Agilent Technologies U2761A User Manual

Page 1
Agilent U2761A USB Modular Function/Arbitrary Waveform Generator
User’s Guide
Agilent Technologies
Page 2
Notices
No p art o f this manu al may be reprod uced in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
U2761-90000
Edition
First Edition, April 30, 2008
Agilent Technologies, Inc. 3501 Stevens Creek Blvd. Santa Clara, CA 95052 USA
Trademark Acknowledgements
Pentium is a U.S. registered trademark of Intel Corporation.
Microsoft, Visual Studio, Windows, and MS Windows are trademarks of Microsoft Cor­poration in the United States and/or other countries.
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to t he max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus­tomers. Agilent provides this customary commercial license in Software and techni­cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com­puter Software Documentation).
Safety Notices
CAUTION
A CAUTION notice denotes a haz- ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
II U2761A User’s Guide
Page 3

Safety Symbols

Direct current (DC) Equipment protected throughout by
Alternating current (AC) Off (supply)
Both direct and alternating current On (supply)
Three-phase alternating current Caution, risk of electric shock
Earth (ground) terminal Caution, risk of danger (refer to this
The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
double insulation or reinforced insulation
manual for specific Warning or Caution information)
Protective conductor terminal Caution, hot surface
Frame or chassis terminal Out position of a bi-stable push control
Equipotentiality In position of a bi-stable push control
U2761A User’s Guide III
Page 4

General Safety Information

WARNING
CAUTION
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture and intended use of the instrument. Agilent Technologies Inc. assumes no liability for the customer’s failure to comply with these requirements.
• Do not operate the product in an explosive atmosphere or in the presence of flammable gases or fumes.
• Do not use the equipment if it does not operate properly. Have the equipment inspected by qualified service personnel(s).
Observe all markings on the instrument before connecting any wiring to
the instrument.
Use the device with the cables provided.
Repair or service that is not covered in this manual should only be
performed by qualified personnel(s).
IV U2761A User’s Guide
Page 5

Environment Conditions

CAUTION
This instrument is designed for indoor use and in the area with low condensation. The table below shows the general environmental requirements for this instrument.
Environment conditions Requirements
Operating temperature 0 °C to 50 °C
Operating humidity 20 to 85% RH non-condensing
Storage temperature –20 °C to 70 °C
Storage humidity 5 to 90% RH non-condensing
The U2761A USB modular function/arbitrary waveform generator complies with the following safety and EMC requirements.
IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
Canada: CAN/CSA-C22.2 No. 61010-1-04
USA: ANSI/UL 61010-1:2004
IEC 61326-2002/EN 61326:1997+A1:1998+A2:2001+A3:2003
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS CISPR11:2004
U2761A User’s Guide V
Page 6

Regulatory Markings

The CE mark is a registered trademark of the European Community.This CE mark shows that the product complies with all the relevant European Legal Directives.
ICES/NMB-001 indicates that this ISM device complies with Canadian ICES-001. Cet appareil ISM est confomre a la norme NMB-001 du Canada.
The CSA mark is a registered trademark of the Canadian Standards Association.
The C-tick mark is a registered trademark of the Spectrum Management Agency of Australia. This signifies compliance with the Australia EMC Framework regulations under the terms of the Radio Communication Act of 1992.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.
VI U2761A User’s Guide
Page 7
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is shown as below:
Do not dispose in domestic household waste
To return this unwanted instrument, contact your nearest Agilent office, or visit:
www.agilent.com/environment/product
for more information.
U2761A User’s Guide VII
Page 8

Declaration of Conformity

DECLARATION OF CONFORMITY
According to EN ISO/IEC 17050-1:2004
Generic example
Manufacturer’s Name:
Agilent Technologies Microwave Products (M) Sdn. Bhd
Manufacturer’s Address:
Bayan Lepas Free Industrial Zone, 11900, Bayan Lepas, Penang, Malaysia
Declares under sole responsibility that the product as originally delivered
Product Name:
Agilent USB Modular Function Generator
Models Number:
U2761A
Product Options:
This declaration covers all options of the above product(s)
complies with the essential requirements of the following applicable European Directives, and carries the CE marking accordingly:
Low Voltage Directive (2006/95/EC) EMC Directive (2004/108/EC)
and conforms with the following product standards:
EMC
Standard
StandardStandard
Standard Limit
LimitLimit
Limit
IEC 61326:2002 / EN 61326:1997+A1:1998+A2:2001+A3:2003 CISPR 11:1990 / EN55011:1990 Class A Group 1 IEC 61000-4-2:1995 / EN 61000-4-2:1995 4 kV CD, 8 kV AD IEC 61000-4-3:1995 / EN 61000-4-3:1996 3 V/m, 80-1000 MHz IEC 61000-4-4:1995 / EN 61000-4-4:1995 0.5 kV signal lines, 1 kV power lines IEC 61000-4-5:1995 / EN 61000-4-5:1995 0.5 kV line-line, 1 kV line-ground IEC 61000-4-6:1996 / EN 61000-4-6:1996 3 V, 0.15-80 MHz IEC 61000-4-11:1994 / EN 61000-4-11:1994 1 cycle / 100%
Canada: ICES-001:2004 Australia/New Zealand: AS/NZS CISPR11:2004
The product was tested in a typical configuration with Agilent Technologies test systems.
Safety
IEC 61010-1:2001 / EN 61010-1:2001 Canada: CAN/CSA-C22.2 No. 61010-1-04 USA: ANSI/UL 61010-1:2004
206349
This DoC applies to above-listed products placed on the EU market after:
08-Jan-2008
Date
Tay Eng Su
Quality Manager
For further information, please contact your local Agilent Technologies sales office, agent or distributor,
or Agilent Technologies Deutschland GmbH, Herrenberger Straße 130, 71034 Böblingen, Germany.
Template: A5971-5302-2, Rev. E U2761A DoC Revision 1.0
VIII U2761A User’s Guide
Page 9
Product Regulations
Performance Criteria
IEC 61326-1:2002 / EN 61326-1:1997+A1:1998+A2:2001+A3:2003
CISPR 11:1990 / EN 55011:1990
– Group 1 Class A
IEC 61000-4-2:1995 / EN 61000-4-2:1995
(ESD 4kV CD, 8kV AD) A
IEC 61000-4-3:1995 / EN 61000-4-3:1996
(3V/m, 80% AM) A
IEC 61000-4-4:1995 / EN 61000-4-4:1995
(EFT 0.5kV line-line, 1kV line-earth) B
IEC 61000-4-5:1995 / EN 61000-4-5:1995
(Surge 0.5kV line-line, 1kV line-earth) A
IEC 61000-4-6:1996 / EN 61000-4-6:1996
(3V, 0.15~80 MHz, 80% AM, power line) A
IEC 61000-4-11:1994 / EN 61000-4-11:1994
(Dips 1 cycle, 100%) A Canada: ICES-001:2004 Australia/New Zealand: AS/NZS CISPR11:2004
EMC
IEC 61010-1:2001 / EN 61010-1:2001
Canada: CAN/CSA-C22.2 No. 61010-1-04
USA: ANSI/UL 61010-1:2004
Safety
Additional Information:
The product herewith complies with the essential requirements of the Low Voltage Directive 2006/95/EC and the EMC Directive 2004/108/EC and carries the CE Marking accordingly (European Union).
1
Performance Criteria:
A Pass - Normal ope ration, no effec t. B Pass - Temporary degradation, self recoverable. C Pass - Temporary degradation, operator intervention required. D Fail - Not recoverable, component damage. N/A – Not applicable
Notes:
Regulatory Information for Canada
ICES/NMB-001:2004 This ISM device complies with Canadian ICES-001. Cet appareil ISM est confomre à la norme NMB-001 du Canada.
Regulatory Information for Australia/New Zealand
This ISM device complies with Australian/New Zealand AS/NZS CISPR11:2004
U2761A User’s Guide IX
Page 10

In This Guide…

1 Getting Started
In this chapter, you prepare your system for installation and configuration to get started with the U2761A.
2 Features and Functions
In this chapter, you will learn about the functions and features offered by the U2761A.
3 Characteristics and Specifications
In this chapter, you will observe the product characteristics and specifications.
X U2761A User’s Guide
Page 11

