ABB 9438 User Manual

Page 1
Low and High Level
Dissolved Oxygen Monitor
9438
User Guide
IM/9438_6
Page 2
ABB
We are an established world force in the design and manufacture of instrumentation for industrial process control, flow measurement, gas and liquid analysis and environmental applications.
As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and support.
The quality, accuracy and performance of the Company’s products result from over 100 years experience, combined with a continuous program of innovative design and development to incorporate the latest technology.
The UKAS Calibration Laboratory No. 0255 is just one of the ten flow calibration plants operated by the Company and is indicative of our dedication to quality and accuracy.
Health and Safety
To ensure that our products are safe and without risk to health, the following points must be noted:
1. The relevant sections of these instructions must be read carefully before proceeding.
2. Warning labels on containers and packages must be observed.
3. Installation, operation, maintenance and servicing must only be carried out by suitably trained personnel and in accordance with the information given.
4. Normal safety precautions must be taken to avoid the possibility of an accident occurring when operating in conditions of high pressure and/ or temperature.
5. Chemicals must be stored away from heat, protected from temperature extremes and powders kept dry. Normal safe handling procedures must be used.
6. When disposing of chemicals ensure that no two chemicals are mixed.
Safety advice concerning the use of the equipment described in this manual or any relevant hazard data sheets (where applicable) may be obtained from the Company address on the back cover, together with servicing and spares information.
EN ISO 9001:2000
Cert. No. Q 05907
R
E
G
I
S
T
E
R
E
D
EN 29001 (ISO 9001)
Lenno, Italy – Cert. No. 9/90A
0255
Stonehouse, U.K.
Warning – Refer to the manual for instructions
Caution – Risk of electric shock
Protective earth (ground) terminal
Earth (ground) terminal
Direct current supply only
Alternating current supply only
Both direct and alternating current supply
The equipment is protected through double insulation
Electrical Safety
This equipment complies with the requirements of CEI/IEC 61010-1:2001-2 "Safety requirements for electrical equipment for measurement, control, and laboratory use". If the equipment is used in a manner NOT specified by the Company, the protection provided by the equipment may be impaired.
Symbols
One or more of the following symbols may appear on the equipment labelling:
Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual for any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of the Technical Publications Department.
Page 3
1
1 INTRODUCTION ........................................................... 2
2 MECHANICAL INSTALLATION .................................... 3
2.1 Siting Requirements ............................................ 3
2.1.1 Instruments ...........................................3
2.1.2 Dissolved Oxygen Flowcell .................... 3
2.2 Mounting the Instrument ..................................... 3
2.2.1 Wall-mounted Instrument ...................... 3
2.2.2 Panel-mounted Instrument .................... 4
2.3 Installing the Dissolved Oxygen Flowcell ............. 5
2.3.1 Flowcell Dimensions (Overall)7 .............. 5
2.3.2 Enclosure Dimensions (Optional) ........... 5
2.3.3 Connecting the Sample Lines ................ 5
3 ELECTRICAL CONNECTIONS .....................................6
3.1 Access to Terminals ............................................ 6
3.1.1 Wall-mounted Instruments .................... 6
3.1.2 Panel-mounted Instruments .................. 6
3.2 Connections, General ......................................... 7
3.2.1 Relay Contact Protection
and Interference Suppression ................ 7
3.2.2 System Wiring Schematic ..................... 8
3.3 Wall-mounted Instrument Connections ............... 8
3.4 Panel-mounted Instrument Connections .............9
3.5 Selecting the Mains Voltage .............................. 10
3.5.1 Wall-mounted Instrument .................... 10
3.5.2 Panel-mounted Instrument .................. 10
3.6 Flowcell Solenoid Valve Connections ................ 11
4 SETTING UP ............................................................... 12
4.1 Fitting the Dissolved Oxygen Sensor ................. 12
4.2 Connecting the Flowcell .................................... 13
4.3 Checking Sample Flow ..................................... 13
5 CONTROLS AND DISPLAYS ..................................... 14
5.1 Displays ............................................................ 14
5.2 Switch Familiarization ........................................ 14
6START UP AND OPERATION ..................................... 15
6.1 Instrument Start-up........................................... 16
6.2 Operation – Dissolved Oxygen
Measurement Mode.......................................... 16
6.2.1 Operation Page ................................... 16
6.2.2 Calibration Page .................................. 17
7 PROGRAMMING AND ELECTRICAL CALIBRATION 18
7.1 Access to Secure Parameters........................... 18
7.2 Language Page ................................................ 18
7.3 Set Up Parameters Page .................................. 18
7.4 Set Up Alarm Page ........................................... 19
7.5 Set Up Retransmission Page ............................ 21
7.6 Electrical Calibration ......................................... 24
7.6.1 Equipment Required ........................... 24
7.6.2 Preparation ......................................... 24
7.7 Factory Settings Page....................................... 25
8 MAINTENANCE .......................................................... 28
8.1 Introduction ...................................................... 28
8.2 Cleaning/Changing the Sensor ......................... 28
8.2.1 Cleaning .............................................. 28
8.2.2 Changing the Sensor .......................... 28
9 SIMPLE FAULT FINDING ........................................... 29
9.1 Diagnostic Messages ........................................ 29
9.2 Low Sensor Efficiency/Slow Sensor Cal.
or no Response to D.O. Changes ..................... 30
9.3 Checking the Temperature Input .......................30
9.4 High Sample Readings ..................................... 30
10 SPECIFICATION ......................................................... 31
11 SPARES ...................................................................... 32
11.1 Strategic Spares ...............................................32
APPENDIX A – 9438 080 24 V DC
POWER SUPPLY UNIT (OPTIONAL).......................... 34
A.1 Description ....................................................... 34
A.2 PSU Dimensions ............................................... 34
A.3 Accessing PSU Terminals ................................. 34
A.4 PSU Connections ............................................. 35
A.5 Wiring Schematic .............................................. 35
A.6 Specification .....................................................35
APPENDIX B – CALIBRATION DIAGNOSTICS ................... 36
B.1 During Calibration ............................................. 36
B.2 Low Sensor Efficiency....................................... 36
CONTENTS
Page 4
2
1 INTRODUCTION
Model 9438
Panel-Mounted Transmitter
9438 080
24 V DC P.S.U.
(optional)
Dissolved Oxygen 9438
Sensor mounted in optional enclosure
Model 9438
Wall-Mounted Transmitter
4
6
0
0
ABB
4600
AB
B
D
.O
.
g/
kg
18 . 1
A
1
A
2
This manual describes how to install and operate the 9438 Low Level Dissolved Oxygen Monitoring system. Fig. 1.1 shows the main elements of the system. Mechanical and electrical installation details of the optional power supply unit are in Appendix A.
The Dissolved Oxygen (D.O.) transmitters and associated flowcell have been designed for continuous monitoring and control of power station boiler feed water/steam condensate.
Calibration of the sensor can be manually initiated when required, or set to automatic with the programmable frequencies: 1 day, 1 week and 4 weeks.
System status can be assessed remotely using programmable alarm and/or current output diagnostic functions.
The 9438 500 transmitter is a wall-mounted instrument and the 9438 501 model is a panel-mounted,
1
/4 DIN-sized instrument. Both instruments have a single programmable D.O. input channel, and a single temperature input channel. The sample temperature is sensed by a Pt1000 resistance thermometer incorporated in the flowcell.
Instrument operation and programming is via four tactile membrane switches located on the front panel. Programs are protected from unauthorized alteration by a five-digit security code.
Fig. 1.1 System Elements
Page 5
3
68
(2.68)
42
(1.65)
Fixing Centers
160 (6.3)
69 (2.72)
Fixing Centers
Allowance for Cable Bends
200 (7.9)
61 (23/8) o.d. Vertical Post
214
(8.43)
232
(9.13)
Three holes Ø6.3 suitable for M6 fasteners
Dimensions in mm (in)
250
(9.84)
Maximum distance
30 metres
C – Within Environmental Limits
55 C Max.
–20 C
Min.
B – Within Temperature Limits
A – Maximum Distance of Instrument to Unit
Locate to suit
IP66
IP65
IP54
Environmental cover
Dissolved Oxygen 9438
IP65
Liquid handling panel
Mark fixing centers
(see Fig. 2.3)
Drill suitable
holes
Fix instrument to wall using
suitable fixings
1
2
3
2.1.2 Dissolved Oxygen Flowcell – Fig 2.7
Allow sufficient clearance (200 mm all around) for easy removal of the flowcell assembly for maintenance when not installed in the optional enclosure – see Section 2.3.1 for overall dimensions of units.
Note. To eliminate the risk of bubbles accumulating at the sensor, and hence giving erroneous readings, the flowcell assembly must be mounted vertically.
2.2 Mounting the Instrument
2.2.1 Wall-mounted Instrument – Figs 2.2 to 2.4
Fig. 2.1 Siting Requirements – Instrument/Sensor
Fig. 2.2 Overall Dimensions
Fig. 2.3 Wall Mounting
2 MECHANICAL INSTALLATION
2.1 Siting Requirements
2.1.1 Instruments – Fig. 2.1
Caution.
• Mount instruments in a location free from excessive vibration.
• Mount away from harmful vapours and/or dripping fluids.
Information. It is preferable to mount the transmitter at eye level thus allowing an unrestricted view of the front panel displays and controls.
