Agilent Technologies U1251A User Manual

Agilent U1251A and U1252A Handheld Digital Multimeter
User’s and Service Guide
Agilent Technologies
Notices
CAUTION
WARNING
No p art o f this manu al may be re produce d in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
U1251-90003
Edition
Tenth Edition, May 4, 2012
Printed in Malaysia
Agilent Technologies, Inc. 3501 Stevens Creek Blvd. Santa Clara, CA 95052 USA
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus­tomers. Agilent provides this customary commercial license in Software and techni­cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com­puter Software Documentation).
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Accessories Warranty
Agilent offers warranty for product’s acces­sories for up to 3 months from the end-user acceptance date.
Standard Calibration Service (optional)
Agilent offers an optional calibration service contract for a period of 3 years from end-user acceptance date.
Safety Notices
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
II Agilent U1251A/U1252A User’s and Service Guide

Safety Symbols

CAT IV
600 V
The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
CAT III 1000 V
Category III 1000 V overvoltage protection Category IV 600 V overvoltage protection
Equipment protected throughout by double insulation or reinforced insulation
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
Earth (ground) terminal
Caution, risk of electric shock
Agilent U1251A/U1252A User’s and Service Guide III

Safety Information

This meter is safety-certified in compliance with EN/IEC 61010-1:2001, UL 61010-1 Second Edition and CAN/CSA
22.2 61010-1 Second Edition, CAT III 1000 V/ CAT IV 600 V Overvoltage Protection, Pollution Degree II. Use with standard or compatible test probes.

General Safety Information

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
IV Agilent U1251A/U1252A User’s and Service Guide
WARNING
• When working above 70 VDC, 33 VAC RMS or 46.7 V peak, exercise caution – such range pose a shock hazard.
• Do not measure more than the rated voltage (as marked on the meter) between terminals, or between terminal and earth ground.
• Double-check the meter’s operation by measuring a known voltage.
• For current measurement, turn off circuit power before connecting the meter to the circuit. Always place the meter in series with the circuit.
• When connecting probes, always connect the common test probe first. When disconnecting probes, always disconnect the live test probe first.
• Detach test probes from the meter before you open the battery cover.
• Do not use the meter with the battery cover or part of the cover removed or loose.
• Replace the battery as soon as the low battery indicator flashes on screen. This is to avoid false readings, which may lead to possible electric shock or personal injury.
• Do not operate the product in an explosive atmosphere or in the presence of flammable gases or fumes.
• Inspect the case for cracks or missing plastic. Pay extra attention to the insulation surrounding the connectors. Do not use the meter if it is damaged.
• Inspect the test probes for damaged insulation or exposed metal, and check for continuity. Do not use the test probe if it is damaged.
• Do not use any other AC charger adaptor apart from the one certified by Agilent with this product.
• Do not use repaired fuses or short-circuited fuse-holders. For continued protection against fire, replace the line fuses only with fuses of the same voltage and current rating and recommended type.
• Do not service or perform adjustments alone. Under certain condition, hazardous voltages may exist, even with the equipment switched off. To avoid dangerous electric shock, service personnel must not attempt internal service or adjustment unless another person, capable of rendering resuscitation or first aid, is present.
• Do not substitute parts or modify equipment to avoid the danger of introducing additional hazards. Return the product to Agilent Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
• Do not operate damaged equipment as the safety protection features built into this product may have been impaired, either through physical damage, excessive moisture, or any other reason. Remove power and do not use the product until safe operation can be verified by service-trained personnel. If necessary, return the product to Agilent Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
Agilent U1251A/U1252A User’s and Service Guide V
CAUTION
Turn off the circuit power and discharge all high-voltage capacitors in the circuit before
you perform resistance, continuity, diodes, or capacitance tests.
Use the correct terminals, function, and range for your measurements.
Never measure voltage when current measurement is selected.
Use only recommended rechargeable battery. Ensure proper insertion of battery in the
meter, and follow the correct polarity.
Disconnect test leads from all the terminals during battery charging.
VI Agilent U1251A/U1252A User’s and Service Guide

Regulatory Markings

The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
The C-tick mark is a registered trademark of the Spectrum Management Agency of Australia. This signifies compliance with the Australia EMC Framework regulations under the terms of the Radio Communication Act of 1992.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001. Cet appareil ISM est confomre a la norme NMB-001 du Canada.
The CSA mark is a registered trademark of the Canadian Standards Association.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.
Agilent U1251A/U1252A User’s and Service Guide VII

Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste
To return this unwanted instrument, contact your nearest Agilent Technologies, or visit: www.agilent.com/environment/product for more information.
Agilent Technologies, through Rechargeable Battery Recycling Corporation (RBRC), offers free and convenient battery recycling options in the U.S. and Canada. Contact RBRC at 877-2-RECYCLE (877.273.2925) or online at:
http://www.call2recycle.org/ for the nearest recycling location.
VIII Agilent U1251A/U1252A User’s and Service Guide

In This Guide…

1Getting Started
This chapter contains information on the Agilent U1251A and U1252A handheld multimeter front panel, rotary switch, keypad, display, terminals and rear panel.
2 Making Measurements
This chapter contains information on how to make measurements using the U1251A and U1252A handheld digital multimeter.
3 Functions and Features
This chapter contains information on the functions and features that are available for the U1251A and U1252A digital multimeter.
4 Changing The Default Setting
This chapter shows you how to change the default factory settings of the U1251A and U1252A and other available setting options
5 Maintenance
This chapter will go through how to troubleshoot the handheld digital multimeter if any problems arise.
6 Performance Tests and Calibration
This chapter contains the performance test and the adjustment procedure.
7 Specifications
This chapter lists the product characteristics, specification assumptions and the specifications of the U1251A and U1252A digital multimeters.
Agilent U1251A/U1252A User’s and Service Guide IX

Declaration of Conformity (DoC)

NOTE
The Declaration of Conformity (DoC) for this instrument is available on the Web site. You can search the DoC by its product model or description.
http://regulations.corporate.agilent.com/DoC/search.htm
If you are unable to search for the respective DoC, please contact your local Agilent representative.
X Agilent U1251A/U1252A User’s and Service Guide

Contents

1 Getting Started
Introducing the U12521A/U1252A Handheld Digital Multimeter 2
Check the shipment 3
Adjusting the tilt-stand 4
The rear panel at a glance 7 The keypad at a glance 9 The display at a glance 11 Selecting display with the Hz button 16 Selecting display with the Dual button 18 Selection display with the Shift button 20 The terminals at a glance 22
2 Making Measurements
Understanding The Measurement Instructions 24
III
Measuring Voltage 24
Measuring AC voltage 25 Measuring DC voltage 26
Measuring Current 27
µA & mA measurement 27 Percentage scale of 4 mA to 20 mA 29 A (ampere) measurement 31
Frequency Counter 32
Measuring Resistance, Conductance and Testing Continuity 34
Tes ti ng Dio de s 38
Measuring Capacitance 41
Measuring Temperature 43
Alerts and Warning During Measurement 47
Abbreviated Book Title Variable (Edit this) XIII
III
Overload alert 47 Input warning 47 Charge terminal alert 48
3 Functions and Features
Dynamic Recording 50
Data Hold (Trigger Hold) 52
Refresh Hold 53
Null (Relative) 55
Decibel Display 57
1 ms Peak Hold 59
Data Logging 61
Manual logging 61 Interval logging 63 Reviewing logged data 65
Square Wave Output (for U1252A) 67
Remote Communication 71
4 Changing The Default Setting
Selecting Setup Mode 74
Setting Data Hold/Refresh Hold Mode 77
Setting Data Logging Mode 78
Setting Thermocouple Types (U1252A only) 79
Setting Reference Impedance for dBm Measurement 80 Setting Minimum Frequency Measurement 81
Setting Temperature Unit 82
Setting Auto Power Saving Mode 84
Setting Percentage (%) Scale Readout 86
XIV Abbreviated Book Title Variable (Edit this)
Setting Beep Frequency 87 Setting Backlight Timer 88 Setting Baud Rate 89
Setting Parity Check 90
Setting Data Bit 91
Setting Echo Mode 92
Setting Print Mode 93
Returning to Default Factory Settings 94
5 Maintenance
Introduction 96
General maintenance 96 Battery replacement 96 Storage considerations 98 Charging the battery 98 Fuse checking procedure 105 Replacing the fuse 106 Troubleshooting 108
III
Replaceable Parts 109
To order replaceable parts 109
6 Performance Tests and Calibration
Calibration Overview 112
Closed-case electronic calibration 112 Agilent Technologies calibration services 112 Calibration interval 113 Adjustment is recommended 113
Recommended Test Equipment 114
Basic Operating Test 115
Backlit test 115
Abbreviated Book Title Variable (Edit this) XV
III
Testing the display 115 Current terminal test 116 Charge terminal alert test 117
Test Considerations 118
Calibration Security 119
Performance Verification Tests 120
Unsecuring the instrument for calibration 128
Calibration process 131
Using the front panel for adjustments 132
Adjustments Consideration 133
Valid adjustment input values 134 Adjustment procedure 135 Finishing the adjustment 142 To read the calibration count 142 Calibration errors 143
7 Specifications
Product Characteristics 146
Measurement Category 149
Measurement category definition 149
Specification Assumptions 150
Electrical Specifications 151
DC Specifications 151 AC Specifications 153 AC+DC Specifications for U1252A 155 Capacitance Specifications 156 Temperature Specifications 156 Frequency Specifications 157 Duty Cycle and Pulse Width Specifications 157 Frequency Sensitivity Specifications 158
XVI Abbreviated Book Title Variable (Edit this)
Peak Hold Specifications 159 Frequency Counter Specifications for U1252A 160 Square Wave Output for U1252A 161
Operating Specifications 162
Display update rate (approximate) 162 Input impedance 163
III
Abbreviated Book Title Variable (Edit this) XVII
III
XVIII Abbreviated Book Title Variable (Edit this)