Contents

List of Figures XV
List of Tables XVII
1 Getting Started 1
Introduction 2
Product at a Glance 3
Product Outlook 3
Product Dimensions 5
Dimensions Without Bumpers 5 Dimensions With Bumpers 6
Standard Purchase Items 7
Inspection and Maintenance 8
Initial Inspection 8 Electrical Check 8 General Maintenance 8
Installation and Configuration 9
Installation 9 A. Check Your System 10 B. Install the IO Libraries Suite 11 C. Install the Module Driver 12 D. Install the Agilent Measurement Manager 13 E. Connect the Module to Your PC 15 F. Verify Your Module Connection 19 G. Launch Your Agilent Measurement Manager 21 55-Pin Backplane Connector Pin Configuration 23 Chassis Installation 24
U2761A User's Guide XI
Page 12
Contents
2 Features and Functions 25
Output Configuration 26
Introduction 26 Output Function 26 Output Frequency 29 Output Amplitude 31 Output Units 35 Output Termination 36 Duty Cycle (Square Waves) 38 Symmetry (Ramp Wave) 40 Output Control 41 Set Output Using SCPI Commands 42
Pulse Waveform 45
Pulse Period 45 Pulse Width 46 Pulse Duty Cycle 47 Generate Pulse Waveform Using SCPI Commands 48
Amplitude Modulation (AM) 49
To S e l ec t A M 49 Carrier Waveform 50 Carrier Frequency 51 Modulating Waveform 52 Modulating Waveform Frequency 53 Modulation Depth 53 Generate AM Using SCPI Commands 54
Frequency Modulation (FM) 55
To S e l ec t F M 55 Carrier Waveform 56 Carrier Frequency 57 Modulating Waveform 58
XII U2761A User's Guide
Page 13
Modulating Waveform Frequency 59 Frequency Deviation 59 Generate FM Using SCPI Commands 60
Phase Modulation (PM) 61
To S e l ec t P M 61 Carrier Waveform 62 Carrier Frequency 63 Modulating Waveform 64 Modulating Waveform Frequency 64 Phase Deviation 65 Generate PM Using SCPI Commands 65
Frequency-Shift Keying (FSK) Modulation 66
To S e l ec t F S K M od ul at i on 66 Carrier Waveform 67 FSK Carrier Frequency 68 FSK “Hop” Frequency 69 Generate FSK Modulation Using SCPI Commands 70
Contents
Phase-Shift Keying (PSK) Modulation 71
To Select PSK Modulation 71 Carrier Waveform 72 PSK Carrier Frequency 73 PSK Rate 74 PSK Deviation 74 Generate PSK Modulation Using SCPI Commands 75
Amplitude-Shift Keying (ASK) Modulation 76
To Select ASK Modulation 76 Carrier Waveform 77 Carrier Frequency 78 ASK Rate 79 Generate ASK Modulation Using SCPI Commands 79
Frequency Sweep 80
U2761A User's Guide XIII
Page 14
Contents
To Select Sweep 80 Start Frequency and Stop Frequency 81 Sweep Mode 82 Sweep Time 82 Sweep Trigger Source 83 Set Frequency Sweep Using SCPI Commands 84
Tri gg er in g 85
Trigger Source Choices 85 Internal Triggering 86 Manual Triggering 86 External Triggering 87 Trigger Input Signal 88 Trigger Output Signal 88 Set Triggering Using SCPI Commands 90
Arbitrary Waveforms 91
To Create and Store an Arbitrary Waveform 91
3 Characteristics and Specifications 95
Product Characteristics 96
Product Specifications and Characteristics 97
Index 103
XIV U2761A User's Guide
Page 15
List of Figures
Figure 1-1 55-pin backplane connector pin configuration 23 Figure 2-1 U2761A soft front panel 28 Figure 2-2 Top panel view of the U2761A 29 Figure 2-3 Panel view of the frequency section 30 Figure 2-4 Panel view of the amplitude section 32 Figure 2-5 Panel view of the DC offset section 34 Figure 2-6 Panel view of the Tools menu 37 Figure 2-7 Square wave duty cycles 38 Figure 2-8 Panel view of the duty cycle section 39 Figure 2-9 Ramp wave duty cycles 40 Figure 2-10 Panel view of the symmetry section 40 Figure 2-11 Panel view of the output section 41 Figure 2-12 Pulse waveform 45 Figure 2-13 Panel view of the pulse width section 46 Figure 2-14 AM waveform 49 Figure 2-15 Panel view of AM 50 Figure 2-16 Panel view of the Arbitrary waveform 51 Figure 2-17 FM waveform 55 Figure 2-18 Panel view of FM 56 Figure 2-19 PM waveform 61 Figure 2-20 Panel view of PM 62 Figure 2-21 FSK modulation waveform 66 Figure 2-22 Panel view of FSK 67 Figure 2-23 PSK modulation waveform 71 Figure 2-24 Panel view of PSK 72 Figure 2-25 ASK modulation waveform 76 Figure 2-26 Panel view of ASK 77 Figure 2-27 Frequency sweep 80 Figure 2-28 Panel view of sweep 81 Figure 2-29 Trigger input pulse 88 Figure 2-30 Trigger output pulse 88 Figure 2-31 Ramp waveform 91
U2761A User’s Guide XV
Page 16
XVI U2761A User’s Guide
Page 17
List of Tables
Ta b l e 1 - 1 Synchronous Simultaneous Interface (SSI) connector
pin description 23
Ta b l e 2 - 1 Output functions 26 Ta b l e 2 - 2 Output frequency range 29 Ta b l e 2 - 3 Carrier frequency for AM 51 Ta b l e 2 - 4 Carrier frequency for FM 57 Ta b l e 2 - 5 Carrier frequency for PM 63 Ta b l e 2 - 6 Carrier frequency for FSK 68 Ta b l e 2 - 7 “Hop” frequency 69 Ta b l e 2 - 8 Carrier frequency for PSK 73 Ta b l e 2 - 9 Carrier frequency for ASK 78
U2761A User’s Guide XVII
Page 18
XVIII U2761A User’s Guide
Page 19
Agilent U2761A USB Modular Function Generator User’s Guide

1 Getting Started

Introduction 2 Product at a Glance 3 Product Dimensions 5 Standard Purchase Items 7 Inspection and Maintenance 8 Installation and Configuration 9
This chapter provides the introduction of the U2761A that helps you get acquainted with the product and the product outlook. This chapter also includes the installation and configuration procedures that will help you get started with the U2761A.
1Agilent Technologies
Page 20
1Getting Started

Introduction

The U2761A is a 20 MHz USB modular function generator with Arbitrary waveform and pulse generation capabilities. It can operate as a standalone or modular unit when used with the U2781A USB modular instrument chassis.
The U2761A adopts the latest direct digital synthesis (DDS) technology that digitally creates Arbitrary waveforms and frequencies from a single, fixed source frequency. DDS offers the precision of digitally control logic; reducing the complexity of the generator while increasing the stability. Thus, creating a stable, accurate output signal for clean, low distortion sine waves with fast rise and fall time up to 20 MHz and linear Ramp waves up to 200 kHz.
Various features of the U2761A
20 MHz Sine and Square waveforms
Hi-Speed 2.0, USBTMC 488.2 standards
Sine, Square, Ramp, Triangle, Pulse, and DC waveforms
14-bit, 50 MSa/s, 64 K-point Arbitrary waveforms
AM, FM, PM, ASK, FSK, and PSK modulation types
40 mVpp to 5 Vpp amplitude range for 50 Ω load, and
80 mVpp to 10 Vpp amplitude range for open circuit
Pulse generation
2 U2761A User’s Guide
Page 21

Product at a Glance

Bumpers

Product Outlook

Getting Started 1
Top V ie w
U2761A User’s Guide 3
Page 22
1Getting Started
USB indicator
Power indicator
10 MHz reference connector
Trigger in connector
Trigger out connector
Output connector
Fan Ventilation
Power inlet
USB inlet
Fastening hole for USB cable with locking mechanism
55-pin backplane connector
Front View
Rear View
4 U2761A User’s Guide
Page 23

Product Dimensions

Top V ie w
105.00 mm
175.00 mm
18.00 mm
Front View
25.00 mm

Dimensions Without Bumpers

Getting Started 1
U2761A User’s Guide 5
Page 24
1Getting Started
Top V ie w
117.00 mm
180.00 mm
15.00 mm
Front View
41.00 mm

Dimensions With Bumpers

6 U2761A User’s Guide
Page 25

Standard Purchase Items

Verify the following items for the standard purchase of the U2761A. If there are any missing or mechanically damaged items, contact the nearest Agilent Sales Office.
12 V, 2 A AC/DC power adapter
Power cord
USB Standard-A to Mini-B interface cable
L-Mount kit (used with modular instrument chassis)
Agilent Automation Ready-CD (contains the IO Libraries
Suite)
Agilent USB Modular Products Quick Start Guide
Agilent USB Modular Products Reference CD-ROM
Agilent Measurement Manager Quick Reference Card
Certificate of Calibration
Getting Started 1
U2761A User’s Guide 7
Page 26
1Getting Started
NOTE

Inspection and Maintenance

Initial Inspection

When you receive your U2761A, inspect the unit for any obvious damage such as broken terminals or cracks, dents, and scratches on the casing that may occur during shipment. If any damage is found, notify the nearest Agilent Sales Office immediately. The front of this manual contains the warranty information.
Keep the original packaging in case the U2761A has to be returned to Agilent in the future. If you return the U2761A for service, attach a tag identifying the owner and model number. Also include a brief description of the problem.

Electrical Check

The U2761A USB Modular Function/Arbitrary Waveform Generator Service Guide will provide the complete
verification and calibration procedures. The procedures will verify to a high level of confidence that the U2761A is operating in accordance with its specifications.

