All rights reserved.
Reproduction, adaption,
or translation without
prior written permission
is prohibited, except as
allowed under the
copyright laws.
Part No. G1311-90003
Edition 01/00
Printed in Germany
Warra nty
The information
contained in this
document is subject to
change without notice.
Agilent Technologies
makes no warranty of
any kind with regard to
this material,
including, but not
limited to, the implied
warranties or
merchantability and
fitness for a particular
purpose.
Agilent Technologies
shall not be liable for
errors contained herein
or for incidental or
consequential damages
in connection with the
furnishing, performance,
or use of this material.
WAR N ING
For details of safety,
see Safety Information
on page 260.
Warning Symbols Used
In This Book
!
The apparatus is marked
with this symbol when
the user should refer to
the instruction manual
in order to protect the
apparatus against
damage.
This manual contains technical reference information about the Agilent 1100
Series quaternary pump. The manual describes the following:
• installation,
• optimizing performance,
• troubleshooting,
• repairing,
• parts and materials,
• theory of operation, and
• specifications.
4
Contents
1 Installing the Pump
How to install the quaternary pump13
Site Requirements14
Unpacking the Quaternary Pump17
Optimizing the Stack Configuration20
Installing the Quaternary Pump22
Flow Connections of the Quaternary Pump25
Priming and Purging the System28
2 Optimizing Performance
How to optimize the quaternary pump to achieve best chromatographic results31
Hints for Successful Use of the Quaternary Pump32
Solvent Information34
Prevent Blocking of Solvent Filters35
Operational Hints for the Vacuum Degasser36
Operational Hints for the Multi Channel Gradient Valve
(MCGV)37
When to use the Continuous Seal Wash Option38
When to Use Alternative Seals39
Optimize the Compressibility Compensation Setting40
3 Troubleshooting and Test Functions
The quaternary pump’s built-in troubleshooting and test
functions43
Status Indicators45
5
Contents
Power Supply Indicator46
Pump Status Indicator46
Error Messages47
Timeout48
Shut-Down49
Remote Timeout50
Synchronization Lost51
Leak52
Leak Sensor Open53
Leak Sensor Short54
Compensation Sensor Open55
Compensation Sensor Short56
Fan Failed57
Open Cover58
Restart Without Cover59
Zero Solvent Counter60
Pressure Above Upper Limit61
Pressure Below Lower Limit62
Pressure Signal Missing63
Missing Pressure Reading64
Pump Configuration65
Valve Fuse66
Inlet-Valve Fuse67
Valve Failed68
Motor-Drive Power69
Encoder Missing70
Inlet-Valve Missing71
Temperature Out of Range72
Temperature Limit Exceeded73
Servo Restart Failed74
6
Contents
Pump Head Missing75
Index Limit76
Index Adjustment77
Index Missing78
Stroke Length79
Initialization Failed80
Wait Timeout81
Pressure Test82
Running the Pressure Test84
Evaluating the Results86
Leak Test88
Running the Leak Test90
Evaluating the Results92
4 Repairing the Pump
Instructions on simple, routine repair procedures as well as
more extensive repairs requiring exchange of internal
parts97
Cleaning the Quaternary Pump99
Using the ESD Strap100
Overview101
Simple Repair Procedures102
Exchanging the Active Inlet Valve Cartridge or the Active Inlet
Valve103
Exchanging the Outlet Ball Valve106
Exchanging the Purge Valve Frit or the Purge Valve108
7
Contents
Removing and Disassembling the Pump Head Assembly110
Exchanging the Pump Seals and Seal Wear-in Procedure112
Exchanging the Plungers115
Installing the Continuous Seal Wash Option116
Exchanging the Wash Seals119
Reassembling the Pump Head Assembly121
Exchanging the Multi-Channel Gradient Valve (MCGV)123
Exchanging the optional Interface Board126
Exchanging Internal Parts127
Removing the Top Cover and Foam128
Exchanging the Low Pressure Pump Main Board (LPM
Board)131
Exchanging the Damper138
Exchanging the Fan141
Exchanging the Pump Drive143
Exchanging the Power Supply147
Exchanging the Leak Sensor152
Exchanging the Status Light Pipe154
Assembling the Main Cover155
Replacing the Top Cover and Foam156
5 Parts and Materials
