To ensure that our products are safe and without risk to health,
the following points must be noted:
• The relevant sections of these instructions must be read
carefully before proceeding.
• Warning labels on containers and packages must be
observed.
• Installation, operation, maintenance and servicing must
only be carried out by suitably trained personnel and in
accordance with the information given.
• Normal safety precautions must be taken to avoid the
possibility of an accident occurring when operating in
conditions of high pressure and/or temperature.
• Chemicals must be stored away from heat, protected from
temperature extremes and powders kept dry. Normal safe
handling procedures must be used.
• When disposing of chemicals ensure that no two chemicals
are mixed.
Safety advice concerning the use of the equipment described in
this manual or any relevant hazard data sheets (where
applicable) may be obtained from the Company address on the
back cover, together with servicing and spares information.
Both direct and alternating current supply
The equipment is protected through double insulation
Information in this manual is intended only to assist our
customers in the efficient operation of our equipment. Use of
this manual for any other purpose is specifically prohibited and
its contents are not to be reproduced in full or part without
prior approval of the Technical Publications Department.
Note. Do not bend the pen arm during removal
and re-fitting as pen clashing may result.
Raise pens
1
Input 1
-999.99
deg C
5
Remove cap
Fit the correct colored pen
capsule ensuring that the arm
locates in the pen
capsule slot
4
Red Pen
(Channel 1)
Green Pen
(Channel 2)
Blue Pen
(Channel 3)
Black Pen
(Channel 4)
2
Identify pens
Note. The number of pens fitted (and
the colors of the pen capsules
supplied) depends on the recorder
model.
Slide pen arm and capsule
assembly onto the appropriate
bracket until it clips into place
6
ensuring that the arm is
positioned above its lifter bar
7
Repeat steps to for
remaining pens as required
36
Advanced circular chart recorder1 Preparation
1 Preparation
1.1 Preparing the Recorder for First Use
Prepare the recorder for first use as follows:
1. Install the recorder – see Section 4, page 60.
2. Connect the recorder – see Section 5, page 64.
Warning. Ensure all connections are made correctly,
especially to the earth (ground) stud – see Section
5.4, page 71.
3. Switch on the power supply to the recorder.
4. Fit the pen capsule(s) – see Fig. 1.1.
Note.
Ensure the correct colored capsule is fitted to
the appropriate pen arm. Each pen arm is
identified by a colored band – see Fig. 1.1.
If the violet, true-time line event option is fitted,
the pen records on the same time line as the red
pen, but on the outer edge of the chart.
Note. On power-up, the pen arm(s) is (are) moved to
an off-chart position for automatic referencing. Pen
chatter may occur on the pen(s) nearest the reference
position. This is a normal function of the recorder.
5. Fit a chart – see Fig. 1.2 on page 3.
Fig. 1.1 Fitting the Pen Capsule(s)
2IM/C1300 Rev. J
C1300
Lift the chart clamp
2
Lower the chart clamp and press down
firmly to ensure that the 2 locating pins
pierce the paper
6
Ensure pen(s) in
raised position
1
Input 1
-999.99
deg C
Fit a chart ensuring
that it is beneath the
pen lifter bars
Locate chart
under guides
Rotate chart to align the
time line with the red pen
3
4
Red Pen
5
Lower pen(s)
7
Input 1
-999.99
deg C
Switch off the power supply to the recorder
8
Advanced circular chart recorder1 Preparation
Fig. 1.2 Fitting a Chart
IM/C1300 Rev. J3
C1300
WWW.abb.com
Self Tests, Software Key and
Module Identification
Instrument Type and Software Version
C1314J
Software Issue
1
Main: CFG Test :Pass
SKey:Type B :Pass
Mod2:Not Fitted
Mod3:Not Fitted
Mod4:Not Fitted
Mod5:Not Fitted
Mod6:Not Fitted
Advanced circular chart recorder1 Preparation
1.2 Powering up the Recorder
To power up the recorder:
1. Switch on the supply to the recorder, any power-operated control circuits and the input signals. Wait for the pens to settle.
2. The start-up self-test sequence shown in Fig. 1.3 is shown on display screen 1 when the power supply is first switched on.