List of Figures

Figure 1-1 Tilt-stand at 60° 4 Figure 1-2 Tilt-stand at 30° 4 Figure 1-3 Tilt-stand at hanging position 5 Figure 1-4 U1252A front panel 6 Figure 1-5 Rear panel 7 Figure 1-6 Rotary switch 8 Figure 1-7 U1252A keypad 9 Figure 1-8 Display symbols 11 Figure 1-9 U1252A connector terminals 22 Figure 2-1 Measuring AC voltage 25 Figure 2-2 Measuring DC voltage 26 Figure 2-3 Measuring mA and mA current 28 Figure 2-4 Measuring scale of 4-20 mA 30 Figure 2-5 A (ampere) current measurement 31 Figure 2-6 Measuring frequency 33 Figure 2-7 Measuring resistance 34 Figure 2-8 Audible continuity, conductance, and resistance
test. 35
Figure 2-9 Conductance measurement 37 Figure 2-10 Measuring forward bias of diode 39 Figure 2-11 Measuring reverse bias of diode 40 Figure 2-12 Capacitance measurements 42 Figure 2-13 Connecting the thermal probe into the non-compensa-
tion transfer adapter 44
Figure 2-14 Connecting the probe with adapter into the
multimeter 44
Figure 2-15 Surface temperature measurement 46 Figure 2-16 Input terminal warning 47 Figure 2-17 Charge terminal alert 48 Figure 3-1 Dynamic recording mode operation 51 Figure 3-2 Data hold mode operation 52 Figure 3-3 Refresh hold mode operation 54 Figure 3-4 Null (relative) mode operation 56 Figure 3-5 dBm/dBV display mode operation 58 Figure 3-6 1 ms peak hold mode operation 60 Figure 3-7
Hand (Manual) logging mode operation 62
Abbreviated Book Title Variable (Edit this) IX
Figure 3-8 Full Log 62 Figure 3-9 Interval (Automatic) logging mode operation 64 Figure 3-10 Log review mode operation 66 Figure 3-11 Frequency adjustment for square wave output 68 Figure 3-12 Pulse width adjustment for square wave output 69 Figure 3-13 Duty cycle adjustment for Square Wave 70 Figure 3-14 Cable connection for remote communication 71 Figure 4-1 Data hold/Refersh hold setup 77 Figure 4-2 Data logging setup 78 Figure 4-3 Thermocouple type setup 79 Figure 4-4 Reference impedance for dBm measurement setup 80 Figure 4-5 Minimum frequency setup 81 Figure 4-6 Temperature unit setup 83 Figure 4-7 Auto power saving setup 85 Figure 4-8 % scale readout setup 86 Figure 4-9 Beep frequency setup 87 Figure 4-10 Backlit timer setup 88 Figure 4-11 Baud rate setup remote control 89 Figure 4-12 Parity check setup 90 Figure 4-13 Data bit setup for remote control 91 Figure 4-14 Echo mode setup for remote control 92 Figure 4-15 Print mode setup for remote control 93 Figure 4-16 Reset setup 94 Figure 5-1 9 V rectangular battery 97 Figure 5-2 Battery capacity display as trickle 100 Figure 5-3 Self-test 100 Figure 5-4 Charging mode 102 Figure 5-5 Charge end and trickle state 103 Figure 5-6 Battery charging procedure 104 Figure 5-7 Fuse replacement 107 Figure 6-1 LCD display 115 Figure 6-2 Input warning 116 Figure 6-3 Charge terminal alert 117
X Abbreviated Book Title Variable (Edit this)

List of Tables

Ta b l e 1 - 1 Rotary switch description and functions 8 Ta b l e 1 - 2 Keypad desciption and functions 9 Ta b l e 1 - 3 General display symbols 12 Ta b l e 1 - 4 Priamry display symbols 13 Ta b l e 1 - 5 Secondary display symbols 14 Ta b l e 1 - 6 Analog bar range and counts 15 Ta b l e 1 - 7 Selecting display with the Hz button 16 Ta b l e 1 - 8 Selecting display with the Dual button 18 Ta b l e 1 - 9 Selecting display with the Shift button 20 Ta b l e 1 - 1 0 Terminal connections for different measuring
functions 22
Ta b l e 2 - 1 Numerical steps descriptions 24 Ta b l e 2 - 2 Percentage scale and measurement range 29 Ta b l e 2 - 3 Audible continuity measurement range 36 Ta b l e 4 - 1 Available setting options in Setup mode 75 Ta b l e 5 - 1 Battery voltage and corresponding percentage of charg-
es in standby and charging modes 100
Ta b l e 5 - 2 Error messages 101 Ta b l e 5 - 3 Measurement readings for fuse checking 106 Ta b l e 5 - 4 Fuse specifications 107 Ta b l e 5 - 5 Basic troubleshooting procedures 108 Ta b l e 6 - 1 Recommended Test Equipment 114 Ta b l e 6 - 2 Performance verification test steps 121 Ta b l e 6 - 3 Valid adjustment input values 134 Ta b l e 6 - 4 Adjustment table 137 Ta b l e 7 - 1 DC Accuracy ± (% of reading + Number of Least Signifi-
cant Digit) 151
Ta b l e 7 - 2 U1251A accuracy specifications ± (% of reading + num-
ber of LSD) for true RMS AC voltage 153
Ta b l e 7 - 3 U1251A accuracy specifications ± (% of reading + num-
ber of LSD) for true RMS AC current 153
Ta b l e 7 - 4 U1252A accuracy specifications ± (% of reading + num-
ber of LSD) for true RMS AC voltage 154
Ta b l e 7 - 5 U1252A accuracy specifications ± (% of reading + num-
ber of LSD) for true RMS AC current 154
Ta b l e 7 - 6 U1252A true RMS ac+dc voltage specifications 155
II
Abbreviated Book Title Variable (Edit this) XI
II
Ta b l e 7 - 7 U1252A true RMS ac+dc current specifications 155 Ta b l e 7 - 8 Capacitance specifications 156 Ta b l e 7 - 9 Temperature specifications 156 Ta b l e 7 - 1 0 Frequency specifications 157 Ta b l e 7 - 1 1 Duty cycle and pulse width specifications 157 Ta b l e 7 - 1 2 Frequency sensitivity and trigger level specifications for
voltage measurements 158
Ta b l e 7 - 1 3 Frequency sensitivity specifications for current
measurements 159
Ta b l e 7 - 1 4 Peak hold specifications for dc voltage and current
measurements 159
Ta b l e 7 - 1 5 Frequency counter (divide by 1) specifications 160 Ta b l e 7 - 1 6 Frequency counter (divide by 100
specifications 160
Ta b l e 7 - 1 7 Square wave output specifications 161 Ta b l e 7 - 1 8 Display update rate (approximate) 162 Ta b l e 7 - 1 9 Input Impedance 163
[4]
)
XII Abbreviated Book Title Variable (Edit this)
Agilent U1251A and U1252A Handheld Digital Multimeter User’s and Service Guide
1 Getting Started
Introducing the U12521A/U1252A Handheld Digital Multimeter 2 Check the shipment 3 Adjusting the tilt-stand 4 The front panel at a glance 6 The rear panel at a glance 7 The rotary switch at a glance 8 The keypad at a glance 9 The display at a glance 11
Selecting display with the Hz button 16 Selecting display with the Dual button 18 Selection display with the Shift button 20
The terminals at a glance 22
This chapter contains information on the Agilent U1251A and U1252A handheld multimeter front panel, rotary switch, keypad, display, terminals and rear panel.
Agilent Technologies
1
1 Getting Started Tutorial

