1-1. About This Manual .......................................................................................................................... 5
1-2. Features ........................................................................................................................................... 5
3-1. Before Use ..................................................................................................................................... 10
3-2. During Use ...................................................................................................................................... 11
3-3. After Use ........................................................................................................................................ 12
3-4. Power Supply ................................................................................................................................ 12
4. Display Symbols and Key Operation ................................................................................................ 13
7-2. Calibration Using the Internal Mass
7-3. Calibration Test Using the Internal Mass
7-4. Calibration Using an External Weight .......................................................................................... 20
7-5. Calibration Test Using an External Weight .................................................................................. 21
7-6. Correcting the Internal Mass Value 1 (Only for the FZ-CT Series)
7-7. Correcting the Internal Mass Value 2
8. Function Switch and Initialization ...................................................................................................... 24
8-1. Permit or Inhibit .............................................................................................................................. 24
8-2. Initializing the Balance ................................................................................................................... 26
9. Function Table .................................................................................................................................... 27
9-1. Structure and Sequence of the Function Table ........................................................................... 27
9-2. Display and Keys ........................................................................................................................... 27
9-3. Details of the Function Table ......................................................................................................... 28
9-4. Description of the Class "Environment, Display" ......................................................................... 32
9-5. Description of the Item "Data Output Mode" ............................................................................... 34
9-6. Description of the Item "Data Format".......................................................................................... 36
9-7. Data Format Examples ................................................................................................................. 38
9-8. Weighing Units ............................................................................................................................... 40
9-9. Storing Units ................................................................................................................................... 42
13. Password Function ............................................................................................................................ 62
13-1. Balance Software Version1.70 to 1.85 ......................................................................................... 62
13-2. Balance Software Version1.860 or later ...................................................................................... 62
13-3. Using the Password Function ...................................................................................................... 64
13-4. Entering the Password Before Weighing ..................................................................................... 65
13-5. Logging Out ................................................................................................................................... 67
13-6. Storing or Changing the Password .............................................................................................. 67
13-7. How to Change the Password ..................................................................................................... 68
20-3. Other Display ................................................................................................................................. 82
20-4. Asking for Repair ........................................................................................................................... 82
21. Options and Peripherals (Sold Separately) ...................................................................................... 83
24-2. Index ............................................................................................................................................... 90
4
Page 5
1.Introduction
This manual describes how the FZ-CT / FX-CT series balance works and how to get the most out of it
in terms of performance. Read this manual thoroughly before using the balance and keep it at hand
for future reference.
※ Operation may vary depending on the software version of the balance. To check the software version of
your balance, refer to “18. How to Check Software Version of Balance”.
1-1.About This Manual
This manual consists of the following five parts:
Basic operation
Adapting to the environment
Functions
Interface and communication
Maintenance
1-2. Features
.................................
........
...........................................
.......
......................................
Describes precautions on handling the balance, balance
construction and basic balance operation.
Describes response adjustment and calibration (sensitivity
adjustment).
Describes various functions of the balance.
Describes the interface which transmits data and controls the
balance.
Describes maintenance, error codes, troubleshooting,
specifications and options.
Compact general-purpose balance, can be installed almost anywhere.
Large breeze break, provided as a standard accessory for more accurate weighing.
The breeze break can be removed with one touch, so cleaning is simple.
Casing construction, strong protection against dust and moisture.
Carat pans for weighing jewelry, provided as standard accessories.
Multiple weighing units with most of the common units used around the world.
The default settings of the unit of weight are the metric carat (ct) and grams (g). The unit can be
switched by pressing the
Standard RS-232C serial interface to communicate with a computer and to output the Good
Laboratory Practice (GLP) / Good Manufacturing Practice (GMP) compliant report.
A built-in clock and calendar that can add the time and date to the output data. (Changing the clock
setting can be limited to the administrator. Refer to “15. Password Function”.)
The password function can be used to restrict balance users or changes to the function settings.The key lock function disables the balance key operation and enables the balance to be operated
only by commands from external devices.
Statistical calculation mode to statistically calculate the weight data, and display or output the sum,
maximum, minimum, range (maximum-minimum), average, standard deviation and coefficient of
variation.
MODE key.
Comparator Indicators, displaying the comparison results.
Underhook for below weighing, provided as a standard accessory.
Hold Function, provided for weighing a moving object such as an animal.
As options, the USB interface (FXi-02), the Ethernet interface (FXi-08) and the built-in battery unit
(FXi-09) are available.
Connecting the USB interface, by cable to a Windows computer, allows transmission of the weight
data to Excel or Word. The Ethernet interface allows connection of the balance to a personal computer
through a LAN. The battery unit allows use of the balance where a power source is not available.
5
Page 6
The AD-8920A remote display, that is sold separately, allows reading the weighing result in a place
away from the balance. It receives power supply directly from the balance instead of the AC adapter
when connected via RS-232C, so the cable arrangement is simplified.
1-3. Compliance
Compliancewith FCC rules
Please note that this device generates, uses and can radiate radio frequency energy. This device has
been tested and has been found to comply with the limits of a Class A computing device pursuant to
Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against
interference when this device is operated in a commercial environment. If this unit is operated in a
residential area, it may cause some interference and under these circumstances the user would be
required to take, at his own expense, whatever measures are necessary to eliminate the interference.
(FCC = Federal Communications Commission in the U.S.A.)
6
Page 7
2. Unpacking the Balance
2-1. Unpacking
The balance is a precision instrument. Unpack the balance carefully. Keep the packing material to be
used for transporting the balance in the future.
See the illustrations to confirm that everything is included.
Install the large breeze break when drafts are present
Top door
Large breeze break
Antistatic coating
Bubble spirit level
Keys
Display
AC adapter
AC adapter ID label
Leveling foot
Weighing pan
Pan support
Breeze break ring
Dust plate
Silver and black colored, one each
How to install the large breeze break
Follow the numbered sequence.
Main unit
Carat pan
Pull out the locking handles, and install the
large breeze break on the main unit.
Push in the locking handles.
Confirm that both of the locking handle
hooks are under the dust plate.
How to open the top door
If removing either of the securing screws
(3) on the top door, the top door may be
pivoted on the remaining screw.
RS-232C serial interface
Main unit rear side
The AC adapter plug is
Grounding terminal
Main unit cover
AX-FXi-31
protected against dust and
may be difficult to insert.
When inserting the plug,
turn the plug while pushing
on it.
AC adapter plug
AC adapter jack
7
Page 8
Notes
Please confirm that the AC adapter type is correct for your local voltage and receptacle type.
Please use the dedicated AC adapter specified for the balance.
Do not use the AC adapter provided with the balance for other models or equipment with which the AC
adapter may not be compatible.
If you use the wrong AC adapter, the balance and other equipment may not operate properly.
8
Page 9
2-2.Installing the Balance
Install the balance as follows:
1. Place the balance on a solid weighing table. Refer to "3. Precautions" for installing the balance.
2. Assemble the pan support, weighing pan, breeze break ring and breeze break, on the balance as
shown in the illustration on page 7.
3. Adjust the leveling feet to level the balance. Confirm it using the bubble spirit level.
4. Confirm that the AC adapter type is correct for the local voltage and power receptacle type.
5. Connect the AC adapter to the balance. Warm up the balance for at least one hour with nothing on the
weighing pan.
Note
The AC adapter plug is protected against dust and may be difficult to insert. When inserting
the plug, turn the plug while pushing on it.
9
Page 10
3.Precautions
To get the optimum performance from the balance and acquire accurate weight data, note the
following:
3-1.Before Use
Install the balance in an environment where the temperature and humidity are not excessive. The best
operating temperature is about 20°C / 68°F at about 50% relative humidity.
Install the balance where it is not exposed to direct sunlight and it is not affected by heaters or air
conditioners.
Install the balance where it is free of dust.
Install the balance away from equipment which produces magnetic fields.
Install the balance in a stable place avoiding vibration and shock. Corners of rooms on the floor at the
ground level are best, as they are less prone to vibration.
The weighing table should be solid, free from vibration and drafts and as level as possible. (An
anti-vibration table or stone table is ideal)
Level the balance by adjusting the leveling feet and check it using the bubble spirit level.
Red circle
Bubble
OK NG
Bubble spirit level
Leveling foot
Make sure that the bubble is in the center of the red circle.
UP
DOWN
DOWN
Leveling foot
UP
When the bubble is off to the left :
Turn the leveling foot on the front right in the
clockwise direction.
When the bubble is off to the backward position :
Turn both leveling feet on the front in the
clockwise direction at the same time.
Ensure a stable power source when using the AC adapter.
Plug in the AC adapter and warm up the balance for at least one hour.
Calibrate the balance before use or after having moved it to another location. Refer to “7. Calibration
(Sensitivity adjustment/Calibration)”.
Caution: Do not install the balance where flammable or corrosive gas is present.
When the bubble is off to the right :
Turn the leveling foot on the front left in the
clockwise direction.
When the bubble is off to the forward position:
Turn both leveling feet on the front in the
counterclockwise direction at the same time.
10
Page 11
3-2.During Use
Discharge static electricity from the material to be weighed
(hereinafter referred to as the sample). When a sample could
have a static charge, the weight data is influenced. Ground
the balance and try the following:
Eliminate static electricity using the optional AD-1683 DC
static eliminator.
Try to keep the ambient humidity above 45%RH.
Use a metal shield case for a charged sample.
Wipe charged plastic samples with a moistened cloth.
This balance uses a strong magnet as part of the balance
assembly, so please use caution when weighing magnetic
materials such as iron. If there is a problem, use the underhook
on the bottom of the balance to suspend the material away
from the influence of the magnet.
Eliminate the temperature difference between a sample and
the environment. When a sample is warmer (cooler) than the
ambient temperature, the sample will be lighter (heavier) than
the true weight. This error is due to a rising (falling) draft
around the sample.
Charged material
Material case
Grounding
Magnetic material
Make each weighing gently and quickly to avoid errors caused
by the evaporation of moisture from the sample or the
absorption of moisture by the sample. Use a pair of long
tweezers to place a sample on the pan to avoid placing a hand
inside the breeze break.
Do not drop things upon the weighing pan, or place a sample
on the pan that is beyond the balance weighing capacity.
Place a sample in the center of the weighing pan.
Do not use a sharp instrument such as a pencil to press the
keys. Use your finger only.
Press the RE-ZERO key before each weighing to eliminate possible errors.
Calibrate the balance periodically so as to eliminate possible errors.
Take into consideration the affect of air buoyancy on a sample when more accuracy is required.
Keep the balance interior free of dust and foreign materials.
The breeze break (The FZ-CT / FX-CT series) is provided as a standard accessory. An anti-static
treatment has been applied to the breeze break components, but they may be charged with static
electricity when they are unpacked or when the humidity is low. If the weighing value is unstable even
when there are no drafts or the balance has a problem with repeatability, remove the breeze break. Or
wipe the clear plates with a moistened cloth, use the optional AD-1683 DC static eliminator or apply an
anti-static spray.
11
Page 12
3-3.After Use
Avoid mechanical shock to the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.
Avoid dust and water so that the balance weighs correctly. Protect the internal parts from liquid spills
and excessive dust.
Do not use organic solvents or chemical cloth to clean the balance.
Clean the balance as described below:
Balance main unit,
Weighing pan
Breeze break
(standard accessory)
Use a soft, lint free cloth that is moistened with a mild detergent to clean.
The edge of the weighing pan is sharp. Use caution when cleaning the pan.
An antistatic treatment has been applied to the breeze break components.
Use a soft, dry, lint free cloth to clean. Cleaning with a cloth that is moistened
with water or a mild detergent repetitively, or washing with water, may
degrade the antistatic effect.
3-4.Power Supply
When the AC adapter is connected, the balance is in the standby mode if the standby indicator is on
(refer to "4. Display Symbols and Key Operation"). This is a normal state and does not harm the
balance. For accurate weighing, allow the balance to reach equilibrium with room temperature, and
then warm up the balance for at least one hour before use.
12
Page 13
4. Display Symbols and Key Operation
Display symbols
Displays the weight data relative to the weighing capacity, in percentage, in
Key operation
Key operation affects how the balance functions. The basic key operations are:
"Press and release the key immediately" or "Press the key" ....................................................
= normal key operation during measurement
"Press and hold the key" .............................................................................................................
Each key, when pressed or when pressed and held, functions as follows:
the weighing mode (Capacity indicator)
Number of statistical data (Statistical calculation mode)
Displays the setting value in the function mode
Processing indicator
Stabilization indicator
Standby indicator
Processing indicator
Blinking
Response indicators
Comparator indicators
Hold mode indicator
Units
Interval output mode
standby indicator
Weight data
Internal output mode
active indicator
Key When pressed When pressed and held (for 2 seconds)
Turns the display on or off. The standby indicator is displayed when the display is turned off.
The weighing mode is enabled when the display is turned on. When the display is turned on,
the password input (
mode is enabled if the password function is not used. Refer to "13-4.Entering the Password Before Weighing".
This key is available anytime. Pressing the key during operation will interrupt the operation and
turn the display off.
In the weighing mode, turns the minimum
weighing value on or off.
In the counting or percent mode, enters the
sample storing mode.
Switches the weighing units stored in the function
table. Refer to "9-8. Weighing Units".
Cancels the operation when performing
settings. With the FZ-CT series, enters the
calibration mode using the internal mass.
(One-touch calibration)
Outputs the weight data to a printer or personal
computer using the RS-232C serial interface,
depending on the function table settings.
Confirms the operation when performing
settings.
) is displayed if the password function is used. The weighing
Enters the function table mode. Refer to "9. Function Table".
Enters the response adjustment mode.
Enters the calibration mode (calibration using
an external weight).