General Maintenance

Any repair that is not covered in your modular product manuals should only be performed by qualified personnel.
1 Power off your module and remove the power cord and
I/O cable from your device.
2 Remove your module from the bumper casing.
3 Shake off any dirt that may have accumulated on the
module.
4 Wipe your module with a dry cloth and install the
bumper back in place.
8 U2761A User’s Guide
Page 27

Installation and Configuration

NOTE
A. Check Your System
B. Install the IO Libraries Suite
C. Install the Module Driver
D. Install the Agilent Measurement Manager
E. Connect the Module to Your PC
F. Verify Your Module Connection
G. Launch Your Agilent Measurement Manager

Installation

Follow the step-by-step instructions shown in the following flowchart to get started with the preparations and installations of your U2761A.
Getting Started 1
You need to install the IVI-COM driver if you are going to use the U2761A with Agilent VEE Pro, LabVIEW, or Microsoft
®
Visual Studio®.
U2761A User’s Guide 9
Page 28
1Getting Started

A. Check Your System

Prior to any installation or configuration, please ensure that your PC meets the following minimum system requirements.
Processor 1.6 GHz Pentium® IV or higher
Operating system Windows® XP Professional or Home Edition
(Service Pack 1 or later), or Windows® 2000 Professional (Service Pack 4 or later)
Browser Microsoft® Internet Explorer 5.01 or higher
Available RAM 512 MB or higher recommended
Hard-disk space 1 GB
Video Super VGA 800×600 (1024×768 recommended)
Prerequisites Agilent IO Libraries Suite 14.2 or higher
(version 15.0 version 2.1
2
patch
, Microsoft® .NET Framework version 1.1 and 2.0
1 Available on Agilent Automation-Ready CD.
1
recommended), Agilent T&M Toolkit Runtime
2
, Agilent T&M Toolkit Redistributable Package 2.1
2
2 Bundled with Agilent Measurement Manager application software installer.
10 U2761A User’s Guide
Page 29

B. Install the IO Libraries Suite

NOTE
NOTE
The IO Libraries Suite 14.2 or higher is available on the Agilent Automation-Ready CD that comes with the standard purchase of the U2761A.
If you do not have the Agilent Automation-Ready CD, obtain the IO Libraries Suite 14.2 or higher at http://www.agilent.com/find/iolib.
Disconnect any USB instruments or connectivity interface from your PC.
1 Close all other applications on your PC, insert the Agilent
Automation-Ready CD into your CD-ROM drive, and follow
the instructions on your screen.
2 If the IO Libraries Suite installation does not start
automatically, go to Start > Run (on the Windows Start menu) and type <drive>:\autorun\auto.exe where
<drive>
3 If you obtain the IO Libraries Suite from the web, save
the self-extracting zip file (*.exe) to any location on your hard disk.
4 Double-click the installation file to launch the installation.
5 Follow the instructions on your screen to proceed with
the installation.
6 After the installation has completed, you will see the IO
Control icon on the Windows taskbar notification area as shown below.
is your CD-ROM drive location.
Getting Started 1
For detailed installation instructions, refer to the Agilent IO Libraries Suite Getting Started Guide at http://www.agilent.com/find/iolib.
U2761A User’s Guide 11
Page 30
1Getting Started
NOTE

C. Install the Module Driver

Please ensure that there are no instruments connected to the PC when installing the driver.
1 Verify that your PC meets the minimum system
requirements as stated in “A. Check Your System” on page 10.
2 Insert the Product Reference CD-ROM into your CD-ROM
drive.
3 The installer will automatically launch the Agilent Modular
Products Installation Menu. Click Hardware Driver to begin the
installation.
4 If the menu does not launch automatically, go to Start >
Run (on the Windows Start menu) and type <drive>:\
Driver\Hardware\setup_Hw.exe where <drive> your CD-ROM drive location. Click OK to begin the installation.
5 Follow the instructions on the screen and click Next to
proceed.
12 U2761A User’s Guide
is
Page 31
6 Click Install to begin the installation. Follow the
instructions on the screen to proceed with the installation.
7 Click Finish when the installation has completed.

D. Install the Agilent Measurement Manager

1 If you have done “C. Install the Module Driver” on
page 12, proceed to Step 2. If not, close all other applications on your PC and insert the Product Reference CD-ROM into your CD-ROM drive.
2 Click Measurement Manager on the U2761A Installation Menu to
begin the installation.
Getting Started 1
3 If the installation menu does not appear after a few
seconds, go to Start > Run and type <drive>:\ Application\Modular Instruments Measurement Manager\setup.exe where <drive> is your CD-ROM
drive location.
4 Click OK to begin the installation.
5 If you do not have any of the prerequisites installed, the
InstallShield Wizard software prerequisite will appear.
U2761A User’s Guide 13
Page 32
1Getting Started
NOTE
6 Click OK to begin the installation of the listed missing
prerequisites.
7 Once the above installation has completed, installation of
the Measurement Manager software will proceed as normal.
8 The Measurement Manager InstallShield Wizard dialog will
appear. Click Next to begin.
9 Read the License Agreement and select I accept the terms in
the License Agreement to proceed. You may click Print to
print a hardcopy of the Agilent License Terms for reference. Click Next to proceed.
10 Fill in the Customer Information Form accordingly and
click Next.
11 Click Next to install to the specified folder or click Change
to install to a different folder.
12 Click Install to begin the installation of the Measurement
Manager.
13 Click Finish when the installation has completed.
14 A shortcut to this software will be created on your
desktop.
USING THE LICENSED MATERIALS INDICATES YOUR ACCEPTANCE OF THE LICENSE TERMS. IF YOU DO NOT AGREE TO ALL OF THESE TERMS,
using
YOU MAY RETURN ANY UNOPENED LICENSED MATERIAL FOR A FULL REFUND. IF THE LICENSED MATERIALS ARE BUNDLED OR PRELOADED WITH ANOTHER PRODUCT, YOU MAY RETURN THE ENTIRE UNUSED PRODUCT FOR A FULL REFUND.
14 U2761A User’s Guide
Page 33

E. Connect the Module to Your PC

NOTE
Ensure that the Agilent Measurement Manager is installed before proceeding.
1 After the installations have completed, connect the power
cord to the AC/DC power adapter. The AC/DC power adapter requirements are 100 to 240 VAC, 50/60 Hz, with an output voltage of +12 VDC.
2 Insert the DC output plug from the AC/DC power adapter
to the power jack on the rear panel of the U2761A.
3 Connect the U2761A to any USB port on your PC with
the bundled USB cable.
4 Your PC will automatically detect the connected unit and
the Found New Hardware Wizard window will appear. Select
Yes, this time only and click Next to proceed.
Getting Started 1
U2761A User’s Guide 15
Page 34
1Getting Started
5 Select Install the software automatically (Recommended) and
click Next.
6 A warning message will appear on the Hardware Installation
window. Click Continue Anyway to proceed with the installation of the U2761A.
16 U2761A User’s Guide
Page 35
Getting Started 1
NOTE
If you do not wish to receive similar warning messages in the future, follow the instructions below.
Go to Start > Control Panel and double-click System.
Select the Hardware tab and click Driver Signing on the Drivers panel. The Driver Signing Options dialog box will appear.
Select Ignore to disable the warning message.
7 Click Finish to complete the installation.
U2761A User’s Guide 17
Page 36
1Getting Started
8 The Assign USB device alias window will appear. Each time
the U2761A is plugged in, this dialog box will appear. To disable this dialog, select the Never show this dialog option in the Show this dialog panel and click
OK.
9 For modules other than the U2300A Series, U2500A
Series, U2600A Series, and U2781A, the system will perform a firmware version check on your connected module.
a If the module firmware version is the same as the
installed version on the PC, it will not perform any firmware download and the U2761A is now ready for use.
b If the module firmware version differs from the
installed version on the PC, the following message box will appear. Firmware versions (V1.00 and V1.01) in the figure are for illustration purpose only, it might vary depending on the device and PC firmware version.
18 U2761A User’s Guide
Page 37
Getting Started 1
NOTE
c Click Ye s to begin the firmware download. The
following message box will appear indicating the download in progress.
Ensure that you do not remove the USB and power connection until the firmware download has completed.
d Your U2761A is ready for use once the firmware
download has completed.

F. Verify Your Module Connection

The Agilent Connection Expert is one of the utilities in the IO Libraries. The Connection Expert configures the connected instruments and enables communication. It is able to automatically detect the U2761A devices plugged into the PC.
1 Go to Start > All Programs > Agilent IO Libraries Suite > Agilent
Connection Expert to launch the Connection Expert.
2 The detected U2761A will be visible on the Instrument I/O
on this PC explorer pane. Right-click on the U2761A
instrument on the explorer pane.
U2761A User’s Guide 19
Page 38
1Getting Started
3 A context menu will appear as shown below and select
Send Commands To This Instrument.
4 The Agilent Interactive IO dialog box will appear. Click Send &
Read to send the *IDN? default command. The instrument’s response should appear in the Instrument Session History panel.
5 If the Connection Expert can successfully communicate
with the U2761A, this indicates that the instrument is installed correctly.
20 U2761A User’s Guide
Page 39

G. Launch Your Agilent Measurement Manager

NOTE
The IO Control will launch automatically when you start your PC.
Launching the Measurement Manager without the IO Control running
will cause failure of the Measurement Manager to detect or establish any connection with the U2761A connected to your PC.
To run the IO Control, go to Start > All Programs > Agilent IO Libraries Suite > Utilities > IO Control.
1 Double-click the Measurement Manager software icon on
your desktop or go to Start > All Programs > Agilent > Modular Products > Agilent Measurement Manager to launch the
software.
2 The Measurement Manager welcome screen will appear.
Getting Started 1
U2761A User’s Guide 21
Page 40
1Getting Started
NOTE
3 The Select USB Device dialog box will appear displaying the
connected U2761A devices. To start the application, select a U2761A device and click OK to establish the connection.
For more information on how to use the Measurement Manager, refer to the Agilent Measurement Manager help file.
22 U2761A User’s Guide
Page 41

55-Pin Backplane Connector Pin Configuration

GND
NC
GND
TRIG4
nBPUB
NC
11
GND
NC
TRIG3
GND
CLK10M
NC
10
GND
NC
GND
TRIG5
GND
NC
9
GND
NC
STAR_TRIG
NC
8
TRIG2
GND
GND
NC
GND
TRIG6
GA2
NC
7
GND
NC
TRIG1
GND
GA1
NC
6
GND
NC
GND
TRIG7
GA0
NC
5
GND
NC
TRIG0
GND
NC
NC
4
GND
VBUS
GND
+12 V
+12 V
+12 V
3
GND
GND
GND
+12 V
+12 V
+12 V
2
GND
USB_D
USB_D+
GND
+12 V
+12 V
1
F
E
D
C
B
A
The 55-pin backplane connector is used when the U2761A module is inserted into the U2781A USB modular instrument chassis. For more details, refer to the Agilent U2781A USB Modular Instrument Chassis User's Guide.
Getting Started 1
Figure 1-1 55-pin backplane connector pin configuration
Ta b l e 1 - 1 Synchronous Simultaneous Interface (SSI) connector pin
description
SSI timing signal Functionality
GND Ground
NC Not connected
VBUS USB bus power sensing input
USB_D+, USB_D– USB differential pair
TRIG0~TRIG7 Trigger bus
+12 V +12 V power with 4 A current
nBPUB USB backplane input detect
CLK10M 10 MHz clock source
STAR_TRIG Star trigger
GA0,GA1,GA2 Geographical address pin
U2761A User’s Guide 23
Page 42
1Getting Started