Introducing the U12521A/U1252A Handheld Digital Multimeter

Key features of this digital multimeter:
DC, AC, and AC + DC (U1252A only) voltage and current
measurements.
True- RMS measurement for both AC voltage and current
Rechargeable Ni- MH battery with built- in charging
capability (U1252A only)
Ambient temperature on second display
Battery capacity indicator
Blue LED backlight
Resistance measurement of up to 50 M (for U1251A)
and 500 M (for U1252A)
Conductance measurement from 0.01 nS (100 GΩ)
Capacitance measurement of up to 100 mF
Frequency counter up of to 20 MHz (U1252A only)
The % scale readout for 4–20 mA or 0–20 mA measurement
dBm with selectable reference impedance
1 ms Peak Hold to catch inrush voltage and current easily
Temperature test with selectable 0
(without ambient temperature compensation).
K- type (for U1251A) and J/K-types temperature
measurement (for U1252A)
Frequency, duty cycle, and pulse width measurements
Dynamic Recording for min, max, and average readings
Data Hold with manual or auto trigger and Null mode
Diode and audible continuity tests
Square wave generator with selectable frequency, pulse
width and duty cycle (U1252A only)
Agilent GUI Application Software (IR- USB cable sold
separately)
Closed case calibration
°C compensation
~ 50 nS
2 Agilent U1251A/U1252A User’s and Service Guide

Check the shipment

Getting Started Tutorial 1
Verify that you have received the following items with your multimeter:
9 V alkaline battery (for U1251A only)
Soft carrying case
4 mm probes
Test leads
Alligator clips
SMT grabbers
Fine tips test probes
Mini grabber (black only)
Rechargeable 7.2 V battery (for U1252A only)
Power cord & AC adapter (for U1252A only)
Quick Start Guide
CD containing the User’s Guide, application software and
instrument drivers
Certificate of Calibration
Contact your nearest Agilent Sales Office if any of the above are missing.
Inspect the shipping containing for damage. Signs of damage may include a dented or torn shipping container or cushioning material that indicates signs of unusual stress or compacting. Save the packaging material in case the multimeter needs to be returned.
Please refer to the Agilent Handheld Tools brochure (5989- 7340EN) for the full and latest list of handheld accessories available.
Agilent U1251A/U1252A User’s and Service Guide 3
1 Getting Started Tutorial

Adjusting the tilt-stand

To adjust the meter to a 60° standing position, pull the tilt- stand outwards to its maximum reach.
IR-USB cable
To P C (h ost )
Tilt-stand at 60 °
Figure 1-1 Tilt-stand at 60
To adjust the meter to a 30° standing position, bend the tip of the stand so that it is parallel to the ground before pulling the stand outwards to its maximum reach.
Figure 1-2 Tilt-stand at 30
4 Agilent U1251A/U1252A User’s and Service Guide
°
IR-USB cable
To P C (h ost )
Tilt-stand at 30 °
°
Getting Started Tutorial 1
3. Flip the tilt-stand over until
this side of the
stand is facing the device instead of
facing you.
2. Detach the tilt-stand
1. Extend the tilt-stand until its
maximum reach
4. Re-attach the tilt-stand to an
upright position
To adjust the meter to a hanging position follow the steps shown in Figure 1- 3 below.
Agilent U1251A/U1252A User’s and Service Guide 5
Figure 1-3 Tilt-stand at hanging position
1 Getting Started Tutorial

The front panel at a glance

Display
Keypad
Rotary switch
Te r m i n a l Connectors
Figure 1-4 U1252A front panel
6 Agilent U1251A/U1252A User’s and Service Guide

The rear panel at a glance

Getting Started Tutorial 1
IR communication port
Tes t pr obe holders
Tilt-stand
Battery access cover
Figure 1-5 Rear panel
Agilent U1251A/U1252A User’s and Service Guide 7
1 Getting Started Tutorial
1
2
3
4
5
6 7 8
9
10

The rotary switch at a glance

Figure 1-6 Rotary switch
Tab l e 1 - 1 Rotary switch description and functions
No. Description/Function
1Charge mode (U1252A only) or OFF
2 AC voltage
3 AC voltage, DC voltage or DC+AC voltage (U1252A only)
4 DC mV, AC mV, AC+DC mV (U1252A only)
5 Resistance (), Continuity, and Conductance ( nS )
6 Frequency counter (U1252A only) or Diode
7 Capacitance or Temperature
8 DC µA, AC µA, and AC+DC µA
8 Agilent U1251A/U1252A User’s and Service Guide
9 DC mA, DC current, AC mA, AC current, or AC+DC current
10 Square-wave output, Duty cycle, or Pulse width output (for
U1252A) and OFF (for U1251A)

The keypad at a glance

1
2
3
4
5
6
7
The operation of each key is shown below. Pressing a key illuminates a related symbol on the display and emits a beep from the multimeter. Turning the rotary switch to another position resets the current operation of the key.
Figure 1- 7 below shows the keypad of the U1252A. The ms%
(Pulse width/Duty cycle), Hz , and frequency counter functions are only available on the U1252A.
Getting Started Tutorial 1
Figure 1-7 U1252A keypad
Tab l e 1 - 2 Keypad desciption and functions
Button Function when pressed for less than 1 second Function when pressed for more than 1 second
1 acts as a toggle switch to turn backlit ON/OFF.
Backlit automatically turns off after 30s (default)
2 freezes the measured value. In Data Hold
mode, press again to trigger hold of the next measured value. In Refresh Hold mode, reading updates automatically once reading is stable and count setting is exceeded
3 saves displayed value as a reference to be
subtracted from subsequent measurements. Press again to see the relative value that has been saved.
[1]
.
displays the battery capacity for 3 seconds.
[1]
.
enters Dynamic Recording mode. Press again to scroll through Max, Min, Avg, and present readings (indicated by MAXMINAVG on display).
enters 1 ms Peak Hold mode. Press
to scroll through Max and Min peak readings.
Agilent U1251A/U1252A User’s and Service Guide 9
1 Getting Started Tutorial
Button Function when pressed for less than 1 second Function when pressed for more than 1 second
4 scrolls through the measuring function(s) at a
particular rotary switch position.
enters Log Review mode. Press to switch to manual or interval logging data. Press or to view first or last logged data respectively. Press or to scroll up or down logged data.
Press for more than 1 second to exit mode.
5 scrolls through available measuring ranges
(except when rotary switch is set at TEMP or at Hz [for U1252A] position)
[2]
.
6 scrolls through available dual-combination
displays (except when rotary switch is set at TEMP, Hz or [for U1252A] position, or when meter is in 1 ms peak hold or dynamic recording mode)
[3]
.
7 enters Frequency Test mode for current or
voltage measurements. Press to scroll through frequency (Hz), duty cycle (%) and pulse width (ms) functions. In duty cycle (%) and pulse width (ms) tests, press to switch to positive or negative
sets to Auto Range mode.
exits Hold, Null, Dynamic Recording, 1 ms
Peak Hold and dual display modes.
enters logging mode. In manual data logging, press to log data manually into memory. In automatic data logging, data logs automatically
[1]
. Press for more than 1 second to exit auto data logging mode.
pulse.
Notes for keypad descriptions and functions: 1 See Ta b l e 4 - 1 on page 75, “Available setting options in Setup mode,” for details of available options.
2 When rotary switch is at TEMP, press to switch to ºC or ºF display. When rotary switch is at Hz, press to
switch to division of signal frequency by 1 or 100.
3 When rotary switch is at TEMP, ETC is ON by default. You may press to disable ETC (Environment Temperature
Compensation), will appear on display. For pulse and duty cycle measurement, press to switch trigger slope to positive or negative. When meter is in peak or dynamic-recording mode, press to restart 1 ms peak hold or dynamic recording mode.
10 Agilent U1251A/U1252A User’s and Service Guide

