Agilent Technologies SpectrAA User Manual

Practical Use of SpectrAA Series for Multielement Analysis
Author
Brian D. Frary
Application Note
Atomic Absorption
Introduction
When operating in the furnace mode the system can be left unattended as there are no flammable gases and no exposed ignition source hence it can be left to operate overnight.
A scenario is described for analysis by flame and shows how even a very complex analytical program is easily handled by the SpectrAA.
The components of the system are:
SpectrAA-30 or SpectrAA-40 Atomic Absorption Spectrophotometer PSC-56 Programmable Sample Changer GTA-96 Graphite Tube Atomizer VGA-76 Vapor Generator Accessory DS-15 Data Station
The above units except for the DS-15, are only fitted with an on/off power switch. Programming and control of each unit in the system is through the DS-15 keyboard but no computer programming knowledge is required by the operator. In fact the system is designed for one finger operation using the “fill in the form” (F.I.T.F) method developed by Agilent Technologies.
2
required (from 0.1% to 99.9%) and set a default time (1 to 500 seconds) for the system to achieve this. If the system is unable to achieve the precision within the default time speci­fied, measurement will stop and the precision achieved printed out with the result. (The smaller the absorbance signal the longer the measurement time required to obtain a given precision).
This facility is extremely useful when high precision is required but the element concentration (and thus absorbance) is expected to vary widely. The operator can set the default time for the lowest expected absorbance knowing that higher absorbances will be read in a shorter time with no loss of precision or waste of analysis time.
The concentrations of lead are expected to be low thus the air-acetylene burner will be used to obtain higher sensitivity. (NOTE When high sensitivity is not required a burner change can be avoided by using a nitrous oxide-acetylene burner for an air-acetylene flame. This eliminates programming a “Pause” to change burners). PROMT mode could be used for measurement but in this case the mean of three replicate readings of 7-second integration is sufficient for the analytical requirements.
Chromium forms ferro-chrome compounds in the air-acety­lene flame and will therefore be determined using the nitrous oxide-acetylene flame. This flame will also be used to deter­mine molybdenum and silicon. All three elements will be determined using five second integration and three replicates.
A delay of five seconds is programmed when using continu­ous aspiration so that each solution will be aspirated for five seconds before any readings are taken. This ensures that the previous solution is washed out of the nebulizer/burner system. Cross contamination is prevented and the determina­tions will be more accurate. In addition a RINSE RATE and RINSE TIME may be programmed if required (where sample concentrations cover a wide range).
RECALIBRATION RATE and/or RESLOPE RATE are selected on previous experience with the samples and elements. Typical values could be reslope 10, recalibration 20.
Up to five standards may be used for each element calibra­tion. 5 mixed standards will be loaded into the sampler carousel, but the number used will vary from element to element (see Table 1).
Versatility and ease of use has been achieved in the system design. The user friendly programming is designed to lead the operator through the correct procedures.
At power up the DS-15 screen displays 3 modes of operation. The operator selects one mode using the soft keys labelled:
DEVELOP METHOD MODIFY METHOD AUTOMATIC RUN
DEVELOP METHOD used by a chemist to set up and store methods for the use of other staff.
MODIFY METHOD used in a similar manner.
AUTOMATIC RUN used for routine automatic analysis.
Flame Analytical Procedure
The following scenario shows how aII the facilities of the system may be used to perform a complex analytical program.
A laboratory has the following samples for analysis:
20 steel samples for the determination of chromium, copper, lead, manganese, molybdenum, nickel and silicon.
10 water samples for the determination of iron, sodium and arsenic.
10 slag samples for the determination of calcium, iron and silicon.
Steel Samples
Analytical method development has shown that most of the elements in the steel samples can be determined by normal calibration using mixed standards containing iron. The SpectrAA- will be calibrated in the concentration mode using the integration measurement mode for these elements.
Chromium and nickel concentrations must be determined accurately over a narrow range and bracketing standards cali­bration will be used for these determinations. High precision of measurement is also required thus PROMT mode will be used.
PROMT is Precision Optimised Measurement Time. This allows the operator to specify the precision of measurement
3
Water Samples
EPA methods specify standard additions calibration for water analysis. Iron will be determined by this method using the air­acetylene flame burning on a nitrous oxide-acetylene burner. Continuous aspiration would normally be used for water sam­ples but in this case the program “Fe in Slag” will be used which uses the micro-sampling technique. Thus two different sample types can be run on one program, the recalibration rate being defined by the number of water samples.
