No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Agilent Technologies, Inc. as governed by
United States and international copyright
laws.
Manual Part Number
70-9062
Edition
Rev A, March 2011
Printed in USA
Agilent Technologies, Inc.
3501 Stevens Creek Blvd.
Santa Clara, CA 95052 USA
Warranty
The material contained in this document is provided “as is,” and is
subject to being changed, without
notice, in future editions. Further,
to the maximum extent permitted
by applicable law, Agilent disclaims
all warranties, either express or
implied, with regard to this manual
and any information contained
herein, including but not limited to
the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or
performance of this document or of
any information contained herein.
Should Agilent and the user have a
separate written agreement with
warranty terms covering the material in this document that conflict
with these terms, the warranty
terms in the separate agreement
shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to
the federal government include only those
rights customarily provided to end user
customers. Agilent provides this customary commercial license in Software and
technical data pursuant to FAR 12.211
(Technical Data) and 12.212 (Computer
Software) and, for the Department of
Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS
227.7202-3 (Rights in Commercial Computer Software or Computer Software
Documentation).
Safety Notices
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in damage to the product or loss
of important data. Do not proceed
beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice,
or the like that, if not correctly
performed or adhered to, could
result in personal injury or
death. Do not proceed beyond a
WARNING notice until the
indicated conditions are fully
understood and met.
Rotating Bottle Apparatus Operator’s Manual
Contents
Figures5
1Safety 7
Electrical Hazards7
Warni ng9
Caution9
Note9
Information Symbols10
General11
WEEE Directive12
2Introduction13
Conventions Used in this Manual15
3Setting Up the Rotating Bottle Apparatus17
Unpacking Your Equipment18
Setting up the Water Bath and 75019
Connecting the Temperature Probe20
Initial Power Up20
Power Cord Connections20
Powering Up the 750 Heater / Circulator21
Powering Up the Rotating Bottle Apparatus21
Exclusions and Limitations34
Obtaining Warranty Service34
Warranty Limitations35
Exclusive Remedies35
Index37
4Rotating Bottle Apparatus Operator’s Manual
Figures
Figure 1. Rotating Bottle Apparatus14
Rotating Bottle Apparatus Operator’s Manual5
Figures
This page was intentionally left blank, except for this message.
6Rotating Bottle Apparatus Operator’s Manual
Rotating Bottle Apparatus
Operator’s Manual
1
Safety
Electrical Hazards 7
WEEE Directive 12
The Rotating Bottle Apparatus has been carefully designed so that when
used properly you have an accurate, fast, flexible, and safe instrument.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
Operation of a Rotating Bottle Apparatus involves the use of solid
dosage forms and aqueous liquids. Unskilled, improper, or careless use
of this instrument can create shock hazards, fire hazards, or other
hazards which can cause death, serious injury to personnel, or severe
damage to equipment and property.
Information on safety practices is provided with your instrument and
operation manuals. Before using your instrument or accessories, you
must thoroughly read these safety practices.
Electrical Hazards
Observe all relevant safety practices at all times.
The Rotating Bottle Apparatus contains electrical circuits, devices, and
components operating at dangerous voltages. Contact with these
circuits, devices, and components can cause death, serious injury, or
painful electric shock.
Panels or covers that are retained by fasteners which require the use of a
tool for removal may be opened only by Agilent-trained,
Agilent-qualified, or Agilent-authorized service engineers. Consult the
manuals or product labels supplied with the 8020 to determine which
parts are operator-accessible.
Agilent Technologies
7
1Safety
Application of the wrong supply voltage, connection of the instrument
to an incorrectly wired supply outlet, or lack of proper electrical
grounding can create a fire hazard or a potentially serious shock hazard
and could seriously damage the instrument and any attached ancillary
equipment.
Always use a three-wire outlet with ground connection which is
adequately rated for the load. The installation must comply with local,
state, and federal safety regulations.
Do not connect the instrument to the main power supply until you have
made sure that the operating voltage is correctly set for the main power
supply in the specific outlet in your laboratory to which the equipment
will be connected.
8Rotating Bottle Apparatus Operator’s Manual
Warning
WARNING
CAUTION
NOTE
Caution
Safety1
A ‘Warning’ message appears in the manual when failure to observe
instructions or precautions could result in death or injury.