The display at a glance

To view the full display (with all segments illuminated),
press and hold the button while turning the rotary switch from OFF to any non- OFF position. After you have finished viewing the full display, press any button to resume normal functionality based on the rotary switch position. This is followed by a wake- up feature.
The meter will then enter power save mode once auto power off (APF) is enabled. To wake the meter up:
1 Turn the rotary switch to OFF position then ON again.
2 Press any button for rotary switch position that is not at
square wave output position. (U1252A only)
3 To set the rotary switch at square wave out position,
press only the Dual, Range and Hold buttons or turn the rotary switch to another position. (U1252A only)
The LCD signs are explained in the following tables.
Getting Started Tutorial 1
Figure 1-8 Display symbols
Agilent U1251A/U1252A User’s and Service Guide 11
1 Getting Started Tutorial
Tab l e 1 - 3 General display symbols
LCD symbol Description
Remote control
Thermocouple types: (K-type) (J-type)
NULL Null math function
Diode / Audible continuity
Audible continuity for resistance
View mode for checking logged data
Data logging indication
Square wave output (U1252A only)
Positive slope for pulse width (ms) and duty cycle (%) measurement
Charging capacitor as capacitance measurement
Negative slope for pulse width (ms) and duty cycle (%) measurement
Discharging capacitor as capacitance measurement
Low battery indication
Auto power off enable
Refresh (auto) Hold
TRIG
MAXMINAVG Dynamic Recording mode: Present value on primary display
MAX Dynamic Recording mode: Maximum value on primary display
MIN Dynamic Recording mode: Minimum value on primary display
AVG
MAX
MIN
Trigger (manual) Hold
Dynamic Recording mode: Average value on primary display
1 ms Peak Hold mode: Positive peak value on primary display
1 ms Peak Hold mode: Negative peak value on primary display
12 Agilent U1251A/U1252A User’s and Service Guide
Tab l e 1 - 4 Priamry display symbols
LCD symbol Description
Auto range
AC + DC
DC
AC
Polarity, digits, and decimal points for primary display
Getting Started Tutorial 1
The primary display signs are explained below.
dBm
dBV
MkHz
MkΩ Resistance units: Ω, kΩ, MΩ
nS
mV
µmA
%
ms
µmnF
ºC
ºF
%
%
Decibel unit relative to 1 mW
Decibel unit relative to 1 V
Frequency units: Hz, kHz, MHz
Conductance unit
Voltage units: mV, V
Current units: µA, mA, A
Duty cycle measurement
Pulse width unit
Capacitance units: nF, µF, mF
Celsius temperature unit
Fahrenheit temperature unit
Percentage scale readout proportional to DC 0–20 mA
Percentage scale readout proportional to DC 4–20 mA
Agilent U1251A/U1252A User’s and Service Guide 13
1 Getting Started Tutorial
The secondary display signs are explained below.
Tab l e 1 - 5 Secondary display symbols
LCD symbol Description
AC + DC
DC
AC
Polarity, digits and decimal points for secondary display
kHz Frequency units: Hz, kHz
No ambient temperature compensation, just thermocouple measurement
ºC
ºF
mV
µmA
s
Celsius ambient temperature unit
Fahrenheit ambient temperature unit
Voltage units: mV, V
Current units:
Elapsed time unit: s (second) for Dynamic Recording and 1 ms Peak Hold modes
µA, mA, A
The analog bar emulates the needle on an analog multimeter, without displaying the overshoot. When measuring peak or null adjustments and viewing fast- changing inputs, the bar graph provides a useful indication because it has quicker update rate for fast response applications.
The bar graph is not used for square wave output, frequency, duty cycle, pulse width, 4–20 mA% scale, 0–20 mA% scale and temperature measurements. When frequency, duty cycle, and pulse width are indicated on the primary display during voltage or current measurement, the bar graph represents the voltage or current value. When 4–20 mA% scale or 0–20 mA% scale is indicated on the primary display, the bar graph represents the current value and not the percentage.
14 Agilent U1251A/U1252A User’s and Service Guide
Tab l e 1 - 6 Analog bar range and counts
Range Counts/segments Used for the function
Getting Started Tutorial 1
The “+” or “–” sign is indicated when the positive or negative value has been measured or calculated. Each segment represents 2500 or 500 counts depending on the range indicated on the peak bar graph. See the Table 1- 6 below.
2500 V, , Diode
2500 V, A,
2500 V, A, , nS
500
V,
500
500
Agilent U1251A/U1252A User’s and Service Guide 15
1 Getting Started Tutorial

Selecting display with the Hz button

The frequency measurement feature helps to detect the presence of harmonic currents in neutral conductors and determines whether these neutral currents are the result of unbalanced phases or non- linear loads. Press to access the frequency measurement mode for current or voltage measurements — voltage or current on the secondary display and frequency on the primary display. Alternatively, pulse width (ms) or duty cycle (%) can appear on the primary display by pressing again. This enables the simultaneous monitoring of real- time voltage and current with frequency, duty cycle, or pulse width. Voltage or current resumes on primary display after you press and hold
for more than 1 second.
Tab l e 1 - 7 Selecting display with the Hz button
Rotary switch position (Function) Primary display Secondary display
for U1252A
(AC voltage)
V for U1251A
for U1252A
(DC voltage)
for U1252A
(AC + DC voltage)
(AC voltage)
(DC voltage)
Frequency (Hz) AC V
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) DC V
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) AC + DC V
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) AC mV
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) DC mV
Pulse width (ms)
Duty cycle (%)
16 Agilent U1251A/U1252A User’s and Service Guide
Tab l e 1 - 7 Selecting display with the Hz button (continued)
Frequency (Hz) AC + DC mV
(AC + DC voltage)
[for U1252A]
(AC Current)
(DC current)
(AC + DC current)
[for U1252A]
(AC current)
(DC current)
[for U1252A]
(AC + DC current)
Hz (Frequency counter) - press
to select frequency division by 1 [for
U1252A]
Hz (Frequency counter) - press
to select frequency division by 100 [for
U1252A]
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) AC µA
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) DC µA
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) AC + DC µA
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) AC mA or A
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) DC mA or A
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) AC + DC mA or A
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) - 1 -
Pulse width (ms)
Duty cycle (%)
Frequency (Hz) - 100 -
Getting Started Tutorial 1
Agilent U1251A/U1252A User’s and Service Guide 17
1 Getting Started Tutorial

Selecting display with the Dual button

Press to select different combinations of dual display.
Normal single display resumes after you press and hold
for more than 1 second. See Table 1- 8 below.
Tab l e 1 - 8 Selecting display with the Dual button
Rotary switch position (Function) Primary display Secondary display
AC V Hz (AC coupling)
(AC voltage)
for U1252A
(AC voltage)
V for U1251A/
for U1252A
(DC voltage)
for U1252A
(AC + DC voltage)
(AC voltage)
(DC voltage)
dBm or dBV (select
by pressing )
AC V
Ambient temperature °C or °F
AC V Hz (AC coupling)
dBm or dBV
[1]
AC V DC V AC V
Ambient temperature °C or °F
DC V Hz (DC coupling)
dBm or dBV
[1]
DC V AC V [for U1252A] DC V
Ambient temperature °C or °F
AC + DC V Hz (AC coupling)
dBm or dBV
[1]
AC + DC V AC V AC + DC V DC V AC + DC V
Ambient temperature °C or °F
AC mV Hz (AC coupling)
dBm or dBV
[1]
AC mV DC mV AC mV
Ambient temperature °C or °F
DC mV Hz (DC coupling)
dBm or dBV
[1]
DC mV AC mV DC mV
Ambient temperature °C or °F
AC V
AC V
DC V
AC + DC V
AC mV
DC mV
18 Agilent U1251A/U1252A User’s and Service Guide
Tab l e 1 - 8 Selecting display with the Dual button
(AC + DC voltage)
(DC current)
(AC current)
(AC + DC current)
[for U1252A]
(DC current)
(AC current)
(AC + DC current)
[for U1252A]
(DC current)
(AC current)
Getting Started Tutorial 1
AC + DC mV Hz (AC coupling) dBm or dBV AC + DC mV AC + DC mV AC mV AC + DC mV DC mV AC + DC mV
DC μA
DC μAAC μA
DC μA Ambient temperature °C or °F
AC μA
AC μADC μA
AC μA Ambient temperature °C or °F
AC + DC μA
AC + DC μAAC μA
AC + DC μADC μA
AC + DC μA Ambient temperature °C or °F
DC mA Hz (DC coupling) DC mA AC mA
% (0–20 or 4–20) DC mA
DC mA
AC mA Hz (AC coupling) AC mA DC mA AC mA
AC + DC mA Hz (AC coupling) AC + DC mA AC mA AC + DC mA DC mA AC + DC mA
DC A Hz (DC coupling)
DC A AC A DC A
AC A Hz (AC coupling) AC A DC A AC A
Ambient temperature °C or °F
Hz (DC coupling)
Hz (AC coupling)
Hz (AC coupling)
Ambient temperature °C or °F
Ambient temperature °C or °F
Ambient temperature °C or °F
Ambient temperature °C or °F
Ambient temperature °C or °F
Agilent U1251A/U1252A User’s and Service Guide 19
1 Getting Started Tutorial
Tab l e 1 - 8 Selecting display with the Dual button
AC + DC A Hz (AC coupling) AC + DC A AC A
(AC + DC current)
[for U1252A]
(Capacitance)
AC + DC A DC A AC + DC A
nF / V / / nS
Ambient temperature °C or °F
Ambient temperature °C or °F
(Diode)/
(Resistance)/
nS (Conductance)
TEMP (Temperature)
°C (°F) Ambient temperature °C or °F
°C (°F) Ambient temperature °C or °F /
0 °C compensation (select by
pressing )
Notes for selecting display with the Dual button: 1 Reading of dBm or dBV depends on the last review on AC V. If the last review is in dBV, the following display
will also remain in dBV.