Page 6
4
…2 MECHANICAL INSTALLATION
Position ‘U’ bolts on pipe
Position plates over ‘U’ bolts
Secure transmitter to mounting plate
Secure plates
1
2
3
4
191 (7.52)
12 (0.47)
Panel Cut-out
96 (3.78)
96
(3.78)
+0.8 –0
92
(3.62 )
+0.03 –0
+0.8 –0
92 (3.62 )
+0.03 –0
Dimensions in mm (in)
Cut a hole in the panel (see Fig. 2.5 for dimensions). Instruments may be close stacked to DIN 43835.
Insert the instrument into the panel cut-out.
Refit the panel clamps to the case, ensuring that the panel clamp anchors are located correctly in their slot.
Secure the instrument by tightening the panel clamp retaining screws.
Loosen the retaining screw on each panel clamp.
Remove the panel clamp and anchors from the instrument case.
4
5
6
1
3
2
3
2.2.2 Panel-mounted Instrument – Figs 2.5 and 2.6
Fig. 2.4 Pipe Mounting
Fig. 2.5 Overall Dimensions
Fig. 2.6 Panel Mounting
…2.2.1 Wall-mounted Instrument – Fig 2.4
Page 7
5
2MECHANICAL INSTALLATION
Dissolved Oxygen 9438
Ø 8.5 for M8 fastener in four positions
Dimensions in mm.
25
25
25
160 mm space required below sensor
panel, to allow for the opening of the
optional environmental cover.
160
440
200
325
Secure the enclosure to a vertical surface using the four fixing holes and suitable screws/bolts.
100
85
175
190
116
approximately
140
310
142
approximately
Ø 5.5 for M5 fastener in four positions
Dimensions in mm.
Flow Gauge
(ml/min)
Needle
Valve
(to set flow
rate)
Solenoid
Sample
Outlet
Solenoid-
operated
Valve
Sample
Drain
User supplied
10 mm i.d.
rubber tubing
User supplied
10 mm i.d.
rubber tubing
1 m maximum
Drain – see
Note
Sample
Outlet
Sample inlet via shut off valve if required.
Sample drain during automatic calibration
2.3 Installing the Dissolved Oxygen Flowcell
2.3.1 Flowcell Dimensions (Overall) – Fig. 2.7
2.3.2 Enclosure Dimensions (Optional) – Fig. 2.8
2.3.3 Connecting the Sample Lines – Fig. 2.9
Mount the flowcell vertically (with or without the enclosure) as shown in Figs 2.7 and 2.8. Connect the sample inlet and outlet tubes as shown in Fig. 2.9.
Note.
The sample flowrate must be between 100 and
400 ml min
–1
.
• The Company recommends that stainless steel tubing is used for sample inlet lines.
• All sample drains should be kept as short as possible and be vertical to allow the sample to drain freely.
Fig. 2.7 Flowcell Dimensions
Fig. 2.8 Enclosure Dimensions
Fig. 2.9 Connecting the Sample Lines
Note. Drain tubes must be straight and
vertical to allow the sample to flow freely.
Page 8
6
3 ELECTRICAL CONNECTIONS
1
2
3
4
2
Earth Studs
slide down
Pull out slightly. . .
. . . and slide off
Remove protection cover
Slacken captive screws
Remove nuts and protection cover
Remove mains cover
Mains Cover
Earth Stud
1
2
Warning.
• Before making any connections, ensure that the power supply, any high voltage-operated control circuits and high common mode voltage are switched off.
• Although certain instruments are fitted with internal fuse protection, a suitably rated external protection device, e.g. fuse or miniature circuit breaker (m.c.b.), must also be fitted by the installer.
3.1 Access to Terminals
3.1.1 Wall-mounted Instruments – Fig. 3.1
3.1.2 Panel-mounted Instruments – Fig. 3.2
Fig. 3.1 Access to Terminals –
Wall-mounted Instrument
Fig. 3.2 Access to Terminals – Panel-mounted
Instrument (Rear View)
Page 9
7
3 ELECTRICAL CONNECTIONS…
NC C NO
External
DC Supply
+–
Relay Contacts
Load
Diode
NC C NO
External
AC Supply
LN
Relay Contacts
C
R
Load
A – AC Applications B – DC Applications
3.2 Connections, General
Information.
Earthing (grounding) – stud terminals are fitted to the transmitter case for bus-bar earth (ground) connection – see Fig. 3.1 or
3.2.
Cable lengths – The cable length between the flowcell and the electronics unit is provided as ordered, and suitably terminated
at both ends.
Cable routing – always route the signal cable and mains-carrying/relay cables separately, ideally in earthed metal conduit.
Ensure that the cables enter the transmitter through the glands nearest the appropriate screw terminals and are short and direct. Do not tuck excess cable into the terminal compartment.
Cable glands & conduit fittings – ensure a moisture-tight fit when using cable glands, conduit fittings and blanking plugs/
bungs (M20 holes). The M16 glands ready-fitted to wall-mounted instruments accept cable of between 4 and 7 mm diameter.
Alarm Relay –the relay contacts are voltage-free and must be appropriately connected in series with the power supply and the
alarm/control device which they are to actuate. Ensure that the contact rating is not exceeded. Refer also to Section 3.2.1 for relay contact protection details when the relays are to be used for switching loads.
Retransmission output – Do not exceed the maximum load specification for the selected current retransmission range – see
Section 7.
Since the retransmission output is isolated the –ve terminal must be connected to earth (ground) if connecting to the isolated input of another device.
3.2.1 Relay Contact Protection and Interference Suppression – Fig. 3.3
If the relays are used to switch loads on and off, the relay contacts can become eroded due to arcing. Arcing also generates radio frequency interference (RFI) which can result in instrument malfunction and incorrect readings. To minimize the effects of RFI, arc suppression components are required; resistor/capacitor networks for AC applications or diodes for DC applications. These components can be connected either across the load or directly across the relay contacts. On 4600 Series instruments the RFI components must be fitted to the relay terminal block along with the supply and load wires – see Fig. 3.3.
For AC applications the value of the resistor/capacitor network depends on the load current and inductance that is switched. Initially, fit a 100R/0.022 µF RC suppressor unit (part no. B9303) as shown in Fig. 3.3A. If the instrument malfunctions (incorrect readings) or resets (display shows 88888) the value of the RC network is too low for suppression – an alternative value must be used. If the correct value cannot be obtained, contact the manufacturer of the switched device for details on the RC unit required.
For DC applications fit a diode as shown in Fig. 3.3B. For general applications use an IN5406 type ( 600 V peak inverse voltage at 3A – part no. B7363)
Note. For reliable switching the minimum voltage must be greater than 12 V and the minimum current greater than 100 mA.
Fig. 3.3 Relay Contact Protection
Page 10
8
…3 ELECTRICAL CONNECTIONS
Power
Supply
Retransmission
Relay 1 Relay 2
Serial
Power Supply
Retrans.
Relay 1 Relay 2
12 34567 +– N L
Serial
(If fitted)
12 3
45 6
12 3
45 6
1 2 3 4 5 6
4 5
– – – – – –
– –
Rx+ Rx– Tx+ Tx– 0V
Retrans. Output
Relays
1 2 3 5 6
– – – – –
NC C NO C NO
Relay 1
Solenoid valve Calibration Relay
Mains Supply
NL––Neutral
Line
Earth
Output
RS422/ RS485
Earth Stud (on case) – see Fig. 3.1
NC C NO
Normally Closed Common Normally Open
=
=
Channel 2Channel 1
13––+ve (Red)
–ve (Black) Braid
D.O. Sensor
Earth Stud (on case) – see Fig. 3.1
Channel 2Channel 1
PT1000
Temperature Compensator
5 6
7
– –
White Green Link
+
Retrans 2 Output (if fitted)
-ve +ve
3.2.2 System Wiring Schematic – Fig. 3.4
The wiring of a single solenoid/sensor system from a user­supplied 24 V DC supply is shown in Fig. 3.4.
If the 9438 080 power supply unit is employed, refer to Appendix A for wiring details.
3.3 Wall-mounted Instrument Connections – Fig. 3.5
Note. Refer to Fig. 3.1 for access to terminals.
Caution. Slacken terminal screws fully before making
connections.
Warning. The power supply earth (ground) must be connected to ensure safety to personnel, reduction of the effects of RFI and correct operation of the power supply interference filter.
Fig. 3.4 Wiring Schematic
(see Appendix A for wiring using optional PSU)
Fig. 3.5 Wall-mounted Instrument Connections
9438 Transmitter
Customer 24 V DC supply
Relay 2
C NO
Solenoid Valve
Solenoid valve powered from user supplied 24 V DC source
+ve
–ve
Page 11
9
3 ELECTRICAL CONNECTIONS…
+
Normally Closed
Common
Normally Open
Common
Normally Open
Neutral
Live
Earth
TBA
Relay 1
Solenoid Valve
Calibration
Relay
1 2 3 4 5 6
8 9 E N L
1 2 3 4 5 6 7 8
10 12
Mains Supply
Retransmission
Output
0V Rx– Rx+ Tx– Tx+ Link Green White
–ve (Black)
+ve (Red)
TBB
RS422/RS485 Serial Interface (if fitted)
D.O. Sensor
1 2 3 4 5 6 7 8 9 E N L
1 2 3 4 5 6 7 8 9 10 11 12
Earth Stud
Earth Stud (on case)
Earth Stud (on case)
Pt1000 Temperature compensator
Braid
Retrans 2 Output (if fitted)
–ve +ve
4 5
3.4 Panel-mounted Instrument Connections – Fig. 3.6
Note. Refer to Fig. 3.2 for Access to Terminals.