Fig. 1.3 Recorder Start-up Self-test Sequence
4IM/C1300 Rev. J
C1300
Main: CFG Test :Pass
SKey:Type B :Pass
Mod2:Not Fitted
Mod3:Not Fitted
Mod4:Not Fitted
Mod5:Dig Out :Pass
Mod6:RS485 :Pass
Analog input and/or analog output calibration is corrupt.
Power down and then up again.
If the fault persists, contact the Company.
Configuration stored in non-volatilememory is corrupt.
Check and correct program data.
If the fault persists, contact the Company.
Note. The software key is optional hardware that is fitted to enable optional software features – Totalizers, Math
Blocks, System Clock and Timers.
Note. Acknowledging an Error Message clears the error state but does not rectify the fault. After acknowledging an
error, carry out the appropriate action detailed in the above table.
Advanced circular chart recorder1 Preparation
1.2.1 Recorder Status and Error Page
If any of the start-up self-tests (see Fig. 1.3) fail, the error message 'Fail' is displayed. Refer to Fig. 1.4 for possible cause and remedial
action information.
Fig. 1.4 Start-up Error Messages
IM/C1300 Rev. J5
C1300
Making a Selection
Enter/Edit User-defined Information
Adjusting and Storing a Parameter Value
Select
Save selection and
advance to next screen
Raise/Lower Pens
If Auto Pen Drop is set to Yes (Section 3.4), the
key
raises/lowers the pen(s) on alternate operations when
Operating Level is selected from the Main Menu.
Use the and keys to select the character to be displayed
at the cursor position see
Note below.
Use the
key to move the cursor.
Press to save the new value and advance to the next screen.
Pen 1 Eng. Hi
100.00 deg C
The first number flashes. Use the and keys to
change the value as required see
Note below.
Each press of the
key selects the next number to
the right. Use the
and keys to change the value
as required see
Note below.
Press to save the new value and advance to the
next screen.
Display
Screen
Operator Keys
Operating Level
Alarm Setpoints
Security Code
Common
Channels
Alarms
Relays
Digital I/O
Highlight a
menu item
Operating Level
Alarm Setpoints
Security Code
Common
Channels
Alarms
Relays
Digital I/O
Display Screen and Operator Keys
Cursor
Note. Continued pressure on the and keys causes the rate of change of the displayed value to increase.
Press the keys momentarily to make small adjustments.
Advanced circular chart recorder2 Operation
2Operation
2.1 Display Screens and Operator Keys
Up to two faceplates are fitted to the front of the recorder, each comprising a display screen and associated operator keys – see
Fig. 2.1.
Each high resolution, dot matrix display screen shows the operating and programming information for up to two input channels in a
variety of formats.
Alarm states are indicated by icons in the lower left corner of the display screen(s) – see Fig. 2.3 on page 7.
Fig. 2.1 Display Screen and Operator Keys
6IM/C1300 Rev. J
C1300
Hardware failure on
Process Variable 1
Examples
Input out of range on
Process Variable 1
Boiler 1
Error3
gal/d
Boiler 1
F-inpt
gal/d
MessageReasonAction
Error1
Error2
Error3
Finpt
Unspecified error from the Analog to Digital Converter
Corrupt data from the Analog to Digital Converter
No reply from the Analog to Digital Converter
Input is above or below fault detection level
or
Input exceeds the limits for the linearizer selected
Check that the input/output board is located correctly in its socket
see Section 5.1
Power down and up
If the error message is still present, contact the Company
Check input source for possible broken sensor
Check input connections
Check input link position
Check input configuration see Section 3.5
Note. Error messages are cleared automatically when the fault condition no longer exists.
Global Alarm Indicator displayed if any alarm in any channel is active
High Process
Low Process
relevant only to the channel currently displayed
Fast Rate
Slow Rate
Advanced circular chart recorder2 Operation
The recorder has dedicated Operating Pages in the Operating Level – see Sections 2.3 to 2.6. These pages display the process
measurements and are not affected by the security system which inhibits access to the Configuration Pages – see Section 3.3,
page 19.
2.2 Input Error Messages and Alarm Icons
Fig. 2.2 Input Error Messages Displayed in the Operating Page
Fig. 2.3 Alarm Icons
IM/C1300 Rev. J7
C1300
Operating Level
Totalizer Log
Totalizer Control
Alarm Setpoints
Security Code
Common
Channels
Alarms
Input 1
-999.99
deg C
Advanced circular chart recorder2 Operation
2.3 Operating Pages
Overview.