Selection display with the Shift button

The table below shows selection of primary display, with respect to measuring function (rotary switch position), using the SHIFT button.
Tab l e 1 - 9 Selecting display with the Shift button
Rotary switch position (Function) Primary display
AC V
(AC Voltage)
dBm (in dual display mode)
dBV (in dual display mode)
V for U1251A
for U1252A
(AC + DC Voltage)
DC V
DC V AC V
AC + DC V
20 Agilent U1251A/U1252A User’s and Service Guide
[1]
[1]
Tab l e 1 - 9 Selecting display with the Shift button
Getting Started Tutorial 1
for U1252A
(AC + DC Voltage)
Ω
(Resistance)
(Diode Test and Frequency)
/ TEMP
(Capacitance and Temperature)
(AC Current)
(DC Current)
(AC + DC Current)
DC mV AC mV
AC + DC mV
Ω
Ω
nS
Diode
Hz
Capacitance Temperature
DC μA
AC μA
AC + DC μA [for U1252A]
DC mA AC mA
AC + DC mA
%(0–20 or 4–20)
DC A AC A
AC + DC A [for U1252A]
Duty cycle (%)
Pulse width (ms)
(Square wave output for U1252A)
Notes for selecting display with the Shift button:
1 Press to switch between dBm and dBV measurement.
Press for more than 1 second to return to AC V measurement only.
Agilent U1251A/U1252A User’s and Service Guide 21
1 Getting Started Tutorial
WARNING

The terminals at a glance

To avoid damaging the multimeter, do not exceed the input limit.
Figure 1-9 U1252A connector terminals
Table 1-10 Terminal connections for different measuring functions
22 Agilent U1251A/U1252A User’s and Service Guide
Rotary switch position Input terminals Overload protection
V . mV . Ω .
for U1252A
V for U1251A
Ω
for U1252A COM
.TEMP
μA . mA COM 440 mA / 1000 V 30 kA fast-acting
A COM 11 A / 1000 V 30 kA fast-acting fuse
COM 1000 V R.M.S.
1000 V R.M.S. for short circuit <0.3 A
fuse
COM 440 mA / 1000 V fast-acting fuse
Agilent U1251A and U1252A Handheld Digital Multimeter User’s and Service Guide
2 Making Measurements
Understanding The Measurement Instructions 24 Measuring Voltage 24
Measuring AC voltage 25 Measuring DC voltage 26
Measuring Current 27
µA & mA measurement 27 Percentage scale of 4 mA to 20 mA 29
A (ampere) measurement 31 Frequency Counter 32 Measuring Resistance, Conductance and Testing Continuity 34 Tes tin g Di od es 38 Measuring Capacitance 41 Measuring Temperature 43 Alerts and Warning During Measurement 47
Overload alert 47
Input warning 47
Charge terminal alert 48
This chapter contains information on how to make measurements using the U1251A and U1252A handheld digital multimeter.
Agilent Technologies
23
2 Making Measurements
WARNING

Understanding The Measurement Instructions

When making measurements, follow the numerical steps labelled in the diagrams. Refer to Table 2- 1 below for a description of the steps.
Tab l e 2 - 1 Numerical steps descriptions
No. Instructions
Turn the rotary switch to the measurement option shown in the diagram
Connect the test leads into the input terminals shown in the diagram
Probe the test points
Read the results on the display

Measuring Voltage

24 Agilent U1251A/U1252A User’s and Service Guide
The U1251A and U1252A offer true- RMS readings for AC
measurements that are accurate for sine waves, square
waves, triangle waves, staircase waves, and other waveforms
without any DC offset.
For AC with DC offset, use AC + DC measurement on or rotary switch location. This applies only to U1252A.
Ensure that the terminal connections are correct for that particular measurement before proceeding with the measurement. To avoid damaging the device, do not exceed the input limit.

Measuring AC voltage

Red Lead
Black Lead
NOTE
Set up the multimeter to measure AC voltage as shown in
Figure 2- 1. Probe the test points and read the display.
Making Measurements 2
Figure 2-1 Measuring AC voltage
Press to display frequency on secondary display. See Ta b l e 1 - 8 of
“Selecting display with the Dual button” on page 18 for a list of the
different combinations available on the secondary display.
Agilent U1251A/U1252A User’s and Service Guide 25
2 Making Measurements
Red Lead
Black Lead

Measuring DC voltage

Set up the multimeter to measure DC voltage as shown in
Figure 2- 2. Probe the test points and read the display.
MAX
Figure 2-2 Measuring DC voltage
26 Agilent U1251A/U1252A User’s and Service Guide

Measuring Current

NOTE

µA & mA measurement

Making Measurements 2
Set up the multimeter to measure μA & mA as shown in
Figure 2- 3. Probe the test points and read the display.
Press if necessary to ensure is shown on the display.
For μA measurement, set the rotary switch to and connect
the positive test lead to μA.mA.
For mA measurement, set the rotary switch to and
connect the positive test lead to μA.mA.
For A (ampere) measurement, set the rotary switch to and
connect the positive test lead to A.
Press to display dual measurements. See Ta b l e 1 - 8 of “Selecting
display with the Dual button” on page 18 for a list of dual
measurements available. Press and hold for more than 1 second to exit dual display mode.
Agilent U1251A/U1252A User’s and Service Guide 27
2 Making Measurements
Red Lead
Black Lead
Figure 2-3 Measuring μA and mA current
28 Agilent U1251A/U1252A User’s and Service Guide

Percentage scale of 4 mA to 20 mA

NOTE
Set up the multimeter to measure percentage scale as shown in Figure 2- 4. Probe the test points and read the display.
Press to select percentage scale display. Ensure that or
is shown on the display.
The percentage scale for 4 mA to 20 mA or 0 mA to 20 mA is calculated
using its corresponding DC mA measurement. The U1251A/U1252A will automatically optimize the best resolution according to Ta b l e 2 - 2 below.
Press to change the measurement range.
The percentage scale for 4 mA to 20 mA or 0 mA to 20 mA is set to two ranges as follows:
Tab l e 2 - 2 Percentage scale and measurement range
Making Measurements 2
Percentage scale (4 mA to 20 or 0 mA to 20 mA)
Always auto range
999.99%
9999.9%
Agilent U1251A/U1252A User’s and Service Guide 29
DC mA auto or manual
range
50 mA, 500 mA
2 Making Measurements
Red Lead
Black Lead
Figure 2-4 Measuring scale of 4-20 mA
30 Agilent U1251A/U1252A User’s and Service Guide

A (ampere) measurement

NOTE
Red Lead
Black Lead
Set up the multimeter to measure A (ampere) as shown in
Figure 2- 5. Probe the test points and read the display.
Making Measurements 2
Connect the red and black test leads to the 10 A input terminal A and COM respectively. The meter is set to A measurement automatically when the red test lead is plugged into the A terminal.
Figure 2-5 A (ampere) current measurement
Agilent U1251A/U1252A User’s and Service Guide 31
2 Making Measurements
WARNING
NOTE

Frequency Counter

Use the frequency counter for low voltage applications only. Never use
the frequency counter for line power system.
For inputs more than 30 Vpp, you are required to use frequency
measurement mode available under the current or voltage measurement instead of frequency counter.
Set up the multimeter to measure frequency as shown in
Figure 2- 6. Probe the test points and read the display.
Press to select the Frequency counter (Hz) function. “-1-“ on
the secondary display means the input signal frequency is divided by 1. This allows signals of up to a maximum frequency of 985 kHz to be measured.
If the reading is unstable or is zero, press to select the division
of the input signal frequency by 100. This allows for a higher frequency range of up to 20 MHz to be measured.
32 Agilent U1251A/U1252A User’s and Service Guide
The signal is out of range if the reading is still unstable after the above
step.
While the secondary display shows “-1-“, you may scroll through the
pulse width (ms), duty cycle (%), and frequency (Hz) measurements by pressing .
Making Measurements 2
Hz
Red Lead
Black Lead
Figure 2-6 Measuring frequency
Agilent U1251A/U1252A User’s and Service Guide 33
2 Making Measurements
CAUTION
Red Lead
Black Lead
nS
Ω