No function at the factory setting
By changing the function table:
Outputs "Title block" and "End block" for
GLP/GMP compliant report. (Refer to "10-2.
GLP Report".)
Sets the display to zero.
13
Page 14
5.Weighing
The default settings of the unit of weight are the metric carat () and grams ( ).
Pressing the MODE key switches the displayed unit, between and .
To use other units, refer to “9-9. Storing Units” on page 37 to store the units.
5-1.Basic Operation (Gram Mode)
1 Place a container (Carat pan) on the weighing pan, if necessary.
Press the RE-ZERO key to cancel the weight (tare). The balance
displays 0.000 .
2 Place a sample on the pan or in the container.
3 Wait for the stabilization indicator to turn on. Read the weighing
value.
While the stabilization indicator is on, pressing the PRINT key
will output the weighing value, using the RS-232C serial interface.
4 Remove the sample and container from the pan.
Note
Peripheral equipment, that is sold separately , such as a
printer or a personal computer is required.
When other units of weight are stored as described in
“9-9. Storing Units”, pressing the MODE key switches the
displayed unit.
Press the SAMPLE key to turn on or off the minimum
weighing value.
When the ON:OFF key is pressed with a container placed on
the weighing pan, the balance displays 0.000 and
weighing is started,
Weighing pan
Carat pan
Sample
Remove the sample
Display off
Weighing pan
Container
14
Page 15
Zero operation, tare operation and weighing range
When weighing is started
The balance will determine the reference zero point when the ON:OFF key is pressed to enter the
weighing mode.
Depending on the load condition at that time, the balance will automatically judge whether to perform zero or
tare operation. The condition for determining which is used is "power on zero range", and when power on
zero range is exceeded, the tare operation is performed.
Re-zero operation
By pressing the RE-ZERO key, the display can be set to zero.
The re-zero operation with the RE-ZERO key will automatically judge whether to perform zero or tare
operation.
The condition for determining which is used is "zero range", and when zero range is exceeded, the tare
operation is performed.
Weighing range
The range that the balance can weigh varies depending on the model.
When the total amount (net weight + tare weight) displayed for each model exceeds the maximum
display, E is displayed to indicate that the weighing range is exceeded.
When exceeded in the negative direction, -E is displayed.
Model
FZ-1200CT, FX-1200CT
FZ-700CT, FX-700CT
FZ-500CT, FX-500CT
Power on
zero range
Zero range -E display range
Approx. ±125 ct Approx. ±25 ct Approx. -100 ct
Approx. ±25 g Approx. ±5 g Approx. -20g or less
Approx. ±75 ct Approx. ±15 ct Approx. -100 ct
Approx. ±15 g Approx. ±3 g Approx. -20 g or less
Approx. ±50 ct Approx. ±10 ct Approx. -100 ct
Approx. ±10 g Approx. ±2 g Approx. -20 g or less
15
Page 16
6.Response Adjustment
This function stabilizes the weighing value, reducing the influence on weighing that is caused by drafts
and vibration at the place where the balance is installed.
The function has three stages as follows and can be changed by simple key operation.
Indicator Parameter Response Stability
FAST Fast response, Sensitive value
MID.
SLOW
Slow response, Stable value
Operation
1 Press and hold the MODE key until RESPONSE is
displayed. And then, release the key.
2 Select a setting for the response adjustment by pressing
the MODE key. Either FAST , MID. or SLOW can
be selected.
3 After waiting for a moment, or when the PRINT key is pressed,
the balance displays eend . Then, it returns to the weighing
mode and displays the updated response indicator. The
response indicator remains displayed for a while.
Note
Setting the response rate automatically changes the
"Condition (Cond)" and "Display refresh rate (5pd)"
parameters of "Environment, Display (ba5fnc)" in the
Since the balance's resolution is high, weighing values may change due to gravity and daily environmental
changes. It is necessary to perform calibration (sensitivity adjustment) with the weight in order to keep the weighing
values from changing even if gravity or the environment changes.
Calibrate the balance if it is installed for the first time or relocated, or when the weighing values change significantly
in daily inspection.
Calibration means to adjust the weighing value of the balance using the reference weight or internal mass.
Calibration test means to weigh with the reference weight and compare how much the result deviates from the
reference value. (Calibration test does not perform adjustment.)
Calibration (Sensitivity adjustment)
Calibration using the internal mass ------------------- Using the internal mass, adjust the balance with a single
touch. (FZ-CT series)
Calibration using an external weight ----------------- Using an external weight, adjust the balance.
Calibration test (Sensitivity check)
Calibration test using an external weight ----------- Checks the accuracy of weighing using an external weight
(target weight) and outputs the result.
* No adjustment is made.
Notes
When calibrating, be sure to install the provided large breeze break.
Do not allow vibration, drafts, or temperature change to affect the balance during calibration.
Calibration adjusts the balance for accurate weighing.
Besides periodic calibration and before each use, perform calibration when:
The balance is installed for the first time.
The balance has been moved.
The ambient environment has changed.
To output the GLP/GMP compliant report using the RS-232C serial interface, set "GLP output ( info )"
of "Data output ( dout )". For details, refer to "9. Function Table". To display the GLP output, a printer
or personal computer is required.
Calibration test is available only when "GLP output ( info )" of "Data output ( dout )" is set to "1", "2" or
"3".
Note on using an external weight
The accuracy of an external weight will influence the accuracy of weighing. Select an appropriate
-0.0150 g to +0.0150 g FZ-700CT/ FX-700CT 150 g, 100 g, 50 g, 20 g
FZ-500CT/ FX-500CT 100 g, 50 g, 20 g
The calibration weight in bold type: factory setting.
The weight value can be adjusted within the range above.
Display
This indicator means "the balance is measuring calibration data".
Do not allow vibration or drafts to affect the balance while this indicator is
displayed.
17
Page 18
7-2.Calibration Using the Internal Mass
This function calibrates the balance using the internal mass. The only operation required is to press
the CAL key. (The FX-CT series does not have this function.)
Notes
Level the balance by adjusting the leveling feet and confirm that the bubble is in the center of
the spirit level.
Poor level adjustment may cause a calibration error while using the internal mass.
(One-Touch Calibration, Only for the FZ-CT Series)
Operation
1 Connect the AC adapter and warm up the balance for at least one hour with nothing on the weighing
pan.
2 Press the CAL key.
3 The balance displays Calin and automatically performs calibration using the internal mass. Do not
allow vibration or drafts to affect the balance.
4 The balance displays end after calibration. If the "GLP output (info)" parameter of the function
table is set, the balance displays glp and outputs the calibration report using the RS-232C
interface. Refer to “GLP output (info)” in "10-2. GLP Report".
5 The balance automatically returns to the weighing mode after calibration.
About the internal mass
The internal mass may change due to corrosion or other damage caused by the operating
environment, or due to aging. The internal mass value can be corrected using an external weight. To
maintain the weighing accuracy, perform the internal mass value correction periodically. Refer to "7-4.
Correcting The Internal Mass Value 1" and "7-7. Correcting The Internal Mass Value 2".
18
Page 19
7-3.Calibration Test Using the Internal Mass
This function tests the balance accuracy using the internal mass.
Calibration test does not perform adjustment.
This is available only when the "GLP output (info)" parameter is set to "1" or "2".
(Only for the FZ-CT Series)
Operation
1 Connect the AC adapter and warm up the balance at least one hour
with nothing on the pan.
2 Refer to "9. Function Table" to set "GLP output (info)"
to "1" or "2".
3 Press and hold the CAL key until CC in is
displayed, then release the key.
4 The balance measures the zero point.
Prevent vibration and drafts to affect the balance.
5 The measured zero point data is displayed.
6 The balance measures the internal mass.
Prevent vibration and drafts to affect the balance.
7 The value of the internal mass is displayed.
The normal range of the value is as follows:
If the displayed value is within the normal range, it indicates
that the calibration sensitivity is correctly adjusted by the
internal mass.
Press and hold
Release
Model Internal mass Normal range
FZ-1200CT 200.000 g
±0.002 g FZ-700CT
100.000 g
FZ-500CT
8 If GLP output is set, glp is displayed and the
calibration test report is output using the RS-232C
interface. Refer to "GLP output (info)" of the function
table, "10-2. GLP Report".
end is displayed after the calibration.
9 The balance automatically returns to the weighing mode.
GLP output
19
Page 20
7-4.Calibration Using an External Weight
This function calibrates the balance using an external weight.
Operation
1 Connect the AC adapter and warm up the balance
for at least one hour with nothing on the pan.
2 Press and hold the CAL key until Cal out is
displayed, and then release the key.
3 The balance displays Cal 0 .
If you want to change the calibration weight (a list of usable
weights is shown on page 17), press the SAMPLE key
and proceed to step 4.
If you use the calibration weight value stored in the
balance, proceed to step 5.
4 Specify the calibration weight value as follows:
SAMPLE key To switch between the calibration weight
selection mode (all of the segments
blinking) and the weight value adjustment
mode (the last three digits blinking).
RE-ZERO (+) / MODE (-) keys
To select the calibration weight to use or
adjust the value. (Refer to page 17.)
PRINT key To store the new calibration weight value.
Even if the AC adapter is removed, the data
is maintained in non-volatile memory.
CAL key To cancel the operation and return to Cal 0 .
5 Confirm that there is nothing on the pan and press the PRINT key.
The balance measures the zero point. Do not allow vibration or
drafts to affect the balance. The balance displays the calibration
weight value.
Place the weight
Press and hold
Release
Release
Select a weight
Example:
New value
50.0120g
6 Place the displayed calibration weight on the pan and press
the PRINT key. The balance measures the calibration weight.
Do not allow vibration or drafts to affect the balance.
7 The balance displays end . Remove the weight from the pan.
8 When the GLP output is set, the balance displays glp and
outputs the calibration report.
9 The balance automatically returns to the weighing mode.
Remove
10 Press the MODE key to select the gram mode. Place the
calibration weight on the pan and confirm that the value
GLP output
displayed is within ±0.002 g of the calibration weight value. If it
is not within the range, check the ambient conditions such as
drafts and vibration, also check the weighing pan. Then, repeat
steps 1 to 9.
20
Page 21
S
R
7-5.Calibration Test Using an External Weight
This function tests the balance weighing accuracy using an external weight and outputs the result.
This is available only when the “GLP output (info)” parameter is set to "1" , "2" or "3". (Calibration test
does not perform calibration.) .
Operation
1 Connect the AC adapter and warm up the balance
for at least one hour with nothing on the pan.
Press and hold
2 Press and hold the CAL key until CC out is
displayed, and then release the key.
3 The balance displays CC 0 .
If you want to change the calibration weight (a list of usable
weights is shown on page 17), press the SAMPLE key
and proceed to step 4.
If you use the calibration weight value stored in the
balance, proceed to step 5.
4 Specify the calibration weight value as follows:
SAMPLE key To switch between the calibration weight
selection mode (all of the segments
blinking) and the weight value adjustment
mode (the last three digits blinking).
RE-ZERO (+) / MODE (-) keys
To select the calibration weight to use or
adjust the value. (Refer to page 17.)
PRINT key To store the new calibration weight value.
Even if the AC adapter is removed, the
data is maintained in non-volatile memory.
CAL key To cancel the operation and return to CC 0 .
Release
Release
elect a weight
Example:
New value
50.0120g
5 Confirm that there is nothing on the pan and press
the PRINT key. The balance measures the zero
point and displays the measured value. Do not allow
vibration or drafts to affect the balance. The balance
displays the target weight value.
6 Place the displayed target weight on the pan and press
the
PRINT key. The balance measures the target weight
and displays the measured value. Do not allow vibration
or drafts to affect the balance.
7 The balance displays end . Remove the weight from
the pan.
8 When the GLP output is set, the balance displays glp
and outputs “Calibration Test Report”.
9 The balance automatically returns to the weighing mode.
Target weight
emove
GLP output
21
Page 22
P
r
7-6.Correcting the Internal Mass Value 1
The FZ-CT series balance can correct the internal mass value using an external weight.
Method 1: Calibrate the balance using the internal mass. Measure the external weight.
Calculate the correction value. And store it in the balance.
The adjustable range of the internal mass value is as follows:
Model Correction Reference Value Adjustable Range
FZ-1200CT 200 g
FZ-500CT
100 g
(Only for the FZ-CT Series)
-0.0150 g to +0.0150 g FZ-700CT
Operation
Example : Correcting the weight value by +0.0030 g in 200 g using the FZ-1200CT.
If correcting the weight value by +0.0030 g in 100 g, use the correction value of +0.0060 g as
the correction reference value is 200 g.
200.0000g 200.0030g
1 Perform calibration using the internal mass (one-touch calibration).
Place the external weight on the weighing pan to obtain the correction value.
Correct the internal mass by +0.0030g in 200g.
Calibrate with the corrected internal mass.
The same mass
Corrected mass value
2 With the factory setting, the balance cannot correct the internal mass
value. Refer to "8. Function Switch And Initialization" and set the
internal mass value correction switch to 1.
3 In the weighing mode, press and hold the SAMPLE key to
display ba5fnc .
4 Press the SAMPLE key several times until C5 in 1 is displayed.
If C5 in 1 is not displayed, perform step 2.
5 Press the PRINT key.
Correct the internal mass value using the following keys:
RE-ZERO key
MODE key
PRINT key
CAL key
6 Press the CAL key to return to the weighing mode.
7 Press the CAL key to perform calibration using the internal mass.
8 Confirm that the internal mass value has been corrected properly by
placing the external weight on the weighing pan.
If the internal mass value has not been corrected properly, repeat the
procedure to adjust the correction value.
...
........
........
...........
To increase the value by one.
To decrease the value by one.
To store the correction value and
display the next menu item of the
function table.