The following information is displayed in the Operating Pages:
– Input (pen) channel readings
– Data logging status (if data logging option is enabled by installation of the appropriate hardware)
– System time/date (if any one of the Totalizer, Math or Timers software options are enabled by installation of the
appropriate software key)
– Totalizer readings (if the totalizer software option is enabled by installation of the appropriate software key)
– Totalizer log (if the totalizer software option is enabled by installation of the appropriate software key)
– Totalizer control (if the totalizer software option is enabled and Totalizer Reset is set to YES – see Section 3.4, page 20.)
– Alarm set points (if Alarm Adjust is set to Yes – see Section 3.4, page 20.)
The Input (pen) channel and totalizer displays are configured separately (see Pen Ch Display and Totals Display in Section
3.4) and each can be displayed in one of several ways. Input channel information can be displayed without totalizer information
and vice-versa.
Input channel 1 and 2 information and system time/date is shown on display screen 1.
If the recorder is fitted with the optional second faceplate:
– Input channel 3 and 4 information, together with system time/date is shown on display screen 2
– Totalizer logs, totalizer control and alarm set point adjustment for pens 3 and 4 and associated totalizers are accessed
from the menu shown on display screen 2 when the key is pressed on faceplate 2.
Totalizer information is shown on the same display screen as the pen to which the totalizer is assigned.
All channel and totalizer configuration is done using faceplate 1.
2.3.1 Accessing the Operating Pages
Press the key to open the Main Menu.
Highlight Operating Level.
see Section 2.3.3, page 10.
8IM/C1300 Rev. J
C1300
Input 1
-999.99
deg C
Autoscroll On
Slow Rate
Autoscroll On
Fast Rate
Autoscroll Off
Autoscroll On
Slow Rate
Advanced circular chart recorder2 Operation
2.3.2 Autoscroll
Autoscroll
When Autoscroll is set to On, each Operating Page screen is displayed in turn for a period of
time determined by the rate selected:
Slow Rate – each screen is displayed in turn for 10 seconds
Fast Rate – each screen is displayed in turn for 5 seconds
Press the key from any Operating Page screen to initiate Autoscroll in Slow Rate mode.
Press the key again to switch to Fast Rate mode.
When in Fast Rate mode, press the key to switch to Slow Rate mode.
When in Slow Rate mode, press the key to switch Autoscroll off.
Press the key at any time to switch Autoscroll off.
IM/C1300 Rev. J9
C1300
Input 1
-999.99
deg C
System Time
Tuesday
25th May 2004
14:33:48
Input 2
O.OOO
m3/s
Totalizer 1
000000000
Input 1
-999.99
deg C
Input 1
-999.99
deg C
System Time
Tuesday
25th May 2004
14:33:48
Data Logging
Status: Online
Advanced circular chart recorder2 Operation
2.3.3 Input (Pen) Channel Display – Separate
The following screens are displayed only if Pen Ch Display is set to Separate – see Section 3.4, page 20.
Input Channel 1 (Pen 1)
Channel tag.
Measured value.
Alarm icons and measured value units.
Note. If Pen 1 Source is set to None (Section 3.14), only a measured value of 0 is displayed.
Input Channel 2 (Pen 2)
Note. If Pen 2 Source is set to None (Section 3.14), only a measured value of 0 is displayed.
Totalizer, Math or Timers software options not enabled by installation of the appropriate software
key and optional Data Logging hardware not installed – return to top of page.
Totalizer software option enabled by installation of the appropriate software key and To ta l sDisplay not set to Off (Section 3.4) – see Sections 2.3.6 to 2.3.10.
Optional Data Logging hardware installed – see Section 2.3.11, page 15.
Optional Data Logging hardware not installed and Totalizer software option enabled by
installation of the appropriate software key and Totals Display set to Off (Section 3.4) –
continued below.
System Time/Date Display
Notes.
Available only if any one of the Totalizer, Math or Timers software options are enabled by
installation of the appropriate software key, or the optional Data Logging hardware is
installed.
If Pen Ch DisplayandTotals Display are set to Off (Section 3.4) and any of the above
software options are enabled and the optional Data Logging hardware is not installed, the
System Time Display is the only Operating Page screen available.
Return to top of page.