Measuring Resistance, Conductance and Testing Continuity

Disconnect the circuit power and discharge all high-voltage capacitors before measuring the resistance to prevent any possible damage to the multimeter or the device under test.
Set up the multimeter to measure resistance as shown in
Figure 2- 7. Then probe the test points (by shunting the
resistor) and read the display.
34 Agilent U1251A/U1252A User’s and Service Guide
Figure 2-7 Measuring resistance
Audible continuity
Making Measurements 2
Press to scroll through audible continuity, conductance and resistance tests as shown in Table 2- 8.
Press SHIFT
MAX
Press SHIFT
Figure 2-8 Audible continuity, conductance, and resistance test.
Agilent U1251A/U1252A User’s and Service Guide 35
Press
2 Making Measurements
Audible continuity
In the range of 0–500 Ω, the beeper will sound if the resistance value falls below 10 Ω. For other ranges, the beeper will sound if the resistance falls below the typical values indicated in Table 2- 3 below.
Tab l e 2 - 3 Audible continuity measurement range
Measurement range Beeper sound threshold
500.00 Ω < 10 Ω
5.0000 kΩ < 100 Ω
50.000 kΩ < 1 kΩ
500.00 kΩ < 10 kΩ
5.0000 MΩ < 100 kΩ
50.000 MΩ < 1 MΩ
500.00 MΩ < 10 MΩ
Conductance
Set up the multimeter to measure conductance as shown in
Figure 2- 9. Probe the test points and read the display.
The conductance measurement enables the measurement of very high resistance of up to 100 GΩ.
As the high- resistance readings are susceptible to noise, you can capture the average readings by using the Dynamic Recording mode. Refer to the section “Dynamic
Recording” on page 50 for more information.
36 Agilent U1251A/U1252A User’s and Service Guide
Making Measurements 2
Ω
nS
Red Lead
Black Lead
MAX
Figure 2-9 Conductance measurement
Agilent U1251A/U1252A User’s and Service Guide 37
2 Making Measurements
CAUTION
NOTE

Testin g D iodes

Disconnect the circuit power and discharge all the high-voltage capacitors before testing the diodes to prevent any possible damage to the meter.
To test a diode, turn the power off to the circuit and remove the diode from the circuit. Set up the multimeter as shown in Figure 2- 10, then use the red probe lead on the positive terminal (anode) and use the black probe lead on the negative terminal (cathode) and read the display.
The cathode is the side with band(s).
The meter can display the diode’s forward bias of up to approximately
2.1 V. A typical diode’s forward bias is between the range of 0.3 V to
0.8 V.
38 Agilent U1251A/U1252A User’s and Service Guide
Next, reverse the probes and measure the voltage across the diodes again as shown in Figure 2- 11 on page 40. The diode’s test result is based on the following:
The diode is considered good if the meter displays “OL” in reverse bias mode.
The diode is considered shorted if the meter displays approximately 0 V in both forward and reverse bias modes, and the meter beeps continuously.
The diode is considered open if the meter displays “OL” in both forward and reverse bias modes.
Making Measurements 2
Red Lead
Black Lead
Hz
Figure 2-10 Measuring forward bias of diode
Agilent U1251A/U1252A User’s and Service Guide 39
2 Making Measurements
Red Lead
Black Lead
Hz
Figure 2-11 Measuring reverse bias of diode
40 Agilent U1251A/U1252A User’s and Service Guide

Measuring Capacitance

CAUTION
NOTE
Disconnect the circuit power and discharge all the high-voltage capacitors before measuring the capacitance to prevent any possible damage to the meter or the device under test. To confirm that the capacitors have discharged, use the DC voltage function.
The meter measures capacitance by charging the capacitor with a known current for a period of time, measuring the voltage and then calculating the capacitance. The larger the capacitor, the longer the charge time. Below are some tips for measuring capacitance:
Making Measurements 2
For measuring capacitance greater than 10,000 µF, discharge the capacitor first, then select a suitable range for the measurement. This will speed up measuring time in order to obtain the correct capacitance value.
For measuring smaller capacitance ranges, press with the test leads open to subtract any residual capacitance from the meter or the leads.
Agilent U1251A/U1252A User’s and Service Guide 41
means the capacitor is charging. means the capacitor is
discharging.
Set up the multimeter as shown in Figure 2- 12. Use the red probe lead on the positive terminal of the capacitor and the black probe lead on the negative terminal and read the display.
2 Making Measurements
Red Lead
Black Lead
TEMP
Figure 2-12 Capacitance measurements
42 Agilent U1251A/U1252A User’s and Service Guide

Measuring Temperature

CAUTION
Do not bend the thermocouple leads at sharp angles. Repeated bending over a period of time may break the leads.
The bead type thermocouple probe is suitable for making temperature measurements between –20 °C to 200 °C in PTFE compatible environments.
Do not use the bead- type thermocouple probe beyond the recommended operating temperature range. Do not immerse this thermocouple probe in liquids. For best results, use a thermocouple probe designed for each application — an immersion probe for liquid or gel and an air probe for air measurements.
Set up the multimeter to measure temperature as shown in
Figure 2- 15 or observe the following steps:
1 Press to select the temperature measurement.
2 Connect the miniture thermal probe into the
3 Connect the thermal probe with the adapter into the
4 For best performance, place the meter in the operating
5 Clean the measurement surface and make sure the probe
6 When measuring above the ambient temperature, move the
7 When measuring below ambient temperature, move the
Making Measurements 2
non- compensation transfer adaptor as shown in
Figure 2- 13.
meter input terminals as shown in Figure 2- 14.
environment for at least one hour to stabilize the unit to environment temperatures.
is securely touching the surface. Remember to disable the applied power.
thermocouple along the surface until you get the highest temperature reading.
thermocouple along the surface until you get the lowest temperature reading.
Agilent U1251A/U1252A User’s and Service Guide 43
2 Making Measurements
8 For a quick measurement, use the 0 °C compensation
adapter to see the temperature variation of the thermocouple sensor. The 0 °C compensation adapter assists in measuring the relative temperature immediately.
Figure 2-13 Connecting the thermal probe into the non-compensation
transfer adapter
Figure 2-14 Connecting the probe with adapter into the multimeter
44 Agilent U1251A/U1252A User’s and Service Guide
Making Measurements 2
If you are working in a constantly varying environment, where ambient temperatures are not constant, do the following:
1 Press to select 0 °C compensation. This gives a
quick measurement of the relative temperature.
2 Avoid contact between the thermocouple probe and the
measurement surface.
3 After a constant reading is obtained, press to set
the reading as the relative reference temperature.
4 Touch the measurement surface with the thermocouple
probe.
5 Read the display for the relative temperature.
Agilent U1251A/U1252A User’s and Service Guide 45
2 Making Measurements
TEMP
Figure 2-15 Surface temperature measurement
46 Agilent U1251A/U1252A User’s and Service Guide

Alerts and Warning During Measurement

WARNING

Overload alert

For your safety, look out for this alert. When you are alerted, immediately remove the test leads from the measuring source.
The meter provides an overload alert for voltage measurements in both auto and manual range modes. The meter beeps periodically once the measuring voltage exceeds 1010 V. For your safety, please be aware of this alert.

Input warning

The meter sounds an alert beep when the test lead is inserted into the A input terminal but the rotary switch is not set to the corresponding mA.A location. The primary display indicates a flashing “A- Err” until the test lead is removed from the A input terminal. Refer to Figure 2- 16.
Making Measurements 2
Agilent U1251A/U1252A User’s and Service Guide 47
Figure 2-16 Input terminal warning
2 Making Measurements

Charge terminal alert

The meter sounds an alert beep when the terminal
detects a voltage level of more than 5 V and the rotary
switch is not set to the corresponding location. The
primary display indicates a flashing “Ch.Err” until the lead
is removed from the input terminal.
Refer to Figure 2- 17 below.
Figure 2-17 Charge terminal alert
48 Agilent U1251A/U1252A User’s and Service Guide
Agilent U1251A and U1252A Handheld Digital Multimeter User’s and Service Guide
3 Functions and Features
Dynamic Recording 50 Data Hold (Trigger Hold) 52 Refresh Hold 53 Null (Relative) 55 Decibel Display 57 1 ms Peak Hold 59 Data Logging 61
Manual logging 61 Interval logging 63
Reviewing logged data 65 Square Wave Output (for U1252A) 67 Remote Communication 71
This chapter contains information on the functions and features that are available for the U1251A and U1252A digital multimeter.
Agilent Technologies
49
3 Features and Functions
NOTE

