Agilent Technologies U1251A User Manual

Agilent U1251A and U1252A Handheld Digital Multimeter
User’s and Service Guide
Agilent Technologies
Notices
CAUTION
WARNING
No p art o f this manu al may be re produce d in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
U1251-90003
Edition
Tenth Edition, May 4, 2012
Printed in Malaysia
Agilent Technologies, Inc. 3501 Stevens Creek Blvd. Santa Clara, CA 95052 USA
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus­tomers. Agilent provides this customary commercial license in Software and techni­cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com­puter Software Documentation).
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Accessories Warranty
Agilent offers warranty for product’s acces­sories for up to 3 months from the end-user acceptance date.
Standard Calibration Service (optional)
Agilent offers an optional calibration service contract for a period of 3 years from end-user acceptance date.
Safety Notices
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
II Agilent U1251A/U1252A User’s and Service Guide

Safety Symbols

CAT IV
600 V
The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
CAT III 1000 V
Category III 1000 V overvoltage protection Category IV 600 V overvoltage protection
Equipment protected throughout by double insulation or reinforced insulation
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
Earth (ground) terminal
Caution, risk of electric shock
Agilent U1251A/U1252A User’s and Service Guide III

Safety Information

This meter is safety-certified in compliance with EN/IEC 61010-1:2001, UL 61010-1 Second Edition and CAN/CSA
22.2 61010-1 Second Edition, CAT III 1000 V/ CAT IV 600 V Overvoltage Protection, Pollution Degree II. Use with standard or compatible test probes.

General Safety Information

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
IV Agilent U1251A/U1252A User’s and Service Guide
WARNING
• When working above 70 VDC, 33 VAC RMS or 46.7 V peak, exercise caution – such range pose a shock hazard.
• Do not measure more than the rated voltage (as marked on the meter) between terminals, or between terminal and earth ground.
• Double-check the meter’s operation by measuring a known voltage.
• For current measurement, turn off circuit power before connecting the meter to the circuit. Always place the meter in series with the circuit.
• When connecting probes, always connect the common test probe first. When disconnecting probes, always disconnect the live test probe first.
• Detach test probes from the meter before you open the battery cover.
• Do not use the meter with the battery cover or part of the cover removed or loose.
• Replace the battery as soon as the low battery indicator flashes on screen. This is to avoid false readings, which may lead to possible electric shock or personal injury.
• Do not operate the product in an explosive atmosphere or in the presence of flammable gases or fumes.
• Inspect the case for cracks or missing plastic. Pay extra attention to the insulation surrounding the connectors. Do not use the meter if it is damaged.
• Inspect the test probes for damaged insulation or exposed metal, and check for continuity. Do not use the test probe if it is damaged.
• Do not use any other AC charger adaptor apart from the one certified by Agilent with this product.
• Do not use repaired fuses or short-circuited fuse-holders. For continued protection against fire, replace the line fuses only with fuses of the same voltage and current rating and recommended type.
• Do not service or perform adjustments alone. Under certain condition, hazardous voltages may exist, even with the equipment switched off. To avoid dangerous electric shock, service personnel must not attempt internal service or adjustment unless another person, capable of rendering resuscitation or first aid, is present.
• Do not substitute parts or modify equipment to avoid the danger of introducing additional hazards. Return the product to Agilent Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
• Do not operate damaged equipment as the safety protection features built into this product may have been impaired, either through physical damage, excessive moisture, or any other reason. Remove power and do not use the product until safe operation can be verified by service-trained personnel. If necessary, return the product to Agilent Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
Agilent U1251A/U1252A User’s and Service Guide V
CAUTION
Turn off the circuit power and discharge all high-voltage capacitors in the circuit before
you perform resistance, continuity, diodes, or capacitance tests.
Use the correct terminals, function, and range for your measurements.
Never measure voltage when current measurement is selected.
Use only recommended rechargeable battery. Ensure proper insertion of battery in the
meter, and follow the correct polarity.
Disconnect test leads from all the terminals during battery charging.
VI Agilent U1251A/U1252A User’s and Service Guide

Regulatory Markings

The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
The C-tick mark is a registered trademark of the Spectrum Management Agency of Australia. This signifies compliance with the Australia EMC Framework regulations under the terms of the Radio Communication Act of 1992.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001. Cet appareil ISM est confomre a la norme NMB-001 du Canada.
The CSA mark is a registered trademark of the Canadian Standards Association.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.
Agilent U1251A/U1252A User’s and Service Guide VII

Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste
To return this unwanted instrument, contact your nearest Agilent Technologies, or visit: www.agilent.com/environment/product for more information.
Agilent Technologies, through Rechargeable Battery Recycling Corporation (RBRC), offers free and convenient battery recycling options in the U.S. and Canada. Contact RBRC at 877-2-RECYCLE (877.273.2925) or online at:
http://www.call2recycle.org/ for the nearest recycling location.
VIII Agilent U1251A/U1252A User’s and Service Guide

In This Guide…

1Getting Started
This chapter contains information on the Agilent U1251A and U1252A handheld multimeter front panel, rotary switch, keypad, display, terminals and rear panel.
2 Making Measurements
This chapter contains information on how to make measurements using the U1251A and U1252A handheld digital multimeter.
3 Functions and Features
This chapter contains information on the functions and features that are available for the U1251A and U1252A digital multimeter.
4 Changing The Default Setting
This chapter shows you how to change the default factory settings of the U1251A and U1252A and other available setting options
5 Maintenance
This chapter will go through how to troubleshoot the handheld digital multimeter if any problems arise.
6 Performance Tests and Calibration
This chapter contains the performance test and the adjustment procedure.
7 Specifications
This chapter lists the product characteristics, specification assumptions and the specifications of the U1251A and U1252A digital multimeters.
Agilent U1251A/U1252A User’s and Service Guide IX

Declaration of Conformity (DoC)

NOTE
The Declaration of Conformity (DoC) for this instrument is available on the Web site. You can search the DoC by its product model or description.
http://regulations.corporate.agilent.com/DoC/search.htm
If you are unable to search for the respective DoC, please contact your local Agilent representative.
X Agilent U1251A/U1252A User’s and Service Guide

Contents

1 Getting Started
Introducing the U12521A/U1252A Handheld Digital Multimeter 2
Check the shipment 3
Adjusting the tilt-stand 4
The rear panel at a glance 7 The keypad at a glance 9 The display at a glance 11 Selecting display with the Hz button 16 Selecting display with the Dual button 18 Selection display with the Shift button 20 The terminals at a glance 22
2 Making Measurements
Understanding The Measurement Instructions 24
III
Measuring Voltage 24
Measuring AC voltage 25 Measuring DC voltage 26
Measuring Current 27
µA & mA measurement 27 Percentage scale of 4 mA to 20 mA 29 A (ampere) measurement 31
Frequency Counter 32
Measuring Resistance, Conductance and Testing Continuity 34
Tes ti ng Dio de s 38
Measuring Capacitance 41
Measuring Temperature 43
Alerts and Warning During Measurement 47
Abbreviated Book Title Variable (Edit this) XIII
III
Overload alert 47 Input warning 47 Charge terminal alert 48
3 Functions and Features
Dynamic Recording 50
Data Hold (Trigger Hold) 52
Refresh Hold 53
Null (Relative) 55
Decibel Display 57
1 ms Peak Hold 59
Data Logging 61
Manual logging 61 Interval logging 63 Reviewing logged data 65
Square Wave Output (for U1252A) 67
Remote Communication 71
4 Changing The Default Setting
Selecting Setup Mode 74
Setting Data Hold/Refresh Hold Mode 77
Setting Data Logging Mode 78
Setting Thermocouple Types (U1252A only) 79
Setting Reference Impedance for dBm Measurement 80 Setting Minimum Frequency Measurement 81
Setting Temperature Unit 82
Setting Auto Power Saving Mode 84
Setting Percentage (%) Scale Readout 86
XIV Abbreviated Book Title Variable (Edit this)
Setting Beep Frequency 87 Setting Backlight Timer 88 Setting Baud Rate 89
Setting Parity Check 90
Setting Data Bit 91
Setting Echo Mode 92
Setting Print Mode 93
Returning to Default Factory Settings 94
5 Maintenance
Introduction 96
General maintenance 96 Battery replacement 96 Storage considerations 98 Charging the battery 98 Fuse checking procedure 105 Replacing the fuse 106 Troubleshooting 108
III
Replaceable Parts 109
To order replaceable parts 109
6 Performance Tests and Calibration
Calibration Overview 112
Closed-case electronic calibration 112 Agilent Technologies calibration services 112 Calibration interval 113 Adjustment is recommended 113
Recommended Test Equipment 114
Basic Operating Test 115
Backlit test 115
Abbreviated Book Title Variable (Edit this) XV
III
Testing the display 115 Current terminal test 116 Charge terminal alert test 117
Test Considerations 118
Calibration Security 119
Performance Verification Tests 120
Unsecuring the instrument for calibration 128
Calibration process 131
Using the front panel for adjustments 132
Adjustments Consideration 133
Valid adjustment input values 134 Adjustment procedure 135 Finishing the adjustment 142 To read the calibration count 142 Calibration errors 143
7 Specifications
Product Characteristics 146
Measurement Category 149
Measurement category definition 149
Specification Assumptions 150
Electrical Specifications 151
DC Specifications 151 AC Specifications 153 AC+DC Specifications for U1252A 155 Capacitance Specifications 156 Temperature Specifications 156 Frequency Specifications 157 Duty Cycle and Pulse Width Specifications 157 Frequency Sensitivity Specifications 158
XVI Abbreviated Book Title Variable (Edit this)
Peak Hold Specifications 159 Frequency Counter Specifications for U1252A 160 Square Wave Output for U1252A 161
Operating Specifications 162
Display update rate (approximate) 162 Input impedance 163
III
Abbreviated Book Title Variable (Edit this) XVII
III
XVIII Abbreviated Book Title Variable (Edit this)