To cancel the operation and return to the
weighing mode.
The balance displays
-0.0150 g after +0.0150 g.
Press and hold
ess several
times
22
Page 23
A
7-7.Correcting the Internal Mass Value 2
The FZ-CT series balance can correct the internal mass value using an external weight.
Method 2: Calibrate the balance using the external weight. The balance performs automatic
calibration using the internal mass, corrects the internal mass value and stores it in the
balance. The corrected mass value is maintained in non-volatile memory even if the AC
adapter is removed.
The usable calibration weights and the adjustable range are as follows:
Model Correction Reference Value Adjustable Range
FZ-1200CT 250 g, 200 g, 100 g, 50 g, 20 g
FZ-500CT 100 g, 50 g, 20 g
The calibration weight in bold type: factory setting.
The calibration weight value can be adjusted within the range above.
(Only for the FZ-CT Series)
-0.0150 g to +0.0150 g FZ-700CT 150 g, 100 g, 50 g, 20 g
Operation
1 With the factory setting, the balance cannot correct the internal mass
value. Refer to "8. Function Switch And Initialization" and set the
internal mass value correction switch to 1.
Press and hold
2 In the weighing mode, press and hold the SAMPLE key to
display ba5fnc .
3 Press the SAMPLE key several times until C5 in 21 is
displayed. If C5 in 2 is not displayed, perform step 1.
4 Press the PRINT key to display Cal 0 .
Refer to "7-4. Calibration Using an External Weight", to
perform the calibration.
5 After calibration, remove the weight.
The balance displays . Press the PRINT key.
6 Cal. is displayed and the balance corrects the internal
mass value automatically
7 After correcting the internal mass value, the balance
displays Cal.
automatically using the corrected internal mass value.
8 The balance displays end and returns to the weighing
mode.
9 Confirm that the internal mass value has been
corrected properly by placing the external weight
used for the correction procedure on the
weighing pan.
If the internal mass value has not been corrected
properly, repeat the procedure to adjust the
correction value.
in
and performs the calibration
djusting the internal mass value
Calibrating with internal mass
Press several times
Calibration using an
external weight
Place the weight
Remove the weight
23
Page 24
8. Function Switch and Initialization
8-1. Permit or Inhibit
The balance stores parameters that must not be changed unintentionally. There are two switches for
the purpose of protecting these parameters. Each switch can select either “permit” or “inhibit”.
“Inhibit” protects parameters against unintentional operations.
Function selection can be restricted with the password function. When using balances with software
version 1.860 or later, the password function can be used to restrict calibration (sensitivity adjustment)
functions.
The switches for function selection" are as follows:
FZ-CT series FX-CT series
Function switch
- Function table
- Calibration using the internal mass
- Calibration using an external weight
- Internal mass correction
- Function table
- Calibration using an external weight
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key.
The balance displays p5 .
3 Press the PRINT key. Then the balance displays the function switches.
4 Set the switches using the following keys.
SAMPLE key
RE-ZERO key
PRINT key
CAL key
....
...
........
...........
To select the switch to change the parameter. The selected switch blinks.
To change the parameter of the switch selected.
0 : To inhibit changes. 1 : To permit changes
To store the new parameter and return to the weighing mode.
To cancel the operation ( Clr is displayed.)
To return to the weighing mode, press the CAL key once again.
24
Page 25
Switches (FZ-CT series)
(The display shown left indicates the factory settings)
Function table
To inhibit changes on the function table
To permit changes on the function table
Calibration using the internal mass (One-touch calibration)
Software version 1.860 or later
Software version 1.43 to 1.85
To inhibit calibration using the internal mass
To inhibit calibration using the internal mass
1
*
To inhibit calibration using the internal mass when a user
() or a guest () is logged in.
To permit calibration using the internal mass
Calibration using an external weight
Software version 1.43 to 1.85
Software version 1.860 or later
To inhibit calibration using an external weight
To inhibit calibration using an external weight
1
*
To inhibit calibration using the external mass when a user
() or a guest () is logged in.
To permit calibration using an external weight
Not used
Internal mass value correction
To inhibit correction
To permit correction
* 1 The operation is permitted when logged in as an administrator ().
Switches (FX-CT series)
(The display shown left indicates the factory settings)
Function table
To inhibit changes on the function table
To permit changes on the function table
Not used
Calibration using an external weight
To inhibit calibration using an external weight
To permit calibration using an external weight
Not Used
Not Used
Software version 1.43 to 1.85
Software version 1.860 or later
1
*
To inhibit calibration using the external mass when a user
To inhibit calibration using an external weight
() or a guest () is logged in.
25
Page 26
8-2.Initializing the Balance
This function restores the following parameters to factory settings.
Calibration data
Function table
Sample unit mass value (counting mode), 100% reference mass value (percent mode)
External weight value
Function switch settings
Statistical data
Internal mass value (Only for the FZ-CT series)
Note
Be sure to calibrate the balance after initialization.
Operation
1 Press the ON:OFF key to turn off the display.
While pressing and holding the PRINT key and the SAMPLE key,
press the ON:OFF key. The balance displays p5 .
2 Press the SAMPLE key to display Clr .
3 Press the PRINT key to displayClr no .
(To cancel the operation, press the CAL key.)
4 Press the RE-ZERO key to displayClr go .
5 Press the PRINT key to initialize the balance.
The balance will automatically return to the weighing mode.
Press and hold Press
26
Page 27
9.Function Table
The function table reads or rewrites the parameters that are stored in the balance. These parameters
stored, even if the AC adapter is removed, are maintained in non-volatile memory.
9-1.Structure and Sequence of the Function Table
The function table menu consists of two layers. The first layer is the "Class" and the second layer is the
"Item". Each item stores a parameter.
Example
This example sets "Auto print mode A ()" for "Data output mode ()" and "10 digits ()"
for "Auto print difference ()".
Start, From weighing mode
Press several times
Class
Press
End, To weighing mode
Press and hold
Item
Press one time
Press one time
Press two times
Press two times
Item : "Data output mode"
Parameter : "Auto print mode A"
Item : "Auto print difference"
Parameter : "10 digits"
Press one time
9-2. Display and Keys
Display/Key Description
The symbol indicates that the parameter displayed is in effect.
When pressed and held (for approx. 2 seconds) in the weighing mode, enters the
function table mode.
Selects the class or item in the function table mode.
Changes the parameter.
When a class is displayed, moves to an item in the class.
When an item is displayed, stores the new parameter and displays the next class.
When an item is displayed, cancels the new parameter and displays the next class.
When a class is displayed, exits the function table mode and returns to the weighing
mode.
27
Page 28
9-3.Details of the Function Table
Class Item and Parameter Description
0 Fast response, sensitive value
1
2 Slow response, stable value
0 Stable range is ±1 digit
1
2 Stable range is ±3 digits
0 OFF
1 ON
The stabilization indicator illuminates when
the display fluctuation is within the range.
With "
Holds the display when stable in animal
weighing mode. With "
turns on.
0 OFF
1 Normal
2 Strong
Keeps zero display by tracking zero drift.
3 Very strong
0 Approx. 5 times/second
1
Approx.
10 times/second
0 Point (.)
1 Comma (,)
0 OFF
1 ON
0 OFF
1 ON (10 minutes)
0 Displays
Does not display
1
0 Does not sound
1 Sounds
Refer to "9-10. Clock and
calendar function".
Period to refresh the display
Decimal point format
Turns on the weighing mode display when
the AC adapter is connected.
Turns off the display after 10 minutes of
inactivity.
Select whether or not to sound the beep
when operating on keys.
Confirms and sets the time and date. The
time and date are added to output data.
0 No comparison
1
Comparison, excluding "near zero" when stable value or overloaded
2
Comparison, including "near zero" when stable value or overloaded
Number of data, sum, maximum, minimum, average, range
1
(maximum-minimum)
Number of data, sum, maximum, minimum, average, range
2
(maximum-minimum), standard deviation, coefficient of variation
Number of data, sum, maximum, minimum, average, range
3
(maximum-minimum), standard deviation, coefficient of variation,
relative error
0
OFF
locfnc
Password
function
pass
1
*
Password function
ON (limited weighing
1
operation)
ON (basic weighing is
2
Refer to “11. Password function”
possible)
ADMINPassword input by the administrato
loc no.
Password
U5ER 01
U5ER 10
C5in1 *
Internal mass value correction: Method 1
C5 in 2 *
Internal mass value correction: Method 2
Password input by user 1
Password input by user 10
Refer to "7. Calibration".
2
*
Displayed only when the function
switch (internal mass value
correction switch) is set to 1.
: Factory setting Digit is a unit of minimum weighing value
: Only fortheFZ-CT series
*
1
: With the software version 1.70 to 1.85, only OFF (
*
) / ON ( ) can be selected.
If the software version is 1.860 or later, OFF ( ) / ON ( ) / ON ( ) can be selected.
2
: Available with the software version 1.860 or later.
*
If the software version is 1.70 to 1.85, one password (equivalent to the administrator password)
can be registered.
Caution
The balance may not transmit the data completely at the specified refresh rate, depending on the
baud rate.
31
Page 32
9-4. Description of the Class "Environment, Display"
Condition ( Cond )
Cond0 This parameter is for sensitive response to the fluctuation of a weighing value. Used for
powder target weighing, weighing a very light sample or when quick response weighing is
required. After setting, the balance displays FAST .
Cond2 This parameter is for stable weighing with slow response. Used to prevent a weighing value
from drifting due to vibration or drafts. After setting, the balance displays SLOW .
Note
With "Hold function (Hold)" set to "ON (1)", this item is used to set the averaging time.
Stability band width ( 5t-b )
This item controls the width to regard a weighing value as a stable value. When the fluctuation per
second is less than the parameter, the balance displays the stabilization indicator and outputs the data.
The parameter influences the "Auto print mode".
5t-b 0 This parameter is for sensitive response of the stabilization indicator. Used for exact weighing.
5t-b 2 This parameter ignores slight fluctuation of a weighing value. Used to prevent a weighing
value from drifting due to vibration or drafts.
Note
With "Hold function (Hold)" set to "ON (1)", this item is used to set the stabilization range.
Hold function ( Hold ) (Animal weighing mode)
This function is used to weigh a moving object such as an animal. When the weight data is 0.0200 g or
greater and the display fluctuation is within the stabilization range for a fixed period of averaging time,
the processing indicator illuminates and the balance displays the average weight of the animal. When
the animal is removed from the weighing pan, the display returns to zero automatically.
This function is available only when the hold function parameter is set to "1" (the animal mode
indicator
The stabilization range and averaging time are set in "Condition (Cond)" and "Stability band width
(5t-b)".
HOLD illuminates) and any weighing unit other than the counting mode is selected.
Weighing Range Averaging Time Stabilization Range
Cond 0
0.0200 g or greater
Cond 1
Cond 2
2 seconds Faster
4 seconds
8 seconds More accurate
5t-b 0
5t-b 1
5t-b 2
Lesser 6.25%
12.5%
Greater 16.7%
Zero tracking ( trc )
This function tracks zero point drift caused by changes in the environment and stabilizes the zero point.
When the weight data is only a few digits, turn the function off for accurate weighing.
Note
Digit indicates a unit of minimum weighing value.
trc 0 The tracking function is not used. Used for weighing a very light sample.
trc 1 The tracking function is used. Normal zero tracking.
trc 2 The tracking function is used. Strong zero tracking.
trc 3 The tracking function is used. Very strong zero tracking.
32
Page 33
Display refresh rate ( 5pd )
Period to refresh the display. This parameter influences "Stream mode" along with "Baud rate" and
"Data output pause".
Note
This item is selected automatically when the response rate is changed.
Decimal point ( pnt )
The decimal point format can be selected.
Auto display-on ( p-on )
When the AC adapter is connected, the display is automatically turned on in the weighing mode,
without the ON:OFF key operation. Used when the balance is built into an automated system. One
hour warm up is necessary for accurate weighing.
Auto display-off ( poff )
This function automatically turns off the balance display, if the weighing value is 0.5 g or less and the
balance is in the no key operation or no command has been received for approx. 10 minutes.
(The balance has the same power requirement in the weighing state as in the off state {standby
mode}.)
33
Page 34
9-5.Description of the Item "Data Output Mode"
The parameter setting of the "Data output mode ( prt )" applies to the performance when the data is
transmitted using the RS-232C serial interface.
Key mode
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
weighing data and the display blinks one time.
Required setting doutprt0 Key mode
Auto print modes A and B
When the displayed value is stable and the conditions of "Auto print polarity", "Auto print difference"
and reference value are met, the balance outputs the weighing data.
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
data and the display blinks one time.
Mode A: Example "For outputting the weighing data each time a sample is added (or removed),
with "ar-d" set to "1" (to adjust zero after the data is output)."
Required setting dout prt1Auto print mode A (reference = zero)
dout ap-pAuto print polarity
dout ap-bAuto print difference
dout ar-d1 Adjusts zero automatically.
Mode B:
Example "For outputting the weighing data while a sample is added."
Required setting dout prt2Auto print mode B (reference = last stable value)
dout ap-pAuto print polarity
dout ap-bAuto print difference
Stream mode
Whether the stabilization indictor is displayed or not, the balance outputs the weighing data at the
specified display refresh rate. The display does not blink in this mode.
Example "For monitoring data on the AD-8920A remote display
Required setting dout prt3Stream mode
ba5fnc 5pdDisplay refresh rate
5if bp5Baud rate
Caution
The balance may not transmit the data completely at the specified display refresh rate and baud rate.
Set the baud rate higher.
Key mode B
When the PRINT key is pressed, whether the stabilization indictor is displayed or not, the balance
outputs the weighing data. The display does not blink in this mode.