Detailed illustrations and lists for identification of parts and
materials159
Overview of Main Assemblies160
Control Module (B-version)163
Solvent Cabinet164
8
Contents
Bottle Head Assembly165
Hydraulic Path166
Cover Parts167
Sheet Metal Kit168
Foam Parts169
Power and Status Light Pipes170
Leak Parts171
Pump Head Assembly172
Pump Head Assembly with Seal Wash Option174
Outlet Ball Valve Assembly176
Purge Valve Assembly177
Active Inlet Valve Assembly178
Accessory Kit G1311-68705179
Seal Wash Option Kit 01018-68722180
Cable Overview181
Analog Cables183
Remote Cables185
BCD Cables190
Auxiliary Cable192
CAN Cable192
External Contact Cable193
RS-232 Cable Kit194
LAN Cables195
6 Introduction to the Quaternary Pump
An introduction to the pump, instrument overview, theory of
operation, external communication and internal
connectors197
9
Contents
Introduction to the Quaternary Pump198
Overview199
Electrical Connections204
Instrument Layout206
Early Maintenance Feedback (EMF)207
The Electronics209
The Low-Pressure Pump Main Board (LPM)210
Firmware Description214
Optional Interface Boards216
Agilent 1100 Series Interfaces218
Setting the 8-bit Configuration Switch223
The Main Power Supply Assembly228
7 Control Module Screens for the Quaternary Pump
Screens available from the Analysis screen233
Screens available from the System screen242
Screens available from the Records screen244
Diagnostics and Tests251
8 Specifications
Performance specifications of the quaternary pump253
Radio Interference264
Sound Emission264
Solvent Information265
Agilent Technologies on Internet266
11
Contents
12
1
1Installing the Pump
How to install the quaternary pump
Installing the Pump
Site Requirements
Site Requirements
A suitable environment is important to ensure optimum performance of the
quaternary pump.
Power Consideration
The quaternary pump power supply has wideranging capability (see Table 1
on page 16). It accepts any line voltage in the range described in the above
mentioned table. Consequently there is no voltage selector in the rear of the
quaternary pump. There are also no externally accessible fuses, because
automatic electronic fuses are implemented in the power supply.
WARNINGTo disconnect the quaternary pump from line, unplug the power cord.
The power supply still uses some power, even if the power switch on
the front panel is turned off.
WARNINGShock hazard or damage of your instrumentation can result, if the
devices are connected to a line voltage higher than specified.
Power Cords
Different power cords are offered as options with the quaternary pump. The
female end of each of the power cords is identical. It plugs into the
power-input socket at the rear of the quaternary pump. The male end of each
of the power cords is different and designed to match the wall socket of a
particular country or region.
WARNINGNever operate your instrumentation from a power outlet that has no
ground connection. Never use a power cord other than the power cord
designed for your region.
WARNINGNever use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.
14
Installing the Pump
Site Requirements
Bench Space
The quaternary pump dimensions and weight (see Table 1 on page 16) allow
to place the quaternary pump on almost any laboratory bench. It needs an
additional 2.5 cm (1.0 inches) of space on either side and approximately 8 cm
(3.1 inches) in the rear for the circulation of air and electric connections.
If the bench should carry a complete Agilent 1100 Series system, make sure
that the bench is designed to carry the weight of all the modules.
NOTEThe pump should be operated in a horizontal position!
Environment
Your quaternary pump will work within specifications at ambient
temperatures and relative humidity as described in Table 1 on page 16.
CAUTIONDo not store, ship or use your quaternary pump under conditions where
temperature fluctuations could cause condensation within the quaternary
pump. Condensation will damage the system electronics. If your quaternary
pump was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.
