Adam Equipment CKT 16UH, CKT 8UH, CKT 32UH, CKT 48UH Instruction Manual

AdamEquipment
Cruiser
(CKT & CKT-M) SERIES
1.0 CONTENTS
PN 3.02.6.6.14038 Rev 2, July 2018
1.0 INTRODUCTION ................................................................................................................. 2
2.0 SPECIFICATIONS ............................................................................................................... 2
OTHER SPECIFICATIONS ......................................................................................................... 5
Key descriptions .............................................................................................................................. 6
3.0 OPERATION ....................................................................................................................... 7
3.1 SETTING UP AND TURN ON THE SCALE ..................................................................... 7
3.2 ZEROING THE DISPLAY ................................................................................................ 7
3.3 TARING ........................................................................................................................... 8
3.4 WEIGHING ...................................................................................................................... 8
3.5 PARTS COUNTING ......................................................................................................... 9
3.6 CHECK COUNTING ...................................................................................................... 10
3.7 PERCENT WEIGHING .................................................................................................. 11
3.8 ACCUMULATED TOTALS ............................................................................................. 12
4.0 RS-232 INTERFACE ......................................................................................................... 14
4.1 INPUT COMMANDS FORMAT ...................................................................................... 14
4.2 RS-232 SETUP .............................................................................................................. 15
4.3 USB SETUP ................................................................................................................... 15
5.0 CALIBRATION ............................................................................................................... 16
5.1 CKT CALIBRATION ....................................................................................................... 16
6.0 BATTERY .......................................................................................................................... 17
7.0 MODULE OPTION COVERS/ BOARDS ........................................................................... 17
8.0 REAL TIME CLOCK SETUP ............................................................................................. 18
9.0 ERROR CODES ................................................................................................................ 19
10.0 SERVICE INFORMATION ................................................................................................. 20
WARRANTY INFORMATION ....................................................................................................... 21
1.0 INTRODUCTION
The Cruiser Checkweighing (CKT) series provides accurate and reliable weighing for
stock control and general warehouse applications
There are 4 models in the CKT series:
1. CKT
2. CKT-UH
3. CKT-M
4. CKTa
Cruiser checkweighing scales can weigh in pound, gram or kilogram weighing units. The scales have stainless steel weighing platforms on an ABS base assembly. All scales are supplied with an RS-232 bi-directional interface and real time clock. The scales have a sealed keypad with colour coded membrane switches and a large,
easy to read liquid crystal type display (LCD). with backlight.
The scales include automatic zero tracking, LED checkweighing lights with selectable
Hi/Low limits, automatic tare, pre-set tare, an accumulation facility that allows the count to be stored and recalled as an accumulated total.
2.0 SPECIFICATIONS
Kilograms
Maximum
CKT 4 CKT 8H CKT 8 CKT 16 CKT 32 CKT 48
4 kg 8 kg 8 kg 16 kg 32 kg 48 kg
Capacity
Tare Range -4 kg -8 kg -8 kg -16 kg -32 kg -48 kg
Readability 0.0001 kg 0.0001 kg 0.0002 kg 0.0005 kg 0.001 kg 0.002 kg
Repeatability
0.0001 kg 0.0001 kg 0.0002 kg 0.0005 kg 0.001 kg 0.002 kg
(S.D.)
Linearity (±) 0.0002 kg 0.0002 kg 0.0004 kg 0.001 kg 0.002 kg 0.004 kg
Grams
Maximum
4000 g 8000 g 8000 g 16000 g 32000 g 48000 g
Capacity
Tare Range -4000 g -8000 g -8000 g -16000 g -32000 g -48000 g
Readability 0.1 g 0.1 g 0.2 g 0.5 g 1 g 2 g
Repeatability
0.1 g 0.1 g 0.2 g 0.5 g 1 g 2 g
(S.D.)
Linearity (±) 0.2 g 0.2 g 0.4 g 1 g 2 g 4 g
© Adam Equipment Company 2017 2
CKTa Series (USA models)
Pounds
Maximum
CKT 8H CKT 16UH CKT 32UH CKT 48UH
8 lb 16 lb 35 lb 70 lb
Capacity
Tare Range -8 lb -9.9995 lb -35 lb -70 lb
Readability 0.0002 lb 0.0002 lb 0.001 lb 0.002 lb
Repeatability
0.0002 lb 0.0002 lb 0.001 lb 0.002 lb
(S.D.)
Linearity (±) 0.0004 lb 0.0004 lb 0.002 lb 0.004 lb
Ounces
Maximum
128 oz 256 oz 560 oz 1120 oz
Capacity
Readability 0.005 oz 0.005 oz 0.02 oz 0.05 oz
Repeatability
0.005 oz 0.005 oz 0.02 oz 0.05 oz
(S.D.)
Linearity (±) 0.01 oz 0.01 oz 0.04 oz 0.1 oz
Pounds: Ounces
Maximum Capacity
8 lb: 0.0 oz 16 lb:
0.0 oz
35 lb:
0.0 oz
70 lb:
0.0 oz
Display shows 8:_16.00 16:_16.0 35:_16.0 70:_16.0
Readability 0.01 oz 0.1 oz 0.1 oz 0.1 oz
Repeatability
0.01 oz 0.1 oz 0.1 oz 0.1 oz
(S.D.)