Read all warnings and cautions carefully and observe them at all times.
A ‘Caution’ message appears in the manual when failure to observe
instructions could result in damage to equipment (Agilent supplied and /
or other associated equipment).
Rotating Bottle Apparatus Operator’s Manual9
Note
A ‘Note’ appears in the manual to give advice or information.
1Safety
Information Symbols
ISwitches main power on
0Switches main power off
Indicates single-phase alternating current
Indicates the product complies with the
requirements of one or more European Union
(EU) directives
Indicates specific equipment meets
consensus-based standards of safety to provide
assurance, required by OSHA, that these
products are safe for use in the workplace for
North America
Indicates that this product must not be disposed
of as unsorted municipal waste
10Rotating Bottle Apparatus Operator’s Manual
General
Safety1
CE Compliant Products
The Rotating Bottle Apparatus has been designed to comply with the
requirements of the Electro-magnetic Compatibility (EMC) Directive
and the Low Voltage Directive (LVD) of the EU.
Agilent, Inc. has confirmed that each product complies with the relevant
directives by testing a prototype against the prescribed European Norm
(EN) standards.
Proof that a product complies with the directives is indicated by:
• the CE marking appearing on the rear of the product.
• the documentation package that accompanies the product containing
a copy of the declaration of conformity. This declaration is the legal
declaration by Agilent, Inc. that the product complies with the
directives and also shows the EN standards to which the product was
tested to demonstrate compliance. The declaration of conformity is
signed by the representative of the manufacturing plant.
cTUVus - U.S. and Canadian Product Approvals
The Rotating Bottle Apparatus has been designed to comply with North
American safety requirements.
This product has been tested and certified for the North American
market by TUV Rheinland of North America, Inc. The TUVus mark
signifies that this product has been tested to U.S. standards and certified
for the U.S. market. The cTUV mark signifies that this product has been
tested to Canadian standards and certified for the Canadian market.
When the two marks are coupled, the cTUVus mark signifies that this
product has been tested to standards and certified for both markets.
Rotating Bottle Apparatus Operator’s Manual11
1Safety
WEEE Directive
All Agilent products that are subject to the WEEE directive shipped
after August 13, 2005 are compliant with the WEEE marking
requirements. Such products are marked with the “crossed out wheelie
bin” WEEE symbol shown on page 10 in accordance with European
Standard EN 50419.
This symbol on the product or on its packaging indicates that this
product must not be disposed of as unsorted municipal waste. The
separate collection and recycling of your waste equipment at the time of
disposal will help to conserve natural resources and ensure that it is
recycled in a manner that protects human health and the environment.
For more information on collection, reuse, and recycling systems,
please contact your local/regional waste administration, your local
distributor, or Agilent, Inc.
12Rotating Bottle Apparatus Operator’s Manual
Rotating Bottle Apparatus
WARNING
Operator’s Manual
2
Introduction
Conventions Used in this Manual 15
The Rotating Bottle Apparatus is designed for release rate testing of
tablets or capsules. Manual changing of media during a test simulates a
dosage passing through a pH gradient.
The Rotating Bottle Apparatus includes the following items:
• One system drive with easy-to-read digital display
• One clear acrylic water bath, 19 x 12.7 x 15.3 inches, with 750
Heater / Circulator or one clear acrylic water bath, 24 x 12 x 12
inches, for use with an immersion circulator
• One stainless steel drive bar with 14 positions
• 20 standard 100 mL test bottles with caps
• Five decant caps with 40-mesh screen
The Rotating Bottle Apparatus contains electrical circuits,
devices, and components operating at dangerous voltages.
Contact with these circuits, devices, and components can cause
death, serious injury, or painful electric shock
Agilent Technologies
13
2Introduction
CAUTION
Figure 1Rotating Bottle Apparatus
14Rotating Bottle Apparatus Operator’s Manual
Panels or covers that are retained by fasteners which require the use
of a tool for removal may be opened only by Agilent-trained,
Agilent-qualified, or Agilent-authorized service engineers.
Conventions Used in this Manual
NOTE
• Items you are asked to press are in bold. For example, “press H on
the keypad”.
Remember to return the warranty card supplied with this manual.