Chassis Installation

The L-mount kit is to be installed to your U2761A module. The following instructions describe the simple procedure of installing the L-mount kit and your module in the U2781A chassis.
1 Unpack the L-mount kit from its packaging.
2 Remove your U2761A module from the bumper casing.
3 Using a Phillips screwdriver, fasten the L-Mount kit to
your U2761A mo du le.
4 Insert your U2761A module into the U2781A chassis with
the 55-pin backplane connector positioned at the bottom of the module.
5 Once you have slotted the module into the chassis, tighten
the screws of the L-mount kit to secure the connection.
24 U2761A User’s Guide
Page 43
Agilent U2761A USB Modular Function Generator User’s Guide

2 Features and Functions

Output Configuration 26 Pulse Waveform 45 Amplitude Modulation (AM) 49 Frequency Modulation (FM) 55 Phase Modulation (PM) 61 Frequency-Shift Keying (FSK) Modulation 66 Phase-Shift Keying (PSK) Modulation 71 Amplitude-Shift Keying (ASK) Modulation 76 Frequency Sweep 80 Tr ig g er i ng 85 Arbitrary Waveforms 91
In this chapter, you will learn about the functions and features of fered by the U2761A.
25Agilent Technologies
Page 44
2Features and Functions

Output Configuration

Introduction

Output Function

This section contains the information to help you configure the U2761A for outputting waveforms. You may not need to change some of the parameters discussed here, but they are provided so that you will have accessibility when needed.
The U2761A can output five standard waveforms (Sine, Square, Ramp, Triangle, and Pulse), and DC. You can select one of the three built-in Arbitrary waveforms or creat e your own custom waveforms. In addition, you can internally modulate Sine, Square, Ramp, Triangle, and Arbitrary waveforms using AM, FM, PM, FSK, PSK, or ASK. The linear or logarithmic frequency sweeping is available for Sine, Square, Ramp, Triangle, and Arbitrary waveforms.
The table below shows which output functions are allowed with modulation and sweep. Each “V” indicates a valid combination. If you change to a function that is not applicable for modulation, or sweep; then the modulation or mode will be disabled.
Ta b l e 2 - 1 Output functions
Sine Square Ramp Triangle Pulse DC Arbitrary
AM, FM, PM, FSK, PSK, ASK Carrier VVVV V
AM, FM, PM Internal Modulation VVVV V
FSK, PSK, ASK Internal Modulation V
Sweep Mode VVVV V
26 U2761A User’s Guide
Page 45
Features and Functions 2
Function Limitation
If you change to a function where the maximum frequency is less than the current function, the frequency will be adjusted to the maximum value for the new function.
For example, if you are currently outputting a 20 MHz sine wave and then change to the Ramp function, the U2761A will automatically adjusts the output frequency to 200 kHz (the upper limit for Ramp).
Amplitude Limitation
If you change to a function where the maximum amplitude is less than the current function, the amplitude will automatically be adjusted to the maximum value for the new function. This may occur when the output units are Vrms or dBm due to the differences in crest factor for the various output functions.
For example, if you output a 2.5 Vrms Square wave (into 50 Ω) and then change to the Sine wave function, the U2761A will automatically be adjusted the output amplitude to 1.768 Vrms (the upper limit for Sine wave in Vrms).
U2761A User’s Guide 27
Page 46
2Features and Functions
Waveform pattern selection
Waveform parameters
Soft Front Panel Operation
The following figure shows the soft front panel of the U2761A.
28 U2761A User’s Guide
Figure 2-1 U2761A soft front panel
Page 47
To select a function, click any of the functions on the top panel as shown below. When a function is selected, the button will be illuminated.
Figure 2-2 Top panel view of the U2761A
Click Arb to output Arbitrary waveform. Select other Arbitrary waveform choices from the drop down list
Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|PULSe|DC|USER}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.

Output Frequency

Features and Functions 2
.
As shown below, the output frequency range depends on the function currently selected. The default frequency is 1 kHz for all functions.
Ta b l e 2 - 2 Output frequency range
Function Minimum frequency Maximum frequency
Sine 1 μHz 20 MHz
Square 1 μHz 20 MHz
Ramp, Triangle 1 μHz 200 kHz
Pulse 500 μHz 5 MHz
DC Not applicable Not applicable
Arbitrary 1 μHz 200 kHz
U2761A User’s Guide 29
Page 48
2Features and Functions
Function Limitations
If you change to a function where the maximum frequency is less than the current function, the frequency will be adjusted to the maximum value for the new function.
For example, if you are currently outputting a 20 MHz sine wave and then change to the Ramp function, the U2761A will automatically adjusts the output frequency to 200 kHz (the upper limit for Ramp).
Duty Cycle Limitations
For Square waveforms, the U2761A may not be able to use the full range of duty cycle values at higher frequencies as shown below.
20% to 80% (frequency
40% to 60% (frequency
If you change to a frequency that cannot produce the current duty cycle, the duty cycle is automatically adjusted to the maximum value for the new frequency.
10 MHz)
> 10 MHz)
For example, if you currently have the duty cycle set to 70% and then change the frequency to 12 MHz, the U2761A will automatically adjusts the duty cycle to 60% (the upper limit for t his frequency).
Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the desired frequency value and select the unit from the drop down list. To set the waveform period instead, select Period
.
Figure 2-3 Panel view of the frequency section
30 U2761A User’s Guide
Page 49
Remote Interface Operation
Vpp2 V
maxVoffset
()×
You can also use the APPLy command to select the function, frequency, amplitude, and offset.

Output Amplitude

The default amplitude is 1 Vpp (into 50 Ω) for all functions.
Offset Voltage Limitations
The relation between the output amplitude and offset voltage is shown below. Vmax is the maximum peak voltage for the selected output termination (5 V for a 50 Ω load or 10 V for a high-impedance load).
Features and Functions 2
FREQuency {<frequency>}
Limits Due to Output Termination
If you change the output termination setting, the value of the output amplitude will be adjusted (and no error will be generated).
For example, if you set the amplitude to 5 Vpp and then change the output termination from 50 Ω to “high impedance”, the amplitude value will double to 10 Vpp. If you change from “high impedance” to 50 Ω, the displayed amplitude value will drop to half. For more information, see
“Output Termination” on page 36.
U2761A User’s Guide 31
Page 50
2Features and Functions
Limits Due to Units Selection
In some cases, the amplitude limits are determined by the output units selected. This may occur when the units are Vrms or dBm due to the differences in crest factor for the various output functions.
For example, if you output a 2.5 Vrms Square wave (into 50 Ω) and then change to the Sine wave function, the U2761A will automatically adjusts the output amplitude to
1.768 Vrms (the upper limit for Sine wave in Vrms).
You can set the output amplitude in Vpp, Vrms, or dBm. For more information, see “Output Units” on page 35.
You cannot specify the output amplitude in dBm if the output termination is set to “high impedance”.
Soft Front Panel Operation
On the Amplitude panel as shown in Figure 2- 4, input the desired amplitude value and select the unit from the drop down list.
Figure 2-4 Panel view of the amplitude section
32 U2761A User’s Guide
Page 51
Remote Interface Operation
V
offsetVmax
V
pp
2
---------
Or, you can set the amplitude by specifying a high level and low level using the following commands.
You can also use the APPLy command to select the function, frequency, amplitude, and offset.

DC Offset Voltage

The default offset is 0 V for all functions.
Limits Due to Amplitude
The relation between the offset voltage and output amplitude is shown below. Vmax is the maximum peak voltage for the selected output termination (5 V for a 50 Ω load or 10 V for a high-impedance load).
Features and Functions 2
VOLTage {<amplitude>}
VOLTage:HIGH {<voltage>}
VOLTage:LOW {<voltage>}
If the specified offset voltage is not valid, the U2761A will automatically adjusts it to the maximum DC voltage allowed with the specified amplitude.
U2761A User’s Guide 33
Page 52
2Features and Functions
Soft Front Panel Operation
On the Offset panel as shown in Figure 2-5, input the desired offset value and select the unit from the drop down list.
Figure 2-5 Panel view of the DC offset section
Remote Interface Operation
VOLTage:OFFSet {<offset>}
Or, you can set the offset by specifying a high level and low level using the following commands.
VOLTage:HIGH {<voltage>}
VOLTage:LOW {<voltage>}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.
34 U2761A User’s Guide
Page 53

Output Units

Features and Functions 2
This configuration applies to output amplitude only. At power-on, the units for output amplitude are volts peak-to-peak.
The output units consist of Vpp, Vrms, or dBm. The default unit is Vpp.
The unit setting is stored in volatile memory. The units are set to “Vpp” upon power-off or after a remote interface reset.
The output units for amplitude cannot be set to dBm if the output termination is currently set to “high impedance”. The units are automatically converted to Vpp.
Soft Front Panel Operation
On the Amplitude panel as shown in Figure 2- 4, input the desired amplitude value and select the unit from the drop down list.
Remote Interface Operation
VOLTage:UNIT {VPP|VRMS|DBM}
U2761A User’s Guide 35
Page 54
2Features and Functions