Caution. Slacken terminal screws fully before making connections.
Warning. The power supply earth (ground) must be connected to ensure safety to personnel, reduction of the effects of
RFI and correct operation of the power supply interference filter.
Fig. 3.6 Panel-mounted Instrument Connections
Page 12
10
…3 ELECTRICAL CONNECTIONS
Remove cover (see Fig. 3.1)
Remove front panel screws
Remove front
panel
Remove cap and screw
Slacken captive screws and remove protection cover
240 V AC
110 V AC
230
or
1
3
4
3
5
2
5
Undo captive screw
Slide instrument out of case
240 V AC
110 V AC
230
or
3
4
1
2
3.5 Selecting the Mains Voltage
3.5.1 Wall-mounted Instrument – Fig. 3.7
3.5.2 Panel-mounted Instrument – Fig. 3.8
Information. Use a small, flat-
blade screwdriver to remove the screw caps from the case.
Fig. 3.7 Selecting the Mains Voltage –
Wall-mounted Instrument
Fig. 3.8 Selecting the Mains Voltage –
Panel-mounted Instrument
Note. Some versions are fitted
with a switch in place of links. The applied voltage should be as indicated on the switch, when positioned.
Note. Some versions are fitted with a switch in place of links. The applied voltage should be as indicated on the switch, when positioned.
Page 13
11
3 ELECTRICAL CONNECTIONS
Specification to BS 6500
Cross sectional area = 0.5 mm
2
Minimum current rating = 3 A Construction = 16/0.2 mm Nominal diameter = 5.4 mm (minimum 5.0 mm)
Remove cover to expose terminals and thread the cable through the cable gland.
Connect the wires as shown. Ensure that the diode remains in the position indicated (cathode to terminal 1 and anode to terminal 2).
Fit the cover and tighten the cable gland.
Solenoid
Valve
1
2
3
+24 V DC via N/O of relay in 9438 transmitter (see Fig. 3.4)
0 V
Cable Gland
Diode
See Appendix A if 9438 080 PSU is supplied.
Customer supplied cable to transmitter or PSU
2
1
3
3
2
1
3.6 Flowcell Solenoid Valve Connections – Fig. 3.9
Fig. 3.9 Solenoid Connections
Note.
Use 2 core cable with 9438 080 PSU (ABB part number 0233 731).
Use 3 core cable with customer supplied 24 V DC supply.
Page 14
12
4 SETTING UP
1
2
3
4
5
7
Optional enclosure not shown for clarity
Slip the connector nut over the connector body and screw on to the oxygen sensor firmly.
Insert the complete assembly into the flowcell ensuring that the O­ring is in place.
Remove the top from the oxygen sensor container.
Unscrew the protective cap from the rear of the oxygen sensor
Use the clamping screw to secure the assembly. Screw in firmly using finger pressure only.
Slide the thrust washer over the connector body.
Place an O-ring (provided) as shown and locate the connector body on the oxygen sensor.
Flowcell
O-ring
Oxygen Sensor
Connector
Body
O-ring
Clamping
Screw
Thrust Washer
Connector
Nut
6
4.1 Fitting the Dissolved Oxygen Sensor – Fig. 4.1
Caution.
• Only install the oxygen sensor immediately prior to use, otherwise leave it stored in its protective container. The sensor has a limited shelf life and should NOT be stored longer than about 6 months. Store under cool conditions.
•Take special care to line up the two pins in the oxygen sensor with their respective sockets before making the connection and tightening.
•Take care not to damage the delicate membrane on the end of the oxygen sensor.
• Ensure that the mating surfaces (carrying the electrical connection) of the oxygen sensor and connector body are clean and completely dry.
Caution. Do not overtighten the clamping screw.
Fig. 4.1 Fitting the Dissolved Oxygen Sensor
Page 15
13
4 SETTING UP
Flow during normal operation
Solenoid Operated Valve – Closed
Flow Regulating Valve
Sensor
Flow Indicator
Drain
Sample In
Flow during calibration or thermal overload condition
Solenoid Operated Valve – Open
Sample In
Drain
Line up the red spots and push the plug on firmly until the locking ring engages.
Push the sensor connector on firmly and tighten ONE TURN clockwise.
4.2 Connecting the Flowcell – Fig. 4.2
4.3 Checking Sample Flow – Fig. 4.3
Note.
• The plug is a latching type to prevent it's accidental removal. To remove, hold the plug at its widest point and pull out.
• The plug is protected against spillage and corrosion by a sleeve which slides over it.
Fig. 4.2 Electrical Connections at the Flowcell
Fig. 4.3 Sample Flow Schematic
Check that the sample flows correctly in both normal operation and during a calibration or thermal overload. To simulate a calibration manually, open the valve – see Section 6.2.1 Operating Page Carefully remove the dissolved oxygen sensor and check that the flowcell is empty. If sample still flows, check that the installation complies with Section 2.3.3.
Page 16
14
5 CONTROLS AND DISPLAYS
18.1
D.O. g/kg
Alarm LEDs
Upper
Display Line
Lower
Display Line
Membrane Switches
A – Advancing to Next Page
Parameter 1 Parameter 2 Parameter 3 Parameter 4
Page 1
Parameter 1 Parameter 2 Parameter 3
Page 2
Advance to
next page
For majority
of parameters
or
B – Moving Between Parameters
C – Adjusting and Storing a Parameter Value
New value is automatically stored
Parameter Value
Adjust
D – Selecting and Storing a Parameter Choice
Parameter X
Y Z
Select
Parameter 1
Parameter 2 Parameter 3
Page X
Parameter 4
Advance to
next parameter
or
New value is automatically stored
or
5.1 Displays – Fig. 5.1
The display comprises a 5-digit, 7-segment digital upper display line and a 16-character dot-matrix lower display line. The upper display line shows numerical values of dissolved oxygen concentration, temperature, alarm set points or programmable parameters. The lower display line shows the associated units or programming information.
5.2 Switch Familiarization
Fig. 5.1 Location of Controls and Displays
Fig. 5.2 Membrane Switch Functions
Page 17
15
6START UP AND OPERATION
Operating parameters.
Available only when 2
nd
retransmission output is fitted.
Secure parameters.
Linear
FACTORY SETTINGS
–––––
uA Zero
XXXX
FACTORY SET CODE
0 0 0 0 0
ELECTRICAL CAL
–––––
Calibrate YES
NO
–––––
uA Span
XXXX
Temp Zero (1k0)
XXXX
Temp Span (1k5)
XXXX
Adjust RTX Span
–––––
Cal Time 1 (min)
2
Cal Time 2 (min)
30
Alter Fact. Code
0 0 0 0 0
Adjust RTX Zero
–––––
Factory Settings Page
Section 7.6
D.O. g/kg
18 1
.
Calibrate YES
NO
–––––
Calibrating Air
879
.
SET UP PARAMETER
–––––
Pressure mmHg
760
Salinity ppt
0
Temp. Units (
C)
–––––
Disp. Units ug/kg
ug/l
ppb
–––––
SET UP ALARM
–––––
High Spt ug/kg
150.
–––––
A1 Action Hi/Lo
High
Low
A1 Type Status
Temp
DO
Off
–––––
Low Spt ug/kg
50.
Alter Sec. Code
0 0 0 0 0
Alter Cal. Code
0 0 0 0 0
SET UP RETRANS
–––––
SET UP RETRANS 1
–––––
200.
RTX Span mg/kg
20.
RTX Zero mg/kg
100.
Enter Input %
500.
Enter Output %
–––––
RTX Type 4-20
0-20
0-10
RTX. Log
Bi-Linear
Linear
–––––
Diagnostics YES
NO
SET UP RETRANS 2
–––––
Cal. Time s
30
Fail M.time s
30
Fail S.time s
30
RTX. Log
Bi-Linear
Linear
–––––
RTX O/P Temp
D.O.
–––––
200.
RTX Span mg/kg
o
C
20.
RTX Zero mg/kg
o
C
–––––
RTX Type 4-20
0-20
0-10
Bi-Linear
Bi-Linear
Linear
100.
Enter Input %
500.
Enter Output %
Cal. Time s
30
Fail M.time s
30
Fail S.time s
30
Test Retrans. (%)
00
Autocal Off
4Weeks
1Week
1Day
–––––
Temperature (
o
C)
200
.
Sensor O/P (
A)
–––––
Sensor Eff

–––––
High Set
g/kg
150
.
Low Set
g/kg
50
.
Operation Page
Section 6.2.1
Calibration Page
Section 6.2.2
Diagnostics YES
NO
Set Up Retransmission Page
Section 7.5
Set Up Parameters Page
Section 7.3
Set Up Alarm Page
Section 7.4
SECURITY CODE
00000
Access to Secure Parameters
Section 7.1
Secure Parameters
ENGLISH
–––––
Language Page
Section 7.2
Open Valve NO
YES
–––––
To Set up Retrans 2
If NO Retrans 2
Reset Cal. NO
YES
–––––
SENSOR CAL.