Dynamic Recording

The Dynamic Recording mode can be used to detect intermittent turn- on or turn- off voltage, current surges or to verify measurement performance without you being present during the process. While the readings are being recorded, you are free to perform other tasks.
The average reading is useful for smoothing out unstable inputs, estimating the percentage of time a circuit is operating and verifying circuit performance. The elapsed time is shown on the secondary display. The maximum time is 99999 seconds. When this maximum time is exceeded, OL is shown on the display.
1 Press for more than 1 second to enter Dynamic
Recording mode. The meter is now in continuous mode or non- data hold (non- trigger) mode. “MAXMINAVG” and the present value of measurement are displayed. The multimeter will beep when a new maximum or minimum value is recorded.
2 Press to cycle through maximum, minimum,
average, and present readings. The MAX, MIN, AVG , and MAXMINAVG indicators will appear, corresponding to the displayed readings.
3 Press or for more than 1 second to exit
Dynamic Recording mode.
Press to restart the dynamic recording.
The average value is the true average of all the measured values taken
in the Dynamic Recording mode. If an overload is recorded, the averaging function will stop and the average value becomes "OL"(overload). is disabled in Dynamic Recording mode.
50 Agilent U1251A/U1252A User’s and Service Guide
Features and Functions 3
Press Max Min for > 1 sec
or press for > 1 sec
Press Max Min for > 1 sec
or press for > 1 sec
Press Max Min for > 1 sec
or press for > 1 sec
Press Max Min for > 1 sec
Press Max Min
Press Max Min
Press Max Min
Figure 3-1 Dynamic recording mode operation
Agilent U1251A/U1252A User’s and Service Guide 51
3 Features and Functions
Press
Press
Press

Data Hold (Trigger Hold)

The data hold function allows the user to freeze the displayed digital value.
1 Press to freeze the displayed value and to enter the
2 Press again to freeze the next value being
3 Press and hold or for more than 1 second to
manual trigger mode. TRIG is displayed.
measured. TRIG flashes before the new value is updated on the display.
quit the data hold function.
Figure 3-2 Data hold mode operation
52 Agilent U1251A/U1252A User’s and Service Guide

Refresh Hold

Features and Functions 3
The Refresh Hold function allows you to freeze the displayed value. The bar- graph is not held and will continue to reflect the instantaneous measured value. You can use the Setup
mode to enable Refresh Hold mode when you are working with fluctuating values. This function will auto trigger or update the held value with a new measured value and emit a tone as a reminder.
1 Press to enter Refresh Hold mode. The present
value will be held and the symbol will appear.
2 It will be ready to hold a new measured value once the
variation of measured values exceeds the variation count setting. While the multimeter is waiting for a new stable value, the symbol will flash.
3 The symbol will stop f lashing once the new
measured value is stable and the new value will be updated to the display. The symbol will remain on and the multimeter will emit a tone to remind you of this.
4 Press again to quit the Refresh Hold function.
Agilent U1251A/U1252A User’s and Service Guide 53
3 Features and Functions
The HOLD sign flashes until the measured value is stable and updated
Press
Variation counts
> setting
Measurement is updated
Press to return
NOTE
54 Agilent U1251A/U1252A User’s and Service Guide
Figure 3-3 Refresh hold mode operation
For voltage and current measurements, the holding value will not be
updated if the reading is below 500 counts.
For resistance and diode measurements, the holding value will not be
updated if the reading is in “OL” (open state).
The holding value may not be updated when the reading does not reach
a stable state for all measurements.

Null (Relative)

NOTE
Features and Functions 3
The Null function subtracts a stored value from the present
measurement and displays the difference between the two.
1 Press to store the displayed reading as the
reference value to be subtracted from subsequent
measurements and to set the display to zero. NULL is
displayed.
2 Press to see the stored reference value. NULL
flashes for 3 seconds before the display returns to zero.
3 To exit this mode, press while NULL is flashing on
the display.
Null can be set for both auto and manual range settings, but not in the
event of an overload.
When taking a resistance measurement and the meter reads a
non-zero value due to the presence of test leads, use the Null function to adjust the display to zero.
When taking a DC voltage measurement, the thermal effect will
influence the accuracy. Short the test leads and press Null once the displayed value is stable in order to zero out the display.
Agilent U1251A/U1252A User’s and Service Guide 55
3 Features and Functions
Press
Press
Press
Figure 3-4 Null (relative) mode operation
56 Agilent U1251A/U1252A User’s and Service Guide

Decibel Display

NOTE
Features and Functions 3
The dBm operation calculates the power delivered to a reference resistance relative to 1 mW and can be applied to DC V, AC V, and AC + DC V measurements for decibel conversion. The voltage measurement is converted to dBm by using the following formula:
The reference resistance may be selected from 1~9999 Ω in Setup mode. The default value is 50 Ω.
The decibel of voltage is calculated with respect to 1 V. The equation for the voltage measurement is as follows:
dBV = 20 log
1 At the , , or rotary switch position,
press to scroll to the dBm measurement on the primary display. The secondary display indicates the AC voltage measurement.
If the rotary switch is at the “~ V” position, press to switch between the dBV and the dBm measurements. The dBm or the dBV measurements can be selected at the ACV position. The selection will be the reference for other voltage measurements.
2 Press for more than 1 second to exit this mode.
10
Vin
Agilent U1251A/U1252A User’s and Service Guide 57
3 Features and Functions
Press
Press
Press
Press
Press
Press
Press
Press
Press
Press
58 Agilent U1251A/U1252A User’s and Service Guide
Figure 3-5 dBm/dBV display mode operation

1 ms Peak Hold

NOTE
Features and Functions 3
The Peak Hold function allows the measurement of peak
voltage for analysis of components such as power
distribution transformers and power factor correction
capacitors. The peak voltage obtained can be used to deter-
mine the crest factor:
Crest factor = Peak value/True RMS value
1 Press for more than 1 second to toggle 1 ms Peak
Hold mode ON / OFF.
2 Press to scroll through maximum and minimum
peak readings. MAX indicates the maximum peak, while MIN indicates the minimum peak.
If the reading is "OL", press to change the measuring range and
to re-start the peak-recording measurement.
If you need to re-start the peak recording, press .
3 Press and hold or for more than 1 second
to exit this mode.
4 According to the measurements shown in Figure 3- 6 on
page 60, the crest Factor will be 2.5048/1.768 =1.416.
Agilent U1251A/U1252A User’s and Service Guide 59
3 Features and Functions
Press
Press
Press
Press
Press
Press
Figure 3-6 1 ms peak hold mode operation
60 Agilent U1251A/U1252A User’s and Service Guide

Data Logging

NOTE
Features and Functions 3
The data logging function provides the convenience of recording test data for future review or analysis. Since data is stored in nonvolatile memory, the data remains saved even if the multimeter is turned OFF or the battery is changed.
The two options offered are manual (hand) logging and interval (time) logging functions, which is determined in the Setup mode.
Data logging records the values on the primary display only.
To use the data logging function, you will need to connect the multimeter to a PC using the U1173A IR-to-USB cable (purchased separately) and download the data logging software from Agilent’s website. Please go to:
http://www.agilent.com/find/hhTechLib to download the software.

Manual logging

Firstly, ensure that manual (hand) logging is specified in
Setup mode.
1 Press for more than 1 second to store the present
value and function on the primary display to the
non- volatile memory. The and the logging index
will be indicated. The logging index flashes on the
secondary display for 3 seconds before returning to
normal display.
2 Press and hold again for the next value that you
would like to save into the memory.
Agilent U1251A/U1252A User’s and Service Guide 61
3 Features and Functions
Press
Press
NOTE
Press
Figure 3-7 Hand (Manual) logging mode operation
The maximum data that can be stored is 100 entries. When the 100 entries are filled, the secondary display indicates “FULL”, as shown in Figure 3-8.
Figure 3-8 Full Log
62 Agilent U1251A/U1252A User’s and Service Guide

Interval logging

NOTE
NOTE
Firstly, ensure that interval (time) logging is specified in
Setup mode.
1 Press for more than 1 second to store the present
The maximum data that can be stored is 200 entries. When the 200 entries are filled, “FULL” is indicated on the secondary display.
2 Press for more than 1 second to exit this mode.
When interval (automatic) logging is enabled, all keypad operations are disabled, except for the Log function.
Features and Functions 3
value and function on the primary display into the non- volatile memory. The and the logging index will be indicated. The readings are automatically saved into the permanent memory at intervals set using the Setup mode.
Agilent U1251A/U1252A User’s and Service Guide 63
3 Features and Functions
Press
Press
Press
Figure 3-9 Interval (Automatic) logging mode operation
64 Agilent U1251A/U1252A User’s and Service Guide

Reviewing logged data

1 Press for more than 1 second to enter the Log
Review mode. The last recorded entry and the last logging index are displayed.
2 Press to switch between hand (manual) and interval
(automatic) logging review mode.
3 Press to ascend or to descend through the logged
data. Press to select the first record and press to select the last record for quick navigation.
4 Press for more than 1 second at the respective Log
Review mode to clear logged data.
5 Press for more than 1 second to exit mode.
6 During the data review in either manual or interval
logging mode, press for more than one second to clear all the logged values.
Features and Functions 3
Agilent U1251A/U1252A User’s and Service Guide 65
3 Features and Functions
Press
Press
Press
Press
Press
Press
Press
Press
Press
Press
Press
Press
Press
Press
Press
Press
Press Press
Press
Figure 3-10 Log review mode operation
66 Agilent U1251A/U1252A User’s and Service Guide