Output Termination

This configuration applies to output amplitude and offset voltage only. The U2761A has a fixed series output impedance of 50 Ω to the device output connector. If the actual load impedance is different from the specified value, the amplitude and offset levels will be incorrect.
The range of the output termination is 1 Ω to 10 kΩ, or Infinite. The default value is 50
The output termination setting is stored in volatile memory and upon power-off or after a remote interface reset, the setting will return to a default value.
If you specify a 50 Ω termination but are actually terminating into an open circuit, the actual output will be twice the value specified. For example, if you set the offset to 100 mVDC (and specify a 50 Ω load) but are terminating the output into an open circuit, the actual offset will be 200 mVDC.
If you change the output termination setting, the output amplitude and offset levels are automatically adjusted (no error will be generated). For example, if you set the amplitude to 5 Vpp and then change the output termination from 50 Ω to “high impedance”, the amplitude value will double to 10 Vpp. If you change from “high impedance” to 50 Ω, the displayed amplitude value will drop to half.
You cannot specify the output amplitude in dBm if the output termination is currently set to “high impedance”. The units are automatically converted to Vpp.
Ω
.
36 U2761A User’s Guide
Page 55
Features and Functions 2
Soft Front Panel Operation
Click Tools and select Waveform Gen as shown in the following.
Figure 2-6 Panel view of the Tools menu
Then, select the Output Setup tab, input the desired load impedance value on the Impedance Load panel and select the unit from the drop down list, or select High Z for high impedance load.
Remote Interface Operation
OUTPut:LOAD {<ohms>|INFinity|}
U2761A User’s Guide 37
Page 56
2Features and Functions
20% Duty Cycle 80% Duty Cycle

Duty Cycle (Square Waves)

The duty cycle of a Square wave represents the amount of time per cycle that the Square wave is at a high level (assuming that the waveform is not inverted).
Figure 2-7 Square wave duty cycles
Refer to “Pulse Waveform” on page 45 for the information on the duty cycle for Pulse waveforms.
Duty Cycle
20% to 80% (frequency 10 MHz)
40% to 60% (frequency > 10 MHz)
The duty cycle is stored in volatile memory where the duty cycle is set to default 50% upon power-off or after a remote interface reset.
The duty cycle setting is stored when you change from Square wave to another function. Thus, when you return to the Square function, the previous duty cycle is used.
Limits Due to Frequency
If the Square wave function is selected and you change to a frequency that cannot produce the current duty cycle, the duty cycle is automatically adjusted to the maximum value for the new frequency.
For example, if you currently have the duty cycle set to 70% and then change the frequency to 12 MHz, the U2761A will automatically adjusts the duty cycle to 60% (the upper limit for t his frequency).
38 U2761A User’s Guide
Page 57
Features and Functions 2
The duty cycle setting does not apply to a Square waveform used as the modulating waveform for AM, FM, or PM. A 50% duty cycle is always used for a modulating Square waveform. The duty cycle setting applies only to a Square waveform carrier.
Soft Front Panel Operation
After selecting Square at the top panel as shown in
Figure 2- 2, on the Duty Cycle panel below, input the desired
duty cycle value.
Figure 2-8 Panel view of the duty cycle section
Remote Interface Operation
FUNCtion:SQUare:DCYCle{<percent>}
The APPLy command automatically sets the duty cycle to 50%.
U2761A User’s Guide 39
Page 58
2Features and Functions
0% Symmetry 100% Symmetry

Symmetry (Ramp Wave)

This configuration applies to Ramp wave only. Symmetry represents the amount of time per cycle that the Ramp wave is rising (assuming that the waveform is not inverted).
Figure 2-9 Ramp wave duty cycles
The symmetry percentage is stored in volatile memory where the symmetry is set to default 100% upon power-off or after a remote interface reset.
The symmetry setting is stored when you change from Ramp wave to another function. Thus, when you return to the Ramp function, the previous symmetry is used.
If you select a Ramp waveform as the modulating waveform for AM, FM, or PM, the symm etry setting does not apply.
Soft Front Panel Operation
After selecting Ramp at the top panel as shown in
Figure 2- 2, on the Symmetry panel below, input the desired
symmetry value.
Figure 2-10 Panel view of the symmetry section
40 U2761A User’s Guide
Page 59

Output Control

Features and Functions 2
Remote Interface Operation
FUNCtion:RAMP:SYMMetry{<percent>}
The APPLy command automatically sets the symmetry to 100%.
You can disable or enable the soft front panel output control. By default, the output is disabled at power-on. When enabled, the Output button is illuminated.
If an excessive external voltage is applied to the device output connector, the output will be disabled. To re-enable the output, remove the overload from the output connector and click Output to enable the output.
Soft Front Panel Operation
Click Output to enable or disable the output as shown in Figure 2- 11.
Figure 2-11 Panel view of the output section
U2761A User’s Guide 41
Page 60
2Features and Functions

Set Output Using SCPI Commands

Remote Interface Operation
OUTPut {OFF|ON}
The APPLy command overrides the current setting and automatically enables the output control.
The following SCPI commands show a sample procedure of generating output.
Example 1, To output a DC voltage
-> *CLS; *RST //To reset the U2761A to default
power-on state, this command can be ignored if this operation is not required.
-> FUNC DC //Sets the output to DC.
-> VOLT:OFFS 1 //Sets the output DC offset voltage
to 1 VDC.
-> OUTP:LOAD INF //Changes the output termination
to infinity.
-> OUTP ON //Turns on output.
42 U2761A User’s Guide
Page 61
Features and Functions 2
Example 2, To output a Sine wave
-> *CLS; *RST //To reset the U2761A to default
power-on state, this command can be ignored if this operation is not required.
-> FUNC SIN //Sets the output to Sine wave.
-> VOLT 5 VPP //Sets the output amplitude to 5 Vpp.
-> FREQ 1000 //Sets the output frequency to 1 kHz.
-> VOLT:OFFS 0 //Sets the output offset to 0.
-> OUTP ON //Turns on output.
Example 3, To output a Square wave
-> *CLS; *RST //To reset the U2761A to
default power-on state, this command can be ignored if this operation is not required.
-> APPL:SQU 2000,5 VPP,0 //Sets the output to Square
wave, frequency 2 kHz, amplitude 5 Vpp, offset 0.
-> FUNC:SQU:DCYC 30 //Changes the duty cycle to
30%.
-> OUTP ON //Turns on output.
U2761A User’s Guide 43
Page 62
2Features and Functions
Example 4, To output a Ramp wave
-> *CLS; *RST //To reset the U2761A to default
power-on state, this command can be ignored if this operation is not required.
-> FUNC RAMP // Sets the output to Ramp wave.
-> VOLT 5 VPP //Sets the output amplitude to
5 Vpp.
-> FREQ 10000 //Sets the output frequency to
10 kHz.
-> VOLT:OFFS 0 //Sets the output offset to 0.
-> FUNC:RAMP:SYMM 50 //Changes the symmetry to 50%.
-> OUTP ON //Turns on output.
44 U2761A User’s Guide
Page 63

Pulse Waveform

Pulse width
Period
90% 90%
50% 50%
10%
10%
Rise time Fall time
Features and Functions 2
As shown below, a Pulse waveform consists of a period, a pulse width, a rising edge, and a falling edge.
Figure 2-12 Pulse waveform

Pulse Period

The range of the pulse period is from 200 ns to 2000 s. The default value is 1 ms. The U2761A will adjusts the pulse
width as necessary to accommodate the specified period.
Soft Front Panel Operation
After selecting Pulse at the top panel as shown in
Figure 2- 2, on the Frequency panel as shown in Figure 2- 3,
select Period , input the desired pulse period value, and then select the unit from the drop down list.
U2761A User’s Guide 45
Remote Interface Operation
PULSe:PERiod{<seconds>}
Page 64
2Features and Functions

Pulse Width

The pulse width represents the time from the 50% threshold of the pulse’s rising edge to the 50% threshold of the next falling edge.
The range of the pulse width is 40 ns to <2000 s (see restrictions below). The default pulse width is 500
The minimum pulse width (Wmin) is affected by the period.
Wmin = 40 ns for period 10 s
Wmin = 200 ns for period > 10 s, but 100 s
Wmin = 2 μs for period > 100 s, but 1000 s
Wmin = 20 μs for period > 1000 s
The specified pulse width must also be less than the difference between the period and the minimum pulse width as shown in the equation below. The U2761A will adjusts the pulse width as necessary to accommodate the specified period.
Pulse Width
Period – Wmin
μ
s.
Soft Front Panel Operation
After selecting Pulse at the top panel as shown in
Figure 2- 2, on the Width panel below, input the desired
pulse width value and select the unit from the drop down list.
Figure 2-13 Panel view of the pulse width section
Remote Interface Operation
FUNCtion:PULSe:WIDTh{<seconds>}
46 U2761A User’s Guide
Page 65

Pulse Duty Cycle

The pulse duty cycle is defined as:
Duty Cycle = 100
where the pulse width represents the time from the 50% threshold of the rising edge of the pulse to the 50% threshold of the next falling edge.
Pulse duty cycle: >0% to <100% (see restrictions below). The default is 50%.
The specified pulse duty cycle must conform to the following restrictions determined by the minimum pulse width (Wmin). The U2761A will adjust the pulse duty cycle as needed to accommodate the specified period.
and
where:
Features and Functions 2
×
Pulse Width / Period
Duty Cycle
Duty Cycle
Wmin = 40 ns for period 10 s.
Wmin = 200 ns for period > 10 s, but 100 s.
Wmin = 2 μs for period > 100 s, but 1000 s.
Wmin = 20 μs for period > 1000 s.
100 × Wmin / Period
100 × (1 – Wmin / Period)
Soft Front Panel Operation
After selecting Pulse at the top panel as shown in
Figure 2- 2, select Duty Cycle as shown in Figure 2- 13 and input the desired pulse duty cycle value on
the Duty Cycle panel.
Remote Interface Operation
FUNCtion:PULSe:DCYCle{<percent>}
U2761A User’s Guide 47
Page 66
2Features and Functions