Completing and returning the card ensures your right to protection
under the terms and conditions of your warranty. It also enables us to
better assist you in the event of any problems. Additionally, it
guarantees you will be informed of any issues that arise concerning your
equipment, such as upgrades, retrofits, or regulatory changes.
Introduction2
Rotating Bottle Apparatus Operator’s Manual15
2Introduction
This page was intentionally left blank, except for this message.
16Rotating Bottle Apparatus Operator’s Manual
Rotating Bottle Apparatus
WARNING
Operator’s Manual
3
Setting Up the Rotating Bottle
Apparatus
Unpacking Your Equipment 18
Setting up the Water Bath and 750 19
Initial Power Up 20
Keep your hands clear of the apparatus as it rotates.
Agilent Technologies
17
3Setting Up the Rotating Bottle Apparatus
WARNING
WARNING
Unpacking Your Equipment
1 Open each carton and check the contents for damage which may
have occurred during shipping. Shipping damage rarely occurs, but if
it does contact both the carrier who delivered the instruments and the
Dissolution Systems Service Department. Though claims for damage
should be filed with the carrier, we can help you file a claim.
2 Carefully remove the Rotating Bottle Apparatus from its shipping
carton. Hold the unit firmly to prevent it from dropping.
Hold the apparatus only by the base. To avoid damage, do not
hold the apparatus by the drive unit.
3 Check for items which may have come loose during shipping. Be
sure to remove all parts before discarding or storing the packaging.
4 Place the unit on a clean, dry, level section of the bench top or table.
At least four inches (10 cm) of space should be open at the rear and
on both sides of the instrument.
18Rotating Bottle Apparatus Operator’s Manual
The electrical connection at the back of the apparatus is the
primary disconnect for the instrument. The apparatus should be
positioned to allow accessibility to the power cord for easy
disconnection.
Setting Up the Rotating Bottle Apparatus3
NOTE
NOTE
Setting up the Water Bath and 750
If your equipment is not supplied with a 750, attach the heater /
circulator according to the manufacturer instructions.
Complete the following steps to install the 750:
1 Locate the two lengths of 1/2-inch plastic tubing (one long, one
short) and four stainless steel tubing clamps in the 750 tubing kit.
These instructions assume the 750 is placed to the right of the Apparatus
7 when looking at the front of the equipment. If your 750 is placed to the
left of the Apparatus 7, use the longer tubing when the instructions call
for the shorter tubing and the shorter tubing when the instructions call
for the longer tubing.
2 Slide a tubing clamp over one end of the shorter length of tubing
supplied. Carefully place this end of the tubing over the connector
labeled INLET on the 750 rear panel. Slide the tubing on until it
meets the rear panel.
3 Slide the tubing clamp over the connector and tighten securely.
4 Place a tubing clamp over the opposite end of the same piece of
tubing. Attach the tubing to the open end of the Y connector coming
from the barbed-angle tubing adapter on the right side of the water
bath.
5 Slide the tubing clamp over the connector and tighten securely.
6 Connect the water bath inlet located on the rear of the water bath to
the 750 rear panel connector labeled OUTLET in the same fashion
using the longer piece of tubing.
7 Connect the four-pin DIN cable between the jack labeled INPUT
SIGNAL on the 750 rear panel and the jack labeled HEATER
CIRCL. on the Apparatus 7 right side panel.
Rotating Bottle Apparatus Operator’s Manual19
3Setting Up the Rotating Bottle Apparatus
NOTE
WARNING
8 Ensure the 750 rear panel power switch is in the OFF position.
If possible, preheat the water to speed achievement of the desired water
bath temperature. Use purified water whenever possible to minimize
scale and mineral buildup. Algaecides may be used to inhibit algae and
bacteria growth. Check the label to ensure the formulation is compatible
with the plastic materials used in the water bath construction.
Connecting the Temperature Probe
1 Plug the supplied water bath temperature probe into the jack labeled
BATH TEMP on the Apparatus 7 right side panel.
2 Place the temperature probe through the small hole in the vessel table
on the rear left side so that approximately 3/4 of the probe is below
the surface of the table.