10IM/C1300 Rev. J
C1300
Input 1
O.O l/s
Input 2
O.OOO m3/s
Totalizer 1
000000000
Input 1
O.O l/s
Input 2
O.OOO m3/s
System Time
Tuesday
25th May 2004
14:33:48
Data Logging
Status: Online
Advanced circular chart recorder2 Operation
2.3.4 Input (Pen) Channel Display – Dual
The following screens are displayed only if Pen Ch Display is set to Dual – see Section 3.4, page 20.
Input Channels 1 and 2 (Pens 1 and 2)
Channel tag.
Measured value and units.
Note. If Pen X Source (where X is the pen number) is set to None (Section 3.14), only a
measured value of 0 is displayed for that channel.
Totalizer, Math or Timers software options not enabled by installation of the appropriate software
key and optional Data Logging hardware not installed – return to top of page.
Totalizer software option enabled by installation of the appropriate software key and TotalsDisplay not set to Off (Section 3.4) – see Sections 2.3.6 to 2.3.10.
Optional Data Logging hardware installed – see Section 2.3.11, page 15.
Optional Data Logging hardware not installed and Totalizer software option enabled by
installation of the appropriate software key and Totals Display set to Off (Section 3.4) – see
Section 2.3.3, page 10.
IM/C1300 Rev. J11
C1300
Input 2
O.OOO
m3/s
Input 1
-999.99
deg C
Input 1
O.O l/s
Input 2
O.OOO m3/s
Totalizer 1
000000000
Input 1
-999.99
deg C
System Time
Tuesday
25th May 2004
14:33:48
Data Logging
Status: Online
Advanced circular chart recorder2 Operation
2.3.5 Input (Pen) Channel Display – Separate and Dual
The following screens are displayed only if Pen Ch Display is set to Separate & Dual – see Section 3.4, page 20.
Input Channel 1 (Pen 1)
Channel tag.
Measured value.
Alarm icons and measured value units.
Note. If Pen 1 Source is set to None (Section 3.14), only a measured value of 0 is displayed.
Input Channel 2 (Pen 2)
Note. If Pen 2 Source is set to None (Section 3.14), only a measured value of 0 is displayed.
Input Channels 1 and 2 (Pens 1 and 2)
Channel tag.
Measured value and units.
Note. If Pen X Source (where X is the pen number) is set to None (Section 3.14), only a
measured value of 0 is displayed for that channel.
Totalizer, Math or Timers software options not enabled by installation of the appropriate software
key and optional Data Logging hardware not installed – return to top of page.
Totalizer software option enabled by installation of the appropriate software key and To ta l sDisplay not set to Off (Section 3.4) – see Sections 2.3.6 to 2.3.10.
Optional Data Logging hardware installed – see Section 2.3.11, page 15.
Optional Data Logging hardware not installed and Totalizer software option enabled by
installation of the appropriate software key and Tot a l s D i s p la y set to Off (Section 3.4) – see
Section 2.3.3, page 10.
12IM/C1300 Rev. J
C1300
Totalizer 1
000000000
Ml
Totalizer 2
000000000
gal
System Time
Tuesday
25th May 2004
14:33:48
Data Logging
Status: Online
Input 1
87.6 gal
000005320
gal
Input 2
856.35 Ml
000027839
Ml
System Time
Tuesday
25th May 2004
14:33:48
Data Logging
Status: Online
Advanced circular chart recorder2 Operation
2.3.6 Totalizer Display – Separate
The following screens are displayed only if Totals Display is set to Separate (Section 3.4) and Tot X Source (where X is the totalizer
number) is set to anything other than None – see Section 3.7, page 35.
Totalizer 1
Totalizer tag.
Totalizer value.
Totalizer value units.
Totalizer 2
Optional data logging hardware installed – see Section 2.3.11, page 15.
2.3.7 Totalizer Display – Rate with Total
The following screens are displayed only if Totals Display is set to Rate with Total (Section 3.4) and To t X S o u r c e (where X is the
totalizer number) is set to anything other than None – see Section 3.7, page 35.
Optional data logging hardware not installed – see Section 2.3.3, page 10.
Totalizer 1
Channel tag.
Measured value and measured value units.
Totalizer value.
Totalizer value units.
Totalizer 2
Optional data logging hardware installed – see Section 2.3.11, page 15.
IM/C1300 Rev. J13
Optional data logging hardware not installed – see Section 2.3.3, page 10.