–––––
00000
Cal. User Code
Cal. Aborted
–––––
–––––
Sensor Eff
Recovery Period
520.
–––––
Slow Sensor Cal.
Fig. 6.1 Overall Programming Chart
Note. The values shown on the pages in this
illustration are the factory default values.
Page 18
16
…6 START UP AND OPERATION
D.O. g/kg
Temperature ( C)
SENSOR CAL.
–––––
18 1
200
.
High Set g/kg
150
Open Valve NO
YES
.
°
.
Sensor Eff
–––––
Low Set g/kg
50
.
–––––
Sensor O/P (A)
–––––
6.1 Instrument Start-up – Fig. 6.1
Ensure all electrical connections have been made and switch on the power supply. If the instrument is being commissioned for the first time, calibration and programming of parameters is required.
The overall operating and programming chart is shown in Fig. 6.1.
6.2 Operation – Dissolved Oxygen Measurement Mode
Operation in the Dissolved Oxygen measurement mode comprises an Operating Page and a Calibration Page. The Operating Page is a general use page in which parameters are viewed only and cannot be altered. To alter or program a parameter, refer to the programming pages in Section 7. The Calibration Page allows a calibration to be carried out. A 5-digit calibration code is used to prevent unauthorized access to the sensor calibration page. The value is preset at 00000 to allow access during commissioning, but should be altered to a unique value, known only to authorized operators, in the
Set Up Alarm page – see Section 7.2
6.2.1 Operation Page
Measured Dissolved Oxygen
The measured dissolved oxygen is displayed in µg/l, mg/l, ppb, ppm, mg/kg or µg/kg. Auto ranging: 0.0 to 99.9 µg kg
–1
100 to 999 µg kg
–1
1.00 to 9.99 mg kg
–1
10.0 to 20.0 mg kg
–1
Press to advance to next parameter or press to advance to Calibration Page, Section 6.2.2.
Sample Temperature
The sample temperature is displayed in either °C or °F – see Section 7.1.
Sensor Output
Raw current signal (µA) generated by the sensor.
Sensor Efficiency
Bar graph indication of the sensor performance, based on last calibration – see Section
6.2.2.
High Setpoint
The
High alarm setpoint value is only visible if the alarm is programmed for either D.O. or
Temperature and is set to
High or High/Low.
Low Setpoint
The
Low alarm setpoint value is only visible if the alarm is programmed for either D.O. or
Temperature and is set to
Low or High/Low.
Open Valve
Manually open the calibration valve to drain the flowcell prior to accessing the sensor. The reading displayed when the valve is open does not represent the dissolved oxygen content of air-saturated water at the prevailing ambient temperature.
Advance to Calibration Page, Section 6.2.2.
Page 19
17
6START UP AND OPERATION
YES
NO
Abort
for 2 secs
Calibrating Air
879
.
SENSOR CAL.
–––––
00000
Cal. User Code
Cal. Aborted
–––––
–––––
Sensor Eff
Recovery Period
520.
D.O. g/kg
18.1
Calibrate YES
NO
–––––
–––––
Slow Sensor Cal.
OR
Press to advance to next parameter
or Press to return to Operation Page, Section 6.1
Calibration Access
Enter the required calibration code number, between 00000 and 19999. If an incorrect value is entered, access to calibration is prevented and
Calibration Page is displayed.
Select YES to enable manual calibration and press to start the calibration sequence.
Calibrating in Air
The calibration valve opens exposing the sensor to air. A flashing dot indicates that a calibration is in progress and the displayed value is the sensor reading based on the LAST calibration. See timings below. Wait Period Where the sensor is exposed to air before the stability of the sensor is
checked. This is preset at 2 minutes (Cal Time 1).
Stability Period The readings are monitored for between 1 and 5 minutes until a stable
response is achieved. When stability has been achieved the calibration valve closes allowing sample to flow past the sensor.
Abort Calibration
Pressing
during Calibrating in Air aborts the calibration and the Cal. Aborted message is displayed for 2 seconds. The calibration valve closes allowing the sample to flow past the sensor again. Note that the
A2 LED continues to indicate that a calibration is in progress, and will do so
for the duration of the programmed recovery period.
Sensor Efficiency
A five-bar display provides an indication of the sensor's performance.
 >85% efficiency  >70% "  >60% "
Pass
 >50% " >40% " <40% " Low sensor efficiency
Slow Sensor Response
If the output from the sensor does not stabilize during the 1 to 5 minute stability period the calibration will not be accepted and Slow Cal. will be desplayed.
Recovery Period
Shows the sample reading as the sensor recovers at the lower sample value. The Recovery Period is preset at 30 minutes (Cal Time 2).
The display automatically returns to the Operating Page at the end of this period.
6.2.2 Calibration Page
Calibration involves standardizing the instrument and the sensor by exposing the sensor to air. During a calibration, retransmission and alarm outputs are automatically held to prevent inadvertent operation of ancillary equipment.
Page 20
18
7 PROGRAMMING AND ELECTRICAL CALIBRATION
SECURITY CODE
00000
Espanol
–––––
SET UP PARAMETER
–––––
Pressure mmHg
760
Salinity ppt
0
Temp. Units (C)
–––––
Disp. Units ug/kg
ug/l
ppb
–––––
Autocal 4 Weeks
1 Week
1 Day
Off
–––––
SET UP ALARM
–––––
Reset Cal. NO YES
–––––
_____
SET UP PARAMETER
–––––
Francais
Espanol
˜
7.1 Access to Secure Parameters
A 5-digit security code is used to prevent tampering with the secure parameters.
Security Code
Enter the required code number between 00000 and 19999 to gain access to the secure parameters. If an incorrect value is entered, access to subsequent programming pages is prevented and the display reverts to the
Operation Page.
Advance to Language Page, Section 7.2.
7.2 Language Page
Language Page Select the language to be displayed on all subsequent pages: Español, Francais, Deutsch or English.
Advance to Set Up Parameters Page, Section 7.3
7.3 Set Up Parameters Page
Press to advance to next parameter or
Press to advance to Set Up Alarm Page, Section 7.4. These two switches are used to advance to all subsequent parameters and pages. If a parameter is changed it is automatically stored on operation of either switch.
Display Units
Select the required display units:
µg/kg, µg/l, or ppb.
Barometric Pressure Correction
Set the local barometric pressure in mm Hg (between 500 and 800). If the local barometric pressure is unknown the default value, which is the standard sea­level value of 760 mm Hg, should not be changed.
Salinity Correction
Required when monitoring sea water or other waters containing high concentrations of dissolved salts. Enter the appropriate value between 0 and 80 parts per thousand (ppt). Leave at the default value of 0 ppt if correction is not required.
Temperature Units
Select either °C or °F.
Auto Calibration
Select the frequency of automatic calibrations: 1 Day, 1 Week or 4 Weeks. Select Off to disable automatic calibrations. Only manual calibrations may be carried out.
Reset Auto Calibration
Select
YES to reset the timing for automatic calibrations. To fix the calibration to a specific
time of day select
YES at the correct time of day. (This will be automatically reset following
a power failure.)
Advance to Set Up Alarm Page, Section 7.4.
Page 21
19
7 PROGRAMMING AND ELECTRICAL CALIBRATION…
SET UP ALARM
–––––
High Spt ug/kg
150.
–––––
–––––
A1 Action Hi/Lo
High
Low
A1 Type Status
Temp
DO
Off
To Alter Sec. Code
7.4 Set Up Alarm Page
Press to advance to next parameter or Press to advance to Set Up Retrans Page, Section 7.5.
Alarm Type
Select the type of alarm required. For Status, Temp and DO alarm types, the alarm l.e.d. is off and the relay energized during normal conditions. In a fail condition, the l.e.d. is on and the relay de-energized.
Status The instrument alerts the operator to a power failure, a condition that causes
any of the error messages listed in Table 9.1 to be displayed, or the status of a calibration.
Calibration Fail will cause the Relay and LED to pulse every second.
Temp The instrument alerts the operator if the temperature of the process fluid
exceeds or drops below the set point value parameter, depending on the type of Alarm Action selected below.
D.O. The instrument alerts the operator if the Dissolved Oxygen value of the sample
exceeds or drops below the set point value parameter, depending on the type of Alarm Action selected below.
Off If selected, no alarm is set and the alarm l.e.d. is off and the relay de-energized
at all times.
Alarm Action
For 'Fail-safe' alarm operation the relay's alarm state must be the same as the power­down state, i.e. the relay is de-energized.
For High alarm operation the relay must be energized below the alarm set point.
For Low alarm operation the relay must be energized above the alarm set point.
The alarm LEDs are illuminated in the alarm condition.
Hi/Lo Alarm activates above the
High Set Point or below the Low Set Point.
High Alarm activates above the
High Set Point.
Low Alarm activates below the Low Set Point.
The set point band is defined as the actual value of the set point plus or minus the hysteresis value. The hysteresis value is ±1% of the
Alarm 1 Set Point. Alarm action
occurs if the input value is above or below the set point band. If the input moves within the set point band, the last alarm action is maintained.
