating frequency value and select the unit
from the drop down list.
Remote Interface Operation
AM:INTernal:FREQuency {<frequency>}
Modulation Depth
The modulation depth is expressed in percentage and
represents the extent of the amplitude variation. At 0%
depth, the output amplitude is half of the selected value. At
100% depth, the output amplitude equals the selected value.
Features and Functions2
The range of the modulation depth is 0% to 100%. The
default modulation depth is 100%.
At 100% depth, the maximum output of the U2761A will not exceed ±2.5 V
peak (into a 50 Ω load).
Soft Front Panel Operation
On the Depth panel as shown in Figure 2- 15, select the
desired modulating depth value from the drop down list.
Remote Interface Operation
AM:DEPTh {<depth in percent>}
U2761A User’s Guide53
Page 72
2Features and Functions
Generate AM Using SCPI Commands
Example 1
-> *CLS; *RST//To reset the U2761A to default
power-on state, this command
can be ignored if this operation
is not required.
-> AM:STAT ON//Enables AM.
-> FUNC SQU//Sets the carrier waveform to
Square wave.
-> FREQ 2000// Sets the carrier frequency to
2 kHz.
-> VOLT 5 VPP//Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0//Sets the output offset to 0.
-> AM:INT:FUNC SIN//Changes the modulating
waveform shape to Sine wave.
-> AM:INT:FREQ 500//Changes the modulating
frequency to 500 Hz.
-> AM:DEPT 50//Changes the modulation depth
to 50%.
-> OUTP ON//Turns on output.
54U2761A User’s Guide
Page 73
Frequency Modulation (FM)
Modulating frequency
FM deviation
A modulated waveform consists of a carrier waveform and a
modulating waveform. Below shows an example of the FM
waveform. In FM, the frequency of the carrier is varied by
the instantaneous voltage of the modulating waveform. The
variation in frequency of the modulated waveform from the
carrier frequency is called the frequency deviation.
Features and Functions2
Figure 2-17 FM waveform
To Selec t FM
The U2761A allows only one modulation mode to be enabled
at a time. The U2761A does not allow FM to be enabled at
the same time that sweep is enabled.
U2761A User’s Guide55
Page 74
2Features and Functions
Soft Front Panel Operation
Click Mod and then select FMas shown in
Figure 2- 18. To output the FM waveform, configure the
settings for the carrier frequency, output amplitude, offset
voltage, modulating frequency, deviation, and the desired
waveform.
Figure 2-18 Panel view of FM
Remote Interface Operation
FM:STATe {OFF|ON}
Carrier Waveform
The FM carrier waveform consists of Sine, Square, Ramp,
Triangle, or Arbitrary waveform. The default waveform is Sine wave. You cannot use Pulse or DC as the carrier
waveform.
Soft Front Panel Operation
For Arbitrary waveforms, select Arb at the top panel
as shown in Figure 2-2 and select the desired waveform
from the drop down list as shown in
Figure 2- 16.
56U2761A User’s Guide
Page 75
Remote Interface Operation
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
Carrier Frequency
The maximum carrier frequency depends on the selected
function as shown below. The default is 1 kHz for all
functions.
Ta b l e 2 - 4 Carrier frequency for FM
FunctionMinimum frequencyMaximum frequency
Sine1 μHz20 MHz
Square1 μHz20 MHz
Features and Functions2
FUNCtion {SINusoid|SQUare|RAMP|USER}
Ramp, Triangle1 μHz200 kHz
Arbitrary1 μHz200 kHz
The carrier frequency must always be greater than or equal
to the frequency deviation plus 1 μHz. If you attempt to set
the deviation to a value plus 1 μHz greater than the carrier
frequency (with FM enabled), the U2761A will automatically
adjusts the deviation to the maximum value allowed with the
present carrier frequency.
The sum of the carrier frequency and deviation must be less
than or equal to the maximum frequency for the selected
function (20 MHz for Sine and Square, and 200 kHz for
Ramp and Arbitrary waveforms). If you attempt to set the
deviation to a value that is not valid, the U2761A will
automatically adjusts it to the maximum value allowed with
the present carrier frequency.