15
Installing the Pump
Site Requirements
Table 1Physical Specifications
TypeSpecificationComments
Weight11 kg (25 lbs)
Dimensions
(height × weight × depth)
Line voltage
Line frequency
Power consumption220 VAMaximum
Ambient operating temperature
Ambient non-operating temperature
Humidity
Operating AltitudeUp to 2000 m (6500 ft)
Non-operating altitudeUp to 4600 m (14950 ft)For storing the quaternary
Safety standards: IEC, CSA, ULInstallation Category II, Pollution Degree 2
140 × 345 × 435 mm
(5.5 × 13.5 × 17 inches)
100–120 or 220–240 VAC,
50 or 60 Hz,
4–55
-40–70
< 95%, at 25–40
± 5%
°C (41–131 °F)
°C (-4–158 °F)
°C (77–104 °F)
± 10%
Wide-ranging capability
Non-condensing
pump
16
Installing the Pump
Unpacking the Quaternary Pump
Unpacking the Quaternary Pump
Damaged Packaging
Upon receipt of your quaternary pump, inspect the shipping containers for
any signs of damage. If the containers or cushioning material are damaged,
save them until the contents have been checked for completeness and the
quaternary pump has been mechanically and electrically checked. If the
shipping container or cushioning material is damaged, notify the carrier and
save the shipping material for the carrier’s inspection.
CAUTIONIf there are signs of damage to the quaternary pump, please do not attempt to
install the quaternary pump.
Delivery Checklist
Ensure all parts and materials have been delivered with the quaternary pump.
The delivery checklist is shown in Table 2. To aid in parts identification,
please see Chapter 5 “Parts and Materials”. Please report missing or damaged
parts to your local Agilent Technologies sales and service office.
Reference Manual2 (1 for the pump, 1 for the degasser)
Accessory kit (see Table 3) 1
Accessory Kit Contents — Quaternary Pump
Table 3Accessory Kit Contents G1311-68705
DescriptionPart NumberQuantity
Capillary, pump to injection deviceG1312-673051
Seal insert tool01018-237021
Wrench; 1/4 – 5/16 inch8710-05101
Wrench; 14 mm8710-19241
ESD wrist strap
*
9300-14081
Hex key 4mm8710-23921
Waste tube (reorder number, 5m)5062-24631.2 m
Velocity regulator (reorder number)5062-24862
PTFE Frit01018-227075
*
ESD: Electrostatic Discharge
18
Installing the Pump
Unpacking the Quaternary Pump
Accessory Kit Contents—Vacuum Degasser
Table 4Accessory Kit Contents G1322-68705
DescriptionPart NumberQuantity
Syringe5062-85341
Syringe adapter9301-13371
Waste tube (reorder number, 5m)5062-24631.2 m
Connecting tubes labeled A to DG1322-673004
19
Installing the Pump
Optimizing the Stack Configuration
Optimizing the Stack Configuration
If your quaternary pump is part of a complete 1100 Series system, you can
ensure optimum performance by limiting the configuration of the system
stack to the following configuration. This configuration optimizes the system
flow path, ensuring minimum delay volume.
CAN Bus cable for inter
module communication
5181-1516 (0.5m)
5161-1519 (1.0m)
AC power
Analog signal to
recorder,for
PN see page 181
GPIB or
LAN to ChemStation
for PN see page 181
NOTEIf a single stack configuration becomes too high, e.g. if an additional module
like a G1327A ALS Thermostat is added or if your bench is to high, a two stack
configuration may be a better setup. Separate the stack between pump and
autosampler and place the stack containing the pump on the right side of the
stack containing the autosampler.
21
Installing the Pump
S
Installing the Quaternary Pump
Installing the Quaternary Pump
PreparationsLocate bench space.
Provide power connections.
Unpack the pump.
Parts requiredPump
Power cord, for other cables see text below and “Cable Overview” on page 181
ChemStation and/or Control Module G1323A/B
1 Place the quaternary pump on the bench in a horizontal position.
2 Ensure the power switch on the front of the quaternary pump is OFF (switch
stands out).