Continued on next page…
mralA
noitcA
rofnoitcADEL
tupnI
tnioPteSevobA
rofnoitcADEL
tupnI
tnioPteSwoleB
rofnoitcAyaleR
tupnI
tnioPteSevobA
rofnoitcAyaleR
tupnI
tnioPteSwoleB
hgiHNOFFOdezigrene-eDdezigrenE
woLFFONOdezigrenEdezigrene-eD
Page 22
20
…7 PROGRAMMING AND ELECTRICAL CALIBRATION
High Spt ug/kg
150.
Low Spt ug/kg
50.
Alter Sec. Code
0 0 0 0 0
Alter Cal. Code
0 0 0 0 0
SET UP RETRANS
–––––
...Continued from A1 Action
High Set Point
The
High set point can be set to any value within the full measurement range, with the
units automatically changing. The set point value is subject to hysteresis as detailed above.
Set the alarm set point to the required value.
Displayed only if
Alarm Action is set to High or Hi/Lo.
Low Set Point
The
Low set point can be set to any value within the full measurement range, with the
units automatically changing. The set point value is subject to hysteresis as detailed above.
Set the alarm set point to the required value.
Displayed only if
Alarm Action is set to High or Hi/Lo.
Alter Secure Parameters Security Code
Set the secure parameters security code to a value between 00000 and 19999.
IMPORTANT – YOU MUST MEMORIZE THE NEWLY SET SECURITY CODE. If it is forgotten contact the Company for advice.
Alter D.O. Sensor Calibration Security Code
Set the pH calibration security code to a value between 00000 and 19999.
IMPORTANT – YOU MUST MEMORIZE THE NEWLY SET SECURITY CODE. If it is forgotten contact the Company for advice.
Advance to Set Up Retransmission Page, Section 7.5.
…7.4 Set Up Alarm Page
Page 23
21
7 PROGRAMMING AND ELECTRICAL CALIBRATION…
SET UP RETRANS
–––––
–––––
SET UP RETRANS 1
–––––
Log
Bi-Linear
Yes
Linear
RTX Type 4-20
0-20
0-10
RTX. Log
Bi-Linear
Linear
–––––
200.
RTX Span mg/kg
20.
RTX Zero mg/kg
100.
Enter Input %
500.
Enter Output %
Diagnostics YES
NO
No
7.5 Set Up Retransmission Page
In this section the actual values denoted by 'xxxxx' are unimportant and are used to determine display reading stability when carrying out the electrical calibration procedure.
Press
to advance to next parameter
or
Press
to advance to Factory Settings Page, Section 7.7.
Set Up Retransmission 1
Retransmission 1 Output Range
The retransmission 1 output can be selected from three mA current ranges to ensure compatibility with the peripheral device connected. Select the current range required for retransmission 1 output.
Retransmission 1 Output Scale
Select the retransmission output scale required.
Log (logarithmic) – see Fig. 7.2. Bi-Linear – see Fig. 7.1. Linear
Note for Bi-linear and log scales. The accuracy specification of the instrument should always be given consideration when setting the scale limits to avoid impractical discrimination on the retransmission output.
Retransmission 1 Span
The span current output can be set to any value between:
Linear 20 mg kg–1 and 20 mg kg
–1
Bi-Linear 20 mg kg–1 and 20 mg kg
–1
Log 100 mg kg–1 and 20 mg kg
–1
Retransmission 1 Zero
The zero current output can be set to any value between 1.0 mg kg
–1
and 200 mg kg–1. This
is available only for logarithmic output.
Note. For linear output, the zero value is always 0 mg kg
–1
Enter Input %
Set the percentage of the display span at which the breakpoint occurs: 1.0 to 100% in
0.1% increments. This is point A on Fig. 7.1.
Enter Output %
Set the percentage output at which the breakpoint occurs: 0.0 to 100% in 0.1% steps. This is point B on Fig. 7.1.
Diagnostics
Select whether the current output diagnostics are required. See Appendix B.
Continued on next page…
Page 24
22
…7 PROGRAMMING AND ELECTRICAL CALIBRATION
SET UP RETRANS 2
–––––
Cal. Time s
30
Fail M.time s
30
Fail S.time s
30
–––––
Log
Bi-Linear
Linear
RTX Type 4-20
0-20
0-10
RTX. Log
Bi-Linear
Linear
–––––
RTX O/P Temp
D.O.
–––––
Te mp
D.O.
200.
RTX Span mg/kg
o
C
20.
RTX Zero mg/kg
o
C
Yes
No
If NO Retrans 2, go to Test Retrans (%)
…7.5 Set Up Retransmission Page
Calibration Pulse time
Set the frequency of the calibration pulse signal. Programmable frequency of 15, 30, 45 seconds, 1, 2, 3, 4, 5 minutes. See Appendix B.
Calibration Fail Mark Time
Set the mark time period for the current output to be driven hard upscale. Programmable period of 30 seconds, 1, 2, 3, 4, 5,…,10 minutes. See Appendix B.
Calibration Fail Space Time
Set the space time period for the current output to be driven to 0%. Programmable period of 30 seconds, 1, 2, 3, 4, 5, …,10 minutes. See Appendix B.
Set Up Retransmission 2 – see also Table 7.1.
Note. Available only on 9438 800 series instruments.
Retransmission 2 Output Range
The retransmission 2 output can be selected from three mA current ranges to ensure compatibility with the peripheral device connected. Select the current range required for retransmission 2 output.
Retransmission 2 Output Assignment
Select the Retransmission output required:
Temp – Temperature D.O. – Dissolved Oxygen
Retransmission 2 Output Scale
Select the retransmission output scale required. Only available if D.O. selected.
Log (Logarithmic) – see Fig. 7.2. Bi-Linear – see Fig. 7.1. Linear
Note for Bi-linear and log scales. The accuracy specification of the instrument should always be given consideration when setting the scale limits to avoid impractical discrimination on the retransmission output.
Retransmission 2 Span
Set the span to the required value. See Table 7.1 for details.
Retransmission 2 Zero
Set the zero to the required value. See Table 7.1 for details.
Continued on next page…
Page 25
23
7 PROGRAMMING AND ELECTRICAL CALIBRATION…
Yes
100.
Enter Input %
500.
Enter Output %
Diagnostics YES
NO
FACTORY SETTINGS
–––––
Bi-linear
Linear
Cal. Time s
30
Fail M.time s
30
Fail S.time s
30
Test Retrans. (%)
00
Return to SET UP RETRANS
No Retrans 2
No
tnemngissAtuptuO2noissimsnarteRoreZ2noissimsnarteRnapS2noissimsnarteR
negyxOdevlossiD
gkgm0=raeniL
1–
gkgm0=raeniL-iB
1–
=goLgkgm0.1
1–
gkgm002dna
1–
gkgm02=raeniL
1–
gkgm02dna
1–
gkgm02=raeniL-iB
1–
gkgm02dna
1–
goL= gkgm001
1–
gkgm02dna
1–
(erutarepmeT )C
02foegnarmuminimottcejbuS( )C
)muminim(5)mumixam(55
(erutarepmeT )F
63foegnarmuminimottcejbuS( )F
)muminim(14)mumixam(131
Enter Input %
Set the percentage of the display span at which the breakpoint occurs: 1.0 to 100% in
0.1% increments. This is point A on Fig. 7.1.
Enter Output %
Set the percentage output at which the breakpoint occurs: 0.0 to 100% in 0.1% steps. This is point B on Fig. 7.1.
Diagnostics
Select whether the current output diagnostics are required. See Appendix B.
Calibration Pulse Time
Set the frequency of the calibration pulse signal. Programmable frequency of 15, 30, 45 seconds, 1, 2, 3, 4, 5 minutes. See Appendix B.
Calibration Fail Mark Time
Set the mark time period for the current output to be driven hard upscale. Programmable period of 30 seconds, 1, 2, 3, 4, 5, …,10 minutes. See Appendix B.
Calibration Fail Space Time
Set the space time period for the current output to be driven to 0%. Programmable period of 30 seconds, 1, 2, 3, 4, 5, …,10 minutes. See Appendix B.
Test Retransmission Output
The instrument automatically transmits a test signal of 0, 25, 50, 75 or 100% of the retransmission range selected above. The % test signal selected is shown on the upper display.
Example – for a selected range of 0 to 20 mA and 50% retransmission test signal, 10 mA
is transmitted.
Select the required retransmission test signal.
Advance to Factory Settings Page, Section 7.7.
…7.5 Set Up Retransmission Page
Table 7.1 Retransmission 2
Page 26
24
…7 PROGRAMMING AND ELECTRICAL CALIBRATION
epyTtnemurtsnI lanimreT
detnuoMllaW 1234567
detnuoMlenaP211101 9876
tnerrucev+
tupni
tnerrucev–
tupni
tupni0001TP
otkniL
)6(7lanimret
tupni0001TP
7.6 Electrical Calibration
Note. The instrument is calibrated by the company prior to despatch and an electrical calibration should be carried out only if the accuracy of the instrument is suspect and suitably calibrated test equipment is available.
7.6.1 Equipment Required
a) Current source: 0 to +100 µA.
b) Decade resistance box (temperature input simulator): 0 to 1k5 .
c) Digital milliammeter (current output measurement): 0 to 20 mA.
Note. Resistance boxes have an inherent residual resistance which may range from a few milliohms up to 1 ohm. This value must be taken into account when simulating input levels, as should the overall tolerance of the resistors within the boxes.