U2761A User’s Guide57
Page 76
2Features and Functions
Modulating Waveform
Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the
desired frequency value and select the unit from the drop
down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
The modulating waveform consists of Sine, Square, Ramp,
Negative Ramp (Nramp), Triangle, Noise, or Arbitrary
waveform. The default modulating waveform is Sine wave.
• Square has 50% duty cycle
• Ramp has 100% symmetry
• Triangle has 50% symmetry
• Negative Ramp has 0% symmetry
Soft Front Panel Operation
Select the desired modulating waveform from the drop down
list as shown in Figure 2-18.
Remote Interface Operation
FM:INTernal:FUNCtion
{SINusoid|SQUare|RAMP|NRAMp|TRIangle|USER}
58U2761A User’s Guide
Page 77
Modulating Waveform Frequency
The range of the modulating waveform frequency is 2 mHz
to 20 kHz. The default modulating waveform frequency is 100 Hz.
Soft Front Panel Operation
On the Mod Frequency panel as shown in Figure 2- 18, input
the desired modulating frequency value and select the unit
from the drop down list.
Remote Interface Operation
FM:INTernal:FREQuency {<frequency>}
Frequency Deviation
The frequency deviation setting represents the peak variation
in frequency of the modulated waveform from the carrier
frequency.
Features and Functions2
The range of the frequency deviation is 1 Hz to 500 kHz
(limited to 100 kHz minus 1 μHz for Ramp and Arbitrary
waveforms). The default frequency deviation is 100 Hz.
The carrier frequency must always be greater than or equal
to the frequency deviation plus 1 μHz. If you attempt to set
the deviation to a value plus 1 μHz greater than the carrier
frequency (with FM enabled), the U2761A will automatically
adjusts the deviation to the maximum value allowed with the
present carrier frequency.
The sum of the carrier frequency and deviation must be less
than or equal to the maximum frequency for the selected
function (20 MHz for Sine and Square, and 200 kHz for
Ramp and Arbitrary waveforms). If you attempt to set the
deviation to a value that is not valid, the U2761A will
automatically adjusts it to the maximum value allowed with
the present carrier frequency.
U2761A User’s Guide59
Page 78
2Features and Functions
Generate FM Using SCPI Commands
Soft Front Panel Operation
On the Deviation panel as shown in Figure 2- 18, input the
desired frequency deviation value and select the unit from
the drop down list.
Remote Interface Operation
FM:DEViation {<peak deviation in Hz>}
Example 1
-> *CLS; *RST//To reset the U2761A to default
power-on state, this command
can be ignored if this operation
is not required.
-> FM:STAT ON//Enables FM.
-> FUNC SIN//Sets the carrier waveform to
Sine wave.
-> FREQ 1000// Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP//Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0//Sets the output offset to 0.
-> FM:INT:FUNC SIN//Changes the modulating
waveform shape to Sine wave.
-> FM:INT:FREQ 500//Changes the modulating
frequency to 500 Hz.
-> FM:DEV 100//Changes the frequency
deviation to 100 Hz.
-> OUTP ON//Turns on output.
60U2761A User’s Guide
Page 79
Phase Modulation (PM)
Modulating frequency
Phase deviation
A modulated waveform consists of a carrier waveform and a
modulating waveform . The following figure shows an example
of a PM waveform. PM is very similar to FM, but in PM, the
phase of the modulated waveform is varied by the
instantaneous voltage of the modulating waveform. The
variation in phase of the modulated waveform from the
carrier waveform is called the phase deviation.
Features and Functions2
Figure 2-19 PM waveform
To Selec t P M
The U2761A allows only one modulation mode to be enabled
at a time. The U2761A does not allow PM to be enabled at
the same time that sweep is enabled.
U2761A User’s Guide61
Page 80
2Features and Functions
Soft Front Panel Operation
Click Mod and then select PM as shown in
Figure 2- 20. To output the PM waveform, configure the
settings for the carrier frequency, output amplitude, offset
voltage, modulating frequency, phase deviation, and the
desired waveform.
Figure 2-20 Panel view of PM
Remote Interface Operation
PM:STATe {OFF|ON}
Carrier Waveform
The PM carrier waveform consists of Sine, Square, Ramp,
Triangle, or Arbitrary waveform. The default waveform is Sine wave. You cannot use Pulse or DC as the carrier
waveform.
