A&D HL-300WP, HL-3000LWP User Manual

Digital Scale
HL-WP Series
1WMPD4000754B
HL-1000WP HL-300WP HL-3000WP HL-3000LWP
SAFETY PRECAUTIONS
All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Institute:
Product Safety Signs and Labels). The meanings are as follows:
WARNING
CAUTION
This manual is subject to change without notice at any time to improve the
product.
Product specifications are subject to change without any obligation on the
part of the manufacture.
When using the HL-WP, the following safety precautions should always be
followed.
WARNING Use only the specified AC adapter for the scale. Other
adapters may cause damage (The AC adapter is optional).
A potentially hazardous situation which, if not avoided, could result in death or serious injury. A potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
Internal service or adjustments to this product should be performed by a qualified person.
CAUTION Avoid installing the scale in direct sunlight, which may cause
discoloration or malfunctions.
Do not mix the type of batteries. Replace all the batteries at the same time.
If the scale is not to be used for a long period of time, remove all batteries from the battery compartment to avoid leakage.
Avoid overloading that could cause damage to the scale.
Avoid using the weighing pan to move the scale, as that could cause damage to the scale.
When the scale is used with an AC adapter, the scale does not conform to IP-65 protection.
IP65
Degree of protection against water: Protected against water jets.
Degree of protection against solid foreign objects: Dust-tight.
IP Code (Degree of protection provided by enclosures) of IEC 60529
1
PARTS DESCRIPTION
t
ON/OFF key
Turns the scale power on/off.
RE-ZERO key
Sets display to zero or subtracts weight of container.
UNITS key
(U.S.A. models only) Changes weighing units.
Battery cover
Display
Weighing Platform
Spirit level
Four leveling feet
Rubber cap
AC adapter jack
(Batteries are not included.)
STABLE indicator
Indicates when the reading is stable.
NET indicator
Indicates when NET weigh is displayed. (Tare function is used.)
ZERO indicator
Indicates when the scale zero is correct.
Polarity
01200.0kg
0-1.000k
04-76.3k
04l12.3k
lb
oz
oz
Battery cover lock
Battery compartment
Weighing unit, g
Weighing unit, lb
Weighing unit, oz
U.S.A. models only
Weighing unit, lb-oz
2
PREPARATION
Installing Batteries / Connecting The AC adapter
Installing Batteries
CAUTION When replacing batteries, be careful not to apply too much force to the weighing pan. An excessive force could damage the scale.
Remove the battery cover and insert six batteries (R6P / LR6 / AA size) into the battery compartment, taking extreme care to that the polarities plus and minus are observed. Be sure to lock the battery cover again after installing the batteries.
AC adapter (optional)
New six batteries (R6P/ LR6/ AA size)
Close the rubber cap
Battery cover lock
Optional AC Adapter
CAUTION
When using an AC adapter, the scale does not comply with the IP65.
The AC input requirement could be 100, 120, 220, 230 or 240 volts (50/60Hz) depending on the area where used, so please verify that the adapter is correct.
Plug the AC adapter to the AC adapter socket inside the battery compartment.
Attach the battery cover and fit the rubber cap.
AC Adapter jack
Setting up the HL-WP
CAUTION
Avoid installing the scale in direct sunlight, as that may cause discoloration or a malfunction. Place your HL-WP on a firm weighing table so that the scale is level. (The scale will not perform accurately when it is not level.)
Place the scale on the firm surface and adjust the feet so that the sprit level shows the scale is level for an accurate weighing.
3
WEIGHING
Before weighing
Automatic Power-Off Function
If the scale is left ON and the stable indicator is displayed, the automatic power-off function turns the power off after approximately 5 minutes (factory setting). To toggle the function disabled/enabled, press the [ON/OFF] key while pressing the [RE-ZERO] key. The scale stores the setting and it will be shown by pressing both the [RE-ZERO] and [UNITS] keys in the weighing mode. (See “CALIBRATION” about the [UNITS] key.)
poff0e Function disabled. poff1e Function enabled.
Error message
eeeee Overload
----- The scale zero is out of range ealba Low battery
Calibration may be required before weighing
Read “CALIBRATION” first and if necessary, calibrate your scale for accurate weighing.
Weighing procedures
1. Turn on the scale.
Press the [ON/OFF] key to turn on the scale. Then the scale displays all segments for a few seconds and then 0 will be displayed.
2. Start weighing
If you do not use a container for weighing,
Verify the reading is 0 . If not, press the [RE-ZERO] key to display 0 . Place objects on the weighing platform to weigh. When the reading
becomes stable, the stable indicator is displayed.
If you use a container for weighing, Place an empty container on the platform. Wait for the stability indicator to be displayed and press the [RE-ZERO] key. Place the objects to be weighed in the container.
Select the weighing unit, --- Only for U.S.A. models ---
If you use the U.S.A. model, press the [UNITS] key to select weighing unit. Normally scales power up in “lb” or “oz” when shipped from the factory. To power up in other units, turn the scale off. Press and hold the [UNITS] key, then press the [ON/OFF].key The scale will start with the next unit. Note) HL-300WP: ozÆgÆozÆ
HL-1000WP/3000WP/3000LWP: lbÆozÆlb-ozÆgÆlbÆ
4
CALIBRATION
When to calibrate
Calibration may be required when the scale initially installed, if it is moved a substantial distance, or in accordance with local regulations. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur.
The position of the [UNITS] key for the Calibration
The “g” model has no [UNITS] key, but the key used to enter calibration is located under the overlay.