Generate Pulse Waveform Using SCPI Commands

Example 1
-> *CLS; *RST //To reset the U2761A to default
power-on state, this command can be ignored if this operation is not required.
-> FUNC PULS // Sets the output to Pulse wave.
-> VOLT 5 VPP //Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0 //Sets the output offset to 0.
-> PULS:PER 1 //Sets the pulse period to 1 s.
-> FUNC:PULS:DCYC 50 //Set the pulse duty cycle to 50%.
-> OUTP ON //Turns on output.
48 U2761A User’s Guide
Page 67

Amplitude Modulation (AM)

Modulating frequency
AM depth
A modulated waveform consists of a carrier waveform and a modulating waveform. An example of the AM waveform is shown in Figure 2- 14. In AM, the amplitude of the carrier is varied by the instantaneous voltage of the modulating waveform. The amount of amplitude modulation is called the modulation depth which refers to the portion of the amplitude range that will be used by the modulation.
Features and Functions 2
Figure 2-14 AM waveform

To Select AM

The U2761A allows only one modulation mode at a time. The U2761A does not allow modulation when sweep is enabled.
U2761A User’s Guide 49
Page 68
2Features and Functions
Soft Front Panel Operation
Click Mod and then select AM as shown in
Figure 2- 15. To output the AM waveform, configure the
settings for the carrier frequency, modulating frequency, depth, output amplitude, offset voltage, and the desired waveform.
Figure 2-15 Panel view of AM
Remote Interface Operation
AM:STATe {OFF|ON}

Carrier Waveform

The AM carrier waveform consists of Sine, Square, Ramp, Triangle, or Arbitrary waveform. The default waveform is Sine wave. You cannot use Pulse or DC as the carrier waveform.
Soft Front Panel Operation
For Arbitrary waveform, select Arb at the top panel as shown in Figure 2-2 and then select the desired waveform from the drop down list as shown in
Figure 2- 16.
50 U2761A User’s Guide
Page 69
Features and Functions 2
Figure 2-16 Panel view of the Arbitrary waveform
Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|USER}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.

Carrier Frequency

The maximum carrier frequency depends on the selected function as shown below. The default frequency is 1 kHz for
all functions.
Ta b l e 2 - 3 Carrier frequency for AM
Function Minimum frequency Maximum frequency
Sine 1 μHz 20 MHz
Square 1 μHz 20 MHz
Ramp, Triangle 1 μHz 200 kHz
Arbitrary 1 μHz 200 kHz
U2761A User’s Guide 51
Page 70
2Features and Functions

Modulating Waveform

Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the desired frequency value and select the unit from the drop down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.
The modulating waveform consists of Sine, Square, Ramp, Negative Ramp (Nramp), Triangle, or Arbitrary waveform. The default modulating waveform is Sine wave.
Square has 50% duty cycle
Ramp has 100% symmetry
Triangle has 50% symmetry
Negative Ramp has 0% symmetry
Soft Front Panel Operation
Select the desired modulating waveform from the drop down list as shown in Figure 2- 15.
Remote Interface Operation
AM:INTernal:FUNCtion
{SINusoid|SQUare|RAMP|NRAMp|TRIangle|USER}
52 U2761A User’s Guide
Page 71

Modulating Waveform Frequency

NOTE
The range of the modulating waveform frequency is 2 mHz to 20 kHz. The default modulating waveform frequency is 100 Hz.
Soft Front Panel Operation
On the Mod Frequency panel as shown in Figure 2- 15, input the desired modulating frequency value and select the unit from the drop down list.
Remote Interface Operation
AM:INTernal:FREQuency {<frequency>}

Modulation Depth

The modulation depth is expressed in percentage and represents the extent of the amplitude variation. At 0% depth, the output amplitude is half of the selected value. At 100% depth, the output amplitude equals the selected value.
Features and Functions 2
The range of the modulation depth is 0% to 100%. The default modulation depth is 100%.
At 100% depth, the maximum output of the U2761A will not exceed ±2.5 V peak (into a 50 Ω load).
Soft Front Panel Operation
On the Depth panel as shown in Figure 2- 15, select the desired modulating depth value from the drop down list.
Remote Interface Operation
AM:DEPTh {<depth in percent>}
U2761A User’s Guide 53
Page 72
2Features and Functions

Generate AM Using SCPI Commands

Example 1
-> *CLS; *RST //To reset the U2761A to default
power-on state, this command can be ignored if this operation is not required.
-> AM:STAT ON //Enables AM.
-> FUNC SQU //Sets the carrier waveform to
Square wave.
-> FREQ 2000 // Sets the carrier frequency to
2 kHz.
-> VOLT 5 VPP //Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0 //Sets the output offset to 0.
-> AM:INT:FUNC SIN //Changes the modulating
waveform shape to Sine wave.
-> AM:INT:FREQ 500 //Changes the modulating
frequency to 500 Hz.
-> AM:DEPT 50 //Changes the modulation depth
to 50%.
-> OUTP ON //Turns on output.
54 U2761A User’s Guide
Page 73

Frequency Modulation (FM)

Modulating frequency
FM deviation
A modulated waveform consists of a carrier waveform and a modulating waveform. Below shows an example of the FM waveform. In FM, the frequency of the carrier is varied by the instantaneous voltage of the modulating waveform. The variation in frequency of the modulated waveform from the carrier frequency is called the frequency deviation.
Features and Functions 2
Figure 2-17 FM waveform

To Selec t FM

The U2761A allows only one modulation mode to be enabled at a time. The U2761A does not allow FM to be enabled at the same time that sweep is enabled.
U2761A User’s Guide 55
Page 74
2Features and Functions
Soft Front Panel Operation
Click Mod and then select FM as shown in
Figure 2- 18. To output the FM waveform, configure the
settings for the carrier frequency, output amplitude, offset voltage, modulating frequency, deviation, and the desired waveform.
Figure 2-18 Panel view of FM
Remote Interface Operation
FM:STATe {OFF|ON}

Carrier Waveform

The FM carrier waveform consists of Sine, Square, Ramp, Triangle, or Arbitrary waveform. The default waveform is Sine wave. You cannot use Pulse or DC as the carrier waveform.
Soft Front Panel Operation
For Arbitrary waveforms, select Arb at the top panel as shown in Figure 2-2 and select the desired waveform from the drop down list as shown in
Figure 2- 16.
56 U2761A User’s Guide
Page 75
Remote Interface Operation
You can also use the APPLy command to select the function, frequency, amplitude, and offset.

Carrier Frequency

The maximum carrier frequency depends on the selected function as shown below. The default is 1 kHz for all
functions.
Ta b l e 2 - 4 Carrier frequency for FM
Function Minimum frequency Maximum frequency
Sine 1 μHz 20 MHz
Square 1 μHz 20 MHz
Features and Functions 2
FUNCtion {SINusoid|SQUare|RAMP|USER}
Ramp, Triangle 1 μHz 200 kHz
Arbitrary 1 μHz 200 kHz
The carrier frequency must always be greater than or equal to the frequency deviation plus 1 μHz. If you attempt to set the deviation to a value plus 1 μHz greater than the carrier frequency (with FM enabled), the U2761A will automatically adjusts the deviation to the maximum value allowed with the present carrier frequency.
The sum of the carrier frequency and deviation must be less than or equal to the maximum frequency for the selected function (20 MHz for Sine and Square, and 200 kHz for Ramp and Arbitrary waveforms). If you attempt to set the deviation to a value that is not valid, the U2761A will automatically adjusts it to the maximum value allowed with the present carrier frequency.
U2761A User’s Guide 57
Page 76
2Features and Functions

Modulating Waveform

Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the desired frequency value and select the unit from the drop down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.
The modulating waveform consists of Sine, Square, Ramp, Negative Ramp (Nramp), Triangle, Noise, or Arbitrary waveform. The default modulating waveform is Sine wave.
Square has 50% duty cycle
Ramp has 100% symmetry
Triangle has 50% symmetry
Negative Ramp has 0% symmetry
Soft Front Panel Operation
Select the desired modulating waveform from the drop down list as shown in Figure 2-18.
Remote Interface Operation
FM:INTernal:FUNCtion
{SINusoid|SQUare|RAMP|NRAMp|TRIangle|USER}
58 U2761A User’s Guide
Page 77

Modulating Waveform Frequency

The range of the modulating waveform frequency is 2 mHz to 20 kHz. The default modulating waveform frequency is 100 Hz.
Soft Front Panel Operation
On the Mod Frequency panel as shown in Figure 2- 18, input the desired modulating frequency value and select the unit from the drop down list.
Remote Interface Operation
FM:INTernal:FREQuency {<frequency>}

Frequency Deviation

The frequency deviation setting represents the peak variation in frequency of the modulated waveform from the carrier frequency.
Features and Functions 2
The range of the frequency deviation is 1 Hz to 500 kHz (limited to 100 kHz minus 1 μHz for Ramp and Arbitrary waveforms). The default frequency deviation is 100 Hz.
The carrier frequency must always be greater than or equal to the frequency deviation plus 1 μHz. If you attempt to set the deviation to a value plus 1 μHz greater than the carrier frequency (with FM enabled), the U2761A will automatically adjusts the deviation to the maximum value allowed with the present carrier frequency.
The sum of the carrier frequency and deviation must be less than or equal to the maximum frequency for the selected function (20 MHz for Sine and Square, and 200 kHz for Ramp and Arbitrary waveforms). If you attempt to set the deviation to a value that is not valid, the U2761A will automatically adjusts it to the maximum value allowed with the present carrier frequency.
U2761A User’s Guide 59
Page 78
2Features and Functions