Soft Front Panel Operation
For Arbitrary waveforms, click Arb at the top panel
as shown in Figure 2-2 and select the desired waveform
from the drop down list as shown in Figure 2- 16.
62U2761A User’s Guide
Page 81
Remote Interface Operation
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
Carrier Frequency
The maximum carrier frequency depends on the function
selected as shown below. The default carrier frequency is 1 kHz for all functions.
Ta b l e 2 - 5 Carrier frequency for PM
FunctionMinimum frequencyMaximum frequency
Sine1 μHz20 MHz
Square1 μHz20 MHz
Features and Functions2
FUNCtion {SINusoid|SQUare|RAMP|USER}
Ramp, Triangle1 μHz200 kHz
Arbitrary1 μHz200 kHz
Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the
desired frequency value and select the unit from the drop
down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
U2761A User’s Guide63
Page 82
2Features and Functions
Modulating Waveform
The modulating waveform consists of Sine, Square, Ramp,
Negative Ramp (Nramp), Triangle, or Arbitrary waveform.
Thedefault waveform is Sine wave.
• Square has 50% duty cycle
• Ramp has 100% symmetry
• Triangle has 50% symmetry
• Negative Ramp has 0% symmetry
Soft Front Panel Operation
Select the desired modulating waveform from the drop down
list as shown in Figure 2- 20.
Remote Interface Operation
PM:INTernal:FUNCtion
{SINusoid|SQUare|RAMP|NRAMp|TRIangle|USER}
Modulating Waveform Frequency
The range of the modulating waveform frequency is 2 mHz
to 20 kHz. The default modulating waveform frequency is 10 Hz.
Soft Front Panel Operation
On the Mod Frequency panel as shown in Figure 2- 20, input
the desired modulating frequency value and select the unit
from the drop down list.
Remote Interface Operation
PM:INTernal:FREQuency {<frequency>}
64U2761A User’s Guide
Page 83
Phase Deviation
The phase deviation setting represents the peak variation in
phase of the modulated waveform from the carrier
waveform. The phase deviation setting can be set from
0 to 360 °. The default phase deviation setting is 180
Soft Front Panel Operation
On the Phase Deviation panel as shown in Figure 2- 20,
input the desired phase deviation value.
Remote Interface Operation
PM:DEViation {<deviation in degrees>}
Generate PM Using SCPI Commands
Example 1
Features and Functions2
°
.
-> *CLS; *RST//To reset the U2761A to default
power-on state, this command
can be ignored if this operation
is not required.
-> PM:STAT ON//Enables PM.
-> FUNC SIN//Sets the carrier waveform to
Sine wave.
-> FREQ 1000// Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP//Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0//Sets the output offset to 0.
-> PM:INT:FUNC SIN//Changes the modulating
waveform shape to Sine wave.
-> PM:INT:FREQ 500//Changes the modulating
frequency to 500 Hz.
-> PM:DEV 180//Changes the phase deviation to
180 °.
-> OUTP ON//Turns on output.
U2761A User’s Guide65
Page 84
2Features and Functions
FSK rate
Hop frequency
Frequency-Shift Keying (FSK) Modulation
FSK is similar to FM except that the frequency alternates
between two preset values. You can configure the U2761A to
“shift” its output frequency between two preset values using
FSK modulation. An example of the FSK modulation
waveform is shown below. The rate at which the output
shifts between the two frequencies (called the “carrier
frequency” and the “hop frequency”) is determined by the
internal rate generator.
Figure 2-21 FSK modulation waveform
To Select FSK Modulation
The U2761A allows only one modulation mode to be enabled
at a time. The U2761A does not allow FSK to be enabled at
the same time that sweep is enabled.
66U2761A User’s Guide
Page 85
Features and Functions2
Soft Front Panel Operation
Click Mod and select FSK as shown in
Figure 2- 22. To output the FSK waveform, configure the
settings for the carrier frequency, output amplitude, offset
voltage, “hop frequency”, and FSK rate.
Figure 2-22 Panel view of FSK
Remote Interface Operation
FSKey:STATe {OFF|ON}
Carrier Waveform
The FSK carrier waveform consists of Sine, Square, Ramp,
Triangle, or Arbitrary waveform. The default carrier waveform is Sine wave. You cannot use Pulse or DC as the
carrier waveform.