List of Figures

Figure 1-1 Tilt-stand at 60° 4 Figure 1-2 Tilt-stand at 30° 4 Figure 1-3 Tilt-stand at hanging position 5 Figure 1-4 U1252A front panel 6 Figure 1-5 Rear panel 7 Figure 1-6 Rotary switch 8 Figure 1-7 U1252A keypad 9 Figure 1-8 Display symbols 11 Figure 1-9 U1252A connector terminals 22 Figure 2-1 Measuring AC voltage 25 Figure 2-2 Measuring DC voltage 26 Figure 2-3 Measuring mA and mA current 28 Figure 2-4 Measuring scale of 4-20 mA 30 Figure 2-5 A (ampere) current measurement 31 Figure 2-6 Measuring frequency 33 Figure 2-7 Measuring resistance 34 Figure 2-8 Audible continuity, conductance, and resistance
test. 35
Figure 2-9 Conductance measurement 37 Figure 2-10 Measuring forward bias of diode 39 Figure 2-11 Measuring reverse bias of diode 40 Figure 2-12 Capacitance measurements 42 Figure 2-13 Connecting the thermal probe into the non-compensa-
tion transfer adapter 44
Figure 2-14 Connecting the probe with adapter into the
multimeter 44
Figure 2-15 Surface temperature measurement 46 Figure 2-16 Input terminal warning 47 Figure 2-17 Charge terminal alert 48 Figure 3-1 Dynamic recording mode operation 51 Figure 3-2 Data hold mode operation 52 Figure 3-3 Refresh hold mode operation 54 Figure 3-4 Null (relative) mode operation 56 Figure 3-5 dBm/dBV display mode operation 58 Figure 3-6 1 ms peak hold mode operation 60 Figure 3-7
Hand (Manual) logging mode operation 62
Abbreviated Book Title Variable (Edit this) IX
Figure 3-8 Full Log 62 Figure 3-9 Interval (Automatic) logging mode operation 64 Figure 3-10 Log review mode operation 66 Figure 3-11 Frequency adjustment for square wave output 68 Figure 3-12 Pulse width adjustment for square wave output 69 Figure 3-13 Duty cycle adjustment for Square Wave 70 Figure 3-14 Cable connection for remote communication 71 Figure 4-1 Data hold/Refersh hold setup 77 Figure 4-2 Data logging setup 78 Figure 4-3 Thermocouple type setup 79 Figure 4-4 Reference impedance for dBm measurement setup 80 Figure 4-5 Minimum frequency setup 81 Figure 4-6 Temperature unit setup 83 Figure 4-7 Auto power saving setup 85 Figure 4-8 % scale readout setup 86 Figure 4-9 Beep frequency setup 87 Figure 4-10 Backlit timer setup 88 Figure 4-11 Baud rate setup remote control 89 Figure 4-12 Parity check setup 90 Figure 4-13 Data bit setup for remote control 91 Figure 4-14 Echo mode setup for remote control 92 Figure 4-15 Print mode setup for remote control 93 Figure 4-16 Reset setup 94 Figure 5-1 9 V rectangular battery 97 Figure 5-2 Battery capacity display as trickle 100 Figure 5-3 Self-test 100 Figure 5-4 Charging mode 102 Figure 5-5 Charge end and trickle state 103 Figure 5-6 Battery charging procedure 104 Figure 5-7 Fuse replacement 107 Figure 6-1 LCD display 115 Figure 6-2 Input warning 116 Figure 6-3 Charge terminal alert 117
X Abbreviated Book Title Variable (Edit this)