Required setting doutprt4 Key mode B
34
Page 35
Key mode C
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
weighing value. When the stabilization indicator is not on, the balance waits for the indicator to turn on
and outputs the weighing value. The balance blinks one time in this mode.
Required setting doutprt5 Key mode C
Interval output mode
Whether the stabilization indictor is displayed or not, the balance outputs the weighing value
periodically and the display blinks one time.
At that time the displayed weighing value will blink once to indicate that it had been output (or stored).
When the PRINT key is pressed, the balance starts to output the weighing value at a preset
interval time. When the PRINT key is pressed again, the balance stops the output.
Example "For outputting the weighing value periodically."
Required setting dout prt6Interval output mode
dout intInterval time
Caution
The balance may not transmit the data completely at the specified interval times and baud rate. Set
the baud rate higher.
35
Page 36
t
a
e
e
a
r
r
)
t
r
a
r
r
)
9-6. Description of the Item "Data Format"
A&D standard format 5if type 0
This format is used when the peripheral equipment can receive the A&D format. With an AD-8127
printer, use the external key printing mode, manual printing mode, auto printing mode or interval
printing mode. With an AD-8121B printer, use MODE 1 or 2.
This format consists of fifteen characters excluding the terminator.
A header of two characters indicates the balance condition.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is
used.
The unit, consisting of three characters, follows the data.
Output example:
Print example:
Header Dat
Stable header
Unstable heade
Overload heade
Header for modes other than the counting mod
Header for the counting mod
DP (dump print) format 5if type 1
This format is suitable for the peripheral equipment that prints the received data as is.
With an AD-8127 printer, use the dump printing mode. With an AD-8121B printer, use MODE 3.
This format consists of sixteen characters excluding the terminator.
A header of two characters indicates the balance condition. No overload header is used.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not zero
or overloaded.
The unit, consisting of three characters, follows the data.
Heade
Dat
Stable header
Unstable heade
Uni
Stable header (Counting mode
Unit Header Dat
Stable header (Counting mode
Te r m in a to
Te r m in a tor
Uni
Te r m in a tor
36
Page 37
KF format 5if type 2
This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only
communicate using this format.
This format consists of fourteen characters excluding the terminator.
This format has no header characters.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not zero
or overloaded.
This format outputs the unit only for a stable value.
+ 1.278 ct
DataUnitTerminator
CRL
F
MT format 5if type 3
This format is used when the peripheral equipment of other manufacturer is connected. Please note
that the connection is not guaranteed.
A header of two characters indicates the balance condition.
The polarity sign is used only for negative data.
The weight data uses spaces in place of the leading zeros.
The character length of this format changes dependent upon the unit
S 1.278ct
HeaderUnit Terminator
Data
CRL
F
S
SD
SI
Stable header
Unstable header
Overload header
NU (numerical) format
5if type4
This format outputs only numerical data.
This format consists of nine characters excluding the terminator.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is
used.
+012. 78
00
Data
CRL
F
Terminator
CSV format 5if type 5
Separates the data of A&D standard format and the unit by a comma (,).
Outputs the unit even when the data is overloaded.
ST,+0001.278, c t<CR><LF>
Weight data
S T , + 0 0 0 1 . 2 7 8 , c t
OL,+9999999E+19, c t
When "Pound Ounce" is selected, the data is output with the unit of ounce (oz).
39
Page 40
e
C
9-8.Weighing Units
With the FZ-CT / FX-CT series balance, the following weighing units and weighing modes are
available:
Percent mode
Counting mode
A unit or mode can be selected and stored in the function table as described on page 13.
If a weighing mode (or unit of weight) has been turned off, that mode or unit will be missing in the
sequence. Tael has four varieties, one of which can be selected and installed at the factory.
To select a unit or mode for weighing, press the
For details about the units and modes, see the table below:
Programmable-unit (No unit displayed. For details,
Tael (HK general, Singapore)
Tael (HK jewelry) 37.429 g
Tael (Taiwan) 37.5 g
Tael (China) 31.25 g
Tola (India)
Messghal
Programmable-unit (Multi-unit)
Conversion factor
1 g =
28.349523125 g
31.1034768 g
0.2 g
3.75 g
1.55517384 g
0.06479891 g
37.7994 g
11.6638038 g
4.6875 g
40
Page 41
The tables below indicate the weighing capacity and the minimum display for each unit, depending on
the balance model.
FZ-500CT
FX-500CT
Minimum Display
Unit
FZ-1200CT
FX-1200CT
Capacity
FZ-700CT
FX-700CT
Gram 252 152 102 0.001
Ounce (Avoir)
8.89 5.36 3.59
Pound 0.555 0.335 0.224 0.000005
Pound/Ounce 0L 8.88 oz 0L 5.36 oz 0L 3.60 oz 0L 0.01 oz
Troy Ounce
Metric Carat
Momme
Pennyweight
Grain (UK)
Tael (HK general, Singapore)
Tael (HK jewelry)
Tael (Taiwan)
Tael (China)
Tola (India)
Messghal
8.10 4.88 3.27 0.00005
1260 760 510 0.001
67.2 40.5 27.2 0.001
162.0 97.7 65.5 0.001
3888 2345 1574 0.02
6.66 4.02 2.69 0.00005
6.73 4.06 2.72 0.00005
6.72 4.05 2.72 0.00005
8.06 4.86 3.26 0.00005
21.6 13.0 8.74 0.0001
53.7 32.4 21.7 0.0005
0.00005
41
Page 42
9-9.Storing Units
The units or modes can be selected and stored in the function table. The sequence of displaying the
units or modes can be arranged in the function table so as to fit the frequency of use.
Select a unit or mode and arrange the sequence of display as follows:
1 Press and hold the
table is displayed.
SAMPLE key until ba5fnc of the function
2 Press the
SAMPLE key several times to display Unit .
3 Press the PRINT key to enter the unit selection mode.
4 Specify a unit or mode in the order to be displayed using the
Displays the units
following keys.
SAMPLE key ........ To sequentially display the units.
RE-ZERO key ....... To specify a unit or mode.
The stabilization indicator appears when the displayed unit or
Specify
mode is specified.
5 Press the
PRINT key to store the units or modes. The balance
displays end and then displays the next menu item of the
function table.
6 Press the
CAL key to exit the function table. Then the
balance returns to the weighing mode with the unit specified
first in step 4.
Notes
When the power is turned on, the scale displays the unit specified first in step 4.
In the weighing mode, to select a unit or mode for weighing, press the MODE key.
Select
sequentially.
Store
42
Page 43
C
s
9-10.Counting Mode (PCS)
This is the mode to determine the number of objects in a sample based on the standard sample unit
mass. The unit mass means the mass of one piece of the sample. The smaller the variables in each
sample unit mass are, the more accurate the counting will be. The balance is equipped with the
Automatic Counting Accuracy Improvement (ACAI) function to improve the counting accuracy.
Notes
If the sample unit mass variable is too large, it may cause a counting error. To improve the counting performance, use the ACAI function frequently or divide the samples
into several groups and count each group.
Selecting the Counting Mode
1 Press the MODE key to select (counting mode).
Storing a Sample Unit Mass
2 Press the SAMPLE key to enter the sample unit mass storing
mode.
Even in the storing mode, pressing the
to the next mode.
MODE key will switch
3 To select the number of samples, press the
SAMPLE key
several times. It may be set to 10, 25, 50 or 100.
Note
A greater number of samples will yield more accurate
Weighing pan
ontainer
counting result.
4 Place a container on the weighing pan, if necessary.
Press the
The number specified in step 3 appears.
e.g.:
RE-ZERO key to cancel the weight (tare).
25 0 is displayed if 25 is selected in step 3.
Sample
5 Place the number of samples specified on the pan.
In this example, 25 pieces.
6 Wait for the stabilization indicator to turn on. Press the
PRINT key to calculate and store the unit mass. The
balance displays (counting mode) and is set to
Place sample
count samples with this unit mass. To improve the
accuracy of the unit mass, go to step 8.
The counting result
Notes
If the balance judges that the mass of the samples is too light and is not adequate to be used
as the unit mass, it displays lo . In that case, store the mass by some quantity. For
example, when the model with the minimum weighing value of 0.001 g is used and 10 pieces
of samples weigh 0.005 g. Store 100 pieces of samples as 10 and multiply the weighing result
by 10.
If the balance judges that the mass of the samples is too light to aquire accurate weighing, it
displays an error requiring the addition of more samples to the specified number. For example,
150- appears to require 25 more samples. Add 25 samples and press the PRINT key.
When the unit mass is stored correctly, the balance proceeds to the counting mode.
For a more accurate counting operation, use samples with a unit mass of at least ten or more
times the minimum weighing value of the balance. For example, when a balance with a
25
to be
43
Page 44
mimimum weighing value of 0.001 g is used, the recommended sample unit mass is 0.01 g or
more.
The sample unit mass stored, even if the AC adapter is removed, is maintained in non-volatile
memory.
Counting operation
7 Place the samples to be counted on the pan.
While the stabilization indicator is on, pressing the
output the weighing value (number of objects), using the RS-232C
serial interface.
Note
A printer, a personal computer, or peripheral equipment that is sold
separately , is required.
Counting mode using the ACAI function
The ACAI is a function that improves the accuracy of the unit mass
automatically by increasing the number of samples and averaging
the unit mass variable to minimize the weighing error, as the
counting process proceeds.
8 If a few more samples are added, the processing indicator turns on. To prevent an error, add three or
more. The processing indicator does not turn on if overloaded. Try to add the same number of
samples as displayed.
PRINT key will
Processing indicator
9 The balance re-calculates the unit mass while the processing indicator is blinking. Do not touch the
balance or samples on the pan until the processing indicator turns off.
10 Counting accuracy is improved when the processing indicator turns off.
Each time the above operation is performed, a more accurate unit mass will be obtained. There is no
definite upper limit of ACAI range for the number of samples exceeding 100. Try to add the same
number of samples as displayed.
11 Remove all the samples used in ACAI and proceed with the counting operation using the improved
unit mass.
44
Page 45
C
e
e
9-11.Percent Mode (%)
This is the mode to display the weighing value in percentage compared to a 100% reference mass
and is used for target weighing or checking the sample variable.
Selecting the percent mode
1 Press the MODE key to select % (percent mode).
Note
If the percent mode is not displayed, refer to " 9-8. Weighing Units"
and “9-9. Storing Units” and enable the percent mode.
Storing the 100% reference mass
2 Press the SAMPLE key to enter the 100% reference mass
storing mode.
Even in the storing mode, pressing the
the next mode.
3 Place a container on the weighing pan, if necessary. Press the
RE-ZERO
1000 % .
key to cancel the weight (tare). The balance displays
MODE key will switch to
Weighing pan
ontainer
4 Place the sample to be set as the 100% reference mass on the
pan or in the container.
Sampl
5 Press the
PRINT key to store the reference mass. The balance
displays 100.00 % . (The decimal point position depends on the
reference value.)
Notes
If the balance judges that the mass of the sample is too light to be
used as a reference, it displays
The position of the decimal point varies with 100% reference mass.
100% reference mass Minimum weighing value
lo . Do not use the sample.
Place sampl
to be weighed
0.0100 g to 0.0999 g 1 %
0.1000 g to 0.9999 g 0.1 %
1.0000 g ~ 0.01 %
The reference mass is stored and maintained in non-volatile
memory even if the AC adapter is removed.
6 Remove the sample.
Weighing result compared to
100% reference mass
Reading the percentage
7 Place a sample to be compared to the reference mass on the pan. The displayed percentage is
based on the 100% reference mass.
While the stabilization indicator is on, pressing the
using the RS-232C serial interface.
Note
Peripheral equipment, that is sold separately , such as a printer or a personal computer is required.
PRINT key will output the weighing value,
100%
45
Page 46
9-12. Description of the Item "Application Function"
Capacity indicator ( apf 1)
In the weighing mode, the indicator displays the weight data relative to the weighing capacity in
percentage. (Zero =
Statistical calculation mode ( apf2)
The mode statistically calculates the weight data, and displays or outputs the results. For details, refer
to "6-5 Statistical Calculation Mode".
0%, maximum capacity = 100%)
46
Page 47
9-13.Statistical Calculation Mode
The statistical calculation mode statistically calculates the weight data, and displays or outputs the
results. To use the statistical calculation mode, set the "Application function (
"Application (
ap fnc)" in the function table to "2", as described below.
Statistical items available are number of data, sum, maximum, minimum, range (maximum-minimum),
average, standard deviation, coefficient of variation and relative error. What statistical items to output
can be selected from the four modes in the function table.
The wrong data input can be canceled by the key operation, if immediately after the input.
Turning the balance off will delete the statistical data.
The standard deviation, coefficient of variation and relative error are obtained by the equation below:
Standard deviation = where Xi is the i-th weight data, N is number of data.
Coefficient of variation (CV)= x 100 (%)
N・Σ(Xi)2 - (ΣXi)2
N・(N - 1)
Standard deviation
Average
Relative error of maximum value = x 100 (%)
Relative error of minimum value =x 100 (%)
Maximum value - Average
Average
Maximum value - Average
Average
Note
When data with the minimum weighing value turned off are used in the statistical calculation,
results will be displayed with the minimum weighing value turned off. (The minimum weighing
value will be rounded off.)
9.13.1. Getting Started
apf)" parameter of
Switching to the statistical function mode (Changing the function table)
1 Press and hold the SAMPLE key until of
the function table is displayed, then release the key.
2 Press the
SAMPLE key several times to display
.
3 Press the
PRINT key to display
4 Press the
RE-ZERO key several times to display .
To select statistical items to output, go to step 5.
To store the statistical function mode setting, go to step 7.
To disable the statistical calculation mode, press the
RE-ZERO key to select .
