Agilent Technologies SpectrAA User Manual

Practical Use of SpectrAA Series for Multielement Analysis
Author
Brian D. Frary
Application Note
Atomic Absorption
Introduction
When operating in the furnace mode the system can be left unattended as there are no flammable gases and no exposed ignition source hence it can be left to operate overnight.
A scenario is described for analysis by flame and shows how even a very complex analytical program is easily handled by the SpectrAA.
The components of the system are:
SpectrAA-30 or SpectrAA-40 Atomic Absorption Spectrophotometer PSC-56 Programmable Sample Changer GTA-96 Graphite Tube Atomizer VGA-76 Vapor Generator Accessory DS-15 Data Station
The above units except for the DS-15, are only fitted with an on/off power switch. Programming and control of each unit in the system is through the DS-15 keyboard but no computer programming knowledge is required by the operator. In fact the system is designed for one finger operation using the “fill in the form” (F.I.T.F) method developed by Agilent Technologies.
2
required (from 0.1% to 99.9%) and set a default time (1 to 500 seconds) for the system to achieve this. If the system is unable to achieve the precision within the default time speci­fied, measurement will stop and the precision achieved printed out with the result. (The smaller the absorbance signal the longer the measurement time required to obtain a given precision).
This facility is extremely useful when high precision is required but the element concentration (and thus absorbance) is expected to vary widely. The operator can set the default time for the lowest expected absorbance knowing that higher absorbances will be read in a shorter time with no loss of precision or waste of analysis time.
The concentrations of lead are expected to be low thus the air-acetylene burner will be used to obtain higher sensitivity. (NOTE When high sensitivity is not required a burner change can be avoided by using a nitrous oxide-acetylene burner for an air-acetylene flame. This eliminates programming a “Pause” to change burners). PROMT mode could be used for measurement but in this case the mean of three replicate readings of 7-second integration is sufficient for the analytical requirements.
Chromium forms ferro-chrome compounds in the air-acety­lene flame and will therefore be determined using the nitrous oxide-acetylene flame. This flame will also be used to deter­mine molybdenum and silicon. All three elements will be determined using five second integration and three replicates.
A delay of five seconds is programmed when using continu­ous aspiration so that each solution will be aspirated for five seconds before any readings are taken. This ensures that the previous solution is washed out of the nebulizer/burner system. Cross contamination is prevented and the determina­tions will be more accurate. In addition a RINSE RATE and RINSE TIME may be programmed if required (where sample concentrations cover a wide range).
RECALIBRATION RATE and/or RESLOPE RATE are selected on previous experience with the samples and elements. Typical values could be reslope 10, recalibration 20.
Up to five standards may be used for each element calibra­tion. 5 mixed standards will be loaded into the sampler carousel, but the number used will vary from element to element (see Table 1).
Versatility and ease of use has been achieved in the system design. The user friendly programming is designed to lead the operator through the correct procedures.
At power up the DS-15 screen displays 3 modes of operation. The operator selects one mode using the soft keys labelled:
DEVELOP METHOD MODIFY METHOD AUTOMATIC RUN
DEVELOP METHOD used by a chemist to set up and store methods for the use of other staff.
MODIFY METHOD used in a similar manner.
AUTOMATIC RUN used for routine automatic analysis.
Flame Analytical Procedure
The following scenario shows how aII the facilities of the system may be used to perform a complex analytical program.
A laboratory has the following samples for analysis:
20 steel samples for the determination of chromium, copper, lead, manganese, molybdenum, nickel and silicon.
10 water samples for the determination of iron, sodium and arsenic.
10 slag samples for the determination of calcium, iron and silicon.
Steel Samples
Analytical method development has shown that most of the elements in the steel samples can be determined by normal calibration using mixed standards containing iron. The SpectrAA- will be calibrated in the concentration mode using the integration measurement mode for these elements.
Chromium and nickel concentrations must be determined accurately over a narrow range and bracketing standards cali­bration will be used for these determinations. High precision of measurement is also required thus PROMT mode will be used.
PROMT is Precision Optimised Measurement Time. This allows the operator to specify the precision of measurement
3
Water Samples
EPA methods specify standard additions calibration for water analysis. Iron will be determined by this method using the air­acetylene flame burning on a nitrous oxide-acetylene burner. Continuous aspiration would normally be used for water sam­ples but in this case the program “Fe in Slag” will be used which uses the micro-sampling technique. Thus two different sample types can be run on one program, the recalibration rate being defined by the number of water samples.
Arsenic will also be determined by standard additions calibra­tion but using hydride generation. Sodium will be determined by EMISSION using normal calibration.
Two addition standards will be prepared from sample 1 and used to produce the calibration graph. The other 9 samples will be measured against this calibration because the matri­ces will be very similar.
Slag Samples
As the matrices will be complex and variable all elements will be determined using STANDARD ADDITIONS calibration. The dissolved solids concentration will be very high therefore con­tinuous aspiration will not be used. Instead, the MICRO-SAM­PLING feature of the PSC-56 will be used and measurements made in peak height mode.
In micro-sampling the sampling probe dips into the solution for a programmed time. During this time a certain volume of liquid is drawn into the flame, the longer the time the greater the volume (for a one second dip about 350 µL). An integra­tion “window” of eight seconds is set on the system. This allows enough time for the solution to travel through the neb­ulizer/burner system into the optical path to be measured. A volume of 200 µL or more will give the same signal as contin­uous aspiration, 100 µL will give about 80% and 50 µL about 50% of the continuous aspiration signal. Micro-sampling is also useful when only small volumes of sample are available [1,2,3,4].
A RINSE between each sample and a SAMPLE FLUSH will also be programmed. Using sample flush the probe dips into a solution for half a second, the slug of liquid is drawn through the capillary “flushing out” any traces of the previous solution aspirated: no measurement is made on this solution. After about 10 seconds the probe dips into the solution for the specified dip time and the signal produced is measured.
Iron will be determined using the air-acetylene flame, silicon and calcium using the nitrous oxide-acetylene flame, both flames being run on the nitrous oxide-acetylene burner.
Typical system parameters for some elements are shown in Figures 1 to 6.
Table 1.
Although RECAL will not be used for program 1, it is included here as it is part of the stored program.
AA = Air-Acetylene NA = Nitrous Oxide-Acetyleen
Loading...
+ 7 hidden pages