Arsenic will also be determined by standard additions calibra­tion but using hydride generation. Sodium will be determined by EMISSION using normal calibration.
Two addition standards will be prepared from sample 1 and used to produce the calibration graph. The other 9 samples will be measured against this calibration because the matri­ces will be very similar.
Slag Samples
As the matrices will be complex and variable all elements will be determined using STANDARD ADDITIONS calibration. The dissolved solids concentration will be very high therefore con­tinuous aspiration will not be used. Instead, the MICRO-SAM­PLING feature of the PSC-56 will be used and measurements made in peak height mode.
In micro-sampling the sampling probe dips into the solution for a programmed time. During this time a certain volume of liquid is drawn into the flame, the longer the time the greater the volume (for a one second dip about 350 µL). An integra­tion “window” of eight seconds is set on the system. This allows enough time for the solution to travel through the neb­ulizer/burner system into the optical path to be measured. A volume of 200 µL or more will give the same signal as contin­uous aspiration, 100 µL will give about 80% and 50 µL about 50% of the continuous aspiration signal. Micro-sampling is also useful when only small volumes of sample are available [1,2,3,4].
A RINSE between each sample and a SAMPLE FLUSH will also be programmed. Using sample flush the probe dips into a solution for half a second, the slug of liquid is drawn through the capillary “flushing out” any traces of the previous solution aspirated: no measurement is made on this solution. After about 10 seconds the probe dips into the solution for the specified dip time and the signal produced is measured.
Iron will be determined using the air-acetylene flame, silicon and calcium using the nitrous oxide-acetylene flame, both flames being run on the nitrous oxide-acetylene burner.
Typical system parameters for some elements are shown in Figures 1 to 6.
Table 1.
Although RECAL will not be used for program 1, it is included here as it is part of the stored program.
AA = Air-Acetylene NA = Nitrous Oxide-Acetyleen
4
Figure 1.
Figure 2.
Figure 3.
Figure 4.
Figure 5.
Figure 6.
5
A DELAY of two minutes is programmed at the start of the first program (15 iron) to ensure the burner is stabilized at its operating temperature. FIRST and LAST sample numbers are then entered, in this example 1 and 20.
The FIRST MEASUREMENT can be a calibration, a reslope or sample.
If CALIBRATION is selected the system will calibrate with the number of standards specified. If RESLOPE is selected this will be applied to the stored calibration for that element. If SAMPLE is selected the concentration values will be calcu­lated from the stored calibration data. If there is no stored cal­ibration then only the absorbance is measured. (Only copper will be determined using a previously stored calibration and thus the first measurement will be reslope.)
For the next program (17, calcium) a delay of 1 minute has been entered to allow the burner to reach its operating tem­perature with the nitrous oxide-acetylene flame. First and last sample numbers are entered as before. This procedure is fol­lowed for programs 18 (silicon), 13 (chromium), 19 (silicon) and 23 (molybdenum) but without the delay time.
A delay of three minutes is entered for program 9 (sodium) which will allow the burner to cool down to its operating temperature; first and last sample numbers being entered as before.
Sample numbers are entered for program 1 (copper) and first measurement is altered to RESLOPE. Sample numbers are entered for programs 7 (nickel) and 8 (manganese).
A delay of two minutes is entered for program 2 (lead) to allow the burner to reach its operating temperature as it will be changed before this determination. The relevant sample numbers are then entered for program 2 and 10 (arsenic).
The final entry to be made is PAUSE AFTER PROGRAM which appears at the bottom of the screen. This PAUSE facility can be used several times in a sequence, enabling carousels, standards and reslope standards to be replaced; burners to be exchanged, rotated, or their height adjusted; the VGA-76 to be fitted. Lamps can also be exchanged if nine or more elements are being determined and multi-element lamps are not being used.
The first PAUSE entry will be 18 which will enable the carousels to be exchanged. When the system automatically pauses after the completion of program 18, the operator can insert the next PAUSE request, in our example 8, to change the burner. Finally when this second pause command is acted upon during operation the operator can insert the last PAUSE request, in our example 2, to enable the VGA-76 to be installed.
Programming the Automatic Run
It will be useful to draw up a table (Table 1) when planning the analysis. This can then be used when compiling the sequence control page on the DS-15 (Figure 8).
Elements requiring standard additions calibration will be determined first.