Generate FM Using SCPI Commands

Soft Front Panel Operation
On the Deviation panel as shown in Figure 2- 18, input the desired frequency deviation value and select the unit from the drop down list.
Remote Interface Operation
FM:DEViation {<peak deviation in Hz>}
Example 1
-> *CLS; *RST //To reset the U2761A to default
power-on state, this command can be ignored if this operation is not required.
-> FM:STAT ON //Enables FM.
-> FUNC SIN //Sets the carrier waveform to
Sine wave.
-> FREQ 1000 // Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP //Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0 //Sets the output offset to 0.
-> FM:INT:FUNC SIN //Changes the modulating
waveform shape to Sine wave.
-> FM:INT:FREQ 500 //Changes the modulating
frequency to 500 Hz.
-> FM:DEV 100 //Changes the frequency
deviation to 100 Hz.
-> OUTP ON //Turns on output.
60 U2761A User’s Guide
Page 79

Phase Modulation (PM)

Modulating frequency
Phase deviation
A modulated waveform consists of a carrier waveform and a modulating waveform . The following figure shows an example of a PM waveform. PM is very similar to FM, but in PM, the phase of the modulated waveform is varied by the instantaneous voltage of the modulating waveform. The variation in phase of the modulated waveform from the carrier waveform is called the phase deviation.
Features and Functions 2
Figure 2-19 PM waveform

To Selec t P M

The U2761A allows only one modulation mode to be enabled at a time. The U2761A does not allow PM to be enabled at the same time that sweep is enabled.
U2761A User’s Guide 61
Page 80
2Features and Functions
Soft Front Panel Operation
Click Mod and then select PM as shown in
Figure 2- 20. To output the PM waveform, configure the
settings for the carrier frequency, output amplitude, offset voltage, modulating frequency, phase deviation, and the desired waveform.
Figure 2-20 Panel view of PM
Remote Interface Operation
PM:STATe {OFF|ON}

Carrier Waveform

The PM carrier waveform consists of Sine, Square, Ramp, Triangle, or Arbitrary waveform. The default waveform is Sine wave. You cannot use Pulse or DC as the carrier waveform.
Soft Front Panel Operation
For Arbitrary waveforms, click Arb at the top panel as shown in Figure 2-2 and select the desired waveform from the drop down list as shown in Figure 2- 16.
62 U2761A User’s Guide
Page 81
Remote Interface Operation
You can also use the APPLy command to select the function, frequency, amplitude, and offset.

Carrier Frequency

The maximum carrier frequency depends on the function selected as shown below. The default carrier frequency is 1 kHz for all functions.
Ta b l e 2 - 5 Carrier frequency for PM
Function Minimum frequency Maximum frequency
Sine 1 μHz 20 MHz
Square 1 μHz 20 MHz
Features and Functions 2
FUNCtion {SINusoid|SQUare|RAMP|USER}
Ramp, Triangle 1 μHz 200 kHz
Arbitrary 1 μHz 200 kHz
Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the desired frequency value and select the unit from the drop down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.
U2761A User’s Guide 63
Page 82
2Features and Functions

Modulating Waveform

The modulating waveform consists of Sine, Square, Ramp, Negative Ramp (Nramp), Triangle, or Arbitrary waveform. The default waveform is Sine wave.
Square has 50% duty cycle
Ramp has 100% symmetry
Triangle has 50% symmetry
Negative Ramp has 0% symmetry
Soft Front Panel Operation
Select the desired modulating waveform from the drop down list as shown in Figure 2- 20.
Remote Interface Operation
PM:INTernal:FUNCtion
{SINusoid|SQUare|RAMP|NRAMp|TRIangle|USER}

Modulating Waveform Frequency

The range of the modulating waveform frequency is 2 mHz to 20 kHz. The default modulating waveform frequency is 10 Hz.
Soft Front Panel Operation
On the Mod Frequency panel as shown in Figure 2- 20, input the desired modulating frequency value and select the unit from the drop down list.
Remote Interface Operation
PM:INTernal:FREQuency {<frequency>}
64 U2761A User’s Guide
Page 83

Phase Deviation

The phase deviation setting represents the peak variation in phase of the modulated waveform from the carrier waveform. The phase deviation setting can be set from 0 to 360 °. The default phase deviation setting is 180
Soft Front Panel Operation
On the Phase Deviation panel as shown in Figure 2- 20, input the desired phase deviation value.
Remote Interface Operation
PM:DEViation {<deviation in degrees>}

Generate PM Using SCPI Commands

Example 1
Features and Functions 2
°
.
-> *CLS; *RST //To reset the U2761A to default
power-on state, this command can be ignored if this operation is not required.
-> PM:STAT ON //Enables PM.
-> FUNC SIN //Sets the carrier waveform to
Sine wave.
-> FREQ 1000 // Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP //Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0 //Sets the output offset to 0.
-> PM:INT:FUNC SIN //Changes the modulating
waveform shape to Sine wave.
-> PM:INT:FREQ 500 //Changes the modulating
frequency to 500 Hz.
-> PM:DEV 180 //Changes the phase deviation to
180 °.
-> OUTP ON //Turns on output.
U2761A User’s Guide 65
Page 84
2Features and Functions
FSK rate
Hop frequency

Frequency-Shift Keying (FSK) Modulation

FSK is similar to FM except that the frequency alternates between two preset values. You can configure the U2761A to “shift” its output frequency between two preset values using FSK modulation. An example of the FSK modulation waveform is shown below. The rate at which the output shifts between the two frequencies (called the “carrier frequency” and the “hop frequency”) is determined by the internal rate generator.
Figure 2-21 FSK modulation waveform

To Select FSK Modulation

The U2761A allows only one modulation mode to be enabled at a time. The U2761A does not allow FSK to be enabled at the same time that sweep is enabled.
66 U2761A User’s Guide
Page 85
Features and Functions 2
Soft Front Panel Operation
Click Mod and select FSK as shown in
Figure 2- 22. To output the FSK waveform, configure the
settings for the carrier frequency, output amplitude, offset voltage, “hop frequency”, and FSK rate.
Figure 2-22 Panel view of FSK
Remote Interface Operation
FSKey:STATe {OFF|ON}

Carrier Waveform

The FSK carrier waveform consists of Sine, Square, Ramp, Triangle, or Arbitrary waveform. The default carrier waveform is Sine wave. You cannot use Pulse or DC as the carrier waveform.
Soft Front Panel Operation
For Arbitrary waveforms, click Arb at the top panel as shown in Figure 2-2 and select the desired waveform from the drop down list as shown in
Figure 2- 16.
U2761A User’s Guide 67
Page 86
2Features and Functions

FSK Carrier Frequency

Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|USER}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.
The maximum carrier frequency depends on the function selected as shown below. The default carrier frequency is 1 kHz for all functions.
Ta b l e 2 - 6 Carrier frequency for FSK
Function Minimum frequency Maximum frequency
Sine 1 μHz 20 MHz
Square 1 μHz 20 MHz
Ramp, Triangle 1 μHz 200 kHz
Arbitrary 1 μHz 200 kHz
Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the desired frequency value and select the unit from the drop down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.
68 U2761A User’s Guide
Page 87

FSK “Hop” Frequency

The maximum alternate (or “hop”) frequency depends on the function selected as shown below. The default “hop” frequency is 100 Hz for all functions.
Ta b l e 2 - 7 “Hop” frequency
Function Minimum frequency Maximum frequency
Sine 1 μHz 20 MHz
Square 1 μHz 20 MHz
Ramp, Triangle 1 μHz 200 kHz
Arbitrary 1 μHz 200 kHz
Only Square wave with a 50% duty cycle is available for the Internal FSK modulating waveform.
Soft Front Panel Operation
Features and Functions 2
On the Hop Frequency panel as shown in Figure 2- 22, input the desired hop frequency value and select the unit from the drop down list.
Remote Interface Operation
FSKey:FREQuency {<frequency>}
U2761A User’s Guide 69
Page 88
2Features and Functions

FSK Rate

Generate FSK Modulation Using SCPI Commands

The FSK rate is the rate at which the output frequency “shifts” between the carrier frequency and the hop frequency.
The range of the FSK rate is 2 mHz to 100 kHz. The default FSK rate value is 10 Hz.
Soft Front Panel Operation
On the FSK Rate panel as shown in Figure 2- 22, input the desired FSK rate value and select the unit from the drop down list.
Remote Interface Operation
FSKey:INTernal:RATE {<rate in Hz>}
Example 1
-> *CLS; *RST //To reset the U2761A to default
power-on state, this command can be ignored if this operation is not required.
-> FSK:STAT ON //Enables FSK modulation.
-> FUNC SIN //Sets the carrier waveform to
Sine wave.
-> FREQ 1000 // Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP //Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0 //Sets the output offset to 0.
-> FSK:FREQ 100 //Sets the "hop" frequency to
100 Hz.
-> FSK:INT:RATE 10 //Sets the FSK rate to 10 Hz.
-> OUTP ON //Turns on output.
70 U2761A User’s Guide
Page 89

Phase-Shift Keying (PSK) Modulation

PSK rate
PSK deviation
Phase-shift keying (PSK) is a form of digital modulation in which the phase of the carrier signal is discretely varied. You can configure the U2761A to “shift” its output phase between two preset phases using PSK. The rate at which the output shifts between the two phases is determined by the internal rate generator on the signal level. The following figure shows an example of the PSK modulation waveform.
Features and Functions 2
Figure 2-23 PSK modulation waveform

To Select PSK Modulation

The U2761A allows only one modulation mode to be enabled at a time. The U2761A does not allow PSK to be enabled at the same time that sweep is enabled.
U2761A User’s Guide 71
Page 90
2Features and Functions
Soft Front Panel Operation
Click Mod and then select PSK as shown in
Figure 2- 24. To output the PSK waveform, configure the
settings for the carrier frequency, output amplitude, offset voltage, deviation, and PSK rate.
Figure 2-24 Panel view of PSK
Remote Interface Operation
PSKey:STATe {OFF|ON}

Carrier Waveform

The PSK carrier waveform consists of Sine, Square, Ramp, Triangle, or Arbitrary waveform. The default waveform is Sine wave. You cannot use Pulse or DC as the carrier waveform.
Soft Front Panel Operation
For Arbitrary waveforms, select Arb at the top panel as shown in Figure 2-2 and select the desired waveform from the drop down list as shown in
Figure 2- 16.
72 U2761A User’s Guide
Page 91
Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|USER}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.