Initial Power Up
20Rotating Bottle Apparatus Operator’s Manual
Power Cord Connections
1 Ensure the Rotating Bottle Apparatus is turned off.
Before plugging the Rotating Bottle Apparatus into any power
outlet, ensure the instrument is properly configured for the
voltage provided. Check the serial number tag on the rear panel
of the instrument to confirm the voltage requirement.
The electrical connection at the back of the apparatus is the
primary disconnect for the instrument.
2 Connect power cords to the receptacles at the rear of the Rotating
Bottle Apparatus and the 750.
3 Plug the cords into electrical outlets of the appropriate voltage.
Setting Up the Rotating Bottle Apparatus3
NOTE
Powering Up the 750 Heater / Circulator
Turn on the 750. Flow should begin immediately and the unit should
operate quietly with a low “hum”. Bubbles may appear at the water bath
inlet as air is expelled from the system. After a few minutes flow into
the water bath should be smooth and steady. If not, stop the system and
prime it as described in the 750 Heater / Circulator Operator’s Manual.
Make a final check for leaks. Allow the heater / circulator to run for five
minutes, then check the water bath, tubing, connectors, and the bench
area immediately surrounding the apparatus. If any water leaks are seen,
turn off the power immediately and remove the power cord. Locate and
fix the source of the leak before proceeding.
Powering Up the Rotating Bottle Apparatus
Turn on the Rotating Bottle Apparatus. The red LED on the front panel
illuminates. If it does not, ensure the unit is securely plugged in and
there is power at the outlet. If the unit still fails to respond, see
“Troubleshooting” on page 32 or call the Dissolution Systems Service
Department.
Rotating Bottle Apparatus Operator’s Manual21
3Setting Up the Rotating Bottle Apparatus
This page was intentionally left blank, except for this message.
22Rotating Bottle Apparatus Operator’s Manual
Rotating Bottle Apparatus
WARNING
Operator’s Manual
4
Operating the Rotating Bottle
Apparatus
Operating Instructions 24
Keep fingers and loose clothing away from the lower end of the
drive belt assembly.
Agilent Technologies
23
4Operating the Rotating Bottle Apparatus
NOTE
NOTE
Operating Instructions
The procedure as described in NF XIV is not an official method, but is
suggested as a suitable test that could be helpful in assuring product
uniformity of timed-release tablets and capsules. See General Tests,
“Timed-Release Tablets and Capsules–In Vitro Test Procedure,” pages
985-986, NF XIV, July 1, 1975.
To operate the Rotating Bottle Apparatus, complete the following steps:
1 Add tablets or capsules and extracting fluid to the standard test
bottles.
2 Secure the standard caps on the bottles.
3 Snap the bottles into the clips.
For even speed control, the bottles should be distributed in parallel pairs
on opposite sides of the shaft.
24Rotating Bottle Apparatus Operator’s Manual
4 Center each bottle within the clip.
5 Carefully place the apparatus in the water bath.
6 Turn on the Rotating Bottle Apparatus.
7 Press C on the keypad to select the speed and enter the desired
rotation speed (between 10 RPM and 60 RPM).
Optionally, press B on the keypad to toggle between 30 RPM and
40 RPM.
Key A on the keypad is inactive.
8 Press RUN to begin rotation of the bottles.
9 At the appropriate sampling time, press STOP. The apparatus pauses
and the bottles stop rotating.
10 Remove the Rotating Bottle Apparatus from the water bath.
11 Remove the bottles from the clips.
12 Replace the standard caps with 40-mesh decant screen caps.
Operating the Rotating Bottle Apparatus4
NOTE
13 Decant, retaining the residue.
14 Add extracting fluid to the residue.
15 Replace the 40-mesh decant screen caps with the standard caps.
Ensure the caps are secured tightly.
16 Snap the bottles into the clips.
For even speed control, the bottles should be distributed in parallel pairs on
opposite sides of the shaft.
17 Center each bottle within the clip.
18 Carefully place the Rotating Bottle Apparatus in the water bath.
19 Press RUN to continue the test.
20 When the test is complete, press STOP.
Rotating Bottle Apparatus Operator’s Manual25
4Operating the Rotating Bottle Apparatus
This page was intentionally left blank, except for this message.