Figure 3Front of Quaternary Pump
tatus Lamp
Power Switch
Serial number
3 At the rear of the quaternary pump move the security lever to its maximum
right position.
4 Connect the power cable to the power connector at the rear of the quaternary
pump. The security lever will prevent that the cover is opened while the power
cord is connected to the quaternary pump.
5 Connect the required interface cables to the quaternary pump.
22
Installing the Pump
Installing the Quaternary Pump
NOTEIn an Agilent 1100 Series system, the individual modules are connected
through CAN cables. The Agilent 1100 Series vacuum degasser is an exception
. The vacuum degasser can be connected via the APG remote connector to the
other modules of the stack. The AUX output allows the user to monitor the
vacuum level in the degasser chamber. An Agilent 1100 Series control module
can be connected to the CAN bus at any of the modules in the system except
for the degasser. The Agilent ChemStation can be connected to the system
through one GPIB or LAN (requires the installation of a LAN- board) cable at
any of the modules (except for the degasser), preferably at the detector
(MUST for the DAD). For more information about connecting the control
module or Agilent ChemStation refer to the respective user manual. For
connecting the Agilent 1100 Series equipment to non-Agilent 1100 Series
equipment, see Chapter 6 “Introduction to the Quaternary Pump”
Figure 4Rear of Quaternary Pump
Security lever
Slot for interface board
Analog pressure, 2mV/bar
APG Remote
RS-232C
CAN
GPIBPower
Configuration switch
6 Connect all capillaries, solvent tubes and waste tubing (see “Flow
Connections of the Quaternary Pump” on page 25).
7 Press in the power switch to turn on the quaternary pump.
23
Installing the Pump
Installing the Quaternary Pump
NOTEThe power switch stays pressed in and a green indicator lamp in the power
switch is on when the quaternary pump is turned on. When the line power
switch stands out and the green light is off, the quaternary pump is turned off.
8 Purge the quarternary pump (see “Priming and Purging the System” on page
28).
WARNINGTo disconnect the quaternary pump from line, unplug the power cord.
The power supply still uses some power, even if the power switch on
the front panel is turned off.
NOTEThe pump was shipped with default configuration settings. To change these
settings, see “Setting the 8-bit Configuration Switch” on page 223.
24
Installing the Pump
Flow Connections of the Quaternary Pump
Flow Connections of the Quaternary Pump
PreparationsPump is installed in the LC system.
Parts requiredOther modules
Parts from accessory kit, see “Accessory Kit Contents — Quaternary Pump” on page 18
Two wrenches 1/4–5/16 inch for capillary connections
WARNINGWhen opening capillary or tube fittings solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used.
1 Remove the front cover by pressing the snap fasteners on both sides.
Figure 5Removing the Front Cover
2 Place the vacuum degasser and the solvent cabinet on top of the quaternary
pump.
3 Put the bottle-head assemblies into solvent reservoirs containing your mobile
phase and place the bottle in the solvent cabinet.
4 Connect the inlet tubes from the bottle-head assemblies to the inlet
connectors A to D (typically the left connection of the channel) of the vacuum
degasser. Fix the tubes in the tube clips of the vacuum degasser.
25
Installing the Pump
Flow Connections of the Quaternary Pump
5 Connect the solvent tubes to the outlet connectors (typically right connection
of the channel) of the vacuum degasser.
6 Connect the syringe adapter from the degasser accessory kit to the solvent
tube of channel A.
7 Using a piece of sanding paper connect the waste tubing to the purge valve
and place it into your waste system.
8 If the quaternary pump is not part of a Agilent1100 System stack or placed on
the bottom of a stack, connect the corrugated waste tube to the waste outlet
of the pump leak handling system.
9 Connect the pump outlet capillary (quaternary pump to injection device) to
the outlet of the purge valve.
10 Prime your system before first use (see “Priming and Purging the System” on
page 28).