C1300
Totalizer 1
000005320
gal
87.6 gal
Totalizer 2
000027839
Ml
856.35 Ml
System Time
Tuesday
25th May 2004
14:33:48
Data Logging
Status: Online
Totalizer 1
000005320
gal
Totalizer 2
000027839
Ml
System Time
Tuesday
25th May 2004
14:33:48
Data Logging
Status: Online
Advanced circular chart recorder2 Operation
2.3.8 Totalizer Display – Total with Rate
The following screens are displayed only if Totals Display is set to Total with Rate (Section 3.4) and Tot X Source (where X is the
totalizer number) is set to anything other than None – see Section 3.7, page 35.
Totalizer 1
Totalizer tag.
Totalizer value.
Totalizer units.
Measured value and measured value units.
Totalizer 2
Optional data logging hardware installed – see Section 2.3.11, page 15.
Optional data logging hardware not installed – see Section 2.3.3, page 10.
2.3.9 Totalizer Display – Dual Total
The following screen is displayed only if Totals Display is set to Dual Total (Section 3.4) and Tot X Source (where X is the totalizer
number) is set to anything other than None – see Section 3.7, page 35.
Totalizers 1 and 2
Totalizer tag.
Totalizer value.
Totalizer units.
Optional data logging hardware installed – see Section 2.3.11, page 15.
Optional data logging hardware not installed – see Section 2.3.3, page 10.
14IM/C1300 Rev. J
C1300
87.6
gal/h
000279513
|1
472.6
gal/h
008831908
|2
System Time
Tuesday
25th May 2004
14:33:48
Data Logging
Status: Online
System Time
Tuesday
25th May 2004
14:33:48
Data Logging
Status: Online
Memory Used: 25%
Remove Card
Data Logging
Status: Offline
It is now safe to
remove the
Compact Flash
(59s)
Data Logging
Status: Not Present
Advanced circular chart recorder2 Operation
2.3.10 Totalizer Display – Dual + Flow Rate
The following screen is displayed only if Totals Display is set to Dual + Flow Rate (Section 3.4) and Tot X Source (where X is the
totalizer number) is set to anything other than None – see Section 3.7, page 35.
Totalizers 1 and 2
Measured value and measured value units.
Totalizer value.
Totalizer units and channel number.
Optional data logging hardware installed – see Section 2.3.11, below.
Optional data logging hardware not installed – see Section 2.3.3, page 10.
2.3.11 Data Logging
Data Logging Status
Provides information on current data logging status.
Press the key to place data logging off-line.
When data logging is off-line, the Compact Flash card can be removed safely.
If the card is not removed within 1 minute, data logging is placed on-line automatically. A timer
indicates the number of seconds remaining before data logging resumes.
This screen is displayed when a Compact Flash card is not inserted.
Data logging starts automatically when a Compact Flash card is inserted.
see Section 2.3.3, page 10.
IM/C1300 Rev. J15
C1300
Select Totalizer
Totalizer 1
Select Totalizer
Totalizer 1
19th Apr 04 15:30
Totalizer 1
000047824
gal
Max : 78
Min : 22
Avg : 56
Operating Level
Totalizer Log
Totalizer Control
Alarm Setpoints
Security Code
Common
Channels
Alarms
Advanced circular chart recorder2 Operation
2.4 Totalizer Log
Note.
The totalizer log is available only if the totalizer software option is enabled by installation of the appropriate software key.
The recorder can store up to 21 log entries for each enabled totalizer. When the maximum number of entries has been
reached, the oldest data is overwritten by the newest.
A new log entry is created for each totalizer at the interval selected in the Tot X Log Enable parameter (where X is the
totalizer number) – see Section 3.7, page 35.
Press the key to open the Main Menu.
Highlight Totalizer Log.
Select Totalizer
Select the totalizer log to display.
Totalizer Log
The first log displayed for the selected totalizer is the most recent. Use the key to display
each log in descending date order.
Date and time.
Totalizer tag.
Current value and units.
Maximum, minimum and average values.
Note. If no log entries exist, Log is empty is shown on the display screen.
Return to top of page.
16IM/C1300 Rev. J
C1300
Select Totalizer
Totalizer 1
Totalizer 1
Reset - NO
000000913
Operating Level
Totalizer Log
Totalizer Control
Alarm Setpoints
Security Code
Common
Channels
Alarms
Totalizer 1
Confirm - NO
000000913
NO
Totalizer 1
Reset - NO
000000913
Advanced circular chart recorder2 Operation
2.5 Totalizer Control
Note. Totalizer control is available only if the totalizer software option is enabled by installation of the appropriate software key
andTotalizer Reset is set to YES – see Section 3.4, page 20.