26Rotating Bottle Apparatus Operator’s Manual
Rotating Bottle Apparatus
Operator’s Manual
5
Troubleshooting and Maintenance
Periodic Maintenance 28
Troubleshooting 32
Agilent Technologies
27
5Troubleshooting and Maintenance
WARNING
Periodic Maintenance
The Rotating Bottle Apparatus contains electrical circuits,
devices, and components operating at dangerous voltages.
Contact with these circuits, devices, and components can cause
death, serious injury, or painful electric shock.
Panels or covers that are retained by fasteners which require the
use of a tool for removal may be opened only by Agilent-trained,
Agilent-qualified, or Agilent-authorized service engineers.
Periodic maintenance intervals may vary depending on frequency of
instrument usage.
Daily Maintenance
• Immediately wipe all spills or errant materials from the exterior of
the instrument or any exposed part using a clean cloth and plain
water.
• Carefully wash the bottles and caps after each use.
Weekly Maintenance
See “Water Bath / Acrylic Care” on page 30 as applicable for additional
information on proper maintenance of your equipment.
• Inspect the water bath and heater / circulator tubing for algae or
other materials. If algae is present, change the bath water and add
an algaecide.
• If you use a water bath algaecide or clear bath product, ensure it is
compatible with PETG and acrylic.
28Rotating Bottle Apparatus Operator’s Manual
Monthly Maintenance
NOTE
CAUTION
See “Water Bath / Acrylic Care” on page 30 as applicable for additional
information on proper maintenance of your equipment.
• Clean the outside of the apparatus with a damp cloth.
• Drain the water from the water bath and clean the bath
thoroughly. Refill the water bath and add an algaecide.
• It is recommended that the water bath temperature probe jack is
checked for surface corrosion and wiped clean with a soft cloth or
nonabrasive pad every one to three months.
Depending on the frequency of use, it may be necessary to complete this
maintenance procedure more often.
Annual Maintenance
Troubleshooting and Maintenance5
• Inspect all electrical connections for corrosion and damage.
• Clean the outside of the apparatus with a damp cloth.
• Check the belt tension—total deflection should not exceed 1/2 inch
(1.3 cm).
Do not use cleaning compounds containing ammonia or abrasive
cleaners on your water bath.
Rotating Bottle Apparatus Operator’s Manual29
5Troubleshooting and Maintenance
Water Bath / Acrylic Care
The water bath supplied with the Rotating Bottle Apparatus should be
maintenance free except for an occasional cleaning. If you use a water
bath algaecide or clear bath product, ensure it is compatible with PETG
and acrylic. The flow paths in the immersion circulator are primarily
stainless steel and should tolerate most clear bath formulations. Check
with the product manufacturer to be sure the product is safe for your
water bath.
• All of our water baths are fabricated entirely of commercial grade
acrylic. When using them with corrosive materials such as
hydrochloric acid or media containing salts, be sure to rinse them
thoroughly with deionized water immediately after each use, and dry
thoroughly with a soft towel or cloth.
• Do not clean with abrasive cleansers or cloths. Use deionized water
whenever possible. If you must use a cleanser or solvent, be sure that
it is as mild as possible, non-abrasive, and fully compatible with
PETG and acrylic before use. If in doubt, call the service department
for advice before proceeding.
• Do not use ammonia, window-cleaning sprays, kitchen scouring
compounds, or solvents such as acetone, gasoline, benzene, alcohol,
carbon tetrachloride, or lacquer thinner. These can scratch the
material’s surface and / or weaken it causing small surface cracks
called “crazing”.
30Rotating Bottle Apparatus Operator’s Manual
• Our recommendations include but are not limited to the following:
WARNING
WARNING
• Hot water:< 150 °F
• Vinegar (5% Glacial Acetic Acid)
• Ethyl alcohol: maximum 10%
• Isopropyl alcohol: maximum 25%
Fuse Replacement
The Rotating Bottle Apparatus contains electrical circuits,
devices, and components operating at dangerous voltages.
Contact with these circuits, devices, and components can cause
death, serious injury, or painful electric shock.
Panels or covers that are retained by fasteners which require the
use of a tool for removal may be opened only by Agilent-trained,
Agilent-qualified, or Agilent-authorized service engineers.
Follow these steps to replace the fuse:
1 Before checking or attempting to replace a fuse, remove the power
cord from the unit.
2 Press the release tab on the fuse holder located in the power entry
module with the blade of a small screwdriver or equivalent.