26
Installing the Pump
Flow Connections of the Quaternary Pump
Figure 6Flow Connections of the Quaternary Pump
Bottle-head assembly (G1311-60003)
Solvent cabinet
Vacuum degasser
Inlet
Outlet
Tube clip (1400-1578)
Tubings (G1322-67300)
Purge valve
Waste tubing (5062-2461)
Outlet capillary to autosampler (G1312-67305)
MCGV
Fitting for corrugated waste tubing (5062-2463, reorder pack, 5m)
27
Installing the Pump
Priming and Purging the System
Priming and Purging the System
The system can be primed either by drawing solvent through the degasser
with a syringe or by pumping with the pump.
Priming the system with a syringe is recommended, when:
• vacuum degasser or connected tubings are used for the first time or
vacuum tubes are empty or
• changing to solvents that are immiscible with the solvent currently in the
vacuum tubes.
Priming the system by using the pump at high flow rate
(3–5 ml/min) is recommended, when:
• pumping system was turned off for a length of time (for example,
overnight) and if volatile solvent mixtures are used, or
• solvents have been changed.
WARNINGWhen opening capillary or tube fittings solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used.
Priming with a Syringe
Before using a new degasser or new tubings for the first time:
1 Prime all tubings with at least 30 ml of iso-propanol no matter whether the
channels will be used with organic mobile phase or with water.
If you are changing to a solvent that is immiscible with the solvent currently
in the tubing continue as follows:
2 Replace the current solvent with adequate organic solvent (see Table 5 on
page 30), if current solvent is organic or with water, if current solvent is an
inorganic buffer or contains salt.
3 Disconnect solvent outlet tube of the channel that is supposed to be primed
from your pump.
4 Connect syringe adapter to solvent outlet tube.
28
Installing the Pump
Priming and Purging the System
5 Push syringe adapter onto syringe.
6 Pull syringe plunger to draw at least 30 ml of solvent through degasser and
tubing.
7 Replace the priming solvent with the new solvent of your choice.
8 Pull syringe plunger to draw at least 30 ml of solvent through degasser and
tubing.
9 Disconnect syringe adapter from solvent tube.
10 Connect the solvent tube to the appropriate channel of the MCGV.
11 Repeat step 3 to step 10 for the other solvent channels.
NOTEWhen priming the vacuum degasser with a syringe the solvent is drawn
through the degasser tubes very quickly. The solvent at the degasser outlet will
therefore not be fully degassed. Pump for approximately 10 minutes with your
selected flow rate before starting any application. This will allow the vacuum
degasser to properly degas the solvent in the degasser tubes.
NOTEThe pump should never be used for priming empty tubings (never let the pump
run dry). Use the syringe to draw enough solvent for completely filling the
tubings to the pump inlet before continueing to prime with the pump.
Priming with the Pump
When the pumping system has been turned off for a certain time (for
example, overnight) oxygen will rediffuse into the solvent channels between
the vacuum degasser and the pump. Solvents containing volatile ingredients
will slightly lose these, if left in the degasser without flow for a prolonged
period of time. Therefore priming of the vacuum degasser and the pumping
system is required before starting an application.
1 Open the purge valve of your pump (by turning it counterclockwise) and set
flow rate to 3-5 ml/min.
2 Flush the vacuum degasser and all tubes with at least 30 ml of solvent.
3 Set flow to required value of your application and close the purge valve.
4 Pump for approximately 10 minutes before starting your application.
5 Repeat step 1 to step 4 for other solvent channels, where needed.
29
Installing the Pump
Priming and Purging the System
Table 5Choice of Priming Solvents for Different Purposes
ActivitySolventComments
After an installation
When switching between
reverse phase and normal
phase (both times)
After an installationEthanol or MethanolAlternative to Isopropanol
To clean the system when
using buffers
After a solvent change
After the installation of
normal phase seals (P/N
0905-1420)
Isopropanol
Isopropanol
Bidistilled water
Bidistilled water
Hexane + 5% IsopropanolGood wetting properties
Best solvent to flush air out of
the system
Best solvent to flush air out of
the system
(second choice) if no
Isopropanol is available
Best solvent to re-dissolve
buffer cristals
Best solvent to re-dissolve
buffer cristals
30
Loading...
+ 242 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.