List of Tables

Ta b l e 1 - 1 Rotary switch description and functions 8 Ta b l e 1 - 2 Keypad desciption and functions 9 Ta b l e 1 - 3 General display symbols 12 Ta b l e 1 - 4 Priamry display symbols 13 Ta b l e 1 - 5 Secondary display symbols 14 Ta b l e 1 - 6 Analog bar range and counts 15 Ta b l e 1 - 7 Selecting display with the Hz button 16 Ta b l e 1 - 8 Selecting display with the Dual button 18 Ta b l e 1 - 9 Selecting display with the Shift button 20 Ta b l e 1 - 1 0 Terminal connections for different measuring
functions 22
Ta b l e 2 - 1 Numerical steps descriptions 24 Ta b l e 2 - 2 Percentage scale and measurement range 29 Ta b l e 2 - 3 Audible continuity measurement range 36 Ta b l e 4 - 1 Available setting options in Setup mode 75 Ta b l e 5 - 1 Battery voltage and corresponding percentage of charg-
es in standby and charging modes 100
Ta b l e 5 - 2 Error messages 101 Ta b l e 5 - 3 Measurement readings for fuse checking 106 Ta b l e 5 - 4 Fuse specifications 107 Ta b l e 5 - 5 Basic troubleshooting procedures 108 Ta b l e 6 - 1 Recommended Test Equipment 114 Ta b l e 6 - 2 Performance verification test steps 121 Ta b l e 6 - 3 Valid adjustment input values 134 Ta b l e 6 - 4 Adjustment table 137 Ta b l e 7 - 1 DC Accuracy ± (% of reading + Number of Least Signifi-
cant Digit) 151
Ta b l e 7 - 2 U1251A accuracy specifications ± (% of reading + num-
ber of LSD) for true RMS AC voltage 153
Ta b l e 7 - 3 U1251A accuracy specifications ± (% of reading + num-
ber of LSD) for true RMS AC current 153
Ta b l e 7 - 4 U1252A accuracy specifications ± (% of reading + num-
ber of LSD) for true RMS AC voltage 154
Ta b l e 7 - 5 U1252A accuracy specifications ± (% of reading + num-
ber of LSD) for true RMS AC current 154
Ta b l e 7 - 6 U1252A true RMS ac+dc voltage specifications 155
II
Abbreviated Book Title Variable (Edit this) XI
II
Ta b l e 7 - 7 U1252A true RMS ac+dc current specifications 155 Ta b l e 7 - 8 Capacitance specifications 156 Ta b l e 7 - 9 Temperature specifications 156 Ta b l e 7 - 1 0 Frequency specifications 157 Ta b l e 7 - 1 1 Duty cycle and pulse width specifications 157 Ta b l e 7 - 1 2 Frequency sensitivity and trigger level specifications for
voltage measurements 158
Ta b l e 7 - 1 3 Frequency sensitivity specifications for current
measurements 159
Ta b l e 7 - 1 4 Peak hold specifications for dc voltage and current
measurements 159
Ta b l e 7 - 1 5 Frequency counter (divide by 1) specifications 160 Ta b l e 7 - 1 6 Frequency counter (divide by 100
specifications 160
Ta b l e 7 - 1 7 Square wave output specifications 161 Ta b l e 7 - 1 8 Display update rate (approximate) 162 Ta b l e 7 - 1 9 Input Impedance 163
[4]
)
XII Abbreviated Book Title Variable (Edit this)
Agilent U1251A and U1252A Handheld Digital Multimeter User’s and Service Guide
1 Getting Started
Introducing the U12521A/U1252A Handheld Digital Multimeter 2 Check the shipment 3 Adjusting the tilt-stand 4 The front panel at a glance 6 The rear panel at a glance 7 The rotary switch at a glance 8 The keypad at a glance 9 The display at a glance 11
Selecting display with the Hz button 16 Selecting display with the Dual button 18 Selection display with the Shift button 20
The terminals at a glance 22
This chapter contains information on the Agilent U1251A and U1252A handheld multimeter front panel, rotary switch, keypad, display, terminals and rear panel.
Agilent Technologies
1
1 Getting Started Tutorial