3 Upon release, slide the holder out of the AC power entry module.
4 Remove the old fuses and insert the new fuses.
5 Slide the holder back into the power entry module and push until it
locks into place.
6 Replace the power cord.
Never replace a fuse with one of a higher amperage rating.
Doing so may compromise the safety margin and could result in
damage to the instrument or personal injury.
Troubleshooting and Maintenance5
Rotating Bottle Apparatus Operator’s Manual31
5Troubleshooting and Maintenance
Troubleshooting
The Dissolution Systems Service Department can assist you if you
experience problems or have questions concerning your Rotating Bottle
Apparatus. Many problems can be traced to simple sources and are
easily solved.
Following is a troubleshooting guide which may help you. The
Dissolution Systems Service Department can be reached at
800.229.1108 (inside the US) or 919.677.1108 (outside the US).
Optionally, you can send a fax to 919.677.1138. You can also e-mail the
Dissolution Systems Service Department at
dissolution.service@agilent.com.
SymptomProbable CausePossible Solution
There is no power to the apparatus. The fuses are blown (open).Replace the fuses in the power entry
module. See “Fuse Replacement” on
page 31.
32Rotating Bottle Apparatus Operator’s Manual
There is no power at the electrical
outlet.
The power cord is damaged or is
not plugged in.
The power entry module is faulty.Check the module to ensure proper voltage
Verify that the outlet has power going to it.
Check the power cord for abrasion and
proper connection. Replace if necessary.
delivery. Contact the Dissolution Systems
Service Department for assistance.
Rotating Bottle Apparatus
Operator’s Manual
6
Service and Warranty
Service and Warranty Information 34
The warranty is provided by Agilent Technologies, Inc. or one of its
authorized representatives.
Agilent Technologies
33
6Service and Warranty
Service and Warranty Information
Agilent dissolution products carry a one-year warranty on parts and
labor. The Dissolution Systems Service Department (or one of its
representatives) will, at its option, either repair or replace any
mechanical and electrical components in your instrument which prove
to be defective. During the first year of warranty coverage, there is no
charge for the labor to repair your unit. The Dissolution Systems Service
Department (or one of its representatives) will determine the best site to
repair the unit, either onsite or returned to Agilent Technologies, Inc.
Any onsite warranty services are provided only at the initial installation
point. Installation and onsite warranty services are available only in
Dissolution Systems service travel areas.
Exclusions and Limitations
Excluded from this warranty are expendable or consumable items such
as, but not limited to, paddles, baskets, vessels, and acrylic water baths.
Also excluded are defects from improper or inadequate maintenance by
the customer, user-induced chemical action or contamination,
unauthorized modification or misuse, and improper site preparation and
maintenance.
Operation of software is not warranted to be uninterrupted or error-free.
Obtaining Warranty Service
To obtain warranty service in the United States, contact the Dissolution
Systems Service Department at 800.229.1108 to obtain authorization to
return units for repair. At the option of the customer, onsite warranty
service is available, but travel charges may be incurred. The customer
should prepay all shipping charges for products returned to the
Dissolution Systems Service Department (unless otherwise authorized),
and Agilent Technologies, Inc. will pay all charges for return to the
customer.
34Rotating Bottle Apparatus Operator’s Manual
Warranty Limitations
Agilent Technologies, Inc. makes no other warranty, either express or
implied, with respect to this product. Specifically disclaimed are any
implied warranties of merchantability and fitness for a particular use. In
no event will Agilent Technologies, Inc. be liable for any indirect,
incidental, or consequential damages arising from the use of this
product. This warranty gives you specific legal rights which may vary
from state to state or province to province, so you may have other rights
and some of these exclusions may not apply to you.
Exclusive Remedies
The remedies provided herein are the customer’s sole and exclusive
remedies. In no event shall Agilent Technologies, Inc. or its
representatives be liable for any direct, indirect, special, incidental, or
consequential damages, whether based on contract, tort, or any other
legal theory. Some states or provinces do not allow the exclusion or
limitation of incidental or consequential damages, so the above
limitation or exclusion may not apply to you.
Service and Warranty6
Rotating Bottle Apparatus Operator’s Manual35
6Service and Warranty
This page was intentionally left blank, except for this message.