Udo Huber
Saponification of procaine: Kinetic
measurements with the Agilent high
throughput analysis system
Abstract
In this application note we describe how the cleavage of procaine, a
p-aminobenzoic acid ester, can be monitored using the Agilent 220
microplate sampler (MPS) with the Agilent 1100 Series LC system. The
data of the measurements is transferred to ChemStore C/S, the data-
base module of the Agilent ChemStation Plus, for data analysis. We
show that the data can then be transferred easily to a spreadsheet
program, for example Microsoft®Excel®, for further calculations such
as determination of the rate coefficient.
Application Note
Procaine is a p-aminobenzoic acid
ester, which can be saponificated
into p-aminobenzoic acid (PABA)
and an alcohol. The reaction is
shown in figure 1.
Since the reaction is first order
the rate of reaction can be
described as:
with:
v (rate of reaction)
k (rate coefficient)
[Ester] (concentration of
procaine)
Integration of this formula gives:
The rate coefficient k can be
determined from the slope of the
straight line in the graph
ln([Ester]
t
/[Ester]0) against time.
Introduction
Kinetic measurements play an
important role in pharmaceutical
chemistry. Not only for pharmacokinetics where the rate of active
compound degradation has to be
determined, but also for drug discovery to test the inhibition effect
of a compound on an enzyme. For
very fast reactions special apparatus, for example shock tubes,
have to be used but slower reactions can be monitored by analyzing reaction samples at specific
time intervals. This application
note describes how this is
achieved using the Agilent 220
MPS with the Agilent 1100 Series
LC System and the Agilent ChemStation Plus software. Saponification of procaine at pH=10 was
selected as a model scenario.
Figure 1
Saponification of procaine
v·-==
Esterd
][][Esterk
dt
][
ln
Ester
Ester
t
][
0
tk
·-=
O
H2N
OH
N
O
H2N
O
+
O
HO
N
Equipment
The system included an Agilent
1100 Series vacuum degasser, an
Agilent 1100 Series binary pump,
an Agilent 1100 Series thermostatted column compartment, an Agilent 1100 Series diode array detector and an Agilent 220 micro plate
sampler.
The system was controlled using
the Agilent ChemStation Plus (version A.07.01) and the micro plate
sampling software (version
A.03.01).
System Setup Overview
1. A chromatographic method for
measuring procaine and PABA
was developed on the Agilent
220 MPS and the Agilent 1100
Series LC system.
2. Standards for both compounds
were measured, the method
was calibrated and the run time
was extended to 20 minutes
(figure 2).
3. Three procaine samples were
dissolved in 0.025 M NaH2PO
4,
buffer adjusted to pH=10.
These samples were measured
with the method described
before, which gives an overall
run time of one hour for the
three samples.
4. The measurement was repeated
24 times to give an overall
study run time of 24 hours.
5. The measured data was automatically transferred to the
ChemStation Plus database
module were the Charts
amount against reaction time
was created.
6. To determne the rate coefficient
the data was then automatically
transferred to Microsoft Excel.
Time [min]
012345
Absorbance
[mAU]
0
50
100
150
200
250
300
PABA
Procaine
Mobile Phases: A= 0.025M NaH2PO4in
water (pH=2.5), B = ACN
Gradient: 5 % B for 3.5 min,
flow 1 ml/min
5 % B to 50 % B in 1.5 min,
flow 1 ml/min
50 % B for 0.5 min,
flow 1 ml/min
50 % B to 5 % B in 0.5 min,
flow 1 ml/min
5 % B, flow from 1 ml/min to
0.1 ml/min in 0.1 min
5 % B,
flow 0.1 ml/min for 18.9 min
5 % B, flow 0.1 ml/min to
1 ml/min in 0.1 min
5 % B for 0.9 min,
flow 1 ml/min
Stop time: 20 min
Column: Zorbax SB-C18, 4.6 x 75 mm,
5 µm
Column temp.: 50 ºC
UV detector: DAD 204 nm/16
(reference 360 nm/100)
Figure 2
Measurement of standards