AUTOMATlC RUN soft key is pressed and SEQUENCE SELEC­TION page (Figure 7) is displayed. The relevant programs are selected on this page and the soft key for SEQUENCE CON­TROL pressed. The programs selected are displayed by number and title on the left side of the screen (Figure 8).
Figure 7.
Figure 8.
6
Having completed the sequence control page the NOTES soft key is pressed then the REPORT FORMAT soft key. This screen enables the following to be displayed and printed on the report:
Operator name Date Batch name
and the report parameters selected:
Format Sequential* or Multi-element
Data printed Mean Concentration only* or
Mean Concentration and %RSDor Mean Concentration, %RSD & Mean Absorbance or
Each Reading (Mean Concentration, %RSD, Mean Absorbance and Every Absorbance Reading).
Print During Run* or After Run or Manual
Instrument status Yes or No*
Notes Yes or No*
Calibration results Yes* or No
Calibration graph Yes or No*
Sample labels Yes* or No
Lines per page Select for page length used (usually 66)
See Figure 9. Parameters normally used are indicated by *.
If PRINT DURING RUN is selected then all the other features selected will be printed during the run. All results obtained after the start of the run can be inspected on the print-out. However when using PRINT MANUAL or PRINT AFTER RUN, only the results currently displayed on the screen are available for inspection.
The analytical run can be completed in a slightly shorter time if PRINT MANUAL is selected.
The absorbance values of all solutions are stored on the pro­gram disc as they are read (no matter which PRINT status is selected). If a power failure occurs during the analysis the only data lost will be that for the solution under analysis. No other data is lost or corrupted.
At the end of the analytical run all the data may be archived and a multi-element report produced. It is recommended that the data is archived before another run is performed as the program disc is cleared of data when START is pressed.
If SAMPLE LABELS is selected then a screen as shown in Figure 10 is displayed (up to 12 alphanumeric characters can be used). INCREMENT LABELS is a time saving facility. In Figure 10, W113 is the first water sample and is entered as sample 1. INCREMENT LABELS is pressed and samples 2 to 10 are automatically labelled W114 to W122. The cursor is moved to sample 11, relabelled SL536 and INCREMENT LABELS pressed. Samples 12 to 20 are automatically labelled SL537 to SL545. Finally the cursor is moved to sample 21, relabelled S907 and INCREMENT LABELS pressed. Samples 22 to 42 are automatically labelled S908 to S928. MORE LABELS is pressed, samples 43 to 67 are displayed and are automatically labelled S929 to S953.
Loading Off Carousels
Two carousels will be used, one for STANDARD ADDITIONS calibration (water samples 1 to 10, slag samples 11 to 20), the other for NORMAL calibration (steel samples 21 to 50).
The standard additions carousel will be loaded as follows:
Figure 9.
Figure 10.
7
Sample position 1 Blank
2 Addition 1 3 Addition 2 4 Addition 0 (sample 1) 5 to 13 Addition 0 (samples 2 to 10) 14 Blank 15 Addition 1 16 Addition 2 17 Addition 0 (sample 11) 18 to 26 Addition 0 (samples 12 to 20)
The second carousel, for normal calibration will be loaded:
Multi-element standards in standard positions 1 to 5. Blank and reslope solutions in their respective positions. Water samples in sample positions 1 to 10 and steel samples in positions 21 to 50.
Operation
When the operator has fitted the required burner and carousel the flame is lit and START key pressed.
Operation proceeds as foIlows:
The system recalls the first program in the sequence (15) resets the sampler and checks all system parameters are cor­rectly set. It waits 2 minutes for the burner to warm up while aspirating distilled water. It then carries out the calibration of the system with the addition standards for waters and mea­sures the first 10 samples (waters). It then recalibrates the system with the addition standards for slags and measures samples 11 to 20 (slags).
The next program (17) is recalled and the system set up to these parameters. The burner is allowed to warm up for the programmed 1 minute before calibration proceeds. Samples 1 to 10 are then measured.
The next program (18) in the sequence is recalled and the system set up on these parameters. Calibration proceeds immediately as no delay has been programmed (the burner is already at its operating temperature). Samples 1 to 10 are measured for this element.
After sample 10 is measured the next program (13) is recalled and the system “beeps” to remind the operator that a pro­grammed PAUSE has been reached. The operator presses STOP, changes carousels, alters PAUSE AFTER PROGRAM to 8 and presses START. As no delay has been programmed the system proceeds to calibrate and then measures samples 31 to 40. A RECALIBRATION is then carried out and samples 41 to 50 are measured.