Soft Front Panel Operation
For Arbitrary waveforms, click Arb at the top panel
as shown in Figure 2-2 and select the desired waveform
from the drop down list as shown in
Figure 2- 16.
U2761A User’s Guide67
Page 86
2Features and Functions
FSK Carrier Frequency
Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|USER}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
The maximum carrier frequency depends on the function
selected as shown below. The default carrier frequency is 1 kHz for all functions.
Ta b l e 2 - 6 Carrier frequency for FSK
FunctionMinimum frequencyMaximum frequency
Sine1 μHz20 MHz
Square1 μHz20 MHz
Ramp, Triangle1 μHz200 kHz
Arbitrary1 μHz200 kHz
Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the
desired frequency value and select the unit from the drop
down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
68U2761A User’s Guide
Page 87
FSK “Hop” Frequency
The maximum alternate (or “hop”) frequency depends on the
function selected as shown below. The default “hop” frequency is 100 Hz for all functions.
Ta b l e 2 - 7 “Hop” frequency
FunctionMinimum frequencyMaximum frequency
Sine1 μHz20 MHz
Square1 μHz20 MHz
Ramp, Triangle1 μHz200 kHz
Arbitrary1 μHz200 kHz
Only Square wave with a 50% duty cycle is available for the
Internal FSK modulating waveform.
Soft Front Panel Operation
Features and Functions2
On the Hop Frequency panel as shown in Figure 2- 22, input
the desired hop frequency value and select the unit from the
drop down list.
Remote Interface Operation
FSKey:FREQuency {<frequency>}
U2761A User’s Guide69
Page 88
2Features and Functions
FSK Rate
Generate FSK Modulation Using SCPI Commands
The FSK rate is the rate at which the output frequency
“shifts” between the carrier frequency and the hop
frequency.
• The range of the FSK rate is2 mHz to 100 kHz. The
default FSK rate value is 10 Hz.
Soft Front Panel Operation
On the FSK Rate panel as shown in Figure 2- 22, input the
desired FSK rate value and select the unit from the drop
down list.
Remote Interface Operation
FSKey:INTernal:RATE {<rate in Hz>}
Example 1
-> *CLS; *RST//To reset the U2761A to default
power-on state, this command
can be ignored if this operation
is not required.
-> FSK:STAT ON//Enables FSK modulation.
-> FUNC SIN//Sets the carrier waveform to
Sine wave.
-> FREQ 1000// Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP//Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0//Sets the output offset to 0.
-> FSK:FREQ 100//Sets the "hop" frequency to
100 Hz.
-> FSK:INT:RATE 10//Sets the FSK rate to 10 Hz.
-> OUTP ON//Turns on output.
70U2761A User’s Guide
Page 89
Phase-Shift Keying (PSK) Modulation
PSK rate
PSK deviation
Phase-shift keying (PSK) is a form of digital modulation in
which the phase of the carrier signal is discretely varied.
You can configure the U2761A to “shift” its output phase
between two preset phases using PSK. The rate at which the
output shifts between the two phases is determined by the
internal rate generator on the signal level. The following
figure shows an example of the PSK modulation waveform.
Features and Functions2
Figure 2-23 PSK modulation waveform
To Select PSK Modulation
The U2761A allows only one modulation mode to be enabled
at a time. The U2761A does not allow PSK to be enabled at
the same time that sweep is enabled.
U2761A User’s Guide71
Page 90
2Features and Functions
Soft Front Panel Operation
Click Mod and then select PSK as shown in
Figure 2- 24. To output the PSK waveform, configure the
settings for the carrier frequency, output amplitude, offset
voltage, deviation, and PSK rate.
Figure 2-24 Panel view of PSK
Remote Interface Operation
PSKey:STATe {OFF|ON}
Carrier Waveform
The PSK carrier waveform consists of Sine, Square, Ramp,
Triangle, or Arbitrary waveform. The default waveform is Sine wave. You cannot use Pulse or DC as the carrier
waveform.
Soft Front Panel Operation
For Arbitrary waveforms, select Arb at the top panel
as shown in Figure 2-2 and select the desired waveform
from the drop down list as shown in
Figure 2- 16.