Press the key to open the Main Menu.
Highlight Totalizer Control.
Select Totalizer
Select the totalizer to reset.
Reset Totalizer
Select YES to reset the totalizer.
Confirm Reset Totalizer
Select YES to reset the totalizer.
Note. If YES is selected, the totalizer is reset when the key is pressed.
Return to previous frame.
IM/C1300 Rev. J17
C1300
Select Alarm
A1
A1 Trip
10.000 deg C
Operating Level
Totalizer Log
Totalizer Control
Alarm Setpoints
Security Code
Common
Channels
Alarms
Select Alarm
A1
Advanced circular chart recorder2 Operation
2.6 Alarm Set Point Adjustment
Note. Alarm set point adjustment is available only if Alarm Adjust is set to YES – see Section 3.4, page 20.
Press the key to open the Main Menu.
Highlight Alarm Adjust.
Select Alarm
Select the alarm to adjust.
Alarm Set Point
Adjust the alarm set point.
Return to top of page.
18IM/C1300 Rev. J
C1300
Enter Code
0000
Operating Level
Totalizer Log
Totalizer Control
Alarm Setpoints
Security Code
Common
Channels
Alarms
Operating Level
Totalizer Log
Totalizer Control
Alarm Setpoints
Security Code
Common
Channels
Alarms
Advanced circular chart recorder3 Configuration
3 Configuration
3.1 Introduction
The configuration procedures are used to make changes to the operating parameter values and for scale adjustment. Configuration
of all channels is performed from faceplate 1.
The configuration settings can be backed up to a PC and restored to the recorder as required – see Section 3.16, page 58.
An overview of the Configuration Pages is on the rear cover fold-out.
When changing the input type, it may be necessary to reposition the input selector links accordingly – see Section 5.2.1, page 67.
3.2 Preparation for Configuration
Note.
Isolate any external alarm/control circuits if inadvertent operation during configuration is undesirable.
The recorder responds instantly to parameter changes. These are saved automatically when either the or key is
pressed.
The display screen of faceplate 2 (if fitted) shows Configuration Mode when the configuration pages are accessed by
scrolling past Security Code in the Main Menu on the display screen of faceplate 1.
3.3 Configuration Level Security
Unauthorized access to the configuration pages is prevented by the use of a security code.
The security code, set to '0' when the recorder is despatched, can be set to any value from 0 to 9999 – see Configure Password in
Section 3.4 on pgae 20. When set to anything other than '0', access to all configuration menus (i.e. all menus below Security Code)
is prevented. Enter the correct code to enable access to the configuration menus.
Security Code
Press the key to open the Main Menu.
Highlight Security Code.
Enter Security Code
Enter the correct security code.
Return to Main Menu.
IM/C1300 Rev. J19
C1300
Chart Duration
24 hours
Totalizer Log
Totalizer Control
Alarm Setpoints
Security Code
Common
Channels
Alarms
Totalizers
Stop Chart Srce
None
Auto Pen Drop
YES
Pen Lift Enable
YES
Config Password
0000
Line Rejection
50 Hz
Advanced circular chart recorder3 Configuration
3.4 Common Configuration
Common Configuration
Press the key to open the Main Menu.
Highlight Common.
Chart Duration
Select the chart duration required per revolution of the chart:
between 1 and 167 hours in 1 hour increments
between 7 and 32 days in 1 day increments
Stop Chart Source
Select the source required for stopping the chart.
Refer to Appendix A on page 77 for a description of sources.
Auto Pen Drop
Select YES to enable the pen(s) to drop automatically onto the chart 5 minutes after they are
lifted.
If NO is selected, the pen(s) remain lifted until they are dropped manually by the operator.
Pen Lift Enable
Select YES to enable the pen(s) to be raised and lowered using the pen lift key ().
Configure Password
Select a code between 1 and 9999.
20IM/C1300 Rev. J
Continued on next page.
C1300
Totals Display
Separate
Line Rejection
50 Hz
Pen Ch Display
Separate
Advanced circular chart recorder3 Configuration
Line Rejection
Select the frequency of the mains supply to ensure maximum noise rejection on analog inputs.