7.6.2 Preparation
a) Switch off the supply and disconnect the sensor, temperature compensator and current output from the electronics unit terminal
block – see Fig. 3.5 or Fig. 3.6.
b) Connect the current source / resistance box to the appropriate terminals – see Table 7.2.
Connect the milliammeter to the retransmission output terminals – see Fig. 3.5 or 3.6.
c) Switch on the supply and allow ten minutes for the circuits to stabilize.
d) Select the
Factory Settings Page and carry out the procedure in Section 7.7.
Table 7.2 Transmitter Terminal Functions
Page 27
25
7 PROGRAMMING AND ELECTRICAL CALIBRATION…
FACTORY SETTINGS
–––––
A Zero (0A)
XXXX
FACTORY SET CODE
0 0 0 0 0
ELECTRICAL CAL
–––––
Calibrate YES
NO
–––––
A Span (100A)
XXXX
Temp Zero (1k0)
XXXX
YESNO
Temp Span (1k5)
XXXX
Adjust RTX Zero
–––––
7.7 Factory Settings Page
When carrying out the electrical calibration procedure, the actual values denoted by xxxxx are unimportant and are used only to determine display reading stability.
Press
to advance to next parameter
or
Press
to return to Operating Page, Section 6.2.1.
Parameters in these pages are factory set and should not normally require adjustment. They can be set up only if the necessary equipment is available.
Factory Settings Access Code
Enter the required code number. If an incorrect value is entered, access to subsequent parameters is prevented and the display reverts to the top of the page.
Select YES to access the electrical calibration sequence. Select NO to advance to
Cal Time 1.
Caution. Do not select YES unless instrument calibration is required.
Microamp Zero
Set the current source to 0 µA and allow the instrument display to stabilize.
Microamp Span
Set the current source to +100 µA and allow the instrument display to stabilize.
Calibrate Temperature Zero
Set the temperature simulator resistance box to 1000 and allow the instrument display to stabilise.
Calibrate Temperature Span
Set the temperature simulator resistance box to 1500 and allow the instrument display to stabilise.
Adjust Retransmission Zero
Set the milliammeter reading to 4.00 mA.
Note. Retransmission signal span is calibrated using 20.00 mA. The correct value transmitted depends on the range selected in the
Set Up Outputs Page.
Continued on next page…
Page 28
26
…7 PROGRAMMING AND ELECTRICAL CALIBRATION
Adjust RTX Span
–––––
Cal Time 1 (min)
2
Cal Time 2 (min)
30
Alter Fact. Code
0 0 0 0 0
D.O. g/kg
18.1
Adjust RTX Zero 2
–––––
Adjust RTX Span 2
–––––
…7.7 Factory Settings Page
Adjust Retransmission Span
Set the milliammeter reading to 20.00 mA.
Note. Retransmission signal span is calibrated using 20.00 mA. The correct value transmitted depends on the range selected in the
Set Up Outputs Page.
Adjust Retransmission Zero 2
See
Adjust Retransmission Zero.
Adjust Retransmission Span 2
See
Adjust Retransmission Span.
Calibration Time 1
Wait period before the stability of the sensor is checked during the calibration sequence. Programmable from 1 to 10 minutes (default = 2 minutes). See Section 7.5.
Calibration Time 2
Recovery period where the sample is allowed to flow and the instrument settles on reading, before the instrument is brought back on-line (default = 30 minutes). See Section 7.5.
Alter Factory Setting Security Code
Set the security code to a value between 00000 and 19999.
Return to Operating Page, Section 6.2.1.
Page 29
27
7 PROGRAMMING AND ELECTRICAL CALIBRATION
Dissolved Oxygen Measurement – shown as % of Display Span
% Retransmission Output
0
10
20
30
40
50
60
70
80
90
100
0102030405060708090100
A
B
Dissolved Oxygen Measurement – shown as % of Display Span
% Retransmission Output
1% 100%10%
0
10
20
30
40
50
60
70
80
90
100
Fig. 7.1 Bi-Linear Scaling
Fig. 7.2 Logarithmic Scaling (two decades example)
Page 30
28
8.1 Introduction
No routine maintenance is required for this instrument other than periodic calibration – see Section 6.2.2. However, if following a calibration the sensor output shows one flashing bar, the sensor capsule has therefore become exhausted and needs replacing immediately. If the output shows two bars, replace the sensor capsule in the near future.
Storage
DO: — use sensors in date rotation to prevent them being
stored longer than necessary. — at all times, store sensors in a dry and cool environment. — store sensors in a refrigerator to extend their life, but DO
NOT allow them to freeze. DO NOT: — allow sensors to dry out, either in storage or in use. — leave sensors in vehicles where they are likely to freeze
or be exposed to high temperatures. — leave sensors on-site without protection from direct sun
or high temperatures. — use the sensor if it's sealed environment has dried out.
A dirty membrane may also be the cause of the low sensor output. To clean the sensor proceed with the following.
Caution.
• Only install the oxygen sensor immediately prior to use, otherwise leave it stored in its protective container.
•Take special care to line up the two pins in the oxygen sensor with their respective sockets before making the connection and tightening.
•Take care not to damage the delicate membrane on the end of the oxygen sensor.
• Ensure that the mating surfaces (carrying the electrical connection) of the oxygen sensor and connector body are clean and
completely dry.
8.2 Cleaning/Changing the Sensor
8.2.1 Cleaning
1) Drain the flowcell, by manually opening the solenoid valve ­Select YES to 'Open Valve' on the main operating page of the transmitter. See Section 6.2.1.
2) Unscrew the clamping screw and carefully remove the sensor assembly from the flowcell. Check that O-ring does not fall out.
8 MAINTENANCE
3) Inspect the sensor. If the membrane is clean, refit the sensor as in 5) below.
If deposits are visible on the membrane, remove by gently wiping the membrane with a moist paper tissue; for oily or greasy deposits, the tissue may be moistened with a mild detergent or, if necessary with iso-propyl alcohol (propan-2­ol). After cleaning, dry the interior of the flowcell with a paper tissue or soft cloth, ensure that the O-ring is correctly positioned.
4) Insert the sensor assembly into the flowcell.
5) Use the clamp screw to secure the assembly. Screw in firmly using finger pressure only.
Caution. Do not overtighten the clamping screw.
6) Close the solenoid valve – Select NO to 'Open Valve' on the main operating page of the transmitter. See Section 6.2.1.
7) Carry out a calibration – see Section 6.2.2. If a low sensor efficiency is displayed, see Section 9.2.
8.2.2 Changing the Sensor
1) Drain the flowcell, by manually opening the solenoid valve – Select YES to 'Open Valve' on the main operating page of the transmitter. See Section 6.2.1.
2) Unscrew the clamping screw and remove the sensor assembly from the flowcell.
3) Disconnect the sensor capsule and discard both the sensor and sealing washer.
4) Take out the O-ring from the flowcell; dry the interior of the flowcell with a tissue or soft cloth and insert the new O-ring supplied with the replacement capsule. Ensure that the O­ring is correctly located on the shoulder near the end of the cavity.
5) Remove the new sensor from its container, taking care not to damage the membrane. Unscrew the protective cap from the rear of the sensor.
6) Fit the new sealing washer (supplied) as shown in Fig 4.1 and locate and secure the connector body on the sensor.
7) Insert the complete assembly into the flowcell.
8) Use the clamping screw to secure the assembly. Screw in firmly using finger pressure only.
Caution. Do not overtighten the clamping screw.
9) Close the solenoid valve – Select NO to 'Open Valve' on the main operating page of the transmitter. See section 6.2.1.
10) Carry out a calibration – see Section 6.2.2. If a low sensor efficiency is displayed, see Section 9.2.
Page 31
29
9 SIMPLE FAULT FINDING
egasseMcitsongaiD
1AmralASUTATS
esuaCelbissoPydemeR
DEL
noitcA
noitcAyaleR
)efasliaF(
yalpsiDgnihsalFFFOdezigrenE
gnirusaemehtfoedistuosignidaeR
gKgm02ot0egnar
1–
.
.3.9&2.9snoitceSeeS
.FFEROSNESWOL
FFO/NO
)doireps1(
/dezigrene-eD
dezigrene
)doireps1(
gnirudrosnes.O.DmorftuptuO
fo%04nahtsselnoitarbilac
.tuptuodetcepxe
.2.9noitceSeeS
.LACROSNESWOLS
FFO/NO
)doireps1(
/dezigrene-eD
dezigrene
)doireps1(
gnirudrosnes.O.DmorftuptuO
deriuqergniveihcatonnoitarbilac
.ytilibats
.2.9noitceSeeS
riAnignitarbilaCNOdezigrene-eD
nehwnoitarbilacgniruddeyalpsiD
.riaotdesopxesirosnes
––
doirePyrevoceRNOdezigrene-eD
tslihwnoitarbilacretfadeyalpsiD
noezilibatsotrosnesrofgnitiaw
setunim03fonoitaruD.elpmas
.)2emiTlaC(
––
DLOC
)dlocootnoituloS(
NOdezigrene-eD5<erutarepmetelpmaS .C
5<tonsierutarepmetelpmasfI ,C
ehtfotupnierutarepmetehtkcehc
fI.3.9noitceSeesrettimsnart
ehttcatnocstsisreptluaf
.ynapmoC
TOH
)tohootnoituloS(
NOdezigrene-eD
55>erutarepmetelpmaS sihT.C
nepootevlavnoitarbilacehtsesuac
tneverpotllecwolfehtniarddna
03retfA.rosnesehtotegamad
ehtdnasesolcevlavehtsetunim derusaemsierutarepmetelpmas
litnuseunitnocssecorpsihT.niaga
55<sierutarepmetelpmaseht .C
tonsierutarepmetelpmasfI
55> erutarepmetehtkcehc,C
eesrettimsnartehtfotupni
stsisreptluaffI.3.9noitceS
.ynapmoCehttcatnoc
0001TPYTLUAFNOdezigrene-eD
/rotasnepmocerutarepmeT
rehtieerasnoitcennocdetaicossa
.tiucrictrohsronepo
snoitcennoclangisllatahtkcehC
kcehc,stsisreptluaffI.edamera
erutarepmetaotesnopserarof
.3.9noitceSees–tupni
TUPNIDILAVNINOdezigrene-eD
gnirusaemfoedistuosilangistupnI
.scinortceleehtfoegnar
otsdnopsertnemurtsnitahtkcehC natuogniyrracyblangistupnina debircsedsanoitarbilaclacirtcele
.6.7noitceSni
RORREYROMEMVNNOdezigrene-eD
yromemelitalov-nonfostnetnoC
gnirudyltcerrocdaerneebtonevah
.purewop
tiaw,rettimsnartffohctiwS
.niaganohctiwsdnasdnoces01
ehttcatnocstsisreptluaffI
.ynapmoC
9.1 Diagnostic Messages
If erroneous or unexpected results are obtained the fault may be indicated by an error message. If Alarm A1 has been selected as a STATUS alarm, then the LED and relay operation can be seen in Table 9.1. The STATUS alarm operates as a FAILSAFE alarm (during an alarm condition the relay state is the same as the power-down state, i.e. de-energized).
Table 9.1 Diagnostic Messages
Page 32
30
…9 SIMPLE FAULT FINDING
erutarepmeT
()C
ecnatsiseRtupnI
()
00.0001
010.9301
023.9701
037.6111
044.5511
050.4911
064.2321
077.0721
089.8031
090.7431
0010.5831
5.0310.0051
9.2 Low Sensor Efficiency/Slow Sensor Cal. or no Response to D.O. Changes
a) Check that the sample drains fully from flowcell. If the sample
does NOT drain fully check: i) Operation of solenoid valve. ii) Sample inlet flow rate does not exceed 400 ml min
–1
maximum.
iii) Sample fluid paths are free flowing and clear of partial
blockages.
iv) Solenoid valve drain tube is not kinked, blocked,
excessively long, does no rise along its length.
v) Flow gauge is not blocked or dirty.
b) Replace the sensor (see Section 8.2.2) as an initial check. It
is also important that all program parameters have been set correctly and have not been altered inadvertently – see Section 7.
If the fault persists: c) Carry out an electrical calibration as detailed in Section 7.6
and check that the instrument responds correctly to the current input.
Failure to respond to the input usually indicates a fault with the transmitter, which must be returned to the Company for repair.
d) If the response in a) is correct, select the
Operating Page and
set the current source to a value which gives an on-scale D.O. reading on the transmitter. Make a note of the current source setting and the D.O. reading. Reconnect the sensor cable and connect the current source to the sensor end of the cable. Set the same current value on the source and check that the transmitter displays the noted reading in this configuration.
If check a) is correct but check b) fails, check the cable connections and condition. If the response for both checks is correct, fit a new sensor and calibrate it.
9.3 Checking the Temperature Input
Check that the instrument responds to a temperature input. Disconnect the PT1000 leads and connect a suitable resistance box directly to the transmitter inputs – see Section 7.6. Check that the transmitter displays the correct values as set on the resistance box – see Table 9.2.
Incorrect readings usually indicate an electrical calibration problem. Recalibrate the instrument – see Section 7.6.
9.4 High Sample Readings
If the sample reading is higher than expected, the most likely reason is air ingress into the main sample line. Check and tighten ALL sample connections as it is possible to have an air leak into the sample without sample leaking.
Table 9.2 Temperature Readings for Resistance Inputs
Page 33
31
10 SPECIFICATION
System
Measuring ranges
Programmable within the ranges 0 to 20.0 µg kg
–1
and 0 to
20 mg kg
–1
Scaling
µg kg
–1
, mg kg–1 or ppb, ppm
Accuracy
±5% of reading or ±1 µg kg
–1
, whichever is the greater
Response time
90% of a step change in 1 minute
Resolution
0.1 µg kg
–1
Stability
±5% of reading or ±1 µg kg
–1
per week, whichever is the greater Not applicable when autocalibration is in operation
Temperature compensation
5 to 55°C (41 to 131°F) automatic using Pt1000 resistance thermometer
Salinity correction
Preset within the range 0 to 80 ppt
Barometric pressure correction
Preset within the range 500 to 800 mm Hg
Sample flow
100 to 400 ml min
–1
Sample pressure
Maximum 2 bar
Sample temperature
5 to 55°C (41 to 131°F)
Sensor ambient temperature
0 to 55°C (32 to 131°F)
Environmental Data
Operating temperature limits
–20 to 55°C (–4 to 131°F)
Operating humidity limits
Up to 95% RH non-condensing
Storage temperature limits
Liquid handling panel: –25 to 70°C (–13 to 158°F) Sensor: 0 to 55°C (32 to 131°F) Transmitter: –25 to 70°C (–13 to 158°F) Solenoid valve power supply (optional): –25 to 70°C (–13 to 158°F)
Protection
Liquid handling panel
IP65 IP54 – Liquid handling panel enclosure Solenoid valve power supply (optional): IP65 (optional)
Transmitter
Panel-mounting, IP66/NEMA4X Wall-mounting, IP66/NEMA4X front
Power Requirements
System
Power consumption, <21 VA
Transmitter
Power supply, 100 to 130 V or 200 to 260 V 50/60 Hz Power consumption, <10 VA
Error due to power supply variations
Less than ±2% for +6% –20% variation from nominal supply voltage
Insulation, mains to earth
2 kV r.m.s.
Solenoid valve
Power supply, 90 to 132 V or 180 to 264 V 47/63 Hz Power consumption, <11 VA
Mechanical Data
Mounting
Transmitter, Wall- or Panel-mounting Liquid handling panel/enclosure, Wall-mounting Solenoid valve power supply, Wall-mounting
Overall dimensions
Liquid handling panel
without unions and without environmental enclosure: 100 approx. x 310 x 118 mm
(3.94 approx. x 12.2 x 4.65 in.) with environmental enclosure: 250 x 440 x 160 mm
(9.84 x 17.32 x 6.3 in.)
Transmitter
Wall-mounting: 160 x 214 x 68 mm
(6.29 x 8.43 x 2.68 in.) Panel-mounting: 96 x 96 x 191 mm
(3.78 x 3.78 x 7.52 in.) Panel cutout: 92 x 92 mm
(3.62 x 3.62 in.)
Weights
Liquid handling panel
With sensor fitted, without environmental enclosure: 1.3 kg (2.86 lb) With sensor fitted and with environmental enclosure: 3.9 kg (8.58 lb)
Transmitter
Wall-mounting: 2 kg (4.41 lb) Panel-mounting: 1.5 kg (3.31 lb) Solenoid valve power supply: 0.7 kg (1.54 lb)
Sample connections
Compression fitting to accept either 6 mm or
1
/4 in. O.D.
tubing – to be specified when ordering
Page 34
32
…10 SPECIFICATION 11 SPARES
Specification – Transmitter
Transmitter Display
Measured value
5-digit x 7-segment back-lit l.c.d.
Information
16-character, single line, dot matrix, back-lit l.c.d.
Insulation, contacts to earth
2kVr.m.s.
Set Point and Relay
No. of set points
One
Set point adjustment
Programmable as a concentration or diagnostics alarm
Set point hysteresis
±1% of f.s.d. (fixed)
Local set point annunciation
Red l.e.d.
Number of relays
One
Relay contacts
Single pole changeover Rating: 250 V AC 250 V DC maximum
3AAC 3 A DC maximum
Loading: (non-inductive) 750 VA 30 W maximum
(inductive) 75 VA 3 W maximum
Retransmission
Number of retransmission signals
One, fully isolated
Optional second current output
Output current 0 to 10, 0 to 20 or 4 to 20 mA programmable
Maximum load resistance
500 (20 mA maximum)
Serial communication
RS422/RS485 (optional, with one current output signal)
Normal, replacement spares are shown in Fig. 11.1. Strategic spares are listed below.
11.1 Strategic Spares
Part No. Description ................................................ Qty
9438 080 24V Power Supply Unit ................................... 1
0234 037 Solenoid Valve assembly................................. 1
0216 574 Flow Gauge assembly ..................................... 1
0216 575 Needle Valve Cartridge assembly .................... 1
PCB Assemblies, Wall Mounted
9438 070 Complete Main PCB assembly, for
single current o/p .......................................... 1
9438 071 Complete Main PCB assembly, for
single current o/p + Serial/Modbus ............... 1
9438 072 Complete Main PCB assembly, for
2 current o/p version ..................................... 1
4600 0295 Display PCB assembly .................................... 1
4600 0335 Low Level D.O. Module assembly ................... 1
4600 0405 2nd Retransmission output module
assembly ....................................................... 1
PCB Assemblies, Panel Mounted
9438 075 Complete Main PCB assembly (cropped),
for single current o/p ..................................... 1
9438 076 Complete Main PCB assembly
(cropped), for single current o/p +
Serial/Modbus ............................................... 1
9438 077 Complete Main PCB assembly (cropped),
for 2 current o/p version ............................... 1
4600 0246 Power supply PCB assembly (cropped) .......... 1
4600 0285 Mother PCB assembly .................................... 1
4600 0335 Low Level D.O. Module assembly ................... 1
4600 0405 2nd Retransmission output module
assembly ....................................................... 1
Test Equipment
9439 950 Dissolved Oxygen Test Simulator .................... 1
9439 035 Test Simulator lead.......................................... 1
Page 35
33
11 SPARES
Replacement Seals Pack (9437016)
comprising:
2 x large O-rings 2 x small O-rings 2 x nylon seals 2 x end caps* 1 x protective cover
*The end cap is used to blank off this connector when the plug is not fitted.
End Cap
O-ring
Protective
Cover
Flowcell
Connector
Body
O-ring (
3
/4 in. I.D. x 0.070in cord)
Plug
Handle Assembly 9437025
which includes:
Connector Body Sensor Connector Cable Assembly Connector Nut Thrust Washer Clamping Screw Plug
Clamping Screw
Thrust Washer
Connector Nut
Compression Coupling (to suit 6 mm or
1
/4 in.
o.d. stainles steel pipe)
Sensor Connector Cable Assembly: 9437 029/031/032/034
(1 m/5 m/10 m/30 m respectively)
Oxygen Sensor 9435 300
(including O-rings)
O-ring (
7
/8 in. I.D. x 0.070 in. cord)
Note. Ensure that the correct O-rings are fitted in the appropriate positions as shown. Fit new O-rings when a new sensor is fitted.
Fig. 11.1 Replacement Spares
Page 36
34
APPENDIX A – 9438 080 24 V DC POWER SUPPLY UNIT (OPTIONAL)
23
All dimensions in mm
Note. Make allowance for cable bends
160
63
CTRS
fixing
98
145 CTRS
136
62
M6 stud for
earth
bonding
Fixing screws M4
Slacken captive screws
A.1 Description
The 24 V DC switch mode power supply unit is capable of powering up to four separate 9438 dissolved oxygen system solenoids. The 24 V is switched to the solenoid when required by the operation of the calibration relay in the main 9438 transmitter.
Fig. A.3 shows the connection details in the PSU.
Fig. A.4 shows the interconnection between the PSU and a single 9438 transmitter.
A.2 PSU Dimensions – Fig. A.1
Warning.
• Before making any connections, ensure that the power supply, any high voltage-operated control circuits and high common mode voltage are switched off.
• Although certain instruments are fitted with internal fuse protection, a suitably rated external protection device, e.g. fuse or miniature circuit breaker (m.c.b.), must also be fitted by the installer.
A.3 Accessing PSU Terminals – Fig. A.2
Fig. A.1 Power Supply Unit Dimensions
Fig. A.2 Access to PSU Terminals
Page 37
35
APPENDIX A – 9438 080 24 V DC POWER SUPPLY UNIT (OPTIONAL)
Pre-wired to terminal block
110/115 V AC 230/240 V AC Mains cable in
To earth terminal in lid
To 4641/4646 transmitter solenoid valve relay (maximum: 4)
To solenoid valve on flowcell assembly (maximum: 4 valves)
Earth bonding stud. (Enclosure must be bonded to earth via this stud)
TO RELAY 2 OF TRANSMITTERS
SWITCHED OUTPUT TO SOL VALVE
E N L
0V
+24V
0V
+24V
0V
+24V
0V
+24V
TX4 TX3 TX2 TX1
SOL 4 SOL 3 SOL 2 SOL 1
COM
N/O
COM
N/O
COM
N/O
COM
N/O
Solenoid Valve Powered Via Optional PSU
9438 Transmitter
Relay 2
C NO
Solenoid Valve
90 – 264 V a.c. Mains. supply
Switch Mode PSU
(up to 4 D.O. Systems)
Switched Output to Solenoid Valve
0V +24V
C N/O
A.4 PSU Connections – Fig. A.3
A.5 Wiring Schematic – Fig. A.4 A.6 Specification
Overall dimensions
160 x 98 x 62 mm (6.3 x 3.86 x 2.44 in.)
Output power
24 V @ 2.5 A, 60 W maximum from all outputs
Holdup time
6 ms at full load 115/230 V AC
Line regulation
0.3% over operating range
Load regulation
0.5% from minimum load to full load
Power supply
Voltage requirements
90 to 264 V AC, 47 to 63 Hz
Fig. A.3 PSU Connections
Fig. A.4 Wiring Schematic
Important Note. PSU should be earth bonded.
Note. 0V is internally connected to earth.
Page 38
36
APPENDIX B – CALIBRATION DIAGNOSTICS
Cal Pulse
Expose to Air
Recovery Period
Time
Output Current (%)
50
100
0
Mark
(30 s to 5 min)
Expose to Air
Output Current (%)
50
100
Space
(30 s to 5 min)
0
Time
Fig. B.1 During Calibration
Fig. B.2 Low Sensor Efficiency
The transmitter can be configured to enable the current output signal to indicate certain calibration diagnostic information. If the option for diagnostics is selected within the Set Up Retransmission scrolls, then the current output will indicate when a calibration is taking place, and also will indicate if the sensor is giving Low Sensor Efficiency.
B.1 During Calibration
The current output value will be maintained during a calibration, but the output will pulse from the maintained value to 0%, depending upon a programmable Cal Pulse period. The Calibration Pulse period can be programmed 15, 30, 45 seconds, 1, 2, 3, 4, 5 minutes.
This will continue for the full duration of the calibration, exposing the sensor to air, and the recovery period. At the end of the recovery period, if the response is good, the instrument will go back on line and the current output will become live.
B.2 Low Sensor Efficiency
If the output from a sensor is found to be below a predetermined level during a calibration (i.e. Low Sensor Efficiency) the calibration will not be accepted. The current output immediately goes above the full scale value, and will continue to pulse on a programmable Mark/Space basis. The time for the Mark and Space periods can be programmed separately to 30 seconds, 1, 2, 3…..10 minutes.
Page 39
PRODUCTS & CUSTOMER SUPPORT
Products
Automation Systems
• for the following industries:
– Chemical & Pharmaceutical – Food & Beverage – Manufacturing – Metals and Minerals – Oil, Gas & Petrochemical – Pulp and Paper
Drives and Motors
• AC and DC Drives, AC and DC Machines, AC motors to 1kV
• Drive systems
• Force Measurement
• Servo Drives
Controllers & Recorders
• Single and Multi-loop Controllers
• Circular Chart and Strip Chart and Recorders
• Paperless Recorders
• Process Indicators
Flexible Automation
• Industrial Robots and Robot Systems
Flow Measurement
• Electromagnetic Flowmeters
• Mass Flow Meters
• Turbine Flowmeters
• Flow Elements
Marine Systems & Turbochargers
• Electrical Systems
• Marine Equipment
• Offshore Retrofit and Refurbishment
Process Analytics
• Process Gas Analysis
• Systems Integration
Transmitters
• Pressure
• Temperature
• Level
• Interface Modules
Valves, Actuators and Positioners
• Control Valves
• Actuators
• Positioners
Water, Gas & Industrial Analytics Instrumentation
• pH, conductivity, and dissolved oxygen transmitters and sensors
• ammonia, nitrate, phosphate, silica, sodium, chloride, fluoride, dissolved oxygen and hydrazine analyzers.
• Zirconia oxygen analyzers, katharometers, hydrogen purity and purge-gas monitors, thermal conductivity.
Customer Support
We provide a comprehensive after sales service via a Worldwide Service Organization. Contact one of the following offices for details on your nearest Service and Repair Centre.
United Kingdom
ABB Limited Tel: +44 (0)1453 826661 Fax: +44 (0)1453 829671
United States of America
ABB Inc. Tel: +1 775 850 4800 Fax: +1 775 850 4808
Client Warranty
Prior to installation, the equipment referred to in this manual must be stored in a clean, dry environment, in accordance with the Company's published specification.
Periodic checks must be made on the equipment's condition. In the event of a failure under warranty, the following documentation must be provided as substantiation:
1. A listing evidencing process operation and alarm logs at time of failure.
2. Copies of all storage, installation, operating and maintenance records relating to the alleged faulty unit.
Page 40
IM/9438 Issue 6
The Company’s policy is one of continuous product improvement and the right is reserved to modify the
information contained herein without notice.
Printed in UK (05.05)
© ABB 2005
ABB Limited
Oldends Lane, Stonehouse Gloucestershire GL10 3TA UK Tel: +44 (0)1453 826 661 Fax: +44 (0)1453 829 671
ABB Inc.
Analytical Instruments 9716 S. Virginia St., Ste. E Reno, Nevada 89521 USA Tel: +1 775 850 4800 Fax: +1 775 850 4808
ABB has Sales & Customer Support expertise in over 100 countries worldwide
www.abb.com
Loading...