Square Wave Output (for U1252A)

NOTE
The square wave output function can be used to generate a
Press
0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 80, 100, 120, 150, 200, 240, 300, 400, 480, 600, 800, 1200, 1600, 2400, 4800
PWM (pulse width modulation) output or provide a synchro-
nous clock source (baud rate generator). You can also use
this function to check and calibrate flow- meter displays,
counters, tachometers, oscilloscopes, frequency converters,
frequency transmitters, and other frequency input devices.
1 Turn the rotary switch to position. The default
display setting is 600 Hz on the secondary display and 50% duty cycle on the primary display.
2 Press or to scroll through the available frequencies
(choose from the available 28 frequencies):
Frequency (Hz)
Features and Functions 3
Pressing is the same as pressing .
3 Press to select the duty cycle (%) on the primary
display.
4 Press or to adjust the duty cycle. Duty cycle can
be set for 256 steps and each step is 0.390625%. The display only indicates the best resolution with 0.001%.
Agilent U1251A/U1252A User’s and Service Guide 67
3 Features and Functions
Figure 3-11 Frequency adjustment for square wave output
5 Press to select the pulse width (ms) on the primary
display.
6 Press or to adjust the pulse width. The pulse
width can be set for 256 steps and each step is 1/ (256 x Frequency). The display range automatically adjusts in the range of 9.9999~9999.9 ms.
68 Agilent U1251A/U1252A User’s and Service Guide
Press
Press
Press
Press
Press
Press
Press
Press
Features and Functions 3
Agilent U1251A/U1252A User’s and Service Guide 69
Figure 3-12 Pulse width adjustment for square wave output
3 Features and Functions
Figure 3-13 Duty cycle adjustment for Square Wave
70 Agilent U1251A/U1252A User’s and Service Guide

Remote Communication

to PC
Connect
The meter has a bi- directional (full duplex) communication capability that speeds data storing from the meter to the PC. To use this feature, you are required to use the optional IR- USB cable with the application software included in the CD or downloadable from the Agilent Web site.
For details on performing PC to meter remote communication click on Help after launching the Agilent GUI Data Logger Software or refer to the GUI Data Logger Quick
Start Guide (U1251- 90023) for more information.
Features and Functions 3
Figure 3-14 Cable connection for remote communication
Agilent U1251A/U1252A User’s and Service Guide 71
3 Features and Functions
72 Agilent U1251A/U1252A User’s and Service Guide
Agilent U1251A and U1252A Handheld Digital Multimeter User’s and Service Guide
4 Changing The Default Setting
Selecting Setup Mode 74 Setting Data Logging Mode 78 Setting Thermocouple Types (U1252A only) 79 Setting Reference Impedance for dBm Measurement 80 Setting Minimum Frequency Measurement 81 Setting Temperature Unit 82 Setting Auto Power Saving Mode 84 Setting Percentage (%) Scale Readout 86 Setting Beep Frequency 87 Setting Backlight Timer 88 Setting Baud Rate 89 Setting Parity Check 90 Setting Data Bit 91 Setting Echo Mode 92 Setting Print Mode 93 Returning to Default Factory Settings 94
This chapter shows you how to change the default factory settings of the U1251A and U1252A and other available setting options
Agilent Technologies
73
4 Changing The Default Setting
NOTE

Selecting Setup Mode

To enter the Setup mode, perform the following steps:
1 Turn the meter OFF.
2 From the OFF position, press and hold while
turning the rotary switch to any non- OFF position.
When you hear a beep, the meter is in Setup mode and you can release
To change a menu item setting in Setup mode, perform the
following steps:
1 Press or to scroll through the menu items.
2 Press or to scroll through the available settings.
See Table 4- 1, “Available setting options in Setup mode,” for details of available options.
3 Press to save changes. These parameters will remain
in the non- volatile memory.
4 Press for more than 1 second to exit Setup mode.
74 Agilent U1251A/U1252A User’s and Service Guide
Table 4-1 Available setting options in Setup mode
Menu Item Available setting options Default factory
Display Description Display Description
OFF Enables Data Hold (manual trigger)
100–1000
Sets the variation count that determines Refresh
Hold (auto trigger)
Enables the reset of factory settings by pressing
dEFAU
and holding for more than 1 second
rHoLd
rESEt
[1]
Refresh Hold
Reset
Changing The Default Setting 4
setting
500
dEFAU
Print
ECHO
dAtAb
PAr tY
bAUd
b-Lit
bEEP
PErnt
APF
FrEq
rEF
Print
Echo
Data bits
Parity check
Baud rate
Backlit display
Frequency of beep
sound of meter
Percentage scale
Auto power off
Minimum frequency
that can be measured
Reference impedance
for dBm measurement
ON, OFF
ON, OFF
7-bit, 8-bit
En, Odd, nOnE
2400 Hz, 4800
Hz, 9600 Hz,
19200 Hz
1–99 s
Enables auto send of data to the PC continuously
when set to ON
Enables the return of characters to the PC when
set to ON
Sets the data bit length for the remote
communication (remote control with PC)
Sets the even, odd or no parity check for the
remote communication (remote control with PC)
Sets the baud rate for the remote communication
(remote control with PC)
Sets the timer for the auto turn-off for the backlit
[2]
display
OFF Disables the auto turn-off for the backlit display
2400 Hz, 1200
Hz, 600 Hz,
Sets the frequency of the meter’s beep sound
300 Hz
OFF Disables the meter’s beep sound
0–20 mA, 4–20
mA
[2]
1–99 m
Sets the % scale readout 4–20 mA
Sets the timer for auto power off
OFF Disables auto power off
0.5 Hz, 1 Hz, 2 Hz, 5 Hz
[2]
1–9999 Ω
Sets the minimum frequency that can be
measured
Sets the reference impedance for dBm
measurement
OFF
OFF
8-bit
nOnE
9600 Hz
30 s
2400 Hz
10 m
0.5 Hz
50 Ω
Agilent U1251A/U1252A User’s and Service Guide 75
4 Changing The Default Setting
Table 4-1 Available setting options in Setup mode (continued)
t.CoUP
[3]
Thermocouple
tYPE K Sets the thermocouple type to K-type
tYPE J Sets the thermocouple type to J-type
Hand Enables manual data logging
d-LoG
Data logging
1–9999 s
[2]
Sets the interval for automatic data logging
Sets the temperature measurement to °C but
d-CF
pressing swaps the display to °F
d-F Sets the temperature measurement to °F
[4]
tEMP
Te m p e r a t u r e
d-FC
Sets the temperature measurement to °F but
pressing
swaps display to °C
d-C Sets the temperature measurement to °C
Notes for setting options in Setup mode:
1 This is the first option displayed once the user enters Setup mode. 2 For b-Lit, APF, rEF and d-LoG menu items, the user can select the digit to be adjusted by pressing 3 This menu option is only available for the U1252A. 4 To view the tEMP menu item, press
for more than 1 second.
tYPE K
Hand
d-CF
76 Agilent U1251A/U1252A User’s and Service Guide

Setting Data Hold/Refresh Hold Mode

1. Set OFF to enable the Data Hold mode (manual trigger by
key or bus via the remote control).
2. Set the variation count within the 100 to 1000 range to
enable the Refresh Hold mode (auto trigger). When the variation of the measuring value exceeds the setting of the variation count the Refresh Hold will be ready to trigger.
Press
Changing The Default Setting 4
Press
Press
Press
Press
Figure 4-1 Data hold/Refersh hold setup
Agilent U1251A/U1252A User’s and Service Guide 77
4 Changing The Default Setting

Setting Data Logging Mode

1 Set to “Hand” to enable the manual data logging mode.
2 Set the interval within 0001 to 9999 seconds to enable
interval (automatic) data logging mode.
3 Press and hold or for more than 1 second to switch
between manual and interval data logging setup.
Press
Press
Press
Press
Press
Figure 4-2 Data logging setup
78 Agilent U1251A/U1252A User’s and Service Guide
Press

Setting Thermocouple Types (U1252A only)

Press
The thermocouple sensor types that can be selected are
type- K (default) or type- J. Press or to switch
between the J and the K type.
Changing The Default Setting 4
Figure 4-3 Thermocouple type setup
Agilent U1251A/U1252A User’s and Service Guide 79
4 Changing The Default Setting

Setting Reference Impedance for dBm Measurement

The reference impedance can be set from 1 to 9999 W. The default value is 50 W.
Press
Press
Press
Press
Figure 4-4 Reference impedance for dBm measurement setup
80 Agilent U1251A/U1252A User’s and Service Guide
Loading...