Introducing the U12521A/U1252A Handheld Digital Multimeter

Key features of this digital multimeter:
DC, AC, and AC + DC (U1252A only) voltage and current
measurements.
True- RMS measurement for both AC voltage and current
Rechargeable Ni- MH battery with built- in charging
capability (U1252A only)
Ambient temperature on second display
Battery capacity indicator
Blue LED backlight
Resistance measurement of up to 50 M (for U1251A)
and 500 M (for U1252A)
Conductance measurement from 0.01 nS (100 GΩ)
Capacitance measurement of up to 100 mF
Frequency counter up of to 20 MHz (U1252A only)
The % scale readout for 4–20 mA or 0–20 mA measurement
dBm with selectable reference impedance
1 ms Peak Hold to catch inrush voltage and current easily
Temperature test with selectable 0
(without ambient temperature compensation).
K- type (for U1251A) and J/K-types temperature
measurement (for U1252A)
Frequency, duty cycle, and pulse width measurements
Dynamic Recording for min, max, and average readings
Data Hold with manual or auto trigger and Null mode
Diode and audible continuity tests
Square wave generator with selectable frequency, pulse
width and duty cycle (U1252A only)
Agilent GUI Application Software (IR- USB cable sold
separately)
Closed case calibration
°C compensation
~ 50 nS
2 Agilent U1251A/U1252A User’s and Service Guide

Check the shipment

Getting Started Tutorial 1
Verify that you have received the following items with your multimeter:
9 V alkaline battery (for U1251A only)
Soft carrying case
4 mm probes
Test leads
Alligator clips
SMT grabbers
Fine tips test probes
Mini grabber (black only)
Rechargeable 7.2 V battery (for U1252A only)
Power cord & AC adapter (for U1252A only)
Quick Start Guide
CD containing the User’s Guide, application software and
instrument drivers
Certificate of Calibration
Contact your nearest Agilent Sales Office if any of the above are missing.
Inspect the shipping containing for damage. Signs of damage may include a dented or torn shipping container or cushioning material that indicates signs of unusual stress or compacting. Save the packaging material in case the multimeter needs to be returned.
Please refer to the Agilent Handheld Tools brochure (5989- 7340EN) for the full and latest list of handheld accessories available.
Agilent U1251A/U1252A User’s and Service Guide 3
1 Getting Started Tutorial

Adjusting the tilt-stand

To adjust the meter to a 60° standing position, pull the tilt- stand outwards to its maximum reach.
IR-USB cable
To P C (h ost )
Tilt-stand at 60 °
Figure 1-1 Tilt-stand at 60
To adjust the meter to a 30° standing position, bend the tip of the stand so that it is parallel to the ground before pulling the stand outwards to its maximum reach.
Figure 1-2 Tilt-stand at 30
4 Agilent U1251A/U1252A User’s and Service Guide
°
IR-USB cable
To P C (h ost )
Tilt-stand at 30 °
°
Getting Started Tutorial 1
3. Flip the tilt-stand over until
this side of the
stand is facing the device instead of
facing you.
2. Detach the tilt-stand
1. Extend the tilt-stand until its
maximum reach
4. Re-attach the tilt-stand to an
upright position
To adjust the meter to a hanging position follow the steps shown in Figure 1- 3 below.
Agilent U1251A/U1252A User’s and Service Guide 5
Figure 1-3 Tilt-stand at hanging position
1 Getting Started Tutorial

The front panel at a glance

Display
Keypad
Rotary switch
Te r m i n a l Connectors
Figure 1-4 U1252A front panel
6 Agilent U1251A/U1252A User’s and Service Guide