Linearity (±) 0.02 oz 0.2 oz 0.2 oz 0.2 oz
Kilograms
Maximum
4 kg 8 kg 16 kg 32 kg
Capacity
Readability 0.0001 kg 0.0001 kg 0.0005 kg 0.001 kg
Repeatability
0.0001 kg 0.0001 kg 0.0005 kg 0.001 kg
(S.D.)
Linearity (±) 0.0002 kg 0.0002 kg 0.001 kg 0.002 kg
Grams
Maximum
4000 g 8000 g 16000 g 32000 g
Capacity
Readability 0.1 g 0.1 g 0.5 g 1 g
Repeatability
0.1 g 0.1 g 0.5 g 1 g
(S.D.)
Linearity (±) 0.2 g 0.2 g 1 g 2 g
© Adam Equipment Company 2017 3
Units of measure
Interface
Stabilisation Time
Operating Temperature
Power supply
Battery
Calibration
Display
CKT Series- kg, g
CKTa Series- kg, g, lb, oz, lb:oz
RS-232 bi-directional Interface
2 Seconds typical
-10°C to 40°C 14°F to 104°F
230 VAC 50/60 Hz 115 VAC available
Internal rechargeable battery (~90 hours operation)
Automatic External
6 digits LCD digital display
with capacity tracker and symbols for units
Scale Housing
ABS Plastic, Stainless Steel platform
Pan Size
Overall Dimensions (wxdxh)
315 x 355 x 110 mm
12.4” x 14” x 4.3”
Net Weight
Applications
Functions
Weighing, Parts counting, % weighing, Check weighing,
Weighing Scales
Check counting, Accumulation of weights.
Date/Time
Real Time Clock (RTC),
To print date and time information- battery backed
210 x 300 mm
8.3” x 11.8”
4.4 kg
9.7 lb
© Adam Equipment Company 2017 4
CKT 4M CKT 8M CKT 20M CKT 40M
Grams
Max 4000 g 8000 g 20000 g 40000 g
e = 1 g 2 g 5 g 10 g
n = 3000 3000 3000 3000
Kilograms
Max 4 kg 8 kg 20 kg 40 kg
e = 0.001 kg 0.002 kg 0.005 kg 0.01 kg
OTHER SPECIFICATIONS
Units of measure kg, g
Tare
Full
Interface
RS-232 bi-directional Interface
Stabilisation Time
Operating Temperature
Power supply
-10°C to 40°C / 14°F to 104°F
2 Seconds typical
230 VAC 50/60 Hz 115 VAC available
Battery Internal rechargeable battery (~90 hours operation)
*Calibration *Not permitted
6 digits LCD digital display
Display
with capacity tracker and symbols for units
Scale Housing ABS Plastic, Stainless Steel platform
Pan Size 210 x 300mm
8.3” x 11.8”
Overall Dimensions (wxdxh)
Net Weight
Applications
315 x 355 x 110 mm
12.4” x 14” x 4.3”
4.1 kg / 9 lb
Weighing Scales
Functions
Check counting, Accumulation of weights.
Real Time Clock (RTC),
Weighing, Parts counting, % weighing, Check weighing,
Date/Time
To print date and time information- battery backed
NOTE: For use in approved applications the scales do not
come with the RS-232 interface.
© Adam Equipment Company 2017 5
Key descriptions
[0-9]
Numeric entry keys to set the unit price.
[CE]
Clears an erroneous entry or return to normal operation.
[Tare] or key Sets the scale to display zero weight when an empty container is placed on the platform or removed again. Also works as an [Enter] key
[MR]
The memory key shows the accumulation number and total weight
[SETUP]
Brings up the setup menu for changing date, time, power etc.
[Zero] or key It sets the zero when the platform is empty and no tare value is entered. Also works as an
[Esc] key
[Print M+]
It sends the results of the current displays to the RS-232 interface.
[SMPL]
This is used to enter the number of items of a sample
[Units]
This key is used to enter the weight of a sample manually. It will also change the weighing units when other units are enabled.
[High Limit]
Used to set the ‘high’ limit when checkweighing
[Low Limit]
Used to set the ‘low’ limit when checkweighing
[Lim] It stores and recalls any of 10 preset limits
[Func]This is used to switch between weighing functions e.g parts counting, weighing or percentage weighing. Also used to select RS-232 parameters.
© Adam Equipment Company 2017 6
3.0 OPERATION
3.1 SETTING UP AND TURN ON THE SCALE
Follow this set up when using the scale for first time:
1) Place the platform in the locating holes on the top cover.
2) Level the scale by adjusting the four feet. Ensure the bubble in the spirit level is in the centre of the level and the scale is supported by all four feet.
3) Attach the power cable to the connector located on the base to the left of the scale.
4) Plug the power cable into the mains.
5) Press the power button located on scale keypad.
6) The scale will then display the software revision and run a self-test.
7) Once the self-test has been completed and a stable condition achieved, the scale will display zero weight alongside stable and zero symbol indicators.
Battery compartment
USB
Mains connection
connectivity port
3.2 ZEROING THE DISPLAY
You can press the [Zero]/0
Esc
key at any time to set the zero point. When the zero point
is obtained, the display will show the indicator for zero.
The scale has a re-zeroing function to account for minor drifting or accumulation of material on the platform. Press [Zero]/0
Esc
to re-zero the scale if small amounts of weight are still
shown when the platform is empty.
© Adam Equipment Company 2017 7
Loading...