PSK Carrier Frequency

The maximum carrier frequency depends on the function selected as shown below. The default is 1 kHz for all functions.
Ta b l e 2 - 8 Carrier frequency for PSK
Function Minimum frequency Maximum frequency
Sine 1 μHz 20 MHz
Square 1 μHz 20 MHz
Ramp, Triangle 1 μHz 200 kHz
Features and Functions 2
Arbitrary 1 μHz 200 kHz
Only Square wave with a 50% duty cycle is available for the Internal PSK modulating waveform.
Soft Front Panel Operation
On the Frequency panel shown in Figure 2- 3, input the desired frequency value and select the unit from the drop down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.
U2761A User’s Guide 73
Page 92
2Features and Functions

PSK Rate

PSK Deviation

The PSK rate is the rate at which the output phase “shifts” between two preset phases. The range of the PSK rate is 2 mHz to 100 kHz. The default PSK rate value is 10 Hz.
Soft Front Panel Operation
On the PSK Rate panel as shown in Figure 2- 24, input the desired PSK rate value and select the unit from the drop down list.
Remote Interface Operation
PSKey:INTernal:RATE {<rate in Hz>}
The deviation setting represents the phase variation of the shifted waveform from the carrier waveform. The deviation can be set from 0 to 360 °. The default deviation setting is
180
°
.
Soft Front Panel Operation
On the Deviation panel as shown in Figure 2- 24, input the desired deviation value.
Remote Interface Operation
PSKey:DEViation {<deviation in degrees>}
74 U2761A User’s Guide
Page 93
Features and Functions 2

Generate PSK Modulation Using SCPI Commands

Example 1
-> *CLS; *RST //To reset the U2761A to
default power-on state, this command can be ignored if this operation is not required.
-> PSK:STAT ON //Enables PSK modulation.
-> FUNC SIN //Sets the carrier waveform to
Sine wave.
-> FREQ 1000 //Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP //Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0 //Sets the output offset to 0.
-> PSK:INT:RATE 10 //Sets the PSK rate to
10 Hz.
-> PSK:DEV 180 //Sets the PSK deviation to
180 °.
-> OUTP ON //Turns on output.
U2761A User’s Guide 75
Page 94
2Features and Functions
ASK rate

Amplitude-Shift Keying (ASK) Modulation

ASK is a form of digital modulation in which the modulating signal apply variations in the amplitude of a carrier signal. The amplitude of the carrier signal varies simultaneously with the modulating signal while phase and frequency remain constant. In other words, the carrier signal can be assumed as an on and off switch. You can configure the U2761A to “shift” its output amplitude between two preset amplitudes using ASK. The rate at which the output shifts between the two amplitudes is determined by the internal rate generator. An example of the ASK modulation waveform is shown in the following.
Figure 2-25 ASK modulation waveform

To Select ASK Modulation

The U2761A allows only one modulation mode to be enabled at a time. The U2761A does not allow ASK to be enabled at the same time that sweep is enabled.
76 U2761A User’s Guide
Page 95
Features and Functions 2
Soft Front Panel Operation
Click Mod and then select ASK as shown in
Figure 2- 26. To output the ASK waveform, configure the
settings for the carrier frequency, output amplitude, offset voltage, and ASK rate.
Figure 2-26 Panel view of ASK
Remote Interface Operation
ASKey:STATe {OFF|ON}

Carrier Waveform

The ASK carrier waveform consists of Sine, Square, Ramp, Triangle, or Arbitrary waveform. The default waveform is Sine wave. You cannot use Pulse or DC as the carrier waveform.
Soft Front Panel Operation
For Arbitrary waveforms, select Arb at the top panel as shown in Figure 2-2 and select the desired waveform from the drop down list as shown in
Figure 2- 16.
U2761A User’s Guide 77
Page 96
2Features and Functions

Carrier Frequency

Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|USER}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.
The maximum carrier frequency depends on the function selected as shown below. The default is 1 kHz for all functions.
Ta b l e 2 - 9 Carrier frequency for ASK
Function Minimum frequency Maximum frequency
Sine 1 μHz 20 MHz
Square 1 μHz 20 MHz
Ramp, Triangle 1 μHz 200 kHz
Arbitrary 1 μHz 200 kHz
Only Square wave with a 50% duty cycle is available for the Internal ASK modulating waveform.
Soft Front Panel Operation
On the Frequency panel as shown in Figure 2- 3, input the desired frequency value and select the unit from the drop down list.
Remote Interface Operation
FREQuency {<frequency>}
You can also use the APPLy command to select the function, frequency, amplitude, and offset.
78 U2761A User’s Guide
Page 97
Features and Functions 2

ASK Rate

The ASK rate is the rate at which the output amplitude “shifts” between two preset amplitudes. The range of the ASK rate is 2 mHz to 100 kHz. The default ASK rate value is 10 Hz.
Soft Front Panel Operation
On the ASK Rate panel as shown in Figure 2- 26, input the desired ASK rate value and select the unit from the drop down list.
Remote Interface Operation
ASKey:INTernal:RATE {<rate in Hz>}

Generate ASK Modulation Using SCPI Commands

Example 1
-> *CLS; *RST //To reset the U2761A to
default power-on state, this command can be ignored if this operation is not required.
-> ASK:STAT ON //Enables ASK modulation.
-> FUNC SIN //Sets the carrier waveform to
Sine wave.
-> FREQ 1000 //Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP // Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0 //Sets the output offset to 0.
-> ASK:INT:RATE 10 //Sets the ASK rate to
10 Hz.
-> OUTP ON //Turns on output.
U2761A User’s Guide 79
Page 98
2Features and Functions
Frequency
Linear type
Start frequency
Stop frequency

Frequency Sweep

In the frequency sweep mode, the U2761A “steps” from the start frequency to the stop frequency at a sweep rate which you specify. You can sweep up or down in frequency, and with either linear or logarithmic spacing. You can also configure the U2761A to output a single sweep (one pass from start frequency to stop frequency) by applying an External or Manual (software) trigger. The U2761A can produce a frequency sweep for Sine, Square, Ramp, Triangle, or Arbitrary’s waveform (Pulse and DC are not allowed). The figure below shows an example of the frequency sweep.
Figure 2-27 Frequency sweep

To Select Sweep

The U2761A does not allow the sweep mode to be enabled at the same time that any modulation mode is enabled.
Soft Front Panel Operation
Click Sweep as shown in Figure 2- 28. To output sweep, configure the settings for frequencies, output amplitude, offset, sweep type and time, and trigger setup.
80 U2761A User’s Guide
Page 99
Figure 2-28 Panel view of sweep
Remote Interface Operation
SWEep:STATe {OFF|ON}

Start Frequency and Stop Frequency

Features and Functions 2
The start frequency and stop frequency set the upper and lower frequency bounds for the sweep. The U2761A begins at the start frequency, sweeps to the stop frequency, and then resets back to the start frequency.
The range of the start and stop frequencies is 1 μHz to 20 MHz (limited to 200 kHz for Ramp and Arbitrary waveforms). The sweep is phase-continuous over the full frequency range. The default start frequency is 100 Hz. The
default stop frequency is 1 kHz.
To sweep up in frequency, set the start frequency < stop frequency. To sweep down in frequency, set the start frequency > stop frequency.
Soft Front Panel Operation
On the Start Frequency and Stop Frequency panels as shown in Figure 2- 28, input the desired start frequency and stop frequency values and select the respective units from the drop down lists.
U2761A User’s Guide 81
Page 100
2Features and Functions

Sweep Mode

Remote Interface Operation
FREQuency:STARt {<frequency>}
FREQuency:STOP {<frequency>}
You can sweep with either linear or logarithmic spacing. For a linear sweep, the U2761A varies the output frequency in a linear fashion during the sweep. For a logarithmic sweep, the U2761A varies the output frequency in a logarithmic fashion.
The sweep mode consists of Linear or Logarithmic. The default sweep mode is Linear.
Soft Front Panel Operation
On the Sweep Type panel as shown in Figure 2- 28, select the Linear or Log mode.
Remote Interface Operation
SWEep:SPACing {LINear|LOGarithmic}

Sweep Time

The sweep time specifies the number of seconds required to sweep from the start frequency to the stop frequency. The number of discrete frequency points in the sweep is automatically calculated by the U2761A and is based on the sweep time you have selected. The range of the sweep time is 1 ms to 500 s. The default sweep time is 1 s.
82 U2761A User’s Guide
Loading...