5
Measurement conditions
1. Warm-up time: ≥ 15 min.
2. Ambient temperature: 23 ±5
˚
C
3. Test signal voltage: 1 Vrms
4. Test cable length: 0 meter
5. Open and short corrections performed
6. Measurement time: Medium or Long
(Other test condition data is available in the
operation manual.)
For |Z|, |Y|, L, C, R, X, G, and B accuracy
(Ae), refer to Table 1. Table 1 equations yield
accuracy based on frequency and DUT
characteristic impedance (Zm). Zm is
from Figure 1, Conversion Diagram.
D accuracy(De) = ± Ae/100
Q accuracy(Qe) = ±
where(Qm x De < 1)
u accuracy (ue) = 0.573 x Ae
Ae = Accuracy of |Z|, |Y|, L, C, R, X, G,
and B
De = D accuracy
Dm = Measured value of D
Qe = Q accuracy
Qm = Measured value of Q
ue = u phase angle accuracy
Zm = DUT impedance at test frequency in
Hertz
Other Specifications
Measurement parameters/ranges
Parameter Range
|Z|, R, X 1 mΩ to 100 MΩ
|Y|, G, B 10 nS to 1000S
C 1 pF to 1 F
L 10 nH to 100 kH
D 0.0001 to 9.9999
Q 0.1 to 9999.9
u -180
˚
to + 180
˚
∆ -999.99% to 999.99%
Option 4263B-001: DC resistance 1 mΩ to
100 M
Ω
Mutual inductance 1 µH to 100 H
(typical)
Turns-ratio 0.9 to 200 (typical)
Measurement conditions and functions
Test frequency: 100 Hz, 120 Hz,
1 kHz, 10 kHz, 100 kHz. (Option 4263B-002
adds 20 kHz.)
AC test signal level: 20 m - 1 Vrms,
5 mVrms steps
Bias:
Internal: + 1.5 and +2.0 Vdc
External: 0 to + 3.0 Vdc
Ranging: Auto and Hold
Trigger: Internal, Manual, and External
Trigger delay time: 0 to 9999 ms in
1 ms steps
Test cable lengths:
0, 1 meter @ f ≤ 100 kHz
2 meter @ f ≤ 10 kHz (20 kHz)
4 meter @ f ≤ 1 kHz
Measurement time:
SHORT MEDIUM LONG
25ms 65ms 500ms
Other instrument functions
Test signal level monitor:
Voltage, current
Error Correction: Open, Short, Load
Comparator: HIGH, IN, and LOW for each
displayed parameter
Save/recall: 10 instrument states from
non-volatile memory
Front-end Protection:
Vmax = @ Vmax ≤ 250 V
Vmax = @ Vmax ≤ 1000 V
C in Farads
Handler interface: Negative logic and isolated.
Signals are HIGH/IN/LOW, No-Contact,
EOM, Index, Alarm, Keylock, Ext. Trigger.
GPIB interface: Instrument control, TALKonly mode for LISTEN-only printers using
GPIB or Centronics/GPIB converter
Physical characteristics
Power: 90-132 Vac or 198-264 Vac.
47-66 Hz. 45 VA typical.
Operating temperature: 0 to 45 °C
Dimensions: 320 (W) x 100 (H) x
300 (H) mm
Weight: 4.5 kg (typical)