To store the setting To select statistical
go to step 7. go to step5.
Press and hold
Press several times
Press several times
47
Page 48
Selecting the statistical items to output
5 Press the SAMPLE key to display .
6 Press the
In the example, is selected to output the number of data, sum,
maximum, minimum, range (maximum-minimum) and average.
Parameter Description
0 Number of data, Sum
1
2
3
RE-ZERO key to select the output items.
Number of data, Sum, Maximum, Minimum,
Range (maximum – minimum), average
Number of data, Sum, Maximum, Minimum,
Range (maximum – minimum), Average, Standard deviation,
Coefficient of variation
Number of data, Sum, Maximum, Minimum,
Range (maximum – minimum), Average, Standard deviation,
Coefficient of variation, Relative error of maximum value,
Relative error of minimum value
7 Press the PRINT key to store the setting.
Note
The statistical calculation mode will be enabled using the ON:OFF key or
when the power is turned on.
8 Press the CAL key to return to the weighing mode.
Selecting the unit
9 Press the MODE key to select the unit to be used for the statistical calculation
mode. In the example shown at the right, carat () is selected.
Notes
Selecting the unit using the MODE key is not available after the data is entered. In this case,
clear the data as described on page 50 and select the unit using the MODE key.
When the unit used for the statistical calculation mode is to be enabled upon power-on,
select the unit in "Unit (Unit)" of the function table beforehand.
9.13.2. Using the Statistical Calculation Mode
Entering data for statistical calculation
Use the following keys to operate the statistical calculation mode.
MODE key
When no data has been entered, selects the unit.
SAMPLE key ..... Turns the minimum weighing value on or off, in the weighing mode.
RE-ZERO key
PRINT key ........ Outputs the data number and the weight data and includes the weight data to
CAL key
........
...
............
When the data is entered, moves between the displaying items (weighing mode,
statistical results and data operation) each time the key is pressed.
Sets the display to zero in the weighing mode.
statistical calculation in the weighing mode. (Output is not in the data format
specified in the function table because of the data number added.)
Outputs the statistical results while the statistical results are displayed. (Output is
not in the data format specified in the function table.)
Returns to the weighing mode.
48
Page 49
1 Press the
RE-ZERO key to set the display to zero.
2 Place the sample on the weighing pan and wait for the stabilization
indicator to turn on.
3 Press the
PRINT key to add the data displayed to statistical
calculation. The number of data on the upper left of the display
Weighing pan
Sample
increases by 1.
4 Repeat steps 1 to 3 for each weighing.
Outputting the statistical results
5 Each time the MODE key is pressed, the display
changes: the results as selected in "Statistical function
mode output items (
5taf)", and .
6 While the results are displayed, press the PRINT key to
output the results.
Notes
When the number of data is 1, the coefficient of
variation is displayed as - - - - - - - .
When the average is 0, the coefficient of variation
is displayed as - - - - - - - .
Statistical items are indicated on the upper left of
9.13.4. Clearing The
the display using the following symbols.
Symbol Statistical item
5um Sum
9.13.3. Deleting The
max Maximum
min Minimum
r Range (Maximum – minimum)
ave Average
5d Standard deviation
Cv Coefficient of variation
max%Relative error of maximum value
min%Relative error of minimum value
Output the results
Statistical Data
Latest Data
Output example Function table
1
2
3
Displays the
statistical
results
selected in the
parameter
0
49
Page 50
9.13.3.Deleting the Latest Data
When the wrong data is entered, it can be deleted and excluded from
statistical calculation. Only the latest data can be deleted.
1 In the weighing mode, press the
MODE key several times to
display .
2 Press the
PRINT key to display .
3 Press the
RE-ZERO key to display .
4 Press the
PRINT key to delete the latest data and exclude it
from statistical calculation.
The number of data decreases by 1 when the balance returns
to the weighing mode.
9.13.4.Clearing the Statistical Data
Press several times
All the statistical data will be deleted and the number of
data will be 0 (zero).
1 In the weighing mode, press the
MODE key several
times to display .
2 Press the
PRINT key to display
.
3 Press the
RE-ZERO key to display .
4 Press the
PRINT key to delete the statistical data.
The number of data becomes 0 (zero) when the
balance returns to the weighing mode.
Press several times
50
Page 51
9-14.Comparator Function
The results of the comparison are indicated by HI OK LO on the display.
Operating conditions: No comparison
To compare, use:
Input method:
Note
"Near zero" means that the weighing value is within 10 digits of the minimum weighing value.
For example, using an FZ-1200CT in gram mode, the range of 0.010 CT is "near zero".
For the description of "Comparator (Cpfnc)", refer to "9-3 Details of the Function Table".
Setting example
(Continuous comparison, excluding "near zero", upper limit and lower limit values, digital input)
Selecting a comparator mode
1 Press and hold the
Comparison when the weight data is stable, excluding "near zero"
Comparison when the weight data is stable, including "near zero"
Continuous comparison, excluding "near zero"
Continuous comparison, including "near zero"
Upper limit value and lower limit value
Digital input
SAMPLE key until of the function table is displayed.
2 Press the SAMPLE key several times to display .
3 Press the PRINT key.
4 Press the RE-ZERO key several times to display .
5 Press the PRINT key to store the selected mode.
Entering the upper limit value
6 With displayed, press the PRINT key. The current setting of the upper limit value is
displayed with all the digits blinking.
When the current setting is not to be changed, press the PRINT or CAL key to go to step 7.
When the current setting is to be changed, press the RE-ZERO key. Change the setting using the
following keys.
SAMPLE key To select the digit to change the value. The selected digit blinks.
RE-ZERO key To change the value of the digit selected.
MODE key To switch the polarity.
PRINT key To store the new setting and go to step 7.
CAL key To cancel the new setting and go to step 7.
Entering the lower limit value
7 With displayed, press the PRINT key. The current setting of the lower limit value is
displayed with all the digits blinking.
When the current setting is not to be changed, press the PRINT or CAL key to go to step 8.
When the current setting is to be changed, press the RE-ZERO key. Change the setting using the
following keys.
SAMPLE key To select the digit to change the value. The selected digit blinks.
RE-ZERO key To change the value of the digit selected.
MODE key To switch the polarity.
PRINT key To store the new setting and go to step 8.
CAL key To cancel the new setting and go to step 8.
8 Press the
CAL key to exit the comparator function and return to the weighing mode.
51
Page 52
e
9-15. Clock and Calendar Function
(Only for the FZ-CT Series)
The FZ-CT series balance is equipped with a clock and calendar function. (This function is not
available for the FX-CT series)
When the "GLP output (info)" parameter is set to "1" or "2" and the "Time/Date output (5-td)"
parameter is set to "
1", "2" or "3", the time and date are added to the output data. Set or confirm the
time and date as follows:
Operation
1 Press and hold the SAMPLE key until ba5fnc of the function
table is displayed.
2 Press the
SAMPLE key several times to display Cl adj .
3 Press the PRINT key. The balance enters the mode to confirm
and set the time and date.
Confirming the date
4 The current date is displayed with all the digits blinking.
When the date is correct and the time does not need to be
confirmed, press the
When the date is correct and the time is to be confirmed,
press the
SAMPLE key and go to step 6.
When the date is not correct and is to be changed, press
the
RE-ZERO key and go to step 5.
Note
The year is expressed using a two-digit format. For example,
the year 2010 is expressed as "10".
Setting the date (with part of the digits blinking)
5 Set the date using the following keys.
SAMPLE key To select the digits to change the value. The
selected digits blink.
RE-ZERO key To increase the value by one.
CAL key and go to step 8.
To the time
setting mode
Stor
setting
the
To the time
setting mode
MODE key To decrease the value by one.
PRINT key To store the new setting, display end
and go to step 6.
CAL key To cancel the new setting and go to step 6.
Confirming the time
6 The current time is displayed with all the digits blinking.
When the time is correct and the date does not need to be
reconfirmed, press the
When the time is correct and the date is to be reconfirmed,
press the
SAMPLE key and go back to step 4.
When the time is not correct and is to be changed, press
the
RE-ZERO key and go to step 7.
52
CAL key and go to step 8.
Page 53
e
Setting the time (with part of the digits blinking)
7 Set the time in a 24-hour format using the following keys.
SAMPLE key To select the digits to change the value.
To the date
setting mode
The selected digits blink.
RE-ZERO key To increase the value by one.
MODE key To decrease the value by one.
PRINT key To store the new setting, display end
and go to step 8.
CAL key To cancel the new setting and go to step 4.
Stor
setting
the
Quitting the operation
8 The balance displays the next menu of the function table. Press
the
CAL key to exit the clock and calendar function and return
to the weighing mode.
Note
Do not enter invalid values such as a non-existing date
when setting the time and date.
When the clock backup battery has been depleted, the balance displays .
Under this condition, press any key and set the time and date. The dead battery only affects the clock
and calendar function. Even so, the function works normally as long as the AC adapter is connected
to the balance.
53
Page 54
10.ID Number and GLP Report
The ID number is used to identify the balance when Good Laboratory Practice (GLP) or Good
Manufacturing Practice (GMP) is used.
The ID number is maintained in non-volatile memory even if the AC adapter is removed.
The output format for GLP/GMP compliant report is selected at "GLP output (info)" of the function
table and can be output to a personal computer or printer using the RS-232C serial interface.
The GLP/GMP compliant report includes the balance manufacturer, model, serial number, ID number
and space for signature for weight data, and the weight used and results for calibration or calibration
test data.
The FZ-CT series is equipped with the built-in clock and can output the time and date using a clock
and calendar function.
With the FX-CT series, when an AD-8121B printer is connected to the balance, the time and date can
be printed using the clock and calendar function of the printer. In this case, set the "GLP output
(
info )" parameter to "1".
When an AD-8127 printer is connected to the balance and outputs the GLP report, the time and date
can be added to the report using the clock function of the printer. (Available for balances with software
version 1.660 or later)
This is very useful for managing the time and date using the AD-8127 password lock function to
prevent them from being falsified.
To output the GLP/GMP compliant report, set the AD-8127 printer to the dump printing mode.
The balance can output the following:
"Calibration report" of the calibration, using the internal mass (Calibration due to changes in
temperature and one-touch calibration.)
"Calibration report" of the calibration, using an external weight.
"Calibration test report" of the calibration test, using an external weight and using the internal mass.
"Title block" and "End block" for the weight data.
10-1.Setting the ID Number
1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed.
2 Press the SAMPLE key several times to display id .
3 Press the PRINT key. Set the ID number using the following keys.
SAMPLE key To select the digit to change the value. The selected digit blinks.
RE-ZERO key, MODE key To set the character of the digit selected.
Refer to the display character set shown below.
PRINT key To store the new ID number and display ap fnc
CAL key To cancel the new ID number and display ap fnc
4 Press the CAL key to return to the weighing mode.
Display character set
54
Page 55
10-2.GLP Report
To output the GLP/GMP compliant report to an AD-8121B printer, AD- 8127 printer or a personal
computer, set the "GLP output (
(General format using the balance built-in clock data) or "
clock data).
Notes on outputting the GLP/GMP compliant report to an AD-8121B/AD-8127 printer
Refer to "12. RS-232C Interface" for connection to a printer.
With an AD-8121B printer, use MODE 3. If MODE1 is used, select temporarily the dump print mode
by pressing the
STAT. key of the AD-8121B printer.
With an AD-8127 printer, use the dump printing mode. If the external key printing mode is used, press
and hold the
ENT key of the AD-8127 printer to switch between the external key printing mode and
the dump printing mode.
Set "Data output pause ()" of the function table to "1".
If the time and date are not correct when the FZ-CT series balance built-in clock data is output (info 1,
2), set the correct time and date in "Clock (Cl adj)" of the function table. With the FX-CT series, set the
correct time and date of the AD-8121B clock and calendar.
“info 3” is available for balances with software version 1.860 or later.
info)" parameter of the function table to "1" (AD-8121 format), "2"
3" (General format using the external device
55
Page 56
e
A
t
A
r
r
e
e
e
Calibration report using an internal mass (Only for the FZ-CT series)
When the setting is "info 1" When the setting is "info 2"
D-8121 format General data forma
Manufacturer
Model
Serial number
ID number
Dat
Time
Calibration type
Remarks
Signature
Space, ASCII 20h
Terminator, CR, LF or CR
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
Calibration test report using the internal mass (Only for the FZ-CT series)
(Calibration test does not perform sensitivity adjustment.)
When the setting is "
D-8121 format
info 1" When the setting is "info 2"
General data format
Manufacture
Model
Serial number
ID numbe
Dat
Tim
Calibration test typ
Zero point value
Target weight value
Target weight
Remarks
Signature
Space, ASCII 20h
Terminator, CR, LF or CR
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
56
Page 57
t
A
t
r
r
e
e
e
e
e
t
A
r
r
e
t
Calibration report using an external weight
When the setting is "info 1" When the setting is "info 2"
D-8121 format
Manufacture
Model
Serial number
ID numbe
Dat
Time
Calibration type
Calibration weigh
Remarks
Signature
Space, ASCII 20h
Terminator, CR, LF or CR
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
General data format
With FZ-CT series, the time and
date are added to output data.
Calibration test report using an external weight
(Calibration test does not perform calibration.)
When the setting is "
info 1" When the setting is "info 2"
D-8121 forma
Manufacture
Model
Serial number
ID numbe
Dat
Tim
Calibration test typ
Zero point valu
Target weight valu
Target weigh
Remarks
Signature
Space, ASCII 20h
Terminator, CR, LF or CR
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
General data forma
With FZ-CT series, the time and
date are added to output data.
57
Page 58
t
A
e
k
e
Title block and end block
When a weighing value is recorded as the GLP data, "Title block" and "End block" are inserted at the
beginning and at the end of a group of weighing values, in the GLP report.