Pen Channel Display
Select the style of input (pen) channel display required in the Operating Page – see Sections 2.3.3
to 2.3.5.
Off– input channel information is not displayed in the Operating Page
Separate– input channel information for channels 1 and 2 is shown separately
on display screen 1. Input channel information for channels 3 and 4
(3 and 4 pen recorders only) is shown separately on display screen 2
– see Section 2.3.3, page 10
Dual– input channel information for channels 1 and 2 is shown
simultaneously on display screen 1. Input channel information for
channels 3 and 4 (3 and 4 pen recorders only) is shown
simultaneously on display screen 2 – see Section 2.3.4, page 11
Separate & Dual – input channel information for channels 1 and 2 is first shown
separately, then simultaneously on display screen 1. Input channel
information for channels 3 and 4 (3 and 4 pen recorders only) is first
shown separately, then simultaneously on display screen 2 – see
Section 2.3.5, page 12
Continued on next page.
IM/C1300 Rev. J21
C1300
Totals Display
Separate
Other Ch Display
NO
Alarm Adjust
NO
Totalizer Reset
NO
Advanced circular chart recorder3 Configuration
Totalizers Display
Note. Displayed only if the totalizer software option is enabled by installation of the appropriate
software key.
Select the style of totalizer display required in the Operating Page – see Sections 2.3.6 to 2.3.10.
Off– totalizer values are not displayed in the Operating Page
Separate– totalizer values for totalizers assigned to channels 1 and 2 are
shown separately on display screen 1. Totalizer values for totalizers
assigned to channels 3 and 4 (3 and 4 pen recorders only) are
shown separately on display screen 2 – see Section 2.3.6, page 13
Rate with Total – totalizer values for totalizers assigned to channels 1 and 2, together
with the current measured values for channels 1 and 2, are shown
separately on display screen 1. Totalizer values for totalizers
assigned to channels 3 and 4 (3 and 4 pen recorders only),
together with the current measured values for channels 3 and 4,
are shown separately on display screen 2 – see Section 2.3.7, page
13
Total with Rate – the current measured values for channels 1 and 2, together with
totalizer values for totalizers assigned to channels 1 and 2, are
shown separately on display screen 1. The current measured
values for channels 3 and 4 (3 and 4 pen recorders only), together
with totalizer values for totalizers assigned to channels 3 and 4, are
shown separately on display screen 2 – see Section 2.3.8, page 14
Dual Total– totalizer values for totalizers assigned to channels 1 and 2 are
shown simultaneously on display screen 1. Totalizer values for
totalizers assigned to channels 3 and 4 (3 and 4 pen recorders only)
are shown simultaneously on display screen 2 – see Section 2.3.9,
page 14
Dual + Flow Rate – totalizer values for totalizers assigned to channels 1 and 2, together
with the current measured values for channels 1 and 2, are shown
simultaneously on display screen 1. Totalizer values for totalizers
assigned to channels 3 and 4 (3 and 4 pen recorders only),
together with the current measured values for channels 3 and 4,
are shown simultaneously on display screen 2 – see Section
2.3.10, page 15
Note. Do not select Dual + Flow Rate if a totalizer is not assigned to each of the channels to
be displayed – see Section 3.7, page 35.
Other Channels Display
Select YES to display information for input channels not assigned to pens.
Alarm Adjust
Select YES to enable alarm set point adjustment in the Operating Page – see Section 2.6, page
18.
Continued on next page.
22IM/C1300 Rev. J
C1300
Contrast 1
100%
Contrast 2
100%
Brightness 1
100%
Brightness 2
100%
Chart Duration
24 hours
Totalizer Reset
NO
Advanced circular chart recorder3 Configuration
Totalizer Reset
Note. Displayed only if the totalizer software option is enabled by installation of the appropriate
software key.
Select YES to enable the totalizers to be reset in the Operating Page – see Section 2.5, page 17.
Contrast 1
Select the required contrast setting for display screen 1, between 0 and 100%.
Contrast 2
Note. Shown only if faceplate 2 is fitted.
Select the required contrast setting for display screen 2, between 0 and 100%.
Brightness 1
Select the required brightness setting for display screen 1, between 0 and 100%.
Brightness 2
Note. Shown only if faceplate 2 is fitted.
Select the required brightness setting for display screen 2, between 0 and 100%.