Calibration using a calibration weight
Calibrate the scale with calibration weights at full capacity.
Location of the [UNITS] key
HL-1000WP : 1000g ± 0.1g HL-300WP: 300g ± 0.02g HL-3000WP : 3000g ± 0.2g HL-3000LWP : 3000g ± 0.2g
1. Enter calibration mode.
Press the [ON/OFF] key to turn the power off. While pressing both the [RE-ZERO] key and the [UNITS] key, press the [ON/OFF] key. Cal will be displayed.
2. Zero calibration
Press the [RE-ZERO] key to calibrate zero. Cal 0 will be displayed. Wait for the stable indicator to be displayed and press the [RE-ZERO] key. Calaf will be displayed in a few seconds. If only the zero calibration is to be done, press the [UNITS] key. The scale will display end and automatically return to the weighing mode.
3. Span calibration
Place the weight at the center of the platform for accurate calibration. Wait for the stable indicator to be displayed and then press the [RE-ZERO] key. The display will show end and the scale will automatically return to the weighing mode.
5
Gravity compensation
If you have calibration weights at the capacity of the scale, you can calibrate the scale with the weights instead of compensating for acceleration due to gravity.
The scale was calibrated at 9.798 m/sec to gravity at your location is not same, compensate the scale by adjusting the acceleration due to gravity. (Refer “The Value of Gravity at Various Locations” and “World Map” and find your location value.)
1. Enter calibration mode.
Press the [ON/OFF] key to turn the power off. While pressing both the [RE-ZERO] key and the [UNITS] key, press the [ON/OFF] key. Cal will be displayed.
2. Select a digit to be changed.
Press the [UNITS] key. The previously set value, for example 979,8 , will be displayed. A digit after the comma “,” will be changed by pressing the [RE-ZERO] key. Press the [UNITS] key to move the comma “,” and select a digit to be changed.
2
, at the factory. If the acceleration due
3. Set the new value
Change the value of the gravity acceleration using the [RE-ZERO] and [UNITS] keys.
4. Save the value in memory.
Press and hold the [UNITS] key until end is displayed. The scale will automatically return to the weighing mode.
6
SPECIFICATIONS
MODEL HL-1000WP HL-300WP HL-3000WP HL-3000LWP Capacity x Resolution
Non-linearity ±1g ±0.2g ±2g Repeatability 0.5g 0.1g 1g Span drift ±0.015% / °C TYP (5°C~35°C / 41°F~95°F) Operating temp. -10°C~40°C / 14°F~104°F, Less than 85%RH Display 13 mm / 0.51 inches, 7segment liquid crystal display Display update Approximately 10 times per second Power 6 x R6P / LR6 / “AA” size batteries or AC adapter Battery life Approximately 100 hours with manganese type
Platform size 128 (W) x 128 (D) mm
Dimensions
Weight (approx.) 870 g / 1.9 lb 950 g / 2.1 lb Calibration weight 1000g ± 0.1g 300g ± 0.02g 3000g±0.2g Accessories This manual Options AC adapter
1000 x 0.5 g 300 x 0.1 g 3000 x 1 g
2.2 x 0.001 lb 6.6 x 0.002 lb
35.28 x 0.02 oz 10.58 x 0.01 oz 105.8 x 0.1 oz
2 lb 3.3 x 0.1 oz 6 lb 9.8 x 0.1 oz
200 hours with alkaline cells at 20°C / 68°F
174(W)x137(D) mm
5.0 (W) x 5.0 (D) in.
170 (W) x 219 (D) x 63 (H) mm
6.7 (W) x 8.6 (D) x 2.5 (H) in.
6.8(W)x5.4(D) in.
174(W)x219(D)x63(H) mm
6.8(W)x8.6(D)x2.5(H) in.
The Value of Gravity at Various Locations
Amsterdam 9.813 m/s2 Manila 9.784 m/s2 Athens 9.807 m/s2 Melbourne 9.800 m/s2 Auckland NZ 9.799 m/s2 Mexico City 9.779 m/s2 Bangkok 9.783 m/s2 Milan 9.806 m/s2 Birmingham 9.813 m/s2 New York 9.802 m/s2 Brussels 9.811 m/s2 Oslo 9.819 m/s2 Buenos Aires 9.797 m/s2 Ottawa 9.806 m/s2 Calcutta 9.788 m/s2 Paris 9.809 m/s2 Cape Town 9.796 m/s2 Rio de Janeiro 9.788 m/s2 Chicago 9.803 m/s2 Rome 9.803 m/s2 Copenhagen 9.815 m/s2 San Francisco 9.800 m/s2 Cyprus 9.797 m/s2 Singapore 9.781 m/s2 Djakarta 9.781 m/s2 Stockholm 9.818 m/s2 Frankfurt 9.810 m/s2 Sydney 9.797 m/s2 Glasgow 9.816 m/s2 Taichung 9.789 m/s2 Havana 9.788 m/s2 Taiwan 9.788 m/s2 Helsinki 9.819 m/s2 Taipei 9.790 m/s2 Kuwait 9.793 m/s2 Tokyo 9.798 m/s2 Lisbon 9.801 m/s2 Vancouver, BC 9.809 m/s2 London (Greenwich) 9.812 m/s2 Washington DC 9.801 m/s2 Los Angeles 9.796 m/s2 Wellington NZ 9.803 m/s2 Madrid 9.800 m/s2 Zurich 9.807 m/s2