Program 19 is then recalled, the system set up, calibration is performed and samples 21 to 35 are measured. RESLOPE is performed and samples 36 to 45 measured.
Program 23 is recalled, the system set up, calibration is per­formed and samples 41 to 50 measured.
Program 9 is recalled, the system is set up and changes to the air-acetylene gas mixture waiting for the 3 minutes pro­grammed delay while the burner cools down to its operating temperature. Calibration is performed and samples 1 to 10 are measured.
Program 1 is recalled, the system set up, a RESLOPE is per­formed and applied to the stored calibration. If the reslope is correct, samples 21 to 36 are measured. A RESLOPE is
performed and samples 37 to 50 measured.
Program 7 is recalled, the system set up, calibration is per­formed and samples 31 to 40 measured. A RECALIBRATION is performed and samples 41 to 50 measured.
Program 8 is recalled, the system set up, calibration is per­formed and samples 27 to 45 measured. Program 2 is recalled and the system “beeps” to remind the operator that a pro­grammed PAUSE has been reached. The operator presses STOP, presses FLAME OFF, changes burners, alters PAUSE AFTER PROGRAM to 2 then presses IGNITE and START. The system waits 2 minutes for the burner to warm up then cali­brates and measures samples 31 to 40. A RECALIBRATION is then carried out and samples 41 to 50 are measured.
Program 10 is recalled and the system “beeps” to remind the operator that a programmed PAUSE has been reached. The operator presses STOP and FLAME OFF, connects the VGA-76 hydride generator, replaces the carousel in use by the one for standard addition calibration and presses IGNITE. START is pressed, the system performs the calibration procedure and determines the arsenic concentrations of samples 1 to 10. After the last sample is determined the flame is automatically extinguished and the system “beeps” indicating the analytical run has been completed.
Archiving and multi-element report
The INDEX key is pressed and page 19 UTILITIES selected. From the display ARCHIVE is selected followed by the file (on the utilities disc) for archiving. The ARCHIVE soft key is pressed and the data is permanently stored.
After returning to the UTILITIES index, PRINT REPORTS is selected. From the next display the data set required is chosen and then the elements required. REPORT FORMAT soft key is pressed and the following is displayed:
8
Format Sequential or multi-element*
Data printed Mean concentration only* or
Mean concentration and %RSDor Mean concentration, %RSD and mean absorbance or
Each Reading (Mean Concentration, %RSD, Mean Absorbance and Every Absorbance Reading).
Instrument status Yes or No*
Notes Yes or No*
Calibration results Yes* or No
Calibration graph Yes or No*
Sample labels Yes* or No
Weight correction Yes* or No
Lines per page Select for page length used (usually 66)
Parameters normally used are indicated by *
If WEIGHT CORRECTION is selected the measured concentra­tion of the elements in the samples is automatically calcu­lated before print out. This is a time (and error) saving facility.
PRINT REPORT is pressed and the multi-element report is duly obtained (Figure 11).
Figure 11.
9
As all the analytical data is now archived other reports can be generated at any time if required. For example a report show­ing all data for arsenic (program 25) is required. This is easily produced by:
Accessing the utilities disc, selecting the appropriate archive file, then selecting the element to be printed (in this case arsenic). Proceed to report format via the soft key and select the options required, then print report.
The SpectrAA- system can be used for an automatic run in the same way. However as it has a manually operated turret the operator must bring the next lamp into the operating position when prompted by the DS-15. If multi­element lamps are used then this requirement is reduced or even eliminated.
References
1. L. M. Voth, “Analysis of High Dissolved Solids Solutions by Flame Microsampling”, Varian Instruments At Work Number AA-13, February 1981.
2. L. M. Voth, “Analysis of Battery Acids by Flame Microsampling”, Varian Instruments At Work Number AA-14 April 1981.
3. L. M. Voth, “Determination of Calcium and Magnesium in Blood Serum by Automated Flame Microsampling”, Varian Instruments At Work Number AA-15, April 1981.
4. L. M. Voth, “Determination of Chromium, Lead and Cadmium in Drinking Water by Solvent Extraction and Flame Microsampling”, Varian Instruments At Work Number AA-16, August 1981.
For More Information
For more information on our products and services, visit our Web site at www.agilent.com/chem
www.agilent.com/chem
Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Information, descriptions, and specifications in this publication are subject to change without notice.
© Agilent Technologies, Inc., 1985 Printed in the USA November 1, 2010 AA048
Loading...