72U2761A User’s Guide
Page 91
Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|USER}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
PSK Carrier Frequency
The maximum carrier frequency depends on the function
selected as shown below. The default is 1 kHz for all functions.
Ta b l e 2 - 8 Carrier frequency for PSK
FunctionMinimum frequencyMaximum frequency
Sine1 μHz20 MHz
Square1 μHz20 MHz
Ramp, Triangle1 μHz200 kHz
Features and Functions2
Arbitrary1 μHz200 kHz
Only Square wave with a 50% duty cycle is available for the
Internal PSK modulating waveform.
Soft Front Panel Operation
On the Frequency panel shown in Figure 2- 3, input the
desired frequency value and select the unit from the drop
down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
U2761A User’s Guide73
Page 92
2Features and Functions
PSK Rate
PSK Deviation
The PSK rate is the rate at which the output phase “shifts”
between two preset phases. The range of the PSK rate is 2
mHz to 100 kHz. The default PSK rate value is 10 Hz.
Soft Front Panel Operation
On the PSK Rate panel as shown in Figure 2- 24, input the
desired PSK rate value and select the unit from the drop
down list.
Remote Interface Operation
PSKey:INTernal:RATE {<rate in Hz>}
The deviation setting represents the phase variation of the
shifted waveform from the carrier waveform. The deviation
can be set from 0 to 360 °. The default deviation setting is
180
°
.
Soft Front Panel Operation
On the Deviation panel as shown in Figure 2- 24, input the
desired deviation value.
Remote Interface Operation
PSKey:DEViation {<deviation in degrees>}
74U2761A User’s Guide
Page 93
Features and Functions2
Generate PSK Modulation Using SCPI Commands
Example 1
-> *CLS; *RST//To reset the U2761A to
default power-on state, this
command can be ignored if
this operation is not
required.
-> PSK:STAT ON//Enables PSK modulation.
-> FUNC SIN//Sets the carrier waveform to
Sine wave.
-> FREQ 1000//Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP//Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0//Sets the output offset to 0.
-> PSK:INT:RATE 10//Sets the PSK rate to
10 Hz.
-> PSK:DEV 180//Sets the PSK deviation to
180 °.
-> OUTP ON//Turns on output.
U2761A User’s Guide75
Page 94
2Features and Functions
ASK rate
Amplitude-Shift Keying (ASK) Modulation
ASK is a form of digital modulation in which the modulating
signal apply variations in the amplitude of a carrier signal.
The amplitude of the carrier signal varies simultaneously
with the modulating signal while phase and frequency
remain constant. In other words, the carrier signal can be
assumed as an on and off switch. You can configure the
U2761A to “shift” its output amplitude between two preset
amplitudes using ASK. The rate at which the output shifts
between the two amplitudes is determined by the internal
rate generator. An example of the ASK modulation waveform
is shown in the following.
Figure 2-25 ASK modulation waveform
To Select ASK Modulation
The U2761A allows only one modulation mode to be enabled
at a time. The U2761A does not allow ASK to be enabled at
the same time that sweep is enabled.
76U2761A User’s Guide
Page 95
Features and Functions2
Soft Front Panel Operation
Click Mod and then select ASK as shown in
Figure 2- 26. To output the ASK waveform, configure the
settings for the carrier frequency, output amplitude, offset
voltage, and ASK rate.
Figure 2-26 Panel view of ASK
Remote Interface Operation
ASKey:STATe {OFF|ON}
Carrier Waveform
The ASK carrier waveform consists of Sine, Square, Ramp,
Triangle, or Arbitrary waveform. The default waveform is Sine wave. You cannot use Pulse or DC as the carrier
waveform.
Soft Front Panel Operation
For Arbitrary waveforms, select Arb at the top panel
as shown in Figure 2-2 and select the desired waveform
from the drop down list as shown in
Figure 2- 16.
U2761A User’s Guide77
Page 96
2Features and Functions
Carrier Frequency
Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|USER}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
The maximum carrier frequency depends on the function
selected as shown below. The default is 1 kHz for all functions.
Ta b l e 2 - 9 Carrier frequency for ASK
FunctionMinimum frequencyMaximum frequency
Sine1 μHz20 MHz
Square1 μHz20 MHz
Ramp, Triangle1 μHz200 kHz
Arbitrary1 μHz200 kHz
Only Square wave with a 50% duty cycle is available for the
Internal ASK modulating waveform.
Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the
desired frequency value and select the unit from the drop
down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
78U2761A User’s Guide
Page 97
Features and Functions2
ASK Rate
The ASK rate is the rate at which the output amplitude
“shifts” between two preset amplitudes. The range of the
ASK rate is 2 mHz to 100 kHz. The default ASK rate value is 10 Hz.
Soft Front Panel Operation
On the ASK Rate panel as shown in Figure 2- 26, input the
desired ASK rate value and select the unit from the drop
down list.
Remote Interface Operation
ASKey:INTernal:RATE {<rate in Hz>}
Generate ASK Modulation Using SCPI Commands
Example 1
-> *CLS; *RST//To reset the U2761A to
default power-on state, this
command can be ignored if
this operation is not
required.
-> ASK:STAT ON//Enables ASK modulation.
-> FUNC SIN//Sets the carrier waveform to
Sine wave.
-> FREQ 1000//Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP// Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0//Sets the output offset to 0.
-> ASK:INT:RATE 10//Sets the ASK rate to
10 Hz.
-> OUTP ON//Turns on output.
U2761A User’s Guide79
Page 98
2Features and Functions
Frequency
Linear type
Start frequency
Stop frequency
Frequency Sweep
In the frequency sweep mode, the U2761A “steps” from the
start frequency to the stop frequency at a sweep rate which
you specify. You can sweep up or down in frequency, and
with either linear or logarithmic spacing. You can also
configure the U2761A to output a single sweep (one pass
from start frequency to stop frequency) by applying an
External or Manual (software) trigger. The U2761A can
produce a frequency sweep for Sine, Square, Ramp, Triangle,
or Arbitrary’s waveform (Pulse and DC are not allowed). The
figure below shows an example of the frequency sweep.
Figure 2-27 Frequency sweep
To Select Sweep
The U2761A does not allow the sweep mode to be enabled at
the same time that any modulation mode is enabled.
Soft Front Panel Operation
Click Sweep as shown in Figure 2- 28. To output
sweep, configure the settings for frequencies, output
amplitude, offset, sweep type and time, and trigger setup.
80U2761A User’s Guide
Page 99
Figure 2-28 Panel view of sweep
Remote Interface Operation
SWEep:STATe {OFF|ON}
Start Frequency and Stop Frequency
Features and Functions2
The start frequency and stop frequency set the upper and
lower frequency bounds for the sweep. The U2761A begins at
the start frequency, sweeps to the stop frequency, and then
resets back to the start frequency.
• The range of the start and stop frequencies is 1 μHz to
20 MHz (limited to 200 kHz for Ramp and Arbitrary
waveforms). The sweep is phase-continuous over the full
frequency range. The default start frequency is 100 Hz. The
default stop frequency is 1 kHz.
• To sweep up in frequency, set the start frequency < stop
frequency. To sweep down in frequency, set the start
frequency > stop frequency.
Soft Front Panel Operation
On the Start Frequency and Stop Frequency panels as
shown in Figure 2- 28, input the desired start frequency and stop frequency values and select the respective units from
the drop down lists.
U2761A User’s Guide81
Page 100
2Features and Functions
Sweep Mode
Remote Interface Operation
FREQuency:STARt {<frequency>}
FREQuency:STOP {<frequency>}
You can sweep with either linear or logarithmic spacing. For
a linear sweep, the U2761A varies the output frequency in a
linear fashion during the sweep. For a logarithmic sweep,
the U2761A varies the output frequency in a logarithmic
fashion.
The sweep mode consists of Linear or Logarithmic. The default sweep mode is Linear.
Soft Front Panel Operation
On the Sweep Type panel as shown in Figure 2- 28, select
the Linear or Log mode.
Remote Interface Operation
SWEep:SPACing {LINear|LOGarithmic}
Sweep Time
The sweep time specifies the number of seconds required to
sweep from the start frequency to the stop frequency. The
number of discrete frequency points in the sweep is
automatically calculated by the U2761A and is based on the
sweep time you have selected. The range of the sweep time
is 1 ms to 500 s. The default sweep time is 1 s.
82U2761A User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.