The rear panel at a glance

Getting Started Tutorial 1
IR communication port
Tes t pr obe holders
Tilt-stand
Battery access cover
Figure 1-5 Rear panel
Agilent U1251A/U1252A User’s and Service Guide 7
1 Getting Started Tutorial
1
2
3
4
5
6 7 8
9
10

The rotary switch at a glance

Figure 1-6 Rotary switch
Tab l e 1 - 1 Rotary switch description and functions
No. Description/Function
1Charge mode (U1252A only) or OFF
2 AC voltage
3 AC voltage, DC voltage or DC+AC voltage (U1252A only)
4 DC mV, AC mV, AC+DC mV (U1252A only)
5 Resistance (), Continuity, and Conductance ( nS )
6 Frequency counter (U1252A only) or Diode
7 Capacitance or Temperature
8 DC µA, AC µA, and AC+DC µA
8 Agilent U1251A/U1252A User’s and Service Guide
9 DC mA, DC current, AC mA, AC current, or AC+DC current
10 Square-wave output, Duty cycle, or Pulse width output (for
U1252A) and OFF (for U1251A)

The keypad at a glance

1
2
3
4
5
6
7
The operation of each key is shown below. Pressing a key illuminates a related symbol on the display and emits a beep from the multimeter. Turning the rotary switch to another position resets the current operation of the key.
Figure 1- 7 below shows the keypad of the U1252A. The ms%
(Pulse width/Duty cycle), Hz , and frequency counter functions are only available on the U1252A.
Getting Started Tutorial 1
Figure 1-7 U1252A keypad
Tab l e 1 - 2 Keypad desciption and functions
Button Function when pressed for less than 1 second Function when pressed for more than 1 second
1 acts as a toggle switch to turn backlit ON/OFF.
Backlit automatically turns off after 30s (default)
2 freezes the measured value. In Data Hold
mode, press again to trigger hold of the next measured value. In Refresh Hold mode, reading updates automatically once reading is stable and count setting is exceeded
3 saves displayed value as a reference to be
subtracted from subsequent measurements. Press again to see the relative value that has been saved.
[1]
.
displays the battery capacity for 3 seconds.
[1]
.
enters Dynamic Recording mode. Press again to scroll through Max, Min, Avg, and present readings (indicated by MAXMINAVG on display).
enters 1 ms Peak Hold mode. Press
to scroll through Max and Min peak readings.
Agilent U1251A/U1252A User’s and Service Guide 9
1 Getting Started Tutorial
Button Function when pressed for less than 1 second Function when pressed for more than 1 second
4 scrolls through the measuring function(s) at a
particular rotary switch position.
enters Log Review mode. Press to switch to manual or interval logging data. Press or to view first or last logged data respectively. Press or to scroll up or down logged data.
Press for more than 1 second to exit mode.
5 scrolls through available measuring ranges
(except when rotary switch is set at TEMP or at Hz [for U1252A] position)
[2]
.
6 scrolls through available dual-combination
displays (except when rotary switch is set at TEMP, Hz or [for U1252A] position, or when meter is in 1 ms peak hold or dynamic recording mode)
[3]
.
7 enters Frequency Test mode for current or
voltage measurements. Press to scroll through frequency (Hz), duty cycle (%) and pulse width (ms) functions. In duty cycle (%) and pulse width (ms) tests, press to switch to positive or negative
sets to Auto Range mode.
exits Hold, Null, Dynamic Recording, 1 ms
Peak Hold and dual display modes.
enters logging mode. In manual data logging, press to log data manually into memory. In automatic data logging, data logs automatically
[1]
. Press for more than 1 second to exit auto data logging mode.
pulse.
Notes for keypad descriptions and functions: 1 See Ta b l e 4 - 1 on page 75, “Available setting options in Setup mode,” for details of available options.
2 When rotary switch is at TEMP, press to switch to ºC or ºF display. When rotary switch is at Hz, press to
switch to division of signal frequency by 1 or 100.
3 When rotary switch is at TEMP, ETC is ON by default. You may press to disable ETC (Environment Temperature
Compensation), will appear on display. For pulse and duty cycle measurement, press to switch trigger slope to positive or negative. When meter is in peak or dynamic-recording mode, press to restart 1 ms peak hold or dynamic recording mode.
10 Agilent U1251A/U1252A User’s and Service Guide
Loading...
+ 155 hidden pages