Note
To output the report to an AD-8121B , use MODE 3 of the AD-8121B.
Operation
1 With the weight data displayed, press and hold the PRINT key until S5tart is displayed.
The "Title block" is output.
2 The weight data is output according to the parameter setting of the data output mode.
3 Press and hold the
DP format (
type 1 of the function table) DP format (type 1 of the function table)
When the setting is "
D-8121 format
PRINT key until recend is displayed. The "End block" is output.
info 1": When the setting is "info 2":
Title bloc
Manufacturer
Model
Serial number
ID number
Dat
General data forma
Tim
Weight data
Time
Remarks
Signature
End block
Space, ASCII 20h
Terminator, CR, LF or CR
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
With FZ-CT series, the time and
date are added to output data.
58
Page 59
Outputting the clock data of the external device “info 3” (with software version 1.860 or later)
When the "GLP output (info)" parameter of the function table is set to "3", the clock data of external
devices such as a printer or a personal computer can be used instead of the balance built-in clock
data for the GLP/GMP compliant report. This is useful for standardizing the clock data using the clock
function of an external device.
Except for the date and time, the same format as “
info 2” is used.
Notes
“info 3” is available for balances with software version 1.860 or later.
To output the clock data of an external device, the external device such as an AD-8127 printer
or Windows communication tools software WinCT RsCom must have a function to output the
time and date when <ESC>D or <ESC>T is received.
Even withinfo 3 set, the data memory function saves the balance built-in clock data for the
calibration report.
Calibration report using the internal mass
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Remark
A blank space to write
a note or memo.
Signature
A blank space
to sign.
Function setting "
info 3"
General format using the external device clock data
A&D
MODEL
S/N
ID
DATE
TIME
CALIBRATED(INT.)
REMARKS
SIGNATURE
- - - - - -
FZ-1200CT
012345678
ABCDEFG
2011/12/31
12:34:56
AD-8127 data
59
Page 60
11.Underhook
The underhook can be used for magnetic materials or density measurement. The built-in underhook is
revealed by removing the plastic cap on the bottom of the balance.
Use the underhook as shown below.
Caution
Do not apply excessive force to the underhook.
When not in use, attach the plastic cap to prevent dust from getting into the balance.
Do not push the underhook upward.
Bottom of the balance
Underhook
Cap
60
Page 61
12.Programmable-Unit
This is a programmable unit conversion function. It multiplies the weight data in grams by an arbitrary
coefficient set in the function table and displays the result.
The coefficient must be within the range between the minimum and maximum shown below. If the
coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient
setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory.
1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed.
2 Press the SAMPLE key several times to display j.
3 Press the PRINT key. The balance enters the mode to confirm or set the coefficient.
Confirming the coefficient
4 The current coefficient is displayed with the first digit blinking.
When it is not to be changed, press the CAL key and go to step 6. When it is to be changed, press the RE-ZERO key and go
to step 5.
Setting the coefficient
5 Set the coefficient using the following keys.
SAMPLE key .......... To select a digit to change the value.
The selected digit blinks.
0.000001 1000
Confirm
RE-ZERO key ......... To change the value.
MODE key ............. To change the decimal point position.
Each time the switch is pressed, the decimal
point position changes as follows:
PRINT key .............. To store the new setting, display end
and go to step 6.
CAL key ................. To cancel the new setting and go to step 6.
Quitting the operation
6 The balance displays Unit . Press the CAL key to exit the
programmable-unit function and return to the weighing mode.
0
.000001
.00001
00
.0001
000
.001
0000
.01
00000
000000
0000001
Decimal point
position using
MODE key
.1
Using the function
Press the MODE key to select the programmable-unit (no display on the unit section). Perform
weighing as described in "5-1. Basic Operation (Gram Mode)". After weighing, the balance displays
the result (weight data in grams x coefficient).
61
Page 62
13.Password Function
The password function can be used to restrict balance use or functions. It is useful for preventing the time
and date setting from being falsified and preventing users from changing other function settings.
Password
At factory setting, the password function is not used.
The management of the password function is performed in "
the Function Table".
Note
The password function varies depending on the balance software version.
13-1.Balance Software Version1.70 to 1.85
Two settings are available in “Password function () ” of the function table.
: Four digits (4 x 4 x 4 x 4 = 256 outcomes) using MODE , SAMPLE , PRINT and
RE-ZERO keys.
Password function" of "9-3. Details of
No password function.
Entering a password is required before starting weighing.
(No password function)
Anybody can perform weighing, use all the balance functions and make changes to them.
No password is required.
(Entering a password is required before starting weighing.)
The administrator (
Only one password by the administrator can be set.
The correct password must be entered for the balance to enter the weighing mode.
) sets a unique password to restrict balance users.
13-2.Balance Software Version1.860 or later
Three settings are available in “Password function () ” of the function table.
(No password function)
No password is required.
Anybody can perform weighing, use all the balance functions and make changes to them.
No password function.
Entering a password is required before starting weighing.
Logging in by the administrator password is required before making changes.
(Entering a password is required before starting weighing.)
The administrator (
(Pressing the
The correct password must be entered for the balance to enter the weighing mode.
Two login levels are available: administrator (
Administrator (
User (U5ER 01~10) Initialization and changes to the settings are restricted, including the clock setting.
No password The balance cannot be used.
62
) sets unique passwords to restrict balance users.
ON:OFF key before weighing will prompt you to enter a password.
) and user (U5ER 01~10).
All the functions and settings can be used.
)
A password can be set for each of 10 users.
Page 63
(Logging in by the administrator password is required before making changes.)
Anybody can perform weighing, but initialization and changes to the settings are restricted, including the clock setting.
(Even if the
Two login levels are available: administrator () and guest ().
ON:OFF key is pressed to start weighing, a password will not be required.)
Administrator (
Guest ()
*No password
When weighing is started by pressing the ON:OFF key while holding down the CAL key in the standby mode, the
balance will prompt you to enter the administrator () password.
) All the functions and settings can be used.
Initialization and changes to the settings are restricted, including the clock setting.
Items restricted depending on the login level
Weighing
Login level
Administrator
)
(
User
(U5ER 01 to 10)
Guest
)
(
*1: Response adjustment, repeatability test, function switch and initialization, function settings (clock setting)
*2: Available under ordinary circumstances, but not available to users (U5ER 01 to 10) and guests (
administrator (
Entering password
before weighing
Required
Available or not available
Available
) sets the function switch to “inhibit” shown in “’7-1. Permit or Inhibit”.
Calibration
(Sensitivity adjustment)
Available Available
Changes to the settings
2
*
Not available
1
*
) when the
63
Page 64
13-3.Using the Password Function
The password function can be switched among “Not used ( )”, “Used ( )” and “Used ( )” in “Password
function (
1 In the weighing mode, press and hold the
until is displayed, and then release the key.
2 Press the
3 Press the
(To cancel the operation, press the CAL key.)
4 Press the
Display “ ” or “ ”.
) ” of the function table.
SAMPLE key
SAMPLE key several times to display .
PRINT key to display “ ”.
RE-ZERO key to switch the numbers.
Press and hold
Press
several times
5 Press the
PRINT key to display with "No"
blinking ("No" is selected).
6 "YES" and "No" can be switched with the
RE-ZERO key.
Press the RE-ZERO key to display with
"YES" blinking ("YES" is selected).
7 Press the
PRINT key to store the new setting. The password
function can now be used.
8
is displayed. To store or change the password,
proceed to Step 4 in “13.6. Storing or Changing the Password”.
Not to store or change the password, press the
CAL key to
return to the weighing mode.
Note
When the balance software version is 1.70 to 1.85, only “Not
used ( )” or “Used ( )” is available.
No is
selected.
Switch between YES/No
YES is
selected.
Confirm
x 1
64
Page 65
o
t
13-4. Entering the Password Before Weighing
(Entering a password is required before starting weighing)
1 In the standby mode, press the
2 After , is displayed
to prompt you to enter a password.
3 Enter a four-digit password using the following keys.
The display will turn off automatically after ten minutes of
non-operation.
MODE key ....................... To enter
SAMPLE key.................... To enter
PRINT key ....................... To enter
RE-ZERO key ................ To enter
CAL key .......................... To go back one character
4 When the password is correct, the balance displays
the login level, turns all the display segments on,
and then enters the weighing mode.
ON:OFF key.
Enter four characters
No operation
for 10 minutes
The default password set at the factory is
(press the RE-ZERO key four times), for administrator
rrect
C
password
Incorrec
password
level.
When the password is incorrect, the balance
displays
turns the display off.
, sounds the buzzer three times and
Login level
Full segments
Weighing mode
Display off
Notes
When the balance software version is 1.70 to 1.85, the login level is not displayed when
logged in.
When the balance software version is 1.860 or later, enter the administrator password
to log in as an administrator.
65
Page 66
Press
characters
Nooperationfor
10m
inutes
ts
Weighingmode
f
Weighing mode
(Logging in by the administrator password is required before making changes)
Logging in as an administrator ()
1 In the standby mode, while holding down the CAL key,
press the ON:OFF key.
2 Enter a four-digit password using the following keys.
The display will turn off automatically after ten minutes
of non-operation.
MODE key ....................... To enter
SAMPLE key ................... To enter
PRINT key ....................... To enter
RE-ZERO key ................ To enter
CAL key .......................... To go back one character
3 When the password is correct, the balance displays he
login level, turns all the display segments on, and then
enters the weighing mode.
The default password set at the factory is
(press the RE-ZERO key four times), for
administrator level.
Whileholdingdown
Enterfour
When the password is incorrect, the balance
displays, sounds the buzzer three times
Correctpassword
and turns the display off.
Logging in as a guest ()
1 In the standby mode, press the ON:OFF key.
2 After , the balance enters the weighing
mode.
Incorrectpassword
Loginlevel
Fullsegmen
Displayof
66
Page 67
13-5.Logging Out
1 To log out, press the ON:OFF key to turn the display off.
0.0000
With the function set, the password will need to
be entered again to enter the weighing mode from the
standby mode.
Weighing mode
Display off
13-6.Storing or Changing the Password
The password can be set (changed) in "Password ()" of the function table.
1 In the weighing mode, press and hold the SAMPLE key
until is displayed, and then release the key.
2 Press the SAMPLE key several times to display .
Press and hold
(for 2 seconds)
Press several times
3
Press the PRINT key to display the login level ().
4 Press the SAMPLE key to display the login level
to be changed (/ to ).
The stabilization indicator
level where a password has been stored. (The
appears at a login
Login level
password can be changed.)
Press
several
times
5 To change the password, press the PRINT key.
For details, refer to “13.7. How to Change the Password”.
Notes
To log out, press the ON:OFF key to turn the display off.With the function set, the administrator password is required to log in as an
Refer to “13.7. How to Change
the Password”.
administrator (). No password needs to be stored for to .
When the balance software version is 1.70 to 1.85, only one password can be stored. While the
balance displays , press the PRINT key to display the current password.
67
Page 68
13-7.How to Change the Password
1 Refer to “13.6. Storing or Changing the Password” to
display the login level to be changed.
2 Press the PRINTkey to display the current password.
The default password set at the factory is (press
the RE-ZERO key four times), for administrator level.
3 Enter a new four-digit password using the following
keys. The display will turn off automatically after ten
minutes of non-operation.
MODE key ............................. To enter
SAMPLE key.......................... To enter
PRINT key ............................. To enter
RE-ZERO key ..................... To enter
CAL key ................................ To go back one character
CAL key (press and hold) ...... To delete the password
Refer to “13-8. Deleting the
Password (
to )”.
Back key
Enter the 4th
character
4 The new password is displayed.
5 is displayed with "No" blinking ("No" is
selected).
(Press the CAL key to return to the fourth character
entry position.)
6 Press the RE-ZERO key to display
with “YES” blinking (“YES” is selected).
Confirm
7 While “YES” is blinking, press the PRINT key to store
the new password.
8 When setting is complete, the next login level is
displayed.
To set a password for the next level, follow the
procedure starting from Step 2. To finish the operation,
press the
mode.
CAL key twice to return to the weighing
Notes
If you have forgotten the password, the balance cannot be used. Make a record of
the password and keep it where you can refer to it if needed.
The same password as the administrator () password cannot be used for the user
( to ).
68
Refer to “13.6. Storing or
Changing the Password”
Current password
No operation
for 10 minutes
Display off
Press twice
Page 69
13-8.Deleting the Password(
1 Refer to “13.7. How to Change the Password” to select the
user (
display the password.
to ) whose password to be deleted and
U5ER 01~10
)
Refer to “13.7. How to
Change the Password”
2 Press and hold the CAL key to display
3 Press the PRINT key to display
(blinking).
.
Press and hold
(for 2 seconds)
4 Press the RE-ZERO key to switch between "Go" and
"No".
5 While
is displayed, press the PRINT key to
delete the password. The balance displays and then
displays the login level.
Note
The administrator password cannot be deleted, but it
can be changed. Refer to “13.6. Storing or Changing
the Password” and “13.7. How to Change the
Password” to change it.
13-9.If the Password is lost or forgotten
If you have forgotten the password, the balance cannot be used.
Contact your local A&D dealer to undo the password.
69
Page 70
14.RS-232C Serial Interface
The balance is a Data Communication Equipment (DCE) device. Connect the balance to a personal
computer (DTE) using a straight through cable.
Transmission system : EIA RS-232C (D-Sub 9-pin, female connector)
Transmission form : Asynchronous, bi-directional, half duplex
Transmission rate : Approx. 10 or 5 times/second (same as data refresh rate)
Data format : Baud rate : 600, 1200, 2400, 4800, 9600, 19200 bps
Data bits : 7 or 8 bits
Parity : Even, Odd (Data bits 7 bits)
None (Data bits 8 bits)
Stop bit : 1 bit
Code : ASCII
Data bits Stop bit
Start bit Parity bit
D-Sub 9-Pin Assignments
Pin No. Signal name Direction Description
1 - -
2 TXD Output Transmit data
3 RXD Input Receive data
4 - - No connection
5 SG - Signal ground
6 DSR Output Data set ready
7 RTS Input Request to send
8 CTS Output Clear to send
9 - Output 12V output *1
Signal names of the balance side are the same as the
DTE side with TXD and RXD reversed.
*1 Used with some peripheral devices manufactured by A&D.
Do not wire them when connecting to other companies' products such as PCs and PLCs.
Using the wrong connection cable may damage the device. Be sure to check the compatible cable.
Same potential
as SG *
1
RS-232C
1
0
-5 V to -15 V
+5 V to +15 V
#4-40UNC
9876
54321
70
Page 71
A
A
A
A
15. Connection to Peripheral Equipment
15-1. Connection to the AD-8121B Printer
Set the following parameters to use the AD-8121B printer.
Example of use
To print A&D standard format weight data, using the FZ-CT
PRINT key or auto print mode. (The time and date can be added.)
/
FX-CT
To print A&D standard format weight data, using the AD-8121B DATA key
or AD-8121B built-in timer. (The time and date can be added.)
To print, using the AD-8121B chart printing function.
To print the FZ-CT
/
FX-CT statistical data. MODE 3
To print GLP output. MODE 3
Class Item and Parameter
prt
dout
Data output
Data output mode
pU5e
Data output pause
bp5
Baud rate
5if
Serial
interface
btpr
Data bit, parity bit
Crlf
Terminator
type
Data format
Factory
setting
0 0,1,2,4,5 *1 3 0,1,2,4,5 *1
0 0 0 0,1 *2
2 2 2 2
0 0 0 0
0 0 0 0
0 0 0 1
D-8121B
MODE 1
D-8121B
MODE 2
AD-8121B mode setting
MODE 1
MODE 2
D-8121B
MODE 3
*1 Set appropriate parameters for "ap-p (Auto print polarity)" and "ap-b (Auto print difference) "
when auto print mode A or B (
Set the AD-8121B DIP switch No.3 to ON when unstable data is printed with "
prt 1 or 2) is selected.
prt 4".
*2 Set 1 when multiple lines are printed in the FZ-CT / FX-CT statistical calculation mode.
Notes
Refer to "10-2. GLP Report" for print samples. Settings of AD-8121B DIP switches
MODE
MODE 1
MODE 2
MODE 3
DIP switch No.3 : Handling unstable data
ON Print
OFF Not printed
Set the DIP switch No.4 to OFF.
D-8121B
DIP switch
MODE
ON
1 2 3 4
MODE
ON
1 2 3 4
MODE
ON
1 2 3 4
Description
Print at receiving data.
Standard mode, statistical calculation mode
Print by the AD-8121B
Standard mode, interval mode, chart mode
DATA key operation or AD-8121B built-in timer.
Print at receiving data.
Dump print mode
MODE
ON
1 2 3 4
MODE
ON
1 2 3 4
71
Page 72
15-2.Connection to the AD-8127 Printer
When connecting the AD-8127 printer to the balance for printing weighing data, configure the
printer and the balance as follows according to the usage examples.
Function settings of the printer
Multi printer
Usage example
When printing the balance weighing
data with the balance’s
or its “Auto print” mode.
When printing the balance weighing
data with the printer’s “Printing” key or
its “TIMER ” mode.
When printing charts with the printer.
When printing the balance’s statistical
calculation results.
When printing the balance’s GLP report.
PRINT key
AD-8127
Printing mode settings
EXT.KEY
MANUAL
AUTO
TIMER
CHART
DUMP
Refer to the instruction manual of the AD-8127 printer for how to change the function
settings of the AD-8127 printer.
Function settings of the balance
Balance
Usage example
Data output mode
When printing the balance weighing
data with the balance’s
or its “Auto print” mode.
When printing the balance weighing
data with the printer’s “Printing” key or
its “TIMER ” mode.
When printing charts with the printer.
When printing the balance’s statistical
calculation results.
When printing the balance’s GLP report.
Refer to “8. Function Table” for how to change the function settings of the balance.
PRINT key
0, 1, 2, 4, 5, 6 1
Data format
0, 1, 2, 4, 5 0
3, 6 0
Balance
72
Page 73
15-3.Connection to a Computer
The balance can be connected to a personal computer
using the RS-232C serial interface. As an option, the
FX
i-02 USB interface is available to transmit the
balance data to a personal computer.
The balance is a DCE device. Use a straight through
cable when connecting to a personal computer. A
commercially available modem cable can be used.
When purchasing, confirm the following specifications.
D-Sub 9-pin
Female connector
Inch screw thread
15-4.Using Windows Communication Tools (WinCT)
When Windows is used as an operating system in a personal computer, the WinCT software, that is
downloaded from the A&D website, can be used to transmit the weight data to the personal computer.
The WinCT has three communication methods: "RsCom", "RsKey" and "RsWeight".
Rscom
Can transmit commands to control the balance.
Can make bi-directional communication between the balance and a personal computer using the
RS-232C serial interface.
Can display or store the data using a text file format. Can also print the data using a printer connected
to a personal computer.
When several ports of a personal computer have balances connected, can communicate with each
balance simultaneously.
Can share a personal computer with other application software.
Rskey
Can transmit the weight data output from the balance directly to other application software such as
Microsoft Excel.
Can be used with most application software.
Rsweight
Can retrieve the weight data from the balance and display the data as a graph in real time.
Can calculate and display the maximum, minimum, average, standard deviation and coefficient of
variation.
Using the WinCT software, the balance can do the following:
1 Analyzing the weight data and the statistics with "RsKey"
The weight data can be input directly into an Excel worksheet. Then, Excel can analyze the data to
obtain sum, average, standard deviation, maximum and minimum value, and display them as a graph.
2 Controlling the balance using commands from a personal computer
By using "RsCom", the personal computer sends commands such as "re-zero" or "send weight data"
to the balance and controls the balance.
3 Printing the balance GLP report using your printer
The balance GLP report can be printed using a printer connected to a personal computer.
4 Receiving weight data at a certain interval
The weight data can be received at a certain interval and data characteristic with elapsed time can be
obtained.
73
Page 74
5 Using a personal computer as an external indicator
With the "RsKey" test mode function, a personal computer can be used as an external weight indicator
for the balance. (To do this, set the balance data output mode to stream mode.)
74
Page 75
16. Commands
16-1. Command List
Note
A command has a terminator added, that is specified using the "Ter mi na to r (CRLF or CR)"
parameter of "Serial interface (
Commands to query weight data
C Cancels the S or SIR command.
Q Requests the weight data immediately.
S Requests the weight data when stabilized.
SI Requests the weight data immediately.
SIR Requests the weight data continuously.
E
S
P Requests the weight data when stabilized.
C
Note: The "Q" and "SI" commands, the "S" and "
Commands to control the balance
CAL Same as the CAL key
EXC * Calibration using an external weight
OFF Turns the display off.
ON Turns the display on.
P Same as the ON:OFF key
PRT Same as the PRINT key
R Same as the RE-ZERO key
SMP Same as the SAMPLE key.
T Tare key
Z Same as the RE-ZERO key
E
S
T Same as the RE-ZERO key
C
U Same as the MODE key
?ID Requests the ID number.
?SN Requests the serial number.
?TN Requests the model name.
?PT
PT: ***.*** ctChanges the tare weight.
KL:***KL:000 Unlocks all keys.
?KL Requests locking all keys status.
LK:*****Locks the specified key.
?LK Requests locking the specified keys status.
Requests the tare weight.
The unit added is the current weighing unit in A&D standard format.
KL:001 Locks all keys.
(Refer to 17. Key Lock Function)
Enter numerical values from 00000 to 00063 in place of
(Refer to 17. Key Lock Function.)
(Refer to 17. Key Lock Function.)
)" in the function table, and is sent to the balance.
E
S
P" commands behave the same.
C
.
*****
Note:
E
The "R", "Z" and "
E
S
75
: ASCII code 1Bh
C
: Only fortheFZ-CT series
*
S
T" commands behave the same.
C
Page 76
16-2.Acknowledge Code and Error Codes
When the "AK, Error code ()" parameter of "Serial interface ()" is set to "1 ", the balance
outputs an appropriate response for each command to improve the reliability of communications.
When is set:
When the balance receives a command to request data and cannot process it, the balance transmits
an error code (EC, Exx).
When the balance receives a command to request data and can process it, the balance outputs the data.
When the balance receives a command to control the balance and cannot process it, the balance
transmits an error code (EC, Exx).
When the balance receives a command to control the balance and can process it, the balance
transmits the acknowledge code (AK).
Among commands to control the balance, the following transmit the acknowledge code (AK) both
when the balance receives the command and when the balance has accomplished the command. If
the command cannot be processed properly, the balance transmits an error code (EC, Exx). This error
can be released using the CAL command.
CAL command (When performing the calibration using a calibration weight)
ON command, P command
R command, Z command, T command (When setting the display to zero)
R command example
AK: ASCII code 06h
76
Page 77
T command example
PT command example
When a communication error has occurred due to external noise, or a parity error has occurred due to
transmission error, the balance transmits an error code. In this case, send the command again.
AK: ASCII code 06h
16-3.Settings Related to RS-232C
Concerning the RS-232C, the balance has two functions: "Data output ()" and "Serial interface
(
)". Set each function as necessary.
77
Page 78
17.Key Lock Function
Sending specified commands to the balance will lock the keys on the balance to disable the key
functions. This is useful for controlling the balance only by an external device such as a personal
computer.
Even if the keys are locked, key operations using commands are available. For details on
commands to operate the keys, refer to “16. Commands”.
The key lock state can be checked using the command to request the state of the balance keys.
The key lock state is maintained until the balance receives the key unlocking command or the
balance is turned off by removing the AC adapter.
17-1.Locking All the Keys
Using the KL command, all the keys on the balance can be disabled.
Command string Description
Requests the state of the all key lock function.
?KL
KL:
KL,000 All keys unlocked
KL,001 All keys locked
KL:000 Unlocks all the keys
KL:001 Locks all the keys
Enter 000 or 001 in place of .
17-2.Locking the Specified Keys
Entering numerical values to the LK command will lock specified keys on the balance to disable
their key functions. Set the numerical values
bits assigned to each key as shown below.
The LK command is available for balances with software version 1.860 or later.
2 Send the LK command to the balance with the sum that is obtained in Step 1.
Command string Description
?LK
LK:
by adding the decimal numbers converted from
LK:00047
Requests the state of the specified key lock function.
e.g.1: LK,00047 Locks all the keys excluding the PRINT key.
e.g. 2: LK,00000 All keys unlocked.
Locks the specified keys.
e.g.: LK:00047 Locks all the keys excluding the PRINT key.
Enter the numerical values 00000 to 00063 in place of .
78
Page 79
18.Checking the Balance Software Version
Specifications may vary depending on the software version of the balance.
The software version can be checked by the following procedure.
1 Reinsert the AC adapter of the balance.
2
P- . is displayed.
The number shown in place of “ . “ is the
software version.
LEVEL
P- 1.860
19.Maintenance
Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that is
moistened with warm water and a mild detergent.
Balance main unit.
Weighing pan
Breeze break
(standard accessory)
Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.
Use the original packing material for transportation.
Use a soft, lint free cloth that is moistened with a mild detergent to clean.
The edge of the weighing pan is sharp. Use caution when cleaning the pan.
An antistatic treatment has been applied to the breeze break components.
Use a soft, dry, lint free cloth to clean. Cleaning with a cloth that is moistened
with water or a mild detergent repetitively, or washing with water, may degrade
the antistatic effect.
79
Page 80
20. Troubleshooting
20-1. Checking the Balance Performance and Environment
The balance is a precision instrument. When the operating environment or the operating method is
inadequate, correct weighing cannot be performed. Place a sample on the pan and remove it, and
repeat this several times. If the balance seems to have a problem with repeatability or to perform
improperly, check as described below. Also, visit our website,
"Information Library" and "FAQ". If improper performance persists after checking, contact the local
A&D dealer for repair.
Checking that the balance performs properly
Check the balance repeatability using an external weight. Be sure to place the weight in the center of
the weighing pan.
Check the balance repeatability, linearity and calibrated value using external weights with a known value.
Checking that the operating environment or weighing method is proper
Operating environment
Is the weighing table solid enough?
Is the balance level? Refer to "3-1. Before Use".
Is the operating environment free from vibration and drafts? Has the breeze break been installed?
Is there a strong electrical or magnetic noise source such as a motor near the balance?
Weighing method
Is the weighing pan installed correctly?
Is the RE-ZERO key pressed before placing a sample on the weighing pan?
Is the sample placed in the center of the weighing pan?
Has the balance been calibrated using an external weight? (Or one-touch calibration, only for the
FZ-CT series)
Has the balance been warmed up for 30 minutes before weighing?
Sample and container
Has the sample absorbed or lost moisture due to the ambient conditions such as temperature and
humidity?
Has the temperature of the container been allowed to equalize to the ambient temperature? Refer to
"
3-2. During Use".
Is the sample charged with static electricity? Refer to "3-2. During Use".
The
Is the sample of magnetic material such as iron? Caution is required for weighing magnetic materials.
FZ-CT / FX-CT series balances are prone to be charged with static electricity when the relative
humidity is low.
Refer to "
3-2. During Use".
https://www.aadd.jp/, for
80
Page 81
20-2.Error Codes
Display Error code Description
Stability error
The balance cannot stabilize due to an environmental problem. Prevent
vibration, drafts, temperature changes, static electricity and magnetic
EC, E11
fields. Refer to "
and "
6. Response Adjustment" about adapting the balance to the
environment. To return to the weighing mode, press the
Out of range error
The value entered is beyond the settable range. Re-enter the value.
Internal mass error (Only for the FZ-CT series)
The internal mass application mechanism does not function properly.
EC, E17
Perform the weighing operation from the beginning again.
Calibration weight error
The calibration weight is too heavy. Confirm that the weighing pan is
EC, E20
properly installed. Confirm the calibration weight value.
Press the
Calibration weight error
The calibration weight is too light. Confirm that the weighing pan is
EC, E21
properly installed. Confirm the calibration weight value.
Press the
Overload error
A sample beyond the balance weighing capacity has been placed on
the pan. Remove the sample from the pan.
Weighing pan error
The weighing value is too light. Confirm that the weighing pan and pan
support are properly installed. Press the ON:OFF key two times to
return to the weighing mode. If the error still persists, calibrate the
balance.
Sample mass error
The balance cannot store the sample for the counting mode or for the
percent mode because it is too light. Use a sample that is heavier.
Unit mass error
The sample unit mass for the counting mode is too light. Storing and
using it for counting will cause a counting error. Add samples to reach
the specified number and press the
Pressing the PRINT key without adding samples will shift the balance
to the counting mode. But, to acquire accurate weighing, be sure to add
samples.
Balance internal error
If this error appears persistently, contact the local A&D dealer.
Battery error for clock
The back-up battery power for the clock was lost completely. After
pressing one of the keys, adjust the date and time. Even if the back-up
battery power for the clock was lost and the balance is on or in the
standby state, the balance performance will be normal. When this error
appears frequently, repair is required.
3. Precautions" for details on the operating environment
CAL key.
CAL key to return to the weighing mode.
CAL key to return to the weighing mode.
PRINT key.
81
Page 82
Display Error code Description
Low battery error
Voltage of the power supply drops.
EC, E00
EC, E01
EC, E02
EC, E03
EC, E04
EC, E06
EC, E07
Other errors
Confirm whether the correct AC adapter is used.
Communications error
A protocol error occurred in communications. Confirm the format, baud
rate and parity.
Undefined command error
An undefined command was received. Confirm the command.
Not ready
A received command cannot be processed.
e.g. The balance received a Q command, but not in the weighing
mode.
e.g. The balance received a Q command while processing a RE-ZERO
command. Adjust the delay time to transmit a command.
Timeout error
If the timeout parameter is set to "
the next character of a command within the time limit of one second.
Confirm the communication.
Excess characters error
The balance received excessive characters in a command.
Confirm the command.
Format error
A command includes incorrect data.
e.g. The data is numerically incorrect. Confirm the command.
Parameter setting error
The received data exceeds the range that the balance can accept.
Confirm the parameter range of the command.
If the errors described above cannot be released or other errors are
displayed, contact the local A&D dealer.
t-Up1", the balance did not receive
20-3.Other Display
The display ( ) of the password input is displayed when using the
password function and turning on the balance.
Enter a correct password. Refer to "
13. Password Function".
20-4. Asking for Repair
If the balance needs service or repair, contact the local A&D dealer.
The balance is a precision instrument. Use much care when handling the balance and observe the
following when transporting the balance.
Use the original packing material.
Remove the weighing pan from the main unit
82
Page 83
A
–02
p
X
i
p
21. Options and Peripherals (Sold Separately)
Note
The FXi-02, FXi-08 and FXi-09 cannot be used at the same time.
FXi-02 USB interface
(Installed in the balance, Applicable OS: Windows 98 OSR2 or later)
Used to transmit the balance weight data (numerical value
only) uni-directionally to a personal computer via USB.
Can transmit the balance weight data (numerical value only)
directly to application software such as Microsoft Excel, Word
and memo pad.
Driver installation is not necessary.
Note
To perform bi-directional communication using WinCT
or input statistical data or GLP data to a personal
computer, use the USB converter (AX-USB-9P-EX)
FXi-08 Ethernet interface
Used to connect the balance to a LAN.
The "WinCT-Plus" data communication software is provided as a
standard accessory and can perform the following.
Acquire data from multiple balances connected to a LAN.
Control these balances with commands.
Acquire data transmitted from balances.
Example: When pressing the
is output and is acquired by the computer.
The stored data can be used with Microsoft Excel (if installed).
PRINT key of the balance, data
FXi-09 Built-In battery unit (Ni-MH rechargeable battery pack)
Charging time: Approximately 10 hours.
Continuous operation hours: Approximately 8 hours
FXi-02 Example of use
HR
Exam
F
-08 Exam
leof use
le of use
Note
The charging time depends on the operating environment.
During charging, the balance is not available for use.
FXi-10 Small breeze break
Breeze break shorter than the standard large breeze break.
(Total height of the small breeze break: 105 mm)
FXi-11 Large breeze break
Breeze break provided as standard.
AX-FXi-31 Main unit cover
Main unit protective cover provided as a standard accessory.
83
Page 84
AX-CARATPAN-W Carat pan (Silver colored)
Two silver colored carat pans (the same as the carat pan provided
as a standard accessory)
AX-CARATPAN-B Carat pan (Black colored)
Two black carat pans (the same as the carat pan provided as a
standard accessory)
AX-USB-9P USB converter
Adds a COM port to a PC.
Enables bi-directional communication between the
PC and the balance when a USB driver is installed.
Can use serial communication software such as WinCT
on a PC without COM ports.
AD-1671 Anti-Vibration table
A 27-kg table with vibration-absorbing rubber feet, to reduce vibration
coming from the floor to enable stable weighing.
Use the AD-8922A remote controller to avoid weighing errors that can
be caused by slight table tilting when operating the scale manually.
AD-1672 Tabletop breeze break
Protects the balance from wind from sources such as air conditioning or people
passing by, thereby reducing balance weighing error.
The transparent panel assembly consists of antistatic plastic material that
protects the balance from static electricity.
AD-1683 DC static eliminator
Used to minimize weighing errors due to static electricity on the material.
The AD-1683 is direct-current static eliminator. The ions generated produce
no breeze and are effective over a long distance. Therefore, the balance
can accurately weight powders, etc. by using the AD-1683.
AD-1684A Electrostatic field meter
This option measures the amount of the static charge on the sample, tare
or peripheral equipment and displays the result.
If those are found to be charged, discharge them using the AD-1683 DC
static eliminator.
AD-1687 Weighing environment logger
A data logger equipped with 4 sensors for temperature, humidity,
barometric pressure and vibration that can measure and store
environmental data.
When connected to the RS-232C interface of the balance, the AD-1687 can
store environmental data along with weighing data. Therefore, it is possible to
store data in an environment where a computer cannot be used.
The stored data can be read to a personal computer using USB.
As the AD-1687 is recognized as USB memory, special software is not
required to read the data.
84
Page 85
AD-1688 Data logger
When connected to the RS-232C interface of the balance, the
AD-1688 can store the data in an environment where a
personal computer cannot be used.
The stored data can be read to a personal computer using
USB. As the AD-1688 is recognized as USB memory, special
software is not required to read the data.
AD-1689 Tweezers for calibration weight
A pair of tweezers ideally suited for holding calibration weights
of 1 g to 500 g.
AD-8127 Multi-Printer
Small dot impact printer that connects to the balance via the RS-232C interface.
Statistical calculation mode, calendar/clock function, interval printing mode, chart printing mode,
dump printing mode
AD-8529PC-W Bluetooth converter (for connecting a personal computer)
This option connects a personal computer to the balance wirelessly (via Bluetooth®) up to 10 m.
Driver installation is required.
AD-8529PR-W Bluetooth converter (for connecting a printer)
This option connects a printer to the balance wirelessly (via Bluetooth®) up to 10 m.
AD-8920A Remote display
Connected to the balance using the RS-232C serial interface to display the weight data away
from the balance.
AD-8922A Remote controller
Connected to the balance using the RS-232C serial interface to display the weight data and
to remotely control the balance.
Analog output and comparator output available as an option to be installed.
85
Page 86
2
22.Specifications
Weighing capacity
Maximum display
Minimum weighing value
(1 digit)
Repeatability
Standard deviation
Linearity
Stabilization time
(typical at FAST)
Sensitivity drift
(10 °C - 30 °C / 50 °F - 86 °F)
Internal mass
Clock and calendar function
Operating environment
Display refresh rate
Counting
mode
Percent
mode
Interface
External calibration weight
Weighing pan diameter
External dimensions
AC adapter
Power consumption
Mass of product
Minimum unit mass
3
*
Number of samples
Minimum 100% reference m-ass
3
*
Minimum 100% display
FZ-1200CT FZ-700CT FZ-500CT
1260 ct 760 ct 510 ct
252 g 152 g 102 g
1
*
1260.042 ct 760.042 ct
252.008 g 152.008 g
510.042 ct
1
*
102.008 g
0.001 ct
0.001 g
0.001 ct
0.0005 g
±0.002 ct
±0.001 g
1
*
Approx. 2 second
±2 ppm/°C
Available
Available
5 °C to 40 °C (41 °F to 104 °F) 85 %RH or less (No
condensation)
*
Approx. 5 times/second
(or 10 times/second)
0.001 g
10, 25, 50 or 100 pieces
0.100 g
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
RS-232C
250 g
200 g
100 g
50 g
150 g
100 g
50 g
20 g
100 g
50 g
20 g
25 g
90 mm
198(W) x 294(D) x 315(H) mm
Confirm that the adapter type is correct for the local voltage
and power receptacle type
Approx. 11 VA (supplied to the AC adapter)
Approx. 3.8 kg
1: With MID. , factory setting of the rate of the response adjustment, the stabilization time is approx. 3
*
seconds.
2: Factory setting
*
3: To use the counting mode or the percent mode, storing each mode as described in “9-9. Storing
*
Units” is required.
86
Page 87
2
Weighing capacity
Maximum display
Minimum weighing value
(1 digit)
Repeatability
Standard deviation
Linearity
Stabilization time
(typical at FAST)
Sensitivity drift
(10 °C - 30 °C / 50 °F - 86 °F )
Internal mass
Clock and calendar function
Operating environment
Display refresh rate
Counting
mode
Percent
mode
Minimum unit mass
3
*
Number of samples
Minimum 100% reference m-ass
3
*
Minimum 100% display
Interface
External calibration weight
Weighing pan diameter
External dimensions
AC adapter
Power consumption
Mass of product
FZ-1200CT FZ-700CT FZ-500CT
1260 ct 760 ct 510 ct
252 g 152 g 102 g
1
*
1260.042 ct 760.042 ct
252.008 g 152.008 g
510.042 ct
1
*
102.008 g
0.001 ct
0.001 g
0.001 ct
0.0005 g
±0.002 ct
±0.001 g
1
*
Approx. 2 second
±2 ppm/°C
Unavailable
Unavailable
5 °C to 40 °C (41 °F to 104 °F) 85 %RH or less (No
condensation)
*
Approx. 5 times/second
(or 10 times/second)
0.001 g
10, 25, 50 or 100 pieces
0.100 g
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
RS-232C
250 g
200 g
100 g
50 g
150 g
100 g
50 g
20 g
100 g
50 g
20 g
25 g
90 mm
198(W) x 294(D) x 315(H) mm
Confirm that the adapter type is correct for the local voltage
and power receptacle type
Approx. 11 VA (supplied to the AC adapter)
Approx. 3.4 kg
1: With MID. , factory setting of the rate of the response adjustment, the stabilization time is approx. 3
*
seconds.
2: Factory setting
*
3: To use the counting mode or the percent mode, storing each mode as described in “9-9. Storing
*
Units” is required.
87
Page 88
23. External Dimensions
FZ-1200CT FZ-700CT FZ-500CT
FX-1200CT FX-700CT FX-500CT
Unit: : mm
1: When opening maximum
*
2: Inside dimension
*
3: Weighing pan diameter
*
4: Height up to the weighing pan
*
5: When pulling out the locking handle
*
6: Inside dimension (from up to the weighing pan)
*
7: Underhook position (hole center)
*
270
1
(
)
*
105 (
)
4
*
86.5 (
198
162
5
107(
)
*
3(
*
1
)
*
7
)
1
(
294
1
5
6
)
*
(
*
2
)
6
3)
*
90 (
)
6
*
315
198(
112 (
2
)
*
61.5191.5
192(
7
)
*
88
Page 89
24. Terms/Index
24-1. Terms
Stable value The weight data when the stabilization indicator appears.
Environment Ambient conditions such as vibration, drafts, temperature, static electricity and
magnetic fields which affect the weighing operation.
Calibration Adjustment of the balance so that it can weigh accurately.
Output To output the weight data using the RS-232C serial interface.
Zero point A weighing reference point or the zero display. Usually refers to the value displayed
when nothing is on the weighing pan.
Digit Unit of digital resolution. Used for the balance, a unit of minimum weighing value.
Tare To cancel the weight of a container which is not included in the weight data.
Mode Balance operational function.
Re-zero To set the display to zero.
GLP Good Laboratory Practice.
GMP Good Manufacturing Practice.
Repeatability Variation in measured values obtained when the same weight is placed and
removed repetitively. Usually expressed as a standard deviation.
e.g. Standard deviation=1 digit: This means that measured values fall within
in the frequency of about 68%.
Stabilization time Time required after a sample being placed, until the stabilization indicator illuminates
and the weight data is displayed.
Sensitivity drift An affect that a change in temperature causes to the weight data. Expressed as
temperature coefficient.
e.g. Temperature coefficient = 2 ppm/
changes by 10
0.0002%/
C, the value displayed changes by the following value.