Agilent Technologies Preparative separation User Manual

Preparative separation of a binary compound mixture – recovery of pure compounds and solvent consumption
Abstract
This Application Note describes the separation of two compounds from
a binary compound mixture using the Agilent 1100 Series purification
system. The parameters purity, recovery, analysis run time, solvent con-
sumption and liquid phase composition are monitored and their interre-
mized with regard to these parameters.
Application
Udo Huber
Introduction
Separation of discrete compounds from a mixture is a typical task in preparative liquid chromatogra­phy. The simplest task is isolating two compounds from a binary mixture, however, it is also possi­ble to isolate pure compounds from more complex compound mixtures, such as natural product extracts 1, for example. The com­pounds to isolate are typically iso­mers or enantiomers
2,3
. In this Application Note we demonstrate the separation of two compounds from a binary com­pound mixture using the Agilent 1100 Series purification system. The method developed on an ana­lytical column was scaled up to preparative scale and the com­pounds were separated in mil­ligram quantities. The separation was done repeatedly, varying the composition of the liquid phase to achieve a separation with good recovery and purity with minimum solvent consumption.
Equipment
The system included two Agilent 1100 Series preparative pumps, an Agilent 1100 Series diode array detector, an Agilent 1100 Series column organizer and an Agilent 220 micro plate sampler modified for higher flow rates. The system was controlled using the Agilent ChemStation (revision A.08.04) and the micro plate sampling soft­ware (revision A.03.02).
Results and Discussion
Overloading of the analytical column
The separation of the binary com­pound mixture was first done on an analytical column. Separation was achieved isocratically with a water/ acetonitrile mixture (figure 1).
Figure 1 Analytical separation of binary compound mixtures
Time [min]
24681012
Absorbance [mAU]
0
500
1000
1500
2000
2500
30 % B
20 % B
15 % B
10 % B
Mobile phase: water = A
acetonitrile = B Isocratic: between 10 and 30 % B Stop time: 12 min Column: Zorbax SB-C18
3 x 150 mm, 5 µm Flow: 0.6 ml/min Injection: 5 µl Column temperature: ambient UV detector: DAD 270 nm/16
(reference 360 nm/100)
Standard cell
(10 mm pathlength)
Since the crude product was very soluble concentration overloading was possible. Figure 2 shows that the analytical column could be loaded with up to 500–1000 µg of each compound
Scale up to preparative scale
The scale-up from the analytical to the preparative column was calcu­lated using the formulae shown in figure 3. After the first preparative run the flow rate was changed to 25 ml/min to achieve comparable retention times.
0.025 µg
0.25 µg
2.5 µg
25 µg
250 µg
Isocratic: 10 % B
Time [min]
2 4 6 8 10 12 14
Absorbance [mAU]
0
1000
2000
3000
4000
5000
500 µg
1000 µg
Analytical column
Zorbax SB-C18 3 x 150 mm, 5 µm
Preparative column
Zorbax SB-C18
21.2 x 150 mm, 5 µm
Flow: 0.6 ml/min
Flow ~ 30 ml/min
Volume: 15 µl/injection
Volume: 750 µl/injection
x
1
=
r
2
1
×π
x1 = max. volume column 1 r1 = radius column 1 x2 = max. volume column 2 r2 = radius column 2 CL= ratio lengths of columns
= 15 µl = 1.5 mm = ? = 10.6 mm = 1
x
2
r
2
2
×π
×
1
C
L
V
1
V
2
r
1
r
2
2
2
=
.
.
Figure 2 Overloading of the analytical column
Figure 3 Scale up from analytical preparative column
Loading